Class 2 Device Recall Disposable Aortic Punch

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Disposable Aortic Punchsee related information
Date Initiated by FirmJanuary 13, 2020
Date PostedJanuary 27, 2020
Recall Status1 Terminated 3 on November 24, 2021
Recall NumberZ-1270-2020
Recall Event ID 84722
Product Classification Instruments, Surgical, Cardiovascular - Product Code DWS
ProductTeleflex Medical Disposable Aortic Punch, in the following sizes: a) 2.8 mm, REF DP-28K b) 3.6 mm, REF DP-36K c) 4.0 mm, REF DP-40K d) 4.4 mm, REF DP-44K e) 4.8 mm, REF DP-48K f) 5.2 mm, REF DP-52K g) 5.6 mm, REF DP-56K h) 6.0 mm, REF DP-60K i) 2.8 mm, REF MDP-28K j) 3.6 mm, REF MDP-36K k) 4.0 mm, REF MDP-40K l) 4.4 mm, REF MDP-44K m) 4.8 mm, REF MDP-48K n) 5.2 mm, REF MDP-52K o) 5.6 mm, REF MDP-56K p) 6.0 mm, REF MDP-60K
Code Information a) Teleflex Medical Disposable Aortic Punch, 2.8 mm, REF DP-28K Lot Numbers: 74A1501140, 74B1602364, 74E1700597, 74G1602599, 74H1900357, 74L1501549, 74A1601253, 74B1900832, 74E1800886, 74G1700636, 74J1501218, 74L1501977, 74A1601254, 74C1501113, 74E1802643, 74G1700637, 74J1501754, 74L1501978, 74A1602903, 74C1600966, 74E1901667, 74G1700718, 74J1501755, 74L1600387, 74A1602904, 74C1600972, 74F1502019, 74G1900066, 74J1601180, 74L1600389, 74A1603168, 74C1600976, 74F1600350, 74G1900340, 74J1603181, 74L1602761, 74A1603170, 74C1600977, 74F1602476, 74G1902629, 74J1800844, 74L1700357, 74A1701154, 74C1702793, 74F1602477, 74G1903756, 74K1500475, 74L1800444, 74A1802567, 74C1702794, 74F1702731, 74H1502508, 74K1500476, 74L1803409, 74A1903113, 74C1800357, 74F1800411, 74H1502510, 74K1501684, 74L1901787, 74A1903146, 74D1600123, 74F1802291, 74H1601384, 74K1501685, 74M1601318, 74B1501366, 74D1601266, 74F1802292, 74H1602748, 74K1900124, 74M1602080, 74B1501710, 74D1601524, 74G1500954, 74H1700589, 74K1900125, 74M1800915, 74B1600576, 74D1602673, 74G1500955, 74H1700590, 74K1900133, 74M1800916, 74B1600580, 74D1802317, 74G1503008, 74H1701427, 74K1901814, 74M1801620, 74B1601613, 74E1600837, 74G1503009, 74H1701863, 74L1501239, 74M1900014, 74B1601616, 74E1700596, 74G1602598, 74H1800764, 74L1501548  b) Teleflex Medical Disposable Aortic Punch, 3.6 mm, REF DP-36K Lot Numbers: 74A1601251, 74C1501114, 74E1600361, 74F1901768, 74H1602749, 74K1901790, 74A1602049, 74C1501504, 74E1600362, 74G1501254, 74H1802543, 74L1501933, 74A1602205, 74C1600454, 74E1700383, 74G1501584, 74H1802910, 74L1501934, 74A1603169, 74C1600978, 74E1700598, 74G1502240, 74H1900354, 74L1502651, 74A1801612, 74C1702795, 74E1800887, 74G1502247, 74H1901509, 74L1502668, 74A1901239, 74C1702796, 74E1802644, 74G1601150, 74J1501761, 74L1602762, 74A1902997, 74C1800358, 74E1900552, 74G1601151, 74J1501762, 74L1702282, 74A1903148, 74D1503157, 74E1901665, 74G1700719, 74J1702333, 74L1702978, 74B1501361, 74D1600124, 74F1500725, 74G1702124, 74J1702528, 74L1800993, 74B1600581, 74D1600125, 74F1502017, 74G1702125, 74J1703089, 74L1803139, 74B1601617, 74D1802318, 74F1502713, 74G1802091, 74J1900939, 74L1803408, 74B1602365, 74D1900834, 74F1502714, 74G1902121, 74J1901700, 74M1601319, 74B1602372, 74D1901907, 74F1600351, 74H1500734, 74K1500480, 74M1702625, 74B1800721, 74E1500848, 74F1703304, 74H1500735, 74K1501130, 74M1801251, 74B1800722, 74E1501779, 74F1801505, 74H1502517, 74K1501131, 74M1801622, 74B1902990, 74E1502618, 74F1802293, 74H1600757, 74K1502964, 74M1900012, 74B1903028, 74E1503095, 74F1900940, 74H1601385, 74K1600559  c) Teleflex Medical Disposable Aortic Punch, 4.0 mm, REF DP-40K  Lot Numbers: 74A1501146, 74A1602181, 74A1702660, 74A1901901, 74B1602046, 74C1503228, 74A1501148, 74A1602360, 74A1800595, 74B1501126, 74B1800506, 74C1503232, 74A1502433, 74A1602361, 74A1800596, 74B1501358, 74B1800723, 74C1600455, 74A1502434, 74A1603161, 74A1801613, 74B1502260, 74B1801756, 74C1600460, 74A1502851, 74A1603162, 74A1802148, 74B1502261, 74B1802580, 74C1600967, 74A1502852, 74A1603163, 74A1802150, 74B1503151, 74C1501122, 74C1600979, 74A1503453, 74A1700378, 74A1802568, 74B1503152, 74C1501123, 74C1600980, 74A1601248, 74A1700766, 74A1900243, 74B1600577, 74C1501505, 74C1702893, 74A1601249, 74A1700767, 74A1900244, 74B1600582, 74C1502259, 74C1702894, 74A1601250, 74A1702659, 74A1901240, 74B1601618, 74C1502260, 74C1800359, 74C1800360, 74E1503093, 74F1601983, 74G1900067, 74H1903474, 74K1602893, 74C1801150, 74E1503099, 74F1602478, 74G1902118, 74J1500711, 74K1602894, 74C1900330, 74E1600838, 74F1602479, 74G1902119, 74J1500712, 74K1700910, 74C1900332, 74E1600839, 74F1603239, 74G1902833, 74J1501214, 74K1701639, 74C1900811, 74E1601257, 74F1603240, 74G1903729, 74J1501215, 74K1702121, 74C1900812, 74E1602244, 74F1700828, 74H1500732, 74J1501766, 74K1702122, 74C1901597, 74E1602245, 74F1701130, 74H1501603, 74J1501767, 74K1702866, 74C1902227, 74E1602810, 74F1701131, 74H1502506, 74J1502404, 74K1802890, 74C1902228, 74E1602811, 74F1702095, 74H1502507, 74J1502405, 74K1900758, 74D1500077, 74E1700599, 74F1702096, 74H1502530, 74J1600424, 74K1900759, 74D1500078, 74E1700600, 74F1702733, 74H1600758, 74J1600425, 74K1901792, 74D1501647, 74E1701203, 74F1702734, 74H1600759, 74J1601181, 74K1902250, 74D1501648, 74E1701234, 74F1703305, 74H1601386, 74J1601737, 74L1500801, 74D1503172, 74E1701236, 74F1800418, 74H1602468, 74J1602579, 74L1500810, 74D1600213, 74E1701755, 74F1801506, 74H1602469, 74J1602677, 74L1501230, 74D1600214, 74E1701758, 74F1802294, 74H1602750, 74J1602678, 74L1501231, 74D1600927, 74E1703220, 74F1802710, 74H1602751, 74J1700737, 74L1501983, 74D1601538, 74E1703236, 74F1802711, 74H1700591, 74J1700738, 74L1502663, 74D1602013, 74E1801602, 74F1900935, 74H1700592, 74J1701829, 74L1600400, 74D1602014, 74E1801603, 74F1900936, 74H1701429, 74J1702529, 74L1600440, 74D1602676, 74E1802105, 74F1900937, 74H1701431, 74J1702530, 74L1601327, 74D1700564, 74E1802106, 74F1901099, 74H1701839, 74J1703090, 74L1601328, 74D1700954, 74E1802645, 74F1901117, 74H1701869, 74J1703091, 74L1602113, 74D1700955, 74E1802984, 74F1901761, 74H1702709, 74J1800845, 74L1602763, 74D1701179, 74E1900554, 74G1500951, 74H1702710, 74J1801321, 74L1602764, 74D1702440, 74E1900593, 74G1501744, 74H1703164, 74J1801322, 74L1700358, 74D1801512, 74E1901447, 74G1502237, 74H1703165, 74J1802104, 74L1700359, 74D1802103, 74E1901448, 74G1503006, 74H1800761, 74J1802105, 74L1701122, 74D1802104, 74E1901663, 74G1600720, 74H1800763, 74J1802736, 74L1701124, 74D1802319, 74E1901664, 74G1601769, 74H1800838, 74J1900237, 74L1701786, 74D1802693, 74E1902715, 74G1602120, 74H1801707, 74J1900956, 74L1702283, 74D1802694, 74E1903772, 74G1602121, 74H1802544, 74J1901697, 74L1702979, 74D1900484, 74F1500753, 74G1602600, 74H1802911, 74J1901698, 74L1800558, 74D1900485, 74F1500754, 74G1700638, 74H1802912, 74J1903043, 74L1800559, 74D1900657, 74F1501151, 74G1700639, 74H1900352, 74K1500471, 74L1800994, 74D1900659, 74F1501152, 74G1701299, 74H1900353, 74K1500472, 74L1800995, 74D1901505, 74F1502021, 74G1701314, 74H1901108, 74K1501121, 74L1801431, 74D1902167, 74F1502022, 74G1702119, 74H1901109, 74K1501122, 74L1803140, 74D1902810, 74F1502217, 74G1702120, 74H1901115, 74K1501680, 74L1900199, 74D1902811, 74F1502710, 74G1800767, 74H1901116, 74K1502261, 74L1900200, 74D1902812, 74F1502711, 74G1800768, 74H1901510, 74K1502262, 74L1900201, 74E1501782, 74F1600352, 74G1801515, 74H1902119, 74K1502270, 74L1900202, 74E1501783, 74F1601101, 74G1802092, 74H1902737, 74K1600560, 74L1900361, 74E1502616, 74F1601102, 74G1802093, 74H1902738, 74K1601326, 74L1900362, 74E1502617, 74F1601325, 74G1802678, 74H1903398, 74K1602034, 74L1900858, 74L1900859, 74M1500805, 74M1601314, 74M1701819, 74M1801624, 74M1900367, 74L1901403, 74M1600393, 74M1602082, 74M1701820, 74M1900010, 74M1901094, 74L1901404, 74M1600394, 74M1701016, 74M1801252, 74M1900017, 74M1901095, 74L1901885, 74M1601313, 74M1701017, 74M1801253, 74M1900366    d) Teleflex Medical Disposable Aortic Punch, 4.4 mm, REF DP-44K Lot Numbers: 74A1501153, 74D1600217, 74E1900553, 74H1502520, 74J1700739, 74L1700360, 74A1502848, 74D1601525, 74E1903148, 74H1502909, 74J1701830, 74L1701125, 74A1602180, 74D1602016, 74E1903763, 74H1600761, 74J1701831, 74L1701787, 74A1800598, 74D1602268, 74F1600354, 74H1600762, 74J1702532, 74L1701788, 74A1801615, 74D1602679, 74F1600355, 74H1601224, 74J1703093, 74L1702285, 74A1801616, 74D1700951, 74F1601104, 74H1601387, 74J1800846, 74L1702981, 74A1802152, 74D1701786, 74F1601140, 74H1602470, 74J1801323, 74L1702982, 74A1802569, 74D1702441, 74F1601984, 74H1602471, 74J1802106, 74L1800560, 74A1900362, 74D1802105, 74F1601985, 74H1700594, 74J1900625, 74L1800561, 74A1901242, 74D1802106, 74F1603241, 74H1700595, 74J1901696, 74L1800997, 74A1901899, 74D1900483, 74F1603242, 74H1701435, 74J1903042, 74L1800998, 74A1902221, 74D1900658, 74F1702097, 74H1701864, 74K1500483, 74L1801771, 74A1902222, 74D1901506, 74F1702735, 74H1701865, 74K1501132, 74L1803136, 74A1902354, 74D1902168, 74F1800868, 74H1702712, 74K1501688, 74L1803261, 74B1501363, 74E1500841, 74F1801508, 74H1702713, 74K1600561, 74L1901191, 74B1502275, 74E1500842, 74F1802712, 74H1703166, 74K1601328, 74L1901200, 74B1601620, 74E1501992, 74F1900938, 74H1703282, 74K1601329, 74L1901201, 74B1602048, 74E1503098, 74G1502238, 74H1800765, 74K1700912, 74L1901407, 74B1602049, 74E1600364, 74G1503014, 74H1800840, 74K1700913, 74M1600396, 74B1800508, 74E1601723, 74G1600917, 74H1800841, 74K1701641, 74M1601316, 74B1800725, 74E1601724, 74G1601771, 74H1801709, 74K1701642, 74M1601317, 74B1800726, 74E1602248, 74G1601772, 74H1801710, 74K1702868, 74M1602089, 74B1801757, 74E1602812, 74G1602122, 74H1802546, 74K1702869, 74M1602090, 74B1802581, 74E1700602, 74G1602123, 74H1802914, 74K1900131, 74M1702628, 74B1802582, 74E1701406, 74G1602673, 74H1900355, 74K1901789, 74M1702629, 74C1501117, 74E1702791, 74G1602718, 74H1901113, 74K1901813, 74M1900011, 74C1501506, 74E1702792, 74G1700640, 74H1901511, 74L1501242, 74M1900372, 74C1600973, 74E1702796, 74G1701317, 74H1902120, 74L1502469, 74M1900689, 74C1600975, 74E1702797, 74G1701318, 74H1903403, 74L1600402, 74M1901097, 74C1702792, 74E1800889, 74G1801516, 74J1600426, 74L1600403, 74C1902226, 74E1801605, 74G1802681, 74J1600427, 74L1601647, 74D1500076, 74E1801606, 74G1900338, 74J1601203, 74L1601648, 74D1501516, 74E1802108, 74G1902123, 74J1602434, 74L1602115, 74D1501645, 74E1802109, 74G1902633, 74J1603182, 74L1602766, 74D1600216, 74E1802987, 74G1903757, 74J1603183, 74L1602767  e) Teleflex Medical Disposable Aortic Punch, 4.8 mm, REF DP-48K Lot Numbers: 74A1502850, 74A1801617, 74B1502262, 74B1802583, 74C1801152, 74D1600931, 74A1602182, 74A1802153, 74B1503158, 74B1901661, 74C1900329, 74D1601537, 74A1602901, 74A1901238, 74B1601609, 74C1501119, 74C1900618, 74D1602017, 74A1603164, 74A1901900, 74B1800505, 74C1600462, 74D1600226, 74D1602018, 74A1700763, 74A1902355, 74B1800719, 74C1600974, 74D1600227, 74D1700939, 74A1800600, 74B1501128, 74B1800727, 74C1600982, 74D1600930, 74D1700943, 74D1702442, 74F1600357, 74G1601774, 74H1701867, 74J1703095, 74K1900760, 74D1702443, 74F1600358, 74G1602124, 74H1701868, 74J1800847, 74K1901336, 74D1802107, 74F1601105, 74G1602719, 74H1702714, 74J1801324, 74L1500803, 74D1802321, 74F1601137, 74G1602720, 74H1800767, 74J1802108, 74L1600443, 74D1901508, 74F1601986, 74G1701319, 74H1800843, 74J1901706, 74L1602116, 74D1902169, 74F1602009, 74G1701320, 74H1800844, 74J1903044, 74L1701123, 74E1600842, 74F1603243, 74G1800770, 74H1802548, 74K1501123, 74L1701789, 74E1602249, 74F1702099, 74G1801517, 74H1802549, 74K1501124, 74L1702286, 74E1602813, 74F1702736, 74G1900337, 74H1802915, 74K1600562, 74L1800443, 74E1701837, 74F1702737, 74G1900339, 74H1900356, 74K1600563, 74L1800999, 74E1701838, 74F1703306, 74G1902120, 74H1901512, 74K1602037, 74L1803265, 74E1801607, 74F1800866, 74G1903728, 74J1501216, 74K1602038, 74M1600397, 74E1801608, 74F1801510, 74G1903758, 74J1501759, 74K1700915, 74M1602092, 74E1802646, 74F1802296, 74H1601389, 74J1600428, 74K1701643, 74M1701020, 74E1802647, 74F1900939, 74H1601390, 74J1601183, 74K1701644, 74M1702630, 74E1902716, 74F1901122, 74H1602472, 74J1601783, 74K1702870, 74M1702631, 74E1902717, 74F1902895, 74H1602752, 74J1602085, 74K1702871, 74M1800913, 74F1502712, 74G1501749, 74H1602753, 74J1700740, 74K1900128, 74M1900368, 74F1503522, 74G1601773, 74H1700596, 74J1701832, 74K1900129, 74M1901096  f) Teleflex Medical Disposable Aortic Punch, 5.2 mm, REF DP-52K Lot Numbers: 74A1601258, 74D1700800, 74F1602482, 74H1600764, 74K1602778, 74M1800914, 74A1700762, 74D1700820, 74F1701133, 74H1701443, 74K1700916, 74A1800602, 74D1701167, 74F1701134, 74H1702715, 74K1702872, 74A1903114, 74D1802322, 74F1800869, 74H1800846, 74K1900132, 74B1501364, 74D1802323, 74F1800870, 74H1801712, 74K1900761, 74B1503155, 74E1600843, 74F1801512, 74H1801713, 74K1901795, 74B1600584, 74E1602250, 74F1901098, 74H1802550, 74K1902045, 74B1602050, 74E1602814, 74G1500952, 74J1601733, 74K1902545, 74B1800510, 74E1700899, 74G1700720, 74J1602581, 74L1501241, 74B1800729, 74E1701840, 74G1900341, 74J1701833, 74L1602118, 74B1802584, 74E1701841, 74G1902122, 74J1802110, 74L1700361, 74B1900830, 74E1800897, 74G1902834, 74J1900239, 74L1800555, 74C1501507, 74E1900596, 74G1903730, 74J1901704, 74L1803267, 74C1902229, 74F1500755, 74G1903755, 74K1501689, 74M1600398, 74D1602269, 74F1601106, 74H1600763, 74K1600564, 74M1601320  g) Teleflex Medical Disposable Aortic Punch, 5.6 mm, REF DP-56K Lot Numbers: 74A1501142, 74B1600585, 74D1501515, 74F1500724, 74H1702716, 74K1902252, 74A1601256, 74B1601614, 74D1502568, 74F1502020, 74J1702533, 74L1501976, 74A1602204, 74B1601622, 74E1500845, 74F1702103, 74K1500481, 74L1502672, 74A1603172, 74B1602367, 74E1502580, 74F1800871, 74K1501683, 74M1900365, 74A1800603, 74B1800720, 74E1602815, 74F1801514, 74K1502970, 74M1900840, 74B1500704, 74C1501508, 74E1701843, 74F1801515, 74K1602779, 74M1901098, 74B1502274, 74C1600968, 74E1701844, 74F1802297, 74K1900126, 74M1901275, 74B1600578, 74D1500075, 74E1800898, 74G1902124, 74K1900127  h) Teleflex Medical Disposable Aortic Punch, 6.0 mm, REF DP-60K Lot Numbers: 74A1500172, 74E1502579, 74M1602094, 74E1800901, 74G1902125, 74D1501639, 74A1602907, 74H1700597, 74E1800902, 74K1901815, 74E1500846, 74E1602816, 74H1702717, 74G1900342    i) Teleflex Medical Disposable Aortic Punch, 2.8 mm, REF MDP-28K Lot Numbers: 74A1502835, 74A1902357, 74E1701845, 74F1902101, 74H1801714, 74A1601252, 74B1501360, 74E1901451, 74F1902891, 74H1802551, 74K1901816, 74A1602906, 74B1601615, 74F1500722, 74F1902892, 74J1702534, 74L1702287, 74A1800642, 74B1602374, 74F1501155, 74H1500733, 74J1901705, 74A1802572, 74B1801755, 74F1800872, 74H1702718, 74K1800082, 74A1901243, 74D1700801, 74F1802298, 74H1702719, 74K1900135  j) Teleflex Medical Disposable Aortic Punch, 3.6 mm, REF MDP-36K Lot Numbers: 74A1602184, 74B1602375, 74F1500984, 74H1601392, 74J1802111, 74L1702288, 74A1602185, 74C1702803, 74F1703309, 74H1701444, 74K1500473, 74L1702289, 74A1802573, 74C1702804, 74F1800873, 74H1702720, 74K1602780, 74L1901405, 74A1901218, 74C1802507, 74F1901968, 74H1801715, 74K1701646, 74L1901411, 74A1902223, 74C1802508, 74F1902108, 74H1802552, 74K1900134, 74M1901099, 74B1501261, 74E1600363, 74G1500956, 74H1802553, 74L1500804, 74B1600586, 74E1701846, 74G1502248, 74H1902743, 74L1500805, 74B1601623, 74E1701847, 74H1502513, 74J1500701, 74L1501981, 74B1602368, 74E1902721, 74H1502514, 74J1702535, 74L1501982  k) Teleflex Medical Disposable Aortic Punch, 4.0 mm, REF MDP-40K Lot Numbers: 74A1503439, 74C1802506, 74F1502715, 74G1602722, 74J1502407, 74K1801695, 74A1601257, 74C1900617, 74F1503526, 74G1800771, 74J1601185, 74L1500807, 74A1602908, 74D1600957, 74F1503527, 74G1801518, 74J1601186, 74L1500808, 74A1603166, 74D1600958, 74F1600359, 74G1802094, 74J1700494, 74L1501935, 74A1700734, 74D1601536, 74F1601332, 74G1902832, 74J1700742, 74L1501936, 74A1701848, 74D1602680, 74F1601333, 74H1500736, 74J1802112, 74L1602768, 74A1802574, 74D1700567, 74F1603244, 74H1502518, 74J1901699, 74L1602769, 74B1501131, 74D1700568, 74F1702104, 74H1502519, 74K1500478, 74L1701126, 74B1501365, 74D1800226, 74F1702105, 74H1602475, 74K1500479, 74L1702983, 74B1503154, 74D1800227, 74F1702738, 74H1602476, 74K1502266, 74L1900759, 74B1600587, 74E1502619, 74F1702739, 74H1701446, 74K1502268, 74L1900860, 74B1601624, 74E1503096, 74F1703310, 74H1702721, 74K1600565, 74M1602095, 74B1602369, 74E1601726, 74F1902081, 74H1902739, 74K1601332, 74M1701824, 74B1802585, 74E1602817, 74F1902177, 74H1902740, 74K1602781, 74M1701825, 74C1600456, 74E1602818, 74F1902178, 74H1902741, 74K1602782, 74M1900369, 74C1600984, 74F1500759, 74G1501993, 74H1903402, 74K1701647, 74M1901093, 74C1600985, 74F1501153, 74G1501994, 74J1501756, 74K1702867, 74C1802505, 74F1502018, 74G1602721, 74J1502406, 74K1800961  l) Teleflex Medical Disposable Aortic Punch, 4.4 mm, REF MDP-44K Lot Numbers: 74A1502836, 74B1902383, 74D1801513, 74F1802299, 74H1702722, 74K1702124, 74A1601255, 74C1503233, 74D1801514, 74F1901118, 74H1801716, 74K1800962, 74A1602179, 74C1702806, 74D1900660, 74G1500953, 74H1900358, 74K1801699, 74A1700760, 74C1801154, 74E1500844, 74G1601152, 74H1901513, 74L1501916, 74A1700761, 74C1900328, 74E1601486, 74G1602723, 74H1902742, 74L1502660, 74A1800643, 74C1901596, 74E1602252, 74G1700721, 74J1603175, 74L1601649, 74A1800644, 74D1601526, 74E1602819, 74G1802095, 74J1603176, 74L1700363, 74A1801620, 74D1601535, 74E1701849, 74H1502515, 74J1700743, 74L1700364, 74A1802575, 74D1602681, 74E1900594, 74H1601394, 74J1703098, 74L1702290, 74B1501359, 74D1700916, 74F1600360, 74H1601395, 74J1800849, 74L1702984, 74B1600575, 74D1800228, 74F1601138, 74H1701448, 74J1901702, 74L1801776, 74B1600588, 74D1800229, 74F1703312, 74H1701449, 74K1602783, 74L1900363, 74L1900760, 74M1602086, 74M1701826, 74L1901239, 74M1701022, 74M1900013  m) Teleflex Medical Disposable Aortic Punch, 4.8 mm, REF MDP-48K Lot Numbers: 74A1602178, 74C1902230, 74F1600361, 74H1502516, 74J1800850, 74K1801700, 74A1801621, 74D1503161, 74F1600362, 74H1601396, 74J1800851, 74K1801701, 74A1802576, 74D1602682, 74F1601982, 74H1602477, 74J1802113, 74L1501243, 74A1802577, 74D1700565, 74F1602484, 74H1603433, 74J1802740, 74L1602770, 74A1900255, 74D1700566, 74F1700830, 74H1701450, 74J1900241, 74L1700365, 74A1902358, 74D1800230, 74F1800874, 74H1701840, 74J1901703, 74L1701127, 74B1600589, 74D1800231, 74F1802300, 74H1802554, 74K1600566, 74L1803269, 74B1800512, 74E1501776, 74G1600722, 74H1901514, 74K1600567, 74L1901406, 74B1800730, 74E1601487, 74G1602125, 74H1903400, 74K1601333, 74M1601321, 74B1802586, 74E1601494, 74G1802096, 74J1501758, 74K1601334, 74M1602097, 74C1501509, 74E1703226, 74G1802684, 74J1600431, 74K1602784, 74M1701023, 74C1801155, 74E1801610, 74G1902626, 74J1601187, 74K1700917, 74M1701827, 74C1802503, 74E1802702, 74G1902835, 74J1601188, 74K1702873, 74M1701828, 74C1802504, 74E1901452, 74G1903731, 74J1702563, 74K1800963  n) Teleflex Medical Disposable Aortic Punch, 5.2 mm, REF MDP-52K Lot Numbers: 74A1603171, 74C1502258, 74D1902815, 74F1902890, 74H1701451, 74L1702291, 74A1603177, 74C1600457, 74F1500987, 74F1902893, 74H1701452, 74M1600399, 74A1801622, 74C1600464, 74F1600363, 74G1501260, 74H1801717, 74M1900015, 74A1900570, 74D1602270, 74F1702740, 74G1503013, 74J1502610, 74M1900690, 74A1903115, 74D1700574, 74F1702741, 74G1801519, 74K1502265, 74M1901101, 74B1501133, 74D1700575, 74F1800875, 74H1502512, 74K1602785, 74C1501510, 74D1800232, 74F1902113, 74H1601397, 74L1500806  o) Teleflex Medical Disposable Aortic Punch, 5.6 mm, REF MDP-56K Lot Numbers: 74A1502837, 74A1601260, 74A1602183, 74B1600579, 74C1600458, 74D1901016  p) Teleflex Medical Disposable Aortic Punch, 6.0 mm, REF MDP-60K Lot Numbers: 74A1500226, 74B1503153, 74C1900613, 74F1602485, 74H1500728, 74J1602582, 74A1601259, 74B1602370, 74D1500074, 74F1702742, 74H1502910, 74J1901701, 74A1700732, 74C1502256, 74D1700569, 74F1800876, 74H1601398, 74K1800083, 74A1801623, 74C1600459, 74D1801515, 74G1501745, 74H1702723, 74K1901793, 74A1802579, 74C1600970, 74E1700603, 74G1801520, 74H1702724, 74L1601650, 74A1901241, 74C1800374, 74F1500723, 74G1902631, 74H1801718, 74L1700366, 74B1501262, 74C1802502, 74F1502701, 74G1902818, 74H1802917 
Recalling Firm/
Manufacturer
TELEFLEX MEDICAL INC
3015 Carrington Mill Blvd
Morrisville NC 27560-5437
Manufacturer Reason
for Recall
There is a potential for the device packaging to become compromised; thus, the sterility of the device cannot be guaranteed.
FDA Determined
Cause 2
Under Investigation by firm
ActionThe firm initiated the recall by letter on 01/13/2020. The actions to be taking be the consignees are as follows: Check inventory. If affected stock is found, discontinue use immediately and quarantine. Complete the acknowledgement form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. For additional questions, contact the local sales representative or Customer Service at 1-866-396-2111.
Quantity in Commerce587066 units
DistributionWorldwide
Total Product Life CycleTPLC Device Report