• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall MADA, Hospilak, respirex, Trianim, McKesson, optimist, portex, medichoice

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall MADA, Hospilak, respirex, Trianim, McKesson, optimist, portex, medichoicesee related information
Date Initiated by FirmJanuary 07, 2011
Date PostedFebruary 15, 2011
Recall Status1 Terminated 3 on February 27, 2013
Recall NumberZ-1256-2011
Recall Event ID 57626
Product Classification oxygen mask - Product Code BYG
ProductOxygen Masks- Labeling as follows: MADA, Inc. Adult Medium Concentration Oxygen Mask with Non-crush 7 Coiled Tubing, Connector Both Ends No 1327; RESPIREX Aerosol Mask Adult Elongated Manufactured for PSS World Medical Inc. RESPIREX Nebulizer Kits w/Aerosol Mask Adult 7 ft. (2.1m) Aerosol Tubing Manufactured for PSS World Medical Inc. Unomedical, Inc., Hospilak REF 100-E Aerosol Mask Adult Unomedical, Inc., Hospilak Ref. No. 101-E, Adult Aerosol Mask w/Aerosol Connector, 15 mm; Unomedical, Inc., Hospilak Ref. No. 102-E, Medium Concentration Mask, Adult 7 (2.1m) Tubing Unomedical, Inc., Hospilak Ref. No. 103-E, Medium Concentration Mask, Adult Unomedical, Inc., Hospilak REF 104-E High Concentration Rebreather Oxygen Mask, Adult w/7 (2.1)m) Oxygen Tubing, Unomedical, Inc., Hospilak Ref. No. 105-E, High Concentration Rebreather Oxygen Mask, Adult, No Tubing Unomedical, Inc., Hospilak Ref. No. 106-E, Oxygen Mask, Adult High Concentration Non-Rebreather 7 (2.1m) Oxygen Tubing Unomedical, Inc., Hospilak Ref. No. 108-E, Oxygen Mask, Adult High Concentration Non-Rebreather 7 (2.1m) Oxygen Tubing Unomedical, Inc., Hospilak Ref. No. 118-E, Aerosol Mask Elongated, Adult 6 (1.8 m) Aerosol Tubing Unomedical, Inc., Hospilak Ref. No. 120-E, Venturi Mask Kit, Adult All Diluters No-Crush MADA INC. Aerosol Mask, Adult Size, Elongated MADA, Inc. 100% Latex Free, 1 unit per package/50 units per case Disposable Single Patient Use MADA INC. Aerosol Mask, Pediatric Size, Elongated No 1327 100% Latex Free MADA, Inc. 100% Latex Free, 1 unit per package/50 units per case Disposable Single Patient Use MADA INC. Up- Mist Medication Nebulizer With 7 ft. Oxygen Tubing and Elongated Aerosol Mask No. 1310MA 100% Latex Free 1 unit per package/50 units per case Disposable Single Patient Use MADA INC Up-Mist Medication Nebulizer with 7 ft. Oxygen Tubing and Pediatric Aerosol Mask No. 1310 NMP 100% Latex Free 1 unit per package/50 units per case Disposable Single Patient Use Unomedical, Inc., Hospilak REF No. 131-E, Venturi Mask Kit, Adult with 31% diluter only, MADA, Inc. Child/Pediatric Medium Concentration Oxygen Mask with Non-Crush 7 Coiled Tubing, Connector Both Ends, NO 1327C, Unomedical, Inc., Hospilak REF No. 182-E, 3-in-1 Emergency Mask, Elongated Adult, 7 (2.1m) Oxygen Tubing, Unomedical, Inc., Hospilak REF No. 200-E, Aerosol Mask, Adult Small, No Tubing, Unomedical, Inc., Hospilak REF No. 202-E, Medium Concentration Oxygen Mask, Adult with 7 (2.1m) Oxygen Tubing, Unomedical, Inc., Hospilak REF No. 220-E, Aerosol Mask, Pediatric, No Tubing, Unomedical, Inc., Hospilak REF No, 221-E, Aerosol Mask Elongated, Pediatric 5 (1.5m) Aerosol Tubing Unomedical, Inc., Hospilak REF No. 222-E, Medium Concentration Oxygen Mask, Pediatric with 7 (2.1m) Oxygen Tubing, Unomedical, Inc., Hospilak REF No. 223-E, Medium Concentration Oxygen Mask, Pediatric No Tubing, Unomedical, Inc., Hospilak REF No. 228-E, Up-Mist Nebulizer with 7 (2.1m) Oxygen Tubing, Pediatric Aerosol Mask Elongated Unomedical, Inc., Hospilak REF No. 245-E, Adult Trach Mask Unomedical, Inc., Hospilak REF No. 246-E, Pediatric Trach Mask Unomedical, Inc., Hospilak REF No. 260-E, Aerosol Mask, Infant No Tubing, Unomedical, Inc., Hospilak REF No. 263-E, Medium Concentration Mask Infant, No Tubing, Unomedical, Inc., Hospilak REF No. 282-E, 3 in 1 Emergency Mask, Elongated Pediatric, 7 (2.1m) Oxygen Tubing TRI-ANIM Adult Aerosol Mask Elongated adult aerosol mask with elastic strap and adjustable noseclip Part Number 301-170 50/case eValueMed Tri-anim Pediatric Aerosol Mask Elongated pediatric aerosol mask with elastic strap and adjustable noseclip Part Number 301-171 50/case eValueMed Tri-anim Adult Oxygen Mask Elongated adult medium mask with elastic strap and adjustable noseclip Part Number 301-180 eVal
Code Information Numerous lots produced between June 2009 and September 2010.
FEI Number 3011987967
Recalling Firm/
Manufacturer
Convatec Inc.
200 Headquarters Park Drive
Skillman NJ 08558-2600
For Additional Information ContactMs. Barbara DeLade
908-904-2675
Manufacturer Reason
for Recall
Face tents and Oxygen masks may have a sticky glue-like substance on the inside or outside of the mask consisting of dioctyl phthalate, epoxidized soybean oil and glycerol monoleate.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionUnomedical, Inc issued recall letters with attached return response forms dated January 7, 2011 via UPS. The letters request the customer stop use and quarantine product. The recall response form is to be completed and returned to Unomedical. The letters/return response form were also sent out in Spanish. Another recall letter was sent dated February 16, 2011 to capture additional lots that werent included in the initial letter.
Quantity in Commerce1113200 pcs.
DistributionWorldwide distribution, including USA, Puerto Rico, Canada, Mexico, Colombia, El Salvador, Paraguay, UK, Austria, Belgium, Cyprus, Czech Republic, Denmark, Estonia, Faroe Islands, Finland, France, Germany, Hungary, Iceland, Ireland, Italy, Netherlands, Norway, Poland, Spain, Sweden and Switzerland, Israel, Jordan, Lebanon, Libya, Morocco, Saudi Arabia, United Arab Emirates, Russia, Australia, New Zealand, Japan and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-