 Class 2 Device Recall Signa Infinity Signa Infinity Twinspeed Signa Vibrant Signa HDx Signa HDxt Signa HDxisee related information
Date Initiated by FirmNovember 21, 2011
Date PostedAugust 09, 2013
Recall Status1 Terminated 3 on October 22, 2014
Recall NumberZ-1943-2013
Recall Event ID 65133
510(K)NumberK041476 K040444 K052293 K013636 K030874 
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
ProductSIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM GE SIGNA 3.0T WITH EXCITE MR SYSTEM GE SIGNA EXCITE 1.5T MR SYSTEM GE SIGNA EXCITE 3.0T MR SYSTEM Signa HDx Signa HDxt Signa HDi Signa Vibrant GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM (i.e. Signa Ovation 5 and Ovation HD) GE 1.5T SIGNA HDe MR SYSTEM GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM GE 0.7T SIGNA OPENSPEED (4) WITH EXCITE MR SYSTEM GE 0.7T SIGNA OPENSPEED (5) WITH EXCITE MR SYSTEM The GE Signa Infinity line of products are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio and short scan times. The Signa Infinity System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa Infinity systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Code Information Ref. # 60808  "Products: Signa Infinity MR System with EXCITE Technology, Signa EXCITE 1.5T and Signa EXCITE 3.0T, Signa 3T with EXCITE, Signa 3T Infinity with EXCITE, GE 1.5T AND 3.0T SIGNA HDX MR SYSTEM (Waukesha, WI) FMI 60808" Mfg Lot or Serial # 00000298045MR6 00000237944MR4 00000278301MR7 00000257724MR5 00000029079MR2 00000280678MR4 00000286982MR4 00000283664MR1 00000263686MR8 00000238343MR8 00001011BD6649 00000105XF5011 00000263832MR8 00000256696MR6 00000284409MR0 00000265824MR3 00000284407MR4 00000252930MR3 00000301212MR7 00000277495MR8 00000238364MR4 00000230664MR5 00000262150MR6 00000275961MR1 00000231910MR1 00000284466MR0 00000276678MR0 00000006583YR9 00000262101MR9 00000283815MR9 00000268471MR0 00000028886MR2 00000297726MR2 00000277916MR3 00000290478MR7 00000247217MR3 00000245819MR8 00000284864MR6 00000287026MR9 00000238355MR2 00000265183MR4 00000266460MR5 00000251094MR9 00000266825MR9 00000265718MR7 00000289270MR1 00000288702MR4 00000286851MR1 00000281374MR9 000000HDMR2685 00000HDMR20256 00000236703MR5 00000297201MR6 00000234797MR9 00000265066MR1 00000278766MR1 00000255610MR8 00000280252MR8 00000281156MR0 00000011742YR4 00000011570YR9 00000279479MR0 00000009057YR1 00000284902MR4 00000287649MR7 00000239864MR2 00000235678MR0 00000241145MR2 00000255599MR3 00000028959MR7 00000235242MR5 00000257752MR6 00000256990MR3 00000275647MR6 00000246615MR9 00000300674MR9 00000011579YR0 00000286249MR8 00000279921MR1 00000011882YR8 00000262908MR7 00000271646MR2 00000290192MR4 00000254485MR6 00000259530MR4 00000288192MR8 00000245803MR2 00000246622MR5 00000278624MR2 00000261879MR1 00000937BD5ED5 00000301020MR4 00000289016MR8 00000229365MR2 00000232415MR0 00000256228MR8 00000264529MR9 00000298718MR8 00000284386MR0 00000276588MR1 00000242558MR5 00000253635MR7 00000256984MR6 00000287626MR6 00000280089MR4 00000256699MR0 00000260698MR6 00000262933MR5 00000272896MR2 00000272890MR5 00000237064MR1 00000005232YR4 00000233272MR4 00000299162MR8 00000229474MR2 00000248485MR5 00000268524MR6 00000237059MR1 00000285218MR4 00000011910YR7 00000266754MR1 00000286785MR1 00000301356MR2 00000299828MR4 00000241868MR9 00000273672MR6 00000262458MR3 00000228289MR3 00000257732MR8 00000286701MR8 00000278873MR5 00000244845MR4 00000291805MR0 00000283949MR6 00000230694MR2 00000242548MR6 00000241866MR3 00000289771MR8 00000292635MR0 00000278669MR7 00000279078MR0 00000235690MR5 00000286843MR8 00000270976MR4 00000269336MR4 00000260699MR4 00000012244YR0 00000266828MR3 00000260864MR4 00000272284MR1 00000273632MR0 00000279874MR2 00000284640MR0 00000258090MR0 00000300909MR9 00000249768MR3 00000272283MR3 00000286016MR1 00000283849MR8 00000296144MR9 00000279405MR5 00000287081MR4 00000302511MR1 00000251241MR6 00000283682MR3 00000274745MR9 00000234794MR6 00000255208MR1 00000262151MR4 00000279483MR2 00000933BD5E06 00000239042MR5 00001008BD6517 00000235236MR7 00000280524MR0 00000267180MR8 00000284599MR8 00000270386MR6 00000HDMR20043 00000287846MR0 00000268283MR9 00000269380MR2 00000235680MR6 00000251800MR9 00000294406MR4 00000245809MR9 00000255571MR2 00000264617MR2 00000265964MR7 00000234855MR5 00000HDMR20069 00000273541MR3 00000278688MR7 00000277575MR7 00000284709MR3 00000011508YR9 00000277494MR1 00000248159MR6 00000282788MR9 00000239241MR3 00000278291MR0 00000278391MR8 00000919BD5B98 00000265720MR3 00000231911MR9 00000241865MR5 00000261877MR5 00000255569MR6 00000950BD6297 00000296292MR6 00000243702MR8 00000010163YR4 00000260470MR0 00000290052MR0 00000012162YR4 00000288015MR1 00000280000MR1 00000266799MR6 00000275583MR3 00000272627MR1 00000231913MR5 00000300799MR4 00000276634MR3 00000244846MR2 00000251231MR7 00000024170MR6 00000243700MR2 00000262464MR1 00000008345YR1 00000239857MR6 00000260626MR7 00000275653MR4 00000281157MR8 00000270462MR5 00000281271MR7 00000261328MR9 00000012023YR8 00000283377MR0 00000008059YR8 00000243685MR5 00000011589YR9 00000285927MR0 00000278193MR8 00000287940MR1 00000237949MR3 00000282547MR9 00000280775MR8 0000267533MR8A 00000265827MR6 00000286853MR7 00000241861MR4 00000265596MR7 00000256704MR8 00000266073MR6 00000254665MR3 00000268049MR4 00000903BD55CD 00000242585MR8 00000261068MR1 00000297656MR1 00000243249MR0 00000262735MR4 00000269045MR1 00000286526MR9 00000271753MR6 00000281937MR3 00000223456MR5 151 000S0933BD5DFD 00000266214MR6 00000294524MR4 00000290130MR4 00000291370MR5 00000025636MR4 00000241514MR9 00000237936MR0 00000292922MR2 00000011550YR1 00000285775MR3 00000915BD5AED 00000282371MR4 00000260011MR2 00000287864MR3 00000006556YR5 00000301463MR6 00000287826MR2 00001048XF501C 00000258419MR1 00000255211MR5 00000244850MR4 00000287767MR8 00000234847MR2 00000249758MR4 00000263792MR4 00000233267MR4 00000280436MR7 00000288096MR1 00000231893MR9 00000029128MR0 00000260017MR9 00000265939MR9 00000263972MR2 00000281826MR8 00000234843MR1 00000260462MR7 00000029187MR8 00000299416MR8 00000277784MR5 00000263507MR6 00000291750MR8 00000252437MR9 00000271051MR5 00000237968MR3 00000232419MR2 00000011585YR7 00000011567YR3 00000218753MR2 00000HDMR10899 00000235237MR5 00000252422MR1 00000288852MR7 00000248132MR3 00000290370MR6 00000275967MR8 00000289763MR5 00000265062MR0 00000280249MR4 000000HDMR1086 00000241143MR7 00000237932MR9 00000028674MR1 00000241137MR9 00000253153MR1 00000255612MR4 00000283865MR4 00000278555MR8 00000263313MR9 00000011711YR9 00000919BD5BB0 00000286091MR4 00000284389MR4 00000279753MR8 00000011521YR2 000000HDMR2984 00000269047MR7 00000275648MR4 00000258403MR5 00000256367MR4 00000012355YR4 00000247727MR1 00000294694MR5 00000292964MR4 00000282393MR8 00000261069MR9 00000278678MR8 00000272891MR3 00000301858MR7 00000238867MR6 00000301804MR1 00000273195MR8 00000283321MR8 00000280462MR3 00000011015YR5 00000271561MR3 00000301565MR8 00000272101MR7 00000250019MR7 00000286886MR7 00000278769MR5 00000258089MR2 00000236831MR4 00000011793YR7 00000022936MR8 00000286767MR9 00000276591MR9 00000287398MR2 00000903BD55EB 00000233430MR8 00000252378MR5 00000262842MR8 00000247213MR2 00000280696MR6 00000234673MR2 00000249770MR9 00000241863MR0 00000238361MR0 00000254188MR6 00000264553MR9 00000255197MR6 00000246623MR3 00000266698MR0 00000259122MR0 00000260863MR6 00000235588MR1 00000242588MR2 00000237946MR9 00000285622MR7 00000245172MR2 00000265613MR0 00000238344MR6 00000257755MR9 00000233437MR3 00000251697MR9 00000233299MR7 00000248478MR0 00000284321MR7 00000245171MR4 00000268318MR3 00000267849MR8 00000298841MR8 00000291802MR7 00000254184MR5 00000247225MR6 0000000000XXXX 00000242549MR4 00000288847MR7 00000283736MR7 00000248167MR9 00000249226MR2 00000233434MR0 00000253166MR3 00000239045MR8 00000236833MR0 00000253631MR6 00000293744MR9 00000281232MR9 00000287652MR2 00000256217MR1 00000241101MR5 00000272563MR8 00000291492MR7 00000268137MR7 00000279056MR6 00000284012MR2 00000272880MR6 00000296858MR4 00000253827MR0 00000259131MR1 00000263795MR7 00000251726MR6 00000287427MR9 00000256687MR5 00000258372MR2 00000239528MR3 00000258087MR6 00000254198MR5 00000254183MR7 00000262048MR2 00000248162MR0 00000233443MR1 00000298384MR9 00000251716MR7 00000258079MR3 00000281800MR3 00000245302MR5 00000236826MR4 00000233300MR3 00000256218MR9 00000237700MR0 00000297705MR6 00000237002MR1 00000240107MR3 00000279455MR0 00000245177MR1 00000270516MR8 00000259421MR6 00000280747MR7 00000253624MR1 00000264467MR2 00000230098MR6 00000264799MR8 00000243695MR4 00000273015MR8 00000259424MR0 00000255214MR9 00000264489MR6 00000255592MR8 00000250433MR0 00000290767MR3 00000291932MR2 00000254668MR7 00000248135MR6 00000276654MR1 00000260675MR4 00000287257MR0 00000234858MR9 00000292213MR6 00000299517MR3 00000237698MR6 00000231892MR1 00000238360MR2 00000271645MR4 00000291455MR4 00000278812MR3 00000237703MR4 00000257746MR8 00000252438MR7 00000254192MR8 00000299483MR8 00000261479MR0 00000241133MR8 00000241100MR7 00000273443MR2 00001101XF5089 00000253627MR4 00000272883MR0 00000263261MR0 00000253164MR8 00000248483MR0 00000246620MR9 00000244849MR6 00000256669MR3 00000245175MR5 00000246625MR8 00000271652MR0 00000300503MR0 00000266800MR2 00000271861MR7 00000238362MR8 00000236699MR5 00000263574MR6 00000261382MR6 00000298980MR4 00000290549MR5 00000300669MR9 00000288872MR5 00000239033MR4 00000259524MR7 00000299622MR1 00000248158MR8 00000264012MR6 0000024925BMR5 00000242553MR6 00000255204MR0 00000241518MR0 00000267671MR6 00000291999MR1 00000246609MR2 00000258418MR3 00000243243MR3 00000291205MR3 00000277777MR9 00000257716MR8 00000292210MR2 00000249276MR7 00000290604MR8 00000265102MR4 00000232406MR9 00000241099MR1 00000251725MR8 00000259941MR3 00000290148MR6 00000259575MR9 00000276419MR9 00000250439MR7 00000253150MR7 00000274162MR7 00000300896MR8 00000257415MR0 00000300753MR1 00000290333MR4 00000255215MR6 00002522296MR9 00000283154MR3 00000252911MR3 00000279055MR8 00000290200MR5 00000239868MR3 00000243698MR8 00000259916MR5 00000298852MR5 00000279409MR7 00000257443MR2 00000239538MR2 00000251709MR2 00000292051MR0 00000235695MR4 00000251060MR0 00000273294MR9 00000260465MR0 00000229479MR1 00000290546MR1 00000266696MR4 00000231898MR8 00000239027MR6 00000264464MR9 00000298853MR3 00000237939MR4 00000256700MR6 00000250434MR8 00000271862MR5 00000259521MR3 00000252423MR9 00000259529MR6 00000281945MR6 00000285427MR1 00000256214MR8 00000262506MR9 00000263577MR9 00000250447MR0 00000236625MR0 00000262592MR9 00000233263MR3 00000236503MR9 41404 00000250034MR6 00000298376MR5 00000265096MR8 00000253839MR5 00000233298MR9 00000256708MR9 00000264465MR6 00000256226MR4 00000264796MR4 00000245174MR8 00000237948MR5 00000238357MR8 00000260104MR5 00000231909MR3 00000233297MR1 00000239508MR5 00000289541MR5 00000238353MR7 00000252432MR0 00000262049MR0 00000262589MR5 00000287107MR7 00000247219MR9 00000261383MR4 00000260862MR8 00000280697MR4 00000298823MR6 00000237692MR9 00000233433MR2 00000238179MR6 00000253147MR3 00000250209MR4 00000250451MR2 00000257429MR1 00000233444MR9 00000271866MR6 00000284601MR2 00000248165MR3 00000258367MR2 00000264794MR9 00000271868MR2 00000261731MR4 00000280693MR3 00000284322MR5 00000251728MR2 00000250192MR2 00000291840MR7 00000243691MR3 00000241725MR1 00000249229MR6 00000300466MR0 00000252660MR6 00000259528MR8 00000249275MR9 00000272882MR2 00000261733MR0 00000237070MR8 00000240103MR2 00000291706MR0 00000230103MR4 00000235688MR9 00000241720MR2 00000273445MR7 00000291754MR0 00000260860MR2 00000251797MR7 00000254674MR5 00000257740MR1 00000271864MR1 00000291787MR0 00000252933MR7 00000288915MR2 00000292391MR0 00000271651MR2 00000256220MR5 00000258096MR7 00000259128MR7 00000287595MR3 00000287614MR2 00000024118MR7 00000257729MR4 00000260016MR1 00000245807MR3 00000253825MR4 00000251714MR2 00000251639MR1 00000250216MR9 00000258399MR5 00000290573MR5 00000231890MR5 00000257750MR0 00000268315MR9 00000230665MR2 00000282102MR3 00000254186MR0 00000259426MR5 00000253639MR9 00000276570MR9 00000239246MR2 00000263259MR4 00000245804MR0 00000239865MR9 00000272879MR8 00000230104MR2 00000288003MR7 00000239010MR2 00000291746MR6 00000250021MR3 00000259962MR9 00000291721MR9 00000273019MR0 00000238182MR0 00000262905MR3 00000251718MR3 00000258421MR7 00000251727MR4 00000254199MR3 00000230683MR5 00000237077MR3 00000230681MR9 00000253161MR4 00000229471MR8 00000243693MR9 00000235241MR7 00000273016MR6 00000285429MR7 00000253842MR9 00000291419MR0 00000261470MR9 00000258095MR9 00000250020MR5 00000250219MR3 00000273296MR4 00000267104MR8 00000242561MR9 00000262734MR7 00000283920MR7 00000229476MR7 00000255213MR1 00000243179MR9 00000232421MR8 00000261474MR1 00000278752MR1 00000256697MR4 00000264551MR3 00000258081MR9 00000257748MR4 00000276267MR2 00000262508MR5 00000246617MR5 00000247229MR8 00000294461MR9 00000294527MR7 00000298978MR8 00000247737MR0 00000284061MR9 00000247228MR0 00000249658MR6 00000285428MR9 00000239534MR1 00000256698MR2 00000245176MR3 00000273442MR4 00000250023MR9 00000255597MR7 00000244852MR0 00000256985MR3 00000259942MR1 00000284622MR8 00000250751MR5 00000253148MR1 00000254500MR2 00000291856MR3 00000274103MR1 00000277776MR1 00000290536MR2 00000247729MR7 00000233426MR6 00000274051MR2 00000231905MR1 00000259586MR6 00000298383MR1 00000233268MR2 00000291733MR4 00000260105MR2 00000263499MR6 00000255591MR0 00000260467MR6 00000251243MR2 00000253638MR1 00000264550MR5 00000296313MR0 00000290032MR2 00000025169MR1 00000024847MR2 00000264792MR3 00000276653MR3 00000233436MR5 00000235693MR9 00000237071MR6 00000238167MR1 00000239234MR8 00000242173MR3 00000239227MR2 00000241869MR7 00000252297MR7 00000258411MR8 00000255203MR2 00000255206MR5 00000255209MR9 00000254662MR0 00000259119MR6 00000250208MR6 00000261478MR2 00000263500MR1 00000262154MR8 00000262659MR6 00000263947MR4 00000277774MR6 00000273444MR0 00000273297MR2 00000281290MR7 00000264455MR7 00000291887MR8 00000285430MR5 00000274106MR4 00000280695MR8 00000285431MR3 00000260674MR7 00000254195MR1 00000274053MR8 00000233435MR7 00000262103MR5 00000234844MR9 00000234798MR7 00000236829MR8 00000261384MR2 00000260468MR4 00000231899MR6 00000260859MR4 00000253831MR2 00000284078MR3 00000257425MR9 00000258410MR0 00000233301MR1 00000259423MR2 00000234862MR1 00000260019MR5 00000276420MR7 00000254190MR2 00000258365MR6 00000262594MR5 00000283363MR0 00000242546MR0 00000259525MR4 00000247224MR9 00000238169MR7 00000252409MR8 00000251715MR9 00000239236MR3 00000267852MR2 00000281264MR2 00000281859MR9 00000255607MR4 00000232426MR7 00000262443MR5 00000276352MR2 00000276268MR0 00000276651MR7 00000290516MR4 00000238173MR9 00000282481MR1 00000263498MR8 00000259567MR6 00000238864MR3 00000258078MR5 00000260691MR1 00000260481MR7 000A0000012MR2 00000273265MR9 00000262066MR4 00000263969MR8 00000271173MR7 00000258099MR1 00000250022MR1 00000280418MR5 00000269437MR0 00000264125MR6 00000271054MR9 00000280096MR9 00000280465MR6 00000281886MR2 00000280433MR4 00000265177MR6 00000265870MR6 00000264407MR8 00000274691MR5 00000268285MR4 00000265173MR5 00000234848MR0 00000272348MR4 00000258416MR7 00000266713MR7 00000273196MR6 00000284315MR9 00000270975MR6 00000273228MR7 00000270569MR7 00000264007MR6 00000276732MR5 00000262068MR0 00000265184MR2 00000258393MR8 00000234851MR4 00000280892MR1 00000269044MR4 00000245298MR5 00000280464MR9 00000263572MR0 00000274998MR4 00000281796MR3 00000280674MR3 00000278196MR1 00000249272MR6 00000275015MR6 00000258370MR6 00000262901MR2 00000262442MR7 00000272339MR3 00000275595MR7 00000270244MR7 00000281935MR7 00000270383MR3 00000278626MR7 00000270250MR4 00000262459MR1 00000282909MR1 00000263942MR5 00000283074MR3 00000259572MR6 00000262444MR3 00000274891MR1 00000279922MR9 00000269326MR5 00000282548MR7 00000281267MR5 00000257749MR2 00000265966MR4 00000267808MR4 00000261062MR4 00000273517MR3 00000281020MR8 00000280954MR9 00000262669MR5 00000267653MR4 00000261683MR7 00000282105MR6 00000271250MR3 00000257420MR0 00000283998MR3 00000262904MR6 00000285302MR6 00000275355MR6 00000281346MR7 00000280724MR6 00000279502MR9 00000274997MR6 00000266463MR9 00000263834MR4 00000266344MR1 00000273536MR3 00000281185MR9 00000264402MR9 00000260672MR1 00000281312MR9 00000262067MR2 00000281456MR4 00000237947MR7 00000238346MR1 00000270023MR5 0000000M3000PT 00000272629MR7 00000264744MR4 00000026259MR2 00000271052MR3 00000280234MR6 00002700253MR8 00000283663MR3 00000263314MR7 00000278247MR2 00000273487MR9 00000269038MR6 00000274072MR8 00000269438MR8 00000270459MR1 00000280056MR3 00000280009MR2 00000270218MR1 00000282253MR4 00000268288MR8 00000267111MR3 00000263655MR3 00000263684MR3 00000265801MR1 00000280276MR7 00000279925MR2 00000279887MR4 00000264783MR2 00000266194MR0 00000269343MR0 00000267657MR5 00000278472MR6 00000263977MR1 00000280085MR2 00000281445MR7 00000274835MR8 00000278722MR4 00000267468MR7 00000264093MR6 00000275097MR4 00000269377MR8 00000273459MR8 00000272621MR4 00000263880MR7 00000264406MR0 00000280735MR2 00000259380MR4 00000277185MR5 00000266753MR3 00000271865MR8 00000272430MR0 00000277489MR1 00000267804MR3 00000266461MR3 00000281774MR0 00000283316MR8 00000269384MR4 00000275060MR2 00000275469MR5 00000280385MR6 00000265722MR9 00000275020MR6 00000264554MR7 00000262462MR5 00000239530MR9 00000281889MR6 00000257745MR0 00000268286MR2 00000276728MR3 00000282859MR8 00000282032MR2 00000267978MR5 00000230695MR9 00000266068MR6 00000282695MR6 00000257421MR8 00000269324MR0 00000261677MR9 00000269040MR2 00000280046MR4 00000261867MR6 00000260650MR7 00000265612MR2 00000265965MR4 00000238366MR9 00000277963MR5 00000255651MR2 00000250752MR3 00000252932MR9 00000272271MR8 00000262075MR5 00000234792MR0 00000271663MR7 00000259917MR3 00000283452MR1 00000257742MR7 00000265061MR2 00000271127MR3 00000271083MR8 00000283468MR7 00000265592MR6 00000267805MR0 00000264403MR7 00000263266MR9 00000283345MR7 00000235586MR5 00000284360MR5 00000287579MR7 00000271234MR7 00000278727MR3 00000277939MR5 00000269456MR0 00000275959MR5 00000265914MR2 00000273365MR7 00000264458MR1 00000279628MR2 00000278245MR6 00000012158YR2 00000272341MR9 00000263948MR2 00000272343MR5 00000275846MR4 00000270385MR8 00000245401MR5 00000270388MR2 00000230687MR6 00000234577MR5 00002686687MR1 00000280435MR9 00000270607MR5 00000267208MR7 00000282830MR9 00000283077MR6 00000230696MR7 00000269576MR5 00000260099MR7 00000279500MR3 00000266677MR4 00000263935MR9 00000280057MR1 00000273200MR6 00000283653MR4 00000260649MR9 1 00000227408MRI 00000264599MR2 00000269459MR4 00000227571MR7 00000280698MR2 00000267974MR4 00000257743MR5 00000277607MR8 00000283437MR2 00000281681MR7 00000278622MR6 00000272529MR9 00000260098MR9 00000280620MR6 00000239523MR4 00000270627MR3 00000274711MR1 00000267040MR4 00000281022MR4 00000273621MR3 00000267211MR1 00000257744MR3 00000264408MR6 00000265610MR6 00000274713MR7 00000281314MR5 00000281273MR3 00000270550MR7 00000269342MR2 00000261477MR4 00000273341MR8 00000269382MR8 00000272105MR8 00000273368MR1 00000280274MR2 00000266156MR9 00000272962MR2 00000265616MR3 00000261736MR3 00000270321MR3 00000278768MR7 00000278076MR5 00000276486MR8 00000250454MR6 00000266822MR6 00000263503MR5 00000279798MR3 00000283665MR8 00000278875MR0 00000279755MR3 00000266011MR6 00000269460MR2 00000276152MR6 00000276414MR0 00000273230MR3 00000272406MR0 00000262800MR6 00000269574MR0 00000269457MR8 00000283365MR5 00000230691MR8 00000277599MR7 00000267811MR8 00000262665MR3 00000281265MR9 00000027862MR0 00000283530MR4 00000266668MR3 00000270625MR7 00000268478MR5 00000269042MR8 00000265182MR6 00000273367MR3 00000284786MR1 00000262668MR7 00000264401MR1 00000266747MR5 00000271246MR1 00000275937MR1 00000274937MR2 00000287317MR2 00000267464MR6 00000283624MR5 00000273527MR2 00000241103MR1 00000267181MR6 00000266838MR2 00000274716MR0 00000270971MR5 00000267469MR5 00000243180MR7 00000265687MR4 00000277862MR9 00000251722MR5 00000270243MR9 00000269378MR6 00000234841MR5 0 00000278726MR5 00000281227MR9 00000277550MR0 00000275086MR7 00000266857MR2 00000270968MR1 00000277965MR0 00000280094MR4 00000281347MR5 00000272077MR9 00000270461MR7 00000276330MR8 00000280702MR2 347400 00000263660MR3 00000270011MR0 00000262196MR9 00000263973MR0 00000262841MR0 00000279076MR4 00000270629MR9 00000274717MR8 00000267210MR3 00000273221MR2 00000276594MR9 00002181458MR0 891011 00000265181MR8 00000266009MR0 00000277949MR4 00000263971MR4 00000267533MR8 00000026584MR6 00000268688MR9 00000026572MR1 00000269577MR3 00000276735MR8 00000271209MR9 00000024927MR2 00000274931MR5 00000274170MR0 00000282463MR9 00000286406MR4 00000266827MR5 00000282524MR8 00000281077MR8 00000277788MR6 00000271118MR2 00000276592MR3 00000263257MR8 00000262666MR1 00000255598MR5 00000027888MR3 00000254663MR8 00000266067MR8 00000273680MR9 00000253826MR2 00000268367MR0 00002675655MR9 00000262673MR7 00000269458MR6 00000027242MR4 00000268365MR4 00000267803MR5 00000272531MR5 00000264492MR0 00000225375MR5 00000270392MR4 00000265917MR5 00000267649MR2 00000261201MR8 00000277496MR6 00000275023MR0 00000259963MR7 00000277959MR3 00000271123MR2 00000270247MR0 00000273528MR0 00000268363MR9 00000249762MR6 00000283151MR9 00000280278MR3 00000272342MR7 00000283473MR7 00000279480MR8 00000243696MR2 00000274930MR7 00000277192MR1 00000272103MR3 00000278619MR2 00000237067MR4 00000278246MR4 00000278365MR2 00000277700MR1 00000277923MR9 00000274889MR5 00000269323MR2 00000270008MR6 00000274933MR1 00000245299MR3 00000249931MR7 00000257373MR1 00000262312MR2 00000273295MR6 00000283762MR3 00000282849MR9 00000283471MR1 00000266673MR3 00000249760MR0 00000273514MR0 00000276737MR4 00000278561MR6 00000252441MR1 00000275669MR0 00000273458MR0 00000274939MR8 00000287086MR3 00000282812MR7 00000268324MR1 00000272892MR1 00000267178MR2 00000270460MR9 00000260009MR6 00000279147MR3 00000282492MR8 00000281457MR2 00000267321MR8 00000261066MR5 00000266672MR5 00000271124MR0 00000275920MR7 00000234845MR6 00000249227MR0 00000277955MR1 00000272286MR6 00000280087MR8 2333 00000245861MR0 00000267527MR0 000WNJ07102006 00000281526MR4 00000285664MR9 00000273460MR6 00000281776MR5 00000270249MR6 00000270969MR9 00000253637MR3 104 00000279607MR6 00000264797MR2 00000260696MR0 00000260015MR3 00000266163MR5 00000281703MR9 00000281188MR3 00000281488MR7 00000271269MR3 00000271647MR0 00000280374MR0 00000243186MR4 00000254667MR9 00000264087MR6 00000277188MR9 00000267322MR6 00000286309MR0 00000243182MR3 00000251056MR8 00000263874MR0 00000281442MR4 00000274839MR0 00000267467MR9 00000272620MR6 00000265007MR5 00000268691MR3 00000275597MR3 00000281829MR2 00000272895MR4 00000247222MR3 00000274709MR5 00000275474MR5 00000271268MR5 00000266837MR4 00000287063MR2 00000277193MR9 00000282792MR1 00000262504MR4 00000259584MR1 00000271526MR6 00000267472MR9 00000278623MR4 00000280069MR6 00000275062MR8 00000243251MR6 00000286391MR8 00000000BYV7UN 00000282107MR2 00000235243MR3 00000266755MR8 00000279885MR8 00000267470MR3 00000282106MR4 00000254492MR2 00000278244MR9 00000281644MR5 00000283800MR1 00000282395MR3 00000278871MR9 00000279499MR8 00000280088MR6 00000262930MR1 00000265919MR1 00000266756MR6 00000267114MR7 00000266674MR1 00000267213MR7 00000269381MR0 00000269462MR8 00000269385MR1 00000269579MR9 00000285951MR0 00000274838MR2 00000273462MR2 00000273675MR9 00000276415MR7 00000275936MR3 00000269043MR6 00000272963MR0 00000275938MR9 00000281455MR6 00000266450MR6 00000262297MR5 00000263871MR6 00000278292MR8 00000281605MR6 00000267648MR4 00000262051MR6 00000280321MR1 00000279450MR1 00000256983MR8 00000276249MR0 00000280301MR3 00000277492MR5 00000271055MR6 00000290712MR9 00000272628MR9 00000266764MR0 00000266007MR4 00000259979MR3 00000249277MR5 00000275344MR0 00000255614MR0 00000267646MR8 00000261870MR0 00000272347MR6 00000269483MR4 00000261197MR8 00000264404MR5 00000272104MR1 00000274079MR3 00000282857MR2 00000283506MR4 00000264095MR1 00000265588MR4 00000280095MR1 00000271121MR6 00000282961MR2 00000267651MR8 00000281266MR7 00000286425MR4 00000273008MR3 00000268119MR5 00000273679MR1 00000270007MR8 00000281840MR9 00000262457MR5 00000283435MR6 00000280434MR2 00000279799MR1 00000278131MR8 00000241097MR5 00000266216MR1 00000295102MR8 00000286792MR7 00000012078YR2 00000278881MR0 00000279149MR9 00000011724YR2 00000011569YR1 00000272490MR4 00000001159YR1 00000295216MR6 00000297879MR9 00000300910MR7 00000280363MR3 00000283952MR0 00000294319MR9 00000287908MR8 00000922BD5C3B 00000238178MR8 00000289202MR4 00000244851MR2 00000289023MR4 00000262906MR1 00000287383MR4 00000288100MR1 00000011862YR0 00000026188MR7 00000289340MR2 00000296303MR1 00000295233MR1 00000286305MR8 00000299327MR7 00000250210MR2 00000281340MR0 00000280947MR3 00000260868MR5 00000294554MR1 00000243184MR9 00000243760MR6 00000298504MR2 00000298223MR9 00000301340MR6 00000297185MR1 00000286528MR5 00000267859MR7 00000290056MR1 00000011927YR1 00000280253MR6 00000293343MR0 00000234576MR7 00000010705YR2 00000233303MR7 00000281841MR7 00000279608MR4 00000298930MR9 00000286307MR4 00000007498YR9 00000298945MR7 00000245327MR2 000000279162MR 00000239539MR0 00000833BD4ADE 00000267324MR2 00000006740YR5 00000284575MR8 00000278819MR8 00000815BD3A4A 00000237069MR0 00000930BD5D3B 00000009444YR1 00000283604MR7 00000273224MR6 00000011926YR3 00000028261MR4 00000283606MR2 00000295058MR2 00000006736YR3 00002891887MR7 00000011954YR5 00000246613MR4 00000251711MR8 00000011879YR4 00000295218MR2 00000012194YR7 00000294323MR1 00000298947MR3 00000285499MR0 00000293047MR7 00000262258MR7 00000011947YR9 00000296848MR5 00000289818MR7 00000012234YR1 00000286620MR0 00000HDML10300 00000297371MR7 00000247227MR2 00000286306MR6 00000011571YR7 00000268011MR4 00000012239YR0 00000295140MR8 00000295810MR6 00000295593MR8 00000298946MR5 00000244844MR7 00000294696MR0 00000011867YR9 00000294557MR4 00000011595YR6 00000029464MR7 00000285725MR8 00000283817MR5 00000287417MR0 00000284263MR1 00000280251MR0 00000295191MR1 00000011786YR1 00000922BD5C37 00000011887YR7 00000290113MR0 00000919BD5BBB 00000011203YR7 00000HDMR20258 00000283994MR2 00000276417MR3 00000296223MR1 00000012241YR6 00000288373MR4 00000278041MR9 00000289585MR2 00000011730YR9 00000270674MR5 00000300496MR7 151545 00000HDMR10743 00000280231MR2 00000011746YR5 00000011741YR6 00000247730MR5 00000287569MR8 00000012180YR6 00000011878YR6 00000256225MR4 00000009576YR0 00000282957MR0 00000012237YR4 00000295876MR7 00000011502YR2 00000302186MR2 00000012275YR4 00000252926MR1 00000258084MR3 00000297475MR6 00000262596MR0 00000292980MR0 00000012306YR7 00000281692MR4 00000012351YR3 00000276411MR6 00000290001MR7 00000277443MR8 00000273361MR6 00000011588YR1 00000028867MR3 00000238351MR1 00000289164MR6 00000288121MR7 00000287256MR2 00000288461MR7 00000285383MR6 00000012042YR8 00000285221MR8 00000007582YR0 00000266346MR6 00000287935MR1 00000286885MR9 00000238171MR3 00000008551YR4 00000263505MR0 00000287489MR9 00000263970MR6 00000268302MR7 00000239859MR2 00000274829MR1 00000233432MR4 00000011798YR6 00000286699MR4 00000288590MR3 00000257730MR2 00000281943MR1 00000918BD5B50 00000012163YR2 00000284535MR2 00000289915MR1 00000011596YR4 00000297488MR9 00000282581MR8 00000012341YR4 00000287566MR4 00000267113MR9 00000266043MR9 00000236830MR6 00000288410MR4 00000273515MR7 00000264743MR6 00000283301MR0 00000286410MR6 00000248479MR8 00000948BD61E8 00000294272MR0 00000299309MR5 00000010512YR2 00000300784MR6 00000297698MR3 00000281606MR4 00000288850MR1 00000267039MR6 1 00000289775MR9 00000262899MR8 00000264092MR8 00000282252MR6 00000267656MR7 00000267666MR7 00000011541YR0 00000280579MR4 00000243686MR3 00000286743MR0 00000009005YR0 00000288344MR5 00000281342MR6 00000278772MR9 00000287069MR9 00000902BD55AD 00000300951MR1 00000250455MR3 00000294553MR3 00000298313MR8 00000239854MR3 00000282817MR6 00000276491MR8 00000302258MR9 00000295743MR9 00000266066MR0 00000919BD5B77 00000275359MR8 00000288066MR4 00000270246MR2 00000012196YR2 00000282104MR9 00000298952MR3 00001101XF50B9 00000286985MR7 00000267112MR1 00000011501YR4 00000276495MR9 00000260451MR0 00000012229YR1 00000291169MR1 00000287078MR0 00000289956MR5 00000253632MR4 00000248169MR5 00000006757YR9 00000295806MR4 00000232411MR9 00000286507MR9 00000245672MR1 00000011219YR3 00000245676MR2 00000HDMR10762 00000229480MR9 00000277780MR3 00000011774YR7 00000293007MR1 00000262802MR2 00000007858YR4 00000279424MR6 00000239038MR3 11797 00000295025MR1 00000262935MR0 00000012387YR7 00000251721MR7 00001101XF50AE 00000247721MR4 00001101XF5099 00000010599YR9 00000278851MR1 00000295027MR7 00000288704MR0 00000012222YR6 00000288980MR6 00000298490MR4 00000293293MR7 00000231889MR7 00000288194MR4 00000012388YR5 00000011584YR0 00000289395MR6 00000270251MR2 00000288071MR4 00000293276MR2 00000286964MR2 00000282795MR4 00000283776MR3 00000286865MR1 00000266065MR2 00000266097MR5 00000252408MR0 00001011BD6684 00000284284MR7 00000011950YR3 00000301289MR5 00000292823MR2 00000286382MR7 00000298792MR3 00000011784YR6 00000299660MR1 00000302505MR3 00000247731MR3 00000286429MR6 00000011897YR6 00000284025MR4 00000286558MR2 00000012020YR4 00000294273MR8 00000011744YR0 00000286385MR0 00000281861MR5 00000298220MR5 00000283520MR5 00000287255MR4 00000292432MR2 00000282868MR9 00000281604MR9 00000287808MR0 00000288592MR9 00000294523MR6 00000011920YR6 00000286634MR1 00000286544MR2 00000287899MR9 00000257430MR9 00000011900YR8 00000289331MR1 00000298496MR1 00000282731MR9 00000285727MR4 00000297370MR9 00000285491MR7 00000253170MR5 00000262299MR1 00000236832MR2 00000284037MR9 00000282270MR8 00000288593MR7 00000288690MR1 00000294304MR1 00000287082MR2 00000253149MR9 00000282394MR6 00000273220MR4 00000298974MR7 00000299166MR9 00000239520MR0 00000297419MR4 00000296233MR0 00000275022MR2 00000279918MR7 00000301847MR0 00000279613MR4 00000287870MR0 00000238349MR5 00000285497MR4 00000296100MR1 00000297852MR6 00000916BD5B05 00000299911MR8 00000283845MR6 00000233302MR9 00000919BD5B9D 00000264090MR2 00000283864MR7 00000288123MR3 00000285965MR0 00000295512MR8 00000011958YR6 00000012236YR6 00000260471MR8 00000011892YR7 00000298413MR6 00000288825MR3 00000289792MR4 00000282828MR3 00000254672MR9 00000011885YR1 00000299762MR5 00000289913MR6 00000264091MR0 00000289533MR2 1 00000012063YR4 00000294444MR5 00000296104MR3 00000298936MR6 00000261627MR4 00000286059MR1 00000285693MR8 00000298809MR5 00000276722MR6 00000285966MR8 00000289531MR6 00000012181YR4 00000295518MR1 00000293317MR4 00000013657YR2 00000284576MR6 00000006592YR0 00000289777MR5 00000919BD5B97 00000276489MR2 00000011901YR6 00000238873MR4 00000284234MR2 00000012303YR4 00000012264YR8 00000288771MR9 00000297011MR9 00000011726YR7 00000267622MR9 00000561695MR3 00000011941YR2 00000297205MR7 00000289764MR3 00000231891MR3 00000289197MR6 00000299850MR8 00000288383MR3 00000287476MR6 00000006669YR6 00000286855MR2 00000012240YR8 00000916BD5B07 00000011953YR7 00000276245MR8 00000297627MR2 00000010722YR7 00000236924MR7 00000228285MR3 00000296367MR6 00000295176MR2 000000HDMR3130 00000006734YR8 00000243420MR7 00000298497MR9 00000011718YR4 00000288190MR2 00000292792MR9 00000284235MR9 00001101XF50A1 00000288152MR2 00000262298MR3 00000286765MR3 00000289201MR6 00000288064MR9 00000236504MR7 00000261476MR6 00000011722YR6 00001003BD6432 00000012189YR7 00000283996MR7 00000289163MR8 00000297369MR1 00000298710MR5 00000237702MR6 00000233270MR8 00000293342MR2 00000283402MR6 00000278848MR7 00000276586MR5 00000293324MR0 00000256208MR0 00000230111MR7 00000281869MR8 00000268689MR7 00000268290MR4 00000286927MR9 00000011787YR9 000000HDMR2904 00000296575MR4 00000277763MR9 00000241098MR3 00000283075MR0 00000289024MR2 00000287438MR6 00000008398YR0 00000288649MR7 00000283022MR2 00000011518YR8 00000294945MR1 00000275050MR3 00000286923MR8 00000286907MR1 00000274307MR8 00000248481MR4 00000280388MR0 00000289441MR8 00000282789MR7 00000011886YR9 00000235702MR8 00000288723MR0 00000280275MR9 00000283019MR8 00000288223MR1 00000012080YR8 00000288720MR6 00000005538YR4 00000287554MR0 00000282948MR9 00001048XF500B 00000279999MR7 00000011572YR5 00000267325MR9 00000286967MR5 00000011996YR6 00000011869YR5 00000281775MR7 00000279562MR3 00000286893MR3 00000274696MR4 00000294738MR0 00000011581YR6 00000905BD566C 00000295741MR3 00000273678MR3 00000287079MR8 00000265587MR6 00000275049MR5 00000296224MR9 00000285751MR4 00000011524YR6 00000283467MR9 00000296631MR5 00000294320MR7 00000280998MR6 00000278679MR6 00000297516MR7 00000011890YR1 00000302526MR9 00000287271MR1 000S0933BD5E02 00000270564MR8 00000012402YR4 00000286448MR6 00000294994MR9 00000232424MR2 00000288703MR2 00000286948MR5 00000280007MR6 00000276630MR1 00000247720MR6 00000295173MR9 00000288722MR2 0 00000299471MR3 00000235692MR1 00000295026MR9 00000248488MR9 00000284211MR0 00000296111MR8 00000284055MR1 00000289189MR3 00000278607MR7 00000275649MR2 00000281825MR0 00000289015MR0 00000279412MR1 00000012133YR5 00000231897MR0 00000918BD5B54 00000288346MR0 00000281941MR5 00000012230YR9 00000294144MR1 00000298826MR9 00000284283MR9 00000294990MR7 00000278313MR2 00000294648MR1 00000275650MR0 00000287286MR9 00000266205MR4 00001001BD6328 00000273396MR2 00000270010MR2 00000HDMR20063 00000009525YR7 00000231904MR4 00000279463MR4 00000276593MR1 00000266783MR0 00000240102MR4 00000290141MR1 00000287849MR4 00000251724MR1 00000286947MR7 00000288757MR8 00000011086YR6 00000256237MR9 00000291979MR3 00000235674MR9 00000282082MR7 00000289203MR2 00000271251MR1 00000301818MR1 00000287551MR6 00000011934YR7 00000260480MR9 00000287880MR9 00000300498MR3 00000283514MR8 00000292820MR8 00000274888MR7 00000284715MR0 00000289789MR0 00000265716MR1 00000286018MR7 00000011563YR4 00000027552MR4 00000286560MR8 00000297700MR7 00000281258MR4 2102 00000239242MR1 00000289627MR2 00000028654MR9 00000298428MR4 00000288097MR9 00000263974MR8 00000297383MR2 00000298011MR8 00000282689MR9 00000282863MR0 00000283818MR3 00000283821MR7 00000294118MR5 00000930BD5D3C 00000278102MR9 00000012003YR0 00000284391MR0 00000289042MR4 00000282683MR2 00000263968MR0 00000284563MR4 00000283523MR9 00000278817MR2 00000283175MR8 00000012224YR2 00000919BD5BAA 00000302523MR6 00000273524MR9 00000235675MR6 00000289332MR9 00000288431MR0 00000278559MR0 00000269061MR8 00000287623MR3 00000283027MR1 00000265000MR0 00000295789MR2 00000289218MR0 00000294737MR2 00000011591YR5 00000299882MR1 00000282794MR7 00000288245MR4 00000012385YR1 00001101XF50A4 00000289335MR2 00000280670MR1 00000286492MR4 00000288721MR4 00000288224MR9 00000275670MR8 00000011224YR3 00000249771MR7 00000289347MR7 00000262667MR9 00000232413MR5 00000283434MR9 00000287553MR2 00000282996MR8 00000283126MR1 00000286094MR8 00000295807MR2 00000008380YR8 00000294605MR1 00000283699MR7 00000277491MR7 00000299640MR3 00000282829MR1 00000299422MR6 00000276266MR4 00000011946YR1 00000285754MR8 00000286850MR3 00000286009MR6 00000286433MR8 00000012232YR5 00000283847MR2 00000286776MR0 00000287461MR8 00000263875MR7 00000288004MR5 00000266203MR9 00000025604MR9 00000295898MR1 00000298502MR6 00000287616MR7 00000292290MR4 00000011884YR4 00000283639MR3 00000230108MR3 00000011578YR2 00000235686MR3 00000288110MR0 00000008550YR6 00000238183MR8 00000251229MR1 00000279618MR3 00000286744MR8 00000289027MR5 00000285055MR0 00000010964YR5 00000301615MR1 00000294519MR4 00000289761MR9 00000298976MR2 00000298187MR6 00000240106MR5 00000275463MR8 00000281228MR7 00000300984MR2 00000236702MR7 00000230099MR4 00000288822MR0 00000272897MR0 00000281485MR3 00000277919MR7 00000235681MR4 00000011547YR7 00000261385MR9 00000242180MR8 00000230107MR5 00000261941MR9 00000284903MR2 00000028606MR1 00000012406YR5 00000289762MR7 00000279415MR4 00000011883YR6 00000011504YR8 00000919BD5BC4 00000267529MR6 00000298468MR0 00000283400MR0 00000276435MR5 00000296830MR3 00000275646MR8 00000235238MR3 00000261378MR4 00000011721YR8 00000011801YR8 00000295899MR9 00000283021MR4 00000279614MR2 00000265723MR7 00000013090YR6 00000288023MR5 00000288775MR0 00000011269YR8 00000284458MR7 00000012065YR9 00000012188YR9 00000279914MR6 00000288161MR3 00000273673MR4 00000295701MR7 00000287898MR1 00000296364MR3 00000012047YR7 00000245817MR2 00000237931MR1 00000288159MR7 00000279882MR5 00000298968MR9 00000012223YR4 00000012225YR9 00000283679MR9 00000277186MR3 00000283364MR8 00000281939MR9 00000268512MR1 00000281862MR3 00000287939MR3 00000289525MR8 00000012017YR0 00000278164MR9 00000287855MR1 00000294892MR5 00000011939YR6 00000285724MR1 00000284275MR5 00000294402MR3 00000249929MR1 00000287819MR7 00000283605MR4 00000283040MR4 00000294668MR9 00000011877YR8 00000288061MR5 00000278770MR3 00000283859MR7 00000294342MR1 00000285382MR8 00000275962MR9 00000295476MR6 00000283173MR3 00000299523MR1 00000294959MR2 00000289261MR0 00000289271MR9 00000295052MR5 00000286895MR8 00000278129MR2 00000237945MR1 00000288191MR0 00000288020MR1 00000278816MR4 00000028286MR6 00000241724MR4 00000288341MR1 00000012193YR9 00000295139MR0 00000238358MR6 00000260697MR8 00000294409MR8 00000299449MR9 00000919BD5BA9 00000297447MR5 00000302624MR2 00000289318MR8 00000023524MR8 00000282340MR9 00000012167YR3 00000288377MR5 00000233622MR0 00000234840MR7 00000279070MR7 00000265552MR0 00000287878MR3 00000298810MR3 00000286587MR1 00000285490MR9 00000012198YR8 00000287617MR5 00000236835MR5 00000012185YR5 00000287656MR3 00000012157YR4 00000288021MR9 00000294394MR2 00000247460MR9 00000247725MR5 00000294442MR9 00000261869MR2 00000001206YR5 00000294327MR2 00000288342MR9 00000294158MR1 00000295742MR1 00000298716MR2 00000298657MR8 00000300863MR8 00000237701MR8 00000239028MR4 00000257374MR9 00000266449MR8 00000268419MR9 00000268282MR1 00000276589MR9 00000278167MR2 00000010140YR2 00000282523MR0 00000285796MR9 00000286998MR0 00000919BD5BB5 00000011511YR3 00000289998MR7 00000298044MR9 00000292315MR9 00000266468MR8 00000011575YR8 00000300904MR0 00000299350MR9 00000298687MR5 00000287415MR4 00000011056YR9 00000235682MR2 00000283408MR3 00000011956YR0 00000263255MR2 00000281154MR5 00000262804MR8 00000296410MR4 00000299527MR2 00000298708MR9 00000012161YR6 00000922BD5C38 00000301290MR3 00000287490MR7 00000296441MR9 00000011796YR0 00000287600MR1 00000289205MR7 00000012192YR1 00000296301MR5 00000289160MR4 00000299758MR3 00000281933MR2 00000295088MR9 00000012272YR1 00000012072YR5 00000011544YR4 00000287588MR8 00000011743YR2 00000011873YR7 00000280891MR3 00000009042YR3 00000299525MR6 00000286409MR8 00000274161MR9 00000012132YR7 00000012070YR9 00000272290MR8 00000287400MR6 00000292474MR4 00000011931YR3 00000012231YR7 00000280001MR9 00000009636YR2 00000276718MR4 00000301323MR2 00000243687MR1 00000236834MR8 00000008549YR8 00000297893MR0 00000262672MR9 00000280391MR4 00000263336MR0 00000298709MR7 00000295235MR6 00000286946MR9 000000HDMR2892 00000286966MR7 00000012346YR3 00000296519MR2 00000285626MR8 00000287062MR4 00000261061MR6 00000287391MR7 00000294160MR7 00000289014MR3 00000289262MR8 00000298429MR2 00000241721MR0 00000297382MR4 00000300923MR0 00000012006YR3 00000293299MR4 00000264221MR3 00000283528MR8 00000287363MR6 00000280767MR5 00000284693MR9 00000283622MR9 00000237481MR6 00000288255MR3 00000011932YR1 00000289336MR0 00000011555YR0 00000298949MR9 00000284785MR3 00000299662MR7 00000286906MRS 00000284690MR5 00000011955YR2 00000292901MR6 00000299424MR2 00000297486MR3 00000012015YR4 00000287117MR6 00000287479MR0 00000011717YR6 00000301299MR4 00000288098MR7 00000011506YR3 00000011551YR9 00000302625MR9 00000831BD49EF 00000286898MR2 00000283997MR5 00000011804YR2 00000011874YR5 00000289438MR4 00000011727YR5 00000011875YR2 00000271084MR6 00000276151MR8 00000300608MR7 00000254664MR6 00000299642MR9 00000280461MR5 00000011586YR5 00000011338YR1 00000011549YR3 00000301349MR7 00000284534MR5 00000286478MR3 00000007882YR4 00000007887YR3 00000300517MR0 00000300937MR0 00000301301MR8 00000300811MR7 00000300581MR6 00000283654MR2 00000282007MR4 00000011938YR8 00000266129MR6 00000239537MR7 00000277860MR3 00000012050YR1 00000012010YR5 00000012008YR9 00000241434MR3 00000273537MR1 00000277917MR1 00000280364MR1 00000274692MR3 00000284472MR8 00000272334MR4 00000283812MR6 00000026005MR8 00000286783MR6 00000283515MR5 00000266343MR3 00000300902MR4 000000HDMR2946 00000028135MR2 00000239858MR4 00000251059MR2 00000279503MR7 00000274078MR5 00000271119MR0 00000287941MR9 00000295004MR6 00000294609MR3 00000302424MR7 00000294625MR9 00000288651MR3 00000253834MR6 00000009853YR3 00000007544YR0 00000269340MR6 00000235683MR0 00000008115YR8 00000245806MR5 00000261444MR4 00000280893MR9 00000295482MR4 00000287506MR0 00000261678MR7 00000272344MR3 00000299885MR4 00000028432MR3 00000296210MR8 00000273516MR5 00000301188MR9 000000HDMR2609 00000287596MR1 00000300619MR4 00000265553MR8 00000253157MR2 00000259568MR4 00000254193MR6 00000234853MR0 00000280950MR7 00000273253MR5 00000261329MR7 00000280949MR9 00000250205MR2 00000295229MR9 00000006586YR2 00000287315MR6 00000253838MR7 00000301501MR3 00000280946MR5 00000301105MR3 00000291943MR9 00000286334MR8 00000272562MR0 00000HDMR20269 00000293315MR8 00000287645MR6 00000284262MR3 00000257754MR2 00000302472MR6 00000252410MR6 00000296145MR6 00000270382MR5 00000026717MR0 00001104XF5011 00000300884MR4 00000291681MR5 00000261330MR5 00000261381MR8 00000265805MR2 00000264619MR8 00000300950MR3 00000011872YR9 00000006726YR4 00000300577MR4 00000280328MR6 00000280229MR6 00000295190MR3 00000288459MR1 00000284411MR6 00000298533MR1 00000274744MR2 00000262670MR3 00000009837YR6 00000302510MR3 00000300457MR9 00000299949MR8 00000300970MR1 00000287877MR5 00000299962MR1 00000300769MR7 00000300458MR7 00000294515MR2 00000289222MR2 00000300702MR8 00000GE0210033 00000249266MR8 00000251090MR7 00000256198MR3 00000259105MR5 00000301098MR0 00000275094MR1 00000267807MR6 00000272532MR3 00000298755MR0 00000288075MR5 00000289884MR9 00000298098MR5 00000919BD5B9F 00000286148MR2 00000011739YR0 00000011728YR3 00000270465MR8 00000301223MR4 00000295003MR8 00000290369MR8 00000288222MR3 00000260861MR0 00000274050MR4 00000280912MR7 00000290572MR7 00000288254MR6 00000288257MR9 00001011BD6688 00000287601MR9 00000012019YR6 00000274712MR9 00000278557MR4 00000237000MR5 00000234791MR2 00000259129MR5 00000257433MR3 00000231900MR2 00000242592MR4 00000264557MR0 00000258406MR8 00000230105MR9 00000256968MR9 00000241522MR2 00000251638MR3 00000235239MR1 00000296056MR5 00000267463MR8 00000267323MR4 00000255596MR9 00000282922MR4 00000232409MR3 00000259580MR9 00000268421MR5 00000285956MR9 00000267197MR2 00000275539MR5 00000287268MR7 00000286370MR2 00000012005YR5 00000295055MR8 00000011564YR2 00000012408YR1 00000265063MR8 00000269041MR0 00000268123MR7 00000283407MR5 00000284060MR1 00000011896YR8 00000912BD59B4 00000294305MR8 00000261871MR8 00000284594MR9 00000242176MR6 00000250196MR3 00000256686MR7 00000277762MR1 00000278672MR1 00000281153MR7 00000288815MR4 00000280761MR8 00000268761MR4 00000293334MR9 00000280725MR3 00000233294MR8 00000278548MR3 00000281487MR9 00000007581YR2 00000282526MR3 00000282856MR4 00000286321MR5 00000287552MR4 00000288211MR6 00000246611MR8 00000286892MR5 00000295541MR7 00000264748MR5 000000HDMR3834 00000265594MR2 00000283774MR8 00000254376MR7 00000919BD5B8E 00000281602MR3 00000283650MR0 00000298068MR8 00000264740MR2 00000294988MR1 00000287773MR6 00000283517MR1 00000274180MR9 00000290310MR2 00000011714YR3 00000262040MR9 00000012057YR6 00000292979MR2 00000011783YR8 00000012071YR7 00000268122MR9 00000286368MR6 00000011580YR8 00000289849MR2 00000260628MR3 00000276493MR4 00001011BD6629 00000029749MR3 00000298713MR9 00001052XF5067 00000287615MR9 000000HDMR5803 00000236707MR6 00000241726MR9 00000234859MR7 00000252664MR8 00000268317MR5 00000256209MR8 00000277609MR4 00000273293MR1 00000280092MR8 00000280694MR1 00000255602MR5 00000285432MR1 00000282611MR3 00000284077MR5 00000287509MR4 00000282392MR0 00000291406MR7 00000250452MR0 00000256210MR6 00000256213MR0 00000251020MR4 00000281517MR3 00000281518MR1 00000279478MR2 00000262891MR5 00000282022MR3 00000281698MR1 00000283118MR8 00000284361MR3 00000280705MR4 00000290039MR7 00000290603MR0 00000290005MR8 00000290548MR7 00000263253MR7 00000293009MR7 00000293031MR1 00000295840MR3 00000296628MR1 00000252662MR2 00000282681MR6 00000290547MR9 00000292214MR4 00000239232MR2 00000255931MR8 00000299516MR5 00000250211MR0 00000298072MR0 00000298382MR3 00000238870MR0 00000298840MR0 00000298579MR4 00000299960MR5 00000299562MR9 00000280373MR2 00000252658MR0 00000269482MR6 00000271238MR8 00000270972MR3 00000264558MR8 00000239044MR1 00000260102MR9 00000264747MR7 00000273488MR7 00000273533MR0 00000269572MR4 00000274172MR6 00000277864MR5 00000274938MR0 00000280626MR3 00000282525MR5 00000284717MR6 00000284596MR4 00000285026MR1 00000270973MR1 00000251871MR0 00000266160MR1 00000264742MR8 00000261679MR5 00000264088MR6 00000265866MR4 00000269379MR4 00000272335MR1 00000279924MR5 00000279890MR8 00000281257MR6 00000279869MR2 00000288196MR9 00000284643MR4 00000286778MR6 00000288204MR1 00000288774MR3 00000283893MR6 00000283404MR2 00000281828MR4 00000266158MR5 00000280691MR7 00000277642MR5 00000286925MR3 00000261441MR0 00000257431MR7 00000286930MR3 00000011230YR0 00000252665MR5 00000261590MR4 00000266345MR8 00000280860MR8 00000278800MR8 00000278554MR1 00000007890YR7 00000280921MR8 00000281057MR0 00000286160MR7 00000009575YR2 00000028686MR4 00000295181MR2 00000252377MR7 00000252406MR4 00000250220MR1 00000265591MR8 00000274934MR9 00000278398MR3 00000276680MR6 00000278673MR9 00000280064MR7 00000278211MR8 00000280673MR5 00000281315MR2 00000282512MR3 00000284237MR5 00000280230MR4 00000919BD5B94 00000285591MR4 00000286736MR4 00000011790YR3 00000011792YR9 00000287961MR7 00000011562YR6 00000288934MR3 00000287559MR9 00000012009YR7 00000288460MR9 00000289975MR5 00000011909YR9 00000289912MR8 00000288932MR7 00000289791MR6 00000011898YR4 00000011507YR1 00000289793MR2 00000011598YR0 00000011599YR8 00000011895YR0 00000011948YR7 00000012383YR6 00000012268YR9 00000294910MR5 00000012354YR7 00000296203MR3 00000012266YR3 00000296112MR6 00000294610MR1 00000294604MR4 00000294627MR5 00000294908MR9 00000296359MR3 00000296645MR5 00000296355MR1 00000295230MR7 00000295514MR4 00000297625MR6 00000296633MR1 00000298262MR7 00000296201MR7 00000296832MR9 00000298824MR4 00000298219MR7 00001007BD6496 00000299950MR6 00000006742YR1 00000008327YR9 00001043XF5005 00000300812MR5 00000280580MR2 00000298977MR0 00000299335MR0 00000256206MR4 00000300124MR5 00000300699MR6 00000300554MR3 00000302253MR0 00000299495MR2 00000261576MR3 00000257727MR8 00000275596MR5 00000289527MR4 00000287598MR7 00000270630MR7 00000233623MR8 00000263292MR5 00000233440MR7 00000300553MR5 00000285119MR4 00000010954YR6 00000274168MR4 00000239043MR3 00000238361MR3 00000243412MR4 00000247728MR9 00000291846MR4 00000294918MR8 00000288849MR3 00000274932MR3 00000277779MR5 00000239230MR6 00000248174MR5 00000258085MR0 00000237708MR3 00000233446MR4 00000243409MR0 00000253630MR8 00000256216MR3 00000246610MR0 00000266072MR8 00000245816MR4 00000279482MR4 00000256969MR7 00000256221MR3 00000255212MR3 00000264462MR3 00000274074MR4 00000259527MR0 00000270563MR0 00000252931MR1 00000270631MR5 00000265960MR5 00000266690MR7 00000262593MR7 00000239527MR5 00000271666MR0 00000260013MR8 00000241678MR2 00000283451MR3 00000247732MR1 00000247734MR7 00000276164MR1 00000252381MR9 00000276274MR8 00000239228MR0 00000284316MR7 00000276733MR3 00000299202MR2 00000239046MR6 00000295763MR7 00000277918MR9 00000249761MR8 00000250213MR6 00000241875MR4 00000012340YR6 00000286154MR0 00000011889YR3 00000282908MR3 00000278747MR1 00000267431MR5 00000272107MR4 00000279069MR9 00000270572MR1 00000259915MR7 00000298581MR0 00000232429MR1 00000011944YR6 00000814BD39ED 00000297630MR6 00000278606MR9 00000281793MR0 00000246614MR2 00000007090YR4 00000279909MR6 00000284641MR8 00000242181MR6 00000253156MR4 00000283347MR3 00000301499MR0 00000011861YR2 00000286426MR2 00000284356MR3 00000268525MR3 00000259378MR8 00000268528MR7 00000299538MR9 00000299621MR3 00000300573MR3 00000261206MR7 00000272432MR6 00000278243MR1 00000287918MR7 00000299184MR2 00000295483MR2 00000255613MR2 00000232414MR3 00000271525MR8 00000256986MR1 00000261379MR2 00000230110MR9 00000249273MR4 00000249278MR3 00000239860MR0 00000264009MR2 00000264998MR6 00000254491MR4 00000264015MR9 00000293676MR3 00000236505MR4 00000239041MR7 00000257438MR2 00000261473MR3 00000249764MR2 00000265826MR8 00000242551MR0 00000268527MR9 00000256967MR1 00000260667MR1 00000262661MR2 00000246265MR3 00000250201MR1 00000235679MR8 00000249765MR9 00000257437MR4 00000254670MR3 00000261471MR7 00000267110MR5 00000270522MR6 00000293715MR9 00000243408MR2 00000252431MR2 00000252429MR6 00000255201MR6 00000026066MR9 00000267106MR3 00000270515MR0 00000270523MR4 00000259123MR8 00000273018MR2 00000256224MR7 00000294526MR9 00000238872MR6 00000238865MR0 00000241719MR4 00000243767MR1 00000249766MR7 00000261442MR8 00000260866MR9 00000267851MR4 00000252920MR4 00000241864MR8 00000245808MR1 00000247722MR2 00000251801MR7 00000262900MR4 00000261327MR1 00000259127MR9 00000268134MR4 00000267107MR1 00000268321MR7 00000270517MR6 00000266801MR0 00000257444MR0 00000254488MR0 00000256694MR1 00000255207MR3 00000241102MR3 00000243766MR3 00000250438MR9 00000257725MR2 00000260670MR5 00000261386MR7 00000258101MR5 00000261380MR0 00000264006MR8 00000271648MR8 00000252435MR3 00000235703MR6 00000252663MR0 00000255198MR4 00000255226MR3 00000265095MR0 00000267109MR7 00000239533MR3 00000268135MR1 00000242555MR1 00000241722MR8 00000254201MR7 00000264997MR8 00000262733MR9 00000261573MR0 00000265091MR9 00000259574MR2 00000249279MR1 00000282686MR5 00000281704MR7 00000270384MR1 00000272336MR9 00000276731MR7 00000278842MR0 00000280768MR3 00000283470MR3 00000286392MR6 00000287810MR6 00000265969MR6 00000261200MR0 00000261065MR7 00000262931MR9 00000264741MR0 00000265179MR2 00000267977MR7 00000271126MR5 00000266147MR8 00000272973MR9 00000274166MR8 00000267037MR0 00000279159MR8 00000289004MR4 00000287879MR1 00000280302MR1 00000288901MR2 00000262803MR0 00000277194MR7 00000279751MR2 00000283386MR1 00000282501MR6 00000288667MR9 00000285172MR3 00000282791MR3 00000283526MR2 00000282790MR5 00000288244MR7 00000286559MR0 00000270974MR9 00000272893MR9 00000278749MR7 00000280898MR8 00000280951MR5 00000285380MR2 00000283761MR5 00000286800MR8 00000289798MR1 00000288811MR3 00000273229MR5 00000274996MR8 00000282088MR4 00000284787MR9 00000285222MR6 00000260625MR9 00000265590MR0 00000276635MR0 00000280005MR0 00000283378MR8 00000286538MR4 00000273676MR7 00000272894MR7 00000277583MR1 00000279802MR3 00000280621MR4 00000285303MR4 00000282831MR7 00000283174MR1 00000286897MR4 00000264089MR4 00000274688MR1 00000277487MR5 00000284644MR2 00000265593MR4 00000273223MR8 00000275091MR7 00000275598MR1 00000277189MR7 00000279752MR0 00000280319MR5 00000276370MR4 00000288589MR5 00000279458MR4 00000280299MR9 00000281936MR5 00000284457MR9 00000267809MR2 00000282858MR0 00000261198MR6 00000285219MR2 00000283662MR5 00000267650MR0 00000281317MR8 00000273535MR5 00000287590MR4 00000278566MR5 00000280002MR7 00000008560YR5 00000283076MR8 00000279423MR8 00000281603MR1 000000HDMR5528 00000288812MR1 00000294929MR5 00000296257MR9 00000301762MR1 00000011864YR6 00000291970MR2 00000011937YR0 00000012360YR4 00000301010MR5 00000011772YR1 00000012165YR7 00000012186YR3 00000295022MR8 00000291977MR7 00000295828MR8 00000292982MR6 00000297474MR9 00001008BD652A 00001009BD654A 00000298182MR7 00000298659MR4 00000300477MR7 00000301264MR8 00000301383MR6 00000301612MR8 00001046XF501A 00000267973MR6 00000277214MR3 00000289196MR8 00000011894YR3 00000012183YR0 00000011997YR4 00000012375YR2 00000288874MR1 00000919BD5BB3 00000292591MR5 00000294936MR0 00000300949MR5 00000296439MR3 00000294944MR4 00000297384MR0 00000297466MR5 00000298759MR2 00000297806MR2 00000299273MR3 00000235694MR7 00001011BD664D 00000299269MR1 00000301928MR8 00000301168MR1 00000302700MR0 00000VALUE1162 00000287630MR8 00000298928MR3 00000298670MR1 00000012184YR8 00000011590YR7 00000292974MR3 00000295878MR3 00000299191MR7 00000293587MR2 00000289315MR4 00000293499MR0 00000296225MR6 00000298328MR6 00000300182MR3 00000300701MR0 00000300729MR1 00000295020MR2 00000296093MR8 00000297420MR2 00000290325MR0 00000290488MR6 00000296634MR9 00000297385MR7 00000298059MR7 00000298672MR7 00000011923YR0 00000299163MR6 00000299313MR7 00000011952YR9 00000289026MR7 00000012358YR8 00000919BD5BA8 00000295531MR8 00000296458MR3 00000295561MR5 00000298680MR0 00000298911MR9 00000299946MR4 00000301398MR4 00000301806MR6 00000301929MR6 00000294328MR0 00000293650MR8 00000294547MR5 00000296094MR6 00000299506MR6 00000301930MR4 00000288936MR8 00000011928YR9 00000293298MR6 00000294920MR4 00000302426MR2 00000298935MR8 00000298403MR7 00000300859MR6 00000300183MR1 00000300123MR7 00000300682MR2 00000300617MR8 00000262664MR6 00000302474MR2 00000277190MR5 00000286479MR1 00000288701MR6 00000011574YR1 00000011871YR1 00000012315YR8 00000294284MR5 00000012304YR2 00000293593MR0 00000295529MR2 00000298467MR2 00000299448MR1 00000257738MR5 00000286434MR6 00000300887MR7 00000285353MR9 00000011558YR4 00000289591MR0 00000298314MR6 00000298155MR3 00000299164MR4 00000298793MR1 00000298969MR7 00000299311MR1 00000299192MR5 00000301187MR1 00000299964MR7 00000302356MR1 00000299539MR7 00000242552MR8 00000298329MR4 00000299312MR9 00000288428MR6 00000288099MR5 00000011876YR0 00000012190YR5 00000848BD5311 00000295013MR7 00000299265MR9 00000298774MR1 00000298221MR3 00000298858MR2 00000300879MR4 00000297725MR4 00000301104MR6 00000299167MR7 00000VALUE1179 00000294685MR3 00000299677MR5 00000298330MR2 00000299275MR8 00000300922MR2 00000296468MR2 00000012125YR1 00000012317YR4 00000012270YR5 00000293586MR4 00000294928MR7 00000295059MR0 00000300747MR3 00000295236MR4 00000294051MR8 00000299165MR1 00000299507MR4 00000301392MR7 00001104XF507C 00000302643MR2 00000289523MR3 00000293164MR0 00000293163MR2 00000919BD5BA1 00000296646MR3 00001101XF50A2 00000301498MR2 00000011738YR2 00000298846MR7 00000297202MR4 00000299328MR5 00000302679MR6 00000012352YR1 00000944BD6168 00000295528MR4 00000297014MR3 00000297658MR7 00000300770MR5 00000293297MR8 00000297728MR8 00000257734MR4 00000299643MR7 00000289039MR0 00000296442MR7 00000295219MR0 00000299349MR1 00000302473MR4 00000913BD5A47 00000011314YR2 00000302188MR8 00000011866YR1 00000301397MR6 00001003BD6415 00000300952MR9 00000302568MR1 00000302532MR7 00000298934MR1 00000GE0131001 00000GE0205029 00000GE0206027 00000249264MR3 00000257756MR7 00000291468MR7 00000290648MR5 00000263263MR6 00000267326MR7 00000271122MR4 00000273006MR7 00000273540MR5 00000276418MR1 00000282959MR6 00000298553MR9 00000296302MR3 00000298214MR8 00000299271MR7 00000299883MR9 00000263264MR4 00000262660MR4 00000298412MR8 00000268306MR8 00000259585MR8 00000286696MR0 00000011899YR2 00000278850MR3 00000011936YR2 00000294626MR7 00000296267MR8 00000297835MR1 00000229478MR3 00000238871MR8 00000265916MR7 00000265589MR2 00000231894MR7 00000253625MR8 00000266782MR2 00000272288MR2 00000294628MR3 00000279884MR1 00000278677MR0 00000011865YR3 00000247724MR8 00000279461MR8 00000011054YR4 00000228373MR7 00000288385MR0 00000265611MR4 00000250218MR5 00000267670MR8 00000287850MR2 00000289474MR9 00000265820MR1 00000264011MR8 00000294306MR6 00000299450MR7 00000298686MR7 00000282691MR5 00000266744MR2 00000283822MR5 00000269339MR8 000000HDMR4304 00000276163MR3 00000280777MR4 00000282950MR5 00000283348MR1 00000287019MR4 00000010970YR2 00000275095MR8 00000286896MR6 00000919BD5B8F 00000247719MR8 00000298094MR4 00000300615MR2 00000012199YR6 00000282855MR6 00000284901MR6 00000279616MR7 00000010066YR9 00000250442MR1 00000273833MR4 00000281867MR2 00000253158MR0 00000283350MR7 00000264618MR0 00000265869MR8 00000269325MR7 00000300750MR7 00000012345YR5 00000241872MR1 00000289773MR4 00000268121MR1 00000284757MR2 00000266467MR0 00000258091MR8 00000274836MR6 00000284238MR3 00000275538MR7 00000263339MR4 00000296833MR7 00000284648MR3 00000293318MR2 00000280922MR6 00000283933MR0 00000274073MR6 00000245821MR4 00000253837MR9 00000258414MR2 00000258368MR0 00000250202MR9 00000246621MR7 00000245408MR0 00000264999MR4 00000251798MR5 00000239521MR8 00000254497MR1 00000251643MR3 00000250449MR6 00000255200MR8 00000258100MR7 00000261591MR2 00000268473MR6 00000265920MR9 00000262065MR6 00000280432MR6 00000272340MR1 00000275645MR0 00000259125MR3 00000265060MR4 00000260695MR2 00000275584MR1 00000266144MR5 00000259842MR3 00000242547MR8 00000266646MR9 00000265172MR7 00000279726MR4 00000267182MR4 00000276250MR8 00000274890MR3 00000267212MR9 00000272078MR7 00000261675MR3 00000298568MR7 00000011540YR2 00000012000YR6 00000297851MR8 00000289334MR5 00000918BD5B42 00000299763MR3 00000265868MR0 00000284465MR2 00000272964MR8 00000011543YR6 00000286626MR7 00000026639MR2 00000287631MR6 00000012007YR1 00000283932MR2 00000012064YR2 00000283073MR5 00000283772MR2 00000284872MR9 00000012018YR8 00000276730MR9 00000267327MR5 00000279165MR5 00000283521MR3 00000283508MR0 00000279630MR8 00000286338MR9 00000288043MR3 00000012056YR8 00000295193MR7 00000289820MR3 00000922BD5C3A 00000294735MR6 00000298808MR7 00000279727MR2 00000289348MR5 00000289526MR6 00000289478MR0 00000012322YR4 00001101XF50A8 00000294552MR5 00000295804MR9 00000297822MR9 00000918BD5B38 00000293588MR0 00000297809MR6 00000300125MR2 00000011024YR7 00000250445MR4 000000HDMR3570 00000264643MR8 00000287964MR1 00000296305MR6 00000011945YR3 00000286849MR5 00000011929YR7 00000011998YR2 00001105XF5020 00000299526MR4 00000280732MR9 00000294658MR0 00000294114MR4 00000295182MR0 00000282269MR0 00000011800YR0 00000277441MR2 00000011582YR4 00000283745MR8 00000294117MR7 00000294661MR4 00000286286MR0 00000919BD5B90 00000905BD5675 00000237699MR4 00000286381MR9 00000297892MR2 00000012062YR6 00000012235YR8 00000012305YR9 00000288806MR3 00000011732YR5 00000283401MR8 00000273531MR4 00000027972MR4 00000296440MR1 00000298222MR1 00000011725YR9 00000011736YR6 00000012379YR4 00000297465MR7 00000299661MR9 00000288433MR6 00000012243YR2 00000289782MR5 00000297970MR6 00000289587MR8 00000294529MR3 00000919BD5BB1 00000297834MR4 00000272535MR6 00000297446MR7 00000918BD5B43 00000297853MR4 00000297444MR2 00000288669MR5 00000243183MR1 00000919BD5BA6 00000011120YR3 00000301138MR4 00000012269YR7 00000294343MR9 00000282910MR9 00000300886MR9 00000294919MR6 00000298281MR7 00000275934MR8 00000300673MR1 00000297660MR3 00000275969MR4 00000284688MR9 00000284384MR5 00000011514YR7 00000298070MR4 00000012122YR8 00000255210MR7 00000285496MR6 00000275047MR9 00000298929MR1 00000288109MR2 00000288586MR1 00000300726MR7 000000245811MR 00000301139MR2 00001002BD6349 00000288610MR9 00000262799MR0 00000278105MR2 00000268303MR5 00000007134YR0 00000281683MR3 000000HDMR2862 00000283698MR9 00000256205MR6 00000011445YR4 00001101XF50CE 00000256211MR4 00000243246MR6 00000267466MR1 00000254189MR4 00000294636MR6 00000258409MR2 00000250446MR2 00000262316MR3 00000255603MR3 00000253828MR8 00000264014MR2 00000264549MR7 00000229475MR9 00000237710MR9 00000264222MR1 00000239238MR9 00000247733MR9 00000253168MR9 00000242590MR8 00000245407MR2 00000267101MR4 00000238365MR1 00000246616MR7 00000251641MR7 00000237073MR2 00000258093MR4 00000242591MR6 00000272561MR2 00000245668MR9 00000230689MR2 00000232403MR6 00000232404MR4 00000262932MR7 00000278304MR1 00000268763MR0 00000256703MR0 00000266465MR4 00000270633MR1 00000279080MR6 00000238181MR2 00000262052MR4 00000266448MR0 00000262805MR5 00000274690MR7 00000271139MR8 00000274171MR8 00000265180MR0 00000233295MR5 00000265922MR5 00000274837MR4 00000265006MR7 00000263340MR2 00000265092MR7 00000242545MR2 00000261063MR2 00000250191MR4 00000011888YR5 00000272971MR3 00000012066YR7 00000296050MR8 00000283954MR6 00000012251YR5 00000287632MR4 00000284355MR5 00000297628MR0 00000281702MR1 00000286894MR1 00000284051MR0 00000296403MR9 00000295522MR7 00000278195MR3 00000278394MR2 00000283951MR2 00000012044YR4 00000011520YR4 00000011911YR5 00000295508MR6 00000286922MR0 00000233422MR5 00000012164YR0 00000011546YR9 00000297372MR5 00000283591MR6 00000296146MR4 00000239029MR2 00000287650MR6 00000294419MR7 00000298071MR2 00000275635MR1 00000281794MR8 00000297659MR5 00000253169MR7 00000277579MR9 00000011594YR9 00000295507MR8 00000295699MR3 00000265105MR7 00000288668MR7 00000294624MR2 00000299676MR7 00000252298MR5 00000300939MR6 00000286265MR4 00000264005MR0 00000299664MR3 00000298489MR6 00000255192MR7 00000241144MR5 00000250199MR7 00000288591MR1 00000265724MR5 000000HDMR3707 00000278312MR4 00000280690MR9 00000280578MR6 00000243768MR9 00000236626MR8 00000258082MR7 00000259578MR3 00000246624MR1 00000262907MR9 00000261572MR2 00000268239MR1 00000229473MR4 00000253623MR3 00000255615MR7 00000245406MR4 00000250448MR8 00000260100MR3 00000237072MR4 00000233442MR3 00000252659MR8 00000232323MR6 00000263946MR6 00000241135MR3 00000239861MR8 00000261676MR1 00000271249MR5 00000272970MR5 00000267654MR2 00000264405MR2 00000271660MR3 00000266270MR8 00000270390MR8 00000282080MR1 00000267328MR3 00000265064MR6 00000011863YR8 00000289848MR4 00000272624MR8 00000285797MR7 00000298690MR9 00000006748YR8 00000277444MR6 00000242557MR7 00000272967MR1 00000286159MR9 00000286624MR2 00000237933MR7 00000237942MR8 00000256289MR0 000000HDMR3145 00000012024YR6 00000281679MR1 00000298555MR4 00000287283MR6 00000011870YR3 00000279066MR5 00000279161MR4 00000280672MR7 00000282546MR1 00000258413MR4 00001003BD6419 00000918BD5B34 00000011943YR8 000000HDMR4228 00000287441MR0 00000278393MR4 00000286156MR5 00000278565MR7 00000919BD5B92 00000280526MR5 00000007852YR7 00000008100YR0 00000260651MR5 00000270252MR0 00000284598MR0 00000008399YR8 00000280362MR5 00000278392MR6 00000281376MR4 00000280952MR3 00000278390MR0 00000279870MR0 00000286269MR6 00000287838MR7 00000011720YR0 00000269580MR7 UNK51078490003 00000265957MR1 00001101XF5026 00000277783MR7 00000272966MR3 00000011980YR2 00000297836MR9 00000296236MR3 00000281159MR4 00000268370MR4 00000296409MR6 00000293527MR8 00000280692MR5 00000262463MR3 00000261203MR4 00000300580MR8 00000290255MR9 00000284686MR3 00000283938MR9 00000274052MR0 00000282641MR0 00000283362MR2 00000278721MR6 00000262505MR1 00000252657MR2 00000258412MR6 00000264561MR2 00000276652MR5 00000271653MR8 00000274105MR6 00000280769MR1 00000242562MR7 00000281344MR2 00000012076YR6 00000274689MR9 00000282583MR4 00000276458MR7 00000011567YR5 00000286650MR7 00000950BD62AB 00000300703MR6 00000297906MR0 00000245675MR4 00000282553MR7 00000282682MR4 00000278468MR4 00000274936MR4 00000012075YR8 00000260627MR5 00000262460MR9 00000264616MR4 00000270245MR4 00000262198MR5 00000284390MR2 00000270226MR4 00000278668MR9 00000283486MR9 00000299272MR5 00000006658YR9 00000262407MR0 00000258396MR1 00000254675MR2 00000245409MR8 00000241867MR1 00000264800MR4 00000273343MR4 00000919BD5BA2 00000270970MR7 00000243683MR0 00000234793MR8 00000237004MR7 00000233445MR6 00000240104MR0 00000292439MR7 00000268521MR2 00000283409MR1 00000254671MR1 00000243699MR6 00000299570MR2 00000251018MR8 00000276655MR8 00000253844MR5 00000237704MR2 00000297862MR5 00000262813MR9 00000262814MR7 00000262736MR2 00000265099MR2 00000243414MR0 00000296314MR8 00000241513MR1 00000245300MR9 00000253833MR8 00000256705MR5 00000242177MR4 00000237934MR5 00000272435MR9 00000284297MR9 00000286145MR8 00000272622MR2 00000265178MR4 00000274832MR5 00000270025MR0 00000286281MR1 00000264745MR1 00000265059MR6 00000261205MR9 00000267207MR9 00000273677MR5 00000287550MR8 00000298719MR6 00000298844MR2 00000296103MR5 00000012359YR6 00000011930YR5 00000012319YR0 00000248176MR0 00000284023MR9 00000272433MR4 00000007527YR5 00000287875MR9 00000286766MR1 00000012128YR5 00000012051YR9 00000284739MR0 00000011119YR5 00000280004MR3 00000298466MR4 00000292467MR8 00000011933YR9 00000288253MR8 00000012259YR8 00000300885MR1 00000296215MR7 00000294159MR9 00000011924YR8 00000011893YR5 00000294161MR5 00000298971MR3 00000295808MR0 00000298931MR7 00000287575MR5 00000012342YR2 00000010513YR0 00000296235MR5 00000297203MR2 00000297699MR1 00000286734MR9 00000250436MR3 00000300475MR1 00000295217MR4 00000296204MR1 00000012159YR0 00000012120YR2 00000297417MR8 00000012386YR9 00000294145MR8 00000298311MR2 00000282008MR2 00000274886MR1 00000276459MR5 00000009417YR7 00000284279MR7 00000278621MR8 00000278748MR9 00000280365MR8 00000286613MR5 00000301066MR7 00000296147MR2 00000922BD5C36 00000295028MR5 00000282511MR5 00000276585MR7 00000279408MR9 00000294146MR6 00000296216MR5 00000301350MR5 00000299663MR5 00000273369MR9 00000282266MR6 00000276676MR4 00000296232MR2 00000294659MR8 00000294341MR3 00000294434MR6 00000288821MR2 00000284642MR6 00000279448MR5 00000280858MR2 00000286601MR0 00000011935YR4 00000012166YR5 00000012373YR7 00000294307MR4 00000298224MR7 00000290408MR4 000000HDMR4252 00000279915MR3 00000007894YR9 00000281187MR5 00000275667MR4 00000297380MR8 00000232427MR5 00000260690MR3 00000261204MR2 00000281158MR6 00000282813MR5 00000287313MR1 00000282379MR7 00000011949YR5 00000028870MR1 00000295900MR5 00000296048MR2 00000241146MR0 00000250033MR8 00000258369MR8 00000251092MR3 00000253165MR5 00000252299MR3 00000243763MR0 00000251242MR4 00000241141MR1 00000257728MR6 00000260676MR2 00000280273MR4 00000275668MR2 00000286697MR8 00000287025MR1 00000297629MR8 00000297724MR7 00000297487MR1 00000011745YR7 00000297819MR5 00000296518MR4 00000299197MR4 00000298548MR9 00000299195MR8 00000299204MR8 00000299196MR6 00000283038MR8 00000831BD4A21 00000273538MR9 00000289266MR9 00000249930MR9 00000273530MR6 00000283156MR8 00000279612MR6 00000297657MR9 00000298495MR3 00000276197MR1 00000238347MR9 00000283023MR0 00000252405MR6 00000235244MR1 00000270567MR1 00000012060YR0 00000287360MR2 00000012302YR6 00000289821MR1 00000263872MR4 00000273622MR1 00000289774MR2 00000278799MR2 00000012197YR0 00000255594MR4 00000283574MR2 00000028626MR7 00000008364YR2 00000241136MR1 00000239031MR8 00001011BD6643 00000271659MR5 00000270389MR0 00000283398MR6 00000299425MR9 00000251245MR7 00000276660MR8 00000230106MR7 00000245400MR7 00000254499MR7 UNK2357500-201 UNK2357500-202 00000263575MR3 00000261737MR1 00000255616MR5 UNK2357500-204 00000250197MR1 00000247718MR0 UNK2357500-205 00000238359MR4 00000300710MR1 00000236502MR1 00000237075MR7 00000271650MR4 00000245404MR9 00000245818MR0 00000268323MR3 00000250441MR3 00000266675MR8 00000254202MR5 00000268117MR9 00000272075MR3 00000266204MR7 00000272287MR4 00000276138MR5 00000283792MR0 00000264096MR9 00000266743MR4 00000282730MR1 00000268115MR3 00000271780MR9 00000275738MR3 00000284337MR3 00000281858MR1 00000273226MR1 00000264094MR4 00000265065MR3 00000278613MR5 00000272079MR5 00000283039MR6 00000279452MR7 00000265614MR8 00000273534MR8 00000260452MR8 00000281866MR4 00000278558MR2 00000266070MR2 00000026397MR9 00000277952MR8 00000268114MR6 00000268510MR5 00000279454MR3 00000270628MR1 00000285690MR4 00000280902MR8 00000012068YR3 00000011925YR5 00000298583MR6 00000838BD4EC9 00000024341MR2 00000248134MR9 00000301566MR6 00000289997MR9 00000288711MR5 00000284296MR1 00000283343MR2 00000298010MR0 00000277757MR1 00000011576YR6 00000028405MR2 00000295442MR8 00000011522YR0 00000293589MR8 00000289439MR2 00000287116MR8 00000299252MR7 00000918BD5B33 00000274715MR2 00000253163MR0 00000265867MR2 00000278103MR7 00000274174MR2 00000284408MR2 00000287421MR2 00000276242MR5 00000288247MR0 00000236997MR3 00000300555MR0 00000012069YR1 00000284492MR6 00000250453MR8 00000295484MR0 00000298757MR6 00000298532MR3 00000283753MR2 00000301106MR1 00000281443MR2 00000296110MR0 00000279908MR8 00000294407MR2 00000294443MR7 00000299326MR9 00000282494MR4 00000294723MR2 00000012261YR4 00000919BD5B8D 00000287422MR0 00000267320MR0 00000273461MR4 00000011509YR7 00000283458MR8 00000299849MR0 00000277532MR8 00000297971MR4 00000298673MR5 00000238345MR3 00000247218MR1 00000250460MR3 00000230690MR0 00000258080MR1 00000232408MR5 00000256987MR9 00000238170MR5 00000237001MR3 00000250204MR5 00000247221MR5 00000257435MR8 00000256670MR1 00000239036MR7 00000235594MR9 00000237003MR9 00000235592MR3 00000252921MR2 00000233429MR0 00000252656MR4 00000250444MR7 00000234578MR3 00000245802MR4 00000261730MR6 00000255595MR1 00000253154MR9 00000251698MR7 00000245815MR6 00000247220MR7 00000255593MR6 00000235591MR5 00000241142MR9 00000235593MR1 00000233438MR1 00000231912MR7 00000232423MR4 00000232410MR1 00000234795MR3 00000231907MR7 00000237066MR6 00000235595MR6 00000232418MR4 00000234860MR5 00000237060MR9 00000237706MR7 00000235691MR3 00000235696MR2 00000237078MR1 00000234799MR5 00000241873MR9 00000236828MR0 00000237707MR5 00000237065MR8 00000250461MR1 00000238165MR5 00000250459MR5 00000237709MR1 00000232422MR6 00000234838MR1 00000237941MR0 00000254507MR7 00000238354MR5 00000239039MR1 00000259977MR7 00000238180MR4 00000239030MR0 00000238350MR3 00000249257MR7 00000242172MR5 00000236627MR6 00000239233MR0 00000241520MR6 00000241718MR6 00000243415MR7 00000242175MR8 00000242179MR0 00000240108MR1 00000242504MR9 00000243697MR0 00000243416MR5 00000241679MR0 00000243692MR1 00000248166YM1 00000248486MR3 00000245673MR9 00000248157MR0 00000245862MR8 00000243188MR0 00000246634MR0 00000243417MR3 00000251228MR3 00000249228MR8 00000248170MR3 00000248482MR2 00000248487MR1 00000248161MR2 00000247212MR4 00000251095MR6 00000251694MR6 00000247735MR4 00000243419MR9 00000250215MR1 00000251640MR9 00000252922MR0 00000252934MR5 00000253830MR4 00000254673MR7 00000247210MR8 00000251693MR8 00000247223MR1 00000250195MR5 00000253829MR6 00000251723MR3 00000253146MR5 00000255216MR4 00000253160MR6 00000250450MR4 00000256707MR1 00000257428MR3 00000258092MR6 00000259133MR7 00000259382MR0 00000257726MR0 00000258374MR8 00000258417MR5 00000256989MR5 00000252427MR0 00000259135MR2 00000261592MR0 00000260018MR7 00000261681MR1 00000255608MR2 00000259576MR7 00000262313MR0 00000258397MR9 00000257440MR8 00000262315MR5 00000265821MR9 00000258408MR4 00000259126MR1 00000262408MR8 00000263260MR2 00000270270MR2 00000242178MR2 00000242559MR3 00000234857MR1 00000239532MR5 00000255611MR6 00000294569MR9 00000258398MR7 00000247214MR0 00000296261MR1 00000298103MR3 00000237938MR6 00000273017MR4 00000260673MR9 00000292024MR7 00000296819MR6 00000264556MR2 00000259381MR2 00000263576MR1 00000292215MR1 00000253155MR6 00000237068MR2 00000270568MR9 00000259569MR2 00000260101MR1 00000272528MR1 00000260461MR9 00000269376MR0 00000270570MR5 00000272076MR1 00000266859MR8 00000270566MR3 00000273620MR5 00000273466MR3 00000272431MR8 00000257441MR6 00000262074MR8 00000269435MR4 00000272429MR2 00000237935MR2 00000265781MR5 00000278305MR8 00000261728MR0 00000268762MR2 00000272338MR5 00000270463MR3 00000280525MR7 00000270977MR2 00000277578MR1 00000264552MR1 00000263791MR6 00000272626MR3 00000278363MR7 00000279145MR7 00000269338MR0 00000270683MR6 00000262404MR7 00000238363MR6 00000273239MR4 00000262314MR8 00000278364MR5 00000302302MR5 00000302746MR3 00000271270MR1 00000267200MR4 00000267203MR8 00000268477MR7 00000268515MR4 00000269575MR7 00000270026MR8 00000272969MR7 00000272070MR4 00000273267MR5 00000271662MR9 00000263338MR6 00000263878MR1 00000263873MR2 00000261880MR9 00000265918MR3 00000262077MR1 00000265921MR7 00000266202MR1 00000266676MR6 00000268366MR2 00000263265MR1 00000268010MR6 00000267812MR6 00000268289MR6 00000268693MR9 00000268008MR0 00000268472MR8 00000268420MR7 00000272071MR2 00000269461MR0 00000271661MR1 00000271664MR5 00000272625MR5 00000269341MR4 00000268686MR3 00000266462MR1 00000272073MR8 00000272074MR6 00000267465MR3 00000272100MR9 00000272889MR7 00000272898MR8 00000272999MR4 00000270565MR5 00000273360MR8 00000271665MR2 00000275582MR5 00000273225MR3 00000275735MR9 00000275599MR9 00000272965MR5 00000275087MR5 00000275737MR5 00000276492MR6 00000266146MR0 00000266148MR6 00000276272MR2 00000266745MR9 00000274173MR4 00000269463MR6 00000279404MR8 00000279459MR2 00000270022MR7 00000272337MR7 00000273489MR5 00000272289MR0 00000273464MR8 00000266064MR5 00000275358MR0 00000270624MR0 00000279148MR1 00000278955MR0 00000277924MR7 00000278469MR2 00000279451MR9 00000283358MR0 00000283576MR7 00000276631MR9 00000233428MR2 00000236996MR5 00000237076MR5 00000242503MR1 00000237940MR2 00000236998MR1 00000239245MR4 00000245671MR3 00000251699MR5 00000251713MR4 00000276436MR3 00000260693MR7 00000262078MR9 00000261682MR9 00000262197MR7 00000261060MR8 00000262461MR7 00000260012MR0 00000261684MR5 00000263819MR5 00000012041YR0 00000287916MR1 00000243694MR7 00000287942MR7 00000298411MR0 00000284385MR2 00000294894MR1 00000237062MR5 00000289839MR3 00000011723YR4 00000255606MR6 00000296358MR5 00000283717MR7 00000297810MR4 00000011773YR9 00000294324MR9 00000286007MR0 00000301065MR9 00000282465MR4 00000012320YR8 822 00000288151MR4 00000286272MR0 00000294321MR5 00000298505MR9 00000919BD5B84 00000902BD55AE 00000263940MR9 00000289210MR7 00000011799YR4 00000302425MR4 00000289392MR3 00000249259MR3 00000299912MR6 00000284313MR4 00000296402MR1 00000285628MR4 00000294115MR1 00000267618MR7 00000295745MR4 00000294960MR0 00000298948MR1 00000012309YR1 00000295095MR4 00000296072MR2 00000930BD5D40 00000816BD3AA6 00000278127MR6 00000288073MR0 00000012156YR6 00000296213MR2 00000296273MR6 00000012124YR4 00000011891YR9 00000012002YR2 00000285728MR2 00000294652MR3 00000012318YR2 00000269584MR9 00000284759MR8 00000287629MR0 00000287881MR7 00000298076MR1 00000011737YR4 00000258402MR7 00000300614MR5 00000299844MR1 00000294647MR3 00000285048MR5 00000295748MR8 00000289790MR8 00000286543MR4 00000011922YR2 00000287460MR0 00000268837MR2 00000268047MR8 00000269332MR3 00000268417MR3 00000273266MR7 00000264749MR3 00000268517MR0 00000272098MR5 00000274169MR2 00000277580MR7 00000276244MR1 00000277960MR1 00000278040MR1 00000278801MR6 00000278556MR6 00000281375MR6 00000280675MR0 00000275527MR0 00000273007MR5 00000280366MR6 00000282337MR5 00000009032YR4 00000281316MR0 00000279906MR2 00000283773MR0 00000283485MR1 00000278221MR7 00000284392MR8 00000284464MR5 00000277533MR6 00000285051MR9 00000277643MR3 00000286617MR6 00000285627MR6 00000278608MR5 00000287393MR3 00000283734MR2 00000286691MR1 00000282003MR3 00000286968MR3 00000286596MR2 00000287269MR5 00000287070MR7 00000287399MR0 00000287361MR0 00000287592MR0 00000286469MR2 00000287900MR5 00000287851MR0 00000287866MR8 00000289657MR9 00000289194MR3 00000011597YR2 00000012048YR5 00000012058YR4 00000289320MR4 00000012053YR5 00000012061YR8 00000012055YR0 00000011777YR0 00000012079YR0 00000281934MR0 00000012182YR2 00000012074YR1 00000012130YR1 00000011912YR3 00000011505YR5 00000011807YR5 00000011261YR5 00000814BD39F1 00000012022YR0 00000293562MR5 00000011729YR1 00000012045YR1 00000919BD5BB8 00000292642MR6 00000277445MR3 00000011557YR6 00000294709MR1 00000294905MR5 00000295189MR5 00000011999YR0 00000919BD5BBE 00000012347YR1 00000285165MR7 00000922BD5C3D 00000919BD5BBA 00000292088MR2 00000012238YR2 00000275968MR6 00000278568MR1 00000277646MR6 00000278397MR5 00000278849MR5 00000276719MR2 00000283026MR3 00000281341MR8 00000283406MR7 00000283116MR2 00000276720MR0 00000276434MR8 00000283114MR7 00000281152MR9 00000011200YR3 00000012123YR6 00000011545YR1 00000011735YR8 00000292942MR0 00000284261MR5 00000280780MR8 00000296036MR7 00000294927MR9 00000295697MR7 00000295239MR8 00000298404MR5 00000294926MR1 00000283775MR5 00001101XF50A6 00000295594MR6 00000290324MR3 00000292984MR2 00000282550MR3 00000279407MR1 00000289384MR0 00000286153MR2 00000237697MR8 00000287314MR9 00000287827MR0 00000245674MR7 00000245820MR6 00000295896MR5 00000295234MR9 00000295237MR2 00000295542MR5 00000295702MR5 00000296092MR0 00000295895MR7 00000296202MR5 00000296304MR9 00000293323MR2 00000296411MR2 00000294555MR8 00000295897MR3 00000296368MR4 00000297204MR0 00000298558MR8 00000297463MR2 00000298582MR8 00000292688MR9 00000296644MR8 00000299493MR7 00000301405MR7 00000296457MR5 00000297626MR4 00001102XF506C 00000296306MR4 00001101XF50B3 00001101XF50BA 00000301243MR2 00001105XF5010 00000298186MR8 00000259647MR6 00000261680MR3 00000269436MR2 00000301351MR3 00000301859MR5 00000267728MR4 00000257427MR5 00000301833MR0 00000268690MR5 00000236827MR2 00000298857MR4 00000251700MR1 00000268470MR2 00000298954MR9 00000300495MR9 00000248175MR2 00000260014MR6 00000299473MR9 00000301338MR0 00000299524MR9 00000299829MR2 00000296361MR9 00000299186MR7 00000300697MR0 00000301339MR8 00000248484MR8 00000300185MR6 00000277691MR2 00000297904MR5 00000299827MR6 00000299268MR3 00000265104MR0 00000302652MR3 00000300610MR3 00000300730MR9 00000298538MR0 00000299244MR4 00000232405MR1 00000293974MR2 00000301910MR6 00000297710MR6 00000299553MR8 00000241134MR6 00000232401MR0 00000298541MR4 00000270528MR3 00000270097MR9 00000268301MR9 00000267860MR5 00000272968MR9 00000268692MR1 00000270467MR4 00000272536MR4 00000267652MR6 00000266824MR2 00000267471MR1 00000275964MR5 00000277961MR9 00000277920MR5 00000276679MR8 00000278163MR1 00000281364MR0 00000278674MR7 00000285602MR9 00000011995YR8 00000288160MR5 00000012004YR8 00000012016YR2 00000292288MR8 00000919BD5BB6 00000295192MR9 00000277764MR7 00000285674MR8 00000288939MR2 00000282732MR7 00000286625MR9 00000277211MR9 00000287568MR0 00000292078MR3 00000280782MR4 00000011959YR4 00000012001YR4 00000292983MR4 00000289655MR3 00000295240MR6 00000281078MR6 00000298197MR5 00000288359MR3 00000296037MR5 00000299666MR8 00000299274MR1 00000299315MR2 00000296108MR4 00000296049MR0 00000298410MR2 00000300771MR3 00000298933MR3 00000299316MR0 00000298409MR4 00000285053MR5 00000236701MR9 00000284869MR5 00000257442MR4 00000243250MR8 00000249655MR2 00000249757MR6 00000249230MR4 00000248172MR9 00000295809MR8 00000298549MR7 00000298825MR1 00000298282MR5 00000297903MR7 00000298756MR8 00000296426MR0 00000298715MR4 00000299641MR1 00000299203MR0 00000299470MR5 00000301737MR3 00000299270MR9 00000301816MR5 00000301736MR5 00000301817MR3 00000295475MR8 00000299183MR4 00000301916MR3 00000295901MR3 00000301815MR7 00000299194MR1 00000254504MR4 00000252424MR7 00000254489MR8 00000254196MR9 00000247457MR5 00000254496MR3 00000253633MR2 00000251244MR0 00000266464MR7 00000297489MR7 00000298189MR2 00000298069MR6 00000284866MR1 00000294520MR2 00000298754MR3 00000243244MR1 00000257424MR2 00000263256MR0 00000243410MR8 00000279077MR2 00000262798MR2 00000274708MR7 00000250200MR3 00000249763MR4 00000279563MR1 00000289819MR5 00000282503MR2 00000297863MR3 00000256219MR7 00000274175MR9 00000295540MR9 00000268526MR1 00000011881YR0 00000288823MR8 00000296521MR8 00000286254MR8 00000011942YR0 00000239237MR1 00000262153MR0 00000012077YR4 00000288014MR4 00000284710MR1 00000278389MR2 00000011513YR9 00000009569YR5 00000011500YR6 00000298584MR4 000000HDMR6286 00000242593MR2 00000289588MR6 00000006589YR6 1112 00000286509MR5 00000253167MR1 00000284833MR1 00000254194MR4 00000262791MR7 00000289914MR4 00000281363MR2 00000249254MR4 00000252434MR6 00000257419MR2 00000249280MR9 00000259579MR1 00000280243MR7 00000249260MR1 00000300574MR1 00000297885MR6 00000270468MR2 00000276276MR3 00000276273MR0 00000288212MR4 00000286777MR8 00000259120MR4 00000269344MR8 00000273518MR1 00000286146MR6 00000262257MR9 00000273465MR5 00000270466MR6 00000011318YR3 00000010509YR8 00000011779YR6 00000294116MR9 00000295474MR1 00000296365MR0 00000298213MR0 00000298845MR9 00000294722MR4 00000287023MR6 00000271777MR5 00000273539MR7 00000299947MR2 00000284024MR7 00000286411MR4 00000227450MR7 00000260464MR3 00000295015MR2 00000299757MR5 00000011782YR0 00000283405MR9 00000288772MR7 00000287475MR8 00000241131MR2 00000245669MR7 00000259379MR6 00000239536MR6 00000284578MR2 00000261940MR1 00000235685MR5 00000282383MR9 00000282549MR5 00000285498MR2 00000299830MR0 00000275357MR2 00000285020MR4 00000259584MR3 00000269348MR9 00000243769MR7 00000276659MR0 00000286384MR2 00000271125MR7 00000282810MR1 00000238177MR0 00000238868MR4 00000259581MR7 00000277958MR5 00000267214MR5 00000277531MR0 00000268511MR3 00000266737MR6 00000276265MR6 00000282923MR2 000000000R1997 00000252440MR3 00000278687MR9 00000255205MR7 00000269454MR5 00000280068MR8 00000012253YR1 000000HDMR2522 00000243690MR5 00000009612YR3 00000295829MR6 00000296864MR2 00000297013MR5 00000289589MR4 00000263254MR5 00000269392MR7 00000273344MR2 00000293239MR0 00000284236MR7 00000284867MR9 00000284019MR7 00000011733YR3 00000297808MR8 00000294525MR1 00000280778MR2 00000282865MR5 00000287275MR2 00000291833MR2 00000250440MR5 00000298373MR2 00000291331MR7 00000298851MR7 00000258373MR0 00000291674MR0 00000292364MR7 00000277602MR9 00000262671MR1 00000258400MR1 00000247216MR5 00000918BD5B3A 00000300938MR8 00000919BD5BA4 00000299168MR5 00000011921YR4 00000286158MR1 00000288705MR7 00000289361MR8 00000299185MR9 00000010985YR0 00000283948MR8 00000283127MR9 00000286275MR3 00000285049MR3 00000290000MR9 00000278560MR8 00000011868YR7 00000012052YR7 00000012349YR7 00000919BD5BBD 00000293023MR8 00000296362MR7 00000298276MR7 00000298679MR2 00000298955MR6 00000300798MR6 00000298970MR5 00000301385MR1 00000251057MR6 00000298758MR4 00000278463MR5 00000261574MR8 00000293279MR6 00000235701MR0 00000290947MR1 00000267093MR3 00000268136MR9 00000298377MR3 00000283351MR5 00000290033MR0 00000291875MR3 00000268009MR8 00000287382MR6 00000296237MR1 00000012191YR3 00000296464MR1 00000287230MR7 00000288462MR5 00000233271MR6 00000276591MR5 00000273366MR5 00000279460MR0 00000272533MR1 00000239869MR1 00000268012MR2 00000300982MR6 00000300810MR9 00000281446MR5 00000011187YR2 00000298531MR5 00000301848MR8 00000294660MR6 00000012348YR9 00000255199MR2 00000282897MR8 00000291505MR6 00000290570MR1 00000257735MR1 00000282029MR8 00000251799MR3 00000286263MR9 00000288041MR7 00000011542YR8 00000294692MR9 00000011785YR3 00000229366MR0 00000265607MR2 00000293051MR9 00000265101MR6 00000252924MR6 00000265098MR4 00000260671MR3 00000248171MR1 00000258420MR9 00000267103MR0 00000271658MR7 00000273227MR9 00000274071MR0 00000276494MR2 00000273457MR2 00000268284MR7 00000273009MR1 00000275845MR6 00000289845MR0 00000011781YR2 00000286798MR4 00000286273MR8 00000011771YR3 00000254185MR2 00000278549MR1 00000011795YR2 00000287316MR4 00000288122MR5 00000286471MR8 00000284287MR0 00000918BD5B53 00000011802YR6 00000011554YR3 00000283843MR1 00000011559YR2 00000297012MR7 00000288387MR4 00000011794YR5 00000275017MR2 00000285054MR3 00000012271YR3 00000294693MR7 00000298042MR3 00000248163MR8 00000302358MR7 00000294946MR9 00000293651MR6 00000280923MR4 00000284714MR3 00000289765MR0 00000299253MR5 00000294895MR8 00000295090MR5 00000298684MR2 00000299910MR0 00000011517YR0 00000282382MR1 00000263944MR1 00000233424MR1 00000268522MR0 00000243185MR6 00000256207MR2 00000266670MR9 00000285301MR8 00000283755MR7 00000282464MR7 00000282958MR8 00000282083MR5 00000286687MR9 00000295700MR9 00000295877MR5 00000286965MR9 00000286786MR9 00000295762MR9 00000255202MR4 00000278627MR5 00000011515YR4 00000294130MR0 00000285673MR0 00000011593YR1 00000905BD567D 00000296091MR2 00000289976MR3 00000011307YR6 00000300745MR7 00000298707MR1 00000240111MR5 00000252426MR2 00000011063YR5 00000278844MR6 00000283620MR3 00000288888MR1 00000269046MR9 00000274070MR2 00000285352MR1 00000012049YR3 00000294392MR6 000000HDMR2875 00000011713YR5 00000011560YR0 00000012323YR2 00000272427MR6 00000238341MR2 00000298647MR9 00000284387MR8 00000285018MR8 00001043XF5019 00000250207MR8 00000268529MR5 00000262792MR5 00000270626MR5 00000268474MR4 00000287379MR2 00000288019MR3 00000298972MR1 00000011503YR0 00000275465MR3 00000246619MR1 00000250206MR0 00000257423MR4 00000246635MR7 00000259134MR5 00000233425MR8 00000237005MR4 00000240109MR9 00000246618MR3 00000249657MR8 215804 00000250420MR7 00000250193MR0 00000250457MR9 00000251719MR1 00000251717MR5 00000252925MR3 00000253841MR1 00000256204MR9 00000254661MR2 00000256693MR3 00000261628MR2 00000261732MR2 00000262100MR1 00000264459MR9 00000265608MR0 00000256223MR9 00000263241MR7 00000267853MR0 00000270524MR2 00000271863MR3 00000257741MR9 00000283361MR4 00000287765MR2 00000243407MR4 00000233441MR5 00000288144MR9 00000234805MR0 00000233266MR6 00000237950MR1 00000251712MR6 00000230663MR7 00000239243MR9 00000294475MR9 00000270272MR8 00000293052MR7 00000237691MR1 00000254197MR7 00000298726MR1 00000264008MR4 00000275356MR4 00000276590MR7 00000271322MR0 00000260692MR9 00000261202MR6 00000263341MR0 00000267806MR8 00000270464MR1 00000270391MR6 00000273362MR4 00000274150MR2 00000273522MR3 00000275847MR2 00000275935MR5 00000276332MR4 00000282084MR3 00000283300MR2 00000285914MR8 00000285300MR0 00000286060MR9 00000286058MR3 00000286061MR7 00000291852MR2 00000264746MR9 00000300776MR2 00000284863MR8 00000290008MR2 00000289220MR6 00000274740MR0 00000256706MR3 00000262807MR1 00000273363MR2 00000272530MR7 00000273674MR2 00000276734MR1 00000277215MR0 00000279920MR3 00000280861MR6 00000282685MR7 00000283791MR2 00000286551MR7 00000285435MR4 00000285917MR1 00000287120MR0 00000287576MR3 00000283152MR7 00000288207MR4 00000289360MR0 00000289198MR4 00000011712YR7 00000012311YR7 00000845BD51D2 00000294322MR3 00000293027MR9 00000294420MR5 00000294506MR1 00000294606MR9 00000294961MR8 00000295089MR7 00000295051MR7 00000292643MR4 00000918BD5B55 00000296642MR2 00000296363MR5 00000281255MR0 00000294395MR9 00000297821MR1 00000297008MR5 00000297697MR5 00000297878MR1 00000298487MR0 00000299651MR0 00000300476MR9 00000300504MR8 00000301764MR7 00000302644MR0 000000HDMR5719 00000GE0208031 00000GE0249066 00000288193MR6 00000287425MR3 00000289339MR4 00000265005MR9 00000297421MR0 00000281056MR2 00000243187MR2 00000289319MR6 00000261729MR8 00000270271MR0 00000270269MR4 00000242556MR9 00000268530MR3 00000258394MR6 00000245670MR5 00000284362MR1 00000242584MR1 UNK22943020002 UNK22943020003 00000246627MR4 00000290550MR3 00000239235MR5 00000254498MR9 00000290571MR9 00000287812MR2 00000272881MR4 00000250198MR9 00000281746MR8 00000250194MR8 00000262149MR8 00000282046MR2 00000263796MR5 00000295560MR7 00000218379MR6 00000283359MR8 00000288898MR0 00000258404MR3 00000290355MR7 00000253162MR2 00000248168MR7 00000234831MR6 00000255588MR6 00000262050MR8 00000265818MR5 00000261939MR3 00000245942MR8 00000241517MR2 00000254486MR4 00000241521MR4 00000238176MR2 00000265822MR7 00000250443MR9 00000243762MR2 00000243418MR1 00000235587MR3 00000241516MR4 00000252918MR8 00000290647MR7 00000290038MR9 00000264559MR6 00000268316MR7 00000274164MR3 00000261734MR8 00000272099MR3 00000268476MR9 00000280619MR8 00000279800MR7 00000280713MR9 00000280759MR2 00000280734MR5 00000281313MR7 00000276199MR7 00000267036MR2 00000275473MR7 00000281865MR6 00000275016MR4 00000275965MR2 00000281857MR3 00000278210MR0 00000278307MR4 00000268418MR1 00000272436MR7 00000275642MR7 00000282079MR3 00000268320MR9 00000260026MR0 00000277921MR3 00000268291MR2 00000277582MR3 00000283150MR1 00000011788YR7 00000289357MR6 00000285692MR0 00000286621MR8 00000297905MR2 00000919BD5B93 00000289887MR2 00000298658MR6 00000293416MR4 00000280700MR6 00000300700MR2 00000275466MR1 00000289658MR7 00000262455MR9 00000274288MR0 00000284577MR4 00000280781MR6 00000294285MR2 00000288435MR1 00000286157MR3 00000918BD5B57 00000294396MR7 00000295591MR2 00000283433MR1 00000288012MR8 00000288614MR1 00000012134YR3 00000295530MR0 00000283516MR3 00000012011YR3 00000280582MR8 00000289847MR6 00000299314MR5 00000299254MR3 00000299310MR3 00001048XF5021 00000299648MR6 00000289394MR9 00000294329MR8 00000280327MR8 00000287847MR8 00000012046YR9 00000286287MR8 00000299650MR2 00000299260MR0 00000254375MR9 00000286250MR6 00000237063MR3 00000283320MR0 00000266069MR4 00000027810MR1 00000297404MR6 00000296047MR4 00000292641MR8 00000012187YR1 00000295902MR1 00000012127YR7 00000282999MR2 00000245814MR9 00000258077MR7 00000262064MR9 00000283302MR8 00000290007MR4 00000241874MR7 00000284233MR4 UNK51078490008 00000259130MR3 00000231908MR5 00000240110MR7 00000264547MR1 00000266821MR8 00000265719MR5 00000267810MR0 00000011734YR1 00000242174MR1 00000266820MR0 00000272285MR8 00000241871MR3 00000283866MR2 00000274714MR5 00000263943MR3 00000277200MR2 00000247211MR6 00000259124MR6 00000257439MR0 00000274935MR6 00000298804MR6 00000298265MR0 00000298790MR7 00000298806MR1 00000298791MR5 00000301038MR6 00000233423MR3 00000252430MR4 00000245940MR2 00000250458MR7 00000256236MR1 00000252915MR4 00000257737MR7 00000259425MR7 00000272565MR3 00000257751MR8 00000261475MR8 00000267105MR5 00000261199MR4 00000275014MR9 00000266669MR1 00000266826MR7 00000269581MR5 00000264620MR6 00000274697MR2 00000275699MR7 00000277600MR3 00000271239MR6 00000301750MR6 00000274694MR9 00000912BD59A1 00000268369MR6 00000268685MR5 00000278038MR5 00000275533MR8 00000279464MR2 00000279615MR9 00000284533MR7 00000285677MR1 00000286649MR9 00000289195MR0 00000288613MR3 00000289162MR0 00000288062MR3 00000011553YR5 00000011512YR1 00000289473MR1 00000011561YR8 00000010254YR3 00000012059YR2 00000296238MR9 00000296865MR9 00000298215MR5 00000299871MR4 00000301186MR3 00000281186MR7 00000012121YR0 00000294344MR7 00000298557MR0 00000277535MR1 00001004BD6441 00000259106MR3 00000GE0151026 00000GE0209032 00000267672MR4 00000265058MR8 00000275736MR7 00000283697MR1 00000287306MR5 00000279447MR7 00000293217MR6 00000298534MR9 00000300778MR8 00000254503MR6 00000249927MR5 00000251093MR1 00000296846MR9 00000297381MR6 00000296847MR7 00000296558MR0 00000296863MR4 00000295174MR7 00000241723MR6 00000282690MR7 00000289785MR8 00000298975MR4 00000287024MR4 00000246608MR4 00000256235MR3 00000270514MR3 00000260103MR7 00000264793MR1 00000259582MR5 00000264798MR0 00000270518MR4 00000260097MR1 00000270521MR8 00000260466MR8 00000272564MR6 00000270525MR9 00000256215MR5 00000258395MR3 00000261387MR5 00000264490MR4 00000266697MR2 00000274107MR2 00000268523MR8 00000253152MR3 00000263941MR7 00000264548MR9 00000283317MR6 00000266757MR4 00000279892MR4 00000278667MR1 00000280903MR6 00000273681MR7 00000279083MR0 00000283357MR2 00000280676MR8 00000276248MR2 00000271120MR8 00000274830MR9 00000262456MR7 00000266746MR7 00000283385MR3 00000285384MR4 00000274892MR9 00000283696MR3 00000285023MR8 00000286267MR0 00000275093MR3 00000282911MR7 00000284214MR4 00000261868MR4 00000277195MR4 00000275475MR2 00000275580MR9 00000286149MR0 00000294435MR3 00000011719YR2 00000296360MR1 00000294271MR2 00000286585MR5 00000277641MR7 00000296404MR7 00000294441MR1 00000012014YR7 00000010796YR1 00000299348MR3 00000299451MR5 00000011026YR2 00000299417MR6 00000295056MR6 00000295477MR4 00000294607MR7 00000287122MR6 00000298093MR6 00000011084YR1 00000295014MR5 00000298843MR4 00000297179MR4 00000284904MR0 00000286012MR0 00000287308MR1 00000287396MR6 000000HDMR5258 00000012013YR9 00000294510MR3 00000298097MR7 00000838BD4ECB 00000299494MR5 00000300186MR4 00000294907MR1 00000297973MR0 00000919BD5B86 00000294514MR5 00000011552YR7 000001106XF022 00000295481MR6 00000286290MR2 00000298488MR8 00000296366MR8 00000287389MR1 00000288979MR8 00000288891MR5 00000301039MR4 00000299675MR9 00000273525MR6 00000286282MR9 00000275844MR9 00000262300MR7 00000008340YR2 00000280857MR4 00000286262MR1 00000286686MR1 00000297445MR9 00000299674MR2 00000295513MR6 00001009BD65FE 00000230686MR8 00000257733MR6 00000263945MR8 00000296463MR3 00000266145MR2 00000274693MR1 00000277191MR3 00000277692MR0 00000276153MR4 00000280627MR1 00000276632MR7 00000290191MR6 00000914BD5A7E 00000294436MR1 00001002BD639A 00000286692MR9 00000287113MR5 00000931BD5D7B 00000294274MR6 00000300616MR0 00000300618MR6 00000011731YR7 00000286248MR0 00000288650MR5 00000011775YR4      REF # 67626  "Products: GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM (Signa Ovation 5 and Ovation HD) (Beijing, China) FMI 67626" Mfg Lot or Serial # 000SPI275 138161HM5 154416HM2 167478HM7 169099HM9 171687HM7 176546HM0 188506HM0 192710HM2 193122HM9 202350HM5 213385HM8 226223HM6 238928HM6 241057HM9 266117HM1 139841HM1 210567HM4 150713HM6 160804HM1 162558HM1 164061HM4 165767HM5 172417HM8 172812HM0 172864HM1 173903HM6 177138HM5 177789HM5 178516HM1 179358HM7 180124HM0 181095HM1 181525HM7 182066HM1 182244HM4 182592HM6 183128HM8 187360HM3 188807HM2 189100HM1 191022HM3 191158HM5 196644HM9 200573HM4 202345HM5 204158HM0 205908HM7 206819HM5 210513HM8 213193HM6 214041HM6 214712HM2 215676HM8 216125HM5 217358HM1 218474HM5 219021HM3 219406HM6 222105HM9 223599HM2 224698HM1 224707HM0 228098HM0 229042HM7 230858HM3 231784HM0 234987HM6 242129HM5 243112HM0 244329HM9 245315HM7 245639HM0 250988HM3 252871HM9 256318HM7 257108HM1 262327HM0 265044HM8 267712HM8 270517HM6 270524HM2 271390HM7 271741HM1 157409HM4 148416HM1 174343HM4 175020HM7 199950HM7 205562HM2 218302HM8 155505HM1 212655HM5 155506HM9 195286HM0 223710HM5 231594HM3 233074HM4 157408HM6 147520HM1 189940HM0 198004HM4 198790HM8 224716HM1 245556HM6 224689HM0 229582HM2 199942HM4 181827HM7 227474HM4 226814HM2 214907HM8 239726HM3 241050HM4 223171HM0 247892HM3 271397HM2 271404HM6 271750HM2 276214HM4 146883HM4 158636HM1 147900HM5 169828HM1 170935HM1 190433HM3 191901HM8 193217HM7 193933HM9 195769HM5 197642HM2 224432HM5 259738HM3 270004HM5 144793HM7 150146HM9 221033HM4 170104HM4 181826HM9 195287HM8 208285HM7 219747HM3 220369HM3 247222HM3 257835HM9 221645HM5 168318HM4 181520HM8 220560HM7    Ref # 67815  "Product: GE 0.7T SIGNA OPENSPEED WITH EXCITE MR SYSTEM, (HFO5), (Signa Openspeed 5)  (Hino, Japan) FMI 67815D" Mfg Lot or Serial # 00000000102YR4 00000000127YR1 00000000133YR9 00000000142YR0 00000000144YR6 00000000146YR1 00000000149YR5 00000000151YR1 00000000155YR2 00000003011YR4 00000003022YR1 00000003023YR9 00000003024YR7 00000003025YR4 00000003026YR2 00000003027YR0 00000003028YR8 00000003029YR6 00000003030YR4 00000003031YR2 00000003033YR8 00000003034YR6 00000003035YR3 00000003036YR1 00000003039YR5 00000003041YR1 00000003042YR9 00000003043YR7 00000003044YR5 00000003050YR2 00000003051YR0 00000003052YR8 00000003053YR6 00000003049YR4 00000000122YR2 00000003047YR8  "Product: GE 0.7T SIGNA OPENSPEED WITH EXCITE MR SYSTEM (HFO4), (Signa Openspeed 4) CR No. 9612283-XX/XX/XX-00X-C (Hino, Japan) FMI 67815C" Mfg Lot or Serial # 00000000100YR8 00000000109YR9 00000000113YR1 00000000118YR0 00000000119YR8 00000000120YR6 00000000128YR9 00000000129YR7 00000000132YR1 00000000134YR7 00000000136YR2 00000000150YR3 00000000152YR9 00000943610YM6 00000951021YM5 00000962592YM2 00000000124YR8 00000000137YR0  "Product: GE 0.35T SIGNA OVATION WITH EXCITE (MFO4) MR SYSTEM (Signa Ovation 4) CR No. 9612283-XX/XX/XX-00X-C (Hino, Japan) FMI 67815B" Mfg Lot or Serial # 00000000123YR0 00000000125YR5 00000000130YR5 00000000138YR8 00000000139YR6 00000000561YR1 00000000563YR7 00000000564YR5 00000000586YR6 00000000756YR7 00000000757YR5 00000000758YR3 00000000759YR1 00000000760YR9 00000000768YR2 00000000770YR8 00000000771YR6 00000000772YR4 00000000773YR2 00000000775YR7 00000000784YR9 00000000785YR6 00000000786YR4 00000000788YR0 00000000796YR3 00000000798YR9 00000000809YR4 00000000812YR8 00000000816YR9 00000000818YR5 00000000825YR0 00000000826YR8 00000000827YR6 00000000835YR9 00000963007YM0 00000000804YR5 00000000778YR1 00000000779YR9 00000000780YR7 00000000789YR8 00000000790YR6 00000000793YR0 00000000795YR5 00000000799YR7 00000000800YR3 00000000801YR1 00000000803YR7 00000000805YR2 00000000807YR8 00000000808YR6 00000000810YR2 00000000819YR3 00000000828YR4 00000000830YR0 00000000831YR8 00000000832YR6 00000000834YR2 00000000836YR7 00000000837YR5 00000000838YR3 00000000839YR1 00000000840YR9 00000000841YR7 00000000811YR0 00000000140YR4 00000000141YR2 00000000565YR2 00000000763YR3 00000000766YR6 00000000767YR4 00000000769YR0 00000000787YR2 00000000761YR7 00000000116YR4 00000000117YR2 00000000777YR3 00000000143YR8 00000000781YR5 00000000776YR5 00000000765YR8 00000000782YR3 00000000791YR4 00000000823YR5 00000000824YR3 00000000797YR1 00000000794YR8  "Product: GE 1.5T SIGNA HDe MR SYSTEM  CR No. 9612283-XX/XX/XX-00X-C (Hino, Japan) FMI 67815A" Mfg Lot or Serial # 00000004984YR1 00000005068YR2 00000005079YR9 00000005080YR7 00000005081YR5 00000005648YR1 00000005656YR4 00000005659YR8 00000005664YR8 00000006155YR6 00000006157YR2 00000006172YR1 00000006176YR2 00000006177YR0 00000006188YR7 00000006194YR5 00000006198YR6 00000006200YR0 00000006204YR2 00000006514YR4 00000006515YR1 00000006516YR9 00000006519YR3 00000006522YR7 00000006525YR0 00000006528YR4 00000006532YR6 00000006535YR9 00000006538YR3 00000006544YR1 00000007947YR5 00000007954YR1 00000007969YR9 00000007970YR7 00000007971YR5 00000008692YR6 00000008698YR3 00000008699YR1 00000008701YR5 00000008703YR1 00000008709YR8 00000008715YR5 00000008716YR3 00000008720YR5 00000008726YR2 00000008727YR0 00000008728YR8 00000008729YR6 00000008732YR0 00000008734YR6 00000009833YR5 00000010024YR8 00000010025YR5 00000010114YR7 00000010578YR3 00000010583YR3 00000010686YR4 00000010794YR6 00000010795YR3 00000010830YR8 00000010831YR6 00000010834YR0 00000010856YR3 00000010863YR9 00000011210YR2 00000011428YR0 00000011429YR8 00000011495YR9 00000011499YR1 00000011637YR6 00000011648YR3 00000011650YR9 00000011669YR9 00000011686YR3 00000011826YR5 00000011829YR9 00000011830YR7 00000011842YR2 00000012085YR7 00000012098YR0 00000012099YR8 00000012100YR4 00000012102YR0 00000012142YR6 00000012172YR3 00000012206YR9 00000012207YR7 00000012221YR8 00000012227YR5 00000012286YR1 00000012287YR9 00000012289YR5 00000012330YR7 00000012393YR5 00000012434YR7 00000012435YR4 00000012503YR9 00000012507YR0 00000012576YR5 00000012651YR6 00000012652YR4 00000012727YR4 00000012728YR2 00000012755YR5 00000012756YR3 00000012818YR1 00000012820YR7 00000012940YR3 00000012943YR7 00000012988YR2 00000012992YR4 00000012994YR0 00000012995YR7 00000013056YR7 00000013061YR7 00000013177YR1 00000013180YR5 00000013185YR4 00000013239YR9 00000013241YR5 00000013246YR4 00000013301YR7 00000013306YR6 00000013310YR8 00000013375YR1 00000013379YR3 00000013384YR3 00000013385YR0 00000013399YR1 00000013453YR6 00000013799YR2 00000013800YR8 00000013802YR4 00000013804YR0 00000013878YR4 00000013880YR0 00000014113YR5 00000014114YR3 00000014214YR1 00000014276YR0 00000014277YR8 00000014520YR1 00000014524YR3 00000014801YR5 00000014806YR4 00000014858YR5 00000014859YR3 00000014909YR6 00000014910YR4 00000014956YR7 00000011535YR2 00000012209YR3 00000014414YR7 00000006174YR7 00000006507YR8 00000010827YR4 00000011960YR2 00000012144YR2 00000012497YR4 00000012575YR7 00000012753YR0 00000012991YR6 00000013693YR7 00000013875YR0 00000014415YR4 00000014419YR6 00000014620YR9 00000006547YR4 00000014907YR0 00000011692YR1 00000012367YR9 00000012508YR8 00000012550YR0 00000012722YR5 00000012942YR9 00000013305YR8 00000013451YR0 00000013626YR7 00000013805YR7 00000013876YR8 00000013877YR6 00000012796YR9 00000007964YR0 00000008704YR9 00000009948YR1 00000010023YR0 00000010685YR6 00000010833YR2 00000011827YR3 00000011824YR0 00000013057YR5 00000006181YR2 00000006183YR8 00000006184YR6 00000006185YR3 00000006186YR1 00000007951YR7 00000007952YR5 00000008688YR4 00000008689YR2 00000008705YR6 00000008706YR4 00000008707YR2 00000008708YR0 00000008710YR6 00000008711YR4 00000008712YR2 00000008730YR4 00000008731YR2 00000008733YR8 00000009832YR7 00000010227YR7 00000010231YR9 00000010250YR9 00000010253YR3 00000010579YR1 00000010754YR0 00000010755YR7 00000010857YR1 00000010860YR5 00000010862YR1 00000011186YR4 00000011211YR0 00000011212YR8 00000011238YR3 00000011290YR4 00000011343YR1 00000011349YR8 00000011537YR8 00000011539YR4 00000011638YR4 00000011639YR2 00000011649YR1 00000011652YR5 00000011653YR3 00000011685YR5 00000011689YR7 00000011690YR5 00000011828YR1 00000011840YR6 00000011961YR0 00000011963YR6 00000011964YR4 00000011987YR5 00000011991YR7 00000011992YR5 00000012103YR8 00000012117YR8 00000012118YR6 00000012119YR4 00000012141YR8 00000012147YR5 00000012171YR5 00000012366YR1 00000012368YR7 00000012389YR3 00000012390YR1 00000012391YR9 00000012449YR5 00000012551YR8 00000012579YR9 00000012580YR7 00000012581YR5 00000012582YR3 00000012723YR3 00000012729YR0 00000012875YR1 00000012876YR9 00000012879YR3 00000012880YR1 00000012936YR1 00000012937YR9 00000012993YR2 00000013055YR9 00000013092YR2 00000013135YR9 00000013238YR1 00000013240YR7 00000013244YR9 00000013245YR6 00000013249YR8 00000013303YR3 00000013309YR0 00000013347YR0 00000013377YR7 00000013397YR5 00000013445YR2 00000013446YR0 00000013448YR6 00000013452YR8 00000013576YR4 00000013577YR2 00000013578YR0 00000013579YR8 00000013580YR6 00000013582YR2 00000013585YR5 00000013617YR6 00000013618YR4 00000013619YR2 00000013620YR0 00000013622YR6 00000013695YR2 00000013696YR0 00000013697YR8 00000013699YR4 00000013700YR0 00000013701YR8 00000013702YR6 00000013810YR7 00000013914YR7 00000013915YR4 00000013916YR2 00000013935YR2 00000013936YR0 00000013937YR8 00000014058YR2 00000014059YR0 00000014069YR9 00000014104YR4 00000014105YR1 00000014109YR3 00000014110YR1 00000014112YR7 00000014212YR5 00000014213YR3 00000014215YR8 00000014216YR6 00000014217YR4 00000014278YR6 00000014327YR1 00000014329YR7 00000014330YR5 00000014331YR3 00000014332YR1 00000014349YR5 00000014350YR3 00000014351YR1 00000014402YR2 00000014413YR9 00000014416YR2 00000014457YR6 00000014515YR1 00000014519YR3 00000014525YR0 00000014526YR8 00000014554YR0 00000014555YR7 00000014623YR3 00000014654YR8 00000014655YR5 00000014656YR3 00000014657YR1 00000014658YR9 00000014698YR5 00000014699YR3 00000014700YR9 00000014701YR7 00000014702YR5 00000014703YR3 00000014704YR1 00000014749YR6 00000014750YR4 00000014751YR2 00000014752YR0 00000014754YR6 00000014800YR7 00000014802YR3 00000014803YR1 00000014804YR9 00000014860YR1 00000014861YR9 00000014862YR7 00000014863YR5 00000014911YR2 00000014957YR5 00000014981YR5 00000014982YR3 00000014983YR1 00000014984YR9 00000014999YR7 00000015000YR3 00000015002YR9 00000015041YR7 00000015131YR6 00000012578YR1 00000013806YR5 00000014107YR7 00000009947YR3 00000013183YR9 00000006164YR8 00000006509YR4 00000012759YR7 00000013879YR2 00000006202YR6 00000006170YR5 00000008717YR1 00000013625YR9 00000005078YR1 00000005655YR6 00000008690YR0 00000009954YR9 00000006527YR6 00000006536YR7 00000008693YR4 00000009955YR6 00000010756YR5 00000010828YR2 00000011288YR8 00000011347YR2 00000011497YR5 00000011636YR8 00000011768YR9 00000011825YR7 00000012821YR5 00000013134YR2 00000013181YR3 00000006161YR4 00000006167YR1 00000012291YR1 00000012501YR3 00000012549YR2 00000013311YR6 00000010582YR5 00000011042YR9 00000010832YR4 00000011285YR4 00000011651YR7 00000011966YR9 00000012086YR5 00000012169YR9 00000012288YR7 00000012327YR3 00000012329YR9 00000012506YR2 00000012655YR7 00000012989YR0 00000013060YR9 00000013242YR3 00000013247YR2 00000013248YR0 00000014908YR8 00000011038YR7 00000012654YR0 00000014516YR9 00000013382YR7 00000013623YR4 00000013627YR5 00000013628YR3 00000013703YR4 00000014061YR6 00000014064YR0 00000011766YR3 00000013130YR0 00000005073YR2 00000006190YR3 00000006512YR8 00000006520YR1 00000006549YR0 00000010864YR7 00000011839YR8 00000011965YR1 00000012168YR1 00000012293YR7 00000012436YR2 00000012496YR6 00000013395YR9 00000013398YR3 00000006552YR4 00000004672YR2 00000005065YR8 00000005067YR4 00000005069YR0 00000005076YR5 00000005084YR9 00000005645YR7 00000005650YR7 00000005652YR3 00000005653YR1 00000005654YR9 00000005658YR0 00000005660YR6 00000005662YR2 00000006158YR0 00000006159YR8 00000006160YR6 00000006163YR0 00000006165YR5 00000006166YR3 00000006168YR9 00000006178YR8 00000006179YR6 00000006180YR4 00000006182YR0 00000006187YR9 00000006192YR9 00000006193YR7 00000006201YR8 00000006505YR2 00000006506YR0 00000006508YR6 00000006513YR6 00000006517YR7 00000006518YR5 00000006524YR3 00000006526YR8 00000006529YR2 00000006531YR8 00000006537YR5 00000006540YR9 00000006546YR6 00000006548YR2 00000007945YR9 00000007946YR7 00000007948YR3 00000007949YR1 00000007950YR9 00000007953YR3 00000007955YR8 00000007956YR6 00000007957YR4 00000007958YR2 00000007959YR0 00000007960YR8 00000007961YR6 00000007962YR4 00000007963YR2 00000007965YR7 00000007972YR3 00000007973YR1 00000007974YR9 00000008685YR0 00000008686YR8 00000008687YR6 00000008691YR8 00000008696YR7 00000008718YR9 00000008723YR9 00000008725YR4 00000009949YR9 00000009956YR4 00000010021YR4 00000010251YR7 00000010580YR9 00000010684YR9 00000010687YR2 00000010858YR9 00000010859YR7 00000010861YR3 00000011040YR3 00000011041YR1 00000011289YR6 00000011344YR9 00000011345YR6 00000011346YR4 00000011348YR0 00000011430YR6 00000011672YR3 00000011694YR7 00000011695YR4 00000011764YR8 00000011841YR4 00000011843YR0 00000011988YR3 00000011990YR9 00000011993YR3 00000012082YR4 00000012084YR0 00000012143YR4 00000012148YR3 00000012170YR7 00000012324YR0 00000012325YR7 00000012326YR5 00000012331YR5 00000012653YR2 00000012657YR3 00000012726YR6 00000012730YR8 00000012804YR1 00000012822YR3 00000012939YR5 00000012941YR1 00000013027YR8 00000013028YR6 00000013093YR0 00000013178YR9 00000013182YR1 00000013186YR2 00000013205YR0 00000013250YR6 00000013251YR4 00000013287YR8 00000013344YR7 00000013345YR4 00000013387YR6 00000013644YR0 00000013583YR0 00000014805YR6 00000012292YR9 00000005657YR2 00000006171YR3 00000006173YR9 00000006197YR8 00000007966YR5 00000007967YR3 00000010829YR0 00000010978YR5 00000011039YR5 00000011534YR5 00000011646YR7 00000011691YR3 00000011994YR1 00000012208YR5 00000012392YR7 00000012498YR2 00000012502YR1 00000012505YR4 00000012724YR1 00000013302YR5 00000013447YR8 00000013694YR5 00000013704YR2 00000013803YR2 00000014108YR5 00000014418YR8 00000014454YR3 00000011765YR5 00000006189YR5 00000011286YR2 00000011986YR7 00000012795YR1 00000012817YR3 00000013184YR7 00000013386YR8 00000013396YR7 00000013449YR4 00000013624YR2 00000014062YR4 00000014111YR9 00000014401YR4 00000014403YR0 00000014455YR0 00000014517YR7 00000014912YR0 00000013095YR5 00000015091YR2 00000011767YR1 00000010977YR7 00000011538YR6 00000012226YR7 00000012577YR3 00000012658YR1 00000013094YR8 00000013203YR5 00000012573YR2 00000013381YR9 00000013645YR7 00000012205YR1 00000012824YR9 00000012938YR7 00000013801YR6 00000014063YR2 00000011035YR3 00000011498YR3 00000011962YR8 00000011989YR1 00000013243YR1 00000013343YR9 00000006539YR1 00000006551YR6 00000008719YR7 00000008721YR3 00000010254YR1 00000011037YR9 00000012754YR8 00000014060YR8 00000014071YR5 00000015092YR0 00000006156YR4 00000010225YR1 00000012547YR6 00000012819YR9 00000013807YR3 00000005077YR3 00000006169YR7 00000008694YR2 00000008713YR0 00000008714YR8 00000010226YR9 00000011432YR2 00000012210YR1 00000004773YR8 00000005082YR3 00000006196YR0 00000006521YR9 00000006545YR8 00000006550YR8 00000010230YR1 00000012823YR1 00000013304YR1 00000013307YR4 00000013376YR9 00000010229YR3 00000014211YR7 00000012083YR2 00000013179YR7 00000012328YR1 00000012990YR8 00000013450YR2 00000014753YR8 00000015001YR1 00000010252YR5 00000012797YR7 00000013380YR1 00000005072YR4 00000005074YR0 00000006203YR4 00000006510YR2 00000006511YR0 00000006523YR5 00000006530YR0 00000006533YR4 00000006534YR2 00000006541YR7 00000006542YR5 00000006543YR3 00000006553YR2 00000006554YR0 00000010939YR7 00000010976YR9 00000012552YR6 00000012553YR4 00000012574YR0 00000012793YR6 00000012794YR4 00000012878YR5 00000012996YR5 00000013058YR3 00000013059YR1 00000013204YR3 00000013378YR5 00000013629YR1 00000013757YR0 00000014328YR9 00000014417YR0 00000014456YR8 00000011838YR0 00000006199YR4 00000010115YR4 00000011536YR0 00000012290YR3 00000014518YR5 00000011532YR9 00000006175YR4 00000006195YR2 00000008700YR7 00000008722YR1 00000008724YR7 00000010581YR7 00000011496YR7 00000011668YR1 00000012101YR2 00000012145YR9 00000012146YR7 00000012149YR1 00000012792YR8 00000013308YR2 00000014065YR7 00000014458YR4 00000014622YR5 00000014621YR7 00000013698YR6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            
FEI Number 2126677
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact
262-513-4122
Manufacturer Reason
for Recall
Screen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the Signa HDxT 1.5T system, internal investigation revealed that the issue also impacts multiple GE healthcare MR Systems.
FDA Determined
Cause 2
Software design
ActionGE Healthcare communicated internally to their Service Team the directions for completing the correction on the customers' systems. GE Healthcare's recall strategy was to update all affected systems that were distributed. At this time,Week of 02/06/2012, deployment of this recall is considered complete. Should you have any questions or require additional information, please contact Regulatory Affairs Leader by telephone at 262-548-2608 or email paulcorrigan@ge.com.
Quantity in Commerce7016 (2919 US; 4097 OUS)
DistributionWorldwide distribution: USA (nationwide) including DC and Puerto Rico and countries: ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIDJAN, BELARUS, BELGIUM, BOLIVIA, BOSNIA-HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, FINLAND, FRANCE, GEORGIA, GERMANY, GREAT BRITAIN, GREECE, GUADELOUPE, GUATEMALA, GUYANA, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN,JORDAN, KAZAKHSTAN, KENYA, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIAN, LUXEMBOURG, MALAYSIA, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PANAMA, PARAGUAY, PERU. PHILIPPINES, POLAND, PORTUGAL, REPUBLIC OF MACEDONIA, REUNION, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAK REPUBLIC, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SUDAN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TUNISIA, TUR,EKY, UKRAINE, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIETNAM, YUGOSLAVIA, MALAWI, MAURITANIA, NIGERIA, UKRAINE, IVORY COST, OMAN, ZAMBIA, UZBEKISTAN, ESTONIA, IRAQ, and JAMAICA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.