• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Stratus CS Acute Care Diagnostics System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Stratus CS Acute Care Diagnostics Systemsee related information
Date Initiated by FirmJune 22, 2016
Create DateAugust 11, 2016
Recall Status1 Terminated 3 on November 29, 2018
Recall NumberZ-2525-2016
Recall Event ID 74534
510(K)NumberK984067 
Product Classification Fluorometric method, cpk or isoenzymes - Product Code JHX
ProductStratus CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certain analytes in body fluids for in vitro diagnostic use. Assays include ~hCG, CRP, CKMB, cTNI, D-Dimer, Myoglobin, and pBNP. SMN: 10444834, 10453531
Code Information All software versions and serial numbers
FEI Number 3002637618
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics Inc
2 Edgewater Dr
Norwood MA 02062-4637
For Additional Information Contact
781-269-3000
Manufacturer Reason
for Recall
Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibration and quality control (QC) status when a TestPak is due to expire within 48 hours.
FDA Determined
Cause 2
Software design
ActionSiemens Healthcare issued an Urgent Field Safety Notice by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on June 22, 2016 both in the United States and Outside the US for communication with affected customers. This notice informs customers of the issue and provides mitigation instructions. Customers are to contact Siemens Customer Care Center or their local Siemens technical support representative at 800-405-6473. Siemens Healthcare is currently developing a software update to address this issue and will be providing new information as it becomes available
Quantity in Commerce977 units
DistributionWorldwide Distribution -- USA, U.A.E., Algeria, Armenia, Australia, Austria, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Canada, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Libya, Lithuania, Malaysia, Mexico, Namibia, P.R. China, Poland, Portugal, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = JHX
-
-