 Class 2 Device Recall Sarns SoftFlow Aortic Cannulasee related information
Date Initiated by FirmFebruary 23, 2012
Date PostedMarch 23, 2012
Recall Status1 Terminated 3 on April 03, 2013
Recall NumberZ-1305-2012
Recall Event ID 60718
510(K)NumberK083301 K934127 
Product Classification Catheter, cannula and tubing, vascular, cardiopulmonary bypass - Product Code DWF
ProductSarns Soft-Flow Aortic Cannula with suture flange Angled tip, wire-reinforced with luer port 7.0mm (21 Fr) OD with 3/8" connector, 14" (36cm) long Product Usage: The Sarns-Aortic Cannula is indicated for use In perfusion of the ascending aorta during cardiopulmonary bypass surgery. The X-Coating is intended to reduce the adhesion of platelets to the blood contacting surfaces of the device during cardiopulmonary bypass procedures.
Code Information catalog number: 6394 and lot numbers: 0546814, 0551455, 0554919, 0556605, 0559943, 0562323, 0563863, 0565369, 0569977, 0571822, 0578746, 0584741, 0589428, 0590865, 0595113, 0600216, 0605363, 0609905, 0611582, 0618230, 0620597, 0623643, 0631691, 0632823, 0634949, 0638016, 0641049, 0647966, and 621703C.
FEI Number 1828100
Recalling Firm/
Manufacturer
Terumo Cardiovascular Systems Corporation
6200 Jackson Road
Ann Arbor MI 48103-9586
For Additional Information ContactTracy Bellanca
734-741-6173
Manufacturer Reason
for Recall
During in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionTerumo sent an URGENT MEDICAL DEVICE RECALL letter dated February 23, 2012 via Federal Express to all US customers. Terumo Foreign Affiliates received the recall notification via email on February 24th, 2012. The letters identified the affected products, reson for removal, potential hazard, affected population and actions to be taken. Customers were instructed to review the Medical Device Removal notice, assure that all users are aware of the notice, confirm receipt of this communication by faxing the attached Customer Response Form to the fax number provided or email to tcvs.recall@terumomedical.com and call Terumo CVS Customer Service for a Returned Goods Authorization (RGA) number. For questions call Terumo CVS Customer Service at 1-800-521-2818.
Quantity in Commerce188,282 Cannulae and 248 Custom Tubing Packs
DistributionUSA Nationwide Distribution including the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV and the countries of: CANADA, CHILE, India, Japan, Singapore, AUSTRALIA, Malaysia, BELGIUM, BELGIUM, BRAZIL, CHILE, BRAZIL, BRAZIL, and Philippines.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.