• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BrightView XCT System Tomography Computed Emission

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BrightView XCT System Tomography Computed Emissionsee related information
Date Initiated by FirmOctober 23, 2013
Date PostedDecember 12, 2014
Recall Status1 Terminated 3 on March 01, 2021
Recall NumberZ-0203-2015
Recall Event ID 69304
510(K)NumberK080927 
Product Classification System, tomography, computed, emission - Product Code KPS
ProductBrightView XCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates with an attenuation device consisting of flat panel x-ray imaging components.
Code Information Brightview XCT
FEI Number 1218950
Recalling Firm/
Manufacturer
Philips Medical Systems, Inc.
3000 Minuteman Rd
Andover MA 01810-1032
For Additional Information Contact
978-687-1501
Manufacturer Reason
for Recall
It was discovered the CT portion of the scan was interrupted ( stopping the CT exposure at the time of interruption) and did not complete successfully.
FDA Determined
Cause 2
Radiation Control for Health and Safety Act
ActionPhilips sent an Important Electronic Product Radiation Warning letter dated November 4, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Action to be taken by the customer: Please be aware that if this condition occurs, users will not be able to detect the situation prior to the interruption of the CT acquisition. However, if the INFORMATION message appears during a clinical study indicating to the operator that the scan did not complete successfully, Philips advises users to understand the following available options, and use their clinical judgment to determine which option best suits their clinical needs: The operator chooses not to rescan the patient: Continue with the SPECT portion of the of the scan and use only the portion of the segments that were acquired prior to the interruption for interpretation; or The operator chooses to rescan the patient: Retry the interrupted segment acquisition. Philips Healthcare is initiating a corrective action consisting of : Distribution of this Field Safety Notice 88200485_486, Conducting the appropriate field safety correction will address the above issue that has been identified. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center (1-800-722-9377: follow the prompts).
Quantity in Commerce344 units worldwide, with approximately 114 units in the United States.
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico and the states of AZ, CA, CO, DC, DE, FL, GA, IL, MA, MD, MN, MO, MS, MT, NC, NJ, NY, OH, OK, OR, PA, TX, VT, WA and WI., and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Finland, France, Gabon, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Russia, Saudi Arabia, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom and Viet Nam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = KPS
-
-