  Class 2 Device Recall Philips see related information
Date Initiated by Firm August 11, 2021
Create Date September 03, 2021
Recall Status1 Completed
Recall Number Z-2412-2021
Recall Event ID 88491
510(K)Number K162931  
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
Product Ingenia 1.5T Evolution-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities
Model: 781315
Code Information Model: 78135 Serial Number Unique Device Identifier (UDI#) 61019 (01)00884838099715 61028 (01)00884838099715 61001 (01)00884838099715 61004 (01)00884838099715 61023 (01)00884838099715 61008 (01)00884838099715 61026 (01)00884838099715 61034 (01)00884838099715 61046 (01)00884838099715 61025 (01)00884838099715 61035 (01)00884838099715 61042 (01)00884838099715 61040 (01)00884838099715 61041 (01)00884838099715(21)61041 61007 (01)00884838099715 61014 (01)00884838099715 61018 (01)00884838099715 61000 N/A 61003 (01)00884838099715 61002 N/A 61006 N/A 61009 N/A 61013 N/A 61017 (01)00884838099715(21)61017 61032 (01)00884838099715(21)61032 61015 N/A 61005 (01)00884838099715 61020 (01)00884838099715 61043 (01)00884838099715 61029 (01)00884838099715 61030 (01)00884838099715 61033 (01)00884838099715(21)61033 61052 (01)00884838099715(21)61052 61064 (01)00884838099715(21)61064 61027 (01)00884838099715 61016 (01)00884838099715 61022 (01)00884838099715 61044 (01)00884838099715 61050 (01)00884838099715 61047 (01)00884838099715(21)61047 61053 (01)00884838099715(21)61053 61049 (01)00884838099715(21)61049
Recalling Firm/
Manufacturer
Philips North America Llc
222 Jacobs St
Cambridge MA 02141-2289
Manufacturer Reason
for Recall
When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
FDA Determined
Cause 2
Device Design
Action Philips notified U.S. customers an Urgent Medical Device Correction letter on 8/11/21. Letter states reason for recall, health risk and action to take: Philips advises that customers continue to utilize the primary motorized function for the MR system since the identified issue with the HDU does not occur in the motorized mode. If you utilize the manual function and it is not functioning properly, then you must perform table movement using the motorized function. A Philips Field Service Engineer (FSE) will perform a HDU during the next scheduled preventive maintenance visit. If the issue is detected in the HDU, the Philips FSE will replace the necessary mechanical component(s) to resolve the issue. If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377) and reference FCO78100527.
Quantity in Commerce 42 units
Distribution Worldwide distribution - US Nationwide distribution and the countries of Albania, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan Palestine, State of Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = LNH and Original Applicant = PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
-
-