• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ARCHITECT cSystems

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ARCHITECT cSystemssee related information
Date Initiated by FirmSeptember 29, 2021
Create DateNovember 19, 2021
Recall Status1 Open3, Classified
Recall NumberZ-0272-2022
Recall Event ID 88880
510(K)NumberK983212 
Product Classification Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
ProductARCHITECT c4000 REF 02P24-01/2P24, 1P86, 1R24, and 1R25; ARCHITECT c8000 REF 01G06-11/1G06; ARCHITECT c16000 REF 03L77-01/3L77;
Code Information Product Description/List Number ARCHITECT c4000 Module List Number: 2P24 UDI: (01)00380740003746(21)c4123456(240)02P2401 Serial Numbers: C400000 C400010 C400011 C400012 C400013 c400014 c400015 C400016 C400017 C400018 C400019 C400020 C400023 C400024 C400025 C400026 C400027 C400030 C400031 C400032 C400034 C400035 C400036 C400038 C400039 C400040 C400041 C400042 C400043 C400045 C400046 C400047 C400049 C400050 C400051 C400053 C400054 C400055 C400056 C400057 C400058 C400060 C400061 C400062 C400063 C400064 C400065 c400066 C400067 C400068 C400069 C400070 C400071 C400072 C400073 C400074 C400075 C400077 C400078 C400079 C400080 C400081 C400083 C400084 C400085 C400086 C400087 C400089 C400090 C400091 C400092 C400093 C400095 C400096 C400098 C400099 C400100 C400101 C400102 C400104 C400105 C400106 C400107 C400108 C400109 C400112 C400114 C400115 C400116 C400117 C400118 C400119 C400120 C400121 C400122 C400123 C400124 C400125 C400127 C400129 C400132 C400133 C400134 C400135 C400136 C400137 c400139 C400140 C400141 C400142 C400143 C400144 C400145 C400146 C400147 C400148 C400149 C400150 C400151 C400152 C400153 C400154 C400155 C400156 C400157 C400158 C400159 C400160 C400161 C400162 C400163 C400164 C400165 C400166 C400167 C400168 C400169 C400170 C400171 C400172 C400173 C400174 C400175 C400176 C400177 C400178 C400179 C400182 C400183 C400184 C400185 C400186 C400187 C400188 C400189 C400190 C400191 C400192 C400195 C400196 C400197 C400199 C400200 C400201 C400202 C400203 C400204 C400205 C400206 C400207 C400209 C400210 C400211 C400212 C400213 C400214 C400215 C400217 C400218 C400220 C400222 C400223 C400224 C400227 c400228 C400229 C400230 C400231 C400232 C400234 C400235 C400236 C400237 C400238 C400239 C400240 C400241 C400242 C400243 C400244 C400245 C400246 C400247 C400248 C400250 C400251 C400252 C400253 C400254 C400256 C400258 C400259 C400260 C400261 C400262 C400263 C400264 C400266 C400268 C400269 C400270 C400271 C400273 C400274 C400275 C400276 C400277 C400278 c400279 C400280 C400281 C400282 C400283 C400284 C400285 C400287 C400288 C400289 C400290 C400291 C400292 C400293 C400294 C400295 C400296 C400297 C400298 C400299 C400300 C400301 C400302 C400303 C400304 C400305 C400306 C400307 C400309 C400310 C400312 C400313 C400315 C400316 C400317 C400318 C400319 C400320 C400321 C400322 C400324 C400325 C400326 C400327 C400328 C400329 C400331 C400332 C400333 C400334 C400335 C400337 C400338 C400340 C400341 C400342 C400344 C400345 C400347 C400348 C400349 C400350 C400351 C400352 C400354 C400355 C400356 C400357 C400358 C400359 C400362 C400363 C400364 C400365 C400366 C400367 C400368 C400369 C400370 C400371 C400373 C400374 C400375 C400376 C400377 C400378 C400379 C400380 C400381 C400382 C400383 C400384 C400385 C400386 C400387 C400388 C400389 C400390 C400391 C400392 C400393 C400394 C400395 C400397 C400398 C400399 C400400 C400402 C400403 C400404 C400405 C400406 C400407 C400408 C400409 C400410 C400411 C400412 C400413 C400414 C400415 C400416 c400417 C400418 C400419 C400420 C400422 C400423 C400424 C400425 C400426 C400427 C400428 C400429 C400430 C400431 C400432 C400433 C400434 C400435 C400436 C400437 C400439 C400440 c400442 C400443 C400444 C400445 C400446 C400447 C400448 C400449 C400450 C400451 C400453 C400454 C400455 C400456 C400457 C400459 C400460 C400461 C400463 C400464 C400465 C400466 C400467 C400468 C400469 C400470 C400471 C400472 C400473 C400474 C400475 C400476 C400477 C400478 C400479 C400480 C400481 C400482 C400483 C400484 C400485 C400486 C400487 C400488 C400489 c400490 C400491 C400492 C400493 c400494 C400495 C400496 C400497 C400498 C400499 C400500 C400501 C400502 C400503 C400504 C400505 C400508 C400509 C400510 C400511 C400512 C400513 C400514 C400515 C400516 C400518 c400519 C400520 C400521 C400522 C400523 C400524 C400525 C400526 C400527 C400528 C400529 C400530 C400531 C400532 C400533 C400534 C400535 C400536 C400537 C400538 C400539 C400540 c400541 C400542 C400543 C400544 C400545 C400546 C400547 C400548 C400549 c400550 C400551 C400552 C400554 c400555 C400556 C400557 C400558 C400559 C400560 C400561 C400562 C400563 C400564 C400565 C400566 C400567 C400568 C400569 C400570 C400571 C400572 C400573 C400574 C400575 C400576 C400577 C400578 C400579 C400580 C400581 C400582 C400583 C400584 C400585 C400586 C400587 C400588 C400589 C400590 C400591 C400593 C400594 C400595 C400596 C400597 C400598 C400599 C400600 C400601 C400602 C400603 C400604 C400605 C400606 C400607 C400608 c400609 C400610 C400611 C400612 C400614 C400616 C400617 C400618 C400619 C400620 C400621 C400622 C400623 C400624 C400625 C400626 C400627 C400629 C400630 C400631 C400632 C400633 C400634 C400635 C400636 C400637 C400638 C400639 C400640 C400641 C400642 C400643 C400645 C400646 C400647 C400648 C400649 C400650 C400651 C400652 C400653 C400654 C400655 C400656 C400657 C400658 C400659 C400660 C400661 C400662 C400663 c400664 C400665 C400666 C400667 C400668 C400669 C400670 c400671 c400672 C400673 C400675 C400676 C400677 C400678 C400679 C400680 C400681 C400682 C400683 C400684 C400685 C400686 C400687 C400688 C400689 C400690 C400691 C400692 C400693 C400694 C400695 C400696 C400697 C400699 c400700 C400701 c400702 C400703 C400704 C400705 C400706 C400707 C400708 C400709 C400710 C400711 C400712 C400713 C400714 C400715 C400716 C400717 C400718 C400719 C400720 C400721 C400722 C400723 C400724 C400725 C400726 C400727 C400728 C400729 C400730 C400731 C400732 C400733 C400734 C400735 C400736 C400737 C400739 C400740 C400741 C400742 C400743 C400744 C400745 C400746 C400747 C400748 C400749 C400750 C400751 C400752 C400753 C400754 C400755 C400756 C400757 C400758 C400759 C400760 C400761 C400762 C400764 C400765 C400766 C400767 C400768 C400769 C400770 C400771 C400772 C400773 C400775 C400776 C400777 C400778 C400779 C400780 C400781 C400782 C400784 C400785 C400786 C400787 C400789 C400790 C400791 C400792 C400793 C400794 c400795 C400796 C400797 C400798 C400799 C400800 C400801 C400802 C400803 C400804 C400805 C400806 C400807 C400809 C400810 C400811 C400812 C400813 c400814 C400815 C400816 C400817 C400818 C400819 C400820 C400821 C400822 C400823 C400824 C400825 c400826 C400827 C400828 C400829 C400830 C400831 C400832 C400834 C400835 C400836 c400837 C400838 C400839 C400840 C400841 C400842 C400843 C400844 c400845 C400846 C400847 C400848 c400849 c400850 C400851 C400852 C400853 C400854 C400855 C400856 C400857 C400858 C400859 C400860 C400861 C400862 C400863 C400864 C400866 C400867 C400869 C400870 C400871 C400872 C400873 C400874 C400875 C400876 C400877 C400879 C400880 C400881 C400882 C400883 C400884 C400885 C400886 C400887 C400888 C400889 C400890 C400891 C400892 C400893 C400895 C400897 C400898 C400899 C400900 C400901 C400902 C400903 C400904 C400905 C400906 C400908 C400909 C400910 C400911 C400912 C400913 C400914 C400915 C400916 c400917 C400919 C400920 C400921 C400922 C400923 C400925 C400926 C400927 C400928 C400929 C400930 C400931 C400932 c400933 C400934 C400935 C400936 C400937 C400938 C400939 C400940 C400941 C400943 C400944 C400945 C400946 C400947 C400948 C400949 C400951 C400952 C400953 C400954 C400955 C400956 C400957 C400958 C400959 C400960 C400961 C400962 C400963 C400964 C400965 C400966 C400967 C400968 C400969 C400970 C400971 C400972 C400973 C400974 C400975 C400976 C400977 C400978 C400979 C400980 C400981 C400982 C400983 C400984 C400985 C400986 C400987 C400988 C400989 C400991 C400992 C400993 C400994 C400995 C400996 C400997 C400998 C400999 C401000 C401001 C401002 C401003 C401004 C401005 C401006 C401007 C401008 C401009 C401010 C401011 C401012 C401013 C401014 C401016 C401017 c401018 C401019 C401020 C401021 C401022 C401023 C401025 C401026 C401027 C401028 C401029 C401030 C401031 C401032 C401033 C401036 C401037 C401038 C401039 C401041 C401042 C401043 C401044 C401045 C401046 C401047 c401048 C401049 C401050 C401051 C401052 C401053 C401054 C401055 C401056 C401057 C401058 C401059 C401060 C401061 C401062 C401063 C401064 C401065 C401066 C401067 C401069 C401070 C401071 C401072 C401073 c401074 C401075 C401076 C401077 C401078 c401079 C401080 C401081 C401082 C401083 C401084 C401085 C401086 C401087 C401088 C401089 C401090 C401091 C401092 C401093 C401094 C401095 C401097 C401098 C401099 C401100 C401101 C401102 C401103 C401104 C401105 C401106 C401107 c401108 c401109 C401110 C401112 C401113 C401114 C401115 C401116 C401117 C401118 C401119 C401120 C401121 C401124 C401125 C401126 C401127 C401128 C401129 C401130 C401131 C401132 C401133 C401134 C401135 C401137 C401138 C401139 C401140 C401141 C401142 C401143 C401144 C401145 C401146 C401147 C401148 C401149 C401150 C401151 C401152 C401153 C401154 C401155 C401156 c401157 C401158 C401159 c401160 C401161 C401162 C401163 C401164 C401165 C401166 C401167 c401168 C401169 C401170 C401171 C401172 C401173 C401174 C401175 C401176 C401177 C401178 C401179 C401180 C401181 C401182 c401183 C401185 c401186 C401187 C401188 C401189 C401190 C401191 c401192 C401193 C401194 C401195 C401196 C401197 C401198 C401199 C401200 c401201 C401202 C401203 C401204 C401205 c401206 C401207 C401208 C401209 C401210 C401211 C401212 C401213 C401214 C401215 C401216 C401217 C401218 C401219 C401220 C401221 C401222 C401223 C401224 C401225 C401226 C401227 C401228 C401229 C401230 C401231 C401232 C401233 C401234 C401235 C401236 C401237 C401238 C401239 C401240 C401241 C401242 C401243 C401244 C401245 C401246 C401247 C401248 C401249 C401250 C401251 C401252 C401253 C401254 C401255 C401256 C401257 C401258 C401259 C401260 C401261 c401262 C401263 C401264 C401265 C401266 C401267 C401268 C401269 C401270 C401271 C401272 C401273 C401274 C401275 C401276 C401277 C401278 C401279 C401280 C401281 C401282 C401283 C401284 C401285 C401286 C401287 C401288 C401289 C401290 c401291 C401292 C401293 C401294 C401295 C401296 C401297 C401298 C401299 C401300 C401301 C401302 C401303 C401304 C401305 C401306 C401307 C401308 C401309 C401310 C401311 C401312 C401313 C401314 C401315 C401316 C401317 C401318 C401319 C401320 C401321 C401322 C401323 C401324 C401325 C401326 C401327 C401328 C401329 C401330 C401331 C401332 C401333 C401334 C401335 C401336 C401337 C401338 C401339 C401340 C401341 C401342 C401343 C401344 C401345 C401347 C401348 C401349 C401351 C401352 C401353 C401354 C401355 C401356 C401357 C401358 C401359 C401360 C401361 C401362 C401363 C401364 C401365 C401366 C401367 C401368 C401369 C401370 C401371 C401372 C401373 C401375 C401376 C401377 C401378 C401379 C401380 C401381 C401382 C401383 C401384 C401385 C401386 C401387 C401388 C401389 C401390 C401391 C401392 C401393 C401394 C401395 C401396 C401397 C401398 C401399 C401400 C401401 C401402 C401403 C401404 C401405 C401406 C401407 C401408 C401409 C401411 C401412 C401413 C401414 C401415 C401416 C401417 C401418 C401419 C401420 C401421 C401422 C401423 C401424 C401425 C401426 C401428 C401429 C401430 C401431 C401432 C401433 C401434 C401435 C401436 C401437 C401438 C401439 C401440 C401441 C401442 c401443 C401444 C401446 C401447 C401448 C401449 C401450 C401451 C401452 C401453 C401454 C401455 C401456 C401457 C401458 C401459 C401460 C401461 C401462 C401463 C401464 C401465 C401466 C401467 C401468 c401469 C401470 C401471 C401472 C401473 C401474 C401475 C401476 C401477 C401478 C401479 C401480 C401481 C401482 C401483 C401484 C401485 C401486 C401487 C401488 C401489 C401490 C401491 C401492 C401493 C401494 C401495 C401496 C401498 C401499 C401500 C401502 C401503 C401504 C401505 C401506 C401507 C401508 C401509 C401510 C401512 C401513 C401514 c401515 C401516 C401517 C401518 C401519 C401520 C401521 C401522 C401523 C401524 C401525 C401526 C401527 C401528 C401529 C401530 C401531 C401532 C401533 C401534 C401535 C401536 C401537 C401538 C401539 C401540 C401541 c401542 C401543 C401544 C401545 C401546 C401547 C401548 C401549 C401550 C401551 C401552 C401553 C401554 C401555 C401556 C401557 C401558 C401559 C401560 C401561 C401562 C401563 C401564 C401565 C401566 C401567 C401568 C401569 C401570 C401571 C401572 C401573 C401574 C401575 C401576 C401577 C401579 C401580 C401581 C401582 C401583 C401584 C401585 C401586 C401587 C401588 C401589 C401590 C401591 C401592 c401593 C401594 C401595 C401596 C401597 C401598 C401599 C401600 C401601 C401602 C401603 C401604 C401605 C401606 C401607 C401608 C401609 C401610 C401611 C401612 C401613 C401614 c401615 C401616 C401617 C401618 C401619 C401620 C401621 C401622 C401623 C401624 C401625 C401626 C401627 C401628 c401629 C401630 C401631 c401632 C401633 C401635 C401636 C401637 C401638 C401639 C401640 C401641 C401642 C401643 C401644 C401645 C401646 C401647 C401648 C401649 C401650 C401651 C401652 C401653 C401654 C401655 C401656 C401658 C401659 C401660 C401661 C401662 C401663 C401664 C401665 C401666 C401667 C401668 C401669 C401670 C401671 C401672 C401673 C401674 c401675 C401676 C401677 C401678 C401679 c401680 c401681 C401682 C401683 C401684 C401685 C401686 C401687 C401688 C401689 C401690 C401691 C401692 C401693 C401694 C401695 c401696 C401697 C401698 C401699 C401700 C401701 C401702 C401703 C401704 C401705 C401706 C401707 C401708 C401709 C401710 C401711 C401712 C401713 C401714 C401715 C401716 C401717 c401718 C401719 c401720 C401721 C401722 C401723 C401724 C401725 C401726 C401727 C401728 C401729 C401730 C401731 C401732 C401733 c401734 c401735 C401736 c401737 C401738 C401739 C401740 C401741 C401742 C401743 C401744 C401745 C401746 C401747 C401748 C401749 C401750 C401751 C401752 C401753 C401754 C401755 C401756 C401757 C401758 C401759 C401760 C401761 C401762 C401763 C401764 C401765 C401766 C401767 C401768 C401769 C401770 C401771 C401772 C401773 C401774 C401775 C401776 C401777 C401778 C401779 c401780 C401781 C401782 C401783 C401784 C401785 C401786 C401787 C401788 C401789 C401790 C401791 C401792 C401793 C401794 C401795 C401796 C401797 C401798 C401799 C401800 C401801 C401802 C401804 C401805 C401806 C401807 C401808 C401809 C401810 C401811 C401813 C401814 C401815 C401816 C401817 C401818 C401819 C401820 C401821 C401822 C401823 C401824 C401825 C401826 C401827 C401828 C401829 C401830 C401831 C401832 C401833 C401834 C401835 C401836 C401837 C401838 C401839 C401840 C401841 C401842 C401843 C401844 C401845 C401846 C401847 C401848 C401849 C401850 C401851 C401852 C401853 C401854 C401855 C401856 C401857 C401858 C401859 c401860 C401861 C401863 C401864 C401865 C401866 C401867 C401868 C401869 C401870 C401871 C401872 C401873 C401874 C401875 C401876 C401877 C401878 C401879 C401880 C401881 C401882 C401883 C401884 C401885 C401886 C401887 C401888 C401889 C401890 C401891 C401892 C401893 C401894 C401895 C401896 C401897 c401898 c401899 C401900 C401901 C401902 C401903 C401904 c401905 C401906 C401907 C401908 C401909 C401910 c401911 C401912 C401913 C401914 C401915 C401916 C401917 C401918 C401919 C401920 C401921 C401922 C401923 C401924 C401925 C401926 C401927 C401928 C401929 C401930 C401931 c401932 c401933 C401934 c401935 C401936 C401938 C401940 C401941 C401942 C401943 C401944 C401945 C401946 C401947 C401948 C401949 C401950 C401951 C401952 C401953 C401954 C401955 C401956 C401957 C401958 C401959 C401960 c401961 C401962 C401963 C401964 C401965 C401966 C401967 C401968 C401969 C401970 c401971 C401972 C401973 C401974 C401975 C401976 C401977 C401978 C401979 C401980 C401981 C401982 C401983 C401984 C401985 C401986 C401987 C401988 C401989 C401990 C401991 C401992 C401993 C401994 C401995 C401996 C401997 C401998 C401999 C402000 C402001 C402002 C402003 C402004 C402005 C402006 C402007 C402008 C402009 C402010 C402011 C402012 C402013 C402014 C402015 C402016 C402017 C402018 C402019 C402020 C402021 C402022 C402023 C402024 C402025 C402026 C402027 C402028 C402029 C402030 C402031 C402032 C402033 C402034 C402035 C402036 C402037 C402038 C402039 C402040 C402041 C402042 C402043 C402044 C402045 C402046 C402047 C402048 C402049 C402050 C402051 C402052 C402053 C402054 C402055 C402056 C402057 C402058 C402059 C402060 C402061 C402062 C402063 C402064 C402065 C402066 C402067 C402068 C402069 C402070 C402071 C402072 C402073 C402074 C402075 C402076 C402077 C402078 C402079 C402080 C402081 C402082 C402083 C402084 C402085 C402086 C402087 C402088 C402089 C402090 C402091 C402092 c402093 C402094 C402095 C402096 C402097 C402098 C402099 C402100 C402101 C402102 C402103 C402104 c402105 C402106 C402107 C402108 C402109 C402110 C402111 C402112 C402113 C402114 C402115 C402116 C402117 C402118 C402119 C402120 C402121 C402122 C402123 C402124 C402125 C402126 c402127 C402128 c402129 C402130 C402131 C402132 C402133 C402134 C402135 C402136 C402137 C402138 C402139 C402140 C402141 C402142 C402143 C402144 C402145 C402146 C402147 C402148 C402149 C402150 C402151 c402152 C402153 C402154 C402155 C402156 C402157 C402158 C402159 C402160 c402161 C402162 c402163 C402164 C402165 C402166 C402167 C402168 C402169 C402170 C402171 C402172 C402173 C402174 C402176 C402177 C402178 C402179 C402180 C402181 C402182 C402183 C402184 C402185 C402186 C402187 c402188 C402189 C402190 C402191 C402192 C402193 C402194 C402195 C402196 C402197 C402198 C402199 C402200 C402201 C402202 C402203 C402204 C402205 C402206 C402207 C402208 C402209 C402210 C402211 c402212 c402213 C402214 C402215 C402216 C402217 C402218 C402219 C402220 C402221 C402222 C402225 C402226 c402227 C402228 C402229 C402230 C402231 C402233 C402234 C402235 C402236 C402237 C402238 C402239 C402240 c402241 C402242 C402243 C402244 C402245 C402246 C402247 C402248 C402249 C402250 C402251 C402252 C402253 C402254 C402255 C402256 C402257 C402258 C402259 C402260 C402261 C402262 C402263 C402264 C402265 C402266 C402267 C402268 C402269 C402270 C402271 C402272 C402273 C402274 C402275 C402276 C402277 C402278 C402279 C402280 C402281 C402282 C402283 C402284 C402285 C402286 C402287 C402288 C402289 C402290 C402291 C402292 C402293 C402294 c402295 C402296 C402297 C402298 C402299 C402300 C402301 C402302 C402303 C402304 C402305 C402306 C402307 C402308 C402309 C402310 C402311 C402312 c402313 C402314 C402315 C402316 C402317 C402318 C402319 C402320 C402321 C402323 C402324 C402325 C402326 C402327 C402328 C402329 C402330 C402331 C402332 C402333 C402334 C402335 C402336 C402337 C402338 C402339 C402340 C402341 C402345 C402346 c402347 C402348 C402349 C402350 C402351 C402352 C402353 C402354 C402355 C402356 C402357 C402358 C402359 C402360 C402361 C402362 C402363 C402364 C402365 C402366 C402367 C402368 C402369 C402370 C402371 C402372 C402373 C402374 c402375 C402376 C402378 c402379 c402380 C402382 C402383 C402384 C402385 C402386 C402387 C402388 C402389 C402390 C402391 C402392 C402393 C402394 C402395 C402396 C402397 C402398 C402399 C402400 C402401 C402402 C402403 C402404 C402405 C402406 C402407 C402408 C402409 C402410 C402411 C402412 C402413 C402414 C402415 C402416 C402417 C402418 C402419 C402420 C402421 C402422 C402423 C402424 C402425 C402426 C402427 C402428 C402429 C402430 C402431 C402432 C402433 C402434 C402435 C402436 C402437 C402438 C402439 C402440 C402441 C402442 C402443 c402444 c402445 c402446 C402447 C402448 C402449 C402450 C402451 C402452 C402453 C402454 C402455 C402456 C402457 C402458 C402459 C402460 C402461 C402462 C402463 c402464 C402465 C402466 C402467 C402468 C402469 C402470 C402471 C402472 c402473 C402474 C402475 C402476 C402477 C402478 C402479 C402480 C402481 c402482 C402483 C402484 C402485 C402486 C402487 C402488 C402489 C402490 C402491 C402492 C402493 C402494 C402495 C402496 C402497 C402498 C402499 C402500 C402501 C402502 C402503 C402504 C402505 C402506 C402507 C402508 C402509 C402510 C402511 C402512 C402513 C402514 c402515 c402516 C402517 c402518 C402519 C402520 C402521 c402522 C402523 c402524 C402525 c402526 C402527 C402528 C402529 C402530 C402531 C402532 C402533 C402534 C402535 C402536 C402537 C402538 C402539 C402540 C402541 C402542 C402543 C402544 C402545 C402546 C402547 C402548 C402549 C402550 C402551 C402552 C402553 C402554 C402555 C402556 C402559 C402560 c402561 C402562 C402563 C402564 C402565 C402567 C402568 C402569 C402570 c402571 c402572 c402573 C402574 C402575 C402576 C402577 C402578 C402579 C402580 C402581 c402582 C402583 c402584 c402585 c402586 c402587 c402588 C402589 C402590 C402591 C402592 C402593 C402594 C402595 C402596 C402597 C402598 C402599 C402602 C402603 C402604 C402605 C402606 C402607 C402608 C402609 C402610 C402611 C402612 C402613 C402614 C402615 C402616 C402617 C402618 C402619 C402620 C402621 C402622 C402623 C402624 C402625 C402626 C402627 C402628 C402629 C402630 C402631 C402632 C402633 C402634 C402635 C402636 C402637 C402638 C402639 c402640 C402641 C402642 C402643 c402644 c402645 C402646 c402647 C402648 C402649 C402650 C402651 C402652 C402653 C402654 C402655 C402656 C402657 C402658 C402659 C402660 C402661 C402662 C402663 C402664 C402665 C402666 C402667 C402668 C402669 C402670 C402671 C402672 C402673 C402674 C402675 C402676 C402677 C402678 c402679 C402680 C402681 C402682 C402683 c402684 c402685 c402686 c402687 C402688 C402689 C402690 C402691 C402692 C402693 C402694 C402695 C402696 c402697 c402698 C402699 C402700 C402701 C402702 C402703 C402704 C402705 C402706 C402707 C402708 C402709 C402710 C402711 C402712 C402713 c402714 C402715 C402716 C402717 C402718 C402719 C402720 C402721 C402722 c402723 c402724 C402725 c402726 C402727 c402728 C402729 C402730 c402731 c402732 C402733 C402734 C402735 C402736 c402737 C402738 C402739 C402740 C402741 C402742 C402743 C402744 C402745 C402746 C402747 C402748 C402749 C402750 C402751 C402752 C402753 C402754 C402755 C402756 C402757 C402758 C402759 C402760 C402761 C402762 C402763 C402764 C402765 C402766 C402767 C402768 C402769 C402770 C402771 C402772 C402773 C402774 C402775 C402776 C402777 c402778 c402779 C402780 C402781 C402782 C402783 C402784 C402785 C402786 C402787 C402788 C402789 C402790 C402791 C402792 C402793 C402794 C402795 C402796 C402797 C402798 c402799 C402800 C402801 C402802 c402803 C402804 C402805 C402806 C402807 C402808 C402809 C402810 c402811 C402812 C402813 C402814 C402815 C402816 C402817 C402818 C402819 C402820 C402821 C402822 C402823 C402824 C402825 C402826 C402827 C402828 C402829 C402830 C402831 C402832 C402833 C402834 C402835 C402836 C402837 C402838 C402839 C402840 C402841 C402842 C402843 C402844 C402845 C402846 C402847 C402848 C402849 C402850 C402851 C402852 C402853 C402854 C402855 C402856 C402857 C402858 C402859 C402860 C402861 C402862 C402863 C402864 C402865 C402866 C402867 C402868 C402869 C402870 C402871 C402872 c402873 C402874 C402875 C402876 C402877 C402878 C402879 C402880 C402881 C402882 C402883 C402884 C402885 C402886 C402887 C402888 C402889 c402890 c402891 C402892 c402893 C402895 C402896 C402897 C402898 C402899 C402900 C402901 C402902 C402903 C402904 C402905 C402906 C402907 C402908 C402909 C402910 C402911 C402912 C402913 C402914 C402915 C402916 C402917 C402918 C402919 C402920 C402921 C402922 C402923 C402924 C402925 C402926 C402927 C402928 C402930 C402931 C402932 C402933 C402934 C402935 C402936 C402937 C402938 C402939 C402940 C402941 C402942 C402943 C402944 C402945 C402946 C402947 C402948 c402949 C402950 C402951 C402952 C402953 C402954 C402955 C402956 C402957 C402958 C402959 C402960 C402961 C402962 C402963 C402964 C402965 C402966 C402969 C402970 C402971 C402972 C402973 C402974 C402975 C402976 C402977 C402978 C402979 C402980 C402981 C402982 C402983 C402984 C402985 C402986 C402987 C402988 C402989 C402990 C402991 C402992 C402993 C402994 C402995 C402996 C402997 C402998 C402999 C403000 C403001 C403002 C403003 C403004 c403005 C403006 C403007 C403008 C403009 C403010 C403012 C403013 C403014 C403015 C403016 C403017 C403018 C403019 C403020 C403021 C403022 C403023 C403024 C403025 C403026 C403027 C403028 C403029 c403030 C403031 C403032 C403033 C403036 C403038 C403043 C403044 C403046 C403047 C403048 c403049 C403050 C403051 C403053 C403055 C403056 C403057 C403058 C403059 C403060 C403061 C403062 c403063 C403064 C403065 C403066 C403067 C403068 C403073 C403074 C403075 c403076 C403077 C403081 C403082 C403083 C403084 C403085 C403086 C403088 C403089 C403090 C403091 C460001 C460002 C460003 C460004 C460005 C460006 C460007 C460008 C460009 C460010 C460011 C460012 C460014 C460016 C460017 C460018 C460019 C460020 C460021 C460022 c460023 C460024 C460025 C460026 C460029 C460030 C460031 C460032 C460033 C460034 C460035 C460038 C460040 C460041 C460042 C460043 C460044 C460046 C460047 C460048 C460049 C460050 C460051 C460052 C460053 C460055 C460056 C460057 C460059 C460060 C460061 C460062 C460063 C460064 C460065 C460066 C460067 C460068 C460069 C460070 C460071 C460072 C460073 C460074 C460075 C460076 C460077 C460078 C460080 C460081 c460082 C460083 C460084 C460085 C460086 C460087 C460088 c460089 C460090 C460091 C460092 C460093 C460095 C460096 C460097 C460098 C460099 C460100 C460101 C460102 C460103 C460105 C460106 C460107 C460108 C460109 C460110 C460111 C460112 C460113 C460114 C460115 C460116 C460117 C460118 C460119 C460120 C460121 C460122 C460123 C460124 C460125 c460126 C460127 C460128 C460129 C460130 C460131 C460132 c460133 C460134 C460135 C460136 C460137 C460138 C460139 C460140 C460141 C460142 C460143 C460144 C460145 C460146 C460147 C460148 C460149 C460150 C460151 C460152 C460153 C460154 C460155 C460156 C460158 C460159 C460160 C460161 C460162 C460163 C460164 C460165 C460166 C460167 C460169 C460170 C460172 C460173 C460174 C460175 C460176 C460177 C460178 C460179 C460180 C460181 c460182 C460183 C460184 C460185 C460186 C460187 C460188 C460189 C460190 C460191 C460192 C460193 C460194 C460195 C460196 C460197 C460198 C460199 C460200 C460201 C460202 C460203 C460204 C460207 C460208 C460209 C460210 C460211 C460212 C460213 C460214 C460215 C460216 C460217 C460218 C460219 C460220 C460221 C460222 C460224 C460226 C460227 C460228 C460229 C460230 c460231 C460232 C460233 C460234 C460235 C460236 C460237 C460238 C460239 C460240 C460241 C460242 C460243 C460244 C460245 C460246 C460247 C460248 C460249 C460250 C460251 C460252 C460253 C460254 C460255 C460256 C460257 C460258 C460259 C460260 C460261 C460263 C460264 C460265 C460266 C460267 C460268 C460269 C460270 C460271 C460272 C460273 C460274 C460275 C460276 C460277 C460278 C460279 C460280 C460281 C460282 C460283 C460284 C460285 C460286 C460287 C460288 C460290 C460291 C460292 C460293 c460295 C460296 C460297 C460298 C460299 C460301 C460302 C460303 C460304 C460305 C460306 C460307 C460308 C460309 C460310 C460311 C460312 C460313 C460314 C460315 C460316 C460317 C460318 C460319 C460320 C460321 C460322 C460323 C460324 C460325 C460326 C460327 C460328 C460329 C460330 C460332 C460333 C460334 C460335 C460336 C460337 C460338 C460340 C460341 C460342 C460343 C460344 C460345 C460346 C460347 C460348 C460349 C460350 C460351 C460352 C460353 C460354 C460355 C460356 C460357 C460359 C460360 C460361 C460362 C460364 C460365 C460366 C460367 C460368 C460370 C460371 C460372 C460373 C460374 C460375 c460376 C460378 C460379 C460380 C460381 c460383 C460384 C460385 c460386 C460387 C460389 C460390 C460391 C460392 c460393 C460394 C460395 C460396 C460397 C460398 C460399 C460400 C460401 C460402 c460403 C460405 C460406 C460408 C460409 C460410 C460411 C460414 C460415 C460416 C460417 C460418 C460419 C460420 C460421 C460422 C460423 C460424 C460425 C460426 C460428 C460429 C460430 C460431 C460432 C460433 C460434 C460435 C460436 C460437 C460438 C460439 C460440 C460441 C460442 C460444 C460445 C460446 C460447 C460449 C460450 C460451 C460452 C460453 C460454 C460455 C460456 C460457 C460458 C460460 C460461 C460462 C460464 C460465 C460466 C460467 C460468 C460469 C460470 C460471 C460472 C460473 C460474 C460475 C460476 C460477 C460478 C460479 C460480 C460482 C460483 C460484 C460485 C460486 C460488 C460489 C460490 C460491 C460492 C460493 C460495 C460497 C460499 C460500 C460501 C460502 C460505 C460506 C460507 C460508 C460509 C460510 C460511 C460512 C460513 C460514 C460515 C460516 C460517 C460518 C460519 c460520 C460521 C460522 C460523 C460524 C460525 C460526 C460527 C460528 C460529 C460530 C460531 C460532 C460534 C460535 c460536 C460537 C460538 C460539 C460540 C460541 C460542 C460543 C460544 C460545 C460546 c460547 C460548 C460549 C460550 C460551 C460552 C460553 C460554 C460555 C460556 C460557 C460558 C460559 C460560 C460561 C460562 C460563 c460564 C460565 C460566 C460567 C460568 c460569 c460570 C460571 C460572 C460573 C460574 C460576 C460577 C460578 C460579 C460580 C460581 C460582 C460583 C460584 c460585 C460586 C460587 C460588 C460589 C460590 C460591 C460592 C460593 C460594 c460595 C460596 C460597 C460598 C460599 C460600 C460601 C460602 C460603 C460604 C460605 C460606 C460607 C460608 C460610 C460611 C460612 C460613 C460614 C460615 C460616 C460617 C460618 C460619 C460620 C460621 C460622 C460623 C460624 C460625 C460626 C460627 C460628 C460629 C460630 C460631 C460632 C460633 C460634 C460635 C460636 C460637 C460638 C460639 C460640 C460641 C460642 C460643 C460645 C460646 C460647 C460648 C460649 C460650 C460651 C460652 C460653 C460654 C460656 C460657 C460658 C460659 C460660 C460661 C460662 C460663 C460664 c460666 C460667 c460668 C460669 C460670 C460671 C460672 C460673 C460674 C460675 C460676 C460677 C460678 C460679 C460680 C460681 C460682 C460683 C460684 C460685 C460686 C460687 C460688 C460689 C460690 C460691 C460692 C460693 C460695 C460696 C460697 C460700 C460701 C460702 C460703 C460704 C460705 C460706 C460707 C460708 C460709 C460710 C460711 C460712 C460713 C460714 C460715 C460716 C460717 C460718 c460719 C460720 C460721 C460722 c460723 C460724 C460725 C460726 C460727 C460728 C460729 C460730 C460731 C460732 C460733 C460734 C460735 C460736 C460737 C460738 C460739 c460740 C460741 C460742 C460743 C460744 C460745 C460746 C460747 C460748 C460750 C460751 C460752 C460753 C460756 c460757 C460758 C460759 C460760 C460761 c460762 C460763 C460764 C460765 C460766 C460767 C460768 C460769 C460770 C460772 C460773 C460774 C460775 C460776 C460777 C460778 C460779 C460780 C460781 C460782 C460783 C460784 C460785 C460787 C460788 C460789 C460790 C460791 C460792 C460793 c460794 C460795 C460796 C460797 C460798 C460799 C460800 C460801 C460802 C460803 C460804 C460805 C460806 C460807 C460808 C460809 C460810 C460811 C460812 C460813 C460814 C460815 C460816 c460817 C460818 C460819 C460820 C460821 C460822 C460824 C460825 C460826 C460827 C460828 C460829 C460830 C460831 C460832 C460833 C460834 c460835 C460836 C460837 C460838 C460839 C460840 C460841 C460842 C460843 C460844 C460845 C460846 c460847 C460848 C460849 c460850 C460852 C460853 C460854 C460855 C460856 C460857 C460858 C460859 C460860 c460861 C460862 C460865 C460866 C460868 C460869 C460870 C460871 C460872 C460873 C460874 C460875 C460876 C460877 C460878 C460879 C460880 C460881 C460882 C460883 c460884 C460885 C460886 C460887 C460888 C460889 C460890 C460891 C460892 C460894 C460895 C460896 C460897 C460898 C460899 C460900 C460901 C460902 C460903 C460904 C460905 C460906 C460907 C460908 C460909 C460910 C460911 C460912 C460913 C460914 C460915 C460916 C460917 C460918 C460919 C460920 C460921 C460922 C460923 c460924 C460925 C460926 C460927 C460928 C460929 C460930 C460931 C460932 C460933 C460935 C460936 C460937 C460938 C460939 C460940 C460941 C460942 C460943 C460944 C460945 C460946 C460947 c460948 C460949 C460950 C460951 C460952 C460953 C460954 C460955 C460956 C460957 C460958 C460959 C460960 C460961 C460963 C460964 C460965 C460966 C460967 C460968 C460969 C460970 C460971 C460972 C460973 C460974 C460975 C460976 C460977 C460978 C460979 C460980 C460981 C460982 C460983 C460984 C460985 C460986 C460987 C460988 C460989 C460990 C460991 C460992 C460993 C460994 C460995 C460996 C460997 C460998 c460999 C461000 c461001 C461002 C461003 C461004 c461005 C461006 C461007 C461008 C461009 C461010 C461011 C461012 C461013 C461014 C461015 C461016 C461017 C461018 C461019 C461020 C461021 C461022 C461023 C461024 C461025 c461026 c461027 C461028 C461029 C461030 C461031 C461032 C461033 C461034 C461035 C461036 C461037 C461038 C461039 C461040 C461041 C461042 C461043 C461044 C461045 C461046 C461047 C461048 c461049 C461050 C461051 C461052 C461053 C461054 C461055 C461056 C461057 C461058 C461060 C461061 C461062 C461063 C461064 C461065 C461066 c461067 C461068 C461069 C461070 C461071 C461072 c461073 C461074 C461075 C461076 C461077 C461078 C461079 C461080 C461081 C461082 C461083 C461084 C461085 C461086 C461087 C461088 C461089 C461090 C461091 C461092 C461093 c461094 C461095 C461096 C461097 C461098 c461099 C461100 C461101 C461102 C461103 c461104 C461105 C461107 C461108 C461109 c461110 C461111 C461112 C461113 C461114 C461115 C461117 c461118 c461119 C461120 C461121 C461122 C461123 C461124 C461125 C461126 C461127 C461128 C461129 C461130 C461131 C461132 C461133 c461134 C461135 C461136 C461137 C461138 C461139 C461140 C461141 C461142 c461143 C461144 C461145 C461146 C461147 C461148 C461149 C461150 C461151 C461152 C461153 c461154 C461155 C461156 C461159 C461160 C461161 c461162 C461163 c461164 C461165 C461166 C461167 C461168 C461169 C461170 C461171 C461172 C461173 C461174 C461175 C461176 C461177 c461178 c461179 C461180 C461181 C461182 C461183 C461184 C461185 C461186 C461187 C461188 C461189 C461190 C461191 C461192 C461193 c461194 C461195 C461196 C461197 C461198 C461199 C461200 C461202 C461204 C461205 C461206 C461207 C461208 C461209 C461210 C461212 C461213 C461214 C461215 C461216 c461217 C461218 C461219 C461220 C461221 C461222 C461223 C461224 C461225 C461226 C461227 C461228 C461229 C461230 C461231 C461233 C461235 c461236 C461237 C461238 C461242 C461243 C461244 C461245 C461246 C461247 C461248 C461249 C461250 C461251 C461252 C461254 C461256 C461257 C461258 C461259 C461260 c461261 C461262 C461264 C461266 C461267 C461271 C461272 C461273 C461274 C461277 C461280 C461281 C461283 C461284 C461286 C461295 C461303 C461304 C461306 c461307 C461312 C461316 C461317 C461319 C461320 C461321 C461322 C461323 C461325 C461327 c461335 c461336 C461337 C461338 C461339 c461340 C461341 C461342 C461343 C461348 C461349 C461350 C461351 C461352 C461357 C461358 C461359 C461360 C461361 C461362 c461364 c461367 c461368 C461369 c461370 C461373 C461374 C461375 C461376 C461380 C461384 C461390 C461392 C461401 C461403 c461404 C461405 C461407 C461408 C461410 C461411 C461412 C461413 C461414 C461415 c461416 C461417 c461418 c461419 C461420 C461421 c461422 C461423 C461424 C461425 C461427 C461428 C461429 C461430 C461431 c461432 c461433 C461434 c461435 C461436 C461437 C461438 C461439 C461440 C461441 C461442 c461443 C461445 C461446 C461447 C461448 C461449 C461450 C461451 c461452 C461453 C461454 C461455 c461456 C461457 C461458 C461459 C461461 C461462 C461463 C461464 C461471 C461472 C461473 C461474 C461475 C461476 C461477 C461478 C461479 C461480 C461481 c461482 c461483 C461484 c461485 C461490 C461491 C461492 C461494 C461495 C461496 C461497 C461498 c461504 C461505 C461506 C461511 C461513 C461514 c461515 C461517 C461519 C461522 C461524 C461525 c461526 c461527 C461528 C461529 C461530 C461531 C461532 C461533 C461534 c461535 c461536 C461537 C461538 C461539 C461540 C461541 C461542 C461543 C461544 C461545 C461546 C461547 C461548 C461549 C461550 C461551 c461552 C461553 C461554 C461555 C461556 C461557 C461558 C461559 C461560 C461561 C461562 C461563 C461564 C461565 C461566 C461567 C461568 c461569 C461570 c461572 C461573 C461574 C461575 C461576 C461577 C461578 C461580 C461581 c461582 c461583 c461584 C461585 C461586 C461587 C461588 C461589 C461590 C461592 C461594 C461596 C461597 C461598 C461599 C461600 C461601 C461602 C461603 c461604 C461606 C461608 C461609 C461610 c461611 c461612 C461613 C461614 C461615 C461616 C461617 C461618 C461619 C461620 C461621 C461622 C461623 C461624 C461625 C461628 C461629 C461630 C461631 C461632 C461633 C461634 C461635 C461636 c461637 C461638 C461639 C461640 C461641 C461642 C461643 C461644 C461645 C461646 C461647 C461648 C461649 C461650 C461651 C461652 C461653 C461654 C461655 C461656 C461657 C461658 C461659 C461660 C461662 C461663 C461664 C461665 C461666 C461667 C461668 C461669 C461670 c461672 C461674 C461675 C461676 C461677 C461678 C461679 C461680 C461682 C461683 C461684 c461685 C461686 C461687 C461688 C461689 C461690 C461691 C461692 C461693 C461694 C461697 C461698 C461699 C461700 C461701 C461702 C461705 c461706 C461707 C461708 C461709 C461710 c461711 C461712 C461713 C461714 C461715 C461716 C461717 c461719 C461720 C461721 C461722 C461723 C461724 C461727 C461728 C461729 C461730 C461731 C461734 C461735 C461736 C461738 C461739 C461740 C461741 C461742 C461744 C461745 C461748 C461749 c461750 C461751 C461752 c461754 c461755 C461756 c461759 c461760 C461762 C461763 c461765 C461766 c461767 C461768 C461769 C461770 C461771 C461772 C461773 C461774 C461775 C461776 c461778 C461780 C461781 C461782 C461783 C461784 C461785 C461786 C461787 c461790 C461793 C461794 C461795 C461796 C461797 C461798 C461799 C461800 C461801 C461802 C461803 C461804 C461805 C461806 C461807 C461808 C461809 C461811 C461812 C461813 C461814 c461815 C461816 C461817 c461818 C461819 C461820 C461821 C461822 C461823 c461825 c461826 c461827 c461828 C461829 C461830 c461831 C461832 C461833 c461834 C461835 C461836 C461837 C461838 C461839 C461840 C461841 C461842 C461843 C461844 C461846 C461847 C461848 c461849 c461850 C461851 C461852 C461853 C461854 C461855 C461856 C461857 C461858 C461859 C461860 C461861 C461862 C461863 C461864 C461865 C461866 C461867 C461868 C461869 C461870 C461871 C461872 C461873 C461874 C461875 C461876 C461877 C461878 C461879 C461880 C461881 C461882 C461883 C461884 C461885 C461886 C461887 c461888 C461889 C461890 C461893 c461894 C461895 C461896 C461897 C461898 C461899 C461900 C461901 C461902 c461903 C461904 C461905 C461906 c461907 C461908 C461909 C461910 C461911 C461912 C461913 C461914 C461915 C461916 C461917 C461918 C461919 C461920 C461921 C461922 C461923 C461924 C461925 C461926 C461927 C461928 C461929 C461930 C461931 C461932 C461933 C461934 C461935 C461936 c461937 C461938 C461939 c461940 C461941 C461942 C461943 c461944 c461945 C461946 C461947 c461948 C461949 C461950 C461951 C461952 C461953 C461954 C461955 C461956 c461957 C461958 C461959 C461961 C461962 C461963 C461964 C461965 C461966 C461967 C461968 C461969 C461970 C461971 C461972 C461973 C461974 C461975 C461976 C461977 C461978 C461979 C461980 C461981 C461982 C461983 C461984 C461985 C461986 C461987 C461988 C461989 C461990 C461991 c461992 C461993 C461994 C461995 C461996 C461997 C461998 C461999 C462000 C462001 C462002 C462003 C462004 C462005 C462006 C462007 c462008 C462009 C462010 C462011 c462012 C462013 C462014 C462015 c462016 C462017 C462018 C462019 C462020 C462021 C462022 C462023 C462024 c462025 C462026 c462027 C462028 C462029 C462030 C462031 C462032 C462033 c462034 c462035 c462036 C462037 c462038 C462039 C462040 C462041 C462042 C462043 C462044 C462045 C462046 C462047 C462048 C462049 C462050 C462051 C462052 C462053 C462054 C462055 C462056 C462057 C462058 C462059 C462060 C462061 C462062 C462063 C462064 c462065 c462066 C462067 C462068 C462069 C462070 C462071 C462072 C462073 C462074 C462075 C462076 C462077 C462078 C462079 C462080 C462081 C462083 C462084 C462085 C462086 C462087 C462088 C462089 C462090 C462091 C462092 C462093 C462094 C462095 C462096 C462098 C462099 C462100 c462101 c462102 C462103 C462104 C462105 C462106 C462107 C462108 C462109 C462110 C462111 C462112 C462113 C462114 C462115 C462116 C462117 C462118 C462119 C462120 C462121 C462122 C462123 c462124 C462125 C462126 C462127 C462128 C462129 C462130 C462131 C462132 c462133 C462134 C462135 C462136 C462137 C462138 C462139 C462140 C462141 C462142 C462143 C462144 C462145 C462146 C462147 C462148 c462149 c462150 C462151 C462152 C462153 C462154 C462155 C462156 C462157 c462158 C462159 C462160 C462161 C462162 C462163 C462164 C462165 C462166 C462167 C462168 C462169 C462170 C462171 C462172 C462173 C462174 C462175 C462176 C462177 C462178 C462179 C462180 C462181 C462182 C462183 C462184 C462185 C462186 C462187 C462188 c462189 C462190 C462191 C462192 C462193 C462194 C462195 C462196 c462197 c462198 C462201 c462202 c462204 c462205 C462206 C462207 C462208 C462210 C462212 C462213 C462214 C462215 C462216 C462217 C462218 C462219 C462220 C462221 C462222 C462233 C462234 c462235 C462236 C462237 c462238 C462239 C462240 C462241 c462242 c462243 C462244 C462245 C462246 C462247 C462248 C462250 C462252 c462253 C462254 C462255 C462256 C462257 c462261 C462263 C462264 C462265 C462266 C462267 C462268 C462269 C462272 c462273 C462274 C462276 C462277 C462278 C462279 C462280 C462281 C462282 C462285 C462286 C462287 C462288 C462289 C462290 C462291 C462292 C462293 C462294 C462295 C462296 C462297 C462298 C462299 C462300 C462302 C462314 C462315 C462318 C462319 Module List Number: 1P86 Serial Numbers: C400059 C401015 C401034 C401578 C401634 C402381 C460036 C460294 c460407 C460533 c460694 c461591 c461595 c461605 c461671 c461753 c461777 c461779 c461788 c461824 c461845 c461891 c461892 c462199 Module List Number: 1R24 Serial Numbers: C400037 C400110 C400194 C400763 C460013 C460015 C460027 C460028 C460039 C460045 C460054 C460058 C460094 C460104 C460157 C460168 C460171 C460205 C460206 C460223 C460225 C460262 C460289 C460300 C460331 C460339 C460358 C460363 C460377 C460382 C460388 C460404 C460412 C460413 C460448 C460459 C460463 C460481 C460487 C460494 C460496 C460503 C460504 C460575 C460609 C460644 C460655 C460665 C460749 C460754 C460755 C460771 C460823 C460851 C460863 C460864 C460867 C460893 C460962 C461059 C461106 C461116 C461157 C461201 C461203 C461234 C461239 C461240 C461241 C461253 C461255 C461293 C461294 C461296 C461301 C461302 C461310 C461315 C461318 C461326 C461387 C461397 C461406 C461465 C461681 Module List Number: 1R25 Serial Numbers: C400033 C400048 C400052 C400076 C400082 C400088 C400094 C400097 C400103 C400111 C400113 C400126 C400128 C400130 C400131 C400138 C400180 C400181 C400193 C400198 C400208 C400216 C400219 C400221 C400225 C400226 C400233 C400249 C400255 C400257 C400265 C400267 C400272 C400286 C400308 C400311 C400314 C400323 C400330 C400336 C400339 C400343 C400353 C400360 C400361 c400372 C400396 C400401 C400421 C400438 C400441 C400452 C400458 C400462 C400506 C400507 C400517 C400553 C400592 C400613 C400615 C400628 C400644 C400674 C400698 C400738 C400774 C400783 C400788 C400808 C400833 C400865 C400868 C400878 C400894 C400918 C400924 C400942 C400950 C400990 C401035 C401040 C401068 C401096 C401111 C401122 C401123 C401136 C401184 C401346 C401350 C401374 C401410 C401427 C401445 C401497 C401501 C401511 C401657 C401803 C401812 C401862 C401939 C460037 C460079 C460369 C460427 C460786 C460934 ARCHITECT c8000 Module List Number: 1G06 Serial Numbers: C800019 C800031 C800038 C800040 C800041 C800044 C800048 C800049 C800052 C800054 C800057 C800059 C800061 C800062 C800064 C800065 C800066 C800067 C800068 C800069 C800070 C800071 C800072 C800073 C800074 C800075 C800076 C800077 C800078 C800079 C800080 C800081 C800082 C800083 C800084 C800085 C800086 C800087 C800088 C800089 C800090 C800091 C800092 C800093 C800094 C800095 C800096 C800097 C800098 C800099 C800100 C800101 C800102 C800103 C800104 C800105 C800106 C800107 C800108 C800109 C800110 C800111 C800112 C800113 C800114 C800115 C800116 C800117 C800118 C800119 C800120 C800121 C800122 C800123 C800124 C800125 C800126 C800127 C800128 C800129 C800130 C800131 C800132 C800133 C800134 C800135 C800136 C800137 C800138 C800139 C800140 C800141 C800142 C800143 C800144 C800145 C800146 C800147 C800148 C800149 C800150 C800151 C800152 C800153 C800154 C800155 C800156 C800157 C800158 C800159 C800160 C800161 C800162 C800163 C800164 C800165 C800166 C800167 C800168 C800169 C800170 C800171 C800172 C800173 C800174 C800175 C800176 C800177 C800178 C800179 C800180 C800181 C800182 C800183 C800184 C800185 C800186 C800187 C800188 C800189 C800190 C800191 C800192 C800193 C800194 C800195 C800196 C800197 C800198 C800199 C800200 C800201 C800202 C800203 C800204 C800205 C800206 C800207 C800208 C800209 C800210 C800211 C800212 C800213 C800214 C800215 C800216 C800217 C800218 C800219 C800220 C800221 C800222 C800223 C800224 C800225 C800226 C800227 C800228 C800229 C800230 C800231 C800232 C800233 C800234 C800235 C800236 C800237 C800238 C800239 C800240 C800241 C800242 C800243 C800245 C800246 C800247 C800248 C800249 C800250 C800251 C800252 C800253 C800254 C800255 C800256 C800257 C800258 C800259 C800260 C800261 C800262 C800263 C800264 C800265 C800266 C800267 C800268 C800269 C800270 C800271 C800272 C800273 C800274 C800275 C800276 C800277 C800278 C800279 C800280 C800281 C800282 C800283 C800284 C800285 C800286 C800287 C800288 C800289 C800290 C800291 C800292 C800293 C800294 C800295 C800296 C800297 C800298 C800299 C800300 C800301 C800302 C800303 C800304 C800305 C800306 C800307 C800308 C800309 C800310 C800311 C800312 C800313 C800314 C800315 C800316 C800317 C800318 C800319 C800320 C800321 C800322 C800323 C800324 C800325 C800326 C800327 C800328 C800329 C800330 C800331 C800332 C800333 C800334 C800335 C800336 C800337 C800338 C800339 C800340 C800341 C800342 C800343 C800344 C800345 C800346 C800347 C800348 C800349 C800350 C800351 C800352 C800353 C800354 C800355 C800356 C800357 C800358 C800359 C800360 C800361 C800362 C800363 C800364 C800365 C800367 C800368 C800369 C800370 C800371 C800372 C800373 C800374 C800375 C800376 C800377 C800378 C800379 C800380 C800381 C800382 C800383 C800384 C800385 C800386 C800387 C800388 C800389 C800390 C800391 C800392 C800393 C800394 C800395 C800396 C800397 C800398 C800399 C800400 C800401 C800402 C800403 C800404 C800405 C800406 C800407 C800408 C800409 C800410 C800411 C800412 C800413 C800414 C800415 C800416 C800417 C800418 C800419 C800420 C800421 C800422 C800423 C800424 C800425 C800426 C800427 C800428 C800429 C800430 C800431 C800432 C800433 C800434 C800435 C800436 C800437 C800438 C800439 C800440 C800441 C800442 C800443 C800444 C800445 C800446 C800447 C800448 C800449 C800450 C800451 C800452 C800453 C800454 C800455 C800456 C800457 C800458 C800459 C800460 C800461 C800462 C800463 C800464 C800465 C800466 C800467 C800468 C800469 C800470 C800471 C800472 C800473 C800474 C800475 C800476 C800477 C800478 C800479 C800480 C800481 C800482 C800483 C800484 C800485 C800486 C800487 C800488 C800489 C800490 C800491 C800492 C800493 C800494 C800495 C800496 C800497 C800498 C800499 C800500 C800501 C800502 C800503 C800504 C800505 C800506 C800507 C800508 C800509 C800510 C800511 C800512 C800513 C800514 C800515 C800516 C800517 C800518 C800519 C800520 C800521 C800522 C800523 C800524 C800525 C800526 C800527 C800528 C800529 C800530 C800531 C800532 C800533 C800534 C800535 C800536 C800537 C800538 C800539 C800540 C800541 C800542 C800543 C800544 C800545 C800546 C800547 C800548 C800549 C800550 C800551 C800552 C800553 C800554 C800555 C800556 C800557 C800558 C800559 C800560 C800561 C800562 C800563 C800564 C800565 C800566 C800567 C800568 C800569 C800570 C800571 C800572 C800573 C800574 C800575 C800576 C800577 C800578 C800579 C800580 C800581 C800582 C800583 C800584 C800585 C800586 C800587 C800588 C800589 C800590 C800591 C800592 C800593 C800594 C800595 C800596 C800597 C800598 C800599 C800600 C800601 C800602 C800603 C800604 C800605 C800606 C800607 C800608 C800609 C800610 C800611 C800612 C800613 C800614 C800615 C800616 C800617 C800618 C800619 C800620 C800621 C800622 C800623 C800624 C800625 C800626 C800627 C800628 C800629 C800630 C800631 C800632 C800633 C800634 C800635 C800636 C800637 C800638 C800639 C800640 C800641 C800642 C800643 C800644 C800645 C800646 C800647 C800648 C800649 C800650 C800651 C800652 C800653 C800654 C800655 C800656 C800657 C800658 C800659 C800660 C800661 C800662 C800663 C800664 C800665 C800666 C800667 C800668 C800669 C800670 C800671 C800672 C800673 C800674 C800675 C800676 C800677 C800678 C800679 C800680 C800681 C800682 C800683 C800684 C800685 C800686 C800687 C800688 C800689 C800690 C800691 C800692 C800693 C800694 C800695 C800696 C800697 C800698 C800699 C800700 C800701 C800702 C800703 C800704 C800705 C800706 C800707 C800708 C800709 C800710 C800711 C800712 C800713 C800714 C800715 C800716 C800717 C800718 C800719 C800720 C800721 C800722 C800723 C800724 C800725 C800726 C800727 C800728 C800729 C800730 C800731 C800732 C800733 C800734 C800735 C800736 C800737 C800738 C800739 C800740 C800741 C800742 C800743 C800744 C800745 C800746 C800747 C800748 C800749 C800750 C800751 C800752 C800753 C800754 C800755 C800756 C800757 C800758 C800759 C800760 C800761 C800762 C800763 C800764 C800765 C800766 C800767 C800768 C800769 C800770 C800771 C800772 C800773 C800774 C800775 C800776 C800777 C800778 C800779 C800780 C800781 C800782 C800783 C800784 C800785 C800786 C800787 C800788 C800789 C800790 C800791 C800792 C800793 C800794 C800795 C800796 C800797 C800798 C800799 C800800 C800801 C800802 C800803 C800804 C800805 C800806 C800807 C800808 C800809 C800810 C800811 C800812 C800813 C800814 C800815 C800816 C800817 C800818 C800819 C800820 C800821 C800822 C800823 C800824 C800825 C800826 C800827 C800828 C800829 C800830 C800831 C800832 C800833 C800834 C800835 C800836 C800837 C800838 C800839 C800840 C800841 C800842 C800843 C800844 C800845 C800847 C800848 C800849 C800850 C800851 C800852 C800853 C800854 C800855 C800856 C800857 C800858 C800859 C800860 C800861 C800862 C800863 C800864 C800865 C800867 C800868 C800869 C800870 C800871 C800872 C800873 C800874 C800875 C800876 C800877 C800878 C800879 C800880 C800881 C800882 C800883 C800884 C800885 C800886 C800887 C800888 C800889 C800890 C800891 C800892 C800893 C800894 C800895 C800896 C800897 C800898 C800899 C800900 C800901 C800902 C800903 C800904 C800905 C800906 C800907 C800908 C800909 C800910 C800911 C800912 C800913 C800914 C800915 C800916 C800917 C800918 C800919 C800920 C800921 C800922 C800923 C800924 C800925 C800926 C800927 C800928 C800929 C800930 C800931 C800932 C800933 C800934 C800935 C800936 C800937 C800938 C800939 C800940 C800941 C800942 C800943 C800944 C800945 C800946 C800947 C800948 C800949 C800950 C800951 C800952 C800953 C800954 C800955 C800956 C800957 C800958 C800959 C800960 C800961 C800962 C800963 C800964 C800965 C800966 C800967 C800968 C800969 C800970 C800971 C800972 C800973 C800974 C800975 C800976 C800977 C800978 C800979 C800980 C800981 C800982 C800983 C800984 C800985 C800986 C800987 C800988 C800989 C800990 C800991 C800992 C800993 C800994 C800995 C800996 C800997 C800998 C800999 C801000 C801001 C801002 C801003 C801004 C801005 C801006 C801007 C801008 C801009 C801010 C801011 C801012 C801013 C801014 C801015 C801016 C801017 C801018 C801019 C801020 C801021 C801022 C801023 C801024 C801025 C801026 C801027 C801028 C801029 C801030 C801031 C801032 C801033 C801034 C801035 C801036 C801037 C801038 C801039 C801040 C801041 C801042 C801043 C801044 C801045 C801046 C801047 C801048 C801049 C801050 C801051 C801052 C801053 C801054 C801055 C801056 C801057 C801058 C801059 C801060 C801061 C801062 C801063 C801064 C801065 C801066 C801067 C801068 C801069 C801070 C801071 C801072 C801073 C801074 C801075 C801076 C801077 C801078 C801079 C801080 C801081 C801082 C801083 C801084 C801085 C801086 C801087 C801088 C801089 C801090 C801091 C801092 C801093 C801094 C801095 C801096 C801097 C801098 C801099 C801100 C801101 C801102 C801103 C801104 C801105 C801106 C801107 C801108 C801109 C801110 C801111 C801112 C801113 C801114 C801115 C801116 C801117 C801118 C801119 C801120 C801121 C801122 C801123 C801124 C801125 C801126 C801127 C801128 C801129 C801130 C801131 C801132 C801133 C801134 C801135 C801136 C801137 C801138 C801139 C801140 C801141 C801142 C801143 C801144 C801145 C801146 C801147 C801148 C801149 C801150 C801151 C801152 C801153 C801154 C801155 C801156 C801157 C801158 C801159 C801161 C801162 C801163 C801164 C801165 C801166 C801167 C801168 C801169 C801170 C801171 C801172 C801173 C801174 C801175 C801177 C801178 C801179 C801180 C801181 C801182 C801183 C801184 C801185 C801186 C801187 C801188 C801189 C801190 C801191 C801192 C801193 C801194 C801195 C801196 C801197 C801198 C801199 C801200 C801201 C801202 C801203 C801204 C801205 C801206 C801207 C801208 C801209 C801210 C801211 C801212 C801213 C801214 C801215 C801216 C801217 C801218 C801219 C801220 C801221 C801222 C801223 C801224 C801225 C801226 C801227 C801228 C801229 C801230 C801231 C801232 C801233 C801234 C801235 C801236 C801237 C801238 C801239 C801240 C801241 C801242 C801243 C801244 C801245 C801246 C801247 C801248 C801249 C801250 C801251 C801252 C801253 C801254 C801255 C801256 C801257 C801258 C801259 C801260 C801261 C801262 C801263 C801264 C801265 C801266 C801267 C801268 C801269 C801270 C801271 C801272 C801273 C801274 C801275 C801276 C801277 C801278 C801279 C801280 C801281 C801282 C801283 C801284 C801285 C801286 C801287 C801288 C801289 C801290 C801291 C801292 C801293 C801294 C801295 C801296 C801297 C801298 C801299 C801300 C801301 C801302 C801303 C801304 C801305 C801306 C801307 C801308 C801309 C801310 C801311 C801312 C801313 C801314 C801315 C801316 C801317 C801318 C801319 C801320 C801321 C801322 C801323 C801324 C801325 C801326 C801327 C801328 C801329 C801330 C801331 C801332 C801333 C801334 C801335 C801336 C801337 C801338 C801339 C801340 C801341 C801342 C801343 C801344 C801345 C801346 C801347 C801348 C801349 C801350 C801351 C801352 C801353 C801354 C801355 C801356 C801357 C801358 C801359 C801360 C801361 C801362 C801363 C801364 C801365 C801366 C801367 C801368 C801369 C801370 C801371 C801372 C801373 C801374 C801375 C801376 C801377 C801378 C801379 C801380 C801381 C801382 C801383 C801384 C801385 C801386 C801387 C801388 C801389 C801390 C801391 C801392 C801393 C801394 C801395 C801396 C801397 C801398 C801399 C801400 C801401 C801402 C801403 C801404 C801405 C801406 C801407 C801408 C801409 C801410 C801411 C801412 C801413 C801414 C801415 C801416 C801417 C801418 C801419 C801420 C801421 C801422 C801423 C801424 C801425 C801426 C801427 C801428 C801429 C801430 C801431 C801432 C801433 C801434 C801435 C801436 C801437 C801438 C801439 C801440 C801441 C801442 C801443 C801444 C801445 C801446 C801447 C801448 C801449 C801450 C801451 C801452 C801453 C801454 C801455 C801456 C801457 C801458 C801459 C801460 C801461 C801462 C801463 C801464 C801465 C801466 C801467 C801468 C801469 C801470 C801471 C801472 C801473 C801474 C801475 C801476 C801477 C801478 C801479 C801480 C801481 C801482 C801483 C801484 C801485 C801486 C801487 C801488 C801489 C801490 C801491 C801492 C801493 C801494 C801495 C801496 C801497 C801498 C801499 C801500 C801501 C801502 C801503 C801504 C801505 C801506 C801507 C801508 C801509 C801510 C801511 C801512 C801513 C801514 C801515 C801516 C801517 C801518 C801519 C801520 C801521 C801522 C801523 C801524 C801525 C801526 C801527 C801528 C801529 C801530 C801531 C801532 C801533 C801534 C801535 C801536 C801537 C801538 C801539 C801540 C801541 C801542 C801543 C801544 C801545 C801546 C801547 C801548 C801549 C801550 C801551 C801552 C801553 C801554 C801555 C801556 C801557 C801558 C801559 C801560 C801561 C801562 C801563 C801564 C801565 C801566 C801567 C801568 C801569 C801570 C801571 C801572 C801573 C801574 C801575 C801576 C801577 C801578 C801579 C801580 C801581 C801582 C801583 C801584 C801585 C801586 C801587 C801588 C801589 C801590 C801591 C801592 C801593 C801594 C801595 C801596 C801597 C801598 C801599 C801600 C801601 C801602 C801603 C801604 C801605 C801606 C801607 C801608 C801609 C801610 C801611 C801612 C801613 C801614 C801615 C801616 C801617 C801618 C801619 C801620 C801621 C801622 C801623 C801624 C801625 C801626 C801627 C801628 C801629 C801630 C801631 C801632 C801633 C801634 C801635 C801636 C801637 C801638 C801639 C801640 C801641 C801642 C801643 C801644 C801645 C801646 C801647 C801648 C801649 C801650 C801651 C801652 C801653 C801654 C801655 C801656 C801657 C801658 C801659 C801660 C801661 C801662 C801663 C801664 C801665 C801666 C801667 C801668 C801669 C801670 C801671 C801672 C801673 C801674 C801675 C801676 C801677 C801678 C801679 C801680 C801681 C801682 C801683 C801684 C801685 C801686 C801687 C801688 C801689 C801690 C801691 C801692 C801693 C801694 C801695 C801696 C801697 C801698 C801699 C801700 C801701 C801702 C801703 C801704 C801705 C801706 C801707 C801708 C801709 C801710 C801711 C801712 C801713 C801714 C801715 C801716 C801717 C801718 C801719 C801720 C801721 C801722 C801723 C801724 C801725 C801726 C801727 C801728 C801729 C801730 C801731 C801732 C801733 C801734 C801735 C801736 C801737 C801738 C801739 C801740 C801741 C801742 C801743 C801744 C801745 C801746 C801747 C801748 C801749 C801750 C801751 C801752 C801753 C801754 C801755 C801756 C801757 C801758 C801759 C801760 C801761 C801762 C801763 C801764 C801765 C801766 C801767 C801768 C801769 C801770 C801771 C801772 C801773 C801774 C801775 C801776 C801777 C801778 C801779 C801780 C801781 C801782 C801783 C801784 C801785 C801786 C801787 C801788 C801789 C801790 C801791 C801792 C801793 C801794 C801795 C801796 C801797 C801798 C801799 C801800 C801801 C801802 C801803 C801804 C801805 C801806 C801807 C801808 C801809 C801810 C801811 C801812 C801813 C801814 C801815 C801816 C801817 C801818 C801819 C801820 C801821 C801822 C801823 C801824 C801825 C801826 C801827 C801828 C801829 C801830 C801831 C801832 C801833 C801834 C801835 C801836 C801837 C801838 C801839 C801840 C801841 C801842 C801843 C801844 C801845 C801846 C801847 C801848 C801849 C801850 C801851 C801852 C801853 C801854 C801855 C801856 C801857 C801858 C801859 C801860 C801861 C801862 C801863 C801864 C801865 C801866 C801867 C801868 C801869 C801870 C801871 C801872 C801873 C801874 C801875 C801876 C801877 C801878 C801879 C801880 C801881 C801882 C801883 C801884 C801885 C801886 C801887 C801888 C801889 C801890 C801891 C801892 C801893 C801894 C801895 C801896 C801897 C801898 C801899 C801900 C801901 C801902 C801903 C801904 C801905 C801906 C801907 C801908 C801909 C801910 C801911 C801912 C801913 C801914 C801915 C801916 C801917 C801918 C801919 C801920 C801921 C801922 C801923 C801924 C801925 C801926 C801927 C801928 C801929 C801930 C801931 C801932 C801933 C801934 C801935 C801936 C801937 C801938 C801939 C801940 C801941 C801942 C801943 C801944 C801945 C801946 C801947 C801948 C801949 C801950 C801951 C801952 C801953 C801954 C801955 C801956 C801957 C801958 C801959 C801960 C801961 C801962 C801963 C801964 C801965 C801966 C801967 C801968 C801969 C801970 C801971 C801972 C801973 C801974 C801975 C801976 C801977 C801978 C801979 C801980 C801981 C801982 C801983 C801984 C801985 C801986 C801987 C801988 C801989 C801990 C801991 C801992 C801993 C801994 C801995 C801996 C801997 C801998 C801999 C802000 C802001 C802002 C802003 C802004 C802005 C802006 C802007 C802008 C802009 C802010 C802011 C802012 C802013 C802014 C802015 C802016 C802017 C802018 C802019 C802020 C802021 C802022 C802023 C802024 C802025 C802026 C802027 C802028 C802029 C802030 C802031 C802032 C802033 C802034 C802035 C802036 C802037 C802038 C802039 C802040 C802041 C802042 C802043 C802044 C802045 C802046 C802047 C802048 C802049 C802050 C802051 C802052 C802053 C802054 C802055 C802056 C802057 C802058 C802059 C802060 C802061 C802062 C802063 C802064 C802065 C802066 C802067 C802068 C802069 C802070 C802071 C802072 C802073 C802074 C802075 C802076 C802077 C802078 C802079 C802080 C802081 C802082 C802083 C802084 C802085 C802086 C802087 C802088 C802089 C802090 C802091 C802092 C802093 C802094 C802095 C802096 C802097 C802098 C802099 C802100 C802101 C802102 C802103 C802104 C802105 C802106 C802107 C802108 C802109 C802110 C802111 C802112 C802113 C802114 C802115 C802116 C802117 C802118 C802119 C802120 C802121 C802122 C802123 C802124 C802125 C802126 C802127 C802128 C802129 C802130 C802132 C802133 C802134 C802135 C802136 C802137 C802138 C802140 C802141 C802142 C802143 C802144 C802145 C802146 C802147 C802148 C802149 C802150 C802151 C802152 C802153 C802154 C802155 C802156 C802157 C802158 C802159 C802160 C802161 C802162 C802163 C802164 C802165 C802166 C802167 C802168 C802169 C802170 C802171 C802172 C802173 C802174 C802175 C802176 C802177 C802178 C802179 C802180 C802181 C802182 C802183 C802184 C802185 C802186 C802187 C802188 C802189 C802190 C802191 C802192 C802193 C802194 C802195 C802196 C802197 C802198 C802199 C802200 C802201 C802202 C802203 C802204 C802205 C802206 C802207 C802208 C802209 C802210 C802211 C802212 C802213 C802214 C802215 C802216 C802217 C802218 C802219 C802220 C802221 C802222 C802223 C802224 C802225 C802226 C802227 C802228 C802229 C802230 C802231 C802232 C802233 C802234 C802235 C802236 C802237 C802238 C802239 C802240 C802241 C802242 C802243 C802244 C802245 C802246 C802247 C802248 C802249 C802250 C802251 C802252 C802253 C802254 C802255 C802256 C802257 C802258 C802259 C802260 C802261 C802262 C802263 C802264 C802265 C802266 C802267 C802268 C802269 C802270 C802271 C802272 C802273 C802274 C802275 C802276 C802277 C802278 C802279 C802280 C802281 C802282 C802283 C802284 C802285 C802286 C802287 C802288 C802290 C802291 C802292 C802293 C802294 C802295 C802296 C802297 C802298 C802299 C802300 C802301 C802302 C802303 C802304 C802305 C802306 C802307 C802308 C802309 C802310 C802311 C802312 C802313 C802314 C802315 C802316 C802317 C802318 C802319 C802320 C802321 C802322 C802323 C802324 C802325 C802326 C802327 C802328 C802329 C802330 C802331 C802332 C802333 C802334 C802335 C802336 C802337 C802338 C802339 C802340 C802341 C802342 C802343 C802344 C802345 C802346 C802347 C802348 C802349 C802350 C802351 C802352 C802353 C802354 C802355 C802356 C802357 C802358 C802359 C802360 C802361 C802362 C802363 C802364 C802365 C802366 C802367 C802368 C802369 C802370 C802371 C802372 C802373 C802374 C802375 C802376 C802377 C802378 C802379 C802380 C802381 C802382 C802383 C802384 C802385 C802386 C802387 C802388 C802389 C802390 C802391 C802392 C802393 C802394 C802395 C802396 C802397 C802398 C802399 C802400 C802401 C802402 C802403 C802404 C802405 C802406 C802407 C802408 C802409 C802410 C802411 C802412 C802413 C802414 C802415 C802416 C802417 C802418 C802419 C802420 C802421 C802422 C802423 C802424 C802425 C802426 C802427 C802428 C802429 C802430 C802431 C802432 C802433 C802434 C802435 C802436 C802437 C802438 C802439 C802440 C802441 C802442 C802443 C802444 C802445 C802446 C802447 C802448 C802449 C802450 C802451 C802452 C802453 C802454 C802455 C802456 C802457 C802458 C802459 C802460 C802461 C802462 C802463 C802464 C802465 C802466 C802467 C802468 C802469 C802470 C802471 C802472 C802473 C802474 C802475 C802476 C802477 C802478 C802479 C802480 C802481 C802482 C802483 C802484 C802485 C802486 C802487 C802488 C802489 C802490 C802491 C802492 C802493 C802494 C802495 C802496 C802497 C802498 C802499 C802500 C802501 C802502 C802503 C802504 C802505 C802506 C802507 C802508 C802509 C802510 C802511 C802512 C802513 C802514 C802515 C802516 C802517 C802518 C802519 C802520 C802521 C802522 C802523 C802524 C802525 C802526 C802527 C802528 C802529 C802530 C802531 C802532 C802533 C802534 C802535 C802536 C802537 C802538 C802539 C802540 C802541 C802542 C802543 C802544 C802545 C802546 C802547 C802548 C802549 C802550 C802551 C802552 C802553 C802554 C802555 C802556 C802557 C802558 C802559 C802560 C802561 C802562 C802563 C802564 C802565 C802566 C802567 C802568 C802569 C802570 C802571 C802572 C802573 C802574 C802575 C802576 C802577 C802578 C802579 C802580 C802581 C802582 C802583 C802584 C802585 C802586 C802587 C802588 C802589 C802590 C802591 C802592 C802593 C802594 C802595 C802596 C802597 C802598 C802599 C802600 C802601 C802602 C802603 C802604 C802605 C802606 C802607 C802608 C802609 C802610 C802611 C802612 C802613 C802614 C802615 C802616 C802617 C802618 C802619 C802620 C802621 C802622 C802623 C802624 C802625 C802626 C802627 C802628 C802629 C802630 C802631 C802632 C802633 C802634 C802635 C802636 C802637 C802638 C802639 C802640 C802641 C802642 C802643 C802644 C802645 C802646 C802647 C802648 C802649 C802650 C802651 C802652 C802653 C802654 C802655 C802656 C802657 C802658 C802659 C802660 C802661 C802662 C802663 C802664 C802665 C802666 C802667 C802668 C802669 C802670 C802671 C802672 C802673 C802674 C802675 C802676 C802677 C802678 C802679 C802680 C802681 C802682 C802683 C802684 C802685 C802686 C802687 C802688 C802689 C802690 C802691 C802692 C802693 C802694 C802695 C802696 C802697 C802698 C802699 C802700 C802701 C802702 C802703 C802704 C802705 C802706 C802707 C802708 C802709 C802710 C802711 C802712 C802713 C802714 C802715 C802716 C802717 C802718 C802719 C802720 C802721 C802722 C802723 C802724 C802725 C802726 C802727 C802728 C802729 C802730 C802731 C802732 C802733 C802734 C802735 C802736 C802737 C802738 C802739 C802740 C802741 C802742 C802743 C802744 C802745 C802746 C802747 C802748 C802749 C802750 C802751 C802752 C802753 C802754 C802755 C802756 C802757 C802758 C802759 C802760 C802761 C802762 C802763 C802764 C802765 C802766 C802767 C802768 C802769 C802770 C802771 C802772 C802773 C802774 C802775 C802776 C802777 C802778 C802779 C802780 C802781 C802782 C802783 C802784 C802785 C802786 C802787 C802788 C802789 C802790 C802791 C802792 C802793 C802794 C802795 C802796 C802797 C802798 C802799 C802800 C802801 C802802 C802803 C802804 C802805 C802806 C802807 C802808 C802809 C802810 C802811 C802812 C802813 C802814 C802815 C802816 C802817 C802818 C802819 C802820 C802821 C802822 C802823 C802824 C802825 C802826 C802827 C802828 C802829 C802830 C802831 C802832 C802833 C802834 C802835 C802836 C802837 C802838 C802839 C802840 C802841 C802842 C802843 C802844 C802845 C802846 C802847 C802848 C802849 C802850 C802851 C802852 C802853 C802854 C802855 C802856 C802857 C802858 C802859 C802860 C802861 C802862 C802863 C802864 C802865 C802866 C802867 C802868 C802869 C802870 C802871 C802872 C802873 C802874 C802875 C802876 C802877 C802878 C802879 C802880 C802881 C802882 C802883 C802884 C802885 C802886 C802887 C802888 C802889 C802890 C802891 C802892 C802893 C802894 C802895 C802896 C802897 C802898 C802899 C802900 C802901 C802902 C802903 C802904 C802905 C802906 C802907 C802908 C802909 C802910 C802911 C802912 C802913 C802914 C802915 C802916 C802917 C802918 C802919 C802920 C802921 C802922 C802923 C802924 C802925 C802926 C802927 C802928 C802929 C802930 C802931 C802932 C802933 C802934 C802935 C802936 C802937 C802938 C802939 C802940 C802941 C802942 C802943 C802944 C802945 C802946 C802947 C802948 C802949 C802950 C802951 C802952 C802953 C802954 C802955 C802956 C802957 C802958 C802959 C802960 C802961 C802962 C802963 C802964 C802965 C802966 C802967 C802968 C802969 C802970 C802971 C802972 C802973 C802974 C802975 C802976 C802977 C802978 C802979 C802980 C802981 C802982 C802983 C802984 C802985 C802986 C802987 C802988 C802989 C802990 C802991 C802992 C802993 C802994 C802995 C802996 C802997 C802998 C802999 C803000 C803001 C803002 C803003 C803004 C803005 C803006 C803007 C803008 C803009 C803010 C803011 C803012 C803013 C803014 C803015 C803016 C803017 C803018 C803019 C803020 C803021 C803022 C803023 C803024 C803025 C803026 C803027 C803028 C803029 C803030 C803031 C803032 C803033 C803034 C803035 C803036 C803037 C803038 C803039 C803040 C803041 C803042 C803043 C803044 C803045 C803046 C803047 C803048 C803049 C803050 C803051 C803052 C803053 C803054 C803055 C803056 C803057 C803058 C803059 C803060 C803061 C803062 C803063 C803064 C803065 C803066 C803067 C803068 C803069 C803070 C803071 C803072 C803073 C803074 C803075 C803076 C803077 C803078 C803079 C803080 C803081 C803082 C803083 C803084 C803085 C803086 C803087 C803088 C803089 C803090 C803091 C803092 C803093 C803094 C803095 C803096 C803097 C803098 C803099 C803100 C803101 C803102 C803103 C803104 C803105 C803106 C803107 C803108 C803109 C803110 C803111 C803112 C803113 C803114 C803115 C803116 C803117 C803118 C803119 C803120 C803121 C803122 C803123 C803124 C803125 C803126 C803127 C803128 C803129 C803130 C803131 C803132 C803133 C803134 C803135 C803136 C803137 C803138 C803139 C803140 C803141 C803142 C803143 C803144 C803145 C803146 C803147 C803148 C803149 C803150 C803151 C803152 C803153 C803154 C803155 C803156 C803157 C803158 C803159 C803160 C803161 C803162 C803163 C803164 C803165 C803166 C803167 C803168 C803169 C803170 C803171 C803172 C803173 C803174 C803175 C803176 C803177 C803178 C803179 C803180 C803181 C803182 C803183 C803184 C803185 C803186 C803187 C803188 C803189 C803190 C803191 C803192 C803193 C803194 C803195 C803196 C803197 C803198 C803199 C803200 C803202 C803203 C803204 C803205 C803206 C803207 C803208 C803209 C803210 C803211 C803212 C803213 C803214 C803215 C803216 C803217 C803218 C803219 C803220 C803221 C803222 C803223 C803224 C803225 C803226 C803227 C803228 C803229 C803230 C803231 C803233 C803234 C803235 C803236 C803237 C803238 C803239 C803240 C803241 C803242 C803243 C803244 C803245 C803246 C803247 C803248 C803249 C803250 C803251 C803252 C803253 C803254 C803255 C803256 C803257 C803258 C803259 C803260 C803261 C803262 C803263 C803264 C803265 C803266 C803267 C803268 C803269 C803270 C803271 C803272 C803273 C803274 C803275 C803276 C803277 C803278 C803279 C803280 C803281 C803282 C803283 C803284 C803285 C803286 C803287 C803288 C803289 C803290 C803291 C803292 C803293 C803294 C803295 C803296 C803297 C803298 C803299 C803300 C803301 C803302 C803303 C803304 C803305 C803306 C803307 C803308 C803309 C803310 C803311 C803312 C803313 C803314 C803315 C803316 C803317 C803318 C803319 C803320 C803321 C803322 C803323 C803324 C803325 C803326 C803327 C803328 C803329 C803330 C803331 C803332 C803333 C803334 C803335 C803336 C803337 C803338 C803339 C803340 C803341 C803342 C803343 C803344 C803345 C803346 C803347 C803348 C803349 C803350 C803351 C803352 C803353 C803354 C803355 C803356 C803357 C803358 C803359 C803360 C803361 C803362 C803363 C803364 C803365 C803366 C803367 C803368 C803369 C803370 C803371 C803372 C803373 C803374 C803375 C803376 C803377 C803378 C803379 C803380 C803381 C803382 C803383 C803384 C803385 C803386 C803387 C803388 C803389 C803390 C803391 C803392 C803393 C803394 C803395 C803396 C803397 C803398 C803399 C803400 C803401 C803402 C803403 C803404 C803405 C803406 C803407 C803408 C803409 C803410 C803411 C803412 C803413 C803414 C803415 C803416 C803417 C803418 C803419 C803420 C803421 C803422 C803423 C803424 C803425 C803426 C803427 C803428 C803429 C803430 C803431 C803432 C803433 C803434 C803435 C803436 C803437 C803438 C803439 C803440 C803441 C803442 C803443 C803444 C803445 C803446 C803447 C803448 C803449 C803450 C803451 C803452 C803453 C803454 C803455 C803456 C803457 C803458 C803459 C803460 C803461 C803462 C803463 C803464 C803465 C803466 C803467 C803468 C803469 C803470 C803471 C803472 C803473 C803474 C803475 C803476 C803477 C803478 C803479 C803480 C803481 C803482 C803483 C803484 C803485 C803486 C803487 C803488 C803489 C803490 C803491 C803492 C803493 C803494 C803495 C803496 C803497 C803498 C803499 C803500 C803501 C803502 C803503 C803504 C803505 C803506 C803507 C803508 C803509 C803510 C803511 C803512 C803513 C803514 C803515 C803516 C803517 C803518 C803519 C803520 C803521 C803522 C803523 C803524 C803525 C803526 C803527 C803528 C803529 C803530 C803531 C803532 C803533 C803534 C803535 C803536 C803537 C803538 C803539 C803540 C803541 C803542 C803543 C803544 C803545 C803546 C803547 C803548 C803549 C803550 C803551 C803552 C803553 C803554 C803555 C803556 C803557 C803558 C803559 C803560 C803561 C803562 C803563 C803564 C803565 C803566 C803567 C803568 C803569 C803570 C803571 C803572 C803573 C803574 C803575 C803576 C803577 C803578 C803579 C803580 C803581 C803582 C803583 C803584 C803585 C803586 C803587 C803588 C803589 C803590 C803591 C803592 C803593 C803594 C803595 C803596 C803597 C803598 C803599 C803600 C803601 C803602 C803603 C803604 C803605 C803606 C803607 C803608 C803609 C803610 C803611 C803612 C803613 C803614 C803615 C803616 C803617 C803618 C803619 C803620 C803621 C803622 C803623 C803624 C803625 C803626 C803627 C803628 C803629 C803630 C803631 C803632 C803633 C803634 C803635 C803636 C803637 C803638 C803639 C803640 C803641 C803642 C803643 C803644 C803645 C803646 C803647 C803648 C803649 C803650 C803651 C803652 C803653 C803654 C803655 C803656 C803657 C803658 C803659 C803660 C803661 C803662 C803663 C803664 C803665 C803666 C803667 C803668 C803669 C803670 C803671 C803672 C803673 C803674 C803675 C803676 C803677 C803678 C803679 C803680 C803681 C803682 C803683 C803684 C803685 C803686 C803687 C803688 C803689 C803690 C803691 C803692 C803693 C803694 C803695 C803696 C803697 C803698 C803699 C803700 C803701 C803702 C803703 C803704 C803705 C803706 C803707 C803708 C803709 C803710 C803711 C803713 C803714 C803715 C803716 C803717 C803718 C803719 C803720 C803721 C803722 C803723 C803725 C803726 C803727 C803728 C803729 C803730 C803731 C803732 C803733 C803734 C803735 C803736 C803737 C803738 C803739 C803740 C803741 C803742 C803743 C803744 C803745 C803746 C803747 C803748 C803749 C803750 C803751 C803752 C803753 C803754 C803755 C803756 C803757 C803758 C803759 C803760 C803761 C803762 C803763 C803764 C803765 C803766 C803768 C803769 C803770 C803771 C803772 C803773 C803774 C803775 C803776 C803777 C803778 C803779 C803780 C803781 C803782 C803783 C803784 C803785 C803786 C803787 C803788 C803789 C803790 C803791 C803792 C803793 C803794 C803795 C803796 C803797 C803798 C803799 C803801 C803802 C803803 C803804 C803805 C803806 C803807 C803808 C803809 C803810 C803811 C803812 C803813 C803814 C803815 C803816 C803817 C803818 C803819 C803820 C803821 C803822 C803823 C803824 C803825 C803826 C803827 C803828 C803829 C803830 C803831 C803832 C803833 C803834 C803835 C803836 C803837 C803838 C803839 C803840 C803841 C803842 C803843 C803844 C803845 C803846 C803847 C803848 C803849 C803850 C803851 C803852 C803853 C803854 C803855 C803856 C803857 C803858 C803859 C803860 C803861 C803862 C803863 C803864 C803865 C803866 C803867 C803868 C803869 C803870 C803871 C803872 C803873 C803874 C803875 C803876 C803877 C803878 C803879 C803880 C803881 C803882 C803883 C803884 C803885 C803886 C803887 C803888 C803889 C803890 C803891 C803892 C803893 C803894 C803895 C803896 C803897 C803898 C803899 C803900 C803901 C803902 C803903 C803904 C803905 C803906 C803907 C803908 C803909 C803911 C803913 C803914 C803915 C803916 C803917 C803918 C803919 C803920 C803921 C803922 C803923 C803924 C803925 C803926 C803927 C803928 C803929 C803930 C803931 C803933 C803934 C803936 C803937 C803938 C803939 C803940 C803941 C803942 C803943 C803944 C803945 C803946 C803947 C803948 C803949 C803950 C803951 C803952 C803953 C803954 C803955 C803956 C803957 C803958 C803959 C803960 C803961 C803962 C803963 C803964 C803965 C803966 C803967 C803968 C803969 C803970 C803971 C803972 C803973 C803974 C803975 C803976 C803977 C803978 C803979 C803980 C803981 C803982 C803983 C803984 C803985 C803986 C803987 C803988 C803989 C803990 C803991 C803992 C803993 C803995 C803998 C804000 C804001 C804002 C804003 C804004 C804005 C804006 C804007 C804008 C804009 C804010 C804011 C804012 C804015 C804019 C804025 C804026 C804027 C804028 C804029 C804030 C804031 C804032 C804034 C804035 C804037 C804038 C804039 C804040 C804041 C804042 C804043 C804044 C804045 C804046 C804047 C804048 C804049 C804050 C804051 C804052 C804053 C804054 C804055 C804056 C804057 C804058 C804059 C804060 C804061 C804062 C804063 C804064 C804065 C804066 C804067 C804068 C804069 C804070 C804071 C804072 C804073 C804074 C804075 C804076 C804077 C804078 C804079 C804080 C804081 C804082 C804083 C804084 C804085 C804086 C804087 C804088 C804089 C804090 C804091 C804092 C804093 C804094 C804095 C804096 C804097 C804098 C804099 C804100 C804101 C804102 C804103 C804104 C804105 C804106 C804107 C804108 C804109 C804110 C804111 C804112 C804113 C804114 C804115 C804116 C804117 C804118 C804119 C804120 C804121 C804122 C804123 C804124 C804125 C804126 C804127 C804128 C804129 C804130 C804131 C804132 C804133 C804134 C804135 C804136 C804137 C804138 C804139 C804140 C804141 C804142 C804143 C804144 C804145 C804146 C804147 C804148 C804149 C804150 C804151 C804152 C804153 C804154 C804155 C804156 C804157 C804158 C804159 C804160 C804161 C804162 C804163 C804164 C804165 C804166 C804167 C804168 C804169 C804170 C804171 C804172 C804173 C804174 C804175 C804176 C804177 C804178 C804179 C804180 C804181 C804182 C804183 C804184 C804185 C804186 C804187 C804188 C804189 C804190 C804191 C804192 C804193 C804194 C804195 C804196 C804197 C804198 C804199 C804200 C804201 C804202 C804203 C804204 C804205 C804206 C804207 C804208 C804209 C804210 C804211 C804212 C804213 C804214 C804215 C804216 C804217 C804218 C804220 C804221 C804223 C804226 C804227 C804228 C804229 C804230 C804231 C804232 C804233 C804234 C804235 C804236 C804237 C804238 C804239 C804240 C804241 C804242 C804243 C804244 C804245 C804246 C804247 C804248 C804249 C804250 C804251 C804252 C804253 C804254 C804255 C804256 C804257 C804258 C804259 C804260 C804261 C804262 C804263 C804264 C804265 C804266 C804267 C804268 C804269 C804270 C804271 C804273 C804274 C804275 C804276 C804277 C804278 C804279 C804280 C804281 C804282 C804283 C804284 C804285 C804286 C804287 C804288 C804289 C804290 C804291 C804292 C804293 C804294 C804295 C804296 C804297 C804298 C804299 C804301 C804302 C804303 C804304 C804305 C804306 C804307 C804308 C804309 C804310 C804311 C804312 C804313 C804314 C804315 C804316 C804317 C804318 C804319 C804320 C804321 C804323 C804324 C804325 C804326 C804327 C804328 C804329 C804330 C804331 C804332 C804333 C804334 C804335 C804336 C804337 C804338 C804339 C804340 C804341 C804342 C804343 C804344 C804345 C804347 C804348 C804349 C804350 C804351 C804352 C804353 C804354 C804355 C804356 C804357 C804358 C804359 C804360 C804363 C804364 C804365 C804366 C804367 C804368 C804369 C804370 C804371 C804372 C804373 C804374 C804375 C804376 C804377 C804378 C804379 C804380 C804381 C804385 C804386 C804387 C804388 C804389 C804390 C804391 C804392 C804393 C804394 C804395 C804396 C804397 C804398 C804399 C804400 C804401 C804402 C804403 C804404 C804406 C804408 C804409 C804410 C804411 C804412 C804413 C804414 C804415 C804416 C804417 C804418 C804419 C804420 C804421 C804422 C804423 C804424 C804425 C804426 C804427 C804428 C804429 C804430 C804431 C804432 C804433 C804434 C804435 C804436 C804437 C804438 C804439 C804440 C804441 C804442 C804443 C804444 C804445 C804446 C804447 C804448 C804449 C804452 C804453 C804454 C804455 C804456 C804457 C804458 C804459 C804460 C804461 C804462 C804463 C804464 C804465 C804466 C804467 C804468 C804469 C804470 C804471 C804472 C804473 C804474 C804475 C804476 C804477 C804478 C804479 C804480 C804481 C804482 C804483 C804484 C804485 C804486 C804487 C804488 C804489 C804490 C804491 C804492 C804493 C804494 C804495 C804496 C804497 C804498 C804499 C804500 C804501 C804502 C804503 C804504 C804505 C804506 C804507 C804508 C804509 C804510 C804511 C804512 C804513 C804514 C804515 C804516 C804517 C804518 C804519 C804520 C804521 C804522 C804523 C804524 C804525 C804526 C804527 C804528 C804529 C804530 C804531 C804532 C804533 C804534 C804535 C804536 C804537 C804538 C804539 C804540 C804541 C804542 C804543 C804544 C804545 C804546 C804547 C804548 C804549 C804550 C804551 C804552 C804553 C804554 C804555 C804556 C804557 C804558 C804559 C804561 C804562 C804563 C804564 C804565 C804566 C804567 C804568 C804570 C804571 C804572 C804573 C804574 C804575 C804576 C804577 C804578 C804579 C804580 C804581 C804582 C804583 C804584 C804585 C804586 C804587 C804588 C804589 C804590 C804591 C804592 C804593 C804594 C804595 C804596 C804597 C804598 C804599 C804600 C804601 C804602 C804603 C804604 C804605 C804606 C804607 C804608 C804609 C804610 C804611 C804612 C804613 C804614 C804615 C804616 C804617 C804618 C804619 C804621 C804622 C804623 C804624 C804626 C804627 C804630 C804631 C804634 C804635 C804636 C804637 C804638 C804639 C804640 C804641 C804642 C804643 C804644 C804645 C804646 C804647 C804648 C804649 C804650 C804652 C804653 C804654 C804655 C804656 C804657 C804658 C804659 C804660 C804661 C804662 C804663 C804664 C804665 C804666 C804667 C804669 C804670 C804671 C804672 C804675 C804676 C804677 C804678 C804679 C804680 C804681 C804682 C804683 C804684 C804685 C804686 C804687 C804688 C804689 C804690 C804691 C804692 C804693 C804694 C804695 C804696 C804697 C804698 C804699 C804700 C804701 C804702 C804703 C804704 C804705 C804706 C804707 C804708 C804709 C804710 C804711 C804712 C804713 C804714 C804715 C804716 C804717 C804718 C804719 C804720 C804721 C804722 C804723 C804724 C804725 C804726 C804727 C804728 C804729 C804730 C804731 C804732 C804733 C804734 C804735 C804736 C804737 C804738 C804739 C804740 C804741 C804742 C804743 C804744 C804745 C804746 C804747 C804748 C804749 C804750 C804751 C804752 C804753 C804754 C804755 C804756 C804757 C804758 C804759 C804760 C804761 C804762 C804763 C804764 C804765 C804766 C804767 C804768 C804769 C804770 C804771 C804772 C804773 C804774 C804775 C804776 C804777 C804778 C804779 C804780 C804781 C804782 C804783 C804784 C804785 C804786 C804787 C804788 C804789 C804790 C804791 C804792 C804793 C804794 C804795 C804796 C804797 C804798 C804799 C804800 C804801 C804802 C804803 C804804 C804805 C804807 C804808 C804809 C804810 C804811 C804812 C804813 C804814 C804815 C804816 C804817 C804818 C804819 C804820 C804821 C804822 C804823 C804824 C804825 C804826 C804827 C804828 C804829 C804830 C804831 C804832 C804833 C804834 C804835 C804836 C804837 C804838 C804839 C804840 C804841 C804842 C804843 C804844 C804845 C804846 C804847 C804848 C804849 C804850 C804851 C804852 C804853 C804854 C804855 C804856 C804857 C804858 C804859 C804863 C804864 C804865 C804868 C804869 C804870 C804871 C804872 C804873 C804874 C804875 C804876 C804877 C804878 C804879 C804880 C804881 C804886 C804888 C804889 C804891 C804895 C804896 C804903 C804915 C809584 C887035 C887036 ARCHITECT c16000 Module List Number: 3L77 UDI: (01)00380740005924(21)C16123456(240)03L7701 Serial Numbers: C1600014 C1600015 C1600016 C1600017 C1600024 C1600025 c1600026 C1600027 C1600028 C1600029 C1600030 C1600031 C1600032 C1600033 C1600034 C1600035 C1600036 C1600037 C1600038 C1600040 C1600041 C1600042 C1600043 C1600044 C1600045 C1600046 C1600047 C1600048 C1600049 C1600050 C1600051 C1600052 C1600053 C1600054 C1600055 C1600056 C1600057 C1600058 C1600059 C1600060 C1600061 C1600062 C1600063 C1600064 C1600065 C1600066 C1600067 C1600068 C1600069 C1600070 C1600071 C1600072 C1600073 C1600074 C1600075 C1600076 C1600077 C1600078 C1600079 C1600080 C1600081 C1600082 C1600083 C1600084 C1600085 C1600086 C1600087 C1600088 C1600089 C1600090 C1600092 C1600093 C1600094 C1600095 C1600096 C1600097 C1600098 C1600099 C1600100 C1600101 C1600102 C1600103 C1600104 C1600105 C1600106 C1600107 C1600108 C1600109 C1600110 C1600111 C1600112 C1600113 C1600114 C1600115 C1600116 C1600117 C1600118 C1600119 C1600120 C1600121 C1600122 C1600123 C1600124 C1600125 C1600126 C1600127 C1600128 C1600129 C1600130 C1600131 C1600132 C1600133 C1600134 C1600135 C1600136 C1600137 C1600138 C1600139 C1600140 C1600141 C1600142 C1600143 C1600144 C1600145 C1600146 C1600147 C1600148 C1600149 C1600150 C1600151 C1600152 C1600153 C1600154 C1600155 C1600156 C1600157 C1600158 C1600159 C1600160 C1600161 C1600162 C1600163 C1600164 C1600165 C1600166 C1600167 C1600168 C1600169 C1600170 C1600171 C1600172 C1600173 C1600174 C1600175 C1600176 C1600177 C1600178 C1600179 C1600180 C1600181 C1600182 C1600183 C1600184 C1600185 C1600186 C1600187 C1600188 C1600189 C1600190 C1600191 C1600192 C1600193 C1600194 C1600195 C1600196 C1600197 C1600198 C1600199 C1600200 C1600201 C1600202 C1600203 C1600204 c1600205 C1600206 C1600207 C1600208 C1600209 C1600210 C1600211 C1600212 C1600213 C1600214 C1600215 C1600216 C1600217 C1600218 C1600219 C1600220 C1600221 c1600222 C1600223 C1600224 C1600225 C1600226 C1600227 C1600228 C1600229 C1600230 C1600231 C1600232 C1600233 c1600234 C1600235 C1600236 C1600237 C1600238 C1600239 C1600240 C1600241 C1600242 c1600243 C1600244 c1600245 C1600246 C1600247 C1600248 C1600249 C1600250 C1600251 C1600252 C1600253 C1600254 c1600255 C1600256 C1600257 c1600258 C1600259 C1600260 C1600261 C1600262 C1600263 C1600264 C1600265 C1600266 C1600267 C1600268 C1600269 C1600270 C1600271 c1600272 C1600273 C1600274 C1600275 C1600276 C1600277 C1600278 C1600279 C1600280 C1600281 C1600282 c1600283 C1600284 C1600285 C1600286 c1600287 C1600288 C1600289 C1600290 C1600291 C1600292 C1600293 C1600294 C1600295 c1600296 C1600297 C1600298 C1600299 C1600300 C1600301 C1600302 C1600303 C1600304 C1600305 C1600306 C1600307 C1600308 C1600309 C1600310 C1600311 C1600312 C1600313 C1600314 c1600315 C1600316 C1600317 C1600318 C1600319 C1600320 C1600321 C1600322 C1600323 C1600324 C1600325 C1600326 C1600327 C1600328 C1600329 C1600330 C1600331 C1600332 C1600333 C1600334 C1600335 c1600336 C1600337 C1600338 C1600339 C1600340 C1600341 C1600342 C1600343 C1600344 C1600345 C1600346 C1600347 C1600348 C1600349 C1600350 C1600351 c1600352 C1600353 C1600354 C1600355 C1600356 C1600357 C1600358 C1600359 C1600360 C1600361 C1600362 C1600363 C1600364 C1600365 C1600367 C1600368 C1600369 c1600370 C1600371 C1600372 C1600373 C1600374 C1600375 C1600376 C1600377 c1600378 C1600379 C1600381 C1600382 C1600383 C1600384 C1600385 C1600386 c1600387 C1600388 C1600389 C1600390 C1600391 C1600392 C1600393 C1600394 C1600395 C1600396 C1600397 C1600398 C1600399 C1600400 C1600401 c1600402 C1600403 C1600404 C1600405 C1600406 C1600407 C1600408 C1600409 c1600410 C1600411 C1600412 C1600413 C1600414 C1600415 C1600416 C1600417 C1600418 C1600419 C1600420 C1600421 C1600422 C1600423 C1600424 C1600425 C1600426 C1600427 C1600428 C1600429 C1600430 C1600431 C1600432 c1600433 C1600434 C1600435 C1600436 C1600437 C1600438 C1600439 C1600440 C1600441 C1600442 C1600443 C1600444 C1600445 C1600446 C1600447 C1600448 C1600449 C1600450 C1600451 C1600452 C1600453 C1600454 C1600455 C1600456 C1600457 C1600458 C1600459 c1600460 C1600461 C1600462 C1600463 C1600464 C1600465 C1600466 C1600467 C1600468 C1600469 C1600470 C1600471 C1600472 C1600473 C1600474 C1600475 C1600476 C1600477 C1600478 C1600479 C1600480 C1600481 C1600482 C1600483 C1600484 c1600485 c1600486 C1600487 C1600488 C1600489 c1600490 c1600491 c1600492 C1600493 C1600494 c1600495 C1600496 C1600497 C1600498 C1600499 C1600500 C1600501 c1600502 C1600503 C1600504 C1600505 C1600506 C1600507 C1600508 C1600509 C1600510 C1600511 c1600512 C1600513 C1600514 C1600515 C1600516 C1600517 C1600518 C1600519 C1600520 C1600521 C1600522 C1600523 C1600524 C1600525 C1600526 c1600527 c1600528 C1600529 C1600530 C1600531 C1600532 C1600533 C1600534 C1600535 C1600536 C1600537 C1600538 C1600539 C1600540 C1600541 C1600542 C1600543 C1600544 C1600545 C1600546 C1600547 C1600548 C1600549 C1600550 C1600552 C1600553 C1600554 C1600555 C1600556 C1600557 C1600558 C1600559 C1600560 C1600561 C1600562 C1600563 C1600564 C1600565 C1600566 C1600567 C1600568 c1600569 c1600570 C1600571 C1600572 C1600573 C1600574 C1600575 C1600576 C1600577 C1600578 C1600579 C1600580 C1600581 C1600582 C1600583 C1600584 C1600585 C1600586 C1600587 C1600588 C1600589 C1600590 C1600591 C1600592 C1600593 C1600594 C1600595 C1600596 C1600597 C1600598 C1600599 C1600600 C1600601 C1600602 C1600603 C1600604 C1600605 c1600606 C1600607 C1600608 C1600609 C1600610 C1600611 C1600612 C1600613 C1600614 C1600615 C1600616 C1600617 C1600618 C1600619 C1600620 C1600621 C1600622 C1600623 C1600624 C1600625 C1600626 C1600627 C1600628 C1600629 C1600630 C1600631 C1600632 C1600633 C1600634 C1600635 C1600636 C1600637 C1600638 C1600639 C1600640 C1600641 C1600642 C1600643 c1600644 C1600645 c1600646 C1600647 C1600648 C1600649 C1600650 C1600651 C1600652 C1600653 C1600654 C1600655 C1600656 c1600657 C1600658 C1600659 C1600660 C1600661 C1600662 C1600663 C1600664 C1600665 C1600666 C1600667 C1600668 C1600669 C1600670 C1600671 C1600672 C1600673 C1600674 C1600675 C1600676 C1600677 C1600678 c1600679 C1600680 C1600681 C1600682 C1600683 C1600684 C1600685 C1600686 C1600687 C1600688 C1600689 c1600690 C1600691 C1600692 C1600693 C1600694 C1600695 C1600696 C1600697 C1600698 C1600699 C1600700 C1600701 C1600702 C1600703 c1600704 c1600705 C1600706 C1600707 C1600708 C1600709 C1600710 C1600711 C1600712 C1600713 C1600714 C1600715 C1600716 C1600717 C1600718 C1600719 C1600720 C1600721 C1600722 C1600723 C1600724 C1600725 C1600726 C1600727 C1600728 C1600729 C1600730 C1600731 C1600732 C1600733 C1600734 C1600735 c1600736 C1600737 C1600738 C1600739 C1600740 C1600741 C1600742 C1600743 C1600744 C1600745 C1600746 C1600747 C1600748 C1600749 C1600750 C1600751 C1600752 C1600753 c1600754 C1600755 C1600756 C1600757 C1600758 C1600759 C1600760 C1600761 C1600762 C1600763 C1600764 c1600765 c1600766 c1600767 C1600768 C1600769 C1600770 C1600771 C1600772 c1600773 c1600774 C1600775 C1600776 C1600777 c1600778 c1600779 C1600780 c1600781 C1600782 C1600783 c1600784 c1600785 c1600786 C1600787 C1600788 C1600789 c1600790 C1600791 C1600792 C1600793 C1600794 C1600795 C1600796 C1600797 C1600798 C1600799 C1600800 C1600801 C1600802 C1600803 C1600804 C1600805 C1600806 C1600807 C1600808 C1600809 C1600810 C1600811 C1600812 C1600813 C1600814 c1600815 C1600816 C1600817 C1600818 C1600819 C1600820 C1600821 C1600822 C1600823 C1600824 C1600825 C1600826 C1600827 c1600828 C1600829 C1600830 C1600831 C1600832 C1600833 C1600834 C1600835 C1600836 C1600837 C1600838 C1600839 C1600840 C1600841 C1600842 C1600843 C1600844 C1600845 C1600846 C1600847 C1600848 C1600849 C1600850 C1600851 C1600852 C1600853 C1600854 C1600855 C1600856 C1600857 C1600858 C1600859 C1600860 C1600861 C1600862 C1600863 C1600864 C1600865 C1600866 C1600867 C1600868 C1600869 C1600870 C1600871 C1600872 C1600873 C1600874 C1600875 C1600876 C1600877 C1600878 c1600879 C1600880 C1600881 C1600882 C1600883 C1600884 C1600885 C1600886 C1600887 C1600888 c1600889 C1600890 C1600891 C1600892 C1600893 C1600894 C1600895 C1600896 C1600897 C1600898 C1600899 C1600900 C1600901 C1600902 C1600903 C1600904 C1600905 C1600906 C1600907 C1600908 C1600909 C1600910 C1600911 C1600912 C1600913 C1600914 c1600915 C1600916 C1600917 C1600918 C1600919 C1600920 C1600921 C1600922 C1600923 c1600924 C1600925 C1600926 C1600927 C1600928 C1600929 C1600930 C1600931 C1600932 C1600933 C1600934 C1600935 C1600936 C1600937 C1600938 C1600939 C1600940 C1600941 C1600942 C1600943 C1600944 C1600945 C1600946 C1600947 C1600948 C1600949 C1600950 C1600951 C1600952 C1600953 C1600954 C1600955 C1600956 C1600957 C1600958 C1600959 C1600960 C1600961 C1600962 c1600963 c1600964 C1600965 C1600966 C1600967 C1600968 C1600969 C1600970 c1600971 C1600972 C1600973 C1600974 C1600975 C1600976 C1600977 C1600978 C1600979 C1600980 C1600981 C1600982 C1600983 C1600984 C1600985 C1600986 C1600987 C1600988 C1600989 C1600990 C1600991 C1600992 C1600993 C1600994 C1600995 c1600996 C1600997 C1600998 C1600999 C1601000 C1601001 C1601002 C1601003 C1601004 C1601005 C1601006 C1601007 C1601008 C1601009 C1601010 C1601011 C1601012 C1601013 C1601014 C1601015 C1601016 c1601017 C1601018 C1601019 C1601020 C1601021 C1601022 C1601023 C1601024 C1601025 C1601026 C1601027 C1601028 c1601029 C1601030 C1601031 c1601032 C1601033 C1601034 C1601035 C1601036 C1601037 c1601038 C1601039 C1601040 C1601041 C1601042 C1601043 C1601044 C1601045 C1601046 C1601047 C1601048 C1601049 C1601050 C1601051 C1601052 c1601053 C1601054 C1601055 C1601056 C1601057 C1601058 C1601059 C1601060 C1601061 c1601062 C1601063 C1601064 C1601065 C1601066 C1601067 C1601068 C1601069 C1601070 C1601071 C1601072 C1601073 C1601074 C1601075 C1601076 C1601077 c1601078 c1601079 c1601080 c1601081 c1601082 c1601083 c1601084 c1601085 c1601086 c1601087 c1601088 c1601089 c1601090 c1601091 C1601092 C1601093 c1601094 C1601095 C1601096 C1601097 C1601098 C1601099 C1601100 C1601101 C1601102 C1601103 C1601104 C1601105 C1601106 C1601107 C1601108 C1601109 C1601110 C1601111 C1601112 C1601113 c1601114 c1601115 c1601116 c1601117 C1601118 C1601119 C1601120 C1601121 C1601122 c1601123 c1601124 C1601125 C1601126 C1601127 c1601128 c1601129 C1601130 c1601131 c1601132 c1601133 c1601134 c1601135 C1601136 C1601137 C1601138 C1601139 C1601140 C1601141 C1601142 C1601143 C1601144 C1601145 c1601146 c1601147 c1601148 c1601149 c1601150 C1601151 C1601152 C1601153 C1601154 C1601155 C1601156 C1601157 C1601158 C1601159 C1601160 C1601161 c1601162 c1601163 c1601164 C1601165 C1601166 C1601167 C1601168 C1601169 C1601170 C1601171 C1601172 c1601173 c1601174 c1601175 c1601176 c1601177 C1601178 C1601179 C1601180 C1601181 c1601182 c1601183 c1601184 c1601185 c1601186 c1601187 c1601188 c1601189 c1601190 c1601191 c1601192 C1601193 C1601194 C1601195 C1601196 C1601197 C1601198 C1601199 c1601200 c1601201 c1601202 c1601203 c1601204 c1601205 c1601206 C1601207 C1601208 C1601209 C1601210 C1601211 C1601212 C1601213 C1601214 C1601215 C1601216 C1601217 C1601218 C1601219 c1601220 c1601221 c1601222 c1601223 c1601224 c1601225 c1601226 c1601227 c1601228 c1601229 c1601230 c1601231 c1601232 c1601233 c1601234 C1601235 C1601236 C1601237 c1601238 c1601239 C1601240 c1601241 c1601242 C1601243 C1601244 c1601245 c1601246 c1601247 C1601248 C1601249 C1601250 C1601251 C1601252 C1601253 C1601254 c1601255 c1601256 c1601257 c1601258 c1601259 c1601260 c1601261 c1601262 c1601263 c1601264 c1601265 c1601266 C1601267 c1601268 c1601269 C1601270 C1601271 C1601272 c1601273 C1601274 C1601275 C1601276 C1601277 C1601278 C1601279 C1601280 C1601281 c1601282 C1601283 C1601284 c1601285 c1601286 c1601287 c1601288 c1601289 c1601290 c1601291 C1601292 C1601293 c1601294 c1601295 c1601296 C1601297 C1601298 C1601299 C1601300 C1601301 C1601302 c1601303 C1601304 C1601305 C1601306 C1601307 C1601308 c1601309 C1601310 c1601311 c1601312 c1601313 c1601314 c1601315 c1601316 C1601317 C1601318 C1601319 C1601320 C1601321 C1601323 C1601324 C1601325 C1601326 C1601327 C1601328 C1601329 C1601330 C1601331 C1601332 C1601333 C1601334 C1601335 C1601336 C1601337 C1601338 C1601339 C1601340 C1601341 C1601342 c1601343 C1601344 C1601345 C1601346 C1601347 C1601348 C1601349 C1601350 C1601351 C1601352 C1601353 C1601354 c1601355 c1601356 c1601357 c1601358 c1601359 c1601360 C1601361 C1601362 C1601363 C1601364 C1601365 c1601366 C1601367 C1601368 C1601369 C1601370 C1601371 C1601372 c1601373 C1601374 C1601375 C1601376 C1601377 C1601378 C1601379 C1601380 C1601381 C1601382 C1601383 C1601385 C1601386 C1601387 C1601388 C1601389 C1601390 c1601391 c1601392 c1601393 c1601394 c1601395 c1601396 c1601397 c1601398 c1601399 c1601400 c1601401 c1601402 c1601403 c1601404 C1601405 C1601406 c1601407 c1601408 c1601409 c1601410 c1601411 c1601412 c1601413 c1601414 c1601415 c1601416 c1601417 c1601418 C1601419 C1601420 C1601421 C1601422 C1601423 C1601424 C1601425 C1601426 C1601427 C1601428 c1601429 c1601430 c1601431 c1601432 c1601433 C1601435 c1601436 c1601437 c1601438 c1601439 c1601440 c1601441 c1601442 c1601443 c1601444 C1601445 C1601446 c1601447 c1601448 c1601449 C1601450 C1601451 c1601452 c1601453 c1601454 c1601455 c1601456 c1601457 c1601458 c1601459 c1601460 c1601461 c1601462 c1601463 c1601464 c1601465 c1601466 C1601467 c1601468 c1601469 c1601470 c1601471 c1601472 c1601473 c1601475 c1601476 c1601478 c1601479 c1601480 c1601481 C1601482 C1601483 C1601484 C1601485 c1601486 c1601487 c1601488 c1601489 C1601490 C1601491 C1601492 C1601493 c1601494 C1601495 C1601496 C1601497 C1601498 c1601499 c1601500 c1601501 c1601502 c1601503 c1601504 c1601505 c1601506 c1601507 c1601508 c1601509 c1601510 c1601511 c1601512 c1601513 C1601514 C1601515 C1601516 C1601517 C1601518 c1601519 c1601520 c1601521 c1601522 c1601523 c1601524 c1601525 c1601526 c1601527 c1601528 c1601529 c1601530 c1601531 c1601532 c1601533 C1601534 C1601535 C1601536 C1601537 C1601538 C1601539 C1601540 C1601541 c1601542 c1601543 c1601544 C1601545 C1601546 C1601547 C1601548 c1601549 c1601550 c1601551 c1601552 c1601553 c1601554 c1601555 c1601556 c1601557 c1601558 c1601559 c1601560 C1601561 C1601562 C1601563 C1601564 c1601565 c1601566 c1601567 c1601568 c1601569 c1601570 c1601571 c1601572 c1601573 c1601574 c1601575 C1601576 C1601577 C1601578 c1601579 C1601580 C1601581 C1601582 C1601583 C1601584 C1601585 c1601586 c1601587 c1601588 c1601589 c1601590 c1601591 c1601592 c1601593 c1601594 c1601595 c1601596 c1601597 C1601598 C1601599 C1601600 C1601601 C1601602 C1601603 c1601604 c1601605 C1601606 C1601607 C1601608 C1601609 C1601610 C1601611 c1601612 c1601613 c1601614 c1601615 c1601616 c1601617 c1601618 c1601619 c1601620 c1601621 c1601623 c1601624 C1601625 C1601626 C1601627 C1601628 C1601629 C1601630 C1601631 C1601632 C1601633 C1601634 c1601635 C1601636 C1601637 C1601638 c1601639 C1601640 C1601641 c1601642 C1601643 c1601644 c1601645 c1601646 c1601647 c1601648 c1601649 c1601650 c1601651 c1601652 c1601653 c1601654 c1601655 c1601656 c1601657 c1601658 c1601659 c1601660 c1601661 c1601662 c1601663 C1601666 C1601667 C1601668 C1601669 c1601670 c1601671 c1601672 c1601673 c1601674 c1601675 c1601676 c1601677 c1601678 c1601679 c1601680 c1601681 c1601682 c1601683 c1601684 C1601685 c1601686 c1601687 C1601688 C1601689 c1601690 c1601691 c1601692 c1601693 c1601694 c1601695 c1601696 c1601697 c1601698 c1601699 c1601700 c1601701 c1601702 C1601703 C1601704 c1601705 c1601706 c1601707 c1601708 c1601709 c1601710 c1601711 c1601712 c1601713 c1601714 c1601715 c1601716 c1601717 C1601718 C1601719 C1601720 C1601721 c1601722 c1601723 c1601724 c1601725 c1601726 c1601727 c1601728 c1601729 c1601730 c1601731 c1601732 c1601733 c1601734 C1601735 C1601736 c1601737 c1601738 c1601739 c1601740 c1601741 c1601742 c1601743 c1601744 c1601745 c1601746 c1601747 c1601748 c1601749 C1601750 C1601751 C1601752 C1601753 C1601754 C1601755 C1601756 C1601757 C1601758 C1601759 C1601760 C1601761 C1601762 C1601763 C1601764 C1601765 C1601766 c1601767 C1601768 C1601769 C1601770 C1601771 C1601772 C1601773 C1601774 C1601775 C1601776 C1601777 C1601778 C1601779 C1601780 C1601781 C1601782 C1601783 C1601784 C1601785 C1601786 C1601787 C1601788 C1601789 C1601790 C1601791 C1601792 C1601793 C1601794 C1601795 C1601796 C1601797 C1601798 C1601799 C1601800 C1601801 C1601802 C1601803 C1601804 c1601805 C1601806 C1601807 C1601808 C1601809 C1601810 c1601811 C1601812 C1601813 C1601814 C1601815 C1601816 C1601817 C1601818 C1601819 C1601820 C1601821 C1601822 C1601823 C1601824 C1601825 C1601826 C1601827 C1601828 c1601829 C1601830 C1601831 c1601832 c1601833 c1601834 c1601835 c1601836 c1601837 c1601838 C1601839 C1601840 C1601842 c1601843 c1601844 c1601845 c1601846 c1601847 c1601848 c1601849 c1601850 c1601851 C1601852 C1601853 C1601854 C1601855 C1601856 c1601857 C1601858 C1601859 C1601860 C1601861 C1601863 C1601864 C1601865 C1601866 C1601867 C1601868 C1601869 C1601870 C1601871 C1601872 C1601873 C1601874 C1601875 C1601876 C1601877 C1601878 C1601879 C1601880 C1601881 C1601882 C1601883 C1601884 C1601885 C1601886 C1601887 C1601888 C1601889 C1601890 C1601891 C1601892 C1601893 C1601894 C1601895 C1601896 C1601897 c1601898 c1601899 C1601900 C1601901 C1601902 C1601903 C1601904 C1601905 C1601906 C1601907 C1601908 C1601909 C1601910 C1601911 C1601912 c1601913 C1601914 C1601917 C1601918 C1601919 C1601920 C1601921 C1601922 C1601923 C1601924 C1601925 c1601926 C1601927 C1601928 C1601929 C1601930 C1601931 C1601932 C1601933 c1601934 C1601935 c1601936 c1601937 C1601938 C1601939 C1601940 C1601941 C1601942 C1601943 C1601944 C1601945 C1601946 C1601947 C1601948 C1601949 C1601950 C1601951 C1601952 C1601953 C1601954 C1601955 C1601956 C1601957 C1601958 C1601959 C1601960 C1601961 C1601962 C1601963 C1601964 C1601965 C1601966 c1601967 C1601968 C1601969 C1601970 C1601971 C1601972 C1601973 C1601974 C1601975 C1601976 C1601977 C1601978 C1601979 C1601980 C1601981 C1601982 C1601983 C1601984 C1601985 C1601986 C1601987 C1601988 C1601989 C1601990 C1601991 C1601992 C1601993 C1601994 C1601995 C1601996 C1601997 C1601998 C1601999 C1602000 C1602001 C1602002 C1602003 C1602004 C1602005 C1602006 C1602007 C1602008 C1602009 C1602010 C1602011 C1602012 C1602013 C1602014 C1602015 C1602016 C1602017 C1602018 C1602019 C1602020 c1602021 C1602022 C1602023 C1602024 C1602025 C1602026 C1602027 C1602028 C1602029 C1602030 C1602031 C1602032 C1602033 C1602034 C1602035 C1602036 C1602037 C1602038 C1602039 C1602040 C1602041 C1602042 C1602043 C1602044 C1602045 C1602046 C1602047 C1602048 C1602049 C1602050 C1602051 c1602052 C1602053 C1602054 C1602055 C1602056 C1602057 C1602058 C1602059 C1602060 C1602061 C1602062 C1602063 C1602064 C1602065 C1602066 C1602067 C1602068 C1602069 C1602070 C1602071 C1602072 C1602073 c1602074 c1602075 c1602076 C1602077 C1602078 C1602079 C1602080 C1602081 C1602082 C1602083 C1602084 C1602085 C1602086 C1602087 C1602088 C1602089 C1602090 C1602091 C1602092 C1602093 C1602094 C1602095 C1602096 C1602097 C1602098 C1602099 c1602100 C1602101 C1602102 C1602103 C1602104 C1602105 C1602106 C1602107 c1602108 C1602109 C1602110 C1602111 C1602112 c1602113 C1602114 C1602115 C1602116 c1602117 C1602120 C1602121 C1602122 C1602123 C1602124 C1602125 C1602126 C1602127 C1602128 C1602129 C1602130 C1602131 C1602132 C1602133 C1602134 C1602135 C1602136 C1602137 C1602138 C1602139 C1602140 C1602141 C1602142 C1602143 C1602144 C1602145 C1602146 C1602147 C1602148 C1602149 C1602150 c1602151
FEI Number 3016438761
Recalling Firm/
Manufacturer
Abbott Laboratories
1915 Hurd Dr
Irving TX 75038
For Additional Information ContactDana Dominguez Schmidt
224-667-8598
Manufacturer Reason
for Recall
Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash buffer containers, missing QC flags, signal spikes, abnormal optics reads, and incorrect sample aspiration. The issues may lead to inaccurate results with an unknown bias direction and magnitude for assays measuring cardiac, metabolic (including diabetes), toxicology, prenatal, therapeutic drug monitoring, infectious disease, oncology, hepatic, pancreatic, hematological, endocrinologic, inflammatory analytes.
FDA Determined
Cause 2
Software design
ActionOn 09/29/2021, the Firm issued an "Urgent Medical Device Correction" Notification Letter via FedEx priority overnight express informing customers that the Firm has identified 3 potential performance issues with Software version 9.41 and earlier. The 3 issues are: 1. When an Error Code 3382, Unable to process test, internal wash pressure (x) error (y) pipettor occurs, the systems are incorrectly placed into a Scheduled Pause status rather than a Stopped status. As a result, the processing module continues to process tests after the hardware error is detected. This may cause incorrect results to be generated. 2. When configuring a Calibrator Sample Volume on the Configure Assays screen, if the user selects multiple assays, the calibrator sample volume from one assay may be carried into the calibrator sample volume for another assay. Incorrect calibrator sample volumes have the potential to generate incorrect calibration curves which may lead to incorrect results and delay of results due to the need to reconfigure assay parameters. 3. When performing a backup on the system while the iARM is replenishing the wash buffer at the same time, the iARM looses communication with the System Control Center. The loss of communication may cause the wash buffer container to overflow. This has the potential to lead to physical and chemical hazards. Customers are instructed to follow the Necessary Actions listed in Appendix A until software version 9.45 or software version 9.50 are installed. Appendix A states: For Issue (listed above): #1-The Necessary Actions to be taken by the customers until mandatory upgrade is completed - Stop the processing module and discard any results produced after an Error Code 3382. Perform Corrective Actions for Error Code 3382. For details, refer to System Operations Manual, Section 10: Troubleshooting and diagnostics. #2-The Necessary Actions to be taken by the customers until mandatory upgrade is completed - Verify the correct calibrator sample v
Quantity in Commerce7,535 systems
DistributionU.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: AFGHANISTAN ALBANIA, ALGERIA, ANDORRA, ANGOLA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHAMAS, BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BELIZE, BENIN, BERMUDA, BOLIVIA, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BRUNEI, BULGARIA, BURKINA FASO, BURUNDI, CAMBODIA, CAMEROON, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEMOCRATIC REP. OF CONGO, DENMARK, FAROE ISLANDS, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ERITREA, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GEORGIA, GERMANY, GHANA, GREECE, GREENLAND, GUATEMALA, GUINEA, HAITI, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, IVORY COAST, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, KYRGYZSTAN, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MADAGASCAR, MALAWI, MALAYSIA, MALDIVES, MALI, MALTA, MAURITANIA, MAURITIUS, MEXICO, MOLDOVA, MONTENEGRO, MONTSERRAT, MOROCCO, MOZAMBIQUE, MYANMAR, NAMIBIA, NEPAL, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, RWANDA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, ST. LUCIA, ST. VINCENT, SURINAME, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TAJIKISTAN, TANZANIA, THAILAND, TRINIDAD & TOBAGO, TUNISIA, TURK & CAICOS, TURKEY, UAE, UGANDA, UK, UKRAINE, URUGUAY, UZBEKISTAN, VIETNAM, WEST BANK/GAZA STRIP, YEMEN, ZAMBIA, ZIMBABWE,
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JJE
-
-