• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Command Module

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Command Modulesee related information
Date Initiated by FirmJuly 22, 2008
Date PostedSeptember 19, 2008
Recall Status1 Terminated 3 on July 25, 2011
Recall NumberZ-2342-2008
Recall Event ID 48759
510(K)NumberK972502 
Product Classification Arrhythmia Detector and alarm - Product Code DSI
ProductUltraview SL Command Module, Model 90496 The product is intended for use with the Patient Care Monitoring System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use. Physiological parameters that may be monitored include cardiac activity, respiration, invasive and noninvasive pressure, temperature, oxygen saturation (SpO2) and cardiac output. Acquired data may then be communicated to an information network for display, recording, editing and analysis.
Code Information Serial Numbers 496-053539, 496-053540, 496-053541, 496-053591, 496-053592, 496-053593, 496-053595, 496-053596, 496-053542, 496-053551, 496-053552, 496-053547, 496-053548, 496-053459, 496-053460, 496-053461, 496-053462, 496-053463, 496-053464, 496-053544, 496-053545, 496-053546, 496-053467, 496-053468, 496-053469, 496-053471, 496-053457, 496-053458, 496-053388, 496-053389, 496-053370, 496-053275, 496-053276, 496-053277, 496-053278, 496-053279, 496-053280, 496-053281, 496-053282, 496-053283, 496-053286, 496-053304, 496-053305, 496-053306, 496-053307, 496-053308, 496-053309, 496-053310, 496-053311, 496-053312, 496-053313, 496-053314, 496-053315, 496-053316, 496-053317, 496-053253, 496-053254, 496-053255, 496-053256, 496-053257, 496-053258, 496-053259, 496-053260, 496-053261, 496-053262, 496-053263, 496-053264, 496-053265, 496-053266, 496-053267, 496-053268, 496-053269, 496-053270, 496-053271, 496-053272, 496-053273, 496-053274, 496-053227, 496-053228, 496-053230, 496-052754, 496-052755, 496-052756, 496-053014, 496-053015, 496-053016, 496-052752, 496-052753, 496-052846, 496-052549, 496-052720, 496-052701, 496-052702, 496-052703, 496-052700, 496-052503, 496-052508, 496-052544, 496-052545, 496-052546, 496-052547, 496-052548, 496-052549, 496-052505, 496-052506, 496-052507, 496-052478, 496-052479, 496-052481, 496-052462, 496-052463, 496-052464, 496-052446, 496-052447, 496-052443, 496-051884, 496-051888, 496-051889, 496-052409, 496-052410, 496-052412, 496-052413, 496-052414, 496-052415, 496-051883, 496-051885, 496-051886, 496-051887, 496-052129, 496-052180, 496-052177, 496-052178, 496-052179, 496-052111, 496-052112, 496-052113, 496-052114, 496-052115, 496-052116, 496-052117, 496-052041, 496-052042, 496-052043, 496-052044, 496-052045, 496-052046, 496-052047, 496-052048, 496-052049, 496-049665, 496-051941, 496-051976, 496-051977, 496-051978, 496-051979, 496-051980, 496-051940, 496-051934, 496-051933, 496-051935, 496-051953, 496-051954, 496-051955, 496-051956, 496-051872, 496-051873, 496-051909, 496-051910, 496-051911, 496-051912, 496-051913, 496-051914, 496-051915, 496-051794, 496-051795, 496-051690, 496-051691, 496-051643, 496-051644, 496-051420, 496-051675, 496-051658, 496-051659, 496-051660, 496-051661, 496-051662, 496-051663, 496-051664, 496-051665, 496-051666, 496-051667, 496-051668, 496-051669, 496-051670, 496-051671, 496-051672, 496-051673, 496-051674, 496-051523, 496-051524, 496-051525, 496-051526, 496-051184, 496-051185, 496-051186, 496-051328, 496-051202, 496-050324, 496-050325, 496-050326, 496-051180, 496-051181, 496-051182, 496-051052, 496-051053, 496-051064, 496-051031, 496-049083, 496-050322, 496-050611, 496-050612, 496-050613, 496-050614, 496-050624, 496-050625, 496-050538, 496-050389, 496-050183, 496-050184, 496-049921, 496-049926, 496-049961, 496-050012, 496-050011, 496-049449, 496-049450, 496-049451, 496-049433, 496-049435, 496-049314, 496-049315, 496-049252, 496-049254, 496-049178, 496-049179, 496-049180, 496-049181, 496-049080, 496-049081, 496-049056, 496-049057, 496-049058, 496-049059, 496-049060, 496-049061, 496-049062, 496-049063, 496-049064, 496-049065, 496-049066, 496-049067, 496-049068, 496-049069, 496-049070, 496-049071, 496-049072, 496-049073, 496-049074, 496-049075, 496-049076, 496-049077, 496-049078, 496-049079, 496-048974, 496-048975, 496-048976, 496-048977, 496-048978, 496-048979, 496-048982, 496-048983, 496-048984, 496-048985, 496-048699, 496-048700, 496-047221, 496-047222, 496-047223, 496-047224, 496-047225, 496-048684, 496-048685, 496-048561, 496-048562, 496-047958, 496-048053, 496-048055, 496-047229, 496-047230, 496-048007, 496-048008, 496-048009, 496-048010, 496-048011, 496-048012, 496-048013, 496-048014, 496-048015, 496-048016, 496-048017, 496-048018, 496-048019, 496-048020, 496-048021, 496-048022, 496-047880, 496-047881, 496-047875, 496-047994, 496-047995, 496-047996, 496-047997, 496-047998, 496-047999, 496-048000, 496-048001, 496-048002, 496-048003, 496-048005, 496-048006, 496-047806, 496-047807, 496-047808, 496-047809, 496-047810, 496-047811, 496-047812, 496-047813, 496-047814, 496-047815, 496-047816, 496-047817, 496-047818, 496-047819, 496-047820, 496-047821, 496-047822, 496-047823, 496-047824, 496-047825, 496-047826, 496-047827, 496-047828, 496-047829, 496-047830, 496-047831, 496-047832, 496-047833, 496-047834, 496-047835, 496-047836, 496-047837, 496-047838, 496-047839, 496-047840, 496-047841, 496-047842, 496-047843, 496-047844, 496-047845, 496-047846, 496-047847, 496-047848, 496-047849, 496-047851, 496-047852, 496-047853, 496-047854, 496-047855, 496-047856, 496-047857, 496-047785, 496-047786, 496-047787, 496-047788, 496-047789, 496-047759, 496-047760, 496-047761, 496-047762, 496-047763, 496-047764, 496-047765, 496-047766, 496-047562, 496-047615, 496-047391, 496-047409, 496-047439, 496-047440, 496-047441, 496-047442, 496-047443, 496-047444, 496-047445, 496-047446, 496-047447, 496-045978, 496-046789, 496-046790, 496-047073, 496-047074, 496-047075, 496-047076, 496-047077, 496-047078, 496-047079, 496-047080, 496-047081, 496-047082, 496-047083, 496-047084, 496-047168, 496-047169, 496-047170, 496-046830, 496-046831, 496-046832, 496-046833, 496-046834, 496-046835, 496-046836, 496-046837, 496-046838, 496-046839, 496-046840, 496-046841, 496-046736, 496-046660, 496-046661, 496-046662, 496-046510, 496-046328, 496-045980, 496-045983, 496-046176, 496-046177, 496-046181, 496-046182, 496-046185, 496-046187, 496-046191, 496-046225, 496-046229, 496-046237, 496-046241, 496-046247, 496-046249, 496-046255, 496-046256, 496-046261, 496-046266, 496-046270, 496-046272, 496-046273, 496-046239, 496-046248, 496-045995, 496-046021, 496-046025, 496-046030, 496-046031, 496-046039, 496-046040, 496-046042, 496-046047, 496-046048, 496-046049, 496-046050, 496-046052, 496-046090, 496-046192, 496-045903, 496-046131, 496-045823, 496-045888, 496-045896, 496-045904, 496-045906, 496-045917, 496-043263, 496-043950, 496-043953, 496-043956, 496-043957, 496-043961, 496-043969, 496-043972, 496-043973, 496-045982, 496-045859, 496-045881, 496-045882, 496-045889, 496-045908, 496-045909, 496-045663, 496-045675, 496-045676, 496-045695, 496-045698, 496-045753, 496-045761, 496-045762, 496-045766, 496-045772, 496-045776, 496-045779, 496-045781, 496-045782, 496-045783, 496-045784, 496-045786, 496-045787, 496-045788, 496-045791, 496-045795, 496-045796, 496-045797, 496-045799, 496-045800, 496-045801, 496-045802, 496-045803, 496-045805, 496-045809, 496-045810, 496-045815, 496-045816, 496-045817, 496-045819, 496-045820, 496-045821, 496-045822, 496-045824, 496-045825, 496-045826, 496-045827, 496-045830, 496-045832, 496-045835, 496-045836, 496-045837, 496-045839, 496-045840, 496-045842, 496-045846, 496-045850, 496-045852, 496-045674, 496-045678, 496-043955, 496-043960, 496-043965, 496-045594, 496-045635, 496-045647, 496-045553, 496-045556, 496-045562, 496-045573, 496-045574, 496-045575, 496-045576, 496-045582, 496-045583, 496-045585, 496-045586, 496-045607, 496-045608, 496-045609, 496-045611, 496-045642, 496-045644, 496-045648, 496-045552, 496-045597, 496-053522, 496-053534, 496-053537, 496-053538, 496-053465, 496-053466, 496-053484, 496-053519, 496-053520, 496-053521, 496-053523, 496-053524, 496-053525, 496-053526, 496-053527, 496-053528, 496-053529, 496-053530, 496-053531, 496-053532, 496-053533, 496-053535, 496-053536, 496-053543, 496-053512, 496-053513, 496-053514, 496-053515, 496-053516, 496-053517, 496-053518, 496-053500, 496-053501, 496-053502, 496-053503, 496-053504, 496-053505, 496-053506, 496-053507, 496-053508, 496-053509, 496-053510, 496-053511, 496-053499, 496-053498, 496-053489, 496-053487, 496-053488, 496-053483, 496-053390, 496-053470, 496-053481, 496-053482, 496-053430, 496-053450, 496-053453, 496-053456, 496-053455, 496-053437, 496-053438, 496-053439, 496-053440, 496-053441, 496-053442, 496-053443, 496-053444, 496-053445, 496-053446, 496-053447, 496-053448, 496-053449, 496-053451, 496-053452, 496-053454, 496-053433, 496-053434, 496-053435, 496-053436, 496-053432, 496-053329, 496-053331, 496-053334, 496-053367, 496-053398, 496-053399, 496-053401, 496-053402, 496-053375, 496-053379, 496-053384, 496-053387, 496-053391, 496-053392, 496-053393, 496-053395, 496-053396, 496-053397, 496-053252, 496-053293, 496-053322, 496-053328, 496-053247, 496-053340, 496-052752, 496-052753, 496-053048, 496-053368, 496-053369, 496-053250, 496-053251, 496-053244, 496-053245, 496-053239, 496-053240, 496-053241, 496-053242, 496-053243, 496-053234, 496-053235, 496-053236, 496-053237, 496-053238, 496-053231, 496-053232, 496-053204, 496-053205, 496-053206, 496-053207, 496-053208, 496-053209, 496-053210, 496-053211, 496-053212, 496-053213, 496-053214, 496-053217, 496-053218, 496-053219, 496-053220, 496-053221, 496-053222, 496-053223, 496-053224, 496-053225, 496-053226, 496-053215, 496-053216, 496-053149, 496-053150, 496-053151, 496-053152, 496-053153, 496-053154, 496-053155, 496-053156, 496-053157, 496-053158, 496-053159, 496-053160, 496-053161, 496-053162, 496-053163, 496-053164, 496-053165, 496-053168, 496-053169, 496-053170, 496-053171, 496-053172, 496-053173, 496-053174, 496-053175, 496-053176, 496-053177, 496-053178, 496-053179, 496-053180, 496-053181, 496-053182, 496-053183, 496-053184, 496-053192, 496-053193, 496-053194, 496-053195, 496-053196, 496-053197, 496-053198, 496-053199, 496-053200, 496-053201, 496-053202, 496-053203, 496-052416, 496-053166, 496-053167, 496-053185, 496-053186, 496-053187, 496-053188, 496-053189, 496-053190, 496-053191, 496-053147, 496-053148, 496-053017, 496-053049, 496-053143, 496-053144, 496-053145, 496-053146, 496-053113, 496-053116, 496-053142, 496-053127, 496-053128, 496-053129, 496-053130, 496-053131, 496-053132, 496-053047, 496-053100, 496-053101, 496-053102, 496-053103, 496-053104, 496-053105, 496-053106, 496-053107, 496-053108, 496-053109, 496-053110, 496-053111, 496-053112, 496-053114, 496-053115, 496-053117, 496-053118, 496-053119, 496-053120, 496-053121, 496-053122, 496-053123, 496-053124, 496-053125, 496-053126, 496-053083, 496-053093, 496-053095, 496-053096, 496-053097, 496-053098, 496-053099, 496-053080, 496-053081, 496-053082, 496-053087, 496-053088, 496-053089, 496-053090, 496-053091, 496-053092, 496-053094, 496-053059, 496-053061, 496-053062, 496-053063, 496-053064, 496-053065, 496-053066, 496-053067, 496-053068, 496-053069, 496-053070, 496-053071, 496-053072, 496-053073, 496-053074, 496-053075, 496-053076, 496-053077, 496-053078, 496-053079, 496-053060, 496-053084, 496-053085, 496-053086, 496-053044, 496-053045, 496-053057, 496-053058, 496-053017, 496-053049, 496-053046, 496-053012, 496-053013, 496-053018, 496-053019, 496-053020, 496-053021, 496-053041, 496-053042, 496-053043, 496-053031, 496-053032, 496-053033, 496-053034, 496-053035, 496-053036, 496-053037, 496-053038, 496-053039, 496-053040, 496-052953, 496-053022, 496-053023, 496-053024, 496-053025, 496-053026, 496-053027, 496-053028, 496-053029, 496-053030, 496-053011, 496-053008, 496-053009, 496-053010, 496-053005, 496-053006, 496-053007, 496-051919, 496-051920, 496-051921, 496-051922, 496-052948, 496-052954, 496-052955, 496-052956, 496-052957, 496-052958, 496-052959, 496-052960, 496-052961, 496-052962, 496-052963, 496-052964, 496-052965, 496-052966, 496-052967, 496-052968, 496-052969, 496-052970, 496-052971, 496-052972, 496-052973, 496-052974, 496-052975, 496-052976, 496-052977, 496-052978, 496-052979, 496-052980, 496-052981, 496-052982, 496-052983, 496-052984, 496-052985, 496-052986, 496-052987, 496-052988, 496-052989, 496-052990, 496-052991, 496-052992, 496-052993, 496-052994, 496-052995, 496-052996, 496-052997, 496-052998, 496-052999, 496-053000, 496-053001, 496-053002, 496-053003, 496-053004, 496-051900, 496-051901, 496-051903, 496-051904, 496-051905, 496-051916, 496-051917, 496-051918, 496-051893, 496-051894, 496-051895, 496-051896, 496-051897, 496-051898, 496-051899, 496-051902, 496-052952, 496-052951, 496-052927, 496-052930, 496-052931, 496-052932, 496-052933, 496-052934, 496-052935, 496-052936, 496-052937, 496-052938, 496-052940, 496-052943, 496-052944, 496-052945, 496-052946, 496-052947, 496-052949, 496-052950, 496-052939, 496-052941, 496-052942, 496-052912, 496-052918, 496-052924, 496-052926, 496-052928, 496-052929, 496-052901, 496-052887, 496-052891, 496-052892, 496-052893, 496-052894, 496-052895, 496-052896, 496-052899, 496-052900, 496-052902, 496-052903, 496-052904, 496-052905, 496-052906, 496-052907, 496-052908, 496-052909, 496-052910, 496-052911, 496-052913, 496-052914, 496-052915, 496-052916, 496-052917, 496-052919, 496-052920, 496-052921, 496-052922, 496-052923, 496-052925, 496-052886, 496-052888, 496-052889, 496-052890, 496-052897, 496-052898, 496-052872, 496-052873, 496-052874, 496-052875, 496-052876, 496-052877, 496-052878, 496-052879, 496-052880, 496-052881, 496-052882, 496-052883, 496-052884, 496-052885, 496-052847, 496-052848, 496-052849, 496-052850, 496-052851, 496-052852, 496-052853, 496-052854, 496-052855, 496-052856, 496-052857, 496-052858, 496-052859, 496-052860, 496-052861, 496-052862, 496-052863, 496-052864, 496-052865, 496-052866, 496-052867, 496-052868, 496-052869, 496-052870, 496-052871, 496-052838, 496-052839, 496-052840, 496-052841, 496-052842, 496-052843, 496-052844, 496-052845, 496-052833, 496-052830, 496-052836, 496-052837, 496-051955, 496-052793, 496-052794, 496-052795, 496-052796, 496-052797, 496-052798, 496-052799, 496-052800, 496-052801, 496-052802, 496-052803, 496-052804, 496-052805, 496-052806, 496-052807, 496-052808, 496-052809, 496-052810, 496-052811, 496-052792, 496-052704, 496-052790, 496-052791, 496-052721, 496-052722, 496-052682, 496-052699, 496-052705, 496-052706, 496-052707, 496-052708, 496-052709, 496-052710, 496-052711, 496-052712, 496-052713, 496-052714, 496-052715, 496-052716, 496-052717, 496-052718, 496-052719, 496-052545, 496-052546, 496-052547, 496-052548, 496-052544, 496-052727, 496-052728, 496-052729, 496-052730, 496-052731, 496-052732, 496-052733, 496-052734, 496-052689, 496-052735, 496-052736, 496-052681, 496-052661, 496-052662, 496-052663, 496-052664, 496-052665, 496-052666, 496-052667, 496-052668, 496-052669, 496-052670, 496-052671, 496-052672, 496-052673, 496-052674, 496-052675, 496-052676, 496-052677, 496-052678, 496-052679, 496-052680, 496-052533, 496-052534, 496-052535, 496-052536, 496-052537, 496-052538, 496-052539, 496-052540, 496-052541, 496-052660, 496-052532, 496-052604, 496-052605, 496-052606, 496-052607, 496-052608, 496-052609, 496-052610, 496-052611, 496-052612, 496-052613, 496-052614, 496-052615, 496-052616, 496-052617, 496-052618, 496-052619, 496-052620, 496-052621, 496-052622, 496-052623, 496-052624, 496-052625, 496-052626, 496-052627, 496-052628, 496-052629, 496-052630, 496-052631, 496-052632, 496-052633, 496-052634, 496-052635, 496-052636, 496-052637, 496-052638, 496-052639, 496-052640, 496-052641, 496-052642, 496-052643, 496-052644, 496-052645, 496-052646, 496-052647, 496-052648, 496-052649, 496-052650, 496-052651, 496-052652, 496-052653, 496-052654, 496-052655, 496-052656, 496-052657, 496-052658, 496-052659, 496-052531, 496-052527, 496-052683, 496-052684, 496-052685, 496-052686, 496-052687, 496-052688, 496-052690, 496-052691, 496-052692, 496-052693, 496-052694, 496-052695, 496-052696, 496-052697, 496-052698, 496-052603, 496-052596, 496-052597, 496-052602, 496-052601, 496-052598, 496-052599, 496-052600, 496-052595, 496-052543, 496-052515, 496-052523, 496-052524, 496-052525, 496-052526, 496-052542, 496-052502, 496-052511, 496-052512, 496-052529, 496-052530, 496-052528, 496-052521, 496-052522, 496-052513, 496-052514, 496-052516, 496-052517, 496-052518, 496-052520, 496-052491, 496-052492, 496-052493, 496-052494, 496-052495, 496-052496, 496-052497, 496-052498, 496-052509, 496-052510, 496-052480, 496-052485, 496-052489, 496-052490, 496-052482, 496-052483, 496-052484, 496-052474, 496-052476, 496-052411, 496-052456, 496-052471, 496-052472, 496-052473, 496-052475, 496-052465, 496-052466, 496-052467, 496-052468, 496-052469, 496-052470, 496-052454, 496-052455, 496-052458, 496-052459, 496-052460, 496-052457, 496-048598, 496-048599, 496-048935, 496-049250, 496-049251, 496-052449, 496-052450, 496-052451, 496-052452, 496-052453, 496-048932, 496-050181, 496-050182, 496-052448, 496-052444, 496-052437, 496-052441, 496-052442, 496-052435, 496-052434, 496-052431, 496-052432, 496-052433, 496-052416, 496-052417, 496-052418, 496-052404, 496-052403, 496-052407, 496-052408, 496-052405, 496-052406, 496-052398, 496-052401, 496-052402, 496-052399, 496-052400, 496-052390, 496-052393, 496-052394, 496-052395, 496-052396, 496-052397, 496-052391, 496-052392, 496-051221, 496-052389, 496-052339, 496-052340, 496-052341, 496-052342, 496-052343, 496-052386, 496-052387, 496-052388, 496-052279, 496-052288, 496-052317, 496-052318, 496-052319, 496-052320, 496-052321, 496-052322, 496-052323, 496-052324, 496-052326, 496-052327, 496-052328, 496-052329, 496-052330, 496-052331, 496-052371, 496-052372, 496-052373, 496-052374, 496-052375, 496-052376, 496-052377, 496-052378, 496-052379, 496-052380, 496-052381, 496-052382, 496-052383, 496-052384, 496-052385, 496-052332, 496-052333, 496-052334, 496-052335, 496-052336, 496-052337, 496-052338, 496-052325, 496-052344, 496-052345, 496-052346, 496-052347, 496-052348, 496-052349, 496-052350, 496-052351, 496-052352, 496-052353, 496-052354, 496-052355, 496-052356, 496-052358, 496-052359, 496-052360, 496-052361, 496-052362, 496-052363, 496-052364, 496-052365, 496-052366, 496-052367, 496-051882, 496-052268, 496-052275, 496-052276, 496-052277, 496-052278, 496-052280, 496-052281, 496-052282, 496-052283, 496-052284, 496-052285, 496-052286, 496-052287, 496-052289, 496-052290, 496-052291, 496-052292, 496-052293, 496-052294, 496-052295, 496-052296, 496-052297, 496-052298, 496-052299, 496-052300, 496-052301, 496-052302, 496-052303, 496-052304, 496-052305, 496-052306, 496-052307, 496-052308, 496-052309, 496-052310, 496-052311, 496-052312, 496-052313, 496-052314, 496-052315, 496-052316, 496-052162, 496-052169, 496-052240, 496-052250, 496-052262, 496-052263, 496-052264, 496-052265, 496-052266, 496-052267, 496-052269, 496-052270, 496-052271, 496-052272, 496-052273, 496-052274, 496-052199, 496-052200, 496-052201, 496-052202, 496-052203, 496-052204, 496-052205, 496-052206, 496-052207, 496-052208, 496-052209, 496-052210, 496-052211, 496-052212, 496-052213, 496-052214, 496-052215, 496-052216, 496-052217, 496-052218, 496-052219, 496-052220, 496-052221, 496-052222, 496-052223, 496-052224, 496-052225, 496-052226, 496-052227, 496-052228, 496-052229, 496-052230, 496-052231, 496-052232, 496-052233, 496-052234, 496-052235, 496-052236, 496-052237, 496-052238, 496-052239, 496-052241, 496-052242, 496-052243, 496-052244, 496-052245, 496-052246, 496-052247, 496-052248, 496-052249, 496-052251, 496-052252, 496-052253, 496-052254, 496-052255, 496-052256, 496-052257, 496-052258, 496-052259, 496-052260, 496-052261, 496-052181, 496-052182, 496-052183, 496-052184, 496-052185, 496-052186, 496-052187, 496-052188, 496-052189, 496-052190, 496-052191, 496-052192, 496-052194, 496-052195, 496-052196, 496-052197, 496-052118, 496-052119, 496-052120, 496-052121, 496-052122, 496-052123, 496-052124, 496-052125, 496-052126, 496-052127, 496-052128, 496-052130, 496-052131, 496-052132, 496-052133, 496-052135, 496-052136, 496-052137, 496-052138, 496-052139, 496-052140, 496-052141, 496-052142, 496-052143, 496-052144, 496-052145, 496-052146, 496-052147, 496-052148, 496-052149, 496-052150, 496-052151, 496-052152, 496-052153, 496-052154, 496-052155, 496-052156, 496-052157, 496-052158, 496-052159, 496-052160, 496-052092, 496-052175, 496-052176, 496-052134, 496-052161, 496-052163, 496-052164, 496-052165, 496-052166, 496-052167, 496-052168, 496-052170, 496-052171, 496-052172, 496-052173, 496-052174, 496-052083, 496-052110, 496-052093, 496-052094, 496-052095, 496-052096, 496-052097, 496-052098, 496-052099, 496-052100, 496-052101, 496-052102, 496-052103, 496-052104, 496-052105, 496-052106, 496-052107, 496-052108, 496-052109, 496-052010, 496-052050, 496-052051, 496-052052, 496-052053, 496-052054, 496-052055, 496-052056, 496-052057, 496-052058, 496-052059, 496-052060, 496-052061, 496-052062, 496-052063, 496-052064, 496-052065, 496-052066, 496-052067, 496-052068, 496-052069, 496-052070, 496-052071, 496-052072, 496-052073, 496-052074, 496-052075, 496-052076, 496-052077, 496-052078, 496-052079, 496-052080, 496-052081, 496-052082, 496-052088, 496-052089, 496-052090, 496-052084, 496-052085, 496-052086, 496-052087, 496-049873, 496-052040, 496-052008, 496-052009, 496-052011, 496-052012, 496-052013, 496-052014, 496-052015, 496-052016, 496-052017, 496-052018, 496-052019, 496-052020, 496-052021, 496-052022, 496-052023, 496-052024, 496-052025, 496-052026, 496-052027, 496-052028, 496-052029, 496-052030, 496-052031, 496-052032, 496-052033, 496-052034, 496-052035, 496-052036, 496-052037, 496-052038, 496-052039, 496-051880, 496-051881, 496-045560, 496-048104, 496-052003, 496-052004, 496-052005, 496-052006, 496-052007, 496-051945, 496-051946, 496-051943, 496-051944, 496-051936, 496-051981, 496-051982, 496-051942, 496-051937, 496-051938, 496-051936, 496-051926, 496-051930, 496-051931, 496-051932, 496-051893, 496-051894, 496-051895, 496-051896, 496-051897, 496-051898, 496-051899, 496-051901, 496-051902, 496-051903, 496-051904, 496-051905, 496-051949, 496-051950, 496-051923, 496-051924, 496-051925, 496-051874, 496-051875, 496-051876, 496-051877, 496-051878, 496-051879, 496-051890, 496-051891, 496-051892, 496-051900, 496-051916, 496-051917, 496-051918, 496-051919, 496-051920, 496-051921, 496-051922, 496-051926, 496-051927, 496-051928, 496-051929, 496-051930, 496-051931, 496-051932, 496-051933, 496-051934, 496-051935, 496-051936, 496-051937, 496-051938, 496-051940, 496-051941, 496-051942, 496-051943, 496-051944, 496-051945, 496-051946, 496-051871, 496-051906, 496-051907, 496-051908, 496-051860, 496-051861, 496-051862, 496-051863, 496-051864, 496-051865, 496-051866, 496-051867, 496-051868, 496-051869, 496-051870, 496-051811, 496-051812, 496-051813, 496-051814, 496-051815, 496-051816, 496-051817, 496-051818, 496-051819, 496-051820, 496-051821, 496-051822, 496-051823, 496-051824, 496-051825, 496-051826, 496-051827, 496-051828, 496-051829, 496-051830, 496-051831, 496-051796, 496-051797, 496-051798, 496-051799, 496-051800, 496-051801, 496-051802, 496-051803, 496-051804, 496-051805, 496-051806, 496-051807, 496-051808, 496-051809, 496-051810, 496-051759, 496-051760, 496-051761, 496-051762, 496-051763, 496-051786, 496-051787, 496-051788, 496-051789, 496-051790, 496-051774, 496-051775, 496-051776, 496-051777, 496-051778, 496-051779, 496-051780, 496-051781, 496-051782, 496-051783, 496-051784, 496-051785, 496-051771, 496-051772, 496-051773, 496-051765, 496-051766, 496-051767, 496-051768, 496-051769, 496-051770, 496-051743, 496-051744, 496-051745, 496-051746, 496-051747, 496-051748, 496-051749, 496-051722, 496-051723, 496-051724, 496-051725, 496-051726, 496-051727, 496-051728, 496-051729, 496-051730, 496-051731, 496-051732, 496-051733, 496-051734, 496-051735, 496-051736, 496-051737, 496-051738, 496-051739, 496-051740, 496-051741, 496-051742, 496-051654, 496-051699, 496-051700, 496-051701, 496-051702, 496-051703, 496-051704, 496-051705, 496-051706, 496-051707, 496-051708, 496-051709, 496-051710, 496-051711, 496-051712, 496-051713, 496-051679, 496-051680, 496-051681, 496-051682, 496-051683, 496-051684, 496-051685, 496-051686, 496-051687, 496-051688, 496-051689, 496-051692, 496-051693, 496-051694, 496-051695, 496-051696, 496-051697, 496-051698, 496-051653, 496-051655, 496-051656, 496-051657, 496-051648, 496-051649, 496-051650, 496-051651, 496-051652, 496-051645, 496-051646, 496-051647, 496-051608, 496-051609, 496-051610, 496-051611, 496-051612, 496-051613, 496-051603, 496-051604, 496-051605, 496-051606, 496-051607, 496-051477, 496-051478, 496-051614, 496-051615, 496-051616, 496-051594, 496-051595, 496-051596, 496-051597, 496-051504, 496-050699, 496-051592, 496-051593, 496-051476, 496-051676, 496-051677, 496-051566, 496-051551, 496-051552, 496-051553, 496-051554, 496-051555, 496-051556, 496-051581, 496-051582, 496-051583, 496-051536, 496-051537, 496-051538, 496-051539, 496-051540, 496-051541, 496-051542, 496-051543, 496-051544, 496-051545, 496-051546, 496-051547, 496-051548, 496-051549, 496-051550, 496-051527, 496-050669, 496-050692, 496-050693, 496-051449, 496-051509, 496-050700, 496-050701, 496-050702, 496-050703, 496-051528, 496-051529, 496-051530, 496-051531, 496-051532, 496-051533, 496-051534, 496-051515, 496-051516, 496-051517, 496-051518, 496-051519, 496-051520, 496-051521, 496-051522, 496-051481, 496-051482, 496-051483, 496-051484, 496-051485, 496-051486, 496-051487, 496-051488, 496-051489, 496-051490, 496-051491, 496-051492, 496-051493, 496-051494, 496-051495, 496-051496, 496-051497, 496-051498, 496-051499, 496-051500, 496-051501, 496-051502, 496-051503, 496-051466, 496-051467, 496-051468, 496-051469, 496-051470, 496-051471, 496-051472, 496-051473, 496-051474, 496-051475, 496-051177, 496-051178, 496-051179, 496-051190, 496-051191, 496-051192, 496-051193, 496-051194, 496-051195, 496-051196, 496-051197, 496-051198, 496-051216, 496-051217, 496-051218, 496-051219, 496-051454, 496-051455, 496-051456, 496-051457, 496-051458, 496-051459, 496-051460, 496-051461, 496-051199, 496-051200, 496-051201, 496-051203, 496-051204, 496-051205, 496-051206, 496-051207, 496-051208, 496-051209, 496-051210, 496-051211, 496-051212, 496-051213, 496-051371, 496-051424, 496-051189, 496-051426, 496-051427, 496-051428, 496-051429, 496-051430, 496-051431, 496-051432, 496-051433, 496-051434, 496-051435, 496-051436, 496-051437, 496-051438, 496-051439, 496-051440, 496-051441, 496-051442, 496-051443, 496-051444, 496-051445, 496-051446, 496-051447, 496-051425, 496-051187, 496-051188, 496-051370, 496-051373, 496-051417, 496-051418, 496-051419, 496-051421, 496-051422, 496-051416, 496-051374, 496-051375, 496-051376, 496-051377, 496-051378, 496-051379, 496-051380, 496-051381, 496-051382, 496-051383, 496-051384, 496-051385, 496-051386, 496-051387, 496-051388, 496-051389, 496-051390, 496-051391, 496-051392, 496-051393, 496-051394, 496-051395, 496-051396, 496-051397, 496-051398, 496-051399, 496-051400, 496-051401, 496-051402, 496-051403, 496-051356, 496-051357, 496-051358, 496-051359, 496-051360, 496-051361, 496-051362, 496-051363, 496-051364, 496-051365, 496-051366, 496-051367, 496-051368, 496-051369, 496-051342, 496-051343, 496-051344, 496-051345, 496-051346, 496-051347, 496-051348, 496-051349, 496-051350, 496-051351, 496-051352, 496-051327, 496-051341, 496-051337, 496-051338, 496-051339, 496-051340, 496-051161, 496-051162, 496-051163, 496-051164, 496-051165, 496-051166, 496-051167, 496-051168, 496-051264, 496-051277, 496-051286, 496-051287, 496-051288, 496-051289, 496-051290, 496-051291, 496-051292, 496-051293, 496-051294, 496-051295, 496-051296, 496-051297, 496-051298, 496-051299, 496-051300, 496-051301, 496-051302, 496-051303, 496-051304, 496-051305, 496-051306, 496-051307, 496-051308, 496-051309, 496-051310, 496-051311, 496-051312, 496-051313, 496-051314, 496-051315, 496-051316, 496-051317, 496-051318, 496-051320, 496-051321, 496-051322, 496-051323, 496-051324, 496-051325, 496-051326, 496-051278, 496-051279, 496-051280, 496-051281, 496-051282, 496-051283, 496-051284, 496-051285, 496-051319, 496-051226, 496-051227, 496-051228, 496-051229, 496-051230, 496-051231, 496-051232, 496-051233, 496-051234, 496-051235, 496-051236, 496-051237, 496-051238, 496-051239, 496-051240, 496-051241, 496-051242, 496-051243, 496-051244, 496-051245, 496-051246, 496-051247, 496-051248, 496-051249, 496-051250, 496-051251, 496-051252, 496-051253, 496-051254, 496-051255, 496-051256, 496-051257, 496-051258, 496-051259, 496-051260, 496-051261, 496-051262, 496-051263, 496-051265, 496-051266, 496-051267, 496-051268, 496-051269, 496-051270, 496-051271, 496-051272, 496-051273, 496-051274, 496-051275, 496-051276, 496-051172, 496-050688, 496-050689, 496-051220, 496-050686, 496-050687, 496-051139, 496-051140, 496-051141, 496-051142, 496-051143, 496-051144, 496-051145, 496-051146, 496-051147, 496-051148, 496-051149, 496-051150, 496-051151, 496-051152, 496-051153, 496-051154, 496-051155, 496-051156, 496-051157, 496-051158, 496-051159, 496-051160, 496-051183, 496-051169, 496-051170, 496-051171, 496-051134, 496-051135, 496-051136, 496-051137, 496-051138, 496-051123, 496-051124, 496-051125, 496-051126, 496-051127, 496-051128, 496-051129, 496-051130, 496-051131, 496-051011, 496-050912, 496-050913, 496-050915, 496-051118, 496-051119, 496-051120, 496-051121, 496-051122, 496-051055, 496-051045, 496-051046, 496-051047, 496-051074, 496-051075, 496-051076, 496-051077, 496-051078, 496-051079, 496-051080, 496-051081, 496-051082, 496-051083, 496-051084, 496-051085, 496-051086, 496-051087, 496-051088, 496-051089, 496-051090, 496-051091, 496-051092, 496-050905, 496-050906, 496-050909, 496-050910, 496-050911, 496-051057, 496-051065, 496-051066, 496-051067, 496-051068, 496-051069, 496-051070, 496-051071, 496-051072, 496-051073, 496-051054, 496-051056, 496-051058, 496-051059, 496-051060, 496-051061, 496-051062, 496-051063, 496-051049, 496-051050, 496-051051, 496-051041, 496-051044, 496-050914, 496-050929, 496-050930, 496-050931, 496-050932, 496-050933, 496-050952, 496-050953, 496-050954, 496-050955, 496-050956, 496-050957, 496-050958, 496-050959, 496-050960, 496-050961, 496-050962, 496-050963, 496-050964, 496-050965, 496-050966, 496-050967, 496-050968, 496-050969, 496-050970, 496-050971, 496-050972, 496-050973, 496-050974, 496-050975, 496-050976, 496-050977, 496-050978, 496-050979, 496-050980, 496-050981, 496-050982, 496-050983, 496-050985, 496-050986, 496-050987, 496-050988, 496-050989, 496-050990, 496-050991, 496-050992, 496-050993, 496-050994, 496-050995, 496-050996, 496-050997, 496-050998, 496-050999, 496-051000, 496-051001, 496-051002, 496-051003, 496-051004, 496-051005, 496-051006, 496-051007, 496-051008, 496-051009, 496-051010, 496-051012, 496-051013, 496-051014, 496-051015, 496-051016, 496-051017, 496-051018, 496-051019, 496-051020, 496-051021, 496-051022, 496-051023, 496-051024, 496-051025, 496-051026, 496-051027, 496-051028, 496-051029, 496-051030, 496-050984, 496-051034, 496-051035, 496-051036, 496-050908, 496-050855, 496-050856, 496-050898, 496-050907, 496-050635, 496-050733, 496-050734, 496-050735, 496-050951, 496-050720, 496-050721, 496-050722, 496-050723, 496-050724, 496-050729, 496-050730, 496-050732, 496-050743, 496-050812, 496-050816, 496-050827, 496-050828, 496-050829, 496-050830, 496-050831, 496-050832, 496-050833, 496-050834, 496-050835, 496-050836, 496-050837, 496-050838, 496-050839, 496-050840, 496-050841, 496-050842, 496-050843, 496-050844, 496-050845, 496-050846, 496-050847, 496-050848, 496-050849, 496-050850, 496-050851, 496-050853, 496-050854, 496-050857, 496-050858, 496-050859, 496-050860, 496-050861, 496-050862, 496-050863, 496-050864, 496-050865, 496-050866, 496-050867, 496-050868, 496-050869, 496-050870, 496-050871, 496-050873, 496-050874, 496-050886, 496-050887, 496-050888, 496-050889, 496-050890, 496-050891, 496-050892, 496-050893, 496-050894, 496-050895, 496-050896, 496-050897, 496-050935, 496-050936, 496-050937, 496-050938, 496-050939, 496-050940, 496-050941, 496-050934, 496-050929, 496-050885, 496-050852, 496-050884, 496-050616, 496-050665, 496-050666, 496-050667, 496-050668, 496-050670, 496-050671, 496-050672, 496-050673, 496-050674, 496-050675, 496-050676, 496-050677, 496-050678, 496-050679, 496-050680, 496-050681, 496-050682, 496-050683, 496-050684, 496-050685, 496-050704, 496-050705, 496-050706, 496-050707, 496-050709, 496-050710, 496-050711, 496-050712, 496-050713, 496-050714, 496-050715, 496-050716, 496-050717, 496-050718, 496-050719, 496-050875, 496-050669, 496-050690, 496-050691, 496-050692, 496-050693, 496-050694, 496-050695, 496-050696, 496-050697, 496-050698, 496-050699, 496-050700, 496-050701, 496-050702, 496-050703, 496-049666, 496-049670, 496-049675, 496-050622, 496-050623, 496-050626, 496-050627, 496-050628, 496-050629, 496-050630, 496-050631, 496-050632, 496-050633, 496-050634, 496-050636, 496-050637, 496-050638, 496-050639, 496-050640, 496-050641, 496-050642, 496-050591, 496-050592, 496-050593, 496-050594, 496-050595, 496-050596, 496-050597, 496-050598, 496-050646, 496-050647, 496-050648, 496-050649, 496-050650, 496-050651, 496-050652, 496-050653, 496-050654, 496-050655, 496-050656, 496-050657, 496-050658, 496-050659, 496-050660, 496-050661, 496-050662, 496-050663, 496-050664, 496-050558, 496-050559, 496-050560, 496-050561, 496-050562, 496-050563, 496-050564, 496-050565, 496-050568, 496-050643, 496-050644, 496-050566, 496-050609, 496-050610, 496-050548, 496-050554, 496-050567, 496-050569, 496-050570, 496-050571, 496-050572, 496-050573, 496-050574, 496-050575, 496-050576, 496-050577, 496-050578, 496-050579, 496-050580, 496-050581, 496-050582, 496-050583, 496-050584, 496-050585, 496-050586, 496-050587, 496-050514, 496-050526, 496-050540, 496-050541, 496-050542, 496-050543, 496-050544, 496-050545, 496-050546, 496-050547, 496-050549, 496-050550, 496-050551, 496-050552, 496-050553, 496-050555, 496-050556, 496-050557, 496-050530, 496-050532, 496-050537, 496-050018, 496-050492, 496-050501, 496-050502, 496-050503, 496-050504, 496-050505, 496-050506, 496-050507, 496-050508, 496-050509, 496-050510, 496-050511, 496-050512, 496-050513, 496-050515, 496-050516, 496-050517, 496-050518, 496-050519, 496-050520, 496-050521, 496-050522, 496-050523, 496-050524, 496-050525, 496-050527, 496-050528, 496-050529, 496-050531, 496-050533, 496-050539, 496-050241, 496-050242, 496-050243, 496-050244, 496-050245, 496-050247, 496-050248, 496-050249, 496-050250, 496-050251, 496-050252, 496-050253, 496-050254, 496-050255, 496-050256, 496-050257, 496-050258, 496-050259, 496-050260, 496-050261, 496-050287, 496-050288, 496-050289, 496-050290, 496-050291, 496-050292, 496-050293, 496-050294, 496-050295, 496-050296, 496-050297, 496-050298, 496-050299, 496-050301, 496-050304, 496-050305, 496-050306, 496-050307, 496-050308, 496-050309, 496-050310, 496-050311, 496-050312, 496-050313, 496-050314, 496-050316, 496-050318, 496-050320, 496-050323, 496-050327, 496-050328, 496-050329, 496-050330, 496-050331, 496-050333, 496-050336, 496-050338, 496-050339, 496-050340, 496-050341, 496-050343, 496-050344, 496-050345, 496-050347, 496-050348, 496-050353, 496-050354, 496-050355, 496-050356, 496-050357, 496-050358, 496-050359, 496-050361, 496-050362, 496-050363, 496-050364, 496-050365, 496-050366, 496-050367, 496-050368, 496-050369, 496-050370, 496-050371, 496-050373, 496-050376, 496-050377, 496-050378, 496-050380, 496-050381, 496-050382, 496-050383, 496-050384, 496-050385, 496-050387, 496-050388, 496-050392, 496-050393, 496-050394, 496-050395, 496-050396, 496-050397, 496-050398, 496-050399, 496-050400, 496-050401, 496-050403, 496-049309, 496-049310, 496-049311, 496-049312, 496-049313, 496-050421, 496-050461, 496-050500, 496-050335, 496-050337, 496-050342, 496-050372, 496-050375, 496-050379, 496-050386, 496-050390, 496-050391, 496-050402, 496-050407, 496-050408, 496-050409, 496-050410, 496-050411, 496-050412, 496-050413, 496-050414, 496-050415, 496-050416, 496-050417, 496-050418, 496-050419, 496-050420, 496-050422, 496-050423, 496-050424, 496-050425, 496-050426, 496-050429, 496-050430, 496-050431, 496-050442, 496-050443, 496-050444, 496-050445, 496-050446, 496-050449, 496-050450, 496-050451, 496-050452, 496-050453, 496-050454, 496-050455, 496-050456, 496-050457, 496-050458, 496-050459, 496-050460, 496-050462, 496-050463, 496-050464, 496-050466, 496-050468, 496-050469, 496-050470, 496-050471, 496-050472, 496-050473, 496-050474, 496-050475, 496-050499, 496-050482, 496-050485, 496-050486, 496-050487, 496-050488, 496-050489, 496-050490, 496-050491, 496-050493, 496-050494, 496-050495, 496-050496, 496-050497, 496-050498, 496-049308, 496-050483, 496-050279, 496-050280, 496-050281, 496-050282, 496-050283, 496-050284, 496-050285, 496-050440, 496-050476, 496-050477, 496-050478, 496-050479, 496-050480, 496-050481, 496-050427, 496-050467, 496-050428, 496-050315, 496-050317, 496-050319, 496-050332, 496-050406, 496-050360, 496-050404, 496-050275, 496-050276, 496-050277, 496-050278, 496-050300, 496-050302, 496-050303, 496-050349, 496-050350, 496-050351, 496-050334, 496-050138, 496-050139, 496-050140, 496-050141, 496-050143, 496-050195, 496-050205, 496-050208, 496-050209, 496-050211, 496-050212, 496-050213, 496-050214, 496-050215, 496-050216, 496-050218, 496-050220, 496-050221, 496-050222, 496-050223, 496-050224, 496-050225, 496-050226, 496-050185, 496-050186, 496-050187, 496-050188, 496-050189, 496-050190, 496-050191, 496-050192, 496-050193, 496-050194, 496-050196, 496-050197, 496-050198, 496-050199, 496-050200, 496-050202, 496-050203, 496-050204, 496-050206, 496-050147, 496-050150, 496-050151, 496-050152, 496-050153, 496-050154, 496-050155, 496-050156, 496-050158, 496-050246, 496-050174, 496-050175, 496-050148, 496-050149, 496-046494, 496-047200, 496-050126, 496-050127, 496-050128, 496-050129, 496-050130, 496-050131, 496-050132, 496-050133, 496-050144, 496-050145, 496-050115, 496-050116, 496-050117, 496-050118, 496-050119, 496-050120, 496-050122, 496-050123, 496-050124, 496-050125, 496-050136, 496-050137, 496-050134, 496-050135, 496-050113, 496-050114, 496-050014, 496-050027, 496-050028, 496-050029, 496-050030, 496-050031, 496-050032, 496-050033, 496-050034, 496-050035, 496-050036, 496-050037, 496-050038, 496-050039, 496-050040, 496-050041, 496-050042, 496-050043, 496-050044, 496-050045, 496-050046, 496-050047, 496-050048, 496-050049, 496-050050, 496-050051, 496-050052, 496-050053, 496-050054, 496-050055, 496-050056, 496-050057, 496-050058, 496-050059, 496-050060, 496-050061, 496-050062, 496-050063, 496-050064, 496-050065, 496-050066, 496-050067, 496-050068, 496-050069, 496-050070, 496-050071, 496-050072, 496-050073, 496-050074, 496-050075, 496-050076, 496-050077, 496-050078, 496-050079, 496-050080, 496-050081, 496-050082, 496-050083, 496-050084, 496-050085, 496-050086, 496-050087, 496-050088, 496-050089, 496-050090, 496-050091, 496-050092, 496-050093, 496-050094, 496-050095, 496-050096, 496-050097, 496-050098, 496-050099, 496-050100, 496-050101, 496-050102, 496-050103, 496-050104, 496-050105, 496-050106, 496-050107, 496-050108, 496-050109, 496-050110, 496-050022, 496-050023, 496-049909, 496-049910, 496-049911, 496-049912, 496-049913, 496-049914, 496-049915, 496-050024, 496-050025, 496-050026, 496-049933, 496-049939, 496-049983, 496-049984, 496-049985, 496-049986, 496-049987, 496-049988, 496-049989, 496-049990, 496-049991, 496-049996, 496-049997, 496-049998, 496-049999, 496-050000, 496-050001, 496-050002, 496-050003, 496-050004, 496-049927, 496-049928, 496-049929, 496-049930, 496-049821, 496-049839, 496-049868, 496-049869, 496-049554, 496-049626, 496-049631, 496-049632, 496-049633, 496-049634, 496-049635, 496-049636, 496-049637, 496-049638, 496-049639, 496-049641, 496-049642, 496-049643, 496-049644, 496-049645, 496-049646, 496-049647, 496-049648, 496-049649, 496-049650, 496-049651, 496-049652, 496-049653, 496-049654, 496-049655, 496-049656, 496-049657, 496-049658, 496-049659, 496-049916, 496-049754, 496-049777, 496-049778, 496-049779, 496-049780, 496-049781, 496-049782, 496-049783, 496-049784, 496-049785, 496-049786, 496-049787, 496-049788, 496-049789, 496-049790, 496-049791, 496-049792, 496-049793, 496-049794, 496-049795, 496-049796, 496-049797, 496-049798, 496-049799, 496-049936, 496-049905, 496-049906, 496-049907, 496-049908, 496-049968, 496-049969, 496-049970, 496-049917, 496-049951, 496-049945, 496-049946, 496-049948, 496-049949, 496-049950, 496-049952, 496-049811, 496-049827, 496-049828, 496-049829, 496-049870, 496-049871, 496-049872, 496-049874, 496-049875, 496-049876, 496-049877, 496-049878, 496-049879, 496-049880, 496-049881, 496-049882, 496-049885, 496-049953, 496-049820, 496-049841, 496-049852, 496-049853, 496-049854, 496-049855, 496-049856, 496-049857, 496-049858, 496-049859, 496-049886, 496-049887, 496-049888, 496-049889, 496-049890, 496-049891, 496-049892, 496-049893, 496-049894, 496-049895, 496-049896, 496-049898, 496-049899, 496-049860, 496-049861, 496-049862, 496-049863, 496-049864, 496-049865, 496-049866, 496-049867, 496-049808, 496-049824, 496-049825, 496-049826, 496-049761, 496-049800, 496-049801, 496-049802, 496-049803, 496-049804, 496-049805, 496-049806, 496-049807, 496-049809, 496-049810, 496-049812, 496-049813, 496-049814, 496-049815, 496-049816, 496-049817, 496-049818, 496-049819, 496-049822, 496-049823, 496-049830, 496-049831, 496-049832, 496-049833, 496-049834, 496-049835, 496-049836, 496-049837, 496-049838, 496-049842, 496-049570, 496-049579, 496-049580, 496-049581, 496-049582, 496-049583, 496-049584, 496-049585, 496-049586, 496-049587, 496-049764, 496-049765, 496-049766, 496-049767, 496-049768, 496-049769, 496-049770, 496-049771, 496-049772, 496-049720, 496-049773, 496-049774, 496-049775, 496-049776, 496-049752, 496-049753, 496-049084, 496-049690, 496-049691, 496-049693, 496-049694, 496-049695, 496-049696, 496-049697, 496-049698, 496-049699, 496-049700, 496-049701, 496-049702, 496-049703, 496-049704, 496-049705, 496-049706, 496-049707, 496-049708, 496-049709, 496-049710, 496-049711, 496-049712, 496-049713, 496-049714, 496-049715, 496-049716, 496-049717, 496-049718, 496-049719, 496-049721, 496-049722, 496-049723, 496-049724, 496-049726, 496-049727, 496-049759, 496-049760, 496-049762, 496-049763, 496-049725, 496-049605, 496-049622, 496-049623, 496-049624, 496-049625, 496-049627, 496-049628, 496-049629, 496-049630, 496-049661, 496-049729, 496-049728, 496-049679, 496-049680, 496-049681, 496-049678, 496-049682, 496-049683, 496-049684, 496-049685, 496-049686, 496-049687, 496-049688, 496-049618, 496-049619, 496-049620, 496-049621, 496-049610, 496-049611, 496-049612, 496-049617, 496-049430, 496-049458, 496-049459, 496-049460, 496-049461, 496-049591, 496-049600, 496-049602, 496-049603, 496-049488, 496-049527, 496-049551, 496-049552, 496-049553, 496-049555, 496-049556, 496-049557, 496-049558, 496-049559, 496-049560, 496-049561, 496-049562, 496-049563, 496-049564, 496-049565, 496-049566, 496-049567, 496-049568, 496-049569, 496-049428, 496-049571, 496-049572, 496-049573, 496-049473, 496-049474, 496-049475, 496-049476, 496-049477, 496-049478, 496-049479, 496-049480, 496-049481, 496-049482, 496-049483, 496-049484, 496-049485, 496-049486, 496-049487, 496-049489, 496-049490, 496-049491, 496-049492, 496-049494, 496-049495, 496-049526, 496-049528, 496-049529, 496-049530, 496-049531, 496-049532, 496-049533, 496-049534, 496-049535, 496-049536, 496-049537, 496-049538, 496-049539, 496-049540, 496-049541, 496-049542, 496-049543, 496-049544, 496-049545, 496-049546, 496-049547, 496-049548, 496-049549, 496-049550, 496-049260, 496-049261, 496-049262, 496-049263, 496-049264, 496-049265, 496-049266, 496-049267, 496-049268, 496-049468, 496-049469, 496-049470, 496-049471, 496-049189, 496-049190, 496-049191, 496-049359, 496-049360, 496-049361, 496-049362, 496-049363, 496-049364, 496-049365, 496-049366, 496-049367, 496-049368, 496-049369, 496-049370, 496-049371, 496-049373, 496-049374, 496-049448, 496-049431, 496-049432, 496-049401, 496-049402, 496-049403, 496-049404, 496-049405, 496-049406, 496-049407, 496-049408, 496-049409, 496-049410, 496-049411, 496-049412, 496-049413, 496-049414, 496-049415, 496-049416, 496-049417, 496-049418, 496-049419, 496-049420, 496-049399, 496-049400, 496-049388, 496-049393, 496-049394, 496-049395, 496-049396, 496-049397, 496-049398, 496-049390, 496-049391, 496-049375, 496-049376, 496-049377, 496-049378, 496-049379, 496-049380, 496-049381, 496-049382, 496-049383, 496-049384, 496-049372, 496-049357, 496-049358, 496-049353, 496-049354, 496-049355, 496-049356, 496-049082, 496-049296, 496-049298, 496-049300, 496-049301, 496-049302, 496-049303, 496-049308, 496-049309, 496-049310, 496-049311, 496-049312, 496-049313, 496-049259, 496-049246, 496-049258, 496-049235, 496-049242, 496-049249, 496-049253, 496-049255, 496-049256, 496-049257, 496-049184, 496-049192, 496-049051, 496-049185, 496-049186, 496-049187, 496-049188, 496-049182, 496-049183, 496-049171, 496-048917, 496-049159, 496-049116, 496-049117, 496-049118, 496-049119, 496-049120, 496-049122, 496-049052, 496-049053, 496-049054, 496-049055, 496-049086, 496-049087, 496-048055, 496-048926, 496-048945, 496-048971, 496-048986, 496-048987, 496-048988, 496-048989, 496-048991, 496-048992, 496-048993, 496-048994, 496-048995, 496-048996, 496-048997, 496-048998, 496-048999, 496-049000, 496-049001, 496-049003, 496-049048, 496-048878, 496-048948, 496-048949, 496-048950, 496-048951, 496-048952, 496-048953, 496-048954, 496-048955, 496-048956, 496-048957, 496-048958, 496-048959, 496-048960, 496-048961, 496-048962, 496-048963, 496-048964, 496-048965, 496-048966, 496-048967, 496-048968, 496-048990, 496-049002, 496-049015, 496-049016, 496-049017, 496-049018, 496-049019, 496-049020, 496-049021, 496-049022, 496-049023, 496-049024, 496-049025, 496-049026, 496-049050, 496-048937, 496-048938, 496-048939, 496-048940, 496-048941, 496-048942, 496-048943, 496-048944, 496-048946, 496-048947, 496-048980, 496-048981, 496-049027, 496-049028, 496-048929, 496-048930, 496-049042, 496-049044, 496-049045, 496-049029, 496-049030, 496-049031, 496-049032, 496-049033, 496-049034, 496-049035, 496-049036, 496-049037, 496-049038, 496-049039, 496-049040, 496-049041, 496-049043, 496-049009, 496-049010, 496-048846, 496-048847, 496-048848, 496-048851, 496-048852, 496-048853, 496-048822, 496-048823, 496-048832, 496-048833, 496-048834, 496-048835, 496-048836, 496-048837, 496-048838, 496-048839, 496-048840, 496-048841, 496-048842, 496-048843, 496-048844, 496-048845, 496-048804, 496-048805, 496-048806, 496-048709, 496-048969, 496-048970, 496-048972, 496-048863, 496-048894, 496-048911, 496-048912, 496-048913, 496-048914, 496-048915, 496-048916, 496-048918, 496-048919, 496-048825, 496-048826, 496-048827, 496-048855, 496-048860, 496-048861, 496-048862, 496-048830, 496-048831, 496-048849, 496-048850, 496-048854, 496-048856, 496-048857, 496-048858, 496-048859, 496-048723, 496-048756, 496-048757, 496-048758, 496-048759, 496-048798, 496-048799, 496-048800, 496-048801, 496-048802, 496-048803, 496-048828, 496-048829, 496-048807, 496-048808, 496-048809, 496-048755, 496-048760, 496-048761, 496-048762, 496-048763, 496-048764, 496-048765, 496-048766, 496-048767, 496-048768, 496-048769, 496-048770, 496-048771, 496-048772, 496-048773, 496-048774, 496-048775, 496-048776, 496-048777, 496-048778, 496-048779, 496-048780, 496-048781, 496-048782, 496-048784, 496-048785, 496-048786, 496-048787, 496-048788, 496-048789, 496-048790, 496-048791, 496-048792, 496-048793, 496-048710, 496-048696, 496-048688, 496-048664, 496-048665, 496-048666, 496-048686, 496-048687, 496-047607, 496-048411, 496-048419, 496-048420, 496-048421, 496-048423, 496-048428, 496-048429, 496-048460, 496-048468, 496-048482, 496-048483, 496-048484, 496-048485, 496-048486, 496-048619, 496-048620, 496-048621, 496-048622, 496-048623, 496-048624, 496-048625, 496-048626, 496-048627, 496-048628, 496-048629, 496-048630, 496-048631, 496-048632, 496-048633, 496-048634, 496-048635, 496-048636, 496-048637, 496-048638, 496-048639, 496-048640, 496-048641, 496-048642, 496-048660, 496-048661, 496-048662, 496-048663, 496-048667, 496-048668, 496-048669, 496-048670, 496-048671, 496-048672, 496-048673, 496-048674, 496-048675, 496-048676, 496-048677, 496-048678, 496-048679, 496-048680, 496-048681, 496-048682, 496-048683, 496-048567, 496-047606, 496-048582, 496-048583, 496-048584, 496-048585, 496-048586, 496-048587, 496-048566, 496-048588, 496-048382, 496-048503, 496-048544, 496-048545, 496-048539, 496-048540, 496-048436, 496-048437, 496-048455, 496-048456, 496-048425, 496-048471, 496-048472, 496-048473, 496-048474, 496-048431, 496-048415, 496-048416, 496-048417, 496-048403, 496-048404, 496-048337, 496-048338, 496-048339, 496-048340, 496-048341, 496-048342, 496-048343, 496-048344, 496-048347, 496-048348, 496-048349, 496-048350, 496-048351, 496-048352, 496-048353, 496-048354, 496-048384, 496-048385, 496-048386, 496-048345, 496-048346, 496-048355, 496-048356, 496-048357, 496-048360, 496-048361, 496-048362, 496-048363, 496-048364, 496-048365, 496-048366, 496-048367, 496-048368, 496-048369, 496-048370, 496-048305, 496-048257, 496-048178, 496-048253, 496-048264, 496-048265, 496-048266, 496-048267, 496-048268, 496-048269, 496-048270, 496-048271, 496-048272, 496-048273, 496-048274, 496-048275, 496-048276, 496-048277, 496-048278, 496-048279, 496-048280, 496-048281, 496-048282, 496-048283, 496-048284, 496-048285, 496-048286, 496-048287, 496-048288, 496-048289, 496-048290, 496-048291, 496-048292, 496-048293, 496-048294, 496-048295, 496-048296, 496-048297, 496-048298, 496-048299, 496-048300, 496-048301, 496-048302, 496-048303, 496-048304, 496-048306, 496-048307, 496-048308, 496-048309, 496-048310, 496-048312, 496-048313, 496-048258, 496-048259, 496-048185, 496-048148, 496-048186, 496-048187, 496-048150, 496-048151, 496-048153, 496-048154, 496-048179, 496-048180, 496-048181, 496-048182, 496-048183, 496-048184, 496-048137, 496-048138, 496-048139, 496-048140, 496-048141, 496-048142, 496-048143, 496-048144, 496-048145, 496-048146, 496-048147, 496-048155, 496-048156, 496-048157, 496-048158, 496-048159, 496-048160, 496-048161, 496-048162, 496-048163, 496-048164, 496-047226, 496-047227, 496-047228, 496-047714, 496-047927, 496-047928, 496-047929, 496-047980, 496-047982, 496-048023, 496-048024, 496-048026, 496-048027, 496-048028, 496-048029, 496-048030, 496-048031, 496-048032, 496-047874, 496-047876, 496-047862, 496-047983, 496-047984, 496-047972, 496-047973, 496-047872, 496-047873, 496-047858, 496-047859, 496-047860, 496-047800, 496-047801, 496-045547, 496-047682, 496-047684, 496-047685, 496-047686, 496-047687, 496-047688, 496-047689, 496-047690, 496-047691, 496-047692, 496-047693, 496-047694, 496-047695, 496-047696, 496-047697, 496-047698, 496-047802, 496-047803, 496-047796, 496-047797, 496-047798, 496-047799, 496-047772, 496-047708, 496-047756, 496-047770, 496-047771, 496-047773, 496-047774, 496-047658, 496-047659, 496-047655, 496-047656, 496-047657, 496-047660, 496-047661, 496-047662, 496-047663, 496-047726, 496-047727, 496-047728, 496-047729, 496-047730, 496-047731, 496-047732, 496-047733, 496-047734, 496-047735, 496-047736, 496-047737, 496-047738, 496-047739, 496-047740, 496-047741, 496-047742, 496-047743, 496-047744, 496-047745, 496-047746, 496-047747, 496-047748, 496-047749, 496-047750, 496-047751, 496-047752, 496-047753, 496-047754, 496-047755, 496-047757, 496-047758, 496-047575, 496-047576, 496-047664, 496-047665, 496-047666, 496-047668, 496-047669, 496-047670, 496-047671, 496-047672, 496-047673, 496-047674, 496-047675, 496-047676, 496-047677, 496-047678, 496-047679, 496-047609, 496-047628, 496-047639, 496-047640, 496-047641, 496-047642, 496-047643, 496-047644, 496-047645, 496-047646, 496-047647, 496-047648, 496-047667, 496-047680, 496-047681, 496-047683, 496-047573, 496-047574, 496-047699, 496-047700, 496-047701, 496-047702, 496-047703, 496-047704, 496-047705, 496-047706, 496-047707, 496-047549, 496-047516, 496-047529, 496-047589, 496-047590, 496-047591, 496-047592, 496-047593, 496-047594, 496-047595, 496-047596, 496-047544, 496-047483, 496-047528, 496-047530, 496-047531, 496-047532, 496-047533, 496-047534, 496-047535, 496-047536, 496-047537, 496-047538, 496-047539, 496-047540, 496-047541, 496-047542, 496-047543, 496-046791, 496-047517, 496-047518, 496-047520, 496-047521, 496-047522, 496-047523, 496-047524, 496-047525, 496-047526, 496-047527, 496-047545, 496-047546, 496-047547, 496-047548, 496-047515, 496-047519, 496-047509, 496-047510, 496-047511, 496-047512, 496-047513, 496-047514, 496-047411, 496-047426, 496-047448, 496-047449, 496-047450, 496-047451, 496-047452, 496-047453, 496-047454, 496-047455, 496-047456, 496-047457, 496-047458, 496-047459, 496-047460, 496-047461, 496-047462, 496-047463, 496-047464, 496-047465, 496-047466, 496-047467, 496-047468, 496-047470, 496-047471, 496-047472, 496-047473, 496-047474, 496-047475, 496-047476, 496-047477, 496-047478, 496-047479, 496-047480, 496-047481, 496-047482, 496-047484, 496-047485, 496-047486, 496-047487, 496-047488, 496-047489, 496-047490, 496-047491, 496-047492, 496-047493, 496-047494, 496-047495, 496-047496, 496-047497, 496-047498, 496-046450, 496-046451, 496-046476, 496-046477, 496-047407, 496-047305, 496-047331, 496-047332, 496-047333, 496-047334, 496-047335, 496-047336, 496-047337, 496-047338, 496-047339, 496-047340, 496-047341, 496-047342, 496-047343, 496-047344, 496-047345, 496-047346, 496-047348, 496-047349, 496-047350, 496-047351, 496-047352, 496-047353, 496-047354, 496-047355, 496-047356, 496-047357, 496-047358, 496-047359, 496-047360, 496-047361, 496-047362, 496-047363, 496-047364, 496-047365, 496-047366, 496-047367, 496-047368, 496-047369, 496-047370, 496-047371, 496-047372, 496-047373, 496-047374, 496-047375, 496-047376, 496-047377, 496-047378, 496-047379, 496-047380, 496-047381, 496-047382, 496-047383, 496-047384, 496-047385, 496-047386, 496-047387, 496-047388, 496-047389, 496-047390, 496-047392, 496-047393, 496-047394, 496-047182, 496-047302, 496-047303, 496-047304, 496-047307, 496-047256, 496-047257, 496-047258, 496-047259, 496-047260, 496-047261, 496-047262, 496-047264, 496-047272, 496-047273, 496-047274, 496-047275, 496-047276, 496-047277, 496-047278, 496-047279, 496-047280, 496-047281, 496-047282, 496-047283, 496-047284, 496-047285, 496-047286, 496-047287, 496-047288, 496-047289, 496-047290, 496-047291, 496-047292, 496-047293, 496-047294, 496-047295, 496-047296, 496-047297, 496-047298, 496-047299, 496-047263, 496-047265, 496-047266, 496-047267, 496-047268, 496-047269, 496-047270, 496-047271, 496-047232, 496-047233, 496-047234, 496-047235, 496-047236, 496-047237, 496-047238, 496-047239, 496-047240, 496-047241, 496-047242, 496-047243, 496-047244, 496-047245, 496-047246, 496-047247, 496-047248, 496-047249, 496-047250, 496-047251, 496-047252, 496-047253, 496-047254, 496-047255, 496-047166, 496-047183, 496-047217, 496-047218, 496-047219, 496-047231, 496-047185, 496-047186, 496-047187, 496-047188, 496-047189, 496-047190, 496-047184, 496-047191, 496-047192, 496-047193, 496-047194, 496-047195, 496-047196, 496-047197, 496-047159, 496-046809, 496-046810, 496-047085, 496-047086, 496-047087, 496-047088, 496-047089, 496-047090, 496-047091, 496-047092, 496-047093, 496-047094, 496-047095, 496-047096, 496-047097, 496-047098, 496-047099, 496-047100, 496-047101, 496-047102, 496-047103, 496-047104, 496-047105, 496-047106, 496-047107, 496-047108, 496-047109, 496-047110, 496-047111, 496-047112, 496-047113, 496-047114, 496-045979, 496-047057, 496-047058, 496-047059, 496-047060, 496-047061, 496-047062, 496-047063, 496-046927, 496-046933, 496-047034, 496-047035, 496-047036, 496-047037, 496-047038, 496-047039, 496-047040, 496-047041, 496-047042, 496-047043, 496-047044, 496-047045, 496-047046, 496-047047, 496-047048, 496-047049, 496-047050, 496-047051, 496-047052, 496-047053, 496-047054, 496-046175, 496-046322, 496-046388, 496-046393, 496-046948, 496-046949, 496-046950, 496-046951, 496-046952, 496-046953, 496-046954, 496-046955, 496-046956, 496-046957, 496-046910, 496-046682, 496-046683, 496-046684, 496-046685, 496-046686, 496-046687, 496-046688, 496-046689, 496-046690, 496-046691, 496-046692, 496-046770, 496-046771, 496-046772, 496-046773, 496-046774, 496-046775, 496-046776, 496-046777, 496-046778, 496-046779, 496-046780, 496-046781, 496-046782, 496-046783, 496-046784, 496-046785, 496-046786, 496-046787, 496-046788, 496-046842, 496-046843, 496-046844, 496-046845, 496-046846, 496-046847, 496-046848, 496-046849, 496-046850, 496-046851, 496-046852, 496-046853, 496-046854, 496-046855, 496-046856, 496-046857, 496-046858, 496-046859, 496-046860, 496-046861, 496-046862, 496-046863, 496-046864, 496-046865, 496-046866, 496-046867, 496-046868, 496-046869, 496-046871, 496-046872, 496-046873, 496-046874, 496-046875, 496-046876, 496-046877, 496-046878, 496-046879, 496-046880, 496-046881, 496-046882, 496-046884, 496-046885, 496-046886, 496-046887, 496-046888, 496-046889, 496-046890, 496-046891, 496-046892, 496-046893, 496-046894, 496-046895, 496-046896, 496-046897, 496-046898, 496-046899, 496-046900, 496-046901, 496-046903, 496-046904, 496-046905, 496-046906, 496-046907, 496-046908, 496-046909, 496-046913, 496-046914, 496-046915, 496-046916, 496-046917, 496-046918, 496-046919, 496-046921, 496-046922, 496-046923, 496-046924, 496-046733, 496-046617, 496-046693, 496-046694, 496-046695, 496-046696, 496-046697, 496-046698, 496-046699, 496-046700, 496-046704, 496-046705, 496-046706, 496-046707, 496-046708, 496-046709, 496-046710, 496-046711, 496-046712, 496-046713, 496-046714, 496-046715, 496-046716, 496-046717, 496-046718, 496-046719, 496-046720, 496-046721, 496-046722, 496-046723, 496-046724, 496-046725, 496-046726, 496-046727, 496-046728, 496-046730, 496-046746, 496-046747, 496-046748, 496-046749, 496-046750, 496-046751, 496-046752, 496-046753, 496-046754, 496-046755, 496-046756, 496-046757, 496-046758, 496-046759, 496-046760, 496-046761, 496-046762, 496-046763, 496-046764, 496-046765, 496-046766, 496-046767, 496-046768, 496-046769, 496-046792, 496-046793, 496-046794, 496-046795, 496-046796, 496-046744, 496-046745, 496-046302, 496-046677, 496-046701, 496-046702, 496-046703, 496-046680, 496-046640, 496-046620, 496-046621, 496-046623, 496-046614, 496-046615, 496-046616, 496-046618, 496-046619, 496-046384, 496-046385, 496-046386, 496-046387, 496-046392, 496-046397, 496-046399, 496-046595, 496-046596, 496-046597, 496-046601, 496-046603, 496-046604, 496-046606, 496-046389, 496-046391, 496-046472, 496-046532, 496-046539, 496-046561, 496-046562, 496-046563, 496-046564, 496-046565, 496-046566, 496-046567, 496-046568, 496-046569, 496-046570, 496-046571, 496-046572, 496-046573, 496-046575, 496-046576, 496-046577, 496-046578, 496-046579, 496-046580, 496-046581, 496-046395, 496-046512, 496-046511, 496-046473, 496-046480, 496-046487, 496-046489, 496-046491, 496-046492, 496-046505, 496-046506, 496-046513, 496-046517, 496-046518, 496-046519, 496-046520, 496-046521, 496-046523, 496-046525, 496-046526, 496-046528, 496-046529, 496-046530, 496-046531, 496-046536, 496-046515, 496-046516, 496-046464, 496-046449, 496-046450, 496-046451, 496-046476, 496-046477, 496-046478, 496-046423, 496-045981, 496-046419, 496-046309, 496-046320, 496-046415, 496-046416, 496-046417, 496-046418, 496-046346, 496-046413, 496-046414, 496-046420, 496-046296, 496-046339, 496-046340, 496-046342, 496-046343, 496-046316, 496-046319, 496-037973, 496-038273, 496-039031, 496-039098, 496-039433, 496-043640, 496-044507, 496-044605, 496-044610, 496-044611, 496-044652, 496-044661, 496-044814, 496-045277, 496-045339, 496-045474, 496-045489, 496-045495, 496-045515, 496-034359, 496-037903, 496-037935, 496-038983, 496-040985, 496-040986, 496-045290, 496-045303, 496-046291, 496-046303, 496-046324, 496-046325, 496-046327, 496-039701, 496-039874, 496-039923, 496-045164, 496-045218, 496-045951, 496-045963, 496-046018, 496-046186, 496-043779, 496-044802, 496-045477, 496-046232, 496-040273, 496-040814, 496-043994, 496-044019, 496-044020, 496-044025, 496-044037, 496-044039, 496-044040, 496-044042, 496-044043, 496-044582, 496-044607, 496-044748, 496-044764, 496-044800, 496-045367, 496-043701, 496-044211, 496-044496, 496-044606, 496-044809, 496-045456, 496-045965, 496-046275, 496-046276, 496-046278, 496-046281, 496-046174, 496-046220, 496-046223, 496-046226, 496-046227, 496-046233, 496-046234, 496-046235, 496-046236, 496-046245, 496-046253, 496-046257, 496-046258, 496-046259, 496-046260, 496-046262, 496-046263, 496-046267, 496-046268, 496-046271, 496-046274, 496-046282, 496-046283, 496-044667, 496-044737, 496-044750, 496-044759, 496-044763, 496-044772, 496-044789, 496-044237, 496-046222, 496-046224, 496-046228, 496-046231, 496-046264, 496-046242, 496-046250, 496-046114, 496-046201, 496-046230, 496-046169, 496-046277, 496-046279, 496-046280, 496-046121, 496-046173, 496-046205, 496-046206, 496-046244, 496-046254, 496-046265, 496-045940, 496-045955, 496-045958, 496-045961, 496-045969, 496-045971, 496-046029, 496-045946, 496-045997, 496-045998, 496-046002, 496-046003, 496-046019, 496-046020, 496-046035, 496-046037, 496-046051, 496-045932, 496-045935, 496-045936, 496-045937, 496-045964, 496-046204, 496-046214, 496-046221, 496-046045, 496-046060, 496-046082, 496-046096, 496-046097, 496-046161, 496-046163, 496-046164, 496-046170, 496-046179, 496-046209, 496-046072, 496-046074, 496-046075, 496-046100, 496-046105, 496-046112, 496-046115, 496-046120, 496-046122, 496-046138, 496-046143, 496-045939, 496-045967, 496-046059, 496-046098, 496-046101, 496-046102, 496-046118, 496-046126, 496-046144, 496-046145, 496-046149, 496-046151, 496-046152, 496-046153, 496-046154, 496-045947, 496-045999, 496-046016, 496-046022, 496-046023, 496-045944, 496-046024, 496-046026, 496-045877, 496-045921, 496-045930, 496-045948, 496-045864, 496-045758, 496-045789, 496-045790, 496-045798, 496-045818, 496-045831, 496-045838, 496-045847, 496-045849, 496-045871, 496-045900, 496-045915, 496-045918, 496-045717, 496-045756, 496-045764, 496-045769, 496-045774, 496-045775, 496-045777, 496-045778, 496-045704, 496-045763, 496-045765, 496-045767, 496-045749, 496-045755, 496-045757, 496-045770, 496-045771, 496-045773, 496-045652, 496-045666, 496-045682, 496-045683, 496-045685, 496-045760, 496-045768, 496-045690, 496-045730, 496-045640, 496-045599, 496-045627, 496-045631, 496-045633, 496-045554, 496-045564, 496-045565, 496-045566, 496-045569, 496-045588, 496-045613, 496-045617, 496-045629, and 496-045632.
FEI Number 3010157426
Recalling Firm/
Manufacturer
Spacelabs Healthcare, Incorporated
5150 220th Ave Se
Issaquah WA 98029-6834
For Additional Information Contact
425-657-7200 Ext. 5970
Manufacturer Reason
for Recall
Potential for module to fail to obtain a NIBP reading on hypertension patients.
FDA Determined
Cause 2
Process control
ActionOn July 22, 2008 an "Urgent MEDICAL DEVICE CORRECTION" letter was mailed to all the firm's customers. The letter states the problem at hand, when the problem would likely to occur and instructions of a way to mitigate the risk, until the modules can be recalibrated . The firm's Customer Connectivity Center will confirm all affected customers have been notified. Spacelabs Technical Support will be recalibrating all modules personally and ensuring all corrections have been applied to all affected customers of record. Contact Spacelabs Healthcare at 1-800-522-7025 for assistance.
Quantity in Commerce5,140
DistributionThe product is distributed nationwide and internationally to medical facilities.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DSI
-
-