Date Initiated by Firm |
March 25, 2022 |
Create Date |
May 18, 2022 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1114-2022 |
Recall Event ID |
90047 |
510(K)Number |
K970281
|
Product Classification |
Radioimmunoassay, prolactin (lactogen) - Product Code CFT
|
Product |
VITROS Immunodiagnostic Products Prolactin Reagent Pack Product Code: 1849793 |
Code Information |
Until further notice, future lots will also be affected.
UDI: 10758750000111
Affected Lot Number/ Expiration Date:
2840 13-MAY-2022;
2850 08-JUN-2022;
2860 02-JUL-2022;
2870 17-JUL-2022;
2875 17-JUL-2022;
2880 14-SEP-2022;
2891 28-SEP-2022;
2900 29-OCT-2022;
|
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics, Inc 1001 Us Highway 202 Raritan NJ 08869-1424
|
For Additional Information Contact |
SAME 908-218-8776
|
Manufacturer Reason for Recall |
Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.
|
FDA Determined Cause 2 |
Device Design |
Action |
Ortho CLinical Diagnostics issued "URGENT PRODUCT CORRECTION NOTIFICATION" dated 3/25/22 to consignees who had been shipped VITROS FSH, LH or Prol in the previous 18 months were notified (CL2022-069) of the issue and instructed as follows:
-Run daily low-level QC in duplicate to confirm acceptable performance.
-If their laboratory is experiencing low end imprecision with VITROS FSH, LH or Prol Reagent, follow one of the suggested testing algorithms as described in the enclosure.
o Note: If their laboratory is not experiencing low end imprecision, continue testing patient samples as usual (single replicate).
-Until further notice, all future lots of the products will also be affected.
-On the enclosed Credit Form, track the additional quantities of VITROS FSH, LH and/or Prol tests used due to the testing protocol. Periodically submit the form to Ortho. Ortho will credit their account.
If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311. |
Quantity in Commerce |
1767 units total: 1521 US; 246 OUS |
Distribution |
Worldwide Distribution:
US (Nationwide)
OUS (Foreign):
Australia
Bermuda
Brazil
Canada L3R 4G5
Chile
China
Colombia
Denmark
France
Germany
India
Italy
Japan
Mexico
Netherlands
Norway
Portugal
Russia
Singapore
Spain
Sweden
United Kingdom
|
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = CFT and Original Applicant = JOHNSON & JOHNSON CLINICAL DIAGNOSTICS, INC.
|