| Class 2 Device Recall UniCel DxH 600 Coulter Cellular Analysis System | |
Date Initiated by Firm | November 03, 2017 |
Create Date | December 13, 2017 |
Recall Status1 |
Terminated 3 on June 19, 2018 |
Recall Number | Z-0237-2018 |
Recall Event ID |
78611 |
510(K)Number | K140911 |
Product Classification |
Counter, differential cell - Product Code GKZ
|
Product | UniCel DxH600, Part Number B23858, automated hematology analyzer
(Note: the DxH 600 is the same instrument as the DxH 800 except that it is a bench top unit with no instrument cart, computer stand, and no connectivity capabilities) |
Code Information |
Serial Number AZ06682 |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd Brea CA 92821-6232
|
For Additional Information Contact | 469-961-4941 |
Manufacturer Reason for Recall | The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position. |
FDA Determined Cause 2 | Device Design |
Action | The firm notified its consignees of the recall on 11/03/2017 by letter. The letter stated the issue and provided the following instructions:
"To avoid a biohazardous spill, do not use the right (green) position of the Single-Tube Station for sample aspiration until the shield is removed.
You may remove the shield following the instructions in Attachment 1 or contact your local Beckman Coulter Representative for further assistance." |
Quantity in Commerce | 1 unit |
Distribution | CA, AR, HI, CA, OH, WA, MD, Canada, Japan |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = GKZ
|
|
|
|