 Class 2 Device Recall Stockert S5/Sorin C5 Systemsee related information
Date Initiated by FirmSeptember 19, 2012
Date PostedNovember 08, 2012
Recall Status1 Terminated 3 on January 18, 2013
Recall NumberZ-0271-2013
Recall Event ID 63383
510(K)NumberK071318 K093882 
Product Classification Console, heart-lung machine, cardiopulmonary bypass - Product Code DTQ
ProductS5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
Code Information Serial Numbers: 10E06219,10E06220, 10E06231-10E06234, 10E06239-10E06241, 10E06255-10E06273, 10E06286-10E06289, 10E06291-10E06355, 10E06359-10E06364, 10E06366, 10E06368-10E06383, 10E06385, 10E06387-10E06391,10E06395-10E06401, 10E06403-10E06447, 10E6502.
FEI Number 1718850
Recalling Firm/
Manufacturer
Sorin Group USA, Inc.
14401 W 65th Way
Arvada CO 80004
For Additional Information ContactMr. Jack Ellison
303-467-6306
Manufacturer Reason
for Recall
Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determi
FDA Determined
Cause 2
Mixed-up of materials/components
ActionThe firm, Sorin Group, sent an "URGENT FIELD SAFETY NOTICE" dated September 19, 2012 to all affected customers by certified mail. The notice described the product, problem and actions to be taken. The customers were instructed to follow the provided instructions for the continued safe use of the Perfusion systems until they have been serviced or replaced by Sorin Group. The Sorin Group Service Team will contact the customer to schedule servicing and replacement of the affected product. The customers were also instructed to complete and return the Customer Response Form via fax to 303-467-6502 or by email to yvonne.feyerherm@sorin.com; assure this notice is distributed to all personnel who need to be aware of this notice; and if they have transferred the affected products to a third party, pass this information with them as well as with Sorin Group Customer Service at 1-800-650-2623. For questions regarding this notice, contact Sorin Group USA Customer Service at 1-800-650-2623.
Quantity in Commerce35 units
DistributionWorldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denmark, Finland, Germany, India, Iran-Islamic Republic of, Ireland, Japan, Lebanon, Morocco, Pakistan, Poland, Russian Federation, Saudia Arabia, Spain, South Africa, Thailand, Turkey and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.