Date Initiated by Firm | September 19, 2016 |
Create Date | October 28, 2016 |
Recall Status1 |
Terminated 3 on June 01, 2017 |
Recall Number | Z-0324-2017 |
Recall Event ID |
75221 |
Product Classification |
Tourniquet, pneumatic - Product Code KCY
|
Product | Reprocessed Pressure Tourniquet Cuff (PTC)
Tourniquet cuffs are single- or dual-bladder inflatable cuffs connected to a tourniquet system via a hose assembly. When wrapped around a limb and inflated, tourniquet cuffs apply an adequate amount of pressure on the arterial blood flow in a limb to create a bloodless surgical field. Tourniquet cuffs are available in a variety of sizes to accommodate a wide range of limb circumferences. |
Code Information |
Model Lot Lot Created Expiration 5921-018-135 2990595 5/5/2014 May-16 5921-018-235 2989898 5/5/2014 May-16 5921-018-235 2992868 5/5/2014 May-16 5921-024-235 2992501 5/5/2014 May-16 5921-030-135 2990147 5/5/2014 May-16 5921-030-135 2991575 5/5/2014 May-16 5921-030-235 2993033 5/5/2014 May-16 5921-034-135 2990152 5/5/2014 May-16 5921-034-135 2992555 5/5/2014 May-16 5921-034-235 2989674 5/5/2014 May-16 5921-034-235 2990240 5/5/2014 May-16 5921-034-235 2991095 5/5/2014 May-16 5921-034-235 2991413 5/5/2014 May-16 5921-034-235 2993046 5/5/2014 May-16 5921-034-235 2993509 5/5/2014 May-16 5921-218-235 2990217 5/5/2014 May-16 5921-218-235 2990553 5/5/2014 May-16 5921-218-235 2991743 5/5/2014 May-16 5921-218-135 2990106 5/5/2014 May-16 5921-218-135 2992540 5/5/2014 May-16 60-7070-102 2990697 5/5/2014 May-16 60-7070-103 2984111 5/5/2014 May-16 60-7070-103 2989473 5/5/2014 May-16 60-7070-103 2990110 5/5/2014 May-16 60-7070-103 2991030 5/5/2014 May-16 60-7070-103 2992398 5/5/2014 May-16 60-7070-103 2992636 5/5/2014 May-16 60-7070-105 2989822 5/5/2014 May-16 60-7070-105 2990012 5/5/2014 May-16 60-7070-105 2991271 5/5/2014 May-16 60-7070-105 2993392 5/5/2014 May-16 60-7070-104 2992643 5/5/2014 May-16 60-7070-106 2990433 5/5/2014 May-16 60-7070-106 2991817 5/5/2014 May-16 60-7070-106 2993407 5/5/2014 May-16 60-7075-004 2989867 5/5/2014 May-16 60-7075-103 2990650 5/5/2014 May-16 60-7075-103 2991432 5/5/2014 May-16 60-7075-103 2992627 5/5/2014 May-16 60-7075-105 2992282 5/5/2014 May-16 60-7075-106 2990451 5/5/2014 May-16 60-7075-106 2990958 5/5/2014 May-16 5921-018-135 2996601 5/8/2014 May-16 5921-024-135 2996808 5/8/2014 May-16 5921-030-235 2995775 5/8/2014 May-16 5921-030-235 2998064 5/8/2014 May-16 5921-034-135 2996324 5/8/2014 May-16 5921-034-235 2994193 5/8/2014 May-16 5921-034-235 2995233 5/8/2014 May-16 5921-034-235 2996975 5/8/2014 May-16 5921-218-235 2994186 5/8/2014 May-16 5921-218-235 2996503 5/8/2014 May-16 5921-218-235 2997742 5/8/2014 May-16 5921-218-135 2996590 5/8/2014 May-16 60-7070-103 2993940 5/8/2014 May-16 60-7070-103 2994661 5/8/2014 May-16 60-7070-103 2995417 5/8/2014 May-16 60-7070-103 2995747 5/8/2014 May-16 60-7070-103 2996779 5/8/2014 May-16 60-7070-103 2996896 5/8/2014 May-16 60-7070-105 2994062 5/8/2014 May-16 60-7070-105 2994707 5/8/2014 May-16 60-7070-105 2995279 5/8/2014 May-16 60-7070-105 2996094 5/8/2014 May-16 60-7070-105 2997735 5/8/2014 May-16 60-7070-105 2997902 5/8/2014 May-16 60-7070-104 2994473 5/8/2014 May-16 60-7070-104 2997034 5/8/2014 May-16 60-7070-003 2997757 5/8/2014 May-16 60-7070-006 2997496 5/8/2014 May-16 60-7070-106 2995752 5/8/2014 May-16 60-7070-106 2996232 5/8/2014 May-16 60-7070-106 2997047 5/8/2014 May-16 60-7070-107 2995995 5/8/2014 May-16 60-7075-003 2994460 5/8/2014 May-16 60-7075-006 2994782 5/8/2014 May-16 60-7075-103 2996143 5/8/2014 May-16 60-7075-103 2996981 5/8/2014 May-16 60-7075-104 2994544 5/8/2014 May-16 60-7075-106 2994879 5/8/2014 May-16 60-7075-106 2997132 5/8/2014 May-16 5921-018-235 3001050 5/12/2014 May-16 5921-024-135 3002574 5/12/2014 May-16 5921-024-235 3000098 5/12/2014 May-16 5921-030-135 2999078 5/12/2014 May-16 5921-030-135 3001370 5/12/2014 May-16 5921-030-235 2999888 5/12/2014 May-16 5921-030-235 2999944 5/12/2014 May-16 5921-030-235 3001009 5/12/2014 May-16 5921-034-135 2998628 5/12/2014 May-16 5921-034-135 2999717 5/12/2014 May-16 5921-034-135 3001084 5/12/2014 May-16 5921-034-136 3001358 5/12/2014 May-16 5921-034-235 2996307 5/12/2014 May-16 5921-034-235 2998231 5/12/2014 May-16 5921-034-235 2999013 5/12/2014 May-16 5921-034-235 3001679 5/12/2014 May-16 5921-034-235 3002934 5/12/2014 May-16 5921-034-236 3001738 5/12/2014 May-16 5921-212-235 3000108 5/12/2014 May-16 5921-218-235 3000804 5/12/2014 May-16 5921-218-235 3001508 5/12/2014 May-16 5921-044-135 3001362 5/12/2014 May-16 5921-218-135 3000316 5/12/2014 May-16 5921-218-136 3001389 5/12/2014 May-16 60-7070-103 2998671 5/12/2014 May-16 60-7070-103 2999294 5/12/2014 May-16 60-7070-103 3000363 5/12/2014 May-16 60-7070-103 3001013 5/12/2014 May-16 60-7070-103 3001622 5/12/2014 May-16 60-7070-103 3002136 5/12/2014 May-16 60-7070-105 2999021 5/12/2014 May-16 60-7070-105 2999343 5/12/2014 May-16 60-7070-105 3001633 5/12/2014 May-16 60-7070-105 3001772 5/12/2014 May-16 60-7070-105 3003024 5/12/2014 May-16 60-7070-104 2999622 5/12/2014 May-16 60-7070-104 3003302 5/12/2014 May-16 60-7070-003 3001434 5/12/2014 May-16 60-7070-004 3001680 5/12/2014 May-16 60-7070-106 2998072 5/12/2014 May-16 60-7070-106 2999336 5/12/2014 May-16 60-7070-106 3000702 5/12/2014 May-16 60-7070-106 3001646 5/12/2014 May-16 60-7070-106 3002526 5/12/2014 May-16 60-7070-107 2998980 5/12/2014 May-16 60-7075-003 2999740 5/12/2014 May-16 60-7075-004 3000067 5/12/2014 May-16 60-7075-005 3000344 5/12/2014 May-16 60-7075-006 2998732 5/12/2014 May-16 60-7075-006 2998914 5/12/2014 May-16 60-7075-102 2999633 5/12/2014 May-16 60-7075-103 2999009 5/12/2014 May-16 60-7075-104 2998778 5/12/2014 May-16 60-7075-106 2998922 5/12/2014 May-16 60-7075-106 3001391 5/12/2014 May-16 60-7075-107 3001740 5/12/2014 May-16 5921-018-135 3006954 5/14/2014 May-16 5921-018-135 3007582 5/14/2014 May-16 5921-018-235 3009132 5/14/2014 May-16 5921-024-135 2984861 5/14/2014 May-16 5921-030-135 3005777 5/14/2014 May-16 5921-030-235 3004709 5/14/2014 May-16 5921-030-235 3008589 5/14/2014 May-16 5921-030-235 3008656 5/14/2014 May-16 5921-030-235 3009386 5/14/2014 May-16 5921-034-135 3005442 5/14/2014 May-16 5921-034-235 3003387 5/14/2014 May-16 5921-034-235 3003704 5/14/2014 May-16 5921-034-235 3006724 5/14/2014 May-16 5921-034-235 3008976 5/14/2014 May-16 5921-218-235 2935644 5/14/2014 May-16 5921-218-235 3003450 5/14/2014 May-16 5921-218-235 3003893 5/14/2014 May-16 5921-218-235 3006439 5/14/2014 May-16 5921-218-235 3008090 5/14/2014 May-16 5921-218-235 3009099 5/14/2014 May-16 5921-044-235 3003840 5/14/2014 May-16 5921-218-135 3003957 5/14/2014 May-16 60-7070-103 3003935 5/14/2014 May-16 60-7070-103 3005194 5/14/2014 May-16 60-7070-103 3006603 5/14/2014 May-16 60-7070-103 3008660 5/14/2014 May-16 60-7070-105 2935323 5/14/2014 May-16 60-7070-105 3004489 5/14/2014 May-16 60-7070-105 3006091 5/14/2014 May-16 60-7070-105 3007991 5/14/2014 May-16 60-7070-105 3009113 5/14/2014 May-16 60-7070-104 3006949 5/14/2014 May-16 60-7070-104 3008429 5/14/2014 May-16 60-7070-106 3003552 5/14/2014 May-16 60-7070-106 3005488 5/14/2014 May-16 60-7070-106 3006384 5/14/2014 May-16 60-7070-106 3006487 5/14/2014 May-16 60-7070-106 3007229 5/14/2014 May-16 60-7070-106 3008969 5/14/2014 May-16 60-7070-106 3009109 5/14/2014 May-16 60-7070-106 3009291 5/14/2014 May-16 60-7070-107 3007763 5/14/2014 May-16 60-7075-005 3004539 5/14/2014 May-16 60-7075-006 3003829 5/14/2014 May-16 60-7075-103 3003048 5/14/2014 May-16 60-7075-103 3006494 5/14/2014 May-16 60-7075-104 3005765 5/14/2014 May-16 60-7075-105 3006193 5/14/2014 May-16 60-7075-106 3004573 5/14/2014 May-16 60-7075-106 3004721 5/14/2014 May-16 5921-018-135 3013487 5/16/2014 May-16 5921-024-135 3009925 5/16/2014 May-16 5921-030-135 3011981 5/16/2014 May-16 5921-030-135 3013460 5/16/2014 May-16 5921-030-235 3010399 5/16/2014 May-16 5921-030-235 3012499 5/16/2014 May-16 5921-030-235 3014359 5/16/2014 May-16 5921-034-135 3009420 5/16/2014 May-16 5921-034-135 3010671 5/16/2014 May-16 5921-034-135 3012218 5/16/2014 May-16 5921-034-136 3012819 5/16/2014 May-16 5921-034-235 3011975 5/16/2014 May-16 5921-034-235 3013688 5/16/2014 May-16 5921-034-236 3012826 5/16/2014 May-16 5921-218-235 3011952 5/16/2014 May-16 5921-218-235 3012512 5/16/2014 May-16 5921-044-135 3013456 5/16/2014 May-16 5921-218-135 3011330 5/16/2014 May-16 5921-218-135 3012531 5/16/2014 May-16 5921-218-136 3012835 5/16/2014 May-16 60-7070-102 3010017 5/16/2014 May-16 60-7070-102 3010310 5/16/2014 May-16 60-7070-103 3009412 5/16/2014 May-16 60-7070-103 3010055 5/16/2014 May-16 60-7070-103 3011561 5/16/2014 May-16 60-7070-103 3012362 5/16/2014 May-16 60-7070-103 3012549 5/16/2014 May-16 60-7070-103 3013851 5/16/2014 May-16 60-7070-105 3010352 5/16/2014 May-16 60-7070-105 3012358 5/16/2014 May-16 60-7070-105 3013661 5/16/2014 May-16 60-7070-105 3014870 5/16/2014 May-16 60-7070-104 3010980 5/16/2014 May-16 60-7070-104 3013036 5/16/2014 May-16 60-7070-104 3014578 5/16/2014 May-16 60-7070-003 3009408 5/16/2014 May-16 60-7070-003 3014051 5/16/2014 May-16 60-7070-005 3011353 5/16/2014 May-16 60-7070-006 3010029 5/16/2014 May-16 60-7070-006 3013350 5/16/2014 May-16 60-7070-106 3011705 5/16/2014 May-16 60-7070-106 3012916 5/16/2014 May-16 60-7070-106 3013062 5/16/2014 May-16 60-7070-106 3013898 5/16/2014 May-16 60-7070-107 3013674 5/16/2014 May-16 60-7075-003 3009372 5/16/2014 May-16 60-7075-003 3014584 5/16/2014 May-16 60-7075-004 3012842 5/16/2014 May-16 60-7075-006 3010784 5/16/2014 May-16 60-7075-006 3010922 5/16/2014 May-16 60-7075-104 3012900 5/16/2014 May-16 60-7075-106 3009121 5/16/2014 May-16 60-7075-106 3013439 5/16/2014 May-16 60-7075-107 3014067 5/16/2014 May-16 5921-018-235 3020828 5/20/2014 May-16 5921-024-235 3017331 5/20/2014 May-16 5921-030-135 3015291 5/20/2014 May-16 5921-030-235 3016575 5/20/2014 May-16 5921-030-235 3020317 5/20/2014 May-16 5921-034-135 3014862 5/20/2014 May-16 5921-034-135 3015372 5/20/2014 May-16 5921-034-135 3018293 5/20/2014 May-16 5921-034-135 3018728 5/20/2014 May-16 5921-034-235 3015075 5/20/2014 May-16 5921-034-235 3016568 5/20/2014 May-16 5921-034-235 3019276 5/20/2014 May-16 5921-034-235 3019754 5/20/2014 May-16 5921-034-235 3019932 5/20/2014 May-16 5921-034-235 3020551 5/20/2014 May-16 5921-218-235 3016285 5/20/2014 May-16 5921-218-235 3018110 5/20/2014 May-16 5921-218-235 3019950 5/20/2014 May-16 5921-044-235 3020863 5/20/2014 May-16 5921-218-135 3015081 5/20/2014 May-16 5921-218-135 3018303 5/20/2014 May-16 5921-218-236 3015624 5/20/2014 May-16 60-7070-103 3015068 5/20/2014 May-16 60-7070-103 3016715 5/20/2014 May-16 60-7070-103 3017019 5/20/2014 May-16 60-7070-103 3018335 5/20/2014 May-16 60-7070-103 3020034 5/20/2014 May-16 60-7070-105 3016151 5/20/2014 May-16 60-7070-105 3019501 5/20/2014 May-16 60-7070-104 3018818 5/20/2014 May-16 60-7070-003 3018666 5/20/2014 May-16 60-7070-005 3021080 5/20/2014 May-16 60-7070-106 2985131 5/20/2014 May-16 60-7070-106 3016886 5/20/2014 May-16 60-7070-106 3017706 5/20/2014 May-16 60-7070-106 3019156 5/20/2014 May-16 60-7070-107 3015630 5/20/2014 May-16 60-7070-107 3020844 5/20/2014 May-16 60-7075-006 3020892 5/20/2014 May-16 60-7080-102 3015227 5/20/2014 May-16 60-7075-103 3016445 5/20/2014 May-16 60-7075-103 3018123 5/20/2014 May-16 60-7075-105 3018264 5/20/2014 May-16 60-7075-106 3015364 5/20/2014 May-16 60-7075-106 3015682 5/20/2014 May-16 60-7075-106 3016484 5/20/2014 May-16 60-7075-106 3020598 5/20/2014 May-16 5921-024-135 3021448 5/21/2014 May-16 5921-024-235 3021476 5/21/2014 May-16 5921-024-235 3022897 5/21/2014 May-16 5921-030-135 3022504 5/21/2014 May-16 5921-030-136 3023374 5/21/2014 May-16 5921-030-235 3022074 5/21/2014 May-16 5921-030-235 3024094 5/21/2014 May-16 5921-034-135 3023841 5/21/2014 May-16 5921-034-135 3025072 5/21/2014 May-16 5921-034-235 3021219 5/21/2014 May-16 5921-034-235 3022233 5/21/2014 May-16 5921-034-235 3023943 5/21/2014 May-16 5921-218-235 3021521 5/21/2014 May-16 5921-218-235 3025174 5/21/2014 May-16 5921-044-135 3025038 5/21/2014 May-16 5921-218-135 3024209 5/21/2014 May-16 60-7070-102 3020871 5/21/2014 May-16 60-7070-103 3021430 5/21/2014 May-16 60-7070-103 3022171 5/21/2014 May-16 60-7070-103 3022377 5/21/2014 May-16 60-7070-103 3022866 5/21/2014 May-16 60-7070-103 3024382 5/21/2014 May-16 60-7070-103 3025685 5/21/2014 May-16 60-7070-105 3022019 5/21/2014 May-16 60-7070-105 3023785 5/21/2014 May-16 60-7070-105 3025119 5/21/2014 May-16 60-7070-104 3017320 5/21/2014 May-16 60-7070-104 3022694 5/21/2014 May-16 60-7070-004 3023335 5/21/2014 May-16 60-7070-005 3021344 5/21/2014 May-16 60-7070-006 3023181 5/21/2014 May-16 60-7070-106 3021226 5/21/2014 May-16 60-7070-106 3021929 5/21/2014 May-16 60-7070-106 3022850 5/21/2014 May-16 60-7070-106 3023716 5/21/2014 May-16 60-7070-106 3025029 5/21/2014 May-16 60-7070-106 3025220 5/21/2014 May-16 60-7070-107 3023264 5/21/2014 May-16 60-7075-003 3023989 5/21/2014 May-16 60-7075-005 3016649 5/21/2014 May-16 60-7075-006 3023839 5/21/2014 May-16 60-7080-102 3022591 5/21/2014 May-16 60-7075-103 3022312 5/21/2014 May-16 60-7075-103 3024056 5/21/2014 May-16 60-7075-104 3021203 5/21/2014 May-16 60-7075-106 3022354 5/21/2014 May-16 60-7075-106 3025161 5/21/2014 May-16 5921-018-135 3029394 5/27/2014 May-16 5921-024-135 3028397 5/27/2014 May-16 5921-024-235 3026224 5/27/2014 May-16 5921-024-235 3030968 5/27/2014 May-16 5921-030-135 3026211 5/27/2014 May-16 5921-030-235 3026085 5/27/2014 May-16 5921-030-235 3028129 5/27/2014 May-16 5921-030-235 3031570 5/27/2014 May-16 5921-034-135 3027531 5/27/2014 May-16 5921-034-135 3029095 5/27/2014 May-16 5921-034-135 3030896 5/27/2014 May-16 5921-034-136 3025671 5/27/2014 May-16 5921-034-136 3030590 5/27/2014 May-16 5921-034-235 3026118 5/27/2014 May-16 5921-034-235 3026783 5/27/2014 May-16 5921-034-235 3029510 5/27/2014 May-16 5921-034-235 3029864 5/27/2014 May-16 5921-034-235 3030170 5/27/2014 May-16 5921-034-236 3026540 5/27/2014 May-16 5921-212-235 3029179 5/27/2014 May-16 5921-218-235 3026461 5/27/2014 May-16 5921-218-235 3026777 5/27/2014 May-16 5921-218-235 3030923 5/27/2014 May-16 5921-218-235 3031845 5/27/2014 May-16 5921-044-235 3029523 5/27/2014 May-16 5921-218-135 3027078 5/27/2014 May-16 5921-218-135 3027391 5/27/2014 May-16 5921-218-135 3029957 5/27/2014 May-16 5921-218-135 3031440 5/27/2014 May-16 5921-218-136 3026529 5/27/2014 May-16 60-7070-103 3025914 5/27/2014 May-16 60-7070-103 3026712 5/27/2014 May-16 60-7070-103 3027111 5/27/2014 May-16 60-7070-103 3028091 5/27/2014 May-16 60-7070-103 3028996 5/27/2014 May-16 60-7070-103 3029341 5/27/2014 May-16 60-7070-103 3030542 5/27/2014 May-16 60-7070-105 3026457 5/27/2014 May-16 60-7070-105 3026720 5/27/2014 May-16 60-7070-105 3027520 5/27/2014 May-16 60-7070-105 3030428 5/27/2014 May-16 60-7070-104 3027169 5/27/2014 May-16 60-7070-104 3030947 5/27/2014 May-16 60-7070-003 3025830 5/27/2014 May-16 60-7070-005 3029838 5/27/2014 May-16 60-7070-106 3025980 5/27/2014 May-16 60-7070-106 3027370 5/27/2014 May-16 60-7070-106 3028635 5/27/2014 May-16 60-7070-106 3030162 5/27/2014 May-16 60-7070-107 3025848 5/27/2014 May-16 60-7070-107 3027786 5/27/2014 May-16 60-7070-107 3029942 5/27/2014 May-16 60-7075-003 3028402 5/27/2014 May-16 60-7075-004 3027130 5/27/2014 May-16 60-7075-006 3028321 5/27/2014 May-16 60-7075-006 3031279 5/27/2014 May-16 60-7075-102 3025784 5/27/2014 May-16 60-7075-103 3031438 5/27/2014 May-16 60-7075-104 3027124 5/27/2014 May-16 60-7075-106 3028117 5/27/2014 May-16 60-7075-106 3028313 5/27/2014 May-16 60-7075-106 3029167 5/27/2014 May-16 60-7075-106 3031543 5/27/2014 May-16 60-7075-106 3031768 5/27/2014 May-16 60-7075-107 3029146 5/27/2014 May-16 60-7075-107 3031251 5/27/2014 May-16 5921-018-135 3039220 5/29/2014 May-16 5921-018-235 3034408 5/29/2014 May-16 5921-024-135 3040193 5/29/2014 May-16 5921-024-135 3040313 5/29/2014 May-16 5921-024-235 3035743 5/29/2014 May-16 5921-024-235 3039323 5/29/2014 May-16 5921-030-135 3037252 5/29/2014 May-16 5921-030-135 3037473 5/29/2014 May-16 5921-030-136 3040908 5/29/2014 May-16 5921-030-235 3033662 5/29/2014 May-16 5921-030-235 3035198 5/29/2014 May-16 5921-030-235 3038667 5/29/2014 May-16 5921-030-235 3040384 5/29/2014 May-16 5921-034-135 3032079 5/29/2014 May-16 5921-034-135 3034262 5/29/2014 May-16 5921-034-135 3035497 5/29/2014 May-16 5921-034-135 3039824 5/29/2014 May-16 5921-034-135 3041102 5/29/2014 May-16 5921-034-235 3033187 5/29/2014 May-16 5921-034-235 3034178 5/29/2014 May-16 5921-034-235 3034301 5/29/2014 May-16 5921-034-235 3035926 5/29/2014 May-16 5921-034-235 3037366 5/29/2014 May-16 5921-034-235 3040657 5/29/2014 May-16 5921-034-235 3041092 5/29/2014 May-16 5921-212-235 3040172 5/29/2014 May-16 5921-218-235 3031911 5/29/2014 May-16 5921-218-235 3033379 5/29/2014 May-16 5921-218-235 3034386 5/29/2014 May-16 5921-218-235 3036335 5/29/2014 May-16 5921-218-235 3037642 5/29/2014 May-16 5921-218-235 3039637 5/29/2014 May-16 5921-218-235 3040418 5/29/2014 May-16 5921-218-235 3041295 5/29/2014 May-16 5921-044-135 3032739 5/29/2014 May-16 5921-044-235 3036524 5/29/2014 May-16 5921-218-135 3034669 5/29/2014 May-16 5921-218-135 3037223 5/29/2014 May-16 5921-218-135 3039126 5/29/2014 May-16 60-7070-102 3032702 5/29/2014 May-16 60-7070-103 3031924 5/29/2014 May-16 60-7070-103 3033173 5/29/2014 May-16 60-7070-103 3034372 5/29/2014 May-16 60-7070-103 3034546 5/29/2014 May-16 60-7070-103 3035484 5/29/2014 May-16 60-7070-103 3036936 5/29/2014 May-16 60-7070-103 3037441 5/29/2014 May-16 60-7070-103 3038508 5/29/2014 May-16 60-7070-103 3039671 5/29/2014 May-16 60-7070-103 3040559 5/29/2014 May-16 60-7070-103 3041501 5/29/2014 May-16 60-7070-105 3032482 5/29/2014 May-16 60-7070-105 3032794 5/29/2014 May-16 60-7070-105 3034648 5/29/2014 May-16 60-7070-105 3036217 5/29/2014 May-16 60-7070-105 3038916 5/29/2014 May-16 60-7070-105 3040268 5/29/2014 May-16 60-7070-104 3033380 5/29/2014 May-16 60-7070-104 3035758 5/29/2014 May-16 60-7070-104 3039830 5/29/2014 May-16 60-7070-003 3034994 5/29/2014 May-16 60-7070-003 3037281 5/29/2014 May-16 60-7070-006 3037275 5/29/2014 May-16 60-7070-106 3032448 5/29/2014 May-16 60-7070-106 3032453 5/29/2014 May-16 60-7070-106 3033643 5/29/2014 May-16 60-7070-106 3034859 5/29/2014 May-16 60-7070-106 3035011 5/29/2014 May-16 60-7070-106 3036943 5/29/2014 May-16 60-7070-106 3037880 5/29/2014 May-16 60-7070-106 3037977 5/29/2014 May-16 60-7070-106 3039663 5/29/2014 May-16 60-7070-106 3040571 5/29/2014 May-16 60-7070-106 3040846 5/29/2014 May-16 60-7070-106 3041662 5/29/2014 May-16 60-7070-107 3035608 5/29/2014 May-16 60-7070-107 3039310 5/29/2014 May-16 60-7070-107 3040210 5/29/2014 May-16 60-7075-003 3034885 5/29/2014 May-16 60-7075-003 3036632 5/29/2014 May-16 60-7075-004 3037384 5/29/2014 May-16 60-7075-005 3040155 5/29/2014 May-16 60-7075-006 3036181 5/29/2014 May-16 60-7080-102 3040186 5/29/2014 May-16 60-7075-102 3040138 5/29/2014 May-16 60-7075-103 3032014 5/29/2014 May-16 60-7075-103 3033529 5/29/2014 May-16 60-7075-103 3033708 5/29/2014 May-16 60-7075-103 3037402 5/29/2014 May-16 60-7075-103 3041909 5/29/2014 May-16 60-7075-104 3033796 5/29/2014 May-16 60-7075-104 3034871 5/29/2014 May-16 60-7075-104 3039231 5/29/2014 May-16 60-7075-105 3032046 5/29/2014 May-16 60-7075-106 3033966 5/29/2014 May-16 60-7075-106 3036609 5/29/2014 May-16 60-7075-106 3040223 5/29/2014 May-16 60-7075-106 3041796 5/29/2014 May-16 5921-024-235 3043588 6/3/2014 Jun-16 5921-024-235 3045890 6/3/2014 Jun-16 5921-030-135 3043934 6/3/2014 Jun-16 5921-030-135 3045900 6/3/2014 Jun-16 5921-030-135 3046794 6/3/2014 Jun-16 5921-030-136 3043041 6/3/2014 Jun-16 5921-030-235 3042974 6/3/2014 Jun-16 5921-030-235 3043638 6/3/2014 Jun-16 5921-030-235 3044175 6/3/2014 Jun-16 5921-030-235 3045370 6/3/2014 Jun-16 5921-030-235 3045729 6/3/2014 Jun-16 5921-034-135 3042990 6/3/2014 Jun-16 5921-034-135 3043463 6/3/2014 Jun-16 5921-034-135 3044201 6/3/2014 Jun-16 5921-034-135 3046079 6/3/2014 Jun-16 5921-034-235 3042339 6/3/2014 Jun-16 5921-034-235 3042942 6/3/2014 Jun-16 5921-034-235 3043831 6/3/2014 Jun-16 5921-034-235 3044802 6/3/2014 Jun-16 5921-034-235 3046482 6/3/2014 Jun-16 5921-034-236 3042114 6/3/2014 Jun-16 5921-218-235 2935491 6/3/2014 Jun-16 5921-218-235 3043104 6/3/2014 Jun-16 5921-218-235 3044943 6/3/2014 Jun-16 5921-218-235 3046469 6/3/2014 Jun-16 5921-218-135 3042146 6/3/2014 Jun-16 5921-218-135 3043883 6/3/2014 Jun-16 5921-218-135 3045835 6/3/2014 Jun-16 5921-218-136 3039628 6/3/2014 Jun-16 60-7070-103 3042568 6/3/2014 Jun-16 60-7070-103 3043574 6/3/2014 Jun-16 60-7070-103 3044240 6/3/2014 Jun-16 60-7070-103 3045410 6/3/2014 Jun-16 60-7070-103 3046521 6/3/2014 Jun-16 60-7070-105 3042179 6/3/2014 Jun-16 60-7070-105 3043519 6/3/2014 Jun-16 60-7070-105 3044255 6/3/2014 Jun-16 60-7070-105 3045218 6/3/2014 Jun-16 60-7070-105 3046261 6/3/2014 Jun-16 60-7070-105 3046509 6/3/2014 Jun-16 60-7070-104 3041488 6/3/2014 Jun-16 60-7070-104 3043629 6/3/2014 Jun-16 60-7070-104 3045633 6/3/2014 Jun-16 60-7070-003 3043136 6/3/2014 Jun-16 60-7070-004 3043002 6/3/2014 Jun-16 60-7070-005 3043266 6/3/2014 Jun-16 60-7070-006 3043073 6/3/2014 Jun-16 60-7070-006 3045916 6/3/2014 Jun-16 60-7070-106 3043416 6/3/2014 Jun-16 60-7070-106 3044065 6/3/2014 Jun-16 60-7070-106 3044530 6/3/2014 Jun-16 60-7070-106 3046067 6/3/2014 Jun-16 60-7070-107 3042961 6/3/2014 Jun-16 60-7075-006 3043767 6/3/2014 Jun-16 60-7075-102 3046668 6/3/2014 Jun-16 60-7075-103 3042133 6/3/2014 Jun-16 60-7075-103 3043162 6/3/2014 Jun-16 60-7075-103 3044272 6/3/2014 Jun-16 60-7075-105 3043820 6/3/2014 Jun-16 60-7075-105 3044054 6/3/2014 Jun-16 60-7075-106 3045764 6/3/2014 Jun-16 5921-018-135 3049884 6/5/2014 Jun-16 5921-018-235 3048533 6/5/2014 Jun-16 5921-024-135 3047998 6/5/2014 Jun-16 5921-024-235 3049907 6/5/2014 Jun-16 5921-030-135 3046972 6/5/2014 Jun-16 5921-030-135 3051965 6/5/2014 Jun-16 5921-030-235 3046862 6/5/2014 Jun-16 5921-030-235 3049515 6/5/2014 Jun-16 5921-030-235 3051418 6/5/2014 Jun-16 5921-034-135 3049589 6/5/2014 Jun-16 5921-034-135 3051460 6/5/2014 Jun-16 5921-034-235 3048258 6/5/2014 Jun-16 5921-034-235 3049267 6/5/2014 Jun-16 5921-034-235 3050694 6/5/2014 Jun-16 5921-034-235 3051070 6/5/2014 Jun-16 5921-218-235 3049279 6/5/2014 Jun-16 5921-218-235 3050976 6/5/2014 Jun-16 5921-030-236 3046870 6/5/2014 Jun-16 5921-044-135 3051748 6/5/2014 Jun-16 5921-044-235 3041834 6/5/2014 Jun-16 5921-218-135 3047491 6/5/2014 Jun-16 5921-218-135 3050504 6/5/2014 Jun-16 5921-218-135 3051920 6/5/2014 Jun-16 5921-218-136 3051949 6/5/2014 Jun-16 60-7070-103 3048113 6/5/2014 Jun-16 60-7070-103 3048859 6/5/2014 Jun-16 60-7070-103 3048871 6/5/2014 Jun-16 60-7070-103 3049527 6/5/2014 Jun-16 60-7070-103 3050251 6/5/2014 Jun-16 60-7070-103 3051487 6/5/2014 Jun-16 60-7070-105 3047518 6/5/2014 Jun-16 60-7070-105 3049205 6/5/2014 Jun-16 60-7070-105 3050222 6/5/2014 Jun-16 60-7070-105 3051425 6/5/2014 Jun-16 60-7070-105 3051851 6/5/2014 Jun-16 60-7070-104 3046843 6/5/2014 Jun-16 60-7070-104 3049660 6/5/2014 Jun-16 60-7070-104 3051908 6/5/2014 Jun-16 60-7070-003 3051004 6/5/2014 Jun-16 60-7070-006 3050589 6/5/2014 Jun-16 60-7070-007 3048514 6/5/2014 Jun-16 60-7070-106 3046606 6/5/2014 Jun-16 60-7070-106 3047103 6/5/2014 Jun-16 60-7070-106 3048790 6/5/2014 Jun-16 60-7070-106 3049747 6/5/2014 Jun-16 60-7070-106 3049969 6/5/2014 Jun-16 60-7070-106 3051186 6/5/2014 Jun-16 60-7075-003 3048372 6/5/2014 Jun-16 60-7075-006 3048449 6/5/2014 Jun-16 60-7075-103 3047671 6/5/2014 Jun-16 60-7075-103 3050781 6/5/2014 Jun-16 60-7075-104 3031562 6/5/2014 Jun-16 60-7075-104 3051912 6/5/2014 Jun-16 60-7075-105 3047142 6/5/2014 Jun-16 60-7075-105 3048317 6/5/2014 Jun-16 60-7075-105 3051657 6/5/2014 Jun-16 60-7075-106 3047541 6/5/2014 Jun-16 60-7075-106 3049710 6/5/2014 Jun-16 5921-018-135 3053778 6/9/2014 Jun-16 5921-024-135 3056944 6/9/2014 Jun-16 5921-024-235 3052099 6/9/2014 Jun-16 5921-024-235 3054704 6/9/2014 Jun-16 5921-030-135 3056391 6/9/2014 Jun-16 5921-030-235 3052121 6/9/2014 Jun-16 5921-030-235 3053220 6/9/2014 Jun-16 5921-030-235 3054727 6/9/2014 Jun-16 5921-034-135 3053811 6/9/2014 Jun-16 5921-034-135 3055012 6/9/2014 Jun-16 5921-034-135 3056389 6/9/2014 Jun-16 5921-034-235 3052132 6/9/2014 Jun-16 5921-034-235 3053462 6/9/2014 Jun-16 5921-034-235 3054478 6/9/2014 Jun-16 5921-034-236 3054840 6/9/2014 Jun-16 5921-218-235 3052145 6/9/2014 Jun-16 5921-218-235 3053103 6/9/2014 Jun-16 5921-218-235 3054942 6/9/2014 Jun-16 5921-218-135 3055008 6/9/2014 Jun-16 5921-218-236 3054497 6/9/2014 Jun-16 60-7070-102 3053914 6/9/2014 Jun-16 60-7070-103 3052150 6/9/2014 Jun-16 60-7070-103 3052546 6/9/2014 Jun-16 60-7070-103 3053541 6/9/2014 Jun-16 60-7070-103 3054855 6/9/2014 Jun-16 60-7070-103 3056651 6/9/2014 Jun-16 60-7070-105 3052166 6/9/2014 Jun-16 60-7070-105 3053064 6/9/2014 Jun-16 60-7070-105 3054356 6/9/2014 Jun-16 60-7070-105 3055096 6/9/2014 Jun-16 60-7070-104 3052412 6/9/2014 Jun-16 60-7070-104 3052477 6/9/2014 Jun-16 60-7070-104 3054373 6/9/2014 Jun-16 60-7070-005 3054838 6/9/2014 Jun-16 60-7070-106 3052178 6/9/2014 Jun-16 60-7070-106 3052837 6/9/2014 Jun-16 60-7070-106 3053980 6/9/2014 Jun-16 60-7070-106 3054677 6/9/2014 Jun-16 60-7070-106 3055873 6/9/2014 Jun-16 60-7075-003 3054961 6/9/2014 Jun-16 60-7075-103 3052505 6/9/2014 Jun-16 60-7075-103 3054919 6/9/2014 Jun-16 60-7075-105 3054387 6/9/2014 Jun-16 60-7075-106 3052431 6/9/2014 Jun-16 60-7075-106 3053882 6/9/2014 Jun-16 60-7075-106 3054988 6/9/2014 Jun-16 5921-018-135 3050464 6/11/2014 Jun-16 5921-018-235 3060341 6/11/2014 Jun-16 5921-024-235 3057461 6/11/2014 Jun-16 5921-024-235 3063218 6/11/2014 Jun-16 5921-030-135 3058578 6/11/2014 Jun-16 5921-030-235 3057276 6/11/2014 Jun-16 5921-030-235 3064504 6/11/2014 Jun-16 5921-034-135 3059307 6/11/2014 Jun-16 5921-034-135 3061081 6/11/2014 Jun-16 5921-034-135 3064589 6/11/2014 Jun-16 5921-034-235 3057118 6/11/2014 Jun-16 5921-034-235 3058131 6/11/2014 Jun-16 5921-034-235 3059693 6/11/2014 Jun-16 5921-034-235 3062833 6/11/2014 Jun-16 5921-218-235 3058917 6/11/2014 Jun-16 5921-218-235 3059664 6/11/2014 Jun-16 5921-218-235 3062867 6/11/2014 Jun-16 5921-044-235 3059934 6/11/2014 Jun-16 5921-212-135 3057141 6/11/2014 Jun-16 5921-218-135 3059311 6/11/2014 Jun-16 5921-218-135 3063717 6/11/2014 Jun-16 5921-218-135 3064035 6/11/2014 Jun-16 60-7070-103 3057501 6/11/2014 Jun-16 60-7070-103 3057670 6/11/2014 Jun-16 60-7070-103 3058566 6/11/2014 Jun-16 60-7070-103 3060048 6/11/2014 Jun-16 60-7070-103 3061837 6/11/2014 Jun-16 60-7070-103 3064042 6/11/2014 Jun-16 60-7070-103 3064687 6/11/2014 Jun-16 60-7070-103 3064775 6/11/2014 Jun-16 60-7070-105 3058288 6/11/2014 Jun-16 60-7070-105 3058532 6/11/2014 Jun-16 60-7070-105 3059655 6/11/2014 Jun-16 60-7070-105 3060990 6/11/2014 Jun-16 60-7070-105 3064639 6/11/2014 Jun-16 60-7070-104 3057643 6/11/2014 Jun-16 60-7070-104 3062898 6/11/2014 Jun-16 60-7070-106 3057490 6/11/2014 Jun-16 60-7070-106 3057586 6/11/2014 Jun-16 60-7070-106 3059209 6/11/2014 Jun-16 60-7070-106 3060133 6/11/2014 Jun-16 60-7070-106 3063202 6/11/2014 Jun-16 60-7070-106 3064591 6/11/2014 Jun-16 60-7075-003 3063959 6/11/2014 Jun-16 60-7075-103 3057658 6/11/2014 Jun-16 60-7075-103 3060179 6/11/2014 Jun-16 60-7075-103 3063646 6/11/2014 Jun-16 60-7075-105 3058889 6/11/2014 Jun-16 60-7075-105 3063963 6/11/2014 Jun-16 60-7075-106 3057426 6/11/2014 Jun-16 60-7075-106 3058891 6/11/2014 Jun-16 60-7075-106 3059722 6/11/2014 Jun-16 60-7075-106 3064781 6/11/2014 Jun-16 5921-018-135 3065910 6/13/2014 Jun-16 5921-018-235 3066249 6/13/2014 Jun-16 5921-018-235 3067699 6/13/2014 Jun-16 5921-018-235 3070406 6/13/2014 Jun-16 5921-024-135 3066212 6/13/2014 Jun-16 5921-024-235 3066709 6/13/2014 Jun-16 5921-024-235 3067862 6/13/2014 Jun-16 5921-024-236 3070614 6/13/2014 Jun-16 5921-030-135 3065002 6/13/2014 Jun-16 5921-030-135 3070475 6/13/2014 Jun-16 5921-030-235 3070021 6/13/2014 Jun-16 5921-034-135 3064907 6/13/2014 Jun-16 5921-034-135 3066161 6/13/2014 Jun-16 5921-034-135 3067827 6/13/2014 Jun-16 5921-034-135 3070370 6/13/2014 Jun-16 5921-034-135 3070985 6/13/2014 Jun-16 5921-034-136 3065722 6/13/2014 Jun-16 5921-034-235 3062487 6/13/2014 Jun-16 5921-034-235 3065400 6/13/2014 Jun-16 5921-034-235 3069374 6/13/2014 Jun-16 5921-034-236 3070755 6/13/2014 Jun-16 5921-218-235 3068529 6/13/2014 Jun-16 5921-218-235 3070888 6/13/2014 Jun-16 5921-044-135 3066265 6/13/2014 Jun-16 5921-212-135 3067804 6/13/2014 Jun-16 5921-218-135 3065561 6/13/2014 Jun-16 5921-218-135 3067668 6/13/2014 Jun-16 5921-218-135 3069065 6/13/2014 Jun-16 60-7070-102 3070148 6/13/2014 Jun-16 60-7070-103 3065143 6/13/2014 Jun-16 60-7070-103 3065376 6/13/2014 Jun-16 60-7070-103 3066671 6/13/2014 Jun-16 60-7070-103 3067550 6/13/2014 Jun-16 60-7070-103 3067584 6/13/2014 Jun-16 60-7070-103 3067836 6/13/2014 Jun-16 60-7070-103 3069123 6/13/2014 Jun-16 60-7070-103 3070729 6/13/2014 Jun-16 60-7070-105 3065087 6/13/2014 Jun-16 60-7070-105 3065865 6/13/2014 Jun-16 60-7070-105 3066830 6/13/2014 Jun-16 60-7070-105 3067476 6/13/2014 Jun-16 60-7070-105 3068950 6/13/2014 Jun-16 60-7070-104 3065975 6/13/2014 Jun-16 60-7070-104 3068398 6/13/2014 Jun-16 60-7070-003 3066895 6/13/2014 Jun-16 60-7070-006 3065714 6/13/2014 Jun-16 60-7070-006 3065780 6/13/2014 Jun-16 60-7070-006 3067371 6/13/2014 Jun-16 60-7070-106 3064897 6/13/2014 Jun-16 60-7070-106 3066637 6/13/2014 Jun-16 60-7070-106 3066732 6/13/2014 Jun-16 60-7070-106 3067606 6/13/2014 Jun-16 60-7070-106 3067929 6/13/2014 Jun-16 60-7070-106 3069860 6/13/2014 Jun-16 60-7070-106 3070890 6/13/2014 Jun-16 60-7075-003 3070963 6/13/2014 Jun-16 60-7075-006 3067363 6/13/2014 Jun-16 60-7075-006 3070583 6/13/2014 Jun-16 60-7080-102 3066905 6/13/2014 Jun-16 60-7075-103 3066616 6/13/2014 Jun-16 60-7075-103 3067741 6/13/2014 Jun-16 60-7075-103 3070959 6/13/2014 Jun-16 60-7075-105 3067469 6/13/2014 Jun-16 60-7075-106 3066824 6/13/2014 Jun-16 60-7075-106 3067794 6/13/2014 Jun-16 60-7075-106 3070066 6/13/2014 Jun-16 5921-024-135 3070410 6/16/2014 Jun-16 5921-024-235 3075460 6/16/2014 Jun-16 5921-030-135 3070740 6/16/2014 Jun-16 5921-030-135 3076267 6/16/2014 Jun-16 5921-030-235 3075350 6/16/2014 Jun-16 5921-034-135 3073317 6/16/2014 Jun-16 5921-034-135 3075769 6/16/2014 Jun-16 5921-034-235 3072399 6/16/2014 Jun-16 5921-034-235 3072489 6/16/2014 Jun-16 5921-034-235 3075101 6/16/2014 Jun-16 5921-034-235 3077237 6/16/2014 Jun-16 5921-218-235 3071096 6/16/2014 Jun-16 5921-218-235 3073166 6/16/2014 Jun-16 5921-218-235 3077263 6/16/2014 Jun-16 5921-218-235 3078767 6/16/2014 Jun-16 5921-212-135 3076813 6/16/2014 Jun-16 5921-218-135 3073956 6/16/2014 Jun-16 5921-218-135 3075908 6/16/2014 Jun-16 60-7070-103 3071168 6/16/2014 Jun-16 60-7070-103 3072462 6/16/2014 Jun-16 60-7070-103 3074636 6/16/2014 Jun-16 60-7070-103 3074881 6/16/2014 Jun-16 60-7070-103 3075494 6/16/2014 Jun-16 60-7070-103 3077179 6/16/2014 Jun-16 60-7070-105 3071110 6/16/2014 Jun-16 60-7070-105 3073624 6/16/2014 Jun-16 60-7070-105 3075418 6/16/2014 Jun-16 60-7070-105 3077271 6/16/2014 Jun-16 60-7070-104 3072135 6/16/2014 Jun-16 60-7070-104 3075879 6/16/2014 Jun-16 60-7070-003 3075172 6/16/2014 Jun-16 60-7070-006 3074092 6/16/2014 Jun-16 60-7070-006 3075027 6/16/2014 Jun-16 60-7070-106 3071229 6/16/2014 Jun-16 60-7070-106 3072040 6/16/2014 Jun-16 60-7070-106 3074059 6/16/2014 Jun-16 60-7070-106 3075563 6/16/2014 Jun-16 60-7070-106 3075626 6/16/2014 Jun-16 60-7070-106 3075795 6/16/2014 Jun-16 60-7070-106 3078087 6/16/2014 Jun-16 60-7070-107 3075062 6/16/2014 Jun-16 60-7075-006 3075299 6/16/2014 Jun-16 60-7075-006 3075402 6/16/2014 Jun-16 60-7075-103 3074966 6/16/2014 Jun-16 60-7075-103 3077817 6/16/2014 Jun-16 60-7075-105 3073227 6/16/2014 Jun-16 60-7075-105 3077829 6/16/2014 Jun-16 60-7075-106 3072254 6/16/2014 Jun-16 60-7075-106 3076280 6/16/2014 Jun-16 5921-018-235 3081082 6/19/2014 Jun-16 5921-024-135 3079098 6/19/2014 Jun-16 5921-024-235 3080650 6/19/2014 Jun-16 5921-024-235 3084833 6/19/2014 Jun-16 5921-030-135 3078883 6/19/2014 Jun-16 5921-030-135 3084155 6/19/2014 Jun-16 5921-030-235 3077801 6/19/2014 Jun-16 5921-030-235 3080106 6/19/2014 Jun-16 5921-030-235 3081727 6/19/2014 Jun-16 5921-034-135 3079149 6/19/2014 Jun-16 5921-034-136 3080904 6/19/2014 Jun-16 5921-034-235 3079288 6/19/2014 Jun-16 5921-034-235 3079578 6/19/2014 Jun-16 5921-034-235 3081356 6/19/2014 Jun-16 5921-034-235 3081903 6/19/2014 Jun-16 5921-034-235 3083448 6/19/2014 Jun-16 5921-218-235 3078780 6/19/2014 Jun-16 5921-218-235 3079702 6/19/2014 Jun-16 5921-218-235 3081770 6/19/2014 Jun-16 5921-218-235 3081957 6/19/2014 Jun-16 5921-044-235 3080335 6/19/2014 Jun-16 5921-218-135 3078981 6/19/2014 Jun-16 5921-218-135 3081961 6/19/2014 Jun-16 5921-218-236 3081791 6/19/2014 Jun-16 60-7070-102 3082346 6/19/2014 Jun-16 60-7070-102 3082604 6/19/2014 Jun-16 60-7070-103 3078779 6/19/2014 Jun-16 60-7070-103 3079108 6/19/2014 Jun-16 60-7070-103 3081271 6/19/2014 Jun-16 60-7070-103 3081926 6/19/2014 Jun-16 60-7070-103 3083901 6/19/2014 Jun-16 60-7070-103 3084795 6/19/2014 Jun-16 60-7070-105 3078820 6/19/2014 Jun-16 60-7070-105 3078999 6/19/2014 Jun-16 60-7070-105 3079787 6/19/2014 Jun-16 60-7070-105 3081696 6/19/2014 Jun-16 60-7070-105 3083067 6/19/2014 Jun-16 60-7070-105 3083300 6/19/2014 Jun-16 60-7070-104 3079584 6/19/2014 Jun-16 60-7070-104 3084814 6/19/2014 Jun-16 60-7070-003 3079242 6/19/2014 Jun-16 60-7070-005 3082071 6/19/2014 Jun-16 60-7070-106 3079154 6/19/2014 Jun-16 60-7070-106 3081325 6/19/2014 Jun-16 60-7070-106 3082351 6/19/2014 Jun-16 60-7070-106 3082834 6/19/2014 Jun-16 60-7070-107 3080293 6/19/2014 Jun-16 60-7075-003 3081318 6/19/2014 Jun-16 60-7075-003 3083482 6/19/2014 Jun-16 60-7075-004 3080697 6/19/2014 Jun-16 60-7075-006 3080621 6/19/2014 Jun-16 60-7075-006 3084748 6/19/2014 Jun-16 60-7075-103 3081053 6/19/2014 Jun-16 60-7075-103 3084252 6/19/2014 Jun-16 60-7075-104 3082207 6/19/2014 Jun-16 60-7075-106 3080894 6/19/2014 Jun-16 60-7075-106 3083606 6/19/2014 Jun-16 60-7075-107 3082295 6/19/2014 Jun-16 5921-018-135 3089520 6/20/2014 Jun-16 5921-018-235 3086542 6/20/2014 Jun-16 5921-024-235 3089294 6/20/2014 Jun-16 5921-030-135 3087150 6/20/2014 Jun-16 5921-030-235 3085602 6/20/2014 Jun-16 5921-030-235 3088692 6/20/2014 Jun-16 5921-034-135 3085003 6/20/2014 Jun-16 5921-034-135 3090059 6/20/2014 Jun-16 5921-034-235 3085185 6/20/2014 Jun-16 5921-034-235 3086437 6/20/2014 Jun-16 5921-034-235 3088811 6/20/2014 Jun-16 5921-212-235 3087742 6/20/2014 Jun-16 5921-218-235 3086094 6/20/2014 Jun-16 5921-218-235 3087902 6/20/2014 Jun-16 5921-218-235 3089202 6/20/2014 Jun-16 5921-018-236 3087726 6/20/2014 Jun-16 5921-030-236 3087735 6/20/2014 Jun-16 5921-030-236 3087994 6/20/2014 Jun-16 5921-218-135 3085308 6/20/2014 Jun-16 60-7070-102 3088086 6/20/2014 Jun-16 60-7070-103 3085504 6/20/2014 Jun-16 60-7070-103 3087291 6/20/2014 Jun-16 60-7070-103 3087653 6/20/2014 Jun-16 60-7070-103 3088208 6/20/2014 Jun-16 60-7070-103 3089195 6/20/2014 Jun-16 60-7070-103 3089985 6/20/2014 Jun-16 60-7070-105 3084822 6/20/2014 Jun-16 60-7070-105 3086245 6/20/2014 Jun-16 60-7070-105 3088628 6/20/2014 Jun-16 60-7070-104 3084931 6/20/2014 Jun-16 60-7070-104 3086557 6/20/2014 Jun-16 60-7070-104 3088249 6/20/2014 Jun-16 60-7070-003 3090044 6/20/2014 Jun-16 60-7070-005 3088093 6/20/2014 Jun-16 60-7070-006 3089508 6/20/2014 Jun-16 60-7070-106 3085331 6/20/2014 Jun-16 60-7070-106 3085455 6/20/2014 Jun-16 60-7070-106 3086968 6/20/2014 Jun-16 60-7070-106 3087607 6/20/2014 Jun-16 60-7070-106 3088126 6/20/2014 Jun-16 60-7070-107 3086416 6/20/2014 Jun-16 60-7070-107 3089304 6/20/2014 Jun-16 60-7075-003 3088231 6/20/2014 Jun-16 60-7075-005 3086923 6/20/2014 Jun-16 60-7075-006 3089667 6/20/2014 Jun-16 60-7080-102 3087956 6/20/2014 Jun-16 60-7075-102 3086947 6/20/2014 Jun-16 60-7075-103 3088055 6/20/2014 Jun-16 60-7075-105 3089370 6/20/2014 Jun-16 60-7075-106 3085710 6/20/2014 Jun-16 60-7075-106 3088824 6/20/2014 Jun-16 5921-018-135 3094701 6/23/2014 Jun-16 5921-018-235 3092818 6/23/2014 Jun-16 5921-024-135 3092077 6/23/2014 Jun-16 5921-024-235 3090092 6/23/2014 Jun-16 5921-024-235 3093747 6/23/2014 Jun-16 5921-030-135 3092374 6/23/2014 Jun-16 5921-030-135 3094315 6/23/2014 Jun-16 5921-030-235 3091381 6/23/2014 Jun-16 5921-030-235 3092576 6/23/2014 Jun-16 5921-030-235 3093354 6/23/2014 Jun-16 5921-034-135 3090139 6/23/2014 Jun-16 5921-034-135 3091750 6/23/2014 Jun-16 5921-034-135 3092361 6/23/2014 Jun-16 5921-034-135 3094218 6/23/2014 Jun-16 5921-034-136 3074102 6/23/2014 Jun-16 5921-034-235 3090796 6/23/2014 Jun-16 5921-034-235 3092609 6/23/2014 Jun-16 5921-034-235 3092684 6/23/2014 Jun-16 5921-034-235 3093222 6/23/2014 Jun-16 5921-218-235 3090457 6/23/2014 Jun-16 5921-218-235 3092825 6/23/2014 Jun-16 5921-044-135 3094738 6/23/2014 Jun-16 5921-044-235 3093832 6/23/2014 Jun-16 5921-218-135 3090403 6/23/2014 Jun-16 5921-218-135 3092524 6/23/2014 Jun-16 5921-218-135 3092582 6/23/2014 Jun-16 60-7070-103 3090501 6/23/2014 Jun-16 60-7070-103 3091732 6/23/2014 Jun-16 60-7070-103 3092326 6/23/2014 Jun-16 60-7070-103 3092915 6/23/2014 Jun-16 60-7070-103 3094294 6/23/2014 Jun-16 60-7070-103 3095149 6/23/2014 Jun-16 60-7070-105 3090000 6/23/2014 Jun-16 60-7070-105 3090997 6/23/2014 Jun-16 60-7070-104 3090351 6/23/2014 Jun-16 60-7070-104 3092388 6/23/2014 Jun-16 60-7070-104 3094565 6/23/2014 Jun-16 60-7070-006 3093527 6/23/2014 Jun-16 60-7070-106 3089287 6/23/2014 Jun-16 60-7070-106 3091922 6/23/2014 Jun-16 60-7070-106 3092737 6/23/2014 Jun-16 60-7070-106 3093845 6/23/2014 Jun-16 60-7070-107 3093312 6/23/2014 Jun-16 60-7075-003 3094577 6/23/2014 Jun-16 60-7075-004 3089315 6/23/2014 Jun-16 60-7075-006 3094772 6/23/2014 Jun-16 60-7075-103 3091895 6/23/2014 Jun-16 60-7075-103 3094327 6/23/2014 Jun-16 60-7075-104 3091427 6/23/2014 Jun-16 60-7075-104 3094900 6/23/2014 Jun-16 60-7075-105 3091001 6/23/2014 Jun-16 60-7075-105 3094928 6/23/2014 Jun-16 60-7075-106 3092168 6/23/2014 Jun-16 60-7075-106 3094107 6/23/2014 Jun-16 5921-024-135 3098534 6/25/2014 Jun-16 5921-024-235 3097369 6/25/2014 Jun-16 5921-030-135 3099626 6/25/2014 Jun-16 5921-030-235 3096496 6/25/2014 Jun-16 5921-030-235 3098244 6/25/2014 Jun-16 5921-034-135 3082028 6/25/2014 Jun-16 5921-034-135 3095986 6/25/2014 Jun-16 5921-034-135 3098302 6/25/2014 Jun-16 5921-034-235 3095509 6/25/2014 Jun-16 5921-034-235 3097189 6/25/2014 Jun-16 5921-034-235 3099544 6/25/2014 Jun-16 5921-218-235 3095226 6/25/2014 Jun-16 5921-218-235 3095290 6/25/2014 Jun-16 5921-218-235 3096522 6/25/2014 Jun-16 5921-218-135 3096567 6/25/2014 Jun-16 60-7070-103 3095448 6/25/2014 Jun-16 60-7070-103 3096047 6/25/2014 Jun-16 60-7070-103 3098122 6/25/2014 Jun-16 60-7070-103 3098285 6/25/2014 Jun-16 60-7070-103 3098547 6/25/2014 Jun-16 60-7070-103 3099619 6/25/2014 Jun-16 60-7070-105 3094890 6/25/2014 Jun-16 60-7070-105 3098631 6/25/2014 Jun-16 60-7070-105 3099563 6/25/2014 Jun-16 60-7070-104 3095955 6/25/2014 Jun-16 60-7070-104 3098263 6/25/2014 Jun-16 60-7070-106 3095274 6/25/2014 Jun-16 60-7070-106 3096892 6/25/2014 Jun-16 60-7070-106 3098575 6/25/2014 Jun-16 60-7070-107 3092678 6/25/2014 Jun-16 60-7075-103 3096514 6/25/2014 Jun-16 60-7075-103 3098568 6/25/2014 Jun-16 60-7075-105 3096742 6/25/2014 Jun-16 60-7075-105 3096973 6/25/2014 Jun-16 60-7075-106 3097376 6/25/2014 Jun-16 60-7075-106 3099822 6/25/2014 Jun-16 5921-018-135 3100484 6/27/2014 Jun-16 5921-018-235 3102571 6/27/2014 Jun-16 5921-024-235 3100180 6/27/2014 Jun-16 5921-030-135 3102603 6/27/2014 Jun-16 5921-030-235 3099931 6/27/2014 Jun-16 5921-030-235 3101973 6/27/2014 Jun-16 5921-030-235 3102027 6/27/2014 Jun-16 5921-030-235 3103826 6/27/2014 Jun-16 5921-030-235 3104907 6/27/2014 Jun-16 5921-034-135 3099851 6/27/2014 Jun-16 5921-034-135 3102062 6/27/2014 Jun-16 5921-034-235 3100758 6/27/2014 Jun-16 5921-034-235 3102418 6/27/2014 Jun-16 5921-034-235 3103888 6/27/2014 Jun-16 5921-218-235 3102898 6/27/2014 Jun-16 5921-218-235 3103531 6/27/2014 Jun-16 5921-218-235 3104438 6/27/2014 Jun-16 5921-218-135 3100360 6/27/2014 Jun-16 60-7070-103 3100228 6/27/2014 Jun-16 60-7070-103 3100762 6/27/2014 Jun-16 60-7070-103 3100855 6/27/2014 Jun-16 60-7070-103 3101329 6/27/2014 Jun-16 60-7070-103 3102619 6/27/2014 Jun-16 60-7070-103 3103315 6/27/2014 Jun-16 60-7070-103 3103549 6/27/2014 Jun-16 60-7070-103 3104383 6/27/2014 Jun-16 60-7070-105 3101068 6/27/2014 Jun-16 60-7070-105 3102015 6/27/2014 Jun-16 60-7070-105 3103565 6/27/2014 Jun-16 60-7070-105 3105018 6/27/2014 Jun-16 60-7070-104 3100151 6/27/2014 Jun-16 60-7070-104 3101197 6/27/2014 Jun-16 60-7070-104 3103078 6/27/2014 Jun-16 60-7070-003 3102769 6/27/2014 Jun-16 60-7070-005 3101258 6/27/2014 Jun-16 60-7070-006 3100840 6/27/2014 Jun-16 60-7070-106 3100926 6/27/2014 Jun-16 60-7070-106 3101099 6/27/2014 Jun-16 60-7070-106 3101270 6/27/2014 Jun-16 60-7070-106 3103021 6/27/2014 Jun-16 60-7070-106 3103218 6/27/2014 Jun-16 60-7070-106 3104100 6/27/2014 Jun-16 60-7070-107 3100069 6/27/2014 Jun-16 60-7075-006 3100394 6/27/2014 Jun-16 60-7075-103 3100413 6/27/2014 Jun-16 60-7075-103 3102965 6/27/2014 Jun-16 60-7075-103 3103959 6/27/2014 Jun-16 60-7075-104 3102091 6/27/2014 Jun-16 60-7075-105 3101178 6/27/2014 Jun-16 60-7075-106 3099996 6/27/2014 Jun-16 60-7075-106 3101236 6/27/2014 Jun-16 60-7075-106 3102977 6/27/2014 Jun-16 60-7075-107 3101204 6/27/2014 Jun-16 5921-018-235 3109697 6/28/2014 Jun-16 5921-024-135 3108461 6/28/2014 Jun-16 5921-030-135 3107274 6/28/2014 Jun-16 5921-030-135 3110057 6/28/2014 Jun-16 5921-030-235 3106882 6/28/2014 Jun-16 5921-030-235 3108950 6/28/2014 Jun-16 5921-030-235 3109110 6/28/2014 Jun-16 5921-034-135 3106967 6/28/2014 Jun-16 5921-034-135 3107290 6/28/2014 Jun-16 5921-034-135 3108712 6/28/2014 Jun-16 5921-034-235 3105189 6/28/2014 Jun-16 5921-034-235 3106692 6/28/2014 Jun-16 5921-034-235 3109511 6/28/2014 Jun-16 5921-218-235 3106458 6/28/2014 Jun-16 5921-218-235 3108868 6/28/2014 Jun-16 5921-218-135 3106556 6/28/2014 Jun-16 5921-218-135 3107469 6/28/2014 Jun-16 60-7070-103 3106174 6/28/2014 Jun-16 60-7070-103 3106811 6/28/2014 Jun-16 60-7070-103 3107891 6/28/2014 Jun-16 60-7070-103 3108604 6/28/2014 Jun-16 60-7070-103 3109335 6/28/2014 Jun-16 60-7070-105 3105348 6/28/2014 Jun-16 60-7070-105 3106375 6/28/2014 Jun-16 60-7070-105 3106573 6/28/2014 Jun-16 60-7070-105 3107773 6/28/2014 Jun-16 60-7070-105 3109414 6/28/2014 Jun-16 60-7070-104 3105428 6/28/2014 Jun-16 60-7070-104 3106093 6/28/2014 Jun-16 60-7070-104 3108295 6/28/2014 Jun-16 60-7070-106 3106188 6/28/2014 Jun-16 60-7070-106 3107675 6/28/2014 Jun-16 60-7070-106 3109065 6/28/2014 Jun-16 60-7070-107 3108555 6/28/2014 Jun-16 60-7075-003 3106951 6/28/2014 Jun-16 60-7075-006 3108708 6/28/2014 Jun-16 60-7075-103 3105437 6/28/2014 Jun-16 60-7075-106 3105162 6/28/2014 Jun-16 60-7075-106 3107476 6/28/2014 Jun-16 60-7075-106 3107668 6/28/2014 Jun-16 60-7075-106 3108823 6/28/2014 Jun-16 5921-018-135 3112648 7/2/2014 Jul-16 5921-024-135 3113347 7/2/2014 Jul-16 5921-030-135 3110251 7/2/2014 Jul-16 5921-030-135 3111867 7/2/2014 Jul-16 5921-030-235 3111500 7/2/2014 Jul-16 5921-030-235 3113368 7/2/2014 Jul-16 5921-034-135 3113306 7/2/2014 Jul-16 5921-034-136 3110185 7/2/2014 Jul-16 5921-034-235 3109856 7/2/2014 Jul-16 5921-034-235 3113539 7/2/2014 Jul-16 5921-034-235 3113663 7/2/2014 Jul-16 5921-034-236 3110976 7/2/2014 Jul-16 5921-218-235 3111268 7/2/2014 Jul-16 5921-218-235 3111977 7/2/2014 Jul-16 5921-218-235 3113310 7/2/2014 Jul-16 5921-044-135 3110982 7/2/2014 Jul-16 5921-218-135 3111117 7/2/2014 Jul-16 5921-218-135 3113135 7/2/2014 Jul-16 60-7070-103 3110639 7/2/2014 Jul-16 60-7070-103 3111252 7/2/2014 Jul-16 60-7070-103 3112081 7/2/2014 Jul-16 60-7070-103 3112301 7/2/2014 Jul-16 60-7070-103 3113093 7/2/2014 Jul-16 60-7070-105 3111724 7/2/2014 Jul-16 60-7070-105 3113541 7/2/2014 Jul-16 60-7070-105 3113922 7/2/2014 Jul-16 60-7070-104 3110035 7/2/2014 Jul-16 60-7070-104 3111983 7/2/2014 Jul-16 60-7070-003 3112477 7/2/2014 Jul-16 60-7070-003 3112611 7/2/2014 Jul-16 60-7070-006 3110377 7/2/2014 Jul-16 60-7070-106 3109175 7/2/2014 Jul-16 60-7070-106 3111619 7/2/2014 Jul-16 60-7070-106 3112960 7/2/2014 Jul-16 60-7070-106 3113188 7/2/2014 Jul-16 60-7070-106 3114167 7/2/2014 Jul-16 60-7070-107 3111707 7/2/2014 Jul-16 60-7075-003 3112178 7/2/2014 Jul-16 60-7075-004 3111015 7/2/2014 Jul-16 60-7075-006 3112454 7/2/2014 Jul-16 60-7075-103 3111027 7/2/2014 Jul-16 60-7075-104 3111062 7/2/2014 Jul-16 60-7075-105 3109716 7/2/2014 Jul-16 60-7075-105 3112150 7/2/2014 Jul-16 60-7075-106 3111070 7/2/2014 Jul-16 5921-018-235 3117899 7/7/2014 Jul-16 5921-024-235 3115632 7/7/2014 Jul-16 5921-024-235 3117071 7/7/2014 Jul-16 5921-030-135 3117670 7/7/2014 Jul-16 5921-030-135 3117810 7/7/2014 Jul-16 5921-030-235 3115374 7/7/2014 Jul-16 5921-030-235 3116406 7/7/2014 Jul-16 5921-030-235 3118371 7/7/2014 Jul-16 5921-030-235 3118532 7/7/2014 Jul-16 5921-034-135 3111516 7/7/2014 Jul-16 5921-034-135 3114122 7/7/2014 Jul-16 5921-034-235 3106587 7/7/2014 Jul-16 5921-034-235 3116084 7/7/2014 Jul-16 5921-034-235 3116239 7/7/2014 Jul-16 5921-034-235 3117627 7/7/2014 Jul-16 5921-034-235 3118756 7/7/2014 Jul-16 5921-218-235 3115645 7/7/2014 Jul-16 5921-218-235 3118131 7/7/2014 Jul-16 5921-218-235 3118469 7/7/2014 Jul-16 5921-218-136 3114612 7/7/2014 Jul-16 60-7070-103 3114413 7/7/2014 Jul-16 60-7070-103 3115390 7/7/2014 Jul-16 60-7070-103 3115511 7/7/2014 Jul-16 60-7070-103 3116304 7/7/2014 Jul-16 60-7070-103 3117593 7/7/2014 Jul-16 60-7070-103 3117881 7/7/2014 Jul-16 60-7070-103 3118634 7/7/2014 Jul-16 60-7070-105 3118154 7/7/2014 Jul-16 60-7070-104 3115096 7/7/2014 Jul-16 60-7070-104 3115342 7/7/2014 Jul-16 60-7070-104 3117085 7/7/2014 Jul-16 60-7070-006 3116959 7/7/2014 Jul-16 60-7070-006 3117063 7/7/2014 Jul-16 60-7070-106 3115818 7/7/2014 Jul-16 60-7070-106 3117322 7/7/2014 Jul-16 60-7075-003 3116423 7/7/2014 Jul-16 60-7075-006 3115020 7/7/2014 Jul-16 60-7075-103 3114743 7/7/2014 Jul-16 60-7075-103 3114872 7/7/2014 Jul-16 60-7075-103 3115524 7/7/2014 Jul-16 60-7075-103 3117921 7/7/2014 Jul-16 60-7075-105 3116508 7/7/2014 Jul-16 60-7075-106 3114838 7/7/2014 Jul-16 60-7075-106 3116717 7/7/2014 Jul-16 60-7075-106 3116819 7/7/2014 Jul-16 5921-024-135 3119535 7/9/2014 Jul-16 5921-024-235 3119519 7/9/2014 Jul-16 5921-030-135 3122028 7/9/2014 Jul-16 5921-030-135 3122305 7/9/2014 Jul-16 5921-030-235 3120303 7/9/2014 Jul-16 5921-030-235 3121620 7/9/2014 Jul-16 5921-034-135 3119041 7/9/2014 Jul-16 5921-034-135 3121631 7/9/2014 Jul-16 5921-034-135 3122282 7/9/2014 Jul-16 5921-034-235 3119686 7/9/2014 Jul-16 5921-034-235 3120989 7/9/2014 Jul-16 5921-034-235 3121146 7/9/2014 Jul-16 5921-218-235 3119934 7/9/2014 Jul-16 5921-218-235 3121413 7/9/2014 Jul-16 5921-218-135 3120110 7/9/2014 Jul-16 5921-218-135 3122230 7/9/2014 Jul-16 60-7070-103 3120008 7/9/2014 Jul-16 60-7070-103 3121011 7/9/2014 Jul-16 60-7070-103 3122689 7/9/2014 Jul-16 60-7070-103 3122978 7/9/2014 Jul-16 60-7070-105 3120326 7/9/2014 Jul-16 60-7070-105 3122442 7/9/2014 Jul-16 60-7070-105 3123046 7/9/2014 Jul-16 60-7070-104 3120012 7/9/2014 Jul-16 60-7070-104 3122564 7/9/2014 Jul-16 60-7070-003 3122380 7/9/2014 Jul-16 60-7070-106 3118940 7/9/2014 Jul-16 60-7070-106 3119999 7/9/2014 Jul-16 60-7070-106 3120201 7/9/2014 Jul-16 60-7070-106 3121497 7/9/2014 Jul-16 60-7070-106 3122435 7/9/2014 Jul-16 60-7070-106 3122474 7/9/2014 Jul-16 60-7070-106 3122834 7/9/2014 Jul-16 60-7070-107 3122351 7/9/2014 Jul-16 60-7075-104 3120802 7/9/2014 Jul-16 60-7075-105 3119348 7/9/2014 Jul-16 60-7075-106 3119232 7/9/2014 Jul-16 5921-024-235 3123274 7/11/2014 Jul-16 5921-024-235 3127431 7/11/2014 Jul-16 5921-030-135 3124707 7/11/2014 Jul-16 5921-030-135 3125723 7/11/2014 Jul-16 5921-030-136 3123080 7/11/2014 Jul-16 5921-030-235 3122868 7/11/2014 Jul-16 5921-030-235 3123773 7/11/2014 Jul-16 5921-030-235 3124822 7/11/2014 Jul-16 5921-030-235 3124977 7/11/2014 Jul-16 5921-030-235 3126629 7/11/2014 Jul-16 5921-030-235 3128220 7/11/2014 Jul-16 5921-030-235 3128352 7/11/2014 Jul-16 5921-034-135 3121654 7/11/2014 Jul-16 5921-034-135 3125003 7/11/2014 Jul-16 5921-034-135 3126104 7/11/2014 Jul-16 5921-034-235 3123668 7/11/2014 Jul-16 5921-034-235 3125504 7/11/2014 Jul-16 5921-034-235 3127132 7/11/2014 Jul-16 5921-034-236 3124571 7/11/2014 Jul-16 5921-218-235 3123300 7/11/2014 Jul-16 5921-218-235 3126162 7/11/2014 Jul-16 5921-218-235 3127166 7/11/2014 Jul-16 5921-218-135 3123606 7/11/2014 Jul-16 5921-218-135 3126278 7/11/2014 Jul-16 5921-218-236 3124619 7/11/2014 Jul-16 60-7070-103 3122966 7/11/2014 Jul-16 60-7070-103 3124425 7/11/2014 Jul-16 60-7070-103 3125878 7/11/2014 Jul-16 60-7070-103 3126865 7/11/2014 Jul-16 60-7070-103 3128811 7/11/2014 Jul-16 60-7070-105 3123261 7/11/2014 Jul-16 60-7070-105 3124058 7/11/2014 Jul-16 60-7070-105 3124592 7/11/2014 Jul-16 60-7070-105 3126115 7/11/2014 Jul-16 60-7070-105 3126492 7/11/2014 Jul-16 60-7070-105 3128306 7/11/2014 Jul-16 60-7070-104 3124585 7/11/2014 Jul-16 60-7070-104 3127630 7/11/2014 Jul-16 60-7070-104 3127719 7/11/2014 Jul-16 60-7070-004 3123699 7/11/2014 Jul-16 60-7070-004 3124603 7/11/2014 Jul-16 60-7070-005 3115083 7/11/2014 Jul-16 60-7070-006 3123804 7/11/2014 Jul-16 60-7070-106 3117059 7/11/2014 Jul-16 60-7070-106 3123836 7/11/2014 Jul-16 60-7070-106 3124528 7/11/2014 Jul-16 60-7070-106 3125734 7/11/2014 Jul-16 60-7070-106 3126831 7/11/2014 Jul-16 60-7070-106 3127902 7/11/2014 Jul-16 60-7070-107 3123453 7/11/2014 Jul-16 60-7070-107 3125042 7/11/2014 Jul-16 60-7075-004 3123815 7/11/2014 Jul-16 60-7075-103 3126369 7/11/2014 Jul-16 60-7075-103 3127408 7/11/2014 Jul-16 60-7075-103 3127551 7/11/2014 Jul-16 60-7075-104 3123328 7/11/2014 Jul-16 60-7075-106 3123196 7/11/2014 Jul-16 60-7075-106 3126617 7/11/2014 Jul-16 5921-018-235 3129890 7/14/2014 Jul-16 5921-024-235 3129481 7/14/2014 Jul-16 5921-024-235 3132817 7/14/2014 Jul-16 5921-030-135 3130500 7/14/2014 Jul-16 5921-030-135 3133751 7/14/2014 Jul-16 5921-030-235 3130603 7/14/2014 Jul-16 5921-030-235 3130954 7/14/2014 Jul-16 5921-030-235 3132935 7/14/2014 Jul-16 5921-034-135 3128830 7/14/2014 Jul-16 5921-034-135 3129928 7/14/2014 Jul-16 5921-034-135 3131224 7/14/2014 Jul-16 5921-034-135 3133330 7/14/2014 Jul-16 5921-034-235 3128458 7/14/2014 Jul-16 5921-034-235 3131731 7/14/2014 Jul-16 5921-034-235 3132466 7/14/2014 Jul-16 5921-034-235 3133865 7/14/2014 Jul-16 5921-218-135 3129218 7/14/2014 Jul-16 5921-218-135 3131435 7/14/2014 Jul-16 60-7070-103 3129078 7/14/2014 Jul-16 60-7070-103 3129680 7/14/2014 Jul-16 60-7070-103 3130528 7/14/2014 Jul-16 60-7070-103 3131847 7/14/2014 Jul-16 60-7070-103 3133018 7/14/2014 Jul-16 60-7070-103 3133849 7/14/2014 Jul-16 60-7070-105 3129937 7/14/2014 Jul-16 60-7070-105 3131059 7/14/2014 Jul-16 60-7070-105 3132736 7/14/2014 Jul-16 60-7070-105 3133245 7/14/2014 Jul-16 60-7070-104 3131615 7/14/2014 Jul-16 60-7070-104 3133638 7/14/2014 Jul-16 60-7070-006 3128628 7/14/2014 Jul-16 60-7070-006 3132024 7/14/2014 Jul-16 60-7070-106 3129959 7/14/2014 Jul-16 60-7070-106 3130043 7/14/2014 Jul-16 60-7070-106 3130707 7/14/2014 Jul-16 60-7070-106 3131966 7/14/2014 Jul-16 60-7070-106 3132547 7/14/2014 Jul-16 60-7070-106 3133115 7/14/2014 Jul-16 60-7070-107 3129963 7/14/2014 Jul-16 60-7075-103 3131269 7/14/2014 Jul-16 60-7075-103 3132961 7/14/2014 Jul-16 60-7075-104 3132053 7/14/2014 Jul-16 60-7075-105 3129195 7/14/2014 Jul-16 60-7075-105 3131604 7/14/2014 Jul-16 60-7075-106 3129389 7/14/2014 Jul-16 60-7075-106 3129475 7/14/2014 Jul-16 60-7075-106 3130402 7/14/2014 Jul-16 60-7075-106 3130592 7/14/2014 Jul-16 60-7075-106 3133087 7/14/2014 Jul-16 5921-018-135 3138291 7/16/2014 Jul-16 5921-018-235 3137347 7/16/2014 Jul-16 5921-030-135 3137231 7/16/2014 Jul-16 5921-030-235 3134147 7/16/2014 Jul-16 5921-030-235 3135294 7/16/2014 Jul-16 5921-034-135 3134962 7/16/2014 Jul-16 5921-034-135 3135370 7/16/2014 Jul-16 5921-034-135 3135667 7/16/2014 Jul-16 5921-034-135 3136889 7/16/2014 Jul-16 5921-034-235 3135336 7/16/2014 Jul-16 5921-034-235 3136359 7/16/2014 Jul-16 5921-034-235 3138197 7/16/2014 Jul-16 5921-034-235 3138788 7/16/2014 Jul-16 5921-218-235 3136550 7/16/2014 Jul-16 5921-218-235 3137933 7/16/2014 Jul-16 5921-218-135 3134479 7/16/2014 Jul-16 5921-218-135 3135888 7/16/2014 Jul-16 5921-218-135 3137055 7/16/2014 Jul-16 60-7070-103 3134582 7/16/2014 Jul-16 60-7070-103 3135499 7/16/2014 Jul-16 60-7070-103 3136773 7/16/2014 Jul-16 60-7070-103 3137415 7/16/2014 Jul-16 60-7070-103 3137853 7/16/2014 Jul-16 60-7070-103 3138317 7/16/2014 Jul-16 60-7070-103 3138607 7/16/2014 Jul-16 60-7070-105 3134650 7/16/2014 Jul-16 60-7070-105 3134728 7/16/2014 Jul-16 60-7070-105 3135021 7/16/2014 Jul-16 60-7070-105 3135434 7/16/2014 Jul-16 60-7070-105 3136545 7/16/2014 Jul-16 60-7070-105 3137721 7/16/2014 Jul-16 60-7070-104 3134744 7/16/2014 Jul-16 60-7070-104 3135512 7/16/2014 Jul-16 60-7070-104 3137161 7/16/2014 Jul-16 60-7070-003 3138068 7/16/2014 Jul-16 60-7070-106 3134716 7/16/2014 Jul-16 60-7070-106 3136022 7/16/2014 Jul-16 60-7070-106 3136288 7/16/2014 Jul-16 60-7070-106 3136290 7/16/2014 Jul-16 60-7070-106 3137220 7/16/2014 Jul-16 60-7070-106 3138340 7/16/2014 Jul-16 60-7070-107 3134374 7/16/2014 Jul-16 60-7070-107 3135452 7/16/2014 Jul-16 60-7075-003 3136993 7/16/2014 Jul-16 60-7075-003 3137142 7/16/2014 Jul-16 60-7075-103 3137824 7/16/2014 Jul-16 60-7075-104 3138006 7/16/2014 Jul-16 60-7075-105 3135985 7/16/2014 Jul-16 60-7075-106 3134511 7/16/2014 Jul-16 60-7075-106 3137914 7/16/2014 Jul-16 5921-024-235 3141485 7/18/2014 Jul-16 5921-030-135 3139331 7/18/2014 Jul-16 5921-030-235 3138987 7/18/2014 Jul-16 5921-030-235 3141947 7/18/2014 Jul-16 5921-030-235 3142861 7/18/2014 Jul-16 5921-034-135 3142153 7/18/2014 Jul-16 5921-034-235 3140368 7/18/2014 Jul-16 5921-034-235 3141590 7/18/2014 Jul-16 5921-034-235 3142667 7/18/2014 Jul-16 5921-218-235 3139090 7/18/2014 Jul-16 5921-218-235 3139543 7/18/2014 Jul-16 5921-218-235 3140296 7/18/2014 Jul-16 5921-218-235 3142803 7/18/2014 Jul-16 5921-218-135 3140663 7/18/2014 Jul-16 60-7070-103 3139396 7/18/2014 Jul-16 60-7070-103 3140489 7/18/2014 Jul-16 60-7070-103 3142124 7/18/2014 Jul-16 60-7070-103 3143062 7/18/2014 Jul-16 60-7070-105 3139035 7/18/2014 Jul-16 60-7070-105 3139486 7/18/2014 Jul-16 60-7070-105 3140766 7/18/2014 Jul-16 60-7070-105 3141143 7/18/2014 Jul-16 60-7070-105 3142651 7/18/2014 Jul-16 60-7070-104 3140277 7/18/2014 Jul-16 60-7070-104 3141779 7/18/2014 Jul-16 60-7070-006 3141780 7/18/2014 Jul-16 60-7070-006 3141975 7/18/2014 Jul-16 60-7070-106 3139401 7/18/2014 Jul-16 60-7070-106 3140778 7/18/2014 Jul-16 60-7070-106 3141162 7/18/2014 Jul-16 60-7070-106 3142622 7/18/2014 Jul-16 60-7070-106 3143070 7/18/2014 Jul-16 60-7070-106 3143210 7/18/2014 Jul-16 60-7070-107 3142169 7/18/2014 Jul-16 60-7075-103 3138816 7/18/2014 Jul-16 60-7075-103 3141906 7/18/2014 Jul-16 60-7075-105 3141588 7/18/2014 Jul-16 60-7075-105 3141733 7/18/2014 Jul-16 60-7075-106 3139354 7/18/2014 Jul-16 60-7075-106 3142679 7/18/2014 Jul-16 60-7075-106 3142959 7/18/2014 Jul-16 5921-018-235 3143769 7/22/2014 Jul-16 5921-018-235 3147515 7/22/2014 Jul-16 5921-024-235 3135900 7/22/2014 Jul-16 5921-030-135 3143559 7/22/2014 Jul-16 5921-030-135 3147694 7/22/2014 Jul-16 5921-030-235 3144385 7/22/2014 Jul-16 5921-030-235 3146030 7/22/2014 Jul-16 5921-034-135 3145197 7/22/2014 Jul-16 5921-034-135 3147214 7/22/2014 Jul-16 5921-034-235 3144005 7/22/2014 Jul-16 5921-034-235 3145837 7/22/2014 Jul-16 5921-034-235 3147440 7/22/2014 Jul-16 5921-218-235 3145408 7/22/2014 Jul-16 5921-218-235 3147850 7/22/2014 Jul-16 5921-218-135 3147312 7/22/2014 Jul-16 60-7070-103 3144453 7/22/2014 Jul-16 60-7070-103 3144815 7/22/2014 Jul-16 60-7070-103 3145068 7/22/2014 Jul-16 60-7070-103 3146770 7/22/2014 Jul-16 60-7070-103 3147504 7/22/2014 Jul-16 60-7070-105 3143626 7/22/2014 Jul-16 60-7070-105 3145496 7/22/2014 Jul-16 60-7070-105 3146076 7/22/2014 Jul-16 60-7070-105 3146450 7/22/2014 Jul-16 60-7070-105 3147233 7/22/2014 Jul-16 60-7070-104 3143898 7/22/2014 Jul-16 60-7070-104 3145580 7/22/2014 Jul-16 60-7070-104 3147781 7/22/2014 Jul-16 60-7070-005 3145081 7/22/2014 Jul-16 60-7070-006 3147543 7/22/2014 Jul-16 60-7070-106 3145034 7/22/2014 Jul-16 60-7070-106 3145132 7/22/2014 Jul-16 60-7070-106 3146838 7/22/2014 Jul-16 60-7070-106 3147100 7/22/2014 Jul-16 60-7075-003 3144359 7/22/2014 Jul-16 60-7075-006 3144022 7/22/2014 Jul-16 60-7080-102 3146523 7/22/2014 Jul-16 60-7075-103 3145184 7/22/2014 Jul-16 60-7075-104 3144371 7/22/2014 Jul-16 60-7075-106 3147280 7/22/2014 Jul-16 5921-018-135 3149572 7/25/2014 Jul-16 5921-018-235 3152361 7/25/2014 Jul-16 5921-024-235 3148943 7/25/2014 Jul-16 5921-024-235 3152374 7/25/2014 Jul-16 5921-024-235 3152582 7/25/2014 Jul-16 5921-030-135 3149501 7/25/2014 Jul-16 5921-030-235 3148569 7/25/2014 Jul-16 5921-030-235 3148666 7/25/2014 Jul-16 5921-030-235 3150604 7/25/2014 Jul-16 5921-034-135 3149139 7/25/2014 Jul-16 5921-034-135 3150708 7/25/2014 Jul-16 5921-034-135 3152666 7/25/2014 Jul-16 5921-034-235 3148737 7/25/2014 Jul-16 5921-034-235 3150139 7/25/2014 Jul-16 5921-034-235 3152335 7/25/2014 Jul-16 5921-218-235 3148238 7/25/2014 Jul-16 5921-218-235 3149909 7/25/2014 Jul-16 5921-218-235 3151404 7/25/2014 Jul-16 5921-218-235 3152904 7/25/2014 Jul-16 5921-218-135 3149194 7/25/2014 Jul-16 5921-218-135 3152548 7/25/2014 Jul-16 60-7070-103 3149187 7/25/2014 Jul-16 60-7070-103 3149295 7/25/2014 Jul-16 60-7070-103 3149875 7/25/2014 Jul-16 60-7070-103 3151637 7/25/2014 Jul-16 60-7070-103 3151951 7/25/2014 Jul-16 60-7070-103 3152255 7/25/2014 Jul-16 60-7070-105 3148582 7/25/2014 Jul-16 60-7070-105 3149881 7/25/2014 Jul-16 60-7070-105 3150658 7/25/2014 Jul-16 60-7070-105 3151029 7/25/2014 Jul-16 60-7070-105 3152201 7/25/2014 Jul-16 60-7070-104 3149899 7/25/2014 Jul-16 60-7070-003 3148991 7/25/2014 Jul-16 60-7070-106 3148229 7/25/2014 Jul-16 60-7070-106 3149913 7/25/2014 Jul-16 60-7070-106 3151946 7/25/2014 Jul-16 60-7075-003 3149564 7/25/2014 Jul-16 60-7075-006 3152342 7/25/2014 Jul-16 60-7075-103 3148744 7/25/2014 Jul-16 60-7075-103 3150207 7/25/2014 Jul-16 60-7075-103 3150564 7/25/2014 Jul-16 60-7075-103 3152357 7/25/2014 Jul-16 60-7075-104 3135971 7/25/2014 Jul-16 60-7075-105 3148414 7/25/2014 Jul-16 60-7075-105 3153312 7/25/2014 Jul-16 60-7075-106 3149490 7/25/2014 Jul-16 60-7075-106 3151895 7/25/2014 Jul-16 60-7075-106 3153159 7/25/2014 Jul-16 5921-018-235 3098450 7/28/2014 Jul-16 5921-018-135 3156941 7/31/2014 Jul-16 5921-018-135 3157638 7/31/2014 Jul-16 5921-018-135 3162220 7/31/2014 Jul-16 5921-018-235 3157640 7/31/2014 Jul-16 5921-018-235 3158727 7/31/2014 Jul-16 5921-024-136 3154304 7/31/2014 Jul-16 5921-024-235 3160343 7/31/2014 Jul-16 5921-030-135 3155699 7/31/2014 Jul-16 5921-030-135 3157471 7/31/2014 Jul-16 5921-030-135 3157780 7/31/2014 Jul-16 5921-030-135 3158631 7/31/2014 Jul-16 5921-030-135 3161839 7/31/2014 Jul-16 5921-030-235 3153687 7/31/2014 Jul-16 5921-030-235 3154651 7/31/2014 Jul-16 5921-030-235 3157629 7/31/2014 Jul-16 5921-030-235 3158758 7/31/2014 Jul-16 5921-030-235 3160772 7/31/2014 Jul-16 5921-030-235 3162174 7/31/2014 Jul-16 5921-030-235 3163420 7/31/2014 Jul-16 5921-034-135 3154531 7/31/2014 Jul-16 5921-034-135 3156892 7/31/2014 Jul-16 5921-034-135 3159193 7/31/2014 Jul-16 5921-034-135 3160937 7/31/2014 Jul-16 5921-034-135 3163514 7/31/2014 Jul-16 5921-034-136 3154985 7/31/2014 Jul-16 5921-034-235 3155550 7/31/2014 Jul-16 5921-034-235 3157266 7/31/2014 Jul-16 5921-034-235 3157877 7/31/2014 Jul-16 5921-034-235 3159528 7/31/2014 Jul-16 5921-034-235 3160764 7/31/2014 Jul-16 5921-034-235 3161582 7/31/2014 Jul-16 5921-034-235 3163023 7/31/2014 Jul-16 5921-218-235 3154772 7/31/2014 Jul-16 5921-218-235 3156329 7/31/2014 Jul-16 5921-218-235 3159473 7/31/2014 Jul-16 5921-218-235 3161342 7/31/2014 Jul-16 5921-218-235 3163010 7/31/2014 Jul-16 5921-030-236 3154667 7/31/2014 Jul-16 5921-218-135 3153591 7/31/2014 Jul-16 5921-218-135 3156023 7/31/2014 Jul-16 5921-218-135 3157485 7/31/2014 Jul-16 5921-218-135 3159516 7/31/2014 Jul-16 5921-218-135 3163522 7/31/2014 Jul-16 60-7070-103 3154176 7/31/2014 Jul-16 60-7070-103 3154481 7/31/2014 Jul-16 60-7070-103 3154633 7/31/2014 Jul-16 60-7070-103 3154717 7/31/2014 Jul-16 60-7070-103 3156355 7/31/2014 Jul-16 60-7070-103 3157278 7/31/2014 Jul-16 60-7070-103 3158926 7/31/2014 Jul-16 60-7070-103 3159062 7/31/2014 Jul-16 60-7070-103 3159268 7/31/2014 Jul-16 60-7070-103 3159782 7/31/2014 Jul-16 60-7070-103 3160260 7/31/2014 Jul-16 60-7070-103 3160481 7/31/2014 Jul-16 60-7070-103 3160740 7/31/2014 Jul-16 60-7070-103 3161645 7/31/2014 Jul-16 60-7070-103 3161904 7/31/2014 Jul-16 60-7070-103 3162458 7/31/2014 Jul-16 60-7070-103 3163317 7/31/2014 Jul-16 60-7070-105 3153625 7/31/2014 Jul-16 60-7070-105 3154294 7/31/2014 Jul-16 60-7070-105 3154520 7/31/2014 Jul-16 60-7070-105 3155112 7/31/2014 Jul-16 60-7070-105 3156034 7/31/2014 Jul-16 60-7070-105 3156800 7/31/2014 Jul-16 60-7070-105 3158394 7/31/2014 Jul-16 60-7070-105 3158919 7/31/2014 Jul-16 60-7070-105 3159880 7/31/2014 Jul-16 60-7070-105 3161970 7/31/2014 Jul-16 60-7070-105 3162462 7/31/2014 Jul-16 60-7070-105 3163184 7/31/2014 Jul-16 60-7070-105 3163454 7/31/2014 Jul-16 60-7070-104 3153354 7/31/2014 Jul-16 60-7070-104 3154764 7/31/2014 Jul-16 60-7070-104 3157590 7/31/2014 Jul-16 60-7070-104 3159661 7/31/2014 Jul-16 60-7070-104 3160701 7/31/2014 Jul-16 60-7070-104 3162980 7/31/2014 Jul-16 60-7070-104 3163449 7/31/2014 Jul-16 60-7070-005 3156533 7/31/2014 Jul-16 60-7070-005 3156863 7/31/2014 Jul-16 60-7070-006 3156925 7/31/2014 Jul-16 60-7070-006 3162368 7/31/2014 Jul-16 60-7070-106 3153417 7/31/2014 Jul-16 60-7070-106 3154786 7/31/2014 Jul-16 60-7070-106 3155120 7/31/2014 Jul-16 60-7070-106 3155235 7/31/2014 Jul-16 60-7070-106 3155453 7/31/2014 Jul-16 60-7070-106 3156939 7/31/2014 Jul-16 60-7070-106 3157820 7/31/2014 Jul-16 60-7070-106 3159036 7/31/2014 Jul-16 60-7070-106 3160336 7/31/2014 Jul-16 60-7070-106 3162055 7/31/2014 Jul-16 60-7070-106 3163068 7/31/2014 Jul-16 60-7070-106 3163533 7/31/2014 Jul-16 60-7075-003 3161303 7/31/2014 Jul-16 60-7075-006 3160389 7/31/2014 Jul-16 60-7075-103 3155668 7/31/2014 Jul-16 60-7075-103 3155862 7/31/2014 Jul-16 60-7075-103 3157463 7/31/2014 Jul-16 60-7075-103 3160447 7/31/2014 Jul-16 60-7075-103 3162265 7/31/2014 Jul-16 60-7075-105 3160888 7/31/2014 Jul-16 60-7075-105 3161272 7/31/2014 Jul-16 60-7075-105 3162334 7/31/2014 Jul-16 60-7075-105 3162491 7/31/2014 Jul-16 60-7075-106 3154992 7/31/2014 Jul-16 60-7075-106 3155917 7/31/2014 Jul-16 60-7075-106 3158677 7/31/2014 Jul-16 60-7075-106 3161498 7/31/2014 Jul-16 5921-018-135 3165920 8/5/2014 Aug-16 5921-030-135 3164689 8/5/2014 Aug-16 5921-030-235 3165234 8/5/2014 Aug-16 5921-030-235 3165983 8/5/2014 Aug-16 5921-030-235 3167076 8/5/2014 Aug-16 5921-034-135 3164913 8/5/2014 Aug-16 5921-034-135 3166128 8/5/2014 Aug-16 5921-034-135 3167903 8/5/2014 Aug-16 5921-034-235 3164085 8/5/2014 Aug-16 5921-034-235 3165487 8/5/2014 Aug-16 5921-034-235 3166074 8/5/2014 Aug-16 5921-034-235 3167608 8/5/2014 Aug-16 5921-034-235 3167886 8/5/2014 Aug-16 5921-034-235 3167961 8/5/2014 Aug-16 5921-218-235 3164230 8/5/2014 Aug-16 5921-218-235 3165494 8/5/2014 Aug-16 5921-218-235 3166710 8/5/2014 Aug-16 5921-218-235 3168025 8/5/2014 Aug-16 5921-218-135 3164966 8/5/2014 Aug-16 5921-218-135 3168213 8/5/2014 Aug-16 60-7070-103 3163726 8/5/2014 Aug-16 60-7070-103 3163973 8/5/2014 Aug-16 60-7070-103 3165027 8/5/2014 Aug-16 60-7070-103 3165626 8/5/2014 Aug-16 60-7070-103 3166116 8/5/2014 Aug-16 60-7070-103 3167259 8/5/2014 Aug-16 60-7070-105 3164194 8/5/2014 Aug-16 60-7070-105 3165223 8/5/2014 Aug-16 60-7070-105 3165748 8/5/2014 Aug-16 60-7070-105 3166723 8/5/2014 Aug-16 60-7070-105 3167325 8/5/2014 Aug-16 60-7070-105 3167339 8/5/2014 Aug-16 60-7070-104 3165070 8/5/2014 Aug-16 60-7070-006 3165769 8/5/2014 Aug-16 60-7070-106 3163648 8/5/2014 Aug-16 60-7070-106 3164627 8/5/2014 Aug-16 60-7070-106 3164810 8/5/2014 Aug-16 60-7070-106 3166019 8/5/2014 Aug-16 60-7070-106 3166673 8/5/2014 Aug-16 60-7070-106 3167575 8/5/2014 Aug-16 60-7075-003 3165193 8/5/2014 Aug-16 60-7075-006 3167988 8/5/2014 Aug-16 60-7075-103 3163682 8/5/2014 Aug-16 60-7075-103 3165635 8/5/2014 Aug-16 60-7075-105 3165822 8/5/2014 Aug-16 60-7075-106 3163653 8/5/2014 Aug-16 60-7075-106 3165657 8/5/2014 Aug-16 5921-018-135 3171242 8/7/2014 Aug-16 5921-024-135 3172035 8/7/2014 Aug-16 5921-030-135 3168311 8/7/2014 Aug-16 5921-030-135 3169606 8/7/2014 Aug-16 5921-030-135 3172706 8/7/2014 Aug-16 5921-030-235 3168348 8/7/2014 Aug-16 5921-030-235 3169756 8/7/2014 Aug-16 5921-030-235 3170703 8/7/2014 Aug-16 5921-030-235 3171741 8/7/2014 Aug-16 5921-034-135 3172694 8/7/2014 Aug-16 5921-034-235 3169495 8/7/2014 Aug-16 5921-034-235 3170355 8/7/2014 Aug-16 5921-034-235 3170790 8/7/2014 Aug-16 5921-034-235 3170941 8/7/2014 Aug-16 5921-034-235 3172493 8/7/2014 Aug-16 5921-034-235 3172924 8/7/2014 Aug-16 5921-218-235 3169482 8/7/2014 Aug-16 5921-218-235 3170465 8/7/2014 Aug-16 5921-218-135 3168812 8/7/2014 Aug-16 5921-218-135 3172809 8/7/2014 Aug-16 60-7070-103 3168127 8/7/2014 Aug-16 60-7070-103 3168865 8/7/2014 Aug-16 60-7070-103 3170034 8/7/2014 Aug-16 60-7070-103 3170676 8/7/2014 Aug-16 60-7070-103 3172026 8/7/2014 Aug-16 60-7070-103 3172799 8/7/2014 Aug-16 60-7070-105 3168463 8/7/2014 Aug-16 60-7070-105 3170109 8/7/2014 Aug-16 60-7070-105 3170828 8/7/2014 Aug-16 60-7070-105 3172331 8/7/2014 Aug-16 60-7070-104 3169578 8/7/2014 Aug-16 60-7070-006 3172969 8/7/2014 Aug-16 60-7070-106 3168896 8/7/2014 Aug-16 60-7070-106 3170047 8/7/2014 Aug-16 60-7070-106 3170338 8/7/2014 Aug-16 60-7070-106 3170902 8/7/2014 Aug-16 60-7070-106 3171756 8/7/2014 Aug-16 60-7070-106 3172128 8/7/2014 Aug-16 60-7075-006 3172365 8/7/2014 Aug-16 60-7075-006 3172930 8/7/2014 Aug-16 60-7075-103 3168808 8/7/2014 Aug-16 60-7075-103 3170533 8/7/2014 Aug-16 60-7075-103 3171495 8/7/2014 Aug-16 60-7075-103 3171827 8/7/2014 Aug-16 60-7075-103 3173077 8/7/2014 Aug-16 60-7075-105 3170496 8/7/2014 Aug-16 60-7075-106 3168353 8/7/2014 Aug-16 60-7075-106 3168849 8/7/2014 Aug-16 60-7075-106 3170611 8/7/2014 Aug-16 60-7075-106 3173020 8/7/2014 Aug-16 5921-018-235 3174002 8/11/2014 Aug-16 5921-018-235 3177554 8/11/2014 Aug-16 5921-024-235 3176909 8/11/2014 Aug-16 5921-030-135 3178281 8/11/2014 Aug-16 5921-030-235 3173737 8/11/2014 Aug-16 5921-030-235 3175283 8/11/2014 Aug-16 5921-030-235 3177667 8/11/2014 Aug-16 5921-034-135 3174094 8/11/2014 Aug-16 5921-034-135 3174213 8/11/2014 Aug-16 5921-034-135 3177111 8/11/2014 Aug-16 5921-034-135 3178193 8/11/2014 Aug-16 5921-034-235 3174617 8/11/2014 Aug-16 5921-034-235 3175876 8/11/2014 Aug-16 5921-034-235 3177698 8/11/2014 Aug-16 5921-218-235 3169098 8/11/2014 Aug-16 5921-218-235 3173818 8/11/2014 Aug-16 5921-218-235 3175436 8/11/2014 Aug-16 5921-218-235 3176295 8/11/2014 Aug-16 5921-218-235 3178027 8/11/2014 Aug-16 5921-218-235 3178206 8/11/2014 Aug-16 5921-218-135 3174561 8/11/2014 Aug-16 5921-218-135 3178252 8/11/2014 Aug-16 60-7070-102 3175732 8/11/2014 Aug-16 60-7070-103 3174307 8/11/2014 Aug-16 60-7070-103 3175616 8/11/2014 Aug-16 60-7070-103 3177084 8/11/2014 Aug-16 60-7070-103 3178350 8/11/2014 Aug-16 60-7070-105 3173008 8/11/2014 Aug-16 60-7070-105 3174779 8/11/2014 Aug-16 60-7070-105 3176643 8/11/2014 Aug-16 60-7070-105 3178494 8/11/2014 Aug-16 60-7070-104 3173168 8/11/2014 Aug-16 60-7070-104 3174865 8/11/2014 Aug-16 60-7070-003 3175347 8/11/2014 Aug-16 60-7070-106 3173342 8/11/2014 Aug-16 60-7070-106 3175072 8/11/2014 Aug-16 60-7070-106 3175375 8/11/2014 Aug-16 60-7070-106 3177336 8/11/2014 Aug-16 60-7070-106 3177508 8/11/2014 Aug-16 60-7070-107 3173152 8/11/2014 Aug-16 60-7070-107 3174883 8/11/2014 Aug-16 60-7075-003 3173199 8/11/2014 Aug-16 60-7075-103 3174705 8/11/2014 Aug-16 60-7075-103 3177701 8/11/2014 Aug-16 60-7075-104 3173181 8/11/2014 Aug-16 60-7075-104 3177729 8/11/2014 Aug-16 60-7075-105 3175600 8/11/2014 Aug-16 60-7075-106 3174751 8/11/2014 Aug-16 60-7075-106 3176529 8/11/2014 Aug-16 60-7075-106 3178439 8/11/2014 Aug-16 60-7075-107 3175399 8/11/2014 Aug-16 5921-018-235 3178846 8/13/2014 Aug-16 5921-024-135 3180862 8/13/2014 Aug-16 5921-024-135 3181045 8/13/2014 Aug-16 5921-030-135 3179506 8/13/2014 Aug-16 5921-030-235 3178983 8/13/2014 Aug-16 5921-030-235 3183292 8/13/2014 Aug-16 5921-034-135 3180597 8/13/2014 Aug-16 5921-034-135 3181452 8/13/2014 Aug-16 5921-034-135 3181985 8/13/2014 Aug-16 5921-034-235 3179146 8/13/2014 Aug-16 5921-034-235 3180787 8/13/2014 Aug-16 5921-034-235 3181916 8/13/2014 Aug-16 5921-218-235 3179696 8/13/2014 Aug-16 5921-218-235 3181032 8/13/2014 Aug-16 5921-218-135 3180495 8/13/2014 Aug-16 5921-218-135 3182087 8/13/2014 Aug-16 60-7070-103 3175750 8/13/2014 Aug-16 60-7070-103 3179704 8/13/2014 Aug-16 60-7070-103 3180566 8/13/2014 Aug-16 60-7070-103 3181772 8/13/2014 Aug-16 60-7070-103 3182672 8/13/2014 Aug-16 60-7070-103 3183007 8/13/2014 Aug-16 60-7070-105 3178867 8/13/2014 Aug-16 60-7070-105 3180652 8/13/2014 Aug-16 60-7070-105 3180889 8/13/2014 Aug-16 60-7070-105 3181888 8/13/2014 Aug-16 60-7070-105 3182221 8/13/2014 Aug-16 60-7070-105 3182968 8/13/2014 Aug-16 60-7070-105 3183030 8/13/2014 Aug-16 60-7070-104 3180051 8/13/2014 Aug-16 60-7070-104 3180901 8/13/2014 Aug-16 60-7070-106 3179198 8/13/2014 Aug-16 60-7070-106 3180131 8/13/2014 Aug-16 60-7070-106 3182459 8/13/2014 Aug-16 60-7070-106 3182535 8/13/2014 Aug-16 60-7070-107 3180870 8/13/2014 Aug-16 60-7070-107 3181410 8/13/2014 Aug-16 60-7075-003 3180046 8/13/2014 Aug-16 60-7075-004 3180938 8/13/2014 Aug-16 60-7075-006 3179510 8/13/2014 Aug-16 60-7075-006 3179903 8/13/2014 Aug-16 60-7080-102 3181111 8/13/2014 Aug-16 60-7075-103 3178996 8/13/2014 Aug-16 60-7075-103 3181391 8/13/2014 Aug-16 60-7075-106 3180647 8/13/2014 Aug-16 60-7075-106 3183095 8/13/2014 Aug-16 5921-018-135 3183387 8/15/2014 Aug-16 5921-018-235 3183409 8/15/2014 Aug-16 5921-018-235 3187595 8/15/2014 Aug-16 5921-024-135 3186091 8/15/2014 Aug-16 5921-030-135 3184573 8/15/2014 Aug-16 5921-030-135 3186101 8/15/2014 Aug-16 5921-030-136 3185810 8/15/2014 Aug-16 5921-030-235 3186920 8/15/2014 Aug-16 5921-030-235 3188504 8/15/2014 Aug-16 5921-034-135 3183665 8/15/2014 Aug-16 5921-034-135 3186748 8/15/2014 Aug-16 5921-034-135 3188100 8/15/2014 Aug-16 5921-034-136 3184933 8/15/2014 Aug-16 5921-034-235 3183133 8/15/2014 Aug-16 5921-034-235 3183523 8/15/2014 Aug-16 5921-034-235 3183935 8/15/2014 Aug-16 5921-034-235 3184600 8/15/2014 Aug-16 5921-034-235 3185604 8/15/2014 Aug-16 5921-034-235 3187077 8/15/2014 Aug-16 5921-034-236 3185796 8/15/2014 Aug-16 5921-218-235 3184306 8/15/2014 Aug-16 5921-218-235 3186425 8/15/2014 Aug-16 5921-218-235 3188144 8/15/2014 Aug-16 5921-044-135 3185643 8/15/2014 Aug-16 5921-044-235 3181013 8/15/2014 Aug-16 5921-218-135 3184323 8/15/2014 Aug-16 5921-218-135 3186252 8/15/2014 Aug-16 60-7070-103 3183484 8/15/2014 Aug-16 60-7070-103 3184518 8/15/2014 Aug-16 60-7070-103 3185658 8/15/2014 Aug-16 60-7070-103 3187480 8/15/2014 Aug-16 60-7070-105 3183654 8/15/2014 Aug-16 60-7070-105 3185847 8/15/2014 Aug-16 60-7070-105 3187818 8/15/2014 Aug-16 60-7070-104 3186234 8/15/2014 Aug-16 60-7070-104 3188233 8/15/2014 Aug-16 60-7070-003 3186753 8/15/2014 Aug-16 60-7070-006 3183151 8/15/2014 Aug-16 60-7070-006 3186172 8/15/2014 Aug-16 60-7070-106 3185586 8/15/2014 Aug-16 60-7070-106 3185691 8/15/2014 Aug-16 60-7070-106 3187065 8/15/2014 Aug-16 60-7070-106 3187374 8/15/2014 Aug-16 60-7070-106 3188061 8/15/2014 Aug-16 60-7070-107 3184922 8/15/2014 Aug-16 60-7075-003 3183677 8/15/2014 Aug-16 60-7075-006 3183216 8/15/2014 Aug-16 60-7080-102 3188354 8/15/2014 Aug-16 60-7075-103 3184119 8/15/2014 Aug-16 60-7075-103 3185945 8/15/2014 Aug-16 60-7075-104 3186138 8/15/2014 Aug-16 60-7075-105 3186674 8/15/2014 Aug-16 60-7075-105 3188280 8/15/2014 Aug-16 60-7075-106 3185037 8/15/2014 Aug-16 60-7075-106 3185627 8/15/2014 Aug-16 60-7075-106 3186559 8/15/2014 Aug-16 60-7075-106 3188279 8/15/2014 Aug-16 5921-018-135 3193712 8/18/2014 Aug-16 5921-024-135 3192986 8/18/2014 Aug-16 5921-024-235 3188298 8/18/2014 Aug-16 5921-024-235 3189301 8/18/2014 Aug-16 5921-030-135 3190906 8/18/2014 Aug-16 5921-030-135 3191171 8/18/2014 Aug-16 5921-034-135 3189408 8/18/2014 Aug-16 5921-034-135 3192131 8/18/2014 Aug-16 5921-034-235 3188515 8/18/2014 Aug-16 5921-034-235 3190705 8/18/2014 Aug-16 5921-034-235 3192955 8/18/2014 Aug-16 5921-034-235 3193529 8/18/2014 Aug-16 5921-218-235 3189800 8/18/2014 Aug-16 5921-218-235 3191850 8/18/2014 Aug-16 5921-218-235 3193490 8/18/2014 Aug-16 5921-218-135 3188673 8/18/2014 Aug-16 5921-218-135 3192482 8/18/2014 Aug-16 60-7070-103 3189412 8/18/2014 Aug-16 60-7070-103 3189703 8/18/2014 Aug-16 60-7070-103 3190170 8/18/2014 Aug-16 60-7070-103 3190284 8/18/2014 Aug-16 60-7070-103 3190471 8/18/2014 Aug-16 60-7070-103 3191222 8/18/2014 Aug-16 60-7070-103 3192972 8/18/2014 Aug-16 60-7070-103 3193248 8/18/2014 Aug-16 60-7070-105 3189178 8/18/2014 Aug-16 60-7070-105 3190746 8/18/2014 Aug-16 60-7070-105 3192392 8/18/2014 Aug-16 60-7070-105 3193765 8/18/2014 Aug-16 60-7070-104 3190999 8/18/2014 Aug-16 60-7070-003 3192476 8/18/2014 Aug-16 60-7070-005 3193555 8/18/2014 Aug-16 60-7070-106 3189892 8/18/2014 Aug-16 60-7070-106 3190975 8/18/2014 Aug-16 60-7070-106 3191781 8/18/2014 Aug-16 60-7070-107 3190143 8/18/2014 Aug-16 60-7070-107 3193517 8/18/2014 Aug-16 60-7075-003 3189828 8/18/2014 Aug-16 60-7075-006 3191019 8/18/2014 Aug-16 60-7075-103 3188602 8/18/2014 Aug-16 60-7075-103 3190751 8/18/2014 Aug-16 60-7075-104 3192469 8/18/2014 Aug-16 60-7075-106 3192120 8/18/2014 Aug-16 5921-018-235 3196388 8/20/2014 Aug-16 5921-024-135 3196763 8/20/2014 Aug-16 5921-024-235 3195575 8/20/2014 Aug-16 5921-024-235 3198256 8/20/2014 Aug-16 5921-030-135 3194166 8/20/2014 Aug-16 5921-030-235 3194081 8/20/2014 Aug-16 5921-030-235 3197707 8/20/2014 Aug-16 5921-030-235 3198963 8/20/2014 Aug-16 5921-034-135 3193881 8/20/2014 Aug-16 5921-034-135 3194579 8/20/2014 Aug-16 5921-034-135 3196390 8/20/2014 Aug-16 5921-034-235 3195421 8/20/2014 Aug-16 5921-034-235 3197164 8/20/2014 Aug-16 5921-034-235 3197366 8/20/2014 Aug-16 5921-034-235 3198481 8/20/2014 Aug-16 5921-034-235 3198596 8/20/2014 Aug-16 5921-034-236 3196451 8/20/2014 Aug-16 5921-218-235 3195082 8/20/2014 Aug-16 5921-218-235 3197559 8/20/2014 Aug-16 5921-218-235 3198985 8/20/2014 Aug-16 5921-044-235 3195919 8/20/2014 Aug-16 5921-218-135 3193868 8/20/2014 Aug-16 5921-218-135 3198045 8/20/2014 Aug-16 60-7070-102 3196530 8/20/2014 Aug-16 60-7070-103 3193834 8/20/2014 Aug-16 60-7070-103 3196106 8/20/2014 Aug-16 60-7070-103 3196873 8/20/2014 Aug-16 60-7070-105 3195146 8/20/2014 Aug-16 60-7070-105 3196553 8/20/2014 Aug-16 60-7070-105 3198916 8/20/2014 Aug-16 60-7070-104 3195903 8/20/2014 Aug-16 60-7070-104 3196049 8/20/2014 Aug-16 60-7070-104 3198124 8/20/2014 Aug-16 60-7070-006 3195413 8/20/2014 Aug-16 60-7070-106 3193796 8/20/2014 Aug-16 60-7070-106 3195547 8/20/2014 Aug-16 60-7070-106 3196446 8/20/2014 Aug-16 60-7070-106 3198499 8/20/2014 Aug-16 60-7070-107 3196569 8/20/2014 Aug-16 60-7075-004 3196759 8/20/2014 Aug-16 60-7075-006 3196737 8/20/2014 Aug-16 60-7080-102 3198959 8/20/2014 Aug-16 60-7075-103 3193899 8/20/2014 Aug-16 60-7075-103 3197690 8/20/2014 Aug-16 60-7075-104 3196728 8/20/2014 Aug-16 60-7075-105 3197642 8/20/2014 Aug-16 60-7075-106 3194757 8/20/2014 Aug-16 60-7075-106 3196576 8/20/2014 Aug-16 5921-018-135 3201281 8/22/2014 Aug-16 5921-018-235 3201880 8/22/2014 Aug-16 5921-024-235 3201795 8/22/2014 Aug-16 5921-030-135 3200464 8/22/2014 Aug-16 5921-030-135 3203541 8/22/2014 Aug-16 5921-030-235 3199231 8/22/2014 Aug-16 5921-030-235 3201091 8/22/2014 Aug-16 5921-030-235 3203895 8/22/2014 Aug-16 5921-034-135 3201378 8/22/2014 Aug-16 5921-034-135 3202262 8/22/2014 Aug-16 5921-034-135 3202408 8/22/2014 Aug-16 5921-034-235 3199188 8/22/2014 Aug-16 5921-034-235 3199616 8/22/2014 Aug-16 5921-034-235 3201805 8/22/2014 Aug-16 5921-034-235 3202828 8/22/2014 Aug-16 5921-034-235 3203174 8/22/2014 Aug-16 5921-034-235 3204167 8/22/2014 Aug-16 5921-034-235 3204484 8/22/2014 Aug-16 5921-218-235 3200948 8/22/2014 Aug-16 5921-218-235 3202166 8/22/2014 Aug-16 5921-218-235 3203784 8/22/2014 Aug-16 5921-218-235 3204510 8/22/2014 Aug-16 5921-218-135 3200185 8/22/2014 Aug-16 5921-218-135 3203208 8/22/2014 Aug-16 60-7070-103 3198971 8/22/2014 Aug-16 60-7070-103 3199573 8/22/2014 Aug-16 60-7070-103 3201101 8/22/2014 Aug-16 60-7070-103 3201898 8/22/2014 Aug-16 60-7070-103 3202579 8/22/2014 Aug-16 60-7070-103 3203998 8/22/2014 Aug-16 60-7070-105 3199443 8/22/2014 Aug-16 60-7070-105 3201384 8/22/2014 Aug-16 60-7070-105 3202580 8/22/2014 Aug-16 60-7070-104 3199594 8/22/2014 Aug-16 60-7070-104 3200061 8/22/2014 Aug-16 60-7070-104 3202483 8/22/2014 Aug-16 60-7070-003 3200429 8/22/2014 Aug-16 60-7070-004 3200813 8/22/2014 Aug-16 60-7070-006 3199470 8/22/2014 Aug-16 60-7070-106 3199347 8/22/2014 Aug-16 60-7070-106 3199491 8/22/2014 Aug-16 60-7070-106 3199556 8/22/2014 Aug-16 60-7070-106 3201397 8/22/2014 Aug-16 60-7070-106 3202372 8/22/2014 Aug-16 60-7070-106 3203443 8/22/2014 Aug-16 60-7070-106 3203556 8/22/2014 Aug-16 60-7070-107 3199587 8/22/2014 Aug-16 60-7075-006 3200205 8/22/2014 Aug-16 60-7075-006 3200656 8/22/2014 Aug-16 60-7075-102 3202150 8/22/2014 Aug-16 60-7075-103 3199560 8/22/2014 Aug-16 60-7075-103 3203766 8/22/2014 Aug-16 60-7075-106 3199220 8/22/2014 Aug-16 60-7075-106 3202023 8/22/2014 Aug-16 5921-018-135 3204974 8/25/2014 Aug-16 5921-018-235 3207419 8/25/2014 Aug-16 5921-024-235 3204547 8/25/2014 Aug-16 5921-024-235 3207194 8/25/2014 Aug-16 5921-024-235 3207276 8/25/2014 Aug-16 5921-024-235 3207945 8/25/2014 Aug-16 5921-030-136 3206481 8/25/2014 Aug-16 5921-030-235 3204534 8/25/2014 Aug-16 5921-030-235 3205771 8/25/2014 Aug-16 5921-030-235 3206959 8/25/2014 Aug-16 5921-030-235 3207909 8/25/2014 Aug-16 5921-034-135 3204770 8/25/2014 Aug-16 5921-034-135 3207314 8/25/2014 Aug-16 5921-034-235 3204704 8/25/2014 Aug-16 5921-034-235 3205737 8/25/2014 Aug-16 5921-034-235 3208993 8/25/2014 Aug-16 5921-218-235 3205217 8/25/2014 Aug-16 5921-218-235 3205319 8/25/2014 Aug-16 5921-218-235 3206406 8/25/2014 Aug-16 5921-218-235 3207796 8/25/2014 Aug-16 5921-218-235 3208584 8/25/2014 Aug-16 5921-044-135 3207403 8/25/2014 Aug-16 5921-218-135 3207207 8/25/2014 Aug-16 5921-218-135 3208610 8/25/2014 Aug-16 60-7070-103 3205077 8/25/2014 Aug-16 60-7070-103 3205927 8/25/2014 Aug-16 60-7070-103 3206786 8/25/2014 Aug-16 60-7070-103 3207295 8/25/2014 Aug-16 60-7070-103 3207913 8/25/2014 Aug-16 60-7070-105 3204715 8/25/2014 Aug-16 60-7070-105 3205272 8/25/2014 Aug-16 60-7070-105 3205716 8/25/2014 Aug-16 60-7070-105 3206963 8/25/2014 Aug-16 60-7070-105 3207555 8/25/2014 Aug-16 60-7070-105 3207932 8/25/2014 Aug-16 60-7070-104 3205326 8/25/2014 Aug-16 60-7070-104 3207615 8/25/2014 Aug-16 60-7070-005 3208545 8/25/2014 Aug-16 60-7070-106 3205083 8/25/2014 Aug-16 60-7070-106 3206819 8/25/2014 Aug-16 60-7070-106 3207806 8/25/2014 Aug-16 60-7070-106 3208787 8/25/2014 Aug-16 60-7070-106 3209073 8/25/2014 Aug-16 60-7075-004 3205936 8/25/2014 Aug-16 60-7075-103 3206857 8/25/2014 Aug-16 60-7075-103 3208964 8/25/2014 Aug-16 60-7075-105 3204758 8/25/2014 Aug-16 60-7075-106 3204762 8/25/2014 Aug-16 60-7075-106 3206468 8/25/2014 Aug-16 60-7075-106 3207604 8/25/2014 Aug-16 5921-018-135 3214312 8/27/2014 Aug-16 5921-018-235 3209981 8/27/2014 Aug-16 5921-018-235 3213021 8/27/2014 Aug-16 5921-024-235 3212879 8/27/2014 Aug-16 5921-030-135 3210030 8/27/2014 Aug-16 5921-030-135 3212707 8/27/2014 Aug-16 5921-030-135 3212920 8/27/2014 Aug-16 5921-030-235 3211742 8/27/2014 Aug-16 5921-030-235 3214088 8/27/2014 Aug-16 5921-034-135 3208760 8/27/2014 Aug-16 5921-034-135 3210484 8/27/2014 Aug-16 5921-034-135 3212456 8/27/2014 Aug-16 5921-034-135 3213826 8/27/2014 Aug-16 5921-034-235 3209400 8/27/2014 Aug-16 5921-034-235 3210872 8/27/2014 Aug-16 5921-034-235 3211270 8/27/2014 Aug-16 5921-034-235 3211551 8/27/2014 Aug-16 5921-034-235 3213819 8/27/2014 Aug-16 5921-034-235 3214163 8/27/2014 Aug-16 5921-034-235 3215213 8/27/2014 Aug-16 5921-218-235 3210789 8/27/2014 Aug-16 5921-218-235 3211317 8/27/2014 Aug-16 5921-218-235 3212912 8/27/2014 Aug-16 5921-218-235 3213520 8/27/2014 Aug-16 5921-218-135 3212350 8/27/2014 Aug-16 5921-218-135 3213963 8/27/2014 Aug-16 60-7070-103 3209232 8/27/2014 Aug-16 60-7070-103 3210151 8/27/2014 Aug-16 60-7070-103 3210801 8/27/2014 Aug-16 60-7070-103 3212138 8/27/2014 Aug-16 60-7070-103 3212439 8/27/2014 Aug-16 60-7070-103 3213343 8/27/2014 Aug-16 60-7070-103 3214486 8/27/2014 Aug-16 60-7070-105 3209617 8/27/2014 Aug-16 60-7070-105 3209896 8/27/2014 Aug-16 60-7070-105 3210432 8/27/2014 Aug-16 60-7070-105 3210497 8/27/2014 Aug-16 60-7070-105 3211956 8/27/2014 Aug-16 60-7070-105 3214331 8/27/2014 Aug-16 60-7070-104 3209906 8/27/2014 Aug-16 60-7070-104 3212155 8/27/2014 Aug-16 60-7070-104 3212814 8/27/2014 Aug-16 60-7070-003 3210053 8/27/2014 Aug-16 60-7070-003 3212724 8/27/2014 Aug-16 60-7070-005 3214930 8/27/2014 Aug-16 60-7070-006 3209975 8/27/2014 Aug-16 60-7070-106 3210048 8/27/2014 Aug-16 60-7070-106 3210743 8/27/2014 Aug-16 60-7070-106 3211653 8/27/2014 Aug-16 60-7070-106 3213287 8/27/2014 Aug-16 60-7070-106 3213531 8/27/2014 Aug-16 60-7070-106 3214768 8/27/2014 Aug-16 60-7070-107 3213107 8/27/2014 Aug-16 60-7070-107 3213273 8/27/2014 Aug-16 60-7075-003 3212880 8/27/2014 Aug-16 60-7075-004 3209910 8/27/2014 Aug-16 60-7075-006 3213944 8/27/2014 Aug-16 60-7075-103 3211892 8/27/2014 Aug-16 60-7075-105 3210180 8/27/2014 Aug-16 60-7075-106 3210216 8/27/2014 Aug-16 60-7075-106 3212690 8/27/2014 Aug-16 60-7075-106 3215162 8/27/2014 Aug-16 5921-018-135 3220999 8/29/2014 Aug-16 5921-018-235 3218490 8/29/2014 Aug-16 5921-024-135 3219792 8/29/2014 Aug-16 5921-024-235 3215503 8/29/2014 Aug-16 5921-024-235 3218398 8/29/2014 Aug-16 5921-024-236 3220746 8/29/2014 Aug-16 5921-030-135 3215686 8/29/2014 Aug-16 5921-030-135 3218474 8/29/2014 Aug-16 5921-030-135 3220463 8/29/2014 Aug-16 5921-030-235 3216072 8/29/2014 Aug-16 5921-030-235 3219326 8/29/2014 Aug-16 5921-030-235 3219825 8/29/2014 Aug-16 5921-034-135 3198992 8/29/2014 Aug-16 5921-034-135 3216265 8/29/2014 Aug-16 5921-034-135 3217955 8/29/2014 Aug-16 5921-034-135 3219272 8/29/2014 Aug-16 5921-034-135 3220428 8/29/2014 Aug-16 5921-034-136 3220369 8/29/2014 Aug-16 5921-034-235 3219480 8/29/2014 Aug-16 5921-034-236 3220136 8/29/2014 Aug-16 5921-218-235 3214911 8/29/2014 Aug-16 5921-218-235 3218380 8/29/2014 Aug-16 5921-218-235 3219335 8/29/2014 Aug-16 5921-218-135 3215883 8/29/2014 Aug-16 60-7070-103 3215640 8/29/2014 Aug-16 60-7070-103 3216750 8/29/2014 Aug-16 60-7070-103 3218351 8/29/2014 Aug-16 60-7070-103 3218754 8/29/2014 Aug-16 60-7070-103 3219936 8/29/2014 Aug-16 60-7070-105 3215519 8/29/2014 Aug-16 60-7070-105 3218037 8/29/2014 Aug-16 60-7070-105 3218322 8/29/2014 Aug-16 60-7070-105 3218847 8/29/2014 Aug-16 60-7070-105 3220191 8/29/2014 Aug-16 60-7070-104 3216086 8/29/2014 Aug-16 60-7070-104 3218435 8/29/2014 Aug-16 60-7070-104 3219576 8/29/2014 Aug-16 60-7070-003 3218517 8/29/2014 Aug-16 60-7070-006 3218708 8/29/2014 Aug-16 60-7070-106 3216277 8/29/2014 Aug-16 60-7070-106 3217675 8/29/2014 Aug-16 60-7070-106 3218295 8/29/2014 Aug-16 60-7070-106 3220278 8/29/2014 Aug-16 60-7070-106 3220672 8/29/2014 Aug-16 60-7075-003 3216283 8/29/2014 Aug-16 60-7075-103 3215780 8/29/2014 Aug-16 60-7075-103 3218442 8/29/2014 Aug-16 60-7075-103 3220519 8/29/2014 Aug-16 60-7075-105 3215722 8/29/2014 Aug-16 60-7075-105 3220722 8/29/2014 Aug-16 60-7075-106 3217940 8/29/2014 Aug-16 60-7075-106 3218529 8/29/2014 Aug-16 60-7075-106 3219303 8/29/2014 Aug-16 5921-018-235 3226303 9/5/2014 Sep-16 5921-024-135 3228531 9/5/2014 Sep-16 5921-024-235 3226286 9/5/2014 Sep-16 5921-030-135 3228391 9/5/2014 Sep-16 5921-030-235 3225620 9/5/2014 Sep-16 5921-030-235 3226632 9/5/2014 Sep-16 5921-030-235 3228084 9/5/2014 Sep-16 5921-034-135 3226342 9/5/2014 Sep-16 5921-034-235 3226645 9/5/2014 Sep-16 5921-034-235 3226825 9/5/2014 Sep-16 5921-218-235 3226621 9/5/2014 Sep-16 5921-218-235 3229072 9/5/2014 Sep-16 5921-218-135 3228066 9/5/2014 Sep-16 60-7070-103 3226134 9/5/2014 Sep-16 60-7070-103 3226802 9/5/2014 Sep-16 60-7070-103 3229415 9/5/2014 Sep-16 60-7070-105 3226548 9/5/2014 Sep-16 60-7070-105 3226744 9/5/2014 Sep-16 60-7070-105 3228343 9/5/2014 Sep-16 60-7070-104 3226152 9/5/2014 Sep-16 60-7070-106 3225829 9/5/2014 Sep-16 60-7070-106 3228819 9/5/2014 Sep-16 60-7070-106 3228937 9/5/2014 Sep-16 60-7070-107 3226511 9/5/2014 Sep-16 60-7075-006 3226480 9/5/2014 Sep-16 60-7075-103 3228045 9/5/2014 Sep-16 60-7075-105 3228698 9/5/2014 Sep-16 60-7075-106 3228520 9/5/2014 Sep-16 5921-018-135 3221100 9/8/2014 Sep-16 5921-018-235 3221074 9/8/2014 Sep-16 5921-024-235 3222274 9/8/2014 Sep-16 5921-024-235 3224150 9/8/2014 Sep-16 5921-030-135 3222964 9/8/2014 Sep-16 5921-030-136 3221308 9/8/2014 Sep-16 5921-030-235 3216600 9/8/2014 Sep-16 5921-030-235 3221694 9/8/2014 Sep-16 5921-030-235 3223540 9/8/2014 Sep-16 5921-034-135 3221768 9/8/2014 Sep-16 5921-034-135 3224132 9/8/2014 Sep-16 5921-034-136 3221320 9/8/2014 Sep-16 5921-034-136 3222550 9/8/2014 Sep-16 5921-034-235 3217424 9/8/2014 Sep-16 5921-034-235 3221660 9/8/2014 Sep-16 5921-034-235 3223102 9/8/2014 Sep-16 5921-034-235 3224956 9/8/2014 Sep-16 5921-034-235 3225179 9/8/2014 Sep-16 5921-034-236 3221349 9/8/2014 Sep-16 5921-218-235 3217242 9/8/2014 Sep-16 5921-218-235 3222108 9/8/2014 Sep-16 5921-218-235 3224703 9/8/2014 Sep-16 5921-018-236 3221233 9/8/2014 Sep-16 5921-030-236 3221279 9/8/2014 Sep-16 5921-218-135 3222337 9/8/2014 Sep-16 5921-218-135 3223806 9/8/2014 Sep-16 5921-218-236 3221383 9/8/2014 Sep-16 60-7070-103 3217067 9/8/2014 Sep-16 60-7070-103 3221026 9/8/2014 Sep-16 60-7070-103 3221543 9/8/2014 Sep-16 60-7070-103 3222067 9/8/2014 Sep-16 60-7070-103 3222699 9/8/2014 Sep-16 60-7070-103 3223295 9/8/2014 Sep-16 60-7070-103 3223593 9/8/2014 Sep-16 60-7070-103 3225133 9/8/2014 Sep-16 60-7070-105 3221171 9/8/2014 Sep-16 60-7070-105 3221591 9/8/2014 Sep-16 60-7070-105 3222725 9/8/2014 Sep-16 60-7070-105 3223305 9/8/2014 Sep-16 60-7070-105 3223834 9/8/2014 Sep-16 60-7070-104 3221570 9/8/2014 Sep-16 60-7070-104 3223847 9/8/2014 Sep-16 60-7070-003 3222955 9/8/2014 Sep-16 60-7070-006 3220733 9/8/2014 Sep-16 60-7070-006 3224262 9/8/2014 Sep-16 60-7070-106 3221377 9/8/2014 Sep-16 60-7070-106 3222288 9/8/2014 Sep-16 60-7070-106 3223346 9/8/2014 Sep-16 60-7070-106 3223572 9/8/2014 Sep-16 60-7070-106 3224113 9/8/2014 Sep-16 60-7075-003 3217392 9/8/2014 Sep-16 60-7075-004 3221205 9/8/2014 Sep-16 60-7075-005 3222741 9/8/2014 Sep-16 60-7075-103 3221687 9/8/2014 Sep-16 60-7075-103 3223139 9/8/2014 Sep-16 60-7075-103 3224204 9/8/2014 Sep-16 60-7075-103 3224483 9/8/2014 Sep-16 60-7075-104 3224918 9/8/2014 Sep-16 60-7075-105 3223141 9/8/2014 Sep-16 60-7075-106 3217699 9/8/2014 Sep-16 60-7075-106 3217843 9/8/2014 Sep-16 60-7075-106 3221602 9/8/2014 Sep-16 60-7075-106 3223643 9/8/2014 Sep-16 60-7075-106 3224844 9/8/2014 Sep-16 60-7075-107 3223555 9/8/2014 Sep-16 5921-018-135 3234767 9/10/2014 Sep-16 5921-018-235 3234213 9/10/2014 Sep-16 5921-030-135 3231482 9/10/2014 Sep-16 5921-030-135 3231739 9/10/2014 Sep-16 5921-030-135 3234831 9/10/2014 Sep-16 5921-030-235 3229980 9/10/2014 Sep-16 5921-030-235 3232100 9/10/2014 Sep-16 5921-034-135 3231242 9/10/2014 Sep-16 5921-034-135 3234094 9/10/2014 Sep-16 5921-034-135 3234820 9/10/2014 Sep-16 5921-034-235 3229489 9/10/2014 Sep-16 5921-034-235 3230611 9/10/2014 Sep-16 5921-034-235 3231424 9/10/2014 Sep-16 5921-034-235 3232057 9/10/2014 Sep-16 5921-034-235 3233716 9/10/2014 Sep-16 5921-218-235 3229807 9/10/2014 Sep-16 5921-218-235 3233077 9/10/2014 Sep-16 5921-218-235 3234780 9/10/2014 Sep-16 5921-218-135 3231112 9/10/2014 Sep-16 5921-218-135 3232897 9/10/2014 Sep-16 60-7070-103 3230075 9/10/2014 Sep-16 60-7070-103 3230204 9/10/2014 Sep-16 60-7070-103 3231741 9/10/2014 Sep-16 60-7070-103 3232377 9/10/2014 Sep-16 60-7070-103 3233326 9/10/2014 Sep-16 60-7070-103 3233388 9/10/2014 Sep-16 60-7070-105 3229491 9/10/2014 Sep-16 60-7070-105 3230719 9/10/2014 Sep-16 60-7070-105 3232884 9/10/2014 Sep-16 60-7070-105 3234733 9/10/2014 Sep-16 60-7070-104 3229443 9/10/2014 Sep-16 60-7070-104 3233737 9/10/2014 Sep-16 60-7070-003 3234115 9/10/2014 Sep-16 60-7070-005 3234209 9/10/2014 Sep-16 60-7070-106 3231257 9/10/2014 Sep-16 60-7070-106 3232171 9/10/2014 Sep-16 60-7070-106 3232752 9/10/2014 Sep-16 60-7070-106 3233498 9/10/2014 Sep-16 60-7075-103 3231120 9/10/2014 Sep-16 60-7075-105 3232214 9/10/2014 Sep-16 60-7075-106 3230607 9/10/2014 Sep-16 60-7075-106 3232527 9/10/2014 Sep-16 5921-018-135 3238724 9/12/2014 Sep-16 5921-018-235 3238223 9/12/2014 Sep-16 5921-018-235 3238252 9/12/2014 Sep-16 5921-024-135 3238004 9/12/2014 Sep-16 5921-024-235 3236720 9/12/2014 Sep-16 5921-030-235 3227846 9/12/2014 Sep-16 5921-030-235 3235220 9/12/2014 Sep-16 5921-030-235 3240273 9/12/2014 Sep-16 5921-034-135 3235001 9/12/2014 Sep-16 5921-034-135 3236322 9/12/2014 Sep-16 5921-034-135 3238776 9/12/2014 Sep-16 5921-034-136 3239631 9/12/2014 Sep-16 5921-034-235 3235029 9/12/2014 Sep-16 5921-034-235 3237294 9/12/2014 Sep-16 5921-034-235 3237698 9/12/2014 Sep-16 5921-034-235 3239704 9/12/2014 Sep-16 5921-034-236 3238101 9/12/2014 Sep-16 5921-218-235 3236139 9/12/2014 Sep-16 5921-218-235 3238954 9/12/2014 Sep-16 5921-030-236 3238193 9/12/2014 Sep-16 5921-218-135 3235038 9/12/2014 Sep-16 5921-218-236 3238129 9/12/2014 Sep-16 60-7070-103 3230372 9/12/2014 Sep-16 60-7070-103 3237786 9/12/2014 Sep-16 60-7070-103 3239007 9/12/2014 Sep-16 60-7070-105 3235895 9/12/2014 Sep-16 60-7070-105 3237013 9/12/2014 Sep-16 60-7070-105 3237577 9/12/2014 Sep-16 60-7070-105 3238949 9/12/2014 Sep-16 60-7070-105 3239095 9/12/2014 Sep-16 60-7070-104 3235212 9/12/2014 Sep-16 60-7070-104 3237963 9/12/2014 Sep-16 60-7070-003 3237622 9/12/2014 Sep-16 60-7070-006 3236749 9/12/2014 Sep-16 60-7070-006 3238164 9/12/2014 Sep-16 60-7070-106 3235334 9/12/2014 Sep-16 60-7070-106 3236966 9/12/2014 Sep-16 60-7070-106 3237130 9/12/2014 Sep-16 60-7070-106 3238657 9/12/2014 Sep-16 60-7070-106 3239654 9/12/2014 Sep-16 60-7070-106 3240153 9/12/2014 Sep-16 60-7070-107 3236754 9/12/2014 Sep-16 60-7075-003 3236125 9/12/2014 Sep-16 60-7075-003 3237633 9/12/2014 Sep-16 60-7075-006 3236295 9/12/2014 Sep-16 60-7075-103 3235017 9/12/2014 Sep-16 60-7075-103 3237680 9/12/2014 Sep-16 60-7075-106 3237701 9/12/2014 Sep-16 5921-018-135 3244269 9/16/2014 Sep-16 5921-024-135 3243023 9/16/2014 Sep-16 5921-024-235 3240260 9/16/2014 Sep-16 5921-024-235 3244211 9/16/2014 Sep-16 5921-030-135 3242980 9/16/2014 Sep-16 5921-030-235 3240475 9/16/2014 Sep-16 5921-030-235 3241467 9/16/2014 Sep-16 5921-030-235 3245626 9/16/2014 Sep-16 5921-030-235 3245724 9/16/2014 Sep-16 5921-034-135 3242161 9/16/2014 Sep-16 5921-034-135 3242853 9/16/2014 Sep-16 5921-034-135 3244731 9/16/2014 Sep-16 5921-034-235 3242038 9/16/2014 Sep-16 5921-034-235 3243465 9/16/2014 Sep-16 5921-034-235 3245537 9/16/2014 Sep-16 5921-218-235 3240755 9/16/2014 Sep-16 5921-218-235 3240857 9/16/2014 Sep-16 5921-218-235 3241326 9/16/2014 Sep-16 5921-218-235 3242869 9/16/2014 Sep-16 5921-044-235 3244932 9/16/2014 Sep-16 5921-218-135 3240294 9/16/2014 Sep-16 5921-218-135 3244593 9/16/2014 Sep-16 60-7070-102 3243034 9/16/2014 Sep-16 60-7070-103 3240195 9/16/2014 Sep-16 60-7070-103 3240842 9/16/2014 Sep-16 60-7070-103 3242491 9/16/2014 Sep-16 60-7070-103 3242848 9/16/2014 Sep-16 60-7070-103 3243522 9/16/2014 Sep-16 60-7070-103 3244230 9/16/2014 Sep-16 60-7070-105 3240452 9/16/2014 Sep-16 60-7070-105 3241621 9/16/2014 Sep-16 60-7070-105 3242029 9/16/2014 Sep-16 60-7070-105 3243227 9/16/2014 Sep-16 60-7070-105 3244187 9/16/2014 Sep-16 60-7070-105 3244810 9/16/2014 Sep-16 60-7070-105 3245281 9/16/2014 Sep-16 60-7070-104 3242157 9/16/2014 Sep-16 60-7070-104 3245879 9/16/2014 Sep-16 60-7070-006 3245223 9/16/2014 Sep-16 60-7070-106 3240729 9/16/2014 Sep-16 60-7070-106 3243204 9/16/2014 Sep-16 60-7070-106 3243420 9/16/2014 Sep-16 60-7070-106 3243918 9/16/2014 Sep-16 60-7070-106 3244921 9/16/2014 Sep-16 60-7070-107 3240667 9/16/2014 Sep-16 60-7075-003 3244442 9/16/2014 Sep-16 60-7075-006 3243372 9/16/2014 Sep-16 60-7075-006 3244708 9/16/2014 Sep-16 60-7075-103 3241839 9/16/2014 Sep-16 60-7075-104 3242933 9/16/2014 Sep-16 60-7075-105 3242905 9/16/2014 Sep-16 60-7075-106 3241388 9/16/2014 Sep-16 60-7075-106 3245795 9/16/2014 Sep-16 5921-018-135 3251481 9/19/2014 Sep-16 5921-018-235 3250392 9/19/2014 Sep-16 5921-024-135 3250080 9/19/2014 Sep-16 5921-024-136 3250165 9/19/2014 Sep-16 5921-024-235 3248542 9/19/2014 Sep-16 5921-030-135 3246924 9/19/2014 Sep-16 5921-030-235 3248189 9/19/2014 Sep-16 5921-030-235 3249359 9/19/2014 Sep-16 5921-030-235 3250482 9/19/2014 Sep-16 5921-030-235 3250836 9/19/2014 Sep-16 5921-034-135 3246144 9/19/2014 Sep-16 5921-034-135 3246982 9/19/2014 Sep-16 5921-034-135 3249392 9/19/2014 Sep-16 5921-034-135 3251391 9/19/2014 Sep-16 5921-034-235 3246694 9/19/2014 Sep-16 5921-034-235 3248664 9/19/2014 Sep-16 5921-034-235 3249257 9/19/2014 Sep-16 5921-034-235 3250274 9/19/2014 Sep-16 5921-034-235 3250815 9/19/2014 Sep-16 5921-034-236 3248705 9/19/2014 Sep-16 5921-218-235 3247325 9/19/2014 Sep-16 5921-218-235 3247620 9/19/2014 Sep-16 5921-218-235 3248779 9/19/2014 Sep-16 5921-218-235 3250344 9/19/2014 Sep-16 5921-218-135 3246365 9/19/2014 Sep-16 5921-218-135 3250177 9/19/2014 Sep-16 60-7070-102 3248018 9/19/2014 Sep-16 60-7070-102 3249940 9/19/2014 Sep-16 60-7070-103 3246349 9/19/2014 Sep-16 60-7070-103 3247575 9/19/2014 Sep-16 60-7070-103 3248629 9/19/2014 Sep-16 60-7070-103 3249300 9/19/2014 Sep-16 60-7070-103 3250623 9/19/2014 Sep-16 60-7070-103 3250964 9/19/2014 Sep-16 60-7070-105 3246752 9/19/2014 Sep-16 60-7070-105 3246976 9/19/2014 Sep-16 60-7070-105 3248278 9/19/2014 Sep-16 60-7070-105 3249469 9/19/2014 Sep-16 60-7070-105 3250725 9/19/2014 Sep-16 60-7070-105 3251294 9/19/2014 Sep-16 60-7070-104 3247903 9/19/2014 Sep-16 60-7070-104 3249815 9/19/2014 Sep-16 60-7070-104 3251389 9/19/2014 Sep-16 60-7070-003 3246948 9/19/2014 Sep-16 60-7070-004 3250131 9/19/2014 Sep-16 60-7070-006 3249177 9/19/2014 Sep-16 60-7070-106 3246298 9/19/2014 Sep-16 60-7070-106 3246464 9/19/2014 Sep-16 60-7070-106 3247893 9/19/2014 Sep-16 60-7070-106 3248115 9/19/2014 Sep-16 60-7070-106 3248693 9/19/2014 Sep-16 60-7070-106 3249316 9/19/2014 Sep-16 60-7070-106 3250862 9/19/2014 Sep-16 60-7070-107 3246116 9/19/2014 Sep-16 60-7070-107 3248877 9/19/2014 Sep-16 60-7075-003 3248975 9/19/2014 Sep-16 60-7075-003 3249226 9/19/2014 Sep-16 60-7075-006 3249005 9/19/2014 Sep-16 60-7075-103 3245977 9/19/2014 Sep-16 60-7075-103 3249014 9/19/2014 Sep-16 60-7075-103 3251468 9/19/2014 Sep-16 60-7075-104 3247133 9/19/2014 Sep-16 60-7075-105 3246354 9/19/2014 Sep-16 60-7075-105 3249120 9/19/2014 Sep-16 60-7075-106 3246061 9/19/2014 Sep-16 60-7075-106 3248162 9/19/2014 Sep-16 60-7075-106 3251554 9/19/2014 Sep-16 5921-018-135 3253510 9/23/2014 Sep-16 5921-024-135 3256418 9/23/2014 Sep-16 5921-024-235 3252485 9/23/2014 Sep-16 5921-030-135 3252459 9/23/2014 Sep-16 5921-030-135 3254980 9/23/2014 Sep-16 5921-030-135 3254998 9/23/2014 Sep-16 5921-030-235 3251842 9/23/2014 Sep-16 5921-030-235 3252307 9/23/2014 Sep-16 5921-030-235 3253789 9/23/2014 Sep-16 5921-034-135 3253903 9/23/2014 Sep-16 5921-034-135 3254001 9/23/2014 Sep-16 5921-034-135 3256544 9/23/2014 Sep-16 5921-034-235 3252209 9/23/2014 Sep-16 5921-034-235 3253180 9/23/2014 Sep-16 5921-034-235 3253801 9/23/2014 Sep-16 5921-034-235 3255564 9/23/2014 Sep-16 5921-218-235 3252767 9/23/2014 Sep-16 5921-218-235 3254830 9/23/2014 Sep-16 5921-218-135 3253523 9/23/2014 Sep-16 5921-218-135 3255538 9/23/2014 Sep-16 5921-218-135 3256366 9/23/2014 Sep-16 60-7070-102 3254040 9/23/2014 Sep-16 60-7070-103 3251646 9/23/2014 Sep-16 60-7070-103 3252444 9/23/2014 Sep-16 60-7070-103 3253791 9/23/2014 Sep-16 60-7070-103 3253829 9/23/2014 Sep-16 60-7070-103 3254539 9/23/2014 Sep-16 60-7070-103 3254671 9/23/2014 Sep-16 60-7070-103 3256027 9/23/2014 Sep-16 60-7070-103 3256661 9/23/2014 Sep-16 60-7070-105 3251667 9/23/2014 Sep-16 60-7070-105 3253216 9/23/2014 Sep-16 60-7070-105 3254541 9/23/2014 Sep-16 60-7070-105 3256350 9/23/2014 Sep-16 60-7070-105 3256748 9/23/2014 Sep-16 60-7070-104 3253237 9/23/2014 Sep-16 60-7070-104 3255689 9/23/2014 Sep-16 60-7070-003 3256528 9/23/2014 Sep-16 60-7070-004 3254064 9/23/2014 Sep-16 60-7070-106 3251900 9/23/2014 Sep-16 60-7070-106 3253394 9/23/2014 Sep-16 60-7070-106 3254157 9/23/2014 Sep-16 60-7070-106 3254436 9/23/2014 Sep-16 60-7070-106 3255658 9/23/2014 Sep-16 60-7070-106 3255997 9/23/2014 Sep-16 60-7070-107 3252170 9/23/2014 Sep-16 60-7070-107 3256769 9/23/2014 Sep-16 60-7075-006 3251729 9/23/2014 Sep-16 60-7075-006 3252492 9/23/2014 Sep-16 60-7075-104 3255141 9/23/2014 Sep-16 60-7075-105 3255248 9/23/2014 Sep-16 60-7075-106 3251755 9/23/2014 Sep-16 60-7075-106 3254979 9/23/2014 Sep-16 5921-018-135 3257498 9/24/2014 Sep-16 5921-018-135 3262370 9/24/2014 Sep-16 5921-018-136 3256867 9/24/2014 Sep-16 5921-018-235 3261002 9/24/2014 Sep-16 5921-024-135 3260616 9/24/2014 Sep-16 5921-024-235 3258163 9/24/2014 Sep-16 5921-030-135 3258144 9/24/2014 Sep-16 5921-030-135 3260951 9/24/2014 Sep-16 5921-030-135 3262130 9/24/2014 Sep-16 5921-030-235 3257927 9/24/2014 Sep-16 5921-030-235 3259166 9/24/2014 Sep-16 5921-030-235 3261883 9/24/2014 Sep-16 5921-034-135 3257390 9/24/2014 Sep-16 5921-034-135 3260157 9/24/2014 Sep-16 5921-034-235 3257319 9/24/2014 Sep-16 5921-034-235 3257666 9/24/2014 Sep-16 5921-034-235 3258137 9/24/2014 Sep-16 5921-034-235 3259994 9/24/2014 Sep-16 5921-034-235 3261693 9/24/2014 Sep-16 5921-218-235 3256833 9/24/2014 Sep-16 5921-218-235 3258057 9/24/2014 Sep-16 5921-218-235 3260347 9/24/2014 Sep-16 5921-212-135 3258159 9/24/2014 Sep-16 5921-218-135 3259068 9/24/2014 Sep-16 5921-218-135 3259648 9/24/2014 Sep-16 5921-218-135 3259695 9/24/2014 Sep-16 5921-218-135 3262202 9/24/2014 Sep-16 60-7070-102 3257405 9/24/2014 Sep-16 60-7070-103 3253758 9/24/2014 Sep-16 60-7070-103 3257374 9/24/2014 Sep-16 60-7070-103 3258734 9/24/2014 Sep-16 60-7070-103 3259580 9/24/2014 Sep-16 60-7070-103 3261353 9/24/2014 Sep-16 60-7070-103 3262304 9/24/2014 Sep-16 60-7070-105 3258019 9/24/2014 Sep-16 60-7070-105 3259034 9/24/2014 Sep-16 60-7070-105 3259386 9/24/2014 Sep-16 60-7070-105 3260830 9/24/2014 Sep-16 60-7070-105 3262260 9/24/2014 Sep-16 60-7070-104 3257970 9/24/2014 Sep-16 60-7070-104 3260980 9/24/2014 Sep-16 60-7070-104 3262287 9/24/2014 Sep-16 60-7070-005 3258536 9/24/2014 Sep-16 60-7070-005 3260436 9/24/2014 Sep-16 60-7070-006 3257763 9/24/2014 Sep-16 60-7070-106 3257240 9/24/2014 Sep-16 60-7070-106 3258026 9/24/2014 Sep-16 60-7070-106 3259121 9/24/2014 Sep-16 60-7070-106 3261444 9/24/2014 Sep-16 60-7070-107 3256795 9/24/2014 Sep-16 60-7070-107 3259598 9/24/2014 Sep-16 60-7075-003 3256851 9/24/2014 Sep-16 60-7075-006 3259132 9/24/2014 Sep-16 60-7075-006 3259289 9/24/2014 Sep-16 60-7075-103 3259329 9/24/2014 Sep-16 60-7075-103 3259761 9/24/2014 Sep-16 60-7075-103 3261910 9/24/2014 Sep-16 60-7075-104 3261542 9/24/2014 Sep-16 60-7075-105 3262554 9/24/2014 Sep-16 60-7075-106 3257958 9/24/2014 Sep-16 60-7075-106 3259301 9/24/2014 Sep-16 60-7075-106 3261221 9/24/2014 Sep-16 60-7075-106 3262124 9/24/2014 Sep-16 5921-018-135 3265505 9/26/2014 Sep-16 5921-018-135 3267061 9/26/2014 Sep-16 5921-024-235 3262633 9/26/2014 Sep-16 5921-024-235 3265338 9/26/2014 Sep-16 5921-024-235 3267687 9/26/2014 Sep-16 5921-030-135 3266620 9/26/2014 Sep-16 5921-030-235 3263908 9/26/2014 Sep-16 5921-030-235 3265068 9/26/2014 Sep-16 5921-030-235 3266198 9/26/2014 Sep-16 5921-030-235 3267429 9/26/2014 Sep-16 5921-034-135 3263500 9/26/2014 Sep-16 5921-034-135 3264917 9/26/2014 Sep-16 5921-034-135 3265454 9/26/2014 Sep-16 5921-034-135 3266840 9/26/2014 Sep-16 5921-034-235 3261877 9/26/2014 Sep-16 5921-034-235 3263193 9/26/2014 Sep-16 5921-034-235 3263313 9/26/2014 Sep-16 5921-034-235 3263912 9/26/2014 Sep-16 5921-034-235 3265150 9/26/2014 Sep-16 5921-034-235 3266048 9/26/2014 Sep-16 5921-034-235 3267401 9/26/2014 Sep-16 5921-034-235 3267710 9/26/2014 Sep-16 5921-034-236 3266180 9/26/2014 Sep-16 5921-212-235 3262565 9/26/2014 Sep-16 5921-218-235 3263299 9/26/2014 Sep-16 5921-218-235 3264226 9/26/2014 Sep-16 5921-218-235 3265386 9/26/2014 Sep-16 5921-218-235 3266534 9/26/2014 Sep-16 5921-218-235 3267438 9/26/2014 Sep-16 5921-218-235 3267907 9/26/2014 Sep-16 5921-212-135 3263571 9/26/2014 Sep-16 5921-212-135 3263795 9/26/2014 Sep-16 5921-218-135 3265243 9/26/2014 Sep-16 5921-218-135 3266394 9/26/2014 Sep-16 60-7070-103 3262861 9/26/2014 Sep-16 60-7070-103 3264165 9/26/2014 Sep-16 60-7070-103 3264482 9/26/2014 Sep-16 60-7070-103 3265034 9/26/2014 Sep-16 60-7070-103 3266114 9/26/2014 Sep-16 60-7070-103 3267144 9/26/2014 Sep-16 60-7070-103 3267472 9/26/2014 Sep-16 60-7070-103 3267976 9/26/2014 Sep-16 60-7070-105 3262786 9/26/2014 Sep-16 60-7070-105 3263563 9/26/2014 Sep-16 60-7070-105 3264411 9/26/2014 Sep-16 60-7070-105 3265121 9/26/2014 Sep-16 60-7070-105 3265227 9/26/2014 Sep-16 60-7070-105 3265567 9/26/2014 Sep-16 60-7070-105 3266011 9/26/2014 Sep-16 60-7070-105 3267090 9/26/2014 Sep-16 60-7070-105 3267673 9/26/2014 Sep-16 60-7070-104 3262959 9/26/2014 Sep-16 60-7070-104 3264699 9/26/2014 Sep-16 60-7070-104 3267088 9/26/2014 Sep-16 60-7070-003 3263537 9/26/2014 Sep-16 60-7070-005 3263769 9/26/2014 Sep-16 60-7070-006 3266484 9/26/2014 Sep-16 60-7070-106 3262963 9/26/2014 Sep-16 60-7070-106 3264242 9/26/2014 Sep-16 60-7070-106 3264634 9/26/2014 Sep-16 60-7070-106 3264891 9/26/2014 Sep-16 60-7070-106 3265477 9/26/2014 Sep-16 60-7070-106 3266961 9/26/2014 Sep-16 60-7070-106 3267781 9/26/2014 Sep-16 60-7070-107 3264316 9/26/2014 Sep-16 60-7070-107 3264469 9/26/2014 Sep-16 60-7070-107 3266736 9/26/2014 Sep-16 60-7075-003 3262549 9/26/2014 Sep-16 60-7075-006 3262740 9/26/2014 Sep-16 60-7075-006 3266855 9/26/2014 Sep-16 60-7075-103 3265364 9/26/2014 Sep-16 60-7075-103 3266341 9/26/2014 Sep-16 60-7075-104 3266450 9/26/2014 Sep-16 60-7075-104 3267747 9/26/2014 Sep-16 60-7075-105 3264371 9/26/2014 Sep-16 60-7075-105 3266409 9/26/2014 Sep-16 60-7075-106 3263657 9/26/2014 Sep-16 60-7075-106 3265420 9/26/2014 Sep-16 60-7075-106 3267995 9/26/2014 Sep-16 5921-018-135 3273307 9/30/2014 Sep-16 5921-024-135 3268181 9/30/2014 Sep-16 5921-024-235 3270172 9/30/2014 Sep-16 5921-030-135 3269000 9/30/2014 Sep-16 5921-030-135 3272341 9/30/2014 Sep-16 5921-030-235 3268110 9/30/2014 Sep-16 5921-030-235 3268595 9/30/2014 Sep-16 5921-030-235 3270129 9/30/2014 Sep-16 5921-030-235 3271305 9/30/2014 Sep-16 5921-030-235 3272114 9/30/2014 Sep-16 5921-034-135 3268031 9/30/2014 Sep-16 5921-034-135 3269555 9/30/2014 Sep-16 5921-034-135 3271314 9/30/2014 Sep-16 5921-034-135 3273060 9/30/2014 Sep-16 5921-034-235 3268728 9/30/2014 Sep-16 5921-034-235 3270384 9/30/2014 Sep-16 5921-034-235 3271102 9/30/2014 Sep-16 5921-034-235 3272167 9/30/2014 Sep-16 5921-034-235 3272339 9/30/2014 Sep-16 5921-218-235 3268825 9/30/2014 Sep-16 5921-218-235 3270603 9/30/2014 Sep-16 5921-218-235 3272318 9/30/2014 Sep-16 5921-218-235 3273104 9/30/2014 Sep-16 5921-218-135 3268229 9/30/2014 Sep-16 5921-218-135 3269403 9/30/2014 Sep-16 5921-218-135 3271555 9/30/2014 Sep-16 60-7070-103 3268818 9/30/2014 Sep-16 60-7070-103 3269615 9/30/2014 Sep-16 60-7070-103 3269777 9/30/2014 Sep-16 60-7070-103 3270193 9/30/2014 Sep-16 60-7070-103 3271540 9/30/2014 Sep-16 60-7070-103 3272285 9/30/2014 Sep-16 60-7070-103 3273094 9/30/2014 Sep-16 60-7070-105 3260933 9/30/2014 Sep-16 60-7070-105 3268238 9/30/2014 Sep-16 60-7070-105 3268700 9/30/2014 Sep-16 60-7070-105 3270101 9/30/2014 Sep-16 60-7070-105 3270657 9/30/2014 Sep-16 60-7070-105 3272007 9/30/2014 Sep-16 60-7070-105 3272758 9/30/2014 Sep-16 60-7070-105 3273311 9/30/2014 Sep-16 60-7070-104 3267930 9/30/2014 Sep-16 60-7070-104 3269586 9/30/2014 Sep-16 60-7070-104 3270309 9/30/2014 Sep-16 60-7070-104 3272146 9/30/2014 Sep-16 60-7070-104 3273348 9/30/2014 Sep-16 60-7070-003 3269564 9/30/2014 Sep-16 60-7070-005 3269118 9/30/2014 Sep-16 60-7070-006 3272612 9/30/2014 Sep-16 60-7070-106 3268319 9/30/2014 Sep-16 60-7070-106 3269593 9/30/2014 Sep-16 60-7070-106 3270487 9/30/2014 Sep-16 60-7070-106 3270635 9/30/2014 Sep-16 60-7070-106 3271852 9/30/2014 Sep-16 60-7070-106 3272048 9/30/2014 Sep-16 60-7070-106 3272743 9/30/2014 Sep-16 60-7070-107 3268856 9/30/2014 Sep-16 60-7070-107 3271604 9/30/2014 Sep-16 60-7070-107 3272829 9/30/2014 Sep-16 60-7075-003 3270460 9/30/2014 Sep-16 60-7075-006 3268873 9/30/2014 Sep-16 60-7075-006 3270585 9/30/2014 Sep-16 60-7075-103 3269351 9/30/2014 Sep-16 60-7075-103 3272547 9/30/2014 Sep-16 60-7075-105 3272151 9/30/2014 Sep-16 60-7075-106 3268175 9/30/2014 Sep-16 60-7075-106 3269248 9/30/2014 Sep-16 60-7075-106 3271071 9/30/2014 Sep-16 60-7075-106 3271426 9/30/2014 Sep-16 60-7075-106 3272893 9/30/2014 Sep-16 5921-030-135 3273808 10/2/2014 Oct-16 5921-030-135 3277721 10/2/2014 Oct-16 5921-030-235 3273601 10/2/2014 Oct-16 5921-030-235 3274422 10/2/2014 Oct-16 5921-030-235 3275241 10/2/2014 Oct-16 5921-030-235 3275304 10/2/2014 Oct-16 5921-030-235 3276668 10/2/2014 Oct-16 5921-034-135 3274335 10/2/2014 Oct-16 5921-034-135 3275531 10/2/2014 Oct-16 5921-034-135 3276597 10/2/2014 Oct-16 5921-034-135 3278301 10/2/2014 Oct-16 5921-034-235 3273574 10/2/2014 Oct-16 5921-034-235 3274716 10/2/2014 Oct-16 5921-034-235 3275389 10/2/2014 Oct-16 5921-034-235 3276257 10/2/2014 Oct-16 5921-212-235 3277933 10/2/2014 Oct-16 5921-218-235 3273831 10/2/2014 Oct-16 5921-218-235 3275315 10/2/2014 Oct-16 5921-218-235 3277212 10/2/2014 Oct-16 5921-218-235 3277914 10/2/2014 Oct-16 5921-218-135 3273416 10/2/2014 Oct-16 5921-218-135 3275343 10/2/2014 Oct-16 5921-218-135 3277334 10/2/2014 Oct-16 5921-218-136 3277249 10/2/2014 Oct-16 60-7070-103 3274096 10/2/2014 Oct-16 60-7070-103 3274621 10/2/2014 Oct-16 60-7070-103 3275239 10/2/2014 Oct-16 60-7070-103 3275651 10/2/2014 Oct-16 60-7070-103 3276972 10/2/2014 Oct-16 60-7070-103 3278145 10/2/2014 Oct-16 60-7070-103 3278204 10/2/2014 Oct-16 60-7070-105 3274388 10/2/2014 Oct-16 60-7070-105 3275148 10/2/2014 Oct-16 60-7070-105 3276046 10/2/2014 Oct-16 60-7070-105 3277946 10/2/2014 Oct-16 60-7070-104 3274802 10/2/2014 Oct-16 60-7070-104 3276328 10/2/2014 Oct-16 60-7070-104 3278150 10/2/2014 Oct-16 60-7070-003 3272552 10/2/2014 Oct-16 60-7070-003 3275786 10/2/2014 Oct-16 60-7070-006 3276929 10/2/2014 Oct-16 60-7070-106 3273812 10/2/2014 Oct-16 60-7070-106 3274697 10/2/2014 Oct-16 60-7070-106 3276177 10/2/2014 Oct-16 60-7070-106 3276606 10/2/2014 Oct-16 60-7070-106 3277310 10/2/2014 Oct-16 60-7070-106 3277685 10/2/2014 Oct-16 60-7070-107 3274645 10/2/2014 Oct-16 60-7070-107 3277368 10/2/2014 Oct-16 60-7075-003 3274357 10/2/2014 Oct-16 60-7075-004 3277220 10/2/2014 Oct-16 60-7075-006 3274286 10/2/2014 Oct-16 60-7075-006 3275605 10/2/2014 Oct-16 60-7075-103 3274700 10/2/2014 Oct-16 60-7075-103 3276430 10/2/2014 Oct-16 60-7075-104 3273883 10/2/2014 Oct-16 60-7075-104 3275614 10/2/2014 Oct-16 60-7075-106 3274106 10/2/2014 Oct-16 60-7075-106 3275358 10/2/2014 Oct-16 60-7075-106 3278099 10/2/2014 Oct-16 5921-018-135 3279710 10/6/2014 Oct-16 5921-024-235 3278863 10/6/2014 Oct-16 5921-024-235 3283602 10/6/2014 Oct-16 5921-030-135 3279088 10/6/2014 Oct-16 5921-030-235 3278726 10/6/2014 Oct-16 5921-030-235 3281265 10/6/2014 Oct-16 5921-030-235 3283074 10/6/2014 Oct-16 5921-034-135 3279396 10/6/2014 Oct-16 5921-034-135 3281594 10/6/2014 Oct-16 5921-034-135 3283486 10/6/2014 Oct-16 5921-034-135 3283591 10/6/2014 Oct-16 5921-034-235 3278227 10/6/2014 Oct-16 5921-034-235 3279562 10/6/2014 Oct-16 5921-034-235 3280765 10/6/2014 Oct-16 5921-034-235 3280966 10/6/2014 Oct-16 5921-034-235 3280978 10/6/2014 Oct-16 5921-218-235 3279401 10/6/2014 Oct-16 5921-218-235 3281556 10/6/2014 Oct-16 5921-218-235 3282894 10/6/2014 Oct-16 5921-218-235 3283088 10/6/2014 Oct-16 5921-218-135 3279192 10/6/2014 Oct-16 5921-218-135 3282106 10/6/2014 Oct-16 5921-218-135 3283280 10/6/2014 Oct-16 60-7070-103 3279639 10/6/2014 Oct-16 60-7070-103 3280102 10/6/2014 Oct-16 60-7070-103 3280457 10/6/2014 Oct-16 60-7070-103 3281869 10/6/2014 Oct-16 60-7070-103 3282632 10/6/2014 Oct-16 60-7070-105 3279057 10/6/2014 Oct-16 60-7070-105 3280435 10/6/2014 Oct-16 60-7070-105 3281310 10/6/2014 Oct-16 60-7070-105 3283279 10/6/2014 Oct-16 60-7070-104 3279850 10/6/2014 Oct-16 60-7070-104 3282012 10/6/2014 Oct-16 60-7070-003 3279246 10/6/2014 Oct-16 60-7070-005 3280426 10/6/2014 Oct-16 60-7070-006 3281979 10/6/2014 Oct-16 60-7070-106 3278393 10/6/2014 Oct-16 60-7070-106 3278703 10/6/2014 Oct-16 60-7070-106 3280071 10/6/2014 Oct-16 60-7070-106 3281573 10/6/2014 Oct-16 60-7070-106 3282199 10/6/2014 Oct-16 60-7070-106 3282825 10/6/2014 Oct-16 60-7075-003 3279061 10/6/2014 Oct-16 60-7075-006 3282020 10/6/2014 Oct-16 60-7075-103 3278735 10/6/2014 Oct-16 60-7075-103 3280085 10/6/2014 Oct-16 60-7075-103 3281671 10/6/2014 Oct-16 60-7075-104 3283137 10/6/2014 Oct-16 60-7075-106 3279747 10/6/2014 Oct-16 60-7075-106 3281824 10/6/2014 Oct-16 60-7075-106 3283570 10/6/2014 Oct-16 60-7075-106 3283924 10/6/2014 Oct-16 5921-018-135 3284992 10/10/2014 Oct-16 5921-024-235 3286034 10/10/2014 Oct-16 5921-030-135 3284985 10/10/2014 Oct-16 5921-030-136 3287437 10/10/2014 Oct-16 5921-030-235 3284050 10/10/2014 Oct-16 5921-030-235 3284944 10/10/2014 Oct-16 5921-030-235 3286184 10/10/2014 Oct-16 5921-030-235 3287722 10/10/2014 Oct-16 5921-034-135 3284512 10/10/2014 Oct-16 5921-034-135 3286918 10/10/2014 Oct-16 5921-034-135 3287395 10/10/2014 Oct-16 5921-034-235 3283976 10/10/2014 Oct-16 5921-034-235 3284793 10/10/2014 Oct-16 5921-034-235 3285098 10/10/2014 Oct-16 5921-034-235 3285495 10/10/2014 Oct-16 5921-034-235 3286788 10/10/2014 Oct-16 5921-034-236 3285392 10/10/2014 Oct-16 5921-218-235 3284717 10/10/2014 Oct-16 5921-218-235 3285986 10/10/2014 Oct-16 5921-218-235 3287191 10/10/2014 Oct-16 60-7070-103 3284202 10/10/2014 Oct-16 60-7070-103 3285154 10/10/2014 Oct-16 60-7070-103 3285836 10/10/2014 Oct-16 60-7070-103 3287039 10/10/2014 Oct-16 60-7070-103 3287640 10/10/2014 Oct-16 60-7070-105 3284679 10/10/2014 Oct-16 60-7070-105 3285566 10/10/2014 Oct-16 60-7070-105 3285940 10/10/2014 Oct-16 60-7070-105 3286483 10/10/2014 Oct-16 60-7070-105 3287416 10/10/2014 Oct-16 60-7070-104 3284066 10/10/2014 Oct-16 60-7070-104 3286178 10/10/2014 Oct-16 60-7070-104 3287189 10/10/2014 Oct-16 60-7070-003 3285482 10/10/2014 Oct-16 60-7070-005 3285424 10/10/2014 Oct-16 60-7070-106 3284239 10/10/2014 Oct-16 60-7070-106 3285756 10/10/2014 Oct-16 60-7070-106 3286598 10/10/2014 Oct-16 60-7070-106 3287672 10/10/2014 Oct-16 60-7070-106 3288228 10/10/2014 Oct-16 60-7075-003 3284287 10/10/2014 Oct-16 60-7075-004 3285430 10/10/2014 Oct-16 60-7075-006 3284273 10/10/2014 Oct-16 60-7075-103 3284260 10/10/2014 Oct-16 60-7075-103 3284701 10/10/2014 Oct-16 60-7075-103 3287232 10/10/2014 Oct-16 60-7075-106 3286828 10/10/2014 Oct-16 5921-018-135 3291216 10/13/2014 Oct-16 5921-018-135 3291302 10/13/2014 Oct-16 5921-024-135 3288285 10/13/2014 Oct-16 5921-024-235 3289500 10/13/2014 Oct-16 5921-030-135 3292913 10/13/2014 Oct-16 5921-030-235 3288302 10/13/2014 Oct-16 5921-030-235 3288864 10/13/2014 Oct-16 5921-030-235 3290595 10/13/2014 Oct-16 5921-030-235 3292586 10/13/2014 Oct-16 5921-034-135 3290147 10/13/2014 Oct-16 5921-034-135 3291466 10/13/2014 Oct-16 5921-034-135 3292627 10/13/2014 Oct-16 5921-034-235 3289418 10/13/2014 Oct-16 5921-034-235 3290293 10/13/2014 Oct-16 5921-034-235 3291053 10/13/2014 Oct-16 5921-034-235 3292367 10/13/2014 Oct-16 5921-218-235 3288736 10/13/2014 Oct-16 5921-218-235 3290054 10/13/2014 Oct-16 5921-218-235 3291318 10/13/2014 Oct-16 5921-044-135 3291011 10/13/2014 Oct-16 5921-218-135 3292403 10/13/2014 Oct-16 5921-218-135 3292555 10/13/2014 Oct-16 60-7070-103 3288373 10/13/2014 Oct-16 60-7070-103 3289846 10/13/2014 Oct-16 60-7070-103 3290666 10/13/2014 Oct-16 60-7070-103 3291478 10/13/2014 Oct-16 60-7070-103 3293014 10/13/2014 Oct-16 60-7070-105 3288360 10/13/2014 Oct-16 60-7070-105 3289516 10/13/2014 Oct-16 60-7070-105 3290632 10/13/2014 Oct-16 60-7070-105 3291394 10/13/2014 Oct-16 60-7070-105 3292593 10/13/2014 Oct-16 60-7070-104 3245635 10/13/2014 Oct-16 60-7070-104 3289350 10/13/2014 Oct-16 60-7070-104 3290613 10/13/2014 Oct-16 60-7070-104 3291631 10/13/2014 Oct-16 60-7070-003 3292615 10/13/2014 Oct-16 60-7070-006 3291620 10/13/2014 Oct-16 60-7070-006 3292351 10/13/2014 Oct-16 60-7070-106 3289707 10/13/2014 Oct-16 60-7070-106 3290650 10/13/2014 Oct-16 60-7070-106 3291665 10/13/2014 Oct-16 60-7070-106 3292248 10/13/2014 Oct-16 60-7075-003 3289015 10/13/2014 Oct-16 60-7075-006 3291237 10/13/2014 Oct-16 60-7080-102 3290991 10/13/2014 Oct-16 60-7075-103 3290272 10/13/2014 Oct-16 60-7075-103 3292379 10/13/2014 Oct-16 60-7075-104 3290303 10/13/2014 Oct-16 60-7075-105 3288244 10/13/2014 Oct-16 60-7075-106 3290621 10/13/2014 Oct-16 60-7075-106 3290856 10/13/2014 Oct-16 60-7075-106 3292909 10/13/2014 Oct-16 5921-018-235 3293033 10/15/2014 Oct-16 5921-024-235 3293290 10/15/2014 Oct-16 5921-024-235 3296714 10/15/2014 Oct-16 5921-030-135 3293889 10/15/2014 Oct-16 5921-030-235 3294384 10/15/2014 Oct-16 5921-030-235 3297023 10/15/2014 Oct-16 5921-034-135 3295365 10/15/2014 Oct-16 5921-034-235 3294058 10/15/2014 Oct-16 5921-034-235 3294281 10/15/2014 Oct-16 5921-034-235 3295100 10/15/2014 Oct-16 5921-034-235 3297010 10/15/2014 Oct-16 5921-034-235 3297197 10/15/2014 Oct-16 5921-218-235 3294439 10/15/2014 Oct-16 5921-218-235 3295924 10/15/2014 Oct-16 5921-218-235 3297844 10/15/2014 Oct-16 5921-218-135 3293794 10/15/2014 Oct-16 5921-218-135 3296025 10/15/2014 Oct-16 60-7070-103 3293219 10/15/2014 Oct-16 60-7070-103 3293940 10/15/2014 Oct-16 60-7070-103 3294070 10/15/2014 Oct-16 60-7070-103 3295706 10/15/2014 Oct-16 60-7070-103 3296561 10/15/2014 Oct-16 60-7070-103 3298119 10/15/2014 Oct-16 60-7070-105 3293815 10/15/2014 Oct-16 60-7070-105 3294753 10/15/2014 Oct-16 60-7070-105 3296463 10/15/2014 Oct-16 60-7070-105 3297761 10/15/2014 Oct-16 60-7070-104 3295137 10/15/2014 Oct-16 60-7070-104 3297420 10/15/2014 Oct-16 60-7070-104 3297598 10/15/2014 Oct-16 60-7070-005 3293424 10/15/2014 Oct-16 60-7070-106 3293448 10/15/2014 Oct-16 60-7070-106 3293760 10/15/2014 Oct-16 60-7070-106 3294674 10/15/2014 Oct-16 60-7070-106 3294846 10/15/2014 Oct-16 60-7070-106 3296232 10/15/2014 Oct-16 60-7070-106 3296519 10/15/2014 Oct-16 60-7075-003 3295982 10/15/2014 Oct-16 60-7075-103 3297372 10/15/2014 Oct-16 60-7075-105 3293316 10/15/2014 Oct-16 60-7075-106 3295377 10/15/2014 Oct-16 5921-018-135 3300807 10/17/2014 Oct-16 5921-018-135 3302992 10/17/2014 Oct-16 5921-030-135 3298895 10/17/2014 Oct-16 5921-030-135 3299148 10/17/2014 Oct-16 5921-030-135 3301672 10/17/2014 Oct-16 5921-030-235 3299124 10/17/2014 Oct-16 5921-030-235 3299410 10/17/2014 Oct-16 5921-030-235 3300296 10/17/2014 Oct-16 5921-030-235 3300600 10/17/2014 Oct-16 5921-034-135 3298573 10/17/2014 Oct-16 5921-034-135 3300470 10/17/2014 Oct-16 5921-034-135 3302239 10/17/2014 Oct-16 5921-034-235 3298436 10/17/2014 Oct-16 5921-034-235 3299680 10/17/2014 Oct-16 5921-034-235 3301561 10/17/2014 Oct-16 5921-034-235 3302358 10/17/2014 Oct-16 5921-034-235 3302508 10/17/2014 Oct-16 5921-218-235 3299447 10/17/2014 Oct-16 5921-218-235 3301463 10/17/2014 Oct-16 5921-218-235 3302803 10/17/2014 Oct-16 60-7070-103 3298296 10/17/2014 Oct-16 60-7070-103 3299176 10/17/2014 Oct-16 60-7070-103 3299512 10/17/2014 Oct-16 60-7070-103 3300811 10/17/2014 Oct-16 60-7070-103 3301812 10/17/2014 Oct-16 60-7070-103 3302679 10/17/2014 Oct-16 60-7070-103 3302874 10/17/2014 Oct-16 60-7070-105 3302468 10/17/2014 Oct-16 60-7070-104 3299819 10/17/2014 Oct-16 60-7070-106 3298556 10/17/2014 Oct-16 60-7070-106 3298560 10/17/2014 Oct-16 60-7070-106 3299971 10/17/2014 Oct-16 60-7070-106 3300017 10/17/2014 Oct-16 60-7070-106 3300277 10/17/2014 Oct-16 60-7070-106 3301705 10/17/2014 Oct-16 60-7070-106 3302169 10/17/2014 Oct-16 60-7070-106 3303155 10/17/2014 Oct-16 60-7075-003 3300427 10/17/2014 Oct-16 60-7075-006 3299202 10/17/2014 Oct-16 60-7075-006 3301601 10/17/2014 Oct-16 60-7075-103 3302294 10/17/2014 Oct-16 60-7075-104 3299406 10/17/2014 Oct-16 60-7075-105 3298381 10/17/2014 Oct-16 60-7075-105 3302218 10/17/2014 Oct-16 60-7075-106 3298905 10/17/2014 Oct-16 60-7075-106 3299169 10/17/2014 Oct-16 60-7075-106 3299195 10/17/2014 Oct-16 5921-018-235 3305006 10/20/2014 Oct-16 5921-024-235 3304513 10/20/2014 Oct-16 5921-024-235 3307768 10/20/2014 Oct-16 5921-030-135 3304257 10/20/2014 Oct-16 5921-030-135 3306762 10/20/2014 Oct-16 5921-030-235 3303441 10/20/2014 Oct-16 5921-030-235 3304836 10/20/2014 Oct-16 5921-030-235 3306007 10/20/2014 Oct-16 5921-030-235 3308238 10/20/2014 Oct-16 5921-034-135 3304131 10/20/2014 Oct-16 5921-034-135 3306167 10/20/2014 Oct-16 5921-034-135 3306581 10/20/2014 Oct-16 5921-034-235 3303345 10/20/2014 Oct-16 5921-034-235 3304972 10/20/2014 Oct-16 5921-034-235 3305653 10/20/2014 Oct-16 5921-034-235 3307589 10/20/2014 Oct-16 5921-034-235 3307752 10/20/2014 Oct-16 5921-034-236 3307003 10/20/2014 Oct-16 5921-218-235 3304986 10/20/2014 Oct-16 5921-218-235 3305372 10/20/2014 Oct-16 5921-218-235 3305420 10/20/2014 Oct-16 5921-218-235 3308684 10/20/2014 Oct-16 5921-218-135 3304363 10/20/2014 Oct-16 5921-218-135 3306791 10/20/2014 Oct-16 60-7070-103 3304033 10/20/2014 Oct-16 60-7070-103 3304219 10/20/2014 Oct-16 60-7070-103 3305412 10/20/2014 Oct-16 60-7070-103 3305757 10/20/2014 Oct-16 60-7070-103 3306961 10/20/2014 Oct-16 60-7070-103 3308065 10/20/2014 Oct-16 60-7070-105 3303674 10/20/2014 Oct-16 60-7070-105 3306151 10/20/2014 Oct-16 60-7070-105 3307625 10/20/2014 Oct-16 60-7070-105 3307995 10/20/2014 Oct-16 60-7070-104 3306523 10/20/2014 Oct-16 60-7070-104 3307602 10/20/2014 Oct-16 60-7070-006 3307144 10/20/2014 Oct-16 60-7070-007 3304300 10/20/2014 Oct-16 60-7070-106 3301820 10/20/2014 Oct-16 60-7070-106 3303206 10/20/2014 Oct-16 60-7070-106 3304955 10/20/2014 Oct-16 60-7070-106 3305010 10/20/2014 Oct-16 60-7070-106 3305863 10/20/2014 Oct-16 60-7070-106 3305973 10/20/2014 Oct-16 60-7070-106 3307570 10/20/2014 Oct-16 60-7070-106 3308595 10/20/2014 Oct-16 60-7070-107 3307088 10/20/2014 Oct-16 60-7075-003 3305676 10/20/2014 Oct-16 60-7075-006 3305742 10/20/2014 Oct-16 60-7075-103 3295349 10/20/2014 Oct-16 60-7075-103 3304261 10/20/2014 Oct-16 60-7075-103 3306770 10/20/2014 Oct-16 60-7075-104 3304993 10/20/2014 Oct-16 60-7075-106 3306974 10/20/2014 Oct-16 60-7075-106 3308054 10/20/2014 Oct-16 60-7075-106 3308175 10/20/2014 Oct-16 5921-018-135 3309627 10/22/2014 Oct-16 5921-018-235 3309529 10/22/2014 Oct-16 5921-024-235 3312780 10/22/2014 Oct-16 5921-030-135 3308991 10/22/2014 Oct-16 5921-030-135 3312189 10/22/2014 Oct-16 5921-030-136 3307061 10/22/2014 Oct-16 5921-030-235 3310108 10/22/2014 Oct-16 5921-030-235 3311985 10/22/2014 Oct-16 5921-034-135 3309253 10/22/2014 Oct-16 5921-034-135 3311508 10/22/2014 Oct-16 5921-034-135 3311679 10/22/2014 Oct-16 5921-034-235 3309379 10/22/2014 Oct-16 5921-034-235 3310843 10/22/2014 Oct-16 5921-034-235 3312999 10/22/2014 Oct-16 5921-218-235 3310578 10/22/2014 Oct-16 5921-030-236 3309187 10/22/2014 Oct-16 5921-218-135 3309636 10/22/2014 Oct-16 5921-218-135 3312018 10/22/2014 Oct-16 60-7070-103 3309333 10/22/2014 Oct-16 60-7070-103 3310080 10/22/2014 Oct-16 60-7070-103 3310993 10/22/2014 Oct-16 60-7070-103 3312307 10/22/2014 Oct-16 60-7070-105 3309016 10/22/2014 Oct-16 60-7070-105 3310067 10/22/2014 Oct-16 60-7070-105 3311481 10/22/2014 Oct-16 60-7070-104 3309501 10/22/2014 Oct-16 60-7070-104 3312733 10/22/2014 Oct-16 60-7070-003 3311013 10/22/2014 Oct-16 60-7070-003 3312714 10/22/2014 Oct-16 60-7070-005 3312722 10/22/2014 Oct-16 60-7070-006 3312025 10/22/2014 Oct-16 60-7070-106 3309000 10/22/2014 Oct-16 60-7070-106 3309932 10/22/2014 Oct-16 60-7070-106 3310149 10/22/2014 Oct-16 60-7070-106 3310226 10/22/2014 Oct-16 60-7070-106 3310718 10/22/2014 Oct-16 60-7070-106 3311148 10/22/2014 Oct-16 60-7070-106 3313053 10/22/2014 Oct-16 60-7075-003 3311971 10/22/2014 Oct-16 60-7075-102 3312854 10/22/2014 Oct-16 60-7075-103 3309139 10/22/2014 Oct-16 60-7075-103 3311512 10/22/2014 Oct-16 60-7075-103 3313048 10/22/2014 Oct-16 60-7075-104 3311992 10/22/2014 Oct-16 60-7075-105 3310196 10/22/2014 Oct-16 60-7075-105 3312245 10/22/2014 Oct-16 60-7075-106 3311826 10/22/2014 Oct-16 60-7075-106 3311928 10/22/2014 Oct-16 5921-018-235 3314600 10/24/2014 Oct-16 5921-030-135 3315806 10/24/2014 Oct-16 5921-030-135 3318543 10/24/2014 Oct-16 5921-030-135 3318589 10/24/2014 Oct-16 5921-030-235 3313285 10/24/2014 Oct-16 5921-030-235 3313770 10/24/2014 Oct-16 5921-030-235 3314347 10/24/2014 Oct-16 5921-030-235 3316223 10/24/2014 Oct-16 5921-030-235 3318687 10/24/2014 Oct-16 5921-034-135 3313942 10/24/2014 Oct-16 5921-034-135 3315875 10/24/2014 Oct-16 5921-034-135 3317446 10/24/2014 Oct-16 5921-034-135 3318570 10/24/2014 Oct-16 5921-034-235 3314142 10/24/2014 Oct-16 5921-034-235 3315765 10/24/2014 Oct-16 5921-034-235 3316345 10/24/2014 Oct-16 5921-034-235 3318349 10/24/2014 Oct-16 5921-218-235 3313302 10/24/2014 Oct-16 5921-218-235 3315529 10/24/2014 Oct-16 5921-218-235 3317134 10/24/2014 Oct-16 5921-218-235 3318464 10/24/2014 Oct-16 5921-218-135 3314093 10/24/2014 Oct-16 5921-218-135 3315538 10/24/2014 Oct-16 5921-218-135 3317147 10/24/2014 Oct-16 60-7070-103 3313105 10/24/2014 Oct-16 60-7070-103 3314126 10/24/2014 Oct-16 60-7070-103 3315899 10/24/2014 Oct-16 60-7070-103 3316559 10/24/2014 Oct-16 60-7070-103 3317405 10/24/2014 Oct-16 60-7070-105 3314001 10/24/2014 Oct-16 60-7070-105 3315028 10/24/2014 Oct-16 60-7070-105 3315490 10/24/2014 Oct-16 60-7070-105 3317201 10/24/2014 Oct-16 60-7070-104 3316769 10/24/2014 Oct-16 60-7070-104 3317771 10/24/2014 Oct-16 60-7070-106 3314489 10/24/2014 Oct-16 60-7070-106 3316193 10/24/2014 Oct-16 60-7070-106 3316500 10/24/2014 Oct-16 60-7070-106 3318383 10/24/2014 Oct-16 60-7075-003 3313237 10/24/2014 Oct-16 60-7075-003 3318515 10/24/2014 Oct-16 60-7075-006 3313069 10/24/2014 Oct-16 60-7075-006 3318639 10/24/2014 Oct-16 60-7080-102 3314208 10/24/2014 Oct-16 60-7075-103 3315517 10/24/2014 Oct-16 60-7075-103 3317678 10/24/2014 Oct-16 60-7075-104 3316867 10/24/2014 Oct-16 60-7075-105 3314766 10/24/2014 Oct-16 60-7075-106 3315125 10/24/2014 Oct-16 60-7075-106 3317217 10/24/2014 Oct-16 60-7075-106 3317240 10/24/2014 Oct-16 5921-018-135 3317856 10/27/2014 Oct-16 5921-018-235 3320474 10/27/2014 Oct-16 5921-024-235 3319340 10/27/2014 Oct-16 5921-024-235 3320852 10/27/2014 Oct-16 5921-024-235 3321470 10/27/2014 Oct-16 5921-030-135 3320517 10/27/2014 Oct-16 5921-030-135 3323383 10/27/2014 Oct-16 5921-030-235 3318019 10/27/2014 Oct-16 5921-030-235 3320224 10/27/2014 Oct-16 5921-030-235 3320966 10/27/2014 Oct-16 5921-030-235 3321064 10/27/2014 Oct-16 5921-030-235 3321188 10/27/2014 Oct-16 5921-034-135 3320984 10/27/2014 Oct-16 5921-034-135 3322860 10/27/2014 Oct-16 5921-034-135 3323625 10/27/2014 Oct-16 5921-034-235 3317825 10/27/2014 Oct-16 5921-034-235 3319145 10/27/2014 Oct-16 5921-034-235 3320173 10/27/2014 Oct-16 5921-034-235 3320593 10/27/2014 Oct-16 5921-034-235 3321190 10/27/2014 Oct-16 5921-034-235 3321323 10/27/2014 Oct-16 5921-034-235 3323171 10/27/2014 Oct-16 5921-034-236 3318781 10/27/2014 Oct-16 5921-218-235 3312002 10/27/2014 Oct-16 5921-218-235 3318924 10/27/2014 Oct-16 5921-218-235 3319676 10/27/2014 Oct-16 5921-218-235 3321489 10/27/2014 Oct-16 5921-218-235 3322553 10/27/2014 Oct-16 5921-030-236 3321518 10/27/2014 Oct-16 5921-044-235 3320230 10/27/2014 Oct-16 5921-218-135 3318694 10/27/2014 Oct-16 5921-218-135 3323061 10/27/2014 Oct-16 60-7070-103 3318845 10/27/2014 Oct-16 60-7070-103 3319774 10/27/2014 Oct-16 60-7070-103 3319844 10/27/2014 Oct-16 60-7070-103 3320945 10/27/2014 Oct-16 60-7070-103 3321306 10/27/2014 Oct-16 60-7070-103 3322569 10/27/2014 Oct-16 60-7070-105 3319109 10/27/2014 Oct-16 60-7070-105 3319816 10/27/2014 Oct-16 60-7070-105 3321040 10/27/2014 Oct-16 60-7070-105 3321637 10/27/2014 Oct-16 60-7070-105 3321721 10/27/2014 Oct-16 60-7070-105 3322856 10/27/2014 Oct-16 60-7070-104 3317904 10/27/2014 Oct-16 60-7070-104 3318976 10/27/2014 Oct-16 60-7070-003 3319607 10/27/2014 Oct-16 60-7070-005 3319372 10/27/2014 Oct-16 60-7070-006 3319301 10/27/2014 Oct-16 60-7070-006 3321161 10/27/2014 Oct-16 60-7070-106 3317991 10/27/2014 Oct-16 60-7070-106 3319150 10/27/2014 Oct-16 60-7070-106 3320194 10/27/2014 Oct-16 60-7070-106 3320269 10/27/2014 Oct-16 60-7070-106 3321869 10/27/2014 Oct-16 60-7070-106 3322181 10/27/2014 Oct-16 60-7070-106 3322238 10/27/2014 Oct-16 60-7070-106 3323401 10/27/2014 Oct-16 60-7070-107 3319408 10/27/2014 Oct-16 60-7070-107 3323832 10/27/2014 Oct-16 60-7075-004 3323845 10/27/2014 Oct-16 60-7075-006 3319987 10/27/2014 Oct-16 60-7075-102 3319994 10/27/2014 Oct-16 60-7075-103 3320461 10/27/2014 Oct-16 60-7075-103 3321334 10/27/2014 Oct-16 60-7075-103 3323878 10/27/2014 Oct-16 60-7075-104 3321525 10/27/2014 Oct-16 60-7075-105 3319262 10/27/2014 Oct-16 60-7075-105 3320868 10/27/2014 Oct-16 60-7075-105 3323897 10/27/2014 Oct-16 60-7075-106 3318897 10/27/2014 Oct-16 60-7075-106 3320627 10/27/2014 Oct-16 60-7075-106 3321502 10/27/2014 Oct-16 60-7075-106 3324066 10/27/2014 Oct-16 5921-018-135 3325606 10/29/2014 Oct-16 5921-018-235 3327127 10/29/2014 Oct-16 5921-024-235 3327485 10/29/2014 Oct-16 5921-030-135 3327094 10/29/2014 Oct-16 5921-030-235 3324916 10/29/2014 Oct-16 5921-030-235 3326695 10/29/2014 Oct-16 5921-030-235 3326774 10/29/2014 Oct-16 5921-034-135 3325196 10/29/2014 Oct-16 5921-034-135 3325328 10/29/2014 Oct-16 5921-034-235 3324097 10/29/2014 Oct-16 5921-034-235 3325014 10/29/2014 Oct-16 5921-034-235 3325794 10/29/2014 Oct-16 5921-034-235 3326742 10/29/2014 Oct-16 5921-034-235 3327883 10/29/2014 Oct-16 5921-218-235 3324633 10/29/2014 Oct-16 5921-218-235 3325993 10/29/2014 Oct-16 5921-218-235 3326121 10/29/2014 Oct-16 5921-218-235 3326757 10/29/2014 Oct-16 5921-218-235 3328665 10/29/2014 Oct-16 5921-218-135 3324315 10/29/2014 Oct-16 5921-218-135 3326243 10/29/2014 Oct-16 60-7070-103 3323821 10/29/2014 Oct-16 60-7070-103 3325337 10/29/2014 Oct-16 60-7070-103 3326994 10/29/2014 Oct-16 60-7070-103 3327767 10/29/2014 Oct-16 60-7070-105 3324202 10/29/2014 Oct-16 60-7070-105 3325584 10/29/2014 Oct-16 60-7070-106 3324481 10/29/2014 Oct-16 60-7070-106 3326454 10/29/2014 Oct-16 60-7070-106 3327812 10/29/2014 Oct-16 60-7070-106 3327906 10/29/2014 Oct-16 60-7075-103 3324963 10/29/2014 Oct-16 60-7075-104 3326676 10/29/2014 Oct-16 60-7075-106 3324130 10/29/2014 Oct-16 60-7075-106 3326081 10/29/2014 Oct-16 5921-024-235 3329210 11/3/2014 Nov-16 5921-030-135 3329223 11/3/2014 Nov-16 5921-030-235 3330273 11/3/2014 Nov-16 5921-030-235 3331792 11/3/2014 Nov-16 5921-030-235 3331841 11/3/2014 Nov-16 5921-030-235 3332142 11/3/2014 Nov-16 5921-034-135 3329004 11/3/2014 Nov-16 5921-034-135 3329070 11/3/2014 Nov-16 5921-034-135 3329707 11/3/2014 Nov-16 5921-034-135 3331530 11/3/2014 Nov-16 5921-034-135 3331927 11/3/2014 Nov-16 5921-034-235 3329425 11/3/2014 Nov-16 5921-034-235 3331054 11/3/2014 Nov-16 5921-034-235 3332001 11/3/2014 Nov-16 5921-034-235 3333236 11/3/2014 Nov-16 5921-218-235 3330423 11/3/2014 Nov-16 5921-218-235 3330651 11/3/2014 Nov-16 5921-218-235 3331915 11/3/2014 Nov-16 5921-218-135 3328789 11/3/2014 Nov-16 5921-218-135 3330717 11/3/2014 Nov-16 60-7070-103 3329730 11/3/2014 Nov-16 60-7070-103 3330971 11/3/2014 Nov-16 60-7070-103 3331802 11/3/2014 Nov-16 60-7070-103 3332040 11/3/2014 Nov-16 60-7070-105 3328855 11/3/2014 Nov-16 60-7070-105 3329402 11/3/2014 Nov-16 60-7070-105 3330434 11/3/2014 Nov-16 60-7070-105 3331524 11/3/2014 Nov-16 60-7070-105 3331632 11/3/2014 Nov-16 60-7070-105 3333184 11/3/2014 Nov-16 60-7070-003 3329418 11/3/2014 Nov-16 60-7070-005 3330701 11/3/2014 Nov-16 60-7070-006 3332157 11/3/2014 Nov-16 60-7070-106 3329091 11/3/2014 Nov-16 60-7070-106 3329345 11/3/2014 Nov-16 60-7070-106 3330370 11/3/2014 Nov-16 60-7070-106 3331451 11/3/2014 Nov-16 60-7070-106 3332190 11/3/2014 Nov-16 60-7070-106 3332811 11/3/2014 Nov-16 60-7075-003 3330256 11/3/2014 Nov-16 60-7075-003 3331697 11/3/2014 Nov-16 60-7075-006 3329626 11/3/2014 Nov-16 60-7075-006 3329776 11/3/2014 Nov-16 60-7075-103 3332188 11/3/2014 Nov-16 60-7075-103 3332231 11/3/2014 Nov-16 60-7075-105 3331286 11/3/2014 Nov-16 60-7075-106 3328988 11/3/2014 Nov-16 60-7075-106 3331264 11/3/2014 Nov-16 60-7075-106 3331904 11/3/2014 Nov-16 5921-018-135 3334207 11/5/2014 Nov-16 5921-018-235 3333735 11/5/2014 Nov-16 5921-024-135 3334346 11/5/2014 Nov-16 5921-024-235 3335949 11/5/2014 Nov-16 5921-030-135 3336267 11/5/2014 Nov-16 5921-030-135 3336866 11/5/2014 Nov-16 5921-030-235 3334517 11/5/2014 Nov-16 5921-030-235 3338002 11/5/2014 Nov-16 5921-034-135 3334875 11/5/2014 Nov-16 5921-034-135 3335210 11/5/2014 Nov-16 5921-034-235 3335313 11/5/2014 Nov-16 5921-034-235 3336982 11/5/2014 Nov-16 5921-034-236 3334615 11/5/2014 Nov-16 5921-218-235 3333420 11/5/2014 Nov-16 5921-218-235 3335456 11/5/2014 Nov-16 5921-218-235 3337292 11/5/2014 Nov-16 5921-218-135 3333742 11/5/2014 Nov-16 5921-218-135 3336298 11/5/2014 Nov-16 60-7070-103 3333918 11/5/2014 Nov-16 60-7070-103 3334474 11/5/2014 Nov-16 60-7070-103 3336088 11/5/2014 Nov-16 60-7070-103 3336222 11/5/2014 Nov-16 60-7070-103 3337011 11/5/2014 Nov-16 60-7070-103 3337394 11/5/2014 Nov-16 60-7070-105 3336279 11/5/2014 Nov-16 60-7070-104 3333301 11/5/2014 Nov-16 60-7070-104 3334173 11/5/2014 Nov-16 60-7070-003 3333579 11/5/2014 Nov-16 60-7070-106 3333892 11/5/2014 Nov-16 60-7070-106 3334187 11/5/2014 Nov-16 60-7070-106 3335252 11/5/2014 Nov-16 60-7070-106 3335500 11/5/2014 Nov-16 60-7070-106 3337007 11/5/2014 Nov-16 60-7070-106 3337683 11/5/2014 Nov-16 60-7070-106 3337942 11/5/2014 Nov-16 60-7070-107 3338056 11/5/2014 Nov-16 60-7075-103 3335661 11/5/2014 Nov-16 60-7075-103 3335795 11/5/2014 Nov-16 60-7075-103 3336969 11/5/2014 Nov-16 60-7075-105 3336308 11/5/2014 Nov-16 60-7075-106 3335898 11/5/2014 Nov-16 5921-018-135 3341184 11/7/2014 Nov-16 5921-018-235 3339566 11/7/2014 Nov-16 5921-024-135 3338638 11/7/2014 Nov-16 5921-024-235 3338865 11/7/2014 Nov-16 5921-024-235 3342822 11/7/2014 Nov-16 5921-030-135 3341372 11/7/2014 Nov-16 5921-030-135 3342425 11/7/2014 Nov-16 5921-030-135 3343336 11/7/2014 Nov-16 5921-030-235 3339495 11/7/2014 Nov-16 5921-030-235 3339681 11/7/2014 Nov-16 5921-034-135 3338542 11/7/2014 Nov-16 5921-034-135 3340801 11/7/2014 Nov-16 5921-034-135 3341047 11/7/2014 Nov-16 5921-034-135 3342402 11/7/2014 Nov-16 5921-034-235 3338073 11/7/2014 Nov-16 5921-034-235 3339836 11/7/2014 Nov-16 5921-034-235 3341703 11/7/2014 Nov-16 5921-218-235 3340339 11/7/2014 Nov-16 5921-218-235 3340817 11/7/2014 Nov-16 5921-030-236 3339521 11/7/2014 Nov-16 5921-218-135 3339051 11/7/2014 Nov-16 5921-218-135 3341477 11/7/2014 Nov-16 5921-218-135 3342846 11/7/2014 Nov-16 60-7070-103 3338947 11/7/2014 Nov-16 60-7070-103 3339149 11/7/2014 Nov-16 60-7070-103 3339668 11/7/2014 Nov-16 60-7070-103 3340216 11/7/2014 Nov-16 60-7070-103 3341338 11/7/2014 Nov-16 60-7070-103 3342091 11/7/2014 Nov-16 60-7070-105 3339363 11/7/2014 Nov-16 60-7070-105 3339392 11/7/2014 Nov-16 60-7070-105 3341941 11/7/2014 Nov-16 60-7070-105 3342378 11/7/2014 Nov-16 60-7070-105 3342441 11/7/2014 Nov-16 60-7070-003 3342592 11/7/2014 Nov-16 60-7070-003 3342730 11/7/2014 Nov-16 60-7070-005 3341023 11/7/2014 Nov-16 60-7070-006 3339067 11/7/2014 Nov-16 60-7070-006 3342619 11/7/2014 Nov-16 60-7070-106 3339290 11/7/2014 Nov-16 60-7070-106 3340151 11/7/2014 Nov-16 60-7070-106 3341611 11/7/2014 Nov-16 60-7070-106 3341761 11/7/2014 Nov-16 60-7070-106 3342748 11/7/2014 Nov-16 60-7070-106 3342912 11/7/2014 Nov-16 60-7070-107 3338353 11/7/2014 Nov-16 60-7070-107 3342833 11/7/2014 Nov-16 60-7075-003 3339870 11/7/2014 Nov-16 60-7075-004 3339550 11/7/2014 Nov-16 60-7075-006 3339858 11/7/2014 Nov-16 60-7075-103 3341695 11/7/2014 Nov-16 60-7075-105 3340122 11/7/2014 Nov-16 60-7075-106 3338428 11/7/2014 Nov-16 60-7075-106 3340791 11/7/2014 Nov-16 60-7075-106 3343508 11/7/2014 Nov-16 5921-018-135 3347235 11/10/2014 Nov-16 5921-018-235 3347371 11/10/2014 Nov-16 5921-024-135 3347610 11/10/2014 Nov-16 5921-024-235 3347787 11/10/2014 Nov-16 5921-030-135 3345865 11/10/2014 Nov-16 5921-030-235 3344969 11/10/2014 Nov-16 5921-030-235 3345812 11/10/2014 Nov-16 5921-030-235 3347112 11/10/2014 Nov-16 5921-034-135 3344562 11/10/2014 Nov-16 5921-034-135 3346766 11/10/2014 Nov-16 5921-034-235 3343475 11/10/2014 Nov-16 5921-034-235 3345542 11/10/2014 Nov-16 5921-034-235 3346017 11/10/2014 Nov-16 5921-218-235 3344057 11/10/2014 Nov-16 5921-218-235 3345693 11/10/2014 Nov-16 5921-218-235 3347815 11/10/2014 Nov-16 5921-218-135 3346157 11/10/2014 Nov-16 5921-218-135 3346212 11/10/2014 Nov-16 60-7070-103 3344100 11/10/2014 Nov-16 60-7070-103 3345221 11/10/2014 Nov-16 60-7070-103 3345428 11/10/2014 Nov-16 60-7070-103 3345947 11/10/2014 Nov-16 60-7070-103 3347773 11/10/2014 Nov-16 60-7070-103 3348419 11/10/2014 Nov-16 60-7070-105 3344673 11/10/2014 Nov-16 60-7070-105 3344953 11/10/2014 Nov-16 60-7070-105 3345463 11/10/2014 Nov-16 60-7070-105 3345896 11/10/2014 Nov-16 60-7070-105 3347623 11/10/2014 Nov-16 60-7070-105 3347756 11/10/2014 Nov-16 60-7070-105 3348662 11/10/2014 Nov-16 60-7070-104 3346072 11/10/2014 Nov-16 60-7070-104 3347385 11/10/2014 Nov-16 60-7070-003 3346797 11/10/2014 Nov-16 60-7070-004 3346933 11/10/2014 Nov-16 60-7070-006 3346914 11/10/2014 Nov-16 60-7070-106 3345209 11/10/2014 Nov-16 60-7070-106 3346040 11/10/2014 Nov-16 60-7070-106 3348160 11/10/2014 Nov-16 60-7070-107 3347858 11/10/2014 Nov-16 60-7075-006 3345039 11/10/2014 Nov-16 60-7075-006 3345189 11/10/2014 Nov-16 60-7075-006 3347702 11/10/2014 Nov-16 60-7075-103 3345588 11/10/2014 Nov-16 60-7075-103 3346190 11/10/2014 Nov-16 60-7075-104 3347363 11/10/2014 Nov-16 60-7075-105 3346778 11/10/2014 Nov-16 60-7075-106 3345714 11/10/2014 Nov-16 60-7075-106 3345751 11/10/2014 Nov-16 60-7075-106 3348346 11/10/2014 Nov-16 60-7075-106 3348474 11/10/2014 Nov-16 60-7075-106 3348540 11/10/2014 Nov-16 5921-024-235 3352426 11/12/2014 Nov-16 5921-030-135 3350451 11/12/2014 Nov-16 5921-030-135 3353274 11/12/2014 Nov-16 5921-030-235 3349498 11/12/2014 Nov-16 5921-030-235 3351228 11/12/2014 Nov-16 5921-030-235 3353424 11/12/2014 Nov-16 5921-034-135 3349728 11/12/2014 Nov-16 5921-034-135 3352881 11/12/2014 Nov-16 5921-034-135 3354070 11/12/2014 Nov-16 5921-034-235 3348945 11/12/2014 Nov-16 5921-034-235 3350958 11/12/2014 Nov-16 5921-034-235 3352346 11/12/2014 Nov-16 5921-034-235 3353385 11/12/2014 Nov-16 5921-218-235 3348932 11/12/2014 Nov-16 5921-218-235 3351838 11/12/2014 Nov-16 5921-218-235 3353363 11/12/2014 Nov-16 5921-218-135 3352691 11/12/2014 Nov-16 60-7070-103 3350077 11/12/2014 Nov-16 60-7070-103 3351180 11/12/2014 Nov-16 60-7070-103 3353261 11/12/2014 Nov-16 60-7070-105 3348847 11/12/2014 Nov-16 60-7070-105 3349763 11/12/2014 Nov-16 60-7070-105 3351244 11/12/2014 Nov-16 60-7070-105 3351631 11/12/2014 Nov-16 60-7070-105 3352419 11/12/2014 Nov-16 60-7070-104 3349569 11/12/2014 Nov-16 60-7070-104 3351575 11/12/2014 Nov-16 60-7070-106 3349229 11/12/2014 Nov-16 60-7070-106 3349422 11/12/2014 Nov-16 60-7070-106 3350645 11/12/2014 Nov-16 60-7070-106 3350716 11/12/2014 Nov-16 60-7070-106 3351840 11/12/2014 Nov-16 60-7070-106 3352000 11/12/2014 Nov-16 60-7070-106 3353623 11/12/2014 Nov-16 60-7070-106 3354036 11/12/2014 Nov-16 60-7070-107 3350217 11/12/2014 Nov-16 60-7070-107 3353482 11/12/2014 Nov-16 60-7075-003 3348950 11/12/2014 Nov-16 60-7075-103 3350511 11/12/2014 Nov-16 60-7075-103 3352909 11/12/2014 Nov-16 60-7075-103 3353566 11/12/2014 Nov-16 60-7075-104 3353359 11/12/2014 Nov-16 60-7075-105 3351935 11/12/2014 Nov-16 60-7075-105 3353532 11/12/2014 Nov-16 60-7075-106 3350414 11/12/2014 Nov-16 60-7075-106 3350530 11/12/2014 Nov-16 60-7075-106 3352708 11/12/2014 Nov-16 5921-018-135 3354281 11/14/2014 Nov-16 5921-018-135 3355687 11/14/2014 Nov-16 5921-018-235 3356923 11/14/2014 Nov-16 5921-024-135 3356934 11/14/2014 Nov-16 5921-024-235 3356947 11/14/2014 Nov-16 5921-024-235 3358573 11/14/2014 Nov-16 5921-030-135 3355042 11/14/2014 Nov-16 5921-030-135 3358347 11/14/2014 Nov-16 5921-030-235 3354321 11/14/2014 Nov-16 5921-030-235 3354571 11/14/2014 Nov-16 5921-030-235 3355192 11/14/2014 Nov-16 5921-030-235 3357178 11/14/2014 Nov-16 5921-030-235 3357653 11/14/2014 Nov-16 5921-034-135 3354155 11/14/2014 Nov-16 5921-034-135 3355527 11/14/2014 Nov-16 5921-034-135 3355656 11/14/2014 Nov-16 5921-034-135 3357386 11/14/2014 Nov-16 5921-034-136 3355850 11/14/2014 Nov-16 5921-034-235 3354537 11/14/2014 Nov-16 5921-034-235 3355558 11/14/2014 Nov-16 5921-034-235 3357567 11/14/2014 Nov-16 5921-034-235 3358174 11/14/2014 Nov-16 5921-034-235 3358797 11/14/2014 Nov-16 5921-034-236 3358404 11/14/2014 Nov-16 5921-218-235 3354977 11/14/2014 Nov-16 5921-218-235 3356492 11/14/2014 Nov-16 5921-218-235 3357291 11/14/2014 Nov-16 5921-218-235 3358368 11/14/2014 Nov-16 5921-044-235 3357859 11/14/2014 Nov-16 5921-218-135 3354309 11/14/2014 Nov-16 5921-218-135 3358594 11/14/2014 Nov-16 60-7070-103 3354299 11/14/2014 Nov-16 60-7070-103 3354724 11/14/2014 Nov-16 60-7070-103 3354965 11/14/2014 Nov-16 60-7070-103 3355401 11/14/2014 Nov-16 60-7070-103 3356779 11/14/2014 Nov-16 60-7070-103 3357828 11/14/2014 Nov-16 60-7070-103 3358709 11/14/2014 Nov-16 60-7070-105 3354996 11/14/2014 Nov-16 60-7070-105 3357145 11/14/2014 Nov-16 60-7070-104 3357790 11/14/2014 Nov-16 60-7070-003 3354313 11/14/2014 Nov-16 60-7070-006 3354500 11/14/2014 Nov-16 60-7070-106 3356018 11/14/2014 Nov-16 60-7070-106 3356714 11/14/2014 Nov-16 60-7070-106 3356808 11/14/2014 Nov-16 60-7070-106 3357871 11/14/2014 Nov-16 60-7070-106 3358142 11/14/2014 Nov-16 60-7070-107 3355821 11/14/2014 Nov-16 60-7070-107 3358750 11/14/2014 Nov-16 60-7075-003 3355878 11/14/2014 Nov-16 60-7075-006 3356812 11/14/2014 Nov-16 60-7075-103 3355185 11/14/2014 Nov-16 60-7075-105 3355090 11/14/2014 Nov-16 60-7075-106 3356041 11/14/2014 Nov-16 60-7075-106 3357270 11/14/2014 Nov-16 5921-018-135 3360753 11/17/2014 Nov-16 5921-018-235 3361488 11/17/2014 Nov-16 5921-024-235 3359985 11/17/2014 Nov-16 5921-030-135 3359729 11/17/2014 Nov-16 5921-030-135 3364008 11/17/2014 Nov-16 5921-030-136 3359362 11/17/2014 Nov-16 5921-030-235 3358933 11/17/2014 Nov-16 5921-030-235 3360425 11/17/2014 Nov-16 5921-030-235 3360544 11/17/2014 Nov-16 5921-030-235 3362198 11/17/2014 Nov-16 5921-030-235 3363025 11/17/2014 Nov-16 5921-030-235 3363849 11/17/2014 Nov-16 5921-030-235 3364065 11/17/2014 Nov-16 5921-034-135 3359260 11/17/2014 Nov-16 5921-034-135 3359944 11/17/2014 Nov-16 5921-034-135 3363348 11/17/2014 Nov-16 5921-034-235 3359614 11/17/2014 Nov-16 5921-034-235 3359819 11/17/2014 Nov-16 5921-034-235 3360027 11/17/2014 Nov-16 5921-034-235 3361978 11/17/2014 Nov-16 5921-034-235 3362228 11/17/2014 Nov-16 5921-034-235 3362654 11/17/2014 Nov-16 5921-034-235 3363588 11/17/2014 Nov-16 5921-034-235 3364168 11/17/2014 Nov-16 5921-218-235 3359679 11/17/2014 Nov-16 5921-218-235 3360711 11/17/2014 Nov-16 5921-218-235 3363170 11/17/2014 Nov-16 5921-218-235 3364293 11/17/2014 Nov-16 5921-030-236 3359888 11/17/2014 Nov-16 5921-044-135 3359315 11/17/2014 Nov-16 5921-218-135 3359772 11/17/2014 Nov-16 5921-218-135 3361909 11/17/2014 Nov-16 5921-218-135 3364744 11/17/2014 Nov-16 5921-218-136 3359389 11/17/2014 Nov-16 5921-218-136 3362011 11/17/2014 Nov-16 60-7070-103 3358951 11/17/2014 Nov-16 60-7070-103 3361071 11/17/2014 Nov-16 60-7070-103 3362200 11/17/2014 Nov-16 60-7070-103 3362285 11/17/2014 Nov-16 60-7070-103 3363324 11/17/2014 Nov-16 60-7070-103 3364479 11/17/2014 Nov-16 60-7070-105 3360635 11/17/2014 Nov-16 60-7070-105 3362608 11/17/2014 Nov-16 60-7070-105 3362762 11/17/2014 Nov-16 60-7070-105 3363629 11/17/2014 Nov-16 60-7070-104 3359013 11/17/2014 Nov-16 60-7070-104 3360500 11/17/2014 Nov-16 60-7070-104 3362990 11/17/2014 Nov-16 60-7070-104 3364553 11/17/2014 Nov-16 60-7070-003 3361691 11/17/2014 Nov-16 60-7070-004 3359294 11/17/2014 Nov-16 60-7070-005 3359508 11/17/2014 Nov-16 60-7070-005 3361670 11/17/2014 Nov-16 60-7070-006 3360193 11/17/2014 Nov-16 60-7070-106 3358824 11/17/2014 Nov-16 60-7070-106 3359583 11/17/2014 Nov-16 60-7070-106 3361435 11/17/2014 Nov-16 60-7070-106 3361501 11/17/2014 Nov-16 60-7070-106 3362855 11/17/2014 Nov-16 60-7070-106 3364728 11/17/2014 Nov-16 60-7070-107 3359992 11/17/2014 Nov-16 60-7070-107 3364031 11/17/2014 Nov-16 60-7075-004 3359391 11/17/2014 Nov-16 60-7075-004 3362024 11/17/2014 Nov-16 60-7080-102 3359327 11/17/2014 Nov-16 60-7075-103 3361346 11/17/2014 Nov-16 60-7075-103 3363492 11/17/2014 Nov-16 60-7075-104 3361643 11/17/2014 Nov-16 60-7075-106 3360977 11/17/2014 Nov-16 60-7075-106 3363934 11/17/2014 Nov-16 5921-018-135 3365622 11/19/2014 Nov-16 5921-018-135 3368567 11/19/2014 Nov-16 5921-018-235 3368184 11/19/2014 Nov-16 5921-024-135 3365050 11/19/2014 Nov-16 5921-024-135 3369723 11/19/2014 Nov-16 5921-024-235 3365124 11/19/2014 Nov-16 5921-024-235 3367525 11/19/2014 Nov-16 5921-024-235 3371003 11/19/2014 Nov-16 5921-030-135 3368364 11/19/2014 Nov-16 5921-030-135 3369845 11/19/2014 Nov-16 5921-030-235 3365508 11/19/2014 Nov-16 5921-030-235 3367190 11/19/2014 Nov-16 5921-030-235 3369440 11/19/2014 Nov-16 5921-030-235 3369982 11/19/2014 Nov-16 5921-034-135 3365362 11/19/2014 Nov-16 5921-034-135 3366046 11/19/2014 Nov-16 5921-034-135 3367671 11/19/2014 Nov-16 5921-034-135 3369365 11/19/2014 Nov-16 5921-034-136 3365565 11/19/2014 Nov-16 5921-034-235 3365406 11/19/2014 Nov-16 5921-034-235 3366407 11/19/2014 Nov-16 5921-034-235 3366756 11/19/2014 Nov-16 5921-034-235 3367830 11/19/2014 Nov-16 5921-034-235 3369214 11/19/2014 Nov-16 5921-034-235 3370348 11/19/2014 Nov-16 5921-218-235 3365971 11/19/2014 Nov-16 5921-218-235 3366183 11/19/2014 Nov-16 5921-218-235 3366578 11/19/2014 Nov-16 5921-218-235 3366804 11/19/2014 Nov-16 5921-218-235 3368047 11/19/2014 Nov-16 5921-218-235 3368236 11/19/2014 Nov-16 5921-218-235 3369813 11/19/2014 Nov-16 5921-218-135 3366033 11/19/2014 Nov-16 5921-218-135 3368258 11/19/2014 Nov-16 60-7070-103 3365370 11/19/2014 Nov-16 60-7070-103 3365959 11/19/2014 Nov-16 60-7070-103 3367368 11/19/2014 Nov-16 60-7070-103 3368624 11/19/2014 Nov-16 60-7070-105 3366392 11/19/2014 Nov-16 60-7070-105 3370526 11/19/2014 Nov-16 60-7070-105 3370590 11/19/2014 Nov-16 60-7070-104 3367573 11/19/2014 Nov-16 60-7070-104 3369464 11/19/2014 Nov-16 60-7070-003 3368477 11/19/2014 Nov-16 60-7070-006 3368454 11/19/2014 Nov-16 60-7070-106 3366203 11/19/2014 Nov-16 60-7070-106 3366566 11/19/2014 Nov-16 60-7070-106 3367811 11/19/2014 Nov-16 60-7070-106 3369246 11/19/2014 Nov-16 60-7070-106 3370104 11/19/2014 Nov-16 60-7070-107 3368178 11/19/2014 Nov-16 60-7075-003 3369455 11/19/2014 Nov-16 60-7075-006 3364989 11/19/2014 Nov-16 60-7075-006 3370519 11/19/2014 Nov-16 60-7075-103 3368034 11/19/2014 Nov-16 60-7075-103 3369781 11/19/2014 Nov-16 60-7075-106 3368052 11/19/2014 Nov-16 60-7075-106 3369377 11/19/2014 Nov-16 60-7075-106 3369438 11/19/2014 Nov-16 60-7070-106 3354707 11/20/2014 Nov-16 5921-018-235 3372891 11/21/2014 Nov-16 5921-024-235 3373852 11/21/2014 Nov-16 5921-030-135 3371953 11/21/2014 Nov-16 5921-030-135 3373211 11/21/2014 Nov-16 5921-030-135 3373295 11/21/2014 Nov-16 5921-030-135 3375715 11/21/2014 Nov-16 5921-030-235 3372051 11/21/2014 Nov-16 5921-030-235 3372407 11/21/2014 Nov-16 5921-030-235 3372509 11/21/2014 Nov-16 5921-030-235 3373765 11/21/2014 Nov-16 5921-030-235 3374611 11/21/2014 Nov-16 5921-030-235 3375292 11/21/2014 Nov-16 5921-034-135 3371383 11/21/2014 Nov-16 5921-034-135 3372177 11/21/2014 Nov-16 5921-034-135 3373380 11/21/2014 Nov-16 5921-034-135 3375151 11/21/2014 Nov-16 5921-034-235 3371639 11/21/2014 Nov-16 5921-034-235 3372411 11/21/2014 Nov-16 5921-034-235 3373796 11/21/2014 Nov-16 5921-034-235 3374291 11/21/2014 Nov-16 5921-034-235 3374761 11/21/2014 Nov-16 5921-034-235 3375285 11/21/2014 Nov-16 5921-218-235 3371026 11/21/2014 Nov-16 5921-218-235 3372112 11/21/2014 Nov-16 5921-218-235 3372430 11/21/2014 Nov-16 5921-218-235 3373909 11/21/2014 Nov-16 5921-218-235 3374301 11/21/2014 Nov-16 5921-218-235 3374837 11/21/2014 Nov-16 5921-218-135 3370780 11/21/2014 Nov-16 5921-218-135 3371365 11/21/2014 Nov-16 5921-218-135 3372699 11/21/2014 Nov-16 5921-218-135 3373754 11/21/2014 Nov-16 5921-218-135 3375829 11/21/2014 Nov-16 60-7070-103 3375421 11/21/2014 Nov-16 60-7070-103 3375649 11/21/2014 Nov-16 60-7070-105 3371948 11/21/2014 Nov-16 60-7070-105 3372584 11/21/2014 Nov-16 60-7070-105 3372901 11/21/2014 Nov-16 60-7070-105 3374081 11/21/2014 Nov-16 60-7070-105 3374703 11/21/2014 Nov-16 60-7070-105 3375356 11/21/2014 Nov-16 60-7070-104 3372067 11/21/2014 Nov-16 60-7070-104 3373564 11/21/2014 Nov-16 60-7070-003 3374068 11/21/2014 Nov-16 60-7070-004 3372843 11/21/2014 Nov-16 60-7070-006 3371354 11/21/2014 Nov-16 60-7070-006 3372815 11/21/2014 Nov-16 60-7070-006 3374270 11/21/2014 Nov-16 60-7070-106 3371821 11/21/2014 Nov-16 60-7070-106 3372363 11/21/2014 Nov-16 60-7070-106 3373419 11/21/2014 Nov-16 60-7070-106 3374859 11/21/2014 Nov-16 60-7070-107 3376106 11/21/2014 Nov-16 60-7075-006 3370981 11/21/2014 Nov-16 60-7075-103 3371656 11/21/2014 Nov-16 60-7075-103 3372773 11/21/2014 Nov-16 60-7075-103 3374075 11/21/2014 Nov-16 60-7075-103 3375198 11/21/2014 Nov-16 60-7075-104 3368598 11/21/2014 Nov-16 60-7075-105 3375762 11/21/2014 Nov-16 60-7075-105 3376031 11/21/2014 Nov-16 60-7075-106 3371192 11/21/2014 Nov-16 60-7075-106 3372739 11/21/2014 Nov-16 60-7075-106 3373950 11/21/2014 Nov-16 60-7075-106 3373997 11/21/2014 Nov-16 60-7075-106 3374726 11/21/2014 Nov-16 60-7075-106 3375259 11/21/2014 Nov-16 5921-018-135 3378916 11/24/2014 Nov-16 5921-018-235 3376962 11/24/2014 Nov-16 5921-024-235 3377281 11/24/2014 Nov-16 5921-024-235 3378169 11/24/2014 Nov-16 5921-030-135 3378928 11/24/2014 Nov-16 5921-030-235 3378370 11/24/2014 Nov-16 5921-030-235 3379560 11/24/2014 Nov-16 5921-034-135 3377210 11/24/2014 Nov-16 5921-034-235 3376217 11/24/2014 Nov-16 5921-034-235 3378013 11/24/2014 Nov-16 5921-034-235 3378066 11/24/2014 Nov-16 5921-034-235 3378153 11/24/2014 Nov-16 5921-034-235 3378195 11/24/2014 Nov-16 5921-218-235 3376319 11/24/2014 Nov-16 5921-218-235 3377619 11/24/2014 Nov-16 5921-218-135 3378679 11/24/2014 Nov-16 60-7070-103 3377537 11/24/2014 Nov-16 60-7070-103 3377822 11/24/2014 Nov-16 60-7070-103 3379306 11/24/2014 Nov-16 60-7070-103 3379869 11/24/2014 Nov-16 60-7070-105 3376479 11/24/2014 Nov-16 60-7070-105 3378786 11/24/2014 Nov-16 60-7070-105 3378835 11/24/2014 Nov-16 60-7070-105 3379876 11/24/2014 Nov-16 60-7070-104 3376744 11/24/2014 Nov-16 60-7070-104 3376927 11/24/2014 Nov-16 60-7070-003 3376936 11/24/2014 Nov-16 60-7070-106 3377275 11/24/2014 Nov-16 60-7070-106 3378202 11/24/2014 Nov-16 60-7070-106 3379334 11/24/2014 Nov-16 60-7070-106 3379848 11/24/2014 Nov-16 60-7070-107 3379119 11/24/2014 Nov-16 60-7075-003 3376342 11/24/2014 Nov-16 60-7075-006 3376609 11/24/2014 Nov-16 60-7075-104 3376728 11/24/2014 Nov-16 60-7075-106 3378475 11/24/2014 Nov-16 5921-018-235 3381425 11/26/2014 Nov-16 5921-024-235 3381058 11/26/2014 Nov-16 5921-024-235 3384575 11/26/2014 Nov-16 5921-030-135 3385217 11/26/2014 Nov-16 5921-030-235 3380591 11/26/2014 Nov-16 5921-030-235 3381516 11/26/2014 Nov-16 5921-030-235 3384339 11/26/2014 Nov-16 5921-030-235 3385684 11/26/2014 Nov-16 5921-034-135 3381172 11/26/2014 Nov-16 5921-034-135 3384151 11/26/2014 Nov-16 5921-034-135 3385697 11/26/2014 Nov-16 5921-034-136 3382956 11/26/2014 Nov-16 5921-034-235 3380440 11/26/2014 Nov-16 5921-034-235 3380589 11/26/2014 Nov-16 5921-034-235 3380710 11/26/2014 Nov-16 5921-034-235 3382454 11/26/2014 Nov-16 5921-034-235 3383644 11/26/2014 Nov-16 5921-034-235 3385326 11/26/2014 Nov-16 5921-218-235 3380246 11/26/2014 Nov-16 5921-218-235 3381563 11/26/2014 Nov-16 5921-218-235 3384048 11/26/2014 Nov-16 5921-018-236 3385763 11/26/2014 Nov-16 5921-218-135 3381456 11/26/2014 Nov-16 5921-218-135 3381537 11/26/2014 Nov-16 5921-218-135 3385666 11/26/2014 Nov-16 60-7070-103 3381822 11/26/2014 Nov-16 60-7070-103 3382393 11/26/2014 Nov-16 60-7070-103 3383876 11/26/2014 Nov-16 60-7070-103 3383946 11/26/2014 Nov-16 60-7070-105 3380349 11/26/2014 Nov-16 60-7070-105 3382774 11/26/2014 Nov-16 60-7070-105 3384318 11/26/2014 Nov-16 60-7070-105 3385530 11/26/2014 Nov-16 60-7070-104 3381724 11/26/2014 Nov-16 60-7070-104 3382431 11/26/2014 Nov-16 60-7070-104 3385335 11/26/2014 Nov-16 60-7070-104 3385548 11/26/2014 Nov-16 60-7070-003 3382682 11/26/2014 Nov-16 60-7070-003 3384665 11/26/2014 Nov-16 60-7070-006 3383174 11/26/2014 Nov-16 60-7070-006 3383254 11/26/2014 Nov-16 60-7070-106 3381129 11/26/2014 Nov-16 60-7070-106 3381252 11/26/2014 Nov-16 60-7070-106 3382522 11/26/2014 Nov-16 60-7070-106 3384445 11/26/2014 Nov-16 60-7070-106 3385759 11/26/2014 Nov-16 60-7070-107 3385264 11/26/2014 Nov-16 60-7075-006 3381210 11/26/2014 Nov-16 60-7075-006 3385873 11/26/2014 Nov-16 60-7075-103 3380911 11/26/2014 Nov-16 60-7075-103 3382863 11/26/2014 Nov-16 60-7075-103 3384114 11/26/2014 Nov-16 60-7075-103 3384263 11/26/2014 Nov-16 60-7075-104 3382844 11/26/2014 Nov-16 60-7075-106 3381642 11/26/2014 Nov-16 60-7075-106 3383142 11/26/2014 Nov-16 60-7075-106 3383541 11/26/2014 Nov-16 5921-018-135 3387538 12/1/2014 Nov-16 5921-018-135 3390611 12/1/2014 Nov-16 5921-024-235 3390393 12/1/2014 Nov-16 5921-030-135 3390533 12/1/2014 Nov-16 5921-030-135 3390892 12/1/2014 Nov-16 5921-030-235 3386410 12/1/2014 Nov-16 5921-030-235 3387000 12/1/2014 Nov-16 5921-030-235 3389189 12/1/2014 Nov-16 5921-034-135 3386963 12/1/2014 Nov-16 5921-034-135 3387754 12/1/2014 Nov-16 5921-034-135 3388681 12/1/2014 Nov-16 5921-034-235 3387490 12/1/2014 Nov-16 5921-034-235 3387731 12/1/2014 Nov-16 5921-034-235 3387796 12/1/2014 Nov-16 5921-034-235 3388242 12/1/2014 Nov-16 5921-034-235 3389503 12/1/2014 Nov-16 5921-034-235 3391177 12/1/2014 Nov-16 5921-218-235 3386097 12/1/2014 Nov-16 5921-218-235 3387540 12/1/2014 Nov-16 5921-218-235 3387997 12/1/2014 Nov-16 5921-218-235 3388597 12/1/2014 Nov-16 5921-218-235 3389780 12/1/2014 Nov-16 5921-030-236 3386520 12/1/2014 Nov-16 5921-218-135 3387282 12/1/2014 Nov-16 5921-218-135 3390166 12/1/2014 Nov-16 60-7070-102 3390034 12/1/2014 Nov-16 60-7070-103 3384683 12/1/2014 Nov-16 60-7070-103 3388741 12/1/2014 Nov-16 60-7070-103 3389073 12/1/2014 Nov-16 60-7070-103 3390973 12/1/2014 Nov-16 60-7070-103 3391067 12/1/2014 Nov-16 60-7070-105 3386646 12/1/2014 Nov-16 60-7070-105 3387211 12/1/2014 Nov-16 60-7070-105 3388046 12/1/2014 Nov-16 60-7070-105 3389617 12/1/2014 Nov-16 60-7070-105 3389910 12/1/2014 Nov-16 60-7070-105 3391235 12/1/2014 Nov-16 60-7070-104 3386475 12/1/2014 Nov-16 60-7070-104 3389297 12/1/2014 Nov-16 60-7070-003 3386508 12/1/2014 Nov-16 60-7070-005 3388274 12/1/2014 Nov-16 60-7070-006 3386084 12/1/2014 Nov-16 60-7070-006 3387712 12/1/2014 Nov-16 60-7070-106 3386066 12/1/2014 Nov-16 60-7070-106 3386764 12/1/2014 Nov-16 60-7070-106 3387457 12/1/2014 Nov-16 60-7070-106 3387806 12/1/2014 Nov-16 60-7070-106 3389444 12/1/2014 Nov-16 60-7070-106 3389561 12/1/2014 Nov-16 60-7070-106 3391165 12/1/2014 Nov-16 60-7070-107 3387333 12/1/2014 Nov-16 60-7070-107 3388469 12/1/2014 Nov-16 60-7075-003 3388154 12/1/2014 Nov-16 60-7075-003 3388219 12/1/2014 Nov-16 60-7075-103 3386916 12/1/2014 Nov-16 60-7075-103 3388033 12/1/2014 Nov-16 60-7075-103 3389906 12/1/2014 Nov-16 60-7075-104 3387173 12/1/2014 Nov-16 60-7075-105 3388578 12/1/2014 Nov-16 60-7075-106 3386468 12/1/2014 Nov-16 60-7075-106 3386937 12/1/2014 Nov-16 60-7075-106 3390262 12/1/2014 Nov-16 60-7075-106 3390483 12/1/2014 Nov-16 5921-018-135 3395306 12/3/2014 Dec-16 5921-018-235 3392747 12/3/2014 Dec-16 5921-024-135 3391545 12/3/2014 Dec-16 5921-024-235 3392897 12/3/2014 Dec-16 5921-024-235 3395980 12/3/2014 Dec-16 5921-030-135 3392349 12/3/2014 Dec-16 5921-030-135 3395709 12/3/2014 Dec-16 5921-030-235 3391794 12/3/2014 Dec-16 5921-030-235 3393927 12/3/2014 Dec-16 5921-030-235 3394820 12/3/2014 Dec-16 5921-030-235 3397032 12/3/2014 Dec-16 5921-034-135 3392625 12/3/2014 Dec-16 5921-034-135 3392866 12/3/2014 Dec-16 5921-034-135 3394588 12/3/2014 Dec-16 5921-034-135 3396000 12/3/2014 Dec-16 5921-034-136 3397520 12/3/2014 Dec-16 5921-034-235 3392116 12/3/2014 Dec-16 5921-034-235 3393498 12/3/2014 Dec-16 5921-034-235 3394158 12/3/2014 Dec-16 5921-034-235 3395560 12/3/2014 Dec-16 5921-034-235 3396913 12/3/2014 Dec-16 5921-034-235 3397045 12/3/2014 Dec-16 5921-218-235 3391385 12/3/2014 Dec-16 5921-218-235 3393422 12/3/2014 Dec-16 5921-218-235 3394664 12/3/2014 Dec-16 5921-218-235 3396125 12/3/2014 Dec-16 5921-218-235 3397406 12/3/2014 Dec-16 5921-218-135 3392159 12/3/2014 Dec-16 5921-218-135 3397287 12/3/2014 Dec-16 60-7070-103 3393326 12/3/2014 Dec-16 60-7070-103 3393374 12/3/2014 Dec-16 60-7070-103 3393511 12/3/2014 Dec-16 60-7070-103 3393970 12/3/2014 Dec-16 60-7070-103 3393989 12/3/2014 Dec-16 60-7070-103 3394561 12/3/2014 Dec-16 60-7070-103 3396248 12/3/2014 Dec-16 60-7070-103 3396555 12/3/2014 Dec-16 60-7070-105 3391532 12/3/2014 Dec-16 60-7070-105 3393031 12/3/2014 Dec-16 60-7070-105 3393632 12/3/2014 Dec-16 60-7070-105 3394297 12/3/2014 Dec-16 60-7070-105 3395040 12/3/2014 Dec-16 60-7070-105 3397294 12/3/2014 Dec-16 60-7070-104 3393175 12/3/2014 Dec-16 60-7070-104 3396253 12/3/2014 Dec-16 60-7070-003 3392464 12/3/2014 Dec-16 60-7070-005 3395784 12/3/2014 Dec-16 60-7070-006 3395979 12/3/2014 Dec-16 60-7070-006 3396028 12/3/2014 Dec-16 60-7070-106 3393388 12/3/2014 Dec-16 60-7070-106 3394348 12/3/2014 Dec-16 60-7070-106 3394722 12/3/2014 Dec-16 60-7070-106 3396276 12/3/2014 Dec-16 60-7070-106 3396461 12/3/2014 Dec-16 60-7070-107 3392504 12/3/2014 Dec-16 60-7070-107 3396314 12/3/2014 Dec-16 60-7075-006 3392396 12/3/2014 Dec-16 60-7075-103 3391756 12/3/2014 Dec-16 60-7075-103 3394278 12/3/2014 Dec-16 60-7075-103 3395491 12/3/2014 Dec-16 60-7075-104 3392953 12/3/2014 Dec-16 60-7075-105 3391606 12/3/2014 Dec-16 60-7075-105 3395573 12/3/2014 Dec-16 60-7075-106 3391699 12/3/2014 Dec-16 60-7075-106 3394213 12/3/2014 Dec-16 60-7075-106 3395119 12/3/2014 Dec-16 5921-018-235 3401967 12/11/2014 Dec-16 5921-024-135 3401265 12/11/2014 Dec-16 5921-024-235 3397583 12/11/2014 Dec-16 5921-024-235 3400224 12/11/2014 Dec-16 5921-024-235 3402342 12/11/2014 Dec-16 5921-030-135 3398036 12/11/2014 Dec-16 5921-030-135 3402763 12/11/2014 Dec-16 5921-030-136 3402286 12/11/2014 Dec-16 5921-030-235 3398688 12/11/2014 Dec-16 5921-030-235 3399926 12/11/2014 Dec-16 5921-034-135 3397698 12/11/2014 Dec-16 5921-034-135 3398983 12/11/2014 Dec-16 5921-034-135 3400615 12/11/2014 Dec-16 5921-034-235 3398155 12/11/2014 Dec-16 5921-034-235 3398309 12/11/2014 Dec-16 5921-034-235 3398753 12/11/2014 Dec-16 5921-034-235 3399919 12/11/2014 Dec-16 5921-034-235 3401600 12/11/2014 Dec-16 5921-034-235 3401709 12/11/2014 Dec-16 5921-034-235 3402613 12/11/2014 Dec-16 5921-218-235 3398500 12/11/2014 Dec-16 5921-218-235 3398748 12/11/2014 Dec-16 5921-218-235 3400253 12/11/2014 Dec-16 5921-218-235 3402020 12/11/2014 Dec-16 5921-044-135 3402915 12/11/2014 Dec-16 5921-218-135 3398070 12/11/2014 Dec-16 5921-218-135 3400403 12/11/2014 Dec-16 5921-218-135 3400636 12/11/2014 Dec-16 5921-218-135 3401663 12/11/2014 Dec-16 60-7070-103 3399198 12/11/2014 Dec-16 60-7070-103 3401895 12/11/2014 Dec-16 60-7070-103 3402199 12/11/2014 Dec-16 60-7070-103 3402927 12/11/2014 Dec-16 60-7070-105 3398769 12/11/2014 Dec-16 60-7070-105 3399069 12/11/2014 Dec-16 60-7070-105 3400248 12/11/2014 Dec-16 60-7070-105 3400419 12/11/2014 Dec-16 60-7070-105 3401541 12/11/2014 Dec-16 60-7070-104 3399360 12/11/2014 Dec-16 60-7070-104 3400978 12/11/2014 Dec-16 60-7070-003 3402632 12/11/2014 Dec-16 60-7070-005 3401980 12/11/2014 Dec-16 60-7070-006 3402678 12/11/2014 Dec-16 60-7070-106 3397554 12/11/2014 Dec-16 60-7070-106 3399547 12/11/2014 Dec-16 60-7070-106 3399817 12/11/2014 Dec-16 60-7070-106 3399988 12/11/2014 Dec-16 60-7070-106 3400921 12/11/2014 Dec-16 60-7070-106 3401231 12/11/2014 Dec-16 60-7070-106 3402609 12/11/2014 Dec-16 60-7070-106 3402716 12/11/2014 Dec-16 60-7070-107 3399114 12/11/2014 Dec-16 60-7070-107 3402168 12/11/2014 Dec-16 60-7075-103 3397596 12/11/2014 Dec-16 60-7075-103 3401283 12/11/2014 Dec-16 60-7075-103 3402936 12/11/2014 Dec-16 60-7075-104 3398275 12/11/2014 Dec-16 60-7075-105 3402240 12/11/2014 Dec-16 60-7075-106 3398528 12/11/2014 Dec-16 60-7075-106 3400966 12/11/2014 Dec-16 5921-018-135 3405972 12/13/2014 Dec-16 5921-024-235 3409247 12/13/2014 Dec-16 5921-024-235 3411894 12/13/2014 Dec-16 5921-024-235 3412064 12/13/2014 Dec-16 5921-030-135 3409299 12/13/2014 Dec-16 5921-030-135 3411927 12/13/2014 Dec-16 5921-030-235 3410888 12/13/2014 Dec-16 5921-030-235 3412174 12/13/2014 Dec-16 5921-034-135 3409544 12/13/2014 Dec-16 5921-034-235 3409252 12/13/2014 Dec-16 5921-034-235 3411175 12/13/2014 Dec-16 5921-034-236 3412462 12/13/2014 Dec-16 5921-218-235 3411446 12/13/2014 Dec-16 5921-218-235 3411524 12/13/2014 Dec-16 5921-218-135 3409690 12/13/2014 Dec-16 5921-218-135 3411609 12/13/2014 Dec-16 5921-218-135 3412399 12/13/2014 Dec-16 60-7070-103 3408919 12/13/2014 Dec-16 60-7070-103 3409281 12/13/2014 Dec-16 60-7070-103 3410045 12/13/2014 Dec-16 60-7070-103 3411802 12/13/2014 Dec-16 60-7070-103 3411958 12/13/2014 Dec-16 60-7070-103 3412489 12/13/2014 Dec-16 60-7070-103 3412573 12/13/2014 Dec-16 60-7070-103 3412732 12/13/2014 Dec-16 60-7070-105 3409268 12/13/2014 Dec-16 60-7070-105 3410874 12/13/2014 Dec-16 60-7070-105 3410992 12/13/2014 Dec-16 60-7070-105 3412277 12/13/2014 Dec-16 60-7070-104 3409688 12/13/2014 Dec-16 60-7070-104 3410032 12/13/2014 Dec-16 60-7070-104 3412265 12/13/2014 Dec-16 60-7070-106 3408758 12/13/2014 Dec-16 60-7070-106 3409983 12/13/2014 Dec-16 60-7070-106 3410362 12/13/2014 Dec-16 60-7070-106 3412188 12/13/2014 Dec-16 60-7070-106 3412685 12/13/2014 Dec-16 60-7070-106 3412980 12/13/2014 Dec-16 60-7070-107 3411553 12/13/2014 Dec-16 60-7070-107 3412637 12/13/2014 Dec-16 60-7075-103 3409275 12/13/2014 Dec-16 60-7075-103 3411582 12/13/2014 Dec-16 60-7075-103 3411860 12/13/2014 Dec-16 60-7075-104 3411199 12/13/2014 Dec-16 60-7075-106 3410928 12/13/2014 Dec-16 60-7075-106 3412404 12/13/2014 Dec-16 5921-018-135 3413859 12/15/2014 Dec-16 5921-018-135 3416159 12/15/2014 Dec-16 5921-024-135 3414868 12/15/2014 Dec-16 5921-024-235 3413726 12/15/2014 Dec-16 5921-024-235 3417216 12/15/2014 Dec-16 5921-030-135 3414085 12/15/2014 Dec-16 5921-030-135 3416752 12/15/2014 Dec-16 5921-030-235 3413719 12/15/2014 Dec-16 5921-030-235 3415129 12/15/2014 Dec-16 5921-030-235 3416383 12/15/2014 Dec-16 5921-030-235 3418060 12/15/2014 Dec-16 5921-034-135 3414037 12/15/2014 Dec-16 5921-034-135 3416088 12/15/2014 Dec-16 5921-034-135 3416331 12/15/2014 Dec-16 5921-034-235 3413121 12/15/2014 Dec-16 5921-034-235 3415101 12/15/2014 Dec-16 5921-034-235 3415172 12/15/2014 Dec-16 5921-034-235 3417285 12/15/2014 Dec-16 5921-034-235 3418073 12/15/2014 Dec-16 5921-218-235 3413184 12/15/2014 Dec-16 5921-218-235 3415402 12/15/2014 Dec-16 5921-218-235 3418056 12/15/2014 Dec-16 5921-030-236 3418018 12/15/2014 Dec-16 5921-218-135 3414540 12/15/2014 Dec-16 5921-218-135 3417318 12/15/2014 Dec-16 60-7070-102 3416464 12/15/2014 Dec-16 60-7070-103 3414834 12/15/2014 Dec-16 60-7070-103 3414881 12/15/2014 Dec-16 60-7070-103 3416799 12/15/2014 Dec-16 60-7070-103 3417957 12/15/2014 Dec-16 60-7070-105 3413412 12/15/2014 Dec-16 60-7070-105 3413630 12/15/2014 Dec-16 60-7070-105 3415397 12/15/2014 Dec-16 60-7070-105 3416897 12/15/2014 Dec-16 60-7070-104 3415425 12/15/2014 Dec-16 60-7070-003 3411595 12/15/2014 Dec-16 60-7070-003 3417961 12/15/2014 Dec-16 60-7070-106 3414269 12/15/2014 Dec-16 60-7070-106 3415635 12/15/2014 Dec-16 60-7070-106 3415983 12/15/2014 Dec-16 60-7070-107 3416061 12/15/2014 Dec-16 60-7075-003 3414997 12/15/2014 Dec-16 60-7075-004 3413393 12/15/2014 Dec-16 60-7075-006 3414984 12/15/2014 Dec-16 60-7075-103 3414194 12/15/2014 Dec-16 60-7075-104 3416438 12/15/2014 Dec-16 60-7075-105 3414351 12/15/2014 Dec-16 60-7075-105 3416128 12/15/2014 Dec-16 60-7075-106 3413757 12/15/2014 Dec-16 60-7075-106 3415186 12/15/2014 Dec-16 60-7075-106 3417259 12/15/2014 Dec-16 5921-018-135 3420138 12/18/2014 Dec-16 5921-024-235 3419862 12/18/2014 Dec-16 5921-024-235 3420880 12/18/2014 Dec-16 5921-024-235 3422683 12/18/2014 Dec-16 5921-030-135 3420776 12/18/2014 Dec-16 5921-030-135 3421403 12/18/2014 Dec-16 5921-030-135 3423266 12/18/2014 Dec-16 5921-030-235 3418213 12/18/2014 Dec-16 5921-030-235 3419993 12/18/2014 Dec-16 5921-030-235 3421139 12/18/2014 Dec-16 5921-030-235 3423348 12/18/2014 Dec-16 5921-034-135 3418934 12/18/2014 Dec-16 5921-034-135 3420164 12/18/2014 Dec-16 5921-034-135 3420592 12/18/2014 Dec-16 5921-034-135 3422339 12/18/2014 Dec-16 5921-034-135 3423503 12/18/2014 Dec-16 5921-034-235 3419235 12/18/2014 Dec-16 5921-034-235 3420223 12/18/2014 Dec-16 5921-034-235 3420949 12/18/2014 Dec-16
5921-034-235 3422048 12/18/2014 Dec-16 5921-034-235 3422791 12/18/2014 Dec-16 5921-034-235 3423245 12/18/2014 Dec-16 5921-034-236 3418459 12/18/2014 Dec-16 5921-218-235 3417302 12/18/2014 Dec-16 5921-218-235 3420748 12/18/2014 Dec-16 5921-218-235 3421806 12/18/2014 Dec-16 5921-218-235 3422704 12/18/2014 Dec-16 5921-030-236 3420977 12/18/2014 Dec-16 5921-044-235 3421000 12/18/2014 Dec-16 5921-218-135 3421899 12/18/2014 Dec-16 5921-218-135 3423369 12/18/2014 Dec-16 5921-218-136 3418266 12/18/2014 Dec-16 5921-218-236 3419344 12/18/2014 Dec-16 60-7070-102 3418158 12/18/2014 Dec-16 60-7070-103 3418444 12/18/2014 Dec-16 60-7070-103 3419870 12/18/2014 Dec-16 60-7070-103 3420186 12/18/2014 Dec-16 60-7070-103 3420661 12/18/2014 Dec-16 60-7070-103 3421248 12/18/2014 Dec-16 60-7070-103 3422762 12/18/2014 Dec-16 60-7070-103 3423232 12/18/2014 Dec-16 60-7070-105 3418353 12/18/2014 Dec-16 60-7070-105 3419804 12/18/2014 Dec-16 60-7070-105 3420140 12/18/2014 Dec-16 60-7070-105 3420657 12/18/2014 Dec-16 60-7070-105 3420931 12/18/2014 Dec-16 60-7070-105 3421966 12/18/2014 Dec-16 60-7070-105 3422167 12/18/2014 Dec-16 60-7070-105 3422736 12/18/2014 Dec-16 60-7070-105 3423376 12/18/2014 Dec-16 60-7070-104 3418535 12/18/2014 Dec-16 60-7070-104 3420920 12/18/2014 Dec-16 60-7070-104 3422445 12/18/2014 Dec-16 60-7070-006 3417413 12/18/2014 Dec-16 60-7070-106 3418561 12/18/2014 Dec-16 60-7070-106 3418629 12/18/2014 Dec-16 60-7070-106 3420101 12/18/2014 Dec-16 60-7070-106 3420397 12/18/2014 Dec-16 60-7070-106 3421269 12/18/2014 Dec-16 60-7070-106 3421604 12/18/2014 Dec-16 60-7070-106 3421712 12/18/2014 Dec-16 60-7070-106 3422608 12/18/2014 Dec-16 60-7070-107 3420642 12/18/2014 Dec-16 60-7070-107 3421950 12/18/2014 Dec-16 60-7075-003 3419699 12/18/2014 Dec-16 60-7075-004 3421658 12/18/2014 Dec-16 60-7075-006 3417432 12/18/2014 Dec-16 60-7075-103 3418348 12/18/2014 Dec-16 60-7075-103 3419476 12/18/2014 Dec-16 60-7075-103 3421071 12/18/2014 Dec-16 60-7075-103 3422228 12/18/2014 Dec-16 60-7075-104 3420711 12/18/2014 Dec-16 60-7075-106 3419300 12/18/2014 Dec-16 60-7075-106 3420674 12/18/2014 Dec-16 60-7075-106 3421194 12/18/2014 Dec-16 5921-018-235 3403639 12/19/2014 Dec-16 5921-024-135 3405257 12/19/2014 Dec-16 5921-024-235 3403660 12/19/2014 Dec-16 5921-024-235 3407382 12/19/2014 Dec-16 5921-030-135 3405652 12/19/2014 Dec-16 5921-030-235 3403312 12/19/2014 Dec-16 5921-030-235 3404916 12/19/2014 Dec-16 5921-030-235 3407209 12/19/2014 Dec-16 5921-030-235 3407526 12/19/2014 Dec-16 5921-034-135 3403267 12/19/2014 Dec-16 5921-034-135 3405815 12/19/2014 Dec-16 5921-034-135 3408146 12/19/2014 Dec-16 5921-034-235 3403144 12/19/2014 Dec-16 5921-034-235 3405112 12/19/2014 Dec-16 5921-034-235 3407162 12/19/2014 Dec-16 5921-034-235 3407975 12/19/2014 Dec-16 5921-034-235 3408024 12/19/2014 Dec-16 5921-218-235 3403383 12/19/2014 Dec-16 5921-218-235 3405584 12/19/2014 Dec-16 5921-218-235 3406682 12/19/2014 Dec-16 5921-218-235 3408105 12/19/2014 Dec-16 5921-218-135 3403570 12/19/2014 Dec-16 5921-218-135 3405940 12/19/2014 Dec-16 5921-218-135 3408271 12/19/2014 Dec-16 60-7070-103 3404021 12/19/2014 Dec-16 60-7070-103 3404614 12/19/2014 Dec-16 60-7070-103 3406075 12/19/2014 Dec-16 60-7070-103 3406522 12/19/2014 Dec-16 60-7070-105 3403185 12/19/2014 Dec-16 60-7070-105 3404176 12/19/2014 Dec-16 60-7070-105 3406994 12/19/2014 Dec-16 60-7070-105 3408011 12/19/2014 Dec-16 60-7070-104 3403171 12/19/2014 Dec-16 60-7070-104 3403504 12/19/2014 Dec-16 60-7070-104 3406109 12/19/2014 Dec-16 60-7070-104 3408263 12/19/2014 Dec-16 60-7070-003 3405773 12/19/2014 Dec-16 60-7070-106 3403799 12/19/2014 Dec-16 60-7070-106 3405290 12/19/2014 Dec-16 60-7070-106 3406816 12/19/2014 Dec-16 60-7070-106 3407018 12/19/2014 Dec-16 60-7070-106 3408547 12/19/2014 Dec-16 60-7070-107 3405513 12/19/2014 Dec-16 60-7070-107 3407640 12/19/2014 Dec-16 60-7075-003 3405606 12/19/2014 Dec-16 60-7075-006 3403331 12/19/2014 Dec-16 60-7075-103 3403527 12/19/2014 Dec-16 60-7075-103 3404992 12/19/2014 Dec-16 60-7075-103 3408409 12/19/2014 Dec-16 60-7075-104 3403515 12/19/2014 Dec-16 60-7075-106 3403280 12/19/2014 Dec-16 60-7075-106 3405196 12/19/2014 Dec-16 60-7075-106 3407416 12/19/2014 Dec-16 60-7075-106 3407746 12/19/2014 Dec-16 5921-018-135 3424887 12/22/2014 Dec-16 5921-018-135 3427946 12/22/2014 Dec-16 5921-024-235 3424388 12/22/2014 Dec-16 5921-024-235 3425985 12/22/2014 Dec-16 5921-024-235 3427784 12/22/2014 Dec-16 5921-030-235 3422387 12/22/2014 Dec-16 5921-030-235 3423981 12/22/2014 Dec-16 5921-030-235 3424511 12/22/2014 Dec-16 5921-030-235 3425971 12/22/2014 Dec-16 5921-030-235 3427064 12/22/2014 Dec-16 5921-030-235 3428454 12/22/2014 Dec-16 5921-034-135 3424326 12/22/2014 Dec-16 5921-034-135 3425066 12/22/2014 Dec-16 5921-034-135 3426235 12/22/2014 Dec-16 5921-034-135 3427415 12/22/2014 Dec-16 5921-034-135 3428289 12/22/2014 Dec-16 5921-034-235 3424054 12/22/2014 Dec-16 5921-034-235 3424645 12/22/2014 Dec-16 5921-034-235 3425370 12/22/2014 Dec-16 5921-034-235 3425447 12/22/2014 Dec-16 5921-034-235 3426261 12/22/2014 Dec-16 5921-034-235 3426862 12/22/2014 Dec-16 5921-034-235 3428364 12/22/2014 Dec-16 5921-218-235 3423877 12/22/2014 Dec-16 5921-218-235 3424553 12/22/2014 Dec-16 5921-218-235 3425045 12/22/2014 Dec-16 5921-218-235 3425622 12/22/2014 Dec-16 5921-218-235 3426022 12/22/2014 Dec-16 5921-218-235 3427518 12/22/2014 Dec-16 5921-030-236 3424818 12/22/2014 Dec-16 5921-218-135 3424873 12/22/2014 Dec-16 5921-218-135 3428340 12/22/2014 Dec-16 60-7070-103 3423018 12/22/2014 Dec-16 60-7070-103 3423574 12/22/2014 Dec-16 60-7070-103 3424152 12/22/2014 Dec-16 60-7070-103 3424238 12/22/2014 Dec-16 60-7070-103 3426183 12/22/2014 Dec-16 60-7070-103 3426371 12/22/2014 Dec-16 60-7070-103 3427560 12/22/2014 Dec-16 60-7070-103 3427671 12/22/2014 Dec-16 60-7070-105 3424134 12/22/2014 Dec-16 60-7070-105 3424970 12/22/2014 Dec-16 60-7070-105 3425410 12/22/2014 Dec-16 60-7070-105 3425916 12/22/2014 Dec-16 60-7070-105 3426337 12/22/2014 Dec-16 60-7070-105 3427347 12/22/2014 Dec-16 60-7070-104 3423561 12/22/2014 Dec-16 60-7070-104 3424962 12/22/2014 Dec-16 60-7070-104 3426448 12/22/2014 Dec-16 60-7070-106 3423638 12/22/2014 Dec-16 60-7070-106 3424792 12/22/2014 Dec-16 60-7070-106 3425218 12/22/2014 Dec-16 60-7070-106 3426316 12/22/2014 Dec-16 60-7070-106 3426794 12/22/2014 Dec-16 60-7070-106 3426929 12/22/2014 Dec-16 60-7070-106 3427161 12/22/2014 Dec-16 60-7070-107 3423779 12/22/2014 Dec-16 60-7075-103 3423849 12/22/2014 Dec-16 60-7075-103 3424915 12/22/2014 Dec-16 60-7075-103 3427882 12/22/2014 Dec-16 60-7075-103 3428204 12/22/2014 Dec-16 60-7075-104 3424357 12/22/2014 Dec-16 60-7075-106 3426363 12/22/2014 Dec-16 60-7075-106 3426681 12/22/2014 Dec-16 60-7075-106 3427637 12/22/2014 Dec-16 5921-018-135 3432071 12/23/2014 Dec-16 5921-024-135 3429656 12/23/2014 Dec-16 5921-024-235 3431838 12/23/2014 Dec-16 5921-030-135 3428598 12/23/2014 Dec-16 5921-030-135 3430989 12/23/2014 Dec-16 5921-030-135 3431082 12/23/2014 Dec-16 5921-030-135 3432398 12/23/2014 Dec-16 5921-030-235 3431356 12/23/2014 Dec-16 5921-034-135 3431007 12/23/2014 Dec-16 5921-034-135 3432118 12/23/2014 Dec-16 5921-034-235 3428551 12/23/2014 Dec-16 5921-034-235 3431259 12/23/2014 Dec-16 5921-034-235 3433177 12/23/2014 Dec-16 5921-218-235 3428408 12/23/2014 Dec-16 5921-218-235 3430008 12/23/2014 Dec-16 5921-218-235 3431631 12/23/2014 Dec-16 5921-218-135 3428956 12/23/2014 Dec-16 5921-218-135 3429312 12/23/2014 Dec-16 5921-218-135 3430915 12/23/2014 Dec-16 5921-218-236 3430106 12/23/2014 Dec-16 60-7070-102 3430049 12/23/2014 Dec-16 60-7070-103 3430286 12/23/2014 Dec-16 60-7070-103 3430763 12/23/2014 Dec-16 60-7070-103 3431095 12/23/2014 Dec-16 60-7070-103 3431506 12/23/2014 Dec-16 60-7070-103 3432037 12/23/2014 Dec-16 60-7070-105 3429515 12/23/2014 Dec-16 60-7070-105 3431105 12/23/2014 Dec-16 60-7070-105 3433165 12/23/2014 Dec-16 60-7070-006 3431523 12/23/2014 Dec-16 60-7070-006 3431727 12/23/2014 Dec-16 60-7070-007 3433196 12/23/2014 Dec-16 60-7070-106 3428837 12/23/2014 Dec-16 60-7070-106 3429144 12/23/2014 Dec-16 60-7070-106 3429279 12/23/2014 Dec-16 60-7070-106 3430856 12/23/2014 Dec-16 60-7070-106 3432474 12/23/2014 Dec-16 60-7070-106 3432731 12/23/2014 Dec-16 60-7075-006 3430152 12/23/2014 Dec-16 60-7075-103 3431801 12/23/2014 Dec-16 60-7075-106 3429976 12/23/2014 Dec-16 60-7075-106 3431612 12/23/2014 Dec-16 5921-018-135 3434912 12/29/2014 Dec-16 5921-024-235 3435214 12/29/2014 Dec-16 5921-030-235 3433424 12/29/2014 Dec-16 5921-030-235 3434186 12/29/2014 Dec-16 5921-030-235 3435019 12/29/2014 Dec-16 5921-030-235 3436705 12/29/2014 Dec-16 5921-030-235 3437546 12/29/2014 Dec-16 5921-034-135 3434350 12/29/2014 Dec-16 5921-034-135 3435866 12/29/2014 Dec-16 5921-034-135 3435969 12/29/2014 Dec-16 5921-034-135 3436767 12/29/2014 Dec-16 5921-034-135 3437372 12/29/2014 Dec-16 5921-034-235 3433359 12/29/2014 Dec-16 5921-034-235 3433938 12/29/2014 Dec-16 5921-034-235 3434688 12/29/2014 Dec-16 5921-034-235 3435543 12/29/2014 Dec-16 5921-034-235 3435694 12/29/2014 Dec-16 5921-034-235 3436221 12/29/2014 Dec-16 5921-034-235 3437047 12/29/2014 Dec-16 5921-034-235 3437386 12/29/2014 Dec-16 5921-034-236 3435349 12/29/2014 Dec-16 5921-218-235 3433371 12/29/2014 Dec-16 5921-218-235 3433955 12/29/2014 Dec-16 5921-218-235 3434164 12/29/2014 Dec-16 5921-218-235 3435163 12/29/2014 Dec-16 5921-218-235 3436191 12/29/2014 Dec-16 5921-044-235 3435799 12/29/2014 Dec-16 5921-218-135 3433702 12/29/2014 Dec-16 5921-218-135 3433804 12/29/2014 Dec-16 5921-218-136 3434500 12/29/2014 Dec-16 60-7070-103 3433550 12/29/2014 Dec-16 60-7070-103 3433917 12/29/2014 Dec-16 60-7070-103 3434898 12/29/2014 Dec-16 60-7070-103 3435493 12/29/2014 Dec-16 60-7070-103 3436018 12/29/2014 Dec-16 60-7070-103 3436104 12/29/2014 Dec-16 60-7070-103 3436307 12/29/2014 Dec-16 60-7070-103 3437204 12/29/2014 Dec-16 60-7070-103 3437757 12/29/2014 Dec-16 60-7070-103 3437994 12/29/2014 Dec-16 60-7070-105 3432875 12/29/2014 Dec-16 60-7070-105 3433901 12/29/2014 Dec-16 60-7070-105 3434321 12/29/2014 Dec-16 60-7070-105 3434996 12/29/2014 Dec-16 60-7070-105 3436170 12/29/2014 Dec-16 60-7070-105 3436250 12/29/2014 Dec-16 60-7070-105 3436820 12/29/2014 Dec-16 60-7070-105 3437289 12/29/2014 Dec-16 60-7070-104 3435710 12/29/2014 Dec-16 60-7070-006 3435308 12/29/2014 Dec-16 60-7070-106 3433718 12/29/2014 Dec-16 60-7070-106 3434234 12/29/2014 Dec-16 60-7070-106 3435090 12/29/2014 Dec-16 60-7070-106 3435747 12/29/2014 Dec-16 60-7070-106 3435953 12/29/2014 Dec-16 60-7070-106 3436535 12/29/2014 Dec-16 60-7070-106 3437676 12/29/2014 Dec-16 60-7070-107 3435897 12/29/2014 Dec-16 60-7075-103 3434846 12/29/2014 Dec-16 60-7075-103 3436353 12/29/2014 Dec-16 60-7075-103 3437607 12/29/2014 Dec-16 60-7075-103 3437648 12/29/2014 Dec-16 60-7075-104 3435845 12/29/2014 Dec-16 60-7075-106 3435205 12/29/2014 Dec-16 60-7075-106 3436266 12/29/2014 Dec-16 60-7075-107 3434494 12/29/2014 Dec-16 5921-018-135 3439423 12/30/2014 Dec-16 5921-018-235 3441718 12/30/2014 Dec-16 5921-024-135 3440726 12/30/2014 Dec-16 5921-030-135 3440607 12/30/2014 Dec-16 5921-030-135 3440624 12/30/2014 Dec-16 5921-030-235 3438265 12/30/2014 Dec-16 5921-030-235 3439698 12/30/2014 Dec-16 5921-030-235 3441521 12/30/2014 Dec-16 5921-034-135 3438103 12/30/2014 Dec-16 5921-034-135 3439944 12/30/2014 Dec-16 5921-034-135 3440774 12/30/2014 Dec-16 5921-034-235 3438001 12/30/2014 Dec-16 5921-034-235 3439153 12/30/2014 Dec-16 5921-034-235 3440800 12/30/2014 Dec-16 5921-034-235 3441005 12/30/2014 Dec-16 5921-034-235 3442159 12/30/2014 Dec-16 5921-034-236 3438188 12/30/2014 Dec-16 5921-218-235 3438231 12/30/2014 Dec-16 5921-218-235 3440006 12/30/2014 Dec-16 5921-218-235 3440859 12/30/2014 Dec-16 5921-218-235 3442308 12/30/2014 Dec-16 5921-218-135 3439225 12/30/2014 Dec-16 5921-218-135 3441891 12/30/2014 Dec-16 60-7070-103 3438086 12/30/2014 Dec-16 60-7070-103 3438283 12/30/2014 Dec-16 60-7070-103 3440132 12/30/2014 Dec-16 60-7070-103 3442440 12/30/2014 Dec-16 60-7070-105 3438064 12/30/2014 Dec-16 60-7070-105 3439729 12/30/2014 Dec-16 60-7070-105 3441870 12/30/2014 Dec-16 60-7070-104 3438306 12/30/2014 Dec-16 60-7070-104 3438573 12/30/2014 Dec-16 60-7070-104 3439362 12/30/2014 Dec-16 60-7070-104 3441610 12/30/2014 Dec-16 60-7070-003 3442312 12/30/2014 Dec-16 60-7070-106 3438072 12/30/2014 Dec-16 60-7070-106 3438778 12/30/2014 Dec-16 60-7070-106 3438895 12/30/2014 Dec-16 60-7070-106 3441149 12/30/2014 Dec-16 60-7070-106 3441257 12/30/2014 Dec-16 60-7070-106 3441430 12/30/2014 Dec-16 60-7070-107 3439218 12/30/2014 Dec-16 60-7075-003 3439963 12/30/2014 Dec-16 60-7075-006 3442061 12/30/2014 Dec-16 60-7075-006 3442214 12/30/2014 Dec-16 60-7075-103 3439679 12/30/2014 Dec-16 60-7075-106 3438190 12/30/2014 Dec-16 60-7075-106 3439468 12/30/2014 Dec-16 5921-018-135 3445745 1/8/2015 Jan-17 5921-030-235 3443737 1/8/2015 Jan-17 5921-030-235 3444975 1/8/2015 Jan-17 5921-030-235 3446037 1/8/2015 Jan-17 5921-034-135 3444191 1/8/2015 Jan-17 5921-034-135 3445265 1/8/2015 Jan-17 5921-034-135 3445296 1/8/2015 Jan-17 5921-034-235 3443613 1/8/2015 Jan-17 5921-034-235 3444808 1/8/2015 Jan-17 5921-034-235 3445208 1/8/2015 Jan-17 5921-034-235 3445628 1/8/2015 Jan-17 5921-218-235 3443319 1/8/2015 Jan-17 5921-218-235 3444382 1/8/2015 Jan-17 5921-218-235 3445979 1/8/2015 Jan-17 60-7070-103 3442472 1/8/2015 Jan-17 60-7070-103 3442809 1/8/2015 Jan-17 60-7070-103 3443943 1/8/2015 Jan-17 60-7070-103 3445055 1/8/2015 Jan-17 60-7070-103 3446063 1/8/2015 Jan-17 60-7070-103 3446173 1/8/2015 Jan-17 60-7070-105 3442821 1/8/2015 Jan-17 60-7070-105 3443083 1/8/2015 Jan-17 60-7070-105 3443792 1/8/2015 Jan-17 60-7070-105 3444444 1/8/2015 Jan-17 60-7070-105 3444463 1/8/2015 Jan-17 60-7070-104 3445072 1/8/2015 Jan-17 60-7070-104 3445882 1/8/2015 Jan-17 60-7070-106 3443303 1/8/2015 Jan-17 60-7070-106 3443390 1/8/2015 Jan-17 60-7070-106 3443576 1/8/2015 Jan-17 60-7070-106 3444492 1/8/2015 Jan-17 60-7070-106 3444780 1/8/2015 Jan-17 60-7070-106 3444877 1/8/2015 Jan-17 60-7070-106 3445404 1/8/2015 Jan-17 60-7070-106 3445518 1/8/2015 Jan-17 60-7070-106 3445600 1/8/2015 Jan-17 60-7070-107 3442832 1/8/2015 Jan-17 60-7075-103 3443020 1/8/2015 Jan-17 60-7075-103 3444311 1/8/2015 Jan-17 60-7075-106 3442845 1/8/2015 Jan-17 5921-018-235 3450654 1/14/2015 Jan-17 5921-024-235 3450715 1/14/2015 Jan-17 5921-030-135 3450497 1/14/2015 Jan-17 5921-030-135 3450506 1/14/2015 Jan-17 5921-030-235 3447343 1/14/2015 Jan-17 5921-030-235 3447554 1/14/2015 Jan-17 5921-030-235 3448500 1/14/2015 Jan-17 5921-030-235 3450263 1/14/2015 Jan-17 5921-034-135 3447358 1/14/2015 Jan-17 5921-034-135 3448879 1/14/2015 Jan-17 5921-034-235 3447835 1/14/2015 Jan-17 5921-034-235 3448619 1/14/2015 Jan-17 5921-034-235 3450413 1/14/2015 Jan-17 5921-034-236 3447857 1/14/2015 Jan-17 5921-218-235 3446708 1/14/2015 Jan-17 5921-218-235 3449373 1/14/2015 Jan-17 5921-218-235 3449654 1/14/2015 Jan-17 5921-044-135 3447861 1/14/2015 Jan-17 5921-044-235 3449292 1/14/2015 Jan-17 5921-218-135 3446442 1/14/2015 Jan-17 5921-218-135 3447698 1/14/2015 Jan-17 60-7070-103 3447499 1/14/2015 Jan-17 60-7070-103 3447937 1/14/2015 Jan-17 60-7070-103 3448425 1/14/2015 Jan-17 60-7070-103 3448983 1/14/2015 Jan-17 60-7070-103 3449886 1/14/2015 Jan-17 60-7070-105 3446322 1/14/2015 Jan-17 60-7070-105 3449076 1/14/2015 Jan-17 60-7070-105 3449318 1/14/2015 Jan-17 60-7070-104 3448657 1/14/2015 Jan-17 60-7070-104 3450095 1/14/2015 Jan-17 60-7070-003 3450770 1/14/2015 Jan-17 60-7070-006 3446658 1/14/2015 Jan-17 60-7070-006 3449360 1/14/2015 Jan-17 60-7070-106 3446720 1/14/2015 Jan-17 60-7070-106 3447241 1/14/2015 Jan-17 60-7070-106 3447362 1/14/2015 Jan-17 60-7070-106 3447826 1/14/2015 Jan-17 60-7070-106 3447916 1/14/2015 Jan-17 60-7070-106 3448186 1/14/2015 Jan-17 60-7070-106 3448234 1/14/2015 Jan-17 60-7070-106 3449770 1/14/2015 Jan-17 60-7070-106 3450893 1/14/2015 Jan-17 60-7070-107 3446419 1/14/2015 Jan-17 60-7075-003 3447508 1/14/2015 Jan-17 60-7075-006 3446796 1/14/2015 Jan-17 60-7075-103 3448851 1/14/2015 Jan-17 60-7075-103 3449631 1/14/2015 Jan-17 60-7075-103 3449743 1/14/2015 Jan-17 60-7075-104 3446765 1/14/2015 Jan-17 60-7075-106 3448833 1/14/2015 Jan-17 5921-018-135 3450903 1/15/2015 Jan-17 5921-018-135 3455041 1/15/2015 Jan-17 5921-018-235 3454312 1/15/2015 Jan-17 5921-024-135 3451496 1/15/2015 Jan-17 5921-024-235 3452384 1/15/2015 Jan-17 5921-030-135 3454813 1/15/2015 Jan-17 5921-030-235 3453013 1/15/2015 Jan-17 5921-034-135 3451546 1/15/2015 Jan-17 5921-034-135 3454768 1/15/2015 Jan-17 5921-034-135 3454982 1/15/2015 Jan-17 5921-034-235 3451956 1/15/2015 Jan-17 5921-034-235 3453701 1/15/2015 Jan-17 5921-034-235 3454464 1/15/2015 Jan-17 5921-034-235 3455266 1/15/2015 Jan-17 5921-034-235 3455574 1/15/2015 Jan-17 5921-218-235 3452210 1/15/2015 Jan-17 5921-218-235 3453520 1/15/2015 Jan-17 5921-218-235 3454100 1/15/2015 Jan-17 5921-218-135 3451454 1/15/2015 Jan-17 5921-218-135 3454591 1/15/2015 Jan-17 60-7070-103 3451607 1/15/2015 Jan-17 60-7070-103 3451973 1/15/2015 Jan-17 60-7070-103 3453343 1/15/2015 Jan-17 60-7070-103 3453410 1/15/2015 Jan-17 60-7070-103 3455039 1/15/2015 Jan-17 60-7070-105 3451258 1/15/2015 Jan-17 60-7070-105 3452004 1/15/2015 Jan-17 60-7070-105 3455136 1/15/2015 Jan-17 60-7070-105 3455221 1/15/2015 Jan-17 60-7070-105 3455767 1/15/2015 Jan-17 60-7070-104 3453772 1/15/2015 Jan-17 60-7070-104 3454026 1/15/2015 Jan-17 60-7070-006 3454723 1/15/2015 Jan-17 60-7070-106 3452036 1/15/2015 Jan-17 60-7070-106 3453481 1/15/2015 Jan-17 60-7070-106 3454995 1/15/2015 Jan-17 60-7070-107 3453078 1/15/2015 Jan-17 60-7070-107 3454536 1/15/2015 Jan-17 60-7075-003 3454455 1/15/2015 Jan-17 60-7075-006 3453992 1/15/2015 Jan-17 60-7075-103 3452460 1/15/2015 Jan-17 60-7075-103 3454493 1/15/2015 Jan-17 60-7075-103 3455324 1/15/2015 Jan-17 60-7075-104 3453239 1/15/2015 Jan-17 60-7075-105 3451987 1/15/2015 Jan-17 60-7075-106 3452117 1/15/2015 Jan-17 60-7075-106 3453819 1/15/2015 Jan-17 60-7075-106 3454267 1/15/2015 Jan-17 5921-018-135 3456905 1/19/2015 Jan-17 5921-024-135 3459388 1/19/2015 Jan-17 5921-024-235 3458823 1/19/2015 Jan-17 5921-030-135 3456745 1/19/2015 Jan-17 5921-030-235 3455896 1/19/2015 Jan-17 5921-030-235 3458426 1/19/2015 Jan-17 5921-030-235 3459958 1/19/2015 Jan-17 5921-030-235 3461140 1/19/2015 Jan-17 5921-034-135 3456876 1/19/2015 Jan-17 5921-034-135 3457407 1/19/2015 Jan-17 5921-034-235 3457392 1/19/2015 Jan-17 5921-034-235 3458269 1/19/2015 Jan-17 5921-034-235 3458419 1/19/2015 Jan-17 5921-034-235 3460251 1/19/2015 Jan-17 5921-218-235 3456142 1/19/2015 Jan-17 5921-218-235 3458063 1/19/2015 Jan-17 5921-218-235 3459181 1/19/2015 Jan-17 5921-218-235 3460042 1/19/2015 Jan-17 5921-218-135 3456489 1/19/2015 Jan-17 5921-218-135 3458708 1/19/2015 Jan-17 60-7070-103 3456807 1/19/2015 Jan-17 60-7070-103 3457453 1/19/2015 Jan-17 60-7070-103 3457964 1/19/2015 Jan-17 60-7070-103 3459303 1/19/2015 Jan-17 60-7070-105 3455865 1/19/2015 Jan-17 60-7070-105 3457385 1/19/2015 Jan-17 60-7070-104 3456086 1/19/2015 Jan-17 60-7070-104 3459943 1/19/2015 Jan-17 60-7070-003 3456922 1/19/2015 Jan-17 60-7070-006 3460061 1/19/2015 Jan-17 60-7070-106 3456375 1/19/2015 Jan-17 60-7070-106 3457411 1/19/2015 Jan-17 60-7070-106 3457972 1/19/2015 Jan-17 60-7070-106 3459106 1/19/2015 Jan-17 60-7070-106 3460312 1/19/2015 Jan-17 60-7070-106 3460878 1/19/2015 Jan-17 60-7070-106 3461089 1/19/2015 Jan-17 60-7070-107 3456946 1/19/2015 Jan-17 60-7070-107 3460057 1/19/2015 Jan-17 60-7075-003 3459498 1/19/2015 Jan-17 60-7075-003 3459792 1/19/2015 Jan-17 60-7075-006 3456797 1/19/2015 Jan-17 60-7075-103 3456254 1/19/2015 Jan-17 60-7075-103 3458731 1/19/2015 Jan-17 60-7075-103 3458966 1/19/2015 Jan-17 60-7075-106 3456518 1/19/2015 Jan-17 60-7075-106 3458636 1/19/2015 Jan-17 5921-018-135 3464180 1/20/2015 Jan-17 5921-018-235 3461748 1/20/2015 Jan-17 5921-024-235 3461782 1/20/2015 Jan-17 5921-024-235 3464762 1/20/2015 Jan-17 5921-030-135 3462010 1/20/2015 Jan-17 5921-030-135 3464360 1/20/2015 Jan-17 5921-030-136 3462956 1/20/2015 Jan-17 5921-030-235 3461487 1/20/2015 Jan-17 5921-030-235 3463882 1/20/2015 Jan-17 5921-030-235 3464105 1/20/2015 Jan-17 5921-030-235 3465569 1/20/2015 Jan-17 5921-034-135 3462757 1/20/2015 Jan-17 5921-034-135 3463732 1/20/2015 Jan-17 5921-034-135 3466287 1/20/2015 Jan-17 5921-034-235 3462317 1/20/2015 Jan-17 5921-034-235 3462607 1/20/2015 Jan-17 5921-034-235 3462735 1/20/2015 Jan-17 5921-212-235 3465286 1/20/2015 Jan-17 5921-218-235 3461313 1/20/2015 Jan-17 5921-218-235 3462579 1/20/2015 Jan-17 5921-218-135 3464523 1/20/2015 Jan-17 5921-218-135 3466343 1/20/2015 Jan-17 60-7070-103 3462150 1/20/2015 Jan-17 60-7070-103 3462496 1/20/2015 Jan-17 60-7070-103 3463443 1/20/2015 Jan-17 60-7070-103 3465106 1/20/2015 Jan-17 60-7070-103 3465374 1/20/2015 Jan-17 60-7070-105 3462034 1/20/2015 Jan-17 60-7070-105 3463876 1/20/2015 Jan-17 60-7070-105 3463933 1/20/2015 Jan-17 60-7070-105 3465697 1/20/2015 Jan-17 60-7070-104 3461456 1/20/2015 Jan-17 60-7070-104 3464665 1/20/2015 Jan-17 60-7070-104 3465020 1/20/2015 Jan-17 60-7070-003 3461805 1/20/2015 Jan-17 60-7070-006 3464450 1/20/2015 Jan-17 60-7070-106 3462669 1/20/2015 Jan-17 60-7070-106 3463103 1/20/2015 Jan-17 60-7070-106 3463249 1/20/2015 Jan-17 60-7070-106 3465217 1/20/2015 Jan-17 60-7070-106 3465802 1/20/2015 Jan-17 60-7070-106 3466009 1/20/2015 Jan-17 60-7070-106 3466084 1/20/2015 Jan-17 60-7075-006 3464631 1/20/2015 Jan-17 60-7075-103 3463951 1/20/2015 Jan-17 60-7075-104 3461908 1/20/2015 Jan-17 60-7075-106 3464216 1/20/2015 Jan-17 60-7075-106 3465511 1/20/2015 Jan-17 5921-018-135 3470685 1/22/2015 Jan-17 5921-018-235 3466549 1/22/2015 Jan-17 5921-018-235 3470895 1/22/2015 Jan-17 5921-018-235 3471039 1/22/2015 Jan-17 5921-024-135 3468219 1/22/2015 Jan-17 5921-024-235 3467224 1/22/2015 Jan-17 5921-024-235 3469779 1/22/2015 Jan-17 5921-030-135 3467921 1/22/2015 Jan-17 5921-030-235 3467063 1/22/2015 Jan-17 5921-030-235 3467754 1/22/2015 Jan-17 5921-030-235 3469284 1/22/2015 Jan-17 5921-030-235 3469297 1/22/2015 Jan-17 5921-034-135 3468741 1/22/2015 Jan-17 5921-034-135 3470142 1/22/2015 Jan-17 5921-034-135 3470310 1/22/2015 Jan-17 5921-034-235 3467253 1/22/2015 Jan-17 5921-034-235 3468702 1/22/2015 Jan-17 5921-034-235 3470811 1/22/2015 Jan-17 5921-218-235 3466861 1/22/2015 Jan-17 5921-218-235 3470283 1/22/2015 Jan-17 5921-212-135 3467194 1/22/2015 Jan-17 5921-218-135 3468196 1/22/2015 Jan-17 5921-218-135 3469143 1/22/2015 Jan-17 60-7070-103 3466577 1/22/2015 Jan-17 60-7070-103 3468344 1/22/2015 Jan-17 60-7070-103 3469136 1/22/2015 Jan-17 60-7070-103 3469638 1/22/2015 Jan-17 60-7070-105 3467749 1/22/2015 Jan-17 60-7070-105 3468550 1/22/2015 Jan-17 60-7070-105 3468623 1/22/2015 Jan-17 60-7070-105 3468634 1/22/2015 Jan-17 60-7070-105 3470103 1/22/2015 Jan-17 60-7070-104 3466362 1/22/2015 Jan-17 60-7070-104 3468739 1/22/2015 Jan-17 60-7070-003 3467868 1/22/2015 Jan-17 60-7070-006 3470709 1/22/2015 Jan-17 60-7070-106 3468891 1/22/2015 Jan-17 60-7070-106 3471104 1/22/2015 Jan-17 60-7070-107 3468022 1/22/2015 Jan-17 60-7070-107 3468033 1/22/2015 Jan-17 60-7075-003 3466959 1/22/2015 Jan-17 60-7075-004 3469751 1/22/2015 Jan-17 60-7075-103 3467806 1/22/2015 Jan-17 60-7075-104 3470587 1/22/2015 Jan-17 60-7075-104 3470745 1/22/2015 Jan-17 60-7075-106 3466963 1/22/2015 Jan-17 60-7075-106 3467141 1/22/2015 Jan-17 60-7075-106 3470502 1/22/2015 Jan-17 5921-018-135 3474261 1/26/2015 Jan-17 5921-024-135 3473619 1/26/2015 Jan-17 5921-024-235 3471232 1/26/2015 Jan-17 5921-030-135 3471865 1/26/2015 Jan-17 5921-030-135 3473795 1/26/2015 Jan-17 5921-030-235 3473439 1/26/2015 Jan-17 5921-034-135 3473776 1/26/2015 Jan-17 5921-034-136 3471746 1/26/2015 Jan-17 5921-034-235 3471519 1/26/2015 Jan-17 5921-034-235 3472888 1/26/2015 Jan-17 5921-034-235 3473657 1/26/2015 Jan-17 5921-034-235 3474993 1/26/2015 Jan-17 5921-218-235 3471588 1/26/2015 Jan-17 5921-218-235 3473563 1/26/2015 Jan-17 5921-218-235 3475258 1/26/2015 Jan-17 5921-030-236 3474274 1/26/2015 Jan-17 5921-218-135 3472050 1/26/2015 Jan-17 5921-218-135 3472452 1/26/2015 Jan-17 5921-218-135 3474509 1/26/2015 Jan-17 60-7070-103 3469999 1/26/2015 Jan-17 60-7070-103 3472793 1/26/2015 Jan-17 60-7070-103 3473223 1/26/2015 Jan-17 60-7070-105 3471981 1/26/2015 Jan-17 60-7070-105 3472971 1/26/2015 Jan-17 60-7070-105 3472985 1/26/2015 Jan-17 60-7070-105 3473748 1/26/2015 Jan-17 60-7070-105 3475178 1/26/2015 Jan-17 60-7070-104 3471877 1/26/2015 Jan-17 60-7070-104 3474476 1/26/2015 Jan-17 60-7070-104 3474901 1/26/2015 Jan-17 60-7070-003 3473559 1/26/2015 Jan-17 60-7070-106 3471428 1/26/2015 Jan-17 60-7070-106 3472646 1/26/2015 Jan-17 60-7070-106 3472740 1/26/2015 Jan-17 60-7070-106 3474940 1/26/2015 Jan-17 60-7070-107 3471601 1/26/2015 Jan-17 60-7070-107 3474079 1/26/2015 Jan-17 60-7075-003 3471250 1/26/2015 Jan-17 60-7075-005 3472066 1/26/2015 Jan-17 60-7075-006 3471722 1/26/2015 Jan-17 60-7075-103 3472520 1/26/2015 Jan-17 60-7075-103 3473473 1/26/2015 Jan-17 60-7075-105 3473585 1/26/2015 Jan-17 60-7075-106 3472124 1/26/2015 Jan-17 60-7075-106 3474513 1/26/2015 Jan-17 5921-018-135 3478467 1/27/2015 Jan-17 5921-018-235 3476024 1/27/2015 Jan-17 5921-018-235 3478792 1/27/2015 Jan-17 5921-024-235 3476926 1/27/2015 Jan-17 5921-024-235 3477982 1/27/2015 Jan-17 5921-030-135 3476030 1/27/2015 Jan-17 5921-030-135 3476552 1/27/2015 Jan-17 5921-030-135 3479125 1/27/2015 Jan-17 5921-030-136 3476665 1/27/2015 Jan-17 5921-030-235 3475386 1/27/2015 Jan-17 5921-030-235 3475703 1/27/2015 Jan-17 5921-030-235 3477821 1/27/2015 Jan-17 5921-030-235 3477911 1/27/2015 Jan-17 5921-034-135 3475598 1/27/2015 Jan-17 5921-034-135 3476654 1/27/2015 Jan-17 5921-034-135 3477331 1/27/2015 Jan-17 5921-034-135 3478319 1/27/2015 Jan-17 5921-034-135 3478511 1/27/2015 Jan-17 5921-034-235 3476285 1/27/2015 Jan-17 5921-034-235 3477527 1/27/2015 Jan-17 5921-034-235 3477562 1/27/2015 Jan-17 5921-218-235 3476432 1/27/2015 Jan-17 5921-218-235 3478077 1/27/2015 Jan-17 5921-218-135 3476095 1/27/2015 Jan-17 5921-218-135 3476957 1/27/2015 Jan-17 5921-218-135 3477383 1/27/2015 Jan-17 60-7070-103 3476855 1/27/2015 Jan-17 60-7070-103 3477734 1/27/2015 Jan-17 60-7070-103 3477775 1/27/2015 Jan-17 60-7070-103 3478744 1/27/2015 Jan-17 60-7070-105 3475483 1/27/2015 Jan-17 60-7070-105 3475987 1/27/2015 Jan-17 60-7070-105 3476179 1/27/2015 Jan-17 60-7070-105 3476961 1/27/2015 Jan-17 60-7070-105 3477429 1/27/2015 Jan-17 60-7070-105 3478031 1/27/2015 Jan-17 60-7070-105 3478700 1/27/2015 Jan-17 60-7070-104 3476817 1/27/2015 Jan-17 60-7070-104 3478991 1/27/2015 Jan-17 60-7070-003 3476263 1/27/2015 Jan-17 60-7070-003 3478229 1/27/2015 Jan-17 60-7070-006 3478293 1/27/2015 Jan-17 60-7070-106 3474067 1/27/2015 Jan-17 60-7070-106 3474843 1/27/2015 Jan-17 60-7070-106 3475941 1/27/2015 Jan-17 60-7070-106 3476360 1/27/2015 Jan-17 60-7070-106 3477543 1/27/2015 Jan-17 60-7075-003 3478576 1/27/2015 Jan-17 60-7075-004 3476506 1/27/2015 Jan-17 60-7075-005 3476510 1/27/2015 Jan-17 60-7075-006 3475648 1/27/2015 Jan-17 60-7075-006 3478582 1/27/2015 Jan-17 60-7080-102 3476649 1/27/2015 Jan-17 60-7075-103 3475669 1/27/2015 Jan-17 60-7075-103 3475960 1/27/2015 Jan-17 60-7075-103 3478595 1/27/2015 Jan-17 60-7075-104 3476421 1/27/2015 Jan-17 60-7075-104 3478856 1/27/2015 Jan-17 60-7075-105 3476146 1/27/2015 Jan-17 60-7075-106 3476237 1/27/2015 Jan-17 60-7075-106 3477641 1/27/2015 Jan-17 60-7075-106 3478553 1/27/2015 Jan-17 60-7070-103 3475034 1/28/2015 Jan-17 5921-018-135 3475465 1/29/2015 Jan-17 5921-018-235 3481747 1/29/2015 Jan-17 5921-024-135 3481438 1/29/2015 Jan-17 5921-024-135 3482545 1/29/2015 Jan-17 5921-024-235 3479380 1/29/2015 Jan-17 5921-024-235 3482756 1/29/2015 Jan-17 5921-030-135 3481472 1/29/2015 Jan-17 5921-030-135 3482196 1/29/2015 Jan-17 5921-030-235 3479276 1/29/2015 Jan-17 5921-030-235 3479913 1/29/2015 Jan-17 5921-030-235 3480896 1/29/2015 Jan-17 5921-030-235 3482652 1/29/2015 Jan-17 5921-030-235 3483367 1/29/2015 Jan-17 5921-034-135 3481233 1/29/2015 Jan-17 5921-034-135 3483806 1/29/2015 Jan-17 5921-034-136 3482858 1/29/2015 Jan-17 5921-034-235 3479156 1/29/2015 Jan-17 5921-034-235 3479691 1/29/2015 Jan-17 5921-034-235 3480039 1/29/2015 Jan-17 5921-034-235 3481003 1/29/2015 Jan-17 5921-034-235 3483224 1/29/2015 Jan-17 5921-218-235 3480191 1/29/2015 Jan-17 5921-218-235 3481171 1/29/2015 Jan-17 5921-218-235 3483708 1/29/2015 Jan-17 5921-044-235 3479398 1/29/2015 Jan-17 5921-212-135 3481639 1/29/2015 Jan-17 5921-218-135 3468098 1/29/2015 Jan-17 5921-218-135 3483909 1/29/2015 Jan-17 60-7070-103 3479282 1/29/2015 Jan-17 60-7070-103 3479847 1/29/2015 Jan-17 60-7070-103 3480640 1/29/2015 Jan-17 60-7070-103 3480733 1/29/2015 Jan-17 60-7070-103 3482739 1/29/2015 Jan-17 60-7070-103 3483092 1/29/2015 Jan-17 60-7070-103 3483314 1/29/2015 Jan-17 60-7070-105 3479461 1/29/2015 Jan-17 60-7070-105 3481214 1/29/2015 Jan-17 60-7070-105 3481995 1/29/2015 Jan-17 60-7070-105 3482699 1/29/2015 Jan-17 60-7070-105 3482702 1/29/2015 Jan-17 60-7070-105 3482815 1/29/2015 Jan-17 60-7070-105 3483517 1/29/2015 Jan-17 60-7070-104 3479474 1/29/2015 Jan-17 60-7070-104 3481591 1/29/2015 Jan-17 60-7070-104 3482993 1/29/2015 Jan-17 60-7070-003 3481884 1/29/2015 Jan-17 60-7070-006 3483875 1/29/2015 Jan-17 60-7070-106 3478962 1/29/2015 Jan-17 60-7070-106 3479834 1/29/2015 Jan-17 60-7070-106 3481026 1/29/2015 Jan-17 60-7070-106 3481349 1/29/2015 Jan-17 60-7070-106 3482964 1/29/2015 Jan-17 60-7070-106 3482972 1/29/2015 Jan-17 60-7070-106 3483662 1/29/2015 Jan-17 60-7070-107 3478989 1/29/2015 Jan-17 60-7070-107 3480087 1/29/2015 Jan-17 60-7075-003 3483419 1/29/2015 Jan-17 60-7075-004 3481455 1/29/2015 Jan-17 60-7075-005 3481723 1/29/2015 Jan-17 60-7075-006 3482804 1/29/2015 Jan-17 60-7080-102 3478839 1/29/2015 Jan-17 60-7075-103 3479141 1/29/2015 Jan-17 60-7075-103 3480213 1/29/2015 Jan-17 60-7075-103 3483253 1/29/2015 Jan-17 60-7075-104 3483398 1/29/2015 Jan-17 60-7075-104 3483689 1/29/2015 Jan-17 60-7075-105 3480664 1/29/2015 Jan-17 60-7075-105 3480686 1/29/2015 Jan-17 60-7075-105 3483118 1/29/2015 Jan-17 60-7075-106 3479636 1/29/2015 Jan-17 60-7075-106 3483125 1/29/2015 Jan-17 5921-018-135 3485292 2/2/2015 Feb-17 5921-018-136 3484703 2/2/2015 Feb-17 5921-018-235 3485302 2/2/2015 Feb-17 5921-024-135 3484068 2/2/2015 Feb-17 5921-024-235 3484178 2/2/2015 Feb-17 5921-024-235 3486609 2/2/2015 Feb-17 5921-030-135 3484047 2/2/2015 Feb-17 5921-030-135 3486286 2/2/2015 Feb-17 5921-030-135 3487138 2/2/2015 Feb-17 5921-030-235 3485271 2/2/2015 Feb-17 5921-030-235 3486134 2/2/2015 Feb-17 5921-034-135 3484243 2/2/2015 Feb-17 5921-034-135 3485285 2/2/2015 Feb-17 5921-034-135 3486293 2/2/2015 Feb-17 5921-034-235 3485228 2/2/2015 Feb-17 5921-034-235 3485677 2/2/2015 Feb-17 5921-034-235 3486621 2/2/2015 Feb-17 5921-218-235 3484607 2/2/2015 Feb-17 5921-218-235 3485216 2/2/2015 Feb-17 5921-218-235 3485432 2/2/2015 Feb-17 5921-218-235 3485581 2/2/2015 Feb-17 5921-218-235 3486123 2/2/2015 Feb-17 5921-030-236 3484010 2/2/2015 Feb-17 5921-212-135 3484291 2/2/2015 Feb-17 5921-218-135 3483966 2/2/2015 Feb-17 5921-218-135 3486759 2/2/2015 Feb-17 60-7070-103 3482108 2/2/2015 Feb-17 60-7070-103 3483921 2/2/2015 Feb-17 60-7070-103 3484579 2/2/2015 Feb-17 60-7070-103 3485360 2/2/2015 Feb-17 60-7070-103 3485762 2/2/2015 Feb-17 60-7070-103 3486405 2/2/2015 Feb-17 60-7070-103 3487058 2/2/2015 Feb-17 60-7070-103 3487563 2/2/2015 Feb-17 60-7070-103 3487619 2/2/2015 Feb-17 60-7070-105 3484514 2/2/2015 Feb-17 60-7070-105 3485356 2/2/2015 Feb-17 60-7070-105 3485631 2/2/2015 Feb-17 60-7070-105 3486077 2/2/2015 Feb-17 60-7070-105 3487043 2/2/2015 Feb-17 60-7070-104 3484317 2/2/2015 Feb-17 60-7070-104 3485770 2/2/2015 Feb-17 60-7070-104 3486175 2/2/2015 Feb-17 60-7070-003 3484695 2/2/2015 Feb-17 60-7070-106 3484522 2/2/2015 Feb-17 60-7070-106 3484885 2/2/2015 Feb-17 60-7070-106 3485620 2/2/2015 Feb-17 60-7070-106 3485736 2/2/2015 Feb-17 60-7070-106 3486414 2/2/2015 Feb-17 60-7070-106 3486860 2/2/2015 Feb-17 60-7070-106 3487275 2/2/2015 Feb-17 60-7070-107 3461299 2/2/2015 Feb-17 60-7070-107 3485387 2/2/2015 Feb-17 60-7075-003 3484204 2/2/2015 Feb-17 60-7075-006 3486771 2/2/2015 Feb-17 60-7080-102 3483847 2/2/2015 Feb-17 60-7075-103 3484580 2/2/2015 Feb-17 60-7075-103 3485864 2/2/2015 Feb-17 60-7075-103 3486054 2/2/2015 Feb-17 60-7075-104 3486802 2/2/2015 Feb-17 60-7075-105 3484166 2/2/2015 Feb-17 60-7075-105 3486152 2/2/2015 Feb-17 60-7075-106 3484270 2/2/2015 Feb-17 60-7075-106 3485373 2/2/2015 Feb-17 60-7075-106 3485886 2/2/2015 Feb-17 5921-018-135 3488622 2/4/2015 Feb-17 5921-024-235 3488045 2/4/2015 Feb-17 5921-024-235 3488434 2/4/2015 Feb-17 5921-024-235 3489732 2/4/2015 Feb-17 5921-024-235 3491050 2/4/2015 Feb-17 5921-030-135 3488605 2/4/2015 Feb-17 5921-030-135 3490385 2/4/2015 Feb-17 5921-030-135 3491481 2/4/2015 Feb-17 5921-030-235 3488021 2/4/2015 Feb-17 5921-030-235 3488717 2/4/2015 Feb-17 5921-030-235 3491218 2/4/2015 Feb-17 5921-030-235 3491680 2/4/2015 Feb-17 5921-034-135 3487880 2/4/2015 Feb-17 5921-034-135 3488679 2/4/2015 Feb-17 5921-034-135 3490257 2/4/2015 Feb-17 5921-034-135 3491391 2/4/2015 Feb-17 5921-034-235 3488050 2/4/2015 Feb-17 5921-034-235 3488239 2/4/2015 Feb-17 5921-034-235 3489055 2/4/2015 Feb-17 5921-034-235 3489296 2/4/2015 Feb-17 5921-034-235 3489352 2/4/2015 Feb-17 5921-034-235 3490606 2/4/2015 Feb-17 5921-034-235 3491370 2/4/2015 Feb-17 5921-218-235 3487833 2/4/2015 Feb-17 5921-218-235 3488900 2/4/2015 Feb-17 5921-218-235 3490300 2/4/2015 Feb-17 5921-218-235 3491202 2/4/2015 Feb-17 5921-218-135 3487234 2/4/2015 Feb-17 5921-218-135 3488698 2/4/2015 Feb-17 5921-218-135 3490521 2/4/2015 Feb-17 60-7070-103 3488499 2/4/2015 Feb-17 60-7070-103 3488963 2/4/2015 Feb-17 60-7070-103 3489600 2/4/2015 Feb-17 60-7070-103 3490203 2/4/2015 Feb-17 60-7070-103 3490550 2/4/2015 Feb-17 60-7070-103 3491021 2/4/2015 Feb-17 60-7070-103 3491294 2/4/2015 Feb-17 60-7070-105 3487782 2/4/2015 Feb-17 60-7070-105 3488512 2/4/2015 Feb-17 60-7070-105 3489064 2/4/2015 Feb-17 60-7070-105 3489811 2/4/2015 Feb-17 60-7070-105 3490108 2/4/2015 Feb-17 60-7070-105 3490929 2/4/2015 Feb-17 60-7070-105 3491434 2/4/2015 Feb-17 60-7070-104 3488032 2/4/2015 Feb-17 60-7070-104 3489745 2/4/2015 Feb-17 60-7070-104 3491256 2/4/2015 Feb-17 60-7070-003 3487920 2/4/2015 Feb-17 60-7070-106 3487688 2/4/2015 Feb-17 60-7070-106 3488520 2/4/2015 Feb-17 60-7070-106 3488985 2/4/2015 Feb-17 60-7070-106 3489323 2/4/2015 Feb-17 60-7070-106 3490584 2/4/2015 Feb-17 60-7070-106 3491182 2/4/2015 Feb-17 60-7075-003 3488583 2/4/2015 Feb-17 60-7075-003 3490411 2/4/2015 Feb-17 60-7075-006 3490407 2/4/2015 Feb-17 60-7075-103 3488549 2/4/2015 Feb-17 60-7075-103 3489334 2/4/2015 Feb-17 60-7075-103 3490337 2/4/2015 Feb-17 60-7075-105 3490261 2/4/2015 Feb-17 60-7075-106 3487912 2/4/2015 Feb-17 60-7075-106 3488148 2/4/2015 Feb-17 60-7075-106 3489249 2/4/2015 Feb-17 60-7075-106 3490843 2/4/2015 Feb-17 60-7075-106 3491531 2/4/2015 Feb-17 5921-018-135 3492041 2/6/2015 Feb-17 5921-018-135 3493665 2/6/2015 Feb-17 5921-018-235 3495295 2/6/2015 Feb-17 5921-024-235 3493198 2/6/2015 Feb-17 5921-024-235 3496234 2/6/2015 Feb-17 5921-030-135 3492906 2/6/2015 Feb-17 5921-030-135 3493696 2/6/2015 Feb-17 5921-030-135 3495555 2/6/2015 Feb-17 5921-030-235 3492104 2/6/2015 Feb-17 5921-030-235 3492542 2/6/2015 Feb-17 5921-030-235 3493167 2/6/2015 Feb-17 5921-030-235 3493466 2/6/2015 Feb-17 5921-030-235 3494064 2/6/2015 Feb-17 5921-030-235 3495059 2/6/2015 Feb-17 5921-030-235 3495806 2/6/2015 Feb-17 5921-034-135 3491992 2/6/2015 Feb-17 5921-034-135 3492526 2/6/2015 Feb-17 5921-034-135 3493620 2/6/2015 Feb-17 5921-034-135 3494283 2/6/2015 Feb-17 5921-034-235 3492087 2/6/2015 Feb-17 5921-034-235 3492798 2/6/2015 Feb-17 5921-034-235 3493095 2/6/2015 Feb-17 5921-034-235 3493478 2/6/2015 Feb-17 5921-034-235 3494277 2/6/2015 Feb-17 5921-034-235 3495224 2/6/2015 Feb-17 5921-034-235 3495403 2/6/2015 Feb-17 5921-218-235 3487252 2/6/2015 Feb-17 5921-218-235 3494914 2/6/2015 Feb-17 5921-030-236 3493872 2/6/2015 Feb-17 5921-218-135 3491577 2/6/2015 Feb-17 5921-218-135 3492471 2/6/2015 Feb-17 5921-218-135 3493961 2/6/2015 Feb-17 5921-218-136 3494905 2/6/2015 Feb-17 5921-218-236 3494946 2/6/2015 Feb-17 60-7070-103 3492115 2/6/2015 Feb-17 60-7070-103 3492640 2/6/2015 Feb-17 60-7070-103 3493146 2/6/2015 Feb-17 60-7070-103 3493838 2/6/2015 Feb-17 60-7070-103 3494220 2/6/2015 Feb-17 60-7070-103 3494811 2/6/2015 Feb-17 60-7070-103 3495488 2/6/2015 Feb-17 60-7070-103 3495689 2/6/2015 Feb-17 60-7070-105 3492056 2/6/2015 Feb-17 60-7070-105 3492437 2/6/2015 Feb-17 60-7070-105 3492780 2/6/2015 Feb-17 60-7070-105 3493409 2/6/2015 Feb-17 60-7070-105 3493893 2/6/2015 Feb-17 60-7070-105 3494594 2/6/2015 Feb-17 60-7070-105 3495461 2/6/2015 Feb-17 60-7070-105 3496210 2/6/2015 Feb-17 60-7070-104 3492705 2/6/2015 Feb-17 60-7070-104 3494824 2/6/2015 Feb-17 60-7070-003 3491793 2/6/2015 Feb-17 60-7070-006 3491633 2/6/2015 Feb-17 60-7070-006 3493011 2/6/2015 Feb-17 60-7070-106 3491857 2/6/2015 Feb-17 60-7070-106 3492444 2/6/2015 Feb-17 60-7070-106 3492910 2/6/2015 Feb-17 60-7070-106 3493575 2/6/2015 Feb-17 60-7070-106 3494231 2/6/2015 Feb-17 60-7070-106 3494436 2/6/2015 Feb-17 60-7070-106 3495529 2/6/2015 Feb-17 60-7070-106 3496140 2/6/2015 Feb-17 60-7070-107 3495173 2/6/2015 Feb-17 60-7075-006 3494882 2/6/2015 Feb-17 60-7075-103 3492601 2/6/2015 Feb-17 60-7075-103 3493122 2/6/2015 Feb-17 60-7075-103 3495269 2/6/2015 Feb-17 60-7075-104 3491406 2/6/2015 Feb-17 60-7075-104 3492975 2/6/2015 Feb-17 60-7075-105 3493263 2/6/2015 Feb-17 60-7075-105 3493432 2/6/2015 Feb-17 60-7075-106 3492746 2/6/2015 Feb-17 60-7075-106 3493654 2/6/2015 Feb-17 60-7075-106 3494587 2/6/2015 Feb-17 60-7075-106 3496441 2/6/2015 Feb-17 5921-018-135 3497433 2/9/2015 Feb-17 5921-018-135 3501205 2/9/2015 Feb-17 5921-018-235 3497825 2/9/2015 Feb-17 5921-024-135 3498349 2/9/2015 Feb-17 5921-030-135 3498866 2/9/2015 Feb-17 5921-030-135 3501026 2/9/2015 Feb-17 5921-030-235 3497303 2/9/2015 Feb-17 5921-030-235 3497422 2/9/2015 Feb-17 5921-030-235 3497962 2/9/2015 Feb-17 5921-030-235 3501279 2/9/2015 Feb-17 5921-034-135 3496657 2/9/2015 Feb-17 5921-034-135 3497645 2/9/2015 Feb-17 5921-034-135 3500158 2/9/2015 Feb-17 5921-034-135 3501298 2/9/2015 Feb-17 5921-034-235 3496494 2/9/2015 Feb-17 5921-034-235 3497374 2/9/2015 Feb-17 5921-034-235 3497697 2/9/2015 Feb-17 5921-034-235 3497785 2/9/2015 Feb-17 5921-034-235 3498074 2/9/2015 Feb-17 5921-034-235 3498976 2/9/2015 Feb-17 5921-034-235 3500221 2/9/2015 Feb-17 5921-034-235 3500307 2/9/2015 Feb-17 5921-218-235 3497152 2/9/2015 Feb-17 5921-218-235 3497873 2/9/2015 Feb-17 5921-218-235 3498845 2/9/2015 Feb-17 5921-218-235 3500060 2/9/2015 Feb-17 5921-218-135 3496912 2/9/2015 Feb-17 5921-218-135 3497077 2/9/2015 Feb-17 5921-218-135 3499291 2/9/2015 Feb-17 5921-218-135 3501019 2/9/2015 Feb-17 60-7070-103 3497106 2/9/2015 Feb-17 60-7070-103 3497511 2/9/2015 Feb-17 60-7070-103 3497958 2/9/2015 Feb-17 60-7070-103 3498821 2/9/2015 Feb-17 60-7070-103 3499719 2/9/2015 Feb-17 60-7070-103 3499885 2/9/2015 Feb-17 60-7070-103 3499941 2/9/2015 Feb-17 60-7070-103 3500330 2/9/2015 Feb-17 60-7070-103 3501625 2/9/2015 Feb-17 60-7070-105 3497264 2/9/2015 Feb-17 60-7070-105 3498026 2/9/2015 Feb-17 60-7070-105 3499270 2/9/2015 Feb-17 60-7070-105 3500191 2/9/2015 Feb-17 60-7070-105 3500808 2/9/2015 Feb-17 60-7070-104 3496500 2/9/2015 Feb-17 60-7070-104 3497650 2/9/2015 Feb-17 60-7070-104 3498897 2/9/2015 Feb-17 60-7070-003 3497530 2/9/2015 Feb-17 60-7070-006 3493024 2/9/2015 Feb-17 60-7070-006 3501280 2/9/2015 Feb-17 60-7070-106 3498061 2/9/2015 Feb-17 60-7070-106 3499918 2/9/2015 Feb-17 60-7070-106 3500087 2/9/2015 Feb-17 60-7070-106 3500849 2/9/2015 Feb-17 60-7070-106 3500975 2/9/2015 Feb-17 60-7070-107 3499393 2/9/2015 Feb-17 60-7075-003 3498924 2/9/2015 Feb-17 60-7075-103 3497576 2/9/2015 Feb-17 60-7075-103 3501144 2/9/2015 Feb-17 60-7075-104 3497989 2/9/2015 Feb-17 60-7075-105 3497970 2/9/2015 Feb-17 60-7075-106 3496707 2/9/2015 Feb-17 60-7075-106 3497744 2/9/2015 Feb-17 60-7075-106 3501185 2/9/2015 Feb-17 60-7075-106 3501723 2/9/2015 Feb-17 5921-018-135 3503978 2/11/2015 Feb-17 5921-018-235 3505445 2/11/2015 Feb-17 5921-024-135 3505105 2/11/2015 Feb-17 5921-030-135 3502773 2/11/2015 Feb-17 5921-030-135 3502986 2/11/2015 Feb-17 5921-030-135 3505510 2/11/2015 Feb-17 5921-030-235 3502940 2/11/2015 Feb-17 5921-030-235 3503810 2/11/2015 Feb-17 5921-030-235 3504973 2/11/2015 Feb-17 5921-030-235 3506264 2/11/2015 Feb-17 5921-034-135 3502858 2/11/2015 Feb-17 5921-034-135 3504184 2/11/2015 Feb-17 5921-034-135 3505506 2/11/2015 Feb-17 5921-034-235 3502392 2/11/2015 Feb-17 5921-034-235 3502566 2/11/2015 Feb-17 5921-034-235 3503248 2/11/2015 Feb-17 5921-034-235 3504243 2/11/2015 Feb-17 5921-034-235 3505179 2/11/2015 Feb-17 5921-034-235 3506012 2/11/2015 Feb-17 5921-034-236 3501878 2/11/2015 Feb-17 5921-034-236 3502288 2/11/2015 Feb-17 5921-218-235 3502699 2/11/2015 Feb-17 5921-218-235 3504034 2/11/2015 Feb-17 5921-218-235 3505450 2/11/2015 Feb-17 5921-218-135 3503028 2/11/2015 Feb-17 5921-218-135 3504132 2/11/2015 Feb-17 5921-218-135 3505497 2/11/2015 Feb-17 5921-218-236 3505547 2/11/2015 Feb-17 60-7070-103 3502131 2/11/2015 Feb-17 60-7070-103 3502235 2/11/2015 Feb-17 60-7070-103 3502495 2/11/2015 Feb-17 60-7070-103 3503627 2/11/2015 Feb-17 60-7070-103 3504386 2/11/2015 Feb-17 60-7070-103 3505649 2/11/2015 Feb-17 60-7070-105 3502328 2/11/2015 Feb-17 60-7070-105 3503211 2/11/2015 Feb-17 60-7070-105 3504364 2/11/2015 Feb-17 60-7070-105 3505011 2/11/2015 Feb-17 60-7070-105 3505789 2/11/2015 Feb-17 60-7070-104 3501639 2/11/2015 Feb-17 60-7070-104 3504902 2/11/2015 Feb-17 60-7070-104 3504994 2/11/2015 Feb-17 60-7070-006 3503501 2/11/2015 Feb-17 60-7070-106 3503426 2/11/2015 Feb-17 60-7070-106 3504580 2/11/2015 Feb-17 60-7070-106 3504695 2/11/2015 Feb-17 60-7070-106 3504859 2/11/2015 Feb-17 60-7070-106 3505817 2/11/2015 Feb-17 60-7070-107 3505216 2/11/2015 Feb-17 60-7075-003 3504215 2/11/2015 Feb-17 60-7075-006 3501641 2/11/2015 Feb-17 60-7075-006 3505534 2/11/2015 Feb-17 60-7075-103 3502652 2/11/2015 Feb-17 60-7075-103 3503909 2/11/2015 Feb-17 60-7075-103 3506335 2/11/2015 Feb-17 60-7075-103 3506613 2/11/2015 Feb-17 60-7075-104 3501656 2/11/2015 Feb-17 60-7075-104 3505413 2/11/2015 Feb-17 60-7075-105 3501660 2/11/2015 Feb-17 60-7075-105 3505200 2/11/2015 Feb-17 60-7075-106 3502154 2/11/2015 Feb-17 60-7075-106 3503658 2/11/2015 Feb-17 60-7075-106 3504579 2/11/2015 Feb-17 60-7075-106 3505325 2/11/2015 Feb-17 5921-018-135 3508423 2/13/2015 Feb-17 5921-018-235 3509296 2/13/2015 Feb-17 5921-018-235 3509882 2/13/2015 Feb-17 5921-024-135 3509951 2/13/2015 Feb-17 5921-024-235 3510695 2/13/2015 Feb-17 5921-030-135 3506433 2/13/2015 Feb-17 5921-030-135 3509853 2/13/2015 Feb-17 5921-030-235 3508084 2/13/2015 Feb-17 5921-030-235 3509594 2/13/2015 Feb-17 5921-030-235 3509967 2/13/2015 Feb-17 5921-034-135 3507795 2/13/2015 Feb-17 5921-034-135 3507879 2/13/2015 Feb-17 5921-034-135 3509778 2/13/2015 Feb-17 5921-034-235 3507366 2/13/2015 Feb-17 5921-034-235 3508633 2/13/2015 Feb-17 5921-034-235 3509616 2/13/2015 Feb-17 5921-034-235 3509922 2/13/2015 Feb-17 5921-218-235 3507707 2/13/2015 Feb-17 5921-218-235 3509265 2/13/2015 Feb-17 5921-218-235 3509628 2/13/2015 Feb-17 5921-218-235 3510010 2/13/2015 Feb-17 5921-218-235 3510068 2/13/2015 Feb-17 5921-044-235 3508680 2/13/2015 Feb-17 5921-218-135 3507563 2/13/2015 Feb-17 5921-218-135 3510533 2/13/2015 Feb-17 60-7070-103 3507164 2/13/2015 Feb-17 60-7070-103 3507345 2/13/2015 Feb-17 60-7070-103 3507931 2/13/2015 Feb-17 60-7070-103 3508195 2/13/2015 Feb-17 60-7070-103 3509310 2/13/2015 Feb-17 60-7070-103 3509539 2/13/2015 Feb-17 60-7070-103 3509797 2/13/2015 Feb-17 60-7070-103 3509869 2/13/2015 Feb-17 60-7070-103 3510262 2/13/2015 Feb-17 60-7070-105 3506728 2/13/2015 Feb-17 60-7070-105 3507487 2/13/2015 Feb-17 60-7070-105 3507626 2/13/2015 Feb-17 60-7070-105 3508583 2/13/2015 Feb-17 60-7070-105 3509685 2/13/2015 Feb-17 60-7070-105 3510522 2/13/2015 Feb-17 60-7070-104 3506744 2/13/2015 Feb-17 60-7070-104 3509347 2/13/2015 Feb-17 60-7070-003 3510735 2/13/2015 Feb-17 60-7070-004 3509876 2/13/2015 Feb-17 60-7070-005 3510956 2/13/2015 Feb-17 60-7070-006 3509637 2/13/2015 Feb-17 60-7070-106 3506873 2/13/2015 Feb-17 60-7070-106 3507822 2/13/2015 Feb-17 60-7070-106 3508009 2/13/2015 Feb-17 60-7070-106 3509980 2/13/2015 Feb-17 60-7070-106 3510243 2/13/2015 Feb-17 60-7070-107 3510669 2/13/2015 Feb-17 60-7075-003 3508698 2/13/2015 Feb-17 60-7075-003 3508971 2/13/2015 Feb-17 60-7075-004 3506530 2/13/2015 Feb-17 60-7075-006 3508992 2/13/2015 Feb-17 60-7075-103 3509644 2/13/2015 Feb-17 60-7075-103 3510166 2/13/2015 Feb-17 60-7075-106 3507711 2/13/2015 Feb-17 60-7075-106 3509157 2/13/2015 Feb-17 60-7075-106 3510052 2/13/2015 Feb-17 5921-018-135 3511578 2/16/2015 Feb-17 5921-018-135 3515616 2/16/2015 Feb-17 5921-018-235 3513553 2/16/2015 Feb-17 5921-024-135 3515399 2/16/2015 Feb-17 5921-024-235 3513650 2/16/2015 Feb-17 5921-024-235 3514574 2/16/2015 Feb-17 5921-024-235 3514849 2/16/2015 Feb-17 5921-030-135 3512832 2/16/2015 Feb-17 5921-030-235 3512570 2/16/2015 Feb-17 5921-030-235 3514629 2/16/2015 Feb-17 5921-034-135 3511363 2/16/2015 Feb-17 5921-034-135 3512821 2/16/2015 Feb-17 5921-034-135 3513970 2/16/2015 Feb-17 5921-034-135 3515254 2/16/2015 Feb-17 5921-034-235 3511242 2/16/2015 Feb-17 5921-034-235 3512312 2/16/2015 Feb-17 5921-034-235 3513632 2/16/2015 Feb-17 5921-034-235 3514767 2/16/2015 Feb-17 5921-034-235 3514981 2/16/2015 Feb-17 5921-034-236 3514588 2/16/2015 Feb-17 5921-218-235 3511509 2/16/2015 Feb-17 5921-218-235 3512515 2/16/2015 Feb-17 5921-218-235 3512694 2/16/2015 Feb-17 5921-218-235 3514369 2/16/2015 Feb-17 5921-044-135 3512775 2/16/2015 Feb-17 5921-218-135 3512752 2/16/2015 Feb-17 5921-218-135 3513860 2/16/2015 Feb-17 5921-218-135 3515001 2/16/2015 Feb-17 5921-218-136 3512809 2/16/2015 Feb-17 60-7070-102 3511014 2/16/2015 Feb-17 60-7070-103 3511448 2/16/2015 Feb-17 60-7070-103 3511739 2/16/2015 Feb-17 60-7070-103 3512074 2/16/2015 Feb-17 60-7070-103 3512438 2/16/2015 Feb-17 60-7070-103 3512562 2/16/2015 Feb-17 60-7070-103 3513737 2/16/2015 Feb-17 60-7070-103 3514854 2/16/2015 Feb-17 60-7070-103 3515644 2/16/2015 Feb-17 60-7070-105 3511862 2/16/2015 Feb-17 60-7070-105 3513286 2/16/2015 Feb-17 60-7070-105 3513958 2/16/2015 Feb-17 60-7070-105 3514999 2/16/2015 Feb-17 60-7070-104 3509135 2/16/2015 Feb-17 60-7070-104 3511261 2/16/2015 Feb-17 60-7070-104 3513175 2/16/2015 Feb-17 60-7070-104 3513785 2/16/2015 Feb-17 60-7070-104 3515231 2/16/2015 Feb-17 60-7070-003 3514798 2/16/2015 Feb-17 60-7070-004 3514503 2/16/2015 Feb-17 60-7070-106 3511584 2/16/2015 Feb-17 60-7070-106 3511597 2/16/2015 Feb-17 60-7070-106 3512365 2/16/2015 Feb-17 60-7070-106 3513856 2/16/2015 Feb-17 60-7070-106 3514158 2/16/2015 Feb-17 60-7070-106 3514910 2/16/2015 Feb-17 60-7070-106 3515119 2/16/2015 Feb-17 60-7070-107 3512019 2/16/2015 Feb-17 60-7070-107 3513390 2/16/2015 Feb-17 60-7070-107 3514526 2/16/2015 Feb-17 60-7075-003 3513498 2/16/2015 Feb-17 60-7075-006 3511634 2/16/2015 Feb-17 60-7075-006 3512035 2/16/2015 Feb-17 60-7075-103 3512163 2/16/2015 Feb-17 60-7075-103 3513841 2/16/2015 Feb-17 60-7075-103 3515628 2/16/2015 Feb-17 60-7075-104 3513405 2/16/2015 Feb-17 60-7075-105 3511033 2/16/2015 Feb-17 60-7075-105 3514400 2/16/2015 Feb-17 60-7075-105 3514561 2/16/2015 Feb-17 60-7075-106 3512159 2/16/2015 Feb-17 60-7075-106 3515375 2/16/2015 Feb-17 5921-018-135 3519683 2/18/2015 Feb-17 5921-018-235 3518091 2/18/2015 Feb-17 5921-018-235 3519696 2/18/2015 Feb-17 5921-024-135 3519620 2/18/2015 Feb-17 5921-024-235 3518528 2/18/2015 Feb-17 5921-030-135 3515811 2/18/2015 Feb-17 5921-030-135 3517475 2/18/2015 Feb-17 5921-030-135 3519510 2/18/2015 Feb-17 5921-030-135 3520064 2/18/2015 Feb-17 5921-030-235 3506255 2/18/2015 Feb-17 5921-030-235 3515946 2/18/2015 Feb-17 5921-030-235 3516555 2/18/2015 Feb-17 5921-030-235 3518321 2/18/2015 Feb-17 5921-030-235 3519466 2/18/2015 Feb-17 5921-034-135 3516224 2/18/2015 Feb-17 5921-034-135 3517512 2/18/2015 Feb-17 5921-034-135 3518247 2/18/2015 Feb-17 5921-034-135 3519677 2/18/2015 Feb-17 5921-034-235 3516071 2/18/2015 Feb-17 5921-034-235 3517124 2/18/2015 Feb-17 5921-034-235 3518036 2/18/2015 Feb-17 5921-034-235 3518782 2/18/2015 Feb-17 5921-034-235 3519151 2/18/2015 Feb-17 5921-034-235 3519523 2/18/2015 Feb-17 5921-034-235 3520174 2/18/2015 Feb-17 5921-218-235 3515766 2/18/2015 Feb-17 5921-218-235 3516490 2/18/2015 Feb-17 5921-218-235 3516997 2/18/2015 Feb-17 5921-218-235 3517842 2/18/2015 Feb-17 5921-218-235 3518674 2/18/2015 Feb-17 5921-218-235 3519259 2/18/2015 Feb-17 5921-218-235 3520161 2/18/2015 Feb-17 5921-218-135 3517097 2/18/2015 Feb-17 5921-218-135 3518688 2/18/2015 Feb-17 60-7070-103 3515895 2/18/2015 Feb-17 60-7070-103 3516207 2/18/2015 Feb-17 60-7070-103 3516662 2/18/2015 Feb-17 60-7070-103 3517273 2/18/2015 Feb-17 60-7070-103 3518004 2/18/2015 Feb-17 60-7070-103 3518544 2/18/2015 Feb-17 60-7070-103 3519302 2/18/2015 Feb-17 60-7070-103 3519743 2/18/2015 Feb-17 60-7070-103 3519893 2/18/2015 Feb-17 60-7070-103 3519957 2/18/2015 Feb-17 60-7070-103 3520190 2/18/2015 Feb-17 60-7070-105 3515848 2/18/2015 Feb-17 60-7070-105 3516538 2/18/2015 Feb-17 60-7070-105 3516932 2/18/2015 Feb-17 60-7070-105 3517452 2/18/2015 Feb-17 60-7070-105 3518313 2/18/2015 Feb-17 60-7070-105 3518965 2/18/2015 Feb-17 60-7070-105 3519292 2/18/2015 Feb-17 60-7070-105 3519990 2/18/2015 Feb-17 60-7070-105 3520249 2/18/2015 Feb-17 60-7070-104 3517583 2/18/2015 Feb-17 60-7070-104 3517667 2/18/2015 Feb-17 60-7070-104 3519484 2/18/2015 Feb-17 60-7070-006 3516875 2/18/2015 Feb-17 60-7070-006 3518129 2/18/2015 Feb-17 60-7070-106 3516125 2/18/2015 Feb-17 60-7070-106 3516269 2/18/2015 Feb-17 60-7070-106 3517336 2/18/2015 Feb-17 60-7070-106 3517468 2/18/2015 Feb-17 60-7070-106 3518494 2/18/2015 Feb-17 60-7070-106 3519341 2/18/2015 Feb-17 60-7070-106 3519988 2/18/2015 Feb-17 60-7075-003 3515876 2/18/2015 Feb-17 60-7075-004 3516765 2/18/2015 Feb-17 60-7075-006 3516194 2/18/2015 Feb-17 60-7075-103 3517888 2/18/2015 Feb-17 60-7075-103 3519855 2/18/2015 Feb-17 60-7075-103 3520208 2/18/2015 Feb-17 60-7075-104 3517241 2/18/2015 Feb-17 60-7075-106 3517900 2/18/2015 Feb-17 60-7075-106 3518563 2/18/2015 Feb-17 60-7075-106 3519961 2/18/2015 Feb-17 5921-018-135 3521798 2/20/2015 Feb-17 5921-018-135 3523402 2/20/2015 Feb-17 5921-018-235 3523537 2/20/2015 Feb-17 5921-024-135 3522336 2/20/2015 Feb-17 5921-024-235 3522358 2/20/2015 Feb-17 5921-030-135 3521968 2/20/2015 Feb-17 5921-030-135 3524079 2/20/2015 Feb-17 5921-030-235 3521025 2/20/2015 Feb-17 5921-030-235 3522916 2/20/2015 Feb-17 5921-030-235 3523977 2/20/2015 Feb-17 5921-030-235 3525871 2/20/2015 Feb-17 5921-030-235 3526179 2/20/2015 Feb-17 5921-034-135 3521209 2/20/2015 Feb-17 5921-034-135 3522315 2/20/2015 Feb-17 5921-034-135 3523418 2/20/2015 Feb-17 5921-034-135 3523795 2/20/2015 Feb-17 5921-034-135 3526030 2/20/2015 Feb-17 5921-034-136 3525227 2/20/2015 Feb-17 5921-034-235 3520541 2/20/2015 Feb-17 5921-034-235 3521191 2/20/2015 Feb-17 5921-034-235 3521767 2/20/2015 Feb-17 5921-034-235 3522888 2/20/2015 Feb-17 5921-034-235 3523441 2/20/2015 Feb-17 5921-034-235 3525364 2/20/2015 Feb-17 5921-034-235 3525372 2/20/2015 Feb-17 5921-034-235 3525757 2/20/2015 Feb-17 5921-218-235 3521330 2/20/2015 Feb-17 5921-218-235 3522667 2/20/2015 Feb-17 5921-218-235 3523172 2/20/2015 Feb-17 5921-218-235 3523867 2/20/2015 Feb-17 5921-218-235 3525828 2/20/2015 Feb-17 5921-218-235 3526048 2/20/2015 Feb-17 5921-044-235 3521574 2/20/2015 Feb-17 5921-218-135 3520212 2/20/2015 Feb-17 5921-218-135 3521733 2/20/2015 Feb-17 5921-218-135 3521923 2/20/2015 Feb-17 5921-218-135 3523058 2/20/2015 Feb-17 5921-218-135 3525449 2/20/2015 Feb-17 5921-218-136 3525270 2/20/2015 Feb-17 60-7070-102 3522434 2/20/2015 Feb-17 60-7070-102 3524046 2/20/2015 Feb-17 60-7070-103 3520732 2/20/2015 Feb-17 60-7070-103 3521751 2/20/2015 Feb-17 60-7070-103 3522447 2/20/2015 Feb-17 60-7070-103 3523015 2/20/2015 Feb-17 60-7070-103 3523384 2/20/2015 Feb-17 60-7070-103 3523571 2/20/2015 Feb-17 60-7070-103 3524917 2/20/2015 Feb-17 60-7070-103 3525533 2/20/2015 Feb-17 60-7070-105 3520914 2/20/2015 Feb-17 60-7070-105 3521640 2/20/2015 Feb-17 60-7070-105 3521910 2/20/2015 Feb-17 60-7070-105 3522826 2/20/2015 Feb-17 60-7070-105 3523378 2/20/2015 Feb-17 60-7070-105 3525340 2/20/2015 Feb-17 60-7070-105 3525939 2/20/2015 Feb-17 60-7070-105 3525973 2/20/2015 Feb-17 60-7070-104 3520905 2/20/2015 Feb-17 60-7070-104 3522755 2/20/2015 Feb-17 60-7070-104 3523833 2/20/2015 Feb-17 60-7070-104 3525925 2/20/2015 Feb-17 60-7070-003 3521353 2/20/2015 Feb-17 60-7070-005 3524453 2/20/2015 Feb-17 60-7070-006 3523931 2/20/2015 Feb-17 60-7070-007 3522651 2/20/2015 Feb-17 60-7070-106 3522239 2/20/2015 Feb-17 60-7070-106 3522985 2/20/2015 Feb-17 60-7070-106 3524359 2/20/2015 Feb-17 60-7070-106 3525735 2/20/2015 Feb-17 60-7070-107 3522362 2/20/2015 Feb-17 60-7070-107 3524235 2/20/2015 Feb-17 60-7075-003 3520628 2/20/2015 Feb-17 60-7075-006 3524597 2/20/2015 Feb-17 60-7075-006 3524741 2/20/2015 Feb-17 60-7075-103 3520462 2/20/2015 Feb-17 60-7075-103 3521722 2/20/2015 Feb-17 60-7075-103 3522874 2/20/2015 Feb-17 60-7075-103 3524469 2/20/2015 Feb-17 60-7075-104 3522680 2/20/2015 Feb-17 60-7075-105 3520458 2/20/2015 Feb-17 60-7075-105 3520709 2/20/2015 Feb-17 60-7075-106 3520231 2/20/2015 Feb-17 60-7075-106 3521278 2/20/2015 Feb-17 60-7075-106 3523186 2/20/2015 Feb-17 60-7075-106 3526146 2/20/2015 Feb-17 5921-018-135 3529322 2/23/2015 Feb-17 5921-018-235 3526599 2/23/2015 Feb-17 5921-018-235 3530843 2/23/2015 Feb-17 5921-024-135 3526612 2/23/2015 Feb-17 5921-030-135 3527280 2/23/2015 Feb-17 5921-030-135 3528943 2/23/2015 Feb-17 5921-030-136 3527866 2/23/2015 Feb-17 5921-030-235 3528511 2/23/2015 Feb-17 5921-030-235 3529538 2/23/2015 Feb-17 5921-034-135 3527995 2/23/2015 Feb-17 5921-034-135 3528927 2/23/2015 Feb-17 5921-034-135 3529501 2/23/2015 Feb-17 5921-034-136 3527312 2/23/2015 Feb-17 5921-034-235 3527035 2/23/2015 Feb-17 5921-034-235 3528319 2/23/2015 Feb-17 5921-034-235 3528613 2/23/2015 Feb-17 5921-034-235 3531111 2/23/2015 Feb-17 5921-034-236 3527878 2/23/2015 Feb-17 5921-218-235 3528390 2/23/2015 Feb-17 5921-218-235 3528632 2/23/2015 Feb-17 5921-218-235 3531013 2/23/2015 Feb-17 5921-044-236 3526864 2/23/2015 Feb-17 5921-218-135 3526534 2/23/2015 Feb-17 5921-218-135 3528839 2/23/2015 Feb-17 5921-218-135 3529490 2/23/2015 Feb-17 60-7070-103 3526523 2/23/2015 Feb-17 60-7070-103 3528123 2/23/2015 Feb-17 60-7070-103 3529016 2/23/2015 Feb-17 60-7070-103 3530513 2/23/2015 Feb-17 60-7070-103 3530756 2/23/2015 Feb-17 60-7070-105 3527558 2/23/2015 Feb-17 60-7070-105 3528451 2/23/2015 Feb-17 60-7070-105 3529028 2/23/2015 Feb-17 60-7070-105 3530235 2/23/2015 Feb-17 60-7070-104 3526654 2/23/2015 Feb-17 60-7070-104 3528621 2/23/2015 Feb-17 60-7070-104 3530929 2/23/2015 Feb-17 60-7070-003 3529474 2/23/2015 Feb-17 60-7070-106 3527298 2/23/2015 Feb-17 60-7070-106 3527821 2/23/2015 Feb-17 60-7070-106 3527850 2/23/2015 Feb-17 60-7070-106 3528548 2/23/2015 Feb-17 60-7070-106 3528860 2/23/2015 Feb-17 60-7070-106 3529000 2/23/2015 Feb-17 60-7070-106 3530634 2/23/2015 Feb-17 60-7070-107 3527026 2/23/2015 Feb-17 60-7070-107 3529823 2/23/2015 Feb-17 60-7075-003 3526990 2/23/2015 Feb-17 60-7075-103 3526506 2/23/2015 Feb-17 60-7075-103 3528894 2/23/2015 Feb-17 60-7075-103 3531084 2/23/2015 Feb-17 60-7075-104 3526373 2/23/2015 Feb-17 60-7075-105 3526198 2/23/2015 Feb-17 60-7075-105 3529810 2/23/2015 Feb-17 60-7075-106 3528065 2/23/2015 Feb-17 60-7075-106 3529346 2/23/2015 Feb-17 60-7075-106 3530411 2/23/2015 Feb-17 5921-018-135 3533935 2/25/2015 Feb-17 5921-018-135 3534142 2/25/2015 Feb-17 5921-018-235 3535059 2/25/2015 Feb-17 5921-024-135 3531928 2/25/2015 Feb-17 5921-024-135 3534637 2/25/2015 Feb-17 5921-024-235 3532896 2/25/2015 Feb-17 5921-030-135 3531336 2/25/2015 Feb-17 5921-030-135 3534103 2/25/2015 Feb-17 5921-030-235 3531124 2/25/2015 Feb-17 5921-030-235 3532382 2/25/2015 Feb-17 5921-030-235 3532849 2/25/2015 Feb-17 5921-034-135 3531423 2/25/2015 Feb-17 5921-034-135 3533302 2/25/2015 Feb-17 5921-034-135 3535085 2/25/2015 Feb-17 5921-034-235 3531786 2/25/2015 Feb-17 5921-034-235 3532266 2/25/2015 Feb-17 5921-034-235 3532981 2/25/2015 Feb-17 5921-218-235 3531874 2/25/2015 Feb-17 5921-218-235 3532640 2/25/2015 Feb-17 5921-218-235 3533727 2/25/2015 Feb-17 5921-218-135 3532158 2/25/2015 Feb-17 5921-218-135 3534249 2/25/2015 Feb-17 60-7070-103 3531772 2/25/2015 Feb-17 60-7070-103 3533466 2/25/2015 Feb-17 60-7070-103 3533547 2/25/2015 Feb-17 60-7070-103 3534381 2/25/2015 Feb-17 60-7070-103 3534922 2/25/2015 Feb-17 60-7070-105 3531362 2/25/2015 Feb-17 60-7070-105 3532245 2/25/2015 Feb-17 60-7070-105 3533053 2/25/2015 Feb-17 60-7070-105 3533200 2/25/2015 Feb-17 60-7070-105 3533696 2/25/2015 Feb-17 60-7070-105 3534745 2/25/2015 Feb-17 60-7070-104 3534188 2/25/2015 Feb-17 60-7070-003 3534208 2/25/2015 Feb-17 60-7070-006 3534713 2/25/2015 Feb-17 60-7070-106 3531294 2/25/2015 Feb-17 60-7070-106 3531992 2/25/2015 Feb-17 60-7070-106 3533484 2/25/2015 Feb-17 60-7070-106 3533608 2/25/2015 Feb-17 60-7070-106 3534644 2/25/2015 Feb-17 60-7070-106 3534905 2/25/2015 Feb-17 60-7070-107 3531512 2/25/2015 Feb-17 60-7075-003 3533146 2/25/2015 Feb-17 60-7075-006 3532812 2/25/2015 Feb-17 60-7075-103 3533919 2/25/2015 Feb-17 60-7075-104 3534556 2/25/2015 Feb-17 60-7075-105 3531605 2/25/2015 Feb-17 60-7075-106 3534750 2/25/2015 Feb-17 5921-018-135 3538159 3/2/2015 Mar-17 5921-024-135 3538660 3/2/2015 Mar-17 5921-024-235 3536299 3/2/2015 Mar-17 5921-024-235 3539109 3/2/2015 Mar-17 5921-030-135 3536309 3/2/2015 Mar-17 5921-030-135 3539113 3/2/2015 Mar-17 5921-030-136 3537595 3/2/2015 Mar-17 5921-030-235 3535749 3/2/2015 Mar-17 5921-030-235 3537569 3/2/2015 Mar-17 5921-030-235 3539075 3/2/2015 Mar-17 5921-030-235 3539236 3/2/2015 Mar-17 5921-030-235 3539726 3/2/2015 Mar-17 5921-034-135 3535282 3/2/2015 Mar-17 5921-034-135 3537856 3/2/2015 Mar-17 5921-034-135 3538090 3/2/2015 Mar-17 5921-034-235 3532910 3/2/2015 Mar-17 5921-034-235 3535643 3/2/2015 Mar-17 5921-034-235 3536129 3/2/2015 Mar-17 5921-034-235 3537175 3/2/2015 Mar-17 5921-034-235 3538953 3/2/2015 Mar-17 5921-218-235 3535765 3/2/2015 Mar-17 5921-218-235 3537238 3/2/2015 Mar-17 5921-218-235 3537433 3/2/2015 Mar-17 5921-218-235 3538472 3/2/2015 Mar-17 5921-218-135 3535211 3/2/2015 Mar-17 5921-218-135 3538214 3/2/2015 Mar-17 60-7070-103 3535276 3/2/2015 Mar-17 60-7070-103 3535593 3/2/2015 Mar-17 60-7070-103 3536661 3/2/2015 Mar-17 60-7070-103 3537744 3/2/2015 Mar-17 60-7070-103 3538969 3/2/2015 Mar-17 60-7070-103 3539372 3/2/2015 Mar-17 60-7070-105 3535367 3/2/2015 Mar-17 60-7070-105 3537524 3/2/2015 Mar-17 60-7070-105 3537697 3/2/2015 Mar-17 60-7070-105 3537728 3/2/2015 Mar-17 60-7070-105 3539081 3/2/2015 Mar-17 60-7070-104 3535586 3/2/2015 Mar-17 60-7070-104 3538192 3/2/2015 Mar-17 60-7070-003 3538527 3/2/2015 Mar-17 60-7070-006 3536321 3/2/2015 Mar-17 60-7070-006 3537077 3/2/2015 Mar-17 60-7070-106 3535230 3/2/2015 Mar-17 60-7070-106 3535572 3/2/2015 Mar-17 60-7070-106 3538100 3/2/2015 Mar-17 60-7070-106 3538365 3/2/2015 Mar-17 60-7070-106 3539941 3/2/2015 Mar-17 60-7070-106 3540096 3/2/2015 Mar-17 60-7070-107 3536397 3/2/2015 Mar-17 60-7075-003 3535194 3/2/2015 Mar-17 60-7075-103 3536062 3/2/2015 Mar-17 60-7075-104 3538845 3/2/2015 Mar-17 60-7075-105 3536592 3/2/2015 Mar-17 60-7075-106 3535063 3/2/2015 Mar-17 60-7075-106 3537134 3/2/2015 Mar-17 60-7075-106 3538813 3/2/2015 Mar-17 5921-018-235 3542615 3/3/2015 Mar-17 5921-024-136 3544015 3/3/2015 Mar-17 5921-024-235 3542540 3/3/2015 Mar-17 5921-030-135 3542796 3/3/2015 Mar-17 5921-030-235 3541053 3/3/2015 Mar-17 5921-034-135 3542037 3/3/2015 Mar-17 5921-034-135 3544309 3/3/2015 Mar-17 5921-034-235 3541011 3/3/2015 Mar-17 5921-034-235 3542322 3/3/2015 Mar-17 5921-034-235 3545171 3/3/2015 Mar-17 5921-218-235 3540684 3/3/2015 Mar-17 5921-218-235 3542269 3/3/2015 Mar-17 5921-218-235 3542950 3/3/2015 Mar-17 5921-218-235 3543112 3/3/2015 Mar-17 5921-218-235 3543337 3/3/2015 Mar-17 5921-018-236 3543993 3/3/2015 Mar-17 5921-218-135 3540451 3/3/2015 Mar-17 5921-218-135 3542909 3/3/2015 Mar-17 60-7070-103 3540405 3/3/2015 Mar-17 60-7070-103 3541114 3/3/2015 Mar-17 60-7070-103 3541387 3/3/2015 Mar-17 60-7070-103 3542691 3/3/2015 Mar-17 60-7070-103 3543235 3/3/2015 Mar-17 60-7070-105 3540825 3/3/2015 Mar-17 60-7070-105 3541024 3/3/2015 Mar-17 60-7070-105 3542139 3/3/2015 Mar-17 60-7070-105 3543219 3/3/2015 Mar-17 60-7070-105 3544867 3/3/2015 Mar-17 60-7070-104 3544977 3/3/2015 Mar-17 60-7070-104 3545128 3/3/2015 Mar-17 60-7070-005 3545035 3/3/2015 Mar-17 60-7070-006 3543827 3/3/2015 Mar-17 60-7070-106 3540293 3/3/2015 Mar-17 60-7070-106 3541048 3/3/2015 Mar-17 60-7070-106 3542765 3/3/2015 Mar-17 60-7070-106 3544281 3/3/2015 Mar-17 60-7075-003 3543550 3/3/2015 Mar-17 60-7075-006 3539816 3/3/2015 Mar-17 60-7075-006 3540936 3/3/2015 Mar-17 60-7075-103 3540422 3/3/2015 Mar-17 60-7075-103 3541736 3/3/2015 Mar-17 60-7075-104 3544494 3/3/2015 Mar-17 60-7075-104 3544559 3/3/2015 Mar-17 60-7075-105 3544619 3/3/2015 Mar-17 60-7075-105 3544846 3/3/2015 Mar-17 60-7075-106 3540621 3/3/2015 Mar-17 60-7075-106 3542173 3/3/2015 Mar-17 60-7075-106 3544724 3/3/2015 Mar-17 5921-018-135 3546492 3/5/2015 Mar-17 5921-018-235 3547790 3/5/2015 Mar-17 5921-024-135 3545625 3/5/2015 Mar-17 5921-024-235 3545562 3/5/2015 Mar-17 5921-024-235 3548637 3/5/2015 Mar-17 5921-030-135 3548103 3/5/2015 Mar-17 5921-030-135 3548427 3/5/2015 Mar-17 5921-030-235 3542823 3/5/2015 Mar-17 5921-030-235 3545781 3/5/2015 Mar-17 5921-030-235 3548126 3/5/2015 Mar-17 5921-030-235 3548489 3/5/2015 Mar-17 5921-034-135 3545641 3/5/2015 Mar-17 5921-034-135 3547505 3/5/2015 Mar-17 5921-034-235 3545267 3/5/2015 Mar-17 5921-034-235 3549153 3/5/2015 Mar-17 5921-034-235 3549963 3/5/2015 Mar-17 5921-034-236 3546923 3/5/2015 Mar-17 5921-212-235 3545464 3/5/2015 Mar-17 5921-212-235 3548807 3/5/2015 Mar-17 5921-218-235 3546519 3/5/2015 Mar-17 5921-218-235 3546664 3/5/2015 Mar-17 5921-218-235 3548616 3/5/2015 Mar-17 5921-212-135 3546638 3/5/2015 Mar-17 5921-218-135 3545752 3/5/2015 Mar-17 5921-218-135 3549225 3/5/2015 Mar-17 60-7070-103 3545331 3/5/2015 Mar-17 60-7070-103 3546934 3/5/2015 Mar-17 60-7070-103 3547420 3/5/2015 Mar-17 60-7070-103 3549447 3/5/2015 Mar-17 60-7070-103 3549508 3/5/2015 Mar-17 60-7070-103 3549857 3/5/2015 Mar-17 60-7070-105 3546640 3/5/2015 Mar-17 60-7070-105 3548946 3/5/2015 Mar-17 60-7070-105 3549916 3/5/2015 Mar-17 60-7070-104 3547465 3/5/2015 Mar-17 60-7070-104 3548895 3/5/2015 Mar-17 60-7070-003 3547994 3/5/2015 Mar-17 60-7070-106 3545673 3/5/2015 Mar-17 60-7070-106 3547546 3/5/2015 Mar-17 60-7070-106 3547726 3/5/2015 Mar-17 60-7070-106 3549423 3/5/2015 Mar-17 60-7070-106 3550009 3/5/2015 Mar-17 60-7070-107 3545660 3/5/2015 Mar-17 60-7080-102 3546376 3/5/2015 Mar-17 60-7075-103 3547178 3/5/2015 Mar-17 60-7075-103 3547522 3/5/2015 Mar-17 60-7075-103 3549406 3/5/2015 Mar-17 60-7075-104 3549148 3/5/2015 Mar-17 60-7075-106 3547761 3/5/2015 Mar-17 60-7075-106 3549633 3/5/2015 Mar-17 5921-018-135 3550554 3/9/2015 Mar-17 5921-018-135 3550772 3/9/2015 Mar-17 5921-018-135 3553837 3/9/2015 Mar-17 5921-024-135 3550963 3/9/2015 Mar-17 5921-024-235 3552797 3/9/2015 Mar-17 5921-030-135 3551927 3/9/2015 Mar-17 5921-030-135 3552072 3/9/2015 Mar-17 5921-030-235 3550793 3/9/2015 Mar-17 5921-030-235 3553653 3/9/2015 Mar-17 5921-034-135 3552212 3/9/2015 Mar-17 5921-034-235 3550304 3/9/2015 Mar-17 5921-034-235 3552967 3/9/2015 Mar-17 5921-034-235 3553301 3/9/2015 Mar-17 5921-034-235 3553355 3/9/2015 Mar-17 5921-218-235 3551414 3/9/2015 Mar-17 5921-218-235 3552830 3/9/2015 Mar-17 5921-218-235 3553068 3/9/2015 Mar-17 5921-218-135 3551887 3/9/2015 Mar-17 5921-218-135 3552017 3/9/2015 Mar-17 5921-218-135 3553630 3/9/2015 Mar-17 60-7070-102 3551841 3/9/2015 Mar-17 60-7070-102 3552443 3/9/2015 Mar-17 60-7070-103 3550985 3/9/2015 Mar-17 60-7070-103 3552462 3/9/2015 Mar-17 60-7070-103 3552637 3/9/2015 Mar-17 60-7070-103 3553800 3/9/2015 Mar-17 60-7070-105 3552126 3/9/2015 Mar-17 60-7070-105 3554488 3/9/2015 Mar-17 60-7070-104 3550391 3/9/2015 Mar-17 60-7070-104 3550406 3/9/2015 Mar-17 60-7070-003 3554572 3/9/2015 Mar-17 60-7070-005 3551374 3/9/2015 Mar-17 60-7070-106 3550803 3/9/2015 Mar-17 60-7070-106 3551134 3/9/2015 Mar-17 60-7070-106 3551451 3/9/2015 Mar-17 60-7070-106 3552038 3/9/2015 Mar-17 60-7070-106 3552220 3/9/2015 Mar-17 60-7070-106 3554211 3/9/2015 Mar-17 60-7070-106 3554474 3/9/2015 Mar-17 60-7070-107 3550423 3/9/2015 Mar-17 60-7070-107 3553514 3/9/2015 Mar-17 60-7075-003 3551645 3/9/2015 Mar-17 60-7075-005 3550549 3/9/2015 Mar-17 60-7075-006 3551405 3/9/2015 Mar-17 60-7080-102 3553505 3/9/2015 Mar-17 60-7075-106 3550343 3/9/2015 Mar-17 60-7075-106 3551894 3/9/2015 Mar-17 60-7075-106 3554586 3/9/2015 Mar-17 5921-018-135 3559051 3/11/2015 Mar-17 5921-018-136 3558812 3/11/2015 Mar-17 5921-024-135 3555309 3/11/2015 Mar-17 5921-024-235 3556144 3/11/2015 Mar-17 5921-024-235 3558617 3/11/2015 Mar-17 5921-030-135 3556527 3/11/2015 Mar-17 5921-030-235 3556907 3/11/2015 Mar-17 5921-030-235 3557608 3/11/2015 Mar-17 5921-030-235 3557727 3/11/2015 Mar-17 5921-030-235 3557736 3/11/2015 Mar-17 5921-034-135 3555070 3/11/2015 Mar-17 5921-034-135 3556417 3/11/2015 Mar-17 5921-034-135 3558041 3/11/2015 Mar-17 5921-034-135 3558307 3/11/2015 Mar-17 5921-034-235 3556128 3/11/2015 Mar-17 5921-034-235 3556884 3/11/2015 Mar-17 5921-034-235 3557421 3/11/2015 Mar-17 5921-218-235 3555460 3/11/2015 Mar-17 5921-218-235 3555795 3/11/2015 Mar-17 5921-218-235 3556752 3/11/2015 Mar-17 5921-218-235 3557696 3/11/2015 Mar-17 5921-044-235 3557568 3/11/2015 Mar-17 5921-218-135 3555867 3/11/2015 Mar-17 5921-218-236 3558849 3/11/2015 Mar-17 60-7070-103 3554921 3/11/2015 Mar-17 60-7070-103 3556171 3/11/2015 Mar-17 60-7070-103 3558056 3/11/2015 Mar-17 60-7070-103 3558459 3/11/2015 Mar-17 60-7070-105 3554810 3/11/2015 Mar-17 60-7070-105 3556429 3/11/2015 Mar-17 60-7070-105 3556558 3/11/2015 Mar-17 60-7070-105 3558638 3/11/2015 Mar-17 60-7070-105 3558865 3/11/2015 Mar-17 60-7070-104 3554978 3/11/2015 Mar-17 60-7070-104 3555172 3/11/2015 Mar-17 60-7070-104 3559067 3/11/2015 Mar-17 60-7070-006 3554945 3/11/2015 Mar-17 60-7070-106 3555473 3/11/2015 Mar-17 60-7070-106 3555674 3/11/2015 Mar-17 60-7070-106 3557631 3/11/2015 Mar-17 60-7070-106 3557974 3/11/2015 Mar-17 60-7070-106 3559196 3/11/2015 Mar-17 60-7070-107 3556687 3/11/2015 Mar-17 60-7070-107 3556747 3/11/2015 Mar-17 60-7075-004 3557484 3/11/2015 Mar-17 60-7075-006 3558002 3/11/2015 Mar-17 60-7075-103 3556813 3/11/2015 Mar-17 60-7075-104 3557523 3/11/2015 Mar-17 60-7075-105 3557957 3/11/2015 Mar-17 60-7075-106 3555313 3/11/2015 Mar-17 60-7075-106 3558189 3/11/2015 Mar-17 5921-018-135 3561560 3/13/2015 Mar-17 5921-018-235 3560490 3/13/2015 Mar-17 5921-024-135 3561573 3/13/2015 Mar-17 5921-030-135 3561103 3/13/2015 Mar-17 5921-030-136 3561671 3/13/2015 Mar-17 5921-030-235 3560071 3/13/2015 Mar-17 5921-030-235 3563455 3/13/2015 Mar-17 5921-030-235 3563673 3/13/2015 Mar-17 5921-034-135 3562181 3/13/2015 Mar-17 5921-034-135 3563710 3/13/2015 Mar-17 5921-034-235 3559242 3/13/2015 Mar-17 5921-034-235 3559773 3/13/2015 Mar-17 5921-034-235 3560000 3/13/2015 Mar-17 5921-034-235 3562326 3/13/2015 Mar-17 5921-034-235 3563948 3/13/2015 Mar-17 5921-212-235 3561663 3/13/2015 Mar-17 5921-218-235 3559606 3/13/2015 Mar-17 5921-218-235 3560063 3/13/2015 Mar-17 5921-218-235 3563377 3/13/2015 Mar-17 5921-218-135 3559545 3/13/2015 Mar-17 5921-218-135 3560139 3/13/2015 Mar-17 5921-218-135 3562293 3/13/2015 Mar-17 60-7070-103 3558503 3/13/2015 Mar-17 60-7070-103 3560966 3/13/2015 Mar-17 60-7070-103 3562303 3/13/2015 Mar-17 60-7070-103 3562894 3/13/2015 Mar-17 60-7070-103 3563486 3/13/2015 Mar-17 60-7070-105 3559509 3/13/2015 Mar-17 60-7070-105 3560248 3/13/2015 Mar-17 60-7070-105 3560379 3/13/2015 Mar-17 60-7070-105 3562904 3/13/2015 Mar-17 60-7070-105 3563061 3/13/2015 Mar-17 60-7070-105 3563866 3/13/2015 Mar-17 60-7070-104 3561458 3/13/2015 Mar-17 60-7070-003 3560540 3/13/2015 Mar-17 60-7070-004 3560187 3/13/2015 Mar-17 60-7070-004 3560253 3/13/2015 Mar-17 60-7070-106 3560295 3/13/2015 Mar-17 60-7070-106 3560488 3/13/2015 Mar-17 60-7070-106 3562991 3/13/2015 Mar-17 60-7070-107 3559005 3/13/2015 Mar-17 60-7070-107 3560691 3/13/2015 Mar-17 60-7075-003 3562096 3/13/2015 Mar-17 60-7075-006 3562106 3/13/2015 Mar-17 60-7080-102 3559131 3/13/2015 Mar-17 60-7075-103 3561208 3/13/2015 Mar-17 60-7075-103 3562989 3/13/2015 Mar-17 60-7075-104 3560517 3/13/2015 Mar-17 60-7075-105 3561659 3/13/2015 Mar-17 60-7075-106 3560059 3/13/2015 Mar-17 60-7075-106 3561231 3/13/2015 Mar-17 60-7075-107 3561600 3/13/2015 Mar-17 5921-018-135 3564434 3/16/2015 Mar-17 5921-018-235 3564406 3/16/2015 Mar-17 5921-024-135 3566695 3/16/2015 Mar-17 5921-024-235 3563233 3/16/2015 Mar-17 5921-024-235 3566068 3/16/2015 Mar-17 5921-024-235 3568133 3/16/2015 Mar-17 5921-030-135 3564202 3/16/2015 Mar-17 5921-030-235 3566047 3/16/2015 Mar-17 5921-030-235 3566454 3/16/2015 Mar-17 5921-030-235 3566483 3/16/2015 Mar-17 5921-030-235 3568840 3/16/2015 Mar-17 5921-034-135 3565964 3/16/2015 Mar-17 5921-034-135 3567733 3/16/2015 Mar-17 5921-034-235 3565257 3/16/2015 Mar-17 5921-034-235 3566918 3/16/2015 Mar-17 5921-034-235 3568649 3/16/2015 Mar-17 5921-218-235 3563995 3/16/2015 Mar-17 5921-218-235 3565901 3/16/2015 Mar-17 5921-218-235 3567158 3/16/2015 Mar-17 5921-218-235 3568654 3/16/2015 Mar-17 5921-218-135 3564260 3/16/2015 Mar-17 5921-218-135 3565794 3/16/2015 Mar-17 5921-218-135 3567143 3/16/2015 Mar-17 5921-218-236 3566703 3/16/2015 Mar-17 60-7070-102 3564391 3/16/2015 Mar-17 60-7070-102 3568445 3/16/2015 Mar-17 60-7070-103 3564389 3/16/2015 Mar-17 60-7070-103 3565584 3/16/2015 Mar-17 60-7070-103 3566816 3/16/2015 Mar-17 60-7070-103 3568318 3/16/2015 Mar-17 60-7070-105 3565392 3/16/2015 Mar-17 60-7070-105 3566166 3/16/2015 Mar-17 60-7070-105 3567779 3/16/2015 Mar-17 60-7070-105 3567968 3/16/2015 Mar-17 60-7070-105 3569140 3/16/2015 Mar-17 60-7070-104 3564130 3/16/2015 Mar-17 60-7070-104 3565371 3/16/2015 Mar-17 60-7070-104 3567444 3/16/2015 Mar-17 60-7070-003 3564819 3/16/2015 Mar-17 60-7070-004 3567705 3/16/2015 Mar-17 60-7070-106 3564740 3/16/2015 Mar-17 60-7070-106 3565955 3/16/2015 Mar-17 60-7070-106 3566034 3/16/2015 Mar-17 60-7070-106 3567975 3/16/2015 Mar-17 60-7070-106 3568893 3/16/2015 Mar-17 60-7070-106 3569129 3/16/2015 Mar-17 60-7070-107 3564287 3/16/2015 Mar-17 60-7070-107 3566505 3/16/2015 Mar-17 60-7075-003 3568581 3/16/2015 Mar-17 60-7075-004 3567437 3/16/2015 Mar-17 60-7075-006 3566939 3/16/2015 Mar-17 60-7075-006 3568608 3/16/2015 Mar-17 60-7075-103 3564826 3/16/2015 Mar-17 60-7075-103 3566653 3/16/2015 Mar-17 60-7075-104 3565316 3/16/2015 Mar-17 60-7075-105 3565760 3/16/2015 Mar-17 60-7075-106 3563504 3/16/2015 Mar-17 60-7075-106 3566624 3/16/2015 Mar-17 60-7075-106 3568789 3/16/2015 Mar-17 5921-018-135 3571776 3/18/2015 Mar-17 5921-018-235 3571036 3/18/2015 Mar-17 5921-024-135 3571487 3/18/2015 Mar-17 5921-024-235 3572184 3/18/2015 Mar-17 5921-030-135 3570878 3/18/2015 Mar-17 5921-030-235 3571140 3/18/2015 Mar-17 5921-030-235 3573277 3/18/2015 Mar-17 5921-034-135 3570813 3/18/2015 Mar-17 5921-034-135 3572719 3/18/2015 Mar-17 5921-034-235 3569805 3/18/2015 Mar-17 5921-034-235 3572099 3/18/2015 Mar-17 5921-218-235 3569822 3/18/2015 Mar-17 5921-218-235 3571418 3/18/2015 Mar-17 5921-218-235 3572695 3/18/2015 Mar-17 5921-030-236 3571528 3/18/2015 Mar-17 5921-218-135 3570137 3/18/2015 Mar-17 5921-218-135 3570205 3/18/2015 Mar-17 5921-218-135 3571441 3/18/2015 Mar-17 5921-218-136 3573334 3/18/2015 Mar-17 60-7070-103 3570192 3/18/2015 Mar-17 60-7070-103 3571252 3/18/2015 Mar-17 60-7070-103 3572166 3/18/2015 Mar-17 60-7070-103 3573541 3/18/2015 Mar-17 60-7070-105 3569473 3/18/2015 Mar-17 60-7070-105 3571661 3/18/2015 Mar-17 60-7070-105 3571867 3/18/2015 Mar-17 60-7070-104 3569418 3/18/2015 Mar-17 60-7070-104 3571247 3/18/2015 Mar-17 60-7070-003 3572420 3/18/2015 Mar-17 60-7070-005 3571456 3/18/2015 Mar-17 60-7070-106 3570591 3/18/2015 Mar-17 60-7070-106 3571494 3/18/2015 Mar-17 60-7070-107 3571223 3/18/2015 Mar-17 60-7075-005 3571500 3/18/2015 Mar-17 60-7075-103 3570440 3/18/2015 Mar-17 60-7075-103 3570673 3/18/2015 Mar-17 60-7075-103 3572735 3/18/2015 Mar-17 60-7075-104 3569635 3/18/2015 Mar-17 60-7075-104 3572726 3/18/2015 Mar-17 60-7075-106 3571439 3/18/2015 Mar-17 5921-018-135 3577461 3/20/2015 Mar-17 5921-018-235 3577442 3/20/2015 Mar-17 5921-024-135 3575217 3/20/2015 Mar-17 5921-024-235 3577593 3/20/2015 Mar-17 5921-030-135 3576605 3/20/2015 Mar-17 5921-030-235 3575576 3/20/2015 Mar-17 5921-030-235 3576066 3/20/2015 Mar-17 5921-030-235 3577092 3/20/2015 Mar-17 5921-034-135 3575609 3/20/2015 Mar-17 5921-034-136 3573489 3/20/2015 Mar-17 5921-034-136 3577085 3/20/2015 Mar-17 5921-034-235 3574654 3/20/2015 Mar-17 5921-034-235 3576097 3/20/2015 Mar-17 5921-034-235 3576327 3/20/2015 Mar-17 5921-034-235 3576944 3/20/2015 Mar-17 5921-218-235 3575238 3/20/2015 Mar-17 5921-218-235 3576596 3/20/2015 Mar-17 5921-218-235 3577224 3/20/2015 Mar-17 5921-218-235 3578120 3/20/2015 Mar-17 5921-018-236 3577586 3/20/2015 Mar-17 5921-044-235 3577572 3/20/2015 Mar-17 5921-218-135 3574445 3/20/2015 Mar-17 5921-218-135 3577194 3/20/2015 Mar-17 60-7070-102 3577604 3/20/2015 Mar-17 60-7070-103 3573671 3/20/2015 Mar-17 60-7070-103 3573951 3/20/2015 Mar-17 60-7070-103 3575744 3/20/2015 Mar-17 60-7070-103 3576362 3/20/2015 Mar-17 60-7070-103 3577978 3/20/2015 Mar-17 60-7070-105 3573914 3/20/2015 Mar-17 60-7070-105 3575388 3/20/2015 Mar-17 60-7070-105 3576959 3/20/2015 Mar-17 60-7070-104 3574497 3/20/2015 Mar-17 60-7070-104 3577403 3/20/2015 Mar-17 60-7070-104 3577720 3/20/2015 Mar-17 60-7070-003 3575303 3/20/2015 Mar-17 60-7070-006 3575390 3/20/2015 Mar-17 60-7070-106 3573310 3/20/2015 Mar-17 60-7070-106 3573979 3/20/2015 Mar-17 60-7070-106 3574356 3/20/2015 Mar-17 60-7070-106 3576475 3/20/2015 Mar-17 60-7070-106 3576740 3/20/2015 Mar-17 60-7070-106 3576874 3/20/2015 Mar-17 60-7070-106 3578668 3/20/2015 Mar-17 60-7075-004 3574990 3/20/2015 Mar-17 60-7075-103 3578226 3/20/2015 Mar-17 60-7075-103 3578787 3/20/2015 Mar-17 60-7075-106 3574466 3/20/2015 Mar-17 60-7075-106 3575012 3/20/2015 Mar-17 5921-018-135 3580822 3/23/2015 Mar-17 5921-024-135 3579975 3/23/2015 Mar-17 5921-024-235 3582801 3/23/2015 Mar-17 5921-030-135 3580206 3/23/2015 Mar-17 5921-030-135 3580585 3/23/2015 Mar-17 5921-030-135 3583191 3/23/2015 Mar-17 5921-030-136 3581063 3/23/2015 Mar-17 5921-030-235 3579353 3/23/2015 Mar-17 5921-030-235 3581810 3/23/2015 Mar-17 5921-030-235 3582631 3/23/2015 Mar-17 5921-034-135 3579213 3/23/2015 Mar-17 5921-034-135 3580441 3/23/2015 Mar-17 5921-034-135 3580748 3/23/2015 Mar-17 5921-034-135 3581806 3/23/2015 Mar-17 5921-034-235 3579284 3/23/2015 Mar-17 5921-034-235 3580460 3/23/2015 Mar-17 5921-034-235 3580867 3/23/2015 Mar-17 5921-034-235 3582649 3/23/2015 Mar-17 5921-034-235 3583170 3/23/2015 Mar-17 5921-218-235 3579311 3/23/2015 Mar-17 5921-218-235 3581720 3/23/2015 Mar-17 5921-218-235 3581754 3/23/2015 Mar-17 5921-044-135 3580172 3/23/2015 Mar-17 5921-218-135 3579094 3/23/2015 Mar-17 5921-218-135 3582394 3/23/2015 Mar-17 5921-218-236 3579934 3/23/2015 Mar-17 60-7070-103 3578986 3/23/2015 Mar-17 60-7070-103 3580155 3/23/2015 Mar-17 60-7070-103 3581211 3/23/2015 Mar-17 60-7070-103 3581658 3/23/2015 Mar-17 60-7070-103 3582263 3/23/2015 Mar-17 60-7070-103 3583492 3/23/2015 Mar-17 60-7070-105 3578804 3/23/2015 Mar-17 60-7070-105 3579812 3/23/2015 Mar-17 60-7070-105 3580058 3/23/2015 Mar-17 60-7070-105 3580626 3/23/2015 Mar-17 60-7070-105 3581380 3/23/2015 Mar-17 60-7070-105 3581950 3/23/2015 Mar-17 60-7070-105 3583221 3/23/2015 Mar-17 60-7070-104 3579056 3/23/2015 Mar-17 60-7070-104 3581777 3/23/2015 Mar-17 60-7070-106 3578080 3/23/2015 Mar-17 60-7070-106 3579250 3/23/2015 Mar-17 60-7070-106 3581708 3/23/2015 Mar-17 60-7070-106 3582715 3/23/2015 Mar-17 60-7070-107 3579981 3/23/2015 Mar-17 60-7075-003 3579779 3/23/2015 Mar-17 60-7075-003 3583209 3/23/2015 Mar-17 60-7075-004 3545527 3/23/2015 Mar-17 60-7075-006 3580635 3/23/2015 Mar-17 60-7075-006 3580908 3/23/2015 Mar-17 60-7075-103 3579330 3/23/2015 Mar-17 60-7075-103 3581834 3/23/2015 Mar-17 60-7075-104 3579952 3/23/2015 Mar-17 60-7075-105 3579798 3/23/2015 Mar-17 60-7075-105 3580920 3/23/2015 Mar-17 60-7075-106 3579348 3/23/2015 Mar-17 60-7075-106 3580949 3/23/2015 Mar-17 60-7075-106 3581085 3/23/2015 Mar-17 60-7075-106 3581868 3/23/2015 Mar-17 5921-018-135 3584856 3/26/2015 Mar-17 5921-018-135 3585362 3/26/2015 Mar-17 5921-024-135 3585304 3/26/2015 Mar-17 5921-024-235 3586014 3/26/2015 Mar-17 5921-030-135 3585499 3/26/2015 Mar-17 5921-030-235 3585622 3/26/2015 Mar-17 5921-030-235 3587539 3/26/2015 Mar-17 5921-034-135 3583746 3/26/2015 Mar-17 5921-034-135 3585315 3/26/2015 Mar-17 5921-034-135 3587951 3/26/2015 Mar-17 5921-034-235 3584065 3/26/2015 Mar-17 5921-034-235 3584595 3/26/2015 Mar-17 5921-034-235 3586196 3/26/2015 Mar-17 5921-034-236 3585729 3/26/2015 Mar-17 5921-218-235 3583714 3/26/2015 Mar-17 5921-218-235 3583952 3/26/2015 Mar-17 5921-218-235 3584335 3/26/2015 Mar-17 5921-218-235 3584451 3/26/2015 Mar-17 5921-218-235 3585701 3/26/2015 Mar-17 5921-218-235 3587029 3/26/2015 Mar-17 5921-212-135 3585370 3/26/2015 Mar-17 5921-218-135 3587072 3/26/2015 Mar-17 60-7070-103 3583561 3/26/2015 Mar-17 60-7070-103 3584255 3/26/2015 Mar-17 60-7070-103 3585468 3/26/2015 Mar-17 60-7070-103 3585786 3/26/2015 Mar-17 60-7070-103 3586022 3/26/2015 Mar-17 60-7070-103 3586177 3/26/2015 Mar-17 60-7070-103 3586261 3/26/2015 Mar-17 60-7070-103 3586407 3/26/2015 Mar-17 60-7070-103 3587766 3/26/2015 Mar-17 60-7070-105 3584134 3/26/2015 Mar-17 60-7070-105 3585679 3/26/2015 Mar-17 60-7070-105 3586411 3/26/2015 Mar-17 60-7070-105 3586469 3/26/2015 Mar-17 60-7070-105 3587352 3/26/2015 Mar-17 60-7070-104 3584147 3/26/2015 Mar-17 60-7070-104 3585651 3/26/2015 Mar-17 60-7070-104 3586424 3/26/2015 Mar-17 60-7070-005 3585959 3/26/2015 Mar-17 60-7070-106 3583722 3/26/2015 Mar-17 60-7070-106 3583910 3/26/2015 Mar-17 60-7070-106 3584422 3/26/2015 Mar-17 60-7070-106 3585583 3/26/2015 Mar-17 60-7070-106 3586219 3/26/2015 Mar-17 60-7070-106 3586344 3/26/2015 Mar-17 60-7070-106 3586371 3/26/2015 Mar-17 60-7070-106 3587368 3/26/2015 Mar-17 60-7070-107 3583503 3/26/2015 Mar-17 60-7075-003 3587142 3/26/2015 Mar-17 60-7075-006 3585391 3/26/2015 Mar-17 60-7075-103 3586005 3/26/2015 Mar-17 60-7075-103 3586098 3/26/2015 Mar-17 60-7075-103 3587064 3/26/2015 Mar-17 60-7075-106 3585520 3/26/2015 Mar-17 60-7075-106 3586725 3/26/2015 Mar-17 5921-018-235 3588034 3/27/2015 Mar-17 5921-024-135 3589116 3/27/2015 Mar-17 5921-024-235 3589100 3/27/2015 Mar-17 5921-024-235 3590758 3/27/2015 Mar-17 5921-030-135 3591338 3/27/2015 Mar-17 5921-030-135 3591987 3/27/2015 Mar-17 5921-030-235 3588695 3/27/2015 Mar-17 5921-030-235 3590612 3/27/2015 Mar-17 5921-034-135 3589455 3/27/2015 Mar-17 5921-034-135 3591340 3/27/2015 Mar-17 5921-034-235 3588465 3/27/2015 Mar-17 5921-034-235 3590048 3/27/2015 Mar-17 5921-034-235 3591648 3/27/2015 Mar-17 5921-218-235 3588611 3/27/2015 Mar-17 5921-218-235 3588987 3/27/2015 Mar-17 5921-218-235 3591301 3/27/2015 Mar-17 5921-218-135 3590325 3/27/2015 Mar-17 5921-218-135 3590356 3/27/2015 Mar-17 5921-218-136 3588010 3/27/2015 Mar-17 60-7070-103 3588393 3/27/2015 Mar-17 60-7070-103 3589233 3/27/2015 Mar-17 60-7070-103 3590271 3/27/2015 Mar-17 60-7070-103 3591624 3/27/2015 Mar-17 60-7070-103 3592500 3/27/2015 Mar-17 60-7070-103 3592585 3/27/2015 Mar-17 60-7070-105 3588703 3/27/2015 Mar-17 60-7070-105 3588902 3/27/2015 Mar-17 60-7070-105 3589625 3/27/2015 Mar-17 60-7070-105 3589768 3/27/2015 Mar-17 60-7070-104 3588735 3/27/2015 Mar-17 60-7070-104 3591262 3/27/2015 Mar-17 60-7070-106 3588514 3/27/2015 Mar-17 60-7070-106 3589866 3/27/2015 Mar-17 60-7070-106 3591630 3/27/2015 Mar-17 60-7070-106 3591828 3/27/2015 Mar-17 60-7070-107 3589251 3/27/2015 Mar-17 60-7075-103 3589320 3/27/2015 Mar-17 60-7075-103 3590506 3/27/2015 Mar-17 60-7075-103 3591047 3/27/2015 Mar-17 60-7075-104 3592563 3/27/2015 Mar-17 60-7075-106 3589641 3/27/2015 Mar-17 60-7075-106 3590432 3/27/2015 Mar-17 60-7075-106 3590631 3/27/2015 Mar-17 5921-018-135 3595923 3/31/2015 Mar-17 5921-018-235 3592626 3/31/2015 Mar-17 5921-018-235 3594061 3/31/2015 Mar-17 5921-024-135 3595519 3/31/2015 Mar-17 5921-024-235 3594832 3/31/2015 Mar-17 5921-024-235 3596644 3/31/2015 Mar-17 5921-030-235 3593481 3/31/2015 Mar-17 5921-030-235 3593596 3/31/2015 Mar-17 5921-034-135 3595041 3/31/2015 Mar-17 5921-034-235 3592730 3/31/2015 Mar-17 5921-034-235 3593499 3/31/2015 Mar-17 5921-034-235 3593890 3/31/2015 Mar-17 5921-034-235 3595503 3/31/2015 Mar-17 5921-034-235 3595999 3/31/2015 Mar-17 5921-218-235 3593078 3/31/2015 Mar-17 5921-218-235 3595104 3/31/2015 Mar-17 5921-218-235 3595471 3/31/2015 Mar-17 5921-218-235 3595485 3/31/2015 Mar-17 5921-218-135 3592949 3/31/2015 Mar-17 5921-218-135 3594042 3/31/2015 Mar-17 60-7070-102 3593842 3/31/2015 Mar-17 60-7070-103 3593857 3/31/2015 Mar-17 60-7070-103 3594128 3/31/2015 Mar-17 60-7070-103 3595018 3/31/2015 Mar-17 60-7070-103 3596265 3/31/2015 Mar-17 60-7070-103 3596560 3/31/2015 Mar-17 60-7070-105 3592795 3/31/2015 Mar-17 60-7070-105 3593370 3/31/2015 Mar-17 60-7070-105 3594694 3/31/2015 Mar-17 60-7070-105 3595934 3/31/2015 Mar-17 60-7070-105 3596119 3/31/2015 Mar-17 60-7070-105 3596732 3/31/2015 Mar-17 60-7070-104 3594866 3/31/2015 Mar-17 60-7070-104 3594982 3/31/2015 Mar-17 60-7070-104 3596659 3/31/2015 Mar-17 60-7070-003 3594233 3/31/2015 Mar-17 60-7070-005 3595158 3/31/2015 Mar-17 60-7070-006 3592931 3/31/2015 Mar-17 60-7070-006 3595170 3/31/2015 Mar-17 60-7070-106 3509103 3/31/2015 Mar-17 60-7070-106 3520750 3/31/2015 Mar-17 60-7070-106 3593050 3/31/2015 Mar-17 60-7070-106 3593130 3/31/2015 Mar-17 60-7070-106 3594768 3/31/2015 Mar-17 60-7070-106 3594813 3/31/2015 Mar-17 60-7070-106 3596375 3/31/2015 Mar-17 60-7070-106 3596556 3/31/2015 Mar-17 60-7070-106 3596637 3/31/2015 Mar-17 60-7070-107 3592879 3/31/2015 Mar-17 60-7075-006 3594723 3/31/2015 Mar-17 60-7075-103 3594222 3/31/2015 Mar-17 60-7075-106 3592603 3/31/2015 Mar-17 5921-018-235 3604025 4/6/2015 Apr-17 5921-024-135 3601407 4/6/2015 Apr-17 5921-024-235 3601290 4/6/2015 Apr-17 5921-024-235 3603792 4/6/2015 Apr-17 5921-030-135 3604746 4/6/2015 Apr-17 5921-030-235 3601566 4/6/2015 Apr-17 5921-030-235 3603856 4/6/2015 Apr-17 5921-030-235 3603915 4/6/2015 Apr-17 5921-034-135 3599238 4/6/2015 Apr-17 5921-034-135 3603595 4/6/2015 Apr-17 5921-034-135 3604115 4/6/2015 Apr-17 5921-034-135 3604348 4/6/2015 Apr-17 5921-034-235 3601794 4/6/2015 Apr-17 5921-034-235 3602308 4/6/2015 Apr-17 5921-034-235 3602723 4/6/2015 Apr-17 5921-034-235 3603989 4/6/2015 Apr-17 5921-034-235 3603991 4/6/2015 Apr-17 5921-034-235 3604485 4/6/2015 Apr-17 5921-218-235 3599576 4/6/2015 Apr-17 5921-218-235 3599716 4/6/2015 Apr-17 5921-218-235 3602159 4/6/2015 Apr-17 5921-218-235 3602312 4/6/2015 Apr-17 5921-218-235 3604018 4/6/2015 Apr-17 5921-218-135 3602042 4/6/2015 Apr-17 60-7070-102 3603650 4/6/2015 Apr-17 60-7070-103 3600386 4/6/2015 Apr-17 60-7070-103 3601739 4/6/2015 Apr-17 60-7070-103 3602995 4/6/2015 Apr-17 60-7070-103 3603763 4/6/2015 Apr-17 60-7070-105 3601964 4/6/2015 Apr-17 60-7070-105 3602035 4/6/2015 Apr-17 60-7070-105 3602734 4/6/2015 Apr-17 60-7070-105 3604041 4/6/2015 Apr-17 60-7070-105 3604471 4/6/2015 Apr-17 60-7070-105 3604865 4/6/2015 Apr-17 60-7070-104 3602116 4/6/2015 Apr-17 60-7070-104 3603716 4/6/2015 Apr-17 60-7070-003 3602349 4/6/2015 Apr-17 60-7070-106 3599361 4/6/2015 Apr-17 60-7070-106 3601154 4/6/2015 Apr-17 60-7070-106 3601424 4/6/2015 Apr-17 60-7070-106 3601634 4/6/2015 Apr-17 60-7070-106 3602222 4/6/2015 Apr-17 60-7070-106 3602438 4/6/2015 Apr-17 60-7070-106 3603272 4/6/2015 Apr-17 60-7070-106 3603511 4/6/2015 Apr-17 60-7070-106 3604221 4/6/2015 Apr-17 60-7075-003 3602639 4/6/2015 Apr-17 60-7075-103 3600994 4/6/2015 Apr-17 60-7075-103 3603238 4/6/2015 Apr-17 60-7075-106 3599390 4/6/2015 Apr-17 60-7075-106 3602019 4/6/2015 Apr-17 60-7075-106 3603873 4/6/2015 Apr-17 60-7075-107 3597889 4/6/2015 Apr-17 5921-018-135 3605161 4/9/2015 Apr-17 5921-018-135 3608345 4/9/2015 Apr-17 5921-024-135 3605157 4/9/2015 Apr-17 5921-024-135 3608062 4/9/2015 Apr-17 5921-024-235 3608456 4/9/2015 Apr-17 5921-030-135 3607985 4/9/2015 Apr-17 5921-030-235 3605323 4/9/2015 Apr-17 5921-030-235 3606712 4/9/2015 Apr-17 5921-030-235 3608004 4/9/2015 Apr-17 5921-034-135 3606141 4/9/2015 Apr-17 5921-034-135 3606173 4/9/2015 Apr-17 5921-034-135 3606461 4/9/2015 Apr-17 5921-034-135 3606691 4/9/2015 Apr-17 5921-034-235 3605628 4/9/2015 Apr-17 5921-034-235 3605766 4/9/2015 Apr-17 5921-034-235 3606367 4/9/2015 Apr-17 5921-034-235 3607622 4/9/2015 Apr-17 5921-034-235 3608043 4/9/2015 Apr-17 5921-218-235 3605174 4/9/2015 Apr-17 5921-218-235 3605946 4/9/2015 Apr-17 5921-218-235 3607066 4/9/2015 Apr-17 5921-218-235 3607835 4/9/2015 Apr-17 5921-218-135 3604854 4/9/2015 Apr-17 5921-218-135 3605399 4/9/2015 Apr-17 5921-218-135 3605905 4/9/2015 Apr-17 5921-218-135 3606380 4/9/2015 Apr-17 5921-218-135 3608129 4/9/2015 Apr-17 60-7070-103 3604849 4/9/2015 Apr-17 60-7070-103 3606102 4/9/2015 Apr-17 60-7070-103 3607358 4/9/2015 Apr-17 60-7070-103 3608138 4/9/2015 Apr-17 60-7070-103 3609146 4/9/2015 Apr-17 60-7070-105 3606016 4/9/2015 Apr-17 60-7070-105 3606921 4/9/2015 Apr-17 60-7070-105 3607583 4/9/2015 Apr-17 60-7070-105 3608268 4/9/2015 Apr-17 60-7070-105 3608350 4/9/2015 Apr-17 60-7070-104 3606248 4/9/2015 Apr-17 60-7070-104 3608252 4/9/2015 Apr-17 60-7070-006 3607861 4/9/2015 Apr-17 60-7070-006 3608015 4/9/2015 Apr-17 60-7070-006 3609237 4/9/2015 Apr-17 60-7070-106 3605347 4/9/2015 Apr-17 60-7070-106 3605616 4/9/2015 Apr-17 60-7070-106 3606314 4/9/2015 Apr-17 60-7070-106 3607239 4/9/2015 Apr-17 60-7070-106 3607370 4/9/2015 Apr-17 60-7070-106 3609285 4/9/2015 Apr-17 60-7075-006 3606028 4/9/2015 Apr-17 60-7080-102 3608309 4/9/2015 Apr-17 60-7075-102 3605882 4/9/2015 Apr-17 60-7075-104 3608460 4/9/2015 Apr-17 60-7075-104 3609179 4/9/2015 Apr-17 60-7075-105 3608206 4/9/2015 Apr-17 60-7075-106 3605381 4/9/2015 Apr-17 5921-018-235 3610250 4/14/2015 Apr-17 5921-018-235 3610655 4/14/2015 Apr-17 5921-018-235 3611969 4/14/2015 Apr-17 5921-024-135 3612521 4/14/2015 Apr-17 5921-024-235 3611953 4/14/2015 Apr-17 5921-030-135 3609244 4/14/2015 Apr-17 5921-030-235 3609568 4/14/2015 Apr-17 5921-030-235 3610561 4/14/2015 Apr-17 5921-030-235 3612131 4/14/2015 Apr-17 5921-034-135 3609727 4/14/2015 Apr-17 5921-034-135 3610590 4/14/2015 Apr-17 5921-034-235 3610428 4/14/2015 Apr-17 5921-034-235 3610854 4/14/2015 Apr-17 5921-034-235 3611383 4/14/2015 Apr-17 5921-034-235 3612623 4/14/2015 Apr-17 5921-034-235 3612843 4/14/2015 Apr-17 5921-218-235 3610485 4/14/2015 Apr-17 5921-218-235 3611602 4/14/2015 Apr-17 5921-218-235 3611706 4/14/2015 Apr-17 5921-218-235 3613529 4/14/2015 Apr-17 60-7070-103 3609961 4/14/2015 Apr-17 60-7070-103 3610459 4/14/2015 Apr-17 60-7070-103 3611312 4/14/2015 Apr-17 60-7070-103 3612219 4/14/2015 Apr-17 60-7070-103 3612718 4/14/2015 Apr-17 60-7070-105 3610041 4/14/2015 Apr-17 60-7070-105 3610284 4/14/2015 Apr-17 60-7070-105 3610767 4/14/2015 Apr-17 60-7070-105 3611320 4/14/2015 Apr-17 60-7070-105 3612513 4/14/2015 Apr-17 60-7070-104 3612226 4/14/2015 Apr-17 60-7070-104 3612476 4/14/2015 Apr-17 60-7070-106 3609620 4/14/2015 Apr-17 60-7070-106 3610798 4/14/2015 Apr-17 60-7070-106 3611440 4/14/2015 Apr-17 60-7070-106 3612235 4/14/2015 Apr-17 60-7070-106 3612316 4/14/2015 Apr-17 60-7070-106 3613490 4/14/2015 Apr-17 60-7075-003 3612578 4/14/2015 Apr-17 60-7075-003 3612836 4/14/2015 Apr-17 60-7075-103 3610056 4/14/2015 Apr-17 60-7075-103 3610463 4/14/2015 Apr-17 60-7075-105 3610812 4/14/2015 Apr-17 60-7075-106 3610638 4/14/2015 Apr-17 60-7075-106 3611845 4/14/2015 Apr-17 5921-018-135 3615523 4/17/2015 Apr-17 5921-018-135 3615581 4/17/2015 Apr-17 5921-018-135 3616728 4/17/2015 Apr-17 5921-018-235 3615727 4/17/2015 Apr-17 5921-018-235 3616008 4/17/2015 Apr-17 5921-024-235 3614871 4/17/2015 Apr-17 5921-030-135 3615840 4/17/2015 Apr-17 5921-030-235 3613662 4/17/2015 Apr-17 5921-030-235 3616077 4/17/2015 Apr-17 5921-030-235 3616582 4/17/2015 Apr-17 5921-034-135 3614892 4/17/2015 Apr-17 5921-034-135 3614960 4/17/2015 Apr-17 5921-034-135 3616065 4/17/2015 Apr-17 5921-034-135 3616666 4/17/2015 Apr-17 5921-034-135 3617473 4/17/2015 Apr-17 5921-034-235 3614648 4/17/2015 Apr-17 5921-034-235 3615466 4/17/2015 Apr-17 5921-034-235 3616962 4/17/2015 Apr-17 5921-218-235 3614816 4/17/2015 Apr-17 5921-218-235 3614863 4/17/2015 Apr-17 5921-218-235 3616873 4/17/2015 Apr-17 5921-218-235 3618148 4/17/2015 Apr-17 5921-218-135 3614047 4/17/2015 Apr-17 5921-218-135 3614742 4/17/2015 Apr-17 5921-218-135 3615715 4/17/2015 Apr-17 5921-218-135 3617635 4/17/2015 Apr-17 60-7070-102 3616049 4/17/2015 Apr-17 60-7070-103 3613932 4/17/2015 Apr-17 60-7070-103 3614449 4/17/2015 Apr-17 60-7070-103 3615568 4/17/2015 Apr-17 60-7070-103 3616678 4/17/2015 Apr-17 60-7070-103 3617559 4/17/2015 Apr-17 60-7070-103 3617730 4/17/2015 Apr-17 60-7070-105 3613555 4/17/2015 Apr-17 60-7070-105 3614580 4/17/2015 Apr-17 60-7070-105 3615179 4/17/2015 Apr-17 60-7070-105 3616031 4/17/2015 Apr-17 60-7070-105 3616335 4/17/2015 Apr-17 60-7070-105 3617027 4/17/2015 Apr-17 60-7070-104 3614735 4/17/2015 Apr-17 60-7070-104 3617748 4/17/2015 Apr-17 60-7070-003 3613921 4/17/2015 Apr-17 60-7070-003 3614262 4/17/2015 Apr-17 60-7070-106 3614483 4/17/2015 Apr-17 60-7070-106 3615575 4/17/2015 Apr-17 60-7070-106 3616856 4/17/2015 Apr-17 60-7070-106 3616904 4/17/2015 Apr-17 60-7070-106 3617425 4/17/2015 Apr-17 60-7070-106 3617460 4/17/2015 Apr-17 60-7075-103 3615285 4/17/2015 Apr-17 60-7075-103 3616737 4/17/2015 Apr-17 60-7075-105 3615838 4/17/2015 Apr-17 60-7075-106 3614885 4/17/2015 Apr-17 60-7075-106 3615770 4/17/2015 Apr-17 60-7075-106 3616887 4/17/2015 Apr-17 5921-018-135 3620024 4/20/2015 Apr-17 5921-018-235 3621735 4/20/2015 Apr-17 5921-024-135 3618531 4/20/2015 Apr-17 5921-024-235 3618549 4/20/2015 Apr-17 5921-030-135 3620007 4/20/2015 Apr-17 5921-030-135 3620167 4/20/2015 Apr-17 5921-030-135 3622091 4/20/2015 Apr-17 5921-030-135 3622516 4/20/2015 Apr-17 5921-030-235 3618985 4/20/2015 Apr-17 5921-030-235 3621522 4/20/2015 Apr-17 5921-034-235 3618084 4/20/2015 Apr-17 5921-034-235 3619142 4/20/2015 Apr-17 5921-034-235 3619310 4/20/2015 Apr-17 5921-034-235 3621431 4/20/2015 Apr-17 5921-034-235 3621939 4/20/2015 Apr-17 5921-034-235 3622186 4/20/2015 Apr-17 5921-218-235 3618239 4/20/2015 Apr-17 5921-218-235 3618900 4/20/2015 Apr-17 5921-218-235 3620466 4/20/2015 Apr-17 5921-218-235 3620665 4/20/2015 Apr-17 5921-218-135 3619161 4/20/2015 Apr-17 5921-218-135 3622117 4/20/2015 Apr-17 60-7070-103 3618646 4/20/2015 Apr-17 60-7070-103 3619323 4/20/2015 Apr-17 60-7070-103 3621184 4/20/2015 Apr-17 60-7070-103 3622004 4/20/2015 Apr-17 60-7070-105 3618153 4/20/2015 Apr-17 60-7070-105 3619040 4/20/2015 Apr-17 60-7070-105 3619093 4/20/2015 Apr-17 60-7070-105 3621505 4/20/2015 Apr-17 60-7070-105 3622036 4/20/2015 Apr-17 60-7070-104 3621355 4/20/2015 Apr-17 60-7070-104 3621393 4/20/2015 Apr-17 60-7070-003 3618391 4/20/2015 Apr-17 60-7070-003 3622313 4/20/2015 Apr-17 60-7070-006 3618554 4/20/2015 Apr-17 60-7070-006 3618835 4/20/2015 Apr-17 60-7070-106 3618937 4/20/2015 Apr-17 60-7070-106 3620030 4/20/2015 Apr-17 60-7070-106 3622332 4/20/2015 Apr-17 60-7070-107 3619587 4/20/2015 Apr-17 60-7075-006 3621132 4/20/2015 Apr-17 60-7075-103 3619055 4/20/2015 Apr-17 60-7075-103 3619334 4/20/2015 Apr-17 60-7075-103 3622321 4/20/2015 Apr-17 60-7075-103 3623050 4/20/2015 Apr-17 60-7075-104 3618370 4/20/2015 Apr-17 60-7075-105 3619086 4/20/2015 Apr-17 60-7075-107 3619830 4/20/2015 Apr-17 5921-034-235 3621676 4/21/2015 Apr-17 5921-218-235 3621742 4/21/2015 Apr-17 60-7075-106 3621648 4/21/2015 Apr-17 5921-018-135 3626375 4/23/2015 Apr-17 5921-018-135 3627112 4/23/2015 Apr-17 5921-018-235 3626946 4/23/2015 Apr-17 5921-024-135 3623549 4/23/2015 Apr-17 5921-024-135 3625542 4/23/2015 Apr-17 5921-024-235 3623260 4/23/2015 Apr-17 5921-024-235 3625767 4/23/2015 Apr-17 5921-024-235 3625968 4/23/2015 Apr-17 5921-030-135 3625395 4/23/2015 Apr-17 5921-030-136 3625655 4/23/2015 Apr-17 5921-030-235 3623764 4/23/2015 Apr-17 5921-030-235 3625638 4/23/2015 Apr-17 5921-030-235 3625746 4/23/2015 Apr-17 5921-030-235 3627371 4/23/2015 Apr-17 5921-034-135 3623032 4/23/2015 Apr-17 5921-034-135 3625115 4/23/2015 Apr-17 5921-034-135 3625189 4/23/2015 Apr-17 5921-034-135 3626352 4/23/2015 Apr-17 5921-034-235 3623531 4/23/2015 Apr-17 5921-034-235 3625297 4/23/2015 Apr-17 5921-034-235 3626001 4/23/2015 Apr-17 5921-034-235 3626126 4/23/2015 Apr-17 5921-034-235 3626685 4/23/2015 Apr-17 5921-034-236 3624821 4/23/2015 Apr-17 5921-218-235 3623717 4/23/2015 Apr-17 5921-218-235 3625629 4/23/2015 Apr-17 5921-218-235 3627149 4/23/2015 Apr-17 5921-044-135 3626436 4/23/2015 Apr-17 5921-218-135 3623468 4/23/2015 Apr-17 5921-218-135 3624057 4/23/2015 Apr-17 5921-218-135 3625444 4/23/2015 Apr-17 5921-218-136 3625664 4/23/2015 Apr-17 5921-218-136 3626905 4/23/2015 Apr-17 60-7070-102 3624320 4/23/2015 Apr-17 60-7070-103 3623256 4/23/2015 Apr-17 60-7070-103 3623701 4/23/2015 Apr-17 60-7070-103 3625209 4/23/2015 Apr-17 60-7070-103 3625588 4/23/2015 Apr-17 60-7070-103 3626310 4/23/2015 Apr-17 60-7070-103 3627476 4/23/2015 Apr-17 60-7070-105 3623336 4/23/2015 Apr-17 60-7070-105 3625307 4/23/2015 Apr-17 60-7070-105 3625382 4/23/2015 Apr-17 60-7070-105 3625733 4/23/2015 Apr-17 60-7070-105 3627242 4/23/2015 Apr-17 60-7070-104 3623842 4/23/2015 Apr-17 60-7070-004 3624884 4/23/2015 Apr-17 60-7070-006 3625722 4/23/2015 Apr-17 60-7070-006 3626560 4/23/2015 Apr-17 60-7070-106 3624331 4/23/2015 Apr-17 60-7070-106 3625574 4/23/2015 Apr-17 60-7070-106 3626470 4/23/2015 Apr-17 60-7070-107 3625459 4/23/2015 Apr-17 60-7075-003 3624074 4/23/2015 Apr-17 60-7075-003 3625981 4/23/2015 Apr-17 60-7080-102 3624976 4/23/2015 Apr-17 60-7075-103 3623772 4/23/2015 Apr-17 60-7075-105 3624723 4/23/2015 Apr-17 60-7075-106 3623391 4/23/2015 Apr-17 60-7075-106 3625041 4/23/2015 Apr-17 60-7075-106 3626778 4/23/2015 Apr-17 5921-018-135 3634228 4/28/2015 Apr-17 5921-018-235 3632355 4/28/2015 Apr-17 5921-018-235 3634007 4/28/2015 Apr-17 5921-024-135 3632551 4/28/2015 Apr-17 5921-024-135 3636202 4/28/2015 Apr-17 5921-024-135 3636294 4/28/2015 Apr-17 5921-024-235 3633848 4/28/2015 Apr-17 5921-024-235 3636239 4/28/2015 Apr-17 5921-030-135 3635096 4/28/2015 Apr-17 5921-030-235 3632918 4/28/2015 Apr-17 5921-030-235 3633616 4/28/2015 Apr-17 5921-030-235 3634893 4/28/2015 Apr-17 5921-034-135 3634886 4/28/2015 Apr-17 5921-034-135 3636391 4/28/2015 Apr-17 5921-034-235 3632719 4/28/2015 Apr-17 5921-034-235 3633064 4/28/2015 Apr-17 5921-034-235 3634770 4/28/2015 Apr-17 5921-218-235 3632648 4/28/2015 Apr-17 5921-218-235 3633659 4/28/2015 Apr-17 5921-218-235 3634048 4/28/2015 Apr-17 5921-030-236 3632816 4/28/2015 Apr-17 5921-218-135 3632034 4/28/2015 Apr-17 5921-218-135 3633671 4/28/2015 Apr-17 5921-218-135 3635106 4/28/2015 Apr-17 5921-218-136 3633692 4/28/2015 Apr-17 60-7070-103 3632023 4/28/2015 Apr-17 60-7070-103 3632735 4/28/2015 Apr-17 60-7070-103 3634259 4/28/2015 Apr-17 60-7070-103 3634789 4/28/2015 Apr-17 60-7070-103 3635199 4/28/2015 Apr-17 60-7070-105 3632197 4/28/2015 Apr-17 60-7070-105 3633882 4/28/2015 Apr-17 60-7070-105 3635785 4/28/2015 Apr-17 60-7070-104 3632047 4/28/2015 Apr-17 60-7070-104 3634484 4/28/2015 Apr-17 60-7070-003 3633685 4/28/2015 Apr-17 60-7070-004 3634271 4/28/2015 Apr-17 60-7070-006 3635927 4/28/2015 Apr-17 60-7070-106 3632859 4/28/2015 Apr-17 60-7070-106 3634179 4/28/2015 Apr-17 60-7070-106 3635240 4/28/2015 Apr-17 60-7070-106 3636370 4/28/2015 Apr-17 60-7070-107 3634167 4/28/2015 Apr-17 60-7075-003 3634292 4/28/2015 Apr-17 60-7075-004 3635238 4/28/2015 Apr-17 60-7075-005 3634715 4/28/2015 Apr-17 60-7080-102 3634743 4/28/2015 Apr-17 60-7075-103 3632393 4/28/2015 Apr-17 60-7075-103 3633644 4/28/2015 Apr-17 60-7075-104 3635077 4/28/2015 Apr-17 60-7075-106 3633830 4/28/2015 Apr-17 60-7075-106 3635943 4/28/2015 Apr-17 5921-018-135 3630706 4/29/2015 Apr-17 5921-024-235 3629374 4/29/2015 Apr-17 5921-024-235 3630710 4/29/2015 Apr-17 5921-030-135 3629361 4/29/2015 Apr-17 5921-030-135 3629553 4/29/2015 Apr-17 5921-030-136 3629319 4/29/2015 Apr-17 5921-030-235 3629524 4/29/2015 Apr-17 5921-030-235 3631908 4/29/2015 Apr-17 5921-034-135 3627513 4/29/2015 Apr-17 5921-034-135 3627566 4/29/2015 Apr-17 5921-034-135 3630924 4/29/2015 Apr-17 5921-034-135 3630976 4/29/2015 Apr-17 5921-034-135 3631707 4/29/2015 Apr-17 5921-034-235 3627739 4/29/2015 Apr-17 5921-034-235 3628333 4/29/2015 Apr-17 5921-034-235 3629958 4/29/2015 Apr-17 5921-034-235 3630878 4/29/2015 Apr-17 5921-034-235 3631814 4/29/2015 Apr-17 5921-034-235 3632075 4/29/2015 Apr-17 5921-218-235 3629272 4/29/2015 Apr-17 5921-218-235 3631661 4/29/2015 Apr-17 5921-218-135 3629255 4/29/2015 Apr-17 60-7070-102 3629405 4/29/2015 Apr-17 60-7070-103 3629123 4/29/2015 Apr-17 60-7070-103 3630280 4/29/2015 Apr-17 60-7070-103 3631642 4/29/2015 Apr-17 60-7070-105 3627972 4/29/2015 Apr-17 60-7070-105 3628141 4/29/2015 Apr-17 60-7070-105 3630591 4/29/2015 Apr-17 60-7070-105 3630821 4/29/2015 Apr-17 60-7070-105 3630850 4/29/2015 Apr-17 60-7070-104 3627495 4/29/2015 Apr-17 60-7070-104 3627756 4/29/2015 Apr-17 60-7070-104 3629511 4/29/2015 Apr-17 60-7070-104 3630035 4/29/2015 Apr-17 60-7070-104 3630639 4/29/2015 Apr-17 60-7070-004 3628000 4/29/2015 Apr-17 60-7070-005 3627725 4/29/2015 Apr-17 60-7070-006 3627836 4/29/2015 Apr-17 60-7070-006 3628063 4/29/2015 Apr-17 60-7070-106 3627843 4/29/2015 Apr-17 60-7070-106 3629650 4/29/2015 Apr-17 60-7070-106 3630493 4/29/2015 Apr-17 60-7070-106 3631657 4/29/2015 Apr-17 60-7070-107 3628028 4/29/2015 Apr-17 60-7070-107 3629065 4/29/2015 Apr-17 60-7075-006 3627938 4/29/2015 Apr-17 60-7075-006 3628037 4/29/2015 Apr-17 60-7075-006 3631015 4/29/2015 Apr-17 60-7075-006 3631688 4/29/2015 Apr-17 60-7075-103 3627815 4/29/2015 Apr-17 60-7075-103 3630558 4/29/2015 Apr-17 60-7075-104 3629446 4/29/2015 Apr-17 60-7075-105 3629240 4/29/2015 Apr-17 60-7075-106 3627532 4/29/2015 Apr-17 5921-018-135 3636679 5/4/2015 May-17 5921-018-135 3639213 5/4/2015 May-17 5921-018-235 3639733 5/4/2015 May-17 5921-024-135 3639444 5/4/2015 May-17 5921-024-135 3640777 5/4/2015 May-17 5921-024-235 3639590 5/4/2015 May-17 5921-030-135 3636633 5/4/2015 May-17 5921-030-135 3637276 5/4/2015 May-17 5921-030-235 3637398 5/4/2015 May-17 5921-030-235 3638739 5/4/2015 May-17 5921-030-235 3639330 5/4/2015 May-17 5921-034-135 3638836 5/4/2015 May-17 5921-034-235 3636985 5/4/2015 May-17 5921-034-235 3637461 5/4/2015 May-17 5921-034-235 3638328 5/4/2015 May-17 5921-034-235 3639311 5/4/2015 May-17 5921-034-235 3639503 5/4/2015 May-17 5921-034-235 3640593 5/4/2015 May-17 5921-034-236 3639655 5/4/2015 May-17 5921-218-235 3636614 5/4/2015 May-17 5921-218-235 3637564 5/4/2015 May-17 5921-218-235 3638051 5/4/2015 May-17 5921-044-135 3639767 5/4/2015 May-17 5921-218-135 3638196 5/4/2015 May-17 5921-218-135 3638718 5/4/2015 May-17 5921-218-135 3639542 5/4/2015 May-17 5921-218-236 3637248 5/4/2015 May-17 60-7070-103 3639459 5/4/2015 May-17 60-7070-103 3640564 5/4/2015 May-17 60-7070-103 3641072 5/4/2015 May-17 60-7070-105 3637141 5/4/2015 May-17 60-7070-105 3637305 5/4/2015 May-17 60-7070-105 3638177 5/4/2015 May-17 60-7070-105 3639416 5/4/2015 May-17 60-7070-104 3637118 5/4/2015 May-17 60-7070-104 3638550 5/4/2015 May-17 60-7070-003 3637555 5/4/2015 May-17 60-7070-003 3638827 5/4/2015 May-17 60-7070-003 3641161 5/4/2015 May-17 60-7070-005 3639056 5/4/2015 May-17 60-7070-006 3639561 5/4/2015 May-17 60-7070-106 3636740 5/4/2015 May-17 60-7070-106 3637092 5/4/2015 May-17 60-7070-106 3637380 5/4/2015 May-17 60-7070-106 3638862 5/4/2015 May-17 60-7070-106 3639746 5/4/2015 May-17 60-7070-106 3640754 5/4/2015 May-17 60-7070-107 3636698 5/4/2015 May-17 60-7070-107 3639588 5/4/2015 May-17 60-7075-003 3637156 5/4/2015 May-17 60-7075-006 3636701 5/4/2015 May-17 60-7075-006 3639060 5/4/2015 May-17 60-7075-006 3639307 5/4/2015 May-17 60-7080-102 3636963 5/4/2015 May-17 60-7075-103 3637125 5/4/2015 May-17 60-7075-103 3639189 5/4/2015 May-17 60-7075-104 3636971 5/4/2015 May-17 60-7075-104 3640847 5/4/2015 May-17 60-7075-105 3637253 5/4/2015 May-17 60-7075-106 3638300 5/4/2015 May-17 60-7075-107 3637540 5/4/2015 May-17 5921-018-135 3643396 5/6/2015 May-17 5921-018-235 3643775 5/6/2015 May-17 5921-024-135 3645359 5/6/2015 May-17 5921-024-136 3642737 5/6/2015 May-17 5921-030-135 3643401 5/6/2015 May-17 5921-030-235 3643969 5/6/2015 May-17 5921-030-235 3644826 5/6/2015 May-17 5921-034-135 3641745 5/6/2015 May-17 5921-034-135 3645067 5/6/2015 May-17 5921-034-135 3645453 5/6/2015 May-17 5921-034-235 3641922 5/6/2015 May-17 5921-034-235 3644225 5/6/2015 May-17 5921-218-235 3641946 5/6/2015 May-17 5921-218-235 3643866 5/6/2015 May-17 5921-218-235 3644032 5/6/2015 May-17 5921-030-236 3645300 5/6/2015 May-17 5921-218-135 3642744 5/6/2015 May-17 5921-218-135 3644362 5/6/2015 May-17 5921-218-136 3645986 5/6/2015 May-17 5921-218-236 3644256 5/6/2015 May-17 60-7070-103 3641811 5/6/2015 May-17 60-7070-103 3641998 5/6/2015 May-17 60-7070-103 3643003 5/6/2015 May-17 60-7070-103 3643214 5/6/2015 May-17 60-7070-103 3644066 5/6/2015 May-17 60-7070-103 3645079 5/6/2015 May-17 60-7070-105 3640749 5/6/2015 May-17 60-7070-105 3644900 5/6/2015 May-17 60-7070-005 3641306 5/6/2015 May-17 60-7070-005 3641766 5/6/2015 May-17 60-7070-106 3641352 5/6/2015 May-17 60-7070-106 3642217 5/6/2015 May-17 60-7070-106 3642666 5/6/2015 May-17 60-7070-106 3643850 5/6/2015 May-17 60-7070-106 3646113 5/6/2015 May-17 60-7070-107 3643026 5/6/2015 May-17 60-7070-107 3645337 5/6/2015 May-17 60-7075-003 3643464 5/6/2015 May-17 60-7075-003 3643710 5/6/2015 May-17 60-7075-003 3643832 5/6/2015 May-17 60-7075-005 3644239 5/6/2015 May-17 60-7075-006 3645371 5/6/2015 May-17 60-7075-103 3641853 5/6/2015 May-17 60-7075-104 3641869 5/6/2015 May-17 60-7075-105 3643440 5/6/2015 May-17 60-7075-105 3645290 5/6/2015 May-17 60-7075-106 3644294 5/6/2015 May-17 60-7075-106 3644389 5/6/2015 May-17 5921-018-235 3648612 5/8/2015 May-17 5921-030-135 3646653 5/8/2015 May-17 5921-030-135 3647002 5/8/2015 May-17 5921-030-135 3649247 5/8/2015 May-17 5921-030-135 3650846 5/8/2015 May-17 5921-030-235 3646372 5/8/2015 May-17 5921-030-235 3648069 5/8/2015 May-17 5921-030-235 3648677 5/8/2015 May-17 5921-030-235 3650062 5/8/2015 May-17 5921-034-135 3647503 5/8/2015 May-17 5921-034-135 3649983 5/8/2015 May-17 5921-034-135 3650138 5/8/2015 May-17 5921-034-235 3646774 5/8/2015 May-17 5921-034-235 3649155 5/8/2015 May-17 5921-034-235 3649268 5/8/2015 May-17 5921-218-235 3646184 5/8/2015 May-17 5921-218-235 3646800 5/8/2015 May-17 5921-218-235 3649642 5/8/2015 May-17 5921-218-235 3649769 5/8/2015 May-17 5921-218-135 3647831 5/8/2015 May-17 60-7070-103 3646132 5/8/2015 May-17 60-7070-103 3647041 5/8/2015 May-17 60-7070-103 3647297 5/8/2015 May-17 60-7070-103 3647557 5/8/2015 May-17 60-7070-103 3648829 5/8/2015 May-17 60-7070-103 3649402 5/8/2015 May-17 60-7070-103 3649418 5/8/2015 May-17 60-7070-103 3650089 5/8/2015 May-17 60-7070-103 3650528 5/8/2015 May-17 60-7070-105 3646393 5/8/2015 May-17 60-7070-105 3647492 5/8/2015 May-17 60-7070-105 3647865 5/8/2015 May-17 60-7070-105 3648974 5/8/2015 May-17 60-7070-105 3649043 5/8/2015 May-17 60-7070-105 3649657 5/8/2015 May-17 60-7070-106 3646236 5/8/2015 May-17 60-7070-106 3647722 5/8/2015 May-17 60-7070-106 3648421 5/8/2015 May-17 60-7070-106 3649234 5/8/2015 May-17 60-7070-106 3649350 5/8/2015 May-17 60-7075-006 3648620 5/8/2015 May-17 60-7075-103 3646630 5/8/2015 May-17 60-7075-103 3649004 5/8/2015 May-17 60-7075-104 3648445 5/8/2015 May-17 60-7075-105 3650460 5/8/2015 May-17 60-7075-106 3646669 5/8/2015 May-17 60-7075-106 3649091 5/8/2015 May-17 5921-018-135 3652743 5/11/2015 May-17 5921-018-135 3654388 5/11/2015 May-17 5921-018-235 3651457 5/11/2015 May-17 5921-024-135 3653311 5/11/2015 May-17 5921-024-235 3653297 5/11/2015 May-17 5921-024-235 3654238 5/11/2015 May-17 5921-024-235 3656219 5/11/2015 May-17 5921-030-135 3651102 5/11/2015 May-17 5921-030-135 3651540 5/11/2015 May-17 5921-030-135 3651868 5/11/2015 May-17 5921-030-135 3656022 5/11/2015 May-17 5921-030-235 3651881 5/11/2015 May-17 5921-030-235 3652552 5/11/2015 May-17 5921-030-235 3655225 5/11/2015 May-17 5921-034-135 3650977 5/11/2015 May-17 5921-034-135 3651403 5/11/2015 May-17 5921-034-135 3651564 5/11/2015 May-17 5921-034-135 3654096 5/11/2015 May-17 5921-034-135 3655447 5/11/2015 May-17 5921-034-235 3650879 5/11/2015 May-17 5921-034-235 3651351 5/11/2015 May-17 5921-034-235 3652506 5/11/2015 May-17 5921-034-235 3653324 5/11/2015 May-17 5921-034-235 3654374 5/11/2015 May-17 5921-218-235 3651538 5/11/2015 May-17 5921-218-235 3654031 5/11/2015 May-17 5921-218-235 3655520 5/11/2015 May-17 5921-218-135 3650473 5/11/2015 May-17 5921-218-135 3654201 5/11/2015 May-17 5921-218-135 3654229 5/11/2015 May-17 5921-218-135 3655874 5/11/2015 May-17 60-7070-103 3652612 5/11/2015 May-17 60-7070-103 3653041 5/11/2015 May-17 60-7070-103 3654020 5/11/2015 May-17 60-7070-103 3654110 5/11/2015 May-17 60-7070-103 3654498 5/11/2015 May-17 60-7070-103 3655050 5/11/2015 May-17 60-7070-105 3651461 5/11/2015 May-17 60-7070-105 3651712 5/11/2015 May-17 60-7070-105 3653638 5/11/2015 May-17 60-7070-105 3654217 5/11/2015 May-17 60-7070-105 3655260 5/11/2015 May-17 60-7070-105 3655370 5/11/2015 May-17 60-7070-104 3652990 5/11/2015 May-17 60-7070-104 3654175 5/11/2015 May-17 60-7070-006 3652770 5/11/2015 May-17 60-7070-006 3654199 5/11/2015 May-17 60-7070-106 3650996 5/11/2015 May-17 60-7070-106 3651643 5/11/2015 May-17 60-7070-106 3653968 5/11/2015 May-17 60-7070-106 3654357 5/11/2015 May-17 60-7070-106 3654915 5/11/2015 May-17 60-7070-107 3653934 5/11/2015 May-17 60-7070-107 3655985 5/11/2015 May-17 60-7075-003 3650724 5/11/2015 May-17 60-7075-005 3651474 5/11/2015 May-17 60-7075-006 3655327 5/11/2015 May-17 60-7075-103 3651442 5/11/2015 May-17 60-7075-103 3655434 5/11/2015 May-17 60-7075-104 3653002 5/11/2015 May-17 60-7075-105 3651749 5/11/2015 May-17 60-7075-105 3654255 5/11/2015 May-17 60-7075-105 3655304 5/11/2015 May-17 60-7075-106 3651636 5/11/2015 May-17 60-7075-106 3654077 5/11/2015 May-17 5921-018-135 3659832 5/13/2015 May-17 5921-018-235 3657142 5/13/2015 May-17 5921-018-235 3657220 5/13/2015 May-17 5921-018-235 3658902 5/13/2015 May-17 5921-024-135 3657093 5/13/2015 May-17 5921-024-135 3661146 5/13/2015 May-17 5921-024-235 3658669 5/13/2015 May-17 5921-024-235 3659930 5/13/2015 May-17 5921-030-135 3657539 5/13/2015 May-17 5921-030-135 3658682 5/13/2015 May-17 5921-030-135 3660904 5/13/2015 May-17 5921-030-135 3661919 5/13/2015 May-17 5921-030-235 3657126 5/13/2015 May-17 5921-030-235 3657895 5/13/2015 May-17 5921-030-235 3658109 5/13/2015 May-17 5921-030-235 3659486 5/13/2015 May-17 5921-030-235 3660319 5/13/2015 May-17 5921-030-235 3660943 5/13/2015 May-17 5921-034-135 3656938 5/13/2015 May-17 5921-034-135 3656986 5/13/2015 May-17 5921-034-135 3657979 5/13/2015 May-17 5921-034-135 3659948 5/13/2015 May-17 5921-034-135 3660303 5/13/2015 May-17 5921-034-135 3661048 5/13/2015 May-17 5921-034-135 3661466 5/13/2015 May-17 5921-034-135 3661791 5/13/2015 May-17 5921-034-135 3662139 5/13/2015 May-17 5921-034-235 3656014 5/13/2015 May-17 5921-034-235 3656870 5/13/2015 May-17 5921-034-235 3657188 5/13/2015 May-17 5921-034-235 3658317 5/13/2015 May-17 5921-034-235 3659781 5/13/2015 May-17 5921-034-235 3659813 5/13/2015 May-17 5921-034-235 3660175 5/13/2015 May-17 5921-034-235 3661318 5/13/2015 May-17 5921-218-235 3656993 5/13/2015 May-17 5921-218-235 3657462 5/13/2015 May-17 5921-218-235 3658006 5/13/2015 May-17 5921-218-235 3659137 5/13/2015 May-17 5921-218-235 3659821 5/13/2015 May-17 5921-218-235 3660272 5/13/2015 May-17 5921-218-235 3660860 5/13/2015 May-17 5921-218-235 3661478 5/13/2015 May-17 5921-218-135 3657334 5/13/2015 May-17 5921-218-135 3659074 5/13/2015 May-17 5921-218-135 3660077 5/13/2015 May-17 5921-218-135 3660716 5/13/2015 May-17 5921-218-135 3660728 5/13/2015 May-17 60-7070-103 3656359 5/13/2015 May-17 60-7070-103 3657029 5/13/2015 May-17 60-7070-103 3657174 5/13/2015 May-17 60-7070-103 3657489 5/13/2015 May-17 60-7070-103 3659035 5/13/2015 May-17 60-7070-103 3659878 5/13/2015 May-17 60-7070-103 3660217 5/13/2015 May-17 60-7070-103 3660326 5/13/2015 May-17 60-7070-103 3660621 5/13/2015 May-17 60-7070-103 3661356 5/13/2015 May-17 60-7070-103 3661935 5/13/2015 May-17 60-7070-105 3656699 5/13/2015 May-17 60-7070-105 3656917 5/13/2015 May-17 60-7070-105 3657212 5/13/2015 May-17 60-7070-105 3657518 5/13/2015 May-17 60-7070-105 3658648 5/13/2015 May-17 60-7070-105 3659100 5/13/2015 May-17 60-7070-105 3659850 5/13/2015 May-17 60-7070-105 3660293 5/13/2015 May-17 60-7070-105 3660962 5/13/2015 May-17 60-7070-105 3661432 5/13/2015 May-17 60-7070-106 3656371 5/13/2015 May-17 60-7070-106 3657072 5/13/2015 May-17 60-7070-106 3657347 5/13/2015 May-17 60-7070-106 3658329 5/13/2015 May-17 60-7070-106 3658340 5/13/2015 May-17 60-7070-106 3660083 5/13/2015 May-17 60-7070-106 3660374 5/13/2015 May-17 60-7070-106 3661151 5/13/2015 May-17 60-7070-107 3658871 5/13/2015 May-17 60-7070-107 3659982 5/13/2015 May-17 60-7070-107 3660134 5/13/2015 May-17 60-7075-006 3660008 5/13/2015 May-17 60-7075-006 3661872 5/13/2015 May-17 60-7075-103 3658386 5/13/2015 May-17 60-7075-103 3660168 5/13/2015 May-17 60-7075-103 3660388 5/13/2015 May-17 60-7075-104 3661677 5/13/2015 May-17 60-7075-106 3657525 5/13/2015 May-17 60-7075-106 3659493 5/13/2015 May-17 60-7075-106 3659723 5/13/2015 May-17 60-7075-106 3660199 5/13/2015 May-17 60-7075-106 3660802 5/13/2015 May-17 60-7075-006 3650494 5/14/2015 May-17 5921-018-135 3665320 5/15/2015 May-17 5921-018-235 3662978 5/15/2015 May-17 5921-018-235 3665383 5/15/2015 May-17 5921-018-235 3666278 5/15/2015 May-17 5921-024-235 3665472 5/15/2015 May-17 5921-030-135 3662984 5/15/2015 May-17 5921-030-135 3665205 5/15/2015 May-17 5921-030-135 3665504 5/15/2015 May-17 5921-030-235 3662627 5/15/2015 May-17 5921-030-235 3664252 5/15/2015 May-17 5921-030-235 3665116 5/15/2015 May-17 5921-034-135 3662806 5/15/2015 May-17 5921-034-135 3664138 5/15/2015 May-17 5921-034-135 3665331 5/15/2015 May-17 5921-034-235 3662379 5/15/2015 May-17 5921-034-235 3662852 5/15/2015 May-17 5921-034-235 3664299 5/15/2015 May-17 5921-034-235 3665455 5/15/2015 May-17 5921-034-235 3665734 5/15/2015 May-17 5921-034-235 3666104 5/15/2015 May-17 5921-218-235 3662615 5/15/2015 May-17 5921-218-235 3662777 5/15/2015 May-17 5921-218-235 3664172 5/15/2015 May-17 5921-218-235 3664983 5/15/2015 May-17 5921-218-235 3665251 5/15/2015 May-17 5921-218-235 3666209 5/15/2015 May-17 5921-218-135 3665365 5/15/2015 May-17 5921-218-135 3665723 5/15/2015 May-17 60-7070-103 3662230 5/15/2015 May-17 60-7070-103 3662810 5/15/2015 May-17 60-7070-103 3663407 5/15/2015 May-17 60-7070-103 3664062 5/15/2015 May-17 60-7070-103 3665280 5/15/2015 May-17 60-7070-103 3665558 5/15/2015 May-17 60-7070-103 3665747 5/15/2015 May-17 60-7070-103 3667145 5/15/2015 May-17 60-7070-105 3662662 5/15/2015 May-17 60-7070-105 3663149 5/15/2015 May-17 60-7070-105 3664089 5/15/2015 May-17 60-7070-105 3664707 5/15/2015 May-17 60-7070-105 3664977 5/15/2015 May-17 60-7070-105 3665438 5/15/2015 May-17 60-7070-105 3666115 5/15/2015 May-17 60-7070-104 3666189 5/15/2015 May-17 60-7070-104 3667329 5/15/2015 May-17 60-7070-003 3665349 5/15/2015 May-17 60-7070-006 3665354 5/15/2015 May-17 60-7070-106 3662490 5/15/2015 May-17 60-7070-106 3662555 5/15/2015 May-17 60-7070-106 3662868 5/15/2015 May-17 60-7070-106 3664544 5/15/2015 May-17 60-7070-106 3665144 5/15/2015 May-17 60-7070-106 3665214 5/15/2015 May-17 60-7070-106 3665536 5/15/2015 May-17 60-7070-106 3665589 5/15/2015 May-17 60-7070-106 3666087 5/15/2015 May-17 60-7070-107 3662720 5/15/2015 May-17 60-7070-107 3666221 5/15/2015 May-17 60-7075-103 3662847 5/15/2015 May-17 60-7075-103 3664397 5/15/2015 May-17 60-7075-103 3665233 5/15/2015 May-17 60-7075-103 3666297 5/15/2015 May-17 60-7075-104 3664500 5/15/2015 May-17 60-7075-106 3664309 5/15/2015 May-17 60-7075-106 3664661 5/15/2015 May-17 60-7075-106 3665267 5/15/2015 May-17 5921-018-135 3672588 5/18/2015 May-17 5921-018-235 3668713 5/18/2015 May-17 5921-018-235 3671646 5/18/2015 May-17 5921-024-135 3671481 5/18/2015 May-17 5921-024-235 3668573 5/18/2015 May-17 5921-030-135 3669937 5/18/2015 May-17 5921-030-135 3672779 5/18/2015 May-17 5921-030-235 3668556 5/18/2015 May-17 5921-030-235 3671241 5/18/2015 May-17 5921-030-235 3672952 5/18/2015 May-17 5921-034-135 3667973 5/18/2015 May-17 5921-034-235 3666136 5/18/2015 May-17 5921-034-235 3668709 5/18/2015 May-17 5921-034-235 3668811 5/18/2015 May-17 5921-034-235 3671531 5/18/2015 May-17 5921-034-235 3671701 5/18/2015 May-17 5921-034-236 3670495 5/18/2015 May-17 5921-218-235 3668587 5/18/2015 May-17 5921-218-235 3669857 5/18/2015 May-17 5921-218-235 3671256 5/18/2015 May-17 5921-218-135 3669148 5/18/2015 May-17 60-7070-102 3668721 5/18/2015 May-17 60-7070-103 3668502 5/18/2015 May-17 60-7070-103 3668951 5/18/2015 May-17 60-7070-103 3670274 5/18/2015 May-17 60-7070-103 3671434 5/18/2015 May-17 60-7070-105 3667522 5/18/2015 May-17 60-7070-105 3669916 5/18/2015 May-17 60-7070-105 3670203 5/18/2015 May-17 60-7070-105 3671679 5/18/2015 May-17 60-7070-105 3672463 5/18/2015 May-17 60-7070-104 3667761 5/18/2015 May-17 60-7070-104 3670787 5/18/2015 May-17 60-7070-104 3672767 5/18/2015 May-17 60-7070-106 3667956 5/18/2015 May-17 60-7070-106 3668541 5/18/2015 May-17 60-7070-106 3668600 5/18/2015 May-17 60-7070-106 3670509 5/18/2015 May-17 60-7070-106 3671084 5/18/2015 May-17 60-7070-106 3671195 5/18/2015 May-17 60-7075-003 3670261 5/18/2015 May-17 60-7075-103 3668663 5/18/2015 May-17 60-7075-103 3672780 5/18/2015 May-17 60-7075-103 3672906 5/18/2015 May-17 60-7075-104 3670815 5/18/2015 May-17 60-7075-104 3672601 5/18/2015 May-17 60-7075-107 3668616 5/18/2015 May-17 5921-018-135 3676822 5/20/2015 May-17 5921-018-235 3676603 5/20/2015 May-17 5921-024-135 3675796 5/20/2015 May-17 5921-024-135 3676234 5/20/2015 May-17 5921-024-235 3672968 5/20/2015 May-17 5921-024-235 3676769 5/20/2015 May-17 5921-024-235 3677759 5/20/2015 May-17 5921-030-135 3676753 5/20/2015 May-17 5921-030-235 3673105 5/20/2015 May-17 5921-030-235 3675540 5/20/2015 May-17 5921-030-235 3677168 5/20/2015 May-17 5921-034-135 3673200 5/20/2015 May-17 5921-034-135 3673608 5/20/2015 May-17 5921-034-135 3673612 5/20/2015 May-17 5921-034-135 3676247 5/20/2015 May-17 5921-034-135 3676313 5/20/2015 May-17 5921-034-135 3677319 5/20/2015 May-17 5921-034-135 3677451 5/20/2015 May-17 5921-034-135 3677678 5/20/2015 May-17 5921-034-235 3673957 5/20/2015 May-17 5921-034-235 3674064 5/20/2015 May-17 5921-034-235 3675529 5/20/2015 May-17 5921-034-235 3676619 5/20/2015 May-17 5921-034-235 3677240 5/20/2015 May-17 5921-218-235 3674072 5/20/2015 May-17 5921-218-235 3676281 5/20/2015 May-17 5921-218-235 3677943 5/20/2015 May-17 5921-218-135 3673770 5/20/2015 May-17 5921-218-135 3674811 5/20/2015 May-17 5921-218-135 3676252 5/20/2015 May-17 5921-218-135 3676571 5/20/2015 May-17 5921-218-135 3677576 5/20/2015 May-17 5921-218-135 3677856 5/20/2015 May-17 60-7070-103 3673007 5/20/2015 May-17 60-7070-103 3673791 5/20/2015 May-17 60-7070-103 3674671 5/20/2015 May-17 60-7070-103 3676210 5/20/2015 May-17 60-7070-103 3677255 5/20/2015 May-17 60-7070-103 3677802 5/20/2015 May-17 60-7070-105 3673974 5/20/2015 May-17 60-7070-105 3674283 5/20/2015 May-17 60-7070-105 3676268 5/20/2015 May-17 60-7070-105 3677433 5/20/2015 May-17 60-7070-104 3674306 5/20/2015 May-17 60-7070-104 3674824 5/20/2015 May-17 60-7070-106 3672975 5/20/2015 May-17 60-7070-106 3673840 5/20/2015 May-17 60-7070-106 3674709 5/20/2015 May-17 60-7070-106 3674830 5/20/2015 May-17 60-7070-106 3676089 5/20/2015 May-17 60-7070-106 3676898 5/20/2015 May-17 60-7070-106 3677771 5/20/2015 May-17 60-7075-003 3673988 5/20/2015 May-17 60-7075-003 3675380 5/20/2015 May-17 60-7075-004 3673198 5/20/2015 May-17 60-7075-103 3675403 5/20/2015 May-17 60-7075-103 3675662 5/20/2015 May-17 60-7075-106 3673318 5/20/2015 May-17 60-7075-106 3673736 5/20/2015 May-17 60-7075-106 3675899 5/20/2015 May-17 60-7075-106 3676814 5/20/2015 May-17 60-7075-106 3676879 5/20/2015 May-17 5921-018-135 3680874 5/21/2015 May-17 5921-018-136 3678924 5/21/2015 May-17 5921-018-235 3600329 5/21/2015 May-17 5921-018-235 3678429 5/21/2015 May-17 5921-024-235 3680260 5/21/2015 May-17 5921-024-236 3678897 5/21/2015 May-17 5921-030-135 3677958 5/21/2015 May-17 5921-030-135 3678723 5/21/2015 May-17 5921-030-135 3681414 5/21/2015 May-17 5921-030-235 3598194 5/21/2015 May-17 5921-034-135 3678884 5/21/2015 May-17 5921-034-135 3680667 5/21/2015 May-17 5921-034-135 3680916 5/21/2015 May-17 5921-034-235 3600393 5/21/2015 May-17 5921-034-235 3678570 5/21/2015 May-17 5921-034-235 3680273 5/21/2015 May-17 5921-034-235 3681569 5/21/2015 May-17 5921-034-236 3680729 5/21/2015 May-17 5921-218-235 3678706 5/21/2015 May-17 5921-218-235 3680888 5/21/2015 May-17 5921-218-135 3680861 5/21/2015 May-17 5921-218-136 3600514 5/21/2015 May-17 5921-218-136 3680452 5/21/2015 May-17 60-7070-103 3597448 5/21/2015 May-17 60-7070-103 3678694 5/21/2015 May-17 60-7070-103 3679885 5/21/2015 May-17 60-7070-103 3681240 5/21/2015 May-17 60-7070-105 3678417 5/21/2015 May-17 60-7070-105 3678639 5/21/2015 May-17 60-7070-105 3680084 5/21/2015 May-17 60-7070-105 3681802 5/21/2015 May-17 60-7070-104 3598529 5/21/2015 May-17 60-7070-104 3678866 5/21/2015 May-17 60-7070-104 3681524 5/21/2015 May-17 60-7070-003 3679132 5/21/2015 May-17 60-7070-106 3597316 5/21/2015 May-17 60-7070-106 3597550 5/21/2015 May-17 60-7070-106 3678396 5/21/2015 May-17 60-7070-106 3678982 5/21/2015 May-17 60-7070-106 3681498 5/21/2015 May-17 60-7070-107 3597760 5/21/2015 May-17 60-7070-107 3600735 5/21/2015 May-17 60-7075-003 3600099 5/21/2015 May-17 60-7075-003 3678976 5/21/2015 May-17 60-7075-004 3680434 5/21/2015 May-17 60-7080-102 3600726 5/21/2015 May-17 60-7075-103 3678878 5/21/2015 May-17 60-7075-104 3600291 5/21/2015 May-17 60-7075-104 3680447 5/21/2015 May-17 60-7075-105 3679150 5/21/2015 May-17 60-7075-105 3680698 5/21/2015 May-17 60-7075-106 3678354 5/21/2015 May-17 60-7075-106 3681553 5/21/2015 May-17 60-7075-106 3681841 5/21/2015 May-17 5921-018-135 3684731 5/22/2015 May-17 5921-018-136 3682381 5/22/2015 May-17 5921-018-235 3684572 5/22/2015 May-17 5921-024-135 3683166 5/22/2015 May-17 5921-024-135 3686660 5/22/2015 May-17 5921-024-235 3682220 5/22/2015 May-17 5921-024-235 3684586 5/22/2015 May-17 5921-030-135 3683648 5/22/2015 May-17 5921-030-135 3684712 5/22/2015 May-17 5921-034-135 3683301 5/22/2015 May-17 5921-034-135 3683611 5/22/2015 May-17 5921-034-135 3684847 5/22/2015 May-17 5921-034-135 3684997 5/22/2015 May-17 5921-034-135 3686639 5/22/2015 May-17 5921-034-135 3686884 5/22/2015 May-17 5921-034-235 3682399 5/22/2015 May-17 5921-034-235 3683828 5/22/2015 May-17 5921-034-235 3684754 5/22/2015 May-17 5921-034-235 3685814 5/22/2015 May-17 5921-034-235 3685931 5/22/2015 May-17 5921-034-235 3687341 5/22/2015 May-17 5921-212-235 3685313 5/22/2015 May-17 5921-218-235 3681792 5/22/2015 May-17 5921-218-235 3683258 5/22/2015 May-17 5921-218-235 3683871 5/22/2015 May-17 5921-218-235 3684125 5/22/2015 May-17 5921-218-235 3684868 5/22/2015 May-17 5921-218-235 3686185 5/22/2015 May-17 5921-218-135 3684691 5/22/2015 May-17 5921-218-135 3687727 5/22/2015 May-17 60-7070-102 3686687 5/22/2015 May-17 60-7070-103 3681991 5/22/2015 May-17 60-7070-103 3683075 5/22/2015 May-17 60-7070-103 3683227 5/22/2015 May-17 60-7070-103 3684102 5/22/2015 May-17 60-7070-103 3685101 5/22/2015 May-17 60-7070-103 3685441 5/22/2015 May-17 60-7070-103 3686953 5/22/2015 May-17 60-7070-103 3687736 5/22/2015 May-17 60-7070-105 3681927 5/22/2015 May-17 60-7070-105 3683317 5/22/2015 May-17 60-7070-105 3684529 5/22/2015 May-17 60-7070-105 3684593 5/22/2015 May-17 60-7070-105 3685234 5/22/2015 May-17 60-7070-105 3686163 5/22/2015 May-17 60-7070-105 3687133 5/22/2015 May-17 60-7070-105 3687654 5/22/2015 May-17 60-7070-104 3683243 5/22/2015 May-17 60-7070-104 3684765 5/22/2015 May-17 60-7070-104 3686429 5/22/2015 May-17 60-7070-003 3683145 5/22/2015 May-17 60-7070-005 3685062 5/22/2015 May-17 60-7070-106 3681839 5/22/2015 May-17 60-7070-106 3683653 5/22/2015 May-17 60-7070-106 3684474 5/22/2015 May-17 60-7070-106 3685223 5/22/2015 May-17 60-7070-106 3685949 5/22/2015 May-17 60-7070-106 3687122 5/22/2015 May-17 60-7075-003 3683121 5/22/2015 May-17 60-7075-003 3686850 5/22/2015 May-17 60-7075-004 3685043 5/22/2015 May-17 60-7075-006 3682038 5/22/2015 May-17 60-7075-102 3685155 5/22/2015 May-17 60-7075-103 3682029 5/22/2015 May-17 60-7075-103 3686464 5/22/2015 May-17 60-7075-105 3682375 5/22/2015 May-17 60-7075-106 3684085 5/22/2015 May-17 60-7075-106 3684461 5/22/2015 May-17 60-7075-106 3686171 5/22/2015 May-17 60-7075-106 3687743 5/22/2015 May-17 60-7075-107 3686845 5/22/2015 May-17 5921-018-135 3689773 5/27/2015 May-17 5921-018-235 3690967 5/27/2015 May-17 5921-024-135 3689675 5/27/2015 May-17 5921-024-235 3686656 5/27/2015 May-17 5921-024-235 3690951 5/27/2015 May-17 5921-030-135 3688278 5/27/2015 May-17 5921-030-135 3689804 5/27/2015 May-17 5921-034-135 3688073 5/27/2015 May-17 5921-034-135 3690352 5/27/2015 May-17 5921-034-136 3687974 5/27/2015 May-17 5921-034-235 3688094 5/27/2015 May-17 5921-034-235 3689131 5/27/2015 May-17 5921-034-235 3690750 5/27/2015 May-17 5921-034-235 3690797 5/27/2015 May-17 5921-218-235 3687919 5/27/2015 May-17 5921-218-235 3690208 5/27/2015 May-17 5921-218-135 3690807 5/27/2015 May-17 60-7070-102 3687903 5/27/2015 May-17 60-7070-103 3687935 5/27/2015 May-17 60-7070-103 3688170 5/27/2015 May-17 60-7070-103 3688906 5/27/2015 May-17 60-7070-103 3689862 5/27/2015 May-17 60-7070-103 3690637 5/27/2015 May-17 60-7070-103 3691039 5/27/2015 May-17 60-7070-105 3688382 5/27/2015 May-17 60-7070-105 3689098 5/27/2015 May-17 60-7070-105 3690659 5/27/2015 May-17 60-7070-105 3691127 5/27/2015 May-17 60-7070-105 3691330 5/27/2015 May-17 60-7070-104 3687817 5/27/2015 May-17 60-7070-104 3690368 5/27/2015 May-17 60-7070-104 3691136 5/27/2015 May-17 60-7070-005 3687864 5/27/2015 May-17 60-7070-106 3687829 5/27/2015 May-17 60-7070-106 3687942 5/27/2015 May-17 60-7070-106 3688981 5/27/2015 May-17 60-7070-106 3689424 5/27/2015 May-17 60-7070-106 3690375 5/27/2015 May-17 60-7070-106 3691191 5/27/2015 May-17 60-7075-003 3691025 5/27/2015 May-17 60-7080-102 3687886 5/27/2015 May-17 60-7075-103 3688284 5/27/2015 May-17 60-7075-103 3689889 5/27/2015 May-17 60-7075-103 3690853 5/27/2015 May-17 60-7075-105 3688266 5/27/2015 May-17 60-7075-106 3687838 5/27/2015 May-17 60-7075-106 3688557 5/27/2015 May-17 60-7075-106 3690876 5/27/2015 May-17 60-7075-106 3691002 5/27/2015 May-17 5921-018-135 3692315 5/28/2015 May-17 5921-018-135 3694986 5/28/2015 May-17 5921-018-135 3698031 5/28/2015 May-17 5921-018-235 3693516 5/28/2015 May-17 5921-024-135 3693805 5/28/2015 May-17 5921-024-135 3697710 5/28/2015 May-17 5921-024-235 3693753 5/28/2015 May-17 5921-024-235 3695742 5/28/2015 May-17 5921-030-135 3692651 5/28/2015 May-17 5921-030-135 3694938 5/28/2015 May-17 5921-030-135 3697832 5/28/2015 May-17 5921-030-235 3695546 5/28/2015 May-17 5921-030-235 3696829 5/28/2015 May-17 5921-030-235 3699488 5/28/2015 May-17 5921-034-135 3693494 5/28/2015 May-17 5921-034-135 3695355 5/28/2015 May-17 5921-034-135 3695816 5/28/2015 May-17 5921-034-135 3695859 5/28/2015 May-17 5921-034-135 3698904 5/28/2015 May-17 5921-034-136 3697656 5/28/2015 May-17 5921-034-235 3693500 5/28/2015 May-17 5921-034-235 3694328 5/28/2015 May-17 5921-034-235 3694578 5/28/2015 May-17 5921-034-235 3695431 5/28/2015 May-17 5921-034-235 3695800 5/28/2015 May-17 5921-034-235 3698569 5/28/2015 May-17 5921-218-235 3691143 5/28/2015 May-17 5921-218-235 3693004 5/28/2015 May-17 5921-218-235 3693711 5/28/2015 May-17 5921-218-235 3694681 5/28/2015 May-17 5921-218-235 3695364 5/28/2015 May-17 5921-218-235 3696575 5/28/2015 May-17 5921-218-235 3697308 5/28/2015 May-17 5921-218-235 3699211 5/28/2015 May-17 5921-030-236 3697515 5/28/2015 May-17 5921-218-135 3693769 5/28/2015 May-17 5921-218-135 3694675 5/28/2015 May-17 5921-218-135 3696817 5/28/2015 May-17 5921-218-135 3698553 5/28/2015 May-17 60-7070-102 3693880 5/28/2015 May-17 60-7070-102 3694794 5/28/2015 May-17 60-7070-103 3692050 5/28/2015 May-17 60-7070-103 3693027 5/28/2015 May-17 60-7070-103 3693977 5/28/2015 May-17 60-7070-103 3694668 5/28/2015 May-17 60-7070-103 3694964 5/28/2015 May-17 60-7070-103 3695607 5/28/2015 May-17 60-7070-103 3698083 5/28/2015 May-17 60-7070-103 3698264 5/28/2015 May-17 60-7070-103 3698728 5/28/2015 May-17 60-7070-103 3698991 5/28/2015 May-17 60-7070-105 3693487 5/28/2015 May-17 60-7070-105 3694112 5/28/2015 May-17 60-7070-105 3694235 5/28/2015 May-17 60-7070-105 3695372 5/28/2015 May-17 60-7070-105 3695514 5/28/2015 May-17 60-7070-105 3696581 5/28/2015 May-17 60-7070-105 3698530 5/28/2015 May-17 60-7070-104 3692343 5/28/2015 May-17 60-7070-104 3693635 5/28/2015 May-17 60-7070-104 3695023 5/28/2015 May-17 60-7070-004 3697205 5/28/2015 May-17 60-7070-006 3697809 5/28/2015 May-17 60-7070-106 3692622 5/28/2015 May-17 60-7070-106 3692826 5/28/2015 May-17 60-7070-106 3693626 5/28/2015 May-17 60-7070-106 3693748 5/28/2015 May-17 60-7070-106 3694940 5/28/2015 May-17 60-7070-106 3695184 5/28/2015 May-17 60-7070-106 3695695 5/28/2015 May-17 60-7070-106 3696974 5/28/2015 May-17 60-7070-106 3697536 5/28/2015 May-17 60-7070-106 3698894 5/28/2015 May-17 60-7070-107 3697813 5/28/2015 May-17 60-7075-003 3693965 5/28/2015 May-17 60-7075-004 3693833 5/28/2015 May-17 60-7080-102 3694760 5/28/2015 May-17 60-7075-102 3697639 5/28/2015 May-17 60-7075-103 3692013 5/28/2015 May-17 60-7075-103 3693036 5/28/2015 May-17 60-7075-103 3695669 5/28/2015 May-17 60-7075-103 3697185 5/28/2015 May-17 60-7075-104 3696961 5/28/2015 May-17 60-7075-105 3693782 5/28/2015 May-17 60-7075-106 3692646 5/28/2015 May-17 60-7075-106 3694566 5/28/2015 May-17 60-7075-106 3695788 5/28/2015 May-17 5921-018-135 3700155 6/1/2015 May-17 5921-024-135 3704054 6/1/2015 May-17 5921-024-235 3702552 6/1/2015 May-17 5921-030-135 3703348 6/1/2015 May-17 5921-030-135 3705176 6/1/2015 May-17 5921-030-235 3699909 6/1/2015 May-17 5921-030-235 3701796 6/1/2015 May-17 5921-030-235 3702216 6/1/2015 May-17 5921-030-235 3702341 6/1/2015 May-17 5921-030-235 3704958 6/1/2015 May-17 5921-030-235 3706257 6/1/2015 May-17 5921-034-135 3701806 6/1/2015 May-17 5921-034-135 3704962 6/1/2015 May-17 5921-034-235 3700299 6/1/2015 May-17 5921-034-235 3701881 6/1/2015 May-17 5921-034-235 3703170 6/1/2015 May-17 5921-034-235 3703664 6/1/2015 May-17 5921-034-235 3705358 6/1/2015 May-17 5921-034-235 3705717 6/1/2015 May-17 5921-034-235 3705740 6/1/2015 May-17 5921-218-235 3701984 6/1/2015 May-17 5921-218-235 3705013 6/1/2015 May-17 5921-218-135 3700164 6/1/2015 May-17 5921-218-135 3703617 6/1/2015 May-17 5921-218-135 3705182 6/1/2015 May-17 60-7070-103 3699517 6/1/2015 May-17 60-7070-103 3701118 6/1/2015 May-17 60-7070-103 3702179 6/1/2015 May-17 60-7070-103 3702696 6/1/2015 May-17 60-7070-103 3705021 6/1/2015 May-17 60-7070-103 3705169 6/1/2015 May-17 60-7070-105 3700186 6/1/2015 May-17 60-7070-105 3702180 6/1/2015 May-17 60-7070-105 3702840 6/1/2015 May-17 60-7070-105 3705045 6/1/2015 May-17 60-7070-104 3699529 6/1/2015 May-17 60-7070-104 3703311 6/1/2015 May-17 60-7070-007 3701403 6/1/2015 May-17 60-7070-106 3700912 6/1/2015 May-17 60-7070-106 3701852 6/1/2015 May-17 60-7070-106 3704433 6/1/2015 May-17 60-7070-106 3704841 6/1/2015 May-17 60-7075-003 3703705 6/1/2015 May-17 60-7075-103 3700252 6/1/2015 May-17 60-7075-103 3703382 6/1/2015 May-17 60-7075-103 3705050 6/1/2015 May-17 60-7075-106 3704049 6/1/2015 May-17 5921-018-135 3708420 6/4/2015 Jun-17 5921-024-135 3708598 6/4/2015 Jun-17 5921-024-235 3711011 6/4/2015 Jun-17 5921-030-235 3706509 6/4/2015 Jun-17 5921-030-235 3707709 6/4/2015 Jun-17 5921-030-235 3709003 6/4/2015 Jun-17 5921-030-235 3711413 6/4/2015 Jun-17 5921-034-135 3708431 6/4/2015 Jun-17 5921-034-135 3711421 6/4/2015 Jun-17 5921-034-135 3711620 6/4/2015 Jun-17 5921-034-235 3707573 6/4/2015 Jun-17 5921-034-235 3711409 6/4/2015 Jun-17 5921-034-235 3711466 6/4/2015 Jun-17 5921-218-235 3707323 6/4/2015 Jun-17 5921-218-235 3709878 6/4/2015 Jun-17 5921-218-235 3711649 6/4/2015 Jun-17 60-7070-103 3706827 6/4/2015 Jun-17 60-7070-103 3707415 6/4/2015 Jun-17 60-7070-103 3708204 6/4/2015 Jun-17 60-7070-103 3710405 6/4/2015 Jun-17 60-7070-103 3710785 6/4/2015 Jun-17 60-7070-105 3706448 6/4/2015 Jun-17 60-7070-105 3708236 6/4/2015 Jun-17 60-7070-105 3709035 6/4/2015 Jun-17 60-7070-105 3710134 6/4/2015 Jun-17 60-7070-105 3711432 6/4/2015 Jun-17 60-7070-104 3708580 6/4/2015 Jun-17 60-7070-104 3709866 6/4/2015 Jun-17 60-7070-106 3707072 6/4/2015 Jun-17 60-7070-106 3708262 6/4/2015 Jun-17 60-7070-106 3708514 6/4/2015 Jun-17 60-7070-106 3709723 6/4/2015 Jun-17 60-7070-106 3710335 6/4/2015 Jun-17 60-7070-107 3708355 6/4/2015 Jun-17 60-7075-003 3710414 6/4/2015 Jun-17 60-7075-103 3708611 6/4/2015 Jun-17 60-7075-104 3708607 6/4/2015 Jun-17 60-7075-105 3711612 6/4/2015 Jun-17 60-7075-106 3706385 6/4/2015 Jun-17 60-7075-106 3706392 6/4/2015 Jun-17 60-7075-106 3709246 6/4/2015 Jun-17 60-7075-106 3711133 6/4/2015 Jun-17 60-7075-106 3711339 6/4/2015 Jun-17 5921-018-135 3713359 6/5/2015 Jun-17 5921-030-135 3713699 6/5/2015 Jun-17 5921-030-235 3716104 6/5/2015 Jun-17 5921-034-135 3713891 6/5/2015 Jun-17 5921-034-135 3716693 6/5/2015 Jun-17 5921-034-235 3597623 6/5/2015 Jun-17 5921-034-235 3712593 6/5/2015 Jun-17 5921-034-235 3715185 6/5/2015 Jun-17 5921-034-236 3714384 6/5/2015 Jun-17 5921-212-235 3713889 6/5/2015 Jun-17 5921-218-135 3713371 6/5/2015 Jun-17 60-7070-102 3600465 6/5/2015 Jun-17 60-7070-103 3598501 6/5/2015 Jun-17 60-7070-103 3712426 6/5/2015 Jun-17 60-7070-103 3714140 6/5/2015 Jun-17 60-7070-103 3716284 6/5/2015 Jun-17 60-7070-103 3716812 6/5/2015 Jun-17 60-7070-105 3713681 6/5/2015 Jun-17 60-7070-105 3716981 6/5/2015 Jun-17 60-7070-104 3715527 6/5/2015 Jun-17 60-7070-003 3714439 6/5/2015 Jun-17 60-7070-003 3716266 6/5/2015 Jun-17 60-7070-004 3714711 6/5/2015 Jun-17 60-7070-004 3716906 6/5/2015 Jun-17 60-7070-005 3715752 6/5/2015 Jun-17 60-7070-005 3716910 6/5/2015 Jun-17 60-7070-006 3711677 6/5/2015 Jun-17 60-7070-006 3713079 6/5/2015 Jun-17 60-7070-006 3713513 6/5/2015 Jun-17 60-7070-106 3711840 6/5/2015 Jun-17 60-7070-106 3714155 6/5/2015 Jun-17 60-7070-106 3714730 6/5/2015 Jun-17 60-7070-106 3715205 6/5/2015 Jun-17 60-7070-106 3715298 6/5/2015 Jun-17 60-7070-106 3716705 6/5/2015 Jun-17 60-7070-107 3713940 6/5/2015 Jun-17 60-7075-003 3716278 6/5/2015 Jun-17 60-7075-006 3713955 6/5/2015 Jun-17 60-7075-006 3714252 6/5/2015 Jun-17 60-7075-103 3712796 6/5/2015 Jun-17 60-7075-103 3716923 6/5/2015 Jun-17 60-7075-104 3715536 6/5/2015 Jun-17 60-7075-105 3713046 6/5/2015 Jun-17 60-7075-106 3716094 6/5/2015 Jun-17 5921-018-235 3717580 6/9/2015 Jun-17 5921-018-235 3720088 6/9/2015 Jun-17 5921-024-135 3720438 6/9/2015 Jun-17 5921-024-135 3721027 6/9/2015 Jun-17 5921-024-136 3720090 6/9/2015 Jun-17 5921-024-235 3720440 6/9/2015 Jun-17 5921-024-235 3722892 6/9/2015 Jun-17 5921-030-235 3718458 6/9/2015 Jun-17 5921-030-235 3718980 6/9/2015 Jun-17 5921-034-135 3719916 6/9/2015 Jun-17 5921-034-135 3720061 6/9/2015 Jun-17 5921-034-235 3718876 6/9/2015 Jun-17 5921-034-236 3722774 6/9/2015 Jun-17 5921-212-235 3720100 6/9/2015 Jun-17 5921-218-235 3717145 6/9/2015 Jun-17 5921-218-235 3721350 6/9/2015 Jun-17 5921-218-235 3722010 6/9/2015 Jun-17 5921-218-135 3719447 6/9/2015 Jun-17 5921-218-135 3721206 6/9/2015 Jun-17 5921-218-136 3719842 6/9/2015 Jun-17 60-7070-103 3718306 6/9/2015 Jun-17 60-7070-103 3718470 6/9/2015 Jun-17 60-7070-103 3721193 6/9/2015 Jun-17 60-7070-103 3722925 6/9/2015 Jun-17 60-7070-105 3720163 6/9/2015 Jun-17 60-7070-105 3720455 6/9/2015 Jun-17 60-7070-105 3722215 6/9/2015 Jun-17 60-7070-104 3719531 6/9/2015 Jun-17 60-7070-104 3721345 6/9/2015 Jun-17 60-7070-003 3719857 6/9/2015 Jun-17 60-7070-106 3718869 6/9/2015 Jun-17 60-7070-106 3719577 6/9/2015 Jun-17 60-7070-106 3721089 6/9/2015 Jun-17 60-7070-107 3717236 6/9/2015 Jun-17 60-7070-107 3722372 6/9/2015 Jun-17 60-7075-006 3718415 6/9/2015 Jun-17 60-7075-006 3722580 6/9/2015 Jun-17 60-7075-103 3719900 6/9/2015 Jun-17 60-7075-103 3722197 6/9/2015 Jun-17 60-7075-104 3719888 6/9/2015 Jun-17 60-7075-105 3717243 6/9/2015 Jun-17 60-7075-105 3717624 6/9/2015 Jun-17 60-7075-105 3722567 6/9/2015 Jun-17 60-7075-106 3719890 6/9/2015 Jun-17 5921-018-135 3724696 6/11/2015 Jun-17 5921-018-235 3727253 6/11/2015 Jun-17 5921-018-235 3729158 6/11/2015 Jun-17 5921-024-135 3728080 6/11/2015 Jun-17 5921-024-235 3725915 6/11/2015 Jun-17 5921-024-235 3729104 6/11/2015 Jun-17 5921-030-135 3726111 6/11/2015 Jun-17 5921-030-135 3726239 6/11/2015 Jun-17 5921-030-235 3724817 6/11/2015 Jun-17 5921-030-235 3725049 6/11/2015 Jun-17 5921-030-235 3725716 6/11/2015 Jun-17 5921-030-235 3728917 6/11/2015 Jun-17 5921-034-135 3725609 6/11/2015 Jun-17 5921-034-135 3728955 6/11/2015 Jun-17 5921-034-235 3722918 6/11/2015 Jun-17 5921-034-235 3724069 6/11/2015 Jun-17 5921-034-235 3728521 6/11/2015 Jun-17 5921-034-235 3728702 6/11/2015 Jun-17 5921-218-235 3724727 6/11/2015 Jun-17 5921-218-235 3725860 6/11/2015 Jun-17 5921-218-235 3728079 6/11/2015 Jun-17 5921-218-235 3728964 6/11/2015 Jun-17 5921-218-135 3725008 6/11/2015 Jun-17 5921-218-135 3726130 6/11/2015 Jun-17 5921-218-135 3726928 6/11/2015 Jun-17 5921-218-135 3729087 6/11/2015 Jun-17 60-7070-103 3723017 6/11/2015 Jun-17 60-7070-103 3723979 6/11/2015 Jun-17 60-7070-103 3725240 6/11/2015 Jun-17 60-7070-103 3726423 6/11/2015 Jun-17 60-7070-103 3726499 6/11/2015 Jun-17 60-7070-103 3728219 6/11/2015 Jun-17 60-7070-103 3728760 6/11/2015 Jun-17 60-7070-103 3729444 6/11/2015 Jun-17 60-7070-105 3724990 6/11/2015 Jun-17 60-7070-105 3727834 6/11/2015 Jun-17 60-7070-105 3728739 6/11/2015 Jun-17 60-7070-104 3724339 6/11/2015 Jun-17 60-7070-104 3727211 6/11/2015 Jun-17 60-7070-104 3728756 6/11/2015 Jun-17 60-7070-006 3724840 6/11/2015 Jun-17 60-7070-106 3724053 6/11/2015 Jun-17 60-7070-106 3725613 6/11/2015 Jun-17 60-7070-106 3725792 6/11/2015 Jun-17 60-7070-106 3727670 6/11/2015 Jun-17 60-7070-106 3727932 6/11/2015 Jun-17 60-7070-106 3729162 6/11/2015 Jun-17 60-7075-006 3727868 6/11/2015 Jun-17 60-7075-104 3724356 6/11/2015 Jun-17 60-7075-105 3728773 6/11/2015 Jun-17 60-7075-106 3725697 6/11/2015 Jun-17 5921-018-135 3732015 6/15/2015 Jun-17 5921-018-135 3735923 6/15/2015 Jun-17 5921-018-235 3731774 6/15/2015 Jun-17 5921-018-235 3735693 6/15/2015 Jun-17 5921-024-135 3732036 6/15/2015 Jun-17 5921-024-135 3732138 6/15/2015 Jun-17 5921-024-135 3735382 6/15/2015 Jun-17 5921-024-235 3732494 6/15/2015 Jun-17 5921-024-235 3735191 6/15/2015 Jun-17 5921-030-135 3732378 6/15/2015 Jun-17 5921-030-135 3734440 6/15/2015 Jun-17 5921-030-235 3731790 6/15/2015 Jun-17 5921-030-235 3734781 6/15/2015 Jun-17 5921-034-135 3731340 6/15/2015 Jun-17 5921-034-135 3732397 6/15/2015 Jun-17 5921-034-135 3733336 6/15/2015 Jun-17 5921-034-135 3733434 6/15/2015 Jun-17 5921-034-135 3735250 6/15/2015 Jun-17 5921-034-135 3735416 6/15/2015 Jun-17 5921-034-235 3730146 6/15/2015 Jun-17 5921-034-235 3731567 6/15/2015 Jun-17 5921-034-235 3733992 6/15/2015 Jun-17 5921-218-235 3731579 6/15/2015 Jun-17 5921-218-235 3733622 6/15/2015 Jun-17 5921-218-235 3734845 6/15/2015 Jun-17 5921-218-235 3735353 6/15/2015 Jun-17 5921-044-235 3732206 6/15/2015 Jun-17 5921-218-135 3730654 6/15/2015 Jun-17 5921-218-135 3733605 6/15/2015 Jun-17 5921-218-135 3735471 6/15/2015 Jun-17 5921-218-236 3730394 6/15/2015 Jun-17 60-7070-103 3730665 6/15/2015 Jun-17 60-7070-103 3731630 6/15/2015 Jun-17 60-7070-103 3732186 6/15/2015 Jun-17 60-7070-103 3732500 6/15/2015 Jun-17 60-7070-103 3733565 6/15/2015 Jun-17 60-7070-103 3734377 6/15/2015 Jun-17 60-7070-103 3735284 6/15/2015 Jun-17 60-7070-103 3735369 6/15/2015 Jun-17 60-7070-105 3730133 6/15/2015 Jun-17 60-7070-105 3731648 6/15/2015 Jun-17 60-7070-105 3731800 6/15/2015 Jun-17 60-7070-105 3732172 6/15/2015 Jun-17 60-7070-105 3734019 6/15/2015 Jun-17 60-7070-105 3734752 6/15/2015 Jun-17 60-7070-104 3730631 6/15/2015 Jun-17 60-7070-104 3734641 6/15/2015 Jun-17 60-7070-106 3731023 6/15/2015 Jun-17 60-7070-106 3731355 6/15/2015 Jun-17 60-7070-106 3731514 6/15/2015 Jun-17 60-7070-106 3732425 6/15/2015 Jun-17 60-7070-106 3734163 6/15/2015 Jun-17 60-7070-106 3734417 6/15/2015 Jun-17 60-7070-106 3734673 6/15/2015 Jun-17 60-7075-006 3734171 6/15/2015 Jun-17 60-7075-102 3730339 6/15/2015 Jun-17 60-7075-102 3732345 6/15/2015 Jun-17 60-7075-104 3730829 6/15/2015 Jun-17 60-7075-104 3735686 6/15/2015 Jun-17 60-7075-105 3734775 6/15/2015 Jun-17 60-7075-106 3731034 6/15/2015 Jun-17 60-7075-106 3731204 6/15/2015 Jun-17 60-7075-106 3732043 6/15/2015 Jun-17 60-7075-106 3735672 6/15/2015 Jun-17 5921-018-135 3736914 6/17/2015 Jun-17 5921-018-135 3737739 6/17/2015 Jun-17 5921-018-235 3736848 6/17/2015 Jun-17 5921-024-235 3738852 6/17/2015 Jun-17 5921-030-135 3738868 6/17/2015 Jun-17 5921-030-235 3736732 6/17/2015 Jun-17 5921-030-235 3738643 6/17/2015 Jun-17 5921-034-235 3735981 6/17/2015 Jun-17 5921-034-235 3737836 6/17/2015 Jun-17 5921-034-235 3739168 6/17/2015 Jun-17 5921-034-235 3740917 6/17/2015 Jun-17 5921-034-235 3741241 6/17/2015 Jun-17 5921-218-235 3737300 6/17/2015 Jun-17 5921-218-235 3738403 6/17/2015 Jun-17 5921-218-235 3738689 6/17/2015 Jun-17 5921-218-235 3741202 6/17/2015 Jun-17 5921-218-135 3738950 6/17/2015 Jun-17 5921-218-135 3741835 6/17/2015 Jun-17 60-7070-103 3736017 6/17/2015 Jun-17 60-7070-103 3737606 6/17/2015 Jun-17 60-7070-103 3738457 6/17/2015 Jun-17 60-7070-103 3738810 6/17/2015 Jun-17 60-7070-103 3739255 6/17/2015 Jun-17 60-7070-103 3740815 6/17/2015 Jun-17 60-7070-105 3736399 6/17/2015 Jun-17 60-7070-105 3738194 6/17/2015 Jun-17 60-7070-105 3741148 6/17/2015 Jun-17 60-7070-104 3738731 6/17/2015 Jun-17 60-7070-104 3741169 6/17/2015 Jun-17 60-7070-006 3736713 6/17/2015 Jun-17 60-7070-106 3735999 6/17/2015 Jun-17 60-7070-106 3736427 6/17/2015 Jun-17 60-7070-106 3738139 6/17/2015 Jun-17 60-7070-106 3739152 6/17/2015 Jun-17 60-7070-106 3740940 6/17/2015 Jun-17 60-7070-106 3741499 6/17/2015 Jun-17 60-7070-106 3741679 6/17/2015 Jun-17 60-7070-107 3736980 6/17/2015 Jun-17 60-7075-006 3738966 6/17/2015 Jun-17 60-7075-103 3736979 6/17/2015 Jun-17 60-7075-103 3737098 6/17/2015 Jun-17 60-7075-103 3741819 6/17/2015 Jun-17 60-7075-106 3737316 6/17/2015 Jun-17 60-7075-106 3738398 6/17/2015 Jun-17 5921-018-135 3743445 6/19/2015 Jun-17 5921-018-235 3742714 6/19/2015 Jun-17 5921-024-135 3745899 6/19/2015 Jun-17 5921-024-235 3743421 6/19/2015 Jun-17 5921-024-235 3743665 6/19/2015 Jun-17 5921-030-135 3742722 6/19/2015 Jun-17 5921-030-135 3744663 6/19/2015 Jun-17 5921-030-235 3742213 6/19/2015 Jun-17 5921-030-235 3743704 6/19/2015 Jun-17 5921-030-235 3745689 6/19/2015 Jun-17 5921-030-235 3747335 6/19/2015 Jun-17 5921-034-135 3742923 6/19/2015 Jun-17 5921-034-135 3745658 6/19/2015 Jun-17 5921-034-135 3745731 6/19/2015 Jun-17 5921-034-135 3747989 6/19/2015 Jun-17 5921-034-235 3742934 6/19/2015 Jun-17 5921-034-235 3744208 6/19/2015 Jun-17 5921-034-235 3745139 6/19/2015 Jun-17 5921-034-235 3746996 6/19/2015 Jun-17 5921-218-235 3739286 6/19/2015 Jun-17 5921-218-235 3742463 6/19/2015 Jun-17 5921-218-235 3746223 6/19/2015 Jun-17 5921-218-235 3747217 6/19/2015 Jun-17 5921-044-235 3747785 6/19/2015 Jun-17 5921-218-135 3744525 6/19/2015 Jun-17 5921-218-135 3746402 6/19/2015 Jun-17 5921-218-135 3746920 6/19/2015 Jun-17 5921-218-236 3744882 6/19/2015 Jun-17 60-7070-103 3742471 6/19/2015 Jun-17 60-7070-103 3742542 6/19/2015 Jun-17 60-7070-103 3744518 6/19/2015 Jun-17 60-7070-103 3744797 6/19/2015 Jun-17 60-7070-103 3745540 6/19/2015 Jun-17 60-7070-103 3745984 6/19/2015 Jun-17 60-7070-103 3747507 6/19/2015 Jun-17 60-7070-105 3743736 6/19/2015 Jun-17 60-7070-105 3745118 6/19/2015 Jun-17 60-7070-105 3745538 6/19/2015 Jun-17 60-7070-105 3745978 6/19/2015 Jun-17 60-7070-105 3747168 6/19/2015 Jun-17 60-7070-104 3742459 6/19/2015 Jun-17 60-7070-104 3743974 6/19/2015 Jun-17 60-7070-104 3747152 6/19/2015 Jun-17 60-7070-006 3734734 6/19/2015 Jun-17 60-7070-106 3743082 6/19/2015 Jun-17 60-7070-106 3744659 6/19/2015 Jun-17 60-7070-106 3746129 6/19/2015 Jun-17 60-7070-106 3747970 6/19/2015 Jun-17 60-7070-107 3743076 6/19/2015 Jun-17 60-7070-107 3745322 6/19/2015 Jun-17 60-7070-107 3747792 6/19/2015 Jun-17 60-7070-107 3748090 6/19/2015 Jun-17 60-7075-006 3744231 6/19/2015 Jun-17 60-7075-103 3744502 6/19/2015 Jun-17 60-7075-103 3746186 6/19/2015 Jun-17 60-7075-104 3744876 6/19/2015 Jun-17 60-7075-106 3742590 6/19/2015 Jun-17 60-7075-106 3744905 6/19/2015 Jun-17 5921-018-135 3751513 6/22/2015 Jun-17 5921-018-235 3748159 6/22/2015 Jun-17 5921-024-135 3748504 6/22/2015 Jun-17 5921-024-235 3748137 6/22/2015 Jun-17 5921-024-235 3751634 6/22/2015 Jun-17 5921-030-135 3748308 6/22/2015 Jun-17 5921-030-135 3749761 6/22/2015 Jun-17 5921-030-135 3754018 6/22/2015 Jun-17 5921-030-235 3749676 6/22/2015 Jun-17 5921-030-235 3751794 6/22/2015 Jun-17 5921-030-235 3752687 6/22/2015 Jun-17 5921-034-135 3749757 6/22/2015 Jun-17 5921-034-135 3751623 6/22/2015 Jun-17 5921-034-135 3752589 6/22/2015 Jun-17 5921-034-235 3747008 6/22/2015 Jun-17 5921-034-235 3750505 6/22/2015 Jun-17 5921-034-235 3750514 6/22/2015 Jun-17 5921-034-235 3752128 6/22/2015 Jun-17 5921-034-235 3752710 6/22/2015 Jun-17 5921-034-235 3753989 6/22/2015 Jun-17 5921-218-235 3749420 6/22/2015 Jun-17 5921-218-235 3750994 6/22/2015 Jun-17 5921-218-235 3752144 6/22/2015 Jun-17 5921-218-235 3752464 6/22/2015 Jun-17 5921-218-235 3752706 6/22/2015 Jun-17 5921-218-135 3751532 6/22/2015 Jun-17 5921-218-135 3751647 6/22/2015 Jun-17 60-7070-103 3747511 6/22/2015 Jun-17 60-7070-103 3749166 6/22/2015 Jun-17 60-7070-103 3751022 6/22/2015 Jun-17 60-7070-103 3751862 6/22/2015 Jun-17 60-7070-103 3753697 6/22/2015 Jun-17 60-7070-105 3749204 6/22/2015 Jun-17 60-7070-105 3750477 6/22/2015 Jun-17 60-7070-105 3751411 6/22/2015 Jun-17 60-7070-105 3752349 6/22/2015 Jun-17 60-7070-105 3752673 6/22/2015 Jun-17 60-7070-105 3753700 6/22/2015 Jun-17 60-7070-104 3752493 6/22/2015 Jun-17 60-7070-006 3750871 6/22/2015 Jun-17 60-7070-106 3748515 6/22/2015 Jun-17 60-7070-106 3749598 6/22/2015 Jun-17 60-7070-106 3751183 6/22/2015 Jun-17 60-7070-106 3751858 6/22/2015 Jun-17 60-7070-106 3752930 6/22/2015 Jun-17 60-7070-106 3755443 6/22/2015 Jun-17 60-7070-107 3750844 6/22/2015 Jun-17 60-7075-006 3751203 6/22/2015 Jun-17 60-7075-103 3748775 6/22/2015 Jun-17 60-7075-103 3748953 6/22/2015 Jun-17 60-7075-103 3749580 6/22/2015 Jun-17 60-7075-103 3752137 6/22/2015 Jun-17 60-7075-104 3749454 6/22/2015 Jun-17 60-7075-106 3748472 6/22/2015 Jun-17 60-7075-106 3750099 6/22/2015 Jun-17 60-7075-106 3752354 6/22/2015 Jun-17 5921-018-135 3755713 6/24/2015 Jun-17 5921-018-135 3756615 6/24/2015 Jun-17 5921-024-135 3755685 6/24/2015 Jun-17 5921-024-135 3760143 6/24/2015 Jun-17 5921-024-135 3760535 6/24/2015 Jun-17 5921-024-235 3756593 6/24/2015 Jun-17 5921-030-135 3752562 6/24/2015 Jun-17 5921-030-135 3758592 6/24/2015 Jun-17 5921-030-235 3756118 6/24/2015 Jun-17 5921-030-235 3758366 6/24/2015 Jun-17 5921-030-235 3761377 6/24/2015 Jun-17 5921-034-135 3755692 6/24/2015 Jun-17 5921-034-135 3755914 6/24/2015 Jun-17 5921-034-135 3762316 6/24/2015 Jun-17 5921-034-235 3756230 6/24/2015 Jun-17 5921-034-235 3759840 6/24/2015 Jun-17 5921-034-235 3760981 6/24/2015 Jun-17 5921-034-236 3759432 6/24/2015 Jun-17 5921-218-235 3756000 6/24/2015 Jun-17 5921-218-235 3760416 6/24/2015 Jun-17 5921-218-235 3761809 6/24/2015 Jun-17 5921-030-236 3759683 6/24/2015 Jun-17 5921-044-135 3759409 6/24/2015 Jun-17 5921-218-135 3755750 6/24/2015 Jun-17 5921-218-135 3761953 6/24/2015 Jun-17 60-7070-102 3759394 6/24/2015 Jun-17 60-7070-102 3759631 6/24/2015 Jun-17 60-7070-103 3753678 6/24/2015 Jun-17 60-7070-103 3758635 6/24/2015 Jun-17 60-7070-103 3760311 6/24/2015 Jun-17 60-7070-103 3761591 6/24/2015 Jun-17 60-7070-105 3758439 6/24/2015 Jun-17 60-7070-105 3759736 6/24/2015 Jun-17 60-7070-105 3761298 6/24/2015 Jun-17 60-7070-104 3758879 6/24/2015 Jun-17 60-7070-003 3761233 6/24/2015 Jun-17 60-7070-003 3761775 6/24/2015 Jun-17 60-7070-006 3756540 6/24/2015 Jun-17 60-7070-106 3755594 6/24/2015 Jun-17 60-7070-106 3758724 6/24/2015 Jun-17 60-7070-106 3759855 6/24/2015 Jun-17 60-7070-106 3761042 6/24/2015 Jun-17 60-7070-107 3756501 6/24/2015 Jun-17 60-7070-107 3762328 6/24/2015 Jun-17 60-7075-003 3759244 6/24/2015 Jun-17 60-7075-005 3758949 6/24/2015 Jun-17 60-7075-006 3762219 6/24/2015 Jun-17 60-7075-103 3759988 6/24/2015 Jun-17 60-7075-103 3760162 6/24/2015 Jun-17 60-7075-103 3761948 6/24/2015 Jun-17 60-7075-104 3758880 6/24/2015 Jun-17 60-7075-106 3760115 6/24/2015 Jun-17 60-7075-106 3762226 6/24/2015 Jun-17 5921-018-135 3763881 6/29/2015 Jun-17 5921-018-135 3766684 6/29/2015 Jun-17 5921-018-235 3766716 6/29/2015 Jun-17 5921-024-235 3764355 6/29/2015 Jun-17 5921-030-135 3762469 6/29/2015 Jun-17 5921-030-135 3767500 6/29/2015 Jun-17 5921-030-235 3765394 6/29/2015 Jun-17 5921-030-235 3767004 6/29/2015 Jun-17 5921-034-135 3766818 6/29/2015 Jun-17 5921-034-135 3766887 6/29/2015 Jun-17 5921-034-235 3764034 6/29/2015 Jun-17 5921-034-235 3765244 6/29/2015 Jun-17 5921-034-235 3765259 6/29/2015 Jun-17 5921-034-235 3766498 6/29/2015 Jun-17 5921-218-235 3765581 6/29/2015 Jun-17 5921-218-235 3767585 6/29/2015 Jun-17 5921-218-135 3763720 6/29/2015 Jun-17 5921-218-135 3763852 6/29/2015 Jun-17 60-7070-103 3761464 6/29/2015 Jun-17 60-7070-103 3763322 6/29/2015 Jun-17 60-7070-103 3763501 6/29/2015 Jun-17 60-7070-103 3764514 6/29/2015 Jun-17 60-7070-103 3765893 6/29/2015 Jun-17 60-7070-105 3762482 6/29/2015 Jun-17 60-7070-105 3762858 6/29/2015 Jun-17 60-7070-105 3764454 6/29/2015 Jun-17 60-7070-105 3766175 6/29/2015 Jun-17 60-7070-105 3766414 6/29/2015 Jun-17 60-7070-104 3763978 6/29/2015 Jun-17 60-7070-106 3762889 6/29/2015 Jun-17 60-7070-106 3763173 6/29/2015 Jun-17 60-7070-106 3764816 6/29/2015 Jun-17 60-7070-106 3765007 6/29/2015 Jun-17 60-7070-106 3767172 6/29/2015 Jun-17 60-7070-106 3767350 6/29/2015 Jun-17 60-7070-107 3765356 6/29/2015 Jun-17 60-7070-107 3767117 6/29/2015 Jun-17 60-7075-103 3764726 6/29/2015 Jun-17 60-7075-104 3766621 6/29/2015 Jun-17 60-7075-106 3764243 6/29/2015 Jun-17 60-7075-106 3767674 6/29/2015 Jun-17 5921-018-135 3772217 6/30/2015 Jun-17 5921-018-235 3769685 6/30/2015 Jun-17 5921-024-135 3768803 6/30/2015 Jun-17 5921-024-135 3770488 6/30/2015 Jun-17 5921-024-235 3769462 6/30/2015 Jun-17 5921-024-235 3769659 6/30/2015 Jun-17 5921-030-135 3769811 6/30/2015 Jun-17 5921-030-135 3770731 6/30/2015 Jun-17 5921-030-235 3768577 6/30/2015 Jun-17 5921-030-235 3771518 6/30/2015 Jun-17 5921-034-135 3769017 6/30/2015 Jun-17 5921-034-135 3769212 6/30/2015 Jun-17 5921-034-135 3770670 6/30/2015 Jun-17 5921-034-135 3770852 6/30/2015 Jun-17 5921-034-135 3771848 6/30/2015 Jun-17 5921-034-136 3772414 6/30/2015 Jun-17 5921-034-235 3768315 6/30/2015 Jun-17 5921-034-235 3769368 6/30/2015 Jun-17 5921-034-235 3770945 6/30/2015 Jun-17 5921-218-235 3768959 6/30/2015 Jun-17 5921-218-235 3770230 6/30/2015 Jun-17 5921-218-235 3771573 6/30/2015 Jun-17 5921-030-236 3772054 6/30/2015 Jun-17 5921-218-135 3767782 6/30/2015 Jun-17 5921-218-135 3769086 6/30/2015 Jun-17 5921-218-135 3770777 6/30/2015 Jun-17 5921-218-236 3772357 6/30/2015 Jun-17 60-7070-103 3767983 6/30/2015 Jun-17 60-7070-103 3769750 6/30/2015 Jun-17 60-7070-105 3768475 6/30/2015 Jun-17 60-7070-105 3769265 6/30/2015 Jun-17 60-7070-105 3770063 6/30/2015 Jun-17 60-7070-105 3771347 6/30/2015 Jun-17 60-7070-105 3773799 6/30/2015 Jun-17 60-7070-006 3772008 6/30/2015 Jun-17 60-7070-106 3768583 6/30/2015 Jun-17 60-7070-106 3769882 6/30/2015 Jun-17 60-7070-107 3772609 6/30/2015 Jun-17 60-7075-003 3772152 6/30/2015 Jun-17 60-7075-006 3771254 6/30/2015 Jun-17 60-7075-103 3770555 6/30/2015 Jun-17 60-7075-103 3771587 6/30/2015 Jun-17 60-7075-106 3767724 6/30/2015 Jun-17 60-7075-106 3768786 6/30/2015 Jun-17 60-7075-106 3775623 6/30/2015 Jun-17 5921-030-235 3730373 7/2/2015 Jul-17 5921-018-235 3773074 7/6/2015 Jul-17 5921-024-235 3772632 7/6/2015 Jul-17 5921-024-235 3779309 7/6/2015 Jul-17 5921-024-236 3777485 7/6/2015 Jul-17 5921-030-136 3777588 7/6/2015 Jul-17 5921-030-136 3780142 7/6/2015 Jul-17 5921-030-235 3777060 7/6/2015 Jul-17 5921-030-235 3777722 7/6/2015 Jul-17 5921-030-235 3780040 7/6/2015 Jul-17 5921-034-135 3778394 7/6/2015 Jul-17 5921-034-135 3778421 7/6/2015 Jul-17 5921-034-136 3777395 7/6/2015 Jul-17 5921-034-235 3772645 7/6/2015 Jul-17 5921-034-235 3772958 7/6/2015 Jul-17 5921-034-235 3776023 7/6/2015 Jul-17 5921-034-235 3778935 7/6/2015 Jul-17 5921-218-235 3776204 7/6/2015 Jul-17 5921-218-235 3777780 7/6/2015 Jul-17 5921-030-236 3777400 7/6/2015 Jul-17 5921-218-135 3776859 7/6/2015 Jul-17 60-7070-102 3773090 7/6/2015 Jul-17 60-7070-103 3773019 7/6/2015 Jul-17 60-7070-103 3773137 7/6/2015 Jul-17 60-7070-103 3773159 7/6/2015 Jul-17 60-7070-103 3777041 7/6/2015 Jul-17 60-7070-103 3777655 7/6/2015 Jul-17 60-7070-103 3780310 7/6/2015 Jul-17 60-7070-105 3778271 7/6/2015 Jul-17 60-7070-105 3779795 7/6/2015 Jul-17 60-7070-105 3779805 7/6/2015 Jul-17 60-7070-104 3773349 7/6/2015 Jul-17 60-7070-104 3777849 7/6/2015 Jul-17 60-7070-104 3779528 7/6/2015 Jul-17 60-7070-003 3776412 7/6/2015 Jul-17 60-7070-005 3773515 7/6/2015 Jul-17 60-7070-106 3773570 7/6/2015 Jul-17 60-7070-106 3777975 7/6/2015 Jul-17 60-7070-106 3778053 7/6/2015 Jul-17 60-7070-106 3779439 7/6/2015 Jul-17 60-7075-004 3776408 7/6/2015 Jul-17 60-7075-005 3776719 7/6/2015 Jul-17 60-7075-006 3777307 7/6/2015 Jul-17 60-7075-103 3776579 7/6/2015 Jul-17 60-7075-105 3778180 7/6/2015 Jul-17 60-7075-106 3776047 7/6/2015 Jul-17 60-7075-106 3778612 7/6/2015 Jul-17 5921-024-235 3785075 7/8/2015 Jul-17 5921-030-135 3783635 7/8/2015 Jul-17 5921-030-235 3784756 7/8/2015 Jul-17 5921-034-135 3781395 7/8/2015 Jul-17 5921-034-135 3785145 7/8/2015 Jul-17 5921-034-135 3785800 7/8/2015 Jul-17 5921-034-235 3782013 7/8/2015 Jul-17 5921-034-235 3782260 7/8/2015 Jul-17 5921-034-235 3784337 7/8/2015 Jul-17 5921-034-235 3786432 7/8/2015 Jul-17 5921-218-235 3781060 7/8/2015 Jul-17 5921-218-235 3784476 7/8/2015 Jul-17 5921-218-235 3786122 7/8/2015 Jul-17 5921-218-135 3781209 7/8/2015 Jul-17 5921-218-135 3784940 7/8/2015 Jul-17 5921-218-135 3785047 7/8/2015 Jul-17 60-7070-102 3782835 7/8/2015 Jul-17 60-7070-103 3780663 7/8/2015 Jul-17 60-7070-103 3780914 7/8/2015 Jul-17 60-7070-103 3783070 7/8/2015 Jul-17 60-7070-103 3785546 7/8/2015 Jul-17 60-7070-103 3785757 7/8/2015 Jul-17 60-7070-105 3781250 7/8/2015 Jul-17 60-7070-105 3781311 7/8/2015 Jul-17 60-7070-105 3782729 7/8/2015 Jul-17 60-7070-105 3784584 7/8/2015 Jul-17 60-7070-104 3781812 7/8/2015 Jul-17 60-7070-104 3785956 7/8/2015 Jul-17 60-7070-006 3784681 7/8/2015 Jul-17 60-7070-106 3781557 7/8/2015 Jul-17 60-7070-106 3784051 7/8/2015 Jul-17 60-7070-106 3784242 7/8/2015 Jul-17 60-7070-106 3786506 7/8/2015 Jul-17 60-7070-107 3650711 7/8/2015 Jul-17 60-7070-107 3781780 7/8/2015 Jul-17 60-7075-003 3786271 7/8/2015 Jul-17 60-7075-103 3781714 7/8/2015 Jul-17 60-7075-103 3784836 7/8/2015 Jul-17 60-7075-106 3722551 7/8/2015 Jul-17 60-7075-106 3781896 7/8/2015 Jul-17 60-7070-103 3750006 7/9/2015 Jul-17 5921-018-135 3795043 7/10/2015 Jul-17 5921-018-235 3793567 7/10/2015 Jul-17 5921-024-135 3794028 7/10/2015 Jul-17 5921-024-235 3791962 7/10/2015 Jul-17 5921-024-235 3793788 7/10/2015 Jul-17 5921-030-235 3796222 7/10/2015 Jul-17 5921-034-135 3791958 7/10/2015 Jul-17 5921-034-235 3796163 7/10/2015 Jul-17 5921-034-235 3796464 7/10/2015 Jul-17 5921-218-235 3793420 7/10/2015 Jul-17 5921-218-235 3794586 7/10/2015 Jul-17 5921-218-235 3795418 7/10/2015 Jul-17 5921-218-135 3794754 7/10/2015 Jul-17 5921-218-136 3795036 7/10/2015 Jul-17 60-7070-102 3792126 7/10/2015 Jul-17 60-7070-102 3792560 7/10/2015 Jul-17 60-7070-102 3794295 7/10/2015 Jul-17 60-7070-103 3792659 7/10/2015 Jul-17 60-7070-103 3793340 7/10/2015 Jul-17 60-7070-103 3793925 7/10/2015 Jul-17 60-7070-103 3794276 7/10/2015 Jul-17 60-7070-103 3794670 7/10/2015 Jul-17 60-7070-103 3795164 7/10/2015 Jul-17 60-7070-103 3795397 7/10/2015 Jul-17 60-7070-103 3796558 7/10/2015 Jul-17 60-7070-105 3792142 7/10/2015 Jul-17 60-7070-105 3792587 7/10/2015 Jul-17 60-7070-105 3792616 7/10/2015 Jul-17 60-7070-105 3794488 7/10/2015 Jul-17 60-7070-105 3795769 7/10/2015 Jul-17 60-7070-105 3796057 7/10/2015 Jul-17 60-7070-104 3793607 7/10/2015 Jul-17 60-7070-104 3794810 7/10/2015 Jul-17 60-7070-003 3793682 7/10/2015 Jul-17 60-7070-106 3791936 7/10/2015 Jul-17 60-7070-106 3795494 7/10/2015 Jul-17 60-7070-106 3795690 7/10/2015 Jul-17 60-7075-103 3794322 7/10/2015 Jul-17 60-7075-103 3795155 7/10/2015 Jul-17 60-7075-106 3795977 7/10/2015 Jul-17 60-7075-106 3796159 7/10/2015 Jul-17 5921-018-135 3787982 7/13/2015 Jul-17 5921-018-135 3791894 7/13/2015 Jul-17 5921-018-235 3787003 7/13/2015 Jul-17 5921-018-235 3789749 7/13/2015 Jul-17 5921-018-235 3791818 7/13/2015 Jul-17 5921-024-135 3787884 7/13/2015 Jul-17 5921-024-135 3790888 7/13/2015 Jul-17 5921-030-135 3788569 7/13/2015 Jul-17 5921-030-135 3788856 7/13/2015 Jul-17 5921-030-135 3791077 7/13/2015 Jul-17 5921-030-235 3791326 7/13/2015 Jul-17 5921-034-135 3790021 7/13/2015 Jul-17 5921-034-135 3790992 7/13/2015 Jul-17 5921-034-235 3788374 7/13/2015 Jul-17 5921-034-235 3788524 7/13/2015 Jul-17 5921-034-235 3790646 7/13/2015 Jul-17 5921-034-235 3791650 7/13/2015 Jul-17 5921-218-235 3787233 7/13/2015 Jul-17 5921-218-235 3787349 7/13/2015 Jul-17 5921-218-235 3787429 7/13/2015 Jul-17 5921-218-235 3790729 7/13/2015 Jul-17 5921-218-235 3791467 7/13/2015 Jul-17 5921-218-235 3791621 7/13/2015 Jul-17 5921-218-135 3787527 7/13/2015 Jul-17 5921-218-135 3790959 7/13/2015 Jul-17 5921-218-136 3789636 7/13/2015 Jul-17 60-7070-103 3787558 7/13/2015 Jul-17 60-7070-103 3791152 7/13/2015 Jul-17 60-7070-103 3791264 7/13/2015 Jul-17 60-7070-105 3787747 7/13/2015 Jul-17 60-7070-105 3790475 7/13/2015 Jul-17 60-7070-104 3789379 7/13/2015 Jul-17 60-7070-104 3790740 7/13/2015 Jul-17 60-7070-003 3791645 7/13/2015 Jul-17 60-7070-005 3789984 7/13/2015 Jul-17 60-7070-006 3789586 7/13/2015 Jul-17 60-7070-106 3786743 7/13/2015 Jul-17 60-7070-106 3788065 7/13/2015 Jul-17 60-7070-106 3790130 7/13/2015 Jul-17 60-7070-106 3790370 7/13/2015 Jul-17 60-7070-106 3791927 7/13/2015 Jul-17 60-7070-107 3790835 7/13/2015 Jul-17 60-7070-107 3791666 7/13/2015 Jul-17 60-7075-103 3787137 7/13/2015 Jul-17 60-7075-103 3789156 7/13/2015 Jul-17 60-7075-103 3789333 7/13/2015 Jul-17 60-7075-103 3791357 7/13/2015 Jul-17 60-7075-105 3789847 7/13/2015 Jul-17 60-7075-106 3789461 7/13/2015 Jul-17 60-7075-106 3791433 7/13/2015 Jul-17 5921-018-135 3800888 7/15/2015 Jul-17 5921-018-135 3800900 7/15/2015 Jul-17 5921-018-135 3801286 7/15/2015 Jul-17 5921-018-235 3799972 7/15/2015 Jul-17 5921-018-235 3800218 7/15/2015 Jul-17 5921-024-135 3801856 7/15/2015 Jul-17 5921-030-135 3799120 7/15/2015 Jul-17 5921-030-235 3799079 7/15/2015 Jul-17 5921-030-235 3799108 7/15/2015 Jul-17 5921-034-135 3798808 7/15/2015 Jul-17 5921-034-235 3798369 7/15/2015 Jul-17 5921-034-235 3799702 7/15/2015 Jul-17 5921-034-235 3799725 7/15/2015 Jul-17 5921-034-235 3801293 7/15/2015 Jul-17 5921-218-235 3796995 7/15/2015 Jul-17 5921-218-235 3797886 7/15/2015 Jul-17 5921-218-235 3800327 7/15/2015 Jul-17 5921-218-235 3801201 7/15/2015 Jul-17 5921-218-135 3797762 7/15/2015 Jul-17 5921-218-135 3798714 7/15/2015 Jul-17 60-7070-103 3797903 7/15/2015 Jul-17 60-7070-103 3798588 7/15/2015 Jul-17 60-7070-103 3799226 7/15/2015 Jul-17 60-7070-103 3800842 7/15/2015 Jul-17 60-7070-103 3801480 7/15/2015 Jul-17 60-7070-105 3798751 7/15/2015 Jul-17 60-7070-105 3798779 7/15/2015 Jul-17 60-7070-105 3799652 7/15/2015 Jul-17 60-7070-105 3801524 7/15/2015 Jul-17 60-7070-105 3801530 7/15/2015 Jul-17 60-7070-104 3797654 7/15/2015 Jul-17 60-7070-104 3800698 7/15/2015 Jul-17 60-7070-003 3800481 7/15/2015 Jul-17 60-7070-106 3798297 7/15/2015 Jul-17 60-7070-106 3798395 7/15/2015 Jul-17 60-7070-106 3799385 7/15/2015 Jul-17 60-7070-106 3799509 7/15/2015 Jul-17 60-7070-106 3801873 7/15/2015 Jul-17 60-7070-107 3801576 7/15/2015 Jul-17 60-7075-103 3796589 7/15/2015 Jul-17 60-7075-103 3797893 7/15/2015 Jul-17 60-7075-103 3799940 7/15/2015 Jul-17 60-7075-105 3801175 7/15/2015 Jul-17 60-7075-106 3799870 7/15/2015 Jul-17 5921-024-235 3803270 7/16/2015 Jul-17 5921-030-135 3803132 7/16/2015 Jul-17 5921-030-135 3804085 7/16/2015 Jul-17 5921-030-235 3802594 7/16/2015 Jul-17 5921-030-235 3803939 7/16/2015 Jul-17 5921-030-235 3805439 7/16/2015 Jul-17 5921-030-235 3806377 7/16/2015 Jul-17 5921-034-135 3803607 7/16/2015 Jul-17 5921-034-135 3803757 7/16/2015 Jul-17 5921-034-135 3804194 7/16/2015 Jul-17 5921-034-135 3804253 7/16/2015 Jul-17 5921-034-135 3804461 7/16/2015 Jul-17 5921-034-135 3806298 7/16/2015 Jul-17 5921-034-235 3802231 7/16/2015 Jul-17 5921-034-235 3803533 7/16/2015 Jul-17 5921-034-235 3805880 7/16/2015 Jul-17 5921-034-235 3806019 7/16/2015 Jul-17 5921-218-235 3804586 7/16/2015 Jul-17 5921-218-235 3805268 7/16/2015 Jul-17 5921-218-235 3806128 7/16/2015 Jul-17 5921-218-135 3803262 7/16/2015 Jul-17 5921-218-135 3804063 7/16/2015 Jul-17 60-7070-102 3803204 7/16/2015 Jul-17 60-7070-103 3804564 7/16/2015 Jul-17 60-7070-103 3806057 7/16/2015 Jul-17 60-7070-105 3802587 7/16/2015 Jul-17 60-7070-105 3803598 7/16/2015 Jul-17 60-7070-105 3803653 7/16/2015 Jul-17 60-7070-105 3804160 7/16/2015 Jul-17 60-7070-105 3806401 7/16/2015 Jul-17 60-7070-105 3806602 7/16/2015 Jul-17 60-7070-104 3803197 7/16/2015 Jul-17 60-7070-104 3805206 7/16/2015 Jul-17 60-7070-106 3801970 7/16/2015 Jul-17 60-7070-106 3803408 7/16/2015 Jul-17 60-7070-106 3803960 7/16/2015 Jul-17 60-7070-106 3804501 7/16/2015 Jul-17 60-7070-106 3806192 7/16/2015 Jul-17 60-7070-106 3806710 7/16/2015 Jul-17 60-7070-107 3804102 7/16/2015 Jul-17 60-7070-107 3804118 7/16/2015 Jul-17 60-7080-102 3804139 7/16/2015 Jul-17 60-7075-103 3805833 7/16/2015 Jul-17 60-7075-104 3805193 7/16/2015 Jul-17 60-7075-105 3802830 7/16/2015 Jul-17 60-7075-106 3802998 7/16/2015 Jul-17 60-7075-106 3803973 7/16/2015 Jul-17 60-7075-106 3805814 7/16/2015 Jul-17 60-7075-107 3803109 7/16/2015 Jul-17 5921-018-135 3807506 7/20/2015 Jul-17 5921-018-235 3807736 7/20/2015 Jul-17 5921-024-235 3807893 7/20/2015 Jul-17 5921-024-235 3807926 7/20/2015 Jul-17 5921-030-235 3809300 7/20/2015 Jul-17 5921-034-135 3808136 7/20/2015 Jul-17 5921-034-235 3806948 7/20/2015 Jul-17 5921-034-235 3808213 7/20/2015 Jul-17 5921-034-235 3809858 7/20/2015 Jul-17 5921-218-235 3808923 7/20/2015 Jul-17 5921-218-235 3810016 7/20/2015 Jul-17 5921-218-235 3810173 7/20/2015 Jul-17 5921-218-135 3809495 7/20/2015 Jul-17 60-7070-103 3807510 7/20/2015 Jul-17 60-7070-103 3807575 7/20/2015 Jul-17 60-7070-103 3808018 7/20/2015 Jul-17 60-7070-103 3809509 7/20/2015 Jul-17 60-7070-103 3809623 7/20/2015 Jul-17 60-7070-103 3810211 7/20/2015 Jul-17 60-7070-105 3808705 7/20/2015 Jul-17 60-7070-105 3809938 7/20/2015 Jul-17 60-7070-105 3809964 7/20/2015 Jul-17 60-7070-105 3810276 7/20/2015 Jul-17 60-7070-105 3810367 7/20/2015 Jul-17 60-7070-104 3807801 7/20/2015 Jul-17 60-7070-104 3807942 7/20/2015 Jul-17 60-7070-104 3809773 7/20/2015 Jul-17 60-7070-003 3809469 7/20/2015 Jul-17 60-7070-004 3809165 7/20/2015 Jul-17 60-7070-106 3807048 7/20/2015 Jul-17 60-7070-106 3808284 7/20/2015 Jul-17 60-7070-106 3808311 7/20/2015 Jul-17 60-7070-106 3808535 7/20/2015 Jul-17 60-7070-106 3808812 7/20/2015 Jul-17 60-7070-106 3809972 7/20/2015 Jul-17 60-7075-006 3807421 7/20/2015 Jul-17 60-7075-103 3809242 7/20/2015 Jul-17 5921-018-135 3810932 7/22/2015 Jul-17 5921-018-135 3813775 7/22/2015 Jul-17 5921-018-235 3813591 7/22/2015 Jul-17 5921-024-235 3811541 7/22/2015 Jul-17 5921-024-235 3811637 7/22/2015 Jul-17 5921-024-235 3814499 7/22/2015 Jul-17 5921-024-235 3814520 7/22/2015 Jul-17 5921-030-135 3812678 7/22/2015 Jul-17 5921-030-135 3812744 7/22/2015 Jul-17 5921-030-135 3813813 7/22/2015 Jul-17 5921-030-235 3811038 7/22/2015 Jul-17 5921-030-235 3811093 7/22/2015 Jul-17 5921-030-235 3811853 7/22/2015 Jul-17 5921-030-235 3811895 7/22/2015 Jul-17 5921-030-235 3813308 7/22/2015 Jul-17 5921-030-235 3813866 7/22/2015 Jul-17 5921-034-135 3810488 7/22/2015 Jul-17 5921-034-135 3812077 7/22/2015 Jul-17 5921-034-135 3813660 7/22/2015 Jul-17 5921-034-135 3813911 7/22/2015 Jul-17 5921-034-235 3811231 7/22/2015 Jul-17 5921-034-235 3812096 7/22/2015 Jul-17 5921-034-235 3812240 7/22/2015 Jul-17 5921-034-235 3813641 7/22/2015 Jul-17 5921-218-235 3812802 7/22/2015 Jul-17 5921-218-135 3811663 7/22/2015 Jul-17 5921-218-135 3812991 7/22/2015 Jul-17 5921-218-135 3813144 7/22/2015 Jul-17 5921-218-135 3813723 7/22/2015 Jul-17 60-7070-103 3814294 7/22/2015 Jul-17 60-7070-103 3814434 7/22/2015 Jul-17 60-7070-105 3810380 7/22/2015 Jul-17 60-7070-105 3811713 7/22/2015 Jul-17 60-7070-105 3811745 7/22/2015 Jul-17 60-7070-105 3812303 7/22/2015 Jul-17 60-7070-105 3813472 7/22/2015 Jul-17 60-7070-105 3813878 7/22/2015 Jul-17 60-7070-104 3814565 7/22/2015 Jul-17 60-7070-106 3811922 7/22/2015 Jul-17 60-7070-106 3812020 7/22/2015 Jul-17 60-7070-106 3813205 7/22/2015 Jul-17 60-7070-107 3811375 7/22/2015 Jul-17 60-7070-107 3813706 7/22/2015 Jul-17 60-7075-003 3811001 7/22/2015 Jul-17 60-7075-003 3811208 7/22/2015 Jul-17 60-7075-106 3810398 7/22/2015 Jul-17 60-7075-106 3810754 7/22/2015 Jul-17 60-7075-106 3814531 7/22/2015 Jul-17 5921-018-135 3818599 7/23/2015 Jul-17 5921-018-235 3817693 7/23/2015 Jul-17 5921-024-136 3815634 7/23/2015 Jul-17 5921-024-235 3816653 7/23/2015 Jul-17 5921-024-235 3818608 7/23/2015 Jul-17 5921-024-235 3818631 7/23/2015 Jul-17 5921-030-135 3814066 7/23/2015 Jul-17 5921-030-135 3817557 7/23/2015 Jul-17 5921-030-136 3814928 7/23/2015 Jul-17 5921-030-235 3814772 7/23/2015 Jul-17 5921-030-235 3814793 7/23/2015 Jul-17 5921-030-235 3814819 7/23/2015 Jul-17 5921-030-235 3816465 7/23/2015 Jul-17 5921-034-135 3813976 7/23/2015 Jul-17 5921-034-135 3814900 7/23/2015 Jul-17 5921-034-135 3815804 7/23/2015 Jul-17 5921-034-135 3816682 7/23/2015 Jul-17 5921-034-235 3814755 7/23/2015 Jul-17 5921-034-235 3816184 7/23/2015 Jul-17 5921-034-235 3817136 7/23/2015 Jul-17 5921-034-235 3818654 7/23/2015 Jul-17 5921-218-235 3814667 7/23/2015 Jul-17 5921-218-235 3816757 7/23/2015 Jul-17 5921-218-235 3817104 7/23/2015 Jul-17 5921-218-235 3818552 7/23/2015 Jul-17 5921-212-135 3815566 7/23/2015 Jul-17 5921-218-135 3814874 7/23/2015 Jul-17 5921-218-135 3815203 7/23/2015 Jul-17 5921-218-135 3818445 7/23/2015 Jul-17 5921-218-136 3815022 7/23/2015 Jul-17 5921-218-136 3815274 7/23/2015 Jul-17 60-7070-103 3814111 7/23/2015 Jul-17 60-7070-103 3816243 7/23/2015 Jul-17 60-7070-103 3816774 7/23/2015 Jul-17 60-7070-103 3817189 7/23/2015 Jul-17 60-7070-103 3817213 7/23/2015 Jul-17 60-7070-103 3818302 7/23/2015 Jul-17 60-7070-105 3815699 7/23/2015 Jul-17 60-7070-105 3816630 7/23/2015 Jul-17 60-7070-105 3817025 7/23/2015 Jul-17 60-7070-105 3817087 7/23/2015 Jul-17 60-7070-104 3815843 7/23/2015 Jul-17 60-7070-104 3815891 7/23/2015 Jul-17 60-7070-104 3817266 7/23/2015 Jul-17 60-7070-106 3814740 7/23/2015 Jul-17 60-7070-106 3816338 7/23/2015 Jul-17 60-7070-106 3817307 7/23/2015 Jul-17 60-7070-106 3817463 7/23/2015 Jul-17 60-7070-107 3814050 7/23/2015 Jul-17 60-7070-107 3816034 7/23/2015 Jul-17 60-7075-003 3815469 7/23/2015 Jul-17 60-7075-003 3816735 7/23/2015 Jul-17 60-7075-006 3814835 7/23/2015 Jul-17 60-7075-103 3814701 7/23/2015 Jul-17 60-7075-103 3814786 7/23/2015 Jul-17 60-7075-103 3814944 7/23/2015 Jul-17 60-7075-103 3814992 7/23/2015 Jul-17 60-7075-103 3818506 7/23/2015 Jul-17 60-7075-104 3814857 7/23/2015 Jul-17 60-7075-104 3816567 7/23/2015 Jul-17 60-7075-105 3816828 7/23/2015 Jul-17 60-7075-105 3816973 7/23/2015 Jul-17 60-7075-106 3816099 7/23/2015 Jul-17 60-7075-106 3816132 7/23/2015 Jul-17 5921-018-135 3822330 7/28/2015 Jul-17 5921-018-235 3820411 7/28/2015 Jul-17 5921-024-235 3821294 7/28/2015 Jul-17 5921-024-235 3821512 7/28/2015 Jul-17 5921-024-236 3822485 7/28/2015 Jul-17 5921-030-135 3820509 7/28/2015 Jul-17 5921-030-135 3821596 7/28/2015 Jul-17 5921-030-235 3818829 7/28/2015 Jul-17 5921-030-235 3820827 7/28/2015 Jul-17 5921-030-235 3822115 7/28/2015 Jul-17 5921-034-135 3819015 7/28/2015 Jul-17 5921-034-135 3821169 7/28/2015 Jul-17 5921-034-135 3822705 7/28/2015 Jul-17 5921-034-135 3822767 7/28/2015 Jul-17 5921-034-235 3819814 7/28/2015 Jul-17 5921-034-235 3820891 7/28/2015 Jul-17 5921-034-235 3820917 7/28/2015 Jul-17 5921-034-235 3822250 7/28/2015 Jul-17 5921-218-235 3820741 7/28/2015 Jul-17 5921-218-235 3820964 7/28/2015 Jul-17 5921-218-235 3823007 7/28/2015 Jul-17 5921-218-135 3820815 7/28/2015 Jul-17 60-7070-103 3818919 7/28/2015 Jul-17 60-7070-103 3819795 7/28/2015 Jul-17 60-7070-103 3820794 7/28/2015 Jul-17 60-7070-103 3821218 7/28/2015 Jul-17 60-7070-103 3822535 7/28/2015 Jul-17 60-7070-105 3818736 7/28/2015 Jul-17 60-7070-105 3819661 7/28/2015 Jul-17 60-7070-105 3819782 7/28/2015 Jul-17 60-7070-105 3820940 7/28/2015 Jul-17 60-7070-105 3821803 7/28/2015 Jul-17 60-7070-105 3821971 7/28/2015 Jul-17 60-7070-105 3822934 7/28/2015 Jul-17 60-7070-104 3818864 7/28/2015 Jul-17 60-7070-104 3820634 7/28/2015 Jul-17 60-7070-104 3822629 7/28/2015 Jul-17 60-7070-106 3821066 7/28/2015 Jul-17 60-7070-106 3822780 7/28/2015 Jul-17 60-7070-106 3822923 7/28/2015 Jul-17 60-7070-107 3819460 7/28/2015 Jul-17 60-7075-006 3820521 7/28/2015 Jul-17 60-7075-103 3820623 7/28/2015 Jul-17 60-7075-104 3820337 7/28/2015 Jul-17 60-7075-106 3819210 7/28/2015 Jul-17 60-7075-106 3819313 7/28/2015 Jul-17 60-7075-106 3820430 7/28/2015 Jul-17 5921-018-135 3823179 7/29/2015 Jul-17 5921-018-135 3825868 7/29/2015 Jul-17 5921-018-136 3826724 7/29/2015 Jul-17 5921-018-235 3824001 7/29/2015 Jul-17 5921-024-235 3823263 7/29/2015 Jul-17 5921-024-235 3825063 7/29/2015 Jul-17 5921-024-235 3825913 7/29/2015 Jul-17 5921-030-135 3826753 7/29/2015 Jul-17 5921-030-135 3826833 7/29/2015 Jul-17 5921-030-235 3823360 7/29/2015 Jul-17 5921-030-235 3823373 7/29/2015 Jul-17 5921-030-235 3825085 7/29/2015 Jul-17 5921-030-235 3825823 7/29/2015 Jul-17 5921-034-135 3823450 7/29/2015 Jul-17 5921-034-135 3824375 7/29/2015 Jul-17 5921-034-135 3825253 7/29/2015 Jul-17 5921-034-235 3823325 7/29/2015 Jul-17 5921-034-235 3824399 7/29/2015 Jul-17 5921-034-235 3825187 7/29/2015 Jul-17 5921-034-235 3825230 7/29/2015 Jul-17 5921-034-235 3825796 7/29/2015 Jul-17 5921-218-235 3823011 7/29/2015 Jul-17 5921-218-235 3823988 7/29/2015 Jul-17 5921-218-235 3824427 7/29/2015 Jul-17 5921-218-235 3825593 7/29/2015 Jul-17 5921-218-235 3825636 7/29/2015 Jul-17 5921-218-235 3825921 7/29/2015 Jul-17 5921-218-135 3823122 7/29/2015 Jul-17 5921-218-135 3824489 7/29/2015 Jul-17 5921-218-135 3824556 7/29/2015 Jul-17 5921-218-135 3828019 7/29/2015 Jul-17 60-7070-102 3826397 7/29/2015 Jul-17 60-7070-103 3823076 7/29/2015 Jul-17 60-7070-103 3823198 7/29/2015 Jul-17 60-7070-103 3824685 7/29/2015 Jul-17 60-7070-103 3824882 7/29/2015 Jul-17 60-7070-103 3825517 7/29/2015 Jul-17 60-7070-105 3823886 7/29/2015 Jul-17 60-7070-105 3824922 7/29/2015 Jul-17 60-7070-105 3825305 7/29/2015 Jul-17 60-7070-104 3823048 7/29/2015 Jul-17 60-7070-104 3824659 7/29/2015 Jul-17 60-7070-104 3825691 7/29/2015 Jul-17 60-7070-003 3826129 7/29/2015 Jul-17 60-7070-003 3826898 7/29/2015 Jul-17 60-7070-003 3827077 7/29/2015 Jul-17 60-7070-004 3826193 7/29/2015 Jul-17 60-7070-005 3827446 7/29/2015 Jul-17 60-7070-006 3826585 7/29/2015 Jul-17 60-7070-006 3826635 7/29/2015 Jul-17 60-7070-006 3826711 7/29/2015 Jul-17 60-7070-006 3827293 7/29/2015 Jul-17 60-7070-106 3823484 7/29/2015 Jul-17 60-7070-106 3823608 7/29/2015 Jul-17 60-7070-106 3823855 7/29/2015 Jul-17 60-7070-106 3824967 7/29/2015 Jul-17 60-7070-106 3824998 7/29/2015 Jul-17 60-7070-106 3825932 7/29/2015 Jul-17 60-7070-107 3827962 7/29/2015 Jul-17 60-7075-003 3826275 7/29/2015 Jul-17 60-7075-003 3826332 7/29/2015 Jul-17 60-7075-003 3827405 7/29/2015 Jul-17 60-7075-003 3827433 7/29/2015 Jul-17 60-7075-103 3825978 7/29/2015 Jul-17 60-7075-103 3826769 7/29/2015 Jul-17 60-7075-104 3827152 7/29/2015 Jul-17 60-7075-104 3827357 7/29/2015 Jul-17 60-7075-105 3827031 7/29/2015 Jul-17 60-7075-105 3827272 7/29/2015 Jul-17 60-7075-106 3824126 7/29/2015 Jul-17 60-7075-106 3824265 7/29/2015 Jul-17 60-7075-106 3825141 7/29/2015 Jul-17 60-7075-106 3828163 7/29/2015 Jul-17 5921-018-235 3828365 7/30/2015 Jul-17 5921-024-135 3828527 7/30/2015 Jul-17 5921-030-235 3829081 7/30/2015 Jul-17 5921-030-235 3832819 7/30/2015 Jul-17 5921-034-135 3833155 7/30/2015 Jul-17 5921-034-136 3831519 7/30/2015 Jul-17 5921-034-235 3829505 7/30/2015 Jul-17 5921-034-235 3832857 7/30/2015 Jul-17 5921-212-235 3829393 7/30/2015 Jul-17 5921-218-235 3829695 7/30/2015 Jul-17 5921-218-235 3830853 7/30/2015 Jul-17 5921-218-235 3832520 7/30/2015 Jul-17 5921-218-235 3832583 7/30/2015 Jul-17 5921-218-135 3831865 7/30/2015 Jul-17 5921-218-135 3834120 7/30/2015 Jul-17 5921-218-236 3829010 7/30/2015 Jul-17 60-7070-103 3828775 7/30/2015 Jul-17 60-7070-103 3828897 7/30/2015 Jul-17 60-7070-103 3829987 7/30/2015 Jul-17 60-7070-103 3830119 7/30/2015 Jul-17 60-7070-103 3832182 7/30/2015 Jul-17 60-7070-103 3832358 7/30/2015 Jul-17 60-7070-103 3833711 7/30/2015 Jul-17 60-7070-105 3829340 7/30/2015 Jul-17 60-7070-105 3831590 7/30/2015 Jul-17 60-7070-105 3831617 7/30/2015 Jul-17 60-7070-105 3832963 7/30/2015 Jul-17 60-7070-105 3833101 7/30/2015 Jul-17 60-7070-104 3831143 7/30/2015 Jul-17 60-7070-104 3831250 7/30/2015 Jul-17 60-7070-104 3834226 7/30/2015 Jul-17 60-7070-006 3829454 7/30/2015 Jul-17 60-7070-006 3829477 7/30/2015 Jul-17 60-7070-006 3834316 7/30/2015 Jul-17 60-7070-106 3829075 7/30/2015 Jul-17 60-7070-106 3830375 7/30/2015 Jul-17 60-7070-106 3830462 7/30/2015 Jul-17 60-7070-106 3834274 7/30/2015 Jul-17 60-7070-106 3834328 7/30/2015 Jul-17 60-7075-006 3829496 7/30/2015 Jul-17 60-7075-106 3828349 7/30/2015 Jul-17 5921-018-235 3836552 8/5/2015 Aug-17 5921-024-135 3837969 8/5/2015 Aug-17 5921-024-235 3835816 8/5/2015 Aug-17 5921-024-235 3835885 8/5/2015 Aug-17 5921-024-235 3837116 8/5/2015 Aug-17 5921-030-135 3836394 8/5/2015 Aug-17 5921-030-135 3836450 8/5/2015 Aug-17 5921-030-135 3838255 8/5/2015 Aug-17 5921-030-235 3836758 8/5/2015 Aug-17 5921-030-235 3837781 8/5/2015 Aug-17 5921-034-135 3835598 8/5/2015 Aug-17 5921-034-135 3835611 8/5/2015 Aug-17 5921-034-135 3837438 8/5/2015 Aug-17 5921-034-135 3837570 8/5/2015 Aug-17 5921-034-235 3834448 8/5/2015 Aug-17 5921-034-235 3836105 8/5/2015 Aug-17 5921-034-235 3836413 8/5/2015 Aug-17 5921-034-235 3836534 8/5/2015 Aug-17 5921-034-235 3836581 8/5/2015 Aug-17 5921-034-235 3838785 8/5/2015 Aug-17 5921-218-235 3835215 8/5/2015 Aug-17 5921-218-235 3836990 8/5/2015 Aug-17 5921-218-235 3837088 8/5/2015 Aug-17 5921-030-236 3839276 8/5/2015 Aug-17 60-7070-102 3834862 8/5/2015 Aug-17 60-7070-103 3834584 8/5/2015 Aug-17 60-7070-103 3834668 8/5/2015 Aug-17 60-7070-103 3837205 8/5/2015 Aug-17 60-7070-103 3837383 8/5/2015 Aug-17 60-7070-103 3838927 8/5/2015 Aug-17 60-7070-103 3839016 8/5/2015 Aug-17 60-7070-105 3834741 8/5/2015 Aug-17 60-7070-105 3834889 8/5/2015 Aug-17 60-7070-105 3834986 8/5/2015 Aug-17 60-7070-105 3836727 8/5/2015 Aug-17 60-7070-104 3835726 8/5/2015 Aug-17 60-7070-104 3836620 8/5/2015 Aug-17 60-7070-005 3836484 8/5/2015 Aug-17 60-7070-106 3834430 8/5/2015 Aug-17 60-7070-106 3838319 8/5/2015 Aug-17 60-7070-106 3838678 8/5/2015 Aug-17 60-7075-003 3835939 8/5/2015 Aug-17 60-7075-004 3836053 8/5/2015 Aug-17 60-7075-104 3835505 8/5/2015 Aug-17 5921-030-235 3842885 8/7/2015 Aug-17 5921-034-135 3839367 8/7/2015 Aug-17 5921-034-135 3839950 8/7/2015 Aug-17 5921-034-135 3839997 8/7/2015 Aug-17 5921-034-235 3841439 8/7/2015 Aug-17 5921-034-236 3843157 8/7/2015 Aug-17 5921-218-235 3841983 8/7/2015 Aug-17 5921-030-236 3839305 8/7/2015 Aug-17 5921-044-135 3839913 8/7/2015 Aug-17 5921-044-235 3839868 8/7/2015 Aug-17 5921-218-135 3839712 8/7/2015 Aug-17 60-7070-102 3840747 8/7/2015 Aug-17 60-7070-103 3840246 8/7/2015 Aug-17 60-7070-103 3840396 8/7/2015 Aug-17 60-7070-103 3840493 8/7/2015 Aug-17 60-7070-103 3843334 8/7/2015 Aug-17 60-7070-105 3839457 8/7/2015 Aug-17 60-7070-105 3839555 8/7/2015 Aug-17 60-7070-105 3843055 8/7/2015 Aug-17 60-7070-105 3843220 8/7/2015 Aug-17 60-7070-104 3841753 8/7/2015 Aug-17 60-7070-006 3843277 8/7/2015 Aug-17 60-7070-106 3840504 8/7/2015 Aug-17 60-7070-106 3840673 8/7/2015 Aug-17 60-7075-006 3839810 8/7/2015 Aug-17 60-7075-103 3842431 8/7/2015 Aug-17 60-7075-104 3840205 8/7/2015 Aug-17 60-7075-105 3841313 8/7/2015 Aug-17 60-7075-106 3840694 8/7/2015 Aug-17 60-7075-106 3840723 8/7/2015 Aug-17 5921-018-135 3849779 8/10/2015 Aug-17 5921-018-235 3849127 8/10/2015 Aug-17 5921-024-135 3850056 8/10/2015 Aug-17 5921-024-235 3846273 8/10/2015 Aug-17 5921-024-235 3851312 8/10/2015 Aug-17 5921-030-135 3845728 8/10/2015 Aug-17 5921-030-235 3850655 8/10/2015 Aug-17 5921-030-235 3850705 8/10/2015 Aug-17 5921-034-135 3845083 8/10/2015 Aug-17 5921-034-235 3844649 8/10/2015 Aug-17 5921-034-235 3846327 8/10/2015 Aug-17 5921-034-235 3846667 8/10/2015 Aug-17 5921-218-235 3846508 8/10/2015 Aug-17 5921-218-235 3849849 8/10/2015 Aug-17 5921-218-235 3850025 8/10/2015 Aug-17 5921-218-235 3851440 8/10/2015 Aug-17 5921-218-235 3851591 8/10/2015 Aug-17 5921-218-135 3847211 8/10/2015 Aug-17 5921-218-135 3847253 8/10/2015 Aug-17 5921-218-135 3851377 8/10/2015 Aug-17 60-7070-103 3843587 8/10/2015 Aug-17 60-7070-103 3845414 8/10/2015 Aug-17 60-7070-103 3847501 8/10/2015 Aug-17 60-7070-103 3847754 8/10/2015 Aug-17 60-7070-103 3848993 8/10/2015 Aug-17 60-7070-103 3849557 8/10/2015 Aug-17 60-7070-105 3846857 8/10/2015 Aug-17 60-7070-105 3850213 8/10/2015 Aug-17 60-7070-104 3845286 8/10/2015 Aug-17 60-7070-104 3849672 8/10/2015 Aug-17 60-7070-104 3849767 8/10/2015 Aug-17 60-7070-004 3849221 8/10/2015 Aug-17 60-7070-006 3848046 8/10/2015 Aug-17 60-7070-007 3852183 8/10/2015 Aug-17 60-7070-106 3846225 8/10/2015 Aug-17 60-7070-106 3846423 8/10/2015 Aug-17 60-7070-106 3849307 8/10/2015 Aug-17 60-7075-003 3848203 8/10/2015 Aug-17 60-7075-005 3844926 8/10/2015 Aug-17 60-7075-006 3850485 8/10/2015 Aug-17 60-7070-157 3851205 8/10/2015 Aug-17 60-7070-155 3848521 8/10/2015 Aug-17 60-7070-155 3850209 8/10/2015 Aug-17 60-7070-154 3848261 8/10/2015 Aug-17 60-7075-106 3846097 8/10/2015 Aug-17 60-7075-106 3846218 8/10/2015 Aug-17 60-7075-056 3850459 8/10/2015 Aug-17 60-7075-156 3850999 8/10/2015 Aug-17 5921-018-235 3852675 8/12/2015 Aug-17 5921-018-235 3854289 8/12/2015 Aug-17 5921-024-135 3855762 8/12/2015 Aug-17 5921-024-135 3857456 8/12/2015 Aug-17 5921-024-235 3852411 8/12/2015 Aug-17 5921-024-235 3856110 8/12/2015 Aug-17 5921-030-135 3854081 8/12/2015 Aug-17 5921-030-135 3854270 8/12/2015 Aug-17 5921-030-135 3856818 8/12/2015 Aug-17 5921-030-235 3853173 8/12/2015 Aug-17 5921-030-235 3856181 8/12/2015 Aug-17 5921-030-235 3856716 8/12/2015 Aug-17 5921-034-135 3853187 8/12/2015 Aug-17 5921-034-135 3853367 8/12/2015 Aug-17 5921-034-135 3855263 8/12/2015 Aug-17 5921-034-135 3855478 8/12/2015 Aug-17 5921-034-135 3856632 8/12/2015 Aug-17 5921-034-135 3857402 8/12/2015 Aug-17 5921-034-235 3855829 8/12/2015 Aug-17 5921-034-235 3857268 8/12/2015 Aug-17 5921-034-235 3857397 8/12/2015 Aug-17 5921-218-135 3854885 8/12/2015 Aug-17 5921-218-135 3856303 8/12/2015 Aug-17 5921-218-136 3853662 8/12/2015 Aug-17 5921-218-136 3853921 8/12/2015 Aug-17 60-7070-102 3853984 8/12/2015 Aug-17 60-7070-103 3854386 8/12/2015 Aug-17 60-7070-103 3856049 8/12/2015 Aug-17 60-7070-105 3854567 8/12/2015 Aug-17 60-7070-105 3856958 8/12/2015 Aug-17 60-7070-104 3853593 8/12/2015 Aug-17 60-7070-005 3855926 8/12/2015 Aug-17 60-7070-106 3854068 8/12/2015 Aug-17 60-7070-106 3854258 8/12/2015 Aug-17 60-7070-106 3855069 8/12/2015 Aug-17 60-7070-106 3855271 8/12/2015 Aug-17 60-7070-106 3855547 8/12/2015 Aug-17 60-7070-107 3856763 8/12/2015 Aug-17 60-7075-004 3854941 8/12/2015 Aug-17 60-7075-006 3855957 8/12/2015 Aug-17 60-7070-155 3854551 8/12/2015 Aug-17 60-7070-156 3853670 8/12/2015 Aug-17 60-7070-156 3853909 8/12/2015 Aug-17 60-7070-153 3853230 8/12/2015 Aug-17 60-7075-106 3854408 8/12/2015 Aug-17 60-7075-107 3855007 8/12/2015 Aug-17 60-7075-153 3856524 8/12/2015 Aug-17 60-7075-106 3851074 8/14/2015 Aug-17 5921-018-135 3859458 8/17/2015 Aug-17 5921-018-235 3857707 8/17/2015 Aug-17 5921-024-235 3858508 8/17/2015 Aug-17 5921-024-235 3861760 8/17/2015 Aug-17 5921-030-135 3859161 8/17/2015 Aug-17 5921-030-135 3859220 8/17/2015 Aug-17 5921-030-235 3857769 8/17/2015 Aug-17 5921-030-235 3859659 8/17/2015 Aug-17 5921-030-235 3861718 8/17/2015 Aug-17 5921-034-135 3857867 8/17/2015 Aug-17 5921-034-135 3858410 8/17/2015 Aug-17 5921-034-135 3859055 8/17/2015 Aug-17 5921-034-135 3862930 8/17/2015 Aug-17 5921-034-135 3863342 8/17/2015 Aug-17 5921-034-235 3859713 8/17/2015 Aug-17 5921-034-235 3859882 8/17/2015 Aug-17 5921-034-235 3861448 8/17/2015 Aug-17 5921-034-235 3861476 8/17/2015 Aug-17 5921-218-235 3857965 8/17/2015 Aug-17 5921-218-235 3858959 8/17/2015 Aug-17 5921-218-235 3859001 8/17/2015 Aug-17 5921-218-235 3862536 8/17/2015 Aug-17 5921-218-135 3860682 8/17/2015 Aug-17 5921-218-236 3861862 8/17/2015 Aug-17 60-7070-103 3857805 8/17/2015 Aug-17 60-7070-103 3859784 8/17/2015 Aug-17 60-7070-103 3861337 8/17/2015 Aug-17 60-7070-103 3862850 8/17/2015 Aug-17 60-7070-103 3862866 8/17/2015 Aug-17 60-7070-105 3858622 8/17/2015 Aug-17 60-7070-105 3861545 8/17/2015 Aug-17 60-7070-104 3859869 8/17/2015 Aug-17 60-7070-104 3862768 8/17/2015 Aug-17 60-7070-007 3862907 8/17/2015 Aug-17 60-7070-106 3857661 8/17/2015 Aug-17 60-7070-106 3861917 8/17/2015 Aug-17 60-7070-106 3862035 8/17/2015 Aug-17 60-7070-106 3862159 8/17/2015 Aug-17 60-7075-103 3859518 8/17/2015 Aug-17 60-7075-104 3860047 8/17/2015 Aug-17 60-7075-106 3857626 8/17/2015 Aug-17 60-7075-106 3863096 8/17/2015 Aug-17 60-7075-106 3863229 8/17/2015 Aug-17 5921-018-135 3868418 8/19/2015 Aug-17 5921-018-235 3866141 8/19/2015 Aug-17 5921-024-235 3865334 8/19/2015 Aug-17 5921-030-135 3851911 8/19/2015 Aug-17 5921-030-135 3864330 8/19/2015 Aug-17 5921-030-135 3864542 8/19/2015 Aug-17 5921-030-135 3868091 8/19/2015 Aug-17 5921-030-235 3864245 8/19/2015 Aug-17 5921-030-235 3866997 8/19/2015 Aug-17 5921-030-235 3868268 8/19/2015 Aug-17 5921-034-135 3863632 8/19/2015 Aug-17 5921-034-135 3866314 8/19/2015 Aug-17 5921-034-135 3866564 8/19/2015 Aug-17 5921-034-235 3818727 8/19/2015 Aug-17 5921-034-235 3820836 8/19/2015 Aug-17 5921-034-235 3834359 8/19/2015 Aug-17 5921-034-235 3865093 8/19/2015 Aug-17 5921-034-235 3865220 8/19/2015 Aug-17 5921-034-235 3866720 8/19/2015 Aug-17 5921-034-235 3869581 8/19/2015 Aug-17 5921-034-236 3852067 8/19/2015 Aug-17 5921-218-235 3867826 8/19/2015 Aug-17 5921-218-135 3868635 8/19/2015 Aug-17 5921-218-135 3869988 8/19/2015 Aug-17 5921-218-136 3869727 8/19/2015 Aug-17 5921-218-136 3869840 8/19/2015 Aug-17 60-7070-103 3863728 8/19/2015 Aug-17 60-7070-103 3865637 8/19/2015 Aug-17 60-7070-103 3865685 8/19/2015 Aug-17 60-7070-103 3869200 8/19/2015 Aug-17 60-7070-103 3869360 8/19/2015 Aug-17 60-7070-103 3869432 8/19/2015 Aug-17 60-7070-105 3864849 8/19/2015 Aug-17 60-7070-105 3867888 8/19/2015 Aug-17 60-7070-105 3867971 8/19/2015 Aug-17 60-7070-105 3869872 8/19/2015 Aug-17 60-7070-104 3865381 8/19/2015 Aug-17 60-7070-104 3868544 8/19/2015 Aug-17
60-7070-004 3865491 8/19/2015 Aug-17 60-7070-106 3830797 8/19/2015 Aug-17 60-7070-106 3864073 8/19/2015 Aug-17 60-7070-106 3864232 8/19/2015 Aug-17 60-7070-106 3868309 8/19/2015 Aug-17 60-7070-106 3868332 8/19/2015 Aug-17 60-7070-106 3868880 8/19/2015 Aug-17 60-7070-107 3864975 8/19/2015 Aug-17 60-7070-107 3868965 8/19/2015 Aug-17 60-7075-006 3866658 8/19/2015 Aug-17 60-7075-103 3820773 8/19/2015 Aug-17 60-7075-103 3867842 8/19/2015 Aug-17 60-7075-104 3865265 8/19/2015 Aug-17 60-7075-106 3819193 8/19/2015 Aug-17 60-7075-106 3868748 8/19/2015 Aug-17 60-7075-106 3869926 8/19/2015 Aug-17 60-7075-153 3867260 8/19/2015 Aug-17 60-7075-153 3867857 8/19/2015 Aug-17 5921-018-235 3871179 8/21/2015 Aug-17 5921-018-235 3875710 8/21/2015 Aug-17 5921-024-135 3875809 8/21/2015 Aug-17 5921-030-135 3872322 8/21/2015 Aug-17 5921-030-135 3876369 8/21/2015 Aug-17 5921-030-235 3870083 8/21/2015 Aug-17 5921-030-235 3872173 8/21/2015 Aug-17 5921-034-135 3870771 8/21/2015 Aug-17 5921-034-135 3874912 8/21/2015 Aug-17 5921-034-235 3876189 8/21/2015 Aug-17 5921-034-235 3876311 8/21/2015 Aug-17 5921-034-235 3877145 8/21/2015 Aug-17 5921-218-235 3870326 8/21/2015 Aug-17 5921-218-235 3871373 8/21/2015 Aug-17 5921-218-235 3874062 8/21/2015 Aug-17 5921-030-236 3877837 8/21/2015 Aug-17 5921-218-135 3870286 8/21/2015 Aug-17 5921-218-135 3871736 8/21/2015 Aug-17 5921-218-135 3875602 8/21/2015 Aug-17 5921-218-136 3877800 8/21/2015 Aug-17 60-7070-103 3870595 8/21/2015 Aug-17 60-7070-103 3874275 8/21/2015 Aug-17 60-7070-103 3875320 8/21/2015 Aug-17 60-7070-103 3876400 8/21/2015 Aug-17 60-7070-105 3871942 8/21/2015 Aug-17 60-7070-105 3872028 8/21/2015 Aug-17 60-7070-105 3874117 8/21/2015 Aug-17 60-7070-105 3874129 8/21/2015 Aug-17 60-7070-105 3875100 8/21/2015 Aug-17 60-7070-104 3870374 8/21/2015 Aug-17 60-7070-104 3870684 8/21/2015 Aug-17 60-7070-104 3870744 8/21/2015 Aug-17 60-7070-104 3874004 8/21/2015 Aug-17 60-7070-104 3875995 8/21/2015 Aug-17 60-7070-003 3876640 8/21/2015 Aug-17 60-7070-056 3870904 8/21/2015 Aug-17 60-7070-106 3870936 8/21/2015 Aug-17 60-7070-106 3873371 8/21/2015 Aug-17 60-7070-106 3875214 8/21/2015 Aug-17 60-7070-106 3878458 8/21/2015 Aug-17 60-7070-107 3874223 8/21/2015 Aug-17 60-7075-006 3876542 8/21/2015 Aug-17 60-7070-152 3870841 8/21/2015 Aug-17 60-7075-103 3876557 8/21/2015 Aug-17 60-7075-103 3876617 8/21/2015 Aug-17 60-7075-104 3876664 8/21/2015 Aug-17 60-7075-106 3876686 8/21/2015 Aug-17 60-7080-152 3876519 8/21/2015 Aug-17 60-7080-152 3876526 8/21/2015 Aug-17 60-7075-154 3871048 8/21/2015 Aug-17 60-7075-055 3876444 8/21/2015 Aug-17 60-7075-157 3876463 8/21/2015 Aug-17 5921-018-235 3882027 8/24/2015 Aug-17 5921-030-135 3881454 8/24/2015 Aug-17 5921-030-135 3881496 8/24/2015 Aug-17 5921-030-235 3878980 8/24/2015 Aug-17 5921-030-235 3884706 8/24/2015 Aug-17 5921-034-135 3879107 8/24/2015 Aug-17 5921-034-235 3879358 8/24/2015 Aug-17 5921-034-235 3879764 8/24/2015 Aug-17 5921-034-235 3879861 8/24/2015 Aug-17 5921-034-235 3881800 8/24/2015 Aug-17 5921-218-235 3883740 8/24/2015 Aug-17 5921-218-135 3881236 8/24/2015 Aug-17 5921-218-135 3885382 8/24/2015 Aug-17 60-7070-103 3878663 8/24/2015 Aug-17 60-7070-103 3880893 8/24/2015 Aug-17 60-7070-103 3884845 8/24/2015 Aug-17 60-7070-105 3878869 8/24/2015 Aug-17 60-7070-105 3878979 8/24/2015 Aug-17 60-7070-105 3881648 8/24/2015 Aug-17 60-7070-105 3881695 8/24/2015 Aug-17 60-7070-105 3881790 8/24/2015 Aug-17 60-7070-105 3884976 8/24/2015 Aug-17 60-7070-104 3881150 8/24/2015 Aug-17 60-7070-106 3878541 8/24/2015 Aug-17 60-7070-106 3878628 8/24/2015 Aug-17 60-7070-106 3882193 8/24/2015 Aug-17 60-7070-106 3882690 8/24/2015 Aug-17 60-7070-106 3882846 8/24/2015 Aug-17 60-7070-107 3884747 8/24/2015 Aug-17 60-7070-157 3883842 8/24/2015 Aug-17 60-7070-155 3884775 8/24/2015 Aug-17 60-7070-154 3884464 8/24/2015 Aug-17 60-7070-154 3884618 8/24/2015 Aug-17 60-7070-156 3883343 8/24/2015 Aug-17 60-7070-153 3884233 8/24/2015 Aug-17 60-7075-103 3878283 8/24/2015 Aug-17 60-7075-103 3881099 8/24/2015 Aug-17 60-7075-106 3885221 8/24/2015 Aug-17 60-7075-153 3883614 8/24/2015 Aug-17 60-7075-156 3885056 8/24/2015 Aug-17 5921-018-135 3874748 8/26/2015 Aug-17 5921-018-135 3886573 8/26/2015 Aug-17 5921-018-135 3892373 8/26/2015 Aug-17 5921-018-235 3890769 8/26/2015 Aug-17 5921-018-235 3892935 8/26/2015 Aug-17 5921-030-135 3891691 8/26/2015 Aug-17 5921-030-235 3874473 8/26/2015 Aug-17 5921-030-235 3887668 8/26/2015 Aug-17 5921-034-135 3874846 8/26/2015 Aug-17 5921-034-135 3888156 8/26/2015 Aug-17 5921-034-135 3888160 8/26/2015 Aug-17 5921-034-135 3888932 8/26/2015 Aug-17 5921-034-235 3885158 8/26/2015 Aug-17 5921-034-235 3885849 8/26/2015 Aug-17 5921-034-235 3886208 8/26/2015 Aug-17 5921-034-235 3891085 8/26/2015 Aug-17 5921-034-235 3892237 8/26/2015 Aug-17 5921-034-235 3892974 8/26/2015 Aug-17 5921-218-235 3886334 8/26/2015 Aug-17 5921-218-235 3889238 8/26/2015 Aug-17 5921-218-235 3889374 8/26/2015 Aug-17 5921-218-235 3890473 8/26/2015 Aug-17 5921-218-235 3890571 8/26/2015 Aug-17 5921-218-135 3873804 8/26/2015 Aug-17 5921-218-135 3887741 8/26/2015 Aug-17 5921-218-135 3890321 8/26/2015 Aug-17 60-7070-103 3873634 8/26/2015 Aug-17 60-7070-103 3874384 8/26/2015 Aug-17 60-7070-103 3884878 8/26/2015 Aug-17 60-7070-103 3887495 8/26/2015 Aug-17 60-7070-103 3887675 8/26/2015 Aug-17 60-7070-103 3889943 8/26/2015 Aug-17 60-7070-103 3891538 8/26/2015 Aug-17 60-7070-105 3888194 8/26/2015 Aug-17 60-7070-105 3891997 8/26/2015 Aug-17 60-7070-104 3888000 8/26/2015 Aug-17 60-7070-104 3889894 8/26/2015 Aug-17 60-7070-003 3886778 8/26/2015 Aug-17 60-7070-003 3891932 8/26/2015 Aug-17 60-7070-106 3887756 8/26/2015 Aug-17 60-7070-106 3887964 8/26/2015 Aug-17 60-7070-106 3890879 8/26/2015 Aug-17 60-7070-106 3890912 8/26/2015 Aug-17 60-7070-107 3889020 8/26/2015 Aug-17 60-7070-107 3892919 8/26/2015 Aug-17 60-7075-005 3887219 8/26/2015 Aug-17 60-7075-006 3887300 8/26/2015 Aug-17 60-7075-006 3892736 8/26/2015 Aug-17 60-7070-156 3887760 8/26/2015 Aug-17 60-7075-103 3886933 8/26/2015 Aug-17 60-7075-107 3892770 8/26/2015 Aug-17 60-7075-107 3892855 8/26/2015 Aug-17 5921-018-135 3892394 8/28/2015 Aug-17 5921-024-135 3893073 8/28/2015 Aug-17 5921-030-135 3893087 8/28/2015 Aug-17 5921-030-135 3899205 8/28/2015 Aug-17 5921-030-235 3893094 8/28/2015 Aug-17 5921-030-235 3893104 8/28/2015 Aug-17 5921-030-235 3894825 8/28/2015 Aug-17 5921-034-135 3895679 8/28/2015 Aug-17 5921-034-135 3895874 8/28/2015 Aug-17 5921-034-235 3892988 8/28/2015 Aug-17 5921-034-235 3894970 8/28/2015 Aug-17 5921-034-235 3898052 8/28/2015 Aug-17 5921-218-235 3892534 8/28/2015 Aug-17 5921-218-235 3898863 8/28/2015 Aug-17 5921-218-235 3898941 8/28/2015 Aug-17 5921-218-235 3899144 8/28/2015 Aug-17 5921-218-135 3892665 8/28/2015 Aug-17 60-7070-103 3893025 8/28/2015 Aug-17 60-7070-103 3895452 8/28/2015 Aug-17 60-7070-103 3895520 8/28/2015 Aug-17 60-7070-103 3896316 8/28/2015 Aug-17 60-7070-103 3898551 8/28/2015 Aug-17 60-7070-103 3898719 8/28/2015 Aug-17 60-7070-103 3900012 8/28/2015 Aug-17 60-7070-103 3900054 8/28/2015 Aug-17 60-7070-105 3896183 8/28/2015 Aug-17 60-7070-105 3899396 8/28/2015 Aug-17 60-7070-104 3892510 8/28/2015 Aug-17 60-7070-104 3896606 8/28/2015 Aug-17 60-7070-106 3895928 8/28/2015 Aug-17 60-7070-106 3896014 8/28/2015 Aug-17 60-7070-106 3897671 8/28/2015 Aug-17 60-7070-106 3897980 8/28/2015 Aug-17 60-7070-106 3900335 8/28/2015 Aug-17 60-7070-106 3900433 8/28/2015 Aug-17 60-7075-003 3895294 8/28/2015 Aug-17 60-7075-103 3893127 8/28/2015 Aug-17 60-7075-103 3893170 8/28/2015 Aug-17 60-7075-105 3895182 8/28/2015 Aug-17 60-7075-106 3893189 8/28/2015 Aug-17 60-7075-106 3893981 8/28/2015 Aug-17 60-7075-106 3894181 8/28/2015 Aug-17 5921-018-135 3901167 8/31/2015 Aug-17 5921-018-135 3906369 8/31/2015 Aug-17 5921-018-235 3903972 8/31/2015 Aug-17 5921-024-136 3901325 8/31/2015 Aug-17 5921-024-235 3904456 8/31/2015 Aug-17 5921-024-235 3906686 8/31/2015 Aug-17 5921-024-235 3906906 8/31/2015 Aug-17 5921-030-135 3901919 8/31/2015 Aug-17 5921-030-136 3902125 8/31/2015 Aug-17 5921-030-235 3901421 8/31/2015 Aug-17 5921-030-235 3903424 8/31/2015 Aug-17 5921-030-235 3906127 8/31/2015 Aug-17 5921-034-135 3902253 8/31/2015 Aug-17 5921-034-135 3902398 8/31/2015 Aug-17 5921-034-135 3905137 8/31/2015 Aug-17 5921-034-136 3901990 8/31/2015 Aug-17 5921-034-235 3902564 8/31/2015 Aug-17 5921-034-235 3902593 8/31/2015 Aug-17 5921-034-235 3902754 8/31/2015 Aug-17 5921-034-235 3903675 8/31/2015 Aug-17 5921-034-235 3903938 8/31/2015 Aug-17 5921-034-235 3906266 8/31/2015 Aug-17 5921-218-235 3900839 8/31/2015 Aug-17 5921-218-235 3903411 8/31/2015 Aug-17 5921-218-235 3905878 8/31/2015 Aug-17 5921-218-235 3906542 8/31/2015 Aug-17 5921-218-135 3904851 8/31/2015 Aug-17 5921-218-135 3906485 8/31/2015 Aug-17 60-7070-103 3901575 8/31/2015 Aug-17 60-7070-103 3901612 8/31/2015 Aug-17 60-7070-103 3903584 8/31/2015 Aug-17 60-7070-103 3903681 8/31/2015 Aug-17 60-7070-103 3905591 8/31/2015 Aug-17 60-7070-103 3905799 8/31/2015 Aug-17 60-7070-105 3902777 8/31/2015 Aug-17 60-7070-105 3903804 8/31/2015 Aug-17 60-7070-105 3903891 8/31/2015 Aug-17 60-7070-105 3906039 8/31/2015 Aug-17 60-7070-105 3906094 8/31/2015 Aug-17 60-7070-104 3900989 8/31/2015 Aug-17 60-7070-104 3904299 8/31/2015 Aug-17 60-7070-004 3901146 8/31/2015 Aug-17 60-7070-056 3906503 8/31/2015 Aug-17 60-7070-106 3902881 8/31/2015 Aug-17 60-7070-107 3904101 8/31/2015 Aug-17 60-7075-003 3901801 8/31/2015 Aug-17 60-7075-006 3904332 8/31/2015 Aug-17 60-7070-152 3903385 8/31/2015 Aug-17 60-7080-102 3902044 8/31/2015 Aug-17 60-7075-102 3901855 8/31/2015 Aug-17 60-7075-103 3904563 8/31/2015 Aug-17 60-7075-104 3905493 8/31/2015 Aug-17 60-7075-105 3903316 8/31/2015 Aug-17 60-7075-106 3903183 8/31/2015 Aug-17 60-7075-106 3906416 8/31/2015 Aug-17 60-7075-054 3903363 8/31/2015 Aug-17 60-7075-154 3902810 8/31/2015 Aug-17 60-7075-153 3902868 8/31/2015 Aug-17 60-7075-153 3906590 8/31/2015 Aug-17 5921-018-235 3911085 9/2/2015 Sep-17 5921-024-235 3906947 9/2/2015 Sep-17 5921-024-235 3909755 9/2/2015 Sep-17 5921-024-235 3914433 9/2/2015 Sep-17 5921-030-135 3913503 9/2/2015 Sep-17 5921-030-235 3907052 9/2/2015 Sep-17 5921-030-235 3907386 9/2/2015 Sep-17 5921-030-235 3908895 9/2/2015 Sep-17 5921-030-235 3910711 9/2/2015 Sep-17 5921-030-235 3910748 9/2/2015 Sep-17 5921-034-135 3907788 9/2/2015 Sep-17 5921-034-135 3907871 9/2/2015 Sep-17 5921-034-135 3910439 9/2/2015 Sep-17 5921-034-135 3910528 9/2/2015 Sep-17 5921-034-136 3913883 9/2/2015 Sep-17 5921-034-235 3909107 9/2/2015 Sep-17 5921-034-235 3910172 9/2/2015 Sep-17 5921-218-235 3909633 9/2/2015 Sep-17 5921-218-235 3911913 9/2/2015 Sep-17 5921-218-235 3914697 9/2/2015 Sep-17 5921-218-135 3907514 9/2/2015 Sep-17 5921-218-135 3913170 9/2/2015 Sep-17 60-7070-102 3914168 9/2/2015 Sep-17 60-7070-103 3905570 9/2/2015 Sep-17 60-7070-103 3906779 9/2/2015 Sep-17 60-7070-103 3908541 9/2/2015 Sep-17 60-7070-103 3911950 9/2/2015 Sep-17 60-7070-103 3912339 9/2/2015 Sep-17 60-7070-103 3914467 9/2/2015 Sep-17 60-7070-105 3907023 9/2/2015 Sep-17 60-7070-105 3909358 9/2/2015 Sep-17 60-7070-105 3911461 9/2/2015 Sep-17 60-7070-105 3911564 9/2/2015 Sep-17 60-7070-104 3909565 9/2/2015 Sep-17 60-7070-104 3913947 9/2/2015 Sep-17 60-7070-003 3909583 9/2/2015 Sep-17 60-7070-006 3913471 9/2/2015 Sep-17 60-7070-106 3911662 9/2/2015 Sep-17 60-7070-106 3911810 9/2/2015 Sep-17 60-7070-106 3913693 9/2/2015 Sep-17 60-7070-107 3908462 9/2/2015 Sep-17 60-7075-004 3913444 9/2/2015 Sep-17 60-7075-006 3911627 9/2/2015 Sep-17 60-7075-103 3906693 9/2/2015 Sep-17 60-7075-103 3909304 9/2/2015 Sep-17 60-7075-103 3910822 9/2/2015 Sep-17 60-7075-104 3910366 9/2/2015 Sep-17 60-7075-105 3906952 9/2/2015 Sep-17 60-7075-105 3910707 9/2/2015 Sep-17 60-7075-106 3909481 9/2/2015 Sep-17 60-7075-106 3909499 9/2/2015 Sep-17 60-7075-106 3910879 9/2/2015 Sep-17 60-7075-157 3906170 9/2/2015 Sep-17 5921-218-235 3866834 9/3/2015 Sep-17 60-7070-107 3868822 9/3/2015 Sep-17 5921-024-135 3920218 9/4/2015 Sep-17 5921-030-135 3918227 9/4/2015 Sep-17 5921-030-135 3918412 9/4/2015 Sep-17 5921-030-235 3917784 9/4/2015 Sep-17 5921-030-235 3918047 9/4/2015 Sep-17 5921-034-135 3917334 9/4/2015 Sep-17 5921-034-135 3920343 9/4/2015 Sep-17 5921-034-235 3915698 9/4/2015 Sep-17 5921-034-235 3915717 9/4/2015 Sep-17 5921-034-235 3919747 9/4/2015 Sep-17 5921-034-235 3921229 9/4/2015 Sep-17 5921-218-235 3914825 9/4/2015 Sep-17 5921-218-235 3918973 9/4/2015 Sep-17 5921-218-135 3920900 9/4/2015 Sep-17 60-7070-103 3918551 9/4/2015 Sep-17 60-7070-103 3918607 9/4/2015 Sep-17 60-7070-103 3920195 9/4/2015 Sep-17 60-7070-105 3915176 9/4/2015 Sep-17 60-7070-105 3915287 9/4/2015 Sep-17 60-7070-105 3920577 9/4/2015 Sep-17 60-7070-104 3915032 9/4/2015 Sep-17 60-7070-104 3919982 9/4/2015 Sep-17 60-7070-004 3920124 9/4/2015 Sep-17 60-7070-106 3917824 9/4/2015 Sep-17 60-7070-106 3917998 9/4/2015 Sep-17 60-7070-106 3918420 9/4/2015 Sep-17 60-7070-106 3921405 9/4/2015 Sep-17 60-7070-107 3917212 9/4/2015 Sep-17 60-7075-003 3919401 9/4/2015 Sep-17 60-7075-006 3920447 9/4/2015 Sep-17 60-7075-102 3920078 9/4/2015 Sep-17 60-7075-103 3917587 9/4/2015 Sep-17 60-7075-105 3921152 9/4/2015 Sep-17 60-7075-106 3914860 9/4/2015 Sep-17 60-7075-106 3915021 9/4/2015 Sep-17 60-7075-106 3920772 9/4/2015 Sep-17 60-7075-106 3920857 9/4/2015 Sep-17 5921-018-135 3923892 9/8/2015 Sep-17 5921-018-135 3925748 9/8/2015 Sep-17 5921-018-135 3926298 9/8/2015 Sep-17 5921-024-235 3922190 9/8/2015 Sep-17 5921-024-235 3924921 9/8/2015 Sep-17 5921-024-235 3925313 9/8/2015 Sep-17 5921-030-235 3921524 9/8/2015 Sep-17 5921-030-235 3921856 9/8/2015 Sep-17 5921-030-235 3923567 9/8/2015 Sep-17 5921-030-235 3924978 9/8/2015 Sep-17 5921-034-135 3920336 9/8/2015 Sep-17 5921-034-135 3925138 9/8/2015 Sep-17 5921-034-135 3925299 9/8/2015 Sep-17 5921-034-235 3859998 9/8/2015 Sep-17 5921-034-235 3922980 9/8/2015 Sep-17 5921-034-235 3923215 9/8/2015 Sep-17 5921-034-235 3925657 9/8/2015 Sep-17 5921-218-235 3921728 9/8/2015 Sep-17 5921-218-235 3923757 9/8/2015 Sep-17 5921-218-235 3925004 9/8/2015 Sep-17 5921-218-135 3922310 9/8/2015 Sep-17 5921-218-135 3924000 9/8/2015 Sep-17 5921-218-135 3925571 9/8/2015 Sep-17 60-7070-103 3921201 9/8/2015 Sep-17 60-7070-103 3922029 9/8/2015 Sep-17 60-7070-103 3922174 9/8/2015 Sep-17 60-7070-103 3924945 9/8/2015 Sep-17 60-7070-103 3925439 9/8/2015 Sep-17 60-7070-103 3927228 9/8/2015 Sep-17 60-7070-105 3921927 9/8/2015 Sep-17 60-7070-105 3924868 9/8/2015 Sep-17 60-7070-105 3925345 9/8/2015 Sep-17 60-7070-105 3926969 9/8/2015 Sep-17 60-7070-104 3921357 9/8/2015 Sep-17 60-7070-104 3926365 9/8/2015 Sep-17 60-7070-003 3924194 9/8/2015 Sep-17 60-7070-106 3819062 9/8/2015 Sep-17 60-7070-106 3921511 9/8/2015 Sep-17 60-7070-106 3922541 9/8/2015 Sep-17 60-7070-106 3924662 9/8/2015 Sep-17 60-7070-106 3924852 9/8/2015 Sep-17 60-7070-106 3926061 9/8/2015 Sep-17 60-7070-106 3926233 9/8/2015 Sep-17 60-7075-103 3921031 9/8/2015 Sep-17 60-7075-103 3921123 9/8/2015 Sep-17 60-7075-103 3924691 9/8/2015 Sep-17 60-7075-104 3922573 9/8/2015 Sep-17 60-7075-104 3925140 9/8/2015 Sep-17 60-7075-105 3924276 9/8/2015 Sep-17 60-7075-105 3925210 9/8/2015 Sep-17 60-7075-106 3926982 9/8/2015 Sep-17 5921-018-135 3930028 9/10/2015 Sep-17 5921-018-235 3925776 9/10/2015 Sep-17 5921-018-235 3929843 9/10/2015 Sep-17 5921-018-235 3931086 9/10/2015 Sep-17 5921-024-135 3928952 9/10/2015 Sep-17 5921-024-135 3931064 9/10/2015 Sep-17 5921-024-135 3933639 9/10/2015 Sep-17 5921-024-136 3924474 9/10/2015 Sep-17 5921-024-235 3928158 9/10/2015 Sep-17 5921-024-235 3930059 9/10/2015 Sep-17 5921-030-135 3930765 9/10/2015 Sep-17 5921-030-135 3931093 9/10/2015 Sep-17 5921-030-135 3931732 9/10/2015 Sep-17 5921-030-235 3927373 9/10/2015 Sep-17 5921-030-235 3928733 9/10/2015 Sep-17 5921-030-235 3929242 9/10/2015 Sep-17 5921-030-235 3933589 9/10/2015 Sep-17 5921-030-235 3934078 9/10/2015 Sep-17 5921-034-135 3928250 9/10/2015 Sep-17 5921-034-135 3932548 9/10/2015 Sep-17 5921-034-135 3934218 9/10/2015 Sep-17 5921-034-235 3925822 9/10/2015 Sep-17 5921-034-235 3927736 9/10/2015 Sep-17 5921-034-235 3929513 9/10/2015 Sep-17 5921-034-235 3929929 9/10/2015 Sep-17 5921-034-235 3930966 9/10/2015 Sep-17 5921-034-235 3930997 9/10/2015 Sep-17 5921-034-235 3933100 9/10/2015 Sep-17 5921-034-236 3932650 9/10/2015 Sep-17 5921-218-235 3928278 9/10/2015 Sep-17 5921-218-235 3930125 9/10/2015 Sep-17 5921-218-235 3931040 9/10/2015 Sep-17 5921-218-135 3928849 9/10/2015 Sep-17 5921-218-135 3933320 9/10/2015 Sep-17 5921-218-135 3934315 9/10/2015 Sep-17 5921-218-136 3928655 9/10/2015 Sep-17 5921-218-136 3929219 9/10/2015 Sep-17 5921-218-236 3930945 9/10/2015 Sep-17 60-7070-102 3930662 9/10/2015 Sep-17 60-7070-102 3932763 9/10/2015 Sep-17 60-7070-103 3923973 9/10/2015 Sep-17 60-7070-103 3928056 9/10/2015 Sep-17 60-7070-103 3929606 9/10/2015 Sep-17 60-7070-103 3929652 9/10/2015 Sep-17 60-7070-103 3934406 9/10/2015 Sep-17 60-7070-105 3927935 9/10/2015 Sep-17 60-7070-105 3928638 9/10/2015 Sep-17 60-7070-105 3928722 9/10/2015 Sep-17 60-7070-105 3930276 9/10/2015 Sep-17 60-7070-105 3931103 9/10/2015 Sep-17 60-7070-104 3928542 9/10/2015 Sep-17 60-7070-104 3930207 9/10/2015 Sep-17 60-7070-104 3933799 9/10/2015 Sep-17 60-7070-003 3930708 9/10/2015 Sep-17 60-7070-004 3930749 9/10/2015 Sep-17 60-7070-005 3930670 9/10/2015 Sep-17 60-7070-006 3929261 9/10/2015 Sep-17 60-7070-006 3931038 9/10/2015 Sep-17 60-7070-106 3930333 9/10/2015 Sep-17 60-7070-106 3930403 9/10/2015 Sep-17 60-7070-106 3930875 9/10/2015 Sep-17 60-7070-106 3933930 9/10/2015 Sep-17 60-7070-107 3927581 9/10/2015 Sep-17 60-7070-107 3930529 9/10/2015 Sep-17 60-7070-107 3930615 9/10/2015 Sep-17 60-7070-107 3931644 9/10/2015 Sep-17 60-7075-003 3928557 9/10/2015 Sep-17 60-7075-004 3930627 9/10/2015 Sep-17 60-7075-005 3928854 9/10/2015 Sep-17 60-7075-006 3930806 9/10/2015 Sep-17 60-7080-102 3929177 9/10/2015 Sep-17 60-7075-103 3928444 9/10/2015 Sep-17 60-7075-103 3932943 9/10/2015 Sep-17 60-7075-104 3930187 9/10/2015 Sep-17 60-7075-104 3930248 9/10/2015 Sep-17 60-7075-105 3929815 9/10/2015 Sep-17 60-7075-106 3928780 9/10/2015 Sep-17 60-7075-107 3930085 9/10/2015 Sep-17 5921-018-135 3937143 9/14/2015 Sep-17 5921-018-135 3937158 9/14/2015 Sep-17 5921-018-135 3940428 9/14/2015 Sep-17 5921-024-235 3935363 9/14/2015 Sep-17 5921-024-235 3938318 9/14/2015 Sep-17 5921-030-135 3935725 9/14/2015 Sep-17 5921-030-135 3939015 9/14/2015 Sep-17 5921-034-135 3938082 9/14/2015 Sep-17 5921-034-135 3939402 9/14/2015 Sep-17 5921-034-235 3935972 9/14/2015 Sep-17 5921-034-235 3937968 9/14/2015 Sep-17 5921-034-235 3938167 9/14/2015 Sep-17 5921-034-235 3939544 9/14/2015 Sep-17 5921-034-235 3939674 9/14/2015 Sep-17 5921-218-235 3933171 9/14/2015 Sep-17 5921-218-235 3935597 9/14/2015 Sep-17 5921-218-235 3937492 9/14/2015 Sep-17 5921-218-235 3940278 9/14/2015 Sep-17 5921-218-135 3933455 9/14/2015 Sep-17 5921-218-135 3938683 9/14/2015 Sep-17 60-7070-103 3932887 9/14/2015 Sep-17 60-7070-103 3934508 9/14/2015 Sep-17 60-7070-103 3935476 9/14/2015 Sep-17 60-7070-103 3937348 9/14/2015 Sep-17 60-7070-103 3938864 9/14/2015 Sep-17 60-7070-103 3939070 9/14/2015 Sep-17 60-7070-105 3933752 9/14/2015 Sep-17 60-7070-105 3934874 9/14/2015 Sep-17 60-7070-105 3935079 9/14/2015 Sep-17 60-7070-105 3935756 9/14/2015 Sep-17 60-7070-105 3935901 9/14/2015 Sep-17 60-7070-105 3937655 9/14/2015 Sep-17 60-7070-105 3937746 9/14/2015 Sep-17 60-7070-105 3940820 9/14/2015 Sep-17 60-7070-104 3935183 9/14/2015 Sep-17 60-7070-104 3939138 9/14/2015 Sep-17 60-7070-003 3935578 9/14/2015 Sep-17 60-7070-006 3939101 9/14/2015 Sep-17 60-7070-106 3935235 9/14/2015 Sep-17 60-7070-106 3936960 9/14/2015 Sep-17 60-7070-106 3939776 9/14/2015 Sep-17 60-7070-106 3940232 9/14/2015 Sep-17 60-7070-107 3934667 9/14/2015 Sep-17 60-7075-103 3934819 9/14/2015 Sep-17 60-7075-103 3934900 9/14/2015 Sep-17 60-7075-103 3935120 9/14/2015 Sep-17 60-7075-103 3940535 9/14/2015 Sep-17 60-7075-105 3934554 9/14/2015 Sep-17 60-7075-106 3934583 9/14/2015 Sep-17 60-7075-106 3937849 9/14/2015 Sep-17 60-7075-106 3937906 9/14/2015 Sep-17 60-7075-106 3906471 9/15/2015 Sep-17 5921-018-135 3946989 9/16/2015 Sep-17 5921-018-135 3952241 9/16/2015 Sep-17 5921-018-235 3946342 9/16/2015 Sep-17 5921-018-235 3953898 9/16/2015 Sep-17 5921-024-135 3953592 9/16/2015 Sep-17 5921-030-135 3948717 9/16/2015 Sep-17 5921-030-235 3948195 9/16/2015 Sep-17 5921-030-235 3953268 9/16/2015 Sep-17 5921-030-235 3953366 9/16/2015 Sep-17 5921-034-135 3948260 9/16/2015 Sep-17 5921-034-135 3948447 9/16/2015 Sep-17 5921-034-135 3952520 9/16/2015 Sep-17 5921-034-135 3953425 9/16/2015 Sep-17 5921-034-235 3946737 9/16/2015 Sep-17 5921-034-235 3949502 9/16/2015 Sep-17 5921-218-235 3946621 9/16/2015 Sep-17 5921-218-235 3949732 9/16/2015 Sep-17 5921-218-235 3953711 9/16/2015 Sep-17 5921-218-235 3953867 9/16/2015 Sep-17 5921-218-135 3952391 9/16/2015 Sep-17 5921-218-236 3949637 9/16/2015 Sep-17 60-7070-103 3946361 9/16/2015 Sep-17 60-7070-103 3946856 9/16/2015 Sep-17 60-7070-103 3949254 9/16/2015 Sep-17 60-7070-103 3949853 9/16/2015 Sep-17 60-7070-103 3953206 9/16/2015 Sep-17 60-7070-103 3954033 9/16/2015 Sep-17 60-7070-105 3946763 9/16/2015 Sep-17 60-7070-105 3948050 9/16/2015 Sep-17 60-7070-105 3949659 9/16/2015 Sep-17 60-7070-105 3952045 9/16/2015 Sep-17 60-7070-105 3952195 9/16/2015 Sep-17 60-7070-105 3954288 9/16/2015 Sep-17 60-7070-104 3946479 9/16/2015 Sep-17 60-7070-104 3949830 9/16/2015 Sep-17 60-7070-104 3949848 9/16/2015 Sep-17 60-7070-003 3953965 9/16/2015 Sep-17 60-7070-106 3946530 9/16/2015 Sep-17 60-7070-106 3949283 9/16/2015 Sep-17 60-7070-106 3949778 9/16/2015 Sep-17 60-7070-107 3947879 9/16/2015 Sep-17 60-7070-107 3953473 9/16/2015 Sep-17 60-7075-006 3954154 9/16/2015 Sep-17 60-7080-102 3946293 9/16/2015 Sep-17 60-7075-103 3946446 9/16/2015 Sep-17 60-7075-103 3946802 9/16/2015 Sep-17 60-7075-103 3949905 9/16/2015 Sep-17 60-7075-103 3949967 9/16/2015 Sep-17 60-7075-103 3953516 9/16/2015 Sep-17 60-7075-104 3949685 9/16/2015 Sep-17 60-7075-105 3949381 9/16/2015 Sep-17 60-7075-106 3949556 9/16/2015 Sep-17 5921-018-135 3961130 9/21/2015 Sep-17 5921-018-235 3957137 9/21/2015 Sep-17 5921-024-135 3960183 9/21/2015 Sep-17 5921-024-136 3957752 9/21/2015 Sep-17 5921-030-235 3956167 9/21/2015 Sep-17 5921-030-235 3956228 9/21/2015 Sep-17 5921-030-235 3957907 9/21/2015 Sep-17 5921-030-235 3957982 9/21/2015 Sep-17 5921-030-235 3958123 9/21/2015 Sep-17 5921-030-235 3961315 9/21/2015 Sep-17 5921-030-235 3962526 9/21/2015 Sep-17 5921-034-135 3952679 9/21/2015 Sep-17 5921-034-235 3952717 9/21/2015 Sep-17 5921-034-235 3955150 9/21/2015 Sep-17 5921-034-235 3957159 9/21/2015 Sep-17 5921-034-235 3958970 9/21/2015 Sep-17 5921-034-235 3960843 9/21/2015 Sep-17 5921-034-236 3957799 9/21/2015 Sep-17 5921-218-235 3957192 9/21/2015 Sep-17 5921-218-235 3957267 9/21/2015 Sep-17 5921-218-235 3957438 9/21/2015 Sep-17 5921-218-235 3959367 9/21/2015 Sep-17 5921-218-235 3960929 9/21/2015 Sep-17 5921-218-235 3960993 9/21/2015 Sep-17 5921-218-235 3962369 9/21/2015 Sep-17 5921-030-236 3957821 9/21/2015 Sep-17 5921-030-236 3960532 9/21/2015 Sep-17 5921-218-135 3956942 9/21/2015 Sep-17 5921-218-135 3957656 9/21/2015 Sep-17 5921-218-136 3957396 9/21/2015 Sep-17 5921-218-236 3957924 9/21/2015 Sep-17 5921-218-236 3958264 9/21/2015 Sep-17 60-7070-103 3952916 9/21/2015 Sep-17 60-7070-103 3956649 9/21/2015 Sep-17 60-7070-103 3956886 9/21/2015 Sep-17 60-7070-103 3958524 9/21/2015 Sep-17 60-7070-103 3958530 9/21/2015 Sep-17 60-7070-103 3960668 9/21/2015 Sep-17 60-7070-103 3961362 9/21/2015 Sep-17 60-7070-103 3962087 9/21/2015 Sep-17 60-7070-105 3954363 9/21/2015 Sep-17 60-7070-105 3956599 9/21/2015 Sep-17 60-7070-105 3956872 9/21/2015 Sep-17 60-7070-105 3957251 9/21/2015 Sep-17 60-7070-105 3958825 9/21/2015 Sep-17 60-7070-105 3958839 9/21/2015 Sep-17 60-7070-105 3961176 9/21/2015 Sep-17 60-7070-105 3961195 9/21/2015 Sep-17 60-7070-104 3957298 9/21/2015 Sep-17 60-7070-104 3960634 9/21/2015 Sep-17 60-7070-004 3955863 9/21/2015 Sep-17 60-7070-005 3956534 9/21/2015 Sep-17 60-7070-006 3959092 9/21/2015 Sep-17 60-7070-006 3959139 9/21/2015 Sep-17 60-7070-106 3954392 9/21/2015 Sep-17 60-7070-106 3954476 9/21/2015 Sep-17 60-7070-106 3954634 9/21/2015 Sep-17 60-7070-106 3954760 9/21/2015 Sep-17 60-7070-106 3956926 9/21/2015 Sep-17 60-7070-106 3957116 9/21/2015 Sep-17 60-7070-106 3959806 9/21/2015 Sep-17 60-7070-106 3960051 9/21/2015 Sep-17 60-7070-106 3961218 9/21/2015 Sep-17 60-7075-006 3960545 9/21/2015 Sep-17 60-7075-103 3960647 9/21/2015 Sep-17 60-7075-104 3956259 9/21/2015 Sep-17 60-7075-106 3955682 9/21/2015 Sep-17 60-7075-106 3955735 9/21/2015 Sep-17 60-7075-106 3958728 9/21/2015 Sep-17 60-7075-106 3958737 9/21/2015 Sep-17 5921-018-235 3941878 9/22/2015 Sep-17 5921-024-135 3944340 9/22/2015 Sep-17 5921-024-235 3941019 9/22/2015 Sep-17 5921-024-235 3945988 9/22/2015 Sep-17 5921-030-135 3941781 9/22/2015 Sep-17 5921-030-135 3943978 9/22/2015 Sep-17 5921-030-136 3942112 9/22/2015 Sep-17 5921-030-136 3942131 9/22/2015 Sep-17 5921-030-235 3942261 9/22/2015 Sep-17 5921-030-235 3945082 9/22/2015 Sep-17 5921-030-235 3945450 9/22/2015 Sep-17 5921-034-135 3941438 9/22/2015 Sep-17 5921-034-135 3943059 9/22/2015 Sep-17 5921-034-235 3941365 9/22/2015 Sep-17 5921-034-235 3942257 9/22/2015 Sep-17 5921-034-235 3946020 9/22/2015 Sep-17 5921-034-236 3945864 9/22/2015 Sep-17 5921-034-236 3945919 9/22/2015 Sep-17 5921-218-235 3941884 9/22/2015 Sep-17 5921-218-235 3942610 9/22/2015 Sep-17 5921-218-235 3945683 9/22/2015 Sep-17 5921-218-135 3941042 9/22/2015 Sep-17 5921-218-135 3944408 9/22/2015 Sep-17 5921-218-135 3944551 9/22/2015 Sep-17 5921-218-135 3945696 9/22/2015 Sep-17 5921-218-136 3942120 9/22/2015 Sep-17 5921-218-136 3942154 9/22/2015 Sep-17 60-7070-102 3943314 9/22/2015 Sep-17 60-7070-103 3941159 9/22/2015 Sep-17 60-7070-103 3941251 9/22/2015 Sep-17 60-7070-103 3941976 9/22/2015 Sep-17 60-7070-103 3942328 9/22/2015 Sep-17 60-7070-103 3944800 9/22/2015 Sep-17 60-7070-103 3945575 9/22/2015 Sep-17 60-7070-105 3941558 9/22/2015 Sep-17 60-7070-105 3942469 9/22/2015 Sep-17 60-7070-105 3944386 9/22/2015 Sep-17 60-7070-105 3945048 9/22/2015 Sep-17 60-7070-105 3945957 9/22/2015 Sep-17 60-7070-104 3941799 9/22/2015 Sep-17 60-7070-104 3942718 9/22/2015 Sep-17 60-7070-104 3944925 9/22/2015 Sep-17 60-7070-003 3943627 9/22/2015 Sep-17 60-7070-005 3942022 9/22/2015 Sep-17 60-7070-006 3945180 9/22/2015 Sep-17 60-7070-106 3941504 9/22/2015 Sep-17 60-7070-106 3944598 9/22/2015 Sep-17 60-7070-106 3944607 9/22/2015 Sep-17 60-7070-106 3945432 9/22/2015 Sep-17 60-7070-107 3910769 9/22/2015 Sep-17 60-7070-107 3941486 9/22/2015 Sep-17 60-7070-107 3943540 9/22/2015 Sep-17 60-7075-004 3944258 9/22/2015 Sep-17 60-7075-104 3941930 9/22/2015 Sep-17 60-7075-105 3943731 9/22/2015 Sep-17 60-7075-106 3941298 9/22/2015 Sep-17 60-7075-106 3942891 9/22/2015 Sep-17 60-7075-106 3943028 9/22/2015 Sep-17 60-7075-106 3944788 9/22/2015 Sep-17 60-7075-106 3946077 9/22/2015 Sep-17 5921-030-235 3958110 9/23/2015 Sep-17 5921-030-235 3965875 9/23/2015 Sep-17 5921-030-235 3966592 9/23/2015 Sep-17 5921-030-235 3970095 9/23/2015 Sep-17 5921-034-135 3963942 9/23/2015 Sep-17 5921-034-135 3969184 9/23/2015 Sep-17 5921-034-235 3962733 9/23/2015 Sep-17 5921-034-235 3964135 9/23/2015 Sep-17 5921-034-235 3965555 9/23/2015 Sep-17 5921-034-235 3969719 9/23/2015 Sep-17 5921-218-235 3963872 9/23/2015 Sep-17 5921-218-235 3965379 9/23/2015 Sep-17 5921-218-235 3969774 9/23/2015 Sep-17 5921-218-135 3966711 9/23/2015 Sep-17 5921-218-135 3968003 9/23/2015 Sep-17 60-7070-103 3963302 9/23/2015 Sep-17 60-7070-103 3963758 9/23/2015 Sep-17 60-7070-103 3963829 9/23/2015 Sep-17 60-7070-103 3965139 9/23/2015 Sep-17 60-7070-105 3963007 9/23/2015 Sep-17 60-7070-105 3963599 9/23/2015 Sep-17 60-7070-105 3965691 9/23/2015 Sep-17 60-7070-105 3969885 9/23/2015 Sep-17 60-7070-104 3964998 9/23/2015 Sep-17 60-7070-005 3962693 9/23/2015 Sep-17 60-7070-006 3969047 9/23/2015 Sep-17 60-7070-106 3963146 9/23/2015 Sep-17 60-7070-106 3965282 9/23/2015 Sep-17 60-7070-106 3965314 9/23/2015 Sep-17 60-7070-106 3966268 9/23/2015 Sep-17 60-7070-106 3966487 9/23/2015 Sep-17 60-7070-106 3970228 9/23/2015 Sep-17 60-7075-103 3963413 9/23/2015 Sep-17 60-7075-103 3965783 9/23/2015 Sep-17 60-7075-104 3964249 9/23/2015 Sep-17 60-7075-106 3963053 9/23/2015 Sep-17 60-7075-106 3969340 9/23/2015 Sep-17 5921-034-135 3939299 9/24/2015 Sep-17 5921-018-135 3970581 9/25/2015 Sep-17 5921-018-235 3970631 9/25/2015 Sep-17 5921-024-235 3970547 9/25/2015 Sep-17 5921-024-235 3975471 9/25/2015 Sep-17 5921-024-235 3975877 9/25/2015 Sep-17 5921-024-236 3972456 9/25/2015 Sep-17 5921-030-135 3973273 9/25/2015 Sep-17 5921-030-235 3971703 9/25/2015 Sep-17 5921-030-235 3972688 9/25/2015 Sep-17 5921-030-235 3972814 9/25/2015 Sep-17 5921-030-235 3974558 9/25/2015 Sep-17 5921-034-135 3971386 9/25/2015 Sep-17 5921-034-135 3974440 9/25/2015 Sep-17 5921-034-135 3976933 9/25/2015 Sep-17 5921-034-235 3971301 9/25/2015 Sep-17 5921-034-235 3972833 9/25/2015 Sep-17 5921-034-235 3974222 9/25/2015 Sep-17 5921-034-235 3976352 9/25/2015 Sep-17 5921-218-235 3971859 9/25/2015 Sep-17 5921-218-235 3973148 9/25/2015 Sep-17 5921-218-235 3974639 9/25/2015 Sep-17 5921-218-235 3976126 9/25/2015 Sep-17 5921-218-135 3975330 9/25/2015 Sep-17 60-7070-103 3970677 9/25/2015 Sep-17 60-7070-103 3970817 9/25/2015 Sep-17 60-7070-103 3972977 9/25/2015 Sep-17 60-7070-103 3974710 9/25/2015 Sep-17 60-7070-103 3974924 9/25/2015 Sep-17 60-7070-103 3976600 9/25/2015 Sep-17 60-7070-103 3976709 9/25/2015 Sep-17 60-7070-105 3971580 9/25/2015 Sep-17 60-7070-105 3974088 9/25/2015 Sep-17 60-7070-105 3974171 9/25/2015 Sep-17 60-7070-105 3975039 9/25/2015 Sep-17 60-7070-105 3976556 9/25/2015 Sep-17 60-7070-104 3972247 9/25/2015 Sep-17 60-7070-104 3974969 9/25/2015 Sep-17 60-7070-005 3972592 9/25/2015 Sep-17 60-7070-106 3926734 9/25/2015 Sep-17 60-7070-106 3973803 9/25/2015 Sep-17 60-7070-106 3975127 9/25/2015 Sep-17 60-7075-106 3972516 9/25/2015 Sep-17 60-7075-106 3975266 9/25/2015 Sep-17 5921-018-235 3978627 9/28/2015 Sep-17 5921-018-235 3981917 9/28/2015 Sep-17 5921-030-135 3968866 9/28/2015 Sep-17 5921-030-235 3979728 9/28/2015 Sep-17 5921-030-235 3980567 9/28/2015 Sep-17 5921-030-235 3982752 9/28/2015 Sep-17 5921-030-235 3984705 9/28/2015 Sep-17 5921-034-135 3977067 9/28/2015 Sep-17 5921-034-135 3978731 9/28/2015 Sep-17 5921-034-135 3978899 9/28/2015 Sep-17 5921-034-136 3982799 9/28/2015 Sep-17 5921-034-136 3982809 9/28/2015 Sep-17 5921-034-235 3877960 9/28/2015 Sep-17 5921-034-235 3945285 9/28/2015 Sep-17 5921-034-235 3978059 9/28/2015 Sep-17 5921-034-235 3978139 9/28/2015 Sep-17 5921-034-235 3980454 9/28/2015 Sep-17 5921-034-235 3980579 9/28/2015 Sep-17 5921-034-235 3982057 9/28/2015 Sep-17 5921-034-235 3982515 9/28/2015 Sep-17 5921-034-235 3984400 9/28/2015 Sep-17 5921-218-235 3977634 9/28/2015 Sep-17 5921-218-235 3978118 9/28/2015 Sep-17 5921-218-235 3981344 9/28/2015 Sep-17 5921-218-235 3982527 9/28/2015 Sep-17 5921-218-235 3984369 9/28/2015 Sep-17 5921-218-135 3929392 9/28/2015 Sep-17 5921-218-135 3977495 9/28/2015 Sep-17 5921-218-135 3978461 9/28/2015 Sep-17 5921-218-135 3980960 9/28/2015 Sep-17 5921-218-135 3985029 9/28/2015 Sep-17 60-7070-103 3978314 9/28/2015 Sep-17 60-7070-103 3979802 9/28/2015 Sep-17 60-7070-103 3981300 9/28/2015 Sep-17 60-7070-103 3982455 9/28/2015 Sep-17 60-7070-103 3982884 9/28/2015 Sep-17 60-7070-103 3984798 9/28/2015 Sep-17 60-7070-105 3901263 9/28/2015 Sep-17 60-7070-105 3923669 9/28/2015 Sep-17 60-7070-105 3978322 9/28/2015 Sep-17 60-7070-105 3980145 9/28/2015 Sep-17 60-7070-105 3982429 9/28/2015 Sep-17 60-7070-105 3984535 9/28/2015 Sep-17 60-7070-105 3984655 9/28/2015 Sep-17 60-7070-104 3924248 9/28/2015 Sep-17 60-7070-104 3954804 9/28/2015 Sep-17 60-7070-104 3978540 9/28/2015 Sep-17 60-7070-104 3978712 9/28/2015 Sep-17 60-7070-104 3981219 9/28/2015 Sep-17 60-7070-104 3982768 9/28/2015 Sep-17 60-7070-006 3981787 9/28/2015 Sep-17 60-7070-006 3982558 9/28/2015 Sep-17 60-7070-106 3926562 9/28/2015 Sep-17 60-7070-106 3977120 9/28/2015 Sep-17 60-7070-106 3979451 9/28/2015 Sep-17 60-7070-106 3979632 9/28/2015 Sep-17 60-7070-106 3980742 9/28/2015 Sep-17 60-7070-106 3980918 9/28/2015 Sep-17 60-7070-106 3982383 9/28/2015 Sep-17 60-7070-106 3985093 9/28/2015 Sep-17 60-7075-005 3901216 9/28/2015 Sep-17 60-7075-006 3980412 9/28/2015 Sep-17 60-7075-103 3927082 9/28/2015 Sep-17 60-7075-104 3982821 9/28/2015 Sep-17 60-7075-106 3924419 9/28/2015 Sep-17 60-7075-106 3979361 9/28/2015 Sep-17 60-7075-106 3981597 9/28/2015 Sep-17 5921-018-135 3990734 10/1/2015 Oct-17 5921-018-235 3986442 10/1/2015 Oct-17 5921-024-235 3990214 10/1/2015 Oct-17 5921-030-235 3986085 10/1/2015 Oct-17 5921-030-235 3991528 10/1/2015 Oct-17 5921-034-135 3989581 10/1/2015 Oct-17 5921-034-136 3990438 10/1/2015 Oct-17 5921-034-235 3986000 10/1/2015 Oct-17 5921-034-235 3987389 10/1/2015 Oct-17 5921-034-235 3991642 10/1/2015 Oct-17 5921-218-235 3986379 10/1/2015 Oct-17 5921-218-235 3991275 10/1/2015 Oct-17 5921-030-236 3986615 10/1/2015 Oct-17 5921-044-135 3988129 10/1/2015 Oct-17 5921-218-136 3986604 10/1/2015 Oct-17 5921-218-236 3986555 10/1/2015 Oct-17 60-7070-103 3988275 10/1/2015 Oct-17 60-7070-103 3988954 10/1/2015 Oct-17 60-7070-103 3992262 10/1/2015 Oct-17 60-7070-105 3986102 10/1/2015 Oct-17 60-7070-105 3986118 10/1/2015 Oct-17 60-7070-105 3988247 10/1/2015 Oct-17 60-7070-105 3990527 10/1/2015 Oct-17 60-7070-105 3992496 10/1/2015 Oct-17 60-7070-104 3985745 10/1/2015 Oct-17 60-7070-104 3989478 10/1/2015 Oct-17 60-7070-104 3992121 10/1/2015 Oct-17 60-7070-104 3992243 10/1/2015 Oct-17 60-7070-004 3985848 10/1/2015 Oct-17 60-7070-005 3986796 10/1/2015 Oct-17 60-7070-106 3985212 10/1/2015 Oct-17 60-7070-106 3990813 10/1/2015 Oct-17 60-7070-106 3991867 10/1/2015 Oct-17 60-7070-106 3991898 10/1/2015 Oct-17 60-7070-107 3990279 10/1/2015 Oct-17 60-7075-003 3989122 10/1/2015 Oct-17 60-7075-003 3989421 10/1/2015 Oct-17 60-7075-006 3985541 10/1/2015 Oct-17 60-7075-006 3988062 10/1/2015 Oct-17 60-7075-103 3990042 10/1/2015 Oct-17 60-7075-104 3986230 10/1/2015 Oct-17 60-7075-105 3986305 10/1/2015 Oct-17 60-7075-106 3985671 10/1/2015 Oct-17 60-7075-106 3988058 10/1/2015 Oct-17 5921-030-235 3852652 10/2/2015 Oct-17 5921-030-235 3852787 10/2/2015 Oct-17 5921-034-135 3851660 10/2/2015 Oct-17 5921-034-235 3852337 10/2/2015 Oct-17 60-7070-103 3851850 10/2/2015 Oct-17 60-7070-105 3852080 10/2/2015 Oct-17 60-7075-104 3852051 10/2/2015 Oct-17 5921-018-135 4000282 10/5/2015 Oct-17 5921-024-135 3996577 10/5/2015 Oct-17 5921-024-235 3993905 10/5/2015 Oct-17 5921-024-235 3996717 10/5/2015 Oct-17 5921-024-235 4001055 10/5/2015 Oct-17 5921-030-135 3993876 10/5/2015 Oct-17 5921-030-135 3996202 10/5/2015 Oct-17 5921-030-135 3998532 10/5/2015 Oct-17 5921-030-235 3994076 10/5/2015 Oct-17 5921-030-235 3994122 10/5/2015 Oct-17 5921-030-235 3996389 10/5/2015 Oct-17 5921-030-235 3999162 10/5/2015 Oct-17 5921-030-235 3999213 10/5/2015 Oct-17 5921-034-135 3992742 10/5/2015 Oct-17 5921-034-135 3994736 10/5/2015 Oct-17 5921-034-135 3996343 10/5/2015 Oct-17 5921-034-135 3997501 10/5/2015 Oct-17 5921-034-235 3995390 10/5/2015 Oct-17 5921-034-235 3997823 10/5/2015 Oct-17 5921-218-235 3993745 10/5/2015 Oct-17 5921-218-235 3993824 10/5/2015 Oct-17 5921-218-235 3994200 10/5/2015 Oct-17 5921-218-235 3999535 10/5/2015 Oct-17 5921-218-235 3999640 10/5/2015 Oct-17 5921-218-235 4000501 10/5/2015 Oct-17 5921-218-135 3995841 10/5/2015 Oct-17 5921-218-135 4000071 10/5/2015 Oct-17 5921-218-135 4001220 10/5/2015 Oct-17 60-7070-103 3995293 10/5/2015 Oct-17 60-7070-103 3995771 10/5/2015 Oct-17 60-7070-103 3996826 10/5/2015 Oct-17 60-7070-103 3996971 10/5/2015 Oct-17 60-7070-103 3999808 10/5/2015 Oct-17 60-7070-105 3992630 10/5/2015 Oct-17 60-7070-105 3992669 10/5/2015 Oct-17 60-7070-105 3995759 10/5/2015 Oct-17 60-7070-105 3997604 10/5/2015 Oct-17 60-7070-105 3997691 10/5/2015 Oct-17 60-7070-105 3999934 10/5/2015 Oct-17 60-7070-104 3994919 10/5/2015 Oct-17 60-7070-104 3996078 10/5/2015 Oct-17 60-7070-104 3999307 10/5/2015 Oct-17 60-7070-003 3997092 10/5/2015 Oct-17 60-7070-006 3995272 10/5/2015 Oct-17 60-7070-106 3995229 10/5/2015 Oct-17 60-7070-106 3998718 10/5/2015 Oct-17 60-7070-107 3996512 10/5/2015 Oct-17 60-7070-107 3996565 10/5/2015 Oct-17 60-7075-006 3994510 10/5/2015 Oct-17 60-7075-103 3992994 10/5/2015 Oct-17 60-7075-103 3998371 10/5/2015 Oct-17 60-7075-105 3996050 10/5/2015 Oct-17 60-7075-105 4001283 10/5/2015 Oct-17 60-7075-106 3993979 10/5/2015 Oct-17 60-7075-106 3997194 10/5/2015 Oct-17 60-7075-106 3999376 10/5/2015 Oct-17 5921-018-235 4000586 10/8/2015 Oct-17 5921-030-135 4002405 10/8/2015 Oct-17 5921-030-135 4006477 10/8/2015 Oct-17 5921-030-235 4001967 10/8/2015 Oct-17 5921-030-235 4006047 10/8/2015 Oct-17 5921-034-135 4001489 10/8/2015 Oct-17 5921-034-135 4001692 10/8/2015 Oct-17 5921-034-135 4002152 10/8/2015 Oct-17 5921-034-135 4003850 10/8/2015 Oct-17 5921-034-135 4007278 10/8/2015 Oct-17 5921-034-235 4003116 10/8/2015 Oct-17 5921-034-235 4006761 10/8/2015 Oct-17 5921-218-235 4003618 10/8/2015 Oct-17 5921-218-235 4003660 10/8/2015 Oct-17 5921-218-235 4008105 10/8/2015 Oct-17 5921-218-135 4004370 10/8/2015 Oct-17 5921-218-135 4005889 10/8/2015 Oct-17 60-7070-102 4004508 10/8/2015 Oct-17 60-7070-102 4004622 10/8/2015 Oct-17 60-7070-103 4001135 10/8/2015 Oct-17 60-7070-103 4003331 10/8/2015 Oct-17 60-7070-103 4004633 10/8/2015 Oct-17 60-7070-103 4007492 10/8/2015 Oct-17 60-7070-105 4001103 10/8/2015 Oct-17 60-7070-105 4002970 10/8/2015 Oct-17 60-7070-105 4007330 10/8/2015 Oct-17 60-7070-105 4007437 10/8/2015 Oct-17 60-7070-105 4007910 10/8/2015 Oct-17 60-7070-104 4002915 10/8/2015 Oct-17 60-7070-003 4006420 10/8/2015 Oct-17 60-7070-106 4002467 10/8/2015 Oct-17 60-7070-106 4006075 10/8/2015 Oct-17 60-7070-106 4006355 10/8/2015 Oct-17 60-7075-103 4004241 10/8/2015 Oct-17 60-7075-103 4005177 10/8/2015 Oct-17 60-7075-104 4000997 10/8/2015 Oct-17 60-7075-104 4007189 10/8/2015 Oct-17 60-7075-105 4001462 10/8/2015 Oct-17 60-7075-106 4002286 10/8/2015 Oct-17 60-7075-106 4002390 10/8/2015 Oct-17 60-7075-106 4004066 10/8/2015 Oct-17 60-7075-106 4006939 10/8/2015 Oct-17 5921-018-135 4015851 10/9/2015 Oct-17 5921-018-235 4008409 10/9/2015 Oct-17 5921-024-135 4016100 10/9/2015 Oct-17 5921-024-235 4014754 10/9/2015 Oct-17 5921-030-135 4015299 10/9/2015 Oct-17 5921-030-235 4005718 10/9/2015 Oct-17 5921-030-235 4008696 10/9/2015 Oct-17 5921-034-135 4016279 10/9/2015 Oct-17 5921-034-135 4016312 10/9/2015 Oct-17 5921-034-235 4011678 10/9/2015 Oct-17 5921-034-235 4011904 10/9/2015 Oct-17 5921-034-235 4014834 10/9/2015 Oct-17 5921-218-235 4005469 10/9/2015 Oct-17 5921-218-235 4011065 10/9/2015 Oct-17 5921-218-235 4013109 10/9/2015 Oct-17 5921-218-135 4008568 10/9/2015 Oct-17 5921-218-135 4013449 10/9/2015 Oct-17 60-7070-102 4013607 10/9/2015 Oct-17 60-7070-103 4011238 10/9/2015 Oct-17 60-7070-103 4011609 10/9/2015 Oct-17 60-7070-103 4013669 10/9/2015 Oct-17 60-7070-103 4016472 10/9/2015 Oct-17 60-7070-103 4016656 10/9/2015 Oct-17 60-7070-105 4012541 10/9/2015 Oct-17 60-7070-105 4014305 10/9/2015 Oct-17 60-7070-105 4016948 10/9/2015 Oct-17 60-7070-104 4005328 10/9/2015 Oct-17 60-7070-104 4008665 10/9/2015 Oct-17 60-7070-104 4013624 10/9/2015 Oct-17 60-7070-104 4013630 10/9/2015 Oct-17 60-7070-003 4013925 10/9/2015 Oct-17 60-7070-005 4005681 10/9/2015 Oct-17 60-7070-106 4012093 10/9/2015 Oct-17 60-7070-106 4012415 10/9/2015 Oct-17 60-7070-106 4012824 10/9/2015 Oct-17 60-7070-106 4012980 10/9/2015 Oct-17 60-7070-106 4015487 10/9/2015 Oct-17 60-7070-106 4015846 10/9/2015 Oct-17 60-7070-107 4015043 10/9/2015 Oct-17 60-7075-003 4005359 10/9/2015 Oct-17 60-7075-003 4015186 10/9/2015 Oct-17 60-7075-004 4016976 10/9/2015 Oct-17 60-7075-006 4017325 10/9/2015 Oct-17 60-7075-103 4005261 10/9/2015 Oct-17 60-7075-103 4013023 10/9/2015 Oct-17 60-7075-105 4016026 10/9/2015 Oct-17 60-7075-106 4014517 10/9/2015 Oct-17 60-7075-106 4014627 10/9/2015 Oct-17 5921-018-135 4021063 10/12/2015 Oct-17 5921-018-235 4020965 10/12/2015 Oct-17 5921-018-235 4021059 10/12/2015 Oct-17 5921-024-235 4020027 10/12/2015 Oct-17 5921-030-135 4019509 10/12/2015 Oct-17 5921-030-235 4018629 10/12/2015 Oct-17 5921-030-235 4022114 10/12/2015 Oct-17 5921-030-235 4022341 10/12/2015 Oct-17 5921-034-135 4019623 10/12/2015 Oct-17 5921-034-135 4022547 10/12/2015 Oct-17 5921-034-235 4018104 10/12/2015 Oct-17 5921-034-235 4021670 10/12/2015 Oct-17 5921-212-235 4019014 10/12/2015 Oct-17 5921-218-235 4017373 10/12/2015 Oct-17 5921-218-235 4020769 10/12/2015 Oct-17 5921-218-235 4023115 10/12/2015 Oct-17 5921-218-135 4017872 10/12/2015 Oct-17 5921-218-135 4017957 10/12/2015 Oct-17 5921-218-135 4021314 10/12/2015 Oct-17 60-7070-103 4018400 10/12/2015 Oct-17 60-7070-103 4020402 10/12/2015 Oct-17 60-7070-103 4020983 10/12/2015 Oct-17 60-7070-103 4023324 10/12/2015 Oct-17 60-7070-105 4020210 10/12/2015 Oct-17 60-7070-105 4020753 10/12/2015 Oct-17 60-7070-105 4021488 10/12/2015 Oct-17 60-7070-104 4017631 10/12/2015 Oct-17 60-7070-104 4021379 10/12/2015 Oct-17 60-7070-106 4017704 10/12/2015 Oct-17 60-7070-106 4017715 10/12/2015 Oct-17 60-7070-106 4017770 10/12/2015 Oct-17 60-7070-106 4020418 10/12/2015 Oct-17 60-7070-106 4020441 10/12/2015 Oct-17 60-7070-106 4022893 10/12/2015 Oct-17 60-7070-106 4022961 10/12/2015 Oct-17 60-7070-107 4019290 10/12/2015 Oct-17 60-7070-107 4022715 10/12/2015 Oct-17 60-7075-003 4022762 10/12/2015 Oct-17 60-7075-005 4023209 10/12/2015 Oct-17 60-7080-102 4019756 10/12/2015 Oct-17 60-7075-103 4019154 10/12/2015 Oct-17 60-7075-103 4021875 10/12/2015 Oct-17 60-7075-105 4019901 10/12/2015 Oct-17 60-7075-106 4018934 10/12/2015 Oct-17 60-7075-106 4022048 10/12/2015 Oct-17 5921-018-135 4026328 10/14/2015 Oct-17 5921-018-135 4026550 10/14/2015 Oct-17 5921-018-235 4029312 10/14/2015 Oct-17 5921-024-135 4024479 10/14/2015 Oct-17 5921-024-135 4027778 10/14/2015 Oct-17 5921-024-135 4029246 10/14/2015 Oct-17 5921-030-135 4023297 10/14/2015 Oct-17 5921-030-135 4027876 10/14/2015 Oct-17 5921-030-235 4023746 10/14/2015 Oct-17 5921-030-235 4023896 10/14/2015 Oct-17 5921-030-235 4026098 10/14/2015 Oct-17 5921-030-235 4028892 10/14/2015 Oct-17 5921-034-135 4024229 10/14/2015 Oct-17 5921-034-135 4026203 10/14/2015 Oct-17 5921-034-235 4024666 10/14/2015 Oct-17 5921-034-235 4024873 10/14/2015 Oct-17 5921-034-235 4026917 10/14/2015 Oct-17 5921-034-235 4027086 10/14/2015 Oct-17 5921-218-235 4025195 10/14/2015 Oct-17 5921-218-235 4026513 10/14/2015 Oct-17 5921-218-235 4029061 10/14/2015 Oct-17 5921-030-236 4028204 10/14/2015 Oct-17 5921-212-135 4023519 10/14/2015 Oct-17 5921-218-135 4025499 10/14/2015 Oct-17 5921-218-135 4025861 10/14/2015 Oct-17 5921-218-135 4029185 10/14/2015 Oct-17 5921-218-136 4025937 10/14/2015 Oct-17 60-7070-103 4023369 10/14/2015 Oct-17 60-7070-103 4024915 10/14/2015 Oct-17 60-7070-103 4027249 10/14/2015 Oct-17 60-7070-103 4029377 10/14/2015 Oct-17 60-7070-105 4024175 10/14/2015 Oct-17 60-7070-105 4025554 10/14/2015 Oct-17 60-7070-105 4025971 10/14/2015 Oct-17 60-7070-105 4027188 10/14/2015 Oct-17 60-7070-105 4028800 10/14/2015 Oct-17 60-7070-105 4028871 10/14/2015 Oct-17 60-7070-104 4024272 10/14/2015 Oct-17 60-7070-104 4026051 10/14/2015 Oct-17 60-7070-104 4029003 10/14/2015 Oct-17 60-7070-106 4025130 10/14/2015 Oct-17 60-7070-106 4025362 10/14/2015 Oct-17 60-7070-106 4028270 10/14/2015 Oct-17 60-7075-004 4023535 10/14/2015 Oct-17 60-7075-006 4028166 10/14/2015 Oct-17 60-7075-104 4023018 10/14/2015 Oct-17 60-7075-104 4028682 10/14/2015 Oct-17 60-7075-105 4026344 10/14/2015 Oct-17 60-7075-106 4023947 10/14/2015 Oct-17 60-7075-106 4026870 10/14/2015 Oct-17 60-7075-106 4027967 10/14/2015 Oct-17 60-7075-107 4025985 10/14/2015 Oct-17 5921-024-235 4033725 10/19/2015 Oct-17 5921-024-235 4038539 10/19/2015 Oct-17 5921-030-135 4032517 10/19/2015 Oct-17 5921-030-135 4032615 10/19/2015 Oct-17 5921-030-135 4034949 10/19/2015 Oct-17 5921-030-235 4033993 10/19/2015 Oct-17 5921-030-235 4034418 10/19/2015 Oct-17 5921-030-235 4039225 10/19/2015 Oct-17 5921-034-135 4030507 10/19/2015 Oct-17 5921-034-235 4031990 10/19/2015 Oct-17 5921-034-235 4032224 10/19/2015 Oct-17 5921-034-235 4035897 10/19/2015 Oct-17 5921-034-235 4038692 10/19/2015 Oct-17 5921-034-235 4038848 10/19/2015 Oct-17 5921-218-235 4033804 10/19/2015 Oct-17 5921-218-235 4038135 10/19/2015 Oct-17 60-7070-102 4030915 10/19/2015 Oct-17 60-7070-103 4033177 10/19/2015 Oct-17 60-7070-103 4033453 10/19/2015 Oct-17 60-7070-103 4036471 10/19/2015 Oct-17 60-7070-103 4036910 10/19/2015 Oct-17 60-7070-103 4039343 10/19/2015 Oct-17 60-7070-105 4031581 10/19/2015 Oct-17 60-7070-105 4032913 10/19/2015 Oct-17 60-7070-105 4034690 10/19/2015 Oct-17 60-7070-105 4034846 10/19/2015 Oct-17 60-7070-105 4037885 10/19/2015 Oct-17 60-7070-104 4031715 10/19/2015 Oct-17 60-7070-104 4031758 10/19/2015 Oct-17 60-7070-104 4037178 10/19/2015 Oct-17 60-7070-006 4034140 10/19/2015 Oct-17 60-7070-106 4029486 10/19/2015 Oct-17 60-7070-106 4035233 10/19/2015 Oct-17 60-7070-106 4035673 10/19/2015 Oct-17 60-7070-106 4037449 10/19/2015 Oct-17 60-7070-106 4037630 10/19/2015 Oct-17 60-7075-006 4032282 10/19/2015 Oct-17 60-7075-103 4029570 10/19/2015 Oct-17 60-7075-103 4029591 10/19/2015 Oct-17 60-7075-103 4035723 10/19/2015 Oct-17 60-7075-103 4038375 10/19/2015 Oct-17 60-7075-105 4032783 10/19/2015 Oct-17 60-7075-105 4032834 10/19/2015 Oct-17 5921-018-135 4039772 10/20/2015 Oct-17 5921-018-235 4042116 10/20/2015 Oct-17 5921-024-135 4044574 10/20/2015 Oct-17 5921-024-235 4040075 10/20/2015 Oct-17 5921-030-135 4040925 10/20/2015 Oct-17 5921-030-235 4043479 10/20/2015 Oct-17 5921-034-135 4040412 10/20/2015 Oct-17 5921-034-135 4040676 10/20/2015 Oct-17 5921-034-135 4046517 10/20/2015 Oct-17 5921-034-235 4043915 10/20/2015 Oct-17 5921-218-235 4044779 10/20/2015 Oct-17 60-7070-103 4039633 10/20/2015 Oct-17 60-7070-103 4042233 10/20/2015 Oct-17 60-7070-103 4042246 10/20/2015 Oct-17 60-7070-103 4044722 10/20/2015 Oct-17 60-7070-105 4041177 10/20/2015 Oct-17 60-7070-105 4043168 10/20/2015 Oct-17 60-7070-105 4043433 10/20/2015 Oct-17 60-7070-105 4045853 10/20/2015 Oct-17 60-7070-003 4039890 10/20/2015 Oct-17 60-7070-106 4041495 10/20/2015 Oct-17 60-7070-106 4041521 10/20/2015 Oct-17 60-7070-106 4041964 10/20/2015 Oct-17 60-7070-106 4045352 10/20/2015 Oct-17 60-7070-106 4045470 10/20/2015 Oct-17 60-7070-106 4045869 10/20/2015 Oct-17 60-7070-107 4044324 10/20/2015 Oct-17 60-7075-103 4042821 10/20/2015 Oct-17 60-7075-104 4043771 10/20/2015 Oct-17 60-7075-106 4040132 10/20/2015 Oct-17 5921-024-135 4051580 10/22/2015 Oct-17 5921-024-235 4053614 10/22/2015 Oct-17 5921-030-135 4048321 10/22/2015 Oct-17 5921-030-135 4048460 10/22/2015 Oct-17 5921-030-135 4054312 10/22/2015 Oct-17 5921-030-235 4050432 10/22/2015 Oct-17 5921-034-135 4046529 10/22/2015 Oct-17 5921-034-135 4049228 10/22/2015 Oct-17 5921-034-135 4052528 10/22/2015 Oct-17 5921-034-235 4047583 10/22/2015 Oct-17 5921-034-235 4047937 10/22/2015 Oct-17 5921-034-235 4053819 10/22/2015 Oct-17 5921-034-235 4054558 10/22/2015 Oct-17 5921-218-235 4046426 10/22/2015 Oct-17 5921-218-235 4050864 10/22/2015 Oct-17 5921-218-135 4050244 10/22/2015 Oct-17 5921-218-135 4051329 10/22/2015 Oct-17 60-7070-103 4046810 10/22/2015 Oct-17 60-7070-103 4047273 10/22/2015 Oct-17 60-7070-103 4050030 10/22/2015 Oct-17 60-7070-103 4053273 10/22/2015 Oct-17 60-7070-105 4052332 10/22/2015 Oct-17 60-7070-105 4052494 10/22/2015 Oct-17 60-7070-104 4048089 10/22/2015 Oct-17 60-7070-104 4051669 10/22/2015 Oct-17 60-7070-106 4049318 10/22/2015 Oct-17 60-7070-106 4049341 10/22/2015 Oct-17 60-7075-006 3970432 10/22/2015 Oct-17 60-7075-006 3970523 10/22/2015 Oct-17 60-7075-103 4046909 10/22/2015 Oct-17 60-7075-103 4047182 10/22/2015 Oct-17 60-7075-105 4049030 10/22/2015 Oct-17 60-7075-105 4049198 10/22/2015 Oct-17 60-7075-106 4048688 10/22/2015 Oct-17 60-7075-106 4053389 10/22/2015 Oct-17 5921-018-135 4058875 10/26/2015 Oct-17 5921-024-235 4057739 10/26/2015 Oct-17 5921-030-135 4059915 10/26/2015 Oct-17 5921-030-235 4056541 10/26/2015 Oct-17 5921-030-235 4061206 10/26/2015 Oct-17 5921-034-135 4058253 10/26/2015 Oct-17 5921-034-235 4054656 10/26/2015 Oct-17 5921-034-235 4060562 10/26/2015 Oct-17 5921-034-235 4060878 10/26/2015 Oct-17 5921-218-235 4055714 10/26/2015 Oct-17 5921-218-235 4060128 10/26/2015 Oct-17 5921-218-235 4060365 10/26/2015 Oct-17 60-7070-103 4055115 10/26/2015 Oct-17 60-7070-103 4057550 10/26/2015 Oct-17 60-7070-103 4058230 10/26/2015 Oct-17 60-7070-103 4061384 10/26/2015 Oct-17 60-7070-105 4056323 10/26/2015 Oct-17 60-7070-105 4059785 10/26/2015 Oct-17 60-7070-105 4060948 10/26/2015 Oct-17 60-7070-104 4057469 10/26/2015 Oct-17 60-7070-106 4056671 10/26/2015 Oct-17 60-7070-106 4056713 10/26/2015 Oct-17 60-7070-106 4059199 10/26/2015 Oct-17 60-7070-106 4059467 10/26/2015 Oct-17 60-7070-106 4059650 10/26/2015 Oct-17 60-7075-103 4055503 10/26/2015 Oct-17 60-7075-103 4055590 10/26/2015 Oct-17 60-7075-103 4058708 10/26/2015 Oct-17 60-7075-103 4058796 10/26/2015 Oct-17 60-7075-106 4057827 10/26/2015 Oct-17 60-7075-106 4058028 10/26/2015 Oct-17 5921-018-135 4068131 10/28/2015 Oct-17 5921-018-235 4067248 10/28/2015 Oct-17 5921-018-235 4067488 10/28/2015 Oct-17 5921-024-135 4068033 10/28/2015 Oct-17 5921-024-136 4061657 10/28/2015 Oct-17 5921-024-235 4065785 10/28/2015 Oct-17 5921-024-235 4069104 10/28/2015 Oct-17 5921-030-135 4065860 10/28/2015 Oct-17 5921-030-135 4066097 10/28/2015 Oct-17 5921-030-235 4065217 10/28/2015 Oct-17 5921-034-135 4062301 10/28/2015 Oct-17 5921-034-135 4062607 10/28/2015 Oct-17 5921-034-135 4062885 10/28/2015 Oct-17 5921-034-135 4065498 10/28/2015 Oct-17 5921-034-135 4066764 10/28/2015 Oct-17 5921-034-135 4066928 10/28/2015 Oct-17 5921-034-135 4068940 10/28/2015 Oct-17 5921-034-136 4062023 10/28/2015 Oct-17 5921-034-235 4063659 10/28/2015 Oct-17 5921-034-235 4064654 10/28/2015 Oct-17 5921-034-235 4066667 10/28/2015 Oct-17 5921-034-235 4067224 10/28/2015 Oct-17 5921-034-235 4068532 10/28/2015 Oct-17 5921-034-236 4069416 10/28/2015 Oct-17 5921-218-235 4066241 10/28/2015 Oct-17 5921-030-236 4066565 10/28/2015 Oct-17 5921-044-135 4067643 10/28/2015 Oct-17 5921-044-235 4066389 10/28/2015 Oct-17 5921-218-135 4061912 10/28/2015 Oct-17 5921-218-135 4063946 10/28/2015 Oct-17 5921-218-135 4067852 10/28/2015 Oct-17 60-7070-103 4061782 10/28/2015 Oct-17 60-7070-103 4065595 10/28/2015 Oct-17 60-7070-105 4064263 10/28/2015 Oct-17 60-7070-105 4064758 10/28/2015 Oct-17 60-7070-105 4066499 10/28/2015 Oct-17 60-7070-105 4066680 10/28/2015 Oct-17 60-7070-105 4066937 10/28/2015 Oct-17 60-7070-105 4068827 10/28/2015 Oct-17 60-7070-104 4062121 10/28/2015 Oct-17 60-7070-104 4064942 10/28/2015 Oct-17 60-7070-104 4067615 10/28/2015 Oct-17 60-7070-104 4069189 10/28/2015 Oct-17 60-7070-003 4062340 10/28/2015 Oct-17 60-7070-003 4062611 10/28/2015 Oct-17 60-7070-106 4063135 10/28/2015 Oct-17 60-7070-106 4066294 10/28/2015 Oct-17 60-7070-106 4068290 10/28/2015 Oct-17 60-7070-106 4069629 10/28/2015 Oct-17 60-7070-107 4067540 10/28/2015 Oct-17 60-7075-006 4061867 10/28/2015 Oct-17 60-7075-102 4065467 10/28/2015 Oct-17 60-7075-105 4065134 10/28/2015 Oct-17 60-7075-105 4069278 10/28/2015 Oct-17 60-7075-106 4066009 10/28/2015 Oct-17 5921-018-135 4071561 10/30/2015 Oct-17 5921-018-235 4072130 10/30/2015 Oct-17 5921-024-135 4071250 10/30/2015 Oct-17 5921-024-235 4074955 10/30/2015 Oct-17 5921-030-135 4070427 10/30/2015 Oct-17 5921-030-135 4070644 10/30/2015 Oct-17 5921-030-135 4076105 10/30/2015 Oct-17 5921-030-235 4071039 10/30/2015 Oct-17 5921-030-235 4071056 10/30/2015 Oct-17 5921-034-135 4071503 10/30/2015 Oct-17 5921-034-135 4072218 10/30/2015 Oct-17 5921-034-135 4075301 10/30/2015 Oct-17 5921-034-235 4071877 10/30/2015 Oct-17 5921-034-235 4071934 10/30/2015 Oct-17 5921-034-235 4071999 10/30/2015 Oct-17 5921-034-235 4073129 10/30/2015 Oct-17 5921-034-235 4074858 10/30/2015 Oct-17 5921-218-235 4074274 10/30/2015 Oct-17 5921-218-235 4074469 10/30/2015 Oct-17 5921-218-135 4069877 10/30/2015 Oct-17 5921-218-135 4071471 10/30/2015 Oct-17 5921-218-136 4072971 10/30/2015 Oct-17 5921-218-236 4073004 10/30/2015 Oct-17 60-7070-103 4071820 10/30/2015 Oct-17 60-7070-103 4072032 10/30/2015 Oct-17 60-7070-103 4072764 10/30/2015 Oct-17 60-7070-103 4075973 10/30/2015 Oct-17 60-7070-103 4076244 10/30/2015 Oct-17 60-7070-103 4076612 10/30/2015 Oct-17 60-7070-105 4070778 10/30/2015 Oct-17 60-7070-105 4070922 10/30/2015 Oct-17 60-7070-105 4071143 10/30/2015 Oct-17 60-7070-105 4072287 10/30/2015 Oct-17 60-7070-105 4072389 10/30/2015 Oct-17 60-7070-104 4071136 10/30/2015 Oct-17 60-7070-003 4075243 10/30/2015 Oct-17 60-7070-004 4075227 10/30/2015 Oct-17 60-7070-005 4070040 10/30/2015 Oct-17 60-7070-005 4070064 10/30/2015 Oct-17 60-7070-006 4073748 10/30/2015 Oct-17 60-7070-006 4075289 10/30/2015 Oct-17 60-7070-106 4069664 10/30/2015 Oct-17 60-7070-106 4070967 10/30/2015 Oct-17 60-7070-106 4071060 10/30/2015 Oct-17 60-7070-106 4072410 10/30/2015 Oct-17 60-7070-106 4072475 10/30/2015 Oct-17 60-7070-106 4075598 10/30/2015 Oct-17 60-7070-106 4075774 10/30/2015 Oct-17 60-7070-107 4071245 10/30/2015 Oct-17 60-7075-003 4071722 10/30/2015 Oct-17 60-7075-004 4075859 10/30/2015 Oct-17 60-7075-006 4075145 10/30/2015 Oct-17 60-7080-102 4073851 10/30/2015 Oct-17 60-7075-102 4075695 10/30/2015 Oct-17 60-7075-102 4075790 10/30/2015 Oct-17 60-7075-103 4070093 10/30/2015 Oct-17 60-7075-103 4074584 10/30/2015 Oct-17 60-7075-104 4073089 10/30/2015 Oct-17 60-7075-106 4071751 10/30/2015 Oct-17 60-7075-106 4073977 10/30/2015 Oct-17 60-7075-107 4073070 10/30/2015 Oct-17 60-7075-053 4069779 10/30/2015 Oct-17 5921-018-135 4080981 11/2/2015 Nov-17 5921-024-135 4078077 11/2/2015 Nov-17 5921-024-235 4079490 11/2/2015 Nov-17 5921-024-235 4081884 11/2/2015 Nov-17 5921-030-135 4076736 11/2/2015 Nov-17 5921-030-235 4078201 11/2/2015 Nov-17 5921-030-235 4078217 11/2/2015 Nov-17 5921-030-235 4082108 11/2/2015 Nov-17 5921-030-235 4082154 11/2/2015 Nov-17 5921-034-135 4078621 11/2/2015 Nov-17 5921-034-135 4081438 11/2/2015 Nov-17 5921-034-135 4081486 11/2/2015 Nov-17 5921-034-235 4077349 11/2/2015 Nov-17 5921-034-235 4077602 11/2/2015 Nov-17 5921-034-235 4080002 11/2/2015 Nov-17 5921-034-235 4080094 11/2/2015 Nov-17 5921-218-235 4077233 11/2/2015 Nov-17 5921-218-235 4077618 11/2/2015 Nov-17 5921-218-235 4080812 11/2/2015 Nov-17 5921-218-235 4080910 11/2/2015 Nov-17 60-7070-103 4076696 11/2/2015 Nov-17 60-7070-103 4078414 11/2/2015 Nov-17 60-7070-103 4080221 11/2/2015 Nov-17 60-7070-103 4080353 11/2/2015 Nov-17 60-7070-103 4082532 11/2/2015 Nov-17 60-7070-105 4076864 11/2/2015 Nov-17 60-7070-105 4079708 11/2/2015 Nov-17 60-7070-105 4079754 11/2/2015 Nov-17 60-7070-105 4079881 11/2/2015 Nov-17 60-7070-105 4081706 11/2/2015 Nov-17 60-7070-104 4077100 11/2/2015 Nov-17 60-7070-104 4077192 11/2/2015 Nov-17 60-7070-104 4081976 11/2/2015 Nov-17 60-7070-106 4078873 11/2/2015 Nov-17 60-7070-106 4078970 11/2/2015 Nov-17 60-7070-106 4081185 11/2/2015 Nov-17 60-7070-107 4079950 11/2/2015 Nov-17 60-7075-103 4077995 11/2/2015 Nov-17 60-7075-105 4081035 11/2/2015 Nov-17 60-7075-107 4082430 11/2/2015 Nov-17 5921-018-235 4085259 11/4/2015 Nov-17 5921-018-235 4085432 11/4/2015 Nov-17 5921-018-235 4085683 11/4/2015 Nov-17 5921-024-135 4086582 11/4/2015 Nov-17 5921-030-135 4083141 11/4/2015 Nov-17 5921-030-135 4083276 11/4/2015 Nov-17 5921-030-135 4083720 11/4/2015 Nov-17 5921-030-135 4083899 11/4/2015 Nov-17 5921-030-135 4086887 11/4/2015 Nov-17 5921-030-135 4089212 11/4/2015 Nov-17 5921-030-235 4085478 11/4/2015 Nov-17 5921-030-235 4085743 11/4/2015 Nov-17 5921-030-235 4086123 11/4/2015 Nov-17 5921-030-235 4088565 11/4/2015 Nov-17 5921-034-135 4084006 11/4/2015 Nov-17 5921-034-135 4088124 11/4/2015 Nov-17 5921-034-235 4083875 11/4/2015 Nov-17 5921-034-235 4087138 11/4/2015 Nov-17 5921-034-235 4087309 11/4/2015 Nov-17 5921-034-236 4082929 11/4/2015 Nov-17 5921-218-235 4084197 11/4/2015 Nov-17 5921-218-235 4085341 11/4/2015 Nov-17 5921-218-235 4087559 11/4/2015 Nov-17 5921-218-235 4089404 11/4/2015 Nov-17 5921-218-135 4084695 11/4/2015 Nov-17 5921-218-135 4088447 11/4/2015 Nov-17 5921-218-136 4083333 11/4/2015 Nov-17 60-7070-103 4082675 11/4/2015 Nov-17 60-7070-103 4085200 11/4/2015 Nov-17 60-7070-103 4087782 11/4/2015 Nov-17 60-7070-105 4084150 11/4/2015 Nov-17 60-7070-105 4084243 11/4/2015 Nov-17 60-7070-105 4086255 11/4/2015 Nov-17 60-7070-104 4088176 11/4/2015 Nov-17 60-7070-005 4082891 11/4/2015 Nov-17 60-7070-005 4083063 11/4/2015 Nov-17 60-7070-106 4082986 11/4/2015 Nov-17 60-7070-106 4083398 11/4/2015 Nov-17 60-7070-106 4083708 11/4/2015 Nov-17 60-7070-106 4087193 11/4/2015 Nov-17 60-7070-106 4088826 11/4/2015 Nov-17 60-7070-107 4088596 11/4/2015 Nov-17 60-7075-006 4087603 11/4/2015 Nov-17 60-7075-103 4084364 11/4/2015 Nov-17 60-7075-103 4088468 11/4/2015 Nov-17 60-7075-104 4085237 11/4/2015 Nov-17 60-7075-106 4082725 11/4/2015 Nov-17 60-7075-106 4082815 11/4/2015 Nov-17 60-7075-106 4086771 11/4/2015 Nov-17 60-7075-106 4088336 11/4/2015 Nov-17 5921-018-135 4089784 11/6/2015 Nov-17 5921-018-135 4093727 11/6/2015 Nov-17 5921-018-235 4091124 11/6/2015 Nov-17 5921-024-135 4092526 11/6/2015 Nov-17 5921-024-235 4093146 11/6/2015 Nov-17 5921-030-135 4093244 11/6/2015 Nov-17 5921-030-135 4093654 11/6/2015 Nov-17 5921-030-235 4091803 11/6/2015 Nov-17 5921-030-235 4093409 11/6/2015 Nov-17 5921-034-135 4089594 11/6/2015 Nov-17 5921-034-135 4089933 11/6/2015 Nov-17 5921-034-135 4091531 11/6/2015 Nov-17 5921-034-135 4092601 11/6/2015 Nov-17 5921-034-235 4090469 11/6/2015 Nov-17 5921-034-235 4090787 11/6/2015 Nov-17 5921-034-235 4093179 11/6/2015 Nov-17 5921-212-235 4093007 11/6/2015 Nov-17 5921-218-235 4089491 11/6/2015 Nov-17 5921-218-235 4092968 11/6/2015 Nov-17 5921-218-235 4093048 11/6/2015 Nov-17 5921-218-135 4091959 11/6/2015 Nov-17 5921-218-135 4092025 11/6/2015 Nov-17 5921-218-135 4093122 11/6/2015 Nov-17 5921-218-236 4092981 11/6/2015 Nov-17 60-7070-102 4093665 11/6/2015 Nov-17 60-7070-103 4087846 11/6/2015 Nov-17 60-7070-103 4089541 11/6/2015 Nov-17 60-7070-103 4091622 11/6/2015 Nov-17 60-7070-103 4092934 11/6/2015 Nov-17 60-7070-105 4087983 11/6/2015 Nov-17 60-7070-105 4089637 11/6/2015 Nov-17 60-7070-105 4089713 11/6/2015 Nov-17 60-7070-105 4091294 11/6/2015 Nov-17 60-7070-105 4092535 11/6/2015 Nov-17 60-7070-105 4092590 11/6/2015 Nov-17 60-7070-104 4092865 11/6/2015 Nov-17 60-7070-003 4093608 11/6/2015 Nov-17 60-7070-003 4093677 11/6/2015 Nov-17 60-7070-106 4092094 11/6/2015 Nov-17 60-7070-106 4092307 11/6/2015 Nov-17 60-7070-107 4088819 11/6/2015 Nov-17 60-7075-006 4093466 11/6/2015 Nov-17 60-7075-103 4090773 11/6/2015 Nov-17 60-7075-103 4090955 11/6/2015 Nov-17 60-7075-103 4093445 11/6/2015 Nov-17 60-7075-104 4092906 11/6/2015 Nov-17 60-7075-105 4092330 11/6/2015 Nov-17 60-7075-106 4093339 11/6/2015 Nov-17 60-7075-106 4093356 11/6/2015 Nov-17 60-7075-106 4093387 11/6/2015 Nov-17 5921-024-235 4093838 11/9/2015 Nov-17 5921-024-235 4099987 11/9/2015 Nov-17 5921-030-135 4094600 11/9/2015 Nov-17 5921-030-135 4094732 11/9/2015 Nov-17 5921-030-135 4098171 11/9/2015 Nov-17 5921-030-135 4098192 11/9/2015 Nov-17 5921-030-235 4094443 11/9/2015 Nov-17 5921-030-235 4095765 11/9/2015 Nov-17 5921-030-235 4096879 11/9/2015 Nov-17 5921-030-235 4098312 11/9/2015 Nov-17 5921-034-135 4093942 11/9/2015 Nov-17 5921-034-135 4096795 11/9/2015 Nov-17 5921-034-235 4093789 11/9/2015 Nov-17 5921-034-235 4095997 11/9/2015 Nov-17 5921-034-235 4099317 11/9/2015 Nov-17 5921-212-235 4095564 11/9/2015 Nov-17 5921-218-235 4093855 11/9/2015 Nov-17 5921-218-235 4096769 11/9/2015 Nov-17 5921-218-235 4098641 11/9/2015 Nov-17 5921-218-135 4093893 11/9/2015 Nov-17 5921-218-135 4097609 11/9/2015 Nov-17 60-7070-103 4095305 11/9/2015 Nov-17 60-7070-103 4097818 11/9/2015 Nov-17 60-7070-103 4099010 11/9/2015 Nov-17 60-7070-103 4099145 11/9/2015 Nov-17 60-7070-105 4093903 11/9/2015 Nov-17 60-7070-105 4096210 11/9/2015 Nov-17 60-7070-105 4097110 11/9/2015 Nov-17 60-7070-105 4100137 11/9/2015 Nov-17 60-7070-104 4097049 11/9/2015 Nov-17 60-7070-006 4096898 11/9/2015 Nov-17 60-7070-006 4097031 11/9/2015 Nov-17 60-7070-106 4095224 11/9/2015 Nov-17 60-7070-106 4097272 11/9/2015 Nov-17 60-7070-106 4097319 11/9/2015 Nov-17 60-7070-106 4099393 11/9/2015 Nov-17 60-7070-107 4094381 11/9/2015 Nov-17 60-7070-107 4098090 11/9/2015 Nov-17 60-7075-003 4099790 11/9/2015 Nov-17 60-7075-105 4098558 11/9/2015 Nov-17 60-7075-106 4098019 11/9/2015 Nov-17 5921-018-135 4107207 11/11/2015 Nov-17 5921-030-135 4108681 11/11/2015 Nov-17 5921-034-135 4106554 11/11/2015 Nov-17 5921-034-135 4108482 11/11/2015 Nov-17 5921-034-235 4104007 11/11/2015 Nov-17 5921-034-235 4106819 11/11/2015 Nov-17 5921-034-235 4108359 11/11/2015 Nov-17 5921-034-235 4108610 11/11/2015 Nov-17 5921-218-235 4106050 11/11/2015 Nov-17 5921-218-235 4107490 11/11/2015 Nov-17 5921-218-235 4108120 11/11/2015 Nov-17 5921-218-235 4108509 11/11/2015 Nov-17 5921-218-135 4107868 11/11/2015 Nov-17 60-7070-103 4104409 11/11/2015 Nov-17 60-7070-103 4104926 11/11/2015 Nov-17 60-7070-103 4105697 11/11/2015 Nov-17 60-7070-103 4106362 11/11/2015 Nov-17 60-7070-103 4107085 11/11/2015 Nov-17 60-7070-105 4100159 11/11/2015 Nov-17 60-7070-105 4105106 11/11/2015 Nov-17 60-7070-104 4103628 11/11/2015 Nov-17 60-7070-104 4104665 11/11/2015 Nov-17 60-7070-106 4106531 11/11/2015 Nov-17 60-7070-106 4107248 11/11/2015 Nov-17 60-7070-106 4107932 11/11/2015 Nov-17 60-7075-103 4106343 11/11/2015 Nov-17 60-7075-103 4108108 11/11/2015 Nov-17 60-7075-104 4107435 11/11/2015 Nov-17 60-7075-106 4106928 11/11/2015 Nov-17 60-7075-106 4108316 11/11/2015 Nov-17 5921-018-135 4109831 11/13/2015 Nov-17 5921-018-135 4114924 11/13/2015 Nov-17 5921-018-235 4115060 11/13/2015 Nov-17 5921-024-135 4109041 11/13/2015 Nov-17 5921-024-135 4112559 11/13/2015 Nov-17 5921-024-135 4113622 11/13/2015 Nov-17 5921-024-235 4112544 11/13/2015 Nov-17 5921-024-235 4114878 11/13/2015 Nov-17 5921-030-135 4110466 11/13/2015 Nov-17 5921-030-135 4115463 11/13/2015 Nov-17 5921-030-135 4115638 11/13/2015 Nov-17 5921-030-235 4110259 11/13/2015 Nov-17 5921-030-235 4113717 11/13/2015 Nov-17 5921-034-135 4110840 11/13/2015 Nov-17 5921-034-135 4110957 11/13/2015 Nov-17 5921-034-135 4111844 11/13/2015 Nov-17 5921-034-135 4113888 11/13/2015 Nov-17 5921-034-135 4113971 11/13/2015 Nov-17 5921-034-135 4115808 11/13/2015 Nov-17 5921-034-135 4115968 11/13/2015 Nov-17 5921-034-235 4109947 11/13/2015 Nov-17 5921-034-235 4111068 11/13/2015 Nov-17 5921-034-235 4111669 11/13/2015 Nov-17 5921-034-235 4113583 11/13/2015 Nov-17 5921-034-235 4114464 11/13/2015 Nov-17 5921-034-235 4115906 11/13/2015 Nov-17 5921-218-235 4110789 11/13/2015 Nov-17 5921-218-235 4113304 11/13/2015 Nov-17 5921-218-235 4114591 11/13/2015 Nov-17 5921-218-235 4115640 11/13/2015 Nov-17 5921-218-235 4115746 11/13/2015 Nov-17 5921-218-135 4111624 11/13/2015 Nov-17 5921-218-135 4112313 11/13/2015 Nov-17 5921-218-135 4114504 11/13/2015 Nov-17 5921-218-135 4116161 11/13/2015 Nov-17 60-7070-102 4114710 11/13/2015 Nov-17 60-7070-103 4108739 11/13/2015 Nov-17 60-7070-103 4109617 11/13/2015 Nov-17 60-7070-103 4111941 11/13/2015 Nov-17 60-7070-103 4112299 11/13/2015 Nov-17 60-7070-103 4112487 11/13/2015 Nov-17 60-7070-103 4113963 11/13/2015 Nov-17 60-7070-103 4114969 11/13/2015 Nov-17 60-7070-105 4108843 11/13/2015 Nov-17 60-7070-105 4111567 11/13/2015 Nov-17 60-7070-105 4114298 11/13/2015 Nov-17 60-7070-104 4110935 11/13/2015 Nov-17 60-7070-104 4112402 11/13/2015 Nov-17 60-7070-106 4110523 11/13/2015 Nov-17 60-7070-106 4110974 11/13/2015 Nov-17 60-7070-106 4112117 11/13/2015 Nov-17 60-7070-106 4114061 11/13/2015 Nov-17 60-7070-106 4115492 11/13/2015 Nov-17 60-7070-106 4116174 11/13/2015 Nov-17 60-7070-107 4109705 11/13/2015 Nov-17 60-7075-004 4111215 11/13/2015 Nov-17 60-7075-006 4112473 11/13/2015 Nov-17 60-7075-103 4108827 11/13/2015 Nov-17 60-7075-103 4109664 11/13/2015 Nov-17 60-7075-103 4113287 11/13/2015 Nov-17 60-7075-103 4114280 11/13/2015 Nov-17 60-7075-106 4110030 11/13/2015 Nov-17 60-7075-106 4112129 11/13/2015 Nov-17 60-7075-106 4114660 11/13/2015 Nov-17 5921-018-135 4122177 11/16/2015 Nov-17 5921-018-235 4118689 11/16/2015 Nov-17 5921-024-135 4121440 11/16/2015 Nov-17 5921-030-135 4120056 11/16/2015 Nov-17 5921-030-135 4122972 11/16/2015 Nov-17 5921-030-235 4116443 11/16/2015 Nov-17 5921-030-235 4118367 11/16/2015 Nov-17 5921-030-235 4119915 11/16/2015 Nov-17 5921-030-235 4122235 11/16/2015 Nov-17 5921-030-235 4124121 11/16/2015 Nov-17 5921-034-135 4117940 11/16/2015 Nov-17 5921-034-135 4120353 11/16/2015 Nov-17 5921-034-135 4122509 11/16/2015 Nov-17 5921-034-135 4123457 11/16/2015 Nov-17 5921-034-235 4117929 11/16/2015 Nov-17 5921-034-235 4119645 11/16/2015 Nov-17 5921-034-235 4121280 11/16/2015 Nov-17 5921-034-235 4123160 11/16/2015 Nov-17 5921-034-235 4123403 11/16/2015 Nov-17 5921-218-235 4118071 11/16/2015 Nov-17 5921-218-235 4120115 11/16/2015 Nov-17 5921-218-235 4121570 11/16/2015 Nov-17 5921-218-235 4122532 11/16/2015 Nov-17 5921-218-235 4123517 11/16/2015 Nov-17 5921-218-135 4118351 11/16/2015 Nov-17 5921-218-135 4120526 11/16/2015 Nov-17 5921-218-135 4121884 11/16/2015 Nov-17 5921-218-135 4123322 11/16/2015 Nov-17 5921-218-135 4123555 11/16/2015 Nov-17 5921-218-135 4123712 11/16/2015 Nov-17 60-7070-102 4120471 11/16/2015 Nov-17 60-7070-103 4116323 11/16/2015 Nov-17 60-7070-103 4118230 11/16/2015 Nov-17 60-7070-103 4119405 11/16/2015 Nov-17 60-7070-103 4119792 11/16/2015 Nov-17 60-7070-103 4121472 11/16/2015 Nov-17 60-7070-103 4121878 11/16/2015 Nov-17 60-7070-103 4122120 11/16/2015 Nov-17 60-7070-103 4123028 11/16/2015 Nov-17 60-7070-103 4124236 11/16/2015 Nov-17 60-7070-103 4125007 11/16/2015 Nov-17 60-7070-105 4117219 11/16/2015 Nov-17 60-7070-105 4117550 11/16/2015 Nov-17 60-7070-105 4118823 11/16/2015 Nov-17 60-7070-105 4120025 11/16/2015 Nov-17 60-7070-105 4121809 11/16/2015 Nov-17 60-7070-105 4123852 11/16/2015 Nov-17 60-7070-104 4116556 11/16/2015 Nov-17 60-7070-104 4120221 11/16/2015 Nov-17 60-7070-104 4122634 11/16/2015 Nov-17 60-7070-106 4116040 11/16/2015 Nov-17 60-7070-106 4117741 11/16/2015 Nov-17 60-7070-106 4118501 11/16/2015 Nov-17 60-7070-106 4118966 11/16/2015 Nov-17 60-7070-106 4119451 11/16/2015 Nov-17 60-7070-106 4120324 11/16/2015 Nov-17 60-7070-106 4120617 11/16/2015 Nov-17 60-7070-106 4122196 11/16/2015 Nov-17 60-7070-106 4123689 11/16/2015 Nov-17 60-7070-106 4123997 11/16/2015 Nov-17 60-7075-006 4117725 11/16/2015 Nov-17 60-7075-103 4117177 11/16/2015 Nov-17 60-7075-103 4118173 11/16/2015 Nov-17 60-7075-103 4121090 11/16/2015 Nov-17 60-7075-103 4122242 11/16/2015 Nov-17 60-7075-103 4123125 11/16/2015 Nov-17 60-7075-106 4116541 11/16/2015 Nov-17 60-7075-106 4118538 11/16/2015 Nov-17 60-7075-106 4120250 11/16/2015 Nov-17 60-7075-106 4121710 11/16/2015 Nov-17 5921-018-135 4125926 11/18/2015 Nov-17 5921-024-135 4125736 11/18/2015 Nov-17 5921-024-235 4127487 11/18/2015 Nov-17 5921-024-235 4131385 11/18/2015 Nov-17 5921-030-135 4125114 11/18/2015 Nov-17 5921-030-235 4126927 11/18/2015 Nov-17 5921-034-135 4125432 11/18/2015 Nov-17 5921-034-235 4125122 11/18/2015 Nov-17 5921-034-235 4125838 11/18/2015 Nov-17 5921-034-235 4126839 11/18/2015 Nov-17 5921-034-235 4127397 11/18/2015 Nov-17 5921-034-235 4128861 11/18/2015 Nov-17 5921-034-235 4130580 11/18/2015 Nov-17 5921-218-235 4125292 11/18/2015 Nov-17 5921-218-235 4126065 11/18/2015 Nov-17 5921-218-235 4127822 11/18/2015 Nov-17 5921-218-235 4129220 11/18/2015 Nov-17 5921-218-235 4131363 11/18/2015 Nov-17 5921-218-135 4125216 11/18/2015 Nov-17 60-7070-103 4125957 11/18/2015 Nov-17 60-7070-103 4126264 11/18/2015 Nov-17 60-7070-103 4127164 11/18/2015 Nov-17 60-7070-103 4128013 11/18/2015 Nov-17 60-7070-103 4130412 11/18/2015 Nov-17 60-7070-103 4130742 11/18/2015 Nov-17 60-7070-103 4131469 11/18/2015 Nov-17 60-7070-103 4131718 11/18/2015 Nov-17 60-7070-103 4131889 11/18/2015 Nov-17 60-7070-105 4125612 11/18/2015 Nov-17 60-7070-105 4127027 11/18/2015 Nov-17 60-7070-105 4127730 11/18/2015 Nov-17 60-7070-105 4130522 11/18/2015 Nov-17 60-7070-105 4130941 11/18/2015 Nov-17 60-7070-104 4125421 11/18/2015 Nov-17 60-7070-104 4128646 11/18/2015 Nov-17 60-7070-104 4131203 11/18/2015 Nov-17 60-7070-106 4125179 11/18/2015 Nov-17 60-7070-106 4126357 11/18/2015 Nov-17 60-7070-106 4127252 11/18/2015 Nov-17 60-7070-106 4127619 11/18/2015 Nov-17 60-7070-106 4129064 11/18/2015 Nov-17 60-7070-106 4130653 11/18/2015 Nov-17 60-7070-106 4131079 11/18/2015 Nov-17 60-7070-106 4131827 11/18/2015 Nov-17 60-7075-006 4126077 11/18/2015 Nov-17 60-7075-106 4124719 11/18/2015 Nov-17 5921-018-135 4132911 11/20/2015 Nov-17 5921-018-235 4133031 11/20/2015 Nov-17 5921-018-235 4138186 11/20/2015 Nov-17 5921-024-235 4132948 11/20/2015 Nov-17 5921-024-235 4133134 11/20/2015 Nov-17 5921-024-235 4135458 11/20/2015 Nov-17 5921-030-135 4130648 11/20/2015 Nov-17 5921-030-135 4131005 11/20/2015 Nov-17 5921-030-135 4133240 11/20/2015 Nov-17 5921-030-135 4136984 11/20/2015 Nov-17 5921-030-135 4138268 11/20/2015 Nov-17 5921-030-136 4138585 11/20/2015 Nov-17 5921-030-235 4132365 11/20/2015 Nov-17 5921-030-235 4132832 11/20/2015 Nov-17 5921-030-235 4134812 11/20/2015 Nov-17 5921-030-235 4136540 11/20/2015 Nov-17 5921-034-135 4131337 11/20/2015 Nov-17 5921-034-135 4132641 11/20/2015 Nov-17 5921-034-135 4132953 11/20/2015 Nov-17 5921-034-135 4133110 11/20/2015 Nov-17 5921-034-135 4136016 11/20/2015 Nov-17 5921-034-135 4136765 11/20/2015 Nov-17 5921-034-235 4132673 11/20/2015 Nov-17 5921-034-235 4134025 11/20/2015 Nov-17 5921-034-235 4136380 11/20/2015 Nov-17 5921-034-235 4137032 11/20/2015 Nov-17 5921-034-235 4137370 11/20/2015 Nov-17 5921-034-235 4138441 11/20/2015 Nov-17 5921-218-235 4131972 11/20/2015 Nov-17 5921-218-235 4134376 11/20/2015 Nov-17 5921-218-235 4136160 11/20/2015 Nov-17 5921-218-235 4136658 11/20/2015 Nov-17 5921-218-135 4136474 11/20/2015 Nov-17 5921-218-135 4138091 11/20/2015 Nov-17 60-7070-102 4135231 11/20/2015 Nov-17 60-7070-103 4132320 11/20/2015 Nov-17 60-7070-103 4132734 11/20/2015 Nov-17 60-7070-103 4132768 11/20/2015 Nov-17 60-7070-103 4133152 11/20/2015 Nov-17 60-7070-103 4134557 11/20/2015 Nov-17 60-7070-103 4135001 11/20/2015 Nov-17 60-7070-103 4136063 11/20/2015 Nov-17 60-7070-103 4136950 11/20/2015 Nov-17 60-7070-103 4137153 11/20/2015 Nov-17 60-7070-103 4138309 11/20/2015 Nov-17 60-7070-105 4132163 11/20/2015 Nov-17 60-7070-105 4132930 11/20/2015 Nov-17 60-7070-105 4134189 11/20/2015 Nov-17 60-7070-105 4136488 11/20/2015 Nov-17 60-7070-105 4138043 11/20/2015 Nov-17 60-7070-105 4138661 11/20/2015 Nov-17 60-7070-104 4132251 11/20/2015 Nov-17 60-7070-104 4133201 11/20/2015 Nov-17 60-7070-104 4136783 11/20/2015 Nov-17 60-7070-104 4138879 11/20/2015 Nov-17 60-7070-106 4131901 11/20/2015 Nov-17 60-7070-106 4132349 11/20/2015 Nov-17 60-7070-106 4133737 11/20/2015 Nov-17 60-7070-106 4134672 11/20/2015 Nov-17 60-7070-106 4135119 11/20/2015 Nov-17 60-7070-106 4137218 11/20/2015 Nov-17 60-7070-107 4135541 11/20/2015 Nov-17 60-7075-003 4136572 11/20/2015 Nov-17 60-7075-004 4132775 11/20/2015 Nov-17 60-7075-103 4130828 11/20/2015 Nov-17 60-7075-104 4130551 11/20/2015 Nov-17 60-7075-104 4137928 11/20/2015 Nov-17 60-7075-105 4137202 11/20/2015 Nov-17 60-7075-106 4131411 11/20/2015 Nov-17 60-7075-106 4136230 11/20/2015 Nov-17 60-7075-106 4137406 11/20/2015 Nov-17 5921-018-135 3864025 11/23/2015 Nov-17 5921-018-135 4148999 11/24/2015 Nov-17 5921-018-135 4152655 11/24/2015 Nov-17 5921-018-235 4147413 11/24/2015 Nov-17 5921-018-235 4150226 11/24/2015 Nov-17 5921-024-235 4148722 11/24/2015 Nov-17 5921-024-235 4150862 11/24/2015 Nov-17 5921-030-135 4149371 11/24/2015 Nov-17 5921-030-135 4153024 11/24/2015 Nov-17 5921-030-235 4150521 11/24/2015 Nov-17 5921-034-135 4148854 11/24/2015 Nov-17 5921-034-135 4149901 11/24/2015 Nov-17 5921-034-135 4150550 11/24/2015 Nov-17 5921-034-135 4150901 11/24/2015 Nov-17 5921-034-135 4152947 11/24/2015 Nov-17 5921-034-235 4147285 11/24/2015 Nov-17 5921-034-235 4147758 11/24/2015 Nov-17 5921-034-235 4148588 11/24/2015 Nov-17 5921-034-235 4151169 11/24/2015 Nov-17 5921-034-235 4152056 11/24/2015 Nov-17 5921-034-235 4153122 11/24/2015 Nov-17 5921-218-235 4096425 11/24/2015 Nov-17 5921-218-235 4147547 11/24/2015 Nov-17 5921-218-235 4148330 11/24/2015 Nov-17 5921-218-235 4149862 11/24/2015 Nov-17 5921-218-235 4151481 11/24/2015 Nov-17 5921-218-235 4152952 11/24/2015 Nov-17 5921-218-235 4152999 11/24/2015 Nov-17 60-7070-103 4096332 11/24/2015 Nov-17 60-7070-103 4147167 11/24/2015 Nov-17 60-7070-103 4148428 11/24/2015 Nov-17 60-7070-103 4149337 11/24/2015 Nov-17 60-7070-103 4150424 11/24/2015 Nov-17 60-7070-105 4147942 11/24/2015 Nov-17 60-7070-105 4149033 11/24/2015 Nov-17 60-7070-105 4149453 11/24/2015 Nov-17 60-7070-105 4151107 11/24/2015 Nov-17 60-7070-105 4152369 11/24/2015 Nov-17 60-7070-105 4152588 11/24/2015 Nov-17 60-7070-105 4152934 11/24/2015 Nov-17 60-7070-104 4149411 11/24/2015 Nov-17 60-7070-104 4151577 11/24/2015 Nov-17 60-7070-004 4096313 11/24/2015 Nov-17 60-7070-106 4148127 11/24/2015 Nov-17 60-7070-106 4148162 11/24/2015 Nov-17 60-7070-106 4149120 11/24/2015 Nov-17 60-7070-106 4150154 11/24/2015 Nov-17 60-7070-106 4150675 11/24/2015 Nov-17 60-7070-106 4151362 11/24/2015 Nov-17 60-7070-106 4151916 11/24/2015 Nov-17 60-7070-106 4152245 11/24/2015 Nov-17 60-7070-106 4152437 11/24/2015 Nov-17 60-7070-106 4152798 11/24/2015 Nov-17 60-7070-106 4153244 11/24/2015 Nov-17 60-7075-003 4152906 11/24/2015 Nov-17 60-7075-006 4150316 11/24/2015 Nov-17 60-7075-103 4095984 11/24/2015 Nov-17 60-7075-105 4149149 11/24/2015 Nov-17 60-7075-105 4152981 11/24/2015 Nov-17 5921-018-135 4142732 11/25/2015 Nov-17 5921-018-135 4145036 11/25/2015 Nov-17 5921-018-235 4142797 11/25/2015 Nov-17 5921-024-235 4139466 11/25/2015 Nov-17 5921-024-235 4144071 11/25/2015 Nov-17 5921-030-135 4146019 11/25/2015 Nov-17 5921-030-135 4146222 11/25/2015 Nov-17 5921-030-235 4139180 11/25/2015 Nov-17 5921-030-235 4140890 11/25/2015 Nov-17 5921-030-235 4142556 11/25/2015 Nov-17 5921-034-135 4142093 11/25/2015 Nov-17 5921-034-135 4144720 11/25/2015 Nov-17 5921-034-135 4147011 11/25/2015 Nov-17 5921-034-136 4146641 11/25/2015 Nov-17 5921-034-235 4140111 11/25/2015 Nov-17 5921-034-235 4140468 11/25/2015 Nov-17 5921-034-235 4141054 11/25/2015 Nov-17 5921-034-235 4143227 11/25/2015 Nov-17 5921-034-235 4144490 11/25/2015 Nov-17 5921-034-235 4145058 11/25/2015 Nov-17 5921-218-235 4139649 11/25/2015 Nov-17 5921-218-235 4140861 11/25/2015 Nov-17 5921-218-235 4142157 11/25/2015 Nov-17 5921-218-235 4143837 11/25/2015 Nov-17 5921-218-235 4144028 11/25/2015 Nov-17 5921-218-135 4141217 11/25/2015 Nov-17 5921-218-135 4141405 11/25/2015 Nov-17 60-7070-102 4139302 11/25/2015 Nov-17 60-7070-103 4139893 11/25/2015 Nov-17 60-7070-103 4140646 11/25/2015 Nov-17 60-7070-103 4142038 11/25/2015 Nov-17 60-7070-103 4142368 11/25/2015 Nov-17 60-7070-103 4143010 11/25/2015 Nov-17 60-7070-103 4143703 11/25/2015 Nov-17 60-7070-103 4144211 11/25/2015 Nov-17 60-7070-103 4144615 11/25/2015 Nov-17 60-7070-103 4145101 11/25/2015 Nov-17 60-7070-103 4145402 11/25/2015 Nov-17 60-7070-103 4146057 11/25/2015 Nov-17 60-7070-103 4146591 11/25/2015 Nov-17 60-7070-103 4147053 11/25/2015 Nov-17 60-7070-105 4140148 11/25/2015 Nov-17 60-7070-105 4141181 11/25/2015 Nov-17 60-7070-105 4142721 11/25/2015 Nov-17 60-7070-105 4143505 11/25/2015 Nov-17 60-7070-105 4144457 11/25/2015 Nov-17 60-7070-105 4144712 11/25/2015 Nov-17 60-7070-105 4145439 11/25/2015 Nov-17 60-7070-105 4145500 11/25/2015 Nov-17 60-7070-105 4146320 11/25/2015 Nov-17 60-7070-104 4144921 11/25/2015 Nov-17 60-7070-104 4146969 11/25/2015 Nov-17 60-7070-004 4140596 11/25/2015 Nov-17 60-7070-006 4144156 11/25/2015 Nov-17 60-7070-106 4139007 11/25/2015 Nov-17 60-7070-106 4139534 11/25/2015 Nov-17 60-7070-106 4140304 11/25/2015 Nov-17 60-7070-106 4141255 11/25/2015 Nov-17 60-7070-106 4143338 11/25/2015 Nov-17 60-7070-106 4144322 11/25/2015 Nov-17 60-7070-106 4144810 11/25/2015 Nov-17 60-7070-106 4145140 11/25/2015 Nov-17 60-7070-106 4146116 11/25/2015 Nov-17 60-7070-106 4146747 11/25/2015 Nov-17 60-7070-107 4141737 11/25/2015 Nov-17 60-7070-107 4146440 11/25/2015 Nov-17 60-7075-003 4142541 11/25/2015 Nov-17 60-7075-006 4140928 11/25/2015 Nov-17 60-7075-105 4141904 11/25/2015 Nov-17 60-7075-105 4145138 11/25/2015 Nov-17 60-7075-105 4146438 11/25/2015 Nov-17 60-7075-106 4141123 11/25/2015 Nov-17 60-7075-106 4145397 11/25/2015 Nov-17 5921-018-135 4155398 11/30/2015 Nov-17 5921-018-135 4159372 11/30/2015 Nov-17 5921-018-235 4153791 11/30/2015 Nov-17 5921-024-135 4154830 11/30/2015 Nov-17 5921-024-135 4157077 11/30/2015 Nov-17 5921-024-235 4159184 11/30/2015 Nov-17 5921-030-135 4156350 11/30/2015 Nov-17 5921-030-135 4160017 11/30/2015 Nov-17 5921-030-235 4153961 11/30/2015 Nov-17 5921-030-235 4155875 11/30/2015 Nov-17 5921-030-235 4156058 11/30/2015 Nov-17 5921-030-235 4157678 11/30/2015 Nov-17 5921-030-235 4159483 11/30/2015 Nov-17 5921-034-135 4154415 11/30/2015 Nov-17 5921-034-135 4158734 11/30/2015 Nov-17 5921-034-135 4159598 11/30/2015 Nov-17 5921-034-235 4153302 11/30/2015 Nov-17 5921-034-235 4153893 11/30/2015 Nov-17 5921-034-235 4155490 11/30/2015 Nov-17 5921-034-235 4157201 11/30/2015 Nov-17 5921-034-235 4158222 11/30/2015 Nov-17 5921-034-235 4159034 11/30/2015 Nov-17 5921-034-235 4160807 11/30/2015 Nov-17 5921-034-236 4153736 11/30/2015 Nov-17 5921-218-235 4153547 11/30/2015 Nov-17 5921-218-235 4156795 11/30/2015 Nov-17 5921-218-235 4158088 11/30/2015 Nov-17 5921-218-235 4159774 11/30/2015 Nov-17 5921-218-235 4159973 11/30/2015 Nov-17 5921-218-135 4154876 11/30/2015 Nov-17 5921-218-135 4155921 11/30/2015 Nov-17 5921-218-135 4156004 11/30/2015 Nov-17 5921-218-135 4158930 11/30/2015 Nov-17 60-7070-102 4154254 11/30/2015 Nov-17 60-7070-103 4154208 11/30/2015 Nov-17 60-7070-103 4154212 11/30/2015 Nov-17 60-7070-103 4154375 11/30/2015 Nov-17 60-7070-103 4154443 11/30/2015 Nov-17 60-7070-103 4154489 11/30/2015 Nov-17 60-7070-103 4154502 11/30/2015 Nov-17 60-7070-103 4154628 11/30/2015 Nov-17 60-7070-103 4154709 11/30/2015 Nov-17 60-7070-103 4154750 11/30/2015 Nov-17 60-7070-103 4155305 11/30/2015 Nov-17 60-7070-103 4156223 11/30/2015 Nov-17 60-7070-103 4156912 11/30/2015 Nov-17 60-7070-103 4158440 11/30/2015 Nov-17 60-7070-103 4158752 11/30/2015 Nov-17 60-7070-103 4160093 11/30/2015 Nov-17 60-7070-105 4153599 11/30/2015 Nov-17 60-7070-105 4157451 11/30/2015 Nov-17 60-7070-105 4160462 11/30/2015 Nov-17 60-7070-104 4154040 11/30/2015 Nov-17 60-7070-104 4156711 11/30/2015 Nov-17 60-7070-104 4158618 11/30/2015 Nov-17 60-7070-003 4158850 11/30/2015 Nov-17 60-7070-006 4153665 11/30/2015 Nov-17 60-7070-106 4153413 11/30/2015 Nov-17 60-7070-106 4154188 11/30/2015 Nov-17 60-7070-106 4156571 11/30/2015 Nov-17 60-7070-106 4157958 11/30/2015 Nov-17 60-7070-106 4158907 11/30/2015 Nov-17 60-7070-106 4160876 11/30/2015 Nov-17 60-7070-107 4154126 11/30/2015 Nov-17 60-7075-103 4155712 11/30/2015 Nov-17 60-7075-103 4159099 11/30/2015 Nov-17 60-7075-104 4154541 11/30/2015 Nov-17 60-7075-105 4158821 11/30/2015 Nov-17 60-7075-106 4154797 11/30/2015 Nov-17 60-7075-106 4156730 11/30/2015 Nov-17 5921-024-135 4163089 12/2/2015 Dec-17 5921-030-235 4162434 12/2/2015 Dec-17 5921-030-235 4164652 12/2/2015 Dec-17 5921-030-235 4166069 12/2/2015 Dec-17 5921-034-135 4163247 12/2/2015 Dec-17 5921-034-235 4161883 12/2/2015 Dec-17 5921-034-235 4164067 12/2/2015 Dec-17 5921-034-235 4165598 12/2/2015 Dec-17 5921-034-235 4166302 12/2/2015 Dec-17 5921-034-235 4168008 12/2/2015 Dec-17 5921-218-235 4158562 12/2/2015 Dec-17 5921-218-235 4161136 12/2/2015 Dec-17 5921-218-235 4162107 12/2/2015 Dec-17 5921-218-235 4163091 12/2/2015 Dec-17 5921-218-235 4163626 12/2/2015 Dec-17 5921-218-235 4165859 12/2/2015 Dec-17 5921-218-135 4162985 12/2/2015 Dec-17 60-7070-102 4164858 12/2/2015 Dec-17 60-7070-103 4161324 12/2/2015 Dec-17 60-7070-103 4161590 12/2/2015 Dec-17 60-7070-103 4162013 12/2/2015 Dec-17 60-7070-103 4163138 12/2/2015 Dec-17 60-7070-103 4164993 12/2/2015 Dec-17 60-7070-103 4165973 12/2/2015 Dec-17 60-7070-105 4162256 12/2/2015 Dec-17 60-7070-105 4162861 12/2/2015 Dec-17 60-7070-105 4164550 12/2/2015 Dec-17 60-7070-105 4166356 12/2/2015 Dec-17 60-7070-105 4168240 12/2/2015 Dec-17 60-7070-104 4162764 12/2/2015 Dec-17 60-7070-003 4162583 12/2/2015 Dec-17 60-7070-006 4163378 12/2/2015 Dec-17 60-7070-106 4161039 12/2/2015 Dec-17 60-7070-106 4161820 12/2/2015 Dec-17 60-7070-106 4163043 12/2/2015 Dec-17 60-7070-106 4164756 12/2/2015 Dec-17 60-7070-106 4165270 12/2/2015 Dec-17 60-7070-106 4165338 12/2/2015 Dec-17 60-7070-106 4165790 12/2/2015 Dec-17 60-7070-106 4167641 12/2/2015 Dec-17 60-7070-107 4166409 12/2/2015 Dec-17 60-7075-103 4162667 12/2/2015 Dec-17 60-7075-104 4163730 12/2/2015 Dec-17 60-7075-105 4163983 12/2/2015 Dec-17 5921-030-135 4070525 12/3/2015 Dec-17 5921-018-235 4166988 12/4/2015 Dec-17 5921-024-235 4167137 12/4/2015 Dec-17 5921-024-235 4170027 12/4/2015 Dec-17 5921-024-235 4174582 12/4/2015 Dec-17 5921-030-135 4166466 12/4/2015 Dec-17 5921-030-135 4167911 12/4/2015 Dec-17 5921-030-135 4170954 12/4/2015 Dec-17 5921-030-135 4175032 12/4/2015 Dec-17 5921-030-235 4168361 12/4/2015 Dec-17 5921-030-235 4169847 12/4/2015 Dec-17 5921-030-235 4172387 12/4/2015 Dec-17 5921-030-235 4173428 12/4/2015 Dec-17 5921-034-135 4166838 12/4/2015 Dec-17 5921-034-135 4167026 12/4/2015 Dec-17 5921-034-135 4172105 12/4/2015 Dec-17 5921-034-135 4174065 12/4/2015 Dec-17 5921-034-135 4174700 12/4/2015 Dec-17 5921-034-235 4169000 12/4/2015 Dec-17 5921-034-235 4169932 12/4/2015 Dec-17 5921-034-235 4171708 12/4/2015 Dec-17 5921-034-235 4172893 12/4/2015 Dec-17 5921-034-235 4174936 12/4/2015 Dec-17 5921-218-235 4166599 12/4/2015 Dec-17 5921-218-235 4167035 12/4/2015 Dec-17 5921-218-235 4170223 12/4/2015 Dec-17 5921-218-235 4171367 12/4/2015 Dec-17 5921-218-235 4172048 12/4/2015 Dec-17 5921-218-235 4173104 12/4/2015 Dec-17 5921-218-235 4173906 12/4/2015 Dec-17 5921-218-235 4174303 12/4/2015 Dec-17 5921-218-135 4170418 12/4/2015 Dec-17 5921-218-135 4170688 12/4/2015 Dec-17 5921-218-135 4171689 12/4/2015 Dec-17 5921-218-135 4172360 12/4/2015 Dec-17 60-7070-102 4174451 12/4/2015 Dec-17 60-7070-103 4166817 12/4/2015 Dec-17 60-7070-103 4168548 12/4/2015 Dec-17 60-7070-103 4169187 12/4/2015 Dec-17 60-7070-103 4171593 12/4/2015 Dec-17 60-7070-103 4172167 12/4/2015 Dec-17 60-7070-103 4173039 12/4/2015 Dec-17 60-7070-103 4174744 12/4/2015 Dec-17 60-7070-105 4167455 12/4/2015 Dec-17 60-7070-105 4168319 12/4/2015 Dec-17 60-7070-105 4169627 12/4/2015 Dec-17 60-7070-105 4170397 12/4/2015 Dec-17 60-7070-105 4171092 12/4/2015 Dec-17 60-7070-105 4173002 12/4/2015 Dec-17 60-7070-105 4173503 12/4/2015 Dec-17 60-7070-105 4174238 12/4/2015 Dec-17 60-7070-105 4175009 12/4/2015 Dec-17 60-7070-104 4166704 12/4/2015 Dec-17 60-7070-104 4169488 12/4/2015 Dec-17 60-7070-104 4172838 12/4/2015 Dec-17 60-7070-104 4174335 12/4/2015 Dec-17 60-7070-003 4174632 12/4/2015 Dec-17 60-7070-005 4173191 12/4/2015 Dec-17 60-7070-006 4170769 12/4/2015 Dec-17 60-7070-006 4174553 12/4/2015 Dec-17 60-7070-106 4169689 12/4/2015 Dec-17 60-7070-106 4170345 12/4/2015 Dec-17 60-7070-106 4171765 12/4/2015 Dec-17 60-7070-106 4172715 12/4/2015 Dec-17 60-7070-106 4172762 12/4/2015 Dec-17 60-7070-106 4174134 12/4/2015 Dec-17 60-7070-106 4175021 12/4/2015 Dec-17 60-7070-107 4170635 12/4/2015 Dec-17 60-7070-107 4173542 12/4/2015 Dec-17 60-7075-006 4170931 12/4/2015 Dec-17 60-7075-006 4174887 12/4/2015 Dec-17 60-7075-104 4171823 12/4/2015 Dec-17 60-7075-105 4171380 12/4/2015 Dec-17 60-7075-106 4167377 12/4/2015 Dec-17 60-7075-106 4173780 12/4/2015 Dec-17 5921-018-135 4178546 12/7/2015 Dec-17 5921-018-235 4177848 12/7/2015 Dec-17 5921-018-235 4179543 12/7/2015 Dec-17 5921-024-136 4177470 12/7/2015 Dec-17 5921-024-235 4179948 12/7/2015 Dec-17 5921-030-135 4175793 12/7/2015 Dec-17 5921-030-135 4180435 12/7/2015 Dec-17 5921-030-136 4176079 12/7/2015 Dec-17 5921-030-235 4175182 12/7/2015 Dec-17 5921-030-235 4176219 12/7/2015 Dec-17 5921-030-235 4176257 12/7/2015 Dec-17 5921-030-235 4177980 12/7/2015 Dec-17 5921-030-235 4179907 12/7/2015 Dec-17 5921-030-235 4180636 12/7/2015 Dec-17 5921-030-235 4180950 12/7/2015 Dec-17 5921-034-135 4175148 12/7/2015 Dec-17 5921-034-135 4176964 12/7/2015 Dec-17 5921-034-135 4177905 12/7/2015 Dec-17 5921-034-135 4180253 12/7/2015 Dec-17 5921-034-135 4180670 12/7/2015 Dec-17 5921-034-235 4175362 12/7/2015 Dec-17 5921-034-235 4177600 12/7/2015 Dec-17 5921-034-235 4178533 12/7/2015 Dec-17 5921-034-235 4179128 12/7/2015 Dec-17 5921-034-235 4180461 12/7/2015 Dec-17 5921-034-235 4180875 12/7/2015 Dec-17 5921-218-235 4175107 12/7/2015 Dec-17 5921-218-235 4178166 12/7/2015 Dec-17 5921-218-235 4178973 12/7/2015 Dec-17 5921-218-135 4175358 12/7/2015 Dec-17 5921-218-135 4176647 12/7/2015 Dec-17 5921-218-135 4176741 12/7/2015 Dec-17 5921-218-136 4176606 12/7/2015 Dec-17 60-7070-103 4175124 12/7/2015 Dec-17 60-7070-103 4175738 12/7/2015 Dec-17 60-7070-103 4177502 12/7/2015 Dec-17 60-7070-103 4177573 12/7/2015 Dec-17 60-7070-103 4178262 12/7/2015 Dec-17 60-7070-103 4178941 12/7/2015 Dec-17 60-7070-103 4179455 12/7/2015 Dec-17 60-7070-103 4181443 12/7/2015 Dec-17 60-7070-105 4176371 12/7/2015 Dec-17 60-7070-105 4177038 12/7/2015 Dec-17 60-7070-105 4177126 12/7/2015 Dec-17 60-7070-105 4178431 12/7/2015 Dec-17 60-7070-105 4179222 12/7/2015 Dec-17 60-7070-105 4180689 12/7/2015 Dec-17 60-7070-105 4181212 12/7/2015 Dec-17 60-7070-104 4176566 12/7/2015 Dec-17 60-7070-003 4179639 12/7/2015 Dec-17 60-7070-006 4177745 12/7/2015 Dec-17 60-7070-106 4175549 12/7/2015 Dec-17 60-7070-106 4176815 12/7/2015 Dec-17 60-7070-106 4176940 12/7/2015 Dec-17 60-7070-106 4178863 12/7/2015 Dec-17 60-7070-106 4179331 12/7/2015 Dec-17 60-7070-106 4180173 12/7/2015 Dec-17 60-7070-107 4175577 12/7/2015 Dec-17 60-7070-107 4176513 12/7/2015 Dec-17 60-7070-107 4180783 12/7/2015 Dec-17 60-7075-003 4086342 12/7/2015 Dec-17 60-7075-103 4177587 12/7/2015 Dec-17 60-7075-104 4178420 12/7/2015 Dec-17 60-7075-105 4178236 12/7/2015 Dec-17 60-7075-106 4175985 12/7/2015 Dec-17 60-7075-106 4177190 12/7/2015 Dec-17 60-7075-106 4177220 12/7/2015 Dec-17 60-7075-106 4181017 12/7/2015 Dec-17 5921-018-135 4180586 12/9/2015 Dec-17 5921-024-135 4184605 12/9/2015 Dec-17 5921-024-135 4186145 12/9/2015 Dec-17 5921-024-135 4186227 12/9/2015 Dec-17 5921-024-235 4181072 12/9/2015 Dec-17 5921-030-135 4185177 12/9/2015 Dec-17 5921-030-135 4186465 12/9/2015 Dec-17 5921-030-235 4182763 12/9/2015 Dec-17 5921-034-135 4181208 12/9/2015 Dec-17 5921-034-135 4181721 12/9/2015 Dec-17 5921-034-135 4184452 12/9/2015 Dec-17 5921-034-135 4184793 12/9/2015 Dec-17 5921-034-135 4186885 12/9/2015 Dec-17 5921-034-235 4183440 12/9/2015 Dec-17 5921-034-235 4184239 12/9/2015 Dec-17 5921-034-235 4187170 12/9/2015 Dec-17 5921-034-235 4187278 12/9/2015 Dec-17 5921-034-235 4187952 12/9/2015 Dec-17 5921-218-235 4180071 12/9/2015 Dec-17 5921-218-235 4180403 12/9/2015 Dec-17 5921-218-235 4182361 12/9/2015 Dec-17 5921-218-235 4183813 12/9/2015 Dec-17 5921-218-235 4187428 12/9/2015 Dec-17 5921-218-135 4181824 12/9/2015 Dec-17 5921-218-135 4185509 12/9/2015 Dec-17 5921-218-135 4186703 12/9/2015 Dec-17 60-7070-103 4180248 12/9/2015 Dec-17 60-7070-103 4182595 12/9/2015 Dec-17 60-7070-103 4183710 12/9/2015 Dec-17 60-7070-103 4183850 12/9/2015 Dec-17 60-7070-103 4184406 12/9/2015 Dec-17 60-7070-103 4184900 12/9/2015 Dec-17 60-7070-103 4186579 12/9/2015 Dec-17 60-7070-103 4187672 12/9/2015 Dec-17 60-7070-105 4182175 12/9/2015 Dec-17 60-7070-105 4182856 12/9/2015 Dec-17 60-7070-105 4184066 12/9/2015 Dec-17 60-7070-105 4184772 12/9/2015 Dec-17 60-7070-105 4186682 12/9/2015 Dec-17 60-7070-105 4187910 12/9/2015 Dec-17 60-7070-104 4184336 12/9/2015 Dec-17 60-7070-104 4186871 12/9/2015 Dec-17 60-7070-104 4188011 12/9/2015 Dec-17 60-7070-005 4184633 12/9/2015 Dec-17 60-7070-006 4182201 12/9/2015 Dec-17 60-7070-106 4181876 12/9/2015 Dec-17 60-7070-106 4183515 12/9/2015 Dec-17 60-7070-106 4184434 12/9/2015 Dec-17 60-7070-106 4187002 12/9/2015 Dec-17 60-7070-106 4187601 12/9/2015 Dec-17 60-7070-106 4187780 12/9/2015 Dec-17 60-7070-107 4184182 12/9/2015 Dec-17 60-7075-006 4186837 12/9/2015 Dec-17 60-7075-103 4180806 12/9/2015 Dec-17 60-7075-103 4184667 12/9/2015 Dec-17 60-7075-103 4184729 12/9/2015 Dec-17 60-7075-103 4185794 12/9/2015 Dec-17 60-7075-104 4186301 12/9/2015 Dec-17 60-7075-105 4186412 12/9/2015 Dec-17 60-7075-106 4184992 12/9/2015 Dec-17 60-7075-106 4185938 12/9/2015 Dec-17 60-7075-106 4186648 12/9/2015 Dec-17 5921-018-135 4191179 12/11/2015 Dec-17 5921-018-135 4194441 12/11/2015 Dec-17 5921-024-135 4194155 12/11/2015 Dec-17 5921-024-235 4189851 12/11/2015 Dec-17 5921-024-235 4192085 12/11/2015 Dec-17 5921-030-135 4190467 12/11/2015 Dec-17 5921-030-135 4190962 12/11/2015 Dec-17 5921-030-135 4194965 12/11/2015 Dec-17 5921-030-235 4188360 12/11/2015 Dec-17 5921-030-235 4191200 12/11/2015 Dec-17 5921-030-235 4191935 12/11/2015 Dec-17 5921-030-235 4192156 12/11/2015 Dec-17 5921-030-235 4194418 12/11/2015 Dec-17 5921-034-135 4188736 12/11/2015 Dec-17 5921-034-135 4189908 12/11/2015 Dec-17 5921-034-135 4191942 12/11/2015 Dec-17 5921-034-135 4192586 12/11/2015 Dec-17 5921-034-235 4190240 12/11/2015 Dec-17 5921-034-235 4190405 12/11/2015 Dec-17 5921-034-235 4191612 12/11/2015 Dec-17 5921-034-235 4193469 12/11/2015 Dec-17 5921-034-235 4194321 12/11/2015 Dec-17 5921-218-235 4189626 12/11/2015 Dec-17 5921-218-235 4193741 12/11/2015 Dec-17 5921-218-135 4190175 12/11/2015 Dec-17 5921-218-135 4194707 12/11/2015 Dec-17 5921-218-136 4188179 12/11/2015 Dec-17 60-7070-103 4188259 12/11/2015 Dec-17 60-7070-103 4188801 12/11/2015 Dec-17 60-7070-103 4190168 12/11/2015 Dec-17 60-7070-103 4191360 12/11/2015 Dec-17 60-7070-103 4191886 12/11/2015 Dec-17 60-7070-103 4192881 12/11/2015 Dec-17 60-7070-103 4194070 12/11/2015 Dec-17 60-7070-103 4194954 12/11/2015 Dec-17 60-7070-105 4189058 12/11/2015 Dec-17 60-7070-105 4191450 12/11/2015 Dec-17 60-7070-105 4193177 12/11/2015 Dec-17 60-7070-105 4194748 12/11/2015 Dec-17 60-7070-104 4191727 12/11/2015 Dec-17 60-7070-104 4193300 12/11/2015 Dec-17 60-7070-003 4189473 12/11/2015 Dec-17 60-7070-006 4189309 12/11/2015 Dec-17 60-7070-006 4189920 12/11/2015 Dec-17 60-7070-106 4188631 12/11/2015 Dec-17 60-7070-106 4190700 12/11/2015 Dec-17 60-7070-106 4191237 12/11/2015 Dec-17 60-7070-106 4191547 12/11/2015 Dec-17 60-7070-106 4192806 12/11/2015 Dec-17 60-7070-106 4193836 12/11/2015 Dec-17 60-7070-106 4194619 12/11/2015 Dec-17 60-7070-106 4194805 12/11/2015 Dec-17 60-7070-107 4192474 12/11/2015 Dec-17 60-7075-103 4188547 12/11/2015 Dec-17 60-7075-103 4190887 12/11/2015 Dec-17 60-7075-103 4191534 12/11/2015 Dec-17 60-7075-103 4192125 12/11/2015 Dec-17 60-7075-103 4192457 12/11/2015 Dec-17 60-7075-103 4194782 12/11/2015 Dec-17 60-7075-106 4189769 12/11/2015 Dec-17 60-7075-106 4190255 12/11/2015 Dec-17 60-7075-106 4191743 12/11/2015 Dec-17 5921-018-135 4202004 12/14/2015 Dec-17 5921-024-135 4195969 12/14/2015 Dec-17 5921-024-135 4199176 12/14/2015 Dec-17 5921-030-135 4195401 12/14/2015 Dec-17 5921-030-135 4199239 12/14/2015 Dec-17 5921-030-235 4148865 12/14/2015 Dec-17 5921-030-235 4196221 12/14/2015 Dec-17 5921-030-235 4196751 12/14/2015 Dec-17 5921-030-235 4201393 12/14/2015 Dec-17 5921-030-235 4202015 12/14/2015 Dec-17 5921-034-135 4195246 12/14/2015 Dec-17 5921-034-135 4195687 12/14/2015 Dec-17 5921-034-135 4195799 12/14/2015 Dec-17 5921-034-135 4198443 12/14/2015 Dec-17 5921-034-135 4199013 12/14/2015 Dec-17 5921-034-135 4199900 12/14/2015 Dec-17 5921-034-235 4195948 12/14/2015 Dec-17 5921-034-235 4196999 12/14/2015 Dec-17 5921-034-235 4197892 12/14/2015 Dec-17 5921-034-235 4199343 12/14/2015 Dec-17 5921-034-235 4200105 12/14/2015 Dec-17 5921-034-235 4201243 12/14/2015 Dec-17 5921-218-235 4195205 12/14/2015 Dec-17 5921-218-235 4195821 12/14/2015 Dec-17 5921-218-235 4197338 12/14/2015 Dec-17 5921-218-235 4197505 12/14/2015 Dec-17 5921-218-235 4198126 12/14/2015 Dec-17 5921-218-235 4199577 12/14/2015 Dec-17 5921-218-235 4201109 12/14/2015 Dec-17 5921-218-235 4201567 12/14/2015 Dec-17 5921-218-135 4191654 12/14/2015 Dec-17 5921-218-135 4196002 12/14/2015 Dec-17 5921-218-135 4196503 12/14/2015 Dec-17 5921-218-135 4197047 12/14/2015 Dec-17 5921-218-135 4200133 12/14/2015 Dec-17 5921-218-135 4202070 12/14/2015 Dec-17 60-7070-102 4197648 12/14/2015 Dec-17 60-7070-103 4196617 12/14/2015 Dec-17 60-7070-103 4198055 12/14/2015 Dec-17 60-7070-103 4198462 12/14/2015 Dec-17 60-7070-103 4199992 12/14/2015 Dec-17 60-7070-103 4200631 12/14/2015 Dec-17 60-7070-105 4196849 12/14/2015 Dec-17 60-7070-105 4198277 12/14/2015 Dec-17 60-7070-105 4200373 12/14/2015 Dec-17 60-7070-105 4201719 12/14/2015 Dec-17 60-7070-105 4202101 12/14/2015 Dec-17 60-7070-104 4195558 12/14/2015 Dec-17 60-7070-104 4199717 12/14/2015 Dec-17 60-7070-104 4200961 12/14/2015 Dec-17 60-7070-005 4196492 12/14/2015 Dec-17 60-7070-106 4197828 12/14/2015 Dec-17 60-7070-106 4198368 12/14/2015 Dec-17 60-7070-106 4199803 12/14/2015 Dec-17 60-7070-106 4200076 12/14/2015 Dec-17 60-7070-106 4200506 12/14/2015 Dec-17 60-7070-106 4200886 12/14/2015 Dec-17 60-7070-106 4200942 12/14/2015 Dec-17 60-7070-106 4202058 12/14/2015 Dec-17 60-7070-107 4197178 12/14/2015 Dec-17 60-7075-003 4198352 12/14/2015 Dec-17 60-7075-103 4196025 12/14/2015 Dec-17 60-7075-104 4196780 12/14/2015 Dec-17 60-7075-105 4195924 12/14/2015 Dec-17 60-7075-105 4199826 12/14/2015 Dec-17 60-7075-105 4200095 12/14/2015 Dec-17 60-7075-106 4196245 12/14/2015 Dec-17 60-7075-106 4198876 12/14/2015 Dec-17 60-7075-106 4199410 12/14/2015 Dec-17 60-7075-106 4199793 12/14/2015 Dec-17 60-7075-106 4201338 12/14/2015 Dec-17 5921-018-235 4203256 12/16/2015 Dec-17 5921-030-135 4203195 12/16/2015 Dec-17 5921-030-135 4205307 12/16/2015 Dec-17 5921-030-135 4206306 12/16/2015 Dec-17 5921-030-235 4207014 12/16/2015 Dec-17 5921-034-135 4201546 12/16/2015 Dec-17 5921-034-135 4202563 12/16/2015 Dec-17 5921-034-135 4205018 12/16/2015 Dec-17 5921-034-135 4206055 12/16/2015 Dec-17 5921-034-135 4206525 12/16/2015 Dec-17 5921-034-235 4202954 12/16/2015 Dec-17 5921-034-235 4204641 12/16/2015 Dec-17 5921-034-235 4206265 12/16/2015 Dec-17 5921-034-235 4207196 12/16/2015 Dec-17 5921-218-235 4204527 12/16/2015 Dec-17 5921-218-235 4205060 12/16/2015 Dec-17 5921-218-135 4202879 12/16/2015 Dec-17 5921-218-135 4203148 12/16/2015 Dec-17 5921-218-135 4203614 12/16/2015 Dec-17 5921-218-135 4206249 12/16/2015 Dec-17 5921-218-136 4203729 12/16/2015 Dec-17 60-7070-102 4205588 12/16/2015 Dec-17 60-7070-103 4203304 12/16/2015 Dec-17 60-7070-103 4204074 12/16/2015 Dec-17 60-7070-103 4205324 12/16/2015 Dec-17 60-7070-103 4205629 12/16/2015 Dec-17 60-7070-103 4206135 12/16/2015 Dec-17 60-7070-103 4206352 12/16/2015 Dec-17 60-7070-103 4206491 12/16/2015 Dec-17 60-7070-105 4203646 12/16/2015 Dec-17 60-7070-105 4204775 12/16/2015 Dec-17 60-7070-105 4205672 12/16/2015 Dec-17 60-7070-104 4204602 12/16/2015 Dec-17 60-7070-003 4205395 12/16/2015 Dec-17 60-7070-106 4203549 12/16/2015 Dec-17 60-7070-106 4203698 12/16/2015 Dec-17 60-7070-106 4204832 12/16/2015 Dec-17 60-7070-106 4205463 12/16/2015 Dec-17 60-7070-107 4203874 12/16/2015 Dec-17 60-7075-003 4202657 12/16/2015 Dec-17 60-7075-003 4202920 12/16/2015 Dec-17 60-7075-006 4202378 12/16/2015 Dec-17 60-7075-006 4206663 12/16/2015 Dec-17 60-7075-103 4205459 12/16/2015 Dec-17 60-7075-103 4206807 12/16/2015 Dec-17 60-7075-105 4203315 12/16/2015 Dec-17 60-7075-106 4204003 12/16/2015 Dec-17 60-7075-106 4205617 12/16/2015 Dec-17 5921-018-135 4213525 12/18/2015 Dec-17 5921-018-235 4210884 12/18/2015 Dec-17 5921-018-235 4215303 12/18/2015 Dec-17 5921-024-135 4212788 12/18/2015 Dec-17 5921-024-235 4211091 12/18/2015 Dec-17 5921-024-235 4213539 12/18/2015 Dec-17 5921-030-135 4213017 12/18/2015 Dec-17 5921-030-135 4215264 12/18/2015 Dec-17 5921-030-235 4207261 12/18/2015 Dec-17 5921-030-235 4210354 12/18/2015 Dec-17 5921-030-235 4213502 12/18/2015 Dec-17 5921-034-135 4212150 12/18/2015 Dec-17 5921-034-135 4212301 12/18/2015 Dec-17 5921-034-135 4215480 12/18/2015 Dec-17 5921-034-135 4216084 12/18/2015 Dec-17 5921-034-235 4208966 12/18/2015 Dec-17 5921-034-235 4211004 12/18/2015 Dec-17 5921-034-235 4212514 12/18/2015 Dec-17 5921-034-235 4214649 12/18/2015 Dec-17 5921-034-235 4215240 12/18/2015 Dec-17 5921-218-235 4209825 12/18/2015 Dec-17 5921-218-235 4211313 12/18/2015 Dec-17 5921-218-235 4212837 12/18/2015 Dec-17 5921-218-235 4214510 12/18/2015 Dec-17 5921-218-235 4215569 12/18/2015 Dec-17 5921-212-135 4211585 12/18/2015 Dec-17 5921-218-135 4211089 12/18/2015 Dec-17 5921-218-135 4213415 12/18/2015 Dec-17 60-7070-103 4210088 12/18/2015 Dec-17 60-7070-103 4210781 12/18/2015 Dec-17 60-7070-103 4211563 12/18/2015 Dec-17 60-7070-103 4212291 12/18/2015 Dec-17 60-7070-103 4213967 12/18/2015 Dec-17 60-7070-103 4214829 12/18/2015 Dec-17 60-7070-103 4215498 12/18/2015 Dec-17 60-7070-105 4207453 12/18/2015 Dec-17 60-7070-105 4209054 12/18/2015 Dec-17 60-7070-105 4209744 12/18/2015 Dec-17 60-7070-105 4210536 12/18/2015 Dec-17 60-7070-105 4212145 12/18/2015 Dec-17 60-7070-105 4212412 12/18/2015 Dec-17 60-7070-105 4213797 12/18/2015 Dec-17 60-7070-105 4215168 12/18/2015 Dec-17 60-7070-105 4216260 12/18/2015 Dec-17 60-7070-104 4207448 12/18/2015 Dec-17 60-7070-104 4211690 12/18/2015 Dec-17 60-7070-104 4213436 12/18/2015 Dec-17 60-7070-104 4214758 12/18/2015 Dec-17 60-7070-003 4212052 12/18/2015 Dec-17 60-7070-106 4209272 12/18/2015 Dec-17 60-7070-106 4209335 12/18/2015 Dec-17 60-7070-106 4210205 12/18/2015 Dec-17 60-7070-106 4210641 12/18/2015 Dec-17 60-7070-106 4211268 12/18/2015 Dec-17 60-7070-106 4212258 12/18/2015 Dec-17 60-7070-106 4213347 12/18/2015 Dec-17 60-7070-106 4214961 12/18/2015 Dec-17 60-7070-106 4216013 12/18/2015 Dec-17 60-7075-004 4212522 12/18/2015 Dec-17 60-7075-006 4207257 12/18/2015 Dec-17 60-7080-102 4212774 12/18/2015 Dec-17 60-7075-103 4207149 12/18/2015 Dec-17 60-7075-103 4207385 12/18/2015 Dec-17 60-7075-103 4207469 12/18/2015 Dec-17 60-7075-103 4213161 12/18/2015 Dec-17 60-7075-104 4213882 12/18/2015 Dec-17 60-7075-105 4211281 12/18/2015 Dec-17 60-7075-105 4213249 12/18/2015 Dec-17 60-7075-106 4207411 12/18/2015 Dec-17 60-7075-106 4210982 12/18/2015 Dec-17 60-7075-106 4211688 12/18/2015 Dec-17 5921-018-135 4218014 12/21/2015 Dec-17 5921-018-135 4220839 12/21/2015 Dec-17 5921-024-135 4217305 12/21/2015 Dec-17 5921-024-235 4217517 12/21/2015 Dec-17 5921-024-235 4222092 12/21/2015 Dec-17 5921-030-135 4216389 12/21/2015 Dec-17 5921-030-135 4220927 12/21/2015 Dec-17 5921-030-235 4217426 12/21/2015 Dec-17 5921-030-235 4219877 12/21/2015 Dec-17 5921-030-235 4220802 12/21/2015 Dec-17 5921-030-235 4221612 12/21/2015 Dec-17 5921-030-235 4223773 12/21/2015 Dec-17 5921-034-135 4217457 12/21/2015 Dec-17 5921-034-135 4217461 12/21/2015 Dec-17 5921-034-135 4220388 12/21/2015 Dec-17 5921-034-235 4216362 12/21/2015 Dec-17 5921-034-235 4217932 12/21/2015 Dec-17 5921-034-235 4218836 12/21/2015 Dec-17 5921-034-235 4220008 12/21/2015 Dec-17 5921-034-235 4221200 12/21/2015 Dec-17 5921-034-235 4223652 12/21/2015 Dec-17 5921-218-235 4217322 12/21/2015 Dec-17 5921-218-235 4218448 12/21/2015 Dec-17 5921-218-235 4218972 12/21/2015 Dec-17 5921-218-235 4220255 12/21/2015 Dec-17 5921-218-235 4222282 12/21/2015 Dec-17 5921-218-135 4216241 12/21/2015 Dec-17 5921-218-135 4217063 12/21/2015 Dec-17 5921-218-135 4219812 12/21/2015 Dec-17 5921-218-135 4220700 12/21/2015 Dec-17 5921-218-135 4221180 12/21/2015 Dec-17 60-7070-103 4216531 12/21/2015 Dec-17 60-7070-103 4218196 12/21/2015 Dec-17 60-7070-103 4218495 12/21/2015 Dec-17 60-7070-103 4220217 12/21/2015 Dec-17 60-7070-103 4220970 12/21/2015 Dec-17 60-7070-103 4221961 12/21/2015 Dec-17 60-7070-103 4222435 12/21/2015 Dec-17 60-7070-103 4222899 12/21/2015 Dec-17 60-7070-105 4191677 12/21/2015 Dec-17 60-7070-105 4217474 12/21/2015 Dec-17 60-7070-105 4218242 12/21/2015 Dec-17 60-7070-105 4219307 12/21/2015 Dec-17 60-7070-105 4221048 12/21/2015 Dec-17 60-7070-105 4221534 12/21/2015 Dec-17 60-7070-104 4219056 12/21/2015 Dec-17 60-7070-003 4223196 12/21/2015 Dec-17 60-7070-106 4216835 12/21/2015 Dec-17 60-7070-106 4218131 12/21/2015 Dec-17 60-7070-106 4219191 12/21/2015 Dec-17 60-7070-106 4220915 12/21/2015 Dec-17 60-7070-106 4221151 12/21/2015 Dec-17 60-7070-106 4221801 12/21/2015 Dec-17 60-7070-107 4216698 12/21/2015 Dec-17 60-7075-003 4219094 12/21/2015 Dec-17 60-7075-005 4217689 12/21/2015 Dec-17 60-7075-103 4191683 12/21/2015 Dec-17 60-7075-103 4216701 12/21/2015 Dec-17 60-7075-103 4219266 12/21/2015 Dec-17 60-7075-103 4220576 12/21/2015 Dec-17 60-7075-104 4218183 12/21/2015 Dec-17 60-7075-105 4216959 12/21/2015 Dec-17 60-7075-105 4217016 12/21/2015 Dec-17 60-7075-106 4217118 12/21/2015 Dec-17 60-7075-106 4220012 12/21/2015 Dec-17 60-7075-106 4220272 12/21/2015 Dec-17 5921-018-135 4224846 12/23/2015 Dec-17 5921-024-135 4224674 12/23/2015 Dec-17 5921-030-135 4224537 12/23/2015 Dec-17 5921-030-235 4224091 12/23/2015 Dec-17 5921-030-235 4227505 12/23/2015 Dec-17 5921-030-235 4228029 12/23/2015 Dec-17 5921-034-135 4222868 12/23/2015 Dec-17 5921-034-135 4223972 12/23/2015 Dec-17 5921-034-135 4225026 12/23/2015 Dec-17 5921-034-135 4227355 12/23/2015 Dec-17 5921-034-235 4224833 12/23/2015 Dec-17 5921-034-235 4225417 12/23/2015 Dec-17 5921-034-235 4226601 12/23/2015 Dec-17 5921-034-235 4227648 12/23/2015 Dec-17 5921-034-235 4227863 12/23/2015 Dec-17 5921-218-235 4224252 12/23/2015 Dec-17 5921-218-235 4225019 12/23/2015 Dec-17 5921-218-235 4226073 12/23/2015 Dec-17 5921-218-235 4226463 12/23/2015 Dec-17 5921-218-235 4228135 12/23/2015 Dec-17 5921-218-135 4223760 12/23/2015 Dec-17 5921-218-135 4227522 12/23/2015 Dec-17 5921-218-135 4227695 12/23/2015 Dec-17 60-7070-103 4223108 12/23/2015 Dec-17 60-7070-103 4223993 12/23/2015 Dec-17 60-7070-103 4224977 12/23/2015 Dec-17 60-7070-103 4225192 12/23/2015 Dec-17 60-7070-103 4225775 12/23/2015 Dec-17 60-7070-105 4223261 12/23/2015 Dec-17 60-7070-105 4224920 12/23/2015 Dec-17 60-7070-105 4225618 12/23/2015 Dec-17 60-7070-105 4226519 12/23/2015 Dec-17 60-7070-105 4228265 12/23/2015 Dec-17 60-7070-104 4222913 12/23/2015 Dec-17 60-7070-104 4224172 12/23/2015 Dec-17 60-7070-104 4226025 12/23/2015 Dec-17 60-7070-003 4226191 12/23/2015 Dec-17 60-7070-003 4228220 12/23/2015 Dec-17 60-7070-005 4226733 12/23/2015 Dec-17 60-7070-106 4222517 12/23/2015 Dec-17 60-7070-106 4222636 12/23/2015 Dec-17 60-7070-106 4223870 12/23/2015 Dec-17 60-7070-106 4224378 12/23/2015 Dec-17 60-7070-106 4225035 12/23/2015 Dec-17 60-7070-106 4225687 12/23/2015 Dec-17 60-7070-106 4226324 12/23/2015 Dec-17 60-7070-106 4227408 12/23/2015 Dec-17 60-7070-106 4227774 12/23/2015 Dec-17 60-7075-006 4223120 12/23/2015 Dec-17 60-7075-006 4226808 12/23/2015 Dec-17 60-7075-103 4223274 12/23/2015 Dec-17 60-7075-103 4225625 12/23/2015 Dec-17 60-7075-103 4226492 12/23/2015 Dec-17 60-7075-105 4224983 12/23/2015 Dec-17 60-7075-106 4225320 12/23/2015 Dec-17 60-7075-106 4228249 12/23/2015 Dec-17 5921-018-135 4228375 12/28/2015 Dec-17 5921-018-235 4231731 12/28/2015 Dec-17 5921-018-235 4235583 12/28/2015 Dec-17 5921-024-135 4232271 12/28/2015 Dec-17 5921-030-135 4232612 12/28/2015 Dec-17 5921-030-235 4232478 12/28/2015 Dec-17 5921-030-235 4232506 12/28/2015 Dec-17 5921-030-235 4232903 12/28/2015 Dec-17 5921-030-235 4235256 12/28/2015 Dec-17 5921-034-135 4228525 12/28/2015 Dec-17 5921-034-135 4230912 12/28/2015 Dec-17 5921-034-135 4232445 12/28/2015 Dec-17 5921-034-135 4233189 12/28/2015 Dec-17 5921-034-135 4233250 12/28/2015 Dec-17 5921-034-135 4233574 12/28/2015 Dec-17 5921-034-235 4228368 12/28/2015 Dec-17 5921-034-235 4232133 12/28/2015 Dec-17 5921-034-235 4234106 12/28/2015 Dec-17 5921-034-235 4235452 12/28/2015 Dec-17 5921-034-235 4236363 12/28/2015 Dec-17 5921-218-235 4231379 12/28/2015 Dec-17 5921-218-235 4233127 12/28/2015 Dec-17 5921-218-235 4235045 12/28/2015 Dec-17 5921-218-235 4235985 12/28/2015 Dec-17 5921-218-235 4236411 12/28/2015 Dec-17 5921-218-135 4233297 12/28/2015 Dec-17 60-7070-102 4232356 12/28/2015 Dec-17 60-7070-103 4230070 12/28/2015 Dec-17 60-7070-103 4231253 12/28/2015 Dec-17 60-7070-103 4231810 12/28/2015 Dec-17 60-7070-103 4232608 12/28/2015 Dec-17 60-7070-103 4233191 12/28/2015 Dec-17 60-7070-103 4233767 12/28/2015 Dec-17 60-7070-103 4234785 12/28/2015 Dec-17 60-7070-103 4236183 12/28/2015 Dec-17 60-7070-105 4230776 12/28/2015 Dec-17 60-7070-105 4232791 12/28/2015 Dec-17 60-7070-105 4235124 12/28/2015 Dec-17 60-7070-105 4236120 12/28/2015 Dec-17 60-7070-104 4230949 12/28/2015 Dec-17 60-7070-104 4234613 12/28/2015 Dec-17 60-7070-104 4236476 12/28/2015 Dec-17 60-7070-106 4228458 12/28/2015 Dec-17 60-7070-106 4230585 12/28/2015 Dec-17 60-7070-106 4232228 12/28/2015 Dec-17 60-7070-106 4232649 12/28/2015 Dec-17 60-7070-106 4234238 12/28/2015 Dec-17 60-7070-106 4234361 12/28/2015 Dec-17 60-7070-106 4236067 12/28/2015 Dec-17 60-7070-106 4236261 12/28/2015 Dec-17 60-7070-107 4226812 12/28/2015 Dec-17 60-7070-107 4229772 12/28/2015 Dec-17 60-7075-003 4235153 12/28/2015 Dec-17 60-7075-006 4234915 12/28/2015 Dec-17 60-7075-103 4228427 12/28/2015 Dec-17 60-7075-103 4232105 12/28/2015 Dec-17 60-7075-103 4232817 12/28/2015 Dec-17 60-7075-103 4233395 12/28/2015 Dec-17 60-7075-103 4235327 12/28/2015 Dec-17 60-7075-104 4231517 12/28/2015 Dec-17 60-7075-105 4231777 12/28/2015 Dec-17 60-7075-105 4233463 12/28/2015 Dec-17 60-7075-106 4228352 12/28/2015 Dec-17 60-7075-106 4232988 12/28/2015 Dec-17 60-7075-106 4235565 12/28/2015 Dec-17 5921-018-135 4238742 1/4/2016 Jan-18 5921-018-135 4243210 1/4/2016 Jan-18 5921-018-235 4240188 1/4/2016 Jan-18 5921-024-135 4236392 1/4/2016 Jan-18 5921-024-235 4236681 1/4/2016 Jan-18 5921-024-235 4242391 1/4/2016 Jan-18 5921-030-135 4236794 1/4/2016 Jan-18 5921-030-135 4240525 1/4/2016 Jan-18 5921-030-135 4243795 1/4/2016 Jan-18 5921-030-235 4237064 1/4/2016 Jan-18 5921-030-235 4238837 1/4/2016 Jan-18 5921-030-235 4242874 1/4/2016 Jan-18 5921-034-135 4236566 1/4/2016 Jan-18 5921-034-135 4237038 1/4/2016 Jan-18 5921-034-135 4240951 1/4/2016 Jan-18 5921-034-235 4237103 1/4/2016 Jan-18 5921-034-235 4237368 1/4/2016 Jan-18 5921-034-235 4240038 1/4/2016 Jan-18 5921-034-235 4241574 1/4/2016 Jan-18 5921-218-235 4237086 1/4/2016 Jan-18 5921-218-235 4238420 1/4/2016 Jan-18 5921-218-235 4238925 1/4/2016 Jan-18 5921-218-235 4242153 1/4/2016 Jan-18 5921-030-236 4240352 1/4/2016 Jan-18 5921-218-135 4239246 1/4/2016 Jan-18 5921-218-135 4241428 1/4/2016 Jan-18 5921-218-236 4238355 1/4/2016 Jan-18 60-7070-103 4236760 1/4/2016 Jan-18 60-7070-103 4238047 1/4/2016 Jan-18 60-7070-103 4238109 1/4/2016 Jan-18 60-7070-103 4239061 1/4/2016 Jan-18 60-7070-103 4239845 1/4/2016 Jan-18 60-7070-103 4240709 1/4/2016 Jan-18 60-7070-103 4240980 1/4/2016 Jan-18 60-7070-103 4241798 1/4/2016 Jan-18 60-7070-103 4242916 1/4/2016 Jan-18 60-7070-105 4237848 1/4/2016 Jan-18 60-7070-105 4238648 1/4/2016 Jan-18 60-7070-105 4239349 1/4/2016 Jan-18 60-7070-105 4240093 1/4/2016 Jan-18 60-7070-105 4240811 1/4/2016 Jan-18 60-7070-105 4241353 1/4/2016 Jan-18 60-7070-105 4242273 1/4/2016 Jan-18 60-7070-105 4243164 1/4/2016 Jan-18 60-7070-105 4243234 1/4/2016 Jan-18 60-7070-104 4241056 1/4/2016 Jan-18 60-7070-003 4240628 1/4/2016 Jan-18 60-7070-106 4236532 1/4/2016 Jan-18 60-7070-106 4237998 1/4/2016 Jan-18 60-7070-106 4238580 1/4/2016 Jan-18 60-7070-106 4239185 1/4/2016 Jan-18 60-7070-106 4239625 1/4/2016 Jan-18 60-7070-106 4239982 1/4/2016 Jan-18 60-7070-106 4240518 1/4/2016 Jan-18 60-7070-106 4240882 1/4/2016 Jan-18 60-7070-106 4242614 1/4/2016 Jan-18 60-7070-106 4243397 1/4/2016 Jan-18 60-7070-107 4240381 1/4/2016 Jan-18 60-7070-107 4240415 1/4/2016 Jan-18 60-7075-103 4237093 1/4/2016 Jan-18 60-7075-103 4240853 1/4/2016 Jan-18 60-7075-103 4243425 1/4/2016 Jan-18 60-7075-103 4243711 1/4/2016 Jan-18 60-7075-104 4239057 1/4/2016 Jan-18 60-7075-104 4240713 1/4/2016 Jan-18 60-7075-105 4238844 1/4/2016 Jan-18 60-7075-106 4242890 1/4/2016 Jan-18 5921-018-135 4241094 1/5/2016 Jan-18 5921-024-135 4247383 1/5/2016 Jan-18 5921-024-235 4245262 1/5/2016 Jan-18 5921-024-235 4246445 1/5/2016 Jan-18 5921-030-135 4243880 1/5/2016 Jan-18 5921-030-135 4245449 1/5/2016 Jan-18 5921-030-135 4248286 1/5/2016 Jan-18 5921-030-235 4244080 1/5/2016 Jan-18 5921-030-235 4246259 1/5/2016 Jan-18 5921-030-235 4246581 1/5/2016 Jan-18 5921-030-235 4247930 1/5/2016 Jan-18 5921-034-135 4244014 1/5/2016 Jan-18 5921-034-135 4247515 1/5/2016 Jan-18 5921-034-135 4248479 1/5/2016 Jan-18 5921-034-136 4244521 1/5/2016 Jan-18 5921-034-235 4243769 1/5/2016 Jan-18 5921-034-235 4244385 1/5/2016 Jan-18 5921-034-235 4245859 1/5/2016 Jan-18 5921-034-235 4246736 1/5/2016 Jan-18 5921-034-235 4246791 1/5/2016 Jan-18 5921-034-235 4247042 1/5/2016 Jan-18 5921-034-235 4247897 1/5/2016 Jan-18 5921-034-236 4247312 1/5/2016 Jan-18 5921-218-235 4243996 1/5/2016 Jan-18 5921-218-235 4245761 1/5/2016 Jan-18 5921-218-235 4245960 1/5/2016 Jan-18 5921-218-235 4246957 1/5/2016 Jan-18 5921-218-235 4247440 1/5/2016 Jan-18 5921-218-235 4248199 1/5/2016 Jan-18 5921-218-135 4244804 1/5/2016 Jan-18 5921-218-135 4247061 1/5/2016 Jan-18 5921-218-135 4247752 1/5/2016 Jan-18 5921-218-135 4247976 1/5/2016 Jan-18 5921-218-135 4248684 1/5/2016 Jan-18 60-7070-102 4247618 1/5/2016 Jan-18 60-7070-103 4244183 1/5/2016 Jan-18 60-7070-103 4244509 1/5/2016 Jan-18 60-7070-103 4245068 1/5/2016 Jan-18 60-7070-103 4245533 1/5/2016 Jan-18 60-7070-103 4248530 1/5/2016 Jan-18 60-7070-105 4244316 1/5/2016 Jan-18 60-7070-105 4248152 1/5/2016 Jan-18 60-7070-105 4248433 1/5/2016 Jan-18 60-7070-104 4243592 1/5/2016 Jan-18 60-7070-104 4247035 1/5/2016 Jan-18 60-7070-104 4248264 1/5/2016 Jan-18 60-7070-106 4243851 1/5/2016 Jan-18 60-7070-106 4244584 1/5/2016 Jan-18 60-7070-106 4244606 1/5/2016 Jan-18 60-7070-106 4246146 1/5/2016 Jan-18 60-7070-106 4246506 1/5/2016 Jan-18 60-7070-106 4248361 1/5/2016 Jan-18 60-7070-107 4244257 1/5/2016 Jan-18 60-7070-107 4248595 1/5/2016 Jan-18 60-7075-006 4248181 1/5/2016 Jan-18 60-7075-103 4244675 1/5/2016 Jan-18 60-7075-105 4247710 1/5/2016 Jan-18 60-7075-106 4244407 1/5/2016 Jan-18 60-7075-106 4245844 1/5/2016 Jan-18 5921-018-135 4250754 1/7/2016 Jan-18 5921-018-235 4253687 1/7/2016 Jan-18 5921-024-135 4251103 1/7/2016 Jan-18 5921-024-235 4203124 1/7/2016 Jan-18 5921-024-235 4250000 1/7/2016 Jan-18 5921-024-235 4252181 1/7/2016 Jan-18 5921-030-135 4249426 1/7/2016 Jan-18 5921-030-135 4250997 1/7/2016 Jan-18 5921-030-135 4252530 1/7/2016 Jan-18 5921-030-235 4252785 1/7/2016 Jan-18 5921-034-135 4251745 1/7/2016 Jan-18 5921-034-135 4251807 1/7/2016 Jan-18 5921-034-135 4252887 1/7/2016 Jan-18 5921-034-135 4252989 1/7/2016 Jan-18 5921-034-135 4254358 1/7/2016 Jan-18 5921-034-235 4248697 1/7/2016 Jan-18 5921-034-235 4252507 1/7/2016 Jan-18 5921-034-235 4252737 1/7/2016 Jan-18 5921-218-235 4250615 1/7/2016 Jan-18 5921-218-235 4252134 1/7/2016 Jan-18 5921-218-135 4248989 1/7/2016 Jan-18 5921-218-135 4252666 1/7/2016 Jan-18 5921-218-135 4254698 1/7/2016 Jan-18 60-7070-103 4248915 1/7/2016 Jan-18 60-7070-103 4249403 1/7/2016 Jan-18 60-7070-103 4253401 1/7/2016 Jan-18 60-7070-103 4253781 1/7/2016 Jan-18 60-7070-105 4241115 1/7/2016 Jan-18 60-7070-105 4248958 1/7/2016 Jan-18 60-7070-105 4250517 1/7/2016 Jan-18 60-7070-105 4253057 1/7/2016 Jan-18 60-7070-105 4254130 1/7/2016 Jan-18 60-7070-104 4252303 1/7/2016 Jan-18 60-7070-003 4248962 1/7/2016 Jan-18 60-7070-003 4254239 1/7/2016 Jan-18 60-7070-106 4248645 1/7/2016 Jan-18 60-7070-106 4250059 1/7/2016 Jan-18 60-7070-106 4250230 1/7/2016 Jan-18 60-7070-106 4253163 1/7/2016 Jan-18 60-7070-106 4253924 1/7/2016 Jan-18 60-7075-103 4250691 1/7/2016 Jan-18 60-7075-103 4251190 1/7/2016 Jan-18 60-7075-103 4254481 1/7/2016 Jan-18 60-7075-104 4251040 1/7/2016 Jan-18 60-7075-105 4251231 1/7/2016 Jan-18 60-7075-106 4250881 1/7/2016 Jan-18 60-7075-106 4251637 1/7/2016 Jan-18 60-7075-106 4252106 1/7/2016 Jan-18 60-7075-106 4252326 1/7/2016 Jan-18 5921-018-135 4256270 1/8/2016 Jan-18 5921-024-135 4256522 1/8/2016 Jan-18 5921-024-235 4258736 1/8/2016 Jan-18 5921-030-235 4258271 1/8/2016 Jan-18 5921-034-135 4254772 1/8/2016 Jan-18 5921-034-135 4255067 1/8/2016 Jan-18 5921-034-135 4256483 1/8/2016 Jan-18 5921-034-235 4255545 1/8/2016 Jan-18 5921-034-235 4256081 1/8/2016 Jan-18 5921-034-235 4256669 1/8/2016 Jan-18 5921-034-235 4257672 1/8/2016 Jan-18 5921-034-235 4259309 1/8/2016 Jan-18 5921-034-235 4259332 1/8/2016 Jan-18 5921-034-235 4259635 1/8/2016 Jan-18 5921-034-235 4259690 1/8/2016 Jan-18 5921-034-235 4259954 1/8/2016 Jan-18 5921-034-235 4260273 1/8/2016 Jan-18 5921-218-235 4257883 1/8/2016 Jan-18 5921-218-235 4259822 1/8/2016 Jan-18 5921-218-235 4259920 1/8/2016 Jan-18 5921-218-235 4260153 1/8/2016 Jan-18 5921-218-135 4255131 1/8/2016 Jan-18 5921-218-135 4255411 1/8/2016 Jan-18 5921-218-135 4256355 1/8/2016 Jan-18 60-7070-103 4255623 1/8/2016 Jan-18 60-7070-103 4256505 1/8/2016 Jan-18 60-7070-103 4256551 1/8/2016 Jan-18 60-7070-103 4256761 1/8/2016 Jan-18 60-7070-103 4257039 1/8/2016 Jan-18 60-7070-103 4257492 1/8/2016 Jan-18 60-7070-103 4258341 1/8/2016 Jan-18 60-7070-103 4259089 1/8/2016 Jan-18 60-7070-103 4260111 1/8/2016 Jan-18 60-7070-103 4260225 1/8/2016 Jan-18 60-7070-105 4256243 1/8/2016 Jan-18 60-7070-105 4256420 1/8/2016 Jan-18 60-7070-105 4258007 1/8/2016 Jan-18 60-7070-105 4258599 1/8/2016 Jan-18 60-7070-105 4259500 1/8/2016 Jan-18 60-7070-104 4255702 1/8/2016 Jan-18 60-7070-106 4255794 1/8/2016 Jan-18 60-7070-106 4255870 1/8/2016 Jan-18 60-7070-106 4256598 1/8/2016 Jan-18 60-7070-106 4256863 1/8/2016 Jan-18 60-7070-106 4257115 1/8/2016 Jan-18 60-7070-106 4258620 1/8/2016 Jan-18 60-7070-106 4259275 1/8/2016 Jan-18 60-7070-106 4259983 1/8/2016 Jan-18 60-7070-107 4255385 1/8/2016 Jan-18 60-7070-107 4260176 1/8/2016 Jan-18 60-7075-103 4254944 1/8/2016 Jan-18 60-7075-105 4254890 1/8/2016 Jan-18 5921-018-235 4265635 1/11/2016 Jan-18 5921-024-135 4260260 1/11/2016 Jan-18 5921-024-135 4263828 1/11/2016 Jan-18 5921-024-235 4261229 1/11/2016 Jan-18 5921-024-235 4267450 1/11/2016 Jan-18 5921-030-135 4260614 1/11/2016 Jan-18 5921-030-135 4263902 1/11/2016 Jan-18 5921-030-235 4261405 1/11/2016 Jan-18 5921-030-235 4264118 1/11/2016 Jan-18 5921-030-235 4264187 1/11/2016 Jan-18 5921-030-235 4267341 1/11/2016 Jan-18 5921-030-235 4267568 1/11/2016 Jan-18 5921-034-135 4262347 1/11/2016 Jan-18 5921-034-135 4262713 1/11/2016 Jan-18 5921-034-135 4264173 1/11/2016 Jan-18 5921-034-135 4266401 1/11/2016 Jan-18 5921-034-235 4263546 1/11/2016 Jan-18 5921-034-235 4263742 1/11/2016 Jan-18 5921-034-235 4263790 1/11/2016 Jan-18 5921-034-235 4263863 1/11/2016 Jan-18 5921-034-235 4263918 1/11/2016 Jan-18 5921-034-235 4266832 1/11/2016 Jan-18 5921-034-235 4266995 1/11/2016 Jan-18 5921-218-235 4260315 1/11/2016 Jan-18 5921-218-235 4260327 1/11/2016 Jan-18 5921-218-235 4262249 1/11/2016 Jan-18 5921-218-235 4264777 1/11/2016 Jan-18 5921-218-135 4262491 1/11/2016 Jan-18 5921-218-135 4265805 1/11/2016 Jan-18 60-7070-103 4260646 1/11/2016 Jan-18 60-7070-103 4260985 1/11/2016 Jan-18 60-7070-103 4262157 1/11/2016 Jan-18 60-7070-103 4263386 1/11/2016 Jan-18 60-7070-103 4264461 1/11/2016 Jan-18 60-7070-103 4265500 1/11/2016 Jan-18 60-7070-105 4260900 1/11/2016 Jan-18 60-7070-105 4261856 1/11/2016 Jan-18 60-7070-105 4264346 1/11/2016 Jan-18 60-7070-105 4265299 1/11/2016 Jan-18 60-7070-104 4261927 1/11/2016 Jan-18 60-7070-104 4262038 1/11/2016 Jan-18 60-7070-104 4264224 1/11/2016 Jan-18 60-7070-104 4265027 1/11/2016 Jan-18 60-7070-106 4261553 1/11/2016 Jan-18 60-7070-106 4263973 1/11/2016 Jan-18 60-7070-106 4264037 1/11/2016 Jan-18 60-7070-106 4265193 1/11/2016 Jan-18 60-7070-106 4266948 1/11/2016 Jan-18 60-7070-106 4267387 1/11/2016 Jan-18 60-7075-103 4262573 1/11/2016 Jan-18 60-7075-103 4263648 1/11/2016 Jan-18 60-7075-103 4264059 1/11/2016 Jan-18 60-7075-104 4262539 1/11/2016 Jan-18 60-7075-105 4266163 1/11/2016 Jan-18 60-7075-105 4266308 1/11/2016 Jan-18 60-7075-106 4266214 1/11/2016 Jan-18 5921-018-135 4269235 1/14/2016 Jan-18 5921-018-235 4269165 1/14/2016 Jan-18 5921-024-135 4273860 1/14/2016 Jan-18 5921-024-136 4270047 1/14/2016 Jan-18 5921-030-135 4271098 1/14/2016 Jan-18 5921-030-135 4271453 1/14/2016 Jan-18 5921-030-235 4267599 1/14/2016 Jan-18 5921-030-235 4268906 1/14/2016 Jan-18 5921-030-235 4273873 1/14/2016 Jan-18 5921-034-135 4267552 1/14/2016 Jan-18 5921-034-135 4267649 1/14/2016 Jan-18 5921-034-135 4267789 1/14/2016 Jan-18 5921-034-135 4267919 1/14/2016 Jan-18 5921-034-135 4271274 1/14/2016 Jan-18 5921-034-235 4270533 1/14/2016 Jan-18 5921-034-235 4270579 1/14/2016 Jan-18 5921-034-235 4271080 1/14/2016 Jan-18 5921-034-235 4272192 1/14/2016 Jan-18 5921-218-235 4268158 1/14/2016 Jan-18 5921-218-235 4270412 1/14/2016 Jan-18 5921-218-235 4271219 1/14/2016 Jan-18 5921-218-235 4271290 1/14/2016 Jan-18 5921-218-235 4271964 1/14/2016 Jan-18 5921-218-135 4267715 1/14/2016 Jan-18 5921-218-135 4269532 1/14/2016 Jan-18 5921-218-135 4270892 1/14/2016 Jan-18 60-7070-102 4272354 1/14/2016 Jan-18 60-7070-103 4268213 1/14/2016 Jan-18 60-7070-103 4269917 1/14/2016 Jan-18 60-7070-103 4269986 1/14/2016 Jan-18 60-7070-103 4271359 1/14/2016 Jan-18 60-7070-103 4272504 1/14/2016 Jan-18 60-7070-103 4273181 1/14/2016 Jan-18 60-7070-105 4267736 1/14/2016 Jan-18 60-7070-105 4267791 1/14/2016 Jan-18 60-7070-105 4270742 1/14/2016 Jan-18 60-7070-105 4270844 1/14/2016 Jan-18 60-7070-105 4270960 1/14/2016 Jan-18 60-7070-105 4272866 1/14/2016 Jan-18 60-7070-105 4273374 1/14/2016 Jan-18 60-7070-105 4274039 1/14/2016 Jan-18 60-7070-104 4272710 1/14/2016 Jan-18 60-7070-104 4273083 1/14/2016 Jan-18 60-7070-003 4268798 1/14/2016 Jan-18 60-7070-106 4267696 1/14/2016 Jan-18 60-7070-106 4269022 1/14/2016 Jan-18 60-7070-106 4270505 1/14/2016 Jan-18 60-7070-106 4271014 1/14/2016 Jan-18 60-7070-106 4271165 1/14/2016 Jan-18 60-7070-106 4271328 1/14/2016 Jan-18 60-7070-106 4271448 1/14/2016 Jan-18 60-7070-106 4273012 1/14/2016 Jan-18 60-7070-106 4273422 1/14/2016 Jan-18 60-7070-107 4267926 1/14/2016 Jan-18 60-7070-107 4271385 1/14/2016 Jan-18 60-7075-103 4271022 1/14/2016 Jan-18 60-7075-106 4269647 1/14/2016 Jan-18 60-7075-106 4271067 1/14/2016 Jan-18 60-7075-106 4273357 1/14/2016 Jan-18 5921-018-135 4274025 1/18/2016 Jan-18 5921-018-135 4277202 1/18/2016 Jan-18 5921-018-235 4278814 1/18/2016 Jan-18 5921-024-135 4278345 1/18/2016 Jan-18 5921-024-235 4277786 1/18/2016 Jan-18 5921-030-135 4274376 1/18/2016 Jan-18 5921-030-135 4274686 1/18/2016 Jan-18 5921-030-135 4278091 1/18/2016 Jan-18 5921-030-135 4278129 1/18/2016 Jan-18 5921-030-235 4274640 1/18/2016 Jan-18 5921-030-235 4276708 1/18/2016 Jan-18 5921-034-135 4274629 1/18/2016 Jan-18 5921-034-135 4274906 1/18/2016 Jan-18 5921-034-135 4277667 1/18/2016 Jan-18 5921-034-135 4278186 1/18/2016 Jan-18 5921-034-235 4274221 1/18/2016 Jan-18 5921-034-235 4276000 1/18/2016 Jan-18 5921-034-235 4276379 1/18/2016 Jan-18 5921-034-235 4277452 1/18/2016 Jan-18 5921-218-235 4274672 1/18/2016 Jan-18 5921-218-235 4275951 1/18/2016 Jan-18 5921-218-235 4278043 1/18/2016 Jan-18 5921-218-135 4274883 1/18/2016 Jan-18 60-7070-103 4273936 1/18/2016 Jan-18 60-7070-103 4274471 1/18/2016 Jan-18 60-7070-103 4276160 1/18/2016 Jan-18 60-7070-103 4276796 1/18/2016 Jan-18 60-7070-103 4277577 1/18/2016 Jan-18 60-7070-103 4277717 1/18/2016 Jan-18 60-7070-103 4278070 1/18/2016 Jan-18 60-7070-103 4280235 1/18/2016 Jan-18 60-7070-105 4275663 1/18/2016 Jan-18 60-7070-105 4276517 1/18/2016 Jan-18 60-7070-105 4279259 1/18/2016 Jan-18 60-7070-105 4279988 1/18/2016 Jan-18 60-7070-104 4274307 1/18/2016 Jan-18 60-7070-104 4276314 1/18/2016 Jan-18 60-7070-104 4277842 1/18/2016 Jan-18 60-7070-003 4279903 1/18/2016 Jan-18 60-7070-106 4274266 1/18/2016 Jan-18 60-7070-106 4274382 1/18/2016 Jan-18 60-7070-106 4275541 1/18/2016 Jan-18 60-7070-106 4277225 1/18/2016 Jan-18 60-7070-106 4277327 1/18/2016 Jan-18 60-7070-106 4278155 1/18/2016 Jan-18 60-7070-107 4275375 1/18/2016 Jan-18 60-7075-003 4276490 1/18/2016 Jan-18 60-7075-006 4277944 1/18/2016 Jan-18 60-7075-103 4273414 1/18/2016 Jan-18 60-7075-103 4274278 1/18/2016 Jan-18 60-7075-103 4277391 1/18/2016 Jan-18 60-7075-104 4274232 1/18/2016 Jan-18 60-7075-105 4277475 1/18/2016 Jan-18 60-7075-106 4274812 1/18/2016 Jan-18 5921-018-135 4284692 1/20/2016 Jan-18 5921-018-235 4281370 1/20/2016 Jan-18 5921-024-135 4285253 1/20/2016 Jan-18 5921-024-235 4283228 1/20/2016 Jan-18 5921-024-235 4286880 1/20/2016 Jan-18 5921-030-135 4281343 1/20/2016 Jan-18 5921-030-135 4281857 1/20/2016 Jan-18 5921-030-135 4285490 1/20/2016 Jan-18 5921-030-235 4281111 1/20/2016 Jan-18 5921-030-235 4284093 1/20/2016 Jan-18 5921-030-235 4285720 1/20/2016 Jan-18 5921-034-135 4283990 1/20/2016 Jan-18 5921-034-135 4285037 1/20/2016 Jan-18 5921-034-135 4285071 1/20/2016 Jan-18 5921-034-235 4280316 1/20/2016 Jan-18 5921-034-235 4281239 1/20/2016 Jan-18 5921-034-235 4281786 1/20/2016 Jan-18 5921-034-235 4282896 1/20/2016 Jan-18 5921-034-235 4284525 1/20/2016 Jan-18 5921-212-235 4285670 1/20/2016 Jan-18 5921-218-235 4280862 1/20/2016 Jan-18 5921-218-235 4281992 1/20/2016 Jan-18 5921-218-235 4283413 1/20/2016 Jan-18 5921-218-235 4286366 1/20/2016 Jan-18 5921-218-135 4280359 1/20/2016 Jan-18 5921-218-135 4281701 1/20/2016 Jan-18 5921-218-135 4282025 1/20/2016 Jan-18 5921-218-135 4285823 1/20/2016 Jan-18 60-7070-103 4281336 1/20/2016 Jan-18 60-7070-103 4281937 1/20/2016 Jan-18 60-7070-103 4283146 1/20/2016 Jan-18 60-7070-103 4283608 1/20/2016 Jan-18 60-7070-103 4283743 1/20/2016 Jan-18 60-7070-103 4285555 1/20/2016 Jan-18 60-7070-103 4287008 1/20/2016 Jan-18 60-7070-105 4281679 1/20/2016 Jan-18 60-7070-105 4282751 1/20/2016 Jan-18 60-7070-105 4283942 1/20/2016 Jan-18 60-7070-105 4284126 1/20/2016 Jan-18 60-7070-105 4285810 1/20/2016 Jan-18 60-7070-105 4286814 1/20/2016 Jan-18 60-7070-005 4283665 1/20/2016 Jan-18 60-7070-106 4279620 1/20/2016 Jan-18 60-7070-106 4280033 1/20/2016 Jan-18 60-7070-106 4281475 1/20/2016 Jan-18 60-7070-106 4281764 1/20/2016 Jan-18 60-7070-106 4282519 1/20/2016 Jan-18 60-7070-106 4283011 1/20/2016 Jan-18 60-7070-106 4285806 1/20/2016 Jan-18 60-7070-106 4285834 1/20/2016 Jan-18 60-7070-107 4280760 1/20/2016 Jan-18 60-7075-003 4285868 1/20/2016 Jan-18 60-7075-103 4282056 1/20/2016 Jan-18 60-7075-103 4284539 1/20/2016 Jan-18 60-7075-103 4285765 1/20/2016 Jan-18 60-7075-104 4285398 1/20/2016 Jan-18 60-7075-106 4280836 1/20/2016 Jan-18 60-7075-106 4285636 1/20/2016 Jan-18 5921-018-135 4288850 1/22/2016 Jan-18 5921-018-135 4290209 1/22/2016 Jan-18 5921-024-135 4293403 1/22/2016 Jan-18 5921-024-136 4290854 1/22/2016 Jan-18 5921-024-235 4289598 1/22/2016 Jan-18 5921-024-235 4292453 1/22/2016 Jan-18 5921-030-135 4288354 1/22/2016 Jan-18 5921-030-135 4289023 1/22/2016 Jan-18 5921-030-135 4291035 1/22/2016 Jan-18 5921-030-135 4293461 1/22/2016 Jan-18 5921-030-235 4287479 1/22/2016 Jan-18 5921-030-235 4288884 1/22/2016 Jan-18 5921-030-235 4290812 1/22/2016 Jan-18 5921-030-235 4291383 1/22/2016 Jan-18 5921-030-235 4291562 1/22/2016 Jan-18 5921-034-135 4287915 1/22/2016 Jan-18 5921-034-135 4288930 1/22/2016 Jan-18 5921-034-135 4290278 1/22/2016 Jan-18 5921-034-135 4290746 1/22/2016 Jan-18 5921-034-135 4290906 1/22/2016 Jan-18 5921-034-135 4293224 1/22/2016 Jan-18 5921-034-235 4288440 1/22/2016 Jan-18 5921-034-235 4289178 1/22/2016 Jan-18 5921-034-235 4289735 1/22/2016 Jan-18 5921-034-235 4291233 1/22/2016 Jan-18 5921-034-235 4291687 1/22/2016 Jan-18 5921-034-236 4292344 1/22/2016 Jan-18 5921-212-235 4287737 1/22/2016 Jan-18 5921-218-235 4288948 1/22/2016 Jan-18 5921-218-235 4290640 1/22/2016 Jan-18 5921-218-235 4291438 1/22/2016 Jan-18 5921-218-235 4292702 1/22/2016 Jan-18 5921-044-135 4289514 1/22/2016 Jan-18 5921-218-135 4289317 1/22/2016 Jan-18 5921-218-135 4289607 1/22/2016 Jan-18 60-7070-103 4287168 1/22/2016 Jan-18 60-7070-103 4287927 1/22/2016 Jan-18 60-7070-103 4288377 1/22/2016 Jan-18 60-7070-103 4289006 1/22/2016 Jan-18 60-7070-103 4289412 1/22/2016 Jan-18 60-7070-103 4293044 1/22/2016 Jan-18 60-7070-105 4287621 1/22/2016 Jan-18 60-7070-105 4288602 1/22/2016 Jan-18 60-7070-105 4289372 1/22/2016 Jan-18 60-7070-105 4292623 1/22/2016 Jan-18 60-7070-104 4287887 1/22/2016 Jan-18 60-7070-003 4287802 1/22/2016 Jan-18 60-7070-005 4289719 1/22/2016 Jan-18 60-7070-106 4287229 1/22/2016 Jan-18 60-7070-106 4288809 1/22/2016 Jan-18 60-7070-106 4289624 1/22/2016 Jan-18 60-7070-106 4290557 1/22/2016 Jan-18 60-7070-107 4289243 1/22/2016 Jan-18 60-7075-003 4290189 1/22/2016 Jan-18 60-7075-006 4290158 1/22/2016 Jan-18 60-7075-103 4288504 1/22/2016 Jan-18 60-7075-103 4290820 1/22/2016 Jan-18 60-7075-103 4293529 1/22/2016 Jan-18 60-7075-104 4290722 1/22/2016 Jan-18 60-7075-104 4293367 1/22/2016 Jan-18 60-7075-106 4289386 1/22/2016 Jan-18 60-7075-106 4291003 1/22/2016 Jan-18 5921-024-135 4297983 1/25/2016 Jan-18 5921-024-235 4294994 1/25/2016 Jan-18 5921-030-135 4296240 1/25/2016 Jan-18 5921-030-135 4297908 1/25/2016 Jan-18 5921-030-135 4298605 1/25/2016 Jan-18 5921-030-235 4293555 1/25/2016 Jan-18 5921-030-235 4294941 1/25/2016 Jan-18 5921-030-235 4296511 1/25/2016 Jan-18 5921-030-235 4296970 1/25/2016 Jan-18 5921-034-135 4293731 1/25/2016 Jan-18 5921-034-135 4294338 1/25/2016 Jan-18 5921-034-135 4295082 1/25/2016 Jan-18 5921-034-135 4296771 1/25/2016 Jan-18 5921-034-135 4297366 1/25/2016 Jan-18 5921-034-135 4298304 1/25/2016 Jan-18 5921-034-235 4294522 1/25/2016 Jan-18 5921-034-235 4296255 1/25/2016 Jan-18 5921-034-235 4297559 1/25/2016 Jan-18 5921-034-235 4298202 1/25/2016 Jan-18 5921-034-235 4298370 1/25/2016 Jan-18 5921-218-235 4294477 1/25/2016 Jan-18 5921-218-235 4296763 1/25/2016 Jan-18 5921-218-135 4296498 1/25/2016 Jan-18 5921-218-135 4296632 1/25/2016 Jan-18 5921-218-135 4298111 1/25/2016 Jan-18 60-7070-102 4294885 1/25/2016 Jan-18 60-7070-103 4294034 1/25/2016 Jan-18 60-7070-103 4294317 1/25/2016 Jan-18 60-7070-103 4295341 1/25/2016 Jan-18 60-7070-103 4296825 1/25/2016 Jan-18 60-7070-103 4298315 1/25/2016 Jan-18 60-7070-105 4294215 1/25/2016 Jan-18 60-7070-105 4294719 1/25/2016 Jan-18 60-7070-105 4296361 1/25/2016 Jan-18 60-7070-105 4297460 1/25/2016 Jan-18 60-7070-105 4298287 1/25/2016 Jan-18 60-7070-104 4294197 1/25/2016 Jan-18 60-7070-104 4294918 1/25/2016 Jan-18 60-7070-104 4296147 1/25/2016 Jan-18 60-7070-104 4296873 1/25/2016 Jan-18 60-7070-106 4293474 1/25/2016 Jan-18 60-7070-106 4294408 1/25/2016 Jan-18 60-7070-106 4296666 1/25/2016 Jan-18 60-7070-106 4297378 1/25/2016 Jan-18 60-7070-106 4298130 1/25/2016 Jan-18 60-7070-107 4295957 1/25/2016 Jan-18 60-7075-003 4294973 1/25/2016 Jan-18 60-7075-003 4297537 1/25/2016 Jan-18 60-7075-103 4294703 1/25/2016 Jan-18 60-7075-103 4294902 1/25/2016 Jan-18 60-7075-103 4296958 1/25/2016 Jan-18 60-7075-103 4298182 1/25/2016 Jan-18 60-7075-105 4294533 1/25/2016 Jan-18 60-7075-105 4297724 1/25/2016 Jan-18 60-7075-106 4294075 1/25/2016 Jan-18 60-7075-106 4296860 1/25/2016 Jan-18 60-7075-106 4298336 1/25/2016 Jan-18 5921-018-135 4258030 1/27/2016 Jan-18 5921-024-135 4300148 1/27/2016 Jan-18 5921-024-235 4301414 1/27/2016 Jan-18 5921-030-135 4301049 1/27/2016 Jan-18 5921-030-235 4299732 1/27/2016 Jan-18 5921-030-235 4301303 1/27/2016 Jan-18 5921-030-235 4303132 1/27/2016 Jan-18 5921-034-135 4298944 1/27/2016 Jan-18 5921-034-135 4300861 1/27/2016 Jan-18 5921-034-135 4301147 1/27/2016 Jan-18 5921-034-235 4301904 1/27/2016 Jan-18 5921-034-235 4303956 1/27/2016 Jan-18 5921-218-235 4303289 1/27/2016 Jan-18 5921-218-235 4303630 1/27/2016 Jan-18 5921-218-235 4304156 1/27/2016 Jan-18 5921-218-135 4299306 1/27/2016 Jan-18 5921-218-135 4299811 1/27/2016 Jan-18 5921-218-135 4301008 1/27/2016 Jan-18 60-7070-102 4299922 1/27/2016 Jan-18 60-7070-103 4300084 1/27/2016 Jan-18 60-7070-103 4301511 1/27/2016 Jan-18 60-7070-103 4301684 1/27/2016 Jan-18 60-7070-103 4301700 1/27/2016 Jan-18 60-7070-103 4303496 1/27/2016 Jan-18 60-7070-103 4303607 1/27/2016 Jan-18 60-7070-105 4300287 1/27/2016 Jan-18 60-7070-105 4300583 1/27/2016 Jan-18 60-7070-105 4300963 1/27/2016 Jan-18 60-7070-105 4303184 1/27/2016 Jan-18 60-7070-105 4303579 1/27/2016 Jan-18 60-7070-104 4299659 1/27/2016 Jan-18 60-7070-104 4300793 1/27/2016 Jan-18 60-7070-104 4301031 1/27/2016 Jan-18 60-7070-005 4299848 1/27/2016 Jan-18 60-7070-005 4303624 1/27/2016 Jan-18 60-7070-106 4260013 1/27/2016 Jan-18 60-7070-106 4300888 1/27/2016 Jan-18 60-7070-106 4301083 1/27/2016 Jan-18 60-7070-106 4303695 1/27/2016 Jan-18 60-7070-106 4304194 1/27/2016 Jan-18 60-7070-107 4302980 1/27/2016 Jan-18 60-7075-103 4300327 1/27/2016 Jan-18 60-7075-103 4301272 1/27/2016 Jan-18 60-7075-104 4257999 1/27/2016 Jan-18 60-7075-104 4300176 1/27/2016 Jan-18 60-7075-105 4300078 1/27/2016 Jan-18 60-7075-106 4300605 1/27/2016 Jan-18 60-7075-106 4301388 1/27/2016 Jan-18 60-7075-107 4300218 1/27/2016 Jan-18 5921-024-235 4305954 1/29/2016 Jan-18 5921-024-235 4308896 1/29/2016 Jan-18 5921-030-135 4304593 1/29/2016 Jan-18 5921-030-135 4307007 1/29/2016 Jan-18 5921-030-235 4305129 1/29/2016 Jan-18 5921-030-235 4305690 1/29/2016 Jan-18 5921-030-235 4309177 1/29/2016 Jan-18 5921-034-135 4304211 1/29/2016 Jan-18 5921-034-135 4304997 1/29/2016 Jan-18 5921-034-135 4306710 1/29/2016 Jan-18 5921-034-135 4306866 1/29/2016 Jan-18 5921-034-235 4307360 1/29/2016 Jan-18 5921-034-235 4307957 1/29/2016 Jan-18 5921-034-235 4308812 1/29/2016 Jan-18 5921-034-235 4309165 1/29/2016 Jan-18 5921-034-236 4308999 1/29/2016 Jan-18 5921-218-235 4306953 1/29/2016 Jan-18 5921-218-235 4308428 1/29/2016 Jan-18 5921-218-235 4308485 1/29/2016 Jan-18 5921-218-235 4308629 1/29/2016 Jan-18 5921-218-235 4308820 1/29/2016 Jan-18 5921-218-235 4309080 1/29/2016 Jan-18 60-7070-103 4304501 1/29/2016 Jan-18 60-7070-103 4308746 1/29/2016 Jan-18 60-7070-103 4309033 1/29/2016 Jan-18 60-7070-105 4304457 1/29/2016 Jan-18 60-7070-105 4306205 1/29/2016 Jan-18 60-7070-105 4306486 1/29/2016 Jan-18 60-7070-105 4307180 1/29/2016 Jan-18 60-7070-105 4307770 1/29/2016 Jan-18 60-7070-105 4308437 1/29/2016 Jan-18 60-7070-105 4308947 1/29/2016 Jan-18 60-7070-006 4309005 1/29/2016 Jan-18 60-7070-006 4309235 1/29/2016 Jan-18 60-7070-106 4306706 1/29/2016 Jan-18 60-7070-106 4307484 1/29/2016 Jan-18 60-7070-106 4308266 1/29/2016 Jan-18 60-7070-106 4308664 1/29/2016 Jan-18 60-7070-106 4308968 1/29/2016 Jan-18 60-7070-106 4309112 1/29/2016 Jan-18 60-7070-107 4307704 1/29/2016 Jan-18 60-7075-003 4306639 1/29/2016 Jan-18 60-7075-006 4304000 1/29/2016 Jan-18 60-7075-006 4304899 1/29/2016 Jan-18 60-7075-006 4307030 1/29/2016 Jan-18 60-7075-104 4304490 1/29/2016 Jan-18 60-7075-106 4304765 1/29/2016 Jan-18 60-7075-106 4307069 1/29/2016 Jan-18 60-7075-107 4306493 1/29/2016 Jan-18 5921-018-135 4313897 2/1/2016 Feb-18 5921-018-235 4315203 2/1/2016 Feb-18 5921-024-135 4312405 2/1/2016 Feb-18 5921-024-135 4313543 2/1/2016 Feb-18 5921-024-135 4315584 2/1/2016 Feb-18 5921-024-235 4312152 2/1/2016 Feb-18 5921-030-135 4310333 2/1/2016 Feb-18 5921-030-135 4312168 2/1/2016 Feb-18 5921-030-135 4312530 2/1/2016 Feb-18 5921-030-235 4311296 2/1/2016 Feb-18 5921-030-235 4314153 2/1/2016 Feb-18 5921-034-135 4309597 2/1/2016 Feb-18 5921-034-135 4310691 2/1/2016 Feb-18 5921-034-135 4311905 2/1/2016 Feb-18 5921-034-135 4312943 2/1/2016 Feb-18 5921-034-135 4314343 2/1/2016 Feb-18 5921-034-235 4309316 2/1/2016 Feb-18 5921-034-235 4311525 2/1/2016 Feb-18 5921-034-235 4312272 2/1/2016 Feb-18 5921-034-235 4312792 2/1/2016 Feb-18 5921-034-235 4315022 2/1/2016 Feb-18 5921-218-235 4309411 2/1/2016 Feb-18 5921-218-235 4312744 2/1/2016 Feb-18 5921-218-135 4309889 2/1/2016 Feb-18 5921-218-135 4311368 2/1/2016 Feb-18 5921-218-135 4312335 2/1/2016 Feb-18 5921-218-135 4315815 2/1/2016 Feb-18 60-7070-103 4309371 2/1/2016 Feb-18 60-7070-103 4310966 2/1/2016 Feb-18 60-7070-103 4311685 2/1/2016 Feb-18 60-7070-103 4311951 2/1/2016 Feb-18 60-7070-103 4312451 2/1/2016 Feb-18 60-7070-103 4314406 2/1/2016 Feb-18 60-7070-103 4314959 2/1/2016 Feb-18 60-7070-105 4310921 2/1/2016 Feb-18 60-7070-105 4312361 2/1/2016 Feb-18 60-7070-105 4314294 2/1/2016 Feb-18 60-7070-105 4314842 2/1/2016 Feb-18 60-7070-104 4310899 2/1/2016 Feb-18 60-7070-104 4312645 2/1/2016 Feb-18 60-7070-005 4310731 2/1/2016 Feb-18 60-7070-106 4309274 2/1/2016 Feb-18 60-7070-106 4311381 2/1/2016 Feb-18 60-7070-106 4314679 2/1/2016 Feb-18 60-7075-006 4312856 2/1/2016 Feb-18 60-7075-103 4310173 2/1/2016 Feb-18 60-7075-103 4310932 2/1/2016 Feb-18 60-7075-103 4312802 2/1/2016 Feb-18 60-7075-103 4313035 2/1/2016 Feb-18 60-7075-103 4314130 2/1/2016 Feb-18 60-7075-103 4314717 2/1/2016 Feb-18 60-7075-104 4311277 2/1/2016 Feb-18 60-7075-104 4314928 2/1/2016 Feb-18 60-7075-105 4310997 2/1/2016 Feb-18 60-7075-105 4315046 2/1/2016 Feb-18 60-7075-106 4311491 2/1/2016 Feb-18 60-7075-106 4312700 2/1/2016 Feb-18 60-7075-106 4315469 2/1/2016 Feb-18 60-7075-106 4315964 2/1/2016 Feb-18 5921-024-135 4268503 2/2/2016 Feb-18 5921-030-135 4308492 2/2/2016 Feb-18 5921-218-235 4258409 2/2/2016 Feb-18 5921-218-235 4290896 2/2/2016 Feb-18 60-7070-103 4279114 2/2/2016 Feb-18 60-7070-105 4274923 2/2/2016 Feb-18 60-7070-106 4264071 2/2/2016 Feb-18 60-7075-106 4276712 2/2/2016 Feb-18 5921-018-135 4318394 2/3/2016 Feb-18 5921-024-135 4316372 2/3/2016 Feb-18 5921-024-235 4316010 2/3/2016 Feb-18 5921-024-235 4321646 2/3/2016 Feb-18 5921-030-135 4316262 2/3/2016 Feb-18 5921-030-235 4316364 2/3/2016 Feb-18 5921-030-235 4318445 2/3/2016 Feb-18 5921-030-235 4321468 2/3/2016 Feb-18 5921-034-135 4316412 2/3/2016 Feb-18 5921-034-135 4318105 2/3/2016 Feb-18 5921-034-135 4319748 2/3/2016 Feb-18 5921-034-136 4316301 2/3/2016 Feb-18 5921-034-235 4316215 2/3/2016 Feb-18 5921-034-235 4317705 2/3/2016 Feb-18 5921-034-235 4318631 2/3/2016 Feb-18 5921-034-235 4319309 2/3/2016 Feb-18 5921-034-235 4319559 2/3/2016 Feb-18 5921-034-235 4321406 2/3/2016 Feb-18 5921-218-235 4319210 2/3/2016 Feb-18 5921-218-235 4319619 2/3/2016 Feb-18 5921-218-235 4320165 2/3/2016 Feb-18 5921-218-135 4317330 2/3/2016 Feb-18 60-7070-102 4316166 2/3/2016 Feb-18 60-7070-103 4316270 2/3/2016 Feb-18 60-7070-103 4316408 2/3/2016 Feb-18 60-7070-103 4318053 2/3/2016 Feb-18 60-7070-103 4318957 2/3/2016 Feb-18 60-7070-103 4319487 2/3/2016 Feb-18 60-7070-103 4319657 2/3/2016 Feb-18 60-7070-103 4319769 2/3/2016 Feb-18 60-7070-103 4320741 2/3/2016 Feb-18 60-7070-103 4321701 2/3/2016 Feb-18 60-7070-105 4317968 2/3/2016 Feb-18 60-7070-105 4320955 2/3/2016 Feb-18 60-7070-104 4317400 2/3/2016 Feb-18 60-7070-104 4319140 2/3/2016 Feb-18 60-7070-104 4321842 2/3/2016 Feb-18 60-7070-106 4316449 2/3/2016 Feb-18 60-7070-106 4317471 2/3/2016 Feb-18 60-7070-106 4318387 2/3/2016 Feb-18 60-7070-106 4318886 2/3/2016 Feb-18 60-7070-106 4319690 2/3/2016 Feb-18 60-7070-106 4319782 2/3/2016 Feb-18 60-7070-106 4321182 2/3/2016 Feb-18 60-7075-003 4316197 2/3/2016 Feb-18 60-7075-004 4319661 2/3/2016 Feb-18 60-7075-103 4316132 2/3/2016 Feb-18 60-7075-106 4317910 2/3/2016 Feb-18 60-7075-106 4319563 2/3/2016 Feb-18 5921-018-235 4323244 2/5/2016 Feb-18 5921-024-235 4322705 2/5/2016 Feb-18 5921-030-135 4324745 2/5/2016 Feb-18 5921-030-135 4324998 2/5/2016 Feb-18 5921-030-235 4325426 2/5/2016 Feb-18 5921-030-235 4325847 2/5/2016 Feb-18 5921-034-135 4322733 2/5/2016 Feb-18 5921-034-235 4322136 2/5/2016 Feb-18 5921-034-235 4322324 2/5/2016 Feb-18 5921-034-235 4322865 2/5/2016 Feb-18 5921-034-235 4323007 2/5/2016 Feb-18 5921-218-235 4322104 2/5/2016 Feb-18 5921-218-235 4322158 2/5/2016 Feb-18 5921-218-235 4322751 2/5/2016 Feb-18 5921-218-235 4324560 2/5/2016 Feb-18 5921-218-235 4324721 2/5/2016 Feb-18 5921-218-235 4325913 2/5/2016 Feb-18 5921-030-236 4323450 2/5/2016 Feb-18 5921-044-235 4326350 2/5/2016 Feb-18 5921-218-135 4322849 2/5/2016 Feb-18 5921-218-136 4325305 2/5/2016 Feb-18 5921-218-236 4322664 2/5/2016 Feb-18 60-7070-105 4322492 2/5/2016 Feb-18 60-7070-105 4325517 2/5/2016 Feb-18 60-7070-105 4325978 2/5/2016 Feb-18 60-7070-105 4325997 2/5/2016 Feb-18 60-7070-104 4322808 2/5/2016 Feb-18 60-7070-003 4324824 2/5/2016 Feb-18 60-7070-106 4322232 2/5/2016 Feb-18 60-7070-106 4323394 2/5/2016 Feb-18 60-7070-106 4324685 2/5/2016 Feb-18 60-7070-106 4325016 2/5/2016 Feb-18 60-7070-106 4325765 2/5/2016 Feb-18 60-7070-107 4325720 2/5/2016 Feb-18 60-7075-003 4326004 2/5/2016 Feb-18 60-7075-006 4326101 2/5/2016 Feb-18 60-7075-104 4322686 2/5/2016 Feb-18 60-7075-104 4325474 2/5/2016 Feb-18 60-7075-105 4324600 2/5/2016 Feb-18 60-7075-107 4325540 2/5/2016 Feb-18 5921-018-235 4331906 2/8/2016 Feb-18 5921-024-135 4330212 2/8/2016 Feb-18 5921-024-235 4327388 2/8/2016 Feb-18 5921-024-235 4332239 2/8/2016 Feb-18 5921-030-235 4329006 2/8/2016 Feb-18 5921-030-235 4329939 2/8/2016 Feb-18 5921-030-235 4331307 2/8/2016 Feb-18 5921-034-135 4329676 2/8/2016 Feb-18 5921-034-135 4329925 2/8/2016 Feb-18 5921-034-135 4330017 2/8/2016 Feb-18 5921-034-136 4329837 2/8/2016 Feb-18 5921-034-235 4325295 2/8/2016 Feb-18 5921-034-235 4326822 2/8/2016 Feb-18 5921-034-235 4328417 2/8/2016 Feb-18 5921-034-235 4329075 2/8/2016 Feb-18 5921-034-235 4330732 2/8/2016 Feb-18 5921-218-235 4204924 2/8/2016 Feb-18 5921-218-235 4326117 2/8/2016 Feb-18 5921-218-235 4326456 2/8/2016 Feb-18 5921-218-235 4326487 2/8/2016 Feb-18 5921-218-235 4327802 2/8/2016 Feb-18 5921-218-235 4328907 2/8/2016 Feb-18 5921-218-235 4329902 2/8/2016 Feb-18 5921-218-235 4331463 2/8/2016 Feb-18 5921-218-135 4325658 2/8/2016 Feb-18 60-7070-103 4326206 2/8/2016 Feb-18 60-7070-103 4326592 2/8/2016 Feb-18 60-7070-103 4327970 2/8/2016 Feb-18 60-7070-103 4329533 2/8/2016 Feb-18 60-7070-103 4329719 2/8/2016 Feb-18 60-7070-103 4330208 2/8/2016 Feb-18 60-7070-103 4330980 2/8/2016 Feb-18 60-7070-103 4332614 2/8/2016 Feb-18 60-7070-105 4326397 2/8/2016 Feb-18 60-7070-105 4327110 2/8/2016 Feb-18 60-7070-105 4329412 2/8/2016 Feb-18 60-7070-105 4330086 2/8/2016 Feb-18 60-7070-104 4327873 2/8/2016 Feb-18 60-7070-104 4329774 2/8/2016 Feb-18 60-7070-005 4326234 2/8/2016 Feb-18 60-7070-006 4326544 2/8/2016 Feb-18 60-7070-106 4326275 2/8/2016 Feb-18 60-7070-106 4326439 2/8/2016 Feb-18 60-7070-106 4326559 2/8/2016 Feb-18 60-7070-106 4329262 2/8/2016 Feb-18 60-7070-106 4332327 2/8/2016 Feb-18 60-7070-106 4332389 2/8/2016 Feb-18 60-7070-107 4326155 2/8/2016 Feb-18 60-7070-107 4326537 2/8/2016 Feb-18 60-7075-104 4330126 2/8/2016 Feb-18 60-7075-106 4330103 2/8/2016 Feb-18 5921-024-235 4336612 2/10/2016 Feb-18 5921-030-135 4334112 2/10/2016 Feb-18 5921-030-135 4336484 2/10/2016 Feb-18 5921-030-235 4334901 2/10/2016 Feb-18 5921-030-235 4335941 2/10/2016 Feb-18 5921-034-135 4333880 2/10/2016 Feb-18 5921-034-135 4335505 2/10/2016 Feb-18 5921-034-235 4332826 2/10/2016 Feb-18 5921-034-235 4334718 2/10/2016 Feb-18 5921-034-235 4335774 2/10/2016 Feb-18 5921-034-235 4335863 2/10/2016 Feb-18 5921-034-235 4337618 2/10/2016 Feb-18 5921-218-235 4335121 2/10/2016 Feb-18 5921-218-135 4334476 2/10/2016 Feb-18 60-7070-103 4332793 2/10/2016 Feb-18 60-7070-103 4334067 2/10/2016 Feb-18 60-7070-103 4335068 2/10/2016 Feb-18 60-7070-103 4336259 2/10/2016 Feb-18 60-7070-103 4337781 2/10/2016 Feb-18 60-7070-103 4338530 2/10/2016 Feb-18 60-7070-105 4264490 2/10/2016 Feb-18 60-7070-105 4333062 2/10/2016 Feb-18 60-7070-105 4335630 2/10/2016 Feb-18 60-7070-105 4336665 2/10/2016 Feb-18 60-7070-104 4333027 2/10/2016 Feb-18 60-7070-104 4336497 2/10/2016 Feb-18 60-7070-003 4335477 2/10/2016 Feb-18 60-7070-106 4334261 2/10/2016 Feb-18 60-7070-106 4334566 2/10/2016 Feb-18 60-7070-106 4335682 2/10/2016 Feb-18 60-7070-106 4338147 2/10/2016 Feb-18 60-7070-107 4333931 2/10/2016 Feb-18 60-7075-103 4332937 2/10/2016 Feb-18 60-7075-103 4333234 2/10/2016 Feb-18 60-7075-103 4334787 2/10/2016 Feb-18 60-7075-103 4336696 2/10/2016 Feb-18 60-7075-105 4335184 2/10/2016 Feb-18 60-7075-106 4332803 2/10/2016 Feb-18 60-7075-106 4335052 2/10/2016 Feb-18 60-7075-106 4336620 2/10/2016 Feb-18 5921-018-135 4341851 2/12/2016 Feb-18 5921-018-235 4338744 2/12/2016 Feb-18 5921-030-135 4344011 2/12/2016 Feb-18 5921-030-235 4339063 2/12/2016 Feb-18 5921-030-235 4340280 2/12/2016 Feb-18 5921-030-235 4344069 2/12/2016 Feb-18 5921-034-135 4337832 2/12/2016 Feb-18 5921-034-135 4340061 2/12/2016 Feb-18 5921-034-135 4342023 2/12/2016 Feb-18 5921-034-135 4342236 2/12/2016 Feb-18 5921-034-135 4343663 2/12/2016 Feb-18 5921-034-135 4344053 2/12/2016 Feb-18 5921-034-235 4339211 2/12/2016 Feb-18 5921-034-235 4339658 2/12/2016 Feb-18 5921-034-235 4341247 2/12/2016 Feb-18 5921-034-235 4343766 2/12/2016 Feb-18 5921-034-235 4343933 2/12/2016 Feb-18 5921-218-235 4337969 2/12/2016 Feb-18 5921-218-235 4338989 2/12/2016 Feb-18 5921-218-235 4342099 2/12/2016 Feb-18 5921-218-235 4342624 2/12/2016 Feb-18 5921-218-235 4343807 2/12/2016 Feb-18 5921-218-235 4344030 2/12/2016 Feb-18 5921-030-236 4343895 2/12/2016 Feb-18 5921-218-135 4339810 2/12/2016 Feb-18 5921-218-135 4343951 2/12/2016 Feb-18 5921-218-135 4344167 2/12/2016 Feb-18 5921-218-236 4343462 2/12/2016 Feb-18 60-7070-103 4338825 2/12/2016 Feb-18 60-7070-103 4339276 2/12/2016 Feb-18 60-7070-103 4340948 2/12/2016 Feb-18 60-7070-103 4341690 2/12/2016 Feb-18 60-7070-103 4343914 2/12/2016 Feb-18 60-7070-103 4344122 2/12/2016 Feb-18 60-7070-105 4338678 2/12/2016 Feb-18 60-7070-105 4339351 2/12/2016 Feb-18 60-7070-105 4340042 2/12/2016 Feb-18 60-7070-105 4341439 2/12/2016 Feb-18 60-7070-105 4342006 2/12/2016 Feb-18 60-7070-105 4343188 2/12/2016 Feb-18 60-7070-105 4343869 2/12/2016 Feb-18 60-7070-105 4344133 2/12/2016 Feb-18 60-7070-104 4339490 2/12/2016 Feb-18 60-7070-104 4341814 2/12/2016 Feb-18 60-7070-104 4342204 2/12/2016 Feb-18 60-7070-104 4344216 2/12/2016 Feb-18 60-7070-003 4341366 2/12/2016 Feb-18 60-7070-004 4343399 2/12/2016 Feb-18 60-7070-004 4343750 2/12/2016 Feb-18 60-7070-005 4342184 2/12/2016 Feb-18 60-7070-006 4342598 2/12/2016 Feb-18 60-7070-006 4343692 2/12/2016 Feb-18 60-7070-006 4343853 2/12/2016 Feb-18 60-7070-106 4339823 2/12/2016 Feb-18 60-7070-106 4341062 2/12/2016 Feb-18 60-7070-106 4341140 2/12/2016 Feb-18 60-7070-106 4341898 2/12/2016 Feb-18 60-7070-106 4342386 2/12/2016 Feb-18 60-7070-106 4342761 2/12/2016 Feb-18 60-7070-106 4343560 2/12/2016 Feb-18 60-7070-106 4343979 2/12/2016 Feb-18 60-7075-103 4338175 2/12/2016 Feb-18 60-7075-103 4339712 2/12/2016 Feb-18 60-7075-103 4340003 2/12/2016 Feb-18 60-7075-106 4342145 2/12/2016 Feb-18 5921-018-135 4346701 2/16/2016 Feb-18 5921-018-135 4350395 2/16/2016 Feb-18 5921-018-135 4351429 2/16/2016 Feb-18 5921-018-235 4348098 2/16/2016 Feb-18 5921-024-135 4348177 2/16/2016 Feb-18 5921-030-135 4346959 2/16/2016 Feb-18 5921-030-235 4344244 2/16/2016 Feb-18 5921-030-235 4347333 2/16/2016 Feb-18 5921-030-235 4350376 2/16/2016 Feb-18 5921-034-135 4344259 2/16/2016 Feb-18 5921-034-135 4344318 2/16/2016 Feb-18 5921-034-135 4348316 2/16/2016 Feb-18 5921-034-135 4348337 2/16/2016 Feb-18 5921-034-235 4345479 2/16/2016 Feb-18 5921-034-235 4349459 2/16/2016 Feb-18 5921-034-235 4350437 2/16/2016 Feb-18 5921-218-235 4344237 2/16/2016 Feb-18 5921-218-235 4344302 2/16/2016 Feb-18 5921-218-235 4344727 2/16/2016 Feb-18 5921-218-235 4345841 2/16/2016 Feb-18 5921-218-235 4347810 2/16/2016 Feb-18 5921-218-235 4348219 2/16/2016 Feb-18 5921-218-235 4348274 2/16/2016 Feb-18 5921-218-235 4349127 2/16/2016 Feb-18 5921-218-235 4349947 2/16/2016 Feb-18 60-7070-103 4345319 2/16/2016 Feb-18 60-7070-103 4345548 2/16/2016 Feb-18 60-7070-103 4346928 2/16/2016 Feb-18 60-7070-103 4347555 2/16/2016 Feb-18 60-7070-103 4348079 2/16/2016 Feb-18 60-7070-103 4348131 2/16/2016 Feb-18 60-7070-103 4348521 2/16/2016 Feb-18 60-7070-103 4350779 2/16/2016 Feb-18 60-7070-105 4345152 2/16/2016 Feb-18 60-7070-105 4346717 2/16/2016 Feb-18 60-7070-105 4349060 2/16/2016 Feb-18 60-7070-105 4350094 2/16/2016 Feb-18 60-7070-105 4350952 2/16/2016 Feb-18 60-7070-104 4345678 2/16/2016 Feb-18 60-7070-104 4347765 2/16/2016 Feb-18 60-7070-104 4350191 2/16/2016 Feb-18 60-7070-104 4351438 2/16/2016 Feb-18 60-7070-006 4344292 2/16/2016 Feb-18 60-7070-106 4346531 2/16/2016 Feb-18 60-7070-106 4347211 2/16/2016 Feb-18 60-7070-106 4349722 2/16/2016 Feb-18 60-7070-106 4350250 2/16/2016 Feb-18 60-7070-106 4351251 2/16/2016 Feb-18 60-7075-103 4350115 2/16/2016 Feb-18 60-7075-106 4344228 2/16/2016 Feb-18 60-7075-106 4346857 2/16/2016 Feb-18 5921-018-235 4355011 2/17/2016 Feb-18 5921-024-135 4351710 2/17/2016 Feb-18 5921-030-235 4351493 2/17/2016 Feb-18 5921-030-235 4354607 2/17/2016 Feb-18 5921-030-235 4355974 2/17/2016 Feb-18 5921-030-235 4356860 2/17/2016 Feb-18 5921-034-135 4350213 2/17/2016 Feb-18 5921-034-135 4351641 2/17/2016 Feb-18 5921-034-235 4351734 2/17/2016 Feb-18 5921-034-235 4352756 2/17/2016 Feb-18 5921-034-235 4353501 2/17/2016 Feb-18 5921-034-235 4355122 2/17/2016 Feb-18 5921-034-235 4356168 2/17/2016 Feb-18 5921-034-235 4357366 2/17/2016 Feb-18 5921-034-235 4357965 2/17/2016 Feb-18 5921-034-236 4353627 2/17/2016 Feb-18 5921-218-235 4351558 2/17/2016 Feb-18 5921-030-236 4353823 2/17/2016 Feb-18 5921-218-236 4353351 2/17/2016 Feb-18 60-7070-103 4351694 2/17/2016 Feb-18 60-7070-103 4352891 2/17/2016 Feb-18 60-7070-103 4353875 2/17/2016 Feb-18 60-7070-103 4354960 2/17/2016 Feb-18 60-7070-103 4356326 2/17/2016 Feb-18 60-7070-103 4356818 2/17/2016 Feb-18 60-7070-103 4357193 2/17/2016 Feb-18 60-7070-105 4345595 2/17/2016 Feb-18 60-7070-105 4352652 2/17/2016 Feb-18 60-7070-105 4354289 2/17/2016 Feb-18 60-7070-105 4355356 2/17/2016 Feb-18 60-7070-105 4357027 2/17/2016 Feb-18 60-7070-105 4357661 2/17/2016 Feb-18 60-7070-104 4354941 2/17/2016 Feb-18 60-7070-004 4356065 2/17/2016 Feb-18 60-7070-006 4351639 2/17/2016 Feb-18 60-7070-006 4357313 2/17/2016 Feb-18 60-7070-106 4352453 2/17/2016 Feb-18 60-7070-106 4353044 2/17/2016 Feb-18 60-7070-106 4354483 2/17/2016 Feb-18 60-7070-106 4354757 2/17/2016 Feb-18 60-7070-106 4355791 2/17/2016 Feb-18 60-7070-106 4356595 2/17/2016 Feb-18 60-7070-106 4357711 2/17/2016 Feb-18 60-7075-103 4357309 2/17/2016 Feb-18 60-7075-106 4351527 2/17/2016 Feb-18 5921-018-135 4361424 2/19/2016 Feb-18 5921-018-135 4361482 2/19/2016 Feb-18 5921-030-135 4358549 2/19/2016 Feb-18 5921-030-135 4360483 2/19/2016 Feb-18 5921-030-135 4360960 2/19/2016 Feb-18 5921-030-135 4361149 2/19/2016 Feb-18 5921-030-135 4361337 2/19/2016 Feb-18 5921-030-235 4358764 2/19/2016 Feb-18 5921-034-135 4358729 2/19/2016 Feb-18 5921-034-135 4360790 2/19/2016 Feb-18 5921-034-135 4361235 2/19/2016 Feb-18 5921-034-135 4361509 2/19/2016 Feb-18 5921-034-135 4361901 2/19/2016 Feb-18 5921-034-235 4358481 2/19/2016 Feb-18 5921-034-235 4358826 2/19/2016 Feb-18 5921-034-235 4360431 2/19/2016 Feb-18 5921-218-135 4360987 2/19/2016 Feb-18 5921-218-135 4361196 2/19/2016 Feb-18 5921-218-135 4361623 2/19/2016 Feb-18 5921-218-135 4361725 2/19/2016 Feb-18 5921-218-135 4361815 2/19/2016 Feb-18 5921-218-135 4361889 2/19/2016 Feb-18 5921-218-135 4361986 2/19/2016 Feb-18 5921-218-136 4357378 2/19/2016 Feb-18 60-7070-102 4328570 2/19/2016 Feb-18 60-7070-103 4358078 2/19/2016 Feb-18 60-7070-103 4358717 2/19/2016 Feb-18 60-7070-103 4360047 2/19/2016 Feb-18 60-7070-105 4359600 2/19/2016 Feb-18 60-7070-105 4360184 2/19/2016 Feb-18 60-7070-105 4361597 2/19/2016 Feb-18 60-7070-105 4361739 2/19/2016 Feb-18 60-7070-105 4362003 2/19/2016 Feb-18 60-7070-104 4358343 2/19/2016 Feb-18 60-7070-104 4360113 2/19/2016 Feb-18 60-7070-104 4361699 2/19/2016 Feb-18 60-7070-006 4362088 2/19/2016 Feb-18 60-7070-106 4323942 2/19/2016 Feb-18 60-7070-106 4358633 2/19/2016 Feb-18 60-7070-106 4359283 2/19/2016 Feb-18 60-7070-106 4360340 2/19/2016 Feb-18 60-7070-106 4361566 2/19/2016 Feb-18 60-7070-106 4361804 2/19/2016 Feb-18 60-7075-003 4360546 2/19/2016 Feb-18 60-7075-005 4361917 2/19/2016 Feb-18 60-7075-103 4358819 2/19/2016 Feb-18 60-7075-105 4361300 2/19/2016 Feb-18 60-7075-105 4361843 2/19/2016 Feb-18 60-7075-105 4361938 2/19/2016 Feb-18 60-7075-106 4361261 2/19/2016 Feb-18 60-7075-106 4361675 2/19/2016 Feb-18 5921-018-135 4369325 2/22/2016 Feb-18 5921-018-235 4366474 2/22/2016 Feb-18 5921-024-235 4364865 2/22/2016 Feb-18 5921-024-235 4366921 2/22/2016 Feb-18 5921-030-235 4364395 2/22/2016 Feb-18 5921-030-235 4367452 2/22/2016 Feb-18 5921-030-235 4369649 2/22/2016 Feb-18 5921-034-135 4368223 2/22/2016 Feb-18 5921-034-135 4368441 2/22/2016 Feb-18 5921-034-135 4369114 2/22/2016 Feb-18 5921-034-235 4362543 2/22/2016 Feb-18 5921-034-235 4363319 2/22/2016 Feb-18 5921-034-235 4366604 2/22/2016 Feb-18 5921-034-235 4367888 2/22/2016 Feb-18 5921-218-235 4363621 2/22/2016 Feb-18 5921-218-235 4365427 2/22/2016 Feb-18 5921-218-135 4365001 2/22/2016 Feb-18 60-7070-102 4368138 2/22/2016 Feb-18 60-7070-103 4362866 2/22/2016 Feb-18 60-7070-103 4365040 2/22/2016 Feb-18 60-7070-103 4366419 2/22/2016 Feb-18 60-7070-103 4368015 2/22/2016 Feb-18 60-7070-103 4368268 2/22/2016 Feb-18 60-7070-103 4368350 2/22/2016 Feb-18 60-7070-105 4363511 2/22/2016 Feb-18 60-7070-105 4364232 2/22/2016 Feb-18 60-7070-105 4364638 2/22/2016 Feb-18 60-7070-105 4366118 2/22/2016 Feb-18 60-7070-105 4367701 2/22/2016 Feb-18 60-7070-104 4362809 2/22/2016 Feb-18 60-7070-104 4365017 2/22/2016 Feb-18 60-7070-104 4369387 2/22/2016 Feb-18 60-7070-106 4363201 2/22/2016 Feb-18 60-7070-106 4364714 2/22/2016 Feb-18 60-7070-106 4366194 2/22/2016 Feb-18 60-7070-106 4366783 2/22/2016 Feb-18 60-7070-106 4369568 2/22/2016 Feb-18 60-7070-107 4366016 2/22/2016 Feb-18 60-7070-107 4368210 2/22/2016 Feb-18 60-7075-103 4368977 2/22/2016 Feb-18 60-7075-104 4366966 2/22/2016 Feb-18 60-7075-106 4363915 2/22/2016 Feb-18 5921-018-135 4373986 2/24/2016 Feb-18 5921-024-135 4370613 2/24/2016 Feb-18 5921-024-135 4373288 2/24/2016 Feb-18 5921-024-235 4372538 2/24/2016 Feb-18 5921-030-135 4369935 2/24/2016 Feb-18 5921-030-135 4372071 2/24/2016 Feb-18 5921-030-135 4373634 2/24/2016 Feb-18 5921-030-235 4371259 2/24/2016 Feb-18 5921-030-235 4372670 2/24/2016 Feb-18 5921-034-135 4371082 2/24/2016 Feb-18 5921-034-135 4371801 2/24/2016 Feb-18 5921-034-135 4374397 2/24/2016 Feb-18 5921-034-235 4369893 2/24/2016 Feb-18 5921-034-235 4370650 2/24/2016 Feb-18 5921-034-235 4372187 2/24/2016 Feb-18 5921-034-235 4373453 2/24/2016 Feb-18 5921-034-235 4373862 2/24/2016 Feb-18 5921-218-235 4370825 2/24/2016 Feb-18 5921-218-235 4371919 2/24/2016 Feb-18 5921-218-235 4373469 2/24/2016 Feb-18 5921-218-135 4372517 2/24/2016 Feb-18 60-7070-103 4369990 2/24/2016 Feb-18 60-7070-103 4372945 2/24/2016 Feb-18 60-7070-103 4373407 2/24/2016 Feb-18 60-7070-103 4373901 2/24/2016 Feb-18 60-7070-103 4374494 2/24/2016 Feb-18 60-7070-105 4370595 2/24/2016 Feb-18 60-7070-105 4372092 2/24/2016 Feb-18 60-7070-105 4372834 2/24/2016 Feb-18 60-7070-105 4373521 2/24/2016 Feb-18 60-7070-104 4370480 2/24/2016 Feb-18 60-7070-104 4370645 2/24/2016 Feb-18 60-7070-104 4373804 2/24/2016 Feb-18 60-7070-003 4370962 2/24/2016 Feb-18 60-7070-003 4373993 2/24/2016 Feb-18 60-7070-106 4371024 2/24/2016 Feb-18 60-7070-106 4371817 2/24/2016 Feb-18 60-7070-106 4372379 2/24/2016 Feb-18 60-7070-106 4374129 2/24/2016 Feb-18 60-7070-107 4370943 2/24/2016 Feb-18 60-7075-003 4369743 2/24/2016 Feb-18 60-7075-003 4372333 2/24/2016 Feb-18 60-7075-006 4372899 2/24/2016 Feb-18 60-7075-103 4371478 2/24/2016 Feb-18 60-7075-103 4373476 2/24/2016 Feb-18 60-7075-105 4373424 2/24/2016 Feb-18 60-7075-105 4374846 2/24/2016 Feb-18 60-7075-106 4370818 2/24/2016 Feb-18 60-7075-106 4373836 2/24/2016 Feb-18 60-7075-107 4373891 2/24/2016 Feb-18 5921-018-135 4376865 2/26/2016 Feb-18 5921-024-135 4376952 2/26/2016 Feb-18 5921-024-136 4377551 2/26/2016 Feb-18 5921-030-135 4375472 2/26/2016 Feb-18 5921-030-135 4376221 2/26/2016 Feb-18 5921-030-135 4376686 2/26/2016 Feb-18 5921-030-235 4374833 2/26/2016 Feb-18 5921-030-235 4377496 2/26/2016 Feb-18 5921-030-235 4377567 2/26/2016 Feb-18 5921-034-135 4376798 2/26/2016 Feb-18 5921-034-135 4377023 2/26/2016 Feb-18 5921-034-135 4377291 2/26/2016 Feb-18 5921-034-235 4344180 2/26/2016 Feb-18 5921-034-235 4374977 2/26/2016 Feb-18 5921-034-235 4376158 2/26/2016 Feb-18 5921-034-235 4376444 2/26/2016 Feb-18 5921-034-235 4377372 2/26/2016 Feb-18 5921-034-235 4377465 2/26/2016 Feb-18 5921-034-235 4377522 2/26/2016 Feb-18 5921-034-235 4378470 2/26/2016 Feb-18 5921-034-236 4375205 2/26/2016 Feb-18 5921-218-235 4375440 2/26/2016 Feb-18 5921-218-235 4375781 2/26/2016 Feb-18 5921-218-235 4377742 2/26/2016 Feb-18 5921-030-236 4377653 2/26/2016 Feb-18 5921-218-135 4375090 2/26/2016 Feb-18 5921-218-135 4377420 2/26/2016 Feb-18 5921-218-135 4377611 2/26/2016 Feb-18 5921-218-135 4377703 2/26/2016 Feb-18 60-7070-102 4375267 2/26/2016 Feb-18 60-7070-103 4271411 2/26/2016 Feb-18 60-7070-103 4375558 2/26/2016 Feb-18 60-7070-103 4376808 2/26/2016 Feb-18 60-7070-103 4377034 2/26/2016 Feb-18 60-7070-103 4377197 2/26/2016 Feb-18 60-7070-103 4377788 2/26/2016 Feb-18 60-7070-103 4378296 2/26/2016 Feb-18 60-7070-105 4319439 2/26/2016 Feb-18 60-7070-105 4324142 2/26/2016 Feb-18 60-7070-105 4375137 2/26/2016 Feb-18 60-7070-105 4376382 2/26/2016 Feb-18 60-7070-105 4377262 2/26/2016 Feb-18 60-7070-105 4377719 2/26/2016 Feb-18 60-7070-105 4377863 2/26/2016 Feb-18 60-7070-104 4331865 2/26/2016 Feb-18 60-7070-003 4377885 2/26/2016 Feb-18 60-7070-006 4377682 2/26/2016 Feb-18 60-7070-106 4374635 2/26/2016 Feb-18 60-7070-106 4375331 2/26/2016 Feb-18 60-7070-106 4376307 2/26/2016 Feb-18 60-7070-106 4376672 2/26/2016 Feb-18 60-7070-106 4376849 2/26/2016 Feb-18 60-7070-106 4377412 2/26/2016 Feb-18 60-7075-006 4377338 2/26/2016 Feb-18 60-7075-103 4376395 2/26/2016 Feb-18 60-7075-103 4377669 2/26/2016 Feb-18 60-7075-105 4376492 2/26/2016 Feb-18 60-7075-106 4375752 2/26/2016 Feb-18 60-7075-106 4377774 2/26/2016 Feb-18 5921-024-235 4379343 2/29/2016 Mar-18 5921-024-235 4382378 2/29/2016 Mar-18 5921-030-135 4382592 2/29/2016 Mar-18 5921-030-135 4383793 2/29/2016 Mar-18 5921-030-235 4383565 2/29/2016 Mar-18 5921-034-235 4380325 2/29/2016 Mar-18 5921-034-235 4383045 2/29/2016 Mar-18 5921-034-235 4383336 2/29/2016 Mar-18 5921-034-235 4384710 2/29/2016 Mar-18 5921-218-235 4378905 2/29/2016 Mar-18 5921-218-235 4380011 2/29/2016 Mar-18 5921-218-235 4380988 2/29/2016 Mar-18 5921-218-235 4381703 2/29/2016 Mar-18 5921-218-235 4383679 2/29/2016 Mar-18 60-7070-103 4379622 2/29/2016 Mar-18 60-7070-103 4380179 2/29/2016 Mar-18 60-7070-103 4380727 2/29/2016 Mar-18 60-7070-103 4381533 2/29/2016 Mar-18 60-7070-103 4383888 2/29/2016 Mar-18 60-7070-103 4384493 2/29/2016 Mar-18 60-7070-105 4377960 2/29/2016 Mar-18 60-7070-105 4379391 2/29/2016 Mar-18 60-7070-105 4380654 2/29/2016 Mar-18 60-7070-105 4381301 2/29/2016 Mar-18 60-7070-105 4382528 2/29/2016 Mar-18 60-7070-105 4383195 2/29/2016 Mar-18 60-7070-104 4379241 2/29/2016 Mar-18 60-7070-104 4382563 2/29/2016 Mar-18 60-7070-106 4378556 2/29/2016 Mar-18 60-7070-106 4379565 2/29/2016 Mar-18 60-7070-106 4380466 2/29/2016 Mar-18 60-7070-106 4380534 2/29/2016 Mar-18 60-7070-106 4381960 2/29/2016 Mar-18 60-7070-106 4383239 2/29/2016 Mar-18 60-7070-106 4383260 2/29/2016 Mar-18 60-7070-106 4383992 2/29/2016 Mar-18 60-7070-107 4380599 2/29/2016 Mar-18 60-7075-103 4384003 2/29/2016 Mar-18 60-7075-104 4380791 2/29/2016 Mar-18 60-7075-106 4380677 2/29/2016 Mar-18 5921-018-135 4387196 3/3/2016 Mar-18 5921-018-235 4390537 3/3/2016 Mar-18 5921-018-235 4390776 3/3/2016 Mar-18 5921-024-235 4386038 3/3/2016 Mar-18 5921-024-235 4386458 3/3/2016 Mar-18 5921-024-235 4388615 3/3/2016 Mar-18 5921-030-135 4390528 3/3/2016 Mar-18 5921-030-235 4385686 3/3/2016 Mar-18 5921-030-235 4388139 3/3/2016 Mar-18 5921-030-235 4389873 3/3/2016 Mar-18 5921-030-235 4390425 3/3/2016 Mar-18 5921-034-135 4390384 3/3/2016 Mar-18 5921-034-135 4390661 3/3/2016 Mar-18 5921-034-135 4390833 3/3/2016 Mar-18 5921-034-235 4385158 3/3/2016 Mar-18 5921-034-235 4389123 3/3/2016 Mar-18 5921-034-235 4390172 3/3/2016 Mar-18 5921-034-235 4390516 3/3/2016 Mar-18 5921-218-235 4385968 3/3/2016 Mar-18 5921-218-235 4388852 3/3/2016 Mar-18 5921-218-235 4390309 3/3/2016 Mar-18 5921-218-235 4390707 3/3/2016 Mar-18 5921-218-135 4390456 3/3/2016 Mar-18 60-7070-103 4385376 3/3/2016 Mar-18 60-7070-103 4386556 3/3/2016 Mar-18 60-7070-103 4386745 3/3/2016 Mar-18 60-7070-103 4387112 3/3/2016 Mar-18 60-7070-103 4387815 3/3/2016 Mar-18 60-7070-103 4388670 3/3/2016 Mar-18 60-7070-103 4389728 3/3/2016 Mar-18 60-7070-103 4390117 3/3/2016 Mar-18 60-7070-103 4390822 3/3/2016 Mar-18 60-7070-105 4385041 3/3/2016 Mar-18 60-7070-105 4386784 3/3/2016 Mar-18 60-7070-105 4387566 3/3/2016 Mar-18 60-7070-105 4388398 3/3/2016 Mar-18 60-7070-105 4389569 3/3/2016 Mar-18 60-7070-105 4390494 3/3/2016 Mar-18 60-7070-104 4384809 3/3/2016 Mar-18 60-7070-104 4386637 3/3/2016 Mar-18 60-7070-104 4388555 3/3/2016 Mar-18 60-7070-003 4388913 3/3/2016 Mar-18 60-7070-106 4385519 3/3/2016 Mar-18 60-7070-106 4385896 3/3/2016 Mar-18 60-7070-106 4387131 3/3/2016 Mar-18 60-7070-106 4387972 3/3/2016 Mar-18 60-7070-106 4389609 3/3/2016 Mar-18 60-7070-106 4389650 3/3/2016 Mar-18 60-7070-106 4390487 3/3/2016 Mar-18 60-7070-107 4386766 3/3/2016 Mar-18 60-7070-107 4390155 3/3/2016 Mar-18 60-7070-107 4390805 3/3/2016 Mar-18 60-7075-003 4390949 3/3/2016 Mar-18 60-7075-103 4390571 3/3/2016 Mar-18 60-7075-106 4389008 3/3/2016 Mar-18 60-7075-106 4390769 3/3/2016 Mar-18 5921-018-135 4391517 3/7/2016 Mar-18 5921-024-135 4392497 3/7/2016 Mar-18 5921-024-135 4393588 3/7/2016 Mar-18 5921-030-135 4395050 3/7/2016 Mar-18 5921-030-135 4396773 3/7/2016 Mar-18 5921-030-235 4392339 3/7/2016 Mar-18 5921-030-235 4393672 3/7/2016 Mar-18 5921-030-235 4396476 3/7/2016 Mar-18 5921-034-135 4394077 3/7/2016 Mar-18 5921-034-135 4396407 3/7/2016 Mar-18 5921-034-235 4392534 3/7/2016 Mar-18 5921-034-235 4395362 3/7/2016 Mar-18 5921-034-235 4396940 3/7/2016 Mar-18 5921-218-235 4392445 3/7/2016 Mar-18 5921-218-235 4394152 3/7/2016 Mar-18 5921-218-235 4397489 3/7/2016 Mar-18 60-7070-102 4395916 3/7/2016 Mar-18 60-7070-103 4392198 3/7/2016 Mar-18 60-7070-103 4392791 3/7/2016 Mar-18 60-7070-103 4394065 3/7/2016 Mar-18 60-7070-103 4395992 3/7/2016 Mar-18 60-7070-103 4396359 3/7/2016 Mar-18 60-7070-103 4396675 3/7/2016 Mar-18 60-7070-105 4393906 3/7/2016 Mar-18 60-7070-105 4395304 3/7/2016 Mar-18 60-7070-105 4395793 3/7/2016 Mar-18 60-7070-105 4397415 3/7/2016 Mar-18 60-7070-104 4392105 3/7/2016 Mar-18 60-7070-003 4393444 3/7/2016 Mar-18 60-7070-106 4392736 3/7/2016 Mar-18 60-7070-106 4393561 3/7/2016 Mar-18 60-7070-106 4395531 3/7/2016 Mar-18 60-7070-106 4396566 3/7/2016 Mar-18 60-7075-006 4394873 3/7/2016 Mar-18 60-7075-103 4391000 3/7/2016 Mar-18 60-7075-104 4392817 3/7/2016 Mar-18 60-7075-105 4390898 3/7/2016 Mar-18 60-7075-105 4392030 3/7/2016 Mar-18 60-7075-106 4393705 3/7/2016 Mar-18 60-7075-106 4395287 3/7/2016 Mar-18 5921-018-135 4399850 3/8/2016 Mar-18 5921-024-135 4400344 3/8/2016 Mar-18 5921-024-235 4399639 3/8/2016 Mar-18 5921-030-135 4400014 3/8/2016 Mar-18 5921-030-135 4402348 3/8/2016 Mar-18 5921-030-235 4398270 3/8/2016 Mar-18 5921-034-135 4390313 3/8/2016 Mar-18 5921-034-135 4399128 3/8/2016 Mar-18 5921-034-135 4400578 3/8/2016 Mar-18 5921-034-135 4401944 3/8/2016 Mar-18 5921-034-235 4399995 3/8/2016 Mar-18 5921-034-235 4400725 3/8/2016 Mar-18 5921-034-235 4403200 3/8/2016 Mar-18 5921-218-235 4396901 3/8/2016 Mar-18 5921-218-235 4399042 3/8/2016 Mar-18 5921-218-235 4400482 3/8/2016 Mar-18 5921-218-235 4402245 3/8/2016 Mar-18 5921-218-235 4403547 3/8/2016 Mar-18 5921-218-135 4396597 3/8/2016 Mar-18 5921-218-135 4401499 3/8/2016 Mar-18 5921-218-135 4403180 3/8/2016 Mar-18 60-7070-103 4398567 3/8/2016 Mar-18 60-7070-103 4399279 3/8/2016 Mar-18 60-7070-103 4400623 3/8/2016 Mar-18 60-7070-103 4402459 3/8/2016 Mar-18 60-7070-105 4397525 3/8/2016 Mar-18 60-7070-105 4398892 3/8/2016 Mar-18 60-7070-105 4399907 3/8/2016 Mar-18 60-7070-105 4400328 3/8/2016 Mar-18 60-7070-105 4401554 3/8/2016 Mar-18 60-7070-105 4403146 3/8/2016 Mar-18 60-7070-104 4398010 3/8/2016 Mar-18 60-7070-104 4399734 3/8/2016 Mar-18 60-7070-104 4400359 3/8/2016 Mar-18 60-7070-104 4401992 3/8/2016 Mar-18 60-7070-106 4397174 3/8/2016 Mar-18 60-7070-106 4398113 3/8/2016 Mar-18 60-7070-106 4399246 3/8/2016 Mar-18 60-7070-106 4399591 3/8/2016 Mar-18 60-7070-106 4400242 3/8/2016 Mar-18 60-7070-106 4402865 3/8/2016 Mar-18 60-7070-107 4400647 3/8/2016 Mar-18 60-7075-003 4395803 3/8/2016 Mar-18 60-7075-006 4401764 3/8/2016 Mar-18 60-7075-102 4402746 3/8/2016 Mar-18 60-7075-103 4399589 3/8/2016 Mar-18 60-7075-103 4400219 3/8/2016 Mar-18 60-7075-103 4402934 3/8/2016 Mar-18 60-7075-103 4403356 3/8/2016 Mar-18 60-7075-104 4395944 3/8/2016 Mar-18 60-7075-105 4400718 3/8/2016 Mar-18 60-7075-106 4398828 3/8/2016 Mar-18 5921-018-135 4404692 3/9/2016 Mar-18 5921-024-135 4406206 3/9/2016 Mar-18 5921-024-135 4407286 3/9/2016 Mar-18 5921-024-235 4162680 3/9/2016 Mar-18 5921-024-235 4407744 3/9/2016 Mar-18 5921-030-135 4019371 3/9/2016 Mar-18 5921-030-135 4404157 3/9/2016 Mar-18 5921-030-135 4406378 3/9/2016 Mar-18 5921-030-135 4407217 3/9/2016 Mar-18 5921-030-235 4403450 3/9/2016 Mar-18 5921-030-235 4405997 3/9/2016 Mar-18 5921-030-235 4407446 3/9/2016 Mar-18 5921-030-235 4407716 3/9/2016 Mar-18 5921-034-135 4403510 3/9/2016 Mar-18 5921-034-135 4404208 3/9/2016 Mar-18 5921-034-135 4406494 3/9/2016 Mar-18 5921-034-135 4407238 3/9/2016 Mar-18 5921-034-135 4407433 3/9/2016 Mar-18 5921-034-235 4404427 3/9/2016 Mar-18 5921-034-235 4404869 3/9/2016 Mar-18 5921-034-235 4406846 3/9/2016 Mar-18 5921-034-235 4407405 3/9/2016 Mar-18 5921-034-235 4407666 3/9/2016 Mar-18 5921-218-235 4403612 3/9/2016 Mar-18 5921-218-235 4406345 3/9/2016 Mar-18 5921-218-235 4407414 3/9/2016 Mar-18 5921-218-235 4407530 3/9/2016 Mar-18 5921-212-135 4405658 3/9/2016 Mar-18 5921-218-135 4291775 3/9/2016 Mar-18 5921-218-135 4406117 3/9/2016 Mar-18 5921-218-135 4407272 3/9/2016 Mar-18 5921-218-135 4407388 3/9/2016 Mar-18 5921-218-135 4407737 3/9/2016 Mar-18 60-7070-103 4405016 3/9/2016 Mar-18 60-7070-103 4405538 3/9/2016 Mar-18 60-7070-103 4406441 3/9/2016 Mar-18 60-7070-103 4407096 3/9/2016 Mar-18 60-7070-105 4358452 3/9/2016 Mar-18 60-7070-105 4403791 3/9/2016 Mar-18 60-7070-105 4404721 3/9/2016 Mar-18 60-7070-105 4406748 3/9/2016 Mar-18 60-7070-105 4407264 3/9/2016 Mar-18 60-7070-105 4407507 3/9/2016 Mar-18 60-7070-005 4405211 3/9/2016 Mar-18 60-7070-106 4320300 3/9/2016 Mar-18 60-7070-106 4404628 3/9/2016 Mar-18 60-7070-106 4405380 3/9/2016 Mar-18 60-7070-106 4405670 3/9/2016 Mar-18 60-7070-106 4406164 3/9/2016 Mar-18 60-7070-106 4407293 3/9/2016 Mar-18 60-7070-107 4403704 3/9/2016 Mar-18 60-7070-107 4407595 3/9/2016 Mar-18 60-7075-003 4407357 3/9/2016 Mar-18 60-7080-102 4407480 3/9/2016 Mar-18 60-7075-103 4405207 3/9/2016 Mar-18 60-7075-103 4406674 3/9/2016 Mar-18 60-7075-104 4404979 3/9/2016 Mar-18 60-7075-105 4406983 3/9/2016 Mar-18 60-7075-106 4403620 3/9/2016 Mar-18 60-7075-106 4406769 3/9/2016 Mar-18 60-7075-106 4406867 3/9/2016 Mar-18 60-7075-106 4407524 3/9/2016 Mar-18 60-7075-107 4407422 3/9/2016 Mar-18 5921-018-135 4407697 3/11/2016 Mar-18 5921-018-135 4414112 3/11/2016 Mar-18 5921-030-135 4411158 3/11/2016 Mar-18 5921-030-135 4412336 3/11/2016 Mar-18 5921-030-235 4407841 3/11/2016 Mar-18 5921-034-135 4407582 3/11/2016 Mar-18 5921-034-135 4411751 3/11/2016 Mar-18 5921-034-135 4412622 3/11/2016 Mar-18 5921-034-235 4412260 3/11/2016 Mar-18 5921-034-235 4413661 3/11/2016 Mar-18 5921-218-235 4412596 3/11/2016 Mar-18 5921-218-235 4412874 3/11/2016 Mar-18 5921-218-235 4414891 3/11/2016 Mar-18 5921-218-135 4407802 3/11/2016 Mar-18 60-7070-102 4412937 3/11/2016 Mar-18 60-7070-103 4411191 3/11/2016 Mar-18 60-7070-103 4412406 3/11/2016 Mar-18 60-7070-103 4412633 3/11/2016 Mar-18 60-7070-103 4413441 3/11/2016 Mar-18 60-7070-103 4414051 3/11/2016 Mar-18 60-7070-103 4414756 3/11/2016 Mar-18 60-7070-104 4412009 3/11/2016 Mar-18 60-7070-106 4411831 3/11/2016 Mar-18 60-7070-106 4412755 3/11/2016 Mar-18 60-7070-106 4413795 3/11/2016 Mar-18 60-7070-106 4414392 3/11/2016 Mar-18 60-7075-103 4411910 3/11/2016 Mar-18 60-7075-104 4412992 3/11/2016 Mar-18 60-7075-105 4413912 3/11/2016 Mar-18 5921-018-135 4417723 3/14/2016 Mar-18 5921-018-135 4419834 3/14/2016 Mar-18 5921-018-235 4419269 3/14/2016 Mar-18 5921-024-135 4416151 3/14/2016 Mar-18 5921-024-135 4418991 3/14/2016 Mar-18 5921-024-235 4417464 3/14/2016 Mar-18 5921-024-235 4421005 3/14/2016 Mar-18 5921-030-135 4415987 3/14/2016 Mar-18 5921-030-135 4416926 3/14/2016 Mar-18 5921-030-235 4417515 3/14/2016 Mar-18 5921-030-235 4418576 3/14/2016 Mar-18 5921-030-235 4420227 3/14/2016 Mar-18 5921-034-135 4414901 3/14/2016 Mar-18 5921-034-135 4417214 3/14/2016 Mar-18 5921-034-135 4418031 3/14/2016 Mar-18 5921-034-235 4416631 3/14/2016 Mar-18 5921-034-235 4418388 3/14/2016 Mar-18 5921-034-235 4419059 3/14/2016 Mar-18 5921-034-235 4420006 3/14/2016 Mar-18 5921-034-235 4420099 3/14/2016 Mar-18 5921-218-235 4416608 3/14/2016 Mar-18 5921-218-235 4416886 3/14/2016 Mar-18 5921-218-235 4418887 3/14/2016 Mar-18 5921-218-235 4420291 3/14/2016 Mar-18 5921-218-135 4415166 3/14/2016 Mar-18 60-7070-103 4416114 3/14/2016 Mar-18 60-7070-103 4417090 3/14/2016 Mar-18 60-7070-103 4418181 3/14/2016 Mar-18 60-7070-103 4418479 3/14/2016 Mar-18 60-7070-103 4419314 3/14/2016 Mar-18 60-7070-103 4420522 3/14/2016 Mar-18 60-7070-105 4414098 3/14/2016 Mar-18 60-7070-105 4415010 3/14/2016 Mar-18 60-7070-105 4415611 3/14/2016 Mar-18 60-7070-105 4416935 3/14/2016 Mar-18 60-7070-105 4418106 3/14/2016 Mar-18 60-7070-105 4419777 3/14/2016 Mar-18 60-7070-104 4415132 3/14/2016 Mar-18 60-7070-104 4415844 3/14/2016 Mar-18 60-7070-104 4417429 3/14/2016 Mar-18 60-7070-004 4419016 3/14/2016 Mar-18 60-7070-106 4415184 3/14/2016 Mar-18 60-7070-106 4417279 3/14/2016 Mar-18 60-7070-106 4419367 3/14/2016 Mar-18 60-7070-106 4419847 3/14/2016 Mar-18 60-7070-107 4417026 3/14/2016 Mar-18 60-7075-003 4416620 3/14/2016 Mar-18 60-7075-003 4419346 3/14/2016 Mar-18 60-7075-006 4419207 3/14/2016 Mar-18 60-7075-103 4415788 3/14/2016 Mar-18 60-7075-103 4417438 3/14/2016 Mar-18 60-7075-105 4414597 3/14/2016 Mar-18 60-7075-105 4418272 3/14/2016 Mar-18 60-7075-106 4416817 3/14/2016 Mar-18 60-7075-106 4418467 3/14/2016 Mar-18 5921-018-135 4424018 3/18/2016 Mar-18 5921-024-135 4424041 3/18/2016 Mar-18 5921-024-235 4423548 3/18/2016 Mar-18 5921-024-235 4426538 3/18/2016 Mar-18 5921-030-135 4423065 3/18/2016 Mar-18 5921-030-135 4423915 3/18/2016 Mar-18 5921-030-135 4424865 3/18/2016 Mar-18 5921-030-235 4421550 3/18/2016 Mar-18 5921-030-235 4423405 3/18/2016 Mar-18 5921-030-235 4423446 3/18/2016 Mar-18 5921-030-235 4424297 3/18/2016 Mar-18 5921-030-235 4426000 3/18/2016 Mar-18 5921-034-135 4423716 3/18/2016 Mar-18 5921-034-135 4424170 3/18/2016 Mar-18 5921-034-135 4424284 3/18/2016 Mar-18 5921-034-135 4424519 3/18/2016 Mar-18 5921-034-235 4422589 3/18/2016 Mar-18 5921-034-235 4424087 3/18/2016 Mar-18 5921-034-235 4424348 3/18/2016 Mar-18 5921-034-235 4426118 3/18/2016 Mar-18 5921-218-235 4420465 3/18/2016 Mar-18 5921-218-235 4423110 3/18/2016 Mar-18 5921-218-235 4423904 3/18/2016 Mar-18 5921-218-235 4424136 3/18/2016 Mar-18 5921-218-235 4424189 3/18/2016 Mar-18 5921-218-235 4424557 3/18/2016 Mar-18 5921-218-235 4426555 3/18/2016 Mar-18 5921-218-135 4422570 3/18/2016 Mar-18 5921-218-135 4423335 3/18/2016 Mar-18 5921-218-135 4424369 3/18/2016 Mar-18 5921-218-135 4424444 3/18/2016 Mar-18 5921-218-135 4425784 3/18/2016 Mar-18 60-7070-103 4421718 3/18/2016 Mar-18 60-7070-103 4424574 3/18/2016 Mar-18 60-7070-103 4424629 3/18/2016 Mar-18 60-7070-103 4425249 3/18/2016 Mar-18 60-7070-103 4425594 3/18/2016 Mar-18 60-7070-103 4426403 3/18/2016 Mar-18 60-7070-105 4420742 3/18/2016 Mar-18 60-7070-105 4422396 3/18/2016 Mar-18 60-7070-105 4423240 3/18/2016 Mar-18 60-7070-105 4423678 3/18/2016 Mar-18 60-7070-105 4424158 3/18/2016 Mar-18 60-7070-105 4424376 3/18/2016 Mar-18 60-7070-105 4426207 3/18/2016 Mar-18 60-7070-104 4420676 3/18/2016 Mar-18 60-7070-104 4426474 3/18/2016 Mar-18 60-7070-106 4421290 3/18/2016 Mar-18 60-7070-106 4422543 3/18/2016 Mar-18 60-7070-106 4423744 3/18/2016 Mar-18 60-7070-106 4423962 3/18/2016 Mar-18 60-7070-106 4424039 3/18/2016 Mar-18 60-7070-106 4424601 3/18/2016 Mar-18 60-7070-106 4425072 3/18/2016 Mar-18 60-7070-106 4425807 3/18/2016 Mar-18 60-7070-107 4423569 3/18/2016 Mar-18 60-7070-107 4424503 3/18/2016 Mar-18 60-7075-006 4424311 3/18/2016 Mar-18 60-7075-103 4422438 3/18/2016 Mar-18 60-7075-103 4423856 3/18/2016 Mar-18 60-7075-103 4423991 3/18/2016 Mar-18 60-7075-104 4425709 3/18/2016 Mar-18 60-7075-105 4425265 3/18/2016 Mar-18 60-7075-106 4423621 3/18/2016 Mar-18 5921-018-235 4429325 3/21/2016 Mar-18 5921-024-235 4427861 3/21/2016 Mar-18 5921-024-235 4431137 3/21/2016 Mar-18 5921-030-135 4429285 3/21/2016 Mar-18 5921-030-135 4430471 3/21/2016 Mar-18 5921-030-235 4427343 3/21/2016 Mar-18 5921-030-235 4431171 3/21/2016 Mar-18 5921-034-135 4428965 3/21/2016 Mar-18 5921-034-235 4427097 3/21/2016 Mar-18 5921-034-235 4427729 3/21/2016 Mar-18 5921-034-235 4429568 3/21/2016 Mar-18 5921-034-235 4430376 3/21/2016 Mar-18 5921-034-235 4432183 3/21/2016 Mar-18 5921-218-235 4427452 3/21/2016 Mar-18 5921-218-235 4427764 3/21/2016 Mar-18 5921-218-235 4428642 3/21/2016 Mar-18 5921-218-235 4430284 3/21/2016 Mar-18 5921-218-235 4431074 3/21/2016 Mar-18 60-7070-103 4426899 3/21/2016 Mar-18 60-7070-103 4427260 3/21/2016 Mar-18 60-7070-103 4427577 3/21/2016 Mar-18 60-7070-103 4428516 3/21/2016 Mar-18 60-7070-103 4428983 3/21/2016 Mar-18 60-7070-103 4430221 3/21/2016 Mar-18 60-7070-103 4430686 3/21/2016 Mar-18 60-7070-103 4430999 3/21/2016 Mar-18 60-7070-105 4427225 3/21/2016 Mar-18 60-7070-105 4428164 3/21/2016 Mar-18 60-7070-105 4428851 3/21/2016 Mar-18 60-7070-105 4430073 3/21/2016 Mar-18 60-7070-105 4430883 3/21/2016 Mar-18 60-7070-104 4427646 3/21/2016 Mar-18 60-7070-104 4430542 3/21/2016 Mar-18 60-7070-104 4432407 3/21/2016 Mar-18 60-7070-106 4427045 3/21/2016 Mar-18 60-7070-106 4427195 3/21/2016 Mar-18 60-7070-106 4428439 3/21/2016 Mar-18 60-7070-106 4428730 3/21/2016 Mar-18 60-7070-106 4430416 3/21/2016 Mar-18 60-7070-106 4430767 3/21/2016 Mar-18 60-7070-106 4431897 3/21/2016 Mar-18 60-7070-106 4432046 3/21/2016 Mar-18 60-7075-103 4429677 3/21/2016 Mar-18 60-7075-104 4429534 3/21/2016 Mar-18 60-7070-104 4376947 3/22/2016 Mar-18 5921-018-135 4434204 3/23/2016 Mar-18 5921-024-135 4435704 3/23/2016 Mar-18 5921-024-235 4436684 3/23/2016 Mar-18 5921-030-135 4433913 3/23/2016 Mar-18 5921-030-135 4435432 3/23/2016 Mar-18 5921-030-135 4436771 3/23/2016 Mar-18 5921-030-235 4434215 3/23/2016 Mar-18 5921-030-235 4435654 3/23/2016 Mar-18 5921-030-235 4437101 3/23/2016 Mar-18 5921-030-235 4437537 3/23/2016 Mar-18 5921-034-135 4432891 3/23/2016 Mar-18 5921-034-135 4435180 3/23/2016 Mar-18 5921-034-135 4436286 3/23/2016 Mar-18 5921-034-135 4438273 3/23/2016 Mar-18 5921-034-235 4433765 3/23/2016 Mar-18 5921-034-235 4434317 3/23/2016 Mar-18 5921-034-235 4436818 3/23/2016 Mar-18 5921-034-235 4437456 3/23/2016 Mar-18 5921-218-235 4433037 3/23/2016 Mar-18 5921-218-235 4434052 3/23/2016 Mar-18 5921-218-235 4436152 3/23/2016 Mar-18 5921-218-235 4436962 3/23/2016 Mar-18 5921-218-235 4437439 3/23/2016 Mar-18 60-7070-103 4432532 3/23/2016 Mar-18 60-7070-103 4433754 3/23/2016 Mar-18 60-7070-103 4434010 3/23/2016 Mar-18 60-7070-103 4434892 3/23/2016 Mar-18 60-7070-103 4435198 3/23/2016 Mar-18 60-7070-103 4435450 3/23/2016 Mar-18 60-7070-103 4437350 3/23/2016 Mar-18 60-7070-105 4432938 3/23/2016 Mar-18 60-7070-105 4433834 3/23/2016 Mar-18 60-7070-105 4434270 3/23/2016 Mar-18 60-7070-105 4435053 3/23/2016 Mar-18 60-7070-105 4437193 3/23/2016 Mar-18 60-7070-105 4437769 3/23/2016 Mar-18 60-7070-104 4433538 3/23/2016 Mar-18 60-7070-104 4437268 3/23/2016 Mar-18 60-7070-106 4433457 3/23/2016 Mar-18 60-7070-106 4434184 3/23/2016 Mar-18 60-7070-106 4435259 3/23/2016 Mar-18 60-7070-106 4436293 3/23/2016 Mar-18 60-7070-106 4437027 3/23/2016 Mar-18 60-7070-106 4438084 3/23/2016 Mar-18 60-7070-107 4433689 3/23/2016 Mar-18 60-7070-107 4438468 3/23/2016 Mar-18 60-7075-006 4436031 3/23/2016 Mar-18 60-7075-103 4430877 3/23/2016 Mar-18 60-7075-103 4437089 3/23/2016 Mar-18 60-7075-103 4438009 3/23/2016 Mar-18 60-7075-104 4435581 3/23/2016 Mar-18 60-7075-106 4433118 3/23/2016 Mar-18 60-7075-106 4433708 3/23/2016 Mar-18 60-7075-106 4434885 3/23/2016 Mar-18 60-7075-106 4437378 3/23/2016 Mar-18 5921-018-135 4440487 3/24/2016 Mar-18 5921-018-135 4443041 3/24/2016 Mar-18 5921-024-135 4438819 3/24/2016 Mar-18 5921-024-235 4439277 3/24/2016 Mar-18 5921-024-235 4439869 3/24/2016 Mar-18 5921-024-235 4442864 3/24/2016 Mar-18 5921-030-235 4439188 3/24/2016 Mar-18 5921-030-235 4440117 3/24/2016 Mar-18 5921-030-235 4442466 3/24/2016 Mar-18 5921-034-135 4440378 3/24/2016 Mar-18 5921-034-235 4439306 3/24/2016 Mar-18 5921-034-235 4439721 3/24/2016 Mar-18 5921-034-235 4440563 3/24/2016 Mar-18 5921-034-235 4440931 3/24/2016 Mar-18 5921-034-235 4441028 3/24/2016 Mar-18 5921-218-235 4438916 3/24/2016 Mar-18 5921-218-235 4439807 3/24/2016 Mar-18 5921-218-235 4440091 3/24/2016 Mar-18 5921-218-235 4440172 3/24/2016 Mar-18 5921-218-235 4440592 3/24/2016 Mar-18 5921-218-235 4440724 3/24/2016 Mar-18 5921-218-235 4440782 3/24/2016 Mar-18 5921-218-235 4440867 3/24/2016 Mar-18 5921-218-235 4440912 3/24/2016 Mar-18 5921-218-135 4439212 3/24/2016 Mar-18 60-7070-103 4438565 3/24/2016 Mar-18 60-7070-103 4439659 3/24/2016 Mar-18 60-7070-103 4440036 3/24/2016 Mar-18 60-7070-103 4440350 3/24/2016 Mar-18 60-7070-103 4440769 3/24/2016 Mar-18 60-7070-103 4440833 3/24/2016 Mar-18 60-7070-103 4441627 3/24/2016 Mar-18 60-7070-105 4438803 3/24/2016 Mar-18 60-7070-105 4440004 3/24/2016 Mar-18 60-7070-105 4441230 3/24/2016 Mar-18 60-7070-105 4442146 3/24/2016 Mar-18 60-7070-105 4442727 3/24/2016 Mar-18 60-7070-003 4442216 3/24/2016 Mar-18 60-7070-005 4440753 3/24/2016 Mar-18 60-7070-106 4439208 3/24/2016 Mar-18 60-7070-106 4439541 3/24/2016 Mar-18 60-7070-106 4440439 3/24/2016 Mar-18 60-7070-106 4440494 3/24/2016 Mar-18 60-7070-106 4440619 3/24/2016 Mar-18 60-7070-106 4440805 3/24/2016 Mar-18 60-7070-106 4441071 3/24/2016 Mar-18 60-7070-106 4441538 3/24/2016 Mar-18 60-7070-106 4442926 3/24/2016 Mar-18 60-7075-103 4439824 3/24/2016 Mar-18 60-7075-106 4438554 3/24/2016 Mar-18 60-7075-106 4440138 3/24/2016 Mar-18 5921-018-135 4455181 3/29/2016 Mar-18 5921-018-135 4455943 3/29/2016 Mar-18 5921-018-235 4451769 3/29/2016 Mar-18 5921-018-235 4453797 3/29/2016 Mar-18 5921-018-235 4456698 3/29/2016 Mar-18 5921-024-135 4451571 3/29/2016 Mar-18 5921-024-135 4453659 3/29/2016 Mar-18 5921-024-135 4456050 3/29/2016 Mar-18 5921-024-235 4451299 3/29/2016 Mar-18 5921-024-235 4452449 3/29/2016 Mar-18 5921-024-236 4450832 3/29/2016 Mar-18 5921-030-136 4451252 3/29/2016 Mar-18 5921-030-235 4455700 3/29/2016 Mar-18 5921-034-135 4453980 3/29/2016 Mar-18 5921-034-135 4454506 3/29/2016 Mar-18 5921-034-135 4455645 3/29/2016 Mar-18 5921-034-235 4452109 3/29/2016 Mar-18 5921-034-235 4452291 3/29/2016 Mar-18 5921-034-235 4456032 3/29/2016 Mar-18 5921-034-236 4451164 3/29/2016 Mar-18 5921-212-235 4454478 3/29/2016 Mar-18 5921-218-235 4451846 3/29/2016 Mar-18 5921-218-235 4452465 3/29/2016 Mar-18 5921-218-235 4452788 3/29/2016 Mar-18 5921-218-235 4453283 3/29/2016 Mar-18 5921-218-235 4454151 3/29/2016 Mar-18 5921-218-235 4455272 3/29/2016 Mar-18 5921-218-235 4455553 3/29/2016 Mar-18 5921-218-235 4455839 3/29/2016 Mar-18 5921-218-235 4456013 3/29/2016 Mar-18 5921-218-135 4450884 3/29/2016 Mar-18 5921-218-135 4452081 3/29/2016 Mar-18 5921-218-135 4452726 3/29/2016 Mar-18 5921-218-135 4454974 3/29/2016 Mar-18 5921-218-135 4455825 3/29/2016 Mar-18 5921-218-136 4454237 3/29/2016 Mar-18 60-7070-103 4451384 3/29/2016 Mar-18 60-7070-103 4451674 3/29/2016 Mar-18 60-7070-103 4452329 3/29/2016 Mar-18 60-7070-103 4452505 3/29/2016 Mar-18 60-7070-103 4452695 3/29/2016 Mar-18 60-7070-103 4453212 3/29/2016 Mar-18 60-7070-103 4455293 3/29/2016 Mar-18 60-7070-005 4453895 3/29/2016 Mar-18 60-7070-106 4451210 3/29/2016 Mar-18 60-7070-106 4452184 3/29/2016 Mar-18 60-7070-106 4452735 3/29/2016 Mar-18 60-7070-106 4453249 3/29/2016 Mar-18 60-7070-106 4453979 3/29/2016 Mar-18 60-7070-106 4454200 3/29/2016 Mar-18 60-7070-106 4454961 3/29/2016 Mar-18 60-7070-106 4455388 3/29/2016 Mar-18 60-7070-106 4455621 3/29/2016 Mar-18 60-7070-106 4455927 3/29/2016 Mar-18 60-7070-106 4456406 3/29/2016 Mar-18 60-7075-006 4453784 3/29/2016 Mar-18 60-7070-153 4452703 3/29/2016 Mar-18 60-7075-103 4453254 3/29/2016 Mar-18 60-7075-104 4308923 3/29/2016 Mar-18 60-7075-105 4308934 3/29/2016 Mar-18 60-7075-106 4452454 3/29/2016 Mar-18 60-7075-106 4454421 3/29/2016 Mar-18 60-7075-106 4455326 3/29/2016 Mar-18 5921-018-135 4459025 3/31/2016 Apr-18 5921-018-135 4461258 3/31/2016 Apr-18 5921-024-135 4459307 3/31/2016 Apr-18 5921-024-235 4459382 3/31/2016 Apr-18 5921-024-235 4461215 3/31/2016 Apr-18 5921-030-135 4457796 3/31/2016 Apr-18 5921-030-235 4458938 3/31/2016 Apr-18 5921-034-135 4456803 3/31/2016 Apr-18 5921-034-135 4458760 3/31/2016 Apr-18 5921-034-135 4459444 3/31/2016 Apr-18 5921-034-235 4456577 3/31/2016 Apr-18 5921-034-235 4457442 3/31/2016 Apr-18 5921-034-235 4458699 3/31/2016 Apr-18 5921-034-235 4459250 3/31/2016 Apr-18 5921-034-236 4458337 3/31/2016 Apr-18 5921-218-235 4456861 3/31/2016 Apr-18 5921-218-235 4457945 3/31/2016 Apr-18 5921-218-235 4458773 3/31/2016 Apr-18 5921-218-235 4461477 3/31/2016 Apr-18 5921-218-135 4457207 3/31/2016 Apr-18 5921-218-135 4457689 3/31/2016 Apr-18 5921-218-135 4458371 3/31/2016 Apr-18 60-7070-103 4457282 3/31/2016 Apr-18 60-7070-103 4457658 3/31/2016 Apr-18 60-7070-103 4457708 3/31/2016 Apr-18 60-7070-103 4458843 3/31/2016 Apr-18 60-7070-103 4461648 3/31/2016 Apr-18 60-7070-105 4459353 3/31/2016 Apr-18 60-7070-105 4461052 3/31/2016 Apr-18 60-7070-104 4459324 3/31/2016 Apr-18 60-7070-104 4459437 3/31/2016 Apr-18 60-7070-005 4457141 3/31/2016 Apr-18 60-7070-005 4458509 3/31/2016 Apr-18 60-7070-106 4457071 3/31/2016 Apr-18 60-7070-106 4457899 3/31/2016 Apr-18 60-7070-106 4458718 3/31/2016 Apr-18 60-7070-107 4461386 3/31/2016 Apr-18 60-7075-003 4457461 3/31/2016 Apr-18 60-7075-103 4461109 3/31/2016 Apr-18 60-7075-103 4462494 3/31/2016 Apr-18 60-7075-104 4461483 3/31/2016 Apr-18 60-7075-106 4459428 3/31/2016 Apr-18 5921-018-135 4463336 4/4/2016 Apr-18 5921-018-135 4465705 4/4/2016 Apr-18 5921-024-135 4467668 4/4/2016 Apr-18 5921-024-235 4464897 4/4/2016 Apr-18 5921-030-135 4464116 4/4/2016 Apr-18 5921-030-235 4462281 4/4/2016 Apr-18 5921-030-235 4465574 4/4/2016 Apr-18 5921-030-235 4466637 4/4/2016 Apr-18 5921-034-135 4462223 4/4/2016 Apr-18 5921-034-135 4462657 4/4/2016 Apr-18 5921-034-135 4463992 4/4/2016 Apr-18 5921-034-135 4466055 4/4/2016 Apr-18 5921-034-135 4467550 4/4/2016 Apr-18 5921-034-235 4462642 4/4/2016 Apr-18 5921-034-235 4464100 4/4/2016 Apr-18 5921-034-235 4466323 4/4/2016 Apr-18 5921-034-235 4466692 4/4/2016 Apr-18 5921-218-235 4440965 4/4/2016 Apr-18 5921-218-235 4464061 4/4/2016 Apr-18 5921-218-235 4465330 4/4/2016 Apr-18 5921-218-235 4466914 4/4/2016 Apr-18 5921-218-135 4465127 4/4/2016 Apr-18 60-7070-102 4464558 4/4/2016 Apr-18 60-7070-103 4463045 4/4/2016 Apr-18 60-7070-103 4463260 4/4/2016 Apr-18 60-7070-103 4463633 4/4/2016 Apr-18 60-7070-103 4464042 4/4/2016 Apr-18 60-7070-103 4465865 4/4/2016 Apr-18 60-7070-103 4466436 4/4/2016 Apr-18 60-7070-103 4467098 4/4/2016 Apr-18 60-7070-105 4462473 4/4/2016 Apr-18 60-7070-105 4463148 4/4/2016 Apr-18 60-7070-105 4463294 4/4/2016 Apr-18 60-7070-105 4463596 4/4/2016 Apr-18 60-7070-105 4463916 4/4/2016 Apr-18 60-7070-105 4465018 4/4/2016 Apr-18 60-7070-105 4465999 4/4/2016 Apr-18 60-7070-105 4466830 4/4/2016 Apr-18 60-7070-104 4462603 4/4/2016 Apr-18 60-7070-104 4464026 4/4/2016 Apr-18 60-7070-003 4462559 4/4/2016 Apr-18 60-7070-106 4461994 4/4/2016 Apr-18 60-7070-106 4463717 4/4/2016 Apr-18 60-7070-106 4464694 4/4/2016 Apr-18 60-7070-106 4466732 4/4/2016 Apr-18 60-7070-106 4467183 4/4/2016 Apr-18 60-7070-106 4467986 4/4/2016 Apr-18 60-7075-103 4462585 4/4/2016 Apr-18 60-7075-103 4464907 4/4/2016 Apr-18 60-7075-103 4467359 4/4/2016 Apr-18 60-7075-105 4461653 4/4/2016 Apr-18 60-7075-105 4466685 4/4/2016 Apr-18 60-7075-106 4463701 4/4/2016 Apr-18 60-7075-106 4464035 4/4/2016 Apr-18 60-7075-106 4466824 4/4/2016 Apr-18 60-7075-106 4467108 4/4/2016 Apr-18 5921-018-135 4449205 4/7/2016 Apr-18 5921-018-235 4445857 4/7/2016 Apr-18 5921-018-235 4449308 4/7/2016 Apr-18 5921-024-135 4445050 4/7/2016 Apr-18 5921-024-235 4445633 4/7/2016 Apr-18 5921-024-235 4449320 4/7/2016 Apr-18 5921-030-235 4444511 4/7/2016 Apr-18 5921-030-235 4446940 4/7/2016 Apr-18 5921-030-235 4449171 4/7/2016 Apr-18 5921-034-135 4443952 4/7/2016 Apr-18 5921-034-135 4444480 4/7/2016 Apr-18 5921-034-135 4445874 4/7/2016 Apr-18 5921-034-135 4447135 4/7/2016 Apr-18 5921-034-135 4449074 4/7/2016 Apr-18 5921-034-135 4449527 4/7/2016 Apr-18 5921-034-135 4450440 4/7/2016 Apr-18 5921-034-235 4443526 4/7/2016 Apr-18 5921-034-235 4443877 4/7/2016 Apr-18 5921-034-235 4444031 4/7/2016 Apr-18 5921-034-235 4444553 4/7/2016 Apr-18 5921-034-235 4445434 4/7/2016 Apr-18 5921-034-235 4446862 4/7/2016 Apr-18 5921-034-235 4447368 4/7/2016 Apr-18 5921-034-235 4449035 4/7/2016 Apr-18 5921-034-235 4450251 4/7/2016 Apr-18 5921-212-235 4443444 4/7/2016 Apr-18 5921-212-235 4445985 4/7/2016 Apr-18 5921-218-235 4443705 4/7/2016 Apr-18 5921-218-235 4443934 4/7/2016 Apr-18 5921-218-235 4444335 4/7/2016 Apr-18 5921-218-235 4447142 4/7/2016 Apr-18 5921-218-235 4447458 4/7/2016 Apr-18 5921-218-235 4449090 4/7/2016 Apr-18 5921-218-235 4450570 4/7/2016 Apr-18 5921-218-135 4442715 4/7/2016 Apr-18 5921-218-135 4443459 4/7/2016 Apr-18 5921-218-135 4445294 4/7/2016 Apr-18 5921-218-135 4446578 4/7/2016 Apr-18 5921-218-135 4450035 4/7/2016 Apr-18 5921-218-136 4450246 4/7/2016 Apr-18 5921-218-236 4446359 4/7/2016 Apr-18 60-7070-103 4443779 4/7/2016 Apr-18 60-7070-103 4444147 4/7/2016 Apr-18 60-7070-103 4446022 4/7/2016 Apr-18 60-7070-103 4447190 4/7/2016 Apr-18 60-7070-103 4447220 4/7/2016 Apr-18 60-7070-103 4448987 4/7/2016 Apr-18 60-7070-103 4449144 4/7/2016 Apr-18 60-7070-103 4449222 4/7/2016 Apr-18 60-7070-103 4449515 4/7/2016 Apr-18 60-7070-103 4450396 4/7/2016 Apr-18 60-7070-105 4444286 4/7/2016 Apr-18 60-7070-105 4444970 4/7/2016 Apr-18 60-7070-005 4445717 4/7/2016 Apr-18 60-7070-106 4443574 4/7/2016 Apr-18 60-7070-106 4445078 4/7/2016 Apr-18 60-7070-106 4446597 4/7/2016 Apr-18 60-7070-106 4446794 4/7/2016 Apr-18 60-7070-106 4449813 4/7/2016 Apr-18 60-7070-106 4450878 4/7/2016 Apr-18 60-7070-107 4446385 4/7/2016 Apr-18 60-7075-003 4446870 4/7/2016 Apr-18 60-7075-006 4446033 4/7/2016 Apr-18 60-7070-153 4444168 4/7/2016 Apr-18 60-7070-153 4444319 4/7/2016 Apr-18 60-7070-153 4444479 4/7/2016 Apr-18 60-7075-103 4443812 4/7/2016 Apr-18 60-7075-103 4447375 4/7/2016 Apr-18 60-7075-103 4449298 4/7/2016 Apr-18 60-7075-106 4443561 4/7/2016 Apr-18 60-7075-106 4444272 4/7/2016 Apr-18 60-7075-106 4444422 4/7/2016 Apr-18 60-7075-106 4444613 4/7/2016 Apr-18 60-7075-106 4446344 4/7/2016 Apr-18 60-7075-106 4449192 4/7/2016 Apr-18 5921-018-135 4468731 4/8/2016 Apr-18 5921-018-235 4470202 4/8/2016 Apr-18 5921-018-235 4470793 4/8/2016 Apr-18 5921-018-235 4471272 4/8/2016 Apr-18 5921-024-235 4468490 4/8/2016 Apr-18 5921-024-235 4471106 4/8/2016 Apr-18 5921-030-135 4467702 4/8/2016 Apr-18 5921-030-135 4470287 4/8/2016 Apr-18 5921-030-135 4471217 4/8/2016 Apr-18 5921-030-135 4472489 4/8/2016 Apr-18 5921-030-235 4468615 4/8/2016 Apr-18 5921-030-235 4470391 4/8/2016 Apr-18 5921-034-135 4469361 4/8/2016 Apr-18 5921-034-135 4470169 4/8/2016 Apr-18 5921-034-135 4471008 4/8/2016 Apr-18 5921-034-235 4467652 4/8/2016 Apr-18 5921-034-235 4470111 4/8/2016 Apr-18 5921-034-235 4470294 4/8/2016 Apr-18 5921-034-235 4470679 4/8/2016 Apr-18 5921-034-235 4470992 4/8/2016 Apr-18 5921-034-235 4471134 4/8/2016 Apr-18 5921-034-236 4472190 4/8/2016 Apr-18 5921-212-235 4472347 4/8/2016 Apr-18 5921-218-235 4469569 4/8/2016 Apr-18 5921-218-235 4471152 4/8/2016 Apr-18 5921-218-135 4468517 4/8/2016 Apr-18 5921-218-135 4470786 4/8/2016 Apr-18 5921-218-135 4471744 4/8/2016 Apr-18 60-7070-103 4467509 4/8/2016 Apr-18 60-7070-103 4467699 4/8/2016 Apr-18 60-7070-103 4468997 4/8/2016 Apr-18 60-7070-103 4470273 4/8/2016 Apr-18 60-7070-103 4470520 4/8/2016 Apr-18 60-7070-103 4470826 4/8/2016 Apr-18 60-7070-103 4471181 4/8/2016 Apr-18 60-7070-103 4472491 4/8/2016 Apr-18 60-7070-105 4467606 4/8/2016 Apr-18 60-7070-105 4468868 4/8/2016 Apr-18 60-7070-104 4469217 4/8/2016 Apr-18 60-7070-104 4471785 4/8/2016 Apr-18 60-7070-104 4472644 4/8/2016 Apr-18 60-7070-003 4471049 4/8/2016 Apr-18 60-7070-005 4471856 4/8/2016 Apr-18 60-7070-006 4469357 4/8/2016 Apr-18 60-7070-106 4468823 4/8/2016 Apr-18 60-7070-106 4470481 4/8/2016 Apr-18 60-7070-106 4470565 4/8/2016 Apr-18 60-7070-106 4471684 4/8/2016 Apr-18 60-7070-106 4471970 4/8/2016 Apr-18 60-7070-107 4467424 4/8/2016 Apr-18 60-7070-154 4467316 4/8/2016 Apr-18 60-7075-103 4471110 4/8/2016 Apr-18 60-7075-106 4468984 4/8/2016 Apr-18 5921-018-135 4481801 4/11/2016 Apr-18 5921-024-135 4480958 4/11/2016 Apr-18 5921-024-235 4481180 4/11/2016 Apr-18 5921-030-235 4480217 4/11/2016 Apr-18 5921-034-135 4481497 4/11/2016 Apr-18 5921-034-235 4478693 4/11/2016 Apr-18 5921-034-235 4481011 4/11/2016 Apr-18 5921-034-235 4482689 4/11/2016 Apr-18 5921-218-235 4480632 4/11/2016 Apr-18 5921-218-235 4482627 4/11/2016 Apr-18 5921-218-135 4478831 4/11/2016 Apr-18 60-7070-102 4479804 4/11/2016 Apr-18 60-7070-103 4479597 4/11/2016 Apr-18 60-7070-103 4480229 4/11/2016 Apr-18 60-7070-103 4480335 4/11/2016 Apr-18 60-7070-103 4481990 4/11/2016 Apr-18 60-7070-103 4482765 4/11/2016 Apr-18 60-7070-103 4483363 4/11/2016 Apr-18 60-7070-103 4483513 4/11/2016 Apr-18 60-7070-105 4479702 4/11/2016 Apr-18 60-7070-105 4480181 4/11/2016 Apr-18 60-7070-105 4481552 4/11/2016 Apr-18 60-7070-105 4482875 4/11/2016 Apr-18 60-7070-104 4477715 4/11/2016 Apr-18 60-7070-104 4479858 4/11/2016 Apr-18 60-7070-005 4481409 4/11/2016 Apr-18 60-7070-106 4478934 4/11/2016 Apr-18 60-7070-106 4479938 4/11/2016 Apr-18 60-7070-106 4480286 4/11/2016 Apr-18 60-7070-106 4481838 4/11/2016 Apr-18 60-7070-106 4482978 4/11/2016 Apr-18 60-7075-103 4481133 4/11/2016 Apr-18 60-7075-103 4483545 4/11/2016 Apr-18 60-7075-104 4482966 4/11/2016 Apr-18 60-7075-107 4482540 4/11/2016 Apr-18 5921-024-235 4474540 4/13/2016 Apr-18 5921-024-235 4477122 4/13/2016 Apr-18 5921-030-135 4475674 4/13/2016 Apr-18 5921-030-135 4476429 4/13/2016 Apr-18 5921-030-135 4478104 4/13/2016 Apr-18 5921-030-135 4478307 4/13/2016 Apr-18 5921-030-235 4474207 4/13/2016 Apr-18 5921-030-235 4475252 4/13/2016 Apr-18 5921-030-235 4477179 4/13/2016 Apr-18 5921-034-135 4473624 4/13/2016 Apr-18 5921-034-135 4473994 4/13/2016 Apr-18 5921-034-135 4476845 4/13/2016 Apr-18 5921-034-235 4473816 4/13/2016 Apr-18 5921-034-235 4474044 4/13/2016 Apr-18 5921-034-235 4475500 4/13/2016 Apr-18 5921-034-235 4476799 4/13/2016 Apr-18 5921-034-235 4477271 4/13/2016 Apr-18 5921-218-235 4472745 4/13/2016 Apr-18 5921-218-235 4474085 4/13/2016 Apr-18 5921-218-235 4475833 4/13/2016 Apr-18 5921-218-235 4476401 4/13/2016 Apr-18 5921-218-235 4476969 4/13/2016 Apr-18 5921-218-235 4477228 4/13/2016 Apr-18 5921-212-135 4478629 4/13/2016 Apr-18 5921-218-135 4474691 4/13/2016 Apr-18 5921-218-135 4477133 4/13/2016 Apr-18 60-7070-102 4473653 4/13/2016 Apr-18 60-7070-103 4472979 4/13/2016 Apr-18 60-7070-103 4473987 4/13/2016 Apr-18 60-7070-103 4474806 4/13/2016 Apr-18 60-7070-103 4476504 4/13/2016 Apr-18 60-7070-103 4476734 4/13/2016 Apr-18 60-7070-103 4477076 4/13/2016 Apr-18 60-7070-103 4477244 4/13/2016 Apr-18 60-7070-103 4478437 4/13/2016 Apr-18 60-7070-105 4473788 4/13/2016 Apr-18 60-7070-105 4474118 4/13/2016 Apr-18 60-7070-105 4474224 4/13/2016 Apr-18 60-7070-105 4475089 4/13/2016 Apr-18 60-7070-105 4476897 4/13/2016 Apr-18 60-7070-105 4477285 4/13/2016 Apr-18 60-7070-104 4473892 4/13/2016 Apr-18 60-7070-003 4477919 4/13/2016 Apr-18 60-7070-006 4478324 4/13/2016 Apr-18 60-7070-106 4473505 4/13/2016 Apr-18 60-7070-106 4475688 4/13/2016 Apr-18 60-7070-106 4476618 4/13/2016 Apr-18 60-7070-106 4478115 4/13/2016 Apr-18 60-7070-107 4476589 4/13/2016 Apr-18 60-7075-003 4475350 4/13/2016 Apr-18 60-7075-005 4473514 4/13/2016 Apr-18 60-7075-006 4472778 4/13/2016 Apr-18 60-7075-103 4473828 4/13/2016 Apr-18 60-7075-103 4474211 4/13/2016 Apr-18 60-7075-103 4475846 4/13/2016 Apr-18 60-7075-104 4474071 4/13/2016 Apr-18 60-7075-104 4476591 4/13/2016 Apr-18 60-7075-105 4475070 4/13/2016 Apr-18 60-7075-105 4477263 4/13/2016 Apr-18 60-7075-106 4474173 4/13/2016 Apr-18 60-7075-106 4475247 4/13/2016 Apr-18 60-7075-106 4477484 4/13/2016 Apr-18 5921-034-235 4386951 4/14/2016 Apr-18 5921-030-135 4483183 4/15/2016 Apr-18 5921-030-235 4483131 4/15/2016 Apr-18 5921-030-235 4485907 4/15/2016 Apr-18 5921-030-235 4486311 4/15/2016 Apr-18 5921-030-235 4486542 4/15/2016 Apr-18 5921-030-235 4487477 4/15/2016 Apr-18 5921-034-135 4483787 4/15/2016 Apr-18 5921-034-135 4484748 4/15/2016 Apr-18 5921-034-135 4486073 4/15/2016 Apr-18 5921-034-135 4486808 4/15/2016 Apr-18 5921-034-235 4483385 4/15/2016 Apr-18 5921-034-235 4485100 4/15/2016 Apr-18 5921-034-235 4486664 4/15/2016 Apr-18 5921-034-235 4487885 4/15/2016 Apr-18 5921-218-235 4483300 4/15/2016 Apr-18 5921-218-235 4483929 4/15/2016 Apr-18 5921-218-235 4487197 4/15/2016 Apr-18 5921-218-135 4486693 4/15/2016 Apr-18 60-7070-103 4484769 4/15/2016 Apr-18 60-7070-103 4484814 4/15/2016 Apr-18 60-7070-103 4486382 4/15/2016 Apr-18 60-7070-103 4487047 4/15/2016 Apr-18 60-7070-105 4483476 4/15/2016 Apr-18 60-7070-105 4483597 4/15/2016 Apr-18 60-7070-105 4484908 4/15/2016 Apr-18 60-7070-105 4486060 4/15/2016 Apr-18 60-7070-105 4486820 4/15/2016 Apr-18 60-7070-105 4487580 4/15/2016 Apr-18 60-7070-104 4483337 4/15/2016 Apr-18 60-7070-104 4485279 4/15/2016 Apr-18 60-7070-106 4483566 4/15/2016 Apr-18 60-7070-106 4484460 4/15/2016 Apr-18 60-7070-106 4485813 4/15/2016 Apr-18 60-7070-106 4486779 4/15/2016 Apr-18 60-7070-107 4486369 4/15/2016 Apr-18 60-7070-154 4487420 4/15/2016 Apr-18 60-7075-103 4485088 4/15/2016 Apr-18 60-7075-103 4487483 4/15/2016 Apr-18 60-7075-105 4484281 4/15/2016 Apr-18 60-7075-106 4483453 4/15/2016 Apr-18 60-7075-106 4486471 4/15/2016 Apr-18 5921-018-135 4490059 4/19/2016 Apr-18 5921-024-135 4492083 4/19/2016 Apr-18 5921-024-135 4493365 4/19/2016 Apr-18 5921-024-235 4492697 4/19/2016 Apr-18 5921-030-135 4487648 4/19/2016 Apr-18 5921-030-135 4489423 4/19/2016 Apr-18 5921-030-135 4492201 4/19/2016 Apr-18 5921-030-135 4492530 4/19/2016 Apr-18 5921-030-235 4488946 4/19/2016 Apr-18 5921-030-235 4489755 4/19/2016 Apr-18 5921-030-235 4491277 4/19/2016 Apr-18 5921-034-135 4489679 4/19/2016 Apr-18 5921-034-135 4490474 4/19/2016 Apr-18 5921-034-135 4492728 4/19/2016 Apr-18 5921-034-135 4493845 4/19/2016 Apr-18 5921-034-235 4489667 4/19/2016 Apr-18 5921-034-235 4491103 4/19/2016 Apr-18 5921-034-235 4492217 4/19/2016 Apr-18 5921-034-235 4492569 4/19/2016 Apr-18 5921-218-235 4489013 4/19/2016 Apr-18 5921-218-235 4489195 4/19/2016 Apr-18 5921-218-235 4489605 4/19/2016 Apr-18 5921-218-235 4489786 4/19/2016 Apr-18 5921-218-235 4492818 4/19/2016 Apr-18 5921-218-135 4488427 4/19/2016 Apr-18 5921-218-135 4489124 4/19/2016 Apr-18 5921-218-135 4489772 4/19/2016 Apr-18 5921-218-135 4492962 4/19/2016 Apr-18 60-7070-102 4489239 4/19/2016 Apr-18 60-7070-102 4492548 4/19/2016 Apr-18 60-7070-103 4487844 4/19/2016 Apr-18 60-7070-103 4489410 4/19/2016 Apr-18 60-7070-103 4489701 4/19/2016 Apr-18 60-7070-103 4489764 4/19/2016 Apr-18 60-7070-103 4490426 4/19/2016 Apr-18 60-7070-103 4492272 4/19/2016 Apr-18 60-7070-103 4492451 4/19/2016 Apr-18 60-7070-103 4492894 4/19/2016 Apr-18 60-7070-103 4493641 4/19/2016 Apr-18 60-7070-105 4489520 4/19/2016 Apr-18 60-7070-105 4490207 4/19/2016 Apr-18 60-7070-105 4490643 4/19/2016 Apr-18 60-7070-105 4492342 4/19/2016 Apr-18 60-7070-104 4489738 4/19/2016 Apr-18 60-7070-104 4490899 4/19/2016 Apr-18 60-7070-003 4488135 4/19/2016 Apr-18 60-7070-006 4488174 4/19/2016 Apr-18 60-7070-106 4488310 4/19/2016 Apr-18 60-7070-106 4489549 4/19/2016 Apr-18 60-7070-106 4489835 4/19/2016 Apr-18 60-7070-106 4491539 4/19/2016 Apr-18 60-7070-106 4493214 4/19/2016 Apr-18 60-7070-107 4488830 4/19/2016 Apr-18 60-7070-107 4492873 4/19/2016 Apr-18 60-7075-006 4487956 4/19/2016 Apr-18 60-7080-102 4489260 4/19/2016 Apr-18 60-7075-103 4491212 4/19/2016 Apr-18 60-7075-103 4492860 4/19/2016 Apr-18 60-7075-103 4494097 4/19/2016 Apr-18 60-7075-104 4493660 4/19/2016 Apr-18 60-7075-105 4492388 4/19/2016 Apr-18 60-7075-106 4488157 4/19/2016 Apr-18 60-7075-106 4490875 4/19/2016 Apr-18 5921-024-135 4496075 4/21/2016 Apr-18 5921-024-235 4496145 4/21/2016 Apr-18 5921-030-135 4496184 4/21/2016 Apr-18 5921-030-235 4496006 4/21/2016 Apr-18 5921-030-235 4497557 4/21/2016 Apr-18 5921-034-135 4495392 4/21/2016 Apr-18 5921-034-135 4495862 4/21/2016 Apr-18 5921-034-135 4496113 4/21/2016 Apr-18 5921-034-135 4496703 4/21/2016 Apr-18 5921-034-235 4494389 4/21/2016 Apr-18 5921-034-235 4495756 4/21/2016 Apr-18 5921-034-235 4495938 4/21/2016 Apr-18 5921-034-235 4496648 4/21/2016 Apr-18 5921-034-235 4498612 4/21/2016 Apr-18 5921-218-235 4492989 4/21/2016 Apr-18 5921-218-235 4494614 4/21/2016 Apr-18 5921-218-235 4495836 4/21/2016 Apr-18 5921-218-235 4498649 4/21/2016 Apr-18 5921-218-135 4495606 4/21/2016 Apr-18 5921-218-135 4496258 4/21/2016 Apr-18 60-7070-103 4494225 4/21/2016 Apr-18 60-7070-103 4495469 4/21/2016 Apr-18 60-7070-103 4495986 4/21/2016 Apr-18 60-7070-103 4496215 4/21/2016 Apr-18 60-7070-103 4496828 4/21/2016 Apr-18 60-7070-103 4497910 4/21/2016 Apr-18 60-7070-103 4498974 4/21/2016 Apr-18 60-7070-103 4499186 4/21/2016 Apr-18 60-7070-105 4494045 4/21/2016 Apr-18 60-7070-105 4495495 4/21/2016 Apr-18 60-7070-105 4496099 4/21/2016 Apr-18 60-7070-105 4496262 4/21/2016 Apr-18 60-7070-105 4497693 4/21/2016 Apr-18 60-7070-105 4498886 4/21/2016 Apr-18 60-7070-104 4495385 4/21/2016 Apr-18 60-7070-104 4496121 4/21/2016 Apr-18 60-7070-104 4499062 4/21/2016 Apr-18 60-7070-106 4493752 4/21/2016 Apr-18 60-7070-106 4496227 4/21/2016 Apr-18 60-7070-106 4497087 4/21/2016 Apr-18 60-7070-106 4499252 4/21/2016 Apr-18 60-7075-003 4497056 4/21/2016 Apr-18 60-7075-103 4495760 4/21/2016 Apr-18 60-7075-105 4494698 4/21/2016 Apr-18 60-7075-105 4497923 4/21/2016 Apr-18 60-7075-106 4494410 4/21/2016 Apr-18 60-7075-106 4497820 4/21/2016 Apr-18 60-7075-107 4498838 4/21/2016 Apr-18 5921-018-135 4499494 4/22/2016 Apr-18 5921-018-135 4500083 4/22/2016 Apr-18 5921-018-235 4503955 4/22/2016 Apr-18 5921-024-235 4500989 4/22/2016 Apr-18 5921-024-235 4502643 4/22/2016 Apr-18 5921-030-135 4498893 4/22/2016 Apr-18 5921-030-135 4499585 4/22/2016 Apr-18 5921-030-135 4501838 4/22/2016 Apr-18 5921-030-135 4503889 4/22/2016 Apr-18 5921-030-235 4499516 4/22/2016 Apr-18 5921-030-235 4500759 4/22/2016 Apr-18 5921-030-235 4502092 4/22/2016 Apr-18 5921-030-235 4503300 4/22/2016 Apr-18 5921-034-135 4498631 4/22/2016 Apr-18 5921-034-135 4499313 4/22/2016 Apr-18 5921-034-135 4500479 4/22/2016 Apr-18 5921-034-135 4502044 4/22/2016 Apr-18 5921-034-135 4503385 4/22/2016 Apr-18 5921-034-136 4502488 4/22/2016 Apr-18 5921-034-235 4499418 4/22/2016 Apr-18 5921-034-235 4500123 4/22/2016 Apr-18 5921-034-235 4501817 4/22/2016 Apr-18 5921-034-235 4502615 4/22/2016 Apr-18 5921-218-235 4382626 4/22/2016 Apr-18 5921-218-235 4499473 4/22/2016 Apr-18 5921-218-235 4502538 4/22/2016 Apr-18 5921-218-235 4503597 4/22/2016 Apr-18 5921-218-135 4499091 4/22/2016 Apr-18 5921-218-135 4500873 4/22/2016 Apr-18 5921-218-136 4500991 4/22/2016 Apr-18 60-7070-103 4499592 4/22/2016 Apr-18 60-7070-103 4502016 4/22/2016 Apr-18 60-7070-103 4502139 4/22/2016 Apr-18 60-7070-103 4502426 4/22/2016 Apr-18 60-7070-103 4502670 4/22/2016 Apr-18 60-7070-103 4504117 4/22/2016 Apr-18 60-7070-105 4501736 4/22/2016 Apr-18 60-7070-105 4502248 4/22/2016 Apr-18 60-7070-105 4503149 4/22/2016 Apr-18 60-7070-105 4503870 4/22/2016 Apr-18 60-7070-105 4504397 4/22/2016 Apr-18 60-7070-104 4501198 4/22/2016 Apr-18 60-7070-104 4502379 4/22/2016 Apr-18 60-7070-003 4501011 4/22/2016 Apr-18 60-7070-106 4500326 4/22/2016 Apr-18 60-7070-106 4501770 4/22/2016 Apr-18 60-7070-106 4503424 4/22/2016 Apr-18 60-7075-003 4502346 4/22/2016 Apr-18 60-7075-006 4501133 4/22/2016 Apr-18 60-7075-006 4504366 4/22/2016 Apr-18 60-7075-103 4499965 4/22/2016 Apr-18 60-7075-103 4502658 4/22/2016 Apr-18 60-7075-103 4503725 4/22/2016 Apr-18 60-7075-104 4502586 4/22/2016 Apr-18 60-7075-105 4504101 4/22/2016 Apr-18 60-7075-106 4499460 4/22/2016 Apr-18 60-7075-106 4501758 4/22/2016 Apr-18 60-7075-106 4502921 4/22/2016 Apr-18 5921-018-135 4339909 4/25/2016 Apr-18 5921-018-235 4505570 4/26/2016 Apr-18 5921-024-235 4505562 4/26/2016 Apr-18 5921-030-135 4509182 4/26/2016 Apr-18 5921-030-135 4510674 4/26/2016 Apr-18 5921-030-235 4504528 4/26/2016 Apr-18 5921-030-235 4505768 4/26/2016 Apr-18 5921-030-235 4506746 4/26/2016 Apr-18 5921-030-235 4509391 4/26/2016 Apr-18 5921-030-235 4510350 4/26/2016 Apr-18 5921-034-135 4508072 4/26/2016 Apr-18 5921-034-135 4508848 4/26/2016 Apr-18 5921-034-135 4509406 4/26/2016 Apr-18 5921-034-136 4506733 4/26/2016 Apr-18 5921-034-235 4505090 4/26/2016 Apr-18 5921-034-235 4505781 4/26/2016 Apr-18 5921-034-235 4505809 4/26/2016 Apr-18 5921-034-235 4508713 4/26/2016 Apr-18 5921-034-235 4508905 4/26/2016 Apr-18 5921-034-235 4509554 4/26/2016 Apr-18 5921-218-235 4505214 4/26/2016 Apr-18 5921-218-235 4505396 4/26/2016 Apr-18 5921-218-235 4508086 4/26/2016 Apr-18 5921-218-235 4509452 4/26/2016 Apr-18 5921-218-135 4507236 4/26/2016 Apr-18 5921-218-135 4509520 4/26/2016 Apr-18 5921-218-135 4510223 4/26/2016 Apr-18 60-7070-102 4510309 4/26/2016 Apr-18 60-7070-103 4505280 4/26/2016 Apr-18 60-7070-103 4505734 4/26/2016 Apr-18 60-7070-103 4505832 4/26/2016 Apr-18 60-7070-103 4507790 4/26/2016 Apr-18 60-7070-103 4509294 4/26/2016 Apr-18 60-7070-105 4505723 4/26/2016 Apr-18 60-7070-105 4505813 4/26/2016 Apr-18 60-7070-105 4507885 4/26/2016 Apr-18 60-7070-105 4509153 4/26/2016 Apr-18 60-7070-104 4505641 4/26/2016 Apr-18 60-7070-104 4506617 4/26/2016 Apr-18 60-7070-104 4508212 4/26/2016 Apr-18 60-7070-003 4505486 4/26/2016 Apr-18 60-7070-005 4505233 4/26/2016 Apr-18 60-7070-006 4505438 4/26/2016 Apr-18 60-7070-006 4505527 4/26/2016 Apr-18 60-7070-106 4507109 4/26/2016 Apr-18 60-7070-106 4507653 4/26/2016 Apr-18 60-7070-106 4508933 4/26/2016 Apr-18 60-7070-106 4509499 4/26/2016 Apr-18 60-7070-107 4504516 4/26/2016 Apr-18 60-7075-003 4509021 4/26/2016 Apr-18 60-7080-102 4505144 4/26/2016 Apr-18 60-7075-103 4507068 4/26/2016 Apr-18 60-7075-103 4509475 4/26/2016 Apr-18 60-7075-103 4509857 4/26/2016 Apr-18 60-7075-106 4510366 4/26/2016 Apr-18 5921-034-135 4385947 4/27/2016 Apr-18 5921-218-135 4497209 4/27/2016 Apr-18 60-7070-106 4497250 4/27/2016 Apr-18 5921-018-135 4515304 4/28/2016 Apr-18 5921-024-235 4511314 4/28/2016 Apr-18 5921-030-235 4513087 4/28/2016 Apr-18 5921-034-135 4512105 4/28/2016 Apr-18 5921-034-135 4512484 4/28/2016 Apr-18 5921-034-135 4513416 4/28/2016 Apr-18 5921-034-235 4512069 4/28/2016 Apr-18 5921-034-235 4513376 4/28/2016 Apr-18 5921-034-235 4515148 4/28/2016 Apr-18 5921-034-235 4515370 4/28/2016 Apr-18 5921-218-235 4511028 4/28/2016 Apr-18 5921-218-235 4512523 4/28/2016 Apr-18 5921-218-235 4512758 4/28/2016 Apr-18 5921-218-235 4515287 4/28/2016 Apr-18 5921-218-235 4515729 4/28/2016 Apr-18 5921-218-135 4510996 4/28/2016 Apr-18 5921-218-135 4511852 4/28/2016 Apr-18 5921-218-136 4511945 4/28/2016 Apr-18 60-7070-102 4512743 4/28/2016 Apr-18 60-7070-103 4510769 4/28/2016 Apr-18 60-7070-103 4512167 4/28/2016 Apr-18 60-7070-103 4512478 4/28/2016 Apr-18 60-7070-103 4513492 4/28/2016 Apr-18 60-7070-103 4514229 4/28/2016 Apr-18 60-7070-103 4515565 4/28/2016 Apr-18 60-7070-103 4515826 4/28/2016 Apr-18 60-7070-105 4511875 4/28/2016 Apr-18 60-7070-105 4512114 4/28/2016 Apr-18 60-7070-105 4512677 4/28/2016 Apr-18 60-7070-105 4513767 4/28/2016 Apr-18 60-7070-104 4511207 4/28/2016 Apr-18 60-7070-104 4512341 4/28/2016 Apr-18 60-7070-104 4515531 4/28/2016 Apr-18 60-7070-106 4510460 4/28/2016 Apr-18 60-7070-106 4511966 4/28/2016 Apr-18 60-7070-106 4512704 4/28/2016 Apr-18 60-7070-106 4513934 4/28/2016 Apr-18 60-7070-106 4515434 4/28/2016 Apr-18 60-7070-107 4514451 4/28/2016 Apr-18 60-7075-103 4512715 4/28/2016 Apr-18 60-7075-103 4512762 4/28/2016 Apr-18 60-7075-104 4512736 4/28/2016 Apr-18 60-7075-105 4512325 4/28/2016 Apr-18 60-7075-106 4514181 4/28/2016 Apr-18 5921-018-135 4518757 4/29/2016 Apr-18 5921-018-235 4518816 4/29/2016 Apr-18 5921-024-135 4518340 4/29/2016 Apr-18 5921-024-135 4518579 4/29/2016 Apr-18 5921-024-235 4516578 4/29/2016 Apr-18 5921-024-235 4518695 4/29/2016 Apr-18 5921-030-135 4514548 4/29/2016 Apr-18 5921-030-135 4519057 4/29/2016 Apr-18 5921-030-235 4515900 4/29/2016 Apr-18 5921-030-235 4518289 4/29/2016 Apr-18 5921-030-235 4518871 4/29/2016 Apr-18 5921-034-235 4516741 4/29/2016 Apr-18 5921-034-235 4517766 4/29/2016 Apr-18 5921-034-235 4518551 4/29/2016 Apr-18 5921-034-235 4519267 4/29/2016 Apr-18 5921-218-235 4516337 4/29/2016 Apr-18 5921-218-235 4519137 4/29/2016 Apr-18 5921-218-135 4517283 4/29/2016 Apr-18 60-7070-103 4515959 4/29/2016 Apr-18 60-7070-103 4517462 4/29/2016 Apr-18 60-7070-103 4518630 4/29/2016 Apr-18 60-7070-103 4519233 4/29/2016 Apr-18 60-7070-103 4520045 4/29/2016 Apr-18 60-7070-105 4515045 4/29/2016 Apr-18 60-7070-105 4515764 4/29/2016 Apr-18 60-7070-105 4517556 4/29/2016 Apr-18 60-7070-105 4519074 4/29/2016 Apr-18 60-7070-105 4520508 4/29/2016 Apr-18 60-7070-104 4518364 4/29/2016 Apr-18 60-7070-003 4517190 4/29/2016 Apr-18 60-7070-003 4519331 4/29/2016 Apr-18 60-7070-003 4519976 4/29/2016 Apr-18 60-7070-004 4517306 4/29/2016 Apr-18 60-7070-006 4519312 4/29/2016 Apr-18 60-7070-006 4520410 4/29/2016 Apr-18 60-7070-106 4516901 4/29/2016 Apr-18 60-7070-107 4519639 4/29/2016 Apr-18 60-7070-107 4520327 4/29/2016 Apr-18 60-7075-003 4517142 4/29/2016 Apr-18 60-7075-103 4515406 4/29/2016 Apr-18 60-7075-103 4517784 4/29/2016 Apr-18 60-7075-103 4519673 4/29/2016 Apr-18 60-7075-104 4515786 4/29/2016 Apr-18 60-7075-104 4515971 4/29/2016 Apr-18 60-7075-105 4515698 4/29/2016 Apr-18 60-7075-105 4515916 4/29/2016 Apr-18 60-7075-106 4516610 4/29/2016 Apr-18 5921-018-135 4521012 5/2/2016 May-18 5921-018-135 4521595 5/2/2016 May-18 5921-018-135 4522404 5/2/2016 May-18 5921-018-135 4524276 5/2/2016 May-18 5921-018-235 4520667 5/2/2016 May-18 5921-024-135 4522265 5/2/2016 May-18 5921-024-235 4522458 5/2/2016 May-18 5921-030-135 4524426 5/2/2016 May-18 5921-030-235 4522489 5/2/2016 May-18 5921-030-235 4523695 5/2/2016 May-18 5921-034-135 4521927 5/2/2016 May-18 5921-034-135 4526035 5/2/2016 May-18 5921-034-235 4520793 5/2/2016 May-18 5921-034-235 4521763 5/2/2016 May-18 5921-034-235 4522086 5/2/2016 May-18 5921-034-235 4523973 5/2/2016 May-18 5921-034-235 4525635 5/2/2016 May-18 5921-218-235 4520937 5/2/2016 May-18 5921-218-235 4521621 5/2/2016 May-18 5921-218-235 4524615 5/2/2016 May-18 5921-218-235 4526504 5/2/2016 May-18 5921-218-135 4525210 5/2/2016 May-18 60-7070-102 4522368 5/2/2016 May-18 60-7070-103 4521020 5/2/2016 May-18 60-7070-103 4523546 5/2/2016 May-18 60-7070-103 4524852 5/2/2016 May-18 60-7070-103 4526365 5/2/2016 May-18 60-7070-105 4521678 5/2/2016 May-18 60-7070-105 4522157 5/2/2016 May-18 60-7070-105 4522283 5/2/2016 May-18 60-7070-105 4525206 5/2/2016 May-18 60-7070-105 4525908 5/2/2016 May-18 60-7070-104 4436012 5/2/2016 May-18 60-7070-104 4522436 5/2/2016 May-18 60-7070-104 4526486 5/2/2016 May-18 60-7070-003 4522038 5/2/2016 May-18 60-7070-006 4522220 5/2/2016 May-18 60-7070-106 4521904 5/2/2016 May-18 60-7070-106 4522415 5/2/2016 May-18 60-7070-106 4524282 5/2/2016 May-18 60-7070-106 4526057 5/2/2016 May-18 60-7070-106 4526312 5/2/2016 May-18 60-7075-006 4521744 5/2/2016 May-18 60-7075-103 4522443 5/2/2016 May-18 60-7075-103 4523742 5/2/2016 May-18 60-7075-103 4526088 5/2/2016 May-18 60-7075-104 4526464 5/2/2016 May-18 60-7075-106 4522502 5/2/2016 May-18 60-7075-106 4526438 5/2/2016 May-18 5921-018-135 4530253 5/4/2016 May-18 5921-024-135 4529177 5/4/2016 May-18 5921-030-135 4528672 5/4/2016 May-18 5921-030-235 4526878 5/4/2016 May-18 5921-030-235 4527259 5/4/2016 May-18 5921-030-235 4530305 5/4/2016 May-18 5921-034-135 4519982 5/4/2016 May-18 5921-034-135 4527228 5/4/2016 May-18 5921-034-135 4529652 5/4/2016 May-18 5921-034-235 4528213 5/4/2016 May-18 5921-218-235 4527919 5/4/2016 May-18 5921-218-235 4528136 5/4/2016 May-18 5921-218-135 4527727 5/4/2016 May-18 5921-218-135 4529993 5/4/2016 May-18 60-7070-102 4529183 5/4/2016 May-18 60-7070-103 4528877 5/4/2016 May-18 60-7070-103 4529131 5/4/2016 May-18 60-7070-103 4529290 5/4/2016 May-18 60-7070-103 4530141 5/4/2016 May-18 60-7070-105 4528896 5/4/2016 May-18 60-7070-104 4529022 5/4/2016 May-18 60-7070-104 4529316 5/4/2016 May-18 60-7070-005 4528590 5/4/2016 May-18 60-7070-106 4527484 5/4/2016 May-18 60-7070-106 4527829 5/4/2016 May-18 60-7070-106 4529203 5/4/2016 May-18 60-7070-106 4530643 5/4/2016 May-18 60-7070-107 4527133 5/4/2016 May-18 60-7075-003 4529261 5/4/2016 May-18 60-7075-103 4526493 5/4/2016 May-18 60-7075-103 4528906 5/4/2016 May-18 60-7075-105 4526911 5/4/2016 May-18 60-7075-105 4529005 5/4/2016 May-18 60-7075-106 4528170 5/4/2016 May-18 60-7075-106 4530749 5/4/2016 May-18 5921-018-135 4536355 5/5/2016 May-18 5921-018-235 4536412 5/5/2016 May-18 5921-024-135 4531980 5/5/2016 May-18 5921-024-135 4532553 5/5/2016 May-18 5921-024-136 4534358 5/5/2016 May-18 5921-024-235 4536099 5/5/2016 May-18 5921-030-135 4530966 5/5/2016 May-18 5921-030-135 4534554 5/5/2016 May-18 5921-030-135 4537104 5/5/2016 May-18 5921-030-136 4535079 5/5/2016 May-18 5921-030-235 4533144 5/5/2016 May-18 5921-034-135 4532388 5/5/2016 May-18 5921-034-135 4533205 5/5/2016 May-18 5921-034-135 4535647 5/5/2016 May-18 5921-034-135 4536121 5/5/2016 May-18 5921-034-235 4530852 5/5/2016 May-18 5921-034-235 4533543 5/5/2016 May-18 5921-034-235 4535725 5/5/2016 May-18 5921-034-235 4536863 5/5/2016 May-18 5921-218-235 4529862 5/5/2016 May-18 5921-218-235 4532613 5/5/2016 May-18 5921-218-235 4533813 5/5/2016 May-18 5921-218-235 4535929 5/5/2016 May-18 5921-218-235 4536301 5/5/2016 May-18 5921-218-135 4531644 5/5/2016 May-18 5921-218-135 4532927 5/5/2016 May-18 5921-218-135 4534294 5/5/2016 May-18 5921-218-236 4536197 5/5/2016 May-18 60-7070-102 4536243 5/5/2016 May-18 60-7070-103 4532609 5/5/2016 May-18 60-7070-103 4535080 5/5/2016 May-18 60-7070-103 4535550 5/5/2016 May-18 60-7070-103 4536340 5/5/2016 May-18 60-7070-105 4529668 5/5/2016 May-18 60-7070-105 4531093 5/5/2016 May-18 60-7070-105 4532498 5/5/2016 May-18 60-7070-105 4534512 5/5/2016 May-18 60-7070-105 4536010 5/5/2016 May-18 60-7070-104 4535610 5/5/2016 May-18 60-7070-003 4536262 5/5/2016 May-18 60-7070-106 4532569 5/5/2016 May-18 60-7070-106 4534097 5/5/2016 May-18 60-7075-006 4532134 5/5/2016 May-18 60-7075-006 4535261 5/5/2016 May-18 60-7075-102 4534084 5/5/2016 May-18 60-7075-104 4533781 5/5/2016 May-18 60-7075-105 4533948 5/5/2016 May-18 5921-018-135 4542702 5/6/2016 May-18 5921-018-235 4532422 5/6/2016 May-18 5921-018-235 4538654 5/6/2016 May-18 5921-024-135 4538024 5/6/2016 May-18 5921-024-235 4531600 5/6/2016 May-18 5921-024-235 4532083 5/6/2016 May-18 5921-024-235 4539252 5/6/2016 May-18 5921-024-235 4539544 5/6/2016 May-18 5921-030-135 4541331 5/6/2016 May-18 5921-030-235 4537056 5/6/2016 May-18 5921-030-235 4538903 5/6/2016 May-18 5921-030-235 4540278 5/6/2016 May-18 5921-030-235 4541775 5/6/2016 May-18 5921-034-135 4536533 5/6/2016 May-18 5921-034-135 4537297 5/6/2016 May-18 5921-034-135 4540158 5/6/2016 May-18 5921-034-235 4532175 5/6/2016 May-18 5921-034-235 4538179 5/6/2016 May-18 5921-034-235 4540526 5/6/2016 May-18 5921-218-235 4539500 5/6/2016 May-18 5921-218-235 4541308 5/6/2016 May-18 5921-030-236 4538762 5/6/2016 May-18 5921-218-135 4540981 5/6/2016 May-18 5921-218-136 4537849 5/6/2016 May-18 60-7070-103 4531732 5/6/2016 May-18 60-7070-103 4537865 5/6/2016 May-18 60-7070-103 4538581 5/6/2016 May-18 60-7070-103 4539172 5/6/2016 May-18 60-7070-103 4539537 5/6/2016 May-18 60-7070-103 4541562 5/6/2016 May-18 60-7070-105 4537672 5/6/2016 May-18 60-7070-105 4539418 5/6/2016 May-18 60-7070-105 4539425 5/6/2016 May-18 60-7070-105 4539571 5/6/2016 May-18 60-7070-105 4541493 5/6/2016 May-18 60-7070-104 4531692 5/6/2016 May-18 60-7070-104 4539281 5/6/2016 May-18 60-7070-004 4531946 5/6/2016 May-18 60-7070-006 4539062 5/6/2016 May-18 60-7070-106 4537266 5/6/2016 May-18 60-7070-106 4538957 5/6/2016 May-18 60-7070-106 4540779 5/6/2016 May-18 60-7070-107 4537883 5/6/2016 May-18 60-7075-103 4530207 5/6/2016 May-18 60-7075-103 4537557 5/6/2016 May-18 60-7075-103 4539563 5/6/2016 May-18 60-7075-103 4541747 5/6/2016 May-18 60-7075-103 4542495 5/6/2016 May-18 60-7075-104 4539456 5/6/2016 May-18 60-7075-105 4540672 5/6/2016 May-18 60-7075-106 4538872 5/6/2016 May-18 60-7075-106 4542177 5/6/2016 May-18 60-7075-107 4537018 5/6/2016 May-18 5921-024-135 4545356 5/9/2016 May-18 5921-024-235 4542165 5/9/2016 May-18 5921-024-235 4546147 5/9/2016 May-18 5921-030-135 4544355 5/9/2016 May-18 5921-030-235 4543564 5/9/2016 May-18 5921-034-135 4542938 5/9/2016 May-18 5921-034-135 4544757 5/9/2016 May-18 5921-034-135 4545665 5/9/2016 May-18 5921-034-135 4546943 5/9/2016 May-18 5921-034-235 4542385 5/9/2016 May-18 5921-034-235 4543881 5/9/2016 May-18 5921-034-235 4546507 5/9/2016 May-18 5921-034-235 4547516 5/9/2016 May-18 5921-218-235 4543141 5/9/2016 May-18 5921-218-235 4545770 5/9/2016 May-18 5921-218-235 4546451 5/9/2016 May-18 5921-030-236 4544788 5/9/2016 May-18 5921-218-135 4543139 5/9/2016 May-18 5921-218-135 4546511 5/9/2016 May-18 60-7070-103 4543063 5/9/2016 May-18 60-7070-103 4544465 5/9/2016 May-18 60-7070-103 4545677 5/9/2016 May-18 60-7070-103 4546342 5/9/2016 May-18 60-7070-103 4546422 5/9/2016 May-18 60-7070-103 4547441 5/9/2016 May-18 60-7070-105 4542760 5/9/2016 May-18 60-7070-105 4543194 5/9/2016 May-18 60-7070-105 4544844 5/9/2016 May-18 60-7070-105 4545961 5/9/2016 May-18 60-7070-105 4546433 5/9/2016 May-18 60-7070-104 4543028 5/9/2016 May-18 60-7070-104 4545903 5/9/2016 May-18 60-7070-104 4546405 5/9/2016 May-18 60-7070-003 4547036 5/9/2016 May-18 60-7070-106 4542986 5/9/2016 May-18 60-7070-106 4544145 5/9/2016 May-18 60-7070-106 4546229 5/9/2016 May-18 60-7070-106 4547707 5/9/2016 May-18 60-7070-107 4545552 5/9/2016 May-18 60-7075-005 4546357 5/9/2016 May-18 60-7075-104 4543071 5/9/2016 May-18 60-7075-104 4546446 5/9/2016 May-18 60-7075-105 4544132 5/9/2016 May-18 60-7075-106 4543627 5/9/2016 May-18 5921-018-135 4551428 5/11/2016 May-18 5921-018-135 4552622 5/11/2016 May-18 5921-024-135 4548256 5/11/2016 May-18 5921-024-135 4552336 5/11/2016 May-18 5921-024-235 4552729 5/11/2016 May-18 5921-030-135 4547473 5/11/2016 May-18 5921-030-135 4550381 5/11/2016 May-18 5921-030-235 4547155 5/11/2016 May-18 5921-030-235 4549933 5/11/2016 May-18 5921-030-235 4552066 5/11/2016 May-18 5921-030-235 4553043 5/11/2016 May-18 5921-034-135 4549732 5/11/2016 May-18 5921-034-135 4552786 5/11/2016 May-18 5921-034-235 4549103 5/11/2016 May-18 5921-034-235 4549713 5/11/2016 May-18 5921-034-235 4550212 5/11/2016 May-18 5921-034-235 4552148 5/11/2016 May-18 5921-218-235 4548835 5/11/2016 May-18 5921-218-235 4549220 5/11/2016 May-18 5921-218-235 4549659 5/11/2016 May-18 5921-218-235 4550967 5/11/2016 May-18 5921-218-235 4552475 5/11/2016 May-18 5921-218-135 4548717 5/11/2016 May-18 5921-218-135 4549616 5/11/2016 May-18 5921-218-135 4551002 5/11/2016 May-18 60-7070-103 4548202 5/11/2016 May-18 60-7070-103 4549038 5/11/2016 May-18 60-7070-103 4549368 5/11/2016 May-18 60-7070-103 4549556 5/11/2016 May-18 60-7070-103 4551463 5/11/2016 May-18 60-7070-103 4552614 5/11/2016 May-18 60-7070-105 4548327 5/11/2016 May-18 60-7070-105 4548928 5/11/2016 May-18 60-7070-105 4549709 5/11/2016 May-18 60-7070-105 4551348 5/11/2016 May-18 60-7070-105 4552679 5/11/2016 May-18 60-7070-104 4549541 5/11/2016 May-18 60-7070-104 4552391 5/11/2016 May-18 60-7070-003 4550462 5/11/2016 May-18 60-7070-106 4547977 5/11/2016 May-18 60-7070-106 4549296 5/11/2016 May-18 60-7070-106 4550685 5/11/2016 May-18 60-7075-006 4549142 5/11/2016 May-18 60-7075-006 4550616 5/11/2016 May-18 60-7075-103 4552273 5/11/2016 May-18 60-7075-104 4552050 5/11/2016 May-18 60-7075-105 4549525 5/11/2016 May-18 60-7075-106 4547912 5/11/2016 May-18 60-7075-106 4550797 5/11/2016 May-18 60-7075-106 4552130 5/11/2016 May-18 |
Recalling Firm/ Manufacturer |
Stryker Sustainability Solutions 1810 W Drake Dr Tempe AZ 85283-4327
|
For Additional Information Contact | 888-888-3433 |
Manufacturer Reason for Recall | Issue with the process for detection of leaking Pressure Tourniquet Cuffs (PTC). Leaking PTC devices may result in excessive blood loss requiring surgical or medical intervention or systemic leakage of local anesthetic during Bier Block requiring medical intervention. |
FDA Determined Cause 2 | Device Design |
Action | Stryker Sustainability Solutions sent an Customer Notification letter September 19, 2016.
Customers were instructed to discontinue use of the affected products. The letter stated to follow the link below to complete the Recall Effectiveness Check Form and indicate if any devices containing expiration dates from May 2016-May 2018 remain in inventory. This form must be completed even if no affected product is found.
Use the following reference number as you complete the form: [Ship to Account #]
If the form indicates that any affected devices remain in inventory, a prepaid shipping label will be issued for the return of the product. Credit for all affected devices returned will be issued to the facility.
.Adverse reactions or quality problems experienced with the use of this product may be reported to:
o Stryker Sustainability Solutions Complaint Hotline: +1(888) 888-3433 X5555
o http://www.stryker.com/productexperience/
o The FDAs MedWatch Adverse Event Reporting program either online, by
regular mail, or by fax. |
Quantity in Commerce | 1,695,760 |
Distribution | U.S. distribution nationwide.
No foreign distribution |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|