 Class 1 Device Recall Laparotomy Surgical Packsee related information
Date Initiated by FirmOctober 08, 2014
Date PostedDecember 16, 2014
Recall Status1 Terminated 3 on August 14, 2017
Recall NumberZ-0362-2015
Recall Event ID 69448
Product Classification General Surgery Tray - Product Code LRO
ProductLaparotomy Surgical Pack, catalog #900-281A Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
Code Information Lot numbers 140412669 140412796 140613735 140714265 
FEI Number 2648727
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
For Additional Information Contact
787-622-5151
Manufacturer Reason
for Recall
Significant compromised sterility, not limited to potentially damaged packaging. Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.
FDA Determined
Cause 2
Packaging
ActionCustomed initiated a recall on October 8th, 2014 of sterile convenience surgical Packs-trays, bags (known as Convenience Packs) and new lots of products added to the initial June 3, 2014 recall of 233 sterile surgical packs. The products are being recalled because of significant compromised sterility, not limited to potentially damaged packaging. Customers were instructed to quarantine and cease use of the affected product identified. Customers are to return the response form included with the letter, to the contact information provided. Upon receipt, the firm will contact customers with further instructions to obtain credit, replacement product, and for return of the recalled product or any other disposition action. If customers have any questions, they should contact the Customed Recall Coordinator at 407-850-5558 extensions 8001 or 8005, or via e-mail: pablo.diaz@customedhealing.com.
Quantity in Commerce852
DistributionUS Distribution to Florida, New York and Puerto Rico..
Total Product Life CycleTPLC Device Report