| | Class 2 Device Recall ACL TOP CTS |  |
| Date Initiated by Firm | June 03, 2014 |
| Date Posted | July 16, 2014 |
| Recall Status1 |
Terminated 3 on September 21, 2017 |
| Recall Number | Z-2062-2014 |
| Recall Event ID |
68578 |
| 510(K)Number | K091980 |
| Product Classification |
Instrument, coagulation, automated - Product Code GKP
|
| Product | ACL TOP CTS, automated coagulation laboratory instrument. |
| Code Information |
Start SN and Date: 06020101 Feb 2006 End SN and Date 10080662 Aug 2010 |
| FEI Number |
1217183
|
Recalling Firm/ Manufacturer |
Instrumentation Laboratory Co. 180 Hartwell Road Bedford MA 01730-2443
|
| For Additional Information Contact | 800-678-0710 |
Manufacturer Reason for Recall | Potential for sample misidentification. |
FDA Determined Cause 2 | Device Design |
| Action | Consignees were notified via phone starting on June 3, 2014. On June 4, a USPS mailing was sent to all US/Canadian customers. On June 4, 2014, email notifications were sent to the IL Affiliates and International Distributors/Dealers (through the International Group) requesting them to translate the Field Safety Notice and provide notification to their customer base and secondary distributors.
The Urgent Product Correction letter informs consignees of the issue and provides workaround instructions. Consignees are asked to complete a Mandatory Response Tracking Record and return it via Fax (781)861-4207 or email: ra-usa@ilww.com. For technical questions, consignees should contact the IL Technical Support Center at 1-800-678-0710. |
| Quantity in Commerce | 572 |
| Distribution | Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV, and the countries of Canada, Mexico, Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Brazil, Brunei, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, El Salvador, Finland, France, Germany, Gibraltar, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Libya, Lithuania, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Quatar, Romania, Russia, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UAE, Uruguay, Venezuela, Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = GKP
|
|
|
|