|
Class 1 Device Recall LIFEPAK 15 Monitor/Defibrillator |
|
Date Initiated by Firm |
March 04, 2010 |
Date Posted |
April 21, 2010 |
Recall Status1 |
Terminated 3 on January 03, 2012 |
Recall Number |
Z-1410-2010 |
Recall Event ID |
54894 |
510(K)Number |
K082937
|
Product Classification |
Automated external defibrillators (non-wearable) - Product Code MKJ
|
Product |
Biphasic LIFEPAK 15 Monitor/Defibrillator. Manufactured by Physio-Control Inc., a division of Medtronic, Inc. Redmond, Washington USA.
The LIFEPAK 15 monitor/defibrillator is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient protocols.
|
Code Information |
Product part numbers: V15-2-000xxxxxx. Serial Numbers: various, non-contiguous as follows: Device Serial Numbers: 37799911, 37799913, 37799914, 37799915, 37800767, 37800768, 37800769, 37800771, 37803641, 37803642, 37803643, 37803644, 37803645, 37803646, 37803647, 37803648, 37803650, 37803651, 37804604, 37804605, 37804606, 37804607, 37804615, 37805136, 37805137, 37805138, 37805139, 37805140, 37805141, 37805142, 37805143, 37805144, 37805145, 37805146, 37805147, 37805148, 37805149, 37805150, 37805151, 37805152, 37805153, 37805154, 37805155, 37805156, 37805157, 37805158, 37805159, 37805160, 37805161, 37805162, 37805163, 37805164, 37805165, 37805166, 37805167, 37805168, 37805169, 37805170, 37805171, 37805172, 37805173, 37805174, 37805182, 37805183, 37805184, 37805185, 37805186, 37805187, 37805188, 37805189, 37805190, 37805191, 37805192, 37805193, 37805194, 37805195, 37805196, 37805197, 37805198, 37805199, 37805201, 37805202, 37808586, 37808587, 37808588, 37808589, 37808590, 37808591, 37808592, 37808593, 37808595, 37808596, 37808597, 37808598, 37808599, 37808600, 37810775, 37810776, 37810777, 37810778, 37810779, 37810780, 37810781, 37810782, 37810783, 37810784, 37810785, 37810786, 37810787, 37810788, 37810789, 37810819, 37810820, 37810821, 37810823, 37811957, 37811958, 37811959, 37811960, 37811961, 37811962, 37811963, 37811964, 37811966, 37811967, 37811969, 37811970, 37811971, 37811972, 37812496, 37812497, 37812498, 37812499, 37812500, 37812501, 37812502, 37812503, 37812504, 37812505, 37812506, 37812507, 37812508, 37812509, 37812510, 37869570, 37869571, 37869572, 37869574, 37869575, 37869576, 37869578, 37869580, 37869582, 37869583, 37869784, 37869785, 37869786, 37869787, 37869788, 37869789, 37869790, 37869791, 37869792, 37869793, 37869800, 37869801, 37869808, 37869809, 37869810, 37869811, 37869812, 37869814, 37869815, 37869816, 37869817, 37869818, 37869820, 37869821, 37869822, 37869823, 37869824, 37869825, 37869826, 37869827, 37869828, 37869829, 37870210, 37870211, 37870212, 37870213, 37870214, 37870215, 37870216, 37870217, 37870218, 37870219, 37870220, 37870221, 37870222, 37870223, 37870224, 37870225, 37870226, 37870228, 37870229, 37870236, 37870245, 37870247, 37870249, 37870250, 37870252, 37870254, 37870255, 37870258, 37870259, 37870260, 37900478, 37900488, 37900602, 37900603, 37900604, 37900608, 37900610, 37900612, 37900615, 37900616, 37900617, 37900623, 37900624, 37900625, 37900626, 37900628, 37900629, 37900630, 37900631, 37900632, 37900633, 37923078, 37923087, 37923100, 37923112, 37923125, 37923155, 37923156, 37923162, 37923201, 37923203, 37923204, 37923205, 37923206, 37923207, 37923208, 37923210, 37923211, 37923213, 37923214, 37924152, 37924153, 37924154, 37924155, 37924156, 37924160, 37924163, 37924164, 37924165, 37928454, 37928455, 37928456, 37928457, 37928458, 37928459, 37928460, 37928461, 37928462, 37928463, 37928790, 37928791, 37928792, 37928793, 37928794, 37928958, 37928960, 37928961, 37928964, 37928967, 37928968, 37930940, 37930942, 37930944, 37930946, 37930948, 37930953, 37930960, 37930962, 37930964, 37930966, 37930968, 37930970, 37930972, 37930974, 37930976, 37930977, 37930979, 37930982, 37930983, 37930984, 37930986, 37930987, 37930988, 37930989, 37930990, 37930991, 37930992, 37930993, 37930994, 37930995, 37930996, 37930997, 37931000, 37931001, 37931003, 37931004, 37931005, 37933237, 37933238, 37933240, 37933241, 37933242, 37933243, 37933244, 37933245, 37933246, 37933247, 37933248, 37933249, 37933250, 37933251, 37933252, 37933253, 37933254, 37933256, 37933257, 37933259, 37933260, 37933261, 37933262, 37933263, 37933264, 37933265, 37933266, 37933267, 37933268, 37933269, 37933271, 37933273, 37933275, 37933276, 37933277, 37933278, 37933279, 37933280, 37937842, 37937843, 37937844, 37937845, 37937846, 37937847, 37937848, 37937849, 37937850, 37937851, 37937852, 37937854, 37937855, 37937856, 37937862, 37937863, 37937864, 37937865, 37937866, 37937867, 37937868, 37937869, 37937870, 37937871, 37937872, 37937873, 37937874.37937881, 37937883, 37937884, 37937885, 37937886, 37937887, 37937889, 37937890, 37937891, 37937892, 37937893, 37937894, 37937895, 37937896, 37949523, 37949524, 37949525, 37949526, 37949527, 37949528, 37949529, 37949530, 37949531, 37949532, 37949533, 37949534, 37949535, 37949536, 37949537, 37949538, 37949539, 37949540, 37949541, 37949542, 37949543, 37949544, 37949545, 37949546, 37949547, 37949548, 37949549, 37949550, 37949551, 37949552, 37949554, 37949558, 37949559, 37949560, 37949561, 37949562, 37949563, 37949564, 37949565, 37949566, 37952557, 37952558, 37952559, 37952560, 37952561, 37952563, 37952564, 37952565, 37952566, 37952567, 37952568, 37952569, 37952570, 37952571, 37952572, 37952573, 37952574, 37952576, 37952580, 37952581, 37952582, 37952583, 37952584, 37952585, 37952586, 37952587, 37952588, 37952589, 37952590, 37952591, 37952592, 37952593, 37952594, 37952595, 37955465, 37955466, 37955467, 37955468, 37955469, 37955470, 37955471, 37955472, 37955474, 37955475, 37955476, 37955477, 37955478, 37955479, 37955483, 37955491, 37955492, 37955493, 37955494, 37955495, 37955497, 37955499, 37955500, 37955501, 37955502, 37955508, 37955509, 37955510, 37955512, 37955513, 37955514, 37955516, 37955517, 37955518, 37955519, 37955521, 37955523, 37963250, 37963292, 37963295, 37963300, 37963302, 37963303, 37963322, 37963323, 37963324, 37963325, 37963327, 37963328, 37963329, 37963923, 37963926, 37963927, 37963928, 37963929, 37963930, 37963931, 37963932, 37963935, 37964552, 37964554, 37964557, 37964559, 37964560, 37964561, 37964564, 37964565, 37964567, 37964569, 37964575, 37964585, 37964587, 37964588, 37964589, 37964591, 37964594, 37964596, 37964599, 37964600, 37964601, 37964607, 37964608, 37964609, 37964610, 37964613, 37964614, 37964615, 37964617, 37964618, 37964621, 37964622, 37964623, 37964626, 37964628, 37964629, 37964630, 37968304, 37968306, 37968321, 37968323, 37968326, 37968422, 37968424, 37968425, 37968427, 37968429, 37968715, 37968716, 37968717, 37968718, 37968721, 37968722, 37968724, 37976542, 37976544, 37976545, 37976551, 37977472, 37977473, 37977475, 37977476, 37977477, 37977478, 38006814, 38006832, 38006833, 38006834, 38006836, 38006837, 38006838, 38006839, 38006841, 38046540, 38046542, 38046543, 38046544, 38046546, 38046548, 38046549, 38046550, 38046552, 38046553, 38046555, 38046556, 38046557, 38046558, 38046559, 38046561, 38046563, 38046564, 38046565, 38046566, 38046568, 38046569, 38046571, 38046572, 38046573, 38051946, 38051949, 38051950, 38051951, 38051952, 38051954, 38051955, 38051956, 38051957, 38051958, 38051960, 38051961, 38051962, 38051963, 38051964, 38051965, 38051967, 38051969, 38051970, 38051971, 38051972, 38051973, 38051974, 38051975, 38051977, 38051979, 38051980, 38051983, 38051990, 38051992, 38051993, 38051994, 38051995, 38051996, 38052002, 38052003, 38052004, 38052006, 38052008, 38052009, 38052012, 38052014, 38053197, 38053198, 38053199, 38053200, 38053201, 38053203, 38053204, 38053209, 38053210, 38053211, 38053213, 38053215, 38053216, 38053217, 38053218, 38053219, 38053220, 38053221, 38053224, 38053225, 38053229, 38053230, 38053231, 38053233, 38053234, 38053236, 38053237, 38053239, 38053241, 38053246, 38053247, 38053248, 38059562, 38059568, 38059570, 38059571, 38059573, 38059574, 38059575, 38059576, 38059578, 38059582, 38059588, 38059589, 38059591, 38059593, 38059595, 38059596, 38059598, 38059600, 38059602, 38059603, 38059604, 38059605, 38059606, 38059607, 38059608, 38062627, 38062630, 38069610, 38120768, 38120769, 38120770, 38120771, 38120772, 38120773, 38120774, 38120775, 38120776, 38120777, 38120779, 38120780, 38120781, 38120782, 38120783, 38120784, 38120785, 38120786, 38120787, 38120788, 38120789, 38120790, 38120791, 38124016, 38124017, 38124018, 38124019, 38124020, 38124022, 38124027, 38124028, 38124029, 38124030, 38124031, 38124032, 38124033, 38124034, 38124035, 38124036, 38124041, 38124042, 38124043, 38124044, 38124046, 38124047, 38124048, 38124049, 38125969, 38125970, 38125971, 38125972, 38125973, 38125974.38125975, 38125976, 38125977, 38125978, 38125979, 38125980, 38125981, 38125982, 38125983, 38125984, 38125985, 38125986, 38125987, 38125988, 38125989, 38125990, 38125991, 38125994, 38126252, 38126253, 38126254, 38126255, 38126256, 38126257, 38126258, 38126259, 38126260, 38126261, 38126262, 38126263, 38126265, 38126267, 38126268, 38126269, 38126270, 38126271, 38126272, 38126273, 38126274, 38126276, 38126277, 38127128, 38127130, 38127131, 38127132, 38127133, 38127134, 38127135, 38127136, 38127137, 38127138, 38127139, 38127140, 38127141, 38127143, 38127144, 38127145, 38127146, 38127147, 38129758, 38130272, 38130273, 38130275, 38130276, 38130277, 38130278, 38130279, 38130280, 38130281, 38130282, 38130283, 38130284, 38130285, 38130286, 38130287, 38131588, 38131589, 38131591, 38131592, 38131597, 38131598, 38131599, 38131600, 38131602, 38131603, 38131605, 38131606, 38131607, 38131608, 38131756, 38131757, 38131758, 38131759, 38131760, 38131761, 38131762, 38131763, 38131764, 38131765, 38131767, 38131768, 38131769, 38131770, 38131772, 38131773, 38131774, 38131775, 38131776, 38131777, 38131779, 38131781, 38131782, 38131842, 38131843, 38131844, 38131845, 38131937, 38132365, 38132366, 38132368, 38132370, 38132371, 38132653, 38132947, 38132948, 38132950, 38132951, 38132953, 38132954, 38133884, 38133885, 38133886, 38133887, 38133888, 38134508, 38134510, 38134511, 38134512, 38134515, 38134516, 38134517, 38134518, 38134520, 38134521, 38134522, 38134523, 38134524, 38134525, 38135120, 38135127, 38135171, 38135181, 38135186, 38135198, 38135205, 38135210, 38135211, 38135212, 38135213, 38135214, 38135215, 38135216, 38135217, 38135218, 38135219, 38135220, 38135221, 38136504, 38136505, 38136506, 38136507, 38136508, 38136509, 38136510, 38136512, 38136513, 38136514, 38136515, 38136516, 38136517, 38136518, 38136519, 38136520, 38136522, 38136523, 38137115, 38137116, 38137122, 38137134, 38137136, 38137137, 38137138, 38137143, 38137149, 38137155, 38137156, 38137161, 38137168, 38137174, 38137178, 38138844, 38138845, 38138846, 38141001, 38141002, 38141003, 38141004, 38141005, 38141006, 38141007, 38141008, 38141009, 38141010, 38141012, 38141014, 38141015, 38141016, 38141017, 38141018, 38141020, 38141023, 38141025, 38141027, 38141028, 38142062, 38142064, 38142065, 38142067, 38142068, 38142070, 38142071, 38142073, 38142077, 38142079, 38142559, 38142564, 38142565, 38142566, 38142567, 38142568, 38142569, 38142574, 38142575, 38142741, 38142742, 38142743, 38142941, 38142943, 38142944, 38142945, 38142946, 38142947, 38142948, 38142949, 38142950, 38142951, 38142952, 38143265, 38143266, 38143268, 38143269, 38143270, 38143272, 38143274, 38143275, 38143276, 38143277, 38143279, 38143280, 38143282, 38143288, 38143289, 38143290, 38143291, 38143292, 38143294, 38143296, 38143297, 38143298, 38143299, 38143300, 38143301, 38143302, 38143843, 38143844, 38143845, 38145711, 38145723, 38145724, 38147665, 38147670, 38147673, 38147680, 38147681, 38147682, 38147683, 38147684, 38147687, 38147688, 38147689, 38148017, 38148019, 38148020, 38148021, 38148026, 38148028, 38148031, 38148032, 38148033, 38148034, 38148035, 38148040, 38148041, 38148043, 38148045, 38148047, 38148048, 38148050, 38148051, 38148053, 38148054, 38148056, 38148057, 38148059, 38148060, 38148061, 38148062, 38148063, 38148065, 38148066, 38148067, 38148068, 38148069, 38148070, 38148072, 38148073, 38148074, 38148076, 38148079, 38148080, 38148081, 38148082, 38148083, 38148084, 38148085, 38148086, 38148088, 38150685, 38150687, 38150688, 38150691, 38150692, 38150693, 38150694, 38150695, 38150696, 38150697, 38150700, 38150701, 38150704, 38150705, 38150706, 38150707, 38150708, 38150709, 38150710, 38151892, 38151893, 38151894, 38151895, 38151896, 38151897, 38151898, 38151899, 38153123, 38153124, 38153126, 38153127, 38153128, 38153130, 38153131, 38153132, 38153133, 38153134, 38153135, 38153136, 38153138, 38153140, 38153141, 38153142, 38153143, 38153864, 38153865, 38153866, 38153867, 38153870, 38153871, 38153876, 38153877, 38153878.38153879, 38153880, 38153881, 38153884, 38153888, 38153889, 38153890, 38153891, 38153892, 38154511, 38154513, 38154514, 38154515, 38154516, 38154517, 38154518, 38154519, 38154520, 38154521, 38154522, 38154523, 38154524, 38154525, 38155371, 38155372, 38155376, 38155378, 38155379, 38155380, 38155381, 38155382, 38155850, 38155851, 38155856, 38155857, 38155864, 38155865, 38155882, 38155889, 38155900, 38155905, 38155906, 38155907, 38155911, 38155912, 38155913, 38155914, 38155918, 38155919, 38155920, 38155924, 38156569, 38156570, 38156571, 38156573, 38156574, 38156575, 38157896, 38157897, 38157899, 38157900, 38157901, 38157903, 38157904, 38157905, 38157906, 38157907, 38157908, 38157909, 38157910, 38157911, 38157912, 38157913, 38157914, 38157915, 38157916, 38157917, 38158071, 38158072, 38158074, 38158075, 38158076, 38158078, 38158079, 38158080, 38158081, 38158082, 38158083, 38158084, 38158085, 38158086, 38158087, 38158088, 38158089, 38158090, 38158091, 38159137, 38159138, 38159139, 38159141, 38159142, 38159144, 38159145, 38159151, 38159152, 38159153, 38159156, 38159157, 38159158, 38159159, 38159160, 38159162, 38159163, 38159165, 38159167, 38159169, 38159170, 38159171, 38159172, 38159174, 38159180, 38159181, 38159182, 38159183, 38159185, 38159187, 38159188, 38159190, 38159192, 38159195, 38159197, 38159200, 38159201, 38160075, 38160078, 38161892, 38161893, 38161894, 38161895, 38161896, 38161897, 38161898, 38161899, 38161900, 38161901, 38161902, 38161903, 38161904, 38161905, 38161906, 38161907, 38161908, 38161909, 38161910, 38161911, 38161912, 38161913, 38161915, 38161917, 38162528, 38162531, 38162535, 38162538, 38162541, 38162543, 38162545, 38162548, 38162551, 38162553, 38162557, 38162559, 38162561, 38162562, 38162565, 38162565, 38162566, 38162567, 38162573, 38162574, 38162576, 38162578, 38162580, 38162581, 38162583, 38162584, 38162586, 38162588, 38162589, 38162591, 38162592, 38162594, 38162595, 38162597, 38162598, 38162599, 38162602, 38162604, 38162605, 38162606, 38162609, 38162610, 38164390, 38164391, 38164393, 38164394, 38164395, 38164396, 38164397, 38164398, 38164399, 38164400, 38164401, 38164402, 38164403, 38164404, 38164405, 38164406, 38164407, 38164408, 38164409, 38164414, 38164420, 38164421, 38164422, 38164423, 38164424, 38164425, 38164426, 38164427, 38164428, 38164429, 38164430, 38164431, 38164432, 38164433, 38164434, 38164435, 38164436, 38164437, 38164438, 38164439, 38164440, 38164441, 38164442, 38164444, 38164445, 38164446, 38164447, 38164448, 38164449, 38166442, 38166443, 38166445, 38166446, 38166448, 38166449, 38166450, 38166451, 38166452, 38166453, 38166454, 38166455, 38166456, 38166457, 38166458, 38166459, 38166460, 38166461, 38166462, 38166463, 38166464, 38166465, 38166466, 38166467, 38166468, 38166473, 38166474, 38166475, 38166476, 38166477, 38166478, 38166479, 38166481, 38166483, 38166485, 38166486, 38166488, 38166489, 38166490, 38166491, 38166492, 38166493, 38166494, 38166496, 38166497, 38166498, 38166500, 38166507, 38166509, 38166510, 38166511, 38166512, 38166513, 38166515, 38166516, 38166517, 38166518, 38166519, 38166520, 38166521, 38166522, 38166523, 38166524, 38166525, 38166526, 38166527, 38166528, 38166529, 38166530, 38166531, 38166532, 38166533, 38166535, 38166536, 38166541, 38166542, 38166543, 38166544, 38166545, 38166546, 38166547, 38166548, 38166549, 38166550, 38166551, 38166552, 38166554, 38166555, 38166556, 38166558, 38166560, 38166561, 38166562, 38166563, 38166564, 38166565, 38166566, 38166567, 38166569, 38166570, 38166571, 38166573, 38166574, 38166575, 38166576, 38166577, 38166578, 38166579, 38166580, 38166581, 38166582, 38166584, 38166586, 38166587, 38166594, 38166595, 38166596, 38166597, 38166598, 38166599, 38166601, 38166602, 38166603, 38166604, 38166605, 38166606, 38166607, 38166608, 38166609, 38166610, 38166612, 38166613, 38166614, 38166615, 38166616, 38166618, 38166619, 38166621, 38166622, 38166627, 38166628, 38166629, 38166630, 38166631, 38166632, 38166633, 38166634, 38166635, 38166636, 38166639.38166640, 38166641, 38166642, 38166643, 38166645, 38166646, 38166647, 38166648, 38166651, 38166652, 38166653, 38166654, 38166655, 38166656, 38166657, 38166658, 38166662, 38166672, 38166673, 38166674, 38166675, 38166676, 38166677, 38166678, 38166679, 38166681, 38166682, 38166683, 38166684, 38170441, 38170442, 38177261, 38177262, 38177263, 38177264, 38177265, 38177265, 38177266, 38177266, 38177267, 38177268, 38177269, 38177269, 38177270, 38177271, 38177272, 38177273, 38177274, 38177275, 38177276, 38177277, 38191198, 38191200, 38191202, 38191204, 38191205, 38191206, 38191208, 38191209, 38191210, 38191213, 38191214, 38191215, 38191216, 38191217, 38191219, 38191220, 38191221, 38191222, 38191223, 38191226, 38191227, 38191228, 38191230, 38191231, 38191234, 38191235, 38191237, 38192726, 38192727, 38192729, 38200432, 38200433, 38200434, 38200435, 38200436, 38200437, 38200438, 38200439, 38200440, 38200441, 38200442, 38200443, 38200444, 38200445, 38200446, 38200447, 38200448, 38200449, 38200450, 38200451, 38200452, 38200453, 38200454, 38200455, 38200456, 38200457, 38200458, 38200459, 38200460, 38200461, 38200462, 38200463, 38200464, 38200465, 38200466, 38200467, 38200468, 38200469, 38200470, 38200471, 38200472, 38200473, 38200474, 38200475, 38200476, 38202787, 38202790, 38202791, 38202802, 38202803, 38202805, 38202805, 38202806, 38202806, 38202807, 38202807, 38202808, 38202809, 38202811, 38202812, 38202813, 38202814, 38202815, 38202818, 38202819, 38202820, 38202821, 38202822, 38202823, 38202824, 38202825, 38202826, 38202827, 38205393, 38205394, 38205395, 38205396, 38205397, 38205398, 38205399, 38205401, 38205402, 38205403, 38205404, 38205405, 38205406, 38205407, 38205408, 38205410, 38205412, 38205413, 38205767, 38205775, 38205775, 38205776, 38205777, 38205777, 38205778, 38205779, 38205780, 38205781, 38205782, 38205783, 38205784, 38205785, 38205786, 38205786, 38205787, 38205790, 38205791, 38205792, 38205793, 38205793, 38205794, 38205795, 38205796, 38205797, 38205798, 38205799, 38205800, 38206149, 38206150, 38206152, 38206152, 38206154, 38206154, 38206155, 38206155, 38206253, 38206254, 38206255, 38206255, 38206256, 38206257, 38211038, 38211039, 38211041, 38211042, 38211043, 38211044, 38211045, 38211046, 38211049, 38211051, 38211051, 38211052, 38211053, 38211054, 38211056, 38211058, 38211059, 38211059, 38211060, 38211061, 38211231, 38211233, 38211234, 38211236, 38211238, 38211239, 38211240, 38211241, 38211242, 38211243, 38211244, 38211245, 38211246, 38211253, 38211254, 38211255, 38211256, 38211263, 38211266, 38212757, 38212758, 38212761, 38212762, 38214235, 38214236, 38214237, 38214238, 38214239, 38214240, 38214241, 38214242, 38214243, 38214243, 38214244, 38214245, 38214247, 38214248, 38214249, 38214250, 38214251, 38214252, 38214253, 38214254, 38214255, 38214257, 38214259, 38214261, 38214262, 38214263, 38214264, 38214266, 38214268, 38214269, 38214270, 38214271, 38214272, 38214274, 38214275, 38214276, 38214278, 38214279, 38214280, 38214282, 38214283, 38217382, 38217387, 38217388, 38217390, 38217391, 38217393, 38217394, 38217395, 38217396, 38217397, 38217399, 38217400, 38217401, 38217403, 38217406, 38217407, 38217408, 38217410, 38221441, 38221442, 38221443, 38221444, 38221445, 38221452, 38221453, 38221454, 38221455, 38221456, 38221457, 38221458, 38221460, 38221461, 38224391, 38224392, 38224393, 38224394, 38224395, 38224397, 38224398, 38224399, 38224400, 38224995, 38224996, 38224997, 38224998, 38224999, 38225063, 38225064, 38225067, 38225068, 38225069, 38226963, 38226964, 38226965, 38226966, 38226968, 38227019, 38227023, 38227024, 38227025, 38227128, 38227129, 38227131, 38227132, 38227133, 38227461, 38227462, 38227463, 38227464, 38227465, 38228190, 38228195, 38228196, 38228199, 38228658, 38228711, 38228712, 38228713, 38228714, 38230541, 38230543, 38230544, 38230545, 38230546, 38230547, 38230548, 38230549, 38230550, 38230551, 38230552, 38230553, 38230554, 38230555, 38230556, 38230558, 38230559, 38230560, 38230562, 38243523, 38243528, 38243530.38243531, 38243532, 38243534, 38243535, 38243536, 38243537, 38243538, 38243713, 38243723, 38243725, 38243727, 38243728, 38243730, 38243732, 38243733, 38243735, 38243736, 38244440, 38244441, 38244442, 38244443, 38244444, 38244445, 38244446, 38244447, 38244448, 38244449, 38245003, 38245004, 38245005, 38245007, 38245008, 38245009, 38245010, 38245011, 38245012, 38245013, 38245125, 38245126, 38245127, 38245128, 38245129, 38245130, 38245131, 38245132, 38245133, 38245134, 38246864, 38246865, 38246866, 38246867, 38246869, 38246872, 38246873, 38246875, 38246876, 38246877, 38247011, 38247012, 38251135, 38251136, 38251137, 38251138, 38251139, 38251141, 38251142, 38251143, 38251144, 38251145, 38253471, 38253472, 38253473, 38253474, 38253475, 38253476, 38253477, 38253478, 38253479, 38253480, 38253765, 38253766, 38253767, 38253768, 38253769, 38253770, 38253771, 38253772, 38253773, 38253774, 38253775, 38253776, 38253777, 38253778, 38253779, 38253780, 38253781, 38253782, 38253783, 38253786, 38253787, 38253788, 38253789, 38253790, 38253792, 38253793, 38253794, 38253795, 38253797, 38253798, 38253799, 38253800, 38253801, 38253802, 38253803, 38253804, 38253805, 38253806, 38254941, 38254942, 38254944, 38254944, 38254946, 38254949, 38254951, 38254953, 38254955, 38254961, 38254963, 38255180, 38255181, 38255182, 38255189, 38255191, 38255192, 38255193, 38255196, 38255410, 38255411, 38255413, 38255415, 38255416, 38255417, 38255419, 38255420, 38255421, 38255423, 38259445, 38259446, 38259447, 38259448, 38259449, 38259450, 38259451, 38259452, 38259453, 38259454, 38259455, 38259456, 38259457, 38259458, 38259460, 38259461, 38259466, 38259471, 38260845, 38260846, 38260848, 38260849, 38260852, 38263075, 38263078, 38263080, 38263083, 38263085, 38263087, 38263090, 38263092, 38263094, 38263298, 38263300, 38263301, 38263302, 38263304, 38263309, 38263310, 38263312, 38263313, 38263315, 38263317, 38263318, 38263321, 38263324, 38263326, 38263327, 38263328, 38263329, 38263330, 38263331, 38265933, 38265934, 38265935, 38265943, 38265946, 38265949, 38265956, 38265962, 38265965, 38265966, 38266199, 38266200, 38266201, 38266214, 38266228, 38266228, 38266334, 38266335, 38266344, 38266344, 38266345, 38271164, 38271165, 38271166, 38271167, 38271168, 38271169, 38271170, 38272211, 38272212, 38272212, 38272213, 38272214, 38272215, 38272216, 38272217, 38272218, 38272219, 38272220, 38273849, 38273851, 38273852, 38273853, 38273854, 38273855, 38273857, 38273859, 38273860, 38273860, 38273862, 38274175, 38274176, 38274177, 38274178, 38274179, 38274180, 38274181, 38274182, 38274183, 38274183, 38274184, 38274185, 38274186, 38274187, 38274320, 38274331, 38274341, 38274342, 38274343, 38274345, 38274346, 38274390, 38274391, 38274396, 38274398, 38274400, 38274401, 38274402, 38274403, 38274404, 38274502, 38274503, 38274504, 38274506, 38274507, 38274508, 38274509, 38274510, 38274512, 38274513, 38279197, 38279198, 38279199, 38279200, 38279201, 38279202, 38279203, 38279204, 38279205, 38279206, 38279206, 38281756, 38281758, 38281759, 38281760, 38281761, 38281762, 38281771, 38281772, 38281773, 38281774, 38281998, 38281999, 38282000, 38282001, 38282002, 38282003, 38282004, 38282783, 38282784, 38282785, 38282785, 38282786, 38282786, 38282787, 38282788, 38282789, 38282790, 38282791, 38282976, 38282977, 38282978, 38282979, 38282980, 38282982, 38282983, 38282984, 38282985, 38282986, 38282987, 38282988, 38282991, 38282992, 38282993, 38287523, 38287524, 38287525, 38287526, 38287527, 38287528, 38287528, 38287529, 38287530, 38287531, 38287532, 38287831, 38287832, 38287837, 38287839, 38287840, 38287841, 38287843, 38287844, 38287845, 38287845, 38287846, 38288301, 38288302, 38288303, 38288304, 38288305, 38288317, 38288318, 38288318, 38288319, 38288320, 38288321, 38288321, 38288322, 38288322, 38288323, 38288324, 38288325, 38288326, 38288815, 38288817, 38288818, 38288819, 38288822, 38288823, 38288824, 38288825, 38290133, 38290135, 38290136, 38290137, 38290137, 38290138, 38290139, 38290140, 38290141, 38290142.38290533, 38290534, 38290535, 38290536, 38290538, 38290539, 38290540, 38290541, 38290542, 38290543, 38291325, 38291328, 38291329, 38291330, 38291331, 38291333, 38291334, 38291957, 38291958, 38291959, 38291960, 38291961, 38291962, 38291963, 38291964, 38291965, 38291966, 38291969, 38291970, 38291971, 38291973, 38291974, 38291985, 38291986, 38291987, 38303880, 38303881, 38303882, 38303883, 38303884, 38303886, 38303887, 38303890, 38303891, 38303892, 38303894, 38303895, 38303896, 38303897, 38303897, 38303898, 38307498, 38307499, 38307500, 38307501, 38307503, 38307504, 38307505, 38307508, 38307509, 38307510, 38307511, 38307512, 38307513, 38307514, 38307516, 38314161, 38314162, 38314163, 38314164, 38314165, 38315309, 38315310, 38315311, 38315312, 38315312, 38315313, 38315314, 38315315, 38315316, 38315317, 38315318, 38315319, 38315320, 38315321, 38315322, 38315323, 38315324, 38315325, 38315326, 38315327, 38315328, 38315329, 38315330, 38315331, 38315332, 38315333, 38316177, 38316178, 38322503, 38322504, 38322505, 38322506, 38322507, 38322509, 38322510, 38322511, 38322512, 38322514, 38322515, 38322516, 38322517, 38322518, 38322519, 38322520, 38322522, 38322523, 38322524, 38322525, 38323219, 38323233, 38323245, 38323249, 38323251, 38323252, 38323254, 38323255, 38323256, 38323257, 38323262, 38323263, 38323265, 38323268, 38323270, 38323311, 38323330, 38323331, 38324195, 38324196, 38324198, 38324199, 38324200, 38324201, 38324202, 38324203, 38324207, 38324209, 38324210, 38324211, 38324212, 38324213, 38324214, 38324215, 38324216, 38324217, 38327179, 38327180, 38327181, 38327182, 38327183, 38329716, 38329717, 38329718, 38329719, 38329720, 38332345, 38332346, 38332347, 38332348, 38332349, 38332350, 38332351, 38332352, 38332353, 38332354, 38332355, 38332356, 38332358, 38332359, 38332360, 38332361, 38332362, 38332363, 38332364, 38332365, 38332748, 38332749, 38332752, 38332753, 38332754, 38332756, 38332758, 38332759, 38332761, 38332762, 38332764, 38332765, 38332766, 38332767, 38332769, 38332770, 38332771, 38332772, 38332774, 38332940, 38332941, 38332942, 38332943, 38332944, 38335249, 38335250, 38335251, 38335252, 38335253, 38335254, 38335255, 38335256, 38335257, 38335258, 38335260, 38335261, 38335262, 38335263, 38335264, 38335265, 38335266, 38335267, 38335268, 38335712, 38335713, 38335714, 38335715, 38335716, 38335717, 38335718, 38335719, 38335720, 38335722, 38335723, 38335724, 38335725, 38335726, 38335727, 38335729, 38335731, 38335732, 38335733, 38335735, 38337729, 38337730, 38337731, 38337732, 38337733, 38337734, 38337735, 38337736, 38337737, 38337738, 38337860, 38337861, 38337862, 38337864, 38337865, 38337866, 38337867, 38337868, 38337870, 38337871, 38337872, 38337873, 38337874, 38337876, 38337877, 38337878, 38337879, 38337880, 38337881, 38337882, 38338635, 38338636, 38338637, 38338638, 38338640, 38338642, 38338643, 38338644, 38338645, 38338646, 38338648, 38338649, 38338650, 38338651, 38338652, 38338655, 38338656, 38338657, 38338658, 38338659, 38340547, 38340548, 38340550, 38340551, 38340552, 38340553, 38340554, 38340555, 38340556, 38340557, 38340558, 38340559, 38340560, 38340561, 38340562, 38340563, 38340564, 38340565, 38340566, 38340567, 38341858, 38341859, 38341860, 38341974, 38341975, 38341977, 38341978, 38341979, 38341980, 38341984, 38341985, 38341986, 38341988, 38341989, 38341990, 38341991, 38341992, 38341993, 38341994, 38341995, 38341997, 38341999, 38342000, 38342001, 38342002, 38342003, 38342004, 38342006, 38342012, 38343129, 38343130, 38343131, 38343132, 38343133, 38343134, 38343135, 38343136, 38343137, 38343138, 38343140, 38343142, 38343144, 38343145, 38343146, 38343147, 38343148, 38343149, 38343151, 38343152, 38343154, 38343155, 38343156, 38343157, 38343158, 38343159, 38343160, 38343161, 38343162, 38343163, 38343164, 38343165, 38343166, 38343167, 38343168, 38343169, 38343170, 38343171, 38343172, 38343173, 38343230, 38343231, 38343231, 38343232, 38343233, 38343234, 38343235, 38343236, 38343236, 38343237, 38343238, 38343239, 38343240.38343241, 38343242, 38343243, 38343244, 38343245, 38343246, 38343247, 38343248, 38343249, 38344736, 38344737, 38344740, 38344742, 38344743, 38344744, 38344745, 38344747, 38344749, 38344750, 38344752, 38344753, 38344755, 38344757, 38344771, 38344786, 38344801, 38344808, 38344810, 38344811, 38345045, 38345048, 38345050, 38345051, 38345055, 38345178, 38345179, 38345180, 38345181, 38345182, 38346548, 38346549, 38346550, 38346552, 38346554, 38346555, 38346556, 38346557, 38346558, 38346559, 38346560, 38346561, 38346562, 38346563, 38346564, 38346565, 38346566, 38346567, 38346568, 38346570, 38350204, 38350205, 38350206, 38350207, 38350208, 38350209, 38350210, 38350211, 38350212, 38350213, 38350214, 38350215, 38353002, 38353016, 38353017, 38353018, 38353019, 38353020, 38353022, 38353024, 38353025, 38353026, 38353028, 38353029, 38353030, 38353031, 38353033, 38353034, 38353036, 38353037, 38353039, 38353040, 38353530, 38353531, 38353532, 38353533, 38353534, 38353535, 38353536, 38353537, 38353538, 38353539, 38353540, 38353541, 38353542, 38353543, 38353544, 38353545, 38353546, 38353547, 38353548, 38353549, 38354628, 38354629, 38354630, 38354631, 38354632, 38354633, 38354634, 38354635, 38354636, 38354637, 38354790, 38354791, 38354792, 38354793, 38354794, 38354795, 38354796, 38354797, 38354798, 38354799, 38354800, 38354801, 38354802, 38354803, 38354804, 38354805, 38354806, 38354807, 38354809, 38354810, 38355625, 38355626, 38355627, 38355628, 38355629, 38355630, 38355631, 38355632, 38355633, 38355635, 38355636, 38355637, 38355638, 38355639, 38355640, 38355641, 38355642, 38355643, 38355644, 38355645, 38359017, 38359018, 38359019, 38359024, 38359026, 38359027, 38359028, 38359029, 38359030, 38359035, 38359556, 38359557, 38359558, 38359559, 38359560, 38359561, 38359562, 38359563, 38359564, 38359565, 38359816, 38359817, 38359820, 38359821, 38359823, 38359824, 38359827, 38359829, 38359831, 38360323, 38360324, 38360325, 38360326, 38360327, 38360328, 38360329, 38360332, 38360334, 38360336, 38361932, 38361933, 38361934, 38361935, 38361937, 38361938, 38361939, 38361939, 38361940, 38361941, 38361943, 38361943, 38361944, 38361945, 38361946, 38361947, 38361947, 38361987, 38361988, 38361989, 38361990, 38361991, 38362760, 38362760, 38362761, 38362762, 38362763, 38362765, 38362766, 38362767, 38362768, 38362769, 38362771, 38362854, 38362855, 38362856, 38362857, 38362858, 38362859, 38362860, 38362861, 38362862, 38362863, 38363399, 38363401, 38363403, 38363404, 38363405, 38363407, 38363412, 38363426, 38363448, 38363451, 38364965, 38364966, 38364967, 38364968, 38364969, 38364970, 38364971, 38364972, 38364973, 38364974, 38364975, 38364976, 38364977, 38364978, 38364979, 38364980, 38364981, 38364982, 38364983, 38364984, 38365243, 38365244, 38365245, 38365247, 38365248, 38365249, 38365254, 38365255, 38365256, 38365257, 38365567, 38365571, 38365573, 38365574, 38365576, 38365577, 38365578, 38365579, 38365580, 38365581, 38365763, 38365764, 38365765, 38365766, 38365767, 38365768, 38365769, 38365770, 38365771, 38365772, 38365937, 38365938, 38365939, 38365940, 38365941, 38365942, 38365943, 38365944, 38365945, 38365946, 38365946, 38368303, 38368304, 38368305, 38368306, 38368307, 38368308, 38368309, 38368310, 38368311, 38368312, 38368313, 38368314, 38368315, 38368316, 38368318, 38368323, 38368325, 38368326, 38368327, 38368329, 38368607, 38368608, 38368609, 38368610, 38368611, 38368612, 38368613, 38368614, 38368615, 38368616, 38368617, 38368618, 38368619, 38368620, 38368621, 38368622, 38368623, 38368624, 38368625, 38368627, 38369575, 38369576, 38369576, 38369578, 38369579, 38369580, 38369581, 38369582, 38369583, 38369584, 38369585, 38369586, 38369587, 38369588, 38369588, 38369589, 38369590, 38369591, 38369592, 38369593, 38369594, 38369595, 38370269, 38370270, 38370271, 38370272, 38370273, 38370274, 38370275, 38370276, 38370277, 38370278, 38370280, 38370281, 38370282, 38370283, 38370284, 38370285, 38370288, 38370290, 38370291, 38370293, 38370423, 38370424, 38370425, 38370426.38370427, 38370428, 38370429, 38370430, 38370431, 38370432, 38370433, 38370434, 38370435, 38370435, 38370436, 38370437, 38370438, 38370439, 38370440, 38370441, 38370441, 38370442, 38371193, 38371195, 38371196, 38371197, 38371198, 38371199, 38371200, 38371200, 38371201, 38371202, 38371203, 38371205, 38371210, 38371211, 38371212, 38371213, 38371214, 38371215, 38371216, 38371218, 38371219, 38371700, 38371701, 38371702, 38371705, 38371706, 38371707, 38371708, 38371709, 38371710, 38371710, 38371711, 38371712, 38371713, 38371713, 38371714, 38371715, 38371717, 38371718, 38371718, 38371719, 38371720, 38371721, 38371722, 38372518, 38372518, 38372519, 38372520, 38372521, 38372522, 38372523, 38372525, 38372526, 38372527, 38372528, 38372530, 38372531, 38372532, 38372533, 38372534, 38372535, 38372537, 38372538, 38372541, 38372543, 38373599, 38373600, 38373601, 38373602, 38373603, 38373604, 38373605, 38373606, 38373607, 38373608, 38373609, 38373610, 38373611, 38373612, 38373613, 38373614, 38373615, 38373616, 38373617, 38373618, 38374115, 38374116, 38374117, 38374118, 38374119, 38374120, 38374121, 38374122, 38374123, 38374125, 38374126, 38374127, 38374129, 38374130, 38374131, 38374132, 38374134, 38374135, 38374136, 38374138, 38374678, 38374679, 38374680, 38374681, 38374682, 38374683, 38374684, 38374686, 38374687, 38374688, 38374689, 38374690, 38374691, 38374692, 38374693, 38374696, 38374697, 38374698, 38374700, 38374701, 38388120, 38388121, 38388122, 38388123, 38388125, 38388126, 38388127, 38388129, 38388147, 38388149, 38388150, 38388152, 38392628, 38392629, 38392630, 38392632, 38392633, 38392634, 38392635, 38392636, 38392637, 38392638, 38400871, 38400873, 38400874, 38400880, 38400881, 38400882, 38400883, 38400884, 38407036, 38407037, 38407038, 38407039, 38407041, 38407147, 38407148, 38407149, 38407150, 38407151, 38407152, 38407153, 38407154, 38407155, 38407156, 38407157, 38407158, 38407159, 38407160, 38407161, 38407162, 38407163, 38407164, 38407170, 38407172, 38407173, 38407174, 38407175, 38407176, 38407177, 38407178, 38407179, 38407180, 38407181, 38407182, 38407183, 38407185, 38407188, 38407190, 38407191, 38407192, 38407193, 38407194, 38407195, 38407196, 38407530, 38407531, 38407533, 38407534, 38407535, 38410842, 38410843, 38410845, 38410846, 38410847, 38410848, 38410849, 38410850, 38410851, 38410852, 38410854, 38410855, 38410856, 38410857, 38410858, 38410860, 38410861, 38410862, 38410864, 38410865, 38410866, 38410868, 38410869, 38410872, 38410873, 38415389, 38415391, 38415392, 38415394, 38415395, 38415407, 38415408, 38415410, 38415411, 38415412, 38417308, 38418568, 38418570, 38418571, 38418572, 38418581, 38418583, 38418584, 38418585, 38418586, 38418588, 38418589, 38418590, 38418591, 38418592, 38418593, 38418594, 38418595, 38418596, 38418597, 38418598, 38418598, 38418599, 38418601, 38418602, 38418603, 38418604, 38418604, 38418605, 38418606, 38418607, 38418608, 38418609, 38418610, 38418611, 38418612, 38418613, 38428175, 38428176, 38428177, 38428178, 38428179, 38428180, 38428181, 38428182, 38428183, 38428184, 38428185, 38428188, 38428189, 38428190, 38428191, 38428192, 38428193, 38428194, 38428195, 38428196, 38432094, 38432094, 38432100, 38432100, 38432101, 38432102, 38432103, 38432104, 38432105, 38432105, 38441911, 38441912, 38441913, 38441914, 38441915, 38441916, 38441917, 38441918, 38441919, 38441920, 38441921, 38441922, 38441922, 38441923, 38441925, 38441926, 38441927, 38441928, 38441929, 38441930, 38441931, 38441932, 38441933, 38441934, 38441935, 38441936, 38445572, 38445573, 38445574, 38445579, 38445580, 38445581, 38445582, 38445583, 38445584, 38445585, 38445590, 38445591, 38445593, 38445594, 38445595, 38445597, 38445599, 38445601, 38445604, 38445607, 38445611, 38445615, 38445619, 38445624, 38445626, 38445627, 38445630, 38445638, 38445640, 38445641, 38445646, 38445649, 38445650, 38445651, 38445652, 38448508, 38448509, 38448510, 38448511, 38448512, 38448513, 38448514, 38448515, 38448516, 38448517, 38448518, 38448520, 38448521, 38448522.38449344, 38449345, 38449346, 38449347, 38449348, 38449352, 38449357, 38449358, 38449359, 38449360, 38449362, 38449363, 38449364, 38449365, 38449367, 38449369, 38449370, 38449371, 38449374, 38449377, 38449378, 38449379, 38449381, 38449382, 38449384, 38449385, 38449387, 38449388, 38449390, 38449391, 38449392, 38449394, 38449400, 38449410, 38458601, 38458602, 38458603, 38458604, 38458605, 38458606, 38458608, 38458609, 38458610, 38458667, 38458668, 38458669, 38458671, 38458672, 38458673, 38458675, 38458676, 38458677, 38458678, 38458679, 38458680, 38458681, 38458683, 38458684, 38458685, 38458686, 38458687, 38458688, 38458690, 38458691, 38458695, 38458696, 38458697, 38458698, 38458700, 38458702, 38462717, 38472556, 38472557, 38472559, 38472560, 38472561, 38472562, and 38472569. |
Recalling Firm/ Manufacturer |
Physio Control, Inc. 11811 Willows Rd Ne Redmond WA 98052-2003
|
For Additional Information Contact |
425-867-4000
|
Manufacturer Reason for Recall |
Potential for the device to power off then on by itself, or to power off by itself and requiring the operator to turn it back on, or the device doesn't turn off.
|
FDA Determined Cause 2 |
Device Design |
Action |
Physio-Control Inc. issued an "Urgent Medical Device Correction" notification dated March 2010 to domestic consignees. Foreign consignees were notified by overseas representatives. Consignees were advised to keep the affected device in service and to test the units in accordance with operating instructions. They were further advised that a local service representative will call and schedule a service visit within 60 days. Consignees were asked to forward the notification to all of their sites and to notify the recalling firm if they no longer own the device.
For further information, contact Physio-Control Inc. Technical Support at 1-800-442-1142, Option 5, 6:00 am to 4:00 pm (Pacific), Monday-Friday. |
Quantity in Commerce |
3,609 |
Distribution |
Worldwide Distribution -- United States, Netherlands, Germany, Canada, Australia, and Hong Kong. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MKJ and Original Applicant = PHYSIO-CONTROL, INC.
|
|
|
|