Class 2 Device Recall King Systems Universal Flex2 Breathing Circuit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall King Systems Universal Flex2 Breathing Circuitsee related information
Date Initiated by FirmAugust 01, 2008
Date PostedFebruary 03, 2009
Recall Status1 Terminated 3 on November 12, 2009
Recall NumberZ-0764-2009
Recall Event ID 48991
Product Classification Breathing Circuit - Product Code CAI
ProductKing Systems Universal Flex2 Breathing Circuit, Latex Free, 40 per case, King Systems Corporation, Noblesville, IN; REF # DF4115-6121Z. This breathing circuit is a device intended to administer medical gases and/or anesthetic gases to a patient during anesthesia for inhalation, or medical gases to a patient during respiratory care inhalation.
Code Information Lots IKCV9, IKH33, IKIZ5, IKVQ4 and IL2T4.
FEI Number 1824226
Recalling Firm/
Manufacturer
King Systems Corp.
15011 Herriman Blvd
Noblesville IN 46060
For Additional Information Contact
317-776-6823
Manufacturer Reason
for Recall
The inner line may be severed or broken into two pieces, thus hindering the delivery of oxygen and anesthetics to the patient and possibly elevated patient carbon dioxide levels.
FDA Determined
Cause 2
Pending
ActionConsignees were notified by an Urgent Product Recall letter on or about 7/29/2008 to cease sales, notify their customers to the hospital level, and to return product on hand.
Quantity in CommerceApprox. 500,000 units of all products.
DistributionWorldwide Distribution - including USA and countries of Australia, Colombia, Germany, Netherlands, and Peru.