| Class 2 Device Recall STERRAD NX Cassettes | |
Date Initiated by Firm | September 20, 2012 |
Date Posted | January 09, 2013 |
Recall Status1 |
Terminated 3 on October 17, 2013 |
Recall Number | Z-0661-2013 |
Recall Event ID |
63877 |
Product Classification |
Sterilizer, chemical - Product Code MLR
|
Product | STERRAD NX Cassettes, P/N 10133.
The STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD NX Cassettes are only used in conjunction with the STERRAD NX Sterilization system. |
Code Information |
Lot # 11J037, 11L038, 11J033 |
Recalling Firm/ Manufacturer |
Advanced Sterilization Products 33 Technology Dr Irvine CA 92618-2346
|
For Additional Information Contact | 949-453-6400 |
Manufacturer Reason for Recall | Advanced Sterilization Products is recalling three lots of STERRAD NX System Cassettes because it does not have adequate data to support the entire duration of the product's labeled shelf-life. |
FDA Determined Cause 2 | Labeling design |
Action | Advanced Sterilization Product sent an Urgent Medical Device Recall (Removal) letter dated December 4, 2012, to all affected customers. The letter informed the customers that the Sterrad NX Cassettes do not have adequate data to support the entire duration of the product's labeled shelf-life. Customers are instructed of the problems identifed and the actions to be taken. Customers are instructed to notify their customers if they have further distributed the recalled products and to have them contact Stericycle at (877) 650-7686 to arrange for return of the product. Customers with questions are instructed to call Stericycle at (877) 650-7686 Mon-Fri from 8am to 5pm, ET.
For questions regarding this recall call 949-453-6400. |
Quantity in Commerce | 12,280 units |
Distribution | Worldwide Distribution - USA (nationwide) and internationally to the following countries:
AR, CO, TW, SI, AU, MX, MY, HU, PH, BE, FR, DE, IE, IT, ES, NO, CH, and GB. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|