Class 2 Device Recall Dimension Vista Estradiol (E2) Flex reagent cartridge |
|
Date Initiated by Firm |
March 21, 2017 |
Date Posted |
May 05, 2017 |
Recall Status1 |
Terminated 3 on August 03, 2018 |
Recall Number |
Z-2608-2017 |
Recall Event ID |
76915 |
Product Classification |
Radioimmunoassay, estradiol - Product Code CHP
|
Product |
Dimension Vista Estradiol (E2) Flex reagent cartridge |
Code Information |
Device Listing Number # D121541 Lot # 16148AA, 16190BA, 16201BA, 16228BC, 16256BA, 16277BB 16298AB, 16343BE,17005BC |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 500 Gbc Dr Ms 514 PO BOX 6101 Newark DE 19702-2466
|
For Additional Information Contact |
800-441-9250
|
Manufacturer Reason for Recall |
The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Siemens mailed an Urgent Medical Device Correction Letter to affected customers on 03/29/2017 to inform them of the issue. The notice explained the reason for the field action, Risk to Health and actions to be taken by the customer. |
Quantity in Commerce |
1,697 units |
Distribution |
Product was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada
Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria
Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland
India,Italy,Japan,Canary Islands,Republic of Korea,Kuwait,Lebanon,Latvia
Macedonia,Netherlands,Norway,New Zealand,Philippines,Poland,Portugal
Qatar,Saudi Arabia,Slovenia,Slovakia,Turkey,Taiwan,South Africa |