• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall WOLFPAK Self Safe Safety IV Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall WOLFPAK Self Safe Safety IV Catheter see related information
Date Initiated by Firm June 21, 2021
Create Date August 16, 2021
Recall Status1 Open3, Classified
Recall Number Z-2305-2021
Recall Event ID 88234
510(K)Number K200373  
Product Classification Catheter, intravascular, therapeutic, short-term less than 30 days - Product Code FOZ
Product WOLF-PAK Self Safe Safety IV Catheter:
a) REF PIVD2010W
b) REF PIV2210W
c) REF PIV2475W

Code Information a) REF PIVD2010W, Lot 1100256 b) REF PIV2210W, Lot 1100316 c) REF PIV2475W, Lot 1200084  
Recalling Firm/
Manufacturer
Delta Med SpA
Via Guido Rossa 20
Viadana Italy
Manufacturer Reason
for Recall
Problems related to the sterilization of the medical devices; possible missing sterility of the medical device
FDA Determined
Cause 2
Process control
Action The firm disseminated the notices on 06/21/2021 by email. The firm is requesting destruction of the affected devices at the consignee.
Quantity in Commerce 40,000 units
Distribution US Nationwide distribution in the states of FL and MN.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = FOZ and Original Applicant = Delta Med Spa
-
-