• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BrightView XCT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BrightView XCTsee related information
Date Initiated by FirmMay 25, 2016
Create DateSeptember 03, 2016
Recall Status1 Terminated 3 on July 02, 2018
Recall NumberZ-2737-2016
Recall Event ID 74894
510(K)NumberK080927 
Product Classification System, tomography, computed, emission - Product Code KPS
ProductBrightView XCT Upgrade Model 882454
Code Information 11000021, 11000082, 11000111
FEI Number 1525965
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Rd
Cleveland OH 44143-2131
For Additional Information ContactHolly Wright Lee
440-483-7600
Manufacturer Reason
for Recall
The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
FDA Determined
Cause 2
Software design
ActionThe firm is informing the consignee of this issue through Customer Information Letter.
Quantity in Commerce3 units
DistributionWorldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic of Lebanon, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = KPS
-
-