 Class 1 Device Recall Customedsee related information
Date Initiated by FirmMay 20, 2014
Date PostedAugust 29, 2014
Recall Status1 Terminated 3 on August 14, 2017
Recall NumberZ-2436-2014
Recall Event ID 68536
Product Classification Orthopedic tray - Product Code OJH
ProductFEMORAL PACK CONTENTS: (2) TOWELS ABSORBENT 15" X 20" LIF (2) DRAPE SHEET 42" X 57" SMS LIF (1) SAFETY IV INTROCAN 18 X 1X ST. LIF (2) GOWN LGE SMS IMPERVIOUS REINFORCED LIF (2) VESSEL LOOP SURGICAL MAXI BLUE (1) SUTURE AID BOOT STD YELLOW ST. LIF (1) YANKAUER SUCTION TUBE WITHOUT VENT L/F (1) EMESIS BASIN 101N 700cc LIF (2) LIGHT SHIELD (1) NEEDLE HYPODERMIC 18G X 1 L/F (1) CATHETER IV SAFETY 20G X 1X LIF (1) DRAPE T LAPAROT 102" X 78" X 121" SMS STD LIF (10) LAP SPONGE PREWASH 18" X 18" XRD LIF (1) MAYO STAND COVER REINFORCED LIF (1) STOCKINETTE IMPERVIOUS 14" X 48" LIF (2) SYRINGE 20cc WITHOUT NEEDLE LUER I LOCK LIF (1) SUTURE BAG FLORAL LIF (2) UTILITY BOWL 16oz. L/F (1) CAUTERY TIP POLISHER LIF (1) TABLE COVER REINFORCED 50" X 90" LIF (4) DRAPE UTILITY WITH TAPE LIF (1) MAYO TRAY SMALL (1) SCALPEL SAFETY WEIGHTED #15 (1) SCALPEL SAFETY WEIGHTED #11 (2) TUBE SUCTION CONNECT X" x 12' LIF (2) NEEDLE & BLADE COUNTER, 1Oc MAG/CLEAR LIF (1) TIME OUT BEACON NON WOVEN ST. LIF (1) SKIN MARKER INK W/8 LABEL/TIME OUT/RULER L/F Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Code Information Product code 900-2164, 11 lots 112041224 112083167 113036762 113047217 113067914 113078468 113088852 131210646 140312118 140412645 140513125 
FEI Number 2648727
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
For Additional Information ContactRosemari Melendez
787-622-5151 Ext. 7540
Manufacturer Reason
for Recall
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
FDA Determined
Cause 2
Package design/selection
ActionCustomed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity in Commerce11 lots/432 units, (multiple units per lot)
DistributionWorldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-