| Class 2 Device Recall Zoll Medical | |
Date Initiated by Firm | August 16, 2012 |
Date Posted | September 17, 2012 |
Recall Status1 |
Terminated 3 on September 19, 2012 |
Recall Number | Z-2401-2012 |
Recall Event ID |
63004 |
510(K)Number | K112432 |
Product Classification |
Automated external defibrillators (non-wearable) - Product Code MKJ
|
Product | Zoll X Series Defibrillator/Pacemaker/Monitor
Product Usage:
The product is used by qualified medical personnel for the purpose of converting ventricular fibrillation to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats. In addition, the product may be used in synchronized mode to terminate certain atrial and ventricular tachycardias and other arrhythmias resistant to drug therapy. The X Series may be used for cardiac pacing in conscious or unconscious patients for up to a few hours duration as an alternative to endocardial stimulation. The X Series may be used for the monitoring a patients electrocardiogram (ECG). |
Code Information |
Serial Numbers: AR11J000164, AR12E000853, AR12E000856, AR12E000868, AR12E000920, AR12E000922, AR12E000923, AR12E000931, AR12E000944 AR12F001100, AR12F001298. |
Recalling Firm/ Manufacturer |
ZOLL Medical Corporation 269 Mill Road Chelmsford MA 01824
|
For Additional Information Contact | 978-421-9655 |
Manufacturer Reason for Recall | Shipped with incorrect software |
FDA Determined Cause 2 | Process control |
Action | Zoll Medical telephoned affected customers on August 16, 2012. Customers were informed of the affected product, problem and actions to be taken. Customers were instructed to remove and return the affected product for replacement. For question contact ZOLL's Technical Support. |
Quantity in Commerce | 11 units |
Distribution | Worldwide Distribution - US Nationwide including the states of: LA, MA, ME, NC, ND, NE, OH and the country of Germany. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = MKJ
|
|
|
|