 Class 1 Device Recall INRatio Monitorssee related information
Date Initiated by FirmDecember 05, 2014
Date PostedJanuary 03, 2015
Recall Status1 Terminated 3 on September 18, 2017
Recall NumberZ-0881-2015
Recall Event ID 69964
510(K)NumberK020679 K021923 K092987 
Product Classification Test, time, prothrombin - Product Code GJS
ProductINRatio Monitors - 0100004 Alere INRatio PT/INR System Professional - 0100007 INRatio Prothrombin Time (PT) Monitoring System - 0100137 Alere INRatio Replacement Monitor The test systems contain individually packaged components needed to use the product, such as a monitor, user guide, quick reference guide training DVD or CD, Lancets (home monitoring kits contain an Autolet, Unilet, gauze pads, and alcohol wipes), power supply (home monitoring kits contain AA batteries), results log book and a warranty card. None of the components, except for the lancets, are sterile. In vitro diagnostic monitoring system used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood.
Code Information All lots of part numbers: 0100004, 0100007, and 0100137.
FEI Number 1000125596
Recalling Firm/
Manufacturer
Alere San Diego, Inc.
9975 Summers Ridge Rd
San Diego CA 92121-2997
For Additional Information ContactJackie Lustig
781-341-4009
Manufacturer Reason
for Recall
In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn Urgent: Medical Device Correction, dated December 5, 2014, was sent to Healthcare Professionals. The letter identified the affected product and reason for recall. Customers were instructed to: -Ensure they have read and understand the precautions described in the current product labeling (a complete list of product labeling is provided in Appendix A) and the additional precautions in this notice describing medical conditions that may increase the risk of obtaining a (falsely or erroneously) lower than expected INR result. The INRatio PT/INR Monitor system should NOT be used if the patient has any of the medical conditions described. -Verify that the patient has hematocrit within the range of 30% to 55%. If not, immediately transition the patient to an alternate INR monitoring method. -Perform INR verification testing for the patients using a laboratory INR test method. Immediately transition any patient with a significantly discrepant low result on the INRatio PT/INR Monitor system to an alternative INR monitoring method. -If have forwarded product to another customer, please provide a copy of this letter to them. -Please complete and FAX or e-mail the enclosed Reply Form (Appendix B the last page of this notice) within 10 days to confirm your receipt of this notice. Customers can contact Alere INRatio Recall Hotline by phone at 1-877-929-2579, for questions regarding this notification. Additionally, a website has been established providing information and frequently asked questions. www.inr-care.com Please FAX or e-mail the completed Reply Form to: Alere San Diego, Inc. Fax: 1-877-929-2580 E-mail: Alere4319@stericycle.com.
Quantity in Commerce231,967 Test Systems including Monitors
DistributionWorldwide Distribution -- US nationwide, Canada, South America, EU, Asia Pacific, Africa, and Middle East.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = GJS