| | Class 2 Device Recall Ponsky Deluxe "Pull" PEG |  |
| Date Initiated by Firm | December 04, 2008 |
| Date Posted | September 15, 2009 |
| Recall Status1 |
Terminated 3 on November 09, 2011 |
| Recall Number | Z-2200-2009 |
| Recall Event ID |
50678 |
| 510(K)Number | K033562 K061021 K915841 K972025 K972101 |
| Product Classification |
Polypectomy snare for Percutaneous Endoscopic Gastrostomy - Product Code KNT
|
| Product | Bard Access Systems, Inc., Ponsky* Deluxe "Pull" PEG (Percutaneous Endoscopic Gastrostomy) Kit with Soft Silicone Retention Dome, REF 000627, 20F. |
| Code Information |
Lots: 43EQA069, 43FQA069, 43HQA070, 43HQA191, 43IQA074, 43JQA070, 43KQA092, 43LQA094, HURB2394, HURD0626, HURE2809, HURF0391, HURG1372, HURH0858, HURJ0535, HURL0222, HUSH1579. |
| FEI Number |
3006260740
|
Recalling Firm/ Manufacturer |
C.R. Bard Access Systems, Inc 5425 Amelia Earhart Dr Salt Lake City UT 84116-3713
|
| For Additional Information Contact | 801-595-0700 |
Manufacturer Reason for Recall | Snare can detach in patient. The Polypectomy Snares and the Entake Peg System (which contain these snares) are being recalled because the looped portion of the snare can become detached and drop into the stomach or other internal canal during use, necessitating retrieval of the remaining piece. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Bard contacted all 1,883 worldwide customers who received the distributed suspect 467 product code lots, comprising 177, 614 units that were either not expired or were still available in the field. These will be returned by the customer to Bard Access Systems. 1-800-290-1689. |
| Quantity in Commerce | 177,614 units |
| Distribution | Worldwide, including USA, Belgium, Canada, Australia, and Japan. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KNT 510(K)s with Product Code = KNT 510(K)s with Product Code = KNT 510(K)s with Product Code = KNT
|
|
|
|