| | Class 2 Device Recall Table Patient Step on RFX/SFX, Legacy, and Precision 500D systems |  |
| Date Initiated by Firm | February 06, 2017 |
| Date Posted | May 22, 2017 |
| Recall Status1 |
Terminated 3 on March 02, 2018 |
| Recall Number | Z-2179-2017 |
| Recall Event ID |
76517 |
| 510(K)Number | K011624 K081091 K760466 K801320 K862120 K973039 |
| Product Classification |
Table, radiographic, tilting - Product Code IXR
|
| Product | Table Patient Step on RFX/SFX, Legacy, and Precision 500D systems |
| Code Information |
FMI 10908 Initial Consignee List Mfg. Lot or Serial # System ID 00000959002WK7 907276RM4 00000966338WK6 256235ORF1 00000966611WK6 256235ORF2 00000956387WK5 205783LRF2 00000525674WK8 205793FR1 00000964963WK3 205387LRF 964349WK5 334732SRF 00000969028WK0 205345RF3 952032WK1 501968LEG 00000510900WK4 501863MLEG 00000277058TU4 479441RF1 00000974520WK9 501321LEG1 29415VP7 928763RF 00000543656WK3 602856LEG 00000551387WK4 602994DRS 984281WK6 928537RM1 00000983374WK0 928537RM2 00000960138WK6 209386ATWRF 00000130329TX6 RADNETRF119 00000546581WK0 650696M6 00000978807WK6 818884LRF1 00000023282M37 RADNETRF1260 00000499668WK2 818331SFX 538693WK3 RADNETRF118 00000972384WK2 RADNETRF144 00000962460WK2 RADNETRF130 00000980889WK0 RADNETRF193 00000530446WK4 209225LEG2 00000531588WK2 209225LEG3 00000517706WK8 510370RFDG1 00000518141WK7 510370RFDG2 00000968783WK1 RADNETRF138 00000963491WK6 RADNETRF137 00000961786WK1 925370LMRF1 00000973156WK3 RADNETRF146 00000953812WK5 RADNETRF126 00000522090WK0 760446RC1 00000969297WK1 909788LF3 00000967126WK4 909788LF4 1025367WK2 RADNETRF129 00000978314WK3 916784LEG 00000978316WK8 619543RF4 00000975431WK8 619543RF5 00000974139WK8 415923LEGACYGI 00000547501WK7 415923PR5 00000963305WK8 408977RF5 00000552931WK8 RADNETRF122 00000966068WK9 RADNETRF139 00001005270WK2 RADNETRF143 00000523402WK6 707462ULEGACY 00000972383WK4 303440BCHRM4 986114WK7 303440LEG1 986513WK0 303355RF 00000503355WK0 970945RM2 00000959073WK8 303CMCRM1 00000954518WK7 719556RM1 00000977456WK3 203331RF1 00000507625WK2 203785RF1 972381WK8 860350RF1 00000543348WK7 703697RF2 00000967162WK9 302392RF 00000961102WK1 302654RF 00000549578WK3 407830RF4 00000956746WK2 561965RM1 00000957833WK7 561965RM3 00000958809WK6 561965RM4 00000538300WK5 305682RM5 546369WK0 305AHLEG2 00000549923WK1 863533RM1 00000963738WK0 941798RM2 00000981461WK7 941798RM3 00000516439WK7 813681RM3 00000519515WK1 813681SPEC1 00000518140WK9 407566CH2 00000979197WK1 239432RF1 00000969627WK9 239931RM2 00000955947WK7 941437RF1 00000966773WK4 561468RM4 00000550144WK0 850862FR3 00000980499WK8 352344RFLEG 00000112375BU1 904755LRAD 00000974170WK3 239369RM3 387184WK5 386362RF1 87793TY6 321752WLG2 00000972386WK7 850626SR4 00000951422WK5 352351RM2 00000970853WK8 352861RF1 00000971022WK9 352861RF2 00000538689WK1 407896RF5 00000529630WK6 407933FL1 00000524475WK1 904672LGY1 00000529192WK7 904672LGY2 00000963293WK6 850747ER1 00000551893WK1 850111RF1 00000506731WK9 904478WF5 00000531358WK0 941629RF2 00000974525WK8 321752WLG1 00000962369WK5 904819RF2 00000534305WK8 904825LEGACY 00000974350WK1 727341RF2 00000969628WK7 727341RF3 00000969619WK6 813634LEG1 00000975429WK2 813634LEG2 00000978458WK8 850325RF1 00000962420WK6 850656RF2 00000969308WK6 954724LEG 00000979786WK1 561798WRM2 34489VP5 403179 37019VP7 403009 402568BU0 403050 532577WK4 403219 Not Available 403258 00000527085WK5 912246MR3 00000530448WK0 912264LEGACY 00000973188WK6 770382LEG2 00000544942WK6 229758LEGACY 00000956847WK8 706576HL1 00000522605WK5 404571DL2 00000970377WK8 404321LEG2 00000966070WK5 706754HR2 00000963310WK8 478994MR1 967884WK8 403294 00000518884WK2 478633DH3 00000516878WK6 478633R1 00000953103WK9 478633R3 00000532968WK5 478633R4 00000963898WK2 912757RF2 00000955368WK6 912765RMD Not Available 1263640 00000541503WK9 912327LD3 000FMITR002796 1334666 Not Available 1186144 00000963795WK0 706554BFL1 00000501270WK3 404867BRF1 00000519905WK4 712336RF 00000972819WK7 208882GX2A 00000552613WK2 815933RF6 00000976869WK8 773356LEG1 00000969285WK6 815284RM2 00000547503WK3 847570RF6 00000973835WK2 847913RF1 00000957800WK6 847657GR4 00000510837WK8 847362CAL 00000497794WK8 708425LD 00000983490WK4 217562PLEG 00000960280WK6 309662JX3 00000968048WK9 618443SRF1 00000978315WK0 864560BJWR1 00000519907WK0 217337CXF1A 00000954100WK4 217337CXF2 00000506734WK3 217383CRF1 00000975577WK8 260425SJRF6 00000003830BC7 713984SB5 00000552619WK9 812482LRF 00000969574WK3 812237TH3 00000970714WK2 785877RF 00000502618WK2 913345MLEG1 544044WK1 913676SR9 00000532969WK3 913317SRM3 00000960804WK3 785543PRF 00000541727WK4 316688RF4 00000972388WK3 620221RF1 00000536174WK6 502875RM2 00000969865WK5 859255RF1 00000972380WK0 606285LEGACY 536550WK7 606256PRE 00000495060WK6 606789XR1 00000962630WK0 606987LEG1 00000955150WK8 606451LEG1 00000951414WK2 318473RD2 00000967364WK1 985748RF 00000982419WK4 985652LEG 00000976993WK6 985882RF1 00000522608WK9 318981MCADVE 00000957655WK4 337981DRS4 00000970749WK8 504885LEG 00000957419WK5 318728RF1 00000543346WK1 318539LEG 00000971903WK0 504391RF1 00000971825WK5 504391RF2 960074WK3 781848RF1 00000538308WK8 781273L2 962275WK4 413748RM5 00000538934WK1 781756W4 00000534769WK5 301981RF5 549581WK7 RADNETRF1241 00000967770WK9 301896RF 529942WK5 RADNETRF1271 00000978808WK4 207255RF2 00000954383WK6 517269RF1 00000953219WK3 517269RF3 00000979585WK7 231547ADV 988710WK0 313593LEG1 00000978805WK0 313593LEG2 00000965500WK2 517353LEGACY 00000957454WK2 517663LEGACY 00000524394WK4 906524ADV1 00000975421WK9 231845RMB 00000975428WK4 FHEOLG1 00000951592WK5 810989RF5 985400WK1 313295LEG 00000951400WK1 507345LEG 00000964575WK5 507332RF 00000969537WK0 507372LEG 00000983757WK6 417235CMRF 00000982589WK4 417269NX4 00000977149WK4 417882CX6 00000533552WK6 417885SXL1 00000978809WK2 417887SNI 00000973834WK5 662627RF4 00000966516WK7 601933LEG 00000551900WK4 228575XR1 00000553331WK0 228575XR2 00000553330WK2 228575XR3 00000977461WK3 601200RM5 00000967778WK2 601364ARF3 00000956274WK5 601364RM2 00000973233WK0 662289RF1 00000982421WK0 318336LEG 00000975683WK4 228809SR2 00000967161WK1 662324RF1 00000973839WK4 601735RF2 00000959586WK9 406538LEG 00000986314WK3 406826RM1 00000552316WK2 919575XR50 00000552933WK4 828452RF1 00000529195WK0 919736SJL1 00000549921WK5 919736SJL2 00000966562WK1 919934JADV1 00000966599WK3 919934JADV2 00000971396WK7 704873DLEG4 00000963221WK7 252641HRLEG 522609WK7 336794MRF1 00000966024WK2 402367RF 00000970543WK5 308282ADV1 00000976667WK6 402887RF 00000977148WK6 402376RF 964350WK3 732840RF1 00000974903WK7 732321LEG6 956777WK7 RADNETRF1266 00000519519WK3 908821LEG 00000972817WK1 201418RF 00000977745WK9 609953LEG1 00000975653WK7 856778RF2 00000969405WK0 732970LEGACY 526628WK0 856582OPRF 00000973623WK2 732505RF 00000979788WK7 505727RM2 00000972167WK1 505722XRM1 00000971607WK7 505722XRM2 00000964482WK4 775738RM2 00000951165WK0 INSIGHTXR014 00000966674WK4 702731SRF13 00000966673WK6 702731SRF7 00000954233WK3 775982ADV1 00000955304WK1 775982XR2 00000973714WK9 775982RF 00000514169WK2 516968LEG1 00000530160WK1 516968LEG2 00000970748WK0 585393VRF2 00000974904WK5 716934RM2 00000965270WK2 914666DRS 00000546244WK5 212463LEGACY 00000963809WK9 516719RFDRS 00000953108WK8 631476RM1 00000552924WK3 516256LEG 544934WK3 914681RM5 00000973841WK0 RADNETRF1256 00000962617WK7 330270LEG 00000971824WK8 740425RF1 00000525668WK0 419483FLUORO 00000530674WK1 330372LEG 00000508469WK4 216971R8 00000977744WK2 740380LEG 00000964962WK5 330837R4 963707WK5 516663LEG 00000548409WK2 740393LEG1 00000961273WK0 740393LEG2 00000544743WK8 419394RF1 00000967340WK1 330332R18 00000539912WK6 4194475 00000550997WK1 580421VRF1 543018WK6 580251DSP 538306WK2 580251ROOM1 00000553329WK4 405359ER3 00000530675WK8 405355RF1 00000525673WK0 918682MRF2 00000552615WK7 405271URF2 00000972818WK9 405372RFT3 00000544940WK0 918567CRF1 00000541893WK4 918294SCRF2 00000541502WK1 918294SCRF3 00000978185WK7 918523LEG1 00000969351WK6 580772RF1 00000547601WK5 717267RF6 959885WK2 717SHCRF1 00000960602WK1 610237LEG1 00000964525WK0 610237OPLEG 00000964264WK6 814375MAB4 00000971674WK7 610278ADV 00000522603WK0 215707LEGACY 00000969026WK4 215748LEG1 00000540269WK8 215898RF1 00000972168WK9 412624F1 00000964802WK3 814827RF 00000978456WK2 610431LEG 000FMITR002678 864261RM15 00000519903WK9 864716RM2 00000530678WK2 803256R1 00000965604WK2 803637RF1 00000535878WK3 843674RF3 00000535876WK7 843674RF4 00000968782WK3 864487UPR2 00000965747WK9 864295GRRM4 00000544739WK6 843293RF 00000532970WK1 605352LEG 00000551896WK4 507537LEG 00000541723WK3 615792RM1 00000531046WK1 423493PRF3 00000539911WK8 615446HRF 00000953009WK8 615264H4 00000978460WK4 731425GI 00000975141WK3 731541RR 00000532515WK4 423837GRF1 00000553952WK3 865475RF 00000959502WK6 423857RF5 00000951884WK6 615769SKYRF3 00000951816WK8 615769SKYRF5 00000951751WK7 615769SKYRF7 00000521502WK5 615781LEG 00000975682WK6 865481RF3 00000548411WK8 931967SRF 512456WK5 817265NAGXR1 00000961772WK1 817472LEG1 00000962910WK6 817472LEG2 00000510617WK4 512476SRM6 00000983500WK0 512448PORF 980878WK3 512454CNTRF 550350WK3 512454RRRRF3 00000528269WK4 956350VRF1 00000952242WK6 409680BVPRF1 00000966029WK1 361857DL1 00000974347WK7 361857ENDO 00000976065WK3 361985RF2 00000535376WK8 512767NWL 00000959001WK9 214947RF4 00000547502WK5 940384DRF1 00000982588WK6 956388RF1 00000536172WK0 915533RF1 00000535370WK1 915592VRF1 00000532293WK8 915592VRF2 00000534304WK1 915595RF4 982066WK3 817336LEGR1 00000955220WK9 817347LEG 00000955387WK6 817551HMLEG4 00000046832TX2 713455LEG 00000965681WK0 713455LEG1 00000965614WK1 713455LEG4 00000983335WK1 713527PRF1 964162WK2 713790SMEND3 00000963015WK3 409539LEG2 00000514172WK6 972751RF 00000980633WK2 281599LEG 00000978455WK4 856346RF 00000992648WK6 936TCIRF1 992178WK4 936TCIRF2 00000951254WK2 956632RM1 00000537113WK3 972548ENDO 00000535879WK1 972548RF2 00000536171WK2 972548RF3 00000956645WK6 214320RF1 00000974462WK4 915520REDDY 00000543349WK5 817255NHLEG 00000975128WK0 915333GOLDER 00000970330WK7 972596GI 00000966006WK9 972596OPLEG1 00000978812WK6 361552LEGD1 00000533548WK4 210575CRF1 00000535369WK3 210575CRF2 00000981459WK1 210614NRF1 00000975150WK4 210614RF1 00000983499WK5 210614SRF1 00000966030WK9 806256RF 00000537108WK3 817424GEXR1 1035192WK2 409822TRF 00000951975WK2 409938ADTX2 00000966774WK2 361788LEGD 00000952466WK1 281338CL5 00000970544WK3 281338LG2 972230WK7 703849RF1 00000543154WK9 276523ADXLEG 00000538303WK9 540236ARF1 00000552929WK2 804541JR2A 00000976668WK4 540862ARF6 00000532517WK0 703478RF 00000970851WK2 804289HDF1 00000971676WK2 804289HDF4 00000978181WK6 804747DHP1 00000976408WK5 804747HDP4 00000970174WK9 540462ARF 00000971104WK5 434447CMRF1 00000963212WK6 509684MC1A 00000980274WK5 509962KRF1 00000974893WK0 253952RM4 00000504041WK5 206228VMA4 00009881460WK9 206386MJLEG 00000958445WK9 206386RF2 00000952137WK8 253756XRM1 00000974523WK3 509486NX2 00000959853WK3 509575PYX4 00000546584WK4 414266RM5 00000543652WK2 414289GI1 00000553948WK1 920303RF1 00000960561WK9 262321RF1 00000979791WK1 715644RF 00000958345WK1 304256RX1A 00000982401WK2 304256RX3A 00000954928WK8 304235WX1B 00000963055WK9 307332RF 00000550998WK9 1855LEGACY 00000969724WK4 0910083012 00000550355WK2 0910082023 00000957389WK0 780624RF 00000989626WK7 905793LEGACY 00000990025WK9 905336LEGACY 00000958696WK7 705544LEGACY 00000989625WK9 905547LEGACY 00000986506WK4 905624LEGACY 00000989627WK5 905845LEGACY 00000989885WK9 416656LEGACY 00000961277WK1 306778RF1 00000961630WK1 EG1112RX07 00000504045WK6 HC0887XR01 00000513395WK4 PMAB01XR03 543015WK2 XR543015WK2 977150WK2 24592065XR 00000534772WK9 0530867LEG 00000961927WK1 56582LEGACY 532967WK7 21971ADVLEG 00000541725WK8 RU1556RX02 00000541891WK8 RU1556RX03 00000957151WK4 10477FLU01 00000978317WK6 00180FLU01 00000957938WK4 00916FLU02 00000966758WK5 00139FLU03 Not Available 90085FLUA3 00000507626WK0 01060FLU02 00000547494WK5 00510FLU03 00000023357M37 907276P500 00001005375WK9 907729RM2P500 00000012142M36 907762P500 00000110928TX9 907729RM1P500 00000010127M39 907543P500 00001005395WK7 907580RM3 00000982307WK1 907580RM4 00001026147WK7 907442KP500 00000022063M32 907443P500 00000003456M31 907747P500 00000017483M39 256235P500 00000998538WK3 251580RF1 00000087393DP5 251580RF2 00001029710WK9 205592PRF1 00000002874M36 205783PRF4 1036065 205971PRF1 1036040WK2 205971PRF2 Not Available AM13DZ2333 00001030301WK4 256737PRF4 00001028159WK0 256486RF1 Not Available 205877BPRF2 00001026858WK9 205877BPRF1 00000994433WK1 256265RF3 00000012639M31 256517RF1 00000012221M38 256517RF3 00000012920M35 256HWCRF3 00000991363WK3 256MIRF1 00000981005WK2 256782P500 00001000383WK8 205387WRF6 00000134560DP2 251SPRINGP5001 1035402wk5 251SPRINGP5002 998535wk9 251HILLP500 ¿1016595WK9 251743MP1 00000997869WK3 256974P1 00000976622WK1 334705EA2 00000975153WK8 334705EA3 00000008862M35 334826ADIRF 00000991903WK6 256249CVRF 00000013942M38 205487RF1 00001029562WK4 479587P52 00001029560WK8 479587P59 00000998425WK3 479314P500A 00000012185M35 479314P500B 00000997663WK0 479441P500A 00000105787DP6 479441P500B 00000006861M39 479441P500C 00001029838WK8 479684P500 00000996484WK2 870972PRE1 00000997109WK4 870972PRE2 00000987749WK9 501987P500 00000011033M38 870845P500 00001030707WK2 501202P500RM2 00001026976WK9 501202P500SH 00000007304M39 479338P5001 00000992835WK9 479964P500A 00000986295WK4 870673P500 00000008197M36 928763WRF1 00000995343WK1 520228RF1 00001035408WK2 928729P500 00000999057WK3 928758RF1 00000009181M39 623207RF1 00000029778M38 928697RF1 00001002912WK2 928453RF3 1034387WK9 623584RF1 Not Available 602460XRESS 00001003370WK2 928669RF2 00000006095M34 INSIGHTXR106 BYWNBH 602SPVP500D Not Available 602870XRF127 00001027431WK4 928348RF2 Not Available 480882XRF 29785M33 623COBALTP500 00000089726DP4 928283RM1 00001002911WK4 928283RM2 00001026242WK6 510814RM4 00001026861WK3 714827P500 00000006507M38 530886P5RA 00000997110WK2 530888P5SA1 00000008562M31 530888P5SA2 00000982981WK3 661633XR2 00000135163DP4 661MSWRF2 00001033809WK3 AM13BKW479 00001027866WK1 760957P500D 00000007191M30 760873RF1 00001026639WK3 760873RF2 00000995339WK9 JMMCRF1 00000118441DP5 650MPP5001 00001023676WK8 650MPP502 00000023395M37 760725P500 00000008278M34 916RASC500D 00001004454WK3 909464RM1 00000026028M31 925674RM8 00000121170DP5 RADNETRF1273 00000009871M35 951737P500 00000993168WK4 RADNETRF198 00001024676WK7 707464RF2 00000988661WK5 626359RF1 00000989101WK1 626359RF2 00000135165DP9 RADNETRF194 00000001431M36 707443STJ2 00000013191M32 WEED500D 00000996488WK3 707725RWD1 Not Available 714966XR 00000002772M32 510797RM3 00000017415M31 559459500D 00000011032M30 408848SLP500 00000002349M39 818409RF1 00001025372WK2 818409RF2 00000030333M39 RADNETRF197 00000006332M31 559891P500 00000029538M36 510264P500 Not Available 4095612 00000010128M37 951CD500 00000024493M39 714843P500 00000011818M32 RM7GEP500 00000010625M32 RM8GEP500 00000012382M38 209SAM500D 1920754 619457P500D 00000026259M32 714670P500D 00000019346M36 949452RM1 00001000993WK4 RADNETRF173 00000120258DP9 209334LR3 00000012270M35 805737P500 1034873WK8 562799P500 00001032521WK5 213683P500 00000089719DP9 323268RF1 00001025013WK2 323361R8 00001025253WK4 323361RF2 00000990129WK9 209826500D 00000014697M37 RADNETRF1267 00000997867WK7 559353500D 00000030639M39 209966P500D 00001028637WK5 510231P500D 00001025010WK8 RADNETRF1270 00001025049WK6 LLPM500D 00001028602WK9 RADNETRF183 00000002981M39 818RIP500D 00000121252DP1 510437P500D GDXR3118 510437PREC500 00000010935M35 510869RF1 00001026260WK8 RADNETRF182 00000012164M30 661PMCRF 00000006153M31 650497RF2 00001032959WK7 RADNETRF192 00000008307M31 530626P5MH 00000007947M35 RADNETRF1250 00000021570M37 530244P500D 00000999388WK2 909793P5R4 00000999387WK4 909793P5R6 00000006149M39 909CH500D 00000011116M31 559SKDH500D 00000029203M37 RADNETRF196 1038574WK8 951552P500 00001029227WK4 916UCD500D 00000118443DP1 831775P500D 00001030463WK2 RADNETRF190 00000020787M38 949489RM1 00001031314WK6 619543P500D 1032722WK9 415600FLRM6 00000997665WK5 415600PRC1 1035760WK6 415641P500D 00000006254M37 415923PACPRC 00000021714M31 415DAVP500D 00000024065M35 415VASFP500 00000030108M35 AM1405Y37 00001034874WK6 415677P500 000000GDXR1312 415DAVP500FP 1035247WK4 408259FL1 00000007240M35 408559RM11 00000023680M32 408885FL1 00000022255M34 408885FL2 00000007826M31 RADNETRF1251 1032517WK3 RADNETRF191 00000002640M31 805546P500 29881M30 805681RF1 00000993103WK1 310264P500 00000066261DP9 818907RF1 00001030188WK5 209536P500 00000979501WK4 650723RF2 00000995342WK3 650723RF4 00000011549M33 650723SHENDO1 00000013514M35 RADNETRF180 00000024976M33 818364R1 00001031985WK3 818708RF1 Not Available 661TEHRF 00001027858WK8 805497RM3 00000980260WK4 760830500D 00001001054WK4 805648RF1 000FMITR003553 805652PRECISION500 00001038685WK2 GON4035772 00000009948M31 925947RM1 00001023876WK4 925947RM2 00000112680DP4 925947RM3 69798768-1 707FHHP500 00000010490M31 719SLVRF1 00000990839WK3 970544RM2 00000004861M31 303659P500 00000030403M30 303460CGP500 00001025579WK2 719285P500 00000008279M32 719SGPRF1 00000137151DP7 COLPUL1XRF1 00000982892WK2 970565RM1 00000030570M36 970874P500 00000007946M37 303399P5001 00001030465WK7 303399P5002 00000002643M35 303825P500 1036064WK2 303PSLP500RM5 1038850WK2 303765XRF888 00001001863WK8 970867500D 00000994426WK5 970945P500 00000977716WK0 970244AMPRF2 00000007384M31 303651500D 00001028822WK3 303673P500D 00001000995WK9 719560RF1 00000017127M32 970870P5001 00000004533M36 719846P500 00000010626M30 719333RF1 00000011790M33 303426RM4A 00000006558M31 303425P500GI 00001034582WK5 303425RF5 00000005423M39 719687P500 00001024519WK9 203739P500 00001025888WK7 203739RM4 00000011115M33 203288P500 00000097765DP2 203688P500 00000981453WK4 860444P500 00000974036WK6 203838RF1 00000983783WK2 203852RM2 00000984702WK1 203852RM3 00000979784WK6 203852RM4 00000982579WK5 860928RM2 00000989785WK1 860456P500 00000010881M31 202452PRE 00000993040WK5 202745PRE 00000996480WK0 3027442 00000989097WK1 3027444 00001027873WK7 RADNETRF1284 1038546WK6 302629R4 00000997865WK1 561996RF1 29988M33 941782P500 00001025009WK0 813571RM4 00001031445WK8 352596P500 00000002776M33 352596RF1 00000977348WK2 727462RM2 976801WK1 ALLIANCEXR115 00001035246WK6 352394RM2 1035103WK9 954WBECP500D 00001034875WK3 352521P500 00000005425M34 863419P500D 00000010882M39 386FHORF3 00000067515DP7 386TWLRF 00000003824M30 FHORF1 00001032853WK2 386FHDRF1 28948M38 386FHDRF2 00000014344GP6 727734RF4 00000008657M39 850883ER4 00001004451WK9 941473RF1 1035898WK4 904321RF 00000003581M36 904321RF1 00000022698M35 904215RF1 00001029439WK5 239343RF4 00000021093M30 239432RF11 00000021712M35 239432RF14 00000009670M31 239768RF1 00001018400WK0 239768RF2 00000010215M32 LEEOUTP500D 00000992837WK5 850862FR1 00000011864M36 333PR2 00000999920WK2 352331SPINE 00000976055WK4 850934GBR2 00000002641M39 850934P500A Not Available 954JDCHRF3 00000003984M32 904202ORF 00000977041WK3 904202RF 00000002466M31 904202RF5 00000013349M36 904202RF6 30105M31 904202RF7 00000001271M39 904202WSRM1 00000003983M34 904202WSRM3 00000084314HL4 904271WDR1 00000009310M34 904399RF1 00000981295WK9 904399RF2 00000011449M36 904627P500 00000080562DP2 904821RF1 00000080558DP0 904821RF2 00000992392WK1 904858RM1 00000029263M31 904SVSRF4 00001029836WK2 904VACRF 00000030105M31 AM13YHM383 00001035014WK8 904202RF2 1035012WK2 904627RF1 00000003053M36 850675P500 00000031051M36 561745P500D2 00001003073WK2 407518P500 00000017128M30 407518PRE 00001030126WK5 PMCP500 00001004653WK0 904719RF2 00000012331M35 727581DC500D 00000978652WK6 561791RM2 00000004233M33 561791RM3 00000979503WK0 954978RM1 00000025249M34 954NWP500 00001006207WK3 850718JR1 00000001235M31 305325RM10 00000097770DP2 305325RM7 Not Available 386NSBDET 1033334WK2 386NSBRF 1035500WK6 850678TR1 00000002778M39 352401RF1 00000993163WK5 352671RF1 00000975954WK9 352671RF2 00001031638WK8 401RF1 00000064764DP4 407296PRE 00001030709WK8 863763RM1 00000981750WK3 863763RM4 00000003199M37 FHFRF 00000978177WK4 904276RF2 00000975953WK1 904276RF3 00001028162WK4 FHALTP500 00001002434WK7 FHALTRF5 00000995871WK1 386326RF1 00001026860WK5 850747GC3 00000984284WK0 850747RF1 00000029779M36 850416P500 1033581WK8 850494WF3 00000981790WK9 850494WF4 00000979188WK0 813757FL3 00000066019DP1 954476500D 00000004715M39 954476RF6 00001030306WK3 954587RM4 00000023455M39 941629P500 00000085030DP5 941766RM2 00000991653WK7 941766RM4 00000987733WK3 386PORF 00001033339WK1 772335P500D 00000030230M37 772335RFRM5 gdxr3083 904FICRF1 1035002WK3 727521P500D 00000980625WK8 727823RM3 00000065980DP5 407321PRF5 1034356WK4 941342P500 00000982368WK3 772589RM3 00000022747M30 863314P500 00000097962WK2 850325RF3 00000979394WK4 850325RF4 Not Available 954724P500D 00000029938M38 AM13ORK505 00000011569M31 813827P500D 00000005426M32 813844USFRF1 00000981291WK8 813870500D 1035225WK3 813873P51 00000022406M33 FWRF2042 00000023454M32 FWRF3653 00001026256WK6 727848P5D1 00001026255WK8 727848P5D2 00000001447M32 772VRAP500D 00000982399WK8 561863RF1 00001004963WK3 863298RF 00001002435WK4 FHWPP500RF1 00000981791WK7 229312PR500D Not Available 229312PR2 1315 404785CWBF1 00001026712WK8 229924P500 00001029717WK4 706554RF1 00000972840WK3 404686CRF2 00000972514WK4 404686CRF3 00001029234WK0 404686EMRF7 00000979502WK2 404712PR1 00000982742WK9 404712PR2 00000013350M34 404728PR1 00000007881M36 404778PR1 00000981074WK8 404778WC2 00000008194M33 404785EFL3 00000013303M33 404785P500D3 Not Available 404785SRF1 Not Available 404785SRF4 00001034486WK9 404785EFL2 00000005491M36 404785P500D2 1036543wk5 404785SRF3 00001026975WK1 706651D3 00000986118WK8 706651OPI2 00000008849M32 706721MCFL1 00000010809M32 706721MCFL2 00000997660WK6 706733V8 00001004240WK6 706745URF1 00000105788DP4 912466P1207 00001030710WK6 478PCHP500 00000029989M31 770382P5D1 00000983133WK0 706494H500 00000002871M32 706335RF1 00000009311M32 770918P500 00001030582WK9 229276CRF1 00000023358M35 229276CRF2 00000007829M35 770786P500 00000977710WK3 404321P500D 00001027620WK2 912384RM5 00000978174WK1 912274RF2 00000983137WK1 912275RF1 00000981456WK7 706213ER1 1033150WK2 706544RF 00000006859M33 770228RM3 00000002217M38 678454P500D1 00000002353M31 678454P500D2 482983GE4 404785KP5 00001027875WK2 706356TRF 00000988666WK4 478633DH2 00000977376WK3 478633PDI3 00000009401M31 478757RF1 1035013WK0 478765P500 00000994425WK7 229890P500D 00001028821WK5 770400P500D 00000008658M37 912826RF1 00001026899WK3 478327P500 00000985423WK3 706291R7 00000009947M33 912576AURORA 00000011943M38 912819PRO1 00001023983WK8 ALLIANCEXR121 00001023983WK8 770434P500D 00000005942M38 770736RF3 00000005424M37 770736RF4 1026974WK4 912754PREC 00001004965WK8 912486EGA4 00001031577WK8 229228P500 1033966WK1 229228RF1 00001034485WK1 706597RF 00000022015M32 229353TR2 00000021356M31 229353TR3 00000022049M31 229391WC1 Not Available 770307P500R2 00000134559DP4 AM13OPT168 00001031835WK0 671USNHP5D2 00001031633WK9 808HMCP500D 00000027295M35 808HRAEP500D 00000980628WK2 808973RM2A 00000023457M35 808YMP500D 00000238775HP1 0004046239DET 00000023727M31 808HRAWP500 00000080067DP2 808433500D1 982744WK5 808242P500D 00001024889WK6 515295P500 00001003372WK8 641843P500 00001024874WK8 319268P500 00000020300M30 641228P500 00000010124M36 563387500D 00000977232WK8 515226RF 00000975292WK4 563589F500D 00000004227M35 712852P500 00001024518WK1 712362P500 00000010380M34 641743ACP500 00000011434M38 515386P500 00000006692M38 319293P500 1035411WK6 641754P500 00000005498M31 641422RM1 00000005111M30 641422RM3 00000978499WK2 319283M500D 00001004452WK7 402358P500 00000001837M34 712472P500 00000026330M31 319863P500 00001025366WK4 319272P500 00001025806WK9 563568P500 1038352WK9 563568PR500 00000991639WK6 515832P500 00000009532M33 563422P500 00000030229M39 208785XR 00000009120M37 208422P500D 00000002639M33 208678500D 00000005813M31 208455RF1 00000001835M38 208227ROOM6 00000977040WK5 208557RM1 00000081599DP3 208786PREC 00000013493M32 208828P500D 00000005335M35 208265BXR1 00000004231M37 208933500D 00001003369WK4 847842RAD5 00000978175WK8 708783RM1 00001012891WK6 630312P500 00000007584M36 217366CXF3 00000013748M39 312996P500D 00001027080WK9 312CMP500D1 00001027063WK5 312CMP500D2 Not Available 312VAJBP500 00000977231WK0 773633EP500 00001034945WK4 773794P500 00000001658M34 773962PR1 00000982398WK0 773IMP500R10 1035755WK6 773P500RM1 010138WK4 773STJP5D319 00000003578M32 773STJP5D323 00000013915GP4 773IMP500R5 00000986821WK7 217935WF1 00000993348WK2 217443URF3 00001024725WK2 217443URF5 00000027155M31 217442VARF1 00001001057WK7 217876DRF1 00001029007WK0 217876DRF5A 00001004951WK8 815756RM2 al3lbc 847297P500D 00000017195M39 630275G04 00001025054WK6 630275G05 00000996487WK5 847742RM6 00001029430WK4 309467EP500 00000009277M35 847316P500D1 00000980495WK6 847570RF10 00000012865M32 708422P500 00000005275M33 815776P500 00000012025M33 309343CRF1A 00000007505M31 630545RF1 00000014033M35 847249PREC500 00000001967M39 815741P500 1034106WK3 708245P500 00000010286M33 618288ARF3 00000030908M38 618288MIP500 00001030577WK9 618288RM2 00000030538M33 217258SBRF2 00000122578DP8 815300P500 00000012381M30 309454BX5A 00000021270M34 309344MRF1B 1035503WK0 309655FP500B 00001030982WK1 309672P500D2 00001026240WK0 309691PRF1A 00000015914M35 815431OP500 00000022356M30 217222QP500 00000004859M35 815489P500 00000017943M32 618443SP500 00001038500WK3 815664P500 00000983035WK7 217528SCRF3 00000982795WK7 217528SCRF4 00001025012WK4 217732LRF 00000004644M31 217788PRF 00001027085WK8 217333P500D 00000995879WK4 217383CGI1A 00001030704WK9 217383P500A 00000012771M32 815432P500 00000064144DP9 847872RF1 1033577WK6 260665P500 00000979621WK0 260920RF 00001031839WK2 574546P500D 00000009118M31 812343PRE500 00000004716M37 219756PRF 00000005943M36 219757P500 00000005270M34 574523EG2 00000006560M37 574523EG4 00001027616WK0 812450DRF10 00000098100WK7 812450RF12 00000988660WK7 260347RF 00000992217WK0 260373RF4 00000996483WK4 260373RF5 00001001363WK9 260460PRE 00000002774M38 260463PRE 00000013176M33 260355P500 Not Available 317338PMP500 1038789WK2 317338PRECISION500 00000995347WK2 574772RF2 26739101 AM1217VZ5XM 00000989784WK4 812265F1 00000007939M32 765674PRE1 00001002437WK0 812858MRF 00000998422WK0 812858PRF 00000008408M37 812522F1 1035903WK2 574647IR4 00000016799M39 812268P500 1034250WK9 812238FL1 1034393WK7 812238FL2 00000983859WK0 574372RF2 00001027594WK9 785243P500 1034358WK0 620225P500 00000007773M35 620341P500 00000011114M36 785239P500 00000981457WK5 620378RF 1033331WK8 620272P500 00000007336M31 620365P500D 1034105WK5 785776P500 00000980187WK9 913498MR2 00000980497WK2 913541HPR3 1034394WK5 913317P500D 1034102WK2 913REHAB 00000005276M31 913294P500D 1034037WK0 620672P500 00000990840WK1 913676SRM4 00000007243M39 785354SD2A 00000002117M30 785354SD4 000000GDXR3069 316689P500 00000030401M34 316858P500 00001036549WK2 316962P500 00001029712WK5 606833P500D 00000007583M38 270793G4 00000982800WK5 270798PREC2 00000978650WK0 502868RF1 00001030466WK5 270659P500 00001028431WK3 270932P500 00001030121WK6 270756P500 00000120793DP5 606573P500 00001025878WK8 606439P500 00000980498WK0 270827RF 00001029713WK3 270887RF 00000012660M37 270692F1 00000002869M36 606878P500 00000030106M39 502587P500 00000974459WK0 502897F1 00001025375WK5 270251JPRF2 00000002644M33 606759P500 00000991630WK5 606248RM1 00000980263WK8 270389RF 1030415WK2 270417RF1 00000022908M38 270417RF2 00000002120M34 606789PRE 00000004536M39 606337P500 00001029425WK4 270725P500 00000994430WK7 606678RF1 00000990547WK2 606678RF2 00001025887WK9 606237PRE 00000012332M33 859879P500 00001025886WK1 606633P500 00001027589WK9 859745P500 1034386WK1 337898P500 00000986516WK3 318448CRM1 00000135915DP7 318473RD1 00000008860M39 318473RD4 00001017448WK0 225755RF2 30501M31 225761PRE500 00001031837WK6 225924WRF2 00000030501M31 AM14BIM496 Not Available 318346RD1 00000025397M31 985867RF3 00000981584WK6 985875P500 00000010739M31 337457MRF 00000008858M33 337531FRM4 00001024743WK5 225NORF 00000989786WK9 985230DCRF3 00000989096WK3 985230DCRF4 00000008961M35 985230RF5 00000030337M30 985230RF6 00000014273M37 318927RF1 00001027482WK7 504873TR4P 00000983225WK4 504873TR5P 00000024861M37 985LCFB 00000003583M32 504842ENDO 00001024575WK1 504842RM3 00000997104WK5 337261URF1 00000117092DP7 337521RF1 00000004529M34 337521RF2 00000005998M30 225869P500 00001024741WK9 504837P500 12084 318371P500 00000012922M31 318330RF1 00000004648M32 318330RF2 28181M36 318812RF 00000010219M34 504380P500 00000005941M30 504894RF 00000011182M33 504897P513 00000003575M38 504988ENDO 00001023632WK1 504988P500 Not Available 504CMP500 00000031054M30 504CRHP500 Not Available 2978112 Not Available 2978110 1035409WK0 985646NSRF3 00001025258WK3 985646RF4 00000978076WK8 504447P500 Not Available 985447P500 00001001071WK8 337363VRF1 00001031848WK3 225658RF1 00000981583WK8 617636RMD 00000006152M33 617667ERCP1 00000006335M34 617667ERCP2 00000012773M38 617724ENDO 00000984287WK3 617726PR35A 00000992390WK5 617726PR64A 00000995344WK9 617732ENDO 00000984285WK7 617732P15 00000981789WK1 617732P16 00000002779M37 617789RFE 00000011843M30 781895RF1WA 00000005154M30 508941RM5 00000979919WK8 781744RF3 00000989779WK4 617726PR32A 00000002978M35 508457RF3 00000980118WK4 978521MVRF4 00000013129M32 781477RF2 00000020626M38 781306RF1 00000011194M38 781979RF1 00001021880WK8 508693MVXF 00000008138M30 413447BRF1 00000995872WK9 413447BRF2 00000013664M38 413854P500 00000022652M32 978741SRF3 00000979921WK4 978741SRF9 1026264WK0 781624HSSHFP 00001026264WK0 781624HSSHRF 00000001427M35 413748RF2 00000009122M33 413748RM3 00001006188WK5 781756BPRF1A 00001026636WK9 508793MP500 00000004530M32 508856EN1 00000017485M34 508856ORP500 00000017783M32 978368P500 00000997114WK4 443481RF4 00000024278M34 443481SJKP500 00000023220M37 UICGERF1 00001003072WK4 410225RF3 1034359WK8 RADNETRF1298 00000030497M32 RADNETRF1303 Not Available RADNETRF1303FP1 00000981069WK8 301896RF2 00001022261WK0 301295P500 00000020788M36 301295P500A 29112M30 RADNETRF1301 00001014111WK7 410604KIRF 00000013830M35 301877RF2 00000982893WK0 301877RF8 00001028432WK1 RADNETRF1286 00000100539WK2 301292P500 00000020302M36 240566DF1 00001029139WK1 240566RHRF1 2088836 240566RHRF2 00001025006WK6 RADNETRF1268 MISSING_SN_GON4225310 RADNETRF1297 Not Available RADNETRF1296 00000989778WK6 301714RF2 00001026786WK2 301609P500A 00000006329M37 301475P500A 00001025008WK2 RADNETRF1269 00000011756M34 410535P500A 00000017941M36 301765P500 00001030123WK2 301891P500 00001034585WK8 RADNETRF1299 00000026412M37 RADNETRF1295 00001036558WK3 AM13XJ54 29262M33 RADNETRF1302 00000011725M39 207973P500D 00001005334WK6 207454RF1 00000022650M36 207498P500 00001025050WK4 207564P500 00000008600M39 207553P500 00000029356M33 207879P500 00001028999WK9 517543P500D 00001034584WK1 586263P500 00000997659WK8 313593P500 00000008041M36 313576P500A 00000009872M33 248615P500E 00000010377M30 810762P500 00001027592WK3 989246P500 28947M30 810606RF1 00000014476M36 616267P500A 00000119459DP6 616267P500B 00000979619WK4 616391PRECISION 00000022407M31 231225P500 00001030299WK0 269945P500 00000005419M37 517437FL 00000008473M31 906485P500 00000017013M34 269789P500 00000010811M38 906884P500 00001005386WK6 269657P500D 421798GE0 248652ERCP Not Available 248551P500RM2 00001033791WK3 989497P500 00001030701WK5 269637P500 00000010460M34 734671P500 00000135923DP1 248524P500 00000022906M32 320845P500 00000006504M35 507526P500 00000974342WK8 218736RM4A 00000979780WK4 320693PREC 00000998428WK7 507283P500 00000012606M30 763581P500 00000030498M30 320589P500 00000025725M33 218347P500 00000975908WK5 763520N8 00000008660M33 651430P500 00000001445M36 218741P500 00000996479WK2 WH231P500D 00000079050HL1 WH231WDR2 00000980629WK0 816348BRF1 00000989783WK6 660542P500 00001032085WK1 660707PREC1 00000011195M35 573814VARF1 00001029006WK2 573231HCP500 1020690WK2 816380P500D 00001005380WK9 573359PRF1 00000997873WK5 573486HR1 00001006202WK4 816501RF1 00000003980M30 816501RF2 00000004240M38 816501RF3 1033965WK3 573SMHP500 00001032084WK4 417347FEP500 00001026972WK8 417347P500 00001034588WK2 816276RMCRF2 1035015WK5 816478P500D 00000005337M31 417682BRF 1035223WK5 417533MLR2 00000005210M30 816282LSPRES1 00001031846WK7 816259LP500D 00000006693M36 660886FRF1 00001034871WK2 660562P500D 00000003058M35 573686P500 124583HL6 314251MR8 00000978498WK4 314525SARM2 00000976057WK0 314525SPRM3 00000007242M31 314996BW2 00000990831WK0 314996MBRM1 00000981078WK9 636916BRM4 00000994417WK4 417269CMP2 00000994423WK2 417269SXP2 00001001055WK1 417269SXP8 00000013596M32 417269XRER 00000024798M31 417CMXR2 00000004229M31 573468MBS1 00000999056WK5 228467P500 1035088WK2 601645DRF 00001003376WK9 662846P500 00000999919WK4 601740PRF1 00000979359WK7 601936RF1 00000001434M30 601936RF2 00000002356M34 662459P500 00000993352WK4 601288PRF5 00000998427WK9 601288PRF6 1035710WK1 601288PRF7 00001028422WK2 601554P5001 00000990535WK7 601426PRF1 Not Available 601250XRF116 00000002123M38 601853RF2 00000002355M36 601445PRF1 00000975152WK0 228818R3 00000010736M37 228875P500D 00001030584WK5 601663P500 00001027588WK1 601798P500 00000086753DP1 662756RF1 00000014034M33 662756RF2 1033968WK7 601883P501 00000982799WK9 662716RF1 00000002467M39 662751RF1 00000003977M36 662751RF2 00000004236M36 406496P500 00000008280M30 406447VARF2 00000011755M36 406BVARF MDYS3064 406228FMRF 00001003374WK4 406363MX1 00001026473WK7 406293P500 00000011636M38 406883P500 00000002984M33 406863P500 00000010126M31 252209RC500D Not Available 252RCH500DIG 00000001436M35 828213AIRF2 11039 828298RM4 1034111wk3 919FMCP500 00000005494M30 919854P500 00000011724M32 704355RF5 00001030705WK6 704667RF1 00000010627M38 704667RF2 00000010807M36 704863RF1 00000011734M31 704863RF4 00000981916WK0 252466CP500D 00000011723M34 919JMCP500 00001028163WK2 828456P500 00000003825M37 252482C500D 00000011416M35 252384RM3 000FMITR003567 910CFFAYP500 19650 910VACP500RM1 143629TX4 910VACP500RM2 00000030232M33 AM1405Y533 00001034397WK8 910FAYVARF1 00000008311M33 910907P500RM3 00000010737M35 910907WARM4P500 1038717WK3 828524P500 00001026148WK5 704829P500 00001027619WK4 704834P500 00000010280M36 919734WR500D 00001027079WK1 252847GE3 00000002122M30 252847PR7 00000011817M34 252847PRF4 00000022746M32 252847VCH500D 00000023608M33 252VA500D 00000977711WK1 910205RF1 00000995878WK6 252436RF2 00000008135M36 828526P500 00000030638M31 DUPLIN500D 00000014567M32 910671SERF5 00000027716M30 704660P500 00000404295GE8 AM129W5M6V5 00001034599WK9 AM129W5O01K 00000133202DP2 AM129W5PJ8L 136750DP7 252OBH500D 00000023916M30 910FHMP500 00000133526DP4 910CFHOKEP500 00000010081M38 919954RM2P500 00001026705WK2 919RALCPP500M 00001026971WK0 919RALCPP500W 00000030399M30 910417P500 00000014121M38 252962NGRM3 00000013471M38 336599P500 00001004652WK2 828286RH4 00001004967WK4 828286RH5 00000030104M34 704210RX6 00000981754WK5 704487CRRF1 00000980624WK1 919663CRF 00000013513M37 828586MOBRF 00000014706M36 828586RF1 1038733WK0 252641PREC 00001029794WK3 WILMVA500D 1029794WK3 WILMVADET1 00001034396WK0 910667RFD7 1034110WK5 910667RFD8 00000004969M32 336679HMR1 00000005648M31 701477P500 00001025988WK5 701252P500 00000987729WK1 701683P500 00000984223WK8 701723P500 00000980264WK6 402245CRF1 00001027424WK9 402759FCP500 00000980265WK3 308534RF 00001024726WK0 402294EBRF5 00000010282M32 402352SP500 00000989099WK7 402643SMX2 Not Available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ot Available 856247P500UPG 00000984229WK5 856325P500 00000012810M38 609835LMCP500 00000089721DP5 609835P500 00000080527DP5 505262P500 00000002982M37 505888P500 00000023495M35 505WHRM1 00001031444WK1 505291P500 00000004527M38 505841RM4 00001002919WK7 505599P500 00001028169WK9 505609P500 1038764WK5 505863XRF1 00000993344WK1 505521RF1 00001034491WK9 575521PRF3 00001033795WK4 505662P500 1034316WK8 575359P500 00000002642M37 505627RF1 00000007774M33 505368P500 00000010624M35 505368P500B 00000982577WK9 775885500D 00000003460M33 775753P500 00001031629WK7 702228P500 00000981915WK2 702270RF1 00001029839WK6 702304RF1 00000005646M35 702360RF1 00000985029WK8 702388RF1 00000975952WK3 702731DCRF1 00000991651WK1 702732RF1 1032895WK3 702853RF2 00000000BRDDLG 702855RF1 00000982308WK9 702855RF2 00000005497M33 702896RF1 000FMITR003554 AM136BD112 000FMITR003554 SDMIP5SHADOW 1038645WK6 702383P5002 00000007658M38 702794RF1 00000008661M31 702653RF03 00001028635WK9 775982RF7 00000982310WK5 775SMP500D 00000990127WK3 518262RMG 00000982371WK7 518262RMH 00001006192WK7 518262SC1 Not Available 518842P500 00001029716WK6 516796P500 29114M36 718584RF2 00000013013M38 914787500D 00000982576WK1 718240RM11 00000009276M37 718613PRECISION 00000090028DP2 718630PREC2 00000991232WK0 718630PRECISION 00001002436WK2 718963RF2 00001030127WK3 RADNETRF1290 Not Available 718240RM9 28575M39 RADNETRF1300 00001006190WK1 716568RF2 28439M38 716834RM210P500 00001027093WK2 716845DRF1 00000026025M37 716845DRF2 00000996806WK6 716859BGRFLL3 00000982797WK3 716859BGXR12 00001008860WK7 716878WCRF2 00000982578WK7 716878WCRF3 00000005204M33 585463GI 00001005379WK1 845831RF1 00001003379WK3 607547RF1 00000010284M38 518873RF2 00000120246DP4 518926FLRM3 00000017016M37 518926FLRM4 00000990416WK0 518725RMB 00000012923M39 716298RME 00000030107M37 718267P500D2 00000011789M35 718267P500D 00001020193WK7 585798P500 00001025367WK2 ORMCP500 00000007704M30 516663RMP3 00001024746WK8 516663RMP5 00001002913WK0 607535P500 00000008659M35 315624P500D 00000981072WK2 LIJ016996 00000982798WK1 LIJ017457 00000024494M37 212288P500D 00001025509WK9 212420ENDO1 00000005207M36 212420ENDO4 00000007447M36 COR369577 1029003WK9 COR481340 00000021355M33 COR481341 00001029319WK9 COR481342 00001026471WK1 COR481360 00000087660HL7 COR515338 00001028636WK7 COR515339 1028636WK7 COR515340 00000012143M34 RADNETRF1291 00000004719M31 631VAP500D 00000010492M37 516632RM3 00000024194M33 516632RM6 00000020467M37 315363P500 00001003367WK8 607432GE1 00000025566M31 631654P500D2 00000026936M35 516705RM1 1038875WK9 GON4465879 00001025051WK2 315338RF 00000991210WK6 518583P500 00000990280WK0 518347RF3 00001029233WK2 631726P500D 00000982890WK6 775356PD Not Available 4094884 00000021715M38 718226PRM1 00000021571M35 718226PRM10 00001030891WK4 718226VZRM2 00000020668M30 718226VZRM3 00000008861M37 HIP816PREC1 00000001430M38 631444P500D 00000019344M31 315425A320 00000012360M34 315464RM3411 00000012184M38 315464RM3629 00000014081M34 518585P500 00000007502M38 RADNETRF1249 00001026859WK7 516256P500 00000017784M30 585596P500 00000009404M35 914681RM3 00000981071WK4 330729500D 00000008007M37 330363R7 00000982030WK9 330489R4 00000007239M37 330489R5 00000003579M30 513247PRE1 GDXR3151 513584PRE1 1034583WK3 513624PRE2 00000029064M33 216476PREC1 00000987736WK6 216844PREC1 00001006924WK3 614257PRE 00000011486M38 614293PRE1 00000998436WK0 614293PRE4 00001027593WK1 614722PRE 00001004962WK5 614839PRE 00000011243M33 614293PRE2 00001020933WK6 614455PRE 00001031096WK9 937641RMA 00001030462WK4 937641RMF 00000098887HL3 DCHRMADIG 00000022620M39 DCHRMFDIG 00000007449M32 419784P500 00000974344WK4 740368PRE1 00000975509WK1 216343R1 Not Available 937PCP500 00000028879M35 AM13HNY56 00000005655M36 216761PREC1 00001027081WK7 740845PRE 00000009949M39 740383PRE 00001028436WK2 740383PRE2 00001030467WK3 740633P500 00001000387WK9 440816PREC1 00001000810WK0 513523RM3 00000997870WK1 419399RF 00001034586WK6 419557RF6 00000971696WK0 937498PRE 00000997100WK3 419251RF6 00000008848M34 419383FL9 00000008741M31 419407FL1 00001024888WK8 937440RM1 00000001836M36 937440RM2 00000024495M34 419294P500 00001029559WK0 419238P500 00001027618WK6 419933RF 00001029708WK3 580379P500 00000013944M34 918JP500 00001003371WK0 918341C500D 00000004242M34 580323P500 00000008038M32 405359P500D 00000013515M32 918253P500 00000113685DP2 405375P500 00000006617M35 580531RF2 00000004857M39 405610MWRF3 00000006002M30 918682500D 00000006000M34 918273CNNP1 1035089WK0 918535P500 00001024572WK8 405271CCP500 00000003455M33 405271P500 00000007386M36 405271PRF5 00001031443WK3 405271UP500 00000018055M34 405752RF2 00001038632WK4 405949GI8P500 00001038790WK0 405949GI7P500 00000002351M35 918274500D 00000008560M35 918825500D 00000XA44507-8 918434WDR1 00001028633WK4 918323P500 00001027083WK3 918434P500 00001027863WK8 918775P500 00001025371WK4 918224P500D1 00000007656M32 405303P500D 00000991631WK3 405742P500 00000021224M31 405744OSUVET 00000013218M33 918696P500 00000997108WK6 918453RM2 00000011031M32 918458P5002 00000010283M30 918458P5003 00000011835M36 918567P5002 00000023038M33 918286P500 00000004862M39 918294MP500 00000082084DP5 918579500D 00000983533WK1 541523EP500 00001028171WK5 541382CRM1 00001026640WK1 541382RM3 00001003929WK5 541548RM1 27210M34 541412RF1 00000004721M37 541573RM1 00000997111WK0 541396RF1 00001024724WK5 503623RF1 00000021354M36 541472P500 00001006205WK7 541387P500 00000030566M34 541274P500 00001025002WK5 541947RF1 00000005811M35 541475RM1 00000029265M36 541789WICRF 00000023433M36 541276P500 00000999393WK2 503215P500D 00000002354M39 503513RM4 QC28163 541447RM1 00000030402M32 541677MP500R3 00000019761M36 503717RF1 00000063938DP5 541296MC 00000008308M39 503216P500D 00000133205DP5 570759PREC 1034112WK1 412851P500D1 00000008136M34 484884P500 00001028437WK0 SLWP500 1033153WK6 412914P500D 00001032522WK3 215785P500 00000987062WK7 610526PRECIS1 00000025797M32 717217PRE500D 00001031635WK4 814226P500 00000014007M39 814768P500 00000010182M34 570645PREC 29992M35 724626P500D 00001033510WK7 814274P500 1035904WK0 724625FL1 00001030579WK5 724772P500D1 00000982396WK4 570271F1 00000066254DP4 570271F2 00000013352M30 814375P500D 00000018111M35 SLRP500 00000008005M31 717733RF3 00000009711M33 717733RF4 1038771WK0 814877P500R5 00000021222M35 814624RF2 00000998429WK5 814624RF3 00001028165WK7 724983500D 00000990837WK7 814437RF1 00000990834WK4 814437RF2 00001032655WK1 724588P500R4 00000023977M32 724588P500R5 00000996676WK3 724458RF1 00001003937WK8 717637RF1 1038760WK3 717637P500 00000003459M35 814643RF 00000030335M34 814837P500 00000003982M36 717544HCRF6 00001011182WK1 717544HCRF7 00001024578WK5 215710AURORA2 00000978129WK5 215750PRECIS1 00000980626WK6 215750PRECIS2 00000013873M35 724537RF2 00000989934WK5 717272RF2 00001024520WK7 717272RF3 00000026260M30 610377P500 00000990838WK5 570523PRE1 00000007106M38 570523PRE2 00000002867M30 570574PREC2 00000984896WK1 717627RF2 00001031838WK4 412664P500D 00001027096WK5 412372P500D 00000021829M37 215487P500 00001003378WK5 215537ERMCRF1 00001038445WK1 AM15L38452 1033511WK5 412232FL3 00001032943WK1 412367P500D1 00001028434WK7 412367P500D2 00001025575WK0 412623P500D1 00001026709WK4 412623P500D2 00000009403M37 412647P500D1 00000023396M35 412647P500D2 00000024604M31 412647P500D3 00000003458M37 412648P500D 00001031260WK1 412784P500D1 00001005387WK4 412825P500D 00000008199M32 570621PREC1 00001030985WK4 570621SEPREC 00000002635M31 814938RM1 00000997871WK9 215538PRE1 00000004241M36 814224RF2 00000990412WK9 412933RF2 00001033159WK3 814684500D 00001030461WK6 610284P500 00001028601WK1 814723P500 00000001222M39 570574PREC1 00001024745WK0 401782SCP500 00001007260WK1 864366ARF2 00000014478M32 803502RF 00000066259DP3 803641RF5 00001020999WK7 803641RF6 1038716WK5 864716P500 00000008559M37 843228RF2 00001034918WK1 843792RFPED2 00001001062WK7 803581P500D Not Available 803256ESS 00000994131WK1 803776VRF5 00000007192M38 843674P501 00000007195M31 843674P502 00000986937WK1 864487UPR1 00001035404WK1 864299P500D 00000997101WK1 843339P500 00000002870M34 843374P500 00000004235M38 843692GRF3 00000013749M37 843847TR1 00000011657M34 843847TR2 00000013302M35 864560NGR3 00000026216M32 843832RF4 30058M32 843832SRF1 00000030058M32 AM1405Y754 00000993780WK6 803775RF2 00000010497M36 864427RF1 00000008008M35 843549RF3 00000999060WK7 605622RF 00000010181M36 605696P500A 28069M33 605256P500D Not Available AM13IF2205 00001024881WK3 605867P500 00000993781WK4 605644P500 00000014803M31 605747P500 00000974761WK9 605328RF 00000982033WK3 605333RM6 00000002578M33 605333RM8 00000004714M32 605624P500A 00000025475M35 423493P500D 00000974460WK8 423855EX3 00000030334M37 931245P500 00001025052WK0 931520P500 00000002350M37 615446NRF 00001032740WK1 615446P500D 1034602WK1 423542RF3A 00001038849WK4 931438P500D 00000009182M37 615435WP500 00001032855WK7 615435WRF2 00001002185WK5 615451SUMRF1 00001002915WK5 615451SUMRF2 1038896WK5 615452XRF 00001029228WK2 865316RF1 1033157WK7 615316P500D 1035683WK0 731541P500IC 00001032942WK3 731541RH 00000029777M30 731541RR1 482984GE2 731541XR500 00000996481WK8 931879JRF2 00000977346WK6 423224RF1 00000977347WK4 423224RF2 00000992395WK4 423562RF1 00000012359M36 865373RF 00001032653WK6 865373RF2 00000990539WK9 865531RF1 00000992840WK9 423562RF2 00001005392WK4 931766RF1 00000991657WK8 865458RF2 00000013217M35 931823LRF1 1038401WK4 931815PRE500 11801 901516P5002 00001000389WK5 615342P500D3 00000999391WK6 615342P500D4 00000978803WK5 615777500D 00001029005WK4 731644P500 00000005495M37 931363HPD1 00000990407WK9 731926RF1 00000980034WK3 731926RF2 00000980117WK6 615768SCMCFL01 00001034603WK9 931393HP500 00001029320WK7 325670PR7 00000118774DP9 972361P500A 00000991907WK7 972747EP500 1020381WK8 979848P500 00000993161WK9 512397ENDO1 00000985424WK1 512448RF1 00000014477M34 512544SDRF1 00001033792WK1 512ACCRF1 00001034490WK1 512NAMCRF1 30801M35 NAMCRF2 1038982WK3 512751XRF888 99750WK3 281420500D2 00001012895WK7 281420AURORA 00001031095WK1 281420SJM500D1 00000013127M36 281427JMRI500D 00000004230M39 830816RF 00000982036WK6 956350VRF3 Not Available 817HUGP500 135166DP7 FCTXRP500 00000005808M31 512528CPRF1 00000135781DP3 CPICRF Not Available CPICWDR 00000012023M38 936539P500D 00001026257WK4 936539P500D2 00000129949DP4 361694FL2 00000989095WK5 361857DPR2 00000999055WK7 361866MPR 00000007448M34 361980RASP 00000989104WK5 361985RF1 00000066222DP1 956350RFP5 00000064162DP1 361961NR2 00001004245WK5 214345P500D1 00000992393WK9 214352PREC1 00000029879M34 214FBMCP500 00000011245M38 469624P500D1 1031842WK6 972566P50011 00000980033WK5 972566P500D1 00000980165WK5 972566P500D2 00001031091WK0 972566PEDIRF1 00000024497M30 WNHP500 00000022929M34 830703VVRF1 00000989932WK9 903416PD1 00000008300M36 919954RM7P500 1031631WK3 940384P500 00000006096M32 915521ENDO 00000006097M30 915521RF2 00000010217M38 915564PRF3 00000006619M31 915564PRF4 00000012024M36 915569PRF1 00001032733WK6 915569PRF2 00001033794WK7 915595PRF 00000004965M30 972875RF01 00000102210HL2 830393CFP 00000012248M31 972420FMP500 00000002980M31 254288DPRF7 00000005420M35 254288DPRF8 00001034104WK8 254DAHRF1 Not Available 254DAHRF2 00000004528M36 817347P500 00000030057M34 817347P500RM1 00001034917WK3 817ALLP500 1034115WK4 817294P500 00001034036WK2 817BSFWP500 00000005812M33 830997HRF2 00001035797WK8 409772TDC2 00000001437M33 512943GTP500 00000003829M39 940549P500 Not Available 817579XRF888 00000017645M33 806364P500 1033793WK9 254580P500 00001027617WK8 281477MWBP500D 00000986289WK7 281588RF2 Not Available 281FTHP500D 00001024574WK4 281MWP500 00001002235WK8 713394MM500D 00000002352M33 713441MMRF1 00000009822M38 713441MMRF2 00000013305M38 713441MOP500D2 00000013216M37 713441MOP500DP 00000009873M31 713455EH500D 00000983534WK9 713636LBJ7 00000984286WK5 713636LBJ8 00000121600DP1 713791P500FP4 00000010079M32 713MWP500 00001028170WK7 713SCP500D 00001028172WK3 713SCP500D2 00000008773M34 713WMP500D 1034589WK0 713WMP500D2 Not Available MHSWP500D 00001026254WK1 713873P500D 00000001661M38 817545P500 1038347WK9 830896P500 00000008601M37 254519RF1 254METRO500D 254METRO500D 00000986294WK7 281348P5001 00000988658WK1 281348P5002 00000013597M30 210292WHR3 00000008738M37 210292WHRM4 00001025256WK7 512LRMCRF 00000982373WK3 972420LR6 00000982395WK6 972420LR7 00000983136WK3 210646NERF3 00001038481WK6 806725PRECISION 1038883WK3 806725GOLDSEAL 00000978495WK0 956668RF1 00000980496WK4 956632MCRF1 00001030128WK1 956632RF2 00000012548M34 972569MMHRF1 00000980032WK7 972698P500D3 00000029784M36 254MEXP500 1034488WK5 936NMCP500 00001032526WK4 830RHHRF1 00001032738WK5 830RHHRF2 00000974206WK5 817255NHPD1 00000990278WK4 817255NHRM1 00000981070WK6 903731PRP500 00000991635WK4 806663P500 00000999395WK7 713359DRF1 00000997661WK4 713359RF3 00000030136M36 713363P500 1035090WK8 HCAPEARP500 00000997357WK9 972238PRO1 00000001963M38 972981P500 00000014498M30 512341RRRF1 00000987067WK6 325658RF1 00000982187WK7 210208RF2 00001000999WK1 210208RF3 00001000994WK2 210572RF1 00001004456WK8 210575MSTH2 00000985030WK6 210575MSTH4 977377WK1 210616MCTRF 00001024577WK7 210617RF1 00000030637M33 210617RM6 00000009274M32 210638RF1 1033156WK9 210FPRF1 00001025978WK6 210MTRF1 00000030336M32 210WHASCP500 00001038837WK9 210351P500 00000013351M32 210358URF8 00000005205M30 210297BSRF1 00001030459WK0 817865FPSP500 00000029880M32 SPRINGWOODP500 00000008965M36 281274MSP500D 00001027854WK7 281274P5DGR 00001025591WK7 281274WHSLP500D 00001004968WK2 281725P500D1 29878M36 325235RF 00001026469WK5 409938P500 00000112670DP5 903SJ500 00000003201M31 361572RF1 1038492WK3 361788P500 00000007827M39 972923RF1 00000012659M39 817599P500 00000025568M37 281338BAP500 1033337WK5 940764P500 00001000998WK3 940764ROOMD 00000007383M33 361293P500 00000001839M30 801292PR5 00000008412M39 435734P500 00000006151M35 208455RF2 00000023600M30 801657P500 00000095805DP8 801479RM5 00000006556M35 801714RM1 00000011433M30 435658RF 00000987065WK0 435637RM4 00000008195M30 435893RM2 00001026424WK0 435722P500 00000013598M38 801314ENDO 00000999913WK7 801408RM22 00000009804GP6 801408RM28 00000005339M37 801535RM24 00000018236M30 435789RF 00000005208M34 801957R&F 00000011820M38 703849RF3 00000001438M31 703558ROOM3 00000002118M38 MONT500D 00001026027WK1 434654RM3 00001026146WK9 434654RM4 00000987756WK4 434924UVAP1 00000991646WK1 434924UVAP2 00000981075WK5 434982RF1 00001003383WK5 540829RF1 00001026978WK5 540CULRF2 00001029429WK6 434799P500D 1036509wk6 571349P500 00000030800M37 434392PRE500 00000008411M31 757314RF3 00001019945WK3 540498RM1 00000012867M38 540498RM3 00000136742DP4 757722C238 00000006334M37 757722RM281 1034587WK4 757736RM1 1035406WK6 757736RM2 00000012658M31 703481P500 00000003462M39 804541JR4 00001027089WK0 276666P500 1034116WK2 804764P500RM3 00000029202M39 757886RF2 00000023609M31 757261RM10 00000999394WK0 804323P500A 00000013192M30 540921P500 00000008563M39 804HDMCRM1 1038732WK2 757398RM5P 00000002637M37 757314SRM1 00000009278M33 540980P500D 00000983536WK4 703689RM03 1038785WK0 276596P500D4 00001029435WK3 804254RMC 00000013219M31 804330JW4A 00001032956WK3 804967RM5 1038874WK2 804285XRF119 00001033796WK2 540489P500 00000978795WK3 540776RF1 00001033971WK1 540776RM2 00000029355M35 434447CMRM3 00000016933M34 AM13MFK9 Not Available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ot Available 509474SR3FP 00000979396WK9 253627RM1500D 00000003830M37 253ECN133995GI 00000993169WK2 509865TCRF1 1035248WK2 360VRP500 Not Available AM12YTN24 00000018535M35 509575YMRF2 00001002916WK3 509575YP500 00000982743WK7 509575YRX6 000000GDXR3120 715268P500 00001027862WK0 715623P500 00001026466WK1 608364P5002 00000009007M36 608364P501 1035222WK7 262785P500 00001033574WK3 262767P500 1033158WK5 608930P500 1035405WK8 262741P500 00000013848M37 920288P500 00000002119M36 608756P500 00000006310M37 262656AP500 00000006862M37 262656GIP500 00001001859WK6 262656P500 00001028638WK3 608775RF500 00000989937WK8 608785P500 00001025586WK7 715532RF 00000987066WK8 608287PRE500 00000004860M33 608847P500 00000011468M36 715748P500 00000012107M39 262646P500 30543M33 414447P502 00000998430WK3 414805P500 00001034395WK2 414805P502 00001001061WK9 414649P500 00000024662M39 414649P500D1 00001001864WK6 608324P501 00000005035M31 608324P502 00000013174M38 262577P500 00000030640M37 262577P502 00001004242WK2 608745P500 00000983135WK5 715361PRE1 00001031632WK1 608647P500 00000989936WK0 608372P500 00000978178WK2 715453PRE500 00001024744WK3 920262P500 00000006506M30 262569P500 00001035905WK7 262696CCP500 1034108WK9 262544P500 00000999916WK0 715847P500 00000010496M38 414328P500 00000998424WK6 715393PRE500 00000009009M32 715538P500 00001034392WK9 715421P500 00000120792DP7 304255P500 00000012535M31 304256PRE1 00000979395WK1 304379RF 00000983423WK5 304473RF1 00000024067M31 304636P500D 1034096WK6 304367P500D1 1035225WK0 304329P500 00000981006WK0 304792PRE1 00000024966M34 3042631D118A 00000025939M30 304257P500 00001033969WK5 304436P500D 00001027092WK4 307684P500 00000994416WK6 307778500D 00000069614DP6 307634RF Not Available 307332PRE500 00000005997M32 307548RF1 00000006857M37 307856P500 00000005155M37 307362P500 00001027091WK6 307864P500D 00000000031140 XR31140 00000004182M32 250426RF2 00000107133DP1 204239RAD2 00000990409WK5 204785RF1 00000992838WK3 204677PREC500 00000065991DP2 204787RF1 00000984222WK0 506857P500 00000007828M37 506857P500B 00000001435M37 705356PR500 00000005094M38 519824RF 00000005809M39 905572RF 00001002438WK8 416813PR500FL1 00001006926WK8 416813PR500FL2 00001005396WK5 416925PR500 00000973963WK2 519254RM5 00001025587WK5 CANFPRECRX1 00000112679DP6 0566557115XR1 00001004650WK6 22105XF1 00000000000492 2717616XRI-A 00000019487M38 752026XR05 00000010388GP7 124861XR03 00000017918M34 PC0362XR12 00000010958GP7 HC7462XR03 FMI10902DUMMYH 352387XR01 00001031637WK0 XR1031637WK0 1017201WK3 2767709RF 00001000381WK2 0835080021 00001005378WK3 IE1100RX01 00000006616M37 IL1062XR13 00000006690M32 IL1000XR06 00000978176WK6 A5639304 00001028083WK2 A5745526 00001009278WK1 C5123628 00000997105WK2 2573108P500D 00001032524WK9 XR25240M33 1022738WK7 JO1129XR02 1023776WK6 JO1034XR04 1022740WK3 KW1010RX04 00000997500WK4 KW1013RX08 00000001660M30 KW1013RX15 00000002873M38 KW1085RX01 00001021391WK6 KW1017RX05 00001021538WK2 KW1017RX06 00001015275WK9 KW1007RX04 00000996766WK2 KW1007RX03 00001004450WK1 KW1018RX09 00001009914WK1 LY1018RX02 00000993953WK9 MT1014RX08 982575WK3 2514890PREC301 00000997112WK8 2564624PRE 00000028574M32 XR1035410WK8 132814DF5 XR27465M34 00001010556WK7 787655P500D 1002809WK0 787653R&FP5D 00000008137M32 RU1838RX03 00000028177M34 RU2366RX01 00001029000WK5 RU7316RX01 00000025563M38 RU8615RX01 00000028068M35 RU2370RX01 00000007655M34 RU1217RX03 00001015282WK5 RU1217RX04 00000005493M32 SA2135RX02 00000010623M37 SA2166RX01 00000003822M34 SA1028RX01 00000004717M35 SA1023RX01 00000004228M33 SA2068RX02 00000993949WK7 SA1054RX05 00000013048GP4 SA1042RX08 00000004244M30 SA2129RX01 00000001659M32 FP970407 00000024938M33 0856087000 00000016935M39 AE1021RX04 00000989488WK2 05169FLU01 00000003978M34 05228FLU02 00000012826GP4 00105FLU01 00000994135WK2 77257FLU03 00000016714M38 IE1545RF01 00001003382WK7 00090FLU02 00000010494M33 00149FLU03 Not Available 77064FLU01 00000020067M35 01345FLU02 Not Available 10537FLU01 00000984899WK5 05104FLU01 00000003823M32 10177FLU01 00000003976M38 00386FLU02 00000005810M37 00572FLU02 00000002463M38 00471FLU03 00000016074M37 10448FLU03 00000007385M38 00065FLU01 00000007451M38 00065FLU02 00000010959GP5 01171FLU02 00000013663M30 00824FLU02 00001008707WK0 00490FLU01 00000002868M38 05229FLU02 00000007585M33 00044FLU01 00000005490M38 10036FLU01 00000000000690 0859080010 00000000025034 0859080009 00000300797WK8 205663SR1 00000066779EC0 205880CR1 00000983375WK7 256332RF2 371969 501968SM2 00000466739WK0 818963RM4 89850TY2 RADNETRF186 00000272974WK7 714824DR 00000401661WK4 909825R2 00000393202WK7 909825R3 00000384688WK8 407896RF6 00000486316WK3 407335RM2 00000445417WK9 813633RFX 00000265744WK3 404737SA1 00000000QA9218 404283E1 00000217293WK0 404882WRF1 00000322522WK4 404291RRF1 904976 912564SC1 00000306212WK2 708795MC4 00000315105TU7 312792RF2 00000338001WK9 847327RF 00000450292WK8 504867LRF2 00000015276-W5 504652RP2 00000435636WK6 504588TRF6 000FMITR004343 505426GERF1 00000524395WK1 505622ADVANTX 00000252886WK7 918341CR3 00000052615WK7 405271PRF4 00000016976EA3 215248RF5 00000334973WK5 412647ADV 00000327674WK8 615855EX2 00000441389WK4 615871SRF1 00000288968WK1 615443UOP1 00000106885WK7 804541JR3 00000098113WK4 703980PRA 242550TU2 304327RF2 00000104926VP1 EG1055RX01 Not Available M4198113 00000327906WK4 IL1084VA01 00000025392BU2 0834790001 00000507885WK2 0834070014 00000052597ES6 0834080009 00000483582WK3 SK1005RX05 00000409109BU6 0856080143 00000030245EC5 205793FADV 00000202488WK3 334213BORF2 00000507540WK3 334277SFX 00000551006WK0 205761RF1 00000056146EC4 205345RF2 00000481398WK6 501329CRF 00000523827WK4 501663BRF1 00000344139WK1 501782HK1 00000536993WK9 520758RF1A 00000538116WK5 520758RF2A 00000281490WK3 805KERN159 00000281486WK1 805KERN160 00000211028WK6 714735F1 00000498993WK5 707443JR5 00000548588WK3 323669ADV 00000270395WK7 RADNETRF131 00000523826WK6 805497RM7 00000037671TY5 RADNETRF108 00000506737WK6 415982C2A 00000504899WK6 415565DFL2 00000442423WK0 415641ADV 00000215945WK7 RADNETRF184 00000485041WK8 805648RM5 00000332701WK2 714466F1 00000482434WK8 407897RF2 00000215938WK2 305468RM3A 00000210515WK3 904752AMI 16572 813586RF1A 00000277957WK7 813586RF2A 00000322168WK6 904526JR2 00000952479WK4 904272SFX1 00000409026WK2 407830RF1 00000972379WK2 FHFGRF1 00000091562TX9 305587ADVTX3 00000539263WK4 941423RF1 53217678 352597RF1 00000496289WK0 407646RF6 00000039794EC3 1009001 Not Available 706632FRF 00000528289WK2 912264SCPU 00000037778TY8 1323134 00000512913WK5 770382WCRF 00000516908WK1 706695MX3 00000483581WK5 770918RF3 00000524487WK6 912751RF5 00000481390WK3 912982XR1 00000505561WK1 706444XF 00000510381WK7 319293SFX Not Available SHSSH712GEXR36 00000440237WK6 312282RF1 00000551004WK5 708388SFX 00000476638WK2 812232TH4 00000541868WK6 316724CX1 00000497964WK7 316421LR2 00000487874WK0 502781G2 00000510836WK0 504885RF1 00000497130WK5 318377MR3 00000268239WK1 318539SRM2 00000548549WK5 301809RF 00000002524BI7 301868RF1 00000459481WK8 313561RF 00000416096WK6 612689RF2 00000493274WK5 612520N7 00000025362ES5 507284CW49 00000544944WK2 651565RF 00000117749TX2 816836GIXR 00000339304WK8 601226RF1 00000420775WK9 601872OR02 00000497129WK7 601636RF2 00000513709WK6 919443NDRF2 00000551005WK2 336372AR2 00000471083WK6 704586RF2 00000100431EC6 201858RM3 00000288019WK3 RADNETRF1293 00000056893DP1 AMIRF 00000471748WK4 505887GRF1 00000485053WK3 505392RF2 00000227405WK8 702731H09 00000362930WK0 516676AVTX 00000056081TY3 RADNETRF1240 00000117176TX8 RADNETRF1221 00000043767EC3 212697RF1 00000271412WK9 516763RM4 00000536752WK9 918333JRF2 00000359839WK8 918343CR1 00000516247WK4 918225RF1 00000467272WK1 405233SRF1 00000440238WK4 814375R1 Not Available 215335ADVSFX 00000263624WK9 615762C02 00000506739WK2 901352R1 00000427246WK4 615781SH02 00000331137WK0 915646BRF 00000482436WK3 915646RM1 00000034706EX2 214420LR2 00000290089WK2 936WHRF3 00000073172TY9 210SWRF 00000372306WK1 512353CRM2 00000430776WK5 703849SFX 00000481391WK1 804392SCARF1 00000550353WK7 703689RM02 00000500807WK3 414529RF1 00000438578WK7 304256BX4 00000504292WK4 304327B05 00000082726EC1 204324RF1 00000003507BC1 204745RF1 00000461692WK6 506356RF1 00000546364WK1 902224RF2 00000202898WK3 902849ARM3 00000028342EX4 33863SFX 00000549583WK3 75658ADVSFX Not Available A4167308 Not Available 062025002 000000000DY961 M5460625 00000345930WK2 PC0008XR02 00000042800EX3 H2197RAD01 Not Available HU1099RX34 Not Available HU1099RX35 Not Available HU1099RX37 Not Available HU1012RX02 00000397477WK1 0834080002 00000399544WK6 0834080005 00000440234WK3 0834080004 00000393378WK5 6984SFX Not Available 85227SCPUPED 00000483580WK7 85226ADVANTX 00000523825WK8 PER39500 00000489901WK9 SK1005RX07 00000481395WK2 00573FLU02 00000481102WK2 00658FLU01 00000459967WK6 01006FLU02 00000388024WK2 05461FLU02 00000262756WK0 00768FLU01 00000456906WK7 00244FLU01 |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | GE Healthcare Service 800-437-1171 |
Manufacturer Reason for Recall | Reported incidents of a patient step detaching from the table. A fall from a patient step detaching while in use could result in an injury to a patient or operator. |
FDA Determined Cause 2 | Device Design |
| Action | Consignees were sent on 2/6/2017 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# FMI 10908 dated February 3rd, 2017. The letter was addressed to Director of Radiology, Risk Manager/Hospital Administrator & Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact consignee to arrange for the correction.. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. |
| Quantity in Commerce | 2,380 (2,215 US; 165 OUS) |
| Distribution | US: Nationwide including DC and Puerto Rico.
OUS:
Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Egypt,
France, Germany, Greece, Honduras, Hungary, Indonesia, Ireland, Israel,
Italy, Jamaica, Jordan, Kuwait, Libya, Malaysia, Malta, Mexico, Panama,
Portugal, Russia, Saudi Arabia, Saudi Arabia, Slovakia, Spain,
Thailand, United Arab Emirates, United Kingdom, Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IXR 510(K)s with Product Code = IXR 510(K)s with Product Code = IXR
|
|
|
|