Class 1 Device Recall Shiley Cuffed Pediatric Tracheostomy Tube

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Shiley Cuffed Pediatric Tracheostomy Tubesee related information
Date Initiated by FirmApril 13, 2010
Date PostedMay 06, 2010
Recall Status1 Terminated 3 on January 18, 2012
Recall NumberZ-1495-2010
Recall Event ID 55354
510(K)NumberK955680 
Product Classification Tube tracheostomy and tube cuff - Product Code JOH
ProductShiley Cuffed Pediatric Tracheostomy Tube, 5.5 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper airway obstructions, provide long term ventilation, support and/or manage tracheal/bronchial secretions.
Code Information Lot Numbers: 0903000863 through 0908001211
Recalling Firm/
Manufacturer
Nellcor Puritan Bennett Inc. (dba Covidien Ltd)
6135 Gunbarrel Ave
Boulder CO 80301-3214
For Additional Information Contact
303-305-2382
Manufacturer Reason
for Recall
Inflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
FDA Determined
Cause 2
Process design
ActionCovidien issued an "Urgent Medical Device Recall" notification dated April 13, 2010. Consignees were alerted to the affected device. If an affected product is already in use, the firm recommends that the tube be replaced as soon as the patient's physician determines it is clinically appropriate. All affected product is to be returned to the firm. For further Information, contact Covidien at 1-800-635-5267.
Quantity in Commerce4,033 units
DistributionWorldwide Distribution -- United States (including Guam and Puerto Rico), Australia, Belgium, Brazil, Canada, Chile, Colombia, Honduras, Japan, Mexico, Panama, Singapore and Uruguay.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = JOH
-
-