| | Class 2 Device Recall OSTEORAPTOR CRVD 2.3 SA UB COBRD BLACK Absorbable Suture anchor |  |
| Date Initiated by Firm | September 09, 2020 |
| Create Date | October 21, 2020 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0166-2021 |
| Recall Event ID |
86462 |
| 510(K)Number | K151105 |
| Product Classification |
Fastener, fixation, biodegradable, soft tissue - Product Code MAI
|
| Product | OSTEORAPTOR CRVD 2.3 SA UB COBRD BLACK- Absorbable Suture anchor
Product Number: 72203291 |
| Code Information |
Batch Numbers: 2012936 2013911 2016129 2018794 2027463 2029395 2033280 |
Recalling Firm/ Manufacturer |
Smith & Nephew, Inc. 150 Minuteman Rd Andover MA 01810-1031
|
Manufacturer Reason for Recall | Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by the pouch |
FDA Determined Cause 2 | Equipment maintenance |
| Action | Smith & Nephew issued Urgent Medical Device Recall letter (R-2020-19) FSN on September 9, 2020 via email or overnight mail to customers that received affected products. Letter states reason for recall, health risk, and action to take:
Required Actions:
1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately.
a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete.
b. Include the customers account number and name in the table below.
2. If you have no product to return, please put an X in the appropriate location below.
3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below.
4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com.
Please Note even if you have no product to return, this form must be completed, signed and returned.
5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number.
If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com. |
| Quantity in Commerce | 263,874 (total) |
| Distribution | Nationwide
Foreign:
AE, AT, AU, BE, BM, BR, CA, CH, CL, CN, CO, CY,
CZ, DE, DK, DO, EC, EE, ES, FI, FR, GB, GE, GR, HK,
HU, IE, IL, IN, IT, JP,KR, LV,MY, NL, NO, PA,PE, PL,
PR, PT, RU, SE, SG, SI, TH, TR, TW, UA, & ZA |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|