| | Class 2 Device Recall Panther Fusion Reaction Tube |  |
| Date Initiated by Firm | August 04, 2020 |
| Create Date | September 21, 2020 |
| Recall Status1 |
Terminated 3 on December 23, 2024 |
| Recall Number | Z-2990-2020 |
| Recall Event ID |
86248 |
| Product Classification |
Instrumentation for clinical multiplex test systems - Product Code NSU
|
| Product | Panther Fusion Tube Tray, Part: FAB-15004, REF: PRD-04000 |
| Code Information |
Box/ Tray lots: 271436/ 618539, 271438/ 620899, 272089/ 620899, 272959/ 621837, 272970/ 621256, 272971/ 621837, 274670/ 625351 |
| FEI Number |
2024800
|
Recalling Firm/ Manufacturer |
Hologic, Inc
10210 Genetic Center Dr
San Diego
CA
92121-4362
|
| For Additional Information Contact | 858-410-8000 |
Manufacturer Reason for Recall | Tube trays may leak which could invalidate patient results or assay worklists, causing delayed results for any assays with which they are used. A possibility exists of false negative result with flu A/B/RSV assay or paraflu assay. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On 08/04/20, Medical Device Recall Notifications were mailed to Laboratory Managers, Site Administrators, and Operators.
Customers were asked to do the following:
1. Discontinue use.
2. Immediately check inventory and segregate affected devices.
3. Complete the customer response form. Document the remaining inventory. Destroy the segregated inventory. Return the customer response form even if there is no affected inventory present.
4. There is no other specific actions to the assays.
5. Recalling firm will ship replacement product based on the customer response form information. |
| Quantity in Commerce | 607 |
| Distribution | U.S. :ID, SC, VT, CA, IL, UT, NJ, MA, TX, AL, CO, FL, ME, NY, VA, WA, WI, OR, DC, GA, NC, HI, CT, OK, TN,
O.U.S.: CA, SE, GB, DE, DO, IT, ES, NO, BE, NL, FR, DK, AT |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|