 Class 2 Device Recall King Systems 72" Universal Flex2 Extension Kitsee related information
Date Initiated by FirmAugust 01, 2008
Date PostedFebruary 03, 2009
Recall Status1 Terminated 3 on November 12, 2009
Recall NumberZ-0816-2009
Recall Event ID 48991
Product Classification Breathing Circuit - Product Code CAI
ProductKing Systems 72" Universal Flex2 Extension Kit, 50 per case, King Systems Corporation, Noblesville, IN; REF # XF75.
Code Information Lots IKAE1, IKCJ4, IKER6, IKGC0, IKIJ7, IKJK6, IKMI3, IKQ81, IKUX4, IKVS2, IKY06, IL161, IL1Y6, IL383, IL7N9 and IL8Y9.
FEI Number 1824226
Recalling Firm/
Manufacturer
King Systems Corp.
15011 Herriman Blvd
Noblesville IN 46060
For Additional Information Contact
317-776-6823
Manufacturer Reason
for Recall
The inner line may be severed or broken into two pieces, thus hindering the delivery of oxygen and anesthetics to the patient and possibly elevated patient carbon dioxide levels.
FDA Determined
Cause 2
Pending
ActionConsignees were notified by an Urgent Product Recall letter on or about 7/29/2008 to cease sales, notify their customers to the hospital level, and to return product on hand.
Quantity in CommerceApprox. 500,000 units of all products.
DistributionWorldwide Distribution - including USA and countries of Australia, Colombia, Germany, Netherlands, and Peru.