| Date Initiated by Firm | November 17, 2009 |
| Date Posted | January 05, 2010 |
| Recall Status1 |
Terminated 3 on June 04, 2012 |
| Recall Number | Z-0589-2010 |
| Recall Event ID |
53843 |
| 510(K)Number | K871254 K884520 |
| Product Classification |
ventilator - Product Code CBK
|
| Product | Maquet Servo Ventilator SV900C/D/E (SV900)
Siemens-Elema AB, Solna, Sweden |
| Code Information |
Serial Numbers 010001-188499 Ventilators within the specified serial number range, that are marked with one of the following labels on the inside of the lid of the pneumatic unit are exempted since they have already been fitted with a new type of the pressure transducer. 1. Upgrade EMO 12/02/1; 2. Upgrade SV300 6682-745 SV900 66 82 747. |
Recalling Firm/ Manufacturer |
MAQUET Inc. 45 Barbour Pond Drive Wayne NJ 07470
|
| For Additional Information Contact | James Molloy 973-709-7227 |
Manufacturer Reason for Recall | Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers. |
FDA Determined Cause 2 | Device Design |
| Action | Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express. |
| Quantity in Commerce | 29364 (8438 of those in US) |
| Distribution | Worldwide distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CBK 510(K)s with Product Code = CBK
|