| | Class 2 Device Recall PhysioControl LIFEPAK 12 defibrillator/monitor |  |
| Date Initiated by Firm | May 08, 2008 |
| Date Posted | September 11, 2008 |
| Recall Status1 |
Terminated 3 on September 04, 2013 |
| Recall Number | Z-1858-2008 |
| Recall Event ID |
48140 |
| 510(K)Number | K991910 |
| Product Classification |
Automated External Defibrillator - Product Code MKJ
|
| Product | Physio-Control LIFEPAK 12 defibrillator/monitor. The LIFEPAK12 defibrillator/monitor series is a battery operated portable external defibrillator with monitor used by healthcare providers in EMS, hospital and clinic settings. The LIFEPAK 12 has both manual and semi-automatic defibrillation operations. When clinically indicated, the LIFEPAK 12 defibrillator/monitor allows the operator to deliver a brief, high-energy shock to the heart of the patient. Operators may pre-configure the device to reduce complexity during normal operation. |
| Code Information |
Part Nbr 3010178-010, DVLP12-02-000102, UVLP12-02-000022, VLP12-02-000021, VLP12-02-000022, VLP12-02-000023, VLP12-02-000057, VLP12-02-000058, VLP12-02-000059, VLP12-02-000099, VLP12-02-000100, VLP12-02-000107, VLP12-02-000110, VLP12-02-000242, VLP12-02-000287, VLP12-02-000296, VLP12-02-000317, VLP12-02-000338, VLP12-02-000411, VLP12-02-000449, VLP12-02-000577, VLP12-02-000800, VLP12-02-000801, VLP12-02-000876, VLP12-02-000983, VLP12-02-000985, VLP12-02-001029, VLP12-02-001038, VLP12-02-001168, VLP12-02-001253, VLP12-02-001366, VLP12-02-001387, VLP12-02-001430, VLP12-02-001449, VLP12-02-001460, VLP12-02-002163, VLP12-02-002175, VLP12-02-002222, VLP12-02-002250, VLP12-02-002860, VLP12-02-003223, VLP12-02-003585, VLP12-02-003809, VLP12-02-003979, VLP12-02-004055, VLP12-02-004221, VLP12-02-004354, VLP12-02-006000, VLP12-02-006011, VLP12-02-006012, VLP12-02-006013, VLP12-02-006018, VLP12-02-006020, VLP12-02-006021, VLP12-02-006030, VLP12-02-006033, VLP12-02-006036, VLP12-02-006037, VLP12-02-006042, VLP12-02-006044, VLP12-02-006057, VLP12-02-006060, VLP12-02-006064, VLP12-02-006067, VLP12-02-006076, VLP12-02-006077, VLP12-02-006081, VLP12-02-006086, VLP12-02-006087, VLP12-02-006088, VLP12-02-006097, VLP12-02-006101, VLP12-02-006103, VLP12-02-006107, VLP12-02-006110, VLP12-02-006114, VLP12-02-006119, VLP12-02-006159, VLP12-02-006176, VLP12-02-006179, VLP12-02-006181, VLP12-02-006196, VLP12-02-006199, VLP12-02-006202, VLP12-02-006210, VLP12-02-006223, VLP12-02-006226, VLP12-02-006234, VLP12-02-006238, VLP12-02-006256, VLP12-02-006263, VLP12-02-006266, VLP12-02-006267, VLP12-02-006268, VLP12-02-006269, VLP12-02-006275, VLP12-02-006290, VLP12-02-006304, VLP12-02-006306, VLP12-02-006312, VLP12-02-006315, VLP12-02-006322, VLP12-02-006329, VLP12-02-006333, VLP12-02-006343, VLP12-02-006349, VLP12-02-006359, VLP12-02-006365, VLP12-02-006366, VLP12-02-006367, VLP12-02-006373, VLP12-02-006393, VLP12-02-006396, VLP12-02-006398, VLP12-02-006414, VLP12-02-006423, VLP12-02-006437, VLP12-02-006455, VLP12-02-006457, VLP12-02-006475, VLP12-02-006478, VLP12-02-006483, VLP12-02-006487, VLP12-02-006500, VLP12-02-006531, VLP12-02-006542, VLP12-02-006555, VLP12-02-006565, VLP12-02-006576, VLP12-02-006577, VLP12-02-006578, VLP12-02-006593, VLP12-02-006600, VLP12-02-006615, VLP12-02-006623, VLP12-02-006624, VLP12-02-006644, VLP12-02-006647, VLP12-02-006686, VLP12-02-006687, VLP12-02-006723, VLP12-02-006747, VLP12-02-006748, VLP12-02-006757, VLP12-02-006780, VLP12-02-006789, VLP12-02-006823, VLP12-02-006836, VLP12-02-006845, VLP12-02-006863, VLP12-02-006911, VLP12-02-006922, VLP12-02-006941, VLP12-02-006945, VLP12-02-006965, VLP12-02-007013, VLP12-02-007028, VLP12-02-007029, VLP12-02-007066, VLP12-02-007098, VLP12-02-007133, VLP12-02-007143, VLP12-02-007170, VLP12-02-007174, VLP12-02-007189, VLP12-02-007209, VLP12-02-007211, VLP12-02-007212, VLP12-02-007213, VLP12-02-007214, VLP12-02-007215, VLP12-02-007217, VLP12-02-007218, VLP12-02-007220, VLP12-02-007222, VLP12-02-007228, VLP12-02-007231, VLP12-02-007241, VLP12-02-007270, VLP12-02-007280, VLP12-02-007312, VLP12-02-007330, VLP12-02-007340, VLP12-02-007351, VLP12-02-007355, VLP12-02-007357, VLP12-02-007359, VLP12-02-007364, VLP12-02-007368, VLP12-02-007369, VLP12-02-007370, VLP12-02-007373, VLP12-02-007374, VLP12-02-007375, VLP12-02-007379, VLP12-02-007384, VLP12-02-007387, VLP12-02-007394, VLP12-02-007399, VLP12-02-007408 SN: 36285363, 36285396, 36285399, 36255174, 36255175, 36255176, 36255177, 36255178, 36176270, 36176271, 36176272, 36176273, 36176274, 36185900, 36150660, 36183667, 36183668, 36183669, 36183670, 36183671, 36183672, 36183673, 36217503, 36217504, 36175110, 36175111, 36175112, 36175113, 36175114, 36175115, 36175116, 36175117, 36175118, 36175119, 36175120, 36175121, 36175122, 36175123, 36175124, 36175125, 36175126, 36175127, 36175211, 36175212, 36175213, 36176278, 36183819, 36183820, 36183821, 36186386, 36186387, 36186389, 36186390, 36213282, 36213283, 36201095, 36201096, 36266477, 36266478, 36278215, 36278216, 36262903, 36262904, 36262905, 36255134, 36255135, 36255136, 36255139, 36255141, 36265918, 36255061, 36271596, 36271597, 36271598, 36271599, 36262898, 36262899, 36174623, 36174624, 36174625, 36174626, 36174627, 36174628, 36174629, 36174630, 36174631, 36174632, 36174633, 36174634, 36174635, 36174636, 36174637, 36174638, 36174639, 36174640, 36183216, 36183217, 36262662, 36262664, 36262665, 36262666, 36262667, 36262668, 36262669, 36262670, 36262671, 36266244, 36260617, 36260618, 36150802, 36213255, 36213256, 36213257, 36213258, 36213259, 36213260, 36213261, 36213262, 36255990, 36271561, 36271562, 36271563, 36271564, 36271565, 36271566, 36271567, 36271568, 36271569, 36271570, 36271571, 36271572, 36271573, 36271574, 36271575, 36271576, 36271577, 36271578, 36271579, 36271580, 36233648, 36233649, 36278171, 36285371, 36285372, 36285373, 36285374, 36285375, 36285376, 36285377, 36285378, 36285379, 36285380, 36285381, 36285382, 36285383, 36285384, 36285385, 36285386, 36285387, 36285388, 36285389, 36175216, 36175217, 36175218, 36175219, 36175220, 36175221, 36175222, 36175223, 36175224, 36175225, 36175226, 36175227, 36175228, 36175229, 36175230, 36175231, 36175232, 36175233, 36186049, 36186050, 36186051, 36186052, 36186053, 36186054, 36186055, 36186056, 36186057, 36186058, 36186059, 36186060, 36186061, 36186062, 36186063, 36186064, 36186065, 36186066, 36186067, 36186068, 36186069, 36186070, 36186071, 36186072, 36186073, 36186074, 36186075, 36186076, 36186077, 36186078, 36186079, 36186080, 36186081, 36186082, 36186083, 36186084, 36186085, 36186086, 36186087, 36186088, 36186089, 36186090, 36275217, 36275218, 36275219, 36275220, 36275221, 36275222, 36275223, 36275224, 36275225, 36275226, 36275227, 36275228, 36275229, 36275230, 36275231, 36275232, 36275233, 36275234, 36275235, 36275236, 36278943, 36278944, 36286152, 36286153, 36286154, 36286155, 36286156, 36286157, 36286158, 36286159, 36286160, 36286161, 36207354, 36207355, 36213235, 36260588, 36260589, 36260590, 36264237, 36260558, 36260559, 36260560, 36260561, 36260562, 36278072, 36260608, 36260555, 36303448, 36285366, 36285367, 36285368, 36285369, 36285370, 36163725, 36163726, 36163729, 36163731, 36163732, 36164283, 36164284, 36164285, 36233640, 36233641, 36233642, 36233643, 36233644, 36233645, 36233646, 36233647, 36174037, 36258693, 36234012, 36257753, 36303573, 36303574, 36303575, 36303576, 36303577, 36303578, 36303579, 36303580, 36303581, 36303582, 36303583, 36303584, 36303585, 36303586, 36303587, 36303588, 36303589, 36303590, 36303591, 36303592, 36304249, 36304250, 36304251, 36304252, 36304253, 36304254, 36304255, 36304256, 36304257, 36304258, 36304259, 36304260, 36304261, 36304262, 36304263, 36304264, 36304265, 36304266, 36304267, 36304268, 36304269, 36304270, 36304271, 36304272, 36304273, 36304274, 36304275, 36304276, 36304277, 36304278, 36181926, 36181927, 36181928, 36181929, 36233639, 36310266, 36310267, 36310268, 36310269, 36310270, 36310271, 36310272, 36310274, 36310275, 36310276, 36257706, 36204620, 36254098, 36174066, 36304245, 36304246, 36304247, 36304248, 36286147, 36286148, 36286149, 36286150, 36286151, 36204612, 36204613, 36254067, 36254068, 36254069, 36254070, 36254071, 36262699, 36262700, 36278170, 36303443, 36218081, 36218082, 36218083, 36254047, 36232224, 36233627, 36303399, 36303400, 36303401, 36303402, 36303403, 36303404, 36303405, 36303406, 36303407, 36303408, 36303409, 36303410, 36303411, 36303412, 36303413, 36303414, 36303415, 36303416, 36303417, 36303418, 36303419, 36303420, 36303421, 36303422, 36213288, 36302088, 36204624, 36204625, 36204626, 36204655, 36204656, 36204657, 36204658, 36204659, 36204660, 36204661, 36256174, 36256175, 36256176, 36256177, 36256178, 36262681, 13393841, 11789924, 11175777, 36213559, 36213560, 36213561, 36213562, 36213563, 36213564, 36213565, 36213566, 36213567, 36213568, 36213569, 36213570, 36213571, 36213572, 36213573, 36213574, 36213575, 36213576, 36213577, 36213578, 36213579, 36213580, 36213581, 36213582, 36213583, 36213584, 36213585, 36213586, 36213587, 36213588, 36213589, 36213590, 36213591, 36213592, 36213593, 36213594, 36213595, 36213596, 36213597, 36213598, 36150717, 36150718, 36150719, 36150720, 36152481, 36152482, 36152483, 36152484, 36152485, 36153037, 36153038, 36153039, 36153040, 36153041, 36153042, 36153043, 36153044, 36153045, 36153046, 36153047, 36153048, 36153049, 36153050, 36153051, 36153052, 36153053, 36153054, 36153055, 36153056, 36153057, 36153058, 36153059, 36153060, 36153061, 36153062, 36153063, 36153064, 36153065, 36153066, 36153067, 36153068, 36153069, 36153070, 36153071, 36153072, 36153073, 36153074, 36153075, 36243683, 36243684, 36270444, 36270445, 36270446, 36270441, 36243695, 36243696, 36243697, 36243698, 36243699, 36278150, 36265931, 36265921, 36286216, 36262920, 36262921, 36266317, 36266369, 36266316, 36235221, 36235222, 36235223, 36235224, 36235225, 36235226, 36235227, 36235228, 36235229, 36235230, 36235231, 36235232, 36235233, 36235234, 36235235, 36235236, 36235237, 36235238, 36235239, 36235240, 36235241, 36235242, 36235243, 36235244, 36235245, 36235246, 36235247, 36235248, 36232173, 36232174, 36224529, 36224530, 36263676, 36247201, 36247202, 36247203, 36247204, 36247205, 36247206, 36247207, 36247208, 36247209, 36247210, 36247211, 36247212, 36247213, 36199480, 36199481, 36199482, 36199483, 36199484, 36199485, 36199486, 36199487, 36164681, 36201103, 36278910, 36278911, 36278912, 36278913, 36278914, 36278915, 36278916, 36278917, 36278918, 36278919, 36278920, 36278921, 36278922, 36278923, 36278924, 36278925, 36255167, 36255168, 36255169, 36282196, 36224667, 36278982, 36263673, 36281309, 36282214, 36282215, 36282218, 36282219, 36282220, 36282221, 36278968, 36204714, 36255057, 36164692, 36164693, 36164694, 36164695, 36199489, 36222822, 36222823, 36222824, 36233623, 36258300, 36218019, 36218020, 36218021, 36218022, 36218023, 36218024, 36218025, 36218026, 36218027, 36218028, 36218029, 36218030, 36218031, 36218032, 36218033, 36218034, 36218035, 36218036, 36218037, 36218038, 36218039, 36218040, 36218041, 36218042, 36218043, 36218044, 36218045, 36218046, 36218047, 36218048, 36218049, 36218050, 36218051, 36218052, 36218053, 36255014, 36255017, 36255018, 36255019, 36255021, 36258454, 36258455, 36258456, 36258457, 36258458, 36258459, 36258460, 36258462, 36258463, 36258464, 36258465, 36258466, 36258467, 36258468, 36263593, 36174721, 36186731, 36281308, 36186147, 36254368, 36264202, 36266463, 36254362, 36204731, 36254336, 36232180, 36278948, 36278949, 36207368, 36285393, 36286201, 36258477, 36258478, 36258479, 36222767, 36222768, 36278935, 36278936, 36204703, 36234017, 36234831, 36164649, 36286179, 36278132, 36254227, 36232167, 36259087, 36259088, 36259089, 36260549, 36204677, 36201060, 36264233, 36277771, 36258193, 36278069, 36254116, 36266387, 36282201, 36282202, 36282206, 36282207, 36278899, 36278900, 36278901, 36278902, 36278903, 36278904, 36278905, 36278906, 36278907, 36258185, 36286144, 36254053, 36285390, 36285391, 36157248, 36157249, 36157250, 36157511, 36157512, 36157513, 36232184, 36232148, 36274789, 36213160, 36213167, 36213225, 36256389, 36269114, 36269115, 36269118, 36269121, 36269128, 36254051, 36256353, 36256356, 36256364, 36281307, 36213326, 36264471, 36264472, 36232159, 36232189, 36232219, 36232223, 36232227, 36232231, 36233614, 36233617, 36233630, 36266209, 36224691, 36236461, 36236462, 36282195, 36207351, 36207352, 36213319, 36254042, 13910305, 12704440, 12516803, 12516802, 11969626, 11969625, 11790995, 36255610, 36266484, 36266485, 36255596, 36282234, 36282235, 36282236, 36282237, 36282238, 36282239, 36282240, 36282241, 36282242, 36282243, 36282244, 36282245, 36282246, 36282247, 36282248, 36282249, 36282250, 36282251, 36282252, 36282253, 36213321, 36213322, 36255595, 36255599, 36255604, 31576530, 14063313, 13788399, 8594124, 36258262, 36231662, 36231663, 36231664, 36231665, 36231666, 36231667, 36231668, 36231669, 36231670, 36231671, 36231672, 36231673, 36231674, 36231675, 36231676, 36231677, 36231678.36231679, 36231680, 36231681, 36231682, 36234046, 36243716, 36243717, 36243718, 36243719, 36243720, 36243721, 36243722, 36243723, 36243724, 36243725, 36243726, 36243727, 36243728, 36243729, 36243730, 36243731, 36243732, 36243733, 36243734, 36243735, 36251472, 36251473, 36251474, 36251475, 36251476, 36251477, 36251478, 36251479, 36243706, 36270456, 36213477, 36207371, 36157251, 36157252, 36157253, 36157254, 36157255, 36157256, 36163125, 36163126, 36163127, 36163430, 36174665, 36174666, 36174667, 36174694, 36174695, 36174696, 36181303, 36181304, 36181305, 36181306, 36181307, 36181308, 36181309, 36181310, 36181311, 36181312, 36181313, 36181314, 36181315, 36181316, 36181317, 36181318, 36181319, 36181320, 36181321, 36181322, 36181323, 36181324, 36183679, 36183680, 36183681, 36183682, 36183683, 36183684, 36183685, 36183686, 36183687, 36183688, 36183689, 36183690, 36183691, 36183692, 36185777, 36185778, 36185779, 36185780, 36185781, 36185782, 36185980, 36185981, 36185982, 36185983, 36185984, 36185985, 36185986, 36185987, 36185988, 36185989, 36185990, 36185991, 36185992, 36185993, 36185994, 36185995, 36185996, 36185997, 36185998, 36185999, 36186000, 36186001, 36186002, 36186003, 36186004, 36186005, 36186006, 36186007, 36186008, 36186009, 36186010, 36186011, 36186012, 36186013, 36257776, 36257777, 36257778, 36257779, 36257780, 36257781, 36257782, 36257783, 36257784, 36257785, 36257786, 36257787, 36257788, 36257789, 36257790, 36257791, 36257741, 36257742, 36257743, 36257744, 36257745, 36257746, 36257747, 36257748, 36257749, 11316919, 11316918, 8571666, 112807, 36278952, 36278960, 36243688, 36254259, 36258438, 36227565, 36229549, 36229592, 36227345, 36302015, 36302018, 36302036, 36302037, 36212427, 36212429, 36212431, 36212453, 36212466, 36182225, 36234827, 36234868, 36234896, 36195633, 36233905, 36234899, 36234900, 36238197, 36238239, 36238254, 36255550, 36274537, 36274540, 36227565, 36156753, 36156798, 36156799, 36204582, 36200742, 36204604, 11402841, 11402842, 11402843, 36200733, 36200734, 36200738, 36265943, 36265944, 36265945, 36265946, 36265947, 36265948, 36265949, 36265951, 36265952, 36267416, 36176592, 36212441, 36212442, 36229549, 36229592, 36212467, 36234875, 36234877, 36173955, 9642655, 9709174, 9709176, 11656114, 11656115, 36207397, 36207398, 36225371, 36227488, 36227489, 36229602, 36231480, 36238304, 36238264, 36194091, 36176284, 36176285, 36176286, 36176287, 36181921, 36181922, 36181923, 36207440, 36207442, 36207443, 36207452, 36172107, 36172108, 36172109, 36172110, 36172111, 36204605, 36227316, 36227317, 36227318, 36227319, 36227320, 36227321, 36227322, 36227323, 36238185, 36238351, 36241513, 11711945, 11711946, 11711947, 11711948, 11711949, 11711950, 11711951, 11576879, 11576880, 11576881, 11576882, 36265966, 36306679, 36167748, 10996724, 11241880, 36274570, 36234878, 36234880, 36234889, 36195632, 36194147, 36194149, 36194157, 36194146, 36229627, 36229630, 36229631, 36229632, 36229633, 36229634, 36229635, 36229636, 36229637, 36229638, 36229639, 36229640, 36229641, 36229642, 36229643, 36229644, 36229645, 36229646, 36229647, 36229648, 36229649, 36229650, 36229651, 36229652, 36229653, 36229654, 36229655, 36229656, 36229657, 36229658, 36229659, 36229660, 36231846, 36227345, 36266095, 36204743, 36205330, 36205333, 36205337, 36207383, 36207384, 36204744, 36205329, 36205331, 36205338, 36278845, 36278846, 36257971, 36238190, 36238303, 36234898, 36238196, 36238215, 36238218, 36238219, 36238221, 36238249, 36170039, 36170045, 36170046, 36173680, 36173681, 36173682, 36173683, 36173684, 36173685, 36173686, 36173687, 36173688, 36173689, 36173690, 36173692, 36173693, 36173694, 36173695, 36173697, 36173698, 36173699, 36173700, 36173701, 36173702, 36173703, 36173704, 36173705, 36173956, 36173957, 36173958, 36227591, 36227631, 36173986, 36170017, 36170018, 36194133, 36194134, 36231582, 36231583, 36227462, 36234834, 36241076, 36227444, 36274548, 36274549, 36270437, 36302768, 36238329, 36238352, 36227562, 36227590, 36195628, 36303440, 36233889, 36227331, 36227332, 36227333, 36227334, 36227335, 36212501, 36194112, 36194119, 36194120, 36194121, 36194122, 36194123, 36194124, 36194142, 36194145, 36238263, 36241555, 36205332, 36301981, 11246582, 36274742, 36266109, 36266113, 9898739, 11329515, 36278850, 36270429, 36270430, 36212444, 36266103, 36167746, 36302038, 36302039, 36266673, 36266990, 36204603, 36278838, 36173983, 36306697, 36213542, 36213548, 36309944, 36266689, 36266691, 36238198, 36238211, 36270424, 36270425, 36163131, 36163162, 36163439, 36172238, 36172239, 36172240, 36172267, 36207376, 36229599, 36229600, 36169913, 36170014, 36170022, 36192946, 36192947, 36266686, 36204725, 36204726, 36274555, 36234005, 36238302, 36306687, 36183122, 36207404, 36207416, 36207420, 36207441, 36212428, 36212432, 36212436, 36212469, 36212470, 36212487, 36212492, 36212498, 36163163, 36182396, 36231824, 36231825, 36229665, 36182381, 36182388, 36204687, 36204697, 36227341, 36173989, 36173990, 36176103, 36240901, 36240902, 36302015, 36302018, 36302036, 36302037, 12439581, 12439582, 12439649, 36266114, 36266119, 36266120, 36266121, 36266126, 36267307, 36204600, 12437549, 36238230, 36229544, 36194148, 36227574, 36266685, 36205336, 36207417, 36207418, 36207419, 36212427, 36212429, 36212431, 36212453, 36212466, 36212430, 31533324, 36227447, 36212462, 36241546, 36194155, 36195626, 36195627, 36195629, 36195630, 36201116, 36302051, 11108635, 11145972, 36172266, 36172275, 36174568, 36176884, 36302803, 36302804, 36302805, 36170041, 36233908, 12734960, 13092386, 13102395, 36212461, 36234876, 36227679, 36274574, 36234894, 36234895, 36212424, 36212509, 36172261, 36172262, 36172263, 36306693, 36234829, 36238296, 36238297, 36241514, 10487043, 36231828, 36231829, 36176102, 36176105, 36170029, 36266687, 36229684, 36229685, 36229686, 36229687, 36229688, 36229690, 36182397, 36241547, 36238255, 36241558, 36238235, 36233913, 36176106, 36233884, 11735830, 11735832, 11735834, 11735835, 11735836, 11896923, 36172247, 36172248, 36172249, 36172250, 36172251, 36270435, 36270436, 36241556, 36241557, 36213541, 36238206, 36182225, 36309945, 36309946, 36309947, 36309948, 36309952, 36309953, 36238267, 36241531, 36234827, 36234868, 36234896, 36266684, 36212450, 36193275, 36194111, 36234828, 36170021, 36233903, 36233904, 36233961, 36193274, 36194151, 36194213, 36195631, 36195634, 36233891, 36229585, 36229586, 36167671, 36234897, 36156731, 36212468, 36212505, 13601705, 36170044, 36170043, 36266111, 36212507, 36212513, 36227618, 9898671, 13978203, 11427794, 11427795, 12539641, 30572994, 36233911, 36156730, 36156732, 36234879, 36274569, 36303748, 36303749, 36156790, 36152511, 12916278, 36183534, 36183535, 36185799, 36170038, 36276927, 36212451, 36212446, 36213153, 36212437, 36170040, 36205334, 36205335, 36227454, 36227455, 36227456, 36192823, 36192824, 36192856, 12222828, 36212449, 36212500, 11763411, 36192985, 36192986, 36182231, 36182310, 36182313, 36182346, 36227329, 36301978, 36266672, 36267185, 36267308, 36204750, 36204755, 36195633, 36229593, 36229595, 36229689, 36229691, 36229694, 36229695, 36229696, 36229697, 36229698, 36229699, 36229589, 36229588, 36229590, 36229615, 36274541, 36238273, 36207392, 36167667, 36227566, 36265961, 36227463, 36251945, 36251946, 36251947, 36156795, 36156796, 36238234, 36241532, 36241539, 36170042, 9731552, 36192905, 36192931, 36227678, 36227680, 36227681, 36227682, 36227683, 36227684, 36227685, 36227686, 36227687, 36227688, 36229596, 36270411, 36217789, 36266688, 36266704, 36233905, 36234899, 36234900, 36238197, 36238239, 36238254, 11366028, 11366029, 11926413, 11989509, 13086995, 13086996, 13955652, 13963103, 30606450, 36302050, 36182327, 36176889, 36238238, 36306688, 36306700, 11693630, 36173691, 36173696, 36173706, 36173953, 36173954, 36174016, 36238223, 36181559, 12371117, 12371118, 12371119, 36251941, 36251942, 11025131, 11106124, 11106128, 13025505, 11025129, 11256361, 36200722, 36200723, 36200724, 36200725, 36200726, 36200727, 36200728, 36200729, 36200730, 36200731, 36200732, 36200735, 36200736, 36200737, 36200739, 36200740, 36200741, 36200743, 36201975, 36201976, 36201978, 36203035, 36200744, 36266124, 13433377, 36182347, 36192804, 36192805, 36233895, 36233896, 36233897, 36233898, 36233899, 36233900, 36301986, 36301987, 36306665, 36234000, 36175985, 36176101, 36192930, 36192984, 36274580, 12433773, 12437231, 13422371, 36192803, 36192890, 36192904, 36241538, 36167799, 36170030, 36212421, 36212422, 36212423, 36212425, 36212426, 36176104, 36163164, 36170023, 10100952, 36212508, 36214148, 36214149, 36214150, 36214151, 36204608, 36227337, 36212506, 36229542, 36255550, 36302750, 36266107, 36302795, 36302783, 36278841, 36182345, 36182357, 12447739, 36274561, 36274564, 36194080, 36227466, 36227459, and 36207377. |
| FEI Number |
3015876
|
Recalling Firm/ Manufacturer |
Physio Control, Inc.
11811 Willows Rd Ne
Redmond
WA
98052-2003
|
| For Additional Information Contact | Todd Bandy 425-867-4000 |
Manufacturer Reason for Recall | The LIFEPAK12 defibrillator/monitor series (LP12) contain a Biphasic PCBA that may have a solder defect on the H Bridge component that could prevent the device from providing a defibrillation shock. |
FDA Determined Cause 2 | Process control |
| Action | Starting May 8, 2008, letters were hand-delivered by the firm's service representatives as they visited customer sites to perform on-site service for affected units. Customers who had not been visited by May 16th were mailed a Certified Receipt letter on May 16, 2008. The letter explained the recall and notified customers that Physio Control was conducting a Field correction to replace all affected Biphasic PCBA. The firm recommends customers continue to perform therapy cable or Standard Paddles Defibrillation Check (monthly or as needed); to have a backup device available until their device is updated; and if no backup device is available, a loaner would be provided by the firm upon request.
Contact Technical Support at 1-800-442-1142, option 5 - 6:00 am to 4:00 pm (Pacific), Monday -Friday or visit www.Physio-Control-notices.com/LP12. |
| Quantity in Commerce | 1,710 Units |
| Distribution | Product distributed nationwide to 5 Department of Defense accounts, 1 VA account, 5 government accounts, 49 hospitals, 223 first responders, and 1 direct account. 8 foreign accounts include: NETHERLANDS, MEXICO, JAPAN, HONG KONG, GERMANY, CANADA and AUSTRALIA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|