Class 1 Device Recall Closed IV Catheter system

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Closed IV Catheter systemsee related information
Date Initiated by FirmOctober 19, 2009
Date PostedFebruary 02, 2010
Recall Status1 Terminated 3 on October 19, 2010
Recall NumberZ-0836-2010
Recall Event ID 53749
510(K)NumberK032843 
Product Classification Intravascular administration set. - Product Code FOZ
ProductBD Nexiva Closed IV Catheter System, REF 383538, 20GA 1.75IN, Becton Dickinson Infusion Therapy Systems Inc.
Code Information Lot number: 8312220
FEI Number 1710034
Recalling Firm/
Manufacturer
Becton Dickinson Medical Systems
9450 S State St
Sandy UT 84070-3213
For Additional Information Contact
801-565-2341
Manufacturer Reason
for Recall
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
FDA Determined
Cause 2
Process control
ActionConsignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Quantity in Commerce5200 units
DistributionNationwide distribution.
Total Product Life CycleTPLC Device Report