| | Class 1 Device Recall Quantum Pump Console |  |
| Date Initiated by Firm | May 29, 2019 |
| Date Posted | August 23, 2019 |
| Recall Status1 |
Terminated 3 on August 21, 2020 |
| Recall Number | Z-2238-2019 |
| Recall Event ID |
83352 |
| 510(K)Number | K173834 |
| Product Classification |
Console, heart-lung machine, cardiopulmonary bypass - Product Code DTQ
|
| Product | Quantum Pump Console |
| Code Information |
Software version 6.2.1.9; Serial Numbers PH1000005, PH1000020, PH1000027, PH1000030, PH1000077, PH1000081, PH1000082, PH1000086, PH1000087, PH1000093, PH1000100, PH1000101, PH1000102, PH1000103, PH1000109, PH1000110, PH1000112, PH1000113, PH1000114, PH1000115, PH1000120, PH1000127, PH1000129, PH1000131, PH1000132, PH1000134, PH1000135, PH1000137, PH1000139, PH1000140, PH1000141, PH1000144, PH1000145, PH1000150, PH1000156, PH1000158, PH1000159, PH1000160, PH1000161, PH1000162, PH1000165, PH1000168, PH1000170, PH1000171, PH1000172, PH1000175, PH1000176, PH1000177, PH1000178, PH1000179, PH1000180, PH1000182, PH1000201, PH1000208, PH1000209, PH1000210, PH1000211, PH1000212, PH1000213, PH1000215, PH1000233, PH1000256, PH1000257, PH1000259, PH1000438, PH1000439, PH1000440, PH1000441, PH1000442, PH1000445, PH1000455, PH1000495, PH1000498, PH1000522, PH1000615, PH1000644, PH1000645, PH1000646, PH1000647, PH1000648, PH1000649, PH1000668, PH1000669, PH1000670, PH1000682, PH1000683, PH1000684, PH1000686, PH1000687, PH1000688, PH1000689, PH1000692, PH1000722, PH1000726, PH1000727, PH1000740, PH1000742, PH1000743, PH1000744, PH1000879, PH1000880, PH1000882, PH1000885, PH1000886, PH1000895, PH1000897, PH1000939, PH1000940 |
| FEI Number |
3006073153
|
Recalling Firm/ Manufacturer |
Spectrum Medical Ltd. Harrier 4, Meteor Business Park Cheltenham Road East Gloucester United Kingdom
|
Manufacturer Reason for Recall | Potential for Quantum Pump Console, part of the Quantum Perfusion Systems, to unexpectedly shut down while in use. |
FDA Determined Cause 2 | Software design |
| Action | Spectrum Medical notified customers on about 05/29/2019 via "URGENT FIELD SAFETY CORRECTION" letter. Customers were informed that Spectrum Medical is in the process of installing the most recent software update v6.2.3.16 to the Quantum Perfusion Systems, (Quantum Pump Console). This software update corrects an error that could potentially lead to power loss. All Quantum Perfusion Systems are in scope of this issue.
Customers were instructed to be aware that user intervention may be necessary with continued use before the software update, assure that all users receive notice of this issue, confirm receipt of the Urgent Medical Device Correction by returning the customer response form, and a Spectrum Medical service team will contact each customer and arrange a time and date to update the systems with the most recent software update. Additionally, customers were instructed to notify all relevant personnel and to customers if the devices have been further distributed. |
| Quantity in Commerce | 108 units (8 US, 100 OUS) |
| Distribution | US distribution to NY, OH, VA, WA.
International distribution to the Netherlands. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DTQ
|
|
|
|