 Class 2 Device Recall CATARACT PACKsee related information
Date Initiated by FirmJuly 10, 2015
Date PostedSeptember 03, 2015
Recall Status1 Terminated 3 on April 12, 2017
Recall NumberZ-2713-2015
Recall Event ID 71721
Product Classification Eye tray - Product Code OJK
ProductCATARACT PACK 6/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate
Code Information Code: 900944 Lots: 141015772 exp. 10/31/15 141015772 exp. 10/31/15 150116981 exp. 1/31/16 150116827 exp. 2/28/16 150116828 exp. 2/28/16 150116829 exp. 2/28/16 150317866 exp. 4/30/16 150317867 exp. 4/30/16 150317868 exp. 4/30/16 150317869 exp. 4/30/16 150418428 exp. 4/30/16 150418426 exp. 4/30/16 150418427 exp. 4/30/16 150418585 exp. 5/31/16 150418586 exp. 5/31/16 150619298 exp. 6/30/16 150619299 exp. 6/30/16 
FEI Number 2648727
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
Manufacturer Reason
for Recall
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
FDA Determined
Cause 2
Packaging
ActionCustomed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer.
Quantity in Commerce1038
DistributionDistributed Only in Puerto Rico.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.