| | Class 2 Device Recall ACUSON Sequoia System |  |
| Date Initiated by Firm | May 15, 2009 |
| Date Posted | April 29, 2011 |
| Recall Status1 |
Terminated 3 on April 29, 2011 |
| Recall Number | Z-2115-2011 |
| Recall Event ID |
53013 |
| Product Classification |
System, imaging, pulsed doppler ultrasonic - Product Code IYN
|
| Product | ACUSON Sequoia C512/512SS Adv PE 10.0 System, model number 10041127. Distributed by Siemens Medical Solutions USA, Inc., Business Unit Ultrasound, Mountain View, CA.
The ACUSON Sequoia is a general purpose Diagnostic Ultrasound system. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in: B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, Harmonic Imaging or 3D imaging on a FPD Display. The ACUSON Sequoia is intended for the following applications: Abdominal, General Imaging and Cardiolgoy for Fetal, Abdominal, Intraoperative (abdominal and neurological), Pediatrics, Small Organs (breat, testes, thyroid and penis), Neonatal/Adult Cephalic, Cardiac (adult, pediatric and neonatal), Transesophageal, Transrectal, Transvaginal, Peripheral Vessels and Musculo-sceletal (superficial and conventional) applications. The system also provides for the measurement of anatomical structures that provide information that is used for clinical diagnosis purposes. |
| Code Information |
Serial numbers: 66214, 69268, 69302, and 69356. |
| FEI Number |
2936884
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc. 1230 Shorebird Way P.O. Box 7393 Mountain View CA 94043
|
| For Additional Information Contact | Kristen Dorrough 650-694-5993 |
Manufacturer Reason for Recall | Shock Hazard. The manufacturer failed to set the system power switch to "ON" during the line level leakage and high pot current draw testing. |
FDA Determined Cause 2 | Release of Material/Component prior to receiving test results |
| Action | Siemens Medical Solutions USA, Inc. sent a Customer Safety Advisory Notification letter out on May 15, 2009. Service engineers were to visit each customer to conduct necessary testing.
For questions regarding this recall call 650-694-5993. |
| Quantity in Commerce | 4 units |
| Distribution | Worldwide Distribution - USA including CA, IL, MD, NC, OH, and UT and the countries of Australia, China, Denmark, Spain, South Korea, Norway, Poland, Russia and Sweden. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|