• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall GORE TAG, THORACIC ENDOPROSTHESIS, CONFORMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall GORE TAG, THORACIC ENDOPROSTHESIS, CONFORMABLEsee related information
Date Initiated by FirmSeptember 25, 2017
Date PostedOctober 25, 2017
Recall Status1 Terminated 3 on August 27, 2019
Recall NumberZ-0048-2018
Recall Event ID 78132
PMA NumberP040043 
Product Classification System, endovascular graft, aortic aneurysm treatment - Product Code MIH
ProductGORE TAG, THORACIC ENDOPROSTHESIS, CONFORMABLE. The GORE TAG Thoracic Endoprosthesis is intended for endovascular repair of all lesions of the descending thoracic aorta.
Code Information Serial Numbers: TGE212110 TGE212110 TGE212110K TGU212110 TGU212110H TGU212110B TGU212110J TGE262610 TGE262610 TGE262610K TGU262610 TGU262610H TGU262610B TGU262610J TGE282810 TGE282810 TGE282810K TGU282810 TGU282810H TGU282810B TGU282810J TGE282815 TGE282815 TGE282815K TGU282815 TGU282815H TGU282815B TGU282815J TGE313110 TGE313110 TGE313110K TGU313110 TGU313110H TGU313110B TGU313110J TGE313115 TGE313115 TGE313115K TGU313115 TGU313115H TGU313115B TGU313115J TGE343410 TGE343410 TGE343410K TGU343410 TGU343410H TGU343410B TGU343410J TGE343415 TGE343415 TGE343415K TGU343415 TGU343415H TGU343415B TGU343415J TGE343420 TGE343420 TGE343420K TGU343420 TGU343420H TGU343420B TGU343420J TGE373710 TGE373710 TGE373710K TGU373710 TGU373710H TGU373710B TGU373710J TGE373715 TGE373715 TGE373715K TGU373715 TGU373715H TGU373715B TGU373715J TGE373720 TGE373720 TGE373720K TGU373720 TGU373720H TGU373720B TGU373720J TGE404010 TGE404010 TGE404010K TGU404010 TGU404010H TGU404010B TGU404010J TGE404015 TGE404015 TGE404015K TGU404015 TGU404015H TGU404015B TGU404015J TGE404020 TGE404020 TGE404020K TGU404020 TGU404020H TGU404020B TGU404020J TGE454510 TGE454510 TGE454510K TGU454510 TGU454510H TGU454510B TGU454510J TGE454515 TGE454515 TGE454515K TGU454515 TGU454515H TGU454515B TGU454515J TGE454520 TGE454520 TGE454520K TGU454520 TGU454520H TGU454520B TGU454520J TGE262110 TGE262110 TGE262110K TGU262110 TGU262110H TGU262110B TGU262110J TGE312610 TGE312610 TGE312610K TGU312610 TGU312610H TGU312610B TGU312610J TGU212115J TGU212120J TGU262615J TGU262620J TGU282820J TGU313120J  
FEI Number 2017233
Recalling Firm/
Manufacturer
W L Gore & Associates, Inc.
1500 N Fourth St.
Flagstaff AZ 86003-2400
For Additional Information Contactcustomer service
800-528-8763
Manufacturer Reason
for Recall
Incomplete and/or partial deployment of the CTAG Device during the endovascular procedure.
FDA Determined
Cause 2
Unknown/Undetermined by firm
ActionThe firm contact their consignee with a safety notice by letter on 09/25/2017. Gore is updating its Instructions for Use (IFU) to include the following new warnings and precautions: " If abnormal or inconsistent deployment line resistance is felt during deployment initiation, STOP deployment action immediately. If device remains constrained, remove device through the introducer sheath. If resistance is felt during removal through the sheath, stop and withdraw device and introducer sheath together. " If the device is in a partially deployed state and remains attached to the catheter, physicians should strongly consider conversion to immediate open surgical repair to avoid additional procedure time and potential harm from additional endovascular maneuvers.
Quantity in Commerce12,197units
DistributionWorldwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA DatabasePMAs with Product Code = MIH
-
-