| | Class 2 Device Recall Integrated Module Housing |  |
| Date Initiated by Firm | August 27, 2012 |
| Date Posted | September 14, 2012 |
| Recall Status1 |
Terminated 3 on April 15, 2013 |
| Recall Number | Z-2382-2012 |
| Recall Event ID |
63007 |
| 510(K)Number | K842616 |
| Product Classification |
Detector and alarm, arrhythmia - Product Code DSI
|
| Product | Ultraview SL 91493 Integrated Module Housing.
PCIS patient monitors is intended to sound various alarm tones that are generated by the specific PCIS patient parameter modules. Each patient parameter module determines (a) when an alarm condition is violated; (b) the alarm priority (i.e. high, medium or low); (c) alarm limits; and (d) when to generate and terminate alarms. PCIS patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. Spacelabs Medical PCIS patient monitors are designed to communicate with a variety of external devices such as display devices, network devices, serial devices, user input devices, audio system, local/remote recorders and Spacelabs Medical patient parameter modules. In all applications, Spacelabs Medical PCIS patient monitors provide access to that data on the Spacelabs Medical network either via hardwire or wireless transmission. Through this network, a patient's data may be displayed at the bedside monitor, central station or clinical information system. |
| Code Information |
Domestic - 1493-000275, 1493-000276, 1493-000277, 1493-002469, 1493-002616, 1493-003358, 1493-003359, 1493-003667, 1493-003668, 1493-003669, 1493-003670, 1493-003671, 1493-003672, 1493-003673, 1493-003674, 1493-003675, 1493-003676, 1493-003677, 1493-003678, 1493-003679, 1493-003680, 1493-003681, 1493-003682, 1493-003683, 1493-003684, 1493-003685, 1493-003686, 1493-003687, 1493-003688, 1493-003693, 1493-003694, 1493-003695, 1493-003696, 1493-003697, 1493-003879, 1493-003880, 1493-003881, 1493-003882, 1493-003883, 1493-003884, 1493-003885, 1493-003886, 1493-003887, 1493-003888, 1493-003889, 1493-003890, 1493-003891, 1493-003892, 1493-003893, 1493-003894, 1493-003895, 1493-003896, 1493-003897, 1493-003898, 1493-003899, 1493-003900, 1493-003901, 1493-003902, 1493-003903, 1493-003904, 1493-003905, 1493-003906, 1493-003907, 1493-003908, 1493-003909, 1493-003914, 1493-003915, 1493-003916, 1493-003917, 1493-003918, 1493-003919, 1493-003920, 1493-003921, 1493-003922, 1493-003923, 1493-003924, 1493-003925, 1493-003926, 1493-003927, 1493-003928, 1493-003929, 1493-003930, 1493-003931, 1493-003932, 1493-003933, 1493-003934, 1493-003935, 1493-003936, 1493-003937, 1493-003938, 1493-003939, 1493-003940, 1493-003941, 1493-003942, 1493-003943, 1493-003944, 1493-003945, 1493-003957, 1493-003958, 1493-003959, 1493-003960, 1493-003961, 1493-003962, 1493-003963, 1493-003964, 1493-003965, 1493-003985, 1493-004007, 1493-004011, 1493-004023, 1493-004026, 1493-004027, 1493-004028, 1493-004029, 1493-004030, 1493-004031, 1493-004032, 1493-004033, 1493-004034, 1493-004043, 1493-004044, 1493-004047, 1493-004048, 1493-004050, 1493-004051, 1493-004052, 1493-004053, 1493-004054, 1493-004055, 1493-004056, 1493-004058, 1493-004060, 1493-004064, 1493-004067, 1493-004068, 1493-004069, 1493-004070, 1493-004071, 1493-004075, 1493-004084, 1493-004085, 1493-004086, 1493-004106, 1493-004112, 1493-004113, 1493-004114, 1493-004115, 1493-004116, 1493-004117, 1493-004118, 1493-004119, 1493-004120, 1493-004121, 1493-004122, 1493-004123, 1493-004124, 1493-004125, 1493-004126, 1493-004127, 1493-004128, 1493-004130, 1493-004158, 1493-004159, 1493-004160, 1493-004161, 1493-004162, 1493-004163, 1493-004164, 1493-004172, 1493-004173, 1493-004174, 1493-004175, 1493-004176, 1493-004177, 1493-004178, 1493-004179, 1493-004180, 1493-004181, 1493-004182, 1493-004183, 1493-004184, 1493-004185, 1493-004205, 1493-004206, 1493-004207, 1493-004208, 1493-004209, 1493-004210, 1493-004211, 1493-004212, 1493-004214, 1493-004215, 1493-004216, 1493-004217, 1493-004218, 1493-004219, 1493-004220, 1493-004221, 1493-004222, 1493-004223, 1493-004224, 1493-004225, 1493-004226, 1493-004227, 1493-004228, 1493-004229, 1493-004230, 1493-004231, 1493-004232, 1493-004233, 1493-004234, 1493-004235, 1493-004236, 1493-004237, 1493-004238, 1493-004239, 1493-004240, 1493-004241, 1493-004242, 1493-004243, 1493-004256, 1493-004257, 1493-004258, 1493-004259, 1493-004260, 1493-004261, 1493-004262, 1493-004263, 1493-004264, 1493-004265, 1493-004266, 1493-004267, 1493-004268, 1493-004269, 1493-004270, 1493-004271, 1493-004272, 1493-004273, 1493-004274, 1493-004275, 1493-004276, 1493-004277, 1493-004278, 1493-004279, 1493-004280, 1493-004281, 1493-004282, 1493-004283, 1493-004284, 1493-004285, 1493-004286, 1493-004287, 1493-004288, 1493-004289, 1493-004290, 1493-004291, 1493-004292, 1493-004293, 1493-004294, 1493-004295, 1493-004296, 1493-004297, 1493-004298, 1493-004299, 1493-004301, 1493-004302, 1493-004306, 1493-004307, 1493-004308, 1493-004309, 1493-004310, 1493-004311, 1493-004312, 1493-004313, 1493-004314, 1493-004315, 1493-004316, 1493-004317, 1493-004318, 1493-004319, 1493-004320, 1493-004321, 1493-004322, 1493-004323, 1493-004324, 1493-004325, 1493-004326, 1493-004327, 1493-004328, 1493-004329, 1493-004330, 1493-004331, 1493-004332, 1493-004333, 1493-004334, 1493-004335, 1493-004336, 1493-004337, 1493-004338, 1493-004339, 1493-004340, 1493-004341, 1493-004342, 1493-004343, 1493-004344, 1493-004345, 1493-004346, 1493-004347, 1493-004348, 1493-004349, 1493-004350, 1493-004351, 1493-004352, 1493-004353, 1493-004354, 1493-004355, 1493-004356, 1493-004357, 1493-004358, 1493-004359, 1493-004360, 1493-004361, 1493-004362, 1493-004363, 1493-004364, 1493-004365, 1493-004366, 1493-004367, 1493-004368, 1493-004369, 1493-004370, 1493-004371, 1493-004372, 1493-004373, 1493-004374, 1493-004375, 1493-004376, 1493-004377, 1493-004378, 1493-004379, 1493-004380, 1493-004381, 1493-004382, 1493-004383, 1493-004384, 1493-004385, 1493-004405, 1493-004406, 1493-004407, and 1493-004408. International - 1493-002941, 1493-003666, 1493-003689, 1493-003690, 1493-003691, 1493-003692, 1493-003910, 1493-003911, 1493-003912, 1493-003913, 1493-003946, 1493-003947, 1493-003948, 1493-003949, 1493-003950, 1493-003951, 1493-003952, 1493-003953, 1493-003954, 1493-003955, 1493-003956, 1493-003966, 1493-003967, 1493-003968, 1493-003969, 1493-003970, 1493-003971, 1493-003972, 1493-003973, 1493-003974, 1493-003975, 1493-003976, 1493-003977, 1493-003978, 1493-003979, 1493-003980, 1493-003981, 1493-003982, 1493-003983, 1493-003984, 1493-003986, 1493-003987, 1493-003988, 1493-003989, 1493-003990, 1493-003991, 1493-003992, 1493-003993, 1493-003994, 1493-003995, 1493-003996, 1493-003997, 1493-003998, 1493-003999, 1493-004000, 1493-004001, 1493-004002, 1493-004003, 1493-004004, 1493-004005, 1493-004008, 1493-004009, 1493-004010, 1493-004012, 1493-004013, 1493-004014, 1493-004015, 1493-004016, 1493-004017, 1493-004018, 1493-004019, 1493-004020, 1493-004021, 1493-004022, 1493-004024, 1493-004025, 1493-004035, 1493-004036, 1493-004037, 1493-004038, 1493-004039, 1493-004040, 1493-004041, 1493-004042, 1493-004045, 1493-004046, 1493-004065, 1493-004066, 1493-004072, 1493-004073, 1493-004074, 1493-004076, 1493-004078, 1493-004079, 1493-004080, 1493-004081, 1493-004083, 1493-004087, 1493-004088, 1493-004089, 1493-004090, 1493-004091, 1493-004092, 1493-004093, 1493-004094, 1493-004095, 1493-004096, 1493-004097, 1493-004098, 1493-004099, 1493-004100, 1493-004101, 1493-004102, 1493-004103, 1493-004104, 1493-004105, 1493-004107, 1493-004108, 1493-004109, 1493-004110, 1493-004111, 1493-004129, 1493-004131, 1493-004132, 1493-004133, 1493-004134, 1493-004135, 1493-004136, 1493-004137, 1493-004138, 1493-004139, 1493-004140, 1493-004141, 1493-004142, 1493-004143, 1493-004144, 1493-004145, 1493-004146, 1493-004147, 1493-004148, 1493-004149, 1493-004150, 1493-004151, 1493-004152, 1493-004153, 1493-004154, 1493-004155, 1493-004165, 1493-004166, 1493-004167, 1493-004168, 1493-004169, 1493-004170, 1493-004171, 1493-004186, 1493-004187, 1493-004188, 1493-004189, 1493-004190, 1493-004191, 1493-004192, 1493-004193, 1493-004194, 1493-004195, 1493-004196, 1493-004197, 1493-004198, 1493-004199, 1493-004200, 1493-004201, 1493-004202, 1493-004203, 1493-004204, 1493-004300, 1493-004304, 1493-004305, and 670-1350-00. |
Recalling Firm/ Manufacturer |
Spacelabs Healthcare, Llc 5150 220th Ave Se Issaquah WA 98029-6834
|
| For Additional Information Contact | Spacelabs Healthcare 800-522-7025 Ext. 5970 |
Manufacturer Reason for Recall | There is a potential failure of the Integrated Module Housing causing the patient module parameters to drop off the monitor display or the patient module will fail to sign on when inserted into the housing. |
FDA Determined Cause 2 | Mixed-up of materials/components |
| Action | Spacelabs Healthcare sent an "URGENT- MEDICAL DEVICE CORRECTION" letter dated August 27, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-800-522-7025 and select 2 for assistance regarding this recall. |
| Quantity in Commerce | 527 units total; 350 US and 177 International |
| Distribution | Worldwide Distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|