| | Class 1 Device Recall LIFEPAK 15 Monitor/Defibrillator |  |
| Date Initiated by Firm | January 12, 2017 |
| Date Posted | February 11, 2017 |
| Recall Status1 |
Terminated 3 on September 01, 2017 |
| Recall Number | Z-1123-2017 |
| Recall Event ID |
76160 |
| 510(K)Number | K142430 |
| Product Classification |
Automated external defibrillators (non-wearable) - Product Code MKJ
|
| Product | LIFEPAK 15 Monitor, Automated External Defibrillator (AED), Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older. |
| Code Information |
SERIAL NUMBERS: 42738872, 44554102, 42701550, 42702539, 42730815, 42800549, 42803986, 42862109, 42884428, 42944609, 42989327, 43022460, 43040692, 43048791, 43066974, 43116838, 43116994, 43119477, 43132954, 43133203, 43133221, 43133906, 43154308, 43154457, 43156695, 43161425, 43163930, 43164076, 43164158, 43166133, 43206034, 43210019, 43219622, 43226291, 43230032, 43232487, 43232904, 43247389, 43251050, 43253506, 43253916, 43254032, 43258901, 43293290, 43321556, 43388132, 43388137, 43388541, 43405542, 43407947, 43409346, 43429508, 43465372, 43472036, 43472591, 43480824, 43489606, 43492883, 43576021, 43616526, 43641893, 43691775, 43715563, 43735849, 43764249, 43800357, 43805213, 43825370, 43826635, 43843347, 43869096, 43884778, 43922984, 43938506, 43947198, 43962139, 43964195, 43968233, 43981293, 43999541, 44010117, 44010465, 44039473, 44056838, 44060132, 44072885, 44073149, 44090669, 44092118, 44094122, 44108545, 44112183, 44117966, 44150897, 44167052, 44168042, 44179750, 44199424, 44225054, 44225258, 44226227, 44226585, 44226654, 44231094, 44233618, 44234803, 44234840, 44237875, 44255648, 44262283, 44262826, 44264102, 44264573, 44265421, 44271391, 44283558, 44287000, 44289115, 44293297, 44303302, 44312782, 44315479, 44327764, 44342010, 44352664, 44372343, 44416667, 44435825, 44438567, 44439898, 44449825, 44450073, 44452220, 44458445, 44460124, 44460749, 44463853, 44466514, 44473363, 44483809, 44486872, 44487784, 44487992, 44488101, 44489681, 44492189, 44492223, 44492252, 44492760, 44492832, 44494129, 44496734, 44497169, 44497196, 44497761, 44497866, 44498576, 44500827, 44500943, 44501037, 44501047, 44501112, 44501201, 44501465, 44501806, 44501832, 44502009, 44502159, 44502179, 44502310, 44502686, 44502718, 44502776, 44502815, 44502943, 44503096, 44508212, 44508273, 44508330, 44508331, 44508358, 44508440, 44508498, 44508645, 44508868, 44509074, 44509520, 44509538, 44511211, 44511231, 44511273, 44511519, 44511573, 44512604, 44513321, 44513358, 44518244, 44518479, 44518497, 44518703, 44518786, 44518917, 44518981, 44519083, 44519095, 44519255, 44519301, 44519307, 44519381, 44519548, 44524967, 44525130, 44525416, 44527324, 44530479, 44530594, 44530719, 44530800, 44531488, 44531803, 44532088, 44534804, 44534900, 44534991, 44535111, 44535241, 44535452, 44535582, 44535643, 44536176, 44536780, 44536823, 44536897, 44538798, 44538819, 44539781, 44539880, 44541779, 44541917, 44548263, 44548530, 44548558, 44548749, 44548825, 44549064, 44549220, 44549413, 44549964, 44550150, 44550234, 44550467, 44550552, 44550575, 44550613, 44550642, 44550675, 44552029, 44552066, 44552865, 44553122, 44554175, 44554259, 44554351, 44554520, 44554837, 44554941, 44555058, 44555261, 44556688, 44557824, 44558175, 44558252, 44558385, 44559077, 44559369, 44559455, 44560001, 44560185, 44561655, 44561903, 44562074, 44562559, 44562662, 44562754, 44563762, 44564474, 44564491, 44571952, 44572025, 44572171, 44572228, 44572357, 44572374, 44572595, 44572865, 44573101, 44573148, 44573485, 44573956, 44574467, 44574766, 44574771, 44574821, 44574929, 44574939, 44575793, 44575871, 44575898, 44576359, 44576647, 44577004, 44577022, 44577116, 44577219, 44578054, 44578080, 44578192, 44578272, 44579036, 44579299, 44579496, 44579697, 44579713, 44579857, 44581304, 44581382, 44583260, 44583657, 44583701, 44584213, 44586462, 44587097, 44587099, 44587181, 44587302, 44587358, 44587899, and 44587976. |
Recalling Firm/ Manufacturer |
Physio-Control, Inc. 11811 Willows Rd NE Redmond WA 98052-2003
|
Manufacturer Reason for Recall | The firm has become aware of a potential device issue where the LIFEPAK 15 device may no longer be able to deliver defibrillation energy due to a potential failure of an internal component (Relay) installed on the Therapy Printed Circuit Board Assembly (PCBA). |
FDA Determined Cause 2 | Other |
| Action | Physio plans to send the URGENT MEDICAL DEVICE CORRECTION ACTION REQUIRED LIFEPAK 15 Monitor/Defibrillator letter, dated January 2017, to their consignees on January 12, 2017.
They were informed of the problem and how to identify a malfunction during the
User Test of the device as described in the chapter titled Maintaining the Equipment within the LIFEPAK 15 Operating Instructions (or follow page 2 of the notification letter).
If the User Test fails, the device will illuminate the Service indicator on the device. Please contact Physio-Control immediately to schedule the device correction. This field corrective action will include replacement of the Therapy PCB assembly on all devices included in the potentially affected population. This will occur as a combination of field and depot service.
Customers are required to take the following actions:
1. Please forward this letter to all of your sites, trainers and users that have an affected LIFEPAK 15 device as identified on the attached Confirmation Sheet.
2. Follow the instructions on the Confirmation Sheet for each serial number listed in your possession. Promptly return the completed Confirmation Sheet to Physio-Control.
3. If the device fails the User Test, as described above, contact Physio-Control immediately to arrange for correction of your device.
4. Continue to perform the User Test as outlined in the Daily Operators Checklist within the LIFEPAK 15 Operating Instructions.
Consignees should return completed confirmation sheet to Physio:
-By fax to: 1-866-448-9567
-By email to: rsrecall@physio-control.com
-Or by mail to: Physio-Control, Inc. P.O. Box 97006, Dept. 15N Redmond, WA, 98073-9706.
Customers with questions call the firm at 1-800-442-1142, option 7, 6:00 A.M. to 4:00 P.M. (PST), Monday Friday. |
| Quantity in Commerce | 338 units total (US =139 units; International = 199 units) |
| Distribution | Worldwide Distribution-US ( nationwide) in states of: AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TX, VA, WA, WI, and WV and countries of: Australia, Brazil, Canada, Czech Republic, Denmark, Finland, France, French Guiana, Germany, India, Ireland, Israel, Italy, Japan, Latvia, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Qatar, Saudi Arabia, Slovenia, South Africa, Sweden, Turkey, United Arab Emirates, and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MKJ
|
|
|
|