Class 2 Device Recall Envoy 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Envoy 500see related information
Date Initiated by FirmSeptember 22, 2015
Date PostedNovember 05, 2015
Recall Status1 Terminated 3 on March 04, 2016
Recall NumberZ-0217-2016
Recall Event ID 72322
510(K)NumberK100525 
Product Classification Glucose oxidase, glucose - Product Code CGA
ProductEnvoy 500 system includes the following product reference numbers and names: 55345 ENVOY 500 Glucose 55450 ENVOY 500 URIC ACID REAGENT KIT 55435 ENVOY 500 TRIGLYCERIDES REAGENT KIT, 55297 ENVOY 500 CHOLESTEROL REAGENT KIT, 55335 ENVOY 500 CREATININE REAGENT KIT 55301 ENVOY 500 HDL CHOLESTEROL REAGENT 55310 ENVOY 500 LDL CHOLESTEROL REAGENT KIT
Code Information all lots
Recalling Firm/
Manufacturer
ELITech Clinical Systems SAS
Zone lndustrielle
SEES France
For Additional Information ContactCustomer Care
800-345-2822
Manufacturer Reason
for Recall
ELITech clinical Systems reagents based on Trinder reaction used on the Envoy 500 instrument system are recalled due to potential drug interference. No complaints were received.
FDA Determined
Cause 2
Under Investigation by firm
ActionThe firm sent recall letters or Technical Bulletins to US consignees on 9/22/15 via registered mail. Technical Bulletin #1022 (for EON 100 users) Technical Bulletin #2008 (for EON 300 users) Technical Bulletin #5064 (for ENVOY users) Technical Bulletin #3009 (for Selectra users) ELITechGroup is not requesting return of the affected product. The method sheets and package inserts will be updated with the new interference claims. When available, the revised labeling will be provided to the customer for customer's use until the new IFU is included in the product packaging. Consignees are requested to return the Confirmation of Receipt form back to the firm by FAX to 401-642-9001 or send email to customercare@elitechgroup.com or send via US Post: Vital Diagnostics, 27 Wellington Rd., Lincoln, Rl. If you have any questions, please contact Vital Diagnostics Technical Support at 1-800-345-2822 or 855-ELITECH
Quantity in Commerce13455 systems
DistributionUS Nationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = CGA
-
-