• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall MAXIMIS Pedicle Screw Spinal Fixation System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall MAXIMIS Pedicle Screw Spinal Fixation Systemsee related information
Date Initiated by FirmJune 21, 2017
Create DateAugust 10, 2017
Recall Status1 Terminated 3 on November 27, 2018
Recall NumberZ-2983-2017
Recall Event ID 77500
510(K)NumberK140359 
Product Classification Orthosis, spinal pedicle fixation - Product Code MNI
ProductPALLAS M/MAXIMIS MIS Tapper 6.5mm. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.
Code Information Catalog No. 0902-1008; Known Lot No. WNV120320, WNV110830, NHS170405.
FEI Number 3010212187
Recalling Firm/
Manufacturer
Valorem Surgical LLC
3963 W Belmont Ave Ste 9
Chicago IL 60618-5129
For Additional Information ContactJoseph Jin
312-852-2700
Manufacturer Reason
for Recall
Devices are not consistent with Quality System Requirements.
FDA Determined
Cause 2
Device Design
ActionValorem Surgical sent an Urgent Medical Device Recall letter dated June 21, 2017, informing consignees of the product removal and instructing consignees to return any product to Valorem Surgical LLC to minimize potential risk to patients. Consignees are to fill out the attached Medical Device Recall Return Response form and send it back to the firm by email at global.safety@ValoremSurgical.com or by fax at 312-275-7117, or by mail. For questions pertaining to this recall, please submit an email inquiry to global.safety@ValoremSurgical.com
Quantity in Commerce262 units total
DistributionNationwide Distribution - US including CA and VA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MNI
-
-