| | Class 2 Device Recall PowerLOAD Cot Fastening System |  |
| Date Initiated by Firm | March 12, 2018 |
| Date Posted | May 01, 2018 |
| Recall Status1 |
Terminated 3 on May 12, 2020 |
| Recall Number | Z-1389-2018 |
| Recall Event ID |
79630 |
| 510(K)Number | K113598 |
| Product Classification |
Stretcher, wheeled - Product Code FPO
|
| Product | Power-LOAD Cot Fastening System, Model #6390, Catalog #6390-000-000, UDI: 07613327261523
Power-LOAD is intended to assist with loading and unloading of a compatible wheeled stretcher (ambulance cot) to and from a transport vehicle and to secure the ambulance cot during transport. The device has a maximum safe working load of 870 lb (395 kg), which includes the weight of the ambulance cot, patient, and equipment attached to the cot (such as oxygen bottles, monitors, and pumps). The intended users of the device are trained professionals, including emergency medical service and medical care center personnel, as well as medical first responders, service technicians and installers. |
| Code Information |
Serial numbers: 160139700 160139701 160139702 160139704 160139705 160139706 160139707 160139708 160139709 160139710 160139711 160139712 160139713 160139714 160139743 160139744 160139745 160139746 160139747 160139748 160139749 160139750 160139751 160139752 160139753 160139754 160139755 160139756 160139804 160139805 160139806 160139807 160139808 160139809 160139810 160139811 160139812 160139813 160139814 160139815 160139816 160139817 160139818 160139819 160139820 160139821 160139822 160139823 160139824 160139825 160139826 160139827 160139828 160139829 160139830 160139831 160139832 160139833 160139834 160139835 160139836 160139837 160139838 160139839 160139894 160139895 160139896 160139897 160139898 160139899 160139900 160139901 160139902 160139903 160139904 160139905 160139906 160139907 160139908 160139909 160139910 160139911 160139912 160139913 160139914 160139915 160139916 160139917 160139918 160139919 160139920 160139921 160139922 160139923 160139924 160139925 160139926 160139927 160139928 160139929 160139930 160139931 160139982 160139983 160139984 160139985 160139986 160139987 160139988 160139989 160139990 160139991 160139992 160139993 160139994 160139995 160139996 160139997 160139998 160139999 160140027 160140028 160140029 160140030 160140031 160140032 160140033 160140034 160140035 160140036 160140037 160140038 160140039 160140040 160140041 160140042 160140043 160140044 160140045 160140046 160140047 160140048 160140049 160140075 160140076 160140077 160140078 160140079 160140080 160140081 160140082 160140083 160140084 160140085 160140086 160140087 160140088 160140089 160140090 160140091 160140127 160140128 160140129 160140130 160140131 160140132 160140133 160140134 160140135 160140136 160140137 160140138 160140139 160140140 160140141 160140142 160140143 160140144 160140145 160140146 160140340 160140341 160140342 160140343 160140344 160140345 160140346 160140347 160140348 160140349 160140350 160140351 160140352 160140353 160140354 160140395 160140396 160140397 160140398 160140399 160140749 160140751 160140752 160140753 160140754 160140755 160140756 160140757 160140758 160140759 160140760 160140793 160140794 160140795 160140796 160140797 160140798 160140799 160140800 160140801 160140802 160140803 160140804 160140805 160140806 160140807 160140808 160140809 160140810 160140811 160140812 160140813 160140814 160140815 160140816 160140817 160140818 160140819 160140820 160140821 160140822 160140823 160140824 160140825 160140826 160140827 160140828 160140829 160140830 160140831 160140832 160140833 160140834 160140835 160140836 160140837 160140838 160140839 160140840 160140878 160140879 160140880 160140881 160140882 160140883 160140884 160140885 160140886 160140887 160140888 160140889 160140890 160140891 160140892 160140893 160140894 160140922 160140923 160140924 160140925 160140926 160140927 160140928 160140929 160140930 160140931 160140932 160140933 160140934 160140935 160140936 160140937 160140938 160140939 160140940 160140941 160140962 160140963 160140964 160140965 160140966 160140967 160140968 160140969 160140970 160140971 160140972 160140973 160140974 160140975 160140976 160140977 160140978 160141031 160141032 160141033 160141035 160141036 160141037 160141039 160141040 160141041 160141042 160141043 160141044 160141045 160141046 160141047 160141048 160141550 160141551 160141557 160141558 160141559 160141560 160141561 160141562 160141563 160141564 160141565 160141566 160240050 160240051 160240052 160240053 160240054 160240055 160240056 160240057 160240058 160240059 160240060 160240061 160240062 160240063 160240064 160240065 160240066 160240400 160240401 160240402 160240403 160240404 160240405 160240406 160240407 160240408 160240409 160240410 160240411 160240412 160240450 160240462 160240463 160240464 160240465 160240466 160240467 160240468 160240469 160240470 160240471 160240472 160240473 160240474 160240475 160240476 160240496 160240497 160240498 160240499 160240500 160240501 160240502 160240503 160240504 160240505 160240506 160240507 160240508 160240509 160240510 160240511 160240512 160240513 160240514 160240515 160240516 160240517 160240518 160240519 160240520 160240521 160240522 160240523 160240524 160240525 160240526 160240527 160240564 160240565 160240566 160240567 160240568 160240569 160240570 160240571 160240572 160240573 160240574 160240575 160240576 160240577 160240578 160240579 160240580 160240581 160240582 160240583 160240584 160240585 160240586 160240587 160240588 160240589 160240590 160240591 160240592 160240593 160240594 160240595 160240596 160240614 160240707 160240708 160240709 160240710 160240711 160240712 160240713 160240714 160240715 160240716 160240717 160240718 160240719 160240720 160240721 160240722 160240723 160240724 160240725 160240726 160241140 160241141 160241142 160241143 160241144 160241145 160241146 160241147 160241148 160241149 160241150 160241151 160241152 160241153 160241154 160241155 160241156 160241157 160241158 160241159 160241190 160241191 160241192 160241193 160241194 160241195 160241196 160241197 160241198 160241199 160241200 160241201 160241202 160241203 160241204 160241205 160241206 160241207 160241208 160241209 160241210 160241246 160241247 160241248 160241249 160241250 160241251 160241252 160241253 160241254 160241255 160241256 160241257 160241258 160241259 160241260 160241261 160241262 160241263 160241264 160241265 160241314 160241315 160241316 160241317 160241318 160241319 160241320 160241321 160241322 160241323 160241324 160241325 160241326 160241327 160241328 160241329 160241330 160241331 160241332 160241333 160241356 160241357 160241358 160241359 160241360 160241361 160241362 160241363 160241364 160241365 160241366 160241367 160241368 160241369 160241370 160241371 160241372 160241373 160241374 160241375 160241376 160241377 160241378 160241379 160241380 160241381 160241382 160241383 160241384 160241385 160241386 160241387 160241388 160241389 160241390 160241391 160241392 160241393
160241394 160241395 160241396 160241397 160241398 160241462 160241463 160241464 160241465 160241466 160241467 160241468 160241469 160241470 160241471 160241472 160241473 160241474 160241475 160241476 160241477 160241478 160241479 160241480 160241481 160241482 160241483 160241484 160241485 160241486 160241541 160241542 160241543 160241544 160241545 160241546 160241547 160241548 160241549 160241550 160241551 160241552 160241553 160241554 160241555 160241556 160241557 160241558 160241559 160241560 160241561 160241562 160241563 160241564 160241565 160241606 160241607 160241608 160241609 160241610 160241611 160241612 160241613 160241614 160241615 160241616 160241617 160241618 160241619 160241620 160241621 160241622 160241623 160241624 160241625 160241626 160241627 160241628 160241629 160241630 160241631 160241632 160241633 160241634 160241635 160241636 160241637 160241638 160241639 160241640 160241641 160241642 160241643 160241644 160241645 160241646 160241647 160241648 160241649 160241650 160241651 160241652 160241653 160241654 160241655 160339106 160339107 160339108 160339109 160339110 160339111 160339112 160339113 160339114 160339115 160339116 160339117 160339118 160339119 160339120 160339121 160339122 160340158 160340159 160340160 160340161 160340162 160340163 160340164 160340165 160340166 160340167 160340168 160340169 160340170 160340171 160340172 160340173 160340174 160340175 160340176 160340177 160340178 160340179 160340180 160340181 160340182 160340183 160340184 160340185 160340186 160340187 160340188 160340189 160340191 160340192 160340193 160340194 160340195 160340196 160340197 160340198 160340199 160340200 160340201 160340202 160340203 160340204 160340205 160340206 160340207 160340208 160340209 160340210 160340211 160340212 160340213 160340214 160340215 160340216 160340217 160340218 160340219 160340220 160340221 160340222 160340223 160340224 160340225 160340227 160340228 160340229 160340230 160340231 160340232 160340233 160340234 160340236 160340237 160340238 160340239 160340274 160340278 160340279 160340280 160340281 160340282 160340283 160340284 160340285 160340286 160340287 160340289 160340290 160340291 160340292 160340293 160340294 160340295 160340296 160340297 160340300 160340301 160340302 160340303 160340304 160340305 160340306 160340307 160340308 160340309 160340310 160340311 160340312 160340313 160340314 160340315 160340316 160340317 160340318 160340319 160340320 160340321 160340322 160340323 160340324 160340325 160340326 160340327 160340328 160340329 160340330 160340331 160340332 160340333 160340334 160340335 160340336 160340337 160340338 160340339 160340340 160340341 160340342 160340343 160340344 160340345 160340346 160340347 160340348 160340349 160340350 160340351 160340352 160340353 160340354 160340355 160340356 160340357 160340358 160340359 160340360 160340361 160340362 160340363 160340364 160340365 160340366 160340367 160340368 160340369 160340370 160340371 160340374 160340375 160340376 160340377 160340378 160340379 160340380 160340381 160340382 160340383 160340384 160340385 160340386 160340387 160340388 160340389 160340390 160340391 160340392 160340393 160340417 160340418 160340419 160340420 160340421 160340422 160340423 160340424 160340425 160340426 160340427 160340428 160340429 160340430 160340485 160340486 160340487 160340488 160340489 160340490 160340491 160340492 160340493 160340494 160340495 160340496 160340497 160340498 160340499 160340500 160340501 160340502 160340503 160340504 160340505 160340506 160340507 160340508 160340509 160340790 160340791 160340792 160340794 160340795 160340796 160340797 160340798 160340799 160340800 160340801 160340802 160340803 160340804 160340805 160340806 160340807 160340808 160340846 160340847 160340848 160340849 160340850 160340851 160340852 160340853 160340854 160340855 160340856 160340857 160340858 160340859 160340860 160340861 160340862 160341017 160341018 160341019 160341020 160341021 160341022 160341023 160341024 160341025 160341026 160341027 160341028 160341029 160341030 160341031 160341032 160341033 160341034 160341035 160341036 160341037 160341038 160341039 160341040 160341041 160341042 160341043 160341044 160341045 160341046 160341047 160341048 160341049 160341050 160341051 160341052 160341053 160341054 160341055 160341056 160341057 160341058 160341059 160341060 160341061 160341062 160341063 160341064 160341065 160341066 160341067 160341068 160341069 160341070 160341071 160341072 160341150 160341151 160341152 160341153 160341154 160341155 160341156 160341157 160341158 160341159 160341160 160341161 160341162 160341163 160341164 160341165 160341166 160341167 160341168 160341169 160341170 160341171 160341172 160341173 160341174 160341250 160341251 160341252 160341253 160341254 160341255 160341256 160341257 160341258 160341259 160341260 160341261 160341262 160341263 160341264 160341265 160341266 160341267 160341268 160341269 160439110 160439111 160439112 160439113 160439114 160439115 160439116 160439117 160439118 160439119 160439120 160439121 160439122 160439123 160439124 160439125 160439126 160439127 160439128 160439129 160439130 160439131 160439132 160439133 160439134 160439186 160439187 160439188 160439189 160439190 160439191 160439193 160439194 160439195 160439196 160439197 160439198 160439199 160439200 160439201 160439202 160439203 160439204 160439205 160439206 160439207 160439208 160439209 160439210 160439211 160439212 160439213 160439214 160439215 160439216 160439217 160439218 160439219 160439220 160439221 160439222 160439223 160439224 160439225 160439285 160439286 160439287 160439288 160439289 160439290 160439291 160439292 160439293 160439294 160439295 160439296 160439297 160439298 160439299 160439300 160439301 160439302 160439303 160439304 160439305 160439306 160439307 160439308 160439309 160439310 160439311 160439312 160439313 160439314 160439315 160439316 160439317 160439318 160439319 160439320 160439321 160439322 160439323 160439324 160439325 160439326 160439329 160439330 160439332 160439333 160439334 160439340 160439341 160439342 160439343 160439344 160439345 160439346 160439347 160439348 160439349 160440103 160440104 160440105 160440106 160440107 160440108 160440109 160440110 160440111 160440112 160440113 160440114 160440115 160440116 160440117 160440118 160440119 160440120 160440121 160440122 160440123 160440124 160440125 160440126 160440127 160440128 160440129 160440130 160440131 160440132 160440133 160440134 160440135 160440136 160440137 160440138 160440139 160440140 160440141 160440143 160440144 160440145 160440146 160440147 160440148 160440149 160440150 160440151 160440152 160440153 160440154 160440155 160440156 160440157 160440158 160440159 160440160 160440161 160440162 160440163 160440164 160440165 160440166 160440196 160440197 160440198 160440199 160440200 160440201 160440202 160440203 160440204 160440205 160440206 160440207 160440208 160440209 160440277 160440278 160440279 160440280 160440281 160440282 160440283 160440284 160440285 160440287 160440288 160440289 160440290 160440291 160440292 160440293 160440294 160440295 160440296 160440297 160440298 160440299 160440300 160440301 160440790 160440791 160440792 160440793 160440794 160440795 160440796 160440797 160440798 160440799 160440800 160440801 160440802 160440803 160440804 160440805 160440806 160440807 160440808 160440809 160440810 160440811 160440812 160440813 160440814 160440815 160440816 160440817 160440818 160440825 160440826 160440855 160440856 160440857 160440858 160440859 160440860 160440861 160440862 160440863 160440865 160440866 160440867 160440868 160440930 160440936 160440937 160440938 160440939 160440940 160440941 160440942 160440943 160440944 160440945 160440946 160440947 160440948 160440949 160440970 160440971 160440972 160440973 160440974 160440975 160440976 160440977 160440978 160440979 160440980 160440981 160440982 160440983 160440984 160440985 160440986 160440987 160440988 160440989 160441011 160441012 160441013 160441014 160441015 160441016 160441017 160441018 160441019 160441020 160441021 160441022 160441023 160441024 160441025 160441026 160441027 160441028 160441040 160441041 160441042 160441043 160441044 160441045 160441046 160441047 160441048 160441049 160441050 160441051 160441052 160441053 160441054 160441055 160441056 160441057 160441058 160441059 160441060 160441061 160441062 160441063 160441064 160441742 160441743 160441744 160441745 160441781 160441782 160441793 160441794 160539098 160539099 160539100 160539101 160539102 160539350 160539351 160539352 160539353 160539354 160539355 160539356 160539357 160539358 160539359 160539360 160539361 160539362 160539363 160539364 160539365 160539366 160539367 160539368 160539369 160539370 160539371 160539372 160539373 160539374 160539375 160539427 160539428 160539429 160539430 160539431 160539432 160539433 160539434 160539435 160539436 160539437 160539438 160539439 160539440 160539441 160539442 160539443 160539444 160539445 160539446 160539484 160539485 160539486 160539522 160539534 160539535 160539536 160539537 160539538 160539539 160539540 160539541 160539542 160539543 160539544 160539545 160539546 160539547 160539548 160539549 160539550 160539551 160539552 160539553 160539554 160539555 160539556 160539557 160539558 160539559 160539560 160539561 160539562 160539622 160539623 160539624 160539625 160539626 160539627 160539628 160539629 160539630 160539631 160539632 160539633 160539634 160539635 160539636 160539637 160539638 160539639 160539640 160539641 160539688 160539689 160539690 160539691 160539692 160539693 160539694 160539695 160539696 160539697 160539698 160539699 160540502 160540503 160540504 160540505 160540506 160540507 160540508 160540509 160540510 160540511 160540512 160540513 160540514 160540515 160540516 160540517 160540518 160540519 160540520 160540521 160540522 160540523 160540529 160540530 160540531 160540532 160540533 160540534 160540535 160540536 160540537 160540538 160540539 160540540 160540541 160540542 160540543 160540544 160540545 160540546 160540547 160540548 160540549 160540589 160540590 160540591 160540592 160540593 160540594 160540595 160540596 160540597 160540598 160540599 160540600 160540601 160540602 160540603 160540604 160540605 160540606 160540607 160540608 160540609 160540610 160540611 160540612 160540613 160540614 160540615 160540616 160540617 160540618 160540619 160540620 160540621 160540622 160540623 160540624 160540625 160540626 160540627 160540628 160540629 160540630 160540631 160540632 160540633 160540634 160540635 160540636 160540637 160540638 160540639 160540640 160540641 160540642 160540643 160540644 160540645 160540646 160540647 160540648 160540649 160540680 160540681 160540682 160540683 160540684 160540685 160540686 160540687 160540688 160540689 160540690 160540691 160540692 160540693 160540694 160541100 160541101 160541102 160541103 160541104 160541105 160541106 160541107 160541108 160541109 160541110 160541111 160541112 160541113 160541114 160541115 160541116 160541117 160541118 160541119 160541120 160541121 160541122 160541123 160541124 160541125 160541126 160541127 160541128 160541129 160541130 160541131 160541132 160541133 160541134 160541135 160541136 160541450 160541451 160541452 160541453 160541454 160541455 160541456 160541457 160541458 160541459 160541460 160541461 160541462 160541463 160541464 160541465 160541466 160541467 160541468 160541469 160541470 160541471 160541472 160541473 160541474 160541475 160541476 160541477 160541478 160541479 160541480 160541481 160541482 160541483 160541484 160541485 160541486 160541487 160541488 160541489 160541490 160541491 160541492 160541493 160541494 160541495 160541583 160541584 160541585 160541586 160541587 160541588 160541589 160541590 160541591 160541592 160541593 160541594 160541595 160541596 160541597 160541598 160541599 160541600 160541601 160541602 160541603 160541604 160541605 160541606 160541607 160541608 160541646 160541647 160541648 160541649 160541650 160541651 160541652 160541653 160541654 160541655 160541691 160541692 160541693 160541694 160541695 160541696 160541697 160541698 160541699 160541700 160541721 160541722 160541723 160541724 160541725 160541726 160541727 160541729 160541730 160541731 160541732 160541733 160541734 160541735 160541736 160541737 160541738 160541739 160541740 160541741 160541742 160541743 160541744 160541745 160541746 160541747 160541748 160541749 160541750 160541751 160541752 160541753 160541754 160541755 160541756 160541757 160541758 160541759 160541760 160541761 160541762 160541763 160541764 160541765 160541766 160541767 160541768 160541769 160541770 160541771 160541772 160541797 160541798 160541799 160639078 160639079 160639080 160639081 160639082 160639083 160639084 160639085 160639086 160639087 160639088 160639100 160639101 160639102 160639103 160639104 160639105 160639106 160639107 160639108 160639109 160639110 160639111 160639112 160639113 160639114 160639115 160639116 160639117 160639118 160639119 160639120 160639121 160639122 160639123 160639124 160639160 160639161 160639162 160639163 160639164 160639165 160639166 160639167 160639168 160639169 160639170 160639171 160639172 160639173 160639174 160639175 160639176 160639177 160639178 160639179 160639180 160639181 160639182 160639183 160639184 160639198 160639250 160639251 160639252 160639253 160639254 160639255 160639256 160639257 160639258 160639259 160639260 160639261 160639262 160639263 160639264 160639265 160639266 160639267 160639268 160639269 160639318 160639319 160639320 160639322 160639323 160639324 160639325 160639326 160639327 160639328 160639329 160639330 160639331 160639332 160639333 160639334 160639335 160639336 160639337 160639338 160639339 160639340 160639341 160639342 160639343 160639344 160639352 160639353 160639354 160639355 160639356 160639357 160639358 160639359 160639360 160639361 160639362 160639363 160639364 160639365 160639366 160639367 160639368 160639369 160639370 160639371 160639372 160639373 160639374 160639375 160639376 160639377 160639378 160639379 160639380 160639381 160639382 160639383 160639384 160639385 160639386 160639387 160639388 160639389 160639390 160639391 160639781 160639782 160639783 160639848 160639849 160639850 160639851 160639852 160639853 160639854 160639855 160639856 160639857 160639858 160639859 160639860 160639861 160639862 160639863 160639864 160639865 160639866 160639867 160639887 160639888 160639889 160639890 160639891 160639892 160639893 160639894 160639895 160639896 160639897 160639898 160639899 160639900 160639901 160639902 160639903 160639904 160639905 160639906 160639907 160639908 160639909 160639910 160639911 160639912 160639913 160639914 160639915 160639916 160639917 160639918 160639919 160639920 160639921 160639922 160639923 160639924 160639925 160639926 160639927 160639928 160639987 160639988 160639989 160639990 160639991 160639992 160639993 160639994 160639995 160640000 160640001 160640002 160640003 160640004 160640005 160640006 160640007 160640008 160640009 160640010 160640011 160640012 160640013 160640014 160640015 160640016 160640017 160640018 160640019 160640020 160640021 160640022 160640023 160640024 160640025 160640026 160640027 160640028 160640029 160640030 160640031 160640075 160640076 160640077 160640078 160640079 160640080 160640081 160640082 160640083 160640084 160640085 160640086 160640087 160640088 160640089 160640090 160640091 160640092 160640093 160640094 160640095 160640499 160640513 160640514 160640515 160640516 160640517 160640518 160640519 160640520 160640521 160640522 160640523 160640524 160640525 160640526 160640527 160640528 160640529 160640530 160640531 160640532 160640533 160640534 160640535 160640536 160640537 160640538 160640539 160640540 160640541 160640542 160640543 160640544 160640545 160640546 160640547 160640548 160640549 160640600 160640601 160640602 160640603 160640604 160640605 160640606 160640607 160640608 160640700 160640701 160640702 160640703 160640704 160640705 160640706 160640707 160640708 160640709 160640710 160640711 160640712 160640713 160640714 160640715 160640716 160640717 160640718 160640719 160640720 160640752 160640753 160640754 160640755 160640756 160640757 160640758 160640759 160640760 160640761 160640762 160640763 160640764 160640765 160640766 160640767 160640768 160640769 160640770 160640771 160640772 160640773 160640774 160640775 160640776 160640777 160640778 160640779 160640780 160640781 160640782 160640783 160640784 160640785 160640786 160640787 160640788 160640789 160640790 160640791 160640792 160640793 160640794 160641265 160641266 160641267 160641268 160641269 160641270 160641271 160641272 160641273 160641274 160641275 160641276 160641277 160641278 160641279 160641280 160641281 160641282 160641283 160641284 160641285 160641286 160641287 160641288 160641289 160641290 160641291 160641292 160641293 160641294 160641295 160641330 160641331 160641332 160641333 160641334 160641335 160641336 160641337 160641338 160641339 160641348 160641349 160641350 160641351 160641352 160641353 160641354 160641355 160641356 160641357 160641358 160641359 160641360 160641361 160641362 160641363 160641364 160641365 160641366 160641367 160641368 160641369 160641370 160641371 160641372 160641373 160641374 160641375 160641376 160641377 160641378 160641379 160641380 160641381 160641382 160641383 160641384 160641385 160641386 160641387 160641388 160641389 160641390 160641391 160641392 160641393 160641394 160641395 160641396 160641397 160641398 160641399 160739015 160739016 160739017 160739018 160739019 160739020 160739021 160739022 160739023 160739050 160739051 160739052 160739053 160739054 160739055 160739056 160739057 160739058 160739059 160739060 160739061 160739062 160739063 160739064 160739065 160739066 160739067 160739068 160739069 160739070 160739071 160739072 160739073 160739074 160739075 160739076 160739077 160739078 160739079 160739080 160739081 160739082 160739083 160739084 160739085 160739086 160739139 160739140 160739141 160739142 160739143 160739144 160739145 160739146 160739147 160739148 160739150 160739151 160739152 160739153 160739154 160739155 160739156 160739157 160739158 160739159 160739160 160739161 160739162 160739163 160739164 160739165 160739166 160739167 160739168 160739169 160739170 160739171 160739172 160739173 160739174 160739175 160739176 160739177 160739178 160739194 160739195 160739196 160739197 160739198 160739199 160739225 160739226 160739227 160739228 160739229 160739230 160739231 160739232 160739233 160739234 160739235 160739236 160739237 160739238 160739239 160739240 160739241 160739242 160739243 160739244 160739245 160739246 160739247 160739248 160739249 160739275 160739276 160739277 160739278 160739279 160739280 160739281 160739282 160739283 160739284 160739285 160739286 160739287 160739288 160739289 160739290 160739291 160739292 160739293 160739294 160739295 160739296 160739297 160739298 160739299 160739300 160739301 160739302 160739303 160739304 160739305 160739306 160739307 160739308 160739309 160739310 160739311 160739312 160739313 160739314 160739315 160739316 160739317 160739319 160739320 160739321 160739322 160739323 160739324 160739325 160739326 160739327 160739328 160739776 160739777 160739778 160739779 160739780 160739781 160739782 160739783 160739784 160739785 160739786 160739787 160739788 160739789 160739790 160739791 160739792 160739793 160739794 160739795 160739796 160739838 160739839 160739840 160739841 160739843 160739844 160739845 160739846 160739847 160739848 160739849 160739850 160739851 160739852 160739853 160739854 160739855 160739856 160739857 160739858 160739859 160739900 160739901 160739902 160739903 160739904 160739905 160739906 160739907 160739908 160739909 160739910 160739911 160739912 160739913 160739914 160739915 160739916 160739917 160739918 160739919 160739920 160739921 160739922 160739923 160739924 160739925 160739926 160739927 160739928 160739929 160739930 160739931 160739932 160739933 160739934 160739989 160739990 160739991 160739992 160739994 160739995 160739996 160739997 160739998 160739999 160740256 160740257 160740258 160740259 160740260 160740261 160740262 160740263 160740264 160740265 160740343 160740344 160740345 160740346 160740347 160740615 160740616 160740617 160740618 160740619 160740620 160740621 160740622 160740623 160740624 160740625 160740626 160740627 160740628 160740629 160741245 160741246 160741508 160741509 160741510 160741511 160741512 160741513 160741514 160741515 160741516 160741544 160741547 160741550 160741551 160741552 160741553 160741554 160741555 160741556 160741557 160741558 160741559 160741560 160741561 160741562 160741563 160741564 160741565 160741566 160741567 160741568 160741569 160741650 160741651 160741652 160741653 160741654 160741655 160741656 160741657 160741658 160741659 160741660 160741661 160741662 160741663 160741664 160741665 160741666 160741692 160741693 160741694 160741695 160741696 160741697 160741698 160741699 160741745 160741746 160741747 160741750 160741751 160741752 160741753 160741754 160741755 160741756 160741757 160741758 160741759 160741760 160741761 160741762 160741763 160741764 160741765 160741766 160741767 160741768 160741769 160741823 160741824 160741825 160741826 160741827 160741828 160741829 160741830 160741831 160741832 160741833 160741834 160742180 160742181 160742182 160742183 160742184 160742185 160742186 160742187 160742188 160742198 160742199 160839175 160839176 160839177 160839178 160839179 160839180 160839181 160839182 160839183 160839184 160839185 160839186 160839187 160839188 160839189 160839190 160839191 160839192 160839193 160839194 160839195 160839230 160839231 160839232 160839233 160839234 160839249 160839300 160839301 160839302 160839303 160839304 160839305 160839306 160839307 160839308 160839309 160839310 160839311 160839312 160839313 160839314 160839315 160839316 160839317 160839318 160839319 160839320 160839321 160839322 160839323 160839324 160839350 160839351 160839352 160839353 160839354 160839355 160839356 160839357 160839358 160839359 160839360 160839361 160839362 160839363 160839364 160839365 160839366 160839367 160839368 160839369 160839443 160839444 160839445 160839450 160839451 160839452 160839453 160839454 160839455 160839456 160839457 160839458 160839459 160839474 160839475 160839476 160839477 160839478 160839479 160839487 160839488 160839489 160839490 160839491 160839496 160839497 160839498 160839499 160839510 160839511 160839512 160839513 160839514 160839515 160839516 160839517 160839518 160839519 160839520 160839521 160839522 160839523 160839524 160839525 160839526 160839534 160839535 160839536 160839537 160839538 160839539 160839540 160839541 160839542 160839543 160839544 160839546 160839547 160839549 160839638 160839639 160839640 160839641 160839642 160839643 160839644 160839645 160839646 160839930 160839931 160839932 160839933 160839934 160839935 160839936 160839937 160839938 160839939 160839940 160839941 160839942 160839943 160840000 160840001 160840002 160840003 160840004 160840005 160840006 160840007 160840008 160840009 160840010 160840041 160840042 160840043 160840044 160840045 160840047 160840048 160840049 160840050 160840051 160840052 160840053 160840054 160840055 160840056 160840057 160840058 160840059 160840060 160840061 160840062 160840063 160840064 160840065 160840066 160840067 160840068 160840069 160840070 160840071 160840072 160840073 160840074 160840097 160840098 160840100 160840101 160840102 160840103 160840104 160840105 160840106 160840107 160840108 160840109 160840110 160840111 160840112 160840113 160840114 160840115 160840116 160840122 160840123 160840124 160840125 160840126 160840127 160840128 160840129 160840130 160840131 160840132 160840133 160840134 160840135 160840136 160840137 160840180 160840181 160840182 160840183 160840184 160840185 160840186 160840187 160840188 160840189 160840190 160840191 160840192 160840193 160840194 160840195 160840196 160840467 160840468 160840469 160840470 160840471 160840472 160840473 160840474 160840475 160840476 160840477 160840478 160840479 160840544 160840545 160840546 160840547 160840548 160840586 160840587 160840588 160840589 160840590 160840591 160840592 160840593 160840594 160840595 160840596 160840599 160840850 160840851 160840852 160840853 160840854 160840855 160840856 160840857 160840858 160840859 160840860 160840861 160840862 160840863 160840864 160840865 160840866 160840867 160840868 160840869 160840870 160840871 160840872 160840873 160840874 160840936 160840937 160840938 160840939 160840940 160840941 160840942 160840943 160840944 160840945 160840946 160840949 160840980 160840981 160840982 160840983 160840984 160840985 160840986 160840987 160840988 160840989 160840990 160840991 160840992 160840993 160840994 160840995 160840996 160840997 160840998 160840999 160841030 160841031 160841032 160841033 160841034 160841035 160841036 160841037 160841038 160841039 160841040 160841041 160841042 160841097 160841098 160841100 160841101 160841102 160841103 160841104 160841105 160841106 160841107 160841108 160841109 160841110 160841111 160841112 160841113 160841114 160841115 160841116 160841117 160841118 160841119 160841120 160841121 160841122 160841123 160841124 160841125 160841126 160841127 160841128 160841129 160841138 160841139 160841140 160841141 160841149 160841210 160841211 160841212 160841213 160841214 160841215 160841216 160841217 160841218 160841219 160841220 160841221 160841222 160841223 160841224 160841225 160841226 160841227 160841228 160841229 160841230 160841231 160841232 160841233 160841234 160841235 160841244 160841245 160841246 160841247 160841248 160841249 160841600 160841601 160841602 160841603 160841604 160841605 160841606 160841607 160841608 160841609 160841610 160841650 160841651 160841652 160841653 160841654 160841655 160841656 160841657 160841658 160841659 160841660 160841661 160841662 160841663 160841664 160841665 160841666 160841667 160841668 160841669 160841710 160841711 160841712 160841713 160841714 160841715 160841716 160841717 160841718 160841719 160841720 160841721 160841722 160841723 160841724 160841725 160841726 160841727 160841728 160841729 160841730 160841731 160841732 160841733 160841734 160939183 160939184 160939185 160939186 160939187 160939188 160939189 160939190 160939191 160939192 160939193 160939194 160939195 160939196 160939197 160939198 160939199 160939200 160939201 160939202 160939203 160939204 160939205 160939206 160939207 160939208 160939209 160939210 160939211 160939212 160939213 160939214 160939215 160939216 160939250 160939251 160939252 160939253 160939254 160939255 160939256 160939257 160939258 160939259 160939260 160939261 160939262 160939263 160939264 160939265 160939266 160939267 160939268 160939269 160939270 160939271 160939272 160939273 160939274 160939275 160939276 160939277 160939278 160939279 160939289 160939290 160939291 160939292 160939293 160939294 160939295 160939296 160939297 160939298 160939299 160939300 160939301 160939302 160939303 160939304 160939305 160939306 160939335 160939336 160939337 160939338 160939339 160939340 160939341 160939342 160939343 160939344 160939345 160939346 160939347 160939348 160939349 160939490 160939491 160939492 160939493 160939494 160939495 160939496 160939497 160939498 160939650 160939695 160940006 160940007 160940008 160940009 160940010 160940011 160940012 160940013 160940014 160940015 160940016 160940017 160940018 160940019 160940041 160940042 160940043 160940044 160940045 160940046 160940047 160940048 160940049 160940299 160940448 160940449 160940500 160940501 160940502 160940503 160940504 160940505 160940506 160940508 160940509 160940510 160940511 160940512 160940513 160940514 160940515 160940516 160940517 160940535 160940536 160940537 160940538 160940539 160940540 160940563 160940564 160940565 160940566 160940567 160940568 160940569 160940570 160940571 160940572 160940573 160940574 160940575 160940577 160940578 160940579 160940580 160940581 160940582 160940583 160940584 160940585 160940586 160940587 160940588 160940589 160940590 160940591 160940592 160940593 160940594 160940595 160940596 160940597 160940598 160940599 160940600 160940601 160940602 160940603 160940604 160940800 160940801 160940802 160940803 160940804 160940809 160940810 160940811 160940812 160940813 160940814 160940815 160940816 160940817 160940818 160940819 160940820 160940821 160940822 160940823 160940824 160940825 160940826 160940827 160940828 160941950 160941951 160941952 160941953 160941954 160941955 160941956 160941957 160941958 160941959 160941960 160941961 160941962 160941963 160941964 160941965 160941966 161039025 161039026 161039027 161039028 161039029 161039030 161039031 161039032 161039033 161039034 161039035 161039036 161039037 161039038 161039039 161039040 161039041 161039042 161039043 161039044 161039045 161039046 161039047 161039048 161039049 161039050 161039051 161039052 161039053 161039054 161039084 161039085 161039086 161039087 161039088 161039089 161039090 161039091 161039092 161039093 161039094 161039095 161039096 161039097 161039098 161039200 161039201 161039202 161039203 161039204 161039205 161039206 161039207 161039208 161039209 161039210 161039211 161039212 161039213 161039214 161039215 161039216 161039217 161039218 161039219 161039232 161039233 161039234 161039235 161039236 161039237 161039238 161039239 161039240 161039241 161039242 161039243 161039244 161039280 161039281 161039282 161039283 161039284 161039285 161039286 161039287 161039288 161039289 161039290 161039363 161039364 161039365 161039366 161039367 161039368 161039369 161039370 161039371 161039372 161039373 161039374 161039375 161039376 161039377 161039378 161039379 161039380 161039381 161039382 161039383 161039384 161039395 161039396 161039397 161039398 161039399 161039500 161039501 161039502 161039503 161039504 161039505 161039506 161039507 161039508 161039509 161039510 161039511 161039512 161039513 161039514 161039515 161039516 161039517 161039518 161039519 161039520 161039626 161039627 161039628 161039629 161039630 161039631 161039632 161039633 161039634 161039635 161039636 161039637 161039638 161039639 161039640 161039641 161039642 161039643 161039644 161039645 161039646 161039647 161039648 161039649 161039825 161039826 161039827 161039828 161039829 161039830 161039831 161039832 161039833 161039834 161039835 161039836 161039837 161039838 161039839 161039840 161039841 161039842 161039843 161039844 161040100 161040101 161040102 161040103 161040104 161040105 161040106 161040107 161040108 161040109 161040110 161040111 161040112 161040113 161040114 161040115 161040116 161040117 161040118 161040119 161040120 161040121 161040122 161040123 161040124 161040125 161040146 161040147 161040148 161040149 161040150 161040151 161040152 161040153 161040154 161040155 161040178 161040179 161040180 161040181 161040182 161040183 161040184 161040185 161040186 161040187 161040188 161040189 161040190 161040191 161040192 161040193 161040194 161040195 161040196 161040197 161040198 161040199 161040249 161040448 161040463 161040464 161040465 161040466 161040467 161040468 161040469 161040470 161040471 161040472 161040473 161040474 161040475 161040476 161040477 161040478 161040479 161040480 161040486 161040487 161040488 161040489 161040490 161040491 161040492 161040493 161040540 161040541 161040542 161040543 161040544 161040545 161040546 161040547 161040548 161040549 161040555 161040556 161040557 161040558 161040559 161040560 161040561 161040562 161040563 161040564 161040565 161040566 161040567 161040568 161040569 161040570 161040571 161040572 161040573 161040574 161040575 161040576 161040577 161040578 161040579 161040580 161040581 161040595 161040596 161040597 161040598 161040599 161040600 161040601 161040602 161040603 161040604 161040605 161040606 161040607 161040608 161040682 161040683 161040684 161040685 161040686 161040687 161040688 161040689 161040690 161040691 161040692 161040693 161040694 161040695 161040696 161040697 161040698 161040699 161040831 161040832 161040833 161040834 161040835 161040836 161040837 161040838 161040839 161040840 161040841 161040842 161040843 161040844 161041300 161041301 161041302 161041303 161041304 161041305 161041306 161041307 161041308 161041309 161041310 161041311 161041312 161041313 161041314 161041315 161041316 161041317 161041318 161041319 161041320 161041321 161041322 161041323 161041324 161041325 161041326 161041327 161041328 161041329 161041330 161041331 161041332 161041333 161041334 161041335 161041336 161041337 161041338 161041339 161041340 161041341 161041342 161041343 161041344 161041345 161041346 161041347 161041348 161041349 161041360 161041361 161041362 161041363 161041364 161041365 161041366 161041367 161041368 161041369 161041370 161041371 161041372 161041373 161041374 161041375 161041376 161041377 161041378 161041379 161041380 161041381 161041382 161041383 161041384 161041385 161041386 161041395 161041396 161041397 161041398 161041875 161041876 161041877 161041878 161041879 161041880 161041881 161041882 161041883 161041884 161041885 161041886 161041942 161041943 161041944 161041945 161041946 161041947 161041948 161041949 161139357 161139358 161139391 161139392 161139393 161139394 161139395 161139396 161139397 161139819 161139920 161139921 161139922 161139923 161139924 161139925 161139959 161139960 161139961 161139962 161139963 161139964 161139965 161139966 161139967 161139968 161139997 161140000 161140001 161140002 161140003 161140004 161140005 161140006 161140007 161140008 161140009 161140010 161140034 161140035 161140036 161140037 161140038 161140039 161140100 161140101 161140102 161140103 161140104 161140105 161140106 161140107 161140108 161140109 161140110 161140111 161140112 161140113 161140114 161140115 161140116 161140117 161140118 161140119 161140120 161140121 161140122 161140123 161140124 161140125 161140126 161140127 161140128 161140129 161140130 161140131 161140132 161140133 161140134 161140135 161140136 161140157 161140158 161140159 161140160 161140161 161140162 161140163 161140164 161140165 161140166 161140167 161140168 161140169 161140190 161140191 161140192 161140193 161140194 161140195 161140196 161140197 161140198 161140199 161140300 161140301 161140302 161140303 161140304 161140305 161140306 161140307 161140308 161140309 161140310 161140311 161140546 161140547 161140548 161140570 161140571 161140572 161140573 161140574 161140575 161140576 161140577 161140578 161140579 161140580 161140581 161140582 161140583 161140584 161140585 161140586 161140587 161140588 161140589 161140590 161140591 161140592 161140593 161140594 161140595 161140596 161140597 161140598 161140599 161140640 161140641 161140642 161140643 161140644 161140645 161140646 161140647 161140648 161140649 161140700 161140701 161140702 161140703 161140704 161140705 161140706 161140707 161140708 161140709 161140710 161140711 161140712 161140713 161140714 161140715 161140716 161140717 161140718 161140719 161140720 161140721 161140722 161140723 161140724 161140725 161140726 161140727 161140728 161140729 161140730 161140731 161140732 161140733 161140734 161140735 161140736 161140737 161140738 161140739 161140740 161140741 161140746 161140747 161140748 161140749 161140797 161140798 161140799 161140850 161140851 161140852 161140853 161140854 161140855 161140856 161140857 161140858 161140859 161140860 161140861 161140862 161140863 161140864 161140865 161140990 161140991 161140992 161141029 161141030 161141031 161141032 161141033 161141034 161141035 161141036 161141037 161141038 161141039 161141040 161141041 161141042 161141043 161141044 161141045 161141046 161141047 161141048 161141070 161141071 161141072 161141073 161141074 161141075 161141076 161141077 161141078 161141079 161141080 161141081 161141082 161141083 161141084 161141085 161141086 161141087 161141088 161141090 161141091 161141092 161141093 161141094 161141095 161141096 161141097 161141098 161141099 161141100 161141126 161141127 161141128 161141129 161141130 161141131 161141132 161141133 161141134 161141135 161141136 161141137 161141138 161141139 161141140 161141141 161141142 161141143 161141144 161141145 161141146 161141147 161141148 161141149 161141330 161141331 161141332 161141333 161141334 161141335 161141336 161141337 161141338 161141339 161141340 161141341 161141342 161141343 161141344 161141345 161141346 161141347 161141348 161141349 161141500 161141501 161141502 161141503 161141504 161141505 161141506 161141507 161141508 161141509 161141510 161141511 161141512 161141513 161141514 161141515 161141516 161141517 161141518 161141519 161141520 161141521 161141522 161141523 161141524 161141590 161141591 161141592 161141593 161141594 161141595 161141596 161141597 161141598 161141599 161141635 161141636 161141637 161141638 161141639 161141640 161141641 161141642 161141643 161141644 161141645 161141646 161141647 161141648 161141649 161141663 161141664 161141665 161141666 161141667 161141668 161141669 161141670 161141750 161141751 161141752 161141753 161141754 161141755 161141756 161141774 161141775 161141776 161141777 161141778 161141779 161141780 161141781 161141782 161141783 161141784 161141785 161141786 161141787 161141788 161141789 161141790 161141791 161141792 161141793 161141798 161141799 161141805 161141806 161141807 161141808 161141809 161141810 161141811 161141812 161141813 161141814 161141815 161141816 161141817 161141818 161141819 161141820 161141821 161141822 161141823 161141824 161141825 161141826 161141827 161141828 161141829 161141830 161141831 161141832 161141833 161141834 161141835 161141836 161141837 161141838 161141839 161141840 161141841 161141842 161141843 161141844 161141845 161141846 161141847 161141848 161141849 161141850 161141851 161141852 161141853 161141854 161141855 161141886 161141887 161141888 161141889 161141890 161141891 161141894 161141895 161141896 161141897 161141898 161141899 161142165 161142166 161142167 161142168 161142169 161142170 161142171 161142172 161142173 161142174 161142175 161142176 161142177 161142178 161142179 161142180 161142181 161142182 161142183 161142184 161142185 161142186 161142187 161142188 161142189 161142190 161142191 161142193 161239050 161239051 161239052 161239053 161239054 161239055 161239056 161239057 161239058 161239059 161239060 161239061 161239062 161239063 161239064 161239065 161239066 161239067 161239068 161239069 161239070 161239071 161239072 161239073 161239074 161239075 161239076 161239077 161239078 161239079 161239080 161239081 161239082 161239083 161239084 161239085 161239086 161239087 161239088 161239089 161239090 161239136 161239137 161239138 161239139 161239140 161239141 161239142 161239143 161239144 161239145 161239146 161239147 161239148 161239149 161239390 161239391 161239392 161239393 161239394 161239395 161239396 161239397 161239398 161239516 161239517 161239518 161239519 161239520 161239521 161239522 161239523 161239524 161239525 161239526 161239527 161239528 161239529 161239530 161239531 161239532 161239533 161239534 161239535 161239536 161239537 161239538 161239539 161239540 161239541 161239542 161239543 161239544 161239545 161239546 161239547 161239548 161239549 161239550 161239551 161239552 161239553 161239599 161239700 161239701 161239702 161239703 161239704 161239705 161239706 161239707 161239708 161239709 161239710 161239711 161239712 161239713 161239714 161239715 161239716 161239717 161239718 161239719 161239720 161239721 161239722 161239723 161239724 161239725 161239726 161239727 161239728 161239845 161239846 161239847 161239848 161239849 161239850 161239851 161239852 161239853 161239854 161239855 161239856 161239857 161239858 161239859 161239860 161239861 161239862 161239863 161239864 161239865 161239866 161239867 161239868 161239869 161239870 161239871 161239872 161239873 161239874 161239875 161239876 161239877 161239878 161239879 161239880 161239881 161239882 161239883 161239884 161239885 161239886 161239887 161239888 161239889 161239890 161239891 161239892 161239893 161239894 161239895 161239896 161239897 161239898 161239899 161239920 161239921 161239922 161239923 161239924 161239925 161239926 161239927 161239928 161239929 161239945 161239946 161239947 161239948 161239949 161240240 161240241 161240242 161240243 161240244 161240245
161240246 161240247 161240248 161240250 161240251 161240252 161240253 161240254 161240255 161240256 161240257 161240258 161240259 161240260 161240261 161240262 161240263 161240264 161240265 161240266 161240267 161240268 161240269 161240270 161240271 161240272 161240295 161240296 161240297 161240298 161240299 161240350 161240351 161240352 161240353 161240354 161240355 161240356 161240357 161240358 161240359 161240360 161240361 161240362 161240363 161240364 161240365 161240366 161240367 161240368 161240369 161240370 161240399 161240448 161240449 161240536 161240537 161240538 161240539 161240540 161240541 161240542 161240543 161240544 161240545 161240750 161240751 161240752 161240753 161240754 161240755 161240756 161240757 161240758 161240759 161240760 161240761 161240762 161240763 161240764 161240765 161240766 161240767 161240768 161240769 161240770 161240771 161240772 161240773 161240774 161240775 161240776 161240777 161240778 161240779 161241200 161241201 161241202 161241203 161241204 161241205 161241326 161241327 161241328 161241329 161241330 161241331 161241332 161241333 161241334 160739842 |
| FEI Number |
1831750
|
Recalling Firm/ Manufacturer |
Stryker Medical Division of Stryker Corporation
3800 E Centre Ave
Portage
MI
49002-5826
|
| For Additional Information Contact | Victoria Haney 269-389-8306 |
Manufacturer Reason for Recall | The power load floor plates which hold the ambulance cot anchors in place during patient transport, may be broken. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On February 27, 2018, affected customers received an Urgent Medical Device Correction Notification via FedEx notifying them of the issue and potential hazard. For distributed units, the four Floor Plate Attachment Brackets that hold the anchor in place will be replaced in the field by Strykers field service personnel. Customers will be asked to return a business reply form to acknowledge that they have received and understand the notification. For further questions, please call (269) 389-8306. |
| Quantity in Commerce | 4724 |
| Distribution | Worldwide Distribution - USA (nationwide) and to the countries of : Australia, Canada, France, Germany, Hong Kong, Italy, India, Netherlands, Portugal, Saint Martin, Switzerland, United Kingdom |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = FPO
|
|
|
|