• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall SlipCath Beacon Tip Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall SlipCath Beacon Tip Cathetersee related information
Date Initiated by FirmApril 15, 2016
Date PostedAugust 22, 2016
Recall Status1 Terminated 3 on August 04, 2017
Recall NumberZ-2614-2016
Recall Event ID 74010
510(K)NumberK122937 
Product Classification Catheter, intravascular, diagnostic - Product Code DQO
ProductSlip-Cath Beacon Tip Catheter Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques.
Code Information All Lots. Catalog prefixes SCBR5.0, SCBR5.5, and SCBR6.5  Catalog/product number; Global/order number SCBR5.0-38-100-P-NS-0 G12005 SCBR5.0-38-100-P-NS-C2 G12008 SCBR5.0-38-100-P-NS-DAV G11996 SCBR5.0-38-100-P-NS-H1 G11991 SCBR5.0-38-100-P-NS-HN4 G23693 SCBR5.0-38-100-P-NS-JB1 G11997 SCBR5.0-38-100-P-NS-JB2 G11998 SCBR5.0-38-100-P-NS-KMP G28535 SCBR5.0-38-100-P-NS-MAN G11999 SCBR5.0-38-100-P-NS-MPA G12002 SCBR5.0-38-100-P-NS-NTR G23695 SCBR5.0-38-100-P-NS-RVC G43120 SCBR5.0-38-100-P-NS-SIM1 G12000 SCBR5.0-38-100-P-NS-SIM2 G11992 SCBR5.0-38-100-P-NS-SIM3 G12001 SCBR5.0-38-100-P-NS-TEGT G23694 SCBR5.0-38-100-P-NS-VERT G12007 SCBR5.0-38-100-P-NS-VTK G13163 SCBR5.0-38-125-P-NS-0 G32366 SCBR5.0-38-125-P-NS-DAV G13413 SCBR5.0-38-125-P-NS-H1 G23696 SCBR5.0-38-125-P-NS-JB1 G13530 SCBR5.0-38-125-P-NS-SIM1 G32369 SCBR5.0-38-125-P-NS-SIM2 G13200 SCBR5.0-38-125-P-NS-TEGT G32367 SCBR5.0-38-125-P-NS-VERT G12168 SCBR5.0-38-125-P-NS-VTK G12082 SCBR5.0-38-40-P-NS-KMP G11993 SCBR5.0-38-65-P-NS-0 G12009 SCBR5.0-38-65-P-NS-C1 G12003 SCBR5.0-38-65-P-NS-C2 G11994 SCBR5.0-38-65-P-NS-C3 G12004 SCBR5.0-38-65-P-NS-DAV G23688 SCBR5.0-38-65-P-NS-H1 G13319 SCBR5.0-38-65-P-NS-KMP G12156 SCBR5.0-38-65-P-NS-LEV1 G12798 SCBR5.0-38-65-P-NS-MPA G23689 SCBR5.0-38-65-P-NS-RC2 G13272 SCBR5.0-38-65-P-NS-RDC G13092 SCBR5.0-38-65-P-NS-RIM G11995 SCBR5.0-38-65-P-NS-SIM1 G12716 SCBR5.0-38-65-P-NS-SIM2 G23691 SCBR5.0-38-65-P-NS-TC G36320 SCBR5.0-38-65-P-NS-TC-BNK G36322 SCBR5.0-38-65-P-NS-TEGT G12179 SCBR5.0-38-80-P-NS-C2 G12006 SCBR5.0-38-80-P-NS-MPA G23692 SCBR5.0-38-80-P-NS-RH G36214 SCBR5.0-38-80-P-NS-VS G13431 SCBR5.0-38-80-P-NS-VS1 G13432 SCBR5.0-38-80-P-NS-VS2 G13433 SCBR5.0-38-80-P-NS-VS3 G13434 SCBR5.5-35-60-P-NS-PATEL G51570 SCBR6.5-35-60-P-NS-PATEL G51571 
FEI Number 1820334
Recalling Firm/
Manufacturer
Cook Inc.
750 N Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactCook Medical Customer Relations Departme
800-457-4500
Manufacturer Reason
for Recall
Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
FDA Determined
Cause 2
Unknown/Undetermined by firm
ActionCook Inc. sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter dated April 15, 2016, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with any medical questions or concerns, were instructed to contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. For information regarding the recall, please contact Stericycle Expert Solutions at 866.201.9067.
Quantity in Commerce298,340
DistributionWorldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DQO
-
-