 Class 2 Device Recall ALIII W/ male disconnect and 20ft hosesee related information
Date Initiated by FirmOctober 06, 2016
Create DateMay 04, 2017
Recall Status1 Terminated 3 on July 26, 2019
Recall NumberZ-1931-2017
Recall Event ID 76059
510(K)NumberK011444 
Product Classification Motor, surgical instrument, ac-powered - Product Code GEY
ProductAL-III W/ male disconnect and 20ft hose Product Usage: Pneumatic system
Code Information All product shipped prior to 09/27/2016;  DFU 18-0040; Part No. AL-III-DISS-20
FEI Number 1045834
Recalling Firm/
Manufacturer
The Anspach Effort, Inc.
4500 Riverside Dr
Palm Beach Gardens FL 33410-4235
For Additional Information ContactAnne Brisson
800-327-6687
Manufacturer Reason
for Recall
Supplied Directions for Use (DFU) did not contain a recommended service interval.
FDA Determined
Cause 2
Under Investigation by firm
ActionConsignees were sent a recall notice via mail, dated 10/10/2016. In the notice, the firm notifies consignees of the Directions For Use (DFU) correction that will include recommended servicing intervals. All DFUs distributed after 09/28/2016 will have the correct DFU. The recalling firm instructs consignees to review the service intervals provided for product received, forward the notification to necessary parties and/or facilities, and complete the attached Verification Form. Complete Verification Forms are to be sent by fax to 877-546-5069 or by email to Anspach4870@stericycle.com. Consignees are also asked to maintain a copy of the notice for their records. Any questions can be directed to Customer Service at 1-800-327-6887, option 1.
Quantity in Commerce19
DistributionWorldwide Distribution - US Nationwide in the states of HI, DC including PR and the countries of Guam and Japan
Total Product Life CycleTPLC Device Report