• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall NMT Patient Cable (989803174581)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall NMT Patient Cable (989803174581)see related information
Date Initiated by FirmMay 04, 2017
Create DateMay 31, 2017
Recall Status1 Terminated 3 on June 28, 2018
Recall NumberZ-2175-2017
Recall Event ID 77166
510(K)NumberK122439 
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
ProductNMT Patient Cable (989803174581) in use with lntelliVue NMT Module (Part# 865383)
Code Information System Serial Number: DE24702801 DE24702582 DE33103000 DE33103006 DE33104465 DE33104466 DE33104469 DE33104473 DE33104497 DE44306376 DE44305951 DE33105241 DE33105483 DE44306622 DE44306625 DE44306626 DE44306644 DE44306645 DE44306657 DE44306604 DE44306608 DE44306638 DE44306639 DE33103007 DE33103055 DE33103062 DE33103493 DE44307378 DE44307381 DE44307385 DE44307391 DE44307392 DE44307395 DE24702503 DE24702507 DE24702508 DE24702509 DE24702510 DE24702511 DE24702512 DE24702513 DE24702514 DE24702515 DE24702516 DE24702517 DE24702518 DE24702520 DE24702521 DE24702563 DE33104560 DE33104569 DE24702655 DE24702656 DE44305701 DE44305702 DE44305706 DE44305709 DE44305734 DE44305745 DE44305746 DE44305748 DE44305751 DE24702492 DE33105536 DE44305884 DE33105413 DE33105416 DE33105432 DE44305788 DE24702254 DE24702257 DE24702262 DE24702265 DE33103146 DE33103147 DE33103154 DE33103144 DE33103577 DE33103457 DE33103458 DE33103912 DE33104871 DE33104881 DE33104878 DE33104879 DE44307338 DE44307311 DE44307327 DE44307330 DE44306896 DE33105420 DE33105556 DE33105558 DE44307636 DE44307609 DE44307610 DE44307611 DE44307612 DE44307616 DE44307617 DE44307619 DE44307620 DE44307621 DE44307628 DE44307635 DE44307637 DE22801752 DE44305695 DE44305696 DE44305697 DE44305707 DE44305710 DE44305711 DE44305712 DE44305717 DE44305728 DE44305729 DE33103729 DE33103731 DE33103735 DE33103737 DE33103738 DE33103739 DE33103742 DE33103749 DE33103989 DE33103999 DE33103896 DE33105418 DE44306377 DE44306384 DE44306385 DE44306388 DE44306389 DE44306391 DE44306394 DE44306395 DE44306396 DE44306397 DE44306399 DE44306400 DE44306401 DE44306402 DE44306403 DE44306404 DE44306405 DE44306406 DE44306407 DE44306408 DE44306411 DE44306412 DE44306413 DE44306175 DE33103068 DE33103079 DE33103090 DE33103094 DE33103017 DE33103019 DE33103020 DE33103060 DE33103066 DE33103067 DE33103069 DE33103071 DE33103072 DE33103080 DE33103083 DE33103086 DE33103087 DE33103088 DE33103089 DE33103075 DE33103093 DE33103102 DE33102973 DE33102974 DE33102969 DE33102978 DE33102982 DE33102989 DE33102990 DE33102991 DE33102993 DE33102994 DE33102995 DE33102999 DE33103001 DE33103004 DE33103005 DE33103028 DE33103029 DE33103030 DE33103031 DE33103032 DE33103033 DE33103034 DE33103035 DE33103036 DE33103039 DE33103043 DE33102964 DE33102976 DE33102977 DE33102979 DE33102980 DE33102981 DE33102983 DE33102984 DE33102985 DE33102986 DE33102987 DE33102988 DE33102992 DE33103572 DE33103588 DE33103565 DE44307794 DE22801748 DE33103364 DE33103377 DE33103378 DE33103380 DE33103381 DE33103382 DE33103384 DE22801766 DE22801768 DE22801769 DE22801770 DE24702361 DE24702728 DE24702729 DE24702730 DE24702747 DE24702752 DE24702754 DE24702755 DE24702757 DE24702759 DE24702761 DE24702753 DE24702748 DE24702746 DE33104503 DE33104510 DE33104511 DE33104524 DE44307581 DE44307585 DE44307186 DE33105011 DE33105012 DE33105015 DE33105548 DE33103397 DE33103402 DE33105208 DE33105211 DE33105252 DE33105253 DE33105254 DE44307005 DE44307006 DE44307322 DE44307325 DE44307326 DE44307329 DE44307202 DE44307232 DE44307237 DE24702497 DE33104975 DE33105000 DE33105001 DE33105540 DE44307248 DE44307244 DE44307253 DE44307316 DE44307317 DE44307319 DE44307320 DE44307324 DE44307286 DE44306239 DE44306250 DE44306256 DE44306291 DE44306292 DE44307181 DE44306233 DE44306245 DE44306421 DE44306428 DE44306432 DE44306439 DE44306230 DE44306231 DE44306280 DE44306284 DE44306247 DE44306248 DE44306158 DE44306163 DE44306169 DE44306172 DE44306185 DE44306186 DE44306187 DE44306188 DE44306189 DE44306190 DE44306191 DE44306192 DE44306193 DE44306194 DE44306195 DE44306196 DE44306197 DE44306198 DE44306199 DE44306200 DE44306201 DE44306202 DE44306203 DE44306204 DE44306216 DE44306222 DE44306223 DE44306224 DE44306232 DE44306234 DE44306235 DE44306236 DE44306237 DE44306244 DE44306246 DE44306254 DE33103389 DE33103405 DE44305926 DE44305931 DE44305939 DE44305940 DE44305970 DE33103348 DE33103357 DE33103360 DE33103361 DE33103362 DE33103363 DE44306074 DE44306075 DE44306080 DE44306961 DE44306962 DE44306963 DE33103915 DE33103926 DE33105189 DE33105192 DE33105195 DE33105198 DE33105199 DE33105203 DE33105232 DE33105264 DE33105266 DE33105268 DE44305786 DE44305801 DE44305945 DE44305946 DE44305947 DE44305948 DE44305950 DE44305952 DE44305953 DE44305954 DE44305955 DE44305956 DE44305957 DE44305958 DE44305959 DE44305960 DE44305961 DE44305963 DE44305964 DE44305968 DE44305969 DE44307011 DE44307012 DE44307013 DE44307018 DE44307020 DE44307022 DE44307025 DE44307016 DE44307027 DE22801772 DE33103443 DE22801780 DE24702288 DE24702359 DE33103633 DE33103639 DE33103648 DE33103655 DE33103668 DE44306574 DE44306756 DE44306757 DE44306759 DE44306760 DE44306761 DE44306762 DE44306764 DE44306765 DE44306767 DE44306768 DE44306770 DE44306772 DE44306057 DE44306061 DE44306062 DE44306065 DE44306067 DE44306072 DE44306321 DE44306322 DE44306323 DE44306324 DE44306325 DE44306326 DE44306327 DE44306328 DE44306329 DE44306330 DE44306331 DE44306332 DE44306333 DE44306334 DE44306335 DE44306337 DE44306338 DE44306339 DE44306340 DE44306341 DE44306342 DE44306343 DE44306344 DE44306345 DE44306346 DE44306347 DE44306348 DE44306349 DE44306351 DE44306352 DE44306353 DE44306354 DE44306355 DE44306356 DE44306357 DE44306358 DE44306359 DE44306360 DE44306361 DE44306316 DE44306320 DE44306362 DE44306363 DE33105489 DE33105491 DE33105498 DE33105510 DE33105511 DE33105513 DE22801793 DE22801797 DE44306828 DE44306829 DE44306830 DE44306834 DE44306837 DE44306847 DE44306766 DE44306777 DE44306778 DE44306784 DE44306796 DE44306806 DE44307728 DE44307300 DE44306452 DE44306310 DE44306652 DE44306664 DE44306602 DE33103442 DE33103444 DE33103445 DE33103446 DE33103447 DE44307776 DE22801767 DE22801781 DE33103866 DE44306557 DE24702766 DE24702771 DE44307394 DE44306904 DE44307523 DE33105533 DE44306183 DE33105516 DE33105523 DE33105528 DE33105529 DE33105530 DE33105534 DE33105539 DE33105541 DE33105546 DE33103350 DE33103352 DE33103353 DE33103356 DE24702659 DE24702663 DE24702668 DE24702670 DE24702672 DE24702691 DE44306093 DE33103365 DE24702347 DE44307375 DE44307377 DE44307383 DE44307398 DE44306019 DE44306020 DE44306027 DE44306034 DE44306042 DE44305985 DE44306012 DE44306013 DE44305973 DE44305980 DE44305981 DE44305986 DE44305987 DE44305977 DE44307084 DE44307127 DE44306798 DE44306799 DE44306800 DE44306801 DE44306802 DE24702291 DE33103440 DE33103349 DE22801746 DE44307753 DE44307767 DE44307768 DE44307769 DE44307771 DE44307773 DE44307780 DE44307785 DE33102966 DE24702592 DE24702735 DE44306958 DE44307470 DE44306889 DE44307340 DE44307372 DE44307332 DE44307341 DE44307528 DE44307551 DE44307387 DE44307401 DE44307402 DE44307406 DE44307412 DE44307345 DE44306725 DE33105519 DE33105555 DE33105559 DE33105572 DE33105608 DE33104537 DE33103480 DE33104705 DE33104706 DE33104707 DE33104708 DE33104709 DE33104710 DE33104711 DE33104712 DE33104721 DE33104731 DE33104732 DE33104733 DE33104734 DE33104735 DE33104736 DE33104737 DE33104738 DE33104739 DE33104740 DE33104742 DE33104743 DE33104744 DE33103469 DE33103472 DE33103473 DE33103482 DE33104807 DE33104812 DE33104813 DE33104822 DE33104824 DE33104838 DE33104840 DE33104842 DE33104808 DE33104809 DE33104810 DE33104814 DE33104815 DE33104818 DE33104820 DE33104821 DE33104825 DE24702751 DE44305928 DE44305934 DE44305935 DE44305937 DE44305966 DE44305972 DE44305915 DE44305925 DE33103276 DE24702413 DE24702414 DE24702415 DE24702417 DE24702475 DE24702479 DE24702487 DE33104459 DE33104460 DE33104463 DE33104464 DE33104467 DE33104470 DE33104461 DE33104477 DE33104486 DE33104493 DE33104495 DE33104479 DE33104098 DE33104104 DE33104482 DE33104483 DE33103084 DE33103313 DE33103722 DE33104212 DE33104224 DE44307512 DE44307514 DE44307515 DE44307519 DE44307486 DE44307487 DE44307494 DE44307495 DE44307497 DE44307498 DE44307480 DE44307482 DE44307483 DE44307484 DE44307485 DE44307489 DE44307490 DE44307491 DE44307492 DE44307493 DE44307481 DE44307499 DE33103990 DE33104256 DE33104252 DE44306305 DE44306306 DE44306307 DE44306308 DE33104816 DE33105341 DE33105352 DE33105526 DE44305846 DE44305886 DE44306137 DE44306139 DE44306130 DE44306392 DE44307105 DE44306562 DE44306564 DE44306566 DE44306567 DE44306581 DE44306603 DE44306605 DE44306607 DE44306609 DE44306612 DE44306614 DE44306615 DE44306616 DE44306619 DE44306620 DE33104870 DE33104877 DE33104893 DE44305916 DE44305917 DE44305919 DE44305920 DE33103056 DE44306141 DE24702763 DE33104635 DE33104644 DE33104632 DE33104637 DE33104638 DE33104626 DE33104550 DE33104551 DE33104552 DE33104555 DE33104557 DE33104575 DE33104576 DE33104577 DE33104578 DE33104579 DE33104580 DE33104581 DE33104582 DE33104583 DE33104584 DE33104585 DE33104586 DE33104587 DE33104588 DE33104589 DE33104590 DE33104591 DE33104592 DE33104593 DE33104594 DE33104595 DE33104596 DE33104597 DE33104600 DE33104601 DE33104602 DE33104630 DE44307692 DE33103552 DE33103756 DE24702046 DE24702064 DE24702055 DE44307211 DE44307218 DE33104076 DE33104078 DE33103288 DE33103291 DE44306478 DE44306483 DE44306484 DE44306486 DE44306487 DE44306491 DE44306492 DE44306493 DE44306494 DE44306495 DE44306496 DE44306497 DE44306498 DE44306499 DE44306500 DE44306501 DE33104209 DE33104225 DE33104227 DE33104231 DE33104249 DE33104251 DE33104253 DE33104258 DE33104260 DE44307723 DE33104533 DE24702348 DE24702351 DE24702352 DE24702356 DE44307083 DE44307087 DE33104111 DE33104118 DE33104120 DE33104127 DE33104128 DE44306642 DE44306647 DE44306648 DE44306660 DE44306662 DE44306663 DE33103441 DE33103451 DE33105490 DE33105500 DE24702928 DE24702932 DE24702597 DE24702599 DE24702608 DE24702610 DE33103049 DE33103074 DE33103095 DE22801533 DE22801535 DE22801547 DE22801553 DE22801554 DE22801555 DE33104294 DE33103369 DE33103379 DE33103383 DE22801536 DE22801334 DE24702282 DE33102965 DE33105298 DE33105299 DE33105302 DE33105210 DE33105336 DE33105337 DE33104755 DE33104756 DE33104760 DE33104764 DE33104766 DE33104768 DE33104769 DE33104770 DE33104773 DE33104796 DE33104799 DE33104805 DE33104828 DE44306312 DE44306315 DE44306350 DE44306364 DE44306365 DE44306366 DE44306367 DE44306368 DE44306369 DE44306370 DE22801063 DE22801064 DE22801122 DE22801123 DE22801278 DE22801280 DE22801281 DE22801282 DE22801283 DE22801284 DE22801118 DE22801301 DE22801297 DE22801330 DE22801331 DE22801332 DE22801333 DE22801335 DE22801318 DE22801326 DE22801327 DE22801328 DE22801299 DE22801306 DE24702181 DE24702215 DE24702219 DE33103569 DE44305962 DE44305971 DE22801329 DE33104610 DE33104624 DE24701990 DE24702000 DE24702024 DE24702040 DE44305767 DE44305780 DE44305782 DE44307622 DE44307623 DE44307624 DE44307625 DE44307626 DE44307627 DE44307629 DE44307633 DE44307634 DE44307630 DE44307631 DE44307646 DE44307349 DE44307651 DE44307632 DE24702875 DE24702887 DE24702895 DE24702868 DE24702885 DE22801315 DE24702489 DE33105137 DE33105138 DE33105146 DE33105148 DE33105150 DE33105158 DE33105169 DE33105170 DE33105174 DE33105181 DE22801310 DE22801689 DE44307064 DE44307070 DE44307075 DE44307081 DE22801321 DE22801322 DE44307678 DE22801309 DE24702343 DE24702344 DE24702349 DE24702357 DE24702386 DE24702399 DE33105073 DE33105077 DE22801320 DE24702339 DE33105113 DE33105117 DE44306055 DE44306060 DE44306071 DE44306073 DE33103042 DE44307380 DE24702133 DE24702145 DE24702148 DE33105549 DE33105568 DE33105575 DE33105576 DE33104472 DE33104487 DE33104491 DE22801382 DE22801290 DE22801288 DE24702849 DE33103299 DE44306438 DE44306450 DE44306453 DE44306455 DE44306456 DE44306457 DE44306458 DE44306459 DE44306463 DE44306479 DE33105550 DE33105573 DE33105592 DE33105602 DE33105609 DE44307379 DE44307588 DE44307589 DE44307597 DE44307599 DE44307606 DE44307608 DE44305982 DE44305993 DE44307700 DE44307703 DE44307711 DE24702279 DE24702292 DE24702293 DE24702294 DE24702295 DE24702299 DE24702300 DE24702303 DE24702306 DE24702363 DE24702362 DE24702593 DE24702594 DE24702595 DE24702596 DE24702598 DE24702604 DE24702605 DE24702609 DE24702614 DE24702615 DE24702616 DE24702617 DE24702618 DE24702619 DE24702620 DE24702623 DE24702612 DE24702613 DE24702621 DE24702622 DE24702624 DE24702625 DE24702626 DE24702627 DE24702628 DE24702629 DE24702630 DE24702631 DE24702632 DE24702644 DE24702648 DE24702636 DE33104047 DE33104050 DE33103603 DE33104311 DE33104312 DE33104313 DE33104314 DE33104315 DE33104316 DE33104317 DE33104318 DE33104319 DE33104320 DE33104322 DE33104326 DE33104335 DE33104346 DE33104349 DE33104351 DE33104353 DE33104355 DE33104359 DE33104361 DE33104362 DE44305887 DE44305888 DE44305890 DE44305894 DE44305897 DE44305899 DE22801427 DE33104122 DE33103906 DE22801497 DE22801498 DE22801499 DE22801500 DE22801502 DE22801503 DE22801504 DE22801506 DE22801508 DE22801509 DE22801510 DE22801511 DE22801512 DE22801513 DE22801514 DE22801515 DE33103988 DE22801400 DE22801410 DE22801454 DE24701873 DE24701880 DE24701897 DE24701900 DE24701909 DE24701917 DE33104885 DE33104888 DE33104900 DE33105118 DE33105135 DE33105136 DE33105139 DE33105149 DE33105151 DE33105152 DE33105153 DE33105154 DE44307469 DE44307475 DE44306993 DE44306987 DE24702370 DE44307420 DE44307421 DE44307422 DE44307423 DE44307424 DE44307425 DE44307426 DE44307427 DE44307428 DE44307429 DE44307430 DE44307431 DE44307432 DE44307433 DE44307434 DE44307435 DE44307436 DE44307437 DE44307438 DE44307439 DE44307440 DE44307441 DE44307442 DE44307443 DE44307444 DE44307445 DE44307446 DE44307447 DE44307448 DE44307449 DE44307450 DE44307451 DE44307452 DE44307453 DE44307454 DE44307455 DE44307456 DE44307457 DE44307458 DE44307459 DE44307460 DE44307461 DE44307462 DE44307463 DE44307464 DE44307465 DE44307466 DE44307471 DE44307472 DE44307473 DE44307476 DE44307477 DE44307478 DE44307479 DE44306722 DE33104971 DE24702364 DE24702368 DE44307152 DE44307732 DE33105585 DE44305713 DE44305716 DE44305721 DE44305725 DE44305726 DE44305731 DE33104337 DE33104343 DE33104365 DE33104368 DE33104369 DE33104370 DE33104375 DE33104376 DE33104384 DE33104398 DE33104402 DE33104399 DE33105521 DE44306043 DE44306044 DE44306085 DE44306087 DE44306089 DE44306090 DE44306091 DE44306094 DE44306095 DE44306096 DE44306097 DE44306098 DE44306099 DE44306101 DE44306429 DE44306435 DE44306451 DE44306629 DE44306630 DE44307173 DE33104079 DE33104089 DE33104554 DE33104556 DE33104558 DE33104559 DE33104571 DE44307736 DE44307737 DE44307738 DE44307739 DE44307740 DE44307741 DE44307742 DE44307744 DE44307745 DE44307746 DE44307747 DE44307748 DE44307749 DE44307750 DE44307751 DE44307752 DE44307788 DE44307789 DE44307792 DE33105297 DE22801460 DE33102967 DE33104496 DE33104492 DE33104498 DE33103605 DE33105168 DE44306610 DE44306611 DE44305634 DE44305636 DE44305640 DE33104071 DE33104075 DE33103407 DE33104488 DE33104546 DE33105064 DE33105069 DE33105384 DE33105520 DE33105288 DE33105334 DE33105360 DE33105361 DE33104977 DE33105262 DE33105259 DE33105263 DE33105296 DE33105323 DE33105349 DE33105350 DE33105345 DE33105357 DE44305800 DE44305812 DE44305978 DE44305992 DE44306832 DE44307061 DE44307062 DE44307065 DE44307066 DE44307125 DE44307135 DE44307137 DE33103394 DE33103489 DE33103502 DE33103843 DE33103846 DE44307138 DE33103862 DE44307305 DE44307309 DE44307808 DE44307796 DE44307810 DE33103894 DE44307763 DE33104158 DE33105274 DE33105304 DE44307082 DE44306573 DE33104090 DE33104648 DE33105106 DE33105107 DE33105130 DE33105260 DE33105542 DE44306217 DE44306669 DE33105110 DE33105119 DE33105129 DE33105284 DE33105217 DE33105221 DE33104155 DE33104157 DE33104714 DE33104625 DE33104627 DE33104978 DE44306041 DE44306135 DE44306136 DE44306691 DE44306697 DE44306701 DE44306703 DE44306705 DE44306708 DE44306709 DE44306710 DE44306711 DE44306713 DE44307121 DE33105124 DE33105251 DE44306667 DE44306674 DE44306309 DE44306313 DE44307130 DE44307132 DE44307133 DE44307139 DE33104108 DE33104110 DE44307140 DE44307150 DE44306883 DE44306901 DE44306905 DE44307073 DE44307085 DE44307077 DE44307124 DE44307396 DE44306053 DE33104902 DE33104912 DE33104917 DE44306946 DE44306967 DE22801006 DE22801007 DE22801008 DE22801010 DE22801011 DE22801012 DE22801014 DE22801021 DE22801023 DE22801024 DE22801025 DE22801036 DE22801037 DE22801046 DE22801047 DE22801048 DE22801049 DE22801050 DE22801053 DE22801054 DE22801055 DE22801056 DE22801057 DE22801058 DE22801061 DE22801062 DE22801066 DE22801067 DE22801069 DE22801070 DE22801071 DE22801073 DE22801074 DE22801075 DE22801076 DE22801078 DE22801079 DE22801080 DE22801081 DE22801082 DE22801083 DE22801084 DE22801085 DE22801086 DE22801087 DE22801089 DE22801090 DE22801093 DE22801099 DE22801106 DE22801072 DE22801077 DE22801088 DE22801092 DE22801094 DE22801096 DE22801097 DE22801100 DE22801101 DE22801102 DE22801103 DE22801104 DE22801105 DE22801107 DE22801108 DE22801109 DE22801110 DE22801111 DE22801112 DE22801113 DE22801114 DE22801115 DE22801116 DE22801117 DE22801119 DE22801120 DE22801121 DE22801124 DE22801125 DE22801126 DE22801127 DE22801128 DE22801130 DE22801131 DE22801133 DE22801134 DE22801135 DE22801136 DE22801137 DE22801138 DE22801139 DE22801140 DE22801141 DE22801142 DE22801143 DE22801145 DE22801146 DE22801147 DE22801148 DE22801166 DE22801150 DE22801151 DE22801152 DE22801153 DE22801154 DE22801155 DE22801156 DE22801157 DE22801158 DE22801159 DE22801160 DE22801161 DE22801162 DE22801163 DE22801164 DE22801165 DE22801167 DE22801168 DE22801169 DE22801170 DE22801171 DE22801172 DE22801173 DE22801174 DE22801175 DE22801176 DE22801177 DE22801178 DE22801179 DE22801180 DE22801181 DE22801182 DE22801183 DE22801184 DE22801185 DE22801186 DE22801187 DE22801188 DE22801189 DE22801190 DE22801191 DE22801192 DE22801193 DE22801194 DE22801195 DE22801196 DE22801197 DE22801198 DE22801200 DE22801201 DE22801202 DE22801203 DE22801204 DE22801205 DE22801206 DE22801207 DE22801208 DE22801210 DE22801211 DE22801212 DE22801213 DE22801214 DE22801215 DE22801216 DE22801217 DE22801218 DE22801219 DE22801220 DE22801221 DE22801226 DE22801227 DE22801232 DE22801234 DE22801235 DE22801236 DE22801237 DE22801238 DE22801239 DE22801240 DE22801241 DE22801242 DE22801243 DE22801244 DE22801245 DE22801246 DE22801247 DE22801248 DE22801249 DE22801250 DE22801251 DE22801252 DE22801253 DE22801254 DE22801255 DE22801256 DE22801257 DE22801258 DE22801259 DE22801264 DE22801268 DE22801308 DE22801314 DE22801098 DE22801357 DE22801362 DE22801364 DE22801373 DE22801438 DE22801444 DE22801445 DE22801455 DE22801381 DE22801480 DE22801488 DE22801492 DE22801496 DE22801692 DE22801693 DE22801694 DE22801695 DE22801697 DE22801701 DE22801702 DE22801703 DE22801704 DE22801705 DE22801707 DE22801710 DE22801711 DE22801712 DE22801713 DE22801714 DE22801715 DE22801716 DE22801717 DE22801719 DE22801720 DE22801721 DE22801722 DE22801723 DE22801724 DE22801725 DE22801727 DE22801728 DE22801729 DE22801730 DE22801731 DE22801732 DE22801733 DE22801734 DE22801735 DE22801736 DE22801737 DE22801738 DE22801740 DE22801741 DE22801763 DE22801764 DE22801765 DE22801771 DE22801773 DE22801775 DE22801776 DE22801777 DE22801567 DE22801576 DE22801577 DE22801578 DE22801580 DE22801581 DE22801582 DE22801583 DE22801585 DE22801612 DE22801613 DE22801614 DE22801615 DE22801616 DE22801617 DE22801618 DE22801619 DE22801620 DE22801621 DE22801622 DE22801623 DE22801624 DE22801625 DE22801626 DE22801627 DE22801628 DE22801629 DE22801630 DE22801631 DE22801632 DE22801633 DE22801634 DE22801635 DE22801636 DE22801637 DE22801638 DE22801639 DE22801640 DE22801641 DE22801642 DE22801643 DE22801644 DE22801645 DE22801646 DE22801647 DE22801648 DE22801649 DE22801650 DE22801651 DE22801655 DE22801517 DE22801518 DE22801519 DE22801520 DE22801521 DE22801523 DE22801524 DE22801525 DE22801526 DE22801538 DE22801539 DE22801540 DE22801541 DE22801542 DE22801543 DE22801544 DE22801545 DE22801546 DE22801548 DE22801549 DE22801550 DE22801551 DE22801552 DE22801556 DE22801558 DE22801559 DE22801562 DE22801566 DE22801588 DE22801589 DE22801590 DE22801592 DE22801593 DE22801594 DE22801595 DE22801597 DE22801598 DE22801599 DE22801600 DE22801602 DE22801603 DE22801604 DE22801605 DE22801606 DE22801607 DE22801608 DE22801609 DE22801610 DE22801611 DE22801679 DE24701908 DE24701910 DE24701911 DE24701925 DE24701934 DE24701939 DE24701942 DE24701943 DE24701944 DE24701945 DE24701948 DE24701949 DE24701950 DE24701951 DE24701952 DE24701953 DE24701954 DE24701955 DE24701956 DE24701957 DE24701960 DE24701961 DE24701962 DE24701963 DE24701964 DE24701965 DE24701966 DE24701967 DE24701968 DE24701969 DE24701970 DE24701971 DE24701972 DE24701973 DE24701974 DE24701975 DE24701976 DE24701977 DE24701979 DE24701982 DE24701994 DE24701995 DE24701983 DE24701984 DE24701985 DE24701986 DE24701987 DE24701989 DE24701991 DE24701992 DE24701993 DE24701996 DE24701997 DE24701998 DE24701999 DE24702001 DE24702002 DE24702003 DE24702004 DE24702005 DE24702006 DE24702007 DE24702008 DE24702009 DE24702010 DE24702011 DE24702012 DE24702013 DE24702014 DE24702015 DE24702016 DE24702017 DE24702018 DE24702019 DE24702020 DE24702021 DE24702037 DE24702041 DE24702042 DE24702045 DE24702047 DE24702060 DE24702067 DE24702068 DE24702069 DE24702022 DE24702023 DE24702025 DE24702035 DE24702036 DE24702038 DE24702070 DE24702071 DE24702072 DE24702073 DE24702074 DE24702075 DE24702076 DE24702077 DE24702078 DE24702079 DE24702080 DE24702081 DE24702083 DE24702084 DE24702085 DE24702086 DE24702087 DE24702088 DE24702089 DE24702090 DE24702091 DE24702092 DE24702093 DE24702094 DE24702095 DE24702096 DE24702097 DE24702098 DE24702099 DE24702100 DE24702101 DE24702102 DE24702103 DE24702104 DE24702105 DE24702106 DE24702107 DE24702108 DE24702109 DE24702065 DE24702110 DE24702111 DE24702112 DE24702113 DE24702114 DE24702115 DE24702120 DE24702121 DE24702122 DE24702123 DE24702124 DE24702125 DE24702126 DE24702127 DE24702129 DE24702130 DE24702131 DE24702132 DE24702134 DE24702135 DE24702136 DE24702137 DE24702140 DE24702141 DE24702142 DE24702143 DE24702144 DE24702146 DE24702147 DE24702149 DE24702150 DE24702151 DE24702152 DE24702153 DE24702154 DE24702155 DE24702156 DE24702157 DE24702158 DE24702159 DE24702160 DE24702161 DE24702162 DE24702163 DE24702164 DE24702165 DE24702166 DE24702167 DE24702168 DE24702172 DE24702179 DE24702187 DE24702189 DE24702190 DE24702191 DE24702192 DE24702193 DE24702194 DE24702195 DE24702196 DE24702197 DE24702198 DE24702199 DE24702200 DE24702201 DE24702202 DE24702204 DE24702205 DE24702206 DE24702207 DE24702208 DE24702185 DE24702186 DE24702188 DE24702227 DE24702271 DE24702272 DE22801708 DE22801739 DE33105436 DE33105437 DE33105438 DE33105440 DE44307676 DE44306482 DE33105455 DE24702552 DE24702556 DE24702557 DE24702581 DE24702209 DE24702211 DE24702222 DE24702768 DE24702769 DE24701896 DE24701932 DE24702053 DE33103859 DE33104855 DE24702062 DE24702661 DE22801747 DE22801755 DE22801756 DE22801757 DE22801782 DE22801783 DE22801784 DE22801786 DE22801787 DE22801788 DE22801789 DE22801794 DE22801798 DE22801799 DE22801800 DE22801785 DE22801790 DE22801791 DE22801792 DE22801795 DE22801801 DE22801408 DE22801409 DE22801414 DE22801416 DE22801411 DE44306489 DE44306490 DE24702118 DE24701894 DE22801269 DE22801303 DE33103961 DE33103969 DE33103970 DE33103974 DE33103976 DE33103978 DE33103900 DE33103902 DE33103905 DE33103907 DE33103908 DE33103909 DE33103914 DE33103923 DE33103924 DE33103925 DE33103927 DE33103935 DE33103936 DE33103937 DE33103938 DE33103939 DE33103940 DE33103944 DE33103945 DE33103946 DE33103948 DE33103949 DE33103950 DE33103951 DE33103952 DE33103955 DE33103959 DE33103953 DE33103958 DE33103956 DE33103932 DE33103933 DE33103942 DE33103943 DE33103947 DE33103957 DE33103968 DE33103954 DE44306148 DE44307365 DE44307367 DE44307369 DE44307370 DE44307371 DE44307374 DE33103803 DE33104928 DE33104929 DE33104953 DE33103875 DE44305927 DE44305933 DE24702270 DE24701872 DE24701875 DE24701881 DE24702770 DE24702772 DE24702777 DE44307014 DE44307017 DE44307024 DE44307026 DE44307147 DE44306035 DE44306047 DE33104833 DE33104846 DE33103626 DE33103628 DE44306845 DE22801507 DE24702901 DE24702918 DE33104332 DE33104334 DE33104348 DE44306177 DE44306179 DE22801808 DE22801809 DE22801814 DE22801820 DE22801825 DE24702601 DE24702611 DE33104817 DE24702043 DE24702054 DE24702066 DE24702385 DE24702387 DE24702389 DE24702393 DE24702395 DE24702396 DE24702397 DE24702398 DE24702403 DE24702643 DE33104640 DE22801279 DE33103872 DE24702642 DE24702646 DE24702324 DE24701887 DE24702210 DE24702662 DE24702664 DE24702671 DE24702943 DE24702944 DE24702474 DE24702476 DE24702482 DE33103609 DE33103611 DE33103618 DE33103619 DE44308385 DE44308387 DE24702637 DE33104276 DE33104287 DE33104553 DE44305647 DE44305649 DE44305650 DE44305679 DE33104038 DE24701885 DE24701918 DE22801286 DE44307642 DE44307757 DE44307758 DE44307759 DE44307760 DE44307761 DE44307781 DE44307793 DE44307797 DE44307798 DE44307800 DE44307805 DE24702117 DE33103081 DE33103309 DE33103316 DE33103317 DE33103318 DE33103319 DE24701893 DE24701903 DE24701906 DE24701912 DE24701915 DE24701919 DE24701922 DE33104927 DE33104934 DE44305796 DE44306088 DE24702059 DE24702116 DE24702119 DE24702128 DE44306817 DE44306818 DE44306819 DE44306827 DE33103876 DE44306253 DE44306259 DE33104873 DE33104880 DE44307063 DE44307054 DE33103351 DE33103354 DE33103355 DE33103359 DE33103413 DE33103414 DE33103415 DE33103416 DE44307701 DE44307148 DE44307149 DE44307154 DE44307155 DE44307159 DE44306666 DE44306668 DE44306684 DE44307030 DE44307033 DE44307048 DE33104091 DE33104095 DE33104107 DE22801395 DE44307730 DE24702276 DE24702278 DE24702286 DE24702273 DE24702789 DE33103931 DE33105363 DE33105364 DE33105374 DE33105377 DE44306516 DE44306517 DE44306520 DE33103174 DE33103211 DE33103214 DE33103216 DE33103208 DE33103213 DE33103121 DE33103148 DE33103150 DE33103151 DE33103155 DE33103160 DE33103161 DE33103163 DE33103164 DE33103165 DE33103166 DE33103167 DE33103168 DE33103169 DE33103170 DE33103171 DE33103172 DE33103173 DE33103175 DE33103176 DE33103177 DE33103178 DE33103179 DE33103180 DE33103181 DE33103182 DE33103183 DE33103192 DE33103205 DE33103206 DE33103230 DE33103231 DE33104826 DE33103210 DE33103212 DE33103219 DE24701888 DE24701889 DE44307078 DE44307079 DE44307086 DE33103162 DE33103204 DE22801817 DE44307389 DE44307191 DE44306375 DE44306378 DE44306381 DE44306386 DE44305622 DE44306983 DE44306984 DE44306985 DE44306986 DE24702638 DE24702639 DE24702650 DE24701886 DE33103901 DE33103919 DE33105331 DE44306460 DE44306471 DE44306474 DE44306549 DE44306552 DE44306558 DE44306542 DE24702044 DE24702048 DE24702056 DE33102957 DE33102959 DE33102962 DE33102968 DE33102975 DE44306488 DE24702855 DE24702870 DE44306634 DE22801516 DE22801501 DE22801505 DE24702372 DE24702373 DE24702374 DE33105373 DE44307172 DE44307727 DE24702180 DE22801209 DE22801223 DE22801224 DE22801233 DE22801285 DE33103215 DE33103217 DE33103222 DE33103245 DE33103246 DE24702256 DE24702297 DE24702304 DE24702305 DE24702322 DE44307354 DE44307355 DE44307356 DE44307357 DE44307358 DE44307359 DE44307361 DE44307363 DE44307364 DE24702694 DE24702695 DE24702696 DE24702697 DE24702698 DE24702702 DE24701861 DE33103554 DE33103474 DE33104116 DE33104119 DE33104561 DE33104563 DE33104568 DE33104972 DE33104974 DE33104976 DE33103278 DE33103259 DE33103011 DE33103024 DE33103297 DE33103314 DE24702951 DE24702954 DE44307122 DE44307131 DE22801369 DE22801372 DE22801376 DE22801417 DE22801418 DE22801419 DE22801420 DE22801421 DE22801428 DE22801429 DE22801430 DE22801433 DE22801434 DE33104944 DE33104957 DE33104462 DE33104471 DE33104475 DE33104481 DE33104485 DE33104489 DE24702277 DE24702283 DE33104750 DE33104758 DE33104759 DE33104761 DE33104763 DE33104765 DE33104803 DE24702867 DE24702873 DE24702874 DE24702877 DE24702880 DE24702884 DE24702886 DE24702889 DE24702891 DE24702892 DE22801266 DE22801307 DE24702214 DE44307566 DE44307548 DE44307552 DE44307554 DE44307555 DE44307556 DE44307557 DE44307558 DE44307560 DE44307561 DE44307562 DE44307563 DE44307564 DE44307565 DE44307567 DE44307568 DE44307569 DE44307570 DE44307571 DE44307572 DE44307573 DE44307574 DE44307575 DE44307576 DE33103010 DE22801293 DE44307815 DE44307825 DE44307829 DE44307832 DE44307754 DE44307755 DE22801229 DE24702703 DE24702706 DE24702710 DE24702712 DE24702723 DE24702724 DE24702725 DE33104804 DE33104795 DE33104802 DE22801374 DE44305922 DE44305923 DE33103630 DE33103635 DE33103638 DE33103012 DE33103013 DE33103014 DE33103015 DE33103016 DE33103058 DE33104895 DE44305642 DE44305646 DE44305651 DE24702216 DE33105338 DE33105343 DE33105346 DE33105366 DE33105371 DE33105380 DE44307143 DE24702031 DE24702051 DE44306301 DE24702952 DE44306651 DE33105285 DE44306390 DE44306153 DE44305818 DE22801261 DE22801270 DE22801271 DE22801272 DE22801273 DE22801274 DE22801292 DE24702390 DE24702394 DE24702400 DE24702401 DE22801294 DE33103755 DE33103759 DE33103767 DE33103769 DE33103770 DE33103322 DE33103287 DE33105229 DE44307270 DE44307271 DE44307272 DE44307280 DE24702859 DE24702830 DE24702834 DE24702835 DE24702854 DE24702857 DE24702858 DE24702861 DE24702862 DE33103981 DE33103984 DE33103998 DE24702804 DE24702805 DE24702806 DE24702807 DE24702808 DE24702809 DE24702811 DE24702813 DE24702814 DE24702815 DE24702816 DE24702817 DE24702818 DE24702819 DE24702820 DE24702821 DE24702822 DE24702823 DE24702829 DE24702836 DE24702837 DE24702838 DE24702839 DE24702828 DE24702832 DE24702833 DE24702840 DE24702841 DE24702842 DE24702843 DE24702844 DE24702846 DE24702860 DE24702863 DE24702853 DE24702856 DE24702831 DE44305791 DE24702502 DE24702506 DE33105577 DE33105593 DE33105378 DE33105382 DE44307254 DE44307260 DE33103325 DE33103332 DE33103335 DE33103336 DE33103337 DE33103338 DE33103339 DE33103340 DE33103341 DE33103343 DE33103344 DE33103345 DE33103330 DE33103113 DE24702579 DE24702658 DE24702756 DE33103076 DE33103077 DE33103114 DE33103386 DE33103395 DE33103403 DE33103404 DE33103461 DE33104946 DE33104954 DE33104955 DE33104956 DE33104963 DE33105318 DE33105321 DE44307004 DE44307080 DE44307088 DE44307090 DE44307091 DE44307095 DE44307096 DE44307097 DE44307099 DE44307071 DE44307089 DE44307093 DE44307094 DE44306994 DE44306866 DE44306867 DE44307542 DE44307543 DE44307544 DE44307545 DE44307549 DE44307550 DE44309185 DE22801368 DE44306303 DE44306304 DE44306317 DE44306319 DE44306054 DE44306081 DE44306082 DE44306083 DE24702226 DE33104741 DE33105481 DE33105487 DE33105493 DE33105512 DE33105525 DE33105544 DE33105547 DE33105551 DE24702633 DE24702634 DE24702635 DE24702641 DE24702653 DE24702657 DE24702876 DE24702881 DE24702896 DE24702897 DE24702899 DE24702900 DE24702920 DE24702921 DE24702925 DE24702929 DE24702930 DE44306123 DE44306126 DE44306150 DE44306151 DE44306152 DE44306164 DE44306167 DE44306170 DE44306181 DE44306182 DE33105355 DE33105372 DE33105376 DE33105342 DE33104862 DE33104863 DE33104864 DE33104865 DE33104868 DE33104869 DE33104874 DE22801575 DE22801587 DE22801596 DE22801601 DE33103593 DE33103594 DE33103597 DE33103101 DE44305914 DE44305918 DE44305921 DE33104293 DE33104296 DE33105557 DE44306835 DE22801305 DE22801355 DE33103105 DE33103860 DE33103861 DE44306891 DE44306971 DE44306972 DE44306976 DE44307301 DE24702727 DE24702736 DE24702737 DE24702342 DE24702346 DE24702353 DE24702354 DE24702355 DE24702360 DE33103579 DE44306723 DE33103571 DE33103581 DE44306680 DE44306718 DE44306719 DE44306724 DE44306992 DE44306995 DE44307000 DE44307001 DE44307282 DE44307283 DE44307287 DE44307288 DE44307289 DE24702419 DE33103128 DE33103145 DE33103152 DE33103156 DE33103157 DE33103158 DE33103159 DE44305664 DE44305668 DE33105339 DE33105340 DE33105344 DE33105356 DE44306953 DE44306954 DE33103367 DE33103368 DE33103391 DE33103392 DE33103396 DE33103399 DE33103400 DE33103401 DE24701899 DE24702026 DE24702289 DE24702290 DE24702296 DE24702301 DE24702302 DE24702307 DE24702308 DE24702309 DE44305771 DE24702170 DE24702177 DE24702182 DE24702203 DE44306512 DE44306513 DE44306515 DE33103342 DE24702175 DE24702183 DE22801018 DE22801022 DE22801051 DE22801052 DE22801065 DE22801068 DE44306836 DE44307488 DE22801537 DE22801557 DE22801560 DE22801561 DE22801563 DE22801564 DE22801565 DE22801570 DE22801571 DE22801574 DE44307578 DE44307583 DE44307584 DE44307587 DE44307595 DE44307592 DE33103592 DE33103599 DE33103623 DE33103334 DE33103602 DE24702402 DE24702408 DE24702421 DE24702442 DE44307677 DE22801754 DE33104529 DE33104538 DE33104541 DE33104542 DE33104543 DE44306371 DE44306372 DE44306373 DE33103960 DE33103963 DE33103966 DE33103972 DE33103973 DE33103992 DE33103994 DE33103548 DE33103556 DE24702034 DE24702039 DE24702052 DE44306720 DE44306727 DE33105185 DE33105190 DE33105212 DE33105218 DE33105222 DE33105223 DE33105224 DE33105226 DE33105228 DE33104905 DE33104909 DE33104913 DE33104915 DE33104916 DE44307801 DE44305692 DE33104897 DE33104797 DE33104798 DE33105583 DE44307577 DE33104096 DE33104114 DE33104121 DE33104126 DE33104130 DE44307229 DE44307239 DE33104514 DE33104515 DE33104914 DE24702847 DE24702848 DE24702851 DE24702866 DE33103324 DE33103346 DE33103418 DE33103448 DE33103460 DE33103732 DE33103733 DE33103734 DE33103736 DE33103740 DE33103741 DE33103743 DE33103744 DE33103745 DE33103746 DE33103747 DE33103748 DE33103750 DE33104904 DE33103385 DE33103390 DE33103398 DE44307764 DE44307766 DE44307772 DE33104836 DE24702949 DE44307059 DE44306763 DE44306769 DE44306773 DE44306774 DE44306775 DE44306776 DE44306780 DE44306781 DE44306782 DE44306783 DE44306786 DE44306717 DE33103665 DE33103672 DE33103674 DE33103676 DE33103678 DE33103679 DE33103680 DE33103681 DE33103682 DE33103683 DE33103684 DE33103685 DE33103687 DE33103688 DE33103689 DE33103690 DE33103691 DE33103692 DE33103693 DE33103694 DE33103695 DE33103696 DE33103697 DE33103698 DE33103699 DE33103700 DE33103701 DE33103702 DE33103703 DE33103704 DE33103705 DE33103706 DE33103707 DE33103708 DE33103709 DE33103710 DE33103517 DE33103651 DE33103652 DE33103653 DE33103656 DE33103664 DE33104830 DE33104834 DE33104837 DE33104839 DE24702923 DE24702931 DE24702938 DE24702269 DE33104959 DE33105053 DE24702484 DE24702486 DE44305816 DE24702758 DE44307783 DE44307786 DE44307787 DE44307790 DE44307791 DE33103991 DE33103993 DE44307213 DE44307217 DE44306159 DE44306162 DE44307504 DE44307508 DE44307656 DE44307664 DE44307098 DE44307182 DE44307183 DE44307184 DE44307185 DE44307187 DE44307190 DE44307192 DE44307193 DE44307194 DE44307196 DE44307197 DE44307198 DE44307199 DE44307208 DE44307039 DE44307041 DE44307043 DE44307046 DE44307051 DE44307052 DE44307053 DE44307055 DE44307056 DE44307057 DE44307060 DE44307067 DE44307068 DE44307069 DE22801827 DE33103387 DE33103388 DE22801230 DE22801298 DE22801300 DE22801311 DE22801312 DE22801313 DE33103880 DE33103881 DE33103882 DE33103883 DE33103884 DE33103886 DE33103887 DE33103889 DE33103890 DE33103895 DE33103898 DE33103899 DE33103916 DE33105155 DE33105201 DE33105220 DE44305930 DE44305932 DE44305944 DE44305967 DE44305936 DE44305941 DE44305942 DE44305943 DE44305938 DE24702652 DE24702684 DE44305770 DE24702553 DE24702555 DE24702559 DE24702565 DE44307733 DE24702554 DE24702558 DE44305764 DE44305765 DE44305766 DE44305768 DE44305772 DE44305778 DE44305779 DE44305783 DE44305784 DE44305789 DE44305794 DE44305795 DE44305797 DE44305798 DE44305763 DE33103822 DE33103826 DE33103827 DE33103828 DE33103833 DE33103839 DE44306505 DE33105058 DE22801403 DE24702337 DE33103267 DE22801304 DE24702826 DE33105062 DE33103854 DE44307376 DE44307382 DE44307397 DE22801522 DE22801527 DE22801534 DE44306690 DE44306693 DE44307366 DE44307386 DE44307388 DE22801442 DE22801456 DE24702726 DE24702738 DE24702739 DE24702740 DE24702741 DE24702742 DE33103025 DE33103037 DE33103040 DE33103041 DE33103044 DE33103848 DE44306134 DE33105417 DE44306036 DE44306037 DE33104321 DE33104323 DE33104324 DE33104325 DE33104327 DE33104328 DE33104329 DE33104330 DE33104333 DE33104336 DE33104342 DE33104344 DE33104352 DE33104367 DE33104404 DE33104478 DE44305639 DE44306144 DE44306145 DE44306146 DE44306147 DE44306174 DE44306544 DE44306582 DE44306583 DE44306584 DE44306585 DE44306586 DE44306587 DE44306589 DE44306590 DE44306591 DE44306592 DE44306593 DE44306594 DE44306595 DE44306596 DE44306597 DE44306598 DE44306599 DE44306600 DE44306601 DE44306637 DE44306665 DE44306698 DE44306699 DE44306700 DE44306578 DE44306579 DE44306706 DE44306838 DE44306841 DE44306844 DE44306846 DE44306856 DE44307255 DE44307259 DE44307331 DE44307312 DE44307313 DE44307323 DE44307328 DE33103512 DE33103285 DE33103286 DE33103305 DE33103323 DE44305913 DE44306779 DE24702027 DE24702028 DE24702029 DE24702030 DE24702032 DE24702033 DE24702058 DE24702082 DE33103266 DE33103269 DE33103629 DE33103634 DE33103640 DE33103641 DE33103649 DE44306040 DE44306045 DE44306046 DE44306050 DE44306014 DE44306015 DE44306016 DE44306017 DE44306018 DE44306021 DE44306022 DE44306023 DE44306024 DE44306025 DE44306026 DE44306028 DE44306029 DE44306030 DE44306032 DE44306165 DE44306205 DE44306206 DE44306207 DE44306208 DE44306209 DE44306210 DE44306211 DE44306212 DE44306213 DE44306214 DE44306218 DE44306219 DE44306220 DE44306221 DE44306225 DE44306226 DE44306227 DE44306228 DE44306229 DE44306281 DE44306283 DE44306285 DE44306286 DE44306287 DE44306288 DE44306289 DE44306290 DE44306293 DE44306294 DE44306295 DE44306296 DE44306297 DE44306298 DE44306299 DE44306393 DE44306409 DE44306704 DE44306712 DE44306714 DE33103119 DE33103120 DE33103122 DE33103126 DE33103130 DE33103133 DE33103138 DE33103139 DE33103141 DE33103143 DE33103570 DE33104827 DE33104835 DE33104844 DE33104845 DE33104849 DE33104850 DE44307390 DE24702249 DE24702250 DE24702251 DE24702252 DE24702253 DE24702255 DE24702258 DE24702259 DE22801019 DE22801020 DE22801002 DE22801028 DE22801029 DE22801016 DE22801041 DE22801042 DE22801043 DE22801045 DE22801003 DE22801005 DE22801026 DE22801027 DE22801031 DE22801032 DE22801033 DE22801034 DE22801035 DE22801038 DE22801017 DE22801030 DE22801004 DE22801015 DE22801040 DE22801001 DE24702764 DE24702767 DE24702774 DE24702775 DE24702776 DE24702778 DE24702779 DE24702780 DE24702781 DE24702782 DE24702784 DE24702785 DE24702786 DE24702787 DE24702790 DE24702791 DE24702793 DE24702794 DE24702795 DE24702796 DE24702797 DE24702798 DE24702799 DE24702800 DE24702802 DE24702810 DE24702812 DE24702824 DE24702825 DE24702358 DE33104831 DE33104854 DE33104973 DE33104989 DE33104990 DE33104993 DE33105002 DE33105003 DE33105005 DE33105006 DE33105007 DE33105008 DE44307513 DE33103500 DE33104968 DE33104970 DE33104979 DE33104992 DE33104994 DE33104995 DE33104178 DE33104183 DE33104214 DE33104364 DE44306831 DE44306839 DE33105328 DE33105365 DE33105368 DE44306155 DE44306171 DE44306523 DE44306525 DE44306528 DE44306531 DE44306532 DE44306533 DE44306534 DE44306535 DE44306537 DE44306539 DE44306522 DE44306524 DE44306526 DE44306527 DE44306529 DE44306530 DE44306536 DE44306538 DE44306540 DE44306541 DE44306543 DE44306545 DE44306547 DE44306548 DE44306551 DE44306554 DE44306555 DE44306556 DE44306560 DE44306561 DE44306546 DE44306563 DE44306565 DE44306568 DE44306569 DE44306570 DE44306571 DE44306575 DE44306576 DE44306577 DE44306507 DE44306510 DE44306511 DE44306521 DE33105140 DE33105141 DE33105143 DE33105144 DE33105159 DE33105162 DE33105163 DE33105166 DE33105175 DE33105177 DE33105183 DE33105186 DE33105187 DE33105193 DE33105194 DE33105196 DE33105197 DE33105200 DE33105202 DE33105205 DE33105207 DE33105209 DE33105214 DE33105216 DE33105227 DE33105230 DE33105235 DE33105237 DE33105238 DE33105239 DE33105240 DE33105242 DE33105243 DE33105248 DE33105180 DE33105182 DE33105127 DE44307276 DE33105145 DE44306572 DE44306580 DE44306502 DE44306892 DE44306893 DE44306894 DE44306898 DE44306899 DE44307262 DE44307263 DE44307264 DE24701947 DE33103768 DE33105009 DE33105029 DE33105030 DE33105032 DE33105033 DE33105034 DE33105035 DE33105036 DE33105037 DE33105045 DE33105017 DE33105385 DE44306628 DE33105010 DE33103023 DE33103064 DE33103410 DE33103411 DE33103424 DE33103425 DE33104148 DE33104176 DE33104182 DE33104219 DE33104255 DE33104288 DE33104354 DE33104358 DE33104532 DE44305819 DE44305821 DE44305822 DE44305823 DE44305830 DE44307242 DE44307245 DE44307249 DE44307252 DE33104949 DE33104950 DE33104952 DE33104958 DE44306997 DE44307035 DE44306950 DE44306959 DE44306966 DE44306969 DE33104749 DE33105332 DE33104945 DE33105379 DE44306038 DE44306039 DE44306048 DE44306049 DE44306051 DE44306052 DE33103301 DE33103545 DE44307679 DE33104848 DE33103406 DE33103417 DE33103149 DE33103153 DE44305769 DE44305774 DE44305775 DE44305776 DE44305781 DE33103586 DE33103917 DE33105522 DE33105022 DE33103026 DE44305863 DE33104196 DE33104200 DE33104201 DE33104202 DE33103294 DE33103296 DE33103298 DE33103303 DE33104843 DE33104636 DE33104099 DE44305852 DE33105381 DE44307050 DE44307058 DE44305777 DE33103563 DE22801426 DE33105114 DE44306103 DE44306990 DE44307580 DE44307593 DE44307596 DE24702568 DE33103038 DE33103054 DE33103061 DE33103070 DE33103091 DE33103117 DE33103092 DE33103108 DE33103109 DE33103129 DE33103132 DE24702707 DE24702713 DE24702714 DE24702715 DE24702907 DE24702912 DE24702937 DE24702947 DE33103321 DE44305817 DE44305853 DE44306999 DE44307002 DE44307008 DE44307009 DE44307347 DE44307594 DE44306173 DE24702284 DE33104284 DE44305907 DE44305912 DE44307351 DE44307353 DE33103600 DE44306251 DE44306252 DE22801386 DE22801387 DE22801388 DE22801394 DE33104983 DE33104981 DE44306064 DE24702477 DE24702500 DE24702501 DE33104270 DE33104829 DE44305909 DE44305910 DE33103008 DE33103018 DE44306695 DE44306702 DE44305678 DE44305686 DE24701884 DE24701890 DE24701891 DE24701892 DE24701895 DE24701901 DE44305677 DE44305682 DE44305688 DE44305691 DE44305676 DE33102963 DE24702298 DE44306056 DE44306058 DE33105423 DE33105424 DE33105426 DE44307368 DE24702478 DE24702480 DE24702498 DE24702494 DE22801360 DE24702287 DE44306858 DE44306860 DE44306861 DE44306862 DE44306863 DE33102997 DE33102998 DE33103009 DE33103027 DE33103045 DE33103046 DE33103048 DE33103050 DE33103051 DE33103052 DE33103059 DE33103063 DE33103576 DE33103601 DE33103604 DE44305799 DE44307220 DE44307708 DE44307710 DE24702171 DE24702176 DE24702169 DE22801361 DE44305681 DE44305685 DE33104969 DE33104982 DE44305803 DE44305805 DE44305806 DE44305807 DE44305808 DE44305809 DE44305810 DE44305811 DE33104772 DE33104774 DE33104621 DE44306859 DE44306865 DE44306871 DE44306885 DE44306887 DE44306888 DE33103796 DE33103797 DE33103798 DE33103799 DE33103800 DE33103801 DE33103805 DE33103807 DE33103808 DE33103809 DE33103817 DE33105308 DE33103595 DE33103596 DE33104094 DE44306084 DE33105066 DE33105087 DE33105105 DE33105108 DE33105112 DE33105115 DE33105116 DE33105120 DE33105121 DE33105125 DE33105126 DE33105128 DE33105131 DE33105132 DE33105133 DE33105134 DE33105367 DE33105369 DE33105370 DE33105393 DE33105394 DE33105395 DE33105396 DE33105397 DE33104112 DE33104115 DE33104117 DE33104123 DE33104124 DE33104125 DE33104132 DE33104134 DE33104137 DE33104143 DE33104150 DE44307158 DE44306643 DE44306661 DE33105375 DE33103715 DE33105383 DE33105389 DE33105398 DE33105402 DE33105405 DE33105408 DE33105409 DE33105412 DE33105414 DE33105415 DE33105421 DE33105422 DE44306833 DE44306842 DE44306843 DE44307299 DE44307123 DE33104960 DE44307195 DE44307204 DE44307205 DE44307206 DE44307209 DE44307210 DE44307212 DE44307219 DE33103606 DE44305845 DE33105275 DE33105289 DE33105291 DE33105294 DE33105271 DE33105067 DE33105079 DE33105096 DE33105103 DE33105111 DE33105122 DE33103610 DE33103612 DE33103865 DE33105016 DE33103663 DE33104614 DE33104618 DE33104622 DE44306646 DE44306649 DE44306653 DE44306654 DE44306656 DE44306658 DE44306659 DE33104000 DE33104001 DE33104002 DE33104003 DE33104006 DE33105564 DE33105569 DE33105594 DE33105600 DE33105026 DE33105043 DE33105044 DE33105047 DE33105048 DE44307854 DE33104144 DE33104174 DE33104180 DE33104244 DE33104340 DE33104356 DE33104357 DE33104363 DE33104468 DE33104748 DE44307221 DE44307222 DE44307223 DE44307225 DE44307227 DE44307228 DE44307233 DE44307234 DE44307235 DE44307236 DE44307241 DE44307267 DE44307268 DE44307277 DE44306822 DE44306826 DE33104722 DE33104729 DE33104745 DE33104746 DE33104747 DE33104751 DE33104752 DE33104753 DE33104754 DE33104924 DE33104925 DE33104926 DE33104930 DE33104931 DE33104932 DE33104933 DE33104936 DE33104937 DE33104938 DE33104939 DE33104940 DE33104941 DE33104942 DE44307579 DE44307582 DE44307003 DE33105057 DE33105329 DE33105351 DE44306300 DE44306311 DE44307129 DE33104660 DE33103847 DE33105236 DE33105244 DE33105245 DE33105246 DE33105247 DE33105249 DE33105267 DE33103589 DE33103590 DE33103450 DE33103454 DE33103456 DE33103459 DE33103840 DE33103844 DE33103845 DE33103851 DE33103852 DE33103857 DE33103858 DE33103864 DE33103867 DE33103870 DE33103873 DE33103877 DE44307266 DE44307269 DE44307278 DE33103673 DE33103567 DE33103654 DE33103658 DE33103659 DE33103661 DE33103667 DE33103669 DE33103670 DE33103671 DE33103675 DE33103677 DE33103686 DE33105313 DE44307141 DE44307142 DE44307145 DE44307151 DE44307157 DE44307531 DE44307535 DE44307639 DE44307652 DE44307658 DE44307669 DE44307670 DE33103657 DE33103802 DE33103804 DE33103806 DE33103810 DE33103811 DE33103812 DE33103813 DE33103814 DE33103815 DE33103816 DE33103818 DE33103819 DE33104290 DE33104291 DE44307496 DE33105191 DE33105206 DE44305785 DE44306692 DE44306655 DE44307399 DE44307230 DE33105147 DE44305663 DE44305666 DE44305667 DE33103980 DE33103982 DE33103983 DE33103985 DE33103986 DE33103987 DE33103996 DE33103997 DE44307717 DE44305869 DE44307281 DE44307284 DE44307613 DE44305826 DE33103476 DE33103495 DE33105157 DE33105213 DE33105215 DE33105234 DE33105250 DE33105257 DE33105261 DE33105269 DE44306840 DE44306849 DE44307201 DE33103608 DE33103620 DE33103621 DE33103624 DE33103625 DE33103627 DE33103930 DE33103934 DE33103871 DE44305736 DE44305737 DE44305739 DE44305741 DE44305742 DE44305743 DE44305744 DE44305747 DE44305749 DE44305752 DE44305761 DE33103636 DE33103642 DE33103643 DE33103646 DE33103650 DE33103660 DE33103662 DE33104500 DE33104501 DE33104507 DE33104508 DE33104509 DE33104512 DE33104516 DE33105092 DE33105094 DE24702850 DE22801745 DE22801758 DE44307126 DE44307128 DE44307134 DE44307136 DE44305714 DE33103021 DE33103366 DE24702852 DE33103752 DE33105347 DE33105362 DE44307500 DE44307503 DE44307505 DE44307506 DE44307507 DE44307518 DE44307501 DE44307502 DE44307509 DE44307510 DE44307511 DE44307516 DE44307517 DE44307521 DE44307525 DE44307530 DE44307532 DE44307533 DE44307539 DE33104891 DE44305654 DE44305655 DE44305665 DE44305675 DE44305680 DE44305754 DE44305755 DE33105164 DE44307031 DE44307040 DE44307044 DE44307045 DE44307047 DE44307049 DE33103452 DE33103453 DE33103455 DE33104018 DE33104027 DE33104070 DE33104077 DE33105233 DE33105563 DE33103762 DE33103304 DE33103306 DE33103307 DE33103308 6.60402E+15 DE33104847 DE33104851 DE33104852 DE33104853 DE33104277 DE33104278 DE33104283 DE33103888 DE33103892 DE44307534 DE44307536 DE33104856 DE24702869 DE33103449 DE33105282 DE731004420 BX92604 BX92076 DE24702888 DE24702890 DE24702893 DE24702894 DE33104297 DE33104308 DE33105571 DE44306102 DE44307153 DE44307156 DE44307161 DE44307174 DE44307175 DE44307709 DE44307696 DE44307714 DE44307725 DE44307722 DE44307731 DE33104147 DE33104193 DE33104232 DE33104271 DE33104273 DE33104299 DE33104301 DE33104303 DE33104304 DE33104306 DE33104360 DE33104272 DE24702572 DE24702573 DE24702574 DE24702575 DE24702576 DE24702577 DE24702578 DE24702580 DE24702584 DE24702586 DE24702588 DE24702589 DE24702590 DE24702871 DE33103910 DE33103911 DE33103913 DE33103920 DE44305694 DE44305698 DE44305733 DE44305735 DE44306418 DE44306430 DE44306436 DE44306437 DE44306449 DE44306454 DE44306444 DE44306445 DE44306448 DE44306462 DE44306122 DE44306124 DE44306125 DE44306127 DE44306129 DE44306131 DE44306132 DE44306133 DE44306138 DE44306140 DE44306142 DE44306143 DE44307160 DE24702418 DE24702940 DE22801059 DE22801129 DE22801144 DE22801228 DE33103837 DE24702366 DE24702369 DE24702371 DE24702376 DE24702380 DE24702381 DE24702383 DE33103135 DE33103136 DE33103140 DE33103142 DE44306128 DE44306176 DE44306178 DE44306180 DE22801275 DE22801472 DE24702719 DE44305693 DE44306744 DE44306508 DE44306509 DE44306518 DE44306519 DE44305624 DE44305633 DE44305637 DE44305638 DE44305641 DE44305644 DE44305645 DE44305652 DE33103719 DE33103726 DE22801441 DE22801443 DE22801446 DE22801449 DE22801450 DE22801451 DE24702365 DE24702375 DE24702377 DE24702379 DE33105025 DE33105097 DE33105099 DE33105100 DE33105101 DE33105102 DE33105104 DE33105109 DE33105123 DE44306679 DE44306681 DE44306682 DE44306694 DE44306715 DE44306716 DE44306721 DE44306726 DE44306728 DE44306729 DE44306730 DE44306731 DE44306732 DE44306733 DE44306734 DE44306738 DE44306745 DE44306737 DE44306739 DE44306746 DE44306748 DE44306740 DE44306749 DE33103096 DE33103097 DE33103098 DE33103106 DE33103107 DE33103123 DE33103124 DE33103125 DE33103127 DE33103134 DE24701874 DE24701878 DE24701898 DE24701907 DE24701921 DE22801452 DE33104305 DE33104414 DE33104417 DE33104633 DE33104634 DE33104639 DE33104641 DE33104642 DE33104643 DE44306952 DE44306964 DE33104400 DE44307226 DE44307231 DE33104757 DE33104762 DE33104767 DE33104771 DE33104806 DE33104811 DE33104819 DE33104823 DE44305844 DE44305847 DE44305848 DE44305849 DE44305850 DE44305851 DE44305859 DE33103644 DE22801477 DE22801478 DE22801479 DE22801481 DE22801482 DE22801483 DE22801494 DE33104131 DE33104135 DE33104140 DE33104151 DE33104152 DE33104163 DE33104164 DE44306374 DE44306379 DE44306380 DE44306382 DE44307400 DE44307403 DE44307404 DE44307405 DE44307407 DE44307408 DE44307409 DE44307410 DE44307411 DE44307413 DE44307414 DE44307415 DE44307416 DE44307417 DE44307418 DE44307419 DE24702472 DE24702473 DE24702481 DE24702483 DE44307675 DE44307687 DE24701865 DE24702493 DE33103326 DE33105031 DE33105039 DE33105041 DE33105042 DE33105046 DE33105049 DE33105050 DE33105051 DE33105052 DE33105056 DE33105059 DE33105068 DE33105054 DE33105055 DE33105060 DE33105078 DE33105083 DE33105156 DE33105273 DE33105276 DE33105277 DE33105278 DE33105279 DE33105280 DE33105281 DE33105283 DE33105327 DE33105358 DE44305719 DE44305722 DE44305723 DE44305974 DE44305975 DE44305976 DE44305994 DE44305996 DE44305997 DE44305998 DE44305999 DE44306000 DE44306003 DE44306004 DE44306008 DE44306009 DE44306011 DE44305995 DE44306002 DE44306005 DE44306007 DE44306010 DE44305814 DE44305828 DE44305885 DE44305889 DE44305891 DE44305892 DE44305893 DE44305895 DE44305896 DE44305898 DE44305900 DE44305902 DE44305903 DE44305965 DE44306398 DE44306410 DE44306419 DE44306420 DE44306422 DE44306424 DE44306426 DE44306427 DE44306431 DE44306434 DE44306485 DE44306514 DE44306707 DE44306735 DE44306736 DE44306741 DE44306742 DE44306743 DE44306747 DE44306750 DE44306751 DE44306752 DE44306753 DE44306754 DE44306884 DE44306886 DE44306996 DE44307028 DE44307029 DE44307036 DE44307038 DE44307042 DE44307032 DE44307037 DE44307334 DE44307335 DE44307337 DE44307339 DE44307344 DE44307348 DE44307350 DE44307352 DE44307360 DE44307362 DE44307393 DE44307467 DE44307468 DE44307600 DE22801348 DE22801358 DE22801359 DE22801365 DE22801366 DE22801375 DE22801439 DE22801440 DE22801448 DE22801453 DE22801698 DE22801718 DE22801726 DE22801672 DE22801673 DE22801675 DE22801676 DE22801677 DE22801678 DE22801684 DE22801685 DE22801686 DE22801687 DE22801690 DE22801691 DE22801742 DE22801743 DE22801744 DE22801750 DE22801751 DE22801753 DE22801759 DE22801749 DE22801760 DE22801761 DE24701877 DE24701882 DE24702404 DE24702406 DE24702407 DE24702409 DE24702410 DE24702416 DE24702420 DE24702425 DE24702440 DE24702488 DE24702490 DE24702583 DE24702585 DE24702587 DE24702760 DE24702762 DE24702765 DE24702788 DE24702792 DE24702803 DE24702919 DE24702935 DE24702936 DE24702941 DE24702942 DE24702945 DE24702864 DE24702865 DE24702872 DE24702878 DE33102960 DE33102972 DE33103099 DE33103104 DE33103111 DE33103112 DE33103116 DE33103281 DE33103289 DE33103292 DE33103376 DE33103419 DE33103420 DE33103421 DE33103422 DE33103423 DE33103427 DE33103467 DE33103568 DE33103574 DE33103580 DE33103584 DE33103585 DE33103587 DE33103591 DE33103564 DE33103566 DE33103578 DE33103582 DE33103575 DE33103613 DE33103614 DE33103617 DE33103751 DE33103757 DE33103761 DE33103765 DE33103766 DE33103758 DE33103780 DE33103781 DE33103782 DE33103784 DE33104010 DE33104011 DE33104012 DE33104013 DE33104014 DE33104015 DE33104016 DE33104017 DE33104020 DE33104021 DE33104022 DE33104023 DE33104024 DE33104025 DE33104026 DE33104028 DE33104029 DE33104034 DE33104043 DE33104045 DE33104049 DE33104499 DE33104502 DE33104504 DE33104505 DE33104506 DE33104513 DE33104517 DE33104518 DE33104519 DE33104520 DE33104521 DE33104522 DE33104523 DE33104526 DE33104527 DE33104866 DE33104867 DE33104966 DE33104967 DE33104980 DE33104964 DE44306006 DE33105160 DE33105161 DE33105167 DE33105171 DE33105188 DE33105204 DE33105219 DE33105225 DE33105231 DE33105256 DE33105258 DE33105265 DE33105270 DE44306414 DE44306415 DE44306416 DE44306417 DE44306423 DE44306425 DE44306433 DE44307321 DE44306606 DE44306184 DE44307638 DE44307643 DE44307644 DE44307650 DE44307674 DE22801383 DE22801470 DE22801471 DE22801493 DE22801796 DE22801802 DE22801804 DE22801805 DE22801815 DE22801818 DE24701905 DE24702173 DE24702174 DE24702178 DE24702184 DE24702504 DE24702505 DE24702602 DE24702908 DE24702946 DE24702948 DE33103223 DE33103248 DE33103251 DE33103253 DE33103965 DE33104177 DE33104226 DE33104230 DE33104236 DE33104286 DE33104484 DE33104565 DE33104573 DE33105061 DE33105063 DE33105524 DE33105535 DE33105543 DE33105553 DE44305635 DE44305643 DE44305648 DE44305674 DE44305683 DE44305684 DE44305687 DE44305689 DE44305690 DE44305991 DE44305983 DE44305988 DE44306063 DE44306068 DE44306282 DE44306318 DE44306633 DE44306613 DE44306618 DE44306650 DE44306671 DE44306672 DE44306673 DE44306675 DE44306676 DE44306677 DE44306678 DE44306683 DE44306685 DE44306686 DE44306687 DE44306688 DE44306696 DE44307682 DE44306302 DE44306314 DE44306864 DE44306868 DE44306870 DE44306874 DE44306876 DE44306877 DE44306878 DE44306879 DE44306881 DE44306882 DE44306951 DE44306955 DE44306973 DE44306974 DE44306975 DE44306977 DE44306978 DE44306979 DE44306981 DE44306957 DE44306965 DE44306968 DE44307114 DE44307108 DE44307072 DE44307074 DE44307076 DE44307092 DE44307100 DE44307101 DE44307102 DE44307103 DE44307106 DE44307107 DE44307109 DE44307110 DE44307112 DE44307113 DE44307115 DE44307116 DE44307117 DE44307118 DE44307119 DE24702345 DE33102958 DE33102970 DE33102971 DE22801810 DE22801811 DE22801819 DE33103373 DE33103280 DE33103491 DE44305611 DE44305612 DE44305613 DE44305614 DE44305615 DE44305616 DE44305617 DE44305618 DE44305619 DE44305620 DE44305623 DE44305625 DE44305626 DE44305627 DE44305628 DE44305629 DE44305630 DE44305631 DE44305632 DE24702212 DE24702217 DE24702218 DE24702220 DE24702221 DE24702223 DE24702225 DE24702228 DE24702229 DE24702230 DE24702231 DE24702232 DE24702233 DE24702234 DE24702235 DE24702236 DE24702237 DE24702238 DE24702239 DE24702240 DE24702241 DE24702242 DE24702243 DE24702244 DE24702245 DE24702246 DE24702247 DE24702260 DE24702261 DE24702263 DE24702266 DE24702267 DE44305990 DE33103375 DE33104908 DE33104910 DE33104921 DE33104911 DE33104918 DE33104919 DE44305979 DE44306890 DE44306895 DE24702338 DE24702350 DE44307111 DE44307318 DE44307291 DE44307292 DE44307293 DE44307296 DE44307297 DE44307298 DE44307302 DE44307303 DE44307304 DE44307306 DE44307307 DE44307308 DE44307310 DE24701857 DE24701858 DE24701859 DE24701860 DE24701862 DE24701863 DE24701864 DE24701866 DE24701867 DE24701868 DE24701869 DE24701870 DE33104894 DE22801806 DE22801821 DE22801826 DE22801830 DE22801831 DE22801833 DE22801696 DE22801699 DE22801700 DE22801706 DE22801709 DE33104187 DE44306869 DE44306771 DE33103371 DE33103372 DE33103374 DE33104300 DE24702745 DE24702749 DE24702750 DE24702367 DE24702378 DE24702382 DE33105014 DE33105021 DE33105027 DE33105038 DE44307146 DE33103869 DE33103885 DE33103893 DE33103897 DE33103891 DE33103967 DE33103971 DE44305984 DE44307120 DE44307144 DE33104081 DE33104086 DE33104087 DE33104092 DE33104102 DE33104103 DE33104093 DE33104100 DE33104101 DE33104105 DE33104106 DE33104109 DE33104133 DE33104136 DE33104138 DE33104139 DE33104141 DE33104142 DE33104145 DE33104146 DE33104149 DE44307188 DE44307189 DE44307200 DE33103754 DE33104872 DE44307343 DE44307333 DE44307336 DE44307342 DE44307346 DE22801354 DE44307707 DE44307712 DE44307718 DE44307729 DE33103863 DE33103874 DE33104535 DE33105093 DE44306755 DE44306758 DE44306785 DE44306787 DE24702827 DE33103103 DE33103118 DE33103763 DE33103764 DE33105004 DE44306956 DE44306960 DE44306991 DE44307807 DE44307813 DE44307814 DE44307816 DE44307817 DE44307818 DE44307819 DE44307820 DE44307821 DE44307822 DE44307823 DE44307824 DE44307826 DE44307827 DE44307828 DE44307830 DE44307831 DE44307834 DE44307215 DE33103331 DE33103333 DE33103347 DE33103358 DE33103462 DE33103463 DE33103776 DE33103777 DE33103778 DE33103779 DE33103783 DE33103785 DE33103786 DE33103787 DE33103788 DE33103789 DE33103791 DE33103792 DE33103793 DE33103794 DE33103795 DE22801762 DE22801774 DE22801779 DE24702057 DE33105272 DE33105286 DE33105287 DE33105290 DE33105292 DE33105293 DE33105300 DE33105301 DE33105303 DE33105305 DE33105311 DE33105317 DE33105319 DE33105322 DE33105348 DE24702783 DE33104562 DE44305720 DE44305724 DE44305732 DE33105330 DE33103393 DE33104241 DE33104243 DE33104247 DE44307520 DE44307522 DE44307524 DE44307526 DE44307527 DE44307529 DE44307537 DE44307538 DE44307640 DE44307641 DE44307653 DE44307654 DE44307655 DE44307657 DE44307661 DE44307665 DE44307667 DE44307671 DE44307672 DE44307680 DE44307683 DE44307684 DE44307685 DE44307686 DE44307688 DE44307689 DE44307690 DE44307647 DE44307659 DE44307660 DE44307662 DE44307666 DE44307668 DE44307693 DE44307694 DE44307695 DE44307697 DE44307698 DE44307699 DE44307702 DE44307704 DE44307705 DE44307706 DE44307203 DE22801263 DE22801265 DE22801267 DE33104476 DE22801225 DE22801231 DE22801149 DE24701902 DE24701904 DE33103616 DE33103622 DE33103637 DE33103647 DE33103711 DE33103713 DE33103714 DE33103716 DE33103717 DE33103720 DE33103721 DE33103723 DE33103725 DE33103727 DE33103730 DE24702773 DE33103615 DE22801778 DE33103903 DE33103921 DE24702640 DE44307681 DE44305610 DE44307618 DE44305929 DE44306900 DE44306906 DE44306908 DE44306909 DE33103273 DE33103274 DE33103275 DE33103528 DE33103533 DE33103537 DE33103540 DE33103541 DE33103504 DE33103505 DE33103508 DE33103511 DE33103534 DE33103535 DE33103543 DE33103544 DE33103547 DE33103549 DE33103550 DE33103551 DE33103555 DE33103523 DE33103524 DE33103525 DE33103526 DE33103527 DE33103529 DE33103530 DE33103531 DE33103532 DE33103536 DE33103538 DE33103539 DE33103542 DE33103494 DE33103503 DE33103506 DE33103507 DE33103509 DE33103510 DE33103513 DE33103514 DE33103515 DE33103516 DE33103518 DE33103519 DE33103520 DE33103521 DE33103522 DE33103962 DE33103979 DE44307645 DE44307648 DE44307649 DE33104628 DE33104629 DE33104631 DE33104645 DE33104647 DE33104651 DE33104653 DE33104654 DE33104658 DE33104664 DE44305864 DE44305866 DE44305868 DE44305870 DE44305873 DE44305874 DE44305875 DE44305879 DE44305882 DE44305865 DE44305867 DE44305876 DE44305883 DE22801484 DE22801485 DE22801487 DE24702674 DE24702675 DE24702676 DE24702678 DE24702679 DE24702681 DE24702682 DE24702683 DE24702685 DE24702686 DE24702687 DE24702688 DE24702689 DE24702690 DE24702692 DE24702693 DE24702699 DE24702700 DE24702701 DE24702704 DE24702705 DE24702708 DE24702709 DE24702711 DE24702718 DE24702720 DE24702721 DE24702722 DE24702716 DE24702717 DE22801377 DE22801379 DE22801390 DE22801392 DE22801396 DE22801457 DE22801458 DE22801459 DE22801461 DE22801462 DE22801463 DE22801464 DE22801465 DE22801466 DE22801467 DE22801473 DE22801475 DE22801476 DE22801489 DE33102955 DE24701913 DE24701914 DE24701916 DE24701920 DE24701923 DE24701924 DE24701926 DE24701927 DE24701928 DE24701930 DE24701931 DE24701935 DE24701936 DE24701937 DE24701938 DE24701940 DE24701941 DE24701946 DE24701958 DE24701978 DE24701980 DE24701981 DE22801302 DE22801317 DE22801319 DE22801323 DE22801324 DE22801325 DE22801336 DE22801341 DE22801343 DE22801345 DE22801346 DE22801347 DE22801337 DE22801338 DE22801339 DE22801340 DE22801342 DE22801344 DE22801350 DE22801352 DE22801356 DE22801351 DE33104073 DE33104074 DE33104080 DE33104082 DE33104085 DE33104088 DE33104129 DE33103293 DE33103300 DE33103302 DE33103284 DE33103290 DE33103295 DE33103842 DE33103868 DE33103878 DE33103879 DE44306059 DE44306066 DE44306070 DE24702061 DE24702879 DE24702953 DE33103047 DE44306031 DE44306033 DE44307756 DE44307774 DE44307778 DE44307782 DE44307384 DE24702665 DE24702667 DE24702669 DE33102996 DE33103002 DE33103003 DE33103022 DE33103218 DE33105142 DE33105165 DE33105172 DE33105173 DE33105176 DE33105178 DE33105179 DE33105184 DE44307762 DE44307765 DE44305653 DE44305656 DE44305659 DE44305662 DE44305740 DE44305750 DE44305760 DE33102956 DE33105028 DE33105065 DE33105070 DE33105071 DE33105072 DE33105074 DE33105075 DE33105076 DE33105080 DE33105081 DE33105082 DE33105084 DE33105085 DE33105088 DE33105089 DE33105090 DE33105091 DE33105095 DE33104160 DE33104172 DE33104222 DE33104235 DE33104237 DE33104259 DE33104269 DE33104167 DE33104184 DE33104205 DE33104210 DE33104218 DE33104229 DE33104250 DE33104289 DE33104220 DE33104257 DE22801532 DE33104605 DE33104606 DE33104609 DE33104611 DE33104612 DE33104617 DE33104619 DE33104620 DE22801490 DE22801528 DE22801529 DE22801530 DE22801531 DE24702049 DE24702050 DE24702063 DE44305658 DE44305661 DE44305669 DE44305670 DE44305671 DE44305672 DE44305753 DE44305756 DE44305757 DE44305758 DE44305759 DE44307841 DE44307848 DE44307850 DE44307855 DE44307795 DE44307799 DE44307802 DE44307803 DE44307804 DE44307806 DE44307811 DE44307812 DE44307835 DE44307836 DE44307838 DE44307839 DE44307853 DE44307837 DE44307840 DE44307844 DE22801039 DE22801044 DE22801468 DE22801469 DE22801474 DE24702666 DE33103466 DE33103468 DE33103470 DE33103477 DE33103479 DE44306670 DE44306689 DE33104549 DE33104570 DE33104574 DE22801349 DE22801363 DE22801367 DE22801371 DE22801370 DE44306873 DE22801657 DE22801661 DE22801668 DE22801669 DE22801681 DE22801652 DE22801653 DE22801654 DE22801656 DE22801658 DE22801659 DE22801660 DE22801662 DE22801663 DE22801664 DE22801666 DE22801667 DE22801670 DE22801671 DE22801674 DE22801680 DE22801682 DE22801683 DE22801688 DE33102961 DE33103065 DE33103078 DE33103115 DE33103184 DE33103185 DE33103186 DE33103187 DE33103188 DE33103189 DE33103190 DE33103191 DE33103193 DE33103194 DE33103195 DE33103196 DE33103197 DE33103198 DE33103199 DE33103200 DE33103201 DE33103202 DE33103203 DE33103224 DE33103228 DE33103233 DE33103235 DE33103236 DE33103237 DE33103239 DE33103240 DE33103241 DE33103243 DE22801353 DE33103546 DE33104906 DE33104907 DE33104920 DE44306902 DE22801568 DE22801569 DE22801572 DE22801573 DE22801579 DE22801584 DE22801586 DE22801591 DE44307250 DE44307290 DE33103282 DE24702562 DE24702567 DE24702569 DE24702570 DE24702560 DE24702561 DE24702564 DE24702566 DE24702571 DE33104331 DE33104339 DE33104341 DE33104345 DE33104350 DE33104377 DE33104378 DE33104379 DE33104381 DE33104382 DE33104383 DE33104385 DE33104386 DE33104387 DE33104389 DE33104390 DE33104391 DE33104392 DE33104393 DE33104395 DE33104396 DE33104397 DE33104406 DE33104409 DE33104410 DE33104411 DE33104412 DE33104413 DE33104418 DE33104419 DE33104420 DE33104421 DE33104422 DE33104423 DE33104424 DE33104425 DE33104426 DE33104427 DE33104428 DE33104429 DE33104430 DE33104431 DE33104432 DE33104433 DE33104434 DE33104435 DE33104436 DE33104437 DE24702496 DE24702499 DE44306810 DE24702491 DE33105098 DE33105531 DE33105537 DE33105538 DE33105552 DE33105560 DE33105561 DE33105562 DE33105565 DE33105566 DE33105567 DE33105570 DE33105574 DE33105578 DE33105579 DE33105580 DE33105582 DE33105584 DE33105586 DE33105587 DE33105588 DE33105595 DE33105596 DE33105597 DE33105598 DE33105599 DE33105601 DE33105603 DE33105605 DE33105606 DE33105607 DE44306897 DE44306914 DE44306932 DE44306935 DE44306936 DE44306937 DE44306938 DE44306939 DE44306940 DE44306941 DE44306943 DE24702902 DE24702909 DE24702910 DE24702911 DE24702913 DE24702915 DE24702924 DE24702926 DE24702927 DE24702933 DE44307553 DE44307559 DE33103329 DE44306001 DE44306154 DE44307590 DE24702905 DE24702916 DE24702914 DE24702917 DE44306931 DE44306933 DE44306934 DE44306942 DE44306944 DE44306945 DE44306947 DE44306948 DE44305815 DE44305832 DE44305854 DE44305855 DE44305856 DE44305857 DE44305860 DE44305861 DE44306949 DE24702903 DE24702904 DE24702939 DE24702906 DE24702654 DE24702660 DE44306788 DE44306789 DE44306790 DE44306791 DE44306792 DE44306794 DE44306797 DE44306804 DE44306805 DE44306808 DE44306809 DE44306811 DE44306813 DE44306814 DE44306816 DE44307214 DE44306795 DE44306803 DE44306807 DE44306812 DE24702734 DE33104338 DE33104372 DE33104373 DE33104380 DE33104407 DE33104408 DE33104415 DE33104416 DE33104530 DE33104531 DE33104536 DE33104539 DE33104540 DE33104544 DE33104547 DE33104898 DE33104901 DE33104922 DE33104923 DE33105386 DE33105387 DE33105388 DE33105390 DE33105391 DE33105399 DE33105400 DE33105401 DE33105403 DE33105404 DE33105406 DE33105407 DE33105410 DE33105419 DE33105425 DE33105429 DE33105430 DE33105431 DE33105433 DE33105434 DE33105439 DE33105441 DE33105442 DE33105444 DE33105445 DE33105447 DE33105448 DE33105450 DE33105457 DE33105458 DE33105459 DE33105461 DE33105462 DE33105463 DE33105465 DE33105467 DE33105470 DE33105471 DE33105472 DE33105473 DE33105476 DE33105477 DE44306793 DE44305621 DE44306988 DE33105392 DE33105411 DE33105427 DE33105428 DE33105435 DE33105446 DE33105451 DE33105453 DE33105456 DE33105460 DE33105464 DE33105466 DE33105468 DE33104366 DE33104371 DE33104545 DE33104886 DE33104903 DE22801423 DE22801425 DE22801287 DE22801289 DE22801291 DE44306241 DE44306627 DE44306989 DE44306982 DE33103583 DE33104564 DE33104566 DE33104572 DE22801276 DE22801295 DE22801296 DE33103412 DE44306632 DE33104567 DE33103244 DE33103247 DE33103249 DE33103250 DE33103252 DE33103254 DE33103255 DE33103256 DE33103257 DE33103258 DE33103260 DE33103261 DE33103262 DE33103263 DE33103207 DE33103264 DE33103265 DE33103270 DE33103271 DE33103272 DE33103277 DE33103279 DE33103283 DE33103221 DE33103209 DE33103220 DE33103268 DE44307724 DE44307726 DE44307734 DE44307735 DE44307743 DE44307715 DE44307716 DE44307719 DE44307720 DE44307721 DE44305824 DE44305825 DE44305827 DE44305829 DE44305831 DE44305833 DE33104019 DE33104046 DE33104030 DE33104031 DE33104032 DE33104033 DE33104035 DE33104036 DE33104037 DE33104039 DE33104040 DE33104041 DE33104042 DE33104044 DE33104048 DE33104051 DE33104052 DE33104053 DE33104054 DE33104055 DE33104056 DE33104057 DE33104058 DE33104059 DE33104060 DE33104061 DE33104062 DE33104063 DE33104064 DE33104065 DE33104066 DE33104067 DE33104068 DE33104069 DE44306156 DE44306157 DE44306160 DE44306161 DE44306166 DE44306168 DE24702743 DE24702744 DE33103904 DE33103918 DE33103573 DE24702248 DE24702264 DE33104401 DE33104403 DE33104405 DE44306149 DE44306243 DE33103712 DE33103724 DE33104935 DE33104943 DE44306970 DE44306980 DE24701876 DE24701879 DE24701883 DE44307251 DE44307256 DE44307258 DE44307261 DE44307540 DE44307546 DE44307547 DE24702645 DE24702647 DE24702649 DE24702651 DE33103053 DE33103057 DE33103082 DE33103110 DE44306872 DE44306903 DE44306907 DE44307023 DE44305773 DE44306815 DE33103310 DE33103315 DE33103320 DE22801277 DE24702138 DE24702139 DE24702384 DE33103964 DE33103975 DE33104832 DE33104841 DE24702213 DE24702224 DE24702388 DE24702391 DE24702392 DE24702405 DE24702411 DE24702412 DE44306848 DE44306850 DE44306851 DE44306852 DE44306853 DE44306854 DE44306855 DE33105320 DE44305657 DE44305660 DE44306857 DE33103631 DE33103632 DE44305727 DE44305790 DE44305792 DE44305793 DE44305802 DE44305813 DE44307224 DE44307238 DE44307240 DE44307713 DE44307833 DE33105506 DE33105295 DE24702485 DE44305715 DE44306998 DE44307007 DE33103598 DE33103607 DE44306383 DE44306387 DE44307216 DE44306588 DE44306631 DE44306636 DE33105309 DE33105314 DE33105306 DE33105324 DE44305703 DE44305762 DE44305901 DE44305908 DE44305705 DE33103553 DE33103557 DE33103558 DE33103559 DE33103560 DE33103561 DE33103562 DE33103718 DE33103728 DE44307586 DE44307591 DE33103831 DE33103838 DE33103841 DE33103849 DE33103850 DE33103853 DE33103855 DE33103856 DE33105478 DE33105479 DE33105486 DE33105488 DE33105495 DE33105499 DE33105502 DE33105504 DE33105505 DE33103753 DE33103760 DE33104991 DE33104996 DE33104997 DE33104998 DE33104999 DE33105013 DE33105018 DE33105019 DE33105020 DE33105023 DE33105024 DE22801316 DE22801260 DE22801262 DE44305718 DE44305730 DE44307294 DE44307295 DE44307314 DE44307315 DE44307691 DE33104097 DE44305699 DE44305700 DE44305704 DE44305877 DE44305820 DE44305858 DE44305862 DE44305871 DE44305872 DE44305878 DE44305880 DE44305881 DE44306078 DE44306079 DE44306076 DE44307207  
FEI Number 1218950
Recalling Firm/
Manufacturer
Philips Electronics North America Corporation
3000 Minuteman Rd
Andover MA 01810-1032
For Additional Information Contact
800-722-9377
Manufacturer Reason
for Recall
Manufacturing defect may cause localized heating, which may result in localized skin burn.
FDA Determined
Cause 2
Device Design
ActionOn May 4, 2017, Philips sent a field safety notice to their consignees. The Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue: Philips will exchange affected NMT Patient cables at the customer site. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Field Safety Notice: Immediately identify all affected cables, remove them from inventory and hold them for exchange by a Philips Healthcare representative or authorized service provider.
Quantity in Commerce9120 devices
DistributionUS: AK AZ CA CO FL HI IA IL IN KY MA MD MI MN MO MS NC NE NH NJ NM NV NY OH OR PA TX VA WA WI WV OUS: Algeria Argentina Australia Austria Bahrain Belarus Belgium Bolivia Brazil Canada Chile China Colombia Costa Rica Denmark Dominican Rep Egypt Finland France Germany Hong Kong Hungary Iceland India Indonesia Ireland Israel Italy Japan Latvia Lebanon Libya Lithuania Luxembourg Malaysia Mexico Mongolia Netherlands New Zealand Norway Oman Pakistan Peru Philippines Poland Portugal Qatar Romania Russia Russian Fed. Saudi Arabia Singapore Slovenia South Africa South Korea Spain Sweden Switzerland Taiwan Thailand Tunisia Turkey Ukraine United Kingdom Utd.Arab.Emir.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MHX
-
-