 Class 1 Device Recall Customedsee related information
Date Initiated by FirmMay 20, 2014
Date PostedAugust 29, 2014
Recall Status1 Terminated 3 on August 14, 2017
Recall NumberZ-2432-2014
Recall Event ID 68536
Product Classification Orthopedic tray - Product Code OJH
ProductORTHOPEDIC SURGICAL PACK HOSPITAL SAN CRISTOBAL CONTENTS: (I) CAUTERY TIP POLISH ER LIF (1) BAG SUTURE FLORAL (1) TUBE SUCTION CONNECT W' X 1 2' LIF (1) BAG GLASSINE (2) LITE GLOVE (1) TABLE COVER BTC 77" X 11 0" HD FULL COVERAGE LIF (1) MAYO STAND COVER REINFORCED LIF (5) TOWELS ABSORBENT 1 5" X 20" LIF (4) DRAPE UTILITY WITH TAPE LIF ( 1 ) DRAPE SHEET 70" X 100" L/ F (5) LAP SPONGE PREWASH 18" X 1 8" XRD LIF (1 ) MAYOTRAYLARGE (1) DRAPE TOP WITH ADHESIVE 108" X 50" STD SMS LIF (1) SKIN STAPLER WIDE 35 LIF ( 1) DRAPE INCISE ANTI M I CROBIAL 23" X 1 7" LIF (1) SKIN MARKER WITH RULER (1) U-DRAPE 60" X 70" WITH TAPE SPLIT 6" X 2 1 " LIF (1 ) PAD ABDOM INAL 8" X 7"h.'' (1) YAUNKAUER SUCTION TUBE WITHOUT VENT LIF (1) SOLUTI ON SURG ICAL DURAPREP 26ML LIF (1 ) BLADE SURG ICAL #10 STAINLESS STEEL (I) SPLIT SHEET WITH ADHESIVE 108" X 77" STD SMS LIF (2) *Prs. GLOVE # 8.5 SURGEON NEUTRALON LATEX ( I ) *Pr. GLOVE# 8 SURGEON NEUTRALON LATEX ( I ) GOWN SURG. REINFORCED XL TOWEL/WRAP AAMI Ill (3) GOWN XL SMS IM PERVIOUS REINFORCED AAM I III LIF ( I ) CAUTERY PENCIL ROCK ER SWITCH (2) *Prs. GLOVES SURG. # 8 POWDER (4) Prs. GLOVES SURG. 8.5 Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Code Information Product code 900-2039, 7 lots: 112062199 113026170 113057698 113099418 131110232 140311817 140312114 
FEI Number 2648727
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
For Additional Information ContactRosemari Melendez
787-622-5151 Ext. 7540
Manufacturer Reason
for Recall
Customed has determined that there is the possibility that packaging integrity may be compromised on differ catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
FDA Determined
Cause 2
Package design/selection
ActionCustomed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity in Commerce7 lots/236 units, multiple units per lot
DistributionWorldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-