Date Initiated by Firm |
May 20, 2020 |
Date Posted |
June 26, 2020 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-2411-2020 |
Recall Event ID |
85684 |
510(K)Number |
K060309
|
Product Classification |
System, balloon, intra-aortic and control - Product Code DSP
|
Product |
Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE SPANISH, Product Code IAP-0500E |
Code Information |
Serial Number Ranges: AUTOCAT2 WAVE SPANISH IAP-0500E 150302W 150304W 150402W through 150403W 150408W 150514W 150826W 151014W 151107W 151110W 160425W through 160427W 160508W through 160510W 160518W Arrow AutoCAT 2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE SPANISH IAP-0500E 160631W 160637W 160642W 160644W 160705W 160919W 170204W through 170206W 170221W 170308W 170328W 170928W 171116W 180333W 180840W 180939W 181054W 181056W through 181060W 190229W 190512W 190514W 190519W 190631W through 190633W |
Recalling Firm/ Manufacturer |
Arrow International Inc 2400 Bernville Rd Reading PA 19605-9607
|
For Additional Information Contact |
610-378-0131
|
Manufacturer Reason for Recall |
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
|
FDA Determined Cause 2 |
Component design/selection |
Action |
Urgent Medical Device Correction notification letters dated 5/20/20 were sent to customers.
ACTIONS TO BE TAKEN BY FACILITIES (Immediate):
Please immediately check your inventory of Arrow¿ AutoCAT¿2 and Arrow¿ AC3 Optimus¿ IABPs, whether stored or in use, and determine if you have an IABP with a model number listed above. Please see Appendix 1 for product codes and serial numbers.
If an IABP in the scope of this Medical Device Correction displays a System Error 3 or High Baseline alarm, now or at any point in the future until the Long-Term Corrective Action detailed below occurs, immediately quarantine the device and contact Teleflex at 1-855-419-8507 or recalls@teleflex.com to receive support for inspection and servicing of the impacted device.
We recommend against using an IABP for ground or air transport between medical facilities pending implementation of the Customer Action (Long Term). For IABP use during ground or air transport prior to implementation of the Customer Action (Long Term), it is recommended that the risks and benefits of using the IABP be assessed by the medical team treating the patient and that alternative circulatory support devices be considered.
Customer Action (Subsequent):
Teleflex will contact each impacted facility to schedule inspection and servicing of all IABP units that contain this internal component. This inspection and servicing will be performed onsite at facilities, for all IABP units.
Facilities should adhere to the following instructions pending implementation of the Teleflex Long-Term Corrective Action described later in this letter:
1. Ensure that a backup IABP is available as instructed within the Operator Manual. If no such replacement IABP is immediately available, it is recommended that the risks and benefits of using the IABP be assessed by the medical team treating the patient and that alternative circulatory support devices be considered.
2. IABP units should be closely monitored during delivery of IAB |
Quantity in Commerce |
2123 total |
Distribution |
Worldwide distribution. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|