• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Spacelabs Medical Flat Panel Display

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Spacelabs Medical Flat Panel Displaysee related information
Date Initiated by FirmMarch 20, 2009
Date PostedDecember 14, 2010
Recall Status1 Terminated 3 on June 13, 2011
Recall NumberZ-0620-2011
Recall Event ID 57370
510(K)NumberK932724 
Product Classification Display, cathode-ray tube, medical - Product Code DXJ
ProductSpacelabs Medical Flat Panel Display is a touchscreen display, 19-inch thin film transistor liquid crystal display (TFT-LCD). The PNs are 010-1619-00 and 010-1655-00, Models 94260-19 and 94256.
Code Information Date Codes (WK numbers) identified on the affected power supplies:  0647, 0703, 0706, 0707, 0709, 0712, 0724, 0727, 0728 or 0732.  Lot/Serial Numbers on the systems for US Consignees:  ELZ-0011442, ELZ-0012143, ELZ-0013307, ELZ-0013557, ELZ-0014567,  ELZ-0014570, ELZ-0014574, ELZ-0014577, ELZ-0014580, ELZ-0014583,  ELZ-0014585, ELZ-0014586, ELZ-0014587, ELZ-0014590, ELZ-0014591,  ELZ-0014592, ELZ-0014593, ELZ-0014594, ELZ-0014595, ELZ-0014596,  ELZ-0014597, ELZ-0014598, ELZ-0014599, ELZ-0014600, ELZ-0014601,  ELZ-0014603, ELZ-0014604, ELZ-0014605, ELZ-0014606, ELZ-0014607,  ELZ-0014608, ELZ-0014609, ELZ-0014610, ELZ-0014611, ELZ-0014612,  ELZ-0014613, ELZ-0014614, ELZ-0014615, ELZ-0014616, ELZ-0014617,  ELZ-0014618, ELZ-0014619, ELZ-0014620, ELZ-0014621, ELZ-0014622,  ELZ-0014623, ELZ-0014625, ELZ-0014626, ELZ-0014629, ELZ-0014630,  ELZ-0014631, ELZ-0014632, ELZ-0014633, ELZ-0014634, ELZ-0014635,  ELZ-0014636, ELZ-0014637, ELZ-0014638, ELZ-0014639, ELZ-0014640,  ELZ-0014641, ELZ-0014645, ELZ-0014648, ELZ-0014656, ELZ-0014657,  ELZ-0014658, ELZ-0014718, ELZ-0014719, ELZ-0014720, ELZ-0014723,  ELZ-0014724, ELZ-0014726, ELZ-0014727, ELZ-0014729, ELZ-0014731,  ELZ-0014733, ELZ-0014736, ELZ-0014737, ELZ-0014738, ELZ-0014740,  ELZ-0014742, ELZ-0014744, ELZ-0014745, ELZ-0014750, ELZ-0014752,  ELZ-0014754, ELZ-0014755, ELZ-0014756, ELZ-0014758, ELZ-0014759,  ELZ-0014760, ELZ-0014762, ELZ-0014764, ELZ-0014765, ELZ-0014766,  ELZ-0014767, ELZ-0014768, ELZ-0014770, ELZ-0014771, ELZ-0014772, ELZ-0014773, ELZ-0014774, ELZ-0014775, ELZ-0014776, ELZ-0014780, ELZ-0014781, ELZ-0014782, ELZ-0014783, ELZ-0014786, ELZ-0014787, ELZ-0014788, ELZ-0014790, ELZ-0014791, ELZ-0014795, ELZ-0014800, ELZ-0014805, ELZ-0014806, ELZ-0014810, ELZ-0014816, ELZ-0014817, ELZ-0014819, ELZ-0014820, ELZ-0014823, ELZ-0014824, ELZ-0014827, ELZ-0014828, ELZ-0014833, ELZ-0014834, ELZ-0014836, ELZ-0014837, ELZ-0014839, ELZ-0014842, ELZ-0014846, ELZ-0014852, ELZ-0014853, ELZ-0014854, ELZ-0014857, ELZ-0014858, ELZ-0014859, ELZ-0014861, ELZ-0014863, ELZ-0014864, ELZ-0014865, ELZ-0014866, ELZ-0014867, ELZ-0014868, ELZ-0014869, ELZ-0014872, ELZ-0014873, ELZ-0014874, ELZ-0014875, ELZ-0014876, ELZ-0014877, ELZ-0014878, ELZ-0014879, ELZ-0014880, ELZ-0014881, ELZ-0014882, ELZ-0014883, ELZ-0014885, ELZ-0014886, ELZ-0014890, ELZ-0014891, ELZ-0014892, ELZ-0014893, ELZ-0014894, ELZ-0014895, ELZ-0014896, ELZ-0014900, ELZ-0014902, ELZ-0014905, ELZ-0014906, ELZ-0014910, ELZ-0014913, ELZ-0014914, ELZ-0014915, ELZ-0014916, ELZ-0014917, ELZ-0014920, ELZ-0014922, ELZ-0014925, ELZ-0014926, ELZ-0014927, ELZ-0014930, ELZ-0014931, ELZ-0014933, ELZ-0014934, ELZ-0014935, ELZ-0014936, ELZ-0014937, ELZ-0014938, ELZ-0014940, ELZ-0014941, ELZ-0014942, ELZ-0014943, ELZ-0014944, ELZ-0014945, ELZ-0014947, ELZ-0014949, ELZ-0014950, ELZ-0014951, ELZ-0014953, ELZ-0014954, ELZ-0014955, ELZ-0014956, ELZ-0014957, ELZ-0014958, ELZ-0014959, ELZ-0014960, ELZ-0014961, ELZ-0014964, ELZ-0014965, ELZ-0014968, ELZ-0014969, ELZ-0014972, ELZ-0014973, ELZ-0014974, ELZ-0014975, ELZ-0014976, ELZ-0014977, ELZ-0014979, ELZ-0014981, ELZ-0014982, ELZ-0014983, ELZ-0014985, ELZ-0014986, ELZ-0014987, ELZ-0014989, ELZ-0014992, ELZ-0014994, ELZ-0014995, ELZ-0014996, ELZ-0014999, ELZ-0015000, ELZ-0015001, ELZ-0015002, ELZ-0015005, ELZ-0015006, ELZ-0015007, ELZ-0015008, ELZ-0015009, ELZ-0015010, ELZ-0015011, ELZ-0015012, ELZ-0015013, ELZ-0015014, ELZ-0015015, ELZ-0015016, ELZ-0015017, ELZ-0015018, ELZ-0015019, ELZ-0015020, ELZ-0015021, ELZ-0015022, ELZ-0015023, ELZ-0015024, ELZ-0015025, ELZ-0015026, ELZ-0015027, ELZ-0015028, ELZ-0015029, ELZ-0015030, ELZ-0015031, ELZ-0015032, ELZ-0015033, ELZ-0015034, ELZ-0015035, ELZ-0015036, ELZ-0015037, ELZ-0015038, ELZ-0015039, ELZ-0015040, ELZ-0015042, ELZ-0015043, ELZ-0015044, ELZ-0015045, ELZ-0015046, ELZ-0015047, ELZ-0015048, ELZ-0015050, ELZ-0015051, ELZ-0015052, ELZ-0015053, ELZ-0015054, ELZ-0015055, ELZ-0015056, ELZ-0015057, ELZ-0015058, ELZ-0015059, ELZ-0015060, ELZ-0015061, ELZ-0015063, ELZ-0015064, ELZ-0015065, ELZ-0015066, ELZ-0015067, ELZ-0015068, ELZ-0015069, ELZ-0015070, ELZ-0015071, ELZ-0015073, ELZ-0015074, ELZ-0015075, ELZ-0015076, ELZ-0015077, ELZ-0015078, ELZ-0015080, ELZ-0015081, ELZ-0015082, ELZ-0015083, ELZ-0015084, ELZ-0015085, ELZ-0015087, ELZ-0015088, ELZ-0015089, ELZ-0015090, ELZ-0015091, ELZ-0015093, ELZ-0015094, ELZ-0015095, ELZ-0015096, ELZ-0015097, ELZ-0015098, ELZ-0015099, ELZ-0015100, ELZ-0015101, ELZ-0015102, ELZ-0015103, ELZ-0015104, ELZ-0015105, ELZ-0015107, ELZ-0015113, ELZ-0015114, ELZ-0015118, ELZ-0015119, ELZ-0015120, ELZ-0015121, ELZ-0015122, ELZ-0015123, ELZ-0015125, ELZ-0015126, ELZ-0015128, ELZ-0015129, ELZ-0015130, ELZ-0015132, ELZ-0015133, ELZ-0015134, ELZ-0015135, ELZ-0015136, ELZ-0015138, ELZ-0015139, ELZ-0015140, ELZ-0015144, ELZ-0015145, ELZ-0015150, ELZ-0015154, ELZ-0015155, ELZ-0015157, ELZ-0015158, ELZ-0015159, ELZ-0015161, ELZ-0015164, ELZ-0015165, ELZ-0015166, ELZ-0015167, ELZ-0015168, ELZ-0015169, ELZ-0015173, ELZ-0015174, ELZ-0015179, ELZ-0015181, ELZ-0015184, ELZ-0015187, ELZ-0015189, ELZ-0015194, ELZ-0015197, ELZ-0015199, ELZ-0015200, ELZ-0015201, ELZ-0015203, ELZ-0015204, ELZ-0015205, ELZ-0015206, ELZ-0015207, ELZ-0015208, ELZ-0015209, ELZ-0015210, ELZ-0015212, ELZ-0015213, ELZ-0015214, ELZ-0015215, ELZ-0015216, ELZ-0015217, ELZ-0015218, ELZ-0015219, ELZ-0015220, ELZ-0015221, ELZ-0015222, ELZ-0015223, ELZ-0015224, ELZ-0015225, ELZ-0015226, ELZ-0015227, ELZ-0015228, ELZ-0015229, ELZ-0015230, ELZ-0015231, ELZ-0015232, ELZ-0015233, ELZ-0015234, ELZ-0015235, ELZ-0015236, ELZ-0015237, ELZ-0015238, ELZ-0015239, ELZ-0015240, ELZ-0015241, ELZ-0015242, ELZ-0015243, ELZ-0015244, ELZ-0015245, ELZ-0015246, ELZ-0015247, ELZ-0015248, ELZ-0015249, ELZ-0015250, ELZ-0015251, ELZ-0015252, ELZ-0015253, ELZ-0015255, ELZ-0015256, ELZ-0015257, ELZ-0015262, ELZ-0015264, ELZ-0015265, ELZ-0015266, ELZ-0015267, ELZ-0015268, ELZ-0015271, ELZ-0015273, ELZ-0015274, ELZ-0015276, ELZ-0015277, ELZ-0015279, ELZ-0015280, ELZ-0015281, ELZ-0015282, ELZ-0015284, ELZ-0015285, ELZ-0015286, ELZ-0015288, ELZ-0015291, ELZ-0015292, ELZ-0015293, ELZ-0015294, ELZ-0015295, ELZ-0015296, ELZ-0015297, ELZ-0015298, ELZ-0015299, ELZ-0015300, ELZ-0015301, ELZ-0015302, ELZ-0015303, ELZ-0015304, ELZ-0015305, ELZ-0015306, ELZ-0015307, ELZ-0015308, ELZ-0015309, ELZ-0015310, ELZ-0015311, ELZ-0015312, ELZ-0015315, ELZ-0015316, ELZ-0015317, ELZ-0015318, ELZ-0015319, ELZ-0015320, ELZ-0015321, ELZ-0015322, ELZ-0015323, ELZ-0015326, ELZ-0015327, ELZ-0015328, ELZ-0015330, ELZ-0015331, ELZ-0015332, ELZ-0015333, ELZ-0015334, ELZ-0015335, ELZ-0015336, ELZ-0015337, ELZ-0015338, ELZ-0015339, ELZ-0015340, ELZ-0015341, ELZ-0015342, ELZ-0015343, ELZ-0015344, ELZ-0015345, ELZ-0015346, ELZ-0015347, ELZ-0015348, ELZ-0015349, ELZ-0015350, ELZ-0015351, ELZ-0015352, ELZ-0015353, ELZ-0015354, ELZ-0015355, ELZ-0015357, ELZ-0015358, ELZ-0015359, ELZ-0015360, ELZ-0015361, ELZ-0015362, ELZ-0015363, ELZ-0015364, ELZ-0015365, ELZ-0015366, ELZ-0015367, ELZ-0015368, ELZ-0015369, ELZ-0015370, ELZ-0015371, ELZ-0015372, ELZ-0015373, ELZ-0015374, ELZ-0015375, ELZ-0015376, ELZ-0015377, ELZ-0015378, ELZ-0015379, ELZ-0015380, ELZ-0015381, ELZ-0015382, ELZ-0015383, ELZ-0015384, ELZ-0015385, ELZ-0015386, ELZ-0015387, ELZ-0015388, ELZ-0015389, ELZ-0015390, ELZ-0015391, ELZ-0015392, ELZ-0015393, ELZ-0015394, ELZ-0015395, ELZ-0015396, ELZ-0015397, ELZ-0015398, ELZ-0015399, ELZ-0015400, ELZ-0015401, ELZ-0015402, ELZ-0015403, ELZ-0015404, ELZ-0015405, ELZ-0015406, ELZ-0015407, ELZ-0015408, ELZ-0015409, ELZ-0015410, ELZ-0015411, ELZ-0015412, ELZ-0015413, ELZ-0015414, ELZ-0015415, ELZ-0015418, ELZ-0015419, ELZ-0015420, ELZ-0015422, ELZ-0015426, ELZ-0015427, ELZ-0015429, ELZ-0015430, ELZ-0015432, ELZ-0015434, ELZ-0015436, ELZ-0015437, ELZ-0015438, ELZ-0015439, ELZ-0015441, ELZ-0015442, ELZ-0015443, ELZ-0015444, ELZ-0015445, ELZ-0015446, ELZ-0015447, ELZ-0015448, ELZ-0015449, ELZ-0015450, ELZ-0015451, ELZ-0015452, ELZ-0015453, ELZ-0015454, ELZ-0015455, ELZ-0015456, ELZ-0015457, ELZ-0015458, ELZ-0015459, ELZ-0015460, ELZ-0015461, ELZ-0015462, ELZ-0015463, ELZ-0015464, ELZ-0015465, ELZ-0015466, ELZ-0015467, ELZ-0015468, ELZ-0015469, ELZ-0015470, ELZ-0015471, ELZ-0015472, ELZ-0015473, ELZ-0015474, ELZ-0015475, ELZ-0015476, ELZ-0015479, ELZ-0015480, ELZ-0015482, ELZ-0015485, ELZ-0015486, ELZ-0015488, ELZ-0015489, ELZ-0015490, ELZ-0015495, ELZ-0015501, ELZ-0015503, ELZ-0015506, ELZ-0015509, ELZ-0015510, ELZ-0015511, ELZ-0015523, ELZ-0015526, ELZ-0015529, ELZ-0015536, ELZ-0015537, ELZ-0015538, ELZ-0015539, ELZ-0015540, ELZ-0015541, ELZ-0015542, ELZ-0015543, ELZ-0015544, ELZ-0015545, ELZ-0015546, ELZ-0015547, ELZ-0015549, ELZ-0015550, ELZ-0015551, ELZ-0015552, ELZ-0015553, ELZ-0015554, ELZ-0015555, ELZ-0015556, ELZ-0015558, ELZ-0015559, ELZ-0015560, ELZ-0015561, ELZ-0015562, ELZ-0015564, ELZ-0015566, ELZ-0015567, ELZ-0015568, ELZ-0015570, ELZ-0015571, ELZ-0015572, ELZ-0015573, ELZ-0015574, ELZ-0015575, ELZ-0015576, ELZ-0015577, ELZ-0015578, ELZ-0015579, ELZ-0015580, ELZ-0015581, ELZ-0015582, ELZ-0015583, ELZ-0015584, ELZ-0015585, ELZ-0015586, ELZ-0015587, ELZ-0015588, ELZ-0015589, ELZ-0015590, ELZ-0015591, ELZ-0015592, ELZ-0015593, ELZ-0015594, ELZ-0015597, ELZ-0015598, ELZ-0015599, ELZ-0015600, ELZ-0015601, ELZ-0015602, ELZ-0015603, ELZ-0015604, ELZ-0015605, ELZ-0015606, ELZ-0015608, ELZ-0015609, ELZ-0015611, ELZ-0015612, ELZ-0015613, ELZ-0015614, ELZ-0015615, ELZ-0015616, ELZ-0015617, ELZ-0015618, ELZ-0015619, ELZ-0015620, ELZ-0015621, ELZ-0015622, ELZ-0015624, ELZ-0015625, ELZ-0015626, ELZ-0015646, ELZ-0015647, ELZ-0015648, ELZ-0015649, ELZ-0015650, ELZ-0015651, ELZ-0015654, ELZ-0015655, ELZ-0015656, ELZ-0015657, ELZ-0015658, ELZ-0015659, ELZ-0015660, ELZ-0015661, ELZ-0015662, ELZ-0015664, ELZ-0015665, ELZ-0015666, ELZ-0015667, ELZ-0015668, ELZ-0015669, ELZ-0015670, ELZ-0015671, ELZ-0015672, ELZ-0015673, ELZ-0015674, ELZ-0015675, ELZ-0015676, ELZ-0015677, ELZ-0015678, ELZ-0015679, ELZ-0015680, ELZ-0015681, ELZ-0015682, ELZ-0015683, ELZ-0015684, ELZ-0015685, ELZ-0015686, ELZ-0015687, ELZ-0015688, ELZ-0015689, ELZ-0015690, ELZ-0015691, ELZ-0015692, ELZ-0015693, ELZ-0015694, ELZ-0015696, ELZ-0015697, ELZ-0015698, ELZ-0015699, ELZ-0015700, ELZ-0015701, ELZ-0015702, ELZ-0015703, ELZ-0015704, ELZ-0015705, ELZ-0015706, ELZ-0015707, ELZ-0015708, ELZ-0015709, ELZ-0015710, ELZ-0015711, ELZ-0015712, ELZ-0015713, ELZ-0015714, ELZ-0015715, ELZ-0015716, ELZ-0015717, ELZ-0015718, ELZ-0015719, ELZ-0015720, ELZ-0015722, ELZ-0015723, ELZ-0015724, ELZ-0015725, ELZ-0015726, ELZ-0015727, ELZ-0015728, ELZ-0015731, ELZ-0015732, ELZ-0015736, ELZ-0015737, ELZ-0015738, ELZ-0015739, ELZ-0015740, ELZ-0015741, ELZ-0015742, ELZ-0015744, ELZ-0015745, ELZ-0015746, ELZ-0015747, ELZ-0015748, ELZ-0015749, ELZ-0015750, ELZ-0015751, ELZ-0015752, ELZ-0015753, ELZ-0015754, ELZ-0015755, ELZ-0015756, ELZ-0015758, ELZ-0015760, ELZ-0015762, ELZ-0015763, ELZ-0015764, ELZ-0015765, ELZ-0015770, ELZ-0015771, ELZ-0015772, ELZ-0015773, ELZ-0015775, ELZ-0015776, ELZ-0015777, ELZ-0015778, ELZ-0015781, ELZ-0015782, ELZ-0015783, ELZ-0015784, ELZ-0015785, ELZ-0015786, ELZ-0015787, ELZ-0015788, ELZ-0015789, ELZ-0015790, ELZ-0015791, ELZ-0015792, ELZ-0015793, ELZ-0015794, ELZ-0015795, ELZ-0015798, ELZ-0015799, ELZ-0015800, ELZ-0015801, ELZ-0015802, ELZ-0015803, ELZ-0015804, ELZ-0015805, ELZ-0015806, ELZ-0015807, ELZ-0015808, ELZ-0015809, ELZ-0015810, ELZ-0015811, ELZ-0015812, ELZ-0015813, ELZ-0015814, ELZ-0015815, ELZ-0015816, ELZ-0015817, ELZ-0015818, ELZ-0015819, ELZ-0015820, ELZ-0015821, ELZ-0015822, ELZ-0015823, ELZ-0015824, ELZ-0015825, ELZ-0015826, ELZ-0015827, ELZ-0015828, ELZ-0015829, ELZ-0015830, ELZ-0015831, ELZ-0015832, ELZ-0015833, ELZ-0015834, ELZ-0015835, ELZ-0015836, ELZ-0015837, ELZ-0015838, ELZ-0015839, ELZ-0015840, ELZ-0015841, ELZ-0015842, ELZ-0015843, ELZ-0015844, ELZ-0015845, ELZ-0015846, ELZ-0015847, ELZ-0015848, ELZ-0015849, ELZ-0015850, ELZ-0015851, ELZ-0015852, ELZ-0015853, ELZ-0015854, ELZ-0015855, ELZ-0015856, ELZ-0015857, ELZ-0015858, ELZ-0015859, ELZ-0015860, ELZ-0015861, ELZ-0015862, ELZ-0015863, ELZ-0015864, ELZ-0015865, ELZ-0015866, ELZ-0015867, ELZ-0015868, ELZ-0015869, ELZ-0015870, ELZ-0015871, ELZ-0015872, ELZ-0015873, ELZ-0015874, ELZ-0015875, ELZ-0015876, ELZ-0015877, ELZ-0015878, ELZ-0015879, ELZ-0015880, ELZ-0015881, ELZ-0015882, ELZ-0015883, ELZ-0015884, ELZ-0015885, ELZ-0015886, ELZ-0015887, ELZ-0015888, ELZ-0015889, ELZ-0015890, ELZ-0015891, ELZ-0015892, ELZ-0015893, ELZ-0015895, ELZ-0015896, ELZ-0015897, ELZ-0015898, ELZ-0015900, ELZ-0015901, ELZ-0015902, ELZ-0015903, ELZ-0015904, ELZ-0015905, ELZ-0015906, ELZ-0015907, ELZ-0015908, ELZ-0015910, ELZ-0015911, ELZ-0015912, ELZ-0015913, ELZ-0015914, ELZ-0015915, ELZ-0015916, ELZ-0015917, ELZ-0015918, ELZ-0015919, ELZ-0015920, ELZ-0015921, ELZ-0015922, ELZ-0015923, ELZ-0015924, ELZ-0015925, ELZ-0015926, ELZ-0015927, ELZ-0015928, ELZ-0015929, ELZ-0015930, ELZ-0015931, ELZ-0015932, ELZ-0015933, ELZ-0015934, ELZ-0015935, ELZ-0015936, ELZ-0015937, ELZ-0015938, ELZ-0015939, ELZ-0015940, ELZ-0015941, ELZ-0015943, ELZ-0015944, ELZ-0015945, ELZ-0015946, ELZ-0015947, ELZ-0015948, ELZ-0015949, ELZ-0015950, ELZ-0015951, ELZ-0015952, ELZ-0015953, ELZ-0015954, ELZ-0015955, ELZ-0015956, ELZ-0015957, ELZ-0015958, ELZ-0015959, ELZ-0015960, ELZ-0015961, ELZ-0015962, ELZ-0015963, ELZ-0015964, ELZ-0015965, ELZ-0015966, ELZ-0015967, ELZ-0015968, ELZ-0015969, ELZ-0015970, ELZ-0015971, ELZ-0015972, ELZ-0015973, ELZ-0015974, ELZ-0015975, ELZ-0015976, ELZ-0015978, ELZ-0015979, ELZ-0015980, ELZ-0015982, ELZ-0015983, ELZ-0015984, ELZ-0015985, ELZ-0015986, ELZ-0015987, ELZ-0015988, ELZ-0015990, ELZ-0015991, ELZ-0015992, ELZ-0015993, ELZ-0015994, ELZ-0015995, ELZ-0015996, ELZ-0015997, ELZ-0015998, ELZ-0015999, ELZ-0016000, ELZ-0016001, ELZ-0016002, ELZ-0016003, ELZ-0016004, ELZ-0016006, ELZ-0016007, ELZ-0016008, ELZ-0016009, ELZ-0016010, ELZ-0016011, ELZ-0016013, ELZ-0016017, ELZ-0016018, ELZ-0016019, ELZ-0016024, ELZ-0016025, ELZ-0016026, ELZ-0016027, ELZ-0016028, ELZ-0016029, ELZ-0016030, ELZ-0016031, ELZ-0016032, ELZ-0016033, ELZ-0016034, ELZ-0016035, ELZ-0016036, ELZ-0016037, ELZ-0016038, ELZ-0016039, ELZ-0016040, ELZ-0016041, ELZ-0016042, ELZ-0016043, ELZ-0016044, ELZ-0016045, ELZ-0016046, ELZ-0016048, ELZ-0016049, ELZ-0016051, ELZ-0016052, ELZ-0016053, ELZ-0016054, ELZ-0016055, ELZ-0016056, ELZ-0016057, ELZ-0016058, ELZ-0016059, ELZ-0016060, ELZ-0016061, ELZ-0016063, ELZ-0016064, ELZ-0016065, ELZ-0016067, ELZ-0016071, ELZ-0016074, ELZ-0016076, ELZ-0016077, ELZ-0016078, ELZ-0016081, ELZ-0016082, ELZ-0016083, ELZ-0016087, ELZ-0016089, ELZ-0016090, ELZ-0016091, ELZ-0016092, ELZ-0016093, ELZ-0016094, ELZ-0016095, ELZ-0016096, ELZ-0016097, ELZ-0016098, ELZ-0016099, ELZ-0016100, ELZ-0016101, ELZ-0016103, ELZ-0016105, ELZ-0016106, ELZ-0016107, ELZ-0016108, ELZ-0016110, ELZ-0016111, ELZ-0016112, ELZ-0016114, ELZ-0016115, ELZ-0016116, ELZ-0016117, ELZ-0016118, ELZ-0016120, ELZ-0016121, ELZ-0016122, ELZ-0016123, ELZ-0016124, ELZ-0016125, ELZ-0016126, ELZ-0016127, ELZ-0016128, ELZ-0016129, ELZ-0016131, ELZ-0016132, ELZ-0016134, ELZ-0016135, ELZ-0016136, ELZ-0016137, ELZ-0016138, ELZ-0016139, ELZ-0016140, ELZ-0016141, ELZ-0016142, ELZ-0016143, ELZ-0016144, ELZ-0016145, ELZ-0016146, ELZ-0016147, ELZ-0016148, ELZ-0016149, ELZ-0016150, ELZ-0016151, ELZ-0016152, ELZ-0016153, ELZ-0016154, ELZ-0016155, ELZ-0016156, ELZ-0016157, ELZ-0016158, ELZ-0016159, ELZ-0016160, ELZ-0016161, ELZ-0016162, ELZ-0016163, ELZ-0016164, ELZ-0016165, ELZ-0016166, ELZ-0016167, ELZ-0016168, ELZ-0016169, ELZ-0016170, ELZ-0016171, ELZ-0016172, ELZ-0016173, ELZ-0016174, ELZ-0016175, ELZ-0016176, ELZ-0016177, ELZ-0016178, ELZ-0016179, ELZ-0016180, ELZ-0016181, ELZ-0016182, ELZ-0016184, ELZ-0016185, ELZ-0016186, ELZ-0016187, ELZ-0016188, ELZ-0016189, ELZ-0016190, ELZ-0016191, ELZ-0016192, ELZ-0016193, ELZ-0016194, ELZ-0016196, ELZ-0016197, ELZ-0016200, ELZ-0016201, ELZ-0016202, ELZ-0016203, ELZ-0016204, ELZ-0016205, ELZ-0016206, ELZ-0016207, ELZ-0016209, ELZ-0016210, ELZ-0016211, ELZ-0016212, ELZ-0016213, ELZ-0016214, ELZ-0016215, ELZ-0016216, ELZ-0016217, ELZ-0016218, ELZ-0016219, ELZ-0016220, ELZ-0016221, ELZ-0016222, ELZ-0016223, ELZ-0016224, ELZ-0016225, ELZ-0016226, ELZ-0016227, ELZ-0016228, ELZ-0016229, ELZ-0016230, ELZ-0016231, ELZ-0016232, ELZ-0016233, ELZ-0016234, ELZ-0016235, ELZ-0016236, ELZ-0016237, ELZ-0016239, ELZ-0016240, ELZ-0016241, ELZ-0016242, ELZ-0016243, ELZ-0016244, ELZ-0016245, ELZ-0016246, ELZ-0016247, ELZ-0016248, ELZ-0016249, ELZ-0016250, ELZ-0016251, ELZ-0016253, ELZ-0016254, ELZ-0016255, ELZ-0016256, ELZ-0016257, ELZ-0016259, ELZ-0016260, ELZ-0016261, ELZ-0016262, ELZ-0016263, ELZ-0016264, ELZ-0016265, ELZ-0016266, ELZ-0016267, ELZ-0016269, ELZ-0016271, ELZ-0016272, ELZ-0016273, ELZ-0016275, ELZ-0016276, ELZ-0016280, ELZ-0016282, ELZ-0016283, ELZ-0016286, ELZ-0016287, ELZ-0016288, ELZ-0016290, ELZ-0016291, ELZ-0016292, ELZ-0016293, ELZ-0016294, ELZ-0016295, ELZ-0016296, ELZ-0016297, ELZ-0016298, ELZ-0016299, ELZ-0016300, ELZ-0016301, ELZ-0016302, ELZ-0016303, ELZ-0016305, ELZ-0016306, ELZ-0016307, ELZ-0016309, ELZ-0016310, ELZ-0016311, ELZ-0016312, ELZ-0016313, ELZ-0016314, ELZ-0016315, ELZ-0016316, ELZ-0016317, ELZ-0016318, ELZ-0016319, ELZ-0016320, ELZ-0016322, ELZ-0016324, ELZ-0016325, ELZ-0016326, ELZ-0016328, ELZ-0016329, ELZ-0016330, ELZ-0016332, ELZ-0016333, ELZ-0016335, ELZ-0016336, ELZ-0016337, ELZ-0016338, ELZ-0016339, ELZ-0016340, ELZ-0016341, ELZ-0016343, ELZ-0016344, ELZ-0016345, ELZ-0016346, ELZ-0016347, ELZ-0016349, ELZ-0016350, ELZ-0016351, ELZ-0016352, ELZ-0016353, ELZ-0016355, ELZ-0016356, ELZ-0016357, ELZ-0016358, ELZ-0016359, ELZ-0016360, ELZ-0016361, ELZ-0016362, ELZ-0016364, ELZ-0016365, ELZ-0016366, ELZ-0016367, ELZ-0016370, ELZ-0016372, ELZ-0016373, ELZ-0016375, ELZ-0016376, ELZ-0016377, ELZ-0016379, ELZ-0016381, ELZ-0016386, ELZ-0016387, ELZ-0016388, ELZ-0016392, ELZ-0016393, ELZ-0016398, ELZ-0016399, ELZ-0016404, ELZ-0016407, ELZ-0016408, ELZ-0016413, ELZ-0016414, ELZ-0016416, ELZ-0016418, ELZ-0016419, ELZ-0016421, ELZ-0016423, ELZ-0016424, ELZ-0016425, ELZ-0016426, ELZ-0016428, ELZ-0016429, ELZ-0016430, ELZ-0016431, ELZ-0016433, ELZ-0016434, ELZ-0016435, ELZ-0016436, ELZ-0016437, ELZ-0016438, ELZ-0016439, ELZ-0016440, ELZ-0016440, ELZ-0016441, ELZ-0016442, ELZ-0016443, ELZ-0016444, ELZ-0016446, ELZ-0016447, ELZ-0016448, ELZ-0016449, ELZ-0016450, ELZ-0016451, ELZ-0016452, ELZ-0016453, ELZ-0016454, ELZ-0016455, ELZ-0016456, ELZ-0016457, ELZ-0016458, ELZ-0016459, ELZ-0016460, ELZ-0016461, ELZ-0016463, ELZ-0016464, ELZ-0016465, ELZ-0016466, ELZ-0016467, ELZ-0016468, ELZ-0016469, ELZ-0016470, ELZ-0016471, ELZ-0016472, ELZ-0016473, ELZ-0016474, ELZ-0016475, ELZ-0016476, ELZ-0016477, ELZ-0016478, ELZ-0016479, ELZ-0016480, ELZ-0016481, ELZ-0016482, ELZ-0016483, ELZ-0016484, ELZ-0016485, ELZ-0016486, ELZ-0016487, ELZ-0016488, ELZ-0016490, ELZ-0016491, ELZ-0016492, ELZ-0016493, ELZ-0016494, ELZ-0016495, ELZ-0016496, ELZ-0016497, ELZ-0016498, ELZ-0016499, ELZ-0016500, ELZ-0016501, ELZ-0016502, ELZ-0016503, ELZ-0016505, ELZ-0016506, ELZ-0016507, ELZ-0016508, ELZ-0016509, ELZ-0016510, ELZ-0016511, ELZ-0016512, ELZ-0016514, ELZ-0016515, ELZ-0016517, ELZ-0016518, ELZ-0016519, ELZ-0016522, ELZ-0016524, ELZ-0016525, ELZ-0016526, ELZ-0016528, ELZ-0016529, ELZ-0016530, ELZ-0016532, ELZ-0016533, ELZ-0016534, ELZ-0016536, ELZ-0016537, ELZ-0016538, ELZ-0016540, ELZ-0016541, ELZ-0016543, ELZ-0016544, ELZ-0016545, ELZ-0016546, ELZ-0016547, ELZ-0016548, ELZ-0016549, ELZ-0016550, ELZ-0016551, ELZ-0016552, ELZ-0016553, ELZ-0016554, ELZ-0016555, ELZ-0016556, ELZ-0016557, ELZ-0016558, ELZ-0016559, ELZ-0016560, ELZ-0016561, ELZ-0016562, ELZ-0016564, ELZ-0016565, ELZ-0016566, ELZ-0016567, ELZ-0016568, ELZ-0016569, ELZ-0016570, ELZ-0016571, ELZ-0016572, ELZ-0016573, ELZ-0016574, ELZ-0016575, ELZ-0016577, ELZ-0016578, ELZ-0016579, ELZ-0016580, ELZ-0016581, ELZ-0016582, ELZ-0016583, ELZ-0016584, ELZ-0016588, ELZ-0016589, ELZ-0016590, ELZ-0016591, ELZ-0016598, ELZ-0016601, ELZ-0016603, ELZ-0016605, ELZ-0016607, ELZ-0016608, ELZ-0016609, ELZ-0016610, ELZ-0016611, ELZ-0016612, ELZ-0016617, ELZ-0016618, ELZ-0016620, ELZ-0016621, ELZ-0016622, ELZ-0016623, ELZ-0016624, ELZ-0016625, ELZ-0016626, ELZ-0016628, ELZ-0016629, ELZ-0016630, ELZ-0016632, ELZ-0016633, ELZ-0016634, ELZ-0016637, ELZ-0016638, ELZ-0016639, ELZ-0016640, ELZ-0016641, ELZ-0016644, ELZ-0016645, ELZ-0016646, ELZ-0016647, ELZ-0016648, ELZ-0016649, ELZ-0016650, ELZ-0016652, ELZ-0016653, ELZ-0016654, ELZ-0016657, ELZ-0016658, ELZ-0016659, ELZ-0016661, ELZ-0016662, ELZ-0016663, ELZ-0016664, ELZ-0016665, ELZ-0016666, ELZ-0016668, ELZ-0016669, ELZ-0016670, ELZ-0016671, ELZ-0016672, ELZ-0016673, ELZ-0016674, ELZ-0016675, ELZ-0016676, ELZ-0016677, ELZ-0016681, ELZ-0016683, ELZ-0016686, ELZ-0016687, ELZ-0016688, ELZ-0016689, ELZ-0016690, ELZ-0016693, ELZ-0016694, ELZ-0016697, ELZ-0016698, ELZ-0016699, ELZ-0016700, ELZ-0016701, ELZ-0016702, ELZ-0016703, ELZ-0016704, ELZ-0016705, ELZ-0016706, ELZ-0016707, ELZ-0016708, ELZ-0016713, ELZ-0016714, ELZ-0016715, ELZ-0016716, ELZ-0016717, ELZ-0016718, ELZ-0016719, ELZ-0016720, ELZ-0016721, ELZ-0016722, ELZ-0016723, ELZ-0016725, ELZ-0016727, ELZ-0016728, ELZ-0016729, ELZ-0016730, ELZ-0016731, ELZ-0016732, ELZ-0016733, ELZ-0016734, ELZ-0016735, ELZ-0016736, ELZ-0016737, ELZ-0016738, ELZ-0016739, ELZ-0016740, ELZ-0016741, ELZ-0016742, ELZ-0016743, ELZ-0016745, ELZ-0016747, ELZ-0016748, ELZ-0016749, ELZ-0016750, ELZ-0016751, ELZ-0016752, ELZ-0016753, ELZ-0016754, ELZ-0016755, ELZ-0016756, ELZ-0016758, ELZ-0016759, ELZ-0016760, ELZ-0016761, ELZ-0016762, ELZ-0016763, ELZ-0016764, ELZ-0016765, ELZ-0016766, ELZ-0016767, ELZ-0016768, ELZ-0016769, ELZ-0016770, ELZ-0016771, ELZ-0016772, ELZ-0016774, ELZ-0016775, ELZ-0016776, ELZ-0016777, ELZ-0016778, ELZ-0016779, ELZ-0016780, ELZ-0016781, ELZ-0016782, ELZ-0016783, ELZ-0016784, ELZ-0016785, ELZ-0016786, ELZ-0016788, ELZ-0016790, ELZ-0016791, ELZ-0016792, ELZ-0016793, ELZ-0016794, ELZ-0016795, ELZ-0016796, ELZ-0016798, ELZ-0016799, ELZ-0016801, ELZ-0016802, ELZ-0016803, ELZ-0016804, ELZ-0016805, ELZ-0016806, ELZ-0016807, ELZ-0016809, ELZ-0016810, ELZ-0016811, ELZ-0016812, ELZ-0016813, ELZ-0016814, ELZ-0016815, ELZ-0016816, ELZ-0016819, ELZ-0016820, ELZ-0016821, ELZ-0016822, ELZ-0016823, ELZ-0016824, ELZ-0016825, ELZ-0016826, ELZ-0016827, ELZ-0016828, ELZ-0016829, ELZ-0016830, ELZ-0016831, ELZ-0016832, ELZ-0016834, ELZ-0016835, ELZ-0016836, ELZ-0016839, ELZ-0016842, ELZ-0016843, ELZ-0016845, ELZ-0016846, ELZ-0016847, ELZ-0016848, ELZ-0016849, ELZ-0016850, ELZ-0016852, ELZ-0016853, ELZ-0016854, ELZ-0016855, ELZ-0016857, ELZ-0016858, ELZ-0016859, ELZ-0016860, ELZ-0016861, ELZ-0016862, ELZ-0016863, ELZ-0016864, ELZ-0016865, ELZ-0016866, ELZ-0016867, ELZ-0016868, ELZ-0016869, ELZ-0016870, ELZ-0016871, ELZ-0016872, ELZ-0016873, ELZ-0016875, ELZ-0016876, ELZ-0016877, ELZ-0016878, ELZ-0016879, ELZ-0016880, ELZ-0016881, ELZ-0016882, ELZ-0016883, ELZ-0016884, ELZ-0016885, ELZ-0016886, ELZ-0016888, ELZ-0016889, ELZ-0016892, ELZ-0016893, ELZ-0016894, ELZ-0016895, ELZ-0016896, ELZ-0016897, ELZ-0016898, ELZ-0016899, ELZ-0016900, ELZ-0016901, ELZ-0016902, ELZ-0016903, ELZ-0016906, ELZ-0016907, ELZ-0016909, ELZ-0016910, ELZ-0016911, ELZ-0016912, ELZ-0016914, ELZ-0016915, ELZ-0016916, ELZ-0016917, ELZ-0016919, ELZ-0016920, ELZ-0016921, ELZ-0016923, ELZ-0016924, ELZ-0016925, ELZ-0016927, ELZ-0016929, ELZ-0016930, ELZ-0016934, ELZ-0016935, ELZ-0016936, ELZ-0016937, ELZ-0016938, ELZ-0017486, ELZ-0017487, ELZ-0017488, ELZ-0017489, ELZ-0017490, ELZ-0017491, ELZ-0017492, ELZ-0017494, ELZ-0017495, ELZ-0017496, ELZ-0017498, ELZ-0017499, ELZ-0017500, ELZ-0017501, ELZ-0017502, ELZ-0017504, ELZ-0017505, ELZ-0017507, ELZ-0017509, ELZ-0017510, ELZ-0017511, ELZ-0017512, ELZ-0017513, ELZ-0017514, ELZ-0017515, ELZ-0017516, ELZ-0017517, ELZ-0017519, ELZ-0017520, ELZ-0017521, ELZ-0017522, ELZ-0017523, ELZ-0017524, ELZ-0017525, ELZ-0017526, ELZ-0017527, ELZ-0017528, ELZ-0017529, ELZ-0017530, ELZ-0017531, ELZ-0017533, ELZ-0017534, ELZ-0017535, ELZ-0017536, ELZ-0017537, ELZ-0017538, ELZ-0017539, ELZ-0017540, ELZ-0017541, ELZ-0017542, ELZ-0017544, ELZ-0017546, ELZ-0017547, ELZ-0017548, ELZ-0017549, ELZ-0017550, ELZ-0017551, ELZ-0017552, ELZ-0017553, ELZ-0017554, ELZ-0017555, ELZ-0017556, ELZ-0017557, ELZ-0017558, ELZ-0017559, ELZ-0017560, ELZ-0017561, ELZ-0017562, ELZ-0017563, ELZ-0017564, ELZ-0017565, ELZ-0017566, ELZ-0017567, ELZ-0017568, ELZ-0017569, ELZ-0017570, ELZ-0017571, ELZ-0017572, ELZ-0017573, ELZ-0017574, ELZ-0017575, ELZ-0017576, ELZ-0017578, ELZ-0017579, ELZ-0017580, ELZ-0017581, ELZ-0017582, ELZ-0017583, ELZ-0017584, ELZ-0017585, ELZ-0017586, ELZ-0017587, ELZ-0017588, ELZ-0017589, ELZ-0017590, ELZ-0017591, ELZ-0017592, ELZ-0017593, ELZ-0017594, ELZ-0017595, ELZ-0017596, ELZ-0017597, ELZ-0017598, ELZ-0017599, ELZ-0017601, ELZ-0017602, ELZ-0017603, ELZ-0017604, ELZ-0017605, ELZ-0017606, ELZ-0017607, ELZ-0017609, ELZ-0017611, ELZ-0017612, ELZ-0017613, ELZ-0017615, ELZ-0017617, ELZ-0017618, ELZ-0017619, ELZ-0017620, ELZ-0017621, ELZ-0017622, ELZ-0017623, ELZ-0017625, ELZ-0017626, ELZ-0017628, ELZ-0017629, ELZ-0017630, ELZ-0017632, ELZ-0017633, ELZ-0017634, ELZ-0017635, ELZ-0017636, ELZ-0017637, ELZ-0017638, ELZ-0017639, ELZ-0017640, ELZ-0017641, ELZ-0017642, ELZ-0017643, ELZ-0017644, ELZ-0017645, ELZ-0017646, ELZ-0017647, ELZ-0017648, ELZ-0017649, ELZ-0017650, ELZ-0017651, ELZ-0017652, ELZ-0017653, ELZ-0017654, ELZ-0017655, ELZ-0017656, ELZ-0017657, ELZ-0017658, ELZ-0017659, ELZ-0017660, ELZ-0017661, ELZ-0017662, ELZ-0017663, ELZ-0017664, ELZ-0017665, ELZ-0017666, ELZ-0017667, ELZ-0017668, ELZ-0017669, ELZ-0017670, ELZ-0017671, ELZ-0017672, ELZ-0017673, ELZ-0017674, ELZ-0017676, ELZ-0017677, ELZ-0017678, ELZ-0017679, ELZ-0017680, ELZ-0017681, ELZ-0017682, ELZ-0017683, ELZ-0017684, ELZ-0017686, ELZ-0017687, ELZ-0017689, ELZ-0017690, ELZ-0017691, ELZ-0017692, ELZ-0017695, ELZ-0017696, ELZ-0017699, ELZ-0017700, ELZ-0017702, ELZ-0017703, ELZ-0017704, ELZ-0017705, ELZ-0017706, ELZ-0017707, ELZ-0017708, ELZ-0017709, ELZ-0017720, ELZ-0017723, ELZ-0017725, ELZ-0017726, ELZ-0017729, ELZ-0017730, ELZ-0017733, ELZ-0017735, ELZ-0017736, ELZ-0017737, ELZ-0017739, ELZ-0017740, ELZ-0017742, ELZ-0017743, ELZ-0017744, ELZ-0017747, ELZ-0017748, ELZ-0017750, ELZ-0017751, ELZ-0017752, ELZ-0017753, ELZ-0017754, ELZ-0017755, ELZ-0017756, ELZ-0017758, ELZ-0017759, ELZ-0017760, ELZ-0017763, ELZ-0017767, ELZ-0017768, ELZ-0017769, ELZ-0017770, ELZ-0017771, ELZ-0017772, ELZ-0017773, ELZ-0017774, ELZ-0017777, ELZ-0017778, ELZ-0017779, ELZ-0017780, ELZ-0017781, ELZ-0017782, ELZ-0017783, ELZ-0017784, ELZ-0017785, ELZ-0017786, ELZ-0017787, ELZ-0017788, ELZ-0017789, ELZ-0017790, ELZ-0017792, ELZ-0017793, ELZ-0017794, ELZ-0017795, ELZ-0017796, ELZ-0017797, ELZ-0017798, ELZ-0017799, ELZ-0017800, ELZ-0017801, ELZ-0017802, ELZ-0017803, ELZ-0017804, ELZ-0017805, ELZ-0017807, ELZ-0017808, ELZ-0017809, ELZ-0017810, ELZ-0017811, ELZ-0017813, ELZ-0017814, ELZ-0017815, ELZ-0017816, ELZ-0017817, ELZ-0017819, ELZ-0017820, ELZ-0017821, ELZ-0017823, ELZ-0017824, ELZ-0017825, ELZ-0017826, ELZ-0017827, ELZ-0017828, ELZ-0017829, ELZ-0017830, ELZ-0017831, ELZ-0017832, ELZ-0017833, ELZ-0017834, ELZ-0017835, ELZ-0017836, ELZ-0017837, ELZ-0017838, ELZ-0017839, ELZ-0017840, ELZ-0017841, ELZ-0017842, ELZ-0017843, ELZ-0017845, ELZ-0017846, ELZ-0017847, ELZ-0017848, ELZ-0017849, ELZ-0017850, ELZ-0017851, ELZ-0017852, ELZ-0017853, ELZ-0017854, ELZ-0017855, ELZ-0017856, ELZ-0017857, ELZ-0017859, ELZ-0017860, ELZ-0017861, ELZ-0017862, ELZ-0017863, ELZ-0017864, ELZ-0017865, ELZ-0017866, ELZ-0017867, ELZ-0017868, ELZ-0017869, ELZ-0017870, ELZ-0017871, ELZ-0017872, ELZ-0017875, ELZ-0017876, ELZ-0017877, ELZ-0017878, ELZ-0017879, ELZ-0017880, ELZ-0017881, ELZ-0017882, ELZ-0017885, ELZ-0017886, ELZ-0017887, ELZ-0017890, ELZ-0017891, ELZ-0017893, ELZ-0017895, ELZ-0017896, ELZ-0017897, ELZ-0017899, ELZ-0017901, ELZ-0017910, ELZ-0017912, ELZ-0017914, ELZ-0017916, ELZ-0017917, ELZ-0017918, ELZ-0017920, ELZ-0017922, ELZ-0017924, ELZ-0017925, ELZ-0017927, ELZ-0017928, ELZ-0017929, ELZ-0017930, ELZ-0017931, ELZ-0017935, ELZ-0017938, ELZ-0017939, ELZ-0017942, ELZ-0017945, ELZ-0017948, ELZ-0017949, ELZ-0017950, ELZ-0017951, ELZ-0017953, ELZ-0017956, ELZ-0017957, ELZ-0017959, ELZ-0017960, ELZ-0017962, ELZ-0017968, ELZ-0017969, ELZ-0017971, ELZ-0017973, ELZ-0017974, ELZ-0017975, ELZ-0017976, ELZ-0017978, ELZ-0017979, ELZ-0017982, ELZ-0017984, ELZ-0017985, ELZ-0017986, ELZ-0017988, ELZ-0017989, ELZ-0017991, ELZ-0017992, ELZ-0017993, ELZ-0017994, ELZ-0017995, ELZ-0017996, ELZ-0017997, ELZ-0018000, ELZ-0018002, ELZ-0018003, ELZ-0018005, ELZ-0018006, ELZ-0018007, ELZ-0018009, ELZ-0018010, ELZ-0018011, ELZ-0018015, ELZ-0018016, ELZ-0018017, ELZ-0018019, ELZ-0018020, ELZ-0018022, ELZ-0018023, ELZ-0018025, ELZ-0018026, ELZ-0018027, ELZ-0018030, ELZ-0018031, ELZ-0018033, ELZ-0018036, ELZ-0018037, ELZ-0018038, ELZ-0018040, ELZ-0018042, ELZ-0018043, ELZ-0018045, ELZ-0018046, ELZ-0018048, ELZ-0018052, ELZ-0018054, ELZ-0018055, ELZ-0018056, ELZ-0018058, ELZ-0018059, ELZ-0018060, ELZ-0018062, ELZ-0018065, ELZ-0018067, ELZ-0018068, ELZ-0018069, ELZ-0018070, ELZ-0018071, ELZ-0018072, ELZ-0018073, ELZ-0018075, ELZ-0018078, ELZ-0018079, ELZ-0018081, ELZ-0018082, ELZ-0018084, ELZ-0018087, ELZ-0018090, ELZ-0018091, ELZ-0018092, ELZ-0018093, ELZ-0018096, ELZ-0018098, ELZ-0018101, ELZ-0018102, ELZ-0018104, ELZ-0018105, ELZ-0018106, ELZ-0018109, ELZ-0018110, ELZ-0018112, ELZ-0018113, ELZ-0018114, ELZ-0018115, ELZ-0018116, ELZ-0018117, ELZ-0018119, ELZ-0018121, ELZ-0018122, ELZ-0018123, ELZ-0018125, ELZ-0018126, ELZ-0018127, ELZ-0018128, ELZ-0018131, ELZ-0018133, ELZ-0018134, ELZ-0018135, ELZ-77135608, ELZ-77135609, ELZ-77135610, ELZ-77135611, ELZ-77135612, ELZ-77135613, ELZ-77135614, ELZ-77135615, ELZ-77135616, ELZ-77135617, ELZ-77135618, ELZ-77135619, ELZ-77135620, ELZ-77135621, ELZ-77135622, ELZ-77135623, ELZ-77135624, ELZ-77135625, ELZ-77135626, ELZ-77135627, ELZ-77135628, ELZ-77135629, ELZ-77135630, ELZ-77135631, ELZ-77135632, ELZ-77135633, ELZ-77135634, ELZ-77135635, ELZ-77135636, ELZ-77135637, ELZ-77135638, ELZ-77135639, ELZ-77135640, ELZ-77135641, ELZ-77135642, ELZ-77135643, ELZ-77135644, ELZ-77135645, ELZ-77135646, ELZ-77135647, ELZ-77135648, ELZ-77135649, ELZ-77135650, ELZ-77135652, ELZ-77135653, ELZ-77135656, ELZ-77135658, ELZ-77135659, ELZ-77135661, ELZ-77135662, ELZ-77135663, ELZ-77135664, ELZ-77135665, ELZ-77135667, ELZ-77135669, ELZ-77135671, ELZ-77135673, ELZ-77135674, ELZ-77135675, ELZ-77135676, ELZ-77135677, ELZ-77135678, ELZ-77135680, ELZ-77135682, ELZ-77135683, ELZ-77135685, ELZ-77135692, ELZ-77135696, ELZ-77135700, ELZ-77135706, ELZ-77135707, ELZ-77135708, ELZ-77135709, ELZ-77135710, ELZ-77135711, ELZ-77135712, ELZ-77135713, ELZ-77135714, ELZ-77135715, ELZ-77135716, ELZ-77135717, ELZ-77135718, ELZ-77135719, ELZ-77135720, ELZ-77135721, ELZ-77135722, ELZ-77135723, ELZ-77135724, ELZ-77135725, ELZ-77135726, ELZ-77135727, ELZ-77135736, ELZ-77135750, ELZ-77135751, ELZ-77135777, ELZ-77135778, ELZ-77135779, ELZ-77135780, ELZ-77135781, ELZ-77135782, ELZ-77135783, ELZ-77135784, ELZ-77135785, ELZ-77135786, ELZ-77135787, ELZ-77135788, ELZ-77135789, ELZ-77135790, ELZ-77135791, ELZ-77135792, ELZ-77135793, ELZ-77135794, ELZ-77135795, ELZ-77135796, ELZ-77135797, ELZ-77135798, ELZ-77135799, ELZ-77135800, ELZ-77135801, ELZ-77135802, ELZ-77135803, ELZ-77135804, ELZ-77135805, ELZ-77135806, ELZ-77135807, ELZ-77135808, ELZ-77135809, ELZ-77135810, ELZ-77135811, ELZ-77135812, ELZ-77135813, ELZ-77135814, ELZ-77135815, ELZ-77135816, ELZ-77135817, ELZ-77135818, ELZ-77135819, ELZ-77135820, ELZ-77135821, ELZ-77135822, ELZ-77135823, ELZ-77135825, ELZ-77135826, ELZ-77135827, ELZ-77135828, ELZ-77135829, ELZ-77135830, ELZ-77135837, ELZ-77135838, ELZ-77135839, ELZ-77135840, ELZ-77135841, ELZ-77135842, ELZ-77135847, ELZ-77135848, ELZ-77135850, ELZ-77135851, ELZ-77135852, ELZ-77135853, ELZ-77135854, ELZ-77135855, ELZ-77135856, ELZ-77135857, ELZ-77135858, ELZ-77135859, ELZ-77135860, ELZ-77135861, ELZ-77135862, ELZ-77135863, ELZ-77135864, ELZ-77135865, ELZ-77135866, ELZ-77135867, ELZ-77135868, ELZ-77135869, ELZ-77135870, ELZ-77135872, ELZ-77135873, ELZ-77135874, ELZ-77135875, ELZ-77135876, ELZ-77135877, ELZ-77135881, ELZ-77135882, ELZ-77135883, ELZ-77135884, ELZ-77135885, ELZ-77135886, ELZ-77135887, ELZ-77135888, ELZ-77135889, ELZ-77135890, ELZ-77135892, ELZ-77135893, ELZ-77135894, ELZ-77135895, ELZ-77135896, ELZ-77135897, ELZ-77135898, ELZ-77135899, ELZ-77135900, ELZ-77135901, ELZ-77135902, ELZ-77135905, ELZ-77135906, ELZ-77135907, ELZ-77135908, ELZ-77135909, ELZ-77135910, ELZ-77135911, ELZ-77135912, ELZ-77135914, ELZ-77135915, ELZ-77135916, ELZ-77135917, ELZ-77135918, ELZ-77135919, ELZ-77135920, ELZ-77135921, ELZ-77135922, ELZ-77135924, ELZ-77135925, ELZ-77135926, ELZ-77135938, ELZ-77135954, ELZ-77135955, ELZ-77135956, ELZ-77135957, ELZ-77135962, ELZ-77135969, ELZ-77135970, ELZ-77135972, ELZ-77135973, ELZ-77135976, ELZ-77135977, ELZ-77135981, ELZ-77135982, ELZ-77135985, ELZ-77135987, ELZ-77135990, ELZ-77135990, ELZ-77135992, ELZ-77135996, ELZ-77166572, ELZ-77166573, ELZ-77166574, ELZ-77166575, ELZ-77166576, ELZ-77166577, ELZ-77166578, ELZ-77166579, ELZ-77166580, ELZ-77166582, ELZ-77166584, ELZ-77166585, ELZ-77385615, ELZ-77385616, ELZ-77385617, ELZ-77385618, ELZ-77385619, ELZ-77385620, ELZ-77385621, ELZ-77385622, ELZ-77385623, ELZ-77385624, ELZ-77385625, ELZ-77385626, ELZ-77385627, ELZ-77385628, ELZ-77385629, ELZ-77385630, ELZ-77385631, ELZ-77385632, ELZ-77385633, ELZ-77385634, ELZ-77385635, ELZ-77385636, ELZ-77385637, ELZ-77385638, ELZ-77385639, ELZ-77385640, ELZ-77385641, ELZ-77385642, ELZ-77385643, ELZ-77385644, ELZ-77385645, ELZ-77385646, ELZ-77385647, ELZ-77385648, ELZ-77385649, ELZ-77385650, ELZ-77385651, ELZ-77385652, ELZ-77385653, ELZ-77385654, ELZ-77385655, ELZ-77385656, ELZ-77385657, ELZ-77385658, ELZ-77385659, ELZ-77385660, ELZ-77385661, ELZ-77385662, ELZ-77385663, ELZ-77385664, ELZ-77385665, ELZ-77385666, ELZ-77385667, ELZ-77385668, ELZ-77385669, ELZ-77385670, ELZ-77385671, ELZ-77385672, ELZ-77385673, ELZ-77385674, ELZ-77385675, ELZ-77385676, ELZ-77385677, ELZ-77385678, ELZ-77385679, ELZ-77385680, ELZ-77385681, ELZ-77385682, ELZ-77385683, ELZ-77385684, ELZ-77385685, ELZ-77385686, ELZ-77385687, ELZ-77385688, ELZ-77385689, ELZ-77385690, ELZ-77385691, ELZ-77385692, ELZ-77385693, ELZ-77385694, ELZ-77385695, ELZ-77385696, ELZ-77385697, ELZ-77385698, ELZ-77385699, ELZ-77385700, ELZ-77385701, ELZ-77385702, ELZ-77385703, ELZ-77385704, ELZ-77385705, ELZ-77385707, ELZ-77385708, ELZ-77385709, ELZ-77385710, ELZ-77385711, ELZ-77385712, ELZ-77385713, ELZ-77385714, ELZ-77385715, ELZ-77385718, ELZ-77385719, ELZ-77385720, ELZ-77385721, ELZ-77385722, ELZ-77385723, ELZ-77385725, ELZ-77385726, ELZ-77385727, ELZ-77385728, ELZ-77385729, ELZ-77385730, ELZ-77385731, ELZ-77385732, ELZ-77385733, ELZ-77385734, ELZ-77385735, ELZ-77385736, ELZ-77385737, ELZ-77385738, ELZ-77385739, ELZ-77385740, ELZ-77385741, ELZ-77385742, ELZ-77385743, ELZ-77385744, ELZ-77385745, ELZ-77385746, ELZ-77385747, ELZ-77385748, ELZ-77385749, ELZ-77385750, ELZ-77385751, ELZ-77385752, ELZ-77385753, ELZ-77385755, ELZ-77385756, ELZ-77385759, ELZ-77385760, ELZ-77385761, ELZ-77385762, ELZ-77385763, ELZ-77385764, ELZ-77385765, ELZ-77385766, ELZ-77385767, ELZ-77385768, ELZ-77385769, ELZ-77385770, ELZ-77385771, ELZ-77385772, ELZ-77385773, ELZ-77385774, ELZ-77385776, ELZ-77385777, ELZ-77385778, ELZ-77385779, ELZ-77385781, ELZ-77385783, ELZ-77385784, ELZ-77385785, ELZ-77385786, ELZ-77385787, ELZ-77385788, ELZ-77385789, ELZ-77385790, ELZ-77385791, ELZ-77385792, ELZ-77385793, ELZ-77385797, ELZ-77385820, ELZ-77385821, ELZ-77385822, ELZ-77385823, ELZ-77385824, ELZ-77385825, ELZ-77385826, ELZ-77385828, ELZ-77385829, ELZ-77385830, ELZ-77385831, ELZ-77385832, ELZ-77385833, ELZ-77385834, ELZ-77385835, ELZ-77385836, ELZ-77385837, ELZ-77385838, ELZ-77385839, ELZ-77385840, ELZ-77385843, ELZ-77385845, ELZ-77385846, ELZ-77385847, ELZ-77385851, ELZ-77385852, ELZ-77385853, ELZ-77385854, ELZ-77385857, ELZ-77385858, ELZ-77385860, ELZ-77385862, ELZ-77385864, ELZ-77385869, ELZ-77385871, ELZ-77385874, ELZ-77385875, ELZ-77385876, ELZ-77385879, ELZ-77385880, ELZ-77385881, ELZ-77385882, ELZ-77385885, ELZ-77385886, ELZ-77385887, ELZ-77385888, ELZ-77385889, ELZ-77385890, ELZ-77385893, ELZ-77385894, ELZ-77385895, ELZ-77385896, ELZ-77385897, ELZ-77385898, ELZ-77385899, ELZ-77385900, ELZ-77385902, ELZ-77385903, ELZ-77385904, ELZ-77385907, ELZ-77385908, ELZ-77385909, ELZ-77385910, ELZ-77385911, ELZ-77385912, ELZ-77385913, ELZ-77385914, ELZ-77385915, ELZ-77385916, ELZ-77385917, ELZ-77385918, ELZ-77385919, ELZ-77385920, ELZ-77385921, ELZ-77385922, ELZ-77385923, ELZ-77385924, ELZ-77385925, ELZ-77385926, ELZ-77385927, ELZ-77385928, ELZ-77385929, ELZ-77385930, ELZ-77385931, ELZ-77385932, ELZ-77385933, ELZ-77385934, ELZ-77385935, ELZ-77385936, ELZ-77385937, ELZ-77385938, ELZ-77385939, ELZ-77385940, ELZ-77385941, ELZ-77385943, ELZ-77385944, ELZ-77385946, ELZ-77385947, ELZ-77385948, ELZ-77385949, ELZ-77385950, ELZ-77385951, ELZ-77385952, ELZ-77385955, ELZ-77385956, ELZ-77385957, ELZ-77385959, ELZ-77385960, ELZ-77385961, ELZ-77385962, ELZ-77385963, ELZ-77385965, ELZ-77385966, ELZ-77385967, ELZ-77385968, ELZ-77385969, ELZ-77385971, ELZ-77385972, ELZ-77385973, ELZ-77385974, ELZ-77385975, ELZ-77385976, ELZ-77385978, ELZ-77385979, ELZ-77385980, ELZ-77385981, ELZ-77385982, ELZ-77385983, ELZ-77385984, ELZ-77385985, ELZ-77385986, ELZ-77385987, ELZ-77385988, ELZ-77385989, ELZ-77385990, ELZ-77385991, ELZ-77385992, ELZ-77385993, ELZ-77385994, ELZ-77385995, ELZ-77385996, ELZ-77385997, ELZ-77385998, ELZ-77385999, ELZ-77386000, ELZ-77386001, ELZ-77386002, ELZ-77386003, ELZ-77386004, ELZ-77386005, ELZ-77386006, ELZ-77386007, ELZ-77386008, ELZ-77386009, ELZ-77386010, ELZ-77386011, ELZ-77386012, ELZ-77386013, ELZ-77386014, ELZ-77386015, ELZ-77386016, ELZ-77386017, ELZ-77386018, ELZ-77386019, ELZ-77386020, ELZ-77386021, ELZ-77386022, ELZ-77386023, ELZ-77386024, ELZ-77386025, ELZ-77386026, ELZ-77386027, ELZ-77386028, ELZ-77386029, ELZ-77386030, ELZ-77386031, ELZ-77386032, ELZ-77386033, ELZ-77386034, and ELZ-77386035.  Lot/Serial Numbers on the systems for Foreign Consignees:  ELZ-0014568, ELZ-0014569, ELZ-0014571, ELZ-0014572, ELZ-0014573, ELZ-0014575, ELZ-0014576, ELZ-0014578, ELZ-0014579, ELZ-0014582, ELZ-0014584, ELZ-0014588, ELZ-0014589, ELZ-0014602, ELZ-0014624, ELZ-0014627, ELZ-0014628, ELZ-0014643, ELZ-0014644, ELZ-0014646, ELZ-0014649, ELZ-0014650, ELZ-0014651, ELZ-0014652, ELZ-0014655, ELZ-0014659, ELZ-0014660, ELZ-0014661, ELZ-0014721, ELZ-0014722, ELZ-0014725, ELZ-0014728, ELZ-0014730, ELZ-0014732, ELZ-0014734, ELZ-0014739, ELZ-0014741, ELZ-0014743, ELZ-0014746, ELZ-0014747, ELZ-0014748, ELZ-0014749, ELZ-0014751, ELZ-0014753, ELZ-0014757, ELZ-0014761, ELZ-0014763, ELZ-0014769, ELZ-0014777, ELZ-0014778, ELZ-0014779, ELZ-0014784, ELZ-0014789, ELZ-0014792, ELZ-0014794, ELZ-0014796, ELZ-0014797, ELZ-0014798, ELZ-0014799, ELZ-0014801, ELZ-0014802, ELZ-0014803, ELZ-0014804, ELZ-0014807, ELZ-0014808, ELZ-0014809, ELZ-0014811, ELZ-0014812, ELZ-0014813, ELZ-0014814, ELZ-0014815, ELZ-0014818, ELZ-0014821, ELZ-0014822, ELZ-0014825, ELZ-0014826, ELZ-0014829, ELZ-0014830, ELZ-0014831, ELZ-0014832, ELZ-0014835, ELZ-0014838, ELZ-0014840, ELZ-0014841, ELZ-0014843, ELZ-0014844, ELZ-0014845, ELZ-0014847, ELZ-0014848, ELZ-0014849, ELZ-0014851, ELZ-0014856, ELZ-0014862, ELZ-0014870, ELZ-0014884, ELZ-0014889, ELZ-0014897, ELZ-0014898, ELZ-0014899, ELZ-0014901, ELZ-0014903, ELZ-0014904, ELZ-0014907, ELZ-0014908, ELZ-0014909, ELZ-0014911, ELZ-0014912, ELZ-0014918, ELZ-0014919, ELZ-0014921, ELZ-0014923, ELZ-0014928, ELZ-0014939, ELZ-0014946, ELZ-0014948, ELZ-0014952, ELZ-0014962, ELZ-0014963, ELZ-0014966, ELZ-0014967, ELZ-0014970, ELZ-0014971, ELZ-0014978, ELZ-0014980, ELZ-0014984, ELZ-0014988, ELZ-0014990, ELZ-0014991, ELZ-0014993, ELZ-0014997, ELZ-0014998, ELZ-0015003, ELZ-0015004, ELZ-0015041, ELZ-0015072, ELZ-0015079, ELZ-0015086, ELZ-0015092, ELZ-0015106, ELZ-0015108, ELZ-0015109, ELZ-0015110, ELZ-0015111, ELZ-0015112, ELZ-0015115, ELZ-0015116, ELZ-0015117, ELZ-0015124, ELZ-0015137, ELZ-0015141, ELZ-0015142, ELZ-0015143, ELZ-0015147, ELZ-0015148, ELZ-0015149, ELZ-0015151, ELZ-0015153, ELZ-0015156, ELZ-0015160, ELZ-0015162, ELZ-0015171, ELZ-0015172, ELZ-0015175, ELZ-0015176, ELZ-0015177, ELZ-0015178, ELZ-0015180, ELZ-0015182, ELZ-0015183, ELZ-0015185, ELZ-0015186, ELZ-0015190, ELZ-0015191, ELZ-0015192, ELZ-0015193, ELZ-0015195, ELZ-0015196, ELZ-0015198, ELZ-0015202, ELZ-0015211, ELZ-0015254, ELZ-0015258, ELZ-0015259, ELZ-0015260, ELZ-0015261, ELZ-0015263, ELZ-0015269, ELZ-0015270, ELZ-0015272, ELZ-0015275, ELZ-0015278, ELZ-0015283, ELZ-0015287, ELZ-0015289, ELZ-0015290, ELZ-0015313, ELZ-0015314, ELZ-0015324, ELZ-0015325, ELZ-0015329, ELZ-0015356, ELZ-0015417, ELZ-0015421, ELZ-0015423, ELZ-0015424, ELZ-0015431, ELZ-0015433, ELZ-0015440, ELZ-0015477, ELZ-0015478, ELZ-0015481, ELZ-0015483, ELZ-0015484, ELZ-0015487, ELZ-0015491, ELZ-0015492, ELZ-0015493, ELZ-0015494, ELZ-0015497, ELZ-0015498, ELZ-0015499, ELZ-0015500, ELZ-0015502, ELZ-0015504, ELZ-0015505, ELZ-0015507, ELZ-0015508, ELZ-0015512, ELZ-0015513, ELZ-0015514, ELZ-0015516, ELZ-0015517, ELZ-0015518, ELZ-0015519, ELZ-0015520, ELZ-0015521, ELZ-0015522, ELZ-0015524, ELZ-0015527, ELZ-0015528, ELZ-0015530, ELZ-0015531, ELZ-0015532, ELZ-0015533, ELZ-0015534, ELZ-0015535, ELZ-0015548, ELZ-0015557, ELZ-0015563, ELZ-0015565, ELZ-0015569, ELZ-0015595, ELZ-0015596, ELZ-0015607, ELZ-0015610, ELZ-0015623, ELZ-0015636, ELZ-0015637, ELZ-0015638, ELZ-0015639, ELZ-0015640, ELZ-0015641, ELZ-0015642, ELZ-0015643, ELZ-0015644, ELZ-0015645, ELZ-0015652, ELZ-0015653, ELZ-0015695, ELZ-0015721, ELZ-0015729, ELZ-0015730, ELZ-0015733, ELZ-0015734, ELZ-0015735, ELZ-0015757, ELZ-0015761, ELZ-0015766, ELZ-0015767, ELZ-0015768, ELZ-0015769, ELZ-0015774, ELZ-0015779, ELZ-0015780, ELZ-0015796, ELZ-0015797, ELZ-0015894, ELZ-0015942, ELZ-0015977, ELZ-0015981, ELZ-0015989, ELZ-0016012, ELZ-0016014, ELZ-0016015, ELZ-0016016, ELZ-0016020, ELZ-0016021, ELZ-0016022, ELZ-0016023, ELZ-0016047, ELZ-0016050, ELZ-0016066, ELZ-0016068, ELZ-0016069, ELZ-0016070, ELZ-0016072, ELZ-0016073, ELZ-0016075, ELZ-0016079, ELZ-0016080, ELZ-0016084, ELZ-0016085, ELZ-0016086, ELZ-0016088, ELZ-0016102, ELZ-0016104, ELZ-0016109, ELZ-0016113, ELZ-0016119, ELZ-0016130, ELZ-0016133, ELZ-0016195, ELZ-0016199, ELZ-0016208, ELZ-0016238, ELZ-0016252, ELZ-0016258, ELZ-0016268, ELZ-0016270, ELZ-0016274, ELZ-0016277, ELZ-0016278, ELZ-0016279, ELZ-0016281, ELZ-0016284, ELZ-0016285, ELZ-0016304, ELZ-0016308, ELZ-0016321, ELZ-0016323, ELZ-0016327, ELZ-0016331, ELZ-0016334, ELZ-0016354, ELZ-0016363, ELZ-0016368, ELZ-0016369, ELZ-0016371, ELZ-0016374, ELZ-0016378, ELZ-0016380, ELZ-0016382, ELZ-0016383, ELZ-0016385, ELZ-0016389, ELZ-0016390, ELZ-0016391, ELZ-0016394, ELZ-0016395, ELZ-0016396, ELZ-0016397, ELZ-0016400, ELZ-0016401, ELZ-0016403, ELZ-0016405, ELZ-0016406, ELZ-0016409, ELZ-0016410, ELZ-0016411, ELZ-0016412, ELZ-0016415, ELZ-0016417, ELZ-0016420, ELZ-0016422, ELZ-0016427, ELZ-0016432, ELZ-0016462, ELZ-0016489, ELZ-0016504, ELZ-0016513, ELZ-0016516, ELZ-0016520, ELZ-0016521, ELZ-0016523, ELZ-0016527, ELZ-0016531, ELZ-0016535, ELZ-0016539, ELZ-0016542, ELZ-0016563, ELZ-0016576, ELZ-0016585, ELZ-0016586, ELZ-0016587, ELZ-0016592, ELZ-0016593, ELZ-0016594, ELZ-0016595, ELZ-0016596, ELZ-0016597, ELZ-0016599, ELZ-0016600, ELZ-0016602, ELZ-0016604, ELZ-0016606, ELZ-0016613, ELZ-0016614, ELZ-0016615, ELZ-0016616, ELZ-0016619, ELZ-0016627, ELZ-0016635, ELZ-0016636, ELZ-0016642, ELZ-0016643, ELZ-0016651, ELZ-0016655, ELZ-0016656, ELZ-0016667, ELZ-0016678, ELZ-0016679, ELZ-0016680, ELZ-0016691, ELZ-0016695, ELZ-0016696, ELZ-0016712, ELZ-0016744, ELZ-0016746, ELZ-0016757, ELZ-0016787, ELZ-0016800, ELZ-0016808, ELZ-0016818, ELZ-0016833, ELZ-0016837, ELZ-0016840, ELZ-0016841, ELZ-0016856, ELZ-0016874, ELZ-0016887, ELZ-0016890, ELZ-0016904, ELZ-0016908, ELZ-0016913, ELZ-0016918, ELZ-0016922, ELZ-0016926, ELZ-0016928, ELZ-0017493, ELZ-0017497, ELZ-0017503, ELZ-0017506, ELZ-0017508, ELZ-0017518, ELZ-0017532, ELZ-0017543, ELZ-0017600, ELZ-0017608, ELZ-0017610, ELZ-0017614, ELZ-0017616, ELZ-0017624, ELZ-0017627, ELZ-0017631, ELZ-0017675, ELZ-0017685, ELZ-0017688, ELZ-0017693, ELZ-0017694, ELZ-0017697, ELZ-0017698, ELZ-0017701, ELZ-0017710, ELZ-0017711, ELZ-0017712, ELZ-0017713, ELZ-0017714, ELZ-0017715, ELZ-0017716, ELZ-0017717, ELZ-0017718, ELZ-0017719, ELZ-0017721, ELZ-0017722, ELZ-0017727, ELZ-0017728, ELZ-0017731, ELZ-0017732, ELZ-0017734, ELZ-0017741, ELZ-0017745, ELZ-0017746, ELZ-0017749, ELZ-0017757, ELZ-0017761, ELZ-0017762, ELZ-0017765, ELZ-0017766, ELZ-0017775, ELZ-0017776, ELZ-0017806, ELZ-0017812, ELZ-0017818, ELZ-0017822, ELZ-0017844, ELZ-0017858, ELZ-0017873, ELZ-0017874, ELZ-0017883, ELZ-0017884, ELZ-0017888, ELZ-0017889, ELZ-0017892, ELZ-0017894, ELZ-0017900, ELZ-0017902, ELZ-0017903, ELZ-0017904, ELZ-0017905, ELZ-0017906, ELZ-0017907, ELZ-0017908, ELZ-0017909, ELZ-0017911, ELZ-0017913, ELZ-0017915, ELZ-0017919, ELZ-0017923, ELZ-0017932, ELZ-0017933, ELZ-0017934, ELZ-0017937, ELZ-0017940, ELZ-0017941, ELZ-0017943, ELZ-0017946, ELZ-0017947, ELZ-0017952, ELZ-0017954, ELZ-0017955, ELZ-0017958, ELZ-0017961, ELZ-0017963, ELZ-0017965, ELZ-0017966, ELZ-0017967, ELZ-0017970, ELZ-0017972, ELZ-0017977, ELZ-0017980, ELZ-0017981, ELZ-0017983, ELZ-0017987, ELZ-0017998, ELZ-0017999, ELZ-0018001, ELZ-0018008, ELZ-0018012, ELZ-0018013, ELZ-0018014, ELZ-0018018, ELZ-0018024, ELZ-0018028, ELZ-0018029, ELZ-0018032, ELZ-0018034, ELZ-0018035, ELZ-0018039, ELZ-0018041, ELZ-0018047, ELZ-0018049, ELZ-0018050, ELZ-0018053, ELZ-0018061, ELZ-0018063, ELZ-0018064, ELZ-0018074, ELZ-0018076, ELZ-0018077, ELZ-0018080, ELZ-0018083, ELZ-0018086, ELZ-0018088, ELZ-0018089, ELZ-0018094, ELZ-0018097, ELZ-0018100, ELZ-0018103, ELZ-0018107, ELZ-0018108, ELZ-0018111, ELZ-0018118, ELZ-0018129, ELZ-77135651, ELZ-77135654, ELZ-77135655, ELZ-77135657, ELZ-77135660, ELZ-77135666, ELZ-77135668, ELZ-77135670, ELZ-77135672, ELZ-77135679, ELZ-77135681, ELZ-77135684, ELZ-77135686, ELZ-77135687, ELZ-77135688, ELZ-77135689, ELZ-77135690, ELZ-77135691, ELZ-77135693, ELZ-77135694, ELZ-77135695, ELZ-77135697, ELZ-77135698, ELZ-77135699, ELZ-77135701, ELZ-77135702, ELZ-77135703, ELZ-77135704, ELZ-77135705, ELZ-77135728, ELZ-77135729, ELZ-77135730, ELZ-77135731, ELZ-77135732, ELZ-77135733, ELZ-77135734, ELZ-77135735, ELZ-77135737, ELZ-77135738, ELZ-77135739, ELZ-77135740, ELZ-77135741, ELZ-77135742, ELZ-77135743, ELZ-77135744, ELZ-77135745, ELZ-77135746, ELZ-77135747, ELZ-77135748, ELZ-77135749, ELZ-77135752, ELZ-77135824, ELZ-77135831, ELZ-77135832, ELZ-77135833, ELZ-77135834, ELZ-77135835, ELZ-77135836, ELZ-77135843, ELZ-77135844, ELZ-77135845, ELZ-77135846, ELZ-77135849, ELZ-77135878, ELZ-77135879, ELZ-77135880, ELZ-77135891, ELZ-77135903, ELZ-77135904, ELZ-77135923, ELZ-77135927, ELZ-77135928, ELZ-77135968, ELZ-77135971, ELZ-77135974, ELZ-77135975, ELZ-77135979, ELZ-77135980, ELZ-77135984, ELZ-77135986, ELZ-77135991, ELZ-77135993, ELZ-77166581, ELZ-77166583, ELZ-77166586, ELZ-77166587, ELZ-77385706, ELZ-77385716, ELZ-77385754, ELZ-77385757, ELZ-77385758, ELZ-77385780, ELZ-77385794, ELZ-77385795, ELZ-77385796, ELZ-77385798, ELZ-77385799, ELZ-77385800, ELZ-77385801, ELZ-77385802, ELZ-77385803, ELZ-77385804, ELZ-77385805, ELZ-77385806, ELZ-77385807, ELZ-77385808, ELZ-77385809, ELZ-77385810, ELZ-77385811, ELZ-77385812, ELZ-77385813, ELZ-77385814, ELZ-77385815, ELZ-77385816, ELZ-77385817, ELZ-77385818, ELZ-77385841, ELZ-77385842, ELZ-77385844, ELZ-77385848, ELZ-77385849, ELZ-77385865, ELZ-77385866, ELZ-77385867, ELZ-77385868, ELZ-77385870, ELZ-77385872, ELZ-77385873, ELZ-77385883, ELZ-77385884, ELZ-77385891, ELZ-77385942, ELZ-77385945, ELZ-77385953, ELZ-77385954, ELZ-77385958, ELZ-77385964, and ELZ-77385970.
Recalling Firm/
Manufacturer
Spacelabs Healthcare, Llc
5150 220th Ave Se
Issaquah WA 98029-6834
For Additional Information ContactTechnical Support
800-522-7025 Ext. 2
Manufacturer Reason
for Recall
The power supply of the Spacelabs Medical Flat Panel Display may fail, causing the display to go blank.
FDA Determined
Cause 2
Other
ActionSpacelabs Healthcare sent an URGENT: MEDICAL DEVICE RECALL Spacelabs Medical Flat Panel Power Supply PNs: 010-1619-00 & 010-1655-00 letter dated March 20, 2009 to US consignees. On March 27, 2009, Spacelabs sent the same recall letter with translation option to their international consignees.The letter stated the firm would be sending a replacement power supply. While waiting for the replacement of the power supply, Spacelabs Medical recommended their consignees brief the staff on the possibility of the display going blank to help prevent any adverse event from happening. The consignees were advised to dispose the old power supply per their premise protocol or local regulations. The consignees from US can contact Spacelabs Medical at 1-800-522-7025, #2 for Technical Support , and consignees outside the United States can call 1-425-657-7200 x5089 for any questions about the recall.
Quantity in Commerce2877 units in US and 731 units outside US
DistributionWorldwide Distribution: USA in the following states: AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, PA, RI, SC, SD, TN, TX, VA, WA, WI, and WY, and including the following countries: AUSTRALIA, AUSTRIA, CANADA, Caribbean, CZECH REPUBLIC, EGYPT, ESTONIA, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, ISRAEL, ITALY, JORDAN, KUWAIT, LEBANON, MEXICO, NETHERLANDS, PAKISTAN, PANAMA, POLAND, PUERTO RICO, SAUDI ARABIA, SINGAPORE, SWITZERLAND, THAILAND, TRINIDAD AND TOBAGO, TURKEY, UNITED KINGDOM, and VENEZUELA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DXJ
-
-