| | Class 2 Device Recall Sight OLO |  |
| Date Initiated by Firm | January 20, 2022 |
| Create Date | April 21, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0963-2022 |
| Recall Event ID |
89645 |
| 510(K)Number | K190898 |
| Product Classification |
Counter, differential cell - Product Code GKZ
|
| Product | The Sight OLO device is a computer vision based platform for blood analysis. The platform
combines computer-vision algorithms for image processing to identify and quantify blood
components (e.g., red blood cells) and their characteristics (e.g., cell volume) in an
automated fashion. Using dedicated staining, the proposed platform provides complete
blood count analysis. The Sight OLO is a compact device, designed to be automated and
simple to operate, to enable rapid testing and analysis. The Sight OLO consists of a
scanning and analyzing device and a CBC test kit, including disposable cartridges and
sample preparation tools. The disposable cartridge containing the blood sample is loaded
into the device through the loading slot. The device is operated through the touch screen
interface. The Sight OLO provides complete blood count information with 5-part differentials for white blood cell types. Specifically, the CBC parameters measured by the Sight OLO are listed below and include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, NEUT%/#,
LYMPH %/#, MONO %/#, EO%/# and BASO%/#. In addition, the Sight OLO signals specific
WBC abnormal cases by flagging the sample. |
| Code Information |
Software versions up to and including v2.59.3.
Serial Number UDI (US):
Q100 +B787OLOU10/$$+7Q-100G
Q121 +B787OLOU10/$$+7Q-121J
Q138 +B787OLOU10/$$+7Q-138R
Q145 +B787OLOU10/$$+7Q-145P
Q146 +B787OLOU10/$$+7Q-146Q
Q167 +B787OLOU10/$$+7Q-167T
Q212 +B787OLOU10/$$+7Q-212K
Q213 +B787OLOU10/$$+7Q-213L
Q225 +B787OLOU10/$$+7Q-225O
Q228 +B787OLOU10/$$+7Q-228R
Q241 +B787OLOU10/$$+7Q-241M
Q258 +B787OLOU10/$$+7Q-258U
Q264 +B787OLOU10/$$+7Q-264R
Q270 +B787OLOU10/$$+7Q-270O
Q273 +B787OLOU10/$$+7Q-273R
Q275 +B787OLOU10/$$+7Q-275T
Q277 +B787OLOU10/$$+7Q-277V
Q283 +B787OLOU10/$$+7Q-283S
Q303 +B787OLOU10/$$+7Q-303L
Q331 +B787OLOU10/$$+7Q-331M
Q332 +B787OLOU10/$$+7Q-332N
Q334 +B787OLOU10/$$+7Q-334P
Q338 +B787OLOU10/$$+7Q-338T
Q341 +B787OLOU10/$$+7Q-341N
Q347 +B787OLOU10/$$+7Q-347T
Q352 +B787OLOU10/$$+7Q-352P
Q363 +B787OLOU10/$$+7Q-363R
Q365 +B787OLOU10/$$+7Q-365T
Q372 +B787OLOU10/$$+7Q-372R
Q375 +B787OLOU10/$$+7Q-375U
Q380 +B787OLOU10/$$+7Q-380Q
Q383 +B787OLOU10/$$+7Q-383T
Q384 +B787OLOU10/$$+7Q-384U
Q387 +B787OLOU10/$$+7Q-387X
Q391 +B787OLOU10/$$+7Q-391S
Q393 +B787OLOU10/$$+7Q-393U
Q394 +B787OLOU10/$$+7Q-394V
Q398 +B787OLOU10/$$+7Q-398Z
Q400 +B787OLOU10/$$+7Q-400J
Q403 +B787OLOU10/$$+7Q-403M
Q411 +B787OLOU10/$$+7Q-411L
Q417 +B787OLOU10/$$+7Q-417R
Q418 +B787OLOU10/$$+7Q-418S
Q420 +B787OLOU10/$$+7Q-420L
Q424 +B787OLOU10/$$+7Q-424P
Q427 +B787OLOU10/$$+7Q-427S
Q436 +B787OLOU10/$$+7Q-436S
Q437 +B787OLOU10/$$+7Q-437T
Q440 +B787OLOU10/$$+7Q-440N
Q442 +B787OLOU10/$$+7Q-442P
Q451 +B787OLOU10/$$+7Q-451P
Q452 +B787OLOU10/$$+7Q-452Q
Q462 +B787OLOU10/$$+7Q-462R
Q463 +B787OLOU10/$$+7Q-463S
Q465 +B787OLOU10/$$+7Q-465U
Q467 +B787OLOU10/$$+7Q-467W
Q468 +B787OLOU10/$$+7Q-468X
Q469 +B787OLOU10/$$+7Q-469Y
Q470 +B787OLOU10/$$+7Q-470Q
Q472 +B787OLOU10/$$+7Q-472S
Q489 +B787OLOU10/$$+7Q-489-
Q496 +B787OLOU10/$$+7Q-496Y
Q499 +B787OLOU10/$$+7Q-499.
Q510 +B787OLOU10/$$+7Q-510L
Q511 +B787OLOU10/$$+7Q-511M
Q512 +B787OLOU10/$$+7Q-512N
Q513 +B787OLOU10/$$+7Q-513O
Q515 +B787OLOU10/$$+7Q-515Q
Q520 +B787OLOU10/$$+7Q-520M
Q524 +B787OLOU10/$$+7Q-524Q
Q527 +B787OLOU10/$$+7Q-527T
Q532 +B787OLOU10/$$+7Q-532P
Q534 +B787OLOU10/$$+7Q-534R
Q541 +B787OLOU10/$$+7Q-541P
Q543 +B787OLOU10/$$+7Q-543R
Q551 +B787OLOU10/$$+7Q-551Q
Q552 +B787OLOU10/$$+7Q-552R
Q557 +B787OLOU10/$$+7Q-557W
Q559 +B787OLOU10/$$+7Q-559Y
Q561 +B787OLOU10/$$+7Q-561R
Q565 +B787OLOU10/$$+7Q-565V
Q569 +B787OLOU10/$$+7Q-569Z
Q570 +B787OLOU10/$$+7Q-570R
Q571 +B787OLOU10/$$+7Q-571S
Q577 +B787OLOU10/$$+7Q-577Y
Q581 +B787OLOU10/$$+7Q-581T
Q586 +B787OLOU10/$$+7Q-586Y
Q594 +B787OLOU10/$$+7Q-594X
Q614 +B787OLOU10/$$+7Q-614Q
Q620 +B787OLOU10/$$+7Q-620N
Q624 +B787OLOU10/$$+7Q-624R
Q633 +B787OLOU10/$$+7Q-633R
Q634 +B787OLOU10/$$+7Q-634S
Q638 +B787OLOU10/$$+7Q-638W
Q640 +B787OLOU10/$$+7Q-640P
Q659 +B787OLOU10/$$+7Q-659Z
Q661 +B787OLOU10/$$+7Q-661S
Q664 +B787OLOU10/$$+7Q-664V
Q667 +B787OLOU10/$$+7Q-667Y
Q668 +B787OLOU10/$$+7Q-668Z
Q669 +B787OLOU10/$$+7Q-669-
Q673 +B787OLOU10/$$+7Q-673V
Q676 +B787OLOU10/$$+7Q-676Y
Q678 +B787OLOU10/$$+7Q-678-
Q679 +B787OLOU10/$$+7Q-679.
Q681 +B787OLOU10/$$+7Q-681U
Q686 +B787OLOU10/$$+7Q-686Z
Q687 +B787OLOU10/$$+7Q-687-
Q96 +B787OLOU10/$$+7Q-096U.
Serial Number UDI (OUS)
Q113 +B787OLOE10/$$+7Q-1134
Q140 +B787OLOE10/$$+7Q-1404
Q142 +B787OLOE10/$$+7Q-1426
Q157 +B787OLOE10/$$+7Q-157C
Q159 +B787OLOE10/$$+7Q-159E
Q162 +B787OLOE10/$$+7Q-1628
Q166 +B787OLOE10/$$+7Q-166C
Q170 +B787OLOE10/$$+7Q-1707
Q175 +B787OLOE10/$$+7Q-175C
Q176 +B787OLOE10/$$+7Q-176D
Q177 +B787OLOE10/$$+7Q-177E
Q179 +B787OLOE10/$$+7Q-179G
Q180 +B787OLOE10/$$+7Q-1808
Q189 +B787OLOE10/$$+7Q-189H
Q191 +B787OLOE10/$$+7Q-191A
Q204 +B787OLOE10/$$+7Q-2045
Q205 +B787OLOE10/$$+7Q-2056
Q206 +B787OLOE10/$$+7Q-2067
Q208 +B787OLOE10/$$+7Q-2089
Q209 +B787OLOU10/$$+7Q-209Q
Q210 +B787OLOE10/$$+7Q-2102
Q216 +B787OLOE10/$$+7Q-2168
Q220 +B787OLOE10/$$+7Q-2203
Q221 +B787OLOE10/$$+7Q-2214
Q223 +B787OLOE10/$$+7Q-2236
Q224 +B787OLOE10/$$+7Q-2247
Q227 +B787OLOE10/$$+7Q-227A
Q232 +B787OLOE10/$$+7Q-2326
Q233 +B787OLOE10/$$+7Q-2337
Q234 +B787OLOE10/$$+7Q-2348
Q236 +B787OLOE10/$$+7Q-236A
Q237 +B787OLOE10/$$+7Q-237B
Q243 +B787OLOE10/$$+7Q-2438
Q245 +B787OLOE10/$$+7Q-245A
Q246 +B787OLOE10/$$+7Q-246B
Q247 +B787OLOU10/$$+7Q-247S
Q249 +B787OLOE10/$$+7Q-249E
Q250 +B787OLOE10/$$+7Q-2506
Q251 +B787OLOE10/$$+7Q-2517
Q255 +B787OLOE10/$$+7Q-255B
Q256 +B787OLOE10/$$+7Q-256C
Q260 +B787OLOE10/$$+7Q-2607
Q261 +B787OLOE10/$$+7Q-2618
Q262 +B787OLOE10/$$+7Q-2629
Q263 +B787OLOE10/$$+7Q-263A
Q265 +B787OLOE10/$$+7Q-265C
Q267 +B787OLOE10/$$+7Q-267E
Q268 +B787OLOE10/$$+7Q-268F
Q271 +B787OLOE10/$$+7Q-2719
Q272 +B787OLOE10/$$+7Q-272A
Q274 +B787OLOE10/$$+7Q-274C
Q276 +B787OLOE10/$$+7Q-276E
Q278 +B787OLOE10/$$+7Q-278G
Q279 +B787OLOE10/$$+7Q-279H
Q280 +B787OLOE10/$$+7Q-2809
Q281 +B787OLOE10/$$+7Q-281A
Q282 +B787OLOE10/$$+7Q-282B
Q284 +B787OLOE10/$$+7Q-284D
Q285 +B787OLOE10/$$+7Q-285E
Q286 +B787OLOE10/$$+7Q-286F
Q287 +B787OLOE10/$$+7Q-287G
Q288 +B787OLOE10/$$+7Q-288H
Q289 +B787OLOE10/$$+7Q-289I
Q290 +B787OLOE10/$$+7Q-290A
Q291 +B787OLOE10/$$+7Q-291B
Q293 +B787OLOE10/$$+7Q-293D
Q294 +B787OLOE10/$$+7Q-294E
Q295 +B787OLOE10/$$+7Q-295F
Q296 +B787OLOE10/$$+7Q-296G
Q298 +B787OLOE10/$$+7Q-298I
Q300 +B787OLOE10/$$+7Q-3002
Q301 +B787OLOE10/$$+7Q-3013
Q302 +B787OLOE10/$$+7Q-3024
Q304 +B787OLOE10/$$+7Q-3046
Q305 +B787OLOE10/$$+7Q-3057
Q306 +B787OLOE10/$$+7Q-3068
Q307 +B787OLOE10/$$+7Q-3079
Q308 +B787OLOE10/$$+7Q-308A
Q309 +B787OLOE10/$$+7Q-309B
Q313 +B787OLOE10/$$+7Q-3136
Q315 +B787OLOU10/$$+7Q-315O
Q318 +B787OLOE10/$$+7Q-318B
Q320 +B787OLOE10/$$+7Q-3204
Q321 +B787OLOE10/$$+7Q-3215
Q322 +B787OLOE10/$$+7Q-3226
Q323 +B787OLOE10/$$+7Q-3237
Q325 +B787OLOE10/$$+7Q-3259
Q326 +B787OLOE10/$$+7Q-326A
Q327 +B787OLOE10/$$+7Q-327B
Q329 +B787OLOE10/$$+7Q-329D
Q330 +B787OLOE10/$$+7Q-3305
Q337 +B787OLOE10/$$+7Q-337C
Q339 +B787OLOE10/$$+7Q-339E
Q346 +B787OLOE10/$$+7Q-346C
Q354 +B787OLOE10/$$+7Q-354B
Q358 +B787OLOU10/$$+7Q-358V
Q361 +B787OLOE10/$$+7Q-3619
Q366 +B787OLOE10/$$+7Q-366E
Q369 +B787OLOE10/$$+7Q-369H
Q371 +B787OLOE10/$$+7Q-371A
Q376 +B787OLOE10/$$+7Q-376F
Q379 +B787OLOE10/$$+7Q-379I
Q385 +B787OLOE10/$$+7Q-385F
Q386 +B787OLOE10/$$+7Q-386G
Q396 +B787OLOE10/$$+7Q-396H
Q401 +B787OLOE10/$$+7Q-4014
Q402 +B787OLOE10/$$+7Q-4025
Q404 +B787OLOE10/$$+7Q-4047
Q406 +B787OLOE10/$$+7Q-4069
Q407 +B787OLOE10/$$+7Q-407A
Q408 +B787OLOE10/$$+7Q-408B
Q409 +B787OLOE10/$$+7Q-409C
Q410 +B787OLOE10/$$+7Q-4104
Q412 +B787OLOE10/$$+7Q-4126
Q413 +B787OLOE10/$$+7Q-4137
Q415 +B787OLOE10/$$+7Q-4159
Q419 +B787OLOE10/$$+7Q-419D
Q421 +B787OLOE10/$$+7Q-4216
Q426 +B787OLOU10/$$+7Q-426R
Q433 +B787OLOU10/$$+7Q-433P
Q434 +B787OLOE10/$$+7Q-434A
Q435 +B787OLOE10/$$+7Q-435B
Q441 +B787OLOE10/$$+7Q-4418
Q443 +B787OLOE10/$$+7Q-443A
Q445 +B787OLOE10/$$+7Q-445C
Q448 +B787OLOE10/$$+7Q-448F
Q454 +B787OLOE10/$$+7Q-454C
Q455 +B787OLOE10/$$+7Q-455D
Q457 +B787OLOE10/$$+7Q-457F
Q458 +B787OLOE10/$$+7Q-458G
Q460 +B787OLOE10/$$+7Q-4609
Q461 +B787OLOE10/$$+7Q-461A
Q466 +B787OLOE10/$$+7Q-466F
Q473 +B787OLOE10/$$+7Q-473D
Q475 +B787OLOE10/$$+7Q-475F
Q476 +B787OLOE10/$$+7Q-476G
Q480 +B787OLOE10/$$+7Q-480B
Q481 +B787OLOE10/$$+7Q-481C
Q484 +B787OLOE10/$$+7Q-484F
Q487 +B787OLOE10/$$+7Q-487I
Q488 +B787OLOE10/$$+7Q-488J
Q490 +B787OLOE10/$$+7Q-490C
Q493 +B787OLOE10/$$+7Q-493F
Q494 +B787OLOE10/$$+7Q-494G
Q500 +B787OLOE10/$$+7Q-5004
Q503 +B787OLOE10/$$+7Q-5037
Q504 +B787OLOE10/$$+7Q-5048
Q505 +B787OLOE10/$$+7Q-5059
Q508 +B787OLOE10/$$+7Q-508C
Q509 +B787OLOE10/$$+7Q-509D
Q514 +B787OLOE10/$$+7Q-5149
Q516 +B787OLOE10/$$+7Q-516B
Q518 +B787OLOE10/$$+7Q-518D
Q519 +B787OLOE10/$$+7Q-519E
Q523 +B787OLOE10/$$+7Q-5239
Q528 +B787OLOE10/$$+7Q-528E
Q530 +B787OLOE10/$$+7Q-5307
Q531 +B787OLOE10/$$+7Q-5318
Q533 +B787OLOE10/$$+7Q-533A
Q535 +B787OLOE10/$$+7Q-535C
Q538 +B787OLOE10/$$+7Q-538F
Q539 +B787OLOE10/$$+7Q-539G
Q540 +B787OLOE10/$$+7Q-5408
Q546 +B787OLOE10/$$+7Q-546E
Q554 +B787OLOE10/$$+7Q-554D
Q556 +B787OLOE10/$$+7Q-556F
Q558 +B787OLOE10/$$+7Q-558H
Q560 +B787OLOE10/$$+7Q-560A
Q562 +B787OLOE10/$$+7Q-562C
Q563 +B787OLOE10/$$+7Q-563D
Q567 +B787OLOE10/$$+7Q-567H
Q572 +B787OLOE10/$$+7Q-572D
Q576 +B787OLOE10/$$+7Q-576H
Q578 +B787OLOE10/$$+7Q-578J
Q579 +B787OLOE10/$$+7Q-579K
Q584 +B787OLOE10/$$+7Q-584G
Q587 +B787OLOE10/$$+7Q-587J
Q588 +B787OLOE10/$$+7Q-588K
Q591 +B787OLOE10/$$+7Q-591E
Q600 +B787OLOE10/$$+7Q-6005
Q610 +B787OLOE10/$$+7Q-6106
Q612 +B787OLOE10/$$+7Q-6128
Q617 +B787OLOE10/$$+7Q-617D
Q618 +B787OLOE10/$$+7Q-618E
Q62 +B787OLOE10/$$+7Q-0627
Q621 +B787OLOE10/$$+7Q-6218
Q622 +B787OLOE10/$$+7Q-6229
Q625 +B787OLOE10/$$+7Q-625C
Q629 +B787OLOE10/$$+7Q-629G
Q631 +B787OLOE10/$$+7Q-6319
Q635 +B787OLOE10/$$+7Q-635D
Q641 +B787OLOE10/$$+7Q-641A
Q642 +B787OLOE10/$$+7Q-642B
Q643 +B787OLOE10/$$+7Q-643C
Q645 +B787OLOE10/$$+7Q-645E
Q646 +B787OLOE10/$$+7Q-646F
Q647 +B787OLOE10/$$+7Q-647G
Q648 +B787OLOE10/$$+7Q-648H
Q649 +B787OLOE10/$$+7Q-649I
Q652 +B787OLOE10/$$+7Q-652C
Q653 +B787OLOE10/$$+7Q-653D
Q655 +B787OLOE10/$$+7Q-655F
Q656 +B787OLOE10/$$+7Q-656G
Q657 +B787OLOE10/$$+7Q-657H
Q660 +B787OLOE10/$$+7Q-660B
Q662 +B787OLOE10/$$+7Q-662D
Q665 +B787OLOE10/$$+7Q-665G
Q666 +B787OLOE10/$$+7Q-666H
Q671 +B787OLOE10/$$+7Q-671D
Q674 +B787OLOE10/$$+7Q-674G
Q709 +B787OLOE10/$$+7Q-709F
Q79 +B787OLOE10/$$+7Q-079F
Q94 +B787OLOE10/$$+7Q-094C
|
| FEI Number |
3016099075
|
Recalling Firm/ Manufacturer |
SIGHT DIAGNOSTICS LTD
23, Derekh Menakhem Begin
Tel Aviv-Yafo
Israel
|
Manufacturer Reason for Recall | In instances where custom reference ranges were configured on the device post installation,
a possibility to inadvertently apply changes to reference range values was found on software
versions 2.59.3 and all earlier versions, which can potentially lead to displaying and printing
incorrectly-configured reference range values |
FDA Determined Cause 2 | Software design |
| Action | Sight Diagnostics issued Medical Device Recall Notification on 1/20/22 via email to Distributors and End Users. Letter states reason for recall, health risk and action to take:
In case you have configured your own reference ranges, please review the configured
values to ensure all values are set properly. You can then continue to use your OLO
analyzer.
The software version of the OLO analyzer in your possession will be updated to a new
version that resolves this software defect. The update will be performed remotely by
Sights product specialists, who will contact you to schedule the update. Please be
aware that the software version update requires the analyzer to be connected to the
internet.
If you need to make any adjustments to your reference ranges before the software
update is performed, please avoid pressing the Cancel button while editing a
reference range parameter, and always review all parameters after making any
changes, to confirm values are set properly.
If you are a distributor, please share this recall notification letter with any customer
to whom the product was shipped, or may have been shipped.
You are kindly requested to fill the attached Acknowledgement and Receipt Form,
and return it to notifications@sightdx.com within 30 days.
Software version 2.63 was released and includes the fix for the software defect
mentioned above.
Questions please contact us at Support@sightdx.com. For full indications
for use and safety information please refer to the Quality and Compliance page at
www.sightdx.com. |
| Quantity in Commerce | 322 units: U.S 109 Foreign 213 |
| Distribution | Worldwide distribution - US Nationwide and the countries of AUSTRALIA, CANADA, CHILE, CURACAO, FINLAND, GEORGIA, GERMANY, GREECE, HONG KONG, IRELAND, ISRAEL, ITALY,
MEXICO, NETHERLANDS, PORTUGAL, SINGAPORE, SLOVENIA, UNITED ARAB EMIRATES,
UNITED KINGDOM.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = GKZ
|
|
|
|