| | Class 2 Device Recall Unicel DXC and SYNCHRON LX Clinical Systems |  |
| Date Initiated by Firm | October 22, 2010 |
| Date Posted | April 26, 2011 |
| Recall Status1 |
Terminated 3 on September 02, 2014 |
| Recall Number | Z-2080-2011 |
| Recall Event ID |
57779 |
| 510(K)Number | K042291 |
| Product Classification |
Electrode, ion specific, potassium - Product Code CEM
|
| Product | UniCel 600 PRO
The UniCel DxC SynchronClinical Systems are fully automated, computer-controlled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent). |
| Code Information |
Part Number: A11810 Subsequent Product Codes: CEK, CEO, CFJ, CGA, CGX, CJW, DCK, DLZ, JFL, JFP, JGJ, JGS, JIY, JJE, JXM, LFP |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 200 S Kraemer Blvd Brea CA 92821-6208
|
| For Additional Information Contact | Clair K. ODonovan, Ph.D. 714-993-5321 |
Manufacturer Reason for Recall | This recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
|
FDA Determined Cause 2 | Environmental control |
| Action | The firm, Beckman Coulter, sent an "URGENT: PRODUCT CORRECTIVE ACTION (PCA)" letter dated October 20, 2010 with attached PCA Response Form (via US Postal Service for US Customers) to all customers who purchased Synchron LX and UniCel DxC Clinical Systems
Operating Software (all versions). The letter describes the product, problem and actions to be taken. The customers were instructed to do the following:
-After modifying any UDR wavelength, they MUST reboot to apply the update of the polychrome correction.
-If the modified UDR requires calibration, recalibrate after rebooting.
- Complete and return the enclosed RESPONSE FORM within 10 days via fax to (714) 961-4234 or mail to: Beckman Coulter, Inc., Attn: Regulatory Affairs, 250 S. Kraemer Blvd, PO Box 8000, Brea, California, 92282-8000; Mail Code E2.SE.08.
In addition, the customers were instructed to share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. If customers forwarded any of the affected product listed above to another laboratory, please provide a copy of this letter to them.
If customers have any questions regarding this Product Corrective Action letter, please call Beckman's Customer Support Center (Hotline) at (800) 854-3633 in the United States and Canada or www.beckmancoulter.com/customersupport/supportrequests. |
| Quantity in Commerce | 4879 units total (3490 in US) |
| Distribution | Worldwide distribution: USA (nationwide) and countries including:Angola, Australia, Belgium, Bulgaria, China, Colombia, Costa Rica, Czech Republic, Denmark, Ecuador, Egypt, Eritrea, Finland, France, French Polynesia, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Macao, Mexico, Namibia, Netherlands, Nez Zealand, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, South Africa, Sweden, Switzerland, Turkey and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CEM
|
|
|
|